Application of FMEA to a Sterility
Testing Isolator: A Case Study
Dr. Tim Sandle
www.pharmamicreouresources.com
Introduction
 Introduction to risk assessment
 What are risks?
 Advantages and disadvantages of
FMEA
 Applying FMEA to review a sterility
testing isolator – case study
Risk Assessment
Risk Assessment
 Increasingly used tool in the pharmaceutical
sector.
 An expectation of regulatory authorities.
 Important because:
– A proactive tool;
– A reactive tool;
– To explore weaknesses and seek improvements;
– To construct rationales;
– Part of the qualification and validation of
processes.
Risk is ever present
 THERE IS NO SUCH THING AS
“ZERO” RISK
 Need to understand, “quantify” and manage risks in
pharmaceuticals and healthcare.
 Risk is defined as the combination of the probability of
occurrence of harm and the severity of that harm i.e.
– What might go wrong?
– What is the likelihood (probability) it will go wrong?
– What are the consequences (severity)?
 Risks relate to a situation where a recognized hazard may
result in harm.
Risk Assessment
 Different approaches for risk assessment
including:
– HACCP (Hazard Analysis Critical Control Points),
which has its origins in the food industry;
– Fault Tree Analysis (FTA) ;
– Modelling software (such as the Monte Carol
model);
– FMEA (Failure Mode and Effects Analysis), which
originated in the engineering sector.
– Several described in ICH Q9 and EU GMP Annex
20.
Risk Assessment
 Important
steps:
– Define problem;
– Select tool;
– Hazard identification;
– Risk assessment;
– Risk control;
– Risk review;
– Risk communication.
ICH Q9
Risk Assessment
 A more detailed approach involves:
– Gathering data through an audit and analysis;
– Constructing diagrams of work flows;
– Pin-pointing areas of greatest risk;
– Examining potential sources of contamination;
– Deciding on the most appropriate sample methods;
– Helping to establish alert and action levels;
– Taking into account changes to the work process /
seasonal activities;
– Using some type of scoring system so that the risk can be
ranked and the level of risk determined.
FMEA
What is it?
FMEA
 FMEA is “Failure mode and effects analysis”
 It is an analytical tool which was originated by the
US military and is widely used in the engineering
industry
 Very structured approach (ISO/TS 16949)
 FMEA is applied to many areas as a problem solving
tool
 This talk adopts one possible approach based on
the approach of the non-commercial FMEA
Information Centre:
http://www.fmeainfocentre.com/introductions.htm
FMEA
 FMEA looks for failure modes.
 A failure mode is a characterization of
the way a product or process fails.
 The term may be applied to
mechanical failure, structural failure,
electrical failure, biological risks and
systems failure.
FMEA
 Advantages:
– Improve product/process reliability and quality
– Increase customer satisfaction
– Early identification and elimination of potential
product/process failure modes
– Prioritize product/process deficiencies
– Capture engineering/organization knowledge
– Emphasizes problem prevention
– Documents risk and actions taken to reduce risk
– Provide focus for improved testing and development
– Minimizes late changes and associated cost
– Catalyst for teamwork and idea exchange between
functions
FMEA
 Disadvantages:
– It is subjective;
– It has a long-drawn-out approach;
– The focus is on failure / non-conformance
types and not on the chain of events
(cause / effect);
– It tends only to focus on major issues;
– Not ideal for environmental monitoring –
HACCP is better.
FMEA
 FMEA
 Dangers:
– Looks at ‘detection’ as a risk mitigation,
which influences the determined score
 Need to be careful with microbiological data
since our methods have poor detectability
 With microbiological risks, focus on severity
and likelihood
– Application of a ‘score’ is subjective.
FMEA
Determining
failure mode
Assessing
severity
Assigning
probability
number
Assigning
detection
number
Calculating risk
priority
number
FMEA
 FMEA steps:
– Setting the scope;
– Defining the problem;
– Setting scales for factors of severity,
occurrence and detection (see later);
– Process mapping;
– Defining failure modes;
– Listing the potential effects of each failure
mode;
– Assigning severity ratings to each process
step;
FMEA
 Steps continued:
– Listing potential causes of each failure mode;
– Assigning and occurrence rating for each
failure mode;
– Examining current controls;
– Examining mechanisms for detection;
– Calculating the risk;
– Examining outcomes and proposing actions to
minimise risks.
 Ideally it should be team based.
FMEA case study
Sterility Testing Isolator
Isolator Study
 Despite a superiority to cleanrooms, all Isolators
are at risk from contamination
 The approach of regulators, such as the FDA, is:
 “Barrier Isolators cannot prevent contamination
caused by GMP deficiencies such as poor aseptic
procedures and inadequate training of…operators”
(The Gold Sheet, Vol. 32, No.10)
Isolator Study
 Description:
– From a pharmaceutical manufacturer based in the
south-east of England.
– The Isolator was one half-suit Isolator, two
transfer Isolators and a steriliser unit.
– Positive pressure, flexible film Isolators with
stainless steel frames and wood bases designed
for aseptic processes (in this case: sterility testing
to Ph. Eur. 2.6.1).
– Air is using HEPA filters and material is transferred
into and out of the main Isolator using transfer
Isolators connected using Rapid Transfer Ports
(RPT).
– Sanitised by hydrogen peroxide vapour.
– The internal environment is classed as Grade A /
ISO 5
Isolator Study
 Identifying the main risks:
– Leaks;
– Gloves / operator manipulations;
– Filters;
– Other airborne contamination;
– Transfer of material into and out of the Isolator;
– The Isolator room;
– Decontamination cycle;
– Cleaning / environmental monitoring issues.
Isolator Study
 Designing the FMEA scheme
– FMEA schemes vary in their approach, scoring
and categorisation.
– All approaches share in common a numerical
approach. The approach adopted was to assign
a score (from 1 to 5) to each of the following
categories:
i) Severity
ii) Occurrence (or probability)
iii) Detection
Isolator Study
i) Severity is the consequence of a failure,
should it occur;
ii) Occurrence is the likelihood of the failure
happening (based on past experience);
iii) Detection is based on the monitoring
systems in place and on how likely a failure
can be detected.
A good detection system is one that can
detect a failure before it occurs.
Isolator Study
 A scale from 1 to 5. It followed that
the likelihood of high severity would
be rated 5; high occurrence rated 5;
but a good detection system would be
rated 1.
 See over….
Isolator Study
Severity 5 Specification limits exceeded. Probable rejection of
test or shutdown of system.
3 Observed trend takes place, but no critical
excursions. Requires investigation.
1 No excursion has taken place. No upward trends
and no investigation is required.
Occurrence 5 Expected to occur 50% time.
3 Expected to occur ≥10 - ≤50% time.
1 Expected to occur ≤10%.
Detection 5 No way to detect the failure mode.
3 Can be partially detected but detection could be
improved.
1 Good detection systems in place.
Isolator Study
 Using these criteria a final FMEA score is
produced (sometimes called a Risk Priority
Number):
x
125
 The total of 125 is derived from: severity
score x occurrence score x detect score, or:
5 x 5 x 5 = 125
Isolator Study
 A score of 27 was the cut-off value: where
action was required.
 Based on 27 being the score derived when
the mid-score is applied to all three
categories
The numerical value '3' from:
Severity (3) x Occurrence (3) x Detection (3)
The supposition that if the mid-rating (or a higher number) was scored for all three
categories then as a minimum the system should be examined in greater detail.
Application of FMEA
Some examples
Examples
 3 examples
– Potential for sanitisation cycle failure
– Pressure leaks to gloves
– Connection of transfer Isolator to main Isolator and
transfer-in / out of material
Isolator Example 1
Potential for
sanitisation
cycle failure
#1
Process step Failure
Mode
Significance
of failure
Severity
of
conseque
nce
(score)
Performing
sanitisation
cycles on
transfer or main
Isolator
An
Isolator is
not
correctly
sanitised
Contaminated
items enter
main Isolator
or main
Isolator itself is
contaminated
4
Isolator Example 1
Potential for
sanitisation
cycle failure
#2
Measures to
detect failure
Occurrenc
e
(score)
Detection
systems
Detection
(score)
Evaporation rate
/ pre- and post-
lot testing of
acid /
sanitisation
cycles developed
using BIs
1 Steriliser
parameters
checked after
sanitisation
and before
use / acid
potency
checked for
each lot /
post-
sanitisation
environmental
monitoring
performed for
main Isolator
1
Isolator Example 1
Potential for
sanitisation
cycle failure
#3
FMEA score:
4 x 1 x 1 = 4
Isolator Example 2
Pressure
leaks to
gloves
#1
Process step Failure
Mode
Significance
of failure
Severity
of
consequ
ence
(score)
Use of gloves
to transfer
material or to
perform sterility
test (sterile
gloves may be
worn
underneath
Isolator gloves)
Contaminati
on from
technician
into Isolator
or weak
area of
positive
pressure to
allow
contaminati
on in
Contaminati
on present
in Isolator /
compromise
of aseptic
technique
4
Isolator Example 2
Pressure
leaks to
gloves
#2
Measures to
detect failure
Occurre
nce
(score
)
Detection
systems
Detection
(score)
Environment
al monitoring
(post-use
finger
plates) /
pressure
charts
2 Environme
ntal
monitoring
is
performed
post-test
on gloves /
gloves are
wiped with
disinfectan
t / gloves
are visually
examined
weekly and
changed as
appropriat
e
3
Isolator Example 2
Pressure
leaks to
gloves
#3
FMEA score:
4 x 2 x 3 = 24
Isolator Example 2
 However, on
review….
FMEA score:
4 x 2 x 1 = 8
Isolator Example 3
The transfer of material in and out of the Isolator is, arguably, the biggest risk:
Non-sterile area between doors
*
*Area of biggest risk
Transfer isolator
Isolator Example 3
Connection of
transfer Isolator
to main Isolator
and transfer-in
/ out of
material #1
Process step Failure Mode Significance
of failure
Severity
of
conseque
nce
(score)
Connection of
transfer
Isolator to
main Isolator
and moving
material in and
out
Contaminat
ion on
outside of
both
Isolators
may enter
the main
Isolator /
failure to
maintain
positive
pressure
Contamina
tion enters
the
Isolator or
positive
pressure is
not
maintained
4
Isolator Example 3
Connection of
transfer Isolator
to main Isolator
and transfer-in
/ out of
material #2
Measures to
detect failure
Occurrence
(score)
Detection
systems
Detection
(score)
Environmenta
l monitoring /
pressure
monitoring
1 DPTE seal
system / use
of
disinfectant
for
connection
1
Isolator Example 3
Connection of
transfer Isolator
to main Isolator
and transfer-in
/ out of
material #3
FMEA score:
4 x 1 x 1 = 4
Isolator study
 Revisit the ranking
 Define residual risk
 Perform a short summary
– Scope
– Data from the assessment & control
(e.g. no. of identified failure modes)
– Level of accepted risk without actions i.e. residual
risk
(e.g. risk priority Number < 27)
– Recommended actions, responsibilities and due dates
(including approval, if appropriate)
– Person in charge for follow-up of FMEA
Isolator Study
 Summary of the entire study:
Isolator FMEA risk assessment
0
5
10
15
20
25
30
R
oom
C
ycle
Frequency
IntegrotyC
onnectionS
anitisation
P
hysical
G
loves
Category
FMEAscore
Cut off score
Further examples
Sandle, T. ‘The use of a risk assessment
in the pharmaceutical industry – the
application of FMEA to a sterility testing
isolator: a case study’, European Journal
of Parenteral and Pharmaceutical
Sciences, 8(2): 43-49
FMEA
 A risk assessment technique – FMEA
can be readily applied to a key
operation
 This technique did not originate in the
pharmaceutical industry.
 This indicates how the synergy of
different approaches can be achieved.
FMEA
 Regular reviews must take place;
 FMEA is not suitable for everything e.g.
HACCP may be more suitable for suitable for
aseptic filling.
 It is not able to discover complex failure
modes involving multiple failures or
subsystems, or to discover expected failure
intervals of particular failure modes.
Thank you
Dr. Tim Sandle
www.pharmamicroresources.com
Any questions?

Mais conteúdo relacionado

Mais procurados

Fmea alignment aiag_and_vda_-_eng
Fmea alignment aiag_and_vda_-_engFmea alignment aiag_and_vda_-_eng
Fmea alignment aiag_and_vda_-_eng
erkinguler
 
Failure Mode & Effect Analysis
Failure Mode & Effect AnalysisFailure Mode & Effect Analysis
Failure Mode & Effect Analysis
ECC International
 

Mais procurados (20)

FMEA and FMECA
FMEA and FMECAFMEA and FMECA
FMEA and FMECA
 
Fault tree analysis
Fault tree analysisFault tree analysis
Fault tree analysis
 
FMEA Introduction.ppt
FMEA Introduction.pptFMEA Introduction.ppt
FMEA Introduction.ppt
 
Fmea alignment aiag_and_vda_-_eng
Fmea alignment aiag_and_vda_-_engFmea alignment aiag_and_vda_-_eng
Fmea alignment aiag_and_vda_-_eng
 
Hazard assessment and risk management techniques
Hazard assessment and risk management techniquesHazard assessment and risk management techniques
Hazard assessment and risk management techniques
 
Fault Tree Analysis
 Fault Tree Analysis Fault Tree Analysis
Fault Tree Analysis
 
091814 Failure Mode Effect and Analysis (FMEA)
091814 Failure Mode Effect and Analysis (FMEA)091814 Failure Mode Effect and Analysis (FMEA)
091814 Failure Mode Effect and Analysis (FMEA)
 
Preliminary hazard analysis (pha)
Preliminary hazard analysis (pha)Preliminary hazard analysis (pha)
Preliminary hazard analysis (pha)
 
Capa
CapaCapa
Capa
 
Failure Modes & Effects Analysis (FMEA)
Failure Modes & Effects Analysis (FMEA)Failure Modes & Effects Analysis (FMEA)
Failure Modes & Effects Analysis (FMEA)
 
Root Cause Analysis
Root Cause AnalysisRoot Cause Analysis
Root Cause Analysis
 
Root cause analysis
Root cause analysisRoot cause analysis
Root cause analysis
 
Failure Mode & Effect Analysis
Failure Mode & Effect AnalysisFailure Mode & Effect Analysis
Failure Mode & Effect Analysis
 
FMEA Presentation
FMEA PresentationFMEA Presentation
FMEA Presentation
 
Failure modes effect analysis
Failure modes effect analysisFailure modes effect analysis
Failure modes effect analysis
 
Introduction to Failure Mode and Effects Analysis (FMEA) in TQM
Introduction to Failure Mode and Effects Analysis (FMEA) in TQMIntroduction to Failure Mode and Effects Analysis (FMEA) in TQM
Introduction to Failure Mode and Effects Analysis (FMEA) in TQM
 
FMEA
FMEAFMEA
FMEA
 
Spc training
Spc training Spc training
Spc training
 
7 qc tools
7 qc tools7 qc tools
7 qc tools
 
FMEA - Failure mode and effects analysis
FMEA - Failure mode and effects analysisFMEA - Failure mode and effects analysis
FMEA - Failure mode and effects analysis
 

Semelhante a Application of FMEA to a Sterility Testing Isolator: A Case Study

11. faliure mode &amp; effect analysis
11. faliure mode &amp; effect analysis11. faliure mode &amp; effect analysis
11. faliure mode &amp; effect analysis
Hakeem-Ur- Rehman
 
IT 381_Chap hubybybybybybybytggyguh7h7_7.ppt
IT 381_Chap hubybybybybybybytggyguh7h7_7.pptIT 381_Chap hubybybybybybybytggyguh7h7_7.ppt
IT 381_Chap hubybybybybybybytggyguh7h7_7.ppt
ksujith0034
 
Risk minor major critical
Risk minor major criticalRisk minor major critical
Risk minor major critical
yasser mekky
 

Semelhante a Application of FMEA to a Sterility Testing Isolator: A Case Study (20)

Risk analysis
Risk analysis  Risk analysis
Risk analysis
 
Quality risk management
Quality risk managementQuality risk management
Quality risk management
 
Hazards and risk management
Hazards and risk managementHazards and risk management
Hazards and risk management
 
Hazard analysis
Hazard analysisHazard analysis
Hazard analysis
 
Risk Management in Sterile Environments
Risk Management in Sterile Environments Risk Management in Sterile Environments
Risk Management in Sterile Environments
 
Quality risk management
Quality risk managementQuality risk management
Quality risk management
 
ICH Q9 QUALITY RISK MANAGEMENT(QRM)
ICH Q9 QUALITY RISK MANAGEMENT(QRM)ICH Q9 QUALITY RISK MANAGEMENT(QRM)
ICH Q9 QUALITY RISK MANAGEMENT(QRM)
 
11. faliure mode &amp; effect analysis
11. faliure mode &amp; effect analysis11. faliure mode &amp; effect analysis
11. faliure mode &amp; effect analysis
 
FMEA
FMEAFMEA
FMEA
 
IT 381_Chap_7.ppt
IT 381_Chap_7.pptIT 381_Chap_7.ppt
IT 381_Chap_7.ppt
 
Failure mode effects analysis, Computer Integrated Manufacturing, Quality fun...
Failure mode effects analysis, Computer Integrated Manufacturing, Quality fun...Failure mode effects analysis, Computer Integrated Manufacturing, Quality fun...
Failure mode effects analysis, Computer Integrated Manufacturing, Quality fun...
 
FMEA presentation for research and development
FMEA presentation for research and developmentFMEA presentation for research and development
FMEA presentation for research and development
 
FMEA análise e odos de falhas e seus efeitos.ppt
FMEA análise e odos de falhas e seus efeitos.pptFMEA análise e odos de falhas e seus efeitos.ppt
FMEA análise e odos de falhas e seus efeitos.ppt
 
IT 381_Chap hubybybybybybybytggyguh7h7_7.ppt
IT 381_Chap hubybybybybybybytggyguh7h7_7.pptIT 381_Chap hubybybybybybybytggyguh7h7_7.ppt
IT 381_Chap hubybybybybybybytggyguh7h7_7.ppt
 
Risk minor major critical
Risk minor major criticalRisk minor major critical
Risk minor major critical
 
Risk management in pharmaceutical Industry
Risk management in pharmaceutical IndustryRisk management in pharmaceutical Industry
Risk management in pharmaceutical Industry
 
FMEA.pptx
FMEA.pptxFMEA.pptx
FMEA.pptx
 
Quality Risk Management 2 brief note
Quality  Risk Management 2 brief noteQuality  Risk Management 2 brief note
Quality Risk Management 2 brief note
 
Risk management and environmental monitoring
Risk management and environmental monitoringRisk management and environmental monitoring
Risk management and environmental monitoring
 
Failure Mode Effect Analysis in Engineering Failures
Failure Mode Effect Analysis in Engineering FailuresFailure Mode Effect Analysis in Engineering Failures
Failure Mode Effect Analysis in Engineering Failures
 

Mais de Tim Sandle, Ph.D.

Mais de Tim Sandle, Ph.D. (20)

Reviewing environmental monitoring.ppt
Reviewing environmental monitoring.pptReviewing environmental monitoring.ppt
Reviewing environmental monitoring.ppt
 
Open discussion on rapid microbiological methods.pptx
Open discussion on rapid microbiological methods.pptxOpen discussion on rapid microbiological methods.pptx
Open discussion on rapid microbiological methods.pptx
 
Electronic Data Management Systems.ppt
Electronic Data Management Systems.pptElectronic Data Management Systems.ppt
Electronic Data Management Systems.ppt
 
Audit efficiency storyboard.pptx
Audit efficiency storyboard.pptxAudit efficiency storyboard.pptx
Audit efficiency storyboard.pptx
 
Anomalies, complaints and non-compliances
Anomalies, complaints and non-compliancesAnomalies, complaints and non-compliances
Anomalies, complaints and non-compliances
 
Pharma micro myths (sandle)
Pharma micro myths (sandle)Pharma micro myths (sandle)
Pharma micro myths (sandle)
 
Operation of Sterility Testing Isolators and validation issues
Operation of Sterility Testing Isolators and validation issuesOperation of Sterility Testing Isolators and validation issues
Operation of Sterility Testing Isolators and validation issues
 
Pharmaceutical Microbiology: Current and Future Challenges
Pharmaceutical Microbiology: Current and Future Challenges Pharmaceutical Microbiology: Current and Future Challenges
Pharmaceutical Microbiology: Current and Future Challenges
 
The selection and use of reference materials
The selection and use of reference materialsThe selection and use of reference materials
The selection and use of reference materials
 
Sterility assurance and microbiology awareness
Sterility assurance and microbiology awarenessSterility assurance and microbiology awareness
Sterility assurance and microbiology awareness
 
Introduction to GxP
Introduction to GxPIntroduction to GxP
Introduction to GxP
 
Depyrogenation by dry heat
Depyrogenation by dry heatDepyrogenation by dry heat
Depyrogenation by dry heat
 
EU GMP Annex1 Review
EU GMP Annex1 ReviewEU GMP Annex1 Review
EU GMP Annex1 Review
 
The Black death - a re-emerging infectious disease
The Black death - a re-emerging infectious diseaseThe Black death - a re-emerging infectious disease
The Black death - a re-emerging infectious disease
 
USP &lt;1116> and its impact on Microbiology
USP &lt;1116> and its impact on MicrobiologyUSP &lt;1116> and its impact on Microbiology
USP &lt;1116> and its impact on Microbiology
 
ISO 14644 - introducing the revised standard
ISO 14644 - introducing the revised standardISO 14644 - introducing the revised standard
ISO 14644 - introducing the revised standard
 
Myths of pharmaceutical microbiology
Myths of pharmaceutical microbiologyMyths of pharmaceutical microbiology
Myths of pharmaceutical microbiology
 
Lal presentation
Lal presentation Lal presentation
Lal presentation
 
Cleanroom sop slides
Cleanroom sop slidesCleanroom sop slides
Cleanroom sop slides
 
Application of microbiological data
Application of microbiological dataApplication of microbiological data
Application of microbiological data
 

Último

Electrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdfElectrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdf
MedicoseAcademics
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan 087776558899
 
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
Sheetaleventcompany
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Sheetaleventcompany
 
Control of Local Blood Flow: acute and chronic
Control of Local Blood Flow: acute and chronicControl of Local Blood Flow: acute and chronic
Control of Local Blood Flow: acute and chronic
MedicoseAcademics
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
rajnisinghkjn
 
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
Sheetaleventcompany
 
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Sheetaleventcompany
 
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan CytotecJual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
jualobat34
 
👉Chandigarh Call Girl Service📲Niamh 8868886958 📲Book 24hours Now📲👉Sexy Call G...
👉Chandigarh Call Girl Service📲Niamh 8868886958 📲Book 24hours Now📲👉Sexy Call G...👉Chandigarh Call Girl Service📲Niamh 8868886958 📲Book 24hours Now📲👉Sexy Call G...
👉Chandigarh Call Girl Service📲Niamh 8868886958 📲Book 24hours Now📲👉Sexy Call G...
Sheetaleventcompany
 

Último (20)

Electrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdfElectrocardiogram (ECG) physiological basis .pdf
Electrocardiogram (ECG) physiological basis .pdf
 
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
Cara Menggugurkan Kandungan Dengan Cepat Selesai Dalam 24 Jam Secara Alami Bu...
 
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
Premium Call Girls Nagpur {9xx000xx09} ❤️VVIP POOJA Call Girls in Nagpur Maha...
 
Control of Local Blood Flow: acute and chronic
Control of Local Blood Flow: acute and chronicControl of Local Blood Flow: acute and chronic
Control of Local Blood Flow: acute and chronic
 
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
Gastric Cancer: Сlinical Implementation of Artificial Intelligence, Synergeti...
 
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Shahdol Just Call 8250077686 Top Class Call Girl Service Available
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
 
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
❤️Call Girl Service In Chandigarh☎️9814379184☎️ Call Girl in Chandigarh☎️ Cha...
 
Cardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their RegulationCardiac Output, Venous Return, and Their Regulation
Cardiac Output, Venous Return, and Their Regulation
 
Intramuscular & Intravenous Injection.pptx
Intramuscular & Intravenous Injection.pptxIntramuscular & Intravenous Injection.pptx
Intramuscular & Intravenous Injection.pptx
 
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
💰Call Girl In Bangalore☎️63788-78445💰 Call Girl service in Bangalore☎️Bangalo...
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
 
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
 
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
Gorgeous Call Girls Dehradun {8854095900} ❤️VVIP ROCKY Call Girls in Dehradun...
 
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan CytotecJual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
Jual Obat Aborsi Di Dubai UAE Wa 0838-4800-7379 Obat Penggugur Kandungan Cytotec
 
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service AvailableCall Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
Call Girls Mussoorie Just Call 8854095900 Top Class Call Girl Service Available
 
👉Chandigarh Call Girl Service📲Niamh 8868886958 📲Book 24hours Now📲👉Sexy Call G...
👉Chandigarh Call Girl Service📲Niamh 8868886958 📲Book 24hours Now📲👉Sexy Call G...👉Chandigarh Call Girl Service📲Niamh 8868886958 📲Book 24hours Now📲👉Sexy Call G...
👉Chandigarh Call Girl Service📲Niamh 8868886958 📲Book 24hours Now📲👉Sexy Call G...
 

Application of FMEA to a Sterility Testing Isolator: A Case Study

  • 1. Application of FMEA to a Sterility Testing Isolator: A Case Study Dr. Tim Sandle www.pharmamicreouresources.com
  • 2. Introduction  Introduction to risk assessment  What are risks?  Advantages and disadvantages of FMEA  Applying FMEA to review a sterility testing isolator – case study
  • 4. Risk Assessment  Increasingly used tool in the pharmaceutical sector.  An expectation of regulatory authorities.  Important because: – A proactive tool; – A reactive tool; – To explore weaknesses and seek improvements; – To construct rationales; – Part of the qualification and validation of processes.
  • 5. Risk is ever present  THERE IS NO SUCH THING AS “ZERO” RISK  Need to understand, “quantify” and manage risks in pharmaceuticals and healthcare.  Risk is defined as the combination of the probability of occurrence of harm and the severity of that harm i.e. – What might go wrong? – What is the likelihood (probability) it will go wrong? – What are the consequences (severity)?  Risks relate to a situation where a recognized hazard may result in harm.
  • 6. Risk Assessment  Different approaches for risk assessment including: – HACCP (Hazard Analysis Critical Control Points), which has its origins in the food industry; – Fault Tree Analysis (FTA) ; – Modelling software (such as the Monte Carol model); – FMEA (Failure Mode and Effects Analysis), which originated in the engineering sector. – Several described in ICH Q9 and EU GMP Annex 20.
  • 7. Risk Assessment  Important steps: – Define problem; – Select tool; – Hazard identification; – Risk assessment; – Risk control; – Risk review; – Risk communication.
  • 9. Risk Assessment  A more detailed approach involves: – Gathering data through an audit and analysis; – Constructing diagrams of work flows; – Pin-pointing areas of greatest risk; – Examining potential sources of contamination; – Deciding on the most appropriate sample methods; – Helping to establish alert and action levels; – Taking into account changes to the work process / seasonal activities; – Using some type of scoring system so that the risk can be ranked and the level of risk determined.
  • 11. FMEA  FMEA is “Failure mode and effects analysis”  It is an analytical tool which was originated by the US military and is widely used in the engineering industry  Very structured approach (ISO/TS 16949)  FMEA is applied to many areas as a problem solving tool  This talk adopts one possible approach based on the approach of the non-commercial FMEA Information Centre: http://www.fmeainfocentre.com/introductions.htm
  • 12. FMEA  FMEA looks for failure modes.  A failure mode is a characterization of the way a product or process fails.  The term may be applied to mechanical failure, structural failure, electrical failure, biological risks and systems failure.
  • 13. FMEA  Advantages: – Improve product/process reliability and quality – Increase customer satisfaction – Early identification and elimination of potential product/process failure modes – Prioritize product/process deficiencies – Capture engineering/organization knowledge – Emphasizes problem prevention – Documents risk and actions taken to reduce risk – Provide focus for improved testing and development – Minimizes late changes and associated cost – Catalyst for teamwork and idea exchange between functions
  • 14. FMEA  Disadvantages: – It is subjective; – It has a long-drawn-out approach; – The focus is on failure / non-conformance types and not on the chain of events (cause / effect); – It tends only to focus on major issues; – Not ideal for environmental monitoring – HACCP is better.
  • 15. FMEA  FMEA  Dangers: – Looks at ‘detection’ as a risk mitigation, which influences the determined score  Need to be careful with microbiological data since our methods have poor detectability  With microbiological risks, focus on severity and likelihood – Application of a ‘score’ is subjective.
  • 17. FMEA  FMEA steps: – Setting the scope; – Defining the problem; – Setting scales for factors of severity, occurrence and detection (see later); – Process mapping; – Defining failure modes; – Listing the potential effects of each failure mode; – Assigning severity ratings to each process step;
  • 18. FMEA  Steps continued: – Listing potential causes of each failure mode; – Assigning and occurrence rating for each failure mode; – Examining current controls; – Examining mechanisms for detection; – Calculating the risk; – Examining outcomes and proposing actions to minimise risks.  Ideally it should be team based.
  • 19. FMEA case study Sterility Testing Isolator
  • 20. Isolator Study  Despite a superiority to cleanrooms, all Isolators are at risk from contamination  The approach of regulators, such as the FDA, is:  “Barrier Isolators cannot prevent contamination caused by GMP deficiencies such as poor aseptic procedures and inadequate training of…operators” (The Gold Sheet, Vol. 32, No.10)
  • 21. Isolator Study  Description: – From a pharmaceutical manufacturer based in the south-east of England. – The Isolator was one half-suit Isolator, two transfer Isolators and a steriliser unit. – Positive pressure, flexible film Isolators with stainless steel frames and wood bases designed for aseptic processes (in this case: sterility testing to Ph. Eur. 2.6.1). – Air is using HEPA filters and material is transferred into and out of the main Isolator using transfer Isolators connected using Rapid Transfer Ports (RPT). – Sanitised by hydrogen peroxide vapour. – The internal environment is classed as Grade A / ISO 5
  • 22. Isolator Study  Identifying the main risks: – Leaks; – Gloves / operator manipulations; – Filters; – Other airborne contamination; – Transfer of material into and out of the Isolator; – The Isolator room; – Decontamination cycle; – Cleaning / environmental monitoring issues.
  • 23. Isolator Study  Designing the FMEA scheme – FMEA schemes vary in their approach, scoring and categorisation. – All approaches share in common a numerical approach. The approach adopted was to assign a score (from 1 to 5) to each of the following categories: i) Severity ii) Occurrence (or probability) iii) Detection
  • 24. Isolator Study i) Severity is the consequence of a failure, should it occur; ii) Occurrence is the likelihood of the failure happening (based on past experience); iii) Detection is based on the monitoring systems in place and on how likely a failure can be detected. A good detection system is one that can detect a failure before it occurs.
  • 25. Isolator Study  A scale from 1 to 5. It followed that the likelihood of high severity would be rated 5; high occurrence rated 5; but a good detection system would be rated 1.  See over….
  • 26. Isolator Study Severity 5 Specification limits exceeded. Probable rejection of test or shutdown of system. 3 Observed trend takes place, but no critical excursions. Requires investigation. 1 No excursion has taken place. No upward trends and no investigation is required. Occurrence 5 Expected to occur 50% time. 3 Expected to occur ≥10 - ≤50% time. 1 Expected to occur ≤10%. Detection 5 No way to detect the failure mode. 3 Can be partially detected but detection could be improved. 1 Good detection systems in place.
  • 27. Isolator Study  Using these criteria a final FMEA score is produced (sometimes called a Risk Priority Number): x 125  The total of 125 is derived from: severity score x occurrence score x detect score, or: 5 x 5 x 5 = 125
  • 28. Isolator Study  A score of 27 was the cut-off value: where action was required.  Based on 27 being the score derived when the mid-score is applied to all three categories The numerical value '3' from: Severity (3) x Occurrence (3) x Detection (3) The supposition that if the mid-rating (or a higher number) was scored for all three categories then as a minimum the system should be examined in greater detail.
  • 30. Examples  3 examples – Potential for sanitisation cycle failure – Pressure leaks to gloves – Connection of transfer Isolator to main Isolator and transfer-in / out of material
  • 31. Isolator Example 1 Potential for sanitisation cycle failure #1 Process step Failure Mode Significance of failure Severity of conseque nce (score) Performing sanitisation cycles on transfer or main Isolator An Isolator is not correctly sanitised Contaminated items enter main Isolator or main Isolator itself is contaminated 4
  • 32. Isolator Example 1 Potential for sanitisation cycle failure #2 Measures to detect failure Occurrenc e (score) Detection systems Detection (score) Evaporation rate / pre- and post- lot testing of acid / sanitisation cycles developed using BIs 1 Steriliser parameters checked after sanitisation and before use / acid potency checked for each lot / post- sanitisation environmental monitoring performed for main Isolator 1
  • 33. Isolator Example 1 Potential for sanitisation cycle failure #3 FMEA score: 4 x 1 x 1 = 4
  • 34. Isolator Example 2 Pressure leaks to gloves #1 Process step Failure Mode Significance of failure Severity of consequ ence (score) Use of gloves to transfer material or to perform sterility test (sterile gloves may be worn underneath Isolator gloves) Contaminati on from technician into Isolator or weak area of positive pressure to allow contaminati on in Contaminati on present in Isolator / compromise of aseptic technique 4
  • 35. Isolator Example 2 Pressure leaks to gloves #2 Measures to detect failure Occurre nce (score ) Detection systems Detection (score) Environment al monitoring (post-use finger plates) / pressure charts 2 Environme ntal monitoring is performed post-test on gloves / gloves are wiped with disinfectan t / gloves are visually examined weekly and changed as appropriat e 3
  • 36. Isolator Example 2 Pressure leaks to gloves #3 FMEA score: 4 x 2 x 3 = 24
  • 37. Isolator Example 2  However, on review…. FMEA score: 4 x 2 x 1 = 8
  • 38. Isolator Example 3 The transfer of material in and out of the Isolator is, arguably, the biggest risk: Non-sterile area between doors * *Area of biggest risk
  • 40. Isolator Example 3 Connection of transfer Isolator to main Isolator and transfer-in / out of material #1 Process step Failure Mode Significance of failure Severity of conseque nce (score) Connection of transfer Isolator to main Isolator and moving material in and out Contaminat ion on outside of both Isolators may enter the main Isolator / failure to maintain positive pressure Contamina tion enters the Isolator or positive pressure is not maintained 4
  • 41. Isolator Example 3 Connection of transfer Isolator to main Isolator and transfer-in / out of material #2 Measures to detect failure Occurrence (score) Detection systems Detection (score) Environmenta l monitoring / pressure monitoring 1 DPTE seal system / use of disinfectant for connection 1
  • 42. Isolator Example 3 Connection of transfer Isolator to main Isolator and transfer-in / out of material #3 FMEA score: 4 x 1 x 1 = 4
  • 43. Isolator study  Revisit the ranking  Define residual risk  Perform a short summary – Scope – Data from the assessment & control (e.g. no. of identified failure modes) – Level of accepted risk without actions i.e. residual risk (e.g. risk priority Number < 27) – Recommended actions, responsibilities and due dates (including approval, if appropriate) – Person in charge for follow-up of FMEA
  • 44. Isolator Study  Summary of the entire study: Isolator FMEA risk assessment 0 5 10 15 20 25 30 R oom C ycle Frequency IntegrotyC onnectionS anitisation P hysical G loves Category FMEAscore Cut off score
  • 45. Further examples Sandle, T. ‘The use of a risk assessment in the pharmaceutical industry – the application of FMEA to a sterility testing isolator: a case study’, European Journal of Parenteral and Pharmaceutical Sciences, 8(2): 43-49
  • 46. FMEA  A risk assessment technique – FMEA can be readily applied to a key operation  This technique did not originate in the pharmaceutical industry.  This indicates how the synergy of different approaches can be achieved.
  • 47. FMEA  Regular reviews must take place;  FMEA is not suitable for everything e.g. HACCP may be more suitable for suitable for aseptic filling.  It is not able to discover complex failure modes involving multiple failures or subsystems, or to discover expected failure intervals of particular failure modes.
  • 48. Thank you Dr. Tim Sandle www.pharmamicroresources.com Any questions?