Mrs Vandana Sharma
Assistant Professor
 Interference is a phenomena
 that leads to changes (either positive or negative) in intensity of
the analyte signal in spectroscopy.
 Interferences in atomic absorption spectroscopy fall into two
basic categories, namely, non-spectral and spectral.
Interferences in AAS
Spectral
Non-spectral
Matrix Ionization
Chemical
Spectral interferences are caused by presence of:
1.another atomic absorption line,
2.or a molecular absorbance band close to the spectral line of
element of interest.
 Most common spectral interferences are due to molecular
emissions from oxides of other elements in the sample.
 The main cause of background absorption is presence of
undissociated molecules of matrix that have broad band
absorption spectra and tiny solid particles, unvaporized solvent
droplets or molecular species in the flame which may scatter
light over a wide wavelength region. When this type of non-
specific adsorption overlaps the atomic absorption of the
analyte, background absorption occurs.
 The problem is overcome by:
1.measuring and subtracting the background absorption from the
total measured absorption to determine the true atomic
absorption.
 2.by choosing another line for the analyte. For example, a
vanadium line at 3082.11 Ao
interferes with aluminum line at
3082.15 Ao
.
This interference is avoided by employing the aluminum line at
3092.7 Ao
1. Matrix Interference
 When a sample is more viscous or has different surface tension
than the standard, it can result in differences in sample uptake
rate due to changes in nebulization efficiency.
 The problem is overcome by:
 Such interferences are minimized by matching as closely as
possible the matrix composition of standard and sample.
2. Chemical Interference
 If a sample contains a species which forms a thermally stable
compound with the analyte that is not completely decomposed by
the energy available in the flame then chemical interference exists.
 Refractory elements
such as Ti, W, Zr, Mo and Al may combine with oxygen to form --
---thermally stable oxides.
 The problem is overcome by:
 Analysis of such elements can be carried out at higher flame
temperatures using nitrous oxide – acetylene flame ----instead
of air-acetylene to provide higher dissociation energy.
or
 Alternatively, an excess of another element or compound can be
added e.g. Ca in presence of phosphate produces stable calcium
phosphate which ---reduces absorption due to Ca ion.
 If an excess of lanthanum is added it forms a thermally stable
compound with phosphate and ----calcium absorption is not
affected.
3. Ionization Interference
 Ionization interference is more common in hot flames.
 The dissociation process does not stop at formation of ground
state atoms.
 Excess energy of the flame can lead to--- ionization of ground
state atoms ---by loss of electrons thereby ---resulting in
depletion of ground state atoms.
 In cooler flames such interference is encountered with easily
ionized elements such as alkali metals and alkaline earths.
 The problem is overcome by:
 Ionization interference is eliminated by adding an excess of an
element which is easily ionized ----thereby creating a large
number of free electrons in the flame ----and suppressing
ionization of the analyte.
 Salts of such elements as K, Rb and Cs are commonly used as
ionization suppressants.
Ca (g)
Analyte
Ca+
(g) + e-
K(g)
ionization suppressant
K+
(g) + e-
Short note on
Instrumentation (AAS)
and
short answers
 Atomic absorption spectroscopy is a precise analytical
technique.
 It is used for the quantification of metals present in the analyte.
 It is one of the best techniques to detect the sample at trace
levels such as part per billion (ppb).
 Atomic absorption spectroscopy was developed by Alan Walsh in
the 1950s.
 This technique involves the absorption of radiant energy (usually
ultraviolet and visible) by a bunch of free atoms.
 The sample used is in a gaseous state.
 When an atom absorbs light of a specific wavelength, it results in
the excitation of electrons from the ground state to the excited
state.
 The energy absorbed per mole is fixed and can be used for
qualitative and quantitative analysis.
PRINCIPLE
 Atoms of the analyte have a fixed number of electrons.
 If the light of a specific wavelength is passed through a flame containing
that atom, electrons present in different energy levels, known as orbitals,
absorb a certain wavelength and excite to higher energy levels.
 The extent of absorption ά the number of ground-state atoms
in the flame.
 Only for information- The ground state is more stable than the excited
state. The electrons spontaneously return back to the ground state. It emits
the same amount of radiant energy. This process is called fluorescence.
Fluorescence is used in atomic emission spectroscopy.
Instrumentation
 The basic components of atomic absorption are:
1. Light source
2. Chopper
3. Atomizer
4. Burners
5. flames
6. Monochromators
7. Detectors
8. Amplifier
9. Readout devices
Collimating mirror
Focusing mirror
gratings
Entrance slit
Exit slit
1. LIGHT SOURCE
 A hollow cathode lamp is used as the light source.
 As the name suggests, it consists of a cylindrical hollow cathode
made up of the same element under consideration.
 The anode is made up of tungsten.
 Both anode and cathode are enclosed in a glass tube with a
quartz window.
 The glass envelope is filled with inert gases (Ar, Ne) under
reduced pressure (1-5 torr).
 When a high voltage (300-500 V) is applied across the
electrodes, it ionizes noble gases at the anode (+).
 The ions produced as a result are accelerated toward the
cathode (-). When these fast-moving ions strike the surface of
the cathode, they cause some of the metal atoms to sputter.
 Thus, the electrons of vaporized metals are excited to higher
energy levels. It is due to continuous collision with higher energy
gas ions.
2. CHOPPER
 A chopper is a rotating wheel between a hollow cathode lamp and a
flame.
 It removes large droplets and allows uniform size droplets to enter
the flame.
 It breaks the steady light from the lamp into pulsating light. This
gives the pulsating current in the photocell that is amplified and
recorded without any interference.
3. Sample introduction system
 This is used to transfer samples to the atomizer.
 It should produce no interference, must be independent of sample
type, and universal for all atomizers.
 Pneumatic nebulizers are commonly used for introducing solutions.
 A jet of compressed air, nebulization gas, aspirates and nebulize the
solution when the sample is sucked through the capillary tube.
 Types of nebulizers
 Concentric nebulizers
 Angular or crossflow nebulizers
4. Atomizers
 The analyte is broken into the free atoms with the atomizer.
 For this purpose, a very high temperature is required.
 Only a portion of the sample solution is introduced into the
burner chamber through the nebulizer.
 In the bottom of the chamber, large particles are accumulated.
These are then drained off.
5. Atomization sources
 Burners
 These are the basic types of burners used in atomic absorption
spectroscopy.
 Total consumption burners
 In the total consumption burner, the sample solution, the fuel
(usually acetylene), and the oxidizing gases enter through
separate passages at an opening in the base of the flame.
 Here gases create a partial vacuum and force the sample to
move up through capillary action.
 Premix chamber burners
 In premix chamber burners, the sample is aspirated through a
nebulizer and sprayed as a fine aerosol into the mixing chamber.
 Here the sample is mixed with fuel and oxidant gases and carried
to the burner head where atomization takes place.
6. Flames
 Flame is generally produced by using a mixture of gases. There
are two types of flames commonly used in atomic absorption
spectroscopy.
 Air acetylene flame
 Nitrous oxide acetylene flame
7. Monochromators
 The function of a monochromator is to select the specific
wavelength of light that is being absorbed by the sample.
 Types of monochromators
 Prism
 Diffraction gratings
 Dispersion from the grating is more uniform than dispersion from
the prism.
 So, the grating can maintain a higher resolution over a longer
range of wavelengths.
8. Detectors
 Photomultiplier tubes are the most commonly used detectors in
atomic absorption spectroscopy.
 These are used to convert electromagnetic waves into electric
currents.
 They are made up of photoemissive cathodes and dynodes.
 These dynodes provide electron multiplications.
 Photodiode array detectors
 These detectors convert the transmitted light signals into an
electric pulse.
9. AMPLIFIERS
 The output from detectors is fed to amplifiers which amplify the
current several times.
10. Readout devices
 Readout systems used in atomic absorption spectroscopy are
meters, chart recorders, digital display meters, etc. hard copies
can make on chart recorders, printers, or plotters.
WORKING PROCESS
 Generally, the samples for atomic absorption spectroscopy
analysis should be in solution form.
The steps involved in the analysis are -
 Conversion of sample into solution form.
 Preparation of blank solution.
 Preparation of a series of standard solutions.
 Setup instrumental parameters
 Measure the absorbance of standard solutions
 A plot calibration curve (concentration vs absorbance)
 Measure the absorbance of the sample solution.
 The unknown concentration of the element is calculated from
the calibration curve.
Calibration
 The instrument must be calibrated before analysis for better
results.
 There must be five standard solutions containing a known
concentration of metals to be analyzed for calibration of the
instrument and of the same concentration as the sample
solutions.
 Calibrating solution is generally prepared as a sample of salt
dissolved in water or dilute acid.
 The calibration curve is plotted between concentration and
absorbance.
 ABSORPTION SPECTRUM
 It consists of a series of well-defined extremely narrow lines
arising from the electronic transition of the outermost electrons.
 Quantitative analysis in AAS
 Quantification of elements is done by plotting calibration curves.
 At least four standards and a blanks solution are required to plot
a linear calibration curve of absorbance versus concentration.
 In all absorbance measurements, the reading must be taken after
the instrument zero has been adjusted against a blank which
may be distilled water or a solution of similar composition to the
test solution.
 Usually, standards are measured in the order of increasing
concentration.
 After making the measurement with one solution, distilled water
is aspirated into the flame to remove all the traces of the
solution before proceeding to the next solution.
 At least two to three separate readings should be taken and then
averaged.
 Safety measures
 The atomic absorption room must be properly ventilated and
provided with an adequate exhaust system with airtight joints on
discharged side.
 Gas cylinders must be fastened securely outside the room well
away from heat or ignition sources.
 Pipes that carry the gas from cylinders must be securely fixed to
avoid damage.
 Make periodic checks of leaks by applying soap solution to joints
and seals.
 Never run acetylene gas at higher pressure (maximum 15 psi).
 Tubes should be made of brass are other substances which
contain copper of less than 65%.
 Never analyze the flame or hollow cathode lamp directly. (uses
protective eyewear).
 When the atomic absorption spectrophotometer is turned off,
close the gas supply valve tightly, and bleed the gas line to the
atmosphere through exhaust systems.
What is the principle of atomic absorption spectroscopy?
 When monochromatic light is passed through a flame containing the atom
under consideration, then the part of the light will be absorbed by the
electron of the atoms atom that causes the excitation. The extent of
absorption is proportional to the number of ground-state atoms.
What is the significance of the atomic absorption technique?
 Atomic absorption is a technique used for the detection of metals in
solution at trace. It gives relative analysis such that every element is
detected one by one.
Which detector is used in atomic absorption spectroscopy?
 There are two types of detectors used in atomic absorption spectroscopy i.e,
film and photomultipliers.
How accurate is atomic absorption spectroscopy?
 Atomic absorption spectroscopy is a very accurate technique for the
detection of metals. It can detect the elements up to 0.1 µg/ mL. Since It is
100 times more precise than UV visible spectroscopy.
Which is the correct order instrumentation of AAS?
Order of instrumentation of AAS:
1. Light source
2. Chopper
3. Atomizer
4. Monochromators
5. Detectors
6. Amplifier
7. Readout devices
What can atomic absorption spectroscopy be used for?
 Atomic absorption spectroscopy is used for the detection of metal at trace and ultra-
trace levels.
 Atomic absorption spectroscopy is used for geochemical and metallurgical analysis.
 It is used for the detection of metals in biological fluids i.e blood, and urine.
 Atomic absorption is used for pharmaceutical analysis.
After absorption in Atomic Absorption Spectrometry AAS do electrons re-emit the
absorbed light?
 When an electron absorbs energy the goes to a higher energy state. They remain in the
excited state for a while and come back to the ground state by emitting the absorbed
energy. Because the ground state is more stable than the excited state.
Why are nonmetals not detected by atomic absorption spectroscopy?
 Metals are detected through atomic absorption by their ionization whereas, nonmetals
do not ionize. That’s why nonmetals are not detected by atomic absorption
spectroscopy.
What is atomic spectroscopy?
 Spectroscopy is the interaction of light with matter. When we studied this interaction at
the atomic level then this is known as atomic spectroscopy.
What is the difference between a molecular absorption spectrum and an atomic
absorption spectrum?
 A molecular spectrum is a study of molecules. It shows light absorb by the whole
molecule. The molecular spectrum has a broad peak. while atomic absorption has a
sharp distinct peak. It shows the light absorb by the atom.
What is the difference between fluorescence spectroscopy and absorbance spectroscopy?
 Fluorescence spectroscopy deals with the emission of light when an electron returns to
the ground state. Whereas, absorbance spectroscopy deals with the absorbance of light
by electrons and their excitation.
What is molecular fluorescence spectroscopy?
 In molecular fluorescence spectroscopy, light energy is used to excite an
electron of certain molecules. When this electron returns to the ground
state they emit light. This emitted light is directly proportional to the
number of atoms present in the analyte.
What elements can AAS detect?
 Atomic absorption spectroscopy is used to detect metal atoms.
What are the disadvantages of Atomic Absorption Spectroscopy?
 Disadvantages of AAS
 Atomic absorption is used only for metals detection.
 It cannot be used qualitative analysis
 Very high temperature is required for analysis
 AAS is a destructive technique
 It is an expensive procedure
 Toxic and flammable gases are required for analysis
Why are atomic absorption lines very narrow?
 The atomic absorption spectrum is very narrow because AAS is a highly
specific method of analysis. Its spectrum displays the characteristic
wavelength absorbed by metal atoms.
What is the detection limit of AAS?
 The detection limit of atomic absorption spectroscopy is ranging from ppm
to ppb or 0.1 µg/ mL to 1.0 µg/ mL.
What do AAS and FES stand for?
 AAS stands for Atomic absorption spectroscopy whereas FES stands for
flame emission spectroscopy

Mais conteúdo relacionado

Mais procurados

Instrumentation IR Spectroscopy: Detectors
Instrumentation IR Spectroscopy: DetectorsInstrumentation IR Spectroscopy: Detectors
Instrumentation IR Spectroscopy: DetectorsVrushali Tambe
 
Flame Emission Spectroscopy (FES) Instrumentation
Flame Emission Spectroscopy (FES) InstrumentationFlame Emission Spectroscopy (FES) Instrumentation
Flame Emission Spectroscopy (FES) InstrumentationAnkitVasoya5
 
Flame photometry, principle, interferences, instrumentation, applications.pptx
Flame photometry, principle, interferences, instrumentation, applications.pptxFlame photometry, principle, interferences, instrumentation, applications.pptx
Flame photometry, principle, interferences, instrumentation, applications.pptxVandana Devesh Sharma
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopyTukai Kulkarni
 
Fluorimetry phosphorimetry
Fluorimetry phosphorimetryFluorimetry phosphorimetry
Fluorimetry phosphorimetryEnosh Gummadi
 
Flame Photometry, by Dr. Umesh Kumar Sharma & Shyma M. S.
Flame Photometry, by Dr. Umesh Kumar Sharma  & Shyma M. S.Flame Photometry, by Dr. Umesh Kumar Sharma  & Shyma M. S.
Flame Photometry, by Dr. Umesh Kumar Sharma & Shyma M. S.Dr. UMESH KUMAR SHARMA
 
Atomic emission spectroscopy PPT
Atomic emission spectroscopy PPTAtomic emission spectroscopy PPT
Atomic emission spectroscopy PPTHumnaMehmood
 
Flame emission spectroscopy
Flame emission spectroscopyFlame emission spectroscopy
Flame emission spectroscopyMehulJain143
 
UV-Visible spectroscopy
UV-Visible spectroscopyUV-Visible spectroscopy
UV-Visible spectroscopyMehulJain143
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopymanishasoral
 
Uv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentationUv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentationAnusreeAnu11
 
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Sajjad Ullah
 
Deviations from Beers law
Deviations from Beers lawDeviations from Beers law
Deviations from Beers lawVrushali Tambe
 
Factors affecting fluorescence intensity(pharmaceutical analysis)
Factors affecting fluorescence intensity(pharmaceutical analysis)Factors affecting fluorescence intensity(pharmaceutical analysis)
Factors affecting fluorescence intensity(pharmaceutical analysis)MdIrfanUddin2
 
Infrared spectroscopy
Infrared spectroscopy Infrared spectroscopy
Infrared spectroscopy Prachi Pathak
 

Mais procurados (20)

Instrumentation IR Spectroscopy: Detectors
Instrumentation IR Spectroscopy: DetectorsInstrumentation IR Spectroscopy: Detectors
Instrumentation IR Spectroscopy: Detectors
 
Flame Emission Spectroscopy (FES) Instrumentation
Flame Emission Spectroscopy (FES) InstrumentationFlame Emission Spectroscopy (FES) Instrumentation
Flame Emission Spectroscopy (FES) Instrumentation
 
Flame photometry, principle, interferences, instrumentation, applications.pptx
Flame photometry, principle, interferences, instrumentation, applications.pptxFlame photometry, principle, interferences, instrumentation, applications.pptx
Flame photometry, principle, interferences, instrumentation, applications.pptx
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Fluorimetry phosphorimetry
Fluorimetry phosphorimetryFluorimetry phosphorimetry
Fluorimetry phosphorimetry
 
Flame photometry
Flame photometryFlame photometry
Flame photometry
 
Flame Photometry, by Dr. Umesh Kumar Sharma & Shyma M. S.
Flame Photometry, by Dr. Umesh Kumar Sharma  & Shyma M. S.Flame Photometry, by Dr. Umesh Kumar Sharma  & Shyma M. S.
Flame Photometry, by Dr. Umesh Kumar Sharma & Shyma M. S.
 
Atomic emission spectroscopy PPT
Atomic emission spectroscopy PPTAtomic emission spectroscopy PPT
Atomic emission spectroscopy PPT
 
Flame emission spectroscopy
Flame emission spectroscopyFlame emission spectroscopy
Flame emission spectroscopy
 
UV-Visible spectroscopy
UV-Visible spectroscopyUV-Visible spectroscopy
UV-Visible spectroscopy
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Flourimetry factors
Flourimetry factorsFlourimetry factors
Flourimetry factors
 
Uv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentationUv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentation
 
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)
 
Quenching In Fluorimetry.
Quenching In Fluorimetry.Quenching In Fluorimetry.
Quenching In Fluorimetry.
 
Deviations from Beers law
Deviations from Beers lawDeviations from Beers law
Deviations from Beers law
 
Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopy
 
Factors affecting fluorescence intensity(pharmaceutical analysis)
Factors affecting fluorescence intensity(pharmaceutical analysis)Factors affecting fluorescence intensity(pharmaceutical analysis)
Factors affecting fluorescence intensity(pharmaceutical analysis)
 
Infrared spectroscopy
Infrared spectroscopy Infrared spectroscopy
Infrared spectroscopy
 
Flame phtometry
Flame phtometryFlame phtometry
Flame phtometry
 

Semelhante a AAS interference.pptx

Atomic absorption spectroscopy lec
Atomic absorption spectroscopy lecAtomic absorption spectroscopy lec
Atomic absorption spectroscopy lecZainab&Sons
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy Zainab&Sons
 
Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry  Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry Sailee Gurav
 
Atomic Absorption Spectroscopy (www.Redicals.com)
Atomic Absorption Spectroscopy (www.Redicals.com)Atomic Absorption Spectroscopy (www.Redicals.com)
Atomic Absorption Spectroscopy (www.Redicals.com)Goa App
 
Atomic absorption & flame emissionn.pptx
Atomic absorption & flame emissionn.pptxAtomic absorption & flame emissionn.pptx
Atomic absorption & flame emissionn.pptxDr. Vijaya Barge
 
Flame emission Spectroscopy SlideShare mineeta mahra
Flame emission Spectroscopy SlideShare mineeta mahraFlame emission Spectroscopy SlideShare mineeta mahra
Flame emission Spectroscopy SlideShare mineeta mahraMineeta Mahra
 
Flame emission & atomic absorption spectroscopy
Flame emission & atomic absorption spectroscopyFlame emission & atomic absorption spectroscopy
Flame emission & atomic absorption spectroscopyHimal Barakoti
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometry Atomic absorption spectrophotometry
Atomic absorption spectrophotometry ZeravanAli
 
Atomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptxAtomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptxPriyankashah645680
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometrySurabhi Pal
 
Atomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.pptAtomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.pptZORAIZ HAIDER
 
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Parijat Suryawanshi
 
Atomic absorption-spectrometry
Atomic absorption-spectrometryAtomic absorption-spectrometry
Atomic absorption-spectrometryMuhammad Amien
 
Atomic absorption-spectrometry
Atomic absorption-spectrometryAtomic absorption-spectrometry
Atomic absorption-spectrometryMUsama10
 
Lecture 3. Atomic-absorption spectrophotometry.pptx
Lecture 3. Atomic-absorption spectrophotometry.pptxLecture 3. Atomic-absorption spectrophotometry.pptx
Lecture 3. Atomic-absorption spectrophotometry.pptxlyazzatfreedom
 
Atomic Absorption Spectrometer.pptx
Atomic Absorption Spectrometer.pptxAtomic Absorption Spectrometer.pptx
Atomic Absorption Spectrometer.pptxdrpvczback
 

Semelhante a AAS interference.pptx (20)

Instru booklet
Instru bookletInstru booklet
Instru booklet
 
Atomic absorption spectroscopy lec
Atomic absorption spectroscopy lecAtomic absorption spectroscopy lec
Atomic absorption spectroscopy lec
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy
 
Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry  Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry
 
Atomic Absorption Spectroscopy (www.Redicals.com)
Atomic Absorption Spectroscopy (www.Redicals.com)Atomic Absorption Spectroscopy (www.Redicals.com)
Atomic Absorption Spectroscopy (www.Redicals.com)
 
Atomic absorption & flame emissionn.pptx
Atomic absorption & flame emissionn.pptxAtomic absorption & flame emissionn.pptx
Atomic absorption & flame emissionn.pptx
 
Flame emission Spectroscopy SlideShare mineeta mahra
Flame emission Spectroscopy SlideShare mineeta mahraFlame emission Spectroscopy SlideShare mineeta mahra
Flame emission Spectroscopy SlideShare mineeta mahra
 
Flame emission & atomic absorption spectroscopy
Flame emission & atomic absorption spectroscopyFlame emission & atomic absorption spectroscopy
Flame emission & atomic absorption spectroscopy
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometry Atomic absorption spectrophotometry
Atomic absorption spectrophotometry
 
Atomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptxAtomic absorption spectroscopy (AAS).pptx
Atomic absorption spectroscopy (AAS).pptx
 
Atomic absorption spectroscopy.pptx
Atomic absorption spectroscopy.pptxAtomic absorption spectroscopy.pptx
Atomic absorption spectroscopy.pptx
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometry
 
Atomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.pptAtomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.ppt
 
Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)Atomic Absorption Spectroscopy (AAS)
Atomic Absorption Spectroscopy (AAS)
 
Atomic spectroscopy
Atomic spectroscopy Atomic spectroscopy
Atomic spectroscopy
 
Atomic absorption-spectrometry
Atomic absorption-spectrometryAtomic absorption-spectrometry
Atomic absorption-spectrometry
 
Atomic absorption-spectrometry
Atomic absorption-spectrometryAtomic absorption-spectrometry
Atomic absorption-spectrometry
 
Lecture 3. Atomic-absorption spectrophotometry.pptx
Lecture 3. Atomic-absorption spectrophotometry.pptxLecture 3. Atomic-absorption spectrophotometry.pptx
Lecture 3. Atomic-absorption spectrophotometry.pptx
 
Atomic absorption spectroscopy
Atomic absorption spectroscopyAtomic absorption spectroscopy
Atomic absorption spectroscopy
 
Atomic Absorption Spectrometer.pptx
Atomic Absorption Spectrometer.pptxAtomic Absorption Spectrometer.pptx
Atomic Absorption Spectrometer.pptx
 

Mais de Vandana Devesh Sharma

HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...Vandana Devesh Sharma
 
Gel chromatography, Introduction, Theory, Instrumentation, Applications .pptx
Gel chromatography, Introduction, Theory, Instrumentation, Applications  .pptxGel chromatography, Introduction, Theory, Instrumentation, Applications  .pptx
Gel chromatography, Introduction, Theory, Instrumentation, Applications .pptxVandana Devesh Sharma
 
Fluorimetry part 2, Instrumentation, single beam and double beam, light sourc...
Fluorimetry part 2, Instrumentation, single beam and double beam, light sourc...Fluorimetry part 2, Instrumentation, single beam and double beam, light sourc...
Fluorimetry part 2, Instrumentation, single beam and double beam, light sourc...Vandana Devesh Sharma
 
Fluorimetry, principle, Concept of singlet,doublet,and triplet electronic sta...
Fluorimetry, principle, Concept of singlet,doublet,and triplet electronic sta...Fluorimetry, principle, Concept of singlet,doublet,and triplet electronic sta...
Fluorimetry, principle, Concept of singlet,doublet,and triplet electronic sta...Vandana Devesh Sharma
 
Potentiometry, Electrochemical cell, construction and working of indicator an...
Potentiometry, Electrochemical cell, construction and working of indicator an...Potentiometry, Electrochemical cell, construction and working of indicator an...
Potentiometry, Electrochemical cell, construction and working of indicator an...Vandana Devesh Sharma
 
Polarography Principle, instrumentation, Applications.pptx
Polarography Principle, instrumentation, Applications.pptxPolarography Principle, instrumentation, Applications.pptx
Polarography Principle, instrumentation, Applications.pptxVandana Devesh Sharma
 

Mais de Vandana Devesh Sharma (7)

HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
 
Gel chromatography, Introduction, Theory, Instrumentation, Applications .pptx
Gel chromatography, Introduction, Theory, Instrumentation, Applications  .pptxGel chromatography, Introduction, Theory, Instrumentation, Applications  .pptx
Gel chromatography, Introduction, Theory, Instrumentation, Applications .pptx
 
Fluorimetry part 2, Instrumentation, single beam and double beam, light sourc...
Fluorimetry part 2, Instrumentation, single beam and double beam, light sourc...Fluorimetry part 2, Instrumentation, single beam and double beam, light sourc...
Fluorimetry part 2, Instrumentation, single beam and double beam, light sourc...
 
Fluorimetry, principle, Concept of singlet,doublet,and triplet electronic sta...
Fluorimetry, principle, Concept of singlet,doublet,and triplet electronic sta...Fluorimetry, principle, Concept of singlet,doublet,and triplet electronic sta...
Fluorimetry, principle, Concept of singlet,doublet,and triplet electronic sta...
 
Potentiometry, Electrochemical cell, construction and working of indicator an...
Potentiometry, Electrochemical cell, construction and working of indicator an...Potentiometry, Electrochemical cell, construction and working of indicator an...
Potentiometry, Electrochemical cell, construction and working of indicator an...
 
Polarography Principle, instrumentation, Applications.pptx
Polarography Principle, instrumentation, Applications.pptxPolarography Principle, instrumentation, Applications.pptx
Polarography Principle, instrumentation, Applications.pptx
 
Condectometry.pptx
Condectometry.pptxCondectometry.pptx
Condectometry.pptx
 

Último

Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfchloefrazer622
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphThiyagu K
 
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...anjaliyadav012327
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdfQucHHunhnh
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Krashi Coaching
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxSayali Powar
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Sapana Sha
 
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...fonyou31
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsTechSoup
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajanpragatimahajan3
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdfSoniaTolstoy
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...Sapna Thakur
 
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 💞 Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 💞 Full Nigh...Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 💞 Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 💞 Full Nigh...Pooja Nehwal
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxGaneshChakor2
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3JemimahLaneBuaron
 
9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room servicediscovermytutordmt
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAssociation for Project Management
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdfQucHHunhnh
 

Último (20)

Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdf
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot Graph
 
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
JAPAN: ORGANISATION OF PMDA, PHARMACEUTICAL LAWS & REGULATIONS, TYPES OF REGI...
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
 
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
Ecosystem Interactions Class Discussion Presentation in Blue Green Lined Styl...
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
social pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajansocial pharmacy d-pharm 1st year by Pragati K. Mahajan
social pharmacy d-pharm 1st year by Pragati K. Mahajan
 
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdfBASLIQ CURRENT LOOKBOOK  LOOKBOOK(1) (1).pdf
BASLIQ CURRENT LOOKBOOK LOOKBOOK(1) (1).pdf
 
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
BAG TECHNIQUE Bag technique-a tool making use of public health bag through wh...
 
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 💞 Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 💞 Full Nigh...Russian Call Girls in Andheri Airport Mumbai WhatsApp  9167673311 💞 Full Nigh...
Russian Call Girls in Andheri Airport Mumbai WhatsApp 9167673311 💞 Full Nigh...
 
Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1
 
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptxINDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptx
 
Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3Q4-W6-Restating Informational Text Grade 3
Q4-W6-Restating Informational Text Grade 3
 
9548086042 for call girls in Indira Nagar with room service
9548086042  for call girls in Indira Nagar  with room service9548086042  for call girls in Indira Nagar  with room service
9548086042 for call girls in Indira Nagar with room service
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across Sectors
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 

AAS interference.pptx

  • 2.  Interference is a phenomena  that leads to changes (either positive or negative) in intensity of the analyte signal in spectroscopy.  Interferences in atomic absorption spectroscopy fall into two basic categories, namely, non-spectral and spectral. Interferences in AAS Spectral Non-spectral Matrix Ionization Chemical
  • 3. Spectral interferences are caused by presence of: 1.another atomic absorption line, 2.or a molecular absorbance band close to the spectral line of element of interest.  Most common spectral interferences are due to molecular emissions from oxides of other elements in the sample.  The main cause of background absorption is presence of undissociated molecules of matrix that have broad band absorption spectra and tiny solid particles, unvaporized solvent droplets or molecular species in the flame which may scatter light over a wide wavelength region. When this type of non- specific adsorption overlaps the atomic absorption of the analyte, background absorption occurs.  The problem is overcome by: 1.measuring and subtracting the background absorption from the total measured absorption to determine the true atomic absorption.
  • 4.  2.by choosing another line for the analyte. For example, a vanadium line at 3082.11 Ao interferes with aluminum line at 3082.15 Ao . This interference is avoided by employing the aluminum line at 3092.7 Ao 1. Matrix Interference  When a sample is more viscous or has different surface tension than the standard, it can result in differences in sample uptake rate due to changes in nebulization efficiency.  The problem is overcome by:  Such interferences are minimized by matching as closely as possible the matrix composition of standard and sample. 2. Chemical Interference  If a sample contains a species which forms a thermally stable compound with the analyte that is not completely decomposed by the energy available in the flame then chemical interference exists.
  • 5.  Refractory elements such as Ti, W, Zr, Mo and Al may combine with oxygen to form -- ---thermally stable oxides.  The problem is overcome by:  Analysis of such elements can be carried out at higher flame temperatures using nitrous oxide – acetylene flame ----instead of air-acetylene to provide higher dissociation energy. or  Alternatively, an excess of another element or compound can be added e.g. Ca in presence of phosphate produces stable calcium phosphate which ---reduces absorption due to Ca ion.  If an excess of lanthanum is added it forms a thermally stable compound with phosphate and ----calcium absorption is not affected.
  • 6. 3. Ionization Interference  Ionization interference is more common in hot flames.  The dissociation process does not stop at formation of ground state atoms.  Excess energy of the flame can lead to--- ionization of ground state atoms ---by loss of electrons thereby ---resulting in depletion of ground state atoms.  In cooler flames such interference is encountered with easily ionized elements such as alkali metals and alkaline earths.  The problem is overcome by:  Ionization interference is eliminated by adding an excess of an element which is easily ionized ----thereby creating a large number of free electrons in the flame ----and suppressing ionization of the analyte.  Salts of such elements as K, Rb and Cs are commonly used as ionization suppressants.
  • 7. Ca (g) Analyte Ca+ (g) + e- K(g) ionization suppressant K+ (g) + e-
  • 8. Short note on Instrumentation (AAS) and short answers
  • 9.  Atomic absorption spectroscopy is a precise analytical technique.  It is used for the quantification of metals present in the analyte.  It is one of the best techniques to detect the sample at trace levels such as part per billion (ppb).  Atomic absorption spectroscopy was developed by Alan Walsh in the 1950s.  This technique involves the absorption of radiant energy (usually ultraviolet and visible) by a bunch of free atoms.  The sample used is in a gaseous state.  When an atom absorbs light of a specific wavelength, it results in the excitation of electrons from the ground state to the excited state.  The energy absorbed per mole is fixed and can be used for qualitative and quantitative analysis.
  • 10. PRINCIPLE  Atoms of the analyte have a fixed number of electrons.  If the light of a specific wavelength is passed through a flame containing that atom, electrons present in different energy levels, known as orbitals, absorb a certain wavelength and excite to higher energy levels.  The extent of absorption ά the number of ground-state atoms in the flame.  Only for information- The ground state is more stable than the excited state. The electrons spontaneously return back to the ground state. It emits the same amount of radiant energy. This process is called fluorescence. Fluorescence is used in atomic emission spectroscopy. Instrumentation  The basic components of atomic absorption are: 1. Light source 2. Chopper 3. Atomizer 4. Burners 5. flames 6. Monochromators 7. Detectors 8. Amplifier 9. Readout devices
  • 12. 1. LIGHT SOURCE  A hollow cathode lamp is used as the light source.  As the name suggests, it consists of a cylindrical hollow cathode made up of the same element under consideration.  The anode is made up of tungsten.  Both anode and cathode are enclosed in a glass tube with a quartz window.  The glass envelope is filled with inert gases (Ar, Ne) under reduced pressure (1-5 torr).  When a high voltage (300-500 V) is applied across the electrodes, it ionizes noble gases at the anode (+).  The ions produced as a result are accelerated toward the cathode (-). When these fast-moving ions strike the surface of the cathode, they cause some of the metal atoms to sputter.  Thus, the electrons of vaporized metals are excited to higher energy levels. It is due to continuous collision with higher energy gas ions.
  • 13. 2. CHOPPER  A chopper is a rotating wheel between a hollow cathode lamp and a flame.  It removes large droplets and allows uniform size droplets to enter the flame.  It breaks the steady light from the lamp into pulsating light. This gives the pulsating current in the photocell that is amplified and recorded without any interference. 3. Sample introduction system  This is used to transfer samples to the atomizer.  It should produce no interference, must be independent of sample type, and universal for all atomizers.  Pneumatic nebulizers are commonly used for introducing solutions.  A jet of compressed air, nebulization gas, aspirates and nebulize the solution when the sample is sucked through the capillary tube.  Types of nebulizers  Concentric nebulizers  Angular or crossflow nebulizers
  • 14. 4. Atomizers  The analyte is broken into the free atoms with the atomizer.  For this purpose, a very high temperature is required.  Only a portion of the sample solution is introduced into the burner chamber through the nebulizer.  In the bottom of the chamber, large particles are accumulated. These are then drained off. 5. Atomization sources  Burners  These are the basic types of burners used in atomic absorption spectroscopy.  Total consumption burners  In the total consumption burner, the sample solution, the fuel (usually acetylene), and the oxidizing gases enter through separate passages at an opening in the base of the flame.  Here gases create a partial vacuum and force the sample to move up through capillary action.
  • 15.  Premix chamber burners  In premix chamber burners, the sample is aspirated through a nebulizer and sprayed as a fine aerosol into the mixing chamber.  Here the sample is mixed with fuel and oxidant gases and carried to the burner head where atomization takes place. 6. Flames  Flame is generally produced by using a mixture of gases. There are two types of flames commonly used in atomic absorption spectroscopy.  Air acetylene flame  Nitrous oxide acetylene flame 7. Monochromators  The function of a monochromator is to select the specific wavelength of light that is being absorbed by the sample.  Types of monochromators  Prism  Diffraction gratings
  • 16.  Dispersion from the grating is more uniform than dispersion from the prism.  So, the grating can maintain a higher resolution over a longer range of wavelengths. 8. Detectors  Photomultiplier tubes are the most commonly used detectors in atomic absorption spectroscopy.  These are used to convert electromagnetic waves into electric currents.  They are made up of photoemissive cathodes and dynodes.  These dynodes provide electron multiplications.  Photodiode array detectors  These detectors convert the transmitted light signals into an electric pulse. 9. AMPLIFIERS  The output from detectors is fed to amplifiers which amplify the current several times.
  • 17. 10. Readout devices  Readout systems used in atomic absorption spectroscopy are meters, chart recorders, digital display meters, etc. hard copies can make on chart recorders, printers, or plotters. WORKING PROCESS  Generally, the samples for atomic absorption spectroscopy analysis should be in solution form. The steps involved in the analysis are -  Conversion of sample into solution form.  Preparation of blank solution.  Preparation of a series of standard solutions.  Setup instrumental parameters  Measure the absorbance of standard solutions  A plot calibration curve (concentration vs absorbance)  Measure the absorbance of the sample solution.  The unknown concentration of the element is calculated from the calibration curve.
  • 18. Calibration  The instrument must be calibrated before analysis for better results.  There must be five standard solutions containing a known concentration of metals to be analyzed for calibration of the instrument and of the same concentration as the sample solutions.  Calibrating solution is generally prepared as a sample of salt dissolved in water or dilute acid.  The calibration curve is plotted between concentration and absorbance.  ABSORPTION SPECTRUM  It consists of a series of well-defined extremely narrow lines arising from the electronic transition of the outermost electrons.
  • 19.  Quantitative analysis in AAS  Quantification of elements is done by plotting calibration curves.  At least four standards and a blanks solution are required to plot a linear calibration curve of absorbance versus concentration.  In all absorbance measurements, the reading must be taken after the instrument zero has been adjusted against a blank which may be distilled water or a solution of similar composition to the test solution.  Usually, standards are measured in the order of increasing concentration.  After making the measurement with one solution, distilled water is aspirated into the flame to remove all the traces of the solution before proceeding to the next solution.  At least two to three separate readings should be taken and then averaged.
  • 20.
  • 21.  Safety measures  The atomic absorption room must be properly ventilated and provided with an adequate exhaust system with airtight joints on discharged side.  Gas cylinders must be fastened securely outside the room well away from heat or ignition sources.  Pipes that carry the gas from cylinders must be securely fixed to avoid damage.  Make periodic checks of leaks by applying soap solution to joints and seals.  Never run acetylene gas at higher pressure (maximum 15 psi).  Tubes should be made of brass are other substances which contain copper of less than 65%.  Never analyze the flame or hollow cathode lamp directly. (uses protective eyewear).  When the atomic absorption spectrophotometer is turned off, close the gas supply valve tightly, and bleed the gas line to the atmosphere through exhaust systems.
  • 22. What is the principle of atomic absorption spectroscopy?  When monochromatic light is passed through a flame containing the atom under consideration, then the part of the light will be absorbed by the electron of the atoms atom that causes the excitation. The extent of absorption is proportional to the number of ground-state atoms. What is the significance of the atomic absorption technique?  Atomic absorption is a technique used for the detection of metals in solution at trace. It gives relative analysis such that every element is detected one by one. Which detector is used in atomic absorption spectroscopy?  There are two types of detectors used in atomic absorption spectroscopy i.e, film and photomultipliers. How accurate is atomic absorption spectroscopy?  Atomic absorption spectroscopy is a very accurate technique for the detection of metals. It can detect the elements up to 0.1 µg/ mL. Since It is 100 times more precise than UV visible spectroscopy. Which is the correct order instrumentation of AAS? Order of instrumentation of AAS: 1. Light source 2. Chopper 3. Atomizer 4. Monochromators 5. Detectors 6. Amplifier 7. Readout devices
  • 23. What can atomic absorption spectroscopy be used for?  Atomic absorption spectroscopy is used for the detection of metal at trace and ultra- trace levels.  Atomic absorption spectroscopy is used for geochemical and metallurgical analysis.  It is used for the detection of metals in biological fluids i.e blood, and urine.  Atomic absorption is used for pharmaceutical analysis. After absorption in Atomic Absorption Spectrometry AAS do electrons re-emit the absorbed light?  When an electron absorbs energy the goes to a higher energy state. They remain in the excited state for a while and come back to the ground state by emitting the absorbed energy. Because the ground state is more stable than the excited state. Why are nonmetals not detected by atomic absorption spectroscopy?  Metals are detected through atomic absorption by their ionization whereas, nonmetals do not ionize. That’s why nonmetals are not detected by atomic absorption spectroscopy. What is atomic spectroscopy?  Spectroscopy is the interaction of light with matter. When we studied this interaction at the atomic level then this is known as atomic spectroscopy. What is the difference between a molecular absorption spectrum and an atomic absorption spectrum?  A molecular spectrum is a study of molecules. It shows light absorb by the whole molecule. The molecular spectrum has a broad peak. while atomic absorption has a sharp distinct peak. It shows the light absorb by the atom. What is the difference between fluorescence spectroscopy and absorbance spectroscopy?  Fluorescence spectroscopy deals with the emission of light when an electron returns to the ground state. Whereas, absorbance spectroscopy deals with the absorbance of light by electrons and their excitation.
  • 24. What is molecular fluorescence spectroscopy?  In molecular fluorescence spectroscopy, light energy is used to excite an electron of certain molecules. When this electron returns to the ground state they emit light. This emitted light is directly proportional to the number of atoms present in the analyte. What elements can AAS detect?  Atomic absorption spectroscopy is used to detect metal atoms. What are the disadvantages of Atomic Absorption Spectroscopy?  Disadvantages of AAS  Atomic absorption is used only for metals detection.  It cannot be used qualitative analysis  Very high temperature is required for analysis  AAS is a destructive technique  It is an expensive procedure  Toxic and flammable gases are required for analysis Why are atomic absorption lines very narrow?  The atomic absorption spectrum is very narrow because AAS is a highly specific method of analysis. Its spectrum displays the characteristic wavelength absorbed by metal atoms. What is the detection limit of AAS?  The detection limit of atomic absorption spectroscopy is ranging from ppm to ppb or 0.1 µg/ mL to 1.0 µg/ mL. What do AAS and FES stand for?  AAS stands for Atomic absorption spectroscopy whereas FES stands for flame emission spectroscopy