IOSR Journal of Applied Physics (IOSR-JAP)
e-ISSN: 2278-4861.Volume 7, Issue 1 Ver. II (Jan.-Feb. 2015), PP 11-15
www.iosrjournals.org
DOI: 10.9790/4861-07121115 www.iosrjournals.org 11 | Page
Determination ofRadiological Quality Parameters Using Optical
Densitometer andSimple Fabricated Equipment
Kpaku Goriya1
,Abdullahi Mundi1
, Aribisala Abdul-Jamiu2
,
Peter Eshiett3
,Richard Itopa4
,DlamaZira Joseph5
,
1. Physics Department, Nasarawa state university Keffi, Nigeria
2. FCT Secondary Education Board, Abuja, Nigeria
3. Federal Medical Center Keffi, Nigeria
4. Department of Radiography, Bayero University Kano
5. Radiology DepartmentAbubakarTafawaBalewa University Teaching Hospital Bauchi, Nigeria.
CORRESPONDING AUTHOR’S EMAIL5
:josephdlama@gmail.com
Abstract: The etiquette of safe practice for the use of x-rays in medical diagnosis requires that each x-ray
facility has an appropriate quality assurance program in radiation protection and to keep doses as low as
reasonable achievable. An alternative method of evaluating radiological quality parameters of diagnostic x-ray
machine is considered in this study. This study was conducted in Federal Medical Center Keffi, Nasarawa State.
The aim was to determine radiological quality parameters using optical densitometer and simple fabricated
equipment. This research has determined some of these parameters via the wooden step wedge and metal pin
fabricated. The wedge and metal pin were exposed to x-ray and the optical densities of the processed films were
measured with a densitometer. The result indicates that simple fabricated equipment can be used to determine
the following radiological quality parameters such as dose, beam alignment, peak kilo voltage and tube current
consistency in the absence of quality control test equipment. This is an indication that departmental protocol
can be established in x-ray department using simple fabricated equipment .These equipment can be used in the
determination of radiological quality parameters.
Keywords:[Radiation Protection, Parameters, Optical Densitometer,Diagnostic, Equipment.]
I. Introduction
Diagnostic radiology machines have been of great help to health care industry and many people have
benefited from the useful information they provide[1]. Providing standard health care infrastructure is one of the
priorities of modern-day hospital in our country[2]. The investment in state –of the art equipment is vital in
achieving this goal. It is essential to maintain the quality of high-end equipment procured to provide high quality
health care. One of the difficult areas of maintaining this infrastructure is within the diagnostic radiology
department[2],[3].
Radiation cannot be seen, heard or smelt hence, to check the quality of radiation producing equipment,
intuitive methods of detection seldom fail[4]. Checking the quality of the image might give an indication of
optimal performance of radiation equipment, but can be often misleading[4]. The quality of diagnostic radiation
equipment directly affects the image quality and unwanted exposure to staff[5]. Poor diagnostic image quality
can lead to diagnostic dilemma[5].
II. Materials And Methods
The materials used for this study include
 X–ray machine
 Densitometer
 Wooden step wedge
 Uniform metal pin
 Radiographic film (Kodak)
 Micrometer screw gauge
 Verniercaliper
2.1X–ray Machine
Characteristics of the x–ray machine include
 Tube voltage, 125kv
 Total filtration, 2.9mmAl/70kv
Determination Of Radiological Quality Parameters Using Optical Densitometer
DOI: 10.9790/4861-07121115 www.iosrjournals.org 12 | Page
 Tube serial number 8324/12
 Model number XD52–30.50/125–T2A
2.2 Densitometer
Characteristic and specification of densitometer
Point densitometer is straight forward device used for determining the optical density (OD) at a few
points in a film. This device is easiest to QA for absolute OD and is therefore often used as the local standards in
OD measurements. Point densitometer use a silicon photodiode to measure the transmitted flux of light passing
through a film. The international organization for standardization (ISO) has developed standards for the
geometric and spectral conditions for the determination of optical density (ISO,5) Diffuse illumination is
achieved by a broad–spectrum, incandescent lamp whose spectral properties conform to international
commission on illumination (CIE) standards. The collimated light passes through the film and the transmitted
component passes through a detecting aperture that usually can be from 1–3mm in diameter. This light flux
passes through a V- filter (555nm peak – 780nm range) and is then detected by a silicon photodiode with
amplifier electronic capable of measuring signals spanning several orders of magnitude. This system consist of a
backlit tablet with light aperture on which one places the film sample, an arm which extends over the film and
has at its end the light source aligned with the aperture and a digital readout of optical density (Sujatha, 2007).
2.3 Construction
Iroko wood was obtained in the form of rectangular block of dimensions. Light, L = 10cm, width, w =
3.5cm, height, H = 3cm, mass, m = 25g, volume, v = 65cm3
. The construction was carried out at kpakurium
workshop in Keffi, Nasarawa State. The wood was machined to square using the plane jack, to smoothen out
rough edges and make the surface uniformly smooth throughout. Five steps of equal thickness and height were
marked on the wooden block with pencil, to aid accuracy in cutting the steps. The difference in thickness
between each step is 0.6cm (Martins, 2007).
2.4 Exposure Procedure
 The wooden step wedge was placed on a 24 x 30cm cassette, a control exposure was made by keeping the
mAs constant at 16mAs and varying the kVp, thus 45, 50, 55, 60, 65 and 70kVp respectively at constant
film to focus distance (FFD) of 100cm, a total of six exposure were made.
 The same procedure was repeated by keeping the kVp constant at 45kVp and varying the mAs, thus 2, 4, 6,
8, and 10mAs respectively at constant FFD of 100cm, a total of six exposure were also made in this case.
 Film processing, this was done in the dark room with only the safe light on. The exposed films were
removed from the cassettes and processed temperature of 0.3 degree Celsius. After processing and drying,
the film was assessed on the view box, all the steps of the wedge were visible.
2.5 Measurement of optical density
The measurement of optical density (O.D), was carried out using the transmission digital densitometer
x–rite 331 nickel–cadmium (Nicad), battery operated (600/700mAhr). For the constant 16mAs films with the
images of the wooden step wedge, the optical densities of the five steps were measured for the constant 45kVp
film with the images of the wooden step wedge, the optical densities f for five steps were measured.
2.6 Simple beam alignment test
A uniform metal pin of thickness 0.3cm, length 10cm was mounted 900
vertically on a rectangular
block of wood 2 x 4cm and 0.5cm thick. The metal pin was placed on a cassette and expose to x–ray using a
suitable exposure factor (4mAs x 50kVp) and FFD of 100cm (Glen, 1993). The tolerance angle () was
calculated using trigonometry ratio (tan = a/b).
The length of the image of the metal pin was traced on the x–ray film with a divider and a meter rule.
The length of the metal pin and that of the image form two side of a triangle. With these information, the
tolerance angle was calculated as shown in the
2.7Statistical Analysis
In order to determine the effect of kVp and mAs on dose, a graph of these two parameters was plot against
optical density using the Microsoft Excel and the SPSS Statistical Package
III. Results
3.1 Fabricated Wooden Step Wedge and Metal Pin
Five steps wooden wedge and 10cm metal pin were fabricated and used to determine the radiological
quality parameters. The fabricated equipment are shown in figure 3.0 below
Determination Of Radiological Quality Parameters Using Optical Densitometer
DOI: 10.9790/4861-07121115 www.iosrjournals.org 13 | Page
 Wooden step wedge (b) Metal Pin
Figure 3.1: simple fabricated equipment
Figure 3.2: radiograph of wooden step wedge at 16mAs and varying kVp
Figure 3.3: A graph of optical density against thickness.
Figure 3.4: A graphof tube charge against optical density.
Table 3.0: Relationship between Optical Density and Dose .
S/N Optical density Dose(Gy)
1 0.74 0.754
2 0.69 0.748
3 0.67 0.746
4 0.60 0.743
5 0.58 0.721
Table 3.1: Simple Beam Alignment Test
1 2 3 Average
Length of metal pin(a)cm 10.00 10.00 10.00 10.00
Length of image(b)cm 0.20 0.30 0.20 0.20
Tolerance angle (degree) 1.14 1.70 1.14 1.32
Figure 3.5: Radiograph of 10cm metal pin
Determination Of Radiological Quality Parameters Using Optical Densitometer
DOI: 10.9790/4861-07121115 www.iosrjournals.org 14 | Page
IV. Discussion
The result from this study shows that, simple fabricated equipment can be used to determine
radiological quality parameters .This is useful in diagnostic radiology where quality control (Q C) test
equipment are not available.
The step wedge method of checking the consistency of peak kilo voltage and tube charge of diagnostic
x-ray machine does so in terms of optical density which represents the total exposure reaching the film and
producing the image.
A common problem among the hospitals in Nigeria is the lack of regular patients dose monitoring and
quality control in diagnostic radiology. A major reason for this is the cost of running a standard radiation
protection and quality assurance on the facilities.[8],[9],[10].
Figure 3.1 shows that at constant 16 mAs using the fifth step of the wedge ,as the kVp of the x-ray
machine increases ,there is also an increase in optical density ,which confirm the relational equation between
optical density and kVp which is O.D=C(kVp)n
. From the graph (figure3.1) the power n is 2.997 and the
constant value is C is 4×10-6
, which is within the recommended limit[15][16]. This is an indication that the kVp
is consistent. This is in consonance with a study done by using aluminum step wedge in Jos university teaching
hospital[13][15][18].
Figure 3.2 shows that optical density decreases with increase in thickness, the higher steps of the wedge
are dense, thus absorb more radiation than they transmit. The lower steps on the other hand are not as thick and
hence readily transmit radiation, absorbing none at all. This implies that attenuation increase with increase in
thickness.[16],[19].
Figure 3.3 shows that at constant 45kVp as the tube charge increases, the optical density also increases
.Higher values than 8mAs has higher optical density, but poor image contrast, this may be due to excessive
dose. This implies that optical density is directly proportional to tube charge, which confirm the relational
equation between optical density and tube charge.( O.D α mAs).
Table 3 shows the relationship between optical density and dose which is obtained from linear path of
the curve of a graph of optical density against thickness (figure 3.3). Y=0.077X+0.8 is the equation of the linear
path of the curve, where Y is the dose and X is the optical density. The doses were obtained by substituting the
values of optical density into the equation .The result obtained as presented in table 3, shows that as the optical
density increases, dose also increases. This indicate that optical density can be use to estimate dose[16],[17].
Table 3.1 shows the average tolerance angle obtained from the simple beam alignment test which is
3o
.This implies that the x-ray beam of the machine is aligned since the tolerance angle is within the
recommended range ±5 [15],[17].
V. Summary of the Study Finding
 As kVp of the x-ray machine increases, there is also increase in optical density. This implies that optical
density is directly proportional to kVp. It means that change in potential can cause drifts in beam intensity
and thus significantly influence radiation dose and image quality as well as after the film density.
 As the tube charge of the x-ray machine increases, there is also increase in optical density. This implies that
optical density is directly proportional to tube charge.
 The optical density of the higher steps, are less than those of the lower steps. This implies that attenuation
increases with increase in thickness, it indicate that optical density is inversely proportional to thickness.
 As the optical density increases dose also increases. This implies that optical density is directly proportional
to dose
 The tolerance angle obtained from the simple beam alignment test is 3o
,which indicate that the x-ray
machine is properly collimated and the beam is aligned
 The following quality parameters were found to be consistence; kVp ,mAs, dose and beam alignment.
VI. Conclusion
The result from this study shows that simple fabricated equipment can be used to determine
radiological quality parameters, such as; dose, beam alignment, kVp and mAsconsistency.This implies that, if
the diagnostic x-ray machine is not maintaining the consistency of this parameters, it’
s output will be poor.
References
[1]. Ash, D & Bates, T. (1994) Report on the Clinical Effects of Inadvertent Radiation under Dosage in 1045 Patients, Clinical Oncol.
World Journal of Medical Sciences 6:214 – 225.
[2]. Akpa, T.C. (2013). Interaction of Radiation with Matter. A Lecture note on, Radiation physics, Nasarawa State University, Keffi
(unpublished).
[3]. Attix, F. H. (1986). Introduction to Radiological Physics and Radiation Dosimetry. Wiley, New York PP56
[4]. AERB Safety Code. (2001). Code No. AERB/SE/MED- 2, Mumbia PP 1 -20.
Determination Of Radiological Quality Parameters Using Optical Densitometer
DOI: 10.9790/4861-07121115 www.iosrjournals.org 15 | Page
[5]. Bushong, S.C. (1993). Radiography Science for Technologist Physics, Biology and Protection. 5th
Edition. St Lovis; Mmosby Year
Book PP45-70
[6]. Beacomfield, T., Nicholson, R., & Thornton, A. (1999). Would Thyroid and Breast Shielding be Beneficial in CT of the Head? Eur-
Radiol. World Journal of Medical Sciences. 8(4), pp664.
[7]. Christian Nordgvist (2011).Effect of Radiation on Human http://www.medicalnewstoday.com/articles/2.1965.php.
[8]. Curry, T. S., Dowdey, J.E& Marry, R. C. (1984). Introduction to Physics of Diagnostic Radiography 3rd
Edition. Lea and Fibiger,
Philadelphia, PA pp88 – 106.
[9]. Edmonds, I.R. (1984). Calculation of Patient Skin Dose from Diagnostic x-ray Procedure. British Journal of Radiology, 57, 733 –
734.
[10]. Egbe, N.O., Inyang, S. O &Eduwem, D.U. (2008). Doses and Image Quality for Chest Radiograph in three Nigerian Hospitals.
European Journal of Radiography 1, 30 – 36.
[11]. Golani industrial complex waliv, vasai (2013).In house Quality assurance of diagnostic radiology Equipment. www.raysafe.com.
[12]. HananFawaz. (2007). Assessment of Entrance Skin Dose and Effective Dose of some Routine X-ray Examination using Calculation
Technique.www.geogle.
[13]. International Commission on Radiological Protection (1996). Protection and Safety in Medicine, Publication 73, Pergamon Press,
Oxford.
[14]. John & Brenan. (2000). Reference Dose level for Patients undergoing Common Diagnostic x-rays Examination in Irish Hospital.
British Journal of Radiology 73, 396 – 402.
[15]. Mangset, W.E. &Izang, N. (2010). Locally Fabricated Metal Step Wedge for Quality Assurance in Diagnostic Radiology.
Department of Physics, University of Jos. Bayero Journal of Pure and Applied Sciences, 3(1); 156 – 163
[16]. Martin, C.T. (2007). Optimization in General Radiography www.biij.org/2007/2/e/8
[17]. Nworgu, O. D &Bamidele, L. (2010). An Evaluation of Entrance Surgace doses from some Common Diagnostic x-ray
Examinations in two Nigerian University Teaching. Journal of Radiological Protection Vol. 29, N2a, 952-4746.
[18]. Rehani, M.M.(1995). Diagnostic Imaging Quality Assurance. Jaypee Brothers Medical Publisher.New York, pp45.
[19]. Sujatha, P. (2007). Radiographic Film for Mega Voltage Beam Dosimetry. Radiation Therapy Department, Memorial Hermann
Hospital, Houston. Texas 77024, pp2239.
[20]. Sato, O., Iwai, S &Farihata, S. (1995). Calculation of Equivalent Dose and Effective Dose Conservation Coefficients for Photon.
Radiation Protection Dosimetry. Nuclear Technology Publishing, Japan, Vol.62, No. 3, pp119 -130.

Mais conteúdo relacionado

Mais procurados

Imaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationImaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verification
Miami Cancer Institute
 
Biophys j savin05
Biophys j savin05Biophys j savin05
Biophys j savin05
crazyduddy
 
twofold processing for denoising ultrasound medical images
twofold processing for denoising ultrasound medical imagestwofold processing for denoising ultrasound medical images
twofold processing for denoising ultrasound medical images
anil kumar
 
BMES_Poster_Microscope_Comparison_Meyer
BMES_Poster_Microscope_Comparison_MeyerBMES_Poster_Microscope_Comparison_Meyer
BMES_Poster_Microscope_Comparison_Meyer
Andrew Meyer
 
Jin_single_photon(4)
Jin_single_photon(4)Jin_single_photon(4)
Jin_single_photon(4)
Jin Wang
 
Magnetic nanoparticles applications and bioavailability for cancer therapy
Magnetic nanoparticles applications and bioavailability for cancer therapyMagnetic nanoparticles applications and bioavailability for cancer therapy
Magnetic nanoparticles applications and bioavailability for cancer therapy
Pravin Chinchole
 
ASME JOURNAL PAPER - med_008_03_030934
ASME JOURNAL PAPER - med_008_03_030934ASME JOURNAL PAPER - med_008_03_030934
ASME JOURNAL PAPER - med_008_03_030934
Sachin Bijadi
 
Introduction to radiation therapy
Introduction to radiation therapyIntroduction to radiation therapy
Introduction to radiation therapy
Rad Tech
 
Radiochemical Properties of Irradiated PVA\AgNO3 Film by Electron Beam
Radiochemical Properties of Irradiated PVA\AgNO3 Film by Electron BeamRadiochemical Properties of Irradiated PVA\AgNO3 Film by Electron Beam
Radiochemical Properties of Irradiated PVA\AgNO3 Film by Electron Beam
International Journal of Science and Research (IJSR)
 

Mais procurados (20)

Imaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationImaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verification
 
Biophys j savin05
Biophys j savin05Biophys j savin05
Biophys j savin05
 
twofold processing for denoising ultrasound medical images
twofold processing for denoising ultrasound medical imagestwofold processing for denoising ultrasound medical images
twofold processing for denoising ultrasound medical images
 
BMES_Poster_Microscope_Comparison_Meyer
BMES_Poster_Microscope_Comparison_MeyerBMES_Poster_Microscope_Comparison_Meyer
BMES_Poster_Microscope_Comparison_Meyer
 
Jin_single_photon(4)
Jin_single_photon(4)Jin_single_photon(4)
Jin_single_photon(4)
 
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...
Dual energy imaging and digital tomosynthesis: Innovative X-ray based imaging...
 
Magnetic nanoparticles applications and bioavailability for cancer therapy
Magnetic nanoparticles applications and bioavailability for cancer therapyMagnetic nanoparticles applications and bioavailability for cancer therapy
Magnetic nanoparticles applications and bioavailability for cancer therapy
 
Bio-waste-derived nanofibers formed by solution blowing and their application...
Bio-waste-derived nanofibers formed by solution blowing and their application...Bio-waste-derived nanofibers formed by solution blowing and their application...
Bio-waste-derived nanofibers formed by solution blowing and their application...
 
Design of Gabor Filter for Noise Reduction in Betel Vine leaves Disease Segme...
Design of Gabor Filter for Noise Reduction in Betel Vine leaves Disease Segme...Design of Gabor Filter for Noise Reduction in Betel Vine leaves Disease Segme...
Design of Gabor Filter for Noise Reduction in Betel Vine leaves Disease Segme...
 
Thickness and n&k measurement with MProbe
Thickness and n&k measurement with MProbeThickness and n&k measurement with MProbe
Thickness and n&k measurement with MProbe
 
Exploring the Layered Structure of Composite Films Using an FTIR Microscope
Exploring the Layered Structure of Composite Films Using an FTIR MicroscopeExploring the Layered Structure of Composite Films Using an FTIR Microscope
Exploring the Layered Structure of Composite Films Using an FTIR Microscope
 
International Conference On Electrical Engineering
International Conference On Electrical EngineeringInternational Conference On Electrical Engineering
International Conference On Electrical Engineering
 
IRJET- Effect of Energy, Type and Thickness of Insulator on Flaw Detectab...
IRJET-  	  Effect of Energy, Type and Thickness of Insulator on Flaw Detectab...IRJET-  	  Effect of Energy, Type and Thickness of Insulator on Flaw Detectab...
IRJET- Effect of Energy, Type and Thickness of Insulator on Flaw Detectab...
 
INSPECTION OF GFRP COMPOSITES BY MICROWAVE NDE
INSPECTION OF GFRP COMPOSITES BY MICROWAVE NDEINSPECTION OF GFRP COMPOSITES BY MICROWAVE NDE
INSPECTION OF GFRP COMPOSITES BY MICROWAVE NDE
 
Drug delivery and nanomedicine with nanomagnetic material
Drug delivery and nanomedicine with nanomagnetic materialDrug delivery and nanomedicine with nanomagnetic material
Drug delivery and nanomedicine with nanomagnetic material
 
INSPECTION OF GFRP COMPOSITES BY MICROWAVE NDE
INSPECTION OF GFRP COMPOSITES BY MICROWAVE NDEINSPECTION OF GFRP COMPOSITES BY MICROWAVE NDE
INSPECTION OF GFRP COMPOSITES BY MICROWAVE NDE
 
ASME JOURNAL PAPER - med_008_03_030934
ASME JOURNAL PAPER - med_008_03_030934ASME JOURNAL PAPER - med_008_03_030934
ASME JOURNAL PAPER - med_008_03_030934
 
Introduction to radiation therapy
Introduction to radiation therapyIntroduction to radiation therapy
Introduction to radiation therapy
 
2018.10.05. aaps. particles investigation
2018.10.05. aaps. particles investigation2018.10.05. aaps. particles investigation
2018.10.05. aaps. particles investigation
 
Radiochemical Properties of Irradiated PVA\AgNO3 Film by Electron Beam
Radiochemical Properties of Irradiated PVA\AgNO3 Film by Electron BeamRadiochemical Properties of Irradiated PVA\AgNO3 Film by Electron Beam
Radiochemical Properties of Irradiated PVA\AgNO3 Film by Electron Beam
 

Destaque

Destaque (20)

Analysis of Spending Pattern on Credit Card Fraud Detection
Analysis of Spending Pattern on Credit Card Fraud DetectionAnalysis of Spending Pattern on Credit Card Fraud Detection
Analysis of Spending Pattern on Credit Card Fraud Detection
 
Assessment of the Heavy Metal Levels in Feeds and Litters of Chickens Rose wi...
Assessment of the Heavy Metal Levels in Feeds and Litters of Chickens Rose wi...Assessment of the Heavy Metal Levels in Feeds and Litters of Chickens Rose wi...
Assessment of the Heavy Metal Levels in Feeds and Litters of Chickens Rose wi...
 
Car Dynamics using Quarter Model and Passive Suspension, Part VI: Sprung-mass...
Car Dynamics using Quarter Model and Passive Suspension, Part VI: Sprung-mass...Car Dynamics using Quarter Model and Passive Suspension, Part VI: Sprung-mass...
Car Dynamics using Quarter Model and Passive Suspension, Part VI: Sprung-mass...
 
Spatial Scalable Video Compression Using H.264
Spatial Scalable Video Compression Using H.264Spatial Scalable Video Compression Using H.264
Spatial Scalable Video Compression Using H.264
 
Algorithm for the Comparison of Different Types of First Order Edge Detection...
Algorithm for the Comparison of Different Types of First Order Edge Detection...Algorithm for the Comparison of Different Types of First Order Edge Detection...
Algorithm for the Comparison of Different Types of First Order Edge Detection...
 
Short-term Load Forecasting using traditional demand forecasting
Short-term Load Forecasting using traditional demand forecastingShort-term Load Forecasting using traditional demand forecasting
Short-term Load Forecasting using traditional demand forecasting
 
Mathematical Relationships between the Compressive Strength and Some Other St...
Mathematical Relationships between the Compressive Strength and Some Other St...Mathematical Relationships between the Compressive Strength and Some Other St...
Mathematical Relationships between the Compressive Strength and Some Other St...
 
Preventing Malicious Node and Provide Secure Routing In Manet
Preventing Malicious Node and Provide Secure Routing In ManetPreventing Malicious Node and Provide Secure Routing In Manet
Preventing Malicious Node and Provide Secure Routing In Manet
 
Numerical Simulation of 퐒퐢ퟏ−퐱퐆퐞퐱 Thin Film Solar Cell Using AMPS - 1D
Numerical Simulation of 퐒퐢ퟏ−퐱퐆퐞퐱 Thin Film Solar Cell Using AMPS - 1DNumerical Simulation of 퐒퐢ퟏ−퐱퐆퐞퐱 Thin Film Solar Cell Using AMPS - 1D
Numerical Simulation of 퐒퐢ퟏ−퐱퐆퐞퐱 Thin Film Solar Cell Using AMPS - 1D
 
Literature Survey on Detection of Brain Tumor from MRI Images
Literature Survey on Detection of Brain Tumor from MRI Images Literature Survey on Detection of Brain Tumor from MRI Images
Literature Survey on Detection of Brain Tumor from MRI Images
 
New Approach for Determination of Propagation Model Adapted To an Environment...
New Approach for Determination of Propagation Model Adapted To an Environment...New Approach for Determination of Propagation Model Adapted To an Environment...
New Approach for Determination of Propagation Model Adapted To an Environment...
 
The Effect of Imbalance on the Performance of Unpaced Production Line – A Mat...
The Effect of Imbalance on the Performance of Unpaced Production Line – A Mat...The Effect of Imbalance on the Performance of Unpaced Production Line – A Mat...
The Effect of Imbalance on the Performance of Unpaced Production Line – A Mat...
 
On Some Types of Fuzzy Separation Axioms in Fuzzy Topological Space on Fuzzy ...
On Some Types of Fuzzy Separation Axioms in Fuzzy Topological Space on Fuzzy ...On Some Types of Fuzzy Separation Axioms in Fuzzy Topological Space on Fuzzy ...
On Some Types of Fuzzy Separation Axioms in Fuzzy Topological Space on Fuzzy ...
 
Smartphone Based Sensing Driver Behavior Modeling
Smartphone Based Sensing Driver Behavior ModelingSmartphone Based Sensing Driver Behavior Modeling
Smartphone Based Sensing Driver Behavior Modeling
 
Bulk Demand (S, s) Inventory System with Varying Environment
Bulk Demand (S, s) Inventory System with Varying EnvironmentBulk Demand (S, s) Inventory System with Varying Environment
Bulk Demand (S, s) Inventory System with Varying Environment
 
Best Power Surge Using Fuzzy Controlled Genetic Technique
Best Power Surge Using Fuzzy Controlled Genetic TechniqueBest Power Surge Using Fuzzy Controlled Genetic Technique
Best Power Surge Using Fuzzy Controlled Genetic Technique
 
Data mining Algorithm’s Variant Analysis
Data mining Algorithm’s Variant AnalysisData mining Algorithm’s Variant Analysis
Data mining Algorithm’s Variant Analysis
 
Design and Implementation of an Improved Wind Speed Meter (Anemometer)
Design and Implementation of an Improved Wind Speed Meter (Anemometer)Design and Implementation of an Improved Wind Speed Meter (Anemometer)
Design and Implementation of an Improved Wind Speed Meter (Anemometer)
 
Power Flow Control in Grid-Connected Wind Energy Conversion System Using PMSG...
Power Flow Control in Grid-Connected Wind Energy Conversion System Using PMSG...Power Flow Control in Grid-Connected Wind Energy Conversion System Using PMSG...
Power Flow Control in Grid-Connected Wind Energy Conversion System Using PMSG...
 
Simulation of Wind Power Dynamic for Electricity Production in Nassiriyah Dis...
Simulation of Wind Power Dynamic for Electricity Production in Nassiriyah Dis...Simulation of Wind Power Dynamic for Electricity Production in Nassiriyah Dis...
Simulation of Wind Power Dynamic for Electricity Production in Nassiriyah Dis...
 

Semelhante a Determination ofRadiological Quality Parameters Using Optical Densitometer andSimple Fabricated Equipment

Design and Development of a Shortwave near Infrared Spectroscopy using NIR LE...
Design and Development of a Shortwave near Infrared Spectroscopy using NIR LE...Design and Development of a Shortwave near Infrared Spectroscopy using NIR LE...
Design and Development of a Shortwave near Infrared Spectroscopy using NIR LE...
IJECEIAES
 
Impact of detector thickness on imaging characteristics of the Siemens Biogra...
Impact of detector thickness on imaging characteristics of the Siemens Biogra...Impact of detector thickness on imaging characteristics of the Siemens Biogra...
Impact of detector thickness on imaging characteristics of the Siemens Biogra...
Anax Fotopoulos
 
Report In Japan 20060613 Liuxu
Report In Japan 20060613 LiuxuReport In Japan 20060613 Liuxu
Report In Japan 20060613 Liuxu
巍 陆
 
2011-IPP-CT Data Evaluation of Fibre Reinforced Polymers to Determine Fibre L...
2011-IPP-CT Data Evaluation of Fibre Reinforced Polymers to Determine Fibre L...2011-IPP-CT Data Evaluation of Fibre Reinforced Polymers to Determine Fibre L...
2011-IPP-CT Data Evaluation of Fibre Reinforced Polymers to Determine Fibre L...
Kana Arunachalam Kannappan
 
Tesi Master Judilka Bermudez
Tesi Master Judilka Bermudez Tesi Master Judilka Bermudez
Tesi Master Judilka Bermudez
thinfilmsworkshop
 
A biologically inspired cmos image sensor
A biologically inspired cmos image sensorA biologically inspired cmos image sensor
A biologically inspired cmos image sensor
Springer
 
A biologically inspired cmos image sensor
A biologically inspired cmos image sensorA biologically inspired cmos image sensor
A biologically inspired cmos image sensor
Springer
 

Semelhante a Determination ofRadiological Quality Parameters Using Optical Densitometer andSimple Fabricated Equipment (20)

Thin films multilayer
Thin films multilayerThin films multilayer
Thin films multilayer
 
Photoacoustic technology for biological tissues characterization
Photoacoustic technology for biological tissues characterizationPhotoacoustic technology for biological tissues characterization
Photoacoustic technology for biological tissues characterization
 
S IGNAL A ND I MAGE P ROCESSING OF O PTICAL C OHERENCE T OMOGRAPHY AT 1310 NM...
S IGNAL A ND I MAGE P ROCESSING OF O PTICAL C OHERENCE T OMOGRAPHY AT 1310 NM...S IGNAL A ND I MAGE P ROCESSING OF O PTICAL C OHERENCE T OMOGRAPHY AT 1310 NM...
S IGNAL A ND I MAGE P ROCESSING OF O PTICAL C OHERENCE T OMOGRAPHY AT 1310 NM...
 
Neutron Imaging and Tomography with Medipix2 and Dental Microroentgenography:...
Neutron Imaging and Tomography with Medipix2 and Dental Microroentgenography:...Neutron Imaging and Tomography with Medipix2 and Dental Microroentgenography:...
Neutron Imaging and Tomography with Medipix2 and Dental Microroentgenography:...
 
Time-resolved mirrorless scintillation detector @ AAPM2018
Time-resolved mirrorless scintillation detector @ AAPM2018Time-resolved mirrorless scintillation detector @ AAPM2018
Time-resolved mirrorless scintillation detector @ AAPM2018
 
Design and Development of a Shortwave near Infrared Spectroscopy using NIR LE...
Design and Development of a Shortwave near Infrared Spectroscopy using NIR LE...Design and Development of a Shortwave near Infrared Spectroscopy using NIR LE...
Design and Development of a Shortwave near Infrared Spectroscopy using NIR LE...
 
Impact of detector thickness on imaging characteristics of the Siemens Biogra...
Impact of detector thickness on imaging characteristics of the Siemens Biogra...Impact of detector thickness on imaging characteristics of the Siemens Biogra...
Impact of detector thickness on imaging characteristics of the Siemens Biogra...
 
11
1111
11
 
Report In Japan 20060613 Liuxu
Report In Japan 20060613 LiuxuReport In Japan 20060613 Liuxu
Report In Japan 20060613 Liuxu
 
2011-IPP-CT Data Evaluation of Fibre Reinforced Polymers to Determine Fibre L...
2011-IPP-CT Data Evaluation of Fibre Reinforced Polymers to Determine Fibre L...2011-IPP-CT Data Evaluation of Fibre Reinforced Polymers to Determine Fibre L...
2011-IPP-CT Data Evaluation of Fibre Reinforced Polymers to Determine Fibre L...
 
MEASUREMENT OF ENTRANCE SKIN EXPOSURE (ESE) VALUE WITH VARIOUS TYPES OF FILTE...
MEASUREMENT OF ENTRANCE SKIN EXPOSURE (ESE) VALUE WITH VARIOUS TYPES OF FILTE...MEASUREMENT OF ENTRANCE SKIN EXPOSURE (ESE) VALUE WITH VARIOUS TYPES OF FILTE...
MEASUREMENT OF ENTRANCE SKIN EXPOSURE (ESE) VALUE WITH VARIOUS TYPES OF FILTE...
 
Tesi Master Judilka Bermudez
Tesi Master Judilka Bermudez Tesi Master Judilka Bermudez
Tesi Master Judilka Bermudez
 
A biologically inspired cmos image sensor
A biologically inspired cmos image sensorA biologically inspired cmos image sensor
A biologically inspired cmos image sensor
 
A biologically inspired cmos image sensor
A biologically inspired cmos image sensorA biologically inspired cmos image sensor
A biologically inspired cmos image sensor
 
MEMS BASED OPTICAL COHERENCE TOMOGRAPHY IMAGING SYSTEM AND OPTICAL BIOPSY PRO...
MEMS BASED OPTICAL COHERENCE TOMOGRAPHY IMAGING SYSTEM AND OPTICAL BIOPSY PRO...MEMS BASED OPTICAL COHERENCE TOMOGRAPHY IMAGING SYSTEM AND OPTICAL BIOPSY PRO...
MEMS BASED OPTICAL COHERENCE TOMOGRAPHY IMAGING SYSTEM AND OPTICAL BIOPSY PRO...
 
A MEMS BASED OPTICAL COHERENCE TOMOGRAPHY IMAGING SYSTEM AND OPTICAL BIOPSY P...
A MEMS BASED OPTICAL COHERENCE TOMOGRAPHY IMAGING SYSTEM AND OPTICAL BIOPSY P...A MEMS BASED OPTICAL COHERENCE TOMOGRAPHY IMAGING SYSTEM AND OPTICAL BIOPSY P...
A MEMS BASED OPTICAL COHERENCE TOMOGRAPHY IMAGING SYSTEM AND OPTICAL BIOPSY P...
 
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...
Diagnosis of Vertical Root Fracture Using Digital Radiography, Helical Comput...
 
New Noise Reduction Technique for Medical Ultrasound Imaging using Gabor Filt...
New Noise Reduction Technique for Medical Ultrasound Imaging using Gabor Filt...New Noise Reduction Technique for Medical Ultrasound Imaging using Gabor Filt...
New Noise Reduction Technique for Medical Ultrasound Imaging using Gabor Filt...
 
Time-resolved mirrorless scintillation detector @ KSMPRS2018
Time-resolved mirrorless scintillation detector @ KSMPRS2018Time-resolved mirrorless scintillation detector @ KSMPRS2018
Time-resolved mirrorless scintillation detector @ KSMPRS2018
 
Task 239766J.pptx
Task 239766J.pptxTask 239766J.pptx
Task 239766J.pptx
 

Mais de IOSR Journals

Mais de IOSR Journals (20)

A011140104
A011140104A011140104
A011140104
 
M0111397100
M0111397100M0111397100
M0111397100
 
L011138596
L011138596L011138596
L011138596
 
K011138084
K011138084K011138084
K011138084
 
J011137479
J011137479J011137479
J011137479
 
I011136673
I011136673I011136673
I011136673
 
G011134454
G011134454G011134454
G011134454
 
H011135565
H011135565H011135565
H011135565
 
F011134043
F011134043F011134043
F011134043
 
E011133639
E011133639E011133639
E011133639
 
D011132635
D011132635D011132635
D011132635
 
C011131925
C011131925C011131925
C011131925
 
B011130918
B011130918B011130918
B011130918
 
A011130108
A011130108A011130108
A011130108
 
I011125160
I011125160I011125160
I011125160
 
H011124050
H011124050H011124050
H011124050
 
G011123539
G011123539G011123539
G011123539
 
F011123134
F011123134F011123134
F011123134
 
E011122530
E011122530E011122530
E011122530
 
D011121524
D011121524D011121524
D011121524
 

Último

Digital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptxDigital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptx
MohamedFarag457087
 
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune WaterworldsBiogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Sérgio Sacani
 
CYTOGENETIC MAP................ ppt.pptx
CYTOGENETIC MAP................ ppt.pptxCYTOGENETIC MAP................ ppt.pptx
CYTOGENETIC MAP................ ppt.pptx
Silpa
 
POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.
Silpa
 
Phenolics: types, biosynthesis and functions.
Phenolics: types, biosynthesis and functions.Phenolics: types, biosynthesis and functions.
Phenolics: types, biosynthesis and functions.
Silpa
 
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptxTHE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
ANSARKHAN96
 
LUNULARIA -features, morphology, anatomy ,reproduction etc.
LUNULARIA -features, morphology, anatomy ,reproduction etc.LUNULARIA -features, morphology, anatomy ,reproduction etc.
LUNULARIA -features, morphology, anatomy ,reproduction etc.
Silpa
 

Último (20)

Dr. E. Muralinath_ Blood indices_clinical aspects
Dr. E. Muralinath_ Blood indices_clinical  aspectsDr. E. Muralinath_ Blood indices_clinical  aspects
Dr. E. Muralinath_ Blood indices_clinical aspects
 
Atp synthase , Atp synthase complex 1 to 4.
Atp synthase , Atp synthase complex 1 to 4.Atp synthase , Atp synthase complex 1 to 4.
Atp synthase , Atp synthase complex 1 to 4.
 
Digital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptxDigital Dentistry.Digital Dentistryvv.pptx
Digital Dentistry.Digital Dentistryvv.pptx
 
Site Acceptance Test .
Site Acceptance Test                    .Site Acceptance Test                    .
Site Acceptance Test .
 
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune WaterworldsBiogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
Biogenic Sulfur Gases as Biosignatures on Temperate Sub-Neptune Waterworlds
 
CYTOGENETIC MAP................ ppt.pptx
CYTOGENETIC MAP................ ppt.pptxCYTOGENETIC MAP................ ppt.pptx
CYTOGENETIC MAP................ ppt.pptx
 
PSYCHOSOCIAL NEEDS. in nursing II sem pptx
PSYCHOSOCIAL NEEDS. in nursing II sem pptxPSYCHOSOCIAL NEEDS. in nursing II sem pptx
PSYCHOSOCIAL NEEDS. in nursing II sem pptx
 
POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.POGONATUM : morphology, anatomy, reproduction etc.
POGONATUM : morphology, anatomy, reproduction etc.
 
Phenolics: types, biosynthesis and functions.
Phenolics: types, biosynthesis and functions.Phenolics: types, biosynthesis and functions.
Phenolics: types, biosynthesis and functions.
 
Chemistry 5th semester paper 1st Notes.pdf
Chemistry 5th semester paper 1st Notes.pdfChemistry 5th semester paper 1st Notes.pdf
Chemistry 5th semester paper 1st Notes.pdf
 
Cyanide resistant respiration pathway.pptx
Cyanide resistant respiration pathway.pptxCyanide resistant respiration pathway.pptx
Cyanide resistant respiration pathway.pptx
 
FAIRSpectra - Enabling the FAIRification of Spectroscopy and Spectrometry
FAIRSpectra - Enabling the FAIRification of Spectroscopy and SpectrometryFAIRSpectra - Enabling the FAIRification of Spectroscopy and Spectrometry
FAIRSpectra - Enabling the FAIRification of Spectroscopy and Spectrometry
 
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICEPATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
 
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptxTHE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
THE ROLE OF BIOTECHNOLOGY IN THE ECONOMIC UPLIFT.pptx
 
LUNULARIA -features, morphology, anatomy ,reproduction etc.
LUNULARIA -features, morphology, anatomy ,reproduction etc.LUNULARIA -features, morphology, anatomy ,reproduction etc.
LUNULARIA -features, morphology, anatomy ,reproduction etc.
 
Genetics and epigenetics of ADHD and comorbid conditions
Genetics and epigenetics of ADHD and comorbid conditionsGenetics and epigenetics of ADHD and comorbid conditions
Genetics and epigenetics of ADHD and comorbid conditions
 
Genome sequencing,shotgun sequencing.pptx
Genome sequencing,shotgun sequencing.pptxGenome sequencing,shotgun sequencing.pptx
Genome sequencing,shotgun sequencing.pptx
 
FAIRSpectra - Enabling the FAIRification of Analytical Science
FAIRSpectra - Enabling the FAIRification of Analytical ScienceFAIRSpectra - Enabling the FAIRification of Analytical Science
FAIRSpectra - Enabling the FAIRification of Analytical Science
 
Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.Proteomics: types, protein profiling steps etc.
Proteomics: types, protein profiling steps etc.
 
300003-World Science Day For Peace And Development.pptx
300003-World Science Day For Peace And Development.pptx300003-World Science Day For Peace And Development.pptx
300003-World Science Day For Peace And Development.pptx
 

Determination ofRadiological Quality Parameters Using Optical Densitometer andSimple Fabricated Equipment

  • 1. IOSR Journal of Applied Physics (IOSR-JAP) e-ISSN: 2278-4861.Volume 7, Issue 1 Ver. II (Jan.-Feb. 2015), PP 11-15 www.iosrjournals.org DOI: 10.9790/4861-07121115 www.iosrjournals.org 11 | Page Determination ofRadiological Quality Parameters Using Optical Densitometer andSimple Fabricated Equipment Kpaku Goriya1 ,Abdullahi Mundi1 , Aribisala Abdul-Jamiu2 , Peter Eshiett3 ,Richard Itopa4 ,DlamaZira Joseph5 , 1. Physics Department, Nasarawa state university Keffi, Nigeria 2. FCT Secondary Education Board, Abuja, Nigeria 3. Federal Medical Center Keffi, Nigeria 4. Department of Radiography, Bayero University Kano 5. Radiology DepartmentAbubakarTafawaBalewa University Teaching Hospital Bauchi, Nigeria. CORRESPONDING AUTHOR’S EMAIL5 :josephdlama@gmail.com Abstract: The etiquette of safe practice for the use of x-rays in medical diagnosis requires that each x-ray facility has an appropriate quality assurance program in radiation protection and to keep doses as low as reasonable achievable. An alternative method of evaluating radiological quality parameters of diagnostic x-ray machine is considered in this study. This study was conducted in Federal Medical Center Keffi, Nasarawa State. The aim was to determine radiological quality parameters using optical densitometer and simple fabricated equipment. This research has determined some of these parameters via the wooden step wedge and metal pin fabricated. The wedge and metal pin were exposed to x-ray and the optical densities of the processed films were measured with a densitometer. The result indicates that simple fabricated equipment can be used to determine the following radiological quality parameters such as dose, beam alignment, peak kilo voltage and tube current consistency in the absence of quality control test equipment. This is an indication that departmental protocol can be established in x-ray department using simple fabricated equipment .These equipment can be used in the determination of radiological quality parameters. Keywords:[Radiation Protection, Parameters, Optical Densitometer,Diagnostic, Equipment.] I. Introduction Diagnostic radiology machines have been of great help to health care industry and many people have benefited from the useful information they provide[1]. Providing standard health care infrastructure is one of the priorities of modern-day hospital in our country[2]. The investment in state –of the art equipment is vital in achieving this goal. It is essential to maintain the quality of high-end equipment procured to provide high quality health care. One of the difficult areas of maintaining this infrastructure is within the diagnostic radiology department[2],[3]. Radiation cannot be seen, heard or smelt hence, to check the quality of radiation producing equipment, intuitive methods of detection seldom fail[4]. Checking the quality of the image might give an indication of optimal performance of radiation equipment, but can be often misleading[4]. The quality of diagnostic radiation equipment directly affects the image quality and unwanted exposure to staff[5]. Poor diagnostic image quality can lead to diagnostic dilemma[5]. II. Materials And Methods The materials used for this study include  X–ray machine  Densitometer  Wooden step wedge  Uniform metal pin  Radiographic film (Kodak)  Micrometer screw gauge  Verniercaliper 2.1X–ray Machine Characteristics of the x–ray machine include  Tube voltage, 125kv  Total filtration, 2.9mmAl/70kv
  • 2. Determination Of Radiological Quality Parameters Using Optical Densitometer DOI: 10.9790/4861-07121115 www.iosrjournals.org 12 | Page  Tube serial number 8324/12  Model number XD52–30.50/125–T2A 2.2 Densitometer Characteristic and specification of densitometer Point densitometer is straight forward device used for determining the optical density (OD) at a few points in a film. This device is easiest to QA for absolute OD and is therefore often used as the local standards in OD measurements. Point densitometer use a silicon photodiode to measure the transmitted flux of light passing through a film. The international organization for standardization (ISO) has developed standards for the geometric and spectral conditions for the determination of optical density (ISO,5) Diffuse illumination is achieved by a broad–spectrum, incandescent lamp whose spectral properties conform to international commission on illumination (CIE) standards. The collimated light passes through the film and the transmitted component passes through a detecting aperture that usually can be from 1–3mm in diameter. This light flux passes through a V- filter (555nm peak – 780nm range) and is then detected by a silicon photodiode with amplifier electronic capable of measuring signals spanning several orders of magnitude. This system consist of a backlit tablet with light aperture on which one places the film sample, an arm which extends over the film and has at its end the light source aligned with the aperture and a digital readout of optical density (Sujatha, 2007). 2.3 Construction Iroko wood was obtained in the form of rectangular block of dimensions. Light, L = 10cm, width, w = 3.5cm, height, H = 3cm, mass, m = 25g, volume, v = 65cm3 . The construction was carried out at kpakurium workshop in Keffi, Nasarawa State. The wood was machined to square using the plane jack, to smoothen out rough edges and make the surface uniformly smooth throughout. Five steps of equal thickness and height were marked on the wooden block with pencil, to aid accuracy in cutting the steps. The difference in thickness between each step is 0.6cm (Martins, 2007). 2.4 Exposure Procedure  The wooden step wedge was placed on a 24 x 30cm cassette, a control exposure was made by keeping the mAs constant at 16mAs and varying the kVp, thus 45, 50, 55, 60, 65 and 70kVp respectively at constant film to focus distance (FFD) of 100cm, a total of six exposure were made.  The same procedure was repeated by keeping the kVp constant at 45kVp and varying the mAs, thus 2, 4, 6, 8, and 10mAs respectively at constant FFD of 100cm, a total of six exposure were also made in this case.  Film processing, this was done in the dark room with only the safe light on. The exposed films were removed from the cassettes and processed temperature of 0.3 degree Celsius. After processing and drying, the film was assessed on the view box, all the steps of the wedge were visible. 2.5 Measurement of optical density The measurement of optical density (O.D), was carried out using the transmission digital densitometer x–rite 331 nickel–cadmium (Nicad), battery operated (600/700mAhr). For the constant 16mAs films with the images of the wooden step wedge, the optical densities of the five steps were measured for the constant 45kVp film with the images of the wooden step wedge, the optical densities f for five steps were measured. 2.6 Simple beam alignment test A uniform metal pin of thickness 0.3cm, length 10cm was mounted 900 vertically on a rectangular block of wood 2 x 4cm and 0.5cm thick. The metal pin was placed on a cassette and expose to x–ray using a suitable exposure factor (4mAs x 50kVp) and FFD of 100cm (Glen, 1993). The tolerance angle () was calculated using trigonometry ratio (tan = a/b). The length of the image of the metal pin was traced on the x–ray film with a divider and a meter rule. The length of the metal pin and that of the image form two side of a triangle. With these information, the tolerance angle was calculated as shown in the 2.7Statistical Analysis In order to determine the effect of kVp and mAs on dose, a graph of these two parameters was plot against optical density using the Microsoft Excel and the SPSS Statistical Package III. Results 3.1 Fabricated Wooden Step Wedge and Metal Pin Five steps wooden wedge and 10cm metal pin were fabricated and used to determine the radiological quality parameters. The fabricated equipment are shown in figure 3.0 below
  • 3. Determination Of Radiological Quality Parameters Using Optical Densitometer DOI: 10.9790/4861-07121115 www.iosrjournals.org 13 | Page  Wooden step wedge (b) Metal Pin Figure 3.1: simple fabricated equipment Figure 3.2: radiograph of wooden step wedge at 16mAs and varying kVp Figure 3.3: A graph of optical density against thickness. Figure 3.4: A graphof tube charge against optical density. Table 3.0: Relationship between Optical Density and Dose . S/N Optical density Dose(Gy) 1 0.74 0.754 2 0.69 0.748 3 0.67 0.746 4 0.60 0.743 5 0.58 0.721 Table 3.1: Simple Beam Alignment Test 1 2 3 Average Length of metal pin(a)cm 10.00 10.00 10.00 10.00 Length of image(b)cm 0.20 0.30 0.20 0.20 Tolerance angle (degree) 1.14 1.70 1.14 1.32 Figure 3.5: Radiograph of 10cm metal pin
  • 4. Determination Of Radiological Quality Parameters Using Optical Densitometer DOI: 10.9790/4861-07121115 www.iosrjournals.org 14 | Page IV. Discussion The result from this study shows that, simple fabricated equipment can be used to determine radiological quality parameters .This is useful in diagnostic radiology where quality control (Q C) test equipment are not available. The step wedge method of checking the consistency of peak kilo voltage and tube charge of diagnostic x-ray machine does so in terms of optical density which represents the total exposure reaching the film and producing the image. A common problem among the hospitals in Nigeria is the lack of regular patients dose monitoring and quality control in diagnostic radiology. A major reason for this is the cost of running a standard radiation protection and quality assurance on the facilities.[8],[9],[10]. Figure 3.1 shows that at constant 16 mAs using the fifth step of the wedge ,as the kVp of the x-ray machine increases ,there is also an increase in optical density ,which confirm the relational equation between optical density and kVp which is O.D=C(kVp)n . From the graph (figure3.1) the power n is 2.997 and the constant value is C is 4×10-6 , which is within the recommended limit[15][16]. This is an indication that the kVp is consistent. This is in consonance with a study done by using aluminum step wedge in Jos university teaching hospital[13][15][18]. Figure 3.2 shows that optical density decreases with increase in thickness, the higher steps of the wedge are dense, thus absorb more radiation than they transmit. The lower steps on the other hand are not as thick and hence readily transmit radiation, absorbing none at all. This implies that attenuation increase with increase in thickness.[16],[19]. Figure 3.3 shows that at constant 45kVp as the tube charge increases, the optical density also increases .Higher values than 8mAs has higher optical density, but poor image contrast, this may be due to excessive dose. This implies that optical density is directly proportional to tube charge, which confirm the relational equation between optical density and tube charge.( O.D α mAs). Table 3 shows the relationship between optical density and dose which is obtained from linear path of the curve of a graph of optical density against thickness (figure 3.3). Y=0.077X+0.8 is the equation of the linear path of the curve, where Y is the dose and X is the optical density. The doses were obtained by substituting the values of optical density into the equation .The result obtained as presented in table 3, shows that as the optical density increases, dose also increases. This indicate that optical density can be use to estimate dose[16],[17]. Table 3.1 shows the average tolerance angle obtained from the simple beam alignment test which is 3o .This implies that the x-ray beam of the machine is aligned since the tolerance angle is within the recommended range ±5 [15],[17]. V. Summary of the Study Finding  As kVp of the x-ray machine increases, there is also increase in optical density. This implies that optical density is directly proportional to kVp. It means that change in potential can cause drifts in beam intensity and thus significantly influence radiation dose and image quality as well as after the film density.  As the tube charge of the x-ray machine increases, there is also increase in optical density. This implies that optical density is directly proportional to tube charge.  The optical density of the higher steps, are less than those of the lower steps. This implies that attenuation increases with increase in thickness, it indicate that optical density is inversely proportional to thickness.  As the optical density increases dose also increases. This implies that optical density is directly proportional to dose  The tolerance angle obtained from the simple beam alignment test is 3o ,which indicate that the x-ray machine is properly collimated and the beam is aligned  The following quality parameters were found to be consistence; kVp ,mAs, dose and beam alignment. VI. Conclusion The result from this study shows that simple fabricated equipment can be used to determine radiological quality parameters, such as; dose, beam alignment, kVp and mAsconsistency.This implies that, if the diagnostic x-ray machine is not maintaining the consistency of this parameters, it’ s output will be poor. References [1]. Ash, D & Bates, T. (1994) Report on the Clinical Effects of Inadvertent Radiation under Dosage in 1045 Patients, Clinical Oncol. World Journal of Medical Sciences 6:214 – 225. [2]. Akpa, T.C. (2013). Interaction of Radiation with Matter. A Lecture note on, Radiation physics, Nasarawa State University, Keffi (unpublished). [3]. Attix, F. H. (1986). Introduction to Radiological Physics and Radiation Dosimetry. Wiley, New York PP56 [4]. AERB Safety Code. (2001). Code No. AERB/SE/MED- 2, Mumbia PP 1 -20.
  • 5. Determination Of Radiological Quality Parameters Using Optical Densitometer DOI: 10.9790/4861-07121115 www.iosrjournals.org 15 | Page [5]. Bushong, S.C. (1993). Radiography Science for Technologist Physics, Biology and Protection. 5th Edition. St Lovis; Mmosby Year Book PP45-70 [6]. Beacomfield, T., Nicholson, R., & Thornton, A. (1999). Would Thyroid and Breast Shielding be Beneficial in CT of the Head? Eur- Radiol. World Journal of Medical Sciences. 8(4), pp664. [7]. Christian Nordgvist (2011).Effect of Radiation on Human http://www.medicalnewstoday.com/articles/2.1965.php. [8]. Curry, T. S., Dowdey, J.E& Marry, R. C. (1984). Introduction to Physics of Diagnostic Radiography 3rd Edition. Lea and Fibiger, Philadelphia, PA pp88 – 106. [9]. Edmonds, I.R. (1984). Calculation of Patient Skin Dose from Diagnostic x-ray Procedure. British Journal of Radiology, 57, 733 – 734. [10]. Egbe, N.O., Inyang, S. O &Eduwem, D.U. (2008). Doses and Image Quality for Chest Radiograph in three Nigerian Hospitals. European Journal of Radiography 1, 30 – 36. [11]. Golani industrial complex waliv, vasai (2013).In house Quality assurance of diagnostic radiology Equipment. www.raysafe.com. [12]. HananFawaz. (2007). Assessment of Entrance Skin Dose and Effective Dose of some Routine X-ray Examination using Calculation Technique.www.geogle. [13]. International Commission on Radiological Protection (1996). Protection and Safety in Medicine, Publication 73, Pergamon Press, Oxford. [14]. John & Brenan. (2000). Reference Dose level for Patients undergoing Common Diagnostic x-rays Examination in Irish Hospital. British Journal of Radiology 73, 396 – 402. [15]. Mangset, W.E. &Izang, N. (2010). Locally Fabricated Metal Step Wedge for Quality Assurance in Diagnostic Radiology. Department of Physics, University of Jos. Bayero Journal of Pure and Applied Sciences, 3(1); 156 – 163 [16]. Martin, C.T. (2007). Optimization in General Radiography www.biij.org/2007/2/e/8 [17]. Nworgu, O. D &Bamidele, L. (2010). An Evaluation of Entrance Surgace doses from some Common Diagnostic x-ray Examinations in two Nigerian University Teaching. Journal of Radiological Protection Vol. 29, N2a, 952-4746. [18]. Rehani, M.M.(1995). Diagnostic Imaging Quality Assurance. Jaypee Brothers Medical Publisher.New York, pp45. [19]. Sujatha, P. (2007). Radiographic Film for Mega Voltage Beam Dosimetry. Radiation Therapy Department, Memorial Hermann Hospital, Houston. Texas 77024, pp2239. [20]. Sato, O., Iwai, S &Farihata, S. (1995). Calculation of Equivalent Dose and Effective Dose Conservation Coefficients for Photon. Radiation Protection Dosimetry. Nuclear Technology Publishing, Japan, Vol.62, No. 3, pp119 -130.