Production of x-
raysDr. Priyanka rana
X-ray Production
• The production of x-rays requires a rapidly moving stream of electrons that
are suddenly decelerated or stopped. The source of electrons is the cathode,
or negative electrode. Electrons are stopped or decelerated by the anode, or
positive electrode. Electrons move between the cathode and the anode
because there is a potential difference in charge between the electrodes.
X ray tubes
• The x-ray tube produces an environment for x-ray production via
bremsstrahlung and characteristic radiation mechanisms.
Major components are the cathode, anode, rotar/stator, glass (or metal)
envelope, and tube housing .
Electrons from the cathode filament are accelerated towards the anode by a
peak voltage ranging from 20,000 to 150,000 V (20 to 150 kVp).
The tube current is the rate of electron flow from the cathode to the anode,
measured in milliamperes (mA), where 1mA = 6.24 x 10*15 electrons/sec.
CATHODe
• Negatively charged electrode
• The source of electrons in the x ray tube is the cathode, which is a helical filament of tungsten wire
surrounded by a focusing cup.
• This structure is electrically connected to the filament circuit.
• The filament circuit provides a voltage upto about 10 V to the filament, producing a current up to
about 7 A through a filament.
• Electrical resistance heat the filament and releases the electrons via a process called thermionic
emission.
• The electrons liberated from the filament flow through the vacuum of the x ray tube, when a
positive voltage is placed on the anode with respect to the cathode.
• Adjustments in the filament current (and thus in the filament temperature) control the tube current.
• A trace of the thorium in the tungsten filament increases the
efficiency of the of electron emission and prolongs filament life.
• The focusing cup is also called the cathode block, surrounds the
filament and shapes the electron beam width.
• The voltage applied to the cathode block is typically the same as
applied to the filament. This shapes the line of the electrical potential
to focus the electron beam to produce a small interaction area (focal
spot) on the anode.
• Although the width of the focusing cup slot determines the focal
spot width, the filament length determines the focal spot length.
• X ray tube has two filaments for diagnostic imaging small and large
focal spot on the target.
• The filament current determines the filament temperature and thus the rate of the
thermoinic electron emission.
• As the electrical resistance to the filament current heats the filament, electrons are emitted
from its surface.
• When no voltage is applied between the anode and the cathode of the x ray tube, an
electron cloud, also called as space charge cloud, builds around the filament. Applying a
positive high voltage to the anode with respect to the cathode accelerates the electrons
towards the anode and produces a tube current.
• Small changes in the filament current can produces relatively large changes in the tube
current.
• The existence of the space charge cloud shields the electric field for tube voltage of 40 kVp
and lower, and only a portion of free electrons are instantaneously accelerated to the
anode. When this happens, the operation of the x-ray tube is space charge limited, which
places an upper limit on the tube current, regardless of the filament current. Above 4o kVp,
the space charge cloud effect is overcome by the applied potential difference and the tube
current is limited only by the emission of electrons from the filament. Therefore, the
filament current controls the tube current in a predictable way (emission limited operation).
ANODE
• Positively charged electrode
• There are two types of anodes : Stationary anode
Rotating anode
STATIONARY:
• The simplest type of the tube has the stationary (fixed) anode. It consist of tungsten insert imbedded in the
copper block .
• The copper act as a dual role: Its supports the tungsten target and it removes heat efficiently from the tungsten
target.
• Dental x ray unit, portable x ray machine and portable fluoroscopy systems use fixed anode x ray tube.
ROTATING:
• It is mostly used in the diagnostic x ray application, mainly because of their greater heat loading and
consequently higher x ray output capabilities.
• Electrons impart their energy on a continuously rotating target, spreading thermal energy over a large area and
a mass of the anode disc.
• A bearing mounted rotor assembly supports the anode disc within the evacuated x ray tube insert .
• X ray machines are designed so that the x ray tube will not be
energized if the anode is not upto full speed, this is the cause for
the short delay (1-2 seconds) when the x ray tube exposure
button is pushed.
• Rotor bearings are heat sensitive and are often the cause of the
x ray tube failure. Bearings are in the high-vaccum environment
of the insert and require special heat – insensitive , non volatile
lubricants.
• A molybdenum stem attaches the anode to the rotor/ bearing
assembly, because molybdenum is poor heat conductor and
reduces heat transfer from the anode to the bearings. Because
it is thermally isolated, the anode must be cooled by radiative
emission.
X-rays are produced by two main mechanisms and
come in two varieties –
characteristic and bremsstrahlung
SPECTRUM
bremsstrahlung
• The conversion of electron kinetic energy into electromagnetic radiation produces x ray.
• A Large voltage is applied between two electrodes (the cathode and the anode) in a
evacuated envelope. The cathode is negatively charged and is the source of electrons; the
anode is positively charged and is the target of electrons. As electrons from the cathode
travel to the anode, they are accelerated by the electrical potential difference between
these electrodes and attain kinetic energy.
• The kinetic energy gained by an electron is proportional to the potential difference
between the cathode and anode.
• On impact with the target, the kinetic energy of the electrons is converted to the other
form of energy. The vast majority of interactions produce unwanted heat by small
collisional energy exchange with electrons in the target. This intense heating limits the
number of x ray photons that can be produced in a given time without destroying the
target. Occasionally an electron comes within the proximity of a positively charged
nucleus in the target electrode.
• The columbic forces attract and decelerate the electron, causing a
significant loss of kinetic energy and a change in a electron’s trajectory. An
x ray photon with energy equal to the kinetic energy lost by the electron is
produced. This radiation is termed bremsstrahlung.
• The subatomic distance between the bombarding electron and the nucleus
determines the energy lost by each electron during the bremsstrahlung
process because the columbic force of attraction increases with the
inverse square of the interaction distance. At relatively large distances
from the nucleus , the columbic attraction force is weak. These encounters
produce low x ray energy.
• For closer interaction distances the force acting on the electron increases,
causing a more dramatic change in the electron’s trajectory and a larger
loss of energy; these encounters produce high x ray energy.
• A direct impact of an electron with the target nucleus results in loss of all
the electron’s kinetic energy.
• The probability of an electron’s directly impacting a nucleus is
extremely low, simply because at the atomic scale, the atom
comprises mainly empty “space” and the nuclear cross section is
very small. Therefore, lower x ray energies are generated in
greater abundance , and the number of higher energy x rays
decreases approximately linearly with energy upto the maximum
energy of the incident electrons.
characteristic
• Each electron in the target atom has a binding energy that depends on the shell in
which it resides.
• Closest to the nucleus are the two electrons in the K shell, which has the highest
binding energy. The L shell, with 8 electrons has the next highest binding energy,
and so forth.
• When the energy of an electron incident on the target exceeds the binding energy
of the electron of a target atom it is energetically possible for a collisional
interaction to eject the electron and ionize the atom.
• The unfilled shell is energetically unstable and an outer shell electron with less
binding energy will fill the vacancy. As this electron transitions to lower energy
state the excess energy can be released as a characteristic x ray photon with an
energy equal to the difference between the binding energies of the electron shell.
• Binding energies are unique to a given element, and so their
differences; consequently the emitted x rays have discrete
energies that are characteristic of that element.
• Many electron transitions can occur from adjacent and non
adjacent shells in the atom, giving rise to the several discrete
energy peaks superimposed on the continuous bremsstrahlung
spectrum. The most prevalent characteristic x ray in the
diagnostic energy range result from K shell vacancies, which are
filled by electrons from the L, M N shells.
• Characteristic K x rays are emitted only when the electrons
impinging on the target exceeds the binding energy of the K
shell electron.
QUESTION NO.1
• What is the role of copper in stationary anode?
• Its supports the tungsten target and it removes heat efficiently
from the tungsten target.

Mais conteúdo relacionado

Mais procurados

Production of x rays
Production of x raysProduction of x rays
Production of x raysaslam bs
 
Production of x rays
Production of x raysProduction of x rays
Production of x raysDeepaGautam
 
Interaction of x ray with matter
Interaction of x ray with matterInteraction of x ray with matter
Interaction of x ray with matterAnjan Dangal
 
X ray beam rest
X ray beam restX ray beam rest
X ray beam restBuandari
 
Half value layer.pptx
Half value layer.pptxHalf value layer.pptx
Half value layer.pptxGokulAnanth9
 
interaction of xray with matter
interaction of xray with matterinteraction of xray with matter
interaction of xray with matterdypradio
 
Multisection tomography and Transaxial Tomography
Multisection tomography and Transaxial TomographyMultisection tomography and Transaxial Tomography
Multisection tomography and Transaxial TomographySelf
 
xray tubes
xray tubesxray tubes
xray tubesaugpax
 
X ray attenuation
X ray attenuationX ray attenuation
X ray attenuationaslam bs
 
Anode heel effect, line focus principle,
Anode heel effect, line focus principle,Anode heel effect, line focus principle,
Anode heel effect, line focus principle,InosRagan
 
Computed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector ConfigurationComputed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector ConfigurationAnjan Dangal
 
Portable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptxPortable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptxSaritaGaire1
 

Mais procurados (20)

X ray tube
X ray tubeX ray tube
X ray tube
 
Grids
GridsGrids
Grids
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
Production of x rays
Production of x raysProduction of x rays
Production of x rays
 
Interaction of x ray with matter
Interaction of x ray with matterInteraction of x ray with matter
Interaction of x ray with matter
 
X ray beam rest
X ray beam restX ray beam rest
X ray beam rest
 
Half value layer.pptx
Half value layer.pptxHalf value layer.pptx
Half value layer.pptx
 
interaction of xray with matter
interaction of xray with matterinteraction of xray with matter
interaction of xray with matter
 
Multisection tomography and Transaxial Tomography
Multisection tomography and Transaxial TomographyMultisection tomography and Transaxial Tomography
Multisection tomography and Transaxial Tomography
 
Filters
FiltersFilters
Filters
 
xray tubes
xray tubesxray tubes
xray tubes
 
X-RAY TUBE (ANODE)
X-RAY TUBE (ANODE)X-RAY TUBE (ANODE)
X-RAY TUBE (ANODE)
 
X Ray Production
X Ray ProductionX Ray Production
X Ray Production
 
X ray attenuation
X ray attenuationX ray attenuation
X ray attenuation
 
X-ray Tube (Cathode)
X-ray Tube (Cathode)X-ray Tube (Cathode)
X-ray Tube (Cathode)
 
Anode heel effect, line focus principle,
Anode heel effect, line focus principle,Anode heel effect, line focus principle,
Anode heel effect, line focus principle,
 
Radiographic grids
Radiographic gridsRadiographic grids
Radiographic grids
 
Automatic exposure control
Automatic exposure controlAutomatic exposure control
Automatic exposure control
 
Computed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector ConfigurationComputed Tomography Instrumentation and Detector Configuration
Computed Tomography Instrumentation and Detector Configuration
 
Portable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptxPortable and mobile radiographic equipments [Autosaved].pptx
Portable and mobile radiographic equipments [Autosaved].pptx
 

Semelhante a Production of xrays

02 x ray production and its interaction with matter
02 x ray production and its interaction with matter02 x ray production and its interaction with matter
02 x ray production and its interaction with matterAkram Kothwal Syed
 
X RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptxX RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptxSrinath Chowdary
 
Xrayproductionandproperties march 9 2022
Xrayproductionandproperties march 9 2022Xrayproductionandproperties march 9 2022
Xrayproductionandproperties march 9 2022dypradio
 
Radiation protection course for radiologists L1
Radiation protection course for radiologists L1Radiation protection course for radiologists L1
Radiation protection course for radiologists L1Amin Amin
 
X ray tube production rakesh
X ray tube production rakeshX ray tube production rakesh
X ray tube production rakeshSylvia Diamond
 
X ray production and properties
X ray production and propertiesX ray production and properties
X ray production and propertiesHanuman Doke
 
xrayproductionandproperties-171229054704.pdf
xrayproductionandproperties-171229054704.pdfxrayproductionandproperties-171229054704.pdf
xrayproductionandproperties-171229054704.pdfAvneeshKumar164042
 
Diagnostic radiography concepts x ray equipment
Diagnostic radiography concepts x ray equipmentDiagnostic radiography concepts x ray equipment
Diagnostic radiography concepts x ray equipmentVijaykumar Garasiya
 
The x ray beam
The x ray beam The x ray beam
The x ray beam airwave12
 
Atomic structure as applied to generation of X-rays.pptx
Atomic structure as applied to generation of X-rays.pptxAtomic structure as applied to generation of X-rays.pptx
Atomic structure as applied to generation of X-rays.pptxDr. Dheeraj Kumar
 
Radioactivity and production of X-rays - Sachin
Radioactivity and production of X-rays - SachinRadioactivity and production of X-rays - Sachin
Radioactivity and production of X-rays - SachinSACHINS700327
 
SPM Form 5 Physics - Electronics
SPM Form 5 Physics - ElectronicsSPM Form 5 Physics - Electronics
SPM Form 5 Physics - ElectronicsLoo Carmen
 
DLA 1204 Radiation Physics
DLA 1204 Radiation PhysicsDLA 1204 Radiation Physics
DLA 1204 Radiation PhysicsKatieHenkel1
 

Semelhante a Production of xrays (20)

02 x ray production and its interaction with matter
02 x ray production and its interaction with matter02 x ray production and its interaction with matter
02 x ray production and its interaction with matter
 
X RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptxX RAY TUBES & PRODUCTION OF.pptx
X RAY TUBES & PRODUCTION OF.pptx
 
Xrayproductionandproperties march 9 2022
Xrayproductionandproperties march 9 2022Xrayproductionandproperties march 9 2022
Xrayproductionandproperties march 9 2022
 
X ray physics
X ray physicsX ray physics
X ray physics
 
Production of xray
Production  of  xrayProduction  of  xray
Production of xray
 
X-ray Production
X-ray ProductionX-ray Production
X-ray Production
 
X-RAY TUBE.pptx
X-RAY TUBE.pptxX-RAY TUBE.pptx
X-RAY TUBE.pptx
 
Radiation protection course for radiologists L1
Radiation protection course for radiologists L1Radiation protection course for radiologists L1
Radiation protection course for radiologists L1
 
X ray tube production rakesh
X ray tube production rakeshX ray tube production rakesh
X ray tube production rakesh
 
Xray production 2
Xray production 2Xray production 2
Xray production 2
 
X ray production and properties
X ray production and propertiesX ray production and properties
X ray production and properties
 
xrayproductionandproperties-171229054704.pdf
xrayproductionandproperties-171229054704.pdfxrayproductionandproperties-171229054704.pdf
xrayproductionandproperties-171229054704.pdf
 
Diagnostic radiography concepts x ray equipment
Diagnostic radiography concepts x ray equipmentDiagnostic radiography concepts x ray equipment
Diagnostic radiography concepts x ray equipment
 
The x ray beam
The x ray beam The x ray beam
The x ray beam
 
Atomic structure as applied to generation of X-rays.pptx
Atomic structure as applied to generation of X-rays.pptxAtomic structure as applied to generation of X-rays.pptx
Atomic structure as applied to generation of X-rays.pptx
 
GenerationofXRays.ppt
GenerationofXRays.pptGenerationofXRays.ppt
GenerationofXRays.ppt
 
Radioactivity and production of X-rays - Sachin
Radioactivity and production of X-rays - SachinRadioactivity and production of X-rays - Sachin
Radioactivity and production of X-rays - Sachin
 
SPM Form 5 Physics - Electronics
SPM Form 5 Physics - ElectronicsSPM Form 5 Physics - Electronics
SPM Form 5 Physics - Electronics
 
X-Ray Shadowgraph Imaging Technique Lecture
X-Ray Shadowgraph Imaging Technique LectureX-Ray Shadowgraph Imaging Technique Lecture
X-Ray Shadowgraph Imaging Technique Lecture
 
DLA 1204 Radiation Physics
DLA 1204 Radiation PhysicsDLA 1204 Radiation Physics
DLA 1204 Radiation Physics
 

Mais de priyanka rana

Benign Bone Tumors and Tumor Like Conditions
Benign Bone Tumors and Tumor Like Conditions Benign Bone Tumors and Tumor Like Conditions
Benign Bone Tumors and Tumor Like Conditions priyanka rana
 
Inflammatory arthritis
Inflammatory arthritisInflammatory arthritis
Inflammatory arthritispriyanka rana
 
Congenital, Acquired Pyogenic, and Acquired Viral Infection
Congenital, Acquired Pyogenic, and Acquired Viral InfectionCongenital, Acquired Pyogenic, and Acquired Viral Infection
Congenital, Acquired Pyogenic, and Acquired Viral Infectionpriyanka rana
 
Antomy of scrotum and testis
Antomy of scrotum and testisAntomy of scrotum and testis
Antomy of scrotum and testispriyanka rana
 

Mais de priyanka rana (11)

Pelvic floor
Pelvic floorPelvic floor
Pelvic floor
 
MRI safety
MRI safety MRI safety
MRI safety
 
HIFU
HIFUHIFU
HIFU
 
Radiology spotters
Radiology spottersRadiology spotters
Radiology spotters
 
Benign Bone Tumors and Tumor Like Conditions
Benign Bone Tumors and Tumor Like Conditions Benign Bone Tumors and Tumor Like Conditions
Benign Bone Tumors and Tumor Like Conditions
 
Anatomy of pns
Anatomy of pnsAnatomy of pns
Anatomy of pns
 
Inflammatory arthritis
Inflammatory arthritisInflammatory arthritis
Inflammatory arthritis
 
Xeroradiography
XeroradiographyXeroradiography
Xeroradiography
 
Congenital, Acquired Pyogenic, and Acquired Viral Infection
Congenital, Acquired Pyogenic, and Acquired Viral InfectionCongenital, Acquired Pyogenic, and Acquired Viral Infection
Congenital, Acquired Pyogenic, and Acquired Viral Infection
 
Antomy of scrotum and testis
Antomy of scrotum and testisAntomy of scrotum and testis
Antomy of scrotum and testis
 
Facial fracture
Facial fractureFacial fracture
Facial fracture
 

Último

Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)eniolaolutunde
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3JemimahLaneBuaron
 
General AI for Medical Educators April 2024
General AI for Medical Educators April 2024General AI for Medical Educators April 2024
General AI for Medical Educators April 2024Janet Corral
 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13Steve Thomason
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeThiyagu K
 
Student login on Anyboli platform.helpin
Student login on Anyboli platform.helpinStudent login on Anyboli platform.helpin
Student login on Anyboli platform.helpinRaunakKeshri1
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdfQucHHunhnh
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfAdmir Softic
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxiammrhaywood
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdfSoniaTolstoy
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Sapana Sha
 
9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room servicediscovermytutordmt
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAssociation for Project Management
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfJayanti Pande
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityGeoBlogs
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfchloefrazer622
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactdawncurless
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Disha Kariya
 

Último (20)

Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)
 
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3
 
General AI for Medical Educators April 2024
General AI for Medical Educators April 2024General AI for Medical Educators April 2024
General AI for Medical Educators April 2024
 
The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13The Most Excellent Way | 1 Corinthians 13
The Most Excellent Way | 1 Corinthians 13
 
Measures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and ModeMeasures of Central Tendency: Mean, Median and Mode
Measures of Central Tendency: Mean, Median and Mode
 
Student login on Anyboli platform.helpin
Student login on Anyboli platform.helpinStudent login on Anyboli platform.helpin
Student login on Anyboli platform.helpin
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
 
9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room service
 
Advance Mobile Application Development class 07
Advance Mobile Application Development class 07Advance Mobile Application Development class 07
Advance Mobile Application Development class 07
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across Sectors
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdf
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activity
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdf
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impact
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..
 

Production of xrays

  • 1. Production of x- raysDr. Priyanka rana
  • 2. X-ray Production • The production of x-rays requires a rapidly moving stream of electrons that are suddenly decelerated or stopped. The source of electrons is the cathode, or negative electrode. Electrons are stopped or decelerated by the anode, or positive electrode. Electrons move between the cathode and the anode because there is a potential difference in charge between the electrodes.
  • 3.
  • 4. X ray tubes • The x-ray tube produces an environment for x-ray production via bremsstrahlung and characteristic radiation mechanisms. Major components are the cathode, anode, rotar/stator, glass (or metal) envelope, and tube housing . Electrons from the cathode filament are accelerated towards the anode by a peak voltage ranging from 20,000 to 150,000 V (20 to 150 kVp). The tube current is the rate of electron flow from the cathode to the anode, measured in milliamperes (mA), where 1mA = 6.24 x 10*15 electrons/sec.
  • 5.
  • 6. CATHODe • Negatively charged electrode • The source of electrons in the x ray tube is the cathode, which is a helical filament of tungsten wire surrounded by a focusing cup. • This structure is electrically connected to the filament circuit. • The filament circuit provides a voltage upto about 10 V to the filament, producing a current up to about 7 A through a filament. • Electrical resistance heat the filament and releases the electrons via a process called thermionic emission. • The electrons liberated from the filament flow through the vacuum of the x ray tube, when a positive voltage is placed on the anode with respect to the cathode. • Adjustments in the filament current (and thus in the filament temperature) control the tube current.
  • 7. • A trace of the thorium in the tungsten filament increases the efficiency of the of electron emission and prolongs filament life. • The focusing cup is also called the cathode block, surrounds the filament and shapes the electron beam width. • The voltage applied to the cathode block is typically the same as applied to the filament. This shapes the line of the electrical potential to focus the electron beam to produce a small interaction area (focal spot) on the anode. • Although the width of the focusing cup slot determines the focal spot width, the filament length determines the focal spot length. • X ray tube has two filaments for diagnostic imaging small and large focal spot on the target.
  • 8. • The filament current determines the filament temperature and thus the rate of the thermoinic electron emission. • As the electrical resistance to the filament current heats the filament, electrons are emitted from its surface. • When no voltage is applied between the anode and the cathode of the x ray tube, an electron cloud, also called as space charge cloud, builds around the filament. Applying a positive high voltage to the anode with respect to the cathode accelerates the electrons towards the anode and produces a tube current. • Small changes in the filament current can produces relatively large changes in the tube current. • The existence of the space charge cloud shields the electric field for tube voltage of 40 kVp and lower, and only a portion of free electrons are instantaneously accelerated to the anode. When this happens, the operation of the x-ray tube is space charge limited, which places an upper limit on the tube current, regardless of the filament current. Above 4o kVp, the space charge cloud effect is overcome by the applied potential difference and the tube current is limited only by the emission of electrons from the filament. Therefore, the filament current controls the tube current in a predictable way (emission limited operation).
  • 9. ANODE • Positively charged electrode • There are two types of anodes : Stationary anode Rotating anode STATIONARY: • The simplest type of the tube has the stationary (fixed) anode. It consist of tungsten insert imbedded in the copper block . • The copper act as a dual role: Its supports the tungsten target and it removes heat efficiently from the tungsten target. • Dental x ray unit, portable x ray machine and portable fluoroscopy systems use fixed anode x ray tube. ROTATING: • It is mostly used in the diagnostic x ray application, mainly because of their greater heat loading and consequently higher x ray output capabilities. • Electrons impart their energy on a continuously rotating target, spreading thermal energy over a large area and a mass of the anode disc. • A bearing mounted rotor assembly supports the anode disc within the evacuated x ray tube insert .
  • 10. • X ray machines are designed so that the x ray tube will not be energized if the anode is not upto full speed, this is the cause for the short delay (1-2 seconds) when the x ray tube exposure button is pushed. • Rotor bearings are heat sensitive and are often the cause of the x ray tube failure. Bearings are in the high-vaccum environment of the insert and require special heat – insensitive , non volatile lubricants. • A molybdenum stem attaches the anode to the rotor/ bearing assembly, because molybdenum is poor heat conductor and reduces heat transfer from the anode to the bearings. Because it is thermally isolated, the anode must be cooled by radiative emission.
  • 11.
  • 12.
  • 13. X-rays are produced by two main mechanisms and come in two varieties – characteristic and bremsstrahlung SPECTRUM
  • 14. bremsstrahlung • The conversion of electron kinetic energy into electromagnetic radiation produces x ray. • A Large voltage is applied between two electrodes (the cathode and the anode) in a evacuated envelope. The cathode is negatively charged and is the source of electrons; the anode is positively charged and is the target of electrons. As electrons from the cathode travel to the anode, they are accelerated by the electrical potential difference between these electrodes and attain kinetic energy. • The kinetic energy gained by an electron is proportional to the potential difference between the cathode and anode. • On impact with the target, the kinetic energy of the electrons is converted to the other form of energy. The vast majority of interactions produce unwanted heat by small collisional energy exchange with electrons in the target. This intense heating limits the number of x ray photons that can be produced in a given time without destroying the target. Occasionally an electron comes within the proximity of a positively charged nucleus in the target electrode.
  • 15.
  • 16. • The columbic forces attract and decelerate the electron, causing a significant loss of kinetic energy and a change in a electron’s trajectory. An x ray photon with energy equal to the kinetic energy lost by the electron is produced. This radiation is termed bremsstrahlung. • The subatomic distance between the bombarding electron and the nucleus determines the energy lost by each electron during the bremsstrahlung process because the columbic force of attraction increases with the inverse square of the interaction distance. At relatively large distances from the nucleus , the columbic attraction force is weak. These encounters produce low x ray energy. • For closer interaction distances the force acting on the electron increases, causing a more dramatic change in the electron’s trajectory and a larger loss of energy; these encounters produce high x ray energy. • A direct impact of an electron with the target nucleus results in loss of all the electron’s kinetic energy.
  • 17. • The probability of an electron’s directly impacting a nucleus is extremely low, simply because at the atomic scale, the atom comprises mainly empty “space” and the nuclear cross section is very small. Therefore, lower x ray energies are generated in greater abundance , and the number of higher energy x rays decreases approximately linearly with energy upto the maximum energy of the incident electrons.
  • 18.
  • 19. characteristic • Each electron in the target atom has a binding energy that depends on the shell in which it resides. • Closest to the nucleus are the two electrons in the K shell, which has the highest binding energy. The L shell, with 8 electrons has the next highest binding energy, and so forth. • When the energy of an electron incident on the target exceeds the binding energy of the electron of a target atom it is energetically possible for a collisional interaction to eject the electron and ionize the atom. • The unfilled shell is energetically unstable and an outer shell electron with less binding energy will fill the vacancy. As this electron transitions to lower energy state the excess energy can be released as a characteristic x ray photon with an energy equal to the difference between the binding energies of the electron shell.
  • 20. • Binding energies are unique to a given element, and so their differences; consequently the emitted x rays have discrete energies that are characteristic of that element. • Many electron transitions can occur from adjacent and non adjacent shells in the atom, giving rise to the several discrete energy peaks superimposed on the continuous bremsstrahlung spectrum. The most prevalent characteristic x ray in the diagnostic energy range result from K shell vacancies, which are filled by electrons from the L, M N shells. • Characteristic K x rays are emitted only when the electrons impinging on the target exceeds the binding energy of the K shell electron.
  • 21.
  • 22.
  • 23.
  • 24. QUESTION NO.1 • What is the role of copper in stationary anode?
  • 25. • Its supports the tungsten target and it removes heat efficiently from the tungsten target.

Notas do Editor

  1. K shell – higher binding energy The electron interacts with the k shell e- via a repulsive electrical force ----the k shell e- is removed( the energy of e- is > k shell binding energy.------leaving a vacancy in the k shell ----.----The unfilled shell is unstable and an outer shell electron with less binding energy will fill the vacancy------an e- from the adjacent l shell fills the vacancy -----energy can be released as a characteristic x ray photon with an energy equal to the difference between the binding energies of the electron shell.