SlideShare uma empresa Scribd logo
1 de 73
GOOD MORNING….!
D R R AV I K U M A R V
D E P T O F O R A L PAT H O L O G Y
&   MICROBIOLOGY
G D C , K O T TAYA M .
INTRODUCTION
-   The microscope is commonly described as an instrument used for
    seeing small objects.
-   The human eye sees because of two properties of the light entering the
    eye from the objects seen.
-   The eye recognizes only differences in brightness and differences in
    color.
-   Differences in brightness of different objects or their component parts
    give rise to brightness contrast; differences in color cause color
    contrast.
-   Brightness = Amplitude
-   Color = Wavelenght
 The improvement of microscopes since their invention about 350 years
  ago has proceeded along two general lines
 First, there has been improvement of the instrument itself in order to
  make it a better and more convenient image-forming apparatus.
 The second approach is to consider the specimen as an essential part
  of the optical system and, as it were, to build the microscope around
  the specimen.
 Until the invention of the phase microscope no very useful means were
  available for observing differences in optical path in a specimen.
 Optical path is the linear path of light through a transmitting medium
  multiplied by the index of refraction of the medium.
 Refractive index in a microscopic specimen depends upon the
  specimen's physical and chemical properties.
 In optics the refractive index (or index of refraction) n of a substance
  (optical medium) is a number that describes how light, or any other radiation,
  propagates through that medium
 More fundamentally, n is defined as the factor by which the wavelength and
  the velocity of the radiation are reduced with respect to their vacuum values
 The speed of light in a medium is v = c/n, where c is the speed in vacuum.
 In physics, the wavelength of a sinusoidal wave is the spatial period of the
  wave the distance over which the wave's shape repeats.
 It is usually determined by considering the distance between consecutive
  corresponding points of the same phase, such as crests, troughs, or zero
  crossings, and is a characteristic of both traveling waves and standing waves,
  as well as other spatial wave patterns.
 Wavelength is commonly designated by theGreek letter lambda (λ).
 Amplitude is the magnitude of change in the oscillating variable with
  each oscillation within an oscillating system
 Peak amplitude
 Peak to peak amplitude
 Semi amplitude
 RMS amplitude
Air               Glass          Water        Air
       R.I.=1.0          R.I.=1.5       R.I.=1.33    R.I.=1.0




  The speed and wavelength of light changes when it passes
  through media with different refractive indices


Phase contrast microscopy makes these phase difference visible
 The details of many biological and industrial specimens are
  characterized by differences in refractive index rather than by
  differences in light absorption. Under an ordinary microscope such
  details are invisible, unless the aperture of the condenser or objective is
  made so small that the resolving power suffers a serious deterioration
  with resultant loss in the observer's ability to interpret what he sees.
 Light can be considered a form of wave motion consisting of sinusoidal
  waves. When a light wave traverses a medium of different optical path
  the phase of the light wave is altered. This alteration may be visualized
  as simply a displacement of the wave in its direction of propagation. If a
  microscopic specimen contains details that differ from each other in
  optical path, the phase of that portion of the illuminating wave front
  passing through the detail is changed. If a microscope can delineate
  change of phase as a change in brightness or color, the eye,
  photographic plate, or photocell will be able to detect the microscopic
  areas causing the phase changes.
Constructive interference          Destructive interference
                         Tutorial
We can see amplitude (intensity) differences
•Illumination bypasses
  Specimen > no phase shift




•Illumination passes
through thin part of          •Illumination passes
Specimen > small phase        through thick part of
retardation                   Specimen > larger phase
                              retardation
WALK INTO THE PAST….
 Abbe, first reported on the effect of introducing phase changes on light
  waves within the optical system of the microscope.
 (1892),he introduced glass wedges into the rear focal plane of the
  microscope objective and thereby changed the phase relationships
 Later Bratuscheck introduced absorbing strips of soot at the back focal
  plane of the microscope objective to weaken the brightness of the zero-
  order spectrum.
 These observations were made before 1892 at the time when Abbe was
  much concerned with establishing the diffraction theory of image
  formation in the microscope, and they were aimed to substantiate such
  knowledge. It is certain that they did not find their way into practical
  applications of the microscope.
   A. E. Conrady and J. Rheinberg in 1905, published experimental results
    showing that a phase reversal exists in different orders of diffraction spectra
    produced by the light.
   It is to Professor F. Zernike, of the University of Groningen, that we are
    indebted for the first application of phase contrast principles to the
    microscope and for an explanation of these principles.
    Zernike's first published work (1934) on this subject showed the advantage
    of the phase contrast method over the familiar knife-edge method in testing
    the quality of optical systems.
    In 1935 Zernike discussed the application of the phase contrast method to
    microscopy, although it is indicated that Zernike realized the importance of
    the phase contrast method in microscopy as early or earlier than 1932.
   However, it was not until 1941 that Kohler and Loos, of the firm of Carl Zeiss,
    published a paper showing the results of their extensive and convincing
    experiments with the Zernike method.
   This phase contrast technique proved to be such an advancement in
    microscopy that Zernike was awarded the Nobel prize (physics) in 1953.
FRITS ZERNIKE
OPTICAL FUNDAMENTALS
 Imagine light waves passing through a glass window.
 Although we think of a glass window as smooth,
   if we zoomed in we would see tiny imperfections.
 Light passing through different parts of the glass window would actually travel
  through different distances of glass depending on the glass thickness at a
  particular point, as shown in the diagram
 The peaks and troughs of the waves no longer match up
   We say these waves have a difference in 'phase‘
 phase differences arise because the light travels through different distances
  depending on the thickness of the glass at each point.
 a phase contrast microscope is used to transform phase differences into
  intensity differences, increasing contrast.
 Today, two main types of phase contrast are positive and negative. Since the
  observed particles are usually thin and transparent, these polar contrasts
  provide strikingly different images.
 Positive phase contrast reveals medium to dark gray images on a lighter
  grey background; these images often have a bright halo along the edge of the
  sample.
 Negative phase contrast is the opposite. The specimen appears lighter with
  a dark background; they also have a dark halo outlining the image.
 Phase Contrast can be performed in two different ways, on upright
  microscopes and inverted microscopes.
 Upright microscopes are the most common type of microscope, designed with
  the objective lenses positioned above the sample, looking downward and
  usually have shorter working distances
   Inverted microscopes are the most durable and easy to use for cell
    microscopy and tissue culture work including the diagnosis of tumor cells.
 Specialized long-working distance phase contrast optical systems have been
  developed for inverted microscope employed for tissue culture investigations
  and sperm cell motility for in vitro-fertilization.
 Interaction of Light Waves with Phase Specimens


 An incident wavefront present in an illuminating beam of light becomes
  divided into two components upon passing through a phase specimen
 The primary component is an undeviated (or undiffracted; zeroth-order)
  planar wavefront, commonly referred to as the surround (S) wave, which
  passes through and around the specimen, but does not interact with it.
   In addition, a deviated or diffracted spherical wavefront (D-wave) is also
    produced, which becomes scattered over a wide arc (in many directions) that
    passes through the full aperture of the objective.
 After leaving the specimen plane, surround and diffracted light waves enter
  the objective front lens element and are subsequently focused at the
  intermediate image plane where they combine through interference to
  produce a resultant particle wave (often referred to as a P-wave).
 Detection of the specimen image depends on the relative intensity
  differences, and therefore on the amplitudes, of the particle and surround
  (P and S) waves.
    If the amplitudes of the particle and surround waves are significantly different
    in the intermediate image plane, then the specimen acquires a considerable
    amount of contrast and is easily visualized in the microscope eyepieces.
 Otherwise, the specimen remains transparent and appears as it would under
  ordinary bright field conditions (in the absence of phase contrast or other
  contrast-enhancing techniques).
 In terms of optical path variations between the specimen and its surrounding
  medium, the portion of the incident light wavefront that traverses the
  specimen (D-wave), but does not pass through the surrounding medium (S-
  wave), is slightly retarded. For arguments in phase contrast microscopy, the
  role of the specimen in altering the optical path length (in effect, the relative
  phase shift) of waves passing through is of paramount importance
    the optical path length (OPL) through an object or space is the product of the
    refractive index (n) and the thickness (t) of the object or intervening medium
    as described by the relationship:


 When light passes from one medium into another, the velocity is altered
  proportionally to the refractive index differences between the two media.
 when a coherent light wave emitted by the focused microscope filament
  passes through a phase specimen having a specific thickness (t) and
  refractive index (n), the wave is either increased or decreased in velocity
 The difference in location of an emergent wavefront between the specimen
  and surrounding medium is termed the phase shift


 n(2) is the refractive index of the specimen and n(1) is the refractive index of
  the surrounding medium. The optical path difference results from the product
  of two terms: the thickness of the specimen, and its difference in refractive
  index with the surrounding medium.
 For individual cells in tissue culture, the optical path difference is relatively
  small. A typical cell in monolayer culture has a thickness around 5
  micrometers and a refractive index of approximately 1.36.
 The cell is surrounded by a nutrient medium having a refractive index of
  1.335, which yields an optical path difference of 0.125 micrometer, or about a
  quarter wavelength (of green light).
 Subcellular structures generate much smaller retardations. These small
  optical path differences produce a linear reduction in intensity with increasing
  phase shift (the image grows progressively darker) up to a point (depending
  upon phase plate configuration), after which, the specimen image becomes
  brighter through reversal of contrast.
 In phase contrast microscopy, the intensity of an image is dependent on a
  variety of factors including absorption at the phase plate, the degree of phase
  advancement or retardation at the phase plate, and the relative sign of this
  phase shift.
Oxytricha saprobia in brightfield (left) and with phase contrast illumination (right)
 To get a better understanding of how phase
  contrast illumination works, we study two wave
  fronts
    First, the condenser annulus is just a small
    aperture located in the center (see the plane
    labeled '1') and the phase plate is also just
    covering a small aperture (located in the plane
    labeled '3').
    Second, the optical system is greatly simplified by
    showing only two single lenses to represent all
    optical elements.
 The plane labeled '1' is the front focal plane of the
  condenser. The light emanating from the small
  aperture 'S' is captured by the condenser and
  emerges as light with only parallel wavefronts from
  the condenser
    When these plane waves (parallel wave fronts) hit
    the phase object 'O' (located in the object plane
    labeled '2'), some of this light is diffracted (and/or
    refracted) while moving through the specimen.
   Assuming that the specimen does not significantly
    alter the amplitudes of the incoming wavefronts but
    mainly changes phase relations newly generated
    spherical wave fronts that are retarded by 90° (λ/4)
    emanate from 'O
 the purple area that contains now "unperturbed"
  plane waves and spherical wave front
   there are now two types of waves, the surround
    wave or S-wave and the diffracted wave or D-wave,
    which have a relative phase-shift of 90° (λ/4)
 - The objective focuses the D-wave inside the
  primary image plane (labeled '4'), while it focuses
  the S-wave inside the back focal plane (labeled '3').
   The location of the phase plate 'P' has now a
    profound impact on the S-wave while leaving most
    of the D-wave "unharmed"
 In what is known as positive phase contrast optics,
  the phase plate 'P' reduces the amplitude of all light
  rays traveling through the phase annulus (mainly S-
  waves) by 70 to 90% and advances the phase by yet
  another 90° (λ/4).
 Hence the recombination of these two waves (D + S)
  in the primary image plane (labeled '4') results in a
  significant amplitude change at all locations where
  there is a now destructive interference due to a 180°
  (λ/2) phase shifted D-wave
 The net phase shift of 180° (λ/2) results directly from
  the 90° (λ/4) retardation of the D-wave due to the
  phase object and the 90° (λ/4) phase advancement
  of the S-wave due to the phase plate.
 Without the phase plate, there would be no significant destructive interference
  that greatly enhances contrast.
    With phase contrast illumination "invisible" phase variations are hence
    translated into visible amplitude variations. The destructive interference is
    illustrated in the figure. Blue and orange indicate D-wave and S-wave,
    respectively. The resulting wave (D + S), indicated by yellow, has a reduced
    amplitude.
PARTS:
 Presented in Figure is a cut-away diagram of a modern
  upright phase contrast microscope, including a
  schematic illustration of the phase contrast optical train.
 Partially coherent illumination is produced by the
  tungsten-halogen lamp
 Light is directed through a collector lens and focused on
  a specialized annulus (labeled condenser annulus)
  positioned in the substage condenser front focal plane.
 Wavefronts passing through the annulus illuminate the
  specimen and either pass through undeviated or are
  diffracted and retarded in phase by structures and phase
  gradients present in the specimen.
 Rays are segregated at the rear focal plane by a phase
  plate and focused at the intermediate image plane to
  form the final phase contrast image observed in the
  eyepieces.
 Two specialized accessories are required to convert a brightfield microscope
  for phase contrast observation.
    A specially designed annular diaphragm, which is matched in diameter and
    optically conjugate to an internal phase plate residing in the objective rear
    focal plane
   The condenser annulus is typically constructed as an opaque flat-black (light
    absorbing) plate with a transparent annular ring, which is positioned so the
    specimen can be illuminated by defocused, parallel light wavefronts emanating
    from the ring.
   The microscope condenser images the annular diaphragm at infinity, while the
    objective produces an image at the rear focal plane
   The condenser annulus either replaces or resides close to the adjustable iris
    diaphragm in the front aperture of the condenser.
   Under conditions of Köhler illumination, surround light waves that do not interact
    with the specimen are focused as a bright ring in the rear focal plane of the
    objective
   The D waves passes through the diffraction plane at various locations across the
    entire objective rear aperture.
   Thus, the two wavefronts do not overlap to a significant extent, and occupy distinct
    portions of the objective rear focal plane.
 The phase contrast annuli must be specifically matched to a particular
  objective equipped with a corresponding phase plate
 By matching the condenser annulus to the objective phase plate, the
  microscope can be aligned to superimpose illuminating light rays passed
  through the annulus onto the objective phase ring to achieve phase contrast
  illumination.
 A majority of the popular universal condenser systems designed for phase
  contrast microscopy are equipped with three or more removable annular
  diaphragms, which are available for use with 4x, 10x, 20x, 40x, 60x, and 100x
  objectives containing the appropriate phase plates.
 The PH1 - smallest aperture – 10x and 20 x
 The PH2 – 40x and 60x
 The PH3 – 100x
 The PHL - 4x and 5x, in long working distance condensers
 Condenser annulus inserts are circular aluminum plates having a stamped
  ring of varying dimensions in the center
    After the stamping operation, annular disk units are anodized and dyed with
    a flat-black pigment to absorb stray light and ensure that illuminating light
    rays passing through the annulus follow a defined pathway.
 The central light stop, is positioned in the center of the plate and secured by
  three slim tabs spaced at 120-degree intervals.
 Modern universal condenser system turrets usually contain between five and
  eight open slot positions that can be fitted with annular phase contrast plates,
  DIC Nomarski (Wollaston) prism plates, or darkfield light stop plates.
 The phase contrast condenser annular diaphragm plates are inserted into the
  appropriate position in the turret and secured into place with a spring clip
 Whenever a universal microscope condenser is utilized, be certain to ensure
  that the diaphragm is opened wider than the diameter of the condenser
  annulus. In fact, it is a good idea to always open the diaphragm to its widest
  position during phase contrast observations.
 Phase contrast condensers are available in a wide spectrum of optical
  correction levels, ranging from the basic Abbe design to highly corrected
  aplanatic-achromatic systems that feature excellent performance.
   In addition, specialized phase contrast condensers are available with long
    working distance (LWD) optics & extra long working distance (ELWD).
 Single annulus phase condensers were very popular at one time, but have
  been largely supplanted by the universal turret models. The older condensers
  require removal and insertion of different phase annuli as the objective is
  changed
PHASE CONTRAST OBJECTIVES
 The most important attribute of objectives designed for phase contrast
  microscopy is the presence of a specialized phase plate
 Phase plates are not interchangeable between objectives and are often
  permanently etched into one of the internal lens elements
 The etched ring is coated with a partially absorbing metallic film that reduces
  light transmission, and is usually made so that the phase of light passing
  through will be advanced by one quarter-wavelength relative to light that is
  transmitted through the rest of the glass.
 The inset illustrates the phase plate that is positioned in the rear focal plane
 The range of phase contrast objectives available from the major
  manufacturers covers almost the complete gamut of correction factors,
  including achromatic, plan achromatic, fluorite, and apochromatic.
    In addition, phase contrast objectives are manufactured that vary in the
    neutral density and retardation value of the phase plate to produce a wide
    spectrum of contrast levels in both positive and negative phase contrast
    modes.
 Presented below is a comparison of digital images recorded using the
  currently available Nikon
 The images are the same viewfield from a fixed and mounted preparation
  of Zygnema filamentous algae. (a),(b), (c) compare the dark low low, dark
  low and apodized phase contrast objectives.
 The dark and bright medium phase contrast objectives (d) and
  (e))demonstrate positive and negative phase contrast.
 DL (Dark Low - Medium Contrast) - DL objectives produce a dark image
  outline on a light gray background, and are the typical objectives utilized for
  all-purpose phase contrast observation.
 These objectives are designed to furnish the strongest dark contrast in
  specimens having a major difference in refractive index from that of the
  surrounding medium.
 The DL phase contrast objective is the most popular style for examination of
  cells and other semi-transparent living material and is especially suited for
  photomicrography and digital imaging.
 DLL (Dark Low Low - Low Contrast) - Similar in design to the DL objective,
  the DLL series yields better images in brightfield illumination and is often
  employed as a "universal" objective in microscope systems that utilize
  multiple illumination modes such as fluorescence, DIC, brightfield, and
  darkfield.
 The DLL phase contrast objective produces less contrast than the DL
  objective, but features higher light transmission values, optical correction, and
  numerical aperture than the standard DL counterpart.
   A majority of the DLL phase contrast objectives offered by the manufacturers
    have fluorite or apochromatic aberration correction levels.
 ADL (Apodized Dark Low - Medium Contrast) - Recently introduced, the
  apodized phase contrast ADL objectives contain a secondary neutral density
  ring on either side of the central ring in the phase plate.
 Addition of the secondary rings assists in reducing unwanted "halo" effects
  often associated with imaging large particles or specimen features (such as
  nuclei, whole cells and fibers) in phase contrast microscopy.
 Apodized objectives are available in plan achromat optical correction and
  feature contrast levels similar to the DL objective series.
 DM (Dark Medium - High Contrast) - DM objectives produce a dark image
  outline on a medium gray background.
    These objectives are designed to be used for high image contrast with
    specimens having small phase shifts or refractive differences, such as fine
    fibers, flagella, cilia, granules, and very small particles.
    Usually restricted to higher magnification objectives having large numerical
    apertures (fluorites and apochromats), DM phase contrast objectives perform
    well with very thin specimens, but often display a reversal of contrast when
    thick specimens are imaged.
 BM (Bright Medium - High Negative Contrast) - Often referred to as
  negative phase contrast, BM objectives produce a bright image outline on a
  medium gray background.
 BM objectives are ideal for visual examination of bacterial flagella, fibrin
  bundles, minute globules, and for blood cell counting.
 For easy identification manufacturers inscribe important specifications on
  the outer barrel in green letters
    In addition, phase contrast objectives have inscriptions on the barrel to
    indicate the objective is designed for phase contrast and also the matching
    annulus designation.
 A typical series of phase contrast objectives are presented.
    As a general rule, when objective numerical aperture and magnification is
    increased, the phase plate width and diameter both decrease and the
    condenser annulus size increases
 The positive phase plate produces dark contrast and contains a partially
  absorbing film designed to reduce the amplitude of the surround wavefront.
 In addition, this plate contains phase retarding material designed to shift
  (retard) the phase of the diffracted light by 90 degrees
 In negative phase plate both materials are placed within the circular phase
  ring so that the undiffracted surround wavefront becomes the only species
  affected, and is attenuated and retarded in phase by 90 degrees.
 Contrast is modulated in phase objectives by varying the properties of the
  phase plate, including the absorption of the metallic film (or anti-reflective
  coatings), the refractive index of the phase retarding material, and the
  thickness of the phase plate.
 Phase plates are produced by vacuum deposition of thin dielectric and
  metallic films onto a glass plate or directly onto one of the lens surfaces within
  the microscope objective
    The role of the dielectric film is to shift the phase of light, while the metallic
    film attenuates undiffracted light intensity.
    The thickness and refractive indices of the dielectric, metallic, and anti-
    reflective films, as well as those of the optical cement, are carefully selected
    to produce the necessary phase shift.
 In optical terminology, phase plates that alter the phase of surround light
  relative to diffracted light by 90 degrees (either positive or negative) are
  termed quarter wavelength plates because of their effect on the optical path
  difference
THE PHASE TELESCOPE
 If the condenser annulus is not in exact alignment with the fixed phase plate
  in the objective the contrast effect afforded will be dramatically compromised.
 This task can be accomplished using either a phase telescope, which can be
  inserted into one of the eyepiece observation tubes (in place of a normal
  eyepiece), or a Bertrand lens built into the microscope binocular (or
  trinocular) eyepiece tube assembly.
 The phase telescope, also commonly referred to as an auxiliary
  telescope or auxiliary microscope consists of a simple two or three lens
  telescope
 The focal length of the phase telescope ranges from about 150 to 200
  millimeters enabling the device to focus on the objective rear focal plane
  when inserted into eyepiece.
 The Bertrand lens is more sophisticated, and acts as a relay lens, transferring
  an image of the objective focal plane to the microscope intermediate image
  plane located in the eyepiece aperture diaphragm
    On most microscopes equipped with a Bertrand lens, the lens can be rotated
    into and out of the optical pathway by means of a small thumbwheel
    mechanism located beneath the eyepiece tubes
 Both the phase telescope and Bertrand lens must be equipped with a
  mechanism for adjustment of focus, because the location of the objective rear
  focal plane can vary with magnification
 Focusing is accomplished simply by twisting the eye tube until the objective
  rear focal plane is sharply focused.
PHASE CONTRAST MICROSCOPE ALIGNMENT
 ensure that all objectives contain phase plates and are firmly seated in the
  nosepiece.
 The objectives and annular plates should also be sequentially ordered in their
  arrangement from lower to higher magnification
 Shown below is the phase contrast kit used on the National Optical 160
  series microscopes.
   It consists of 4 objective lenses, a centering telescope and Zernike
    phase condenser lens.
 The long adjustment screws on the phase condenser push in to engage set
  screws for proper alignment of the phase ring.
 There is a thumb wheel on the opposite side used to dial in the proper setting
  to match the power of the objective lens.
 There is also a "BF" setting on the thumb wheel for brightfield. This allows
  you to use the phase objectives as standard brightfield lenses.
 Not all phase contrast microscopes are the same but generally they rely on
  similar techniques to set up the system for optimum results.’
   In the system shown, the phase condenser has five settings that you spin
    with your thumb ( 10X, 20X, 40X, 100X and BF) BF is "brightfield", no
    phase.
 To set up your microscope for phase optics, you first set it at BF and focus on
  the specimen.
   Adjust the height of the condenser for optimum image quality.
    Next, set the condenser turret to the phase setting for that particular lens and
    remove the specimen.
 The controls that stick out from both sides on the back are for centering the
  condenser.
 Next, you remove one of the eyepiece lenses and insert the
  centering telescope in its place.
   The set screw is used to focus the centering telescope.
   When looking through the telescope, you will see two rings.
 They may or may not be concentric. By turning centering
  adjustment screws on the condenser, you align the rings so that they
  become concentric.
 Once the microscope has been aligned for phase contrast, it will generally
  hold its centration for a considerable number of objective/annuli changes, but
  should be checked periodically to ensure proper alignment.
   If the microscope starts to slip out of alignment, the images appearing in the
    eyepieces (or on a computer monitor) will appear increasingly more like those
    observed with brightfield illumination.
 Most of the microscope manufacturers provide a green interference or
  absorption filter with their auxiliary phase contrast kits to produce
  monochromatic light having the same wavelength used for the original
  calibration of the objective phase plates.
   As a result, contrast is increased when the filter is inserted into the optical
    pathway
   A majority of the commercial phase plates are designed to produce phase
    shifts of a quarter wavelength in the green (550 nanometers) portion of the
    visible light spectrum.
 Theoretically, if white light is utilized instead of monochromatic light, extinction
  by interference will not be complete for all colors, and contrast will suffer.
    This limitation is particularly important if achromat objectives, which are
    corrected for chromatic aberration only in the green region, are utilized for
    phase contrast observation or image recording.
    However, with the current highly corrected fluorite and apochromatic
    objectives, the difference in contrast often is negligible, and therefore,
    insignificant.
 How to get good images..?
 The key to successful imaging with phase contrast illumination is to properly
  align the microscope, and to ensure that sufficiently thin specimens are
  spread evenly within the mounting medium on the microscope slide.
 Images made with exceedingly thick specimens often suffer from out-of-focus
  blur and contrast inversion artifacts that can be difficult to interpret.
    If the microscope is utilized for an extended period of time, occasionally
    check the objective rear focal plane to verify alignment of the condenser
    annulus with the phase plate in the objective.
INTERPRETATION OF PHASE CONTRAST IMAGES
 Images produced by phase contrast microscopy are relatively simple to
  interpret when the specimen is thin and distributed evenly on the substrate
 When using positive phase contrast optics, which is the traditional, images
  appear darker than the surrounding medium when the refractive index of the
  specimen exceeds that of the medium.
   Phase contrast optics differentially enhance the contrast near the edges
 image density can be utilized as a gauge for approximating relationships
  between various structures.
    In effect, a series of internal cellular organelles having increasing density,
    such as vacuoles, cytoplasm, the interphase nucleus, and the nucleolus (or
    mitotic chromosomes), are typically visualized as progressively darker objects
    relative to a fixed reference, such as the background.
 In order to avoid confusion regarding bright and dark contrast in phase
  contrast images, the optical path differences occurring within the specimen
  preparation should be carefully considered
    As discussed above, the optical path difference is derived from the product of
    the refractive index and the specimen (object) thickness, and is related to the
    relative phase shift between the specimen and background waves.
 For example, a small specimen having a high refractive index can display an
  identical optical path difference to a larger specimen having a lower refractive
  index.
 The external medium can also be replaced with another having either a
  higher or lower refractive index to generate changes in specimen image
  contrast. In fact, the effect on image contrast of refractive index variations in
  the surrounding medium forms the basis of the technique known
  as immersion refractometry.
 Two very common effects in phase contrast images are the
  characteristic halo and shade-off contrast patterns in which the observed
  intensity does not directly correspond to the optical path difference
    In all forms of positive phase contrast, bright phase halos usually surround
    the boundaries between large specimen features and the medium. Identical
    halos appear darker than the specimen with negative phase contrast optical
    systems
 Halos occur in phase contrast microscopy because the circular phase-
  retarding (and neutral density) ring located in the objective phase plate also
  transmits a small degree of diffracted light from the specimen.
 problem is compounded by the fact that the width of the zeroth-order
  surround wavefront projected onto the phase plate by the condenser annulus
  is smaller than the actual width of the phase plate ring
 The halo effect can also be significantly reduced by utilizing specially
  designed phase objectives that contain a small ring of neutral density material
  surrounding the central phase ring material near the objective rear aperture.
  These objectives are termed apodized phase contrast objective
 In practice, halo reduction and an increase in specimen contrast with
  apodized optical systems can be achieved by the utilization of selective
  amplitude filters
 These amplitude filters consist of neutral density thin films applied to the
  phase plate surrounding the phase film.
 Transmittance increase from 25 to 50 % with their use.
 Shade-off is another very common optical artifact in phase contrast
  microscopy, and is often most easily observed in large, extended phase
  specimens
 The intensity profile of a large, uniformly thick positive phase contrast
  specimen often gradually increases from the edges to the center, where the
  light intensity in the central region can approach that of the surrounding
  medium
 This effect is termed shade-off, and is frequently observed when examining
  extended planar specimens, such as material slabs (glass or mica), replicas,
  flattened tissue culture cells, and large organelles.
 The shade-off phenomenon is also commonly termed the zone-of-action
  effect, because central zones having uniform thickness in the specimen
  diffract light differently than the highly refractive zones at edges and
  boundaries.
 Halo and shade-off artifacts depend on both the geometrical and optical
  properties of the phase plate and the specimen being examined.
    Wider phase plates having reduced transmittance tend to produce higher
    intensity halos and shade-off, whereas the ring diameter has a smaller
    influence on these effects
   In addition, these effects are heavily influenced by the objective
    magnification, with lower magnifications producing better images.
ADVANTAGES
 The capacity to observe living cells and, as such, the ability to examine cells
  in a natural state, study bilological processes.
 Observing a living organism in its natural state and/or environment can
  provide far more information than specimens that need to be killed, fixed or
  stain to view under a microscope
 High-contrast, high-resolution images
 Ideal for studying and interpreting thin specimens
 Ability to combine with other means of observation, such as fluorescence
 Modern phase contrast microscopes, with CCD or CMOS computer devices,
  can capture photo and/or video images
 In addition, advances to the phase contrast microscope, especially those that
  incorporate technology, enable a scientist to hone in on minute internal
  structures of a particle and can even detect a mere small number of protein
  molecules.
DISADVANTAGES

 Annuli or rings limit the aperture to some extent, which decreases resolution
 This method of observation is not ideal for thick organisms or particles
 Images may appear grey or green, if white or green lights are used,
  respectively, resulting in poor photomicrograph
 Shade-off and halo effect, referred to a phase artifacts
 Shade-off occurs with larger particles, results in a steady reduction of contrast
  moving from the center of the object toward its edges
 Halo effect, where images are often surrounded by bright areas, which
  obscure details along the perimeter of the specimen
CONCLUSION
 The phase contrast microscope opened up an entire world of microscopy,
  providing incredible definition and clarity of particles never seen before.
 These transparent specimens could not be explored because they do not
  have the capacity to absorb light.
 Zernike found a way to manipulate light paths through the use of strategically
  placed rings and his system is a staple of most modern microscopes.
 Although there are a few disadvantages, such as shade-off and halo
  distortions, phase contrast provides highly detailed, well-contrasted images
 The most important breakthrough is the ability to observe living particles in a
  natural state.
Phase contrast microscope

Mais conteúdo relacionado

Mais procurados

Phase contrast and Fluorescent microscope
Phase contrast and Fluorescent microscopePhase contrast and Fluorescent microscope
Phase contrast and Fluorescent microscopeNithiya Pappuraj
 
Phase contrast & Flurosence microscope
Phase contrast & Flurosence microscopePhase contrast & Flurosence microscope
Phase contrast & Flurosence microscopemicrobiology Notes
 
Microscopy - Magnification, Resolving power, Principles, Types and Applications
Microscopy - Magnification, Resolving power, Principles, Types and ApplicationsMicroscopy - Magnification, Resolving power, Principles, Types and Applications
Microscopy - Magnification, Resolving power, Principles, Types and ApplicationsNethravathi Siri
 
FLUORESCENCE MICROSCOPY
FLUORESCENCE MICROSCOPYFLUORESCENCE MICROSCOPY
FLUORESCENCE MICROSCOPYprachann
 
Confocal microscope presentation pt
Confocal microscope presentation  ptConfocal microscope presentation  pt
Confocal microscope presentation ptShrigouri4605
 
Fluorescence microscope
Fluorescence microscopeFluorescence microscope
Fluorescence microscopeSuganyaPaulraj
 
1. MICROSCOPY - introduction + principle (Basics)
1. MICROSCOPY -  introduction + principle  (Basics)1. MICROSCOPY -  introduction + principle  (Basics)
1. MICROSCOPY - introduction + principle (Basics)Nethravathi Siri
 
Fluorescence and electron microscopy
Fluorescence and electron microscopyFluorescence and electron microscopy
Fluorescence and electron microscopyShmilona Jain
 
Light microscope
Light microscopeLight microscope
Light microscopekhalilkazmi
 
Electron microscope ppt
Electron microscope pptElectron microscope ppt
Electron microscope pptNithyaNandapal
 

Mais procurados (20)

Phase contrast and Fluorescent microscope
Phase contrast and Fluorescent microscopePhase contrast and Fluorescent microscope
Phase contrast and Fluorescent microscope
 
Fluorescence Microscopy
Fluorescence MicroscopyFluorescence Microscopy
Fluorescence Microscopy
 
Phase contrast & Flurosence microscope
Phase contrast & Flurosence microscopePhase contrast & Flurosence microscope
Phase contrast & Flurosence microscope
 
Microscopy - Magnification, Resolving power, Principles, Types and Applications
Microscopy - Magnification, Resolving power, Principles, Types and ApplicationsMicroscopy - Magnification, Resolving power, Principles, Types and Applications
Microscopy - Magnification, Resolving power, Principles, Types and Applications
 
Dark field microscope
Dark field microscopeDark field microscope
Dark field microscope
 
Light Microscopy
Light MicroscopyLight Microscopy
Light Microscopy
 
Dark field microscopy
Dark field microscopyDark field microscopy
Dark field microscopy
 
FLUORESCENCE MICROSCOPY
FLUORESCENCE MICROSCOPYFLUORESCENCE MICROSCOPY
FLUORESCENCE MICROSCOPY
 
Confocal microscope presentation pt
Confocal microscope presentation  ptConfocal microscope presentation  pt
Confocal microscope presentation pt
 
phase contrast microscope
phase contrast microscopephase contrast microscope
phase contrast microscope
 
Fluorescence microscope
Fluorescence microscopeFluorescence microscope
Fluorescence microscope
 
1. MICROSCOPY - introduction + principle (Basics)
1. MICROSCOPY -  introduction + principle  (Basics)1. MICROSCOPY -  introduction + principle  (Basics)
1. MICROSCOPY - introduction + principle (Basics)
 
Nomarski dic
Nomarski dicNomarski dic
Nomarski dic
 
Phase contrast microscope
Phase contrast microscopePhase contrast microscope
Phase contrast microscope
 
Fluorescence and electron microscopy
Fluorescence and electron microscopyFluorescence and electron microscopy
Fluorescence and electron microscopy
 
DARK FIELD MICROSCOPY
DARK FIELD MICROSCOPYDARK FIELD MICROSCOPY
DARK FIELD MICROSCOPY
 
Light microscope
Light microscopeLight microscope
Light microscope
 
Electron microscope ppt
Electron microscope pptElectron microscope ppt
Electron microscope ppt
 
Confocal microscopy
Confocal microscopyConfocal microscopy
Confocal microscopy
 
Light microscope
Light microscopeLight microscope
Light microscope
 

Destaque

Scanning transmission electron microscope (2)
Scanning transmission electron microscope (2)Scanning transmission electron microscope (2)
Scanning transmission electron microscope (2)Gulfam Hussain
 
Fluorescence Microscopy
Fluorescence MicroscopyFluorescence Microscopy
Fluorescence MicroscopyAalap Tripathy
 
Nitrogen Fixation
Nitrogen FixationNitrogen Fixation
Nitrogen Fixationdpdhanera
 
Cultural characteristics
Cultural characteristicsCultural characteristics
Cultural characteristicsDhananjay Desai
 
Dark field microscopy
Dark field microscopyDark field microscopy
Dark field microscopyCHARLIE
 
Blockbusters #soccd
Blockbusters #soccdBlockbusters #soccd
Blockbusters #soccdSteve Bishop
 
History of microbiology
History of microbiologyHistory of microbiology
History of microbiologyprabesh paudel
 
Conjugation
Conjugation  Conjugation
Conjugation ninabakh
 
biodegradation of hydrocarbons
biodegradation of hydrocarbonsbiodegradation of hydrocarbons
biodegradation of hydrocarbonskhuda bakhsh
 
History of microbiology
History of microbiologyHistory of microbiology
History of microbiologyVijay Baria
 
Genetic Analysis and Mapping in Bacteria and Bacteriophages
Genetic Analysis and Mapping in Bacteria and Bacteriophages Genetic Analysis and Mapping in Bacteria and Bacteriophages
Genetic Analysis and Mapping in Bacteria and Bacteriophages mpattani
 
Culture characteristic of bacteria
Culture characteristic of bacteriaCulture characteristic of bacteria
Culture characteristic of bacteriaDr. Samira Fattah
 
NITROGEN CYCLE AND FIXATION
NITROGEN CYCLE AND FIXATIONNITROGEN CYCLE AND FIXATION
NITROGEN CYCLE AND FIXATIONscience24-7
 
Bacterial transformation
Bacterial transformationBacterial transformation
Bacterial transformationipezpagn91
 
Electron microscopy
Electron microscopyElectron microscopy
Electron microscopysuniu
 

Destaque (20)

Scanning transmission electron microscope (2)
Scanning transmission electron microscope (2)Scanning transmission electron microscope (2)
Scanning transmission electron microscope (2)
 
Fluorescence Microscopy
Fluorescence MicroscopyFluorescence Microscopy
Fluorescence Microscopy
 
Fluorescence Microscopy
Fluorescence MicroscopyFluorescence Microscopy
Fluorescence Microscopy
 
Nitrogen Fixation
Nitrogen FixationNitrogen Fixation
Nitrogen Fixation
 
Cultural characteristics
Cultural characteristicsCultural characteristics
Cultural characteristics
 
Dark field microscopy
Dark field microscopyDark field microscopy
Dark field microscopy
 
Blockbusters #soccd
Blockbusters #soccdBlockbusters #soccd
Blockbusters #soccd
 
History of microbiology
History of microbiologyHistory of microbiology
History of microbiology
 
Conjugation
Conjugation  Conjugation
Conjugation
 
biodegradation of hydrocarbons
biodegradation of hydrocarbonsbiodegradation of hydrocarbons
biodegradation of hydrocarbons
 
History of microbiology
History of microbiologyHistory of microbiology
History of microbiology
 
Transformations
TransformationsTransformations
Transformations
 
Genetic Analysis and Mapping in Bacteria and Bacteriophages
Genetic Analysis and Mapping in Bacteria and Bacteriophages Genetic Analysis and Mapping in Bacteria and Bacteriophages
Genetic Analysis and Mapping in Bacteria and Bacteriophages
 
Bright field microscopes
Bright field microscopesBright field microscopes
Bright field microscopes
 
Culture characteristic of bacteria
Culture characteristic of bacteriaCulture characteristic of bacteria
Culture characteristic of bacteria
 
Nitrogen cycle
Nitrogen cycleNitrogen cycle
Nitrogen cycle
 
Sem n tem
Sem n temSem n tem
Sem n tem
 
NITROGEN CYCLE AND FIXATION
NITROGEN CYCLE AND FIXATIONNITROGEN CYCLE AND FIXATION
NITROGEN CYCLE AND FIXATION
 
Bacterial transformation
Bacterial transformationBacterial transformation
Bacterial transformation
 
Electron microscopy
Electron microscopyElectron microscopy
Electron microscopy
 

Semelhante a Phase contrast microscope

Ncert class-12-physics-part-2
Ncert class-12-physics-part-2Ncert class-12-physics-part-2
Ncert class-12-physics-part-2RAHUL SINGH
 
Chapter 7 - Wave optics.pptx
Chapter 7 - Wave optics.pptxChapter 7 - Wave optics.pptx
Chapter 7 - Wave optics.pptxPooja M
 
PHASE CONTRAST MICRSCOPY characteristics of material .pptx
PHASE CONTRAST MICRSCOPY characteristics of material .pptxPHASE CONTRAST MICRSCOPY characteristics of material .pptx
PHASE CONTRAST MICRSCOPY characteristics of material .pptxkannanchellakannu
 
CLASS XII PHYSICS Chapter 7 - wave optics
CLASS XII PHYSICS Chapter 7 - wave optics CLASS XII PHYSICS Chapter 7 - wave optics
CLASS XII PHYSICS Chapter 7 - wave optics Pooja M
 
To Study Principles of Microscopy: Light Microscope, Phase Contrast Microsco...
To Study Principles of Microscopy: Light Microscope, Phase Contrast  Microsco...To Study Principles of Microscopy: Light Microscope, Phase Contrast  Microsco...
To Study Principles of Microscopy: Light Microscope, Phase Contrast Microsco...Om Prakash
 
Optical properties of nanomaterials
Optical properties of nanomaterialsOptical properties of nanomaterials
Optical properties of nanomaterialsudhay roopavath
 
Application Of Biochemical Tools And Techniques.
Application Of Biochemical Tools And Techniques.Application Of Biochemical Tools And Techniques.
Application Of Biochemical Tools And Techniques.Ashley Carter
 
Diffraction-Fraunhofer Diffraction
Diffraction-Fraunhofer DiffractionDiffraction-Fraunhofer Diffraction
Diffraction-Fraunhofer Diffractionjohnisaac21
 
Reflection-refraction.ppt
Reflection-refraction.pptReflection-refraction.ppt
Reflection-refraction.pptsimonckt1
 
Microscopy fin
Microscopy finMicroscopy fin
Microscopy finMUBOSScz
 
Optical phenomena
Optical phenomena Optical phenomena
Optical phenomena Satish Gupta
 
microscopy in bacteriology by Tejpal.pptx
microscopy in bacteriology by Tejpal.pptxmicroscopy in bacteriology by Tejpal.pptx
microscopy in bacteriology by Tejpal.pptxVetico
 
Phase contrast microscope
Phase contrast microscopePhase contrast microscope
Phase contrast microscopeSalman Khan
 

Semelhante a Phase contrast microscope (20)

Phase contrast Microscopy
Phase contrast MicroscopyPhase contrast Microscopy
Phase contrast Microscopy
 
Ncert class-12-physics-part-2
Ncert class-12-physics-part-2Ncert class-12-physics-part-2
Ncert class-12-physics-part-2
 
Chapter 7 - Wave optics.pptx
Chapter 7 - Wave optics.pptxChapter 7 - Wave optics.pptx
Chapter 7 - Wave optics.pptx
 
PHASE CONTRAST MICRSCOPY characteristics of material .pptx
PHASE CONTRAST MICRSCOPY characteristics of material .pptxPHASE CONTRAST MICRSCOPY characteristics of material .pptx
PHASE CONTRAST MICRSCOPY characteristics of material .pptx
 
CLASS XII PHYSICS Chapter 7 - wave optics
CLASS XII PHYSICS Chapter 7 - wave optics CLASS XII PHYSICS Chapter 7 - wave optics
CLASS XII PHYSICS Chapter 7 - wave optics
 
To Study Principles of Microscopy: Light Microscope, Phase Contrast Microsco...
To Study Principles of Microscopy: Light Microscope, Phase Contrast  Microsco...To Study Principles of Microscopy: Light Microscope, Phase Contrast  Microsco...
To Study Principles of Microscopy: Light Microscope, Phase Contrast Microsco...
 
Microscope (1).pptx
Microscope (1).pptxMicroscope (1).pptx
Microscope (1).pptx
 
Optical properties of nanomaterials
Optical properties of nanomaterialsOptical properties of nanomaterials
Optical properties of nanomaterials
 
Ppt f1
Ppt f1Ppt f1
Ppt f1
 
Microscopy
MicroscopyMicroscopy
Microscopy
 
Application Of Biochemical Tools And Techniques.
Application Of Biochemical Tools And Techniques.Application Of Biochemical Tools And Techniques.
Application Of Biochemical Tools And Techniques.
 
Diffraction-Fraunhofer Diffraction
Diffraction-Fraunhofer DiffractionDiffraction-Fraunhofer Diffraction
Diffraction-Fraunhofer Diffraction
 
Reflection-refraction.ppt
Reflection-refraction.pptReflection-refraction.ppt
Reflection-refraction.ppt
 
Reflection-refraction.ppt
Reflection-refraction.pptReflection-refraction.ppt
Reflection-refraction.ppt
 
Microscopy fin
Microscopy finMicroscopy fin
Microscopy fin
 
Optical phenomena
Optical phenomenaOptical phenomena
Optical phenomena
 
Optical phenomena
Optical phenomena Optical phenomena
Optical phenomena
 
microscopy in bacteriology by Tejpal.pptx
microscopy in bacteriology by Tejpal.pptxmicroscopy in bacteriology by Tejpal.pptx
microscopy in bacteriology by Tejpal.pptx
 
Microscope
Microscope Microscope
Microscope
 
Phase contrast microscope
Phase contrast microscopePhase contrast microscope
Phase contrast microscope
 

Último

Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsTechSoup
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingTechSoup
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...christianmathematics
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhikauryashika82
 
Disha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfDisha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfchloefrazer622
 
Measures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDMeasures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDThiyagu K
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...EduSkills OECD
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfAdmir Softic
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3JemimahLaneBuaron
 
Unit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxUnit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxVishalSingh1417
 
General AI for Medical Educators April 2024
General AI for Medical Educators April 2024General AI for Medical Educators April 2024
General AI for Medical Educators April 2024Janet Corral
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxheathfieldcps1
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdfSoniaTolstoy
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfsanyamsingh5019
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajanpragatimahajan3
 
fourth grading exam for kindergarten in writing
fourth grading exam for kindergarten in writingfourth grading exam for kindergarten in writing
fourth grading exam for kindergarten in writingTeacherCyreneCayanan
 

Último (20)

Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy Consulting
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
 
Disha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdfDisha NEET Physics Guide for classes 11 and 12.pdf
Disha NEET Physics Guide for classes 11 and 12.pdf
 
Measures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDMeasures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SD
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3
 
Unit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxUnit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptx
 
General AI for Medical Educators April 2024
General AI for Medical Educators April 2024General AI for Medical Educators April 2024
General AI for Medical Educators April 2024
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdf
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajan
 
fourth grading exam for kindergarten in writing
fourth grading exam for kindergarten in writingfourth grading exam for kindergarten in writing
fourth grading exam for kindergarten in writing
 
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptxINDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
 
Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1
 
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
 

Phase contrast microscope

  • 2. D R R AV I K U M A R V D E P T O F O R A L PAT H O L O G Y & MICROBIOLOGY G D C , K O T TAYA M .
  • 3. INTRODUCTION - The microscope is commonly described as an instrument used for seeing small objects. - The human eye sees because of two properties of the light entering the eye from the objects seen. - The eye recognizes only differences in brightness and differences in color. - Differences in brightness of different objects or their component parts give rise to brightness contrast; differences in color cause color contrast. - Brightness = Amplitude - Color = Wavelenght
  • 4.  The improvement of microscopes since their invention about 350 years ago has proceeded along two general lines  First, there has been improvement of the instrument itself in order to make it a better and more convenient image-forming apparatus.  The second approach is to consider the specimen as an essential part of the optical system and, as it were, to build the microscope around the specimen.  Until the invention of the phase microscope no very useful means were available for observing differences in optical path in a specimen.  Optical path is the linear path of light through a transmitting medium multiplied by the index of refraction of the medium.  Refractive index in a microscopic specimen depends upon the specimen's physical and chemical properties.
  • 5.  In optics the refractive index (or index of refraction) n of a substance (optical medium) is a number that describes how light, or any other radiation, propagates through that medium  More fundamentally, n is defined as the factor by which the wavelength and the velocity of the radiation are reduced with respect to their vacuum values  The speed of light in a medium is v = c/n, where c is the speed in vacuum.
  • 6.  In physics, the wavelength of a sinusoidal wave is the spatial period of the wave the distance over which the wave's shape repeats.  It is usually determined by considering the distance between consecutive corresponding points of the same phase, such as crests, troughs, or zero crossings, and is a characteristic of both traveling waves and standing waves, as well as other spatial wave patterns.  Wavelength is commonly designated by theGreek letter lambda (λ).
  • 7.  Amplitude is the magnitude of change in the oscillating variable with each oscillation within an oscillating system  Peak amplitude  Peak to peak amplitude  Semi amplitude  RMS amplitude
  • 8. Air Glass Water Air R.I.=1.0 R.I.=1.5 R.I.=1.33 R.I.=1.0 The speed and wavelength of light changes when it passes through media with different refractive indices Phase contrast microscopy makes these phase difference visible
  • 9.  The details of many biological and industrial specimens are characterized by differences in refractive index rather than by differences in light absorption. Under an ordinary microscope such details are invisible, unless the aperture of the condenser or objective is made so small that the resolving power suffers a serious deterioration with resultant loss in the observer's ability to interpret what he sees.  Light can be considered a form of wave motion consisting of sinusoidal waves. When a light wave traverses a medium of different optical path the phase of the light wave is altered. This alteration may be visualized as simply a displacement of the wave in its direction of propagation. If a microscopic specimen contains details that differ from each other in optical path, the phase of that portion of the illuminating wave front passing through the detail is changed. If a microscope can delineate change of phase as a change in brightness or color, the eye, photographic plate, or photocell will be able to detect the microscopic areas causing the phase changes.
  • 10. Constructive interference Destructive interference Tutorial We can see amplitude (intensity) differences
  • 11. •Illumination bypasses Specimen > no phase shift •Illumination passes through thin part of •Illumination passes Specimen > small phase through thick part of retardation Specimen > larger phase retardation
  • 12. WALK INTO THE PAST….  Abbe, first reported on the effect of introducing phase changes on light waves within the optical system of the microscope.  (1892),he introduced glass wedges into the rear focal plane of the microscope objective and thereby changed the phase relationships  Later Bratuscheck introduced absorbing strips of soot at the back focal plane of the microscope objective to weaken the brightness of the zero- order spectrum.  These observations were made before 1892 at the time when Abbe was much concerned with establishing the diffraction theory of image formation in the microscope, and they were aimed to substantiate such knowledge. It is certain that they did not find their way into practical applications of the microscope.
  • 13. A. E. Conrady and J. Rheinberg in 1905, published experimental results showing that a phase reversal exists in different orders of diffraction spectra produced by the light.  It is to Professor F. Zernike, of the University of Groningen, that we are indebted for the first application of phase contrast principles to the microscope and for an explanation of these principles.  Zernike's first published work (1934) on this subject showed the advantage of the phase contrast method over the familiar knife-edge method in testing the quality of optical systems.  In 1935 Zernike discussed the application of the phase contrast method to microscopy, although it is indicated that Zernike realized the importance of the phase contrast method in microscopy as early or earlier than 1932.  However, it was not until 1941 that Kohler and Loos, of the firm of Carl Zeiss, published a paper showing the results of their extensive and convincing experiments with the Zernike method.  This phase contrast technique proved to be such an advancement in microscopy that Zernike was awarded the Nobel prize (physics) in 1953.
  • 15. OPTICAL FUNDAMENTALS  Imagine light waves passing through a glass window.  Although we think of a glass window as smooth, if we zoomed in we would see tiny imperfections.  Light passing through different parts of the glass window would actually travel through different distances of glass depending on the glass thickness at a particular point, as shown in the diagram  The peaks and troughs of the waves no longer match up We say these waves have a difference in 'phase‘  phase differences arise because the light travels through different distances depending on the thickness of the glass at each point.  a phase contrast microscope is used to transform phase differences into intensity differences, increasing contrast.
  • 16.
  • 17.  Today, two main types of phase contrast are positive and negative. Since the observed particles are usually thin and transparent, these polar contrasts provide strikingly different images.  Positive phase contrast reveals medium to dark gray images on a lighter grey background; these images often have a bright halo along the edge of the sample.  Negative phase contrast is the opposite. The specimen appears lighter with a dark background; they also have a dark halo outlining the image.
  • 18.  Phase Contrast can be performed in two different ways, on upright microscopes and inverted microscopes.  Upright microscopes are the most common type of microscope, designed with the objective lenses positioned above the sample, looking downward and usually have shorter working distances  Inverted microscopes are the most durable and easy to use for cell microscopy and tissue culture work including the diagnosis of tumor cells.  Specialized long-working distance phase contrast optical systems have been developed for inverted microscope employed for tissue culture investigations and sperm cell motility for in vitro-fertilization.
  • 19.
  • 20.  Interaction of Light Waves with Phase Specimens  An incident wavefront present in an illuminating beam of light becomes divided into two components upon passing through a phase specimen  The primary component is an undeviated (or undiffracted; zeroth-order) planar wavefront, commonly referred to as the surround (S) wave, which passes through and around the specimen, but does not interact with it.  In addition, a deviated or diffracted spherical wavefront (D-wave) is also produced, which becomes scattered over a wide arc (in many directions) that passes through the full aperture of the objective.  After leaving the specimen plane, surround and diffracted light waves enter the objective front lens element and are subsequently focused at the intermediate image plane where they combine through interference to produce a resultant particle wave (often referred to as a P-wave).
  • 21.  Detection of the specimen image depends on the relative intensity differences, and therefore on the amplitudes, of the particle and surround (P and S) waves.  If the amplitudes of the particle and surround waves are significantly different in the intermediate image plane, then the specimen acquires a considerable amount of contrast and is easily visualized in the microscope eyepieces.  Otherwise, the specimen remains transparent and appears as it would under ordinary bright field conditions (in the absence of phase contrast or other contrast-enhancing techniques).  In terms of optical path variations between the specimen and its surrounding medium, the portion of the incident light wavefront that traverses the specimen (D-wave), but does not pass through the surrounding medium (S- wave), is slightly retarded. For arguments in phase contrast microscopy, the role of the specimen in altering the optical path length (in effect, the relative phase shift) of waves passing through is of paramount importance
  • 22. the optical path length (OPL) through an object or space is the product of the refractive index (n) and the thickness (t) of the object or intervening medium as described by the relationship:  When light passes from one medium into another, the velocity is altered proportionally to the refractive index differences between the two media.  when a coherent light wave emitted by the focused microscope filament passes through a phase specimen having a specific thickness (t) and refractive index (n), the wave is either increased or decreased in velocity  The difference in location of an emergent wavefront between the specimen and surrounding medium is termed the phase shift  n(2) is the refractive index of the specimen and n(1) is the refractive index of the surrounding medium. The optical path difference results from the product of two terms: the thickness of the specimen, and its difference in refractive index with the surrounding medium.
  • 23.  For individual cells in tissue culture, the optical path difference is relatively small. A typical cell in monolayer culture has a thickness around 5 micrometers and a refractive index of approximately 1.36.  The cell is surrounded by a nutrient medium having a refractive index of 1.335, which yields an optical path difference of 0.125 micrometer, or about a quarter wavelength (of green light).  Subcellular structures generate much smaller retardations. These small optical path differences produce a linear reduction in intensity with increasing phase shift (the image grows progressively darker) up to a point (depending upon phase plate configuration), after which, the specimen image becomes brighter through reversal of contrast.  In phase contrast microscopy, the intensity of an image is dependent on a variety of factors including absorption at the phase plate, the degree of phase advancement or retardation at the phase plate, and the relative sign of this phase shift.
  • 24.
  • 25. Oxytricha saprobia in brightfield (left) and with phase contrast illumination (right)
  • 26.  To get a better understanding of how phase contrast illumination works, we study two wave fronts  First, the condenser annulus is just a small aperture located in the center (see the plane labeled '1') and the phase plate is also just covering a small aperture (located in the plane labeled '3').  Second, the optical system is greatly simplified by showing only two single lenses to represent all optical elements.
  • 27.  The plane labeled '1' is the front focal plane of the condenser. The light emanating from the small aperture 'S' is captured by the condenser and emerges as light with only parallel wavefronts from the condenser  When these plane waves (parallel wave fronts) hit the phase object 'O' (located in the object plane labeled '2'), some of this light is diffracted (and/or refracted) while moving through the specimen.  Assuming that the specimen does not significantly alter the amplitudes of the incoming wavefronts but mainly changes phase relations newly generated spherical wave fronts that are retarded by 90° (λ/4) emanate from 'O
  • 28.  the purple area that contains now "unperturbed" plane waves and spherical wave front  there are now two types of waves, the surround wave or S-wave and the diffracted wave or D-wave, which have a relative phase-shift of 90° (λ/4)  - The objective focuses the D-wave inside the primary image plane (labeled '4'), while it focuses the S-wave inside the back focal plane (labeled '3').  The location of the phase plate 'P' has now a profound impact on the S-wave while leaving most of the D-wave "unharmed"
  • 29.  In what is known as positive phase contrast optics, the phase plate 'P' reduces the amplitude of all light rays traveling through the phase annulus (mainly S- waves) by 70 to 90% and advances the phase by yet another 90° (λ/4).  Hence the recombination of these two waves (D + S) in the primary image plane (labeled '4') results in a significant amplitude change at all locations where there is a now destructive interference due to a 180° (λ/2) phase shifted D-wave  The net phase shift of 180° (λ/2) results directly from the 90° (λ/4) retardation of the D-wave due to the phase object and the 90° (λ/4) phase advancement of the S-wave due to the phase plate.
  • 30.  Without the phase plate, there would be no significant destructive interference that greatly enhances contrast.  With phase contrast illumination "invisible" phase variations are hence translated into visible amplitude variations. The destructive interference is illustrated in the figure. Blue and orange indicate D-wave and S-wave, respectively. The resulting wave (D + S), indicated by yellow, has a reduced amplitude.
  • 31.
  • 33.  Presented in Figure is a cut-away diagram of a modern upright phase contrast microscope, including a schematic illustration of the phase contrast optical train.  Partially coherent illumination is produced by the tungsten-halogen lamp  Light is directed through a collector lens and focused on a specialized annulus (labeled condenser annulus) positioned in the substage condenser front focal plane.  Wavefronts passing through the annulus illuminate the specimen and either pass through undeviated or are diffracted and retarded in phase by structures and phase gradients present in the specimen.  Rays are segregated at the rear focal plane by a phase plate and focused at the intermediate image plane to form the final phase contrast image observed in the eyepieces.
  • 34.  Two specialized accessories are required to convert a brightfield microscope for phase contrast observation.  A specially designed annular diaphragm, which is matched in diameter and optically conjugate to an internal phase plate residing in the objective rear focal plane
  • 35.
  • 36. The condenser annulus is typically constructed as an opaque flat-black (light absorbing) plate with a transparent annular ring, which is positioned so the specimen can be illuminated by defocused, parallel light wavefronts emanating from the ring.  The microscope condenser images the annular diaphragm at infinity, while the objective produces an image at the rear focal plane  The condenser annulus either replaces or resides close to the adjustable iris diaphragm in the front aperture of the condenser.  Under conditions of Köhler illumination, surround light waves that do not interact with the specimen are focused as a bright ring in the rear focal plane of the objective  The D waves passes through the diffraction plane at various locations across the entire objective rear aperture.  Thus, the two wavefronts do not overlap to a significant extent, and occupy distinct portions of the objective rear focal plane.
  • 37.  The phase contrast annuli must be specifically matched to a particular objective equipped with a corresponding phase plate  By matching the condenser annulus to the objective phase plate, the microscope can be aligned to superimpose illuminating light rays passed through the annulus onto the objective phase ring to achieve phase contrast illumination.  A majority of the popular universal condenser systems designed for phase contrast microscopy are equipped with three or more removable annular diaphragms, which are available for use with 4x, 10x, 20x, 40x, 60x, and 100x objectives containing the appropriate phase plates.
  • 38.  The PH1 - smallest aperture – 10x and 20 x  The PH2 – 40x and 60x  The PH3 – 100x  The PHL - 4x and 5x, in long working distance condensers  Condenser annulus inserts are circular aluminum plates having a stamped ring of varying dimensions in the center  After the stamping operation, annular disk units are anodized and dyed with a flat-black pigment to absorb stray light and ensure that illuminating light rays passing through the annulus follow a defined pathway.  The central light stop, is positioned in the center of the plate and secured by three slim tabs spaced at 120-degree intervals.
  • 39.  Modern universal condenser system turrets usually contain between five and eight open slot positions that can be fitted with annular phase contrast plates, DIC Nomarski (Wollaston) prism plates, or darkfield light stop plates.  The phase contrast condenser annular diaphragm plates are inserted into the appropriate position in the turret and secured into place with a spring clip
  • 40.  Whenever a universal microscope condenser is utilized, be certain to ensure that the diaphragm is opened wider than the diameter of the condenser annulus. In fact, it is a good idea to always open the diaphragm to its widest position during phase contrast observations.  Phase contrast condensers are available in a wide spectrum of optical correction levels, ranging from the basic Abbe design to highly corrected aplanatic-achromatic systems that feature excellent performance.  In addition, specialized phase contrast condensers are available with long working distance (LWD) optics & extra long working distance (ELWD).  Single annulus phase condensers were very popular at one time, but have been largely supplanted by the universal turret models. The older condensers require removal and insertion of different phase annuli as the objective is changed
  • 41.
  • 42. PHASE CONTRAST OBJECTIVES  The most important attribute of objectives designed for phase contrast microscopy is the presence of a specialized phase plate  Phase plates are not interchangeable between objectives and are often permanently etched into one of the internal lens elements  The etched ring is coated with a partially absorbing metallic film that reduces light transmission, and is usually made so that the phase of light passing through will be advanced by one quarter-wavelength relative to light that is transmitted through the rest of the glass.
  • 43.  The inset illustrates the phase plate that is positioned in the rear focal plane  The range of phase contrast objectives available from the major manufacturers covers almost the complete gamut of correction factors, including achromatic, plan achromatic, fluorite, and apochromatic.  In addition, phase contrast objectives are manufactured that vary in the neutral density and retardation value of the phase plate to produce a wide spectrum of contrast levels in both positive and negative phase contrast modes.
  • 44.  Presented below is a comparison of digital images recorded using the currently available Nikon  The images are the same viewfield from a fixed and mounted preparation of Zygnema filamentous algae. (a),(b), (c) compare the dark low low, dark low and apodized phase contrast objectives.  The dark and bright medium phase contrast objectives (d) and (e))demonstrate positive and negative phase contrast.
  • 45.  DL (Dark Low - Medium Contrast) - DL objectives produce a dark image outline on a light gray background, and are the typical objectives utilized for all-purpose phase contrast observation.  These objectives are designed to furnish the strongest dark contrast in specimens having a major difference in refractive index from that of the surrounding medium.  The DL phase contrast objective is the most popular style for examination of cells and other semi-transparent living material and is especially suited for photomicrography and digital imaging.
  • 46.  DLL (Dark Low Low - Low Contrast) - Similar in design to the DL objective, the DLL series yields better images in brightfield illumination and is often employed as a "universal" objective in microscope systems that utilize multiple illumination modes such as fluorescence, DIC, brightfield, and darkfield.  The DLL phase contrast objective produces less contrast than the DL objective, but features higher light transmission values, optical correction, and numerical aperture than the standard DL counterpart.  A majority of the DLL phase contrast objectives offered by the manufacturers have fluorite or apochromatic aberration correction levels.
  • 47.  ADL (Apodized Dark Low - Medium Contrast) - Recently introduced, the apodized phase contrast ADL objectives contain a secondary neutral density ring on either side of the central ring in the phase plate.  Addition of the secondary rings assists in reducing unwanted "halo" effects often associated with imaging large particles or specimen features (such as nuclei, whole cells and fibers) in phase contrast microscopy.  Apodized objectives are available in plan achromat optical correction and feature contrast levels similar to the DL objective series.
  • 48.  DM (Dark Medium - High Contrast) - DM objectives produce a dark image outline on a medium gray background.  These objectives are designed to be used for high image contrast with specimens having small phase shifts or refractive differences, such as fine fibers, flagella, cilia, granules, and very small particles.  Usually restricted to higher magnification objectives having large numerical apertures (fluorites and apochromats), DM phase contrast objectives perform well with very thin specimens, but often display a reversal of contrast when thick specimens are imaged.  BM (Bright Medium - High Negative Contrast) - Often referred to as negative phase contrast, BM objectives produce a bright image outline on a medium gray background.  BM objectives are ideal for visual examination of bacterial flagella, fibrin bundles, minute globules, and for blood cell counting.
  • 49.  For easy identification manufacturers inscribe important specifications on the outer barrel in green letters  In addition, phase contrast objectives have inscriptions on the barrel to indicate the objective is designed for phase contrast and also the matching annulus designation.
  • 50.
  • 51.  A typical series of phase contrast objectives are presented.  As a general rule, when objective numerical aperture and magnification is increased, the phase plate width and diameter both decrease and the condenser annulus size increases  The positive phase plate produces dark contrast and contains a partially absorbing film designed to reduce the amplitude of the surround wavefront.  In addition, this plate contains phase retarding material designed to shift (retard) the phase of the diffracted light by 90 degrees  In negative phase plate both materials are placed within the circular phase ring so that the undiffracted surround wavefront becomes the only species affected, and is attenuated and retarded in phase by 90 degrees.
  • 52.  Contrast is modulated in phase objectives by varying the properties of the phase plate, including the absorption of the metallic film (or anti-reflective coatings), the refractive index of the phase retarding material, and the thickness of the phase plate.  Phase plates are produced by vacuum deposition of thin dielectric and metallic films onto a glass plate or directly onto one of the lens surfaces within the microscope objective  The role of the dielectric film is to shift the phase of light, while the metallic film attenuates undiffracted light intensity.  The thickness and refractive indices of the dielectric, metallic, and anti- reflective films, as well as those of the optical cement, are carefully selected to produce the necessary phase shift.  In optical terminology, phase plates that alter the phase of surround light relative to diffracted light by 90 degrees (either positive or negative) are termed quarter wavelength plates because of their effect on the optical path difference
  • 53. THE PHASE TELESCOPE  If the condenser annulus is not in exact alignment with the fixed phase plate in the objective the contrast effect afforded will be dramatically compromised.  This task can be accomplished using either a phase telescope, which can be inserted into one of the eyepiece observation tubes (in place of a normal eyepiece), or a Bertrand lens built into the microscope binocular (or trinocular) eyepiece tube assembly.  The phase telescope, also commonly referred to as an auxiliary telescope or auxiliary microscope consists of a simple two or three lens telescope  The focal length of the phase telescope ranges from about 150 to 200 millimeters enabling the device to focus on the objective rear focal plane when inserted into eyepiece.
  • 54.  The Bertrand lens is more sophisticated, and acts as a relay lens, transferring an image of the objective focal plane to the microscope intermediate image plane located in the eyepiece aperture diaphragm  On most microscopes equipped with a Bertrand lens, the lens can be rotated into and out of the optical pathway by means of a small thumbwheel mechanism located beneath the eyepiece tubes  Both the phase telescope and Bertrand lens must be equipped with a mechanism for adjustment of focus, because the location of the objective rear focal plane can vary with magnification  Focusing is accomplished simply by twisting the eye tube until the objective rear focal plane is sharply focused.
  • 55. PHASE CONTRAST MICROSCOPE ALIGNMENT  ensure that all objectives contain phase plates and are firmly seated in the nosepiece.  The objectives and annular plates should also be sequentially ordered in their arrangement from lower to higher magnification
  • 56.  Shown below is the phase contrast kit used on the National Optical 160 series microscopes.  It consists of 4 objective lenses, a centering telescope and Zernike phase condenser lens.
  • 57.  The long adjustment screws on the phase condenser push in to engage set screws for proper alignment of the phase ring.  There is a thumb wheel on the opposite side used to dial in the proper setting to match the power of the objective lens.  There is also a "BF" setting on the thumb wheel for brightfield. This allows you to use the phase objectives as standard brightfield lenses.  Not all phase contrast microscopes are the same but generally they rely on similar techniques to set up the system for optimum results.’  In the system shown, the phase condenser has five settings that you spin with your thumb ( 10X, 20X, 40X, 100X and BF) BF is "brightfield", no phase.
  • 58.  To set up your microscope for phase optics, you first set it at BF and focus on the specimen.  Adjust the height of the condenser for optimum image quality.  Next, set the condenser turret to the phase setting for that particular lens and remove the specimen.  The controls that stick out from both sides on the back are for centering the condenser.
  • 59.  Next, you remove one of the eyepiece lenses and insert the centering telescope in its place.  The set screw is used to focus the centering telescope.  When looking through the telescope, you will see two rings.  They may or may not be concentric. By turning centering adjustment screws on the condenser, you align the rings so that they become concentric.
  • 60.  Once the microscope has been aligned for phase contrast, it will generally hold its centration for a considerable number of objective/annuli changes, but should be checked periodically to ensure proper alignment.  If the microscope starts to slip out of alignment, the images appearing in the eyepieces (or on a computer monitor) will appear increasingly more like those observed with brightfield illumination.  Most of the microscope manufacturers provide a green interference or absorption filter with their auxiliary phase contrast kits to produce monochromatic light having the same wavelength used for the original calibration of the objective phase plates.  As a result, contrast is increased when the filter is inserted into the optical pathway
  • 61. A majority of the commercial phase plates are designed to produce phase shifts of a quarter wavelength in the green (550 nanometers) portion of the visible light spectrum.  Theoretically, if white light is utilized instead of monochromatic light, extinction by interference will not be complete for all colors, and contrast will suffer.  This limitation is particularly important if achromat objectives, which are corrected for chromatic aberration only in the green region, are utilized for phase contrast observation or image recording.  However, with the current highly corrected fluorite and apochromatic objectives, the difference in contrast often is negligible, and therefore, insignificant.
  • 62.  How to get good images..?  The key to successful imaging with phase contrast illumination is to properly align the microscope, and to ensure that sufficiently thin specimens are spread evenly within the mounting medium on the microscope slide.  Images made with exceedingly thick specimens often suffer from out-of-focus blur and contrast inversion artifacts that can be difficult to interpret.  If the microscope is utilized for an extended period of time, occasionally check the objective rear focal plane to verify alignment of the condenser annulus with the phase plate in the objective.
  • 63. INTERPRETATION OF PHASE CONTRAST IMAGES  Images produced by phase contrast microscopy are relatively simple to interpret when the specimen is thin and distributed evenly on the substrate  When using positive phase contrast optics, which is the traditional, images appear darker than the surrounding medium when the refractive index of the specimen exceeds that of the medium.  Phase contrast optics differentially enhance the contrast near the edges  image density can be utilized as a gauge for approximating relationships between various structures.  In effect, a series of internal cellular organelles having increasing density, such as vacuoles, cytoplasm, the interphase nucleus, and the nucleolus (or mitotic chromosomes), are typically visualized as progressively darker objects relative to a fixed reference, such as the background.
  • 64.  In order to avoid confusion regarding bright and dark contrast in phase contrast images, the optical path differences occurring within the specimen preparation should be carefully considered  As discussed above, the optical path difference is derived from the product of the refractive index and the specimen (object) thickness, and is related to the relative phase shift between the specimen and background waves.  For example, a small specimen having a high refractive index can display an identical optical path difference to a larger specimen having a lower refractive index.  The external medium can also be replaced with another having either a higher or lower refractive index to generate changes in specimen image contrast. In fact, the effect on image contrast of refractive index variations in the surrounding medium forms the basis of the technique known as immersion refractometry.
  • 65.
  • 66.  Two very common effects in phase contrast images are the characteristic halo and shade-off contrast patterns in which the observed intensity does not directly correspond to the optical path difference  In all forms of positive phase contrast, bright phase halos usually surround the boundaries between large specimen features and the medium. Identical halos appear darker than the specimen with negative phase contrast optical systems  Halos occur in phase contrast microscopy because the circular phase- retarding (and neutral density) ring located in the objective phase plate also transmits a small degree of diffracted light from the specimen.  problem is compounded by the fact that the width of the zeroth-order surround wavefront projected onto the phase plate by the condenser annulus is smaller than the actual width of the phase plate ring
  • 67.  The halo effect can also be significantly reduced by utilizing specially designed phase objectives that contain a small ring of neutral density material surrounding the central phase ring material near the objective rear aperture. These objectives are termed apodized phase contrast objective  In practice, halo reduction and an increase in specimen contrast with apodized optical systems can be achieved by the utilization of selective amplitude filters  These amplitude filters consist of neutral density thin films applied to the phase plate surrounding the phase film.  Transmittance increase from 25 to 50 % with their use.
  • 68.  Shade-off is another very common optical artifact in phase contrast microscopy, and is often most easily observed in large, extended phase specimens  The intensity profile of a large, uniformly thick positive phase contrast specimen often gradually increases from the edges to the center, where the light intensity in the central region can approach that of the surrounding medium  This effect is termed shade-off, and is frequently observed when examining extended planar specimens, such as material slabs (glass or mica), replicas, flattened tissue culture cells, and large organelles.
  • 69.  The shade-off phenomenon is also commonly termed the zone-of-action effect, because central zones having uniform thickness in the specimen diffract light differently than the highly refractive zones at edges and boundaries.  Halo and shade-off artifacts depend on both the geometrical and optical properties of the phase plate and the specimen being examined.  Wider phase plates having reduced transmittance tend to produce higher intensity halos and shade-off, whereas the ring diameter has a smaller influence on these effects  In addition, these effects are heavily influenced by the objective magnification, with lower magnifications producing better images.
  • 70. ADVANTAGES  The capacity to observe living cells and, as such, the ability to examine cells in a natural state, study bilological processes.  Observing a living organism in its natural state and/or environment can provide far more information than specimens that need to be killed, fixed or stain to view under a microscope  High-contrast, high-resolution images  Ideal for studying and interpreting thin specimens  Ability to combine with other means of observation, such as fluorescence  Modern phase contrast microscopes, with CCD or CMOS computer devices, can capture photo and/or video images  In addition, advances to the phase contrast microscope, especially those that incorporate technology, enable a scientist to hone in on minute internal structures of a particle and can even detect a mere small number of protein molecules.
  • 71. DISADVANTAGES  Annuli or rings limit the aperture to some extent, which decreases resolution  This method of observation is not ideal for thick organisms or particles  Images may appear grey or green, if white or green lights are used, respectively, resulting in poor photomicrograph  Shade-off and halo effect, referred to a phase artifacts  Shade-off occurs with larger particles, results in a steady reduction of contrast moving from the center of the object toward its edges  Halo effect, where images are often surrounded by bright areas, which obscure details along the perimeter of the specimen
  • 72. CONCLUSION  The phase contrast microscope opened up an entire world of microscopy, providing incredible definition and clarity of particles never seen before.  These transparent specimens could not be explored because they do not have the capacity to absorb light.  Zernike found a way to manipulate light paths through the use of strategically placed rings and his system is a staple of most modern microscopes.  Although there are a few disadvantages, such as shade-off and halo distortions, phase contrast provides highly detailed, well-contrasted images  The most important breakthrough is the ability to observe living particles in a natural state.