Advanced Haemodynamic
 Monitoring and support
Part 1: Monitoring
Contents

1.    Introduction                                    6. Monitoring Perfusion – Gastric Tonometry
2.    What is haemodynamic monitoring?                7. Monitoring Oxygen Consumption – Mixed
3.    Advanced haemodynamic monitoring                   Venous Saturation
4.    Central Venous Pressure Monitoring              8. Summary
5.    Cardiac Output Measurement
     a)   Introduction
     b)   Derived parameters from cardiac output
     c)   Cardiac Output using CO2
     d)   Cardiac output using dilution technique
     e)   Pulmonary Artery Catheters
     f)   Transpulmonary Thermodilution
     g)   Lithium dilution
     h)   Stroke Volume / Pulse Pressure Variation
     i)   Oesophageal Doppler
     j)   Echocardiography
     k)   Thoracic bioimpedance
     l)   Indications for monitoring cardiac output
     m)   Clinical applications
     n)   Evidence to support cardiac output
          monitoring
Introduction

•   Monitoring is one of the fundamental aspects of intensive care
•   We monitor several physiological parameters regularly:
     – ECG, heart rate, SpO2, arterial blood pressure in real time
     – Other parameters such as temperature, urine output, respiration rate, GCS
       and blood sugars are monitored at frequent, but regular intervals
•   We monitor these parameters:
     – To alert us of deterioration or potential deterioration of patient’s state
     – To intervene when there are signs pointing to a deterioration
     – To assess response to treatment




       ‘Not everything that counts can be counted and not everything that
       can be counted counts’ . . . Albert Einstein
What is haemodynamic monitoring?
•   Haemodynamic monitoring focuses on the circulatory system, its adequacy
    and end organ perfusion
•   Markers such as blood pressure and heart rate give an indirect measure of the
    circulatory system
•   Tachycardia and hypotension may thus indicate either a decreased preload or
    impaired myocardial contractility
•   They are not accurate:
     – The heart rate is influenced by many other factors
     – Tachycardia may be the result of pain, anxiety, drugs or arrhythmias and
       not necessarily a reduced preload or decreased cardiac output
     – Similarly, a low blood pressure may be secondary to drugs, postural
       effects or vasodilatation
     – A ‘normal’ blood pressure in a hypertensive patient may actually indicate a
       shocked state!
What is haemodynamic monitoring?
•    Surrogate markers of organ perfusion include:
      – Skin perfusion - measured by capillary refill
      – Urine output
      – Mental state – usually assessed by the GCS
•    These may do not accurately measure perfusion:
      – Capillary refill may be affected by other factors including ambient
        temperature
      – Urine output may be vary in the setting of previous renal dysfunction
      – Mental alertness may not be assessable in the intubated patient;
        delirium may contribute to a decreased GCS
•    Other indirect markers of perfusion including serum lactate are not exclusive
     to poor perfusion and are elevated in other medical conditions as well
Advanced haemodynamic monitoring
•    The ‘regular’ monitoring of the haemodynamic status provides sufficient
     information in most situations
•    There are situations where this may not sufficient to make decisions
•    Advanced haemodynamic monitoring involves monitoring the circulatory
     system and the components that affect it :
      – Preload
      – Myocardial contractility
      – Afterload
•    It also involves monitoring oxygen consumption and end organ perfusion
Central Venous Pressure Monitoring
Central Venous Pressure
•   The pressure measured in the superior and inferior venacava
•   Reflects right atrial pressure
•   Usually measured through a catheter inserted into the internal jugular or
    subclavian veins as well as through peripherally inserted central venous
    catheters
•   The pressure in the femoral veins also reflects central venous pressure,
    although increased abdominal pressure may modify this sufficiently to render
    it inaccurate
•   The right atrial pressure may be directly measured using the pulmonary artery
    catheter (see section on PA catheters)
•   CVP pressure monitoring is common - catheters in the central veins are
    inserted frequently for administration of vasoactive agents or hypertonic
    solutions and for long-term venous access
History

•     First described in patients undergoing thoracic surgery
•     Hughes and Magovern1 described a fall in CVP with blood loss and a rise in
      CVP with transfusion in 1959
•     CVP was initially used as a guide for fluid resuscitation and monitoring in
      cardiac and thoracic surgery and later on was used in major operative
      procedures
•     This practice then spilled over to the Intensive Care Units and now forms a
      common modality of haemodynamic monitoring in critically ill patients




1. Hughes RE, Magovern GJ. The Relationship Between Right Atrial Pressure and Blood Volume. Arch Surg. 1959;79: 238-243.
Central Venous Pressure
•   Used as a surrogate marker for cardiac filling pressures - rise in CVP
    indicates better cardiac filling
•   Frank- Starling principle states that the force of cardiac contraction is
    proportional to the end-diastolic length of the fibre
•   This muscle length is proportional to the volume of blood at the end of
    diastole: the preload
•   Thus, preload is one of the determinants of cardiac output
•   Increased preload results in proportionately increased stroke volume
•   In clinical practice, it is not possible to measure the end diastolic volume of
    the right ventricle to determine the preload
•   End diastolic right ventricular pressure is reflected in the CVP which is
    therefore used as a surrogate marker of preload
Central Venous Pressure
•   The relationship between ventricular
    end diastolic volume and pressure is
    not linear but curvilinear (see figure)
                                                                                                       Decreased
•   When the ventricle is operating in the                                                             compliance
    flat portion of the curve, there is little




                                                 Left ventricular pressure
    change in pressure with end diastolic
    volume                                                                                                            Normal
                                                                              End diastolic pressure
•   In the steep area of the curve, small
    changes of volume results in greater
    changes in pressure                                                      End diastolic pressure

•   Decreased ventricular compliance                                                                                    Increased
                                                                                                                       compliance
                                                                             End diastolic pressure
    causes greater pressure changes with
    volume                                                                                  Left ventricular volume

•   Ventricular compliance is influenced
    by relaxation, geometry and the
    mechanical characteristics of the
    ventricle
Factors affecting CVP
•   Normal CVP in a healthy adult: 0 -7 mm Hg
•   CVP of 0 may be normal in a healthy person - central veins are capacitance veins,
    potentially able to hold large volumes of blood without appreciable changes in pressure
•   Higher CVP may reflect poor ventricular compliance than good filling pressures
•   Changes in the transmural pressure - difference in the pressure between the cardiac
    chamber and the juxtacardiac pressure – also affects CVP
•   Juxtacardiac pressure is affected by pericardial and pleural pressures
•   In a ventilated patient, pleural pressures are affected by the compliance of the lung,
    pressures exerted by the ventilator, and PEEP
•   These pressure changes are complex and do not affect the CVP in a linear fashion. The
    practice of subtracting PEEP pressure to the CVP may not always reflect actual CVP
•   Pressures in the right atria may also be affected by changes in the ventricles or the atria:
     – Valvular abnormalities of the tricuspid and rarely, the pulmonary valves
     – Pulmonary hypertension
Factors affecting CVP…
  •     The pressures are affected by changes in the respiratory cycle:
         – Decreased intra-thoracic pressure at the beginning of spontaneous
           ventilation decreases the CVP
         – The onset of positive pressure ventilation causes an increase in CVP
  • The CVP is usually measured at the end of expiration when
    juxtacardiac pressures approach atmospheric pressure
  • Technical issues may also affect the actual CVP measurement,
    including:
         –   Improper position of the catheter – abutment against a vessel wall
         –   Thrombus in the catheter
         –   Improper zeroing or placement of the transducer
         –   Simultaneous rapid Infusion of fluids1 (rates > 50 ml/hour tends to
             increase the CVP)



1. Ho AM-H, Dion PW, Karmakar MK and Jenkins CR. Accuracy of central venous pressure monitoring during simultaneous continuous
infusion through the same catheter. Anaesthesia, 2005; 60: 1027–1030
CVP waveform
•   Waveform has a characteristic trace
    coinciding with changes in the cardiac cycle:
     – Prominent positive a wave - atrial
       contraction
     – Smaller, positive c wave – closure of
       tricuspid valve during isovolumetric
       contraction
•   Atrial relaxation decreases in atrial pressure
    (x descent)
•   Filling of the atrium causes pressures to rise
    again (v wave)
•   This is followed by the y descent, a negative
    waveform reflecting ventricular filling as the
    tricuspid valve opens
Monitoring with CVP
•      Used both in resuscitation and as a guide to fluid therapy
•      Low CVP in a haemodynamically unstable patient reflects decreased preload
       and may be an indication for additional fluids
•      Its use to guide fluid resuscitation is recommended and in several guidelines
       including management of septic shock
•      Evidence to support this is however not convincing
•      In a systematic review, Marik1 found no correlation between CVP and blood
       volume
•      This review also found that CVP, or a change in CVP did not predict fluid
       responsiveness
•      The CVP should not be used in isolation and the pressure measured is a
       composite of a number of factors affecting the right atrial pressure
•      Trends in the change in the CVP and using the CVP in conjunction with other
       haemodynamic parameters helps overcome some of these barriers

    1. Marik PE, Baram B, Vahid B. Does Central Venous Pressure Predict Fluid Responsiveness? Chest, 2008; 134: 172 - 178
Cardiac Output Measurement
Introduction

•   Assessment of cardiac output is an important part of monitoring
    haemodynamically unstable patients
•   Until recently, the measurement of cardiac output was considered essential:
     – To characterize the type of shock
     – To monitor response to therapeutic interventions
     – To optimise factors that influence cardiac output
     – To optimise oxygen delivery
Definitions
Derived parameters from cardiac output
           Parameter                                    Calculation                           Normal values

    Stroke Volume (index)                                                                    60 – 100 ml/ beat
                                                                                           (33 - 47 ml/m2/beat)

 Systemic Vascular Resistance                                                             1000 -1500 dyne s/ cm5
           (index)                                                                      (1970 - 2390 dyne s/cm5/m2)


Pulmonary Vascular Resistance                                                                <250 dyne s/cm5
           (index)                                                                        (255 - 285 dyne s/cm5/m2)


 Left Ventricular Stroke Work                                                              58 - 104 gm-m/beat
            (index)                                                                      (50 - 62 gm-m/m2/beat)

 Right Ventricular Stroke Work                                                               8 - 16 gm-m/beat
            (index)                                                                       (5 - 10 gm-m/m2/beat)


    MAP= mean arterial pressure; RAP = right atrial pressure ( or CVP); PAWP = pulmonary artery wedge pressure;
    MPAP = mean pulmonary artery pressure (pulmonary systolic pressure +(2 x pulmonary diastolic pressure)/ 3); 80
    and 0.10136 are the numbers required for conversion to the units of measurement
Derived parameters…

          Parameter                               Calculation                       Normal values


 Coronary Artery Perfusion                     Diastolic BP-PAWP                    60 - 80 mmHg
        pressure

  Oxygen Delivery (index)                  CaO2 x Cardiac Output (CI)             950 - 1150 ml/min
                                                                                (500 - 600 ml/min/m2)

Oxygen Consumption (index)                  (C(a - v)O2) x CO (CI)                 200 - 250 ml/min
                                                                                (120 - 160 ml/min/m2)




CaO2= oxygen content of arterial blood ({1.34 x Hb/L x SaO2} + {0.0031 x PaO2}); (C(a - v)O2) =
difference between arterial and venous oxygen content;
Cardiac output measurement: Fick Principle
Cardiac Output using CO2
Cardiac output using CO2
Cardiac output measurement using CO2
Cardiac output measurement using CO2

•   The potential benefits using this technique include its non-invasive nature and its ability to be
    repeated at relatively frequent intervals
•   It can be only done in intubated and ventilated patients; large gas leaks in non-invasive
    ventilation preclude its use
•   It has not shown consistent correlation with other techniques of measurement of CO
•   Its measurement is unreliable in several clinical situations:
     – Large intra pulmonary shunts
     – Where tidal volumes vary (spontaneously breathing patient) or when ventilator settings
       may alter dead space
     – Chronic lung disease, where ventilator- perfusion mismatch can occur
     – The postoperative cardiac patient where increased pulmonary dead space and shunts
       may occur
     – Acute lung injury (ALI) and ARDS where the heterogeneous nature of the injury causes
       a wide variation of gas exchange. End tidal CO2 may be a poor estimate of CaCO2
Cardiac output using dilution technique
Cardiac output measurement: Indicator-dilution




                         Concentration of indicator
                                                                           Re-circulation




                                                                                  Extrapolated
                                                                                 primary decay
                                                                                     curve
                                                            Transit time

                                                      Time (seconds
                                    Injection
Cardiac output measurement: Thermodilution
Cardiac output with thermodilution
•   Major advantages of using the thermodilution technique include:
     – Non-toxic and can be repeated
     – Repeated measurements, performed correctly, show very low coefficients of variance
     – There is no recirculation
     – Shows good agreement with other indicator dilution methods including indocyanine
       green
     – Has now become the ‘gold standard’ for clinical measurement of cardiac output

•   The values obtained using thermodilution have some limitations:
     – Requires placement of a pulmonary artery catheter
     – Inaccurate in low-output states, tricuspid regurgitation, atrial and ventricular septal
       defects
     – There must be a variation of ~15% between the means of 2 measurements to interpret
       changes as significant
     – Cold saline may cause bradycardia, decreasing cardiac output. This limitation may be
       overcome by using saline at room temperature, compromising on precision.
       Measurements using this may vary if the patient is hypothermic
Pulmonary Artery Catheters
History
•   Bradley, along with Branthwaite, described pulmonary artery catheterisation,
    using a rigid thermistor tipped catheter placed under fluoroscopy in the
    pulmonary artery with a second catheter for injection in the right atrium1.
    Cardiac output was measured by thermodilution

•   Jeremy Swan conceived and developed
    a balloon tipped catheter that could be
    floated into the pulmonary artery. William
    Ganz added a thermistor to the tip to the
    device, enabling measurement of cardiac
    output thermodilution                                                      Jeremy Swan and William Ganz

•   Advances in technology have kept pace with the development of the
    catheter with addition of more lumens and the use of better materials
•   There are now dedicated monitors available with software to calculate a
    broad range of haemodynamic parameters
•   Incorporation of a heating element proximally has eliminated the need for
    injecting cold solutions

          1. Chatterjee K. The Swan-Ganz Catheters: Past, Present, and Future. Circulation. 2009;119:147-152
Principles
•   The pulmonary artery (PA) catheter allows measurement of the pulmonary
    pressures - systolic, diastolic and wedged
•   Enables measurement of cardiac output using the thermodilution method
•   Allows measurement of mixed venous oxygen saturation, from which
    oxygen consumption can be calculated
•   Together, these haemodynamic parameters provide a better understanding
    of preload, myocardial contractility and afterload
•   Information from these parameters may be used to guide fluid and
    vasoactive medication therapy
Principles
•   The balloon at the end of the PA catheter is
    inflated to wedge the catheter in the pulmonary
    artery and isolate it from upstream pressures             Pulmonary
                                                                artery

•   The pressure measured at the tip of the catheter
    reflects the pressure of a distal static column of
    blood which ends at the junction of the pulmonary
    vein with the left atrium
•   As the resistance of the pulmonary venous            RV                 LV
    system is negligible, this pressure reflects
    pulmonary venous and left atrial pressure – a
    measure of the left ventricular filling pressure
•   Sometimes the pulmonary artery diastolic
    pressure is used as a measure of left atrial                Pulmonary
                                                                  artery
    pressure instead of the wedged pressure                      wedged


•   This is acceptable in normal circumstances
    because at the end of diastole, the pulmonary
    artery pressure equilibrates with the downstream
    capillary and venous pressures                       RV                 LV
West’s Zones and pulmonary vascular pressures

•     The pressure within the capillaries are subject to the pressure in the alveoli
      and may influence the wedged pressure
•     West described the interaction of the alveoli and capillary pressures and
      grouped them into three zones: see figure

                                                                    •   Catheters placed in West’s zones 1
                                                                        and 2 are more likely to reflect
                                                                        alveolar (PA) than venous pressures
                                                                        (PV). Catheters placed in zone 3 are
                                                                        not influenced by the alveolar
                                                                        pressures and are more likely to
                                                                        reflect true venous pressures.
                                                                    •   In mechanically ventilated patients,
                                                                        PA is affected by PEEP and this may
                                                                        influence wedge pressures
©2006 by American Physiological Society
       Levitzky M G. Advances in Physiology Education 2006;30:5-8
Normal waveforms
           •    The pressure waveforms in the wedged position resembles a damped CVP
                waveform with ‘a’, ‘c’ and ‘v’ waves along with an ‘x’ and ‘y’ descents
           •    When the catheter is not wedged, it displays the pulmonary artery waveform
           •    There is a phase delay in the waveform compared to the ECG trace
           •    the peak of the a wave follows the P wave of the ECG by about 240
                milliseconds; the v wave follows the T wave of the ECG; the v wave is larger
                than the a wave



                 Right atrium   Right ventricle   Pulmonary artery           wedged


           20
pressure




                                                                         v
                                                                     a
           10

                                                                                      a   v
Normal Values
       Right atrial pressure            2 – 6 mm Hg

Right ventricular systolic pressure    15 – 25 mm Hg

Right ventricular diastolic pressure    0 – 8 mm Hg

Pulmonary artery systolic pressure     15 – 25 mm Hg

Pulmonary artery diastolic pressure    8 – 15 mm Hg

                                       10-20 mm Hg


Pulmonary artery wedged pressure       6 – 12 mm Hg
Inaccuracies with thermodilution
•   Catheter malposition:
                                                                                         Normal curve
     –   Thermistor wedged against a vessel wall

     –   Coiling of the catheter - thermistor close to the injectate port

•   Injection/ technique:
                                                                                 Uneven injection
     –   Incorrect recording of injectate temperature

     –   Slow injection - underestimates cardiac output (see figure)

     –   Large volume – underestimates; small volume over estimates
         cardiac output
                                                                                  Uneven injection
•   Cardiac:
     –   Cardiac dysrhythmias

     –   Extremes of cardiac output                                                        Prolonged
                                                                                           injection time
•   Others:
     –   Abnormal respiratory patterns – respirations cause
         fluctuations in cardiac output
                                                                             Artefacts in thermodilution
     –   Rapid infusion of IV fluids at the same time                       caused by improper injection
     –   Abnormal haematocrit – will affect K value
Semicontinuous thermodilution cardiac output
•   Thermodilution technique requires a injection of cold saline every time one needs
    to measure cardiac output
•   Limits its use as a continuous monitoring tool – particularly in conditions where
    haemodynamic changes can potentially occur rapidly
•   One way to overcome this challenge is to use heat instead of cold as an indicator
•   Achieved by using a heating filament on the proximal portion of the catheter with
    a thermistor distally to measure temperature
•   Heating filament generates low power pulses of heat; decay of heat is measured
    by the thermistor and cardiac output computed
•   Shows good correlation with the Fick and thermodilution techniques
•   Potential disadvantages include:
     – Inaccuracy when cold fluid is boloused through the same line
     – Incompatibility with Magnetic Resonance Imaging
     – Interference from electro-diathermy
     – Inability to detect sudden changes in cardiac output
Transpulmonary Thermodilution
Cardiac Output : Transpulmonary thermodilution


• Transpulmonary thermodilution technique is a variation of the traditional
  thermodilution technique
• Does not require a pulmonary artery catheter - a central venous catheter
  and an arterial line, preferably placed in the femoral vein is required
• Allows calculation of additional parameters: intra thoracic blood volume,
  global end diastolic volume and extra vascular lung water that may be
  help in the management of complex haemodynamic states
• Commercially available packages also uses the arterial pulse waveform
  (pulse contour analysis), referenced to the thermodilution curve, to give a
  beat by beat analysis of cardiac output
• This allows calculation of additional parameters such as pulse pressure
  variation and stroke volume variation that can aid in determining fluid
  status
Transpulmonary thermodilution - principles

•   Following central venous injection, cold saline / dye passes through the right
    heart, the lung, the left heart and the aorta as far as the detection site
•   The individual cardiac chambers and the lung, with the extravascular lung water,
    are the mixing chambers for the cold bolus
•   When a dye bound to albumin is used, it will mix only within the heart chambers
    and pulmonary circulation

                                                      Extra vascular lung water
             Injection of cold saline
                into a central vein

                                                                             Left heart
                                                      Pulmonary
                                                        blood             LA
                                    RA          RV     volume
                                                                                     LV

                                        Right heart
                                                        lungs

                                                                      Temperature changes
                                                                    recorded from a peripheral
                                                                              artery
Transpulmonary dilution - principles

•   If a cold injectate is used, the chambers of the heart and the lung become
    the mixing chamber. The volume of this mixing chamber is the intrathoracic
    thermal volume (ITTV)
•   The lungs, along with the extravascular lung water (EVLW) is the largest
    mixing chamber and its the volume is the pulmonary thermal volume (PTV)




                                    Pulmonary
               RA       RV            blood         LA     LV
                                     volume



                           Intrathoracic thermal volume

                             Pulmonary thermal volume
Transpulmonary dilution - principles
• Indicators bound to albumin (like indocyanine green) do not leave the
  circulation
• It mixes within the chambers of the heart and the pulmonary circulation
• The volume within the chambers of the heart and the pulmonary
  circulation is the intrathoracic blood volume (ITBV)
• The volume of blood within the lung is the pulmonary blood volume
  (PBV)


                                  Pulmonary
               RA      RV           blood           LA   LV
                                   volume



                            Intrathoracic blood volume

                            Pulmonary blood volume
Measurement of cardiac output
•   Cold saline / dye is injected into a central vein and the change in temperature or
    concentration of dye is measured distally in an artery
•   Cardiac output is calculated applying the Stewart – Hamilton formula
•   Mean transit time (MTt) is the mean time taken for the indicator to be detected
•   It is measured as the period between the start of the injection, and the point
    where the decay in the curve has dropped to 75% of its maximum
•   Down slope time (DSt) represents the mixing behaviour of the indicator within the
    pulmonary circulation - the largest mixing chamber
•   It is measured as the time taken for the curve to drop from 75% to 25% of its
    maximum

                                              Downslope time
                          Mean transit time




                                                      Time - seconds
                   injection
Measurement of cardiac output




                               Downslope time
           Mean transit time




                                          Time - seconds
    injection
Cardiac output using double indicators
   •      A second indicator bound to albumin, usually a dye such indocyanine green, in
          addition cold saline can be also used to to obtain a cardiac output
   •      This indicator remains confined to the intravascular compartment
   •      Using the MTt for both indicators, ITBV and ITTV can be computed:
                                                        ITBV = MTtdye x cardiac output
                                                        ITTV = MTtthermal x cardiac output
                                      Extravascular lung water (EVLW) = ITTV - ITBV
   •       EVLW represents the interstitial mixing compartment and has been shown to
          correlate with severity of illness1. It may be used as a guide to reduce positive
          fluid balance



                                                                                                  Downslope time
                                                                        Mean transit time



1. Sakka SG, Klein M, Reinhart K, Meier-Hellmann A.
Prognostic value of extravascular lung water in critically ill
patients. Chest. 2002; 122: 2080- 2086.

                                                                 injection            Time - seconds
Single indicator technique

•   The single indicator technique retains the same approach but uses only cold
    saline as the indicator
•   The advantage is convenience and decreased costs
•   It allows calculation of the same parameters as the double indicator
    technique and correlates well with both the double indicator technique and
    thermodilution technique
•   It also allows referencing cardiac output to the pulse waveform, and using
    proprietary algorithms, to provide a continuous cardiac output using pulse
    contour analysis
Cardiac output – derived values
 • Subtracting PTV from ITTV gives the volume of the four chambers of the
   heart: the global end diastolic volume (GEDV):
                                                   GEDV = ITTV - PTV
 • ITBV has been demonstrated to have a linear relationship with GEDV1:
                                                    ITBV = (GEDV x 1.25) – 28.4 ml

 •      Extravascular lung water can be calculated by subtracting ITTV from ITBV
                                                   EVLW = ITTV - ITBV

 •      Both GEDV and ITBV are indicators of preload. Being volumetric measurements,
        they are better than pressure based measurements
 •      They do not differentiate between right and left cardiac volumes and can be
        unreliable in right heart failure


                                                                                     Pulmonary
                                                                RA        RV           blood     LA   LV
1. Genahr A, McLuckie A. Transpulmonary                                               volume
thermodilution in the critically ill. British J Intensive
Care 2004: 6 - 10
Lithium dilution
• Lithium is used as the indicator: cardiac output is measured using
  the transpulmonary technique
• Lithium is non toxic in small doses and is measured using an ion
  sensitive electrode
• Shows good agreement with thermodilution techniques
• Advantages of using this technique:
    – Lithium can be injected peripherally
    – Blood can be sampled from a peripherally placed arterial line
• Limitations include:
    – Inability to use it in patients on Lithium therapy
    – The need to sample blood and its disposal
Stroke Volume / Pulse Pressure
           Variation
Principles
•   Both techniques rely on the principle that raised intrathoracic pressures in the
    inspiratory cycle of mechanical ventilators decreases venous return and
    therefore, cardiac output
•   This decrease is determined by calculating cardiac output (area under the arterial
    waveform in systole) during inspiration (decreases) and expiration (SVV)
•   The calculated cardiac output is usually          Maximum pulse pressure

    referenced to a previous measurement of                                    Minimum pulse pressure

    cardiac output using a transpulmonary
    dilution technique
•   The Flo trac™ system developed by
    Edwards Life Sciences uses a proprietary           inspiration             expiration
    algorithm to calculate cardiac output using
    pulse waveform analysis without the need
    for calibration
•   The difference between pulse pressures
    (PPV) or the systolic pressures (SPV) is
                                                          Stroke volume max
    also used as a surrogate marker, obviating                                     Stroke volume min

    the need for cardiac output measurement
Application

•   Pulse pressure variation and stroke volume variation is calculated as the
    difference between the maximum and minimum systolic pressure / stroke
    volume during the inspiratory cycle
•   The variation is usually expressed as a percentage of the mean pulse
    pressure/ stroke volume
•   Larger variations have been shown to correlate with fluid responsiveness,
    indicating that the heart is operating on the steep side of the Frank-Starling
    curve
•   In a meta-analysis1, variations between over 11 % appeared to correlate
    well with fluid responsiveness in mechanically ventilated patients
•   Robust data on spontaneously breathing patients is awaited



     1. Marik PE, Cavallazzi R, VasuT , Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness
    in mechanically ventilated patients: A systematic review of the literature. Crit Care Med 2009; 37: 2642 - 2647
Oesophageal Doppler
Oesophageal Doppler




                                                                         oesophagus                          aorta




                                                                            Doppler probe
ft     = transmitted frequency
c      = velocity of sound in tissue (1540 m/s)
                                                                                             f transmitted
cos θ = incident angle between the ultrasound beam and blood flow
direction (if the probe is parallel to the aorta, θ = 0 and cos θ = 1)


                                                                                               f received



                                                                                            Δf = ft - fr
Oesophageal Doppler
•   Multiplying the flow velocity by the cross-sectional area of the aorta and the
    ejection time gives an estimate of the stroke volume
•   Depending on the equipment used, the cross-sectional area of the aorta is
    estimated either from a normogram based on patient height, weight and age or,
    from direct estimates using the ultrasound
•   A correctional factor is applied to account for that proportion of stroke volume
    distributed to the coronary, carotid and subclavian arteries
•   Cardiac output is calculated by multiplying the stroke volume with the heart rate
•   In addition to cardiac output, the system also gives information about the
    preload and contractility of the heart by analysing the waveform
Oesophageal Doppler – waveform analysis
•    Waveform of velocity against time is shown below
•    The base of the triangle represents the systolic ejection time, also called the flow
     time. Because this is dependent on the heart rate, it is corrected using a
     modification of Bazett’s equation for correction of QT interval in the ECG
•    The corrected flow time (FTc) is the systolic ejection time corrected to one
     cardiac cycle per second
•    The area under the curve (shaded orange) is the stroke distance: the distance
     the column of blood has moved forward in the aorta during systole
•    Multiplying the stroke distance with the cross-sectional area of the aorta (direct
     measurement or normogram) gives the stroke volume. A correction is applied to
     account for blood flow to the branches of the aortic arch and coronaries
                       velocity




                                  peak velocity




                                                   AUC =
                                                   Stroke
                                                  distance


                                                  flow time   Time
Oesophageal Doppler – waveform analysis
•   In addition to measuring stroke volume and cardiac output, analysis of the
    Doppler waveform gives information about preload and contractility
•   FTc and peak velocity are the parameters used to indicate preload
•   A shortened FTc with a normal or moderately reduced peak velocity (waveform
    with narrow base and near normal amplitude) is seen in hypovolemic states
•   A shortened FTc with a markedly reduced peak velocity (waveform with a
    narrow base and decreased amplitude) reflects increased afterload
•   Reduced amplitude with a more shallow slope and rounded apex is seen with
    left ventricular failure
•    Peak velocity and mean acceleration reflect contractility
                      velocity




                                 peak velocity




                                                  AUC =
                                                  Stroke
                                                 distance


                                                 flow time   Time
Oesophageal Doppler – waveform analysis
                                                               SV=47 PV=63                      SV=64 PV=63.5
   •     The waveform is also used
         to see response to therapy
         such as fluid boluses or
         inotropes – see figure
                                                               Hypovolemia - ↓ FTc, stroke         Response to fluids
   •     While this does look                                  volume, normal peak velocity
         promising, there are limited
         number of studies1 that have                          SV=42 PV=42 FTc= 232             SV=72 PV=65 FTc= 345

         validated oesophageal
         Doppler as a monitoring tool
   •     Most of the studies have
                                                                Increased afterload - ↓ SV,     Response to vasodilators
         been done in a peri-                                          PV and FTc
         operative setting and its
         applicability to a general ICU                        SV=42 PV=45.5                    SV=59 PV=63
         setting is not clear


1. Ospina-Tascón GA, Cordioli RL Vincent J-L. What type of
monitoring has been shown to improve outcomes in acutely ill     Left ventricular failure - ↓
                                                                                                  Response to inotropes
patients? Intensive Care Med. 2008; 34: 800–820                  SV, PV and normal FTc
Oesophageal Doppler - limitations

•   The measurements assume that flow in the aorta is laminar. This may not
    always be the case and the values may not be accurate in:
     – Coarction of aorta or aortic stenosis
     – Aneurysms or dissection
     – When balloon pump counter pulsation is used

•   Deviations from the assumed angle of insonation (θ, the angle at which the
    Doppler signals are sent) will result in erroneous velocity values
•   The cross sectional area of the aorta is assumed to be a perfect circle. This
    may not be true, especially if there are aortic plaques or compliance is poor
•   Estimation of cardiac output assumes that a fixed proportion of blood goes
    to the coronaries and the branches of the aortic arch. This may not always
    be true. Shock states result in redistribution of cardiac output
Echocardiography
•   Provides accurate assessment of stroke volume and chamber pressures
•   Also enables estimation of myocardial function / dysfunction including
    diastolic dysfunction
•   Severity of valvular dysfunction can be assessed
•   Disadvantages:
     – Requires trained personnel
     – Equipment costs
     – Operator dependent
     – Body habitus, ventilation and position of the patient may preclude obtaining
       good images
     – Examination takes considerable time; real time imaging is not possible – can be
       overcome to a certain extent by limiting assessment to fixed protocols
Thoracic bioimpedance

•   Measures thoracic impedance to a high frequency, low magnitude current applied
    between electrodes in the neck and thorax
•    Bioimpedance of the thorax is influenced by
    the amount of fluid in the thorax
•   Fluid in the thorax (mainly blood in the
    venacavae and aorta) changes with cardiac
    cycle – causing changes in bioimpedance
                                                                                    Thoracic
•   Bioimpedance is inversely proportional to      High frequency
                                                   low magnitude
                                                                                 Bioimpedance
                                                                                   and ECG
    the amount of fluid in the thorax                  current
                                                                                    sensing

•   Stroke volume and cardiac output is
    computed after filtering electric noise
    caused by other fluid movement
•   Results using thoracic bioimpedance have been mixed - some studies have shown
    good correlation with invasive methods currently used, others have not
•   Acute lung injury, pleural effusions, pneumonia and obesity interfere with
    bioimpedance and result in inaccurate measurements
Indications for monitoring cardiac output

               Diagnostic                                    Therapeutic
•   Differential diagnosis of shock             •   Guide therapy in shock
     –   Cardiogenic                                 –   Fluid management
     –   Septic                                      –   Vasopressor / inotropes
     –   Hypovolemic                            •   Guide therapy peri-operatively
•   Evaluation of pulmonary oedema              •   Guide therapy in cardiac failure
     –   Cardiogenic from non cardiogenic
                                                •   Ventilator management
•   Evaluation of cardiac failure
                                                •   Multi - organ failure
     –   Tamponade
•   Evaluation of cardiac structures
     –   Valvular disease and intracardiac
         shunts
•   Others
     –   Shock in acute myocardial infarction
     –   Pulmonary hypertension
Clinical applications
•   Assessment of fluid status:
     – CVP/ Right atrial pressure – non specific, low pressures may be due to venous
       dilatation; high pressures may be reflective of pulmonary/ intra cardiac pathology
     – Right ventricular end diastolic volume – estimated by echo and thermodilution – more
       accurate measurement of preload
     – Global end diastolic volume – transpulmonary thermodilution
     – Decreased peak velocity and FTc on oesophageal Doppler
     – Increased pulse pressure / stroke volume variation in mechanically ventilated patients

•   Assessment of contractility:
     – Direct measurement of ejection fraction, systolic and diastolic function on echo
     – cardiac output / index; stroke volume / index using thermodilution, transpulmonary
       dilution or Doppler
     – Left or right ventricular stroke work index

•   Assessment of afterload:
     – Systemic vascular resistance / index
Evidence to support cardiac output monitoring
     •     Most of the evidence comes from the use of PA catheters
     •     Several reports published1 - suggesting harm from the use of PA catheters
     •     More recent randomized control studies and several meta analyses2 have
           not shown harm from their use
     •     These studies also failed to show a proven benefit (mortality or length of
           ICU/ hospital stay) of using PA catheters compared with standard care3
     •     Paucity of studies on the efficacy of other forms of cardiac output
           monitoring; however they are likely to reflect data from PA catheters
     •     Routine use of cardiac output monitoring may not be beneficial
     •     There may be may be some conditions where it may provide additional data
           aiding in diagnosis and therapy

1. Connors AF, Speroff T, Dawson NV, Thomas CT, Harrell FE, Wagner D et al.The Effectiveness of Right Heart Catheterization in the Initial
Care of Critically III Patients. JAMA. 1996; 276: 889-897
2. Shah MR, Hasselblad V, Stevenson LW, Binanay C, O’Connor CM, Sopko G et al. Impact of the Pulmonary Artery Catheter in Critically Ill
Patients. JAMA. 2005; 294: 1664-1670
3. Harvey S, Young D, BramptonW, Cooper A,Doig GS, SibbaldW, Rowan K. Pulmonary artery catheters for adult patients in intensive care.
Cochrane Database of Systematic Reviews 2006, Issue 3.
Monitoring Perfusion – Gastric
         Tonometry
Introduction

•   The goal of haemodynamic monitoring and support - ensure organs are
    adequately perfused
•   Assessment of the mental state and urine output are important clinical
    markers of organ perfusion in the brain and kidney
•   Often, these organs become involved in the disease process or cannot be
    assessed because of sedation / existing disease
•   Gastric tonometry is one technique used to assess regional perfusion
Gastric tonometry

•    Assesses splanchnic perfusion based on stomach’s mucosal pH by measuring
     gastric luminal PCO2 using a fluid filled balloon permeable to gases
•    The balloon is attached to a nasogastric tube and allowed to equilibrate with
     the luminal carbon dioxide. Luminal CO2 reflects intramucosal CO2
•    PCO2 is measured along with a simultaneous arterial blood gas and the luminal
     pH is calculated
•    Several limitations to using gastric tonometry routinely:
       – takes about 90 minutes for CO2 to equilibrate between the balloon and the lumen
       – Luminal CO2 may be affected by acid secretion (or lack of it if the patient is on acid
         suppressing agents) and feeding
       – No convincing evidence1 to support its routine use in the intensive care as several trials
         have failed to show benefit in using this form of monitoring




    1. Holley A, Lukin W, Paratz J, Hawkins T, Boots R Lipman J. Review article: Part two: Goal-directed resuscitation – Which goals?
    Perfusion targets. Emergency Medicine Australasia (2012) 24, 127–135
Monitoring Oxygen Consumption –
    Mixed Venous Saturation
Mixed venous saturation
•   Oxygen content of venous blood (CvO2) in the pulmonary artery (mixed venous
    blood) is used to estimate oxygen consumption (VO2), using the Fick equation
•   Indicative of the global metabolic requirements of the body
•   Relationship between oxygen delivery (DO2) and VO2 is linear – DO2 increases to
    keep up VO2 - up to a point (critical DO2). Beyond this, consumption exceeds
    supply, forces cells to revert to anaerobic metabolism
•   The point at which critical DO2 occurs is not the same for all organs; it also
    changes with disease states like sepsis
•   Low CvO2 reflects inadequate DO2 or increased consumption with greater oxygen
    extraction
•   Saturation of mixed venous blood (SvO2) can be used instead of CvO2
•   The normal SvO2 is > 75%
•   The central venous saturation (ScvO2) of the superior venacava is used instead of
    SvO2 ; there is some evidence that changes parallel that of SvO2
•   SvO2 is usually lower than ScvO2 by 2-5 %; this can reverse in shocked states
Clinical utility
•    ScvO2 is one of the clinical monitoring tools used to guide fluid resuscitation
     as part of the bundle in ‘early goal-directed therapy’ of septic shock1, 2
•    A ScvO2 < 70 % was used as a trigger to increase DO2 by increasing
     cardiac output or increasing haemoglobin once fluid resuscitation resulted in
     a target CVP of 8 -12
•    Using this bundle, Rivers demonstrated a decrease in mortality; we await
     results of more recent studies
•    Several studies have used SvO2 monitoring in the perioperative setting3.
     Most of the studies involving vascular patients do not demonstrate a
     mortality benefit or reduced length of ICU / hospital stay though there is
     some evidence to support its use in cardiac surgery


1. Holley A, Lukin W, Paratz J, Hawkins T, Boots R, Lipman J. Review article: Part one: Goal-directed resuscitation – Which
   goals? Haemodynamic targets. Emergency Medicine Australasia (2012) 24, 14–22
2. Rhodes A, Bennett DE. Early goal-directed therapy: An evidence-based review. Crit Care Med 2004; 32: S448 -450
3. Shepherd SJ, Pearse R. Role of Central and Mixed Venous Oxygen Saturation Measurement in Perioperative Care.
   Anesthesiology 2009; 111: 649 - 656
Summary
•   Advanced haemodynamic monitoring is useful in critically ill patients with
    haemodynamic instability
•   It provides more information about the circulatory state of the patient
    including
     –   Preload
     –   Myocardial contractility
     –   Afterload
     –   Perfusion and oxygen consumption
•   Cardiac output monitoring provides information about all these parameters
•   There are several modalities of monitoring to choose from
•   Decisions on which modality is best depends on the clinical situation,
    available resources and institutional preference
•   There is no clear evidence that cardiac output monitoring improves outcomes
    – but information from it can be used in clinical decision making
•   Mixed venous saturation is useful to monitor oxygen consumption and is one
    of the parameters used to titrate therapy in septic shock

Mais conteúdo relacionado

Mais procurados

Haemodynamic monitoring
Haemodynamic monitoringHaemodynamic monitoring
Haemodynamic monitoringguest5c708a
 
Anesthesia for children with Congenital Heart Disease
Anesthesia for children with Congenital Heart DiseaseAnesthesia for children with Congenital Heart Disease
Anesthesia for children with Congenital Heart Diseasecairo1957
 
Neuromuscular Monitoring
Neuromuscular MonitoringNeuromuscular Monitoring
Neuromuscular MonitoringMohtasib Madaoo
 
Anaesthesia for cardiopulmonary bypass surgery [autosaved]
Anaesthesia for cardiopulmonary bypass surgery [autosaved]Anaesthesia for cardiopulmonary bypass surgery [autosaved]
Anaesthesia for cardiopulmonary bypass surgery [autosaved]Nida fatima
 
Arterial line analysis
Arterial line analysisArterial line analysis
Arterial line analysissamirelansary
 
Hemodynamic parameters & fluid therapy Asim
Hemodynamic parameters &  fluid therapy AsimHemodynamic parameters &  fluid therapy Asim
Hemodynamic parameters & fluid therapy AsimMuhammad Asim Rana
 
Anaesthetic management of Abdominal aortic aneurysms
Anaesthetic management of Abdominal aortic aneurysmsAnaesthetic management of Abdominal aortic aneurysms
Anaesthetic management of Abdominal aortic aneurysmsAbhijit Nair
 
ANESTHESIA FOR TOF SURGERY
ANESTHESIA FOR TOF SURGERYANESTHESIA FOR TOF SURGERY
ANESTHESIA FOR TOF SURGERYRaju Jadhav
 
Anaesthetic management of mitral valvular heart disease
Anaesthetic management of mitral valvular heart diseaseAnaesthetic management of mitral valvular heart disease
Anaesthetic management of mitral valvular heart diseaseDhritiman Chakrabarti
 
INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING vikramnaidu2311
 
Echocardiography to assess Fluid Responsiveness in icu
Echocardiography to assess Fluid Responsiveness in icuEchocardiography to assess Fluid Responsiveness in icu
Echocardiography to assess Fluid Responsiveness in icuDr.Mahmoud Abbas
 
Inotropes and vasopressors
Inotropes and vasopressorsInotropes and vasopressors
Inotropes and vasopressorspankaj rana
 
Pulmonary artery catherisation
Pulmonary artery catherisationPulmonary artery catherisation
Pulmonary artery catherisationaratimohan
 
Anesthesia For Congenital Diaphragmatic Hernia
Anesthesia For Congenital Diaphragmatic HerniaAnesthesia For Congenital Diaphragmatic Hernia
Anesthesia For Congenital Diaphragmatic Herniakrishna dhakal
 
Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction Arun Shetty
 

Mais procurados (20)

Haemodynamic monitoring
Haemodynamic monitoringHaemodynamic monitoring
Haemodynamic monitoring
 
Anesthesia for children with Congenital Heart Disease
Anesthesia for children with Congenital Heart DiseaseAnesthesia for children with Congenital Heart Disease
Anesthesia for children with Congenital Heart Disease
 
Neuromuscular Monitoring
Neuromuscular MonitoringNeuromuscular Monitoring
Neuromuscular Monitoring
 
Anaesthesia for cardiopulmonary bypass surgery [autosaved]
Anaesthesia for cardiopulmonary bypass surgery [autosaved]Anaesthesia for cardiopulmonary bypass surgery [autosaved]
Anaesthesia for cardiopulmonary bypass surgery [autosaved]
 
Arterial line analysis
Arterial line analysisArterial line analysis
Arterial line analysis
 
Hemodynamic parameters & fluid therapy Asim
Hemodynamic parameters &  fluid therapy AsimHemodynamic parameters &  fluid therapy Asim
Hemodynamic parameters & fluid therapy Asim
 
Anaesthetic management of Abdominal aortic aneurysms
Anaesthetic management of Abdominal aortic aneurysmsAnaesthetic management of Abdominal aortic aneurysms
Anaesthetic management of Abdominal aortic aneurysms
 
ANESTHESIA FOR TOF SURGERY
ANESTHESIA FOR TOF SURGERYANESTHESIA FOR TOF SURGERY
ANESTHESIA FOR TOF SURGERY
 
Anaesthetic management of mitral valvular heart disease
Anaesthetic management of mitral valvular heart diseaseAnaesthetic management of mitral valvular heart disease
Anaesthetic management of mitral valvular heart disease
 
Ivc guided fluid management in the icu
Ivc guided fluid management in the icuIvc guided fluid management in the icu
Ivc guided fluid management in the icu
 
Inotropes
InotropesInotropes
Inotropes
 
Ischemic preconditioning
Ischemic preconditioningIschemic preconditioning
Ischemic preconditioning
 
INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING
 
Echocardiography to assess Fluid Responsiveness in icu
Echocardiography to assess Fluid Responsiveness in icuEchocardiography to assess Fluid Responsiveness in icu
Echocardiography to assess Fluid Responsiveness in icu
 
Hocm mpuh
Hocm mpuhHocm mpuh
Hocm mpuh
 
Inotropes and vasopressors
Inotropes and vasopressorsInotropes and vasopressors
Inotropes and vasopressors
 
Pulmonary artery catherisation
Pulmonary artery catherisationPulmonary artery catherisation
Pulmonary artery catherisation
 
Anesthesia For Congenital Diaphragmatic Hernia
Anesthesia For Congenital Diaphragmatic HerniaAnesthesia For Congenital Diaphragmatic Hernia
Anesthesia For Congenital Diaphragmatic Hernia
 
PiCCO Monitor
PiCCO Monitor PiCCO Monitor
PiCCO Monitor
 
Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction Intro to Hypoxic pulmonary vasoconstriction
Intro to Hypoxic pulmonary vasoconstriction
 

Semelhante a Advanced+Haemodynamic+Monitoring+and+support+-+part+1+vs+1+0.pptx

Cardiac output and venous return
Cardiac output and venous returnCardiac output and venous return
Cardiac output and venous returnErhard Rutashobya
 
Physiology of Blood Pressure - Regulation
Physiology of Blood Pressure - RegulationPhysiology of Blood Pressure - Regulation
Physiology of Blood Pressure - RegulationDeevenaHadassah
 
Blood pressure and Its Applied Physiology In Dentistry (Prosthodontics)
Blood pressure and Its Applied Physiology In Dentistry (Prosthodontics)Blood pressure and Its Applied Physiology In Dentistry (Prosthodontics)
Blood pressure and Its Applied Physiology In Dentistry (Prosthodontics)Self employed
 
Blood Pressure Measurement (2011)
Blood Pressure Measurement (2011)Blood Pressure Measurement (2011)
Blood Pressure Measurement (2011)Liesl Brown
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoringManisha Shakya
 
functional hemodynamic monitoring
functional hemodynamic monitoringfunctional hemodynamic monitoring
functional hemodynamic monitoringanaesthesiaESICMCH
 
Basic hemodynamic monitoring for nurses
Basic hemodynamic monitoring for nurses Basic hemodynamic monitoring for nurses
Basic hemodynamic monitoring for nurses Muhammad Asim Rana
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoringPratik Tantia
 
physiological monitoring of a surgical patient.pptx
physiological monitoring of a surgical patient.pptxphysiological monitoring of a surgical patient.pptx
physiological monitoring of a surgical patient.pptxkiogakimathi
 
Monitoring of critically ill
Monitoring of critically illMonitoring of critically ill
Monitoring of critically illStudyTricksByIshu
 
cardiac physiology
cardiac physiologycardiac physiology
cardiac physiologyHETA PATEL
 
Day 1 -RESERVE CONPENSATION FAILURE2wb.pptx
Day 1 -RESERVE CONPENSATION FAILURE2wb.pptxDay 1 -RESERVE CONPENSATION FAILURE2wb.pptx
Day 1 -RESERVE CONPENSATION FAILURE2wb.pptxMkindi Mkindi
 
Clinical approach to cardiac patients and
Clinical approach to cardiac patients andClinical approach to cardiac patients and
Clinical approach to cardiac patients andJemal64
 
Hypertension.pptx
Hypertension.pptxHypertension.pptx
Hypertension.pptxImtiyaz60
 
Hypertension and its pathophysiology.pptx
Hypertension and its pathophysiology.pptxHypertension and its pathophysiology.pptx
Hypertension and its pathophysiology.pptxImtiyaz60
 
CARDIOVASCULAR PHYSIOLOGY -2.rajesh munigial
CARDIOVASCULAR PHYSIOLOGY -2.rajesh munigialCARDIOVASCULAR PHYSIOLOGY -2.rajesh munigial
CARDIOVASCULAR PHYSIOLOGY -2.rajesh munigialRajesh Munigial
 

Semelhante a Advanced+Haemodynamic+Monitoring+and+support+-+part+1+vs+1+0.pptx (20)

Ibp,pac,cvp
Ibp,pac,cvpIbp,pac,cvp
Ibp,pac,cvp
 
Cardiac output and venous return
Cardiac output and venous returnCardiac output and venous return
Cardiac output and venous return
 
Physiology of Blood Pressure - Regulation
Physiology of Blood Pressure - RegulationPhysiology of Blood Pressure - Regulation
Physiology of Blood Pressure - Regulation
 
Blood pressure and Its Applied Physiology In Dentistry (Prosthodontics)
Blood pressure and Its Applied Physiology In Dentistry (Prosthodontics)Blood pressure and Its Applied Physiology In Dentistry (Prosthodontics)
Blood pressure and Its Applied Physiology In Dentistry (Prosthodontics)
 
Blood Pressure Measurement (2011)
Blood Pressure Measurement (2011)Blood Pressure Measurement (2011)
Blood Pressure Measurement (2011)
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoring
 
functional hemodynamic monitoring
functional hemodynamic monitoringfunctional hemodynamic monitoring
functional hemodynamic monitoring
 
Basic hemodynamic monitoring for nurses
Basic hemodynamic monitoring for nurses Basic hemodynamic monitoring for nurses
Basic hemodynamic monitoring for nurses
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoring
 
physiological monitoring of a surgical patient.pptx
physiological monitoring of a surgical patient.pptxphysiological monitoring of a surgical patient.pptx
physiological monitoring of a surgical patient.pptx
 
Monitoring of critically ill
Monitoring of critically illMonitoring of critically ill
Monitoring of critically ill
 
cardiac physiology
cardiac physiologycardiac physiology
cardiac physiology
 
Cardiogenic shock
Cardiogenic shockCardiogenic shock
Cardiogenic shock
 
Cardiogenic shock
Cardiogenic shockCardiogenic shock
Cardiogenic shock
 
Day 1 -RESERVE CONPENSATION FAILURE2wb.pptx
Day 1 -RESERVE CONPENSATION FAILURE2wb.pptxDay 1 -RESERVE CONPENSATION FAILURE2wb.pptx
Day 1 -RESERVE CONPENSATION FAILURE2wb.pptx
 
Blood Pressure
Blood PressureBlood Pressure
Blood Pressure
 
Clinical approach to cardiac patients and
Clinical approach to cardiac patients andClinical approach to cardiac patients and
Clinical approach to cardiac patients and
 
Hypertension.pptx
Hypertension.pptxHypertension.pptx
Hypertension.pptx
 
Hypertension and its pathophysiology.pptx
Hypertension and its pathophysiology.pptxHypertension and its pathophysiology.pptx
Hypertension and its pathophysiology.pptx
 
CARDIOVASCULAR PHYSIOLOGY -2.rajesh munigial
CARDIOVASCULAR PHYSIOLOGY -2.rajesh munigialCARDIOVASCULAR PHYSIOLOGY -2.rajesh munigial
CARDIOVASCULAR PHYSIOLOGY -2.rajesh munigial
 

Advanced+Haemodynamic+Monitoring+and+support+-+part+1+vs+1+0.pptx

  • 3. Contents 1. Introduction 6. Monitoring Perfusion – Gastric Tonometry 2. What is haemodynamic monitoring? 7. Monitoring Oxygen Consumption – Mixed 3. Advanced haemodynamic monitoring Venous Saturation 4. Central Venous Pressure Monitoring 8. Summary 5. Cardiac Output Measurement a) Introduction b) Derived parameters from cardiac output c) Cardiac Output using CO2 d) Cardiac output using dilution technique e) Pulmonary Artery Catheters f) Transpulmonary Thermodilution g) Lithium dilution h) Stroke Volume / Pulse Pressure Variation i) Oesophageal Doppler j) Echocardiography k) Thoracic bioimpedance l) Indications for monitoring cardiac output m) Clinical applications n) Evidence to support cardiac output monitoring
  • 4. Introduction • Monitoring is one of the fundamental aspects of intensive care • We monitor several physiological parameters regularly: – ECG, heart rate, SpO2, arterial blood pressure in real time – Other parameters such as temperature, urine output, respiration rate, GCS and blood sugars are monitored at frequent, but regular intervals • We monitor these parameters: – To alert us of deterioration or potential deterioration of patient’s state – To intervene when there are signs pointing to a deterioration – To assess response to treatment ‘Not everything that counts can be counted and not everything that can be counted counts’ . . . Albert Einstein
  • 5. What is haemodynamic monitoring? • Haemodynamic monitoring focuses on the circulatory system, its adequacy and end organ perfusion • Markers such as blood pressure and heart rate give an indirect measure of the circulatory system • Tachycardia and hypotension may thus indicate either a decreased preload or impaired myocardial contractility • They are not accurate: – The heart rate is influenced by many other factors – Tachycardia may be the result of pain, anxiety, drugs or arrhythmias and not necessarily a reduced preload or decreased cardiac output – Similarly, a low blood pressure may be secondary to drugs, postural effects or vasodilatation – A ‘normal’ blood pressure in a hypertensive patient may actually indicate a shocked state!
  • 6. What is haemodynamic monitoring? • Surrogate markers of organ perfusion include: – Skin perfusion - measured by capillary refill – Urine output – Mental state – usually assessed by the GCS • These may do not accurately measure perfusion: – Capillary refill may be affected by other factors including ambient temperature – Urine output may be vary in the setting of previous renal dysfunction – Mental alertness may not be assessable in the intubated patient; delirium may contribute to a decreased GCS • Other indirect markers of perfusion including serum lactate are not exclusive to poor perfusion and are elevated in other medical conditions as well
  • 7. Advanced haemodynamic monitoring • The ‘regular’ monitoring of the haemodynamic status provides sufficient information in most situations • There are situations where this may not sufficient to make decisions • Advanced haemodynamic monitoring involves monitoring the circulatory system and the components that affect it : – Preload – Myocardial contractility – Afterload • It also involves monitoring oxygen consumption and end organ perfusion
  • 9. Central Venous Pressure • The pressure measured in the superior and inferior venacava • Reflects right atrial pressure • Usually measured through a catheter inserted into the internal jugular or subclavian veins as well as through peripherally inserted central venous catheters • The pressure in the femoral veins also reflects central venous pressure, although increased abdominal pressure may modify this sufficiently to render it inaccurate • The right atrial pressure may be directly measured using the pulmonary artery catheter (see section on PA catheters) • CVP pressure monitoring is common - catheters in the central veins are inserted frequently for administration of vasoactive agents or hypertonic solutions and for long-term venous access
  • 10. History • First described in patients undergoing thoracic surgery • Hughes and Magovern1 described a fall in CVP with blood loss and a rise in CVP with transfusion in 1959 • CVP was initially used as a guide for fluid resuscitation and monitoring in cardiac and thoracic surgery and later on was used in major operative procedures • This practice then spilled over to the Intensive Care Units and now forms a common modality of haemodynamic monitoring in critically ill patients 1. Hughes RE, Magovern GJ. The Relationship Between Right Atrial Pressure and Blood Volume. Arch Surg. 1959;79: 238-243.
  • 11. Central Venous Pressure • Used as a surrogate marker for cardiac filling pressures - rise in CVP indicates better cardiac filling • Frank- Starling principle states that the force of cardiac contraction is proportional to the end-diastolic length of the fibre • This muscle length is proportional to the volume of blood at the end of diastole: the preload • Thus, preload is one of the determinants of cardiac output • Increased preload results in proportionately increased stroke volume • In clinical practice, it is not possible to measure the end diastolic volume of the right ventricle to determine the preload • End diastolic right ventricular pressure is reflected in the CVP which is therefore used as a surrogate marker of preload
  • 12. Central Venous Pressure • The relationship between ventricular end diastolic volume and pressure is not linear but curvilinear (see figure) Decreased • When the ventricle is operating in the compliance flat portion of the curve, there is little Left ventricular pressure change in pressure with end diastolic volume Normal End diastolic pressure • In the steep area of the curve, small changes of volume results in greater changes in pressure End diastolic pressure • Decreased ventricular compliance Increased compliance End diastolic pressure causes greater pressure changes with volume Left ventricular volume • Ventricular compliance is influenced by relaxation, geometry and the mechanical characteristics of the ventricle
  • 13. Factors affecting CVP • Normal CVP in a healthy adult: 0 -7 mm Hg • CVP of 0 may be normal in a healthy person - central veins are capacitance veins, potentially able to hold large volumes of blood without appreciable changes in pressure • Higher CVP may reflect poor ventricular compliance than good filling pressures • Changes in the transmural pressure - difference in the pressure between the cardiac chamber and the juxtacardiac pressure – also affects CVP • Juxtacardiac pressure is affected by pericardial and pleural pressures • In a ventilated patient, pleural pressures are affected by the compliance of the lung, pressures exerted by the ventilator, and PEEP • These pressure changes are complex and do not affect the CVP in a linear fashion. The practice of subtracting PEEP pressure to the CVP may not always reflect actual CVP • Pressures in the right atria may also be affected by changes in the ventricles or the atria: – Valvular abnormalities of the tricuspid and rarely, the pulmonary valves – Pulmonary hypertension
  • 14. Factors affecting CVP… • The pressures are affected by changes in the respiratory cycle: – Decreased intra-thoracic pressure at the beginning of spontaneous ventilation decreases the CVP – The onset of positive pressure ventilation causes an increase in CVP • The CVP is usually measured at the end of expiration when juxtacardiac pressures approach atmospheric pressure • Technical issues may also affect the actual CVP measurement, including: – Improper position of the catheter – abutment against a vessel wall – Thrombus in the catheter – Improper zeroing or placement of the transducer – Simultaneous rapid Infusion of fluids1 (rates > 50 ml/hour tends to increase the CVP) 1. Ho AM-H, Dion PW, Karmakar MK and Jenkins CR. Accuracy of central venous pressure monitoring during simultaneous continuous infusion through the same catheter. Anaesthesia, 2005; 60: 1027–1030
  • 15. CVP waveform • Waveform has a characteristic trace coinciding with changes in the cardiac cycle: – Prominent positive a wave - atrial contraction – Smaller, positive c wave – closure of tricuspid valve during isovolumetric contraction • Atrial relaxation decreases in atrial pressure (x descent) • Filling of the atrium causes pressures to rise again (v wave) • This is followed by the y descent, a negative waveform reflecting ventricular filling as the tricuspid valve opens
  • 16. Monitoring with CVP • Used both in resuscitation and as a guide to fluid therapy • Low CVP in a haemodynamically unstable patient reflects decreased preload and may be an indication for additional fluids • Its use to guide fluid resuscitation is recommended and in several guidelines including management of septic shock • Evidence to support this is however not convincing • In a systematic review, Marik1 found no correlation between CVP and blood volume • This review also found that CVP, or a change in CVP did not predict fluid responsiveness • The CVP should not be used in isolation and the pressure measured is a composite of a number of factors affecting the right atrial pressure • Trends in the change in the CVP and using the CVP in conjunction with other haemodynamic parameters helps overcome some of these barriers 1. Marik PE, Baram B, Vahid B. Does Central Venous Pressure Predict Fluid Responsiveness? Chest, 2008; 134: 172 - 178
  • 18. Introduction • Assessment of cardiac output is an important part of monitoring haemodynamically unstable patients • Until recently, the measurement of cardiac output was considered essential: – To characterize the type of shock – To monitor response to therapeutic interventions – To optimise factors that influence cardiac output – To optimise oxygen delivery
  • 20. Derived parameters from cardiac output Parameter Calculation Normal values Stroke Volume (index) 60 – 100 ml/ beat (33 - 47 ml/m2/beat) Systemic Vascular Resistance 1000 -1500 dyne s/ cm5 (index) (1970 - 2390 dyne s/cm5/m2) Pulmonary Vascular Resistance <250 dyne s/cm5 (index) (255 - 285 dyne s/cm5/m2) Left Ventricular Stroke Work 58 - 104 gm-m/beat (index) (50 - 62 gm-m/m2/beat) Right Ventricular Stroke Work 8 - 16 gm-m/beat (index) (5 - 10 gm-m/m2/beat) MAP= mean arterial pressure; RAP = right atrial pressure ( or CVP); PAWP = pulmonary artery wedge pressure; MPAP = mean pulmonary artery pressure (pulmonary systolic pressure +(2 x pulmonary diastolic pressure)/ 3); 80 and 0.10136 are the numbers required for conversion to the units of measurement
  • 21. Derived parameters… Parameter Calculation Normal values Coronary Artery Perfusion Diastolic BP-PAWP 60 - 80 mmHg pressure Oxygen Delivery (index) CaO2 x Cardiac Output (CI) 950 - 1150 ml/min (500 - 600 ml/min/m2) Oxygen Consumption (index) (C(a - v)O2) x CO (CI) 200 - 250 ml/min (120 - 160 ml/min/m2) CaO2= oxygen content of arterial blood ({1.34 x Hb/L x SaO2} + {0.0031 x PaO2}); (C(a - v)O2) = difference between arterial and venous oxygen content;
  • 22. Cardiac output measurement: Fick Principle
  • 26. Cardiac output measurement using CO2 • The potential benefits using this technique include its non-invasive nature and its ability to be repeated at relatively frequent intervals • It can be only done in intubated and ventilated patients; large gas leaks in non-invasive ventilation preclude its use • It has not shown consistent correlation with other techniques of measurement of CO • Its measurement is unreliable in several clinical situations: – Large intra pulmonary shunts – Where tidal volumes vary (spontaneously breathing patient) or when ventilator settings may alter dead space – Chronic lung disease, where ventilator- perfusion mismatch can occur – The postoperative cardiac patient where increased pulmonary dead space and shunts may occur – Acute lung injury (ALI) and ARDS where the heterogeneous nature of the injury causes a wide variation of gas exchange. End tidal CO2 may be a poor estimate of CaCO2
  • 27. Cardiac output using dilution technique
  • 28. Cardiac output measurement: Indicator-dilution Concentration of indicator Re-circulation Extrapolated primary decay curve Transit time Time (seconds Injection
  • 29. Cardiac output measurement: Thermodilution
  • 30. Cardiac output with thermodilution • Major advantages of using the thermodilution technique include: – Non-toxic and can be repeated – Repeated measurements, performed correctly, show very low coefficients of variance – There is no recirculation – Shows good agreement with other indicator dilution methods including indocyanine green – Has now become the ‘gold standard’ for clinical measurement of cardiac output • The values obtained using thermodilution have some limitations: – Requires placement of a pulmonary artery catheter – Inaccurate in low-output states, tricuspid regurgitation, atrial and ventricular septal defects – There must be a variation of ~15% between the means of 2 measurements to interpret changes as significant – Cold saline may cause bradycardia, decreasing cardiac output. This limitation may be overcome by using saline at room temperature, compromising on precision. Measurements using this may vary if the patient is hypothermic
  • 32. History • Bradley, along with Branthwaite, described pulmonary artery catheterisation, using a rigid thermistor tipped catheter placed under fluoroscopy in the pulmonary artery with a second catheter for injection in the right atrium1. Cardiac output was measured by thermodilution • Jeremy Swan conceived and developed a balloon tipped catheter that could be floated into the pulmonary artery. William Ganz added a thermistor to the tip to the device, enabling measurement of cardiac output thermodilution Jeremy Swan and William Ganz • Advances in technology have kept pace with the development of the catheter with addition of more lumens and the use of better materials • There are now dedicated monitors available with software to calculate a broad range of haemodynamic parameters • Incorporation of a heating element proximally has eliminated the need for injecting cold solutions 1. Chatterjee K. The Swan-Ganz Catheters: Past, Present, and Future. Circulation. 2009;119:147-152
  • 33. Principles • The pulmonary artery (PA) catheter allows measurement of the pulmonary pressures - systolic, diastolic and wedged • Enables measurement of cardiac output using the thermodilution method • Allows measurement of mixed venous oxygen saturation, from which oxygen consumption can be calculated • Together, these haemodynamic parameters provide a better understanding of preload, myocardial contractility and afterload • Information from these parameters may be used to guide fluid and vasoactive medication therapy
  • 34. Principles • The balloon at the end of the PA catheter is inflated to wedge the catheter in the pulmonary artery and isolate it from upstream pressures Pulmonary artery • The pressure measured at the tip of the catheter reflects the pressure of a distal static column of blood which ends at the junction of the pulmonary vein with the left atrium • As the resistance of the pulmonary venous RV LV system is negligible, this pressure reflects pulmonary venous and left atrial pressure – a measure of the left ventricular filling pressure • Sometimes the pulmonary artery diastolic pressure is used as a measure of left atrial Pulmonary artery pressure instead of the wedged pressure wedged • This is acceptable in normal circumstances because at the end of diastole, the pulmonary artery pressure equilibrates with the downstream capillary and venous pressures RV LV
  • 35. West’s Zones and pulmonary vascular pressures • The pressure within the capillaries are subject to the pressure in the alveoli and may influence the wedged pressure • West described the interaction of the alveoli and capillary pressures and grouped them into three zones: see figure • Catheters placed in West’s zones 1 and 2 are more likely to reflect alveolar (PA) than venous pressures (PV). Catheters placed in zone 3 are not influenced by the alveolar pressures and are more likely to reflect true venous pressures. • In mechanically ventilated patients, PA is affected by PEEP and this may influence wedge pressures ©2006 by American Physiological Society Levitzky M G. Advances in Physiology Education 2006;30:5-8
  • 36. Normal waveforms • The pressure waveforms in the wedged position resembles a damped CVP waveform with ‘a’, ‘c’ and ‘v’ waves along with an ‘x’ and ‘y’ descents • When the catheter is not wedged, it displays the pulmonary artery waveform • There is a phase delay in the waveform compared to the ECG trace • the peak of the a wave follows the P wave of the ECG by about 240 milliseconds; the v wave follows the T wave of the ECG; the v wave is larger than the a wave Right atrium Right ventricle Pulmonary artery wedged 20 pressure v a 10 a v
  • 37. Normal Values Right atrial pressure 2 – 6 mm Hg Right ventricular systolic pressure 15 – 25 mm Hg Right ventricular diastolic pressure 0 – 8 mm Hg Pulmonary artery systolic pressure 15 – 25 mm Hg Pulmonary artery diastolic pressure 8 – 15 mm Hg 10-20 mm Hg Pulmonary artery wedged pressure 6 – 12 mm Hg
  • 38. Inaccuracies with thermodilution • Catheter malposition: Normal curve – Thermistor wedged against a vessel wall – Coiling of the catheter - thermistor close to the injectate port • Injection/ technique: Uneven injection – Incorrect recording of injectate temperature – Slow injection - underestimates cardiac output (see figure) – Large volume – underestimates; small volume over estimates cardiac output Uneven injection • Cardiac: – Cardiac dysrhythmias – Extremes of cardiac output Prolonged injection time • Others: – Abnormal respiratory patterns – respirations cause fluctuations in cardiac output Artefacts in thermodilution – Rapid infusion of IV fluids at the same time caused by improper injection – Abnormal haematocrit – will affect K value
  • 39. Semicontinuous thermodilution cardiac output • Thermodilution technique requires a injection of cold saline every time one needs to measure cardiac output • Limits its use as a continuous monitoring tool – particularly in conditions where haemodynamic changes can potentially occur rapidly • One way to overcome this challenge is to use heat instead of cold as an indicator • Achieved by using a heating filament on the proximal portion of the catheter with a thermistor distally to measure temperature • Heating filament generates low power pulses of heat; decay of heat is measured by the thermistor and cardiac output computed • Shows good correlation with the Fick and thermodilution techniques • Potential disadvantages include: – Inaccuracy when cold fluid is boloused through the same line – Incompatibility with Magnetic Resonance Imaging – Interference from electro-diathermy – Inability to detect sudden changes in cardiac output
  • 41. Cardiac Output : Transpulmonary thermodilution • Transpulmonary thermodilution technique is a variation of the traditional thermodilution technique • Does not require a pulmonary artery catheter - a central venous catheter and an arterial line, preferably placed in the femoral vein is required • Allows calculation of additional parameters: intra thoracic blood volume, global end diastolic volume and extra vascular lung water that may be help in the management of complex haemodynamic states • Commercially available packages also uses the arterial pulse waveform (pulse contour analysis), referenced to the thermodilution curve, to give a beat by beat analysis of cardiac output • This allows calculation of additional parameters such as pulse pressure variation and stroke volume variation that can aid in determining fluid status
  • 42. Transpulmonary thermodilution - principles • Following central venous injection, cold saline / dye passes through the right heart, the lung, the left heart and the aorta as far as the detection site • The individual cardiac chambers and the lung, with the extravascular lung water, are the mixing chambers for the cold bolus • When a dye bound to albumin is used, it will mix only within the heart chambers and pulmonary circulation Extra vascular lung water Injection of cold saline into a central vein Left heart Pulmonary blood LA RA RV volume LV Right heart lungs Temperature changes recorded from a peripheral artery
  • 43. Transpulmonary dilution - principles • If a cold injectate is used, the chambers of the heart and the lung become the mixing chamber. The volume of this mixing chamber is the intrathoracic thermal volume (ITTV) • The lungs, along with the extravascular lung water (EVLW) is the largest mixing chamber and its the volume is the pulmonary thermal volume (PTV) Pulmonary RA RV blood LA LV volume Intrathoracic thermal volume Pulmonary thermal volume
  • 44. Transpulmonary dilution - principles • Indicators bound to albumin (like indocyanine green) do not leave the circulation • It mixes within the chambers of the heart and the pulmonary circulation • The volume within the chambers of the heart and the pulmonary circulation is the intrathoracic blood volume (ITBV) • The volume of blood within the lung is the pulmonary blood volume (PBV) Pulmonary RA RV blood LA LV volume Intrathoracic blood volume Pulmonary blood volume
  • 45. Measurement of cardiac output • Cold saline / dye is injected into a central vein and the change in temperature or concentration of dye is measured distally in an artery • Cardiac output is calculated applying the Stewart – Hamilton formula • Mean transit time (MTt) is the mean time taken for the indicator to be detected • It is measured as the period between the start of the injection, and the point where the decay in the curve has dropped to 75% of its maximum • Down slope time (DSt) represents the mixing behaviour of the indicator within the pulmonary circulation - the largest mixing chamber • It is measured as the time taken for the curve to drop from 75% to 25% of its maximum Downslope time Mean transit time Time - seconds injection
  • 46. Measurement of cardiac output Downslope time Mean transit time Time - seconds injection
  • 47. Cardiac output using double indicators • A second indicator bound to albumin, usually a dye such indocyanine green, in addition cold saline can be also used to to obtain a cardiac output • This indicator remains confined to the intravascular compartment • Using the MTt for both indicators, ITBV and ITTV can be computed: ITBV = MTtdye x cardiac output ITTV = MTtthermal x cardiac output Extravascular lung water (EVLW) = ITTV - ITBV • EVLW represents the interstitial mixing compartment and has been shown to correlate with severity of illness1. It may be used as a guide to reduce positive fluid balance Downslope time Mean transit time 1. Sakka SG, Klein M, Reinhart K, Meier-Hellmann A. Prognostic value of extravascular lung water in critically ill patients. Chest. 2002; 122: 2080- 2086. injection Time - seconds
  • 48. Single indicator technique • The single indicator technique retains the same approach but uses only cold saline as the indicator • The advantage is convenience and decreased costs • It allows calculation of the same parameters as the double indicator technique and correlates well with both the double indicator technique and thermodilution technique • It also allows referencing cardiac output to the pulse waveform, and using proprietary algorithms, to provide a continuous cardiac output using pulse contour analysis
  • 49. Cardiac output – derived values • Subtracting PTV from ITTV gives the volume of the four chambers of the heart: the global end diastolic volume (GEDV): GEDV = ITTV - PTV • ITBV has been demonstrated to have a linear relationship with GEDV1: ITBV = (GEDV x 1.25) – 28.4 ml • Extravascular lung water can be calculated by subtracting ITTV from ITBV EVLW = ITTV - ITBV • Both GEDV and ITBV are indicators of preload. Being volumetric measurements, they are better than pressure based measurements • They do not differentiate between right and left cardiac volumes and can be unreliable in right heart failure Pulmonary RA RV blood LA LV 1. Genahr A, McLuckie A. Transpulmonary volume thermodilution in the critically ill. British J Intensive Care 2004: 6 - 10
  • 50. Lithium dilution • Lithium is used as the indicator: cardiac output is measured using the transpulmonary technique • Lithium is non toxic in small doses and is measured using an ion sensitive electrode • Shows good agreement with thermodilution techniques • Advantages of using this technique: – Lithium can be injected peripherally – Blood can be sampled from a peripherally placed arterial line • Limitations include: – Inability to use it in patients on Lithium therapy – The need to sample blood and its disposal
  • 51. Stroke Volume / Pulse Pressure Variation
  • 52. Principles • Both techniques rely on the principle that raised intrathoracic pressures in the inspiratory cycle of mechanical ventilators decreases venous return and therefore, cardiac output • This decrease is determined by calculating cardiac output (area under the arterial waveform in systole) during inspiration (decreases) and expiration (SVV) • The calculated cardiac output is usually Maximum pulse pressure referenced to a previous measurement of Minimum pulse pressure cardiac output using a transpulmonary dilution technique • The Flo trac™ system developed by Edwards Life Sciences uses a proprietary inspiration expiration algorithm to calculate cardiac output using pulse waveform analysis without the need for calibration • The difference between pulse pressures (PPV) or the systolic pressures (SPV) is Stroke volume max also used as a surrogate marker, obviating Stroke volume min the need for cardiac output measurement
  • 53. Application • Pulse pressure variation and stroke volume variation is calculated as the difference between the maximum and minimum systolic pressure / stroke volume during the inspiratory cycle • The variation is usually expressed as a percentage of the mean pulse pressure/ stroke volume • Larger variations have been shown to correlate with fluid responsiveness, indicating that the heart is operating on the steep side of the Frank-Starling curve • In a meta-analysis1, variations between over 11 % appeared to correlate well with fluid responsiveness in mechanically ventilated patients • Robust data on spontaneously breathing patients is awaited 1. Marik PE, Cavallazzi R, VasuT , Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: A systematic review of the literature. Crit Care Med 2009; 37: 2642 - 2647
  • 55. Oesophageal Doppler oesophagus aorta Doppler probe ft = transmitted frequency c = velocity of sound in tissue (1540 m/s) f transmitted cos θ = incident angle between the ultrasound beam and blood flow direction (if the probe is parallel to the aorta, θ = 0 and cos θ = 1) f received Δf = ft - fr
  • 56. Oesophageal Doppler • Multiplying the flow velocity by the cross-sectional area of the aorta and the ejection time gives an estimate of the stroke volume • Depending on the equipment used, the cross-sectional area of the aorta is estimated either from a normogram based on patient height, weight and age or, from direct estimates using the ultrasound • A correctional factor is applied to account for that proportion of stroke volume distributed to the coronary, carotid and subclavian arteries • Cardiac output is calculated by multiplying the stroke volume with the heart rate • In addition to cardiac output, the system also gives information about the preload and contractility of the heart by analysing the waveform
  • 57. Oesophageal Doppler – waveform analysis • Waveform of velocity against time is shown below • The base of the triangle represents the systolic ejection time, also called the flow time. Because this is dependent on the heart rate, it is corrected using a modification of Bazett’s equation for correction of QT interval in the ECG • The corrected flow time (FTc) is the systolic ejection time corrected to one cardiac cycle per second • The area under the curve (shaded orange) is the stroke distance: the distance the column of blood has moved forward in the aorta during systole • Multiplying the stroke distance with the cross-sectional area of the aorta (direct measurement or normogram) gives the stroke volume. A correction is applied to account for blood flow to the branches of the aortic arch and coronaries velocity peak velocity AUC = Stroke distance flow time Time
  • 58. Oesophageal Doppler – waveform analysis • In addition to measuring stroke volume and cardiac output, analysis of the Doppler waveform gives information about preload and contractility • FTc and peak velocity are the parameters used to indicate preload • A shortened FTc with a normal or moderately reduced peak velocity (waveform with narrow base and near normal amplitude) is seen in hypovolemic states • A shortened FTc with a markedly reduced peak velocity (waveform with a narrow base and decreased amplitude) reflects increased afterload • Reduced amplitude with a more shallow slope and rounded apex is seen with left ventricular failure • Peak velocity and mean acceleration reflect contractility velocity peak velocity AUC = Stroke distance flow time Time
  • 59. Oesophageal Doppler – waveform analysis SV=47 PV=63 SV=64 PV=63.5 • The waveform is also used to see response to therapy such as fluid boluses or inotropes – see figure Hypovolemia - ↓ FTc, stroke Response to fluids • While this does look volume, normal peak velocity promising, there are limited number of studies1 that have SV=42 PV=42 FTc= 232 SV=72 PV=65 FTc= 345 validated oesophageal Doppler as a monitoring tool • Most of the studies have Increased afterload - ↓ SV, Response to vasodilators been done in a peri- PV and FTc operative setting and its applicability to a general ICU SV=42 PV=45.5 SV=59 PV=63 setting is not clear 1. Ospina-Tascón GA, Cordioli RL Vincent J-L. What type of monitoring has been shown to improve outcomes in acutely ill Left ventricular failure - ↓ Response to inotropes patients? Intensive Care Med. 2008; 34: 800–820 SV, PV and normal FTc
  • 60. Oesophageal Doppler - limitations • The measurements assume that flow in the aorta is laminar. This may not always be the case and the values may not be accurate in: – Coarction of aorta or aortic stenosis – Aneurysms or dissection – When balloon pump counter pulsation is used • Deviations from the assumed angle of insonation (θ, the angle at which the Doppler signals are sent) will result in erroneous velocity values • The cross sectional area of the aorta is assumed to be a perfect circle. This may not be true, especially if there are aortic plaques or compliance is poor • Estimation of cardiac output assumes that a fixed proportion of blood goes to the coronaries and the branches of the aortic arch. This may not always be true. Shock states result in redistribution of cardiac output
  • 61. Echocardiography • Provides accurate assessment of stroke volume and chamber pressures • Also enables estimation of myocardial function / dysfunction including diastolic dysfunction • Severity of valvular dysfunction can be assessed • Disadvantages: – Requires trained personnel – Equipment costs – Operator dependent – Body habitus, ventilation and position of the patient may preclude obtaining good images – Examination takes considerable time; real time imaging is not possible – can be overcome to a certain extent by limiting assessment to fixed protocols
  • 62. Thoracic bioimpedance • Measures thoracic impedance to a high frequency, low magnitude current applied between electrodes in the neck and thorax • Bioimpedance of the thorax is influenced by the amount of fluid in the thorax • Fluid in the thorax (mainly blood in the venacavae and aorta) changes with cardiac cycle – causing changes in bioimpedance Thoracic • Bioimpedance is inversely proportional to High frequency low magnitude Bioimpedance and ECG the amount of fluid in the thorax current sensing • Stroke volume and cardiac output is computed after filtering electric noise caused by other fluid movement • Results using thoracic bioimpedance have been mixed - some studies have shown good correlation with invasive methods currently used, others have not • Acute lung injury, pleural effusions, pneumonia and obesity interfere with bioimpedance and result in inaccurate measurements
  • 63. Indications for monitoring cardiac output Diagnostic Therapeutic • Differential diagnosis of shock • Guide therapy in shock – Cardiogenic – Fluid management – Septic – Vasopressor / inotropes – Hypovolemic • Guide therapy peri-operatively • Evaluation of pulmonary oedema • Guide therapy in cardiac failure – Cardiogenic from non cardiogenic • Ventilator management • Evaluation of cardiac failure • Multi - organ failure – Tamponade • Evaluation of cardiac structures – Valvular disease and intracardiac shunts • Others – Shock in acute myocardial infarction – Pulmonary hypertension
  • 64. Clinical applications • Assessment of fluid status: – CVP/ Right atrial pressure – non specific, low pressures may be due to venous dilatation; high pressures may be reflective of pulmonary/ intra cardiac pathology – Right ventricular end diastolic volume – estimated by echo and thermodilution – more accurate measurement of preload – Global end diastolic volume – transpulmonary thermodilution – Decreased peak velocity and FTc on oesophageal Doppler – Increased pulse pressure / stroke volume variation in mechanically ventilated patients • Assessment of contractility: – Direct measurement of ejection fraction, systolic and diastolic function on echo – cardiac output / index; stroke volume / index using thermodilution, transpulmonary dilution or Doppler – Left or right ventricular stroke work index • Assessment of afterload: – Systemic vascular resistance / index
  • 65. Evidence to support cardiac output monitoring • Most of the evidence comes from the use of PA catheters • Several reports published1 - suggesting harm from the use of PA catheters • More recent randomized control studies and several meta analyses2 have not shown harm from their use • These studies also failed to show a proven benefit (mortality or length of ICU/ hospital stay) of using PA catheters compared with standard care3 • Paucity of studies on the efficacy of other forms of cardiac output monitoring; however they are likely to reflect data from PA catheters • Routine use of cardiac output monitoring may not be beneficial • There may be may be some conditions where it may provide additional data aiding in diagnosis and therapy 1. Connors AF, Speroff T, Dawson NV, Thomas CT, Harrell FE, Wagner D et al.The Effectiveness of Right Heart Catheterization in the Initial Care of Critically III Patients. JAMA. 1996; 276: 889-897 2. Shah MR, Hasselblad V, Stevenson LW, Binanay C, O’Connor CM, Sopko G et al. Impact of the Pulmonary Artery Catheter in Critically Ill Patients. JAMA. 2005; 294: 1664-1670 3. Harvey S, Young D, BramptonW, Cooper A,Doig GS, SibbaldW, Rowan K. Pulmonary artery catheters for adult patients in intensive care. Cochrane Database of Systematic Reviews 2006, Issue 3.
  • 66. Monitoring Perfusion – Gastric Tonometry
  • 67. Introduction • The goal of haemodynamic monitoring and support - ensure organs are adequately perfused • Assessment of the mental state and urine output are important clinical markers of organ perfusion in the brain and kidney • Often, these organs become involved in the disease process or cannot be assessed because of sedation / existing disease • Gastric tonometry is one technique used to assess regional perfusion
  • 68. Gastric tonometry • Assesses splanchnic perfusion based on stomach’s mucosal pH by measuring gastric luminal PCO2 using a fluid filled balloon permeable to gases • The balloon is attached to a nasogastric tube and allowed to equilibrate with the luminal carbon dioxide. Luminal CO2 reflects intramucosal CO2 • PCO2 is measured along with a simultaneous arterial blood gas and the luminal pH is calculated • Several limitations to using gastric tonometry routinely: – takes about 90 minutes for CO2 to equilibrate between the balloon and the lumen – Luminal CO2 may be affected by acid secretion (or lack of it if the patient is on acid suppressing agents) and feeding – No convincing evidence1 to support its routine use in the intensive care as several trials have failed to show benefit in using this form of monitoring 1. Holley A, Lukin W, Paratz J, Hawkins T, Boots R Lipman J. Review article: Part two: Goal-directed resuscitation – Which goals? Perfusion targets. Emergency Medicine Australasia (2012) 24, 127–135
  • 69. Monitoring Oxygen Consumption – Mixed Venous Saturation
  • 70. Mixed venous saturation • Oxygen content of venous blood (CvO2) in the pulmonary artery (mixed venous blood) is used to estimate oxygen consumption (VO2), using the Fick equation • Indicative of the global metabolic requirements of the body • Relationship between oxygen delivery (DO2) and VO2 is linear – DO2 increases to keep up VO2 - up to a point (critical DO2). Beyond this, consumption exceeds supply, forces cells to revert to anaerobic metabolism • The point at which critical DO2 occurs is not the same for all organs; it also changes with disease states like sepsis • Low CvO2 reflects inadequate DO2 or increased consumption with greater oxygen extraction • Saturation of mixed venous blood (SvO2) can be used instead of CvO2 • The normal SvO2 is > 75% • The central venous saturation (ScvO2) of the superior venacava is used instead of SvO2 ; there is some evidence that changes parallel that of SvO2 • SvO2 is usually lower than ScvO2 by 2-5 %; this can reverse in shocked states
  • 71. Clinical utility • ScvO2 is one of the clinical monitoring tools used to guide fluid resuscitation as part of the bundle in ‘early goal-directed therapy’ of septic shock1, 2 • A ScvO2 < 70 % was used as a trigger to increase DO2 by increasing cardiac output or increasing haemoglobin once fluid resuscitation resulted in a target CVP of 8 -12 • Using this bundle, Rivers demonstrated a decrease in mortality; we await results of more recent studies • Several studies have used SvO2 monitoring in the perioperative setting3. Most of the studies involving vascular patients do not demonstrate a mortality benefit or reduced length of ICU / hospital stay though there is some evidence to support its use in cardiac surgery 1. Holley A, Lukin W, Paratz J, Hawkins T, Boots R, Lipman J. Review article: Part one: Goal-directed resuscitation – Which goals? Haemodynamic targets. Emergency Medicine Australasia (2012) 24, 14–22 2. Rhodes A, Bennett DE. Early goal-directed therapy: An evidence-based review. Crit Care Med 2004; 32: S448 -450 3. Shepherd SJ, Pearse R. Role of Central and Mixed Venous Oxygen Saturation Measurement in Perioperative Care. Anesthesiology 2009; 111: 649 - 656
  • 72. Summary • Advanced haemodynamic monitoring is useful in critically ill patients with haemodynamic instability • It provides more information about the circulatory state of the patient including – Preload – Myocardial contractility – Afterload – Perfusion and oxygen consumption • Cardiac output monitoring provides information about all these parameters • There are several modalities of monitoring to choose from • Decisions on which modality is best depends on the clinical situation, available resources and institutional preference • There is no clear evidence that cardiac output monitoring improves outcomes – but information from it can be used in clinical decision making • Mixed venous saturation is useful to monitor oxygen consumption and is one of the parameters used to titrate therapy in septic shock