TECHNICAL SPECIFICATIONS SERVO-s
1. DATA SHEET
VENTILATION
SERVO-s
HIGHLIGHTS
® Simple to learn, operate and maintain
SERVO-s offers unique SERVO ventilation
capabilities for adult and pediatric patients Monitors lung mechanics
Features a range of standard and optional Supports invasive as well as Non invasive
ventilation modes ventilation (NIV)
2.
3. | Critical Care | SERVO-s | 3 |
KEY FEATURES
SERVO-s
For adult and pediatric patients Simple to operate and learn
- Large color display with clear graphic presentation
A ventilator system designed to operate in a variety of hospital - Intuitive user interface
ventilatory care settings - Select between touch screen, direct access knobs or dial
Standard ventilation modes: - Fast and simple start up procedure e.g. automated pre-use
check with user configurable default ventilator settings
Controlled ventilation: - PC - Suction Support
- VC - Previous Mode functionality
- High-resolution pressure, flow and volume waveforms
Supported ventilation: - PS/CPAP
Simple to maintain and clean
Combined ventilation: - SIMV (PC) + PS - One-piece interchangeable part for cleaning
- SIMV (VC) + PS - Extended maintenance interval (5000 hours)
- Maintenance information given via graphical user interface
Optional ventilation modes:
Reliable
Controlled ventilation: - PRVC - Battery backup (60 minutes)
- Apnea backup
Combined ventilation: - SIMV (PRVC) + PS - Direct access knobs for vital settings
(included in PRVC) - Enhanced alarm system
- Bi-Vent
Uninterrupted bedside quality ventilatory treatment during
Optional functions: intra-hospital transportation with Mobile Cart
- Lung mechanics parameters
- Alarm output connector for external alarm Can be mounted on both ceiling pendant and shelf
Supports invasive ventilation and the option Non invasive Non-consumable and maintenance-free ultrasonic O2 sensor,
ventilation (NIV) with leakage compensation alternative to consumable O2 cell
High level of sensitivity to your patient´s efforts KEY TO ABBREVIATIONS
- Precise gas delivery system
- Extreme speed in sensing and regulation NIV Non invasive ventilation
Diagnostic trend tools ensure accurate time and detail recording PRVC Pressure Regulated Volume Control
for up to 24 hours after an event SIMV Synchronized Intermittent Mandatory Ventilation
Volume/Pressure and Flow/Volume loops VC Volume Control
PC Pressure Control
PS Pressure Support
CPAP Continuous Positive Airway Pressure
4. | 4 | SERVO-s | Critical Care |
TECHNICAL SPECIFICATIONS
SERVO-s
The system - General Power supply
The device complies with Power supply, automatic range 100 – 120 V AC ±10%, 50 –
requirements of Medical Device selection: 60 Hz, or 220 – 240 V AC ±10%,
Directive 93/42/EEC. 50 – 60 Hz.
Classification: Class I equipment. According External 12 V DC: 12.0 V – 15.0 V DC, 10 A
to IEC/EN 60 601-1. Battery capacity: Rechargeable, 12 V, 7 Ah
Standards: IEC/EN 60 601-1 (Type B). Battery backup time: At least 1 h
IEC/EN 60 601-2-12.
Battery recharge time: Approximately 6 h
EN 794-1.
IP classification IP 20
Max power consumption: At 100 – 120 V: 2 A, 190 VA, 140
Electromagnetic compatibility W.
(EMC): At 220 – 240 V: 1 A, 190 VA, 140
Emission: According to IEC/EN 60601-1-2 W.
(Edition 2:2001).
Immunity- Extended test to According to IEC/EN 60601-1-2.
30 v/m: The ventilator - General
The 'EMC Declaration, Information to the Responsible Dimensions: (See dimensional drawings page
Organization' is available from MAQUET. 10)
User Interface and Patient Unit: W 380 x D 300 x H 520 mm
Patient range: Patient weight 10 – 250 kg User Interface: W 355 x D 53 x H 295 mm
Patient Unit: W 380 x D 300 x H 210 mm
Operating conditions
Operating temperature: +10 to +40ºC Weight: Approximately 18 kg
Relative humidity: 15 to 95% non-condensing Method of triggering: Flow and pressure
Atmospheric pressure: 660 to 1060 hPa Max. operating pressure: Approximately 115 cmH2O
Lowest pressure in breathing –400 cmH2O Bias flow: 2 l/min
system:
Non-operating conditions
Impact: Peak acceleration: 15 g.
Pulse duration: 6 ms.
Number of impacts: 1000.
Storage temperature: –25 to +60º C (–13 to 140º F)
Storage Relative Humidity: < 95%
Storage Atmospheric Pressure: 470 to 1060 hPa
5. | Critical Care | SERVO-s | 5 |
TECHNICAL SPECIFICATIONS
SERVO-s
Gas supply Ventilation Modes – Invasive ventilation
Inlet gas pressure: 200 – 650 kPa / 2.0 – 6.5 bar / Controlled ventilation:
29 – 94 PSI PC
Connection standards available: AGA, DISS, NIST, or French
standard. VC
Unavailable gas/loss of gas The flow from an unavailable PRVC Optional
pressure: gas (air or O2) is automatically
compensated for so that the Supported ventilation:
patient gets the preset volume
and pressure. PS/CPAP
Combined ventilation:
Patient system gas connectors
SIMV (PC) + PS
Conical fittings: Male 22 mm / female 15 mm. In
accordance with ISO 5356-1. SIMV (VC) + PS
Gas exhaust port: Male 30 mm cone SIMV (PRVC) + PS Optional, comes with the
corresponding controlled
ventilation mode.
Screen Optional, pressure controlled
Bi-Vent
Type: TFT-LCD module ventilation on two independently
Size: 31 cm (12.1”) diagonal adjustable levels, allowing
unrestricted spontaneous
Viewing area: 246.0 x 184.5 mm
breathing on both levels.
Inspiratory channel
Pressure drop: Max. 6 cmH2O at a flow of 1 l/s
Ventilation modes – Non invasive ventilation (optional)
Internal compressible factor: Max. 0.1 ml/cmH2O
NIV PC
Gas delivery system: Microprocessor controlled valve NIV PS
Inspiratory flow range: 0 to 3.3 l/s
Expiratory channel
Pressure drop: Max. 3 cmH2O at a flow of 1 l/s
Internal compressible factor: Max. 0.1 ml/cmH2O
PEEP regulation: Microprocessor controlled valve
Rise time, expiratory flow <12 ms for 10 – 90 % response
measurement: at flow of 0.05 – 3.2 l/s
Expiratory flow range: 0 to 3.2 l/s
6. | 6 | SERVO-s | Critical Care |
TECHNICAL SPECIFICATIONS
SERVO-s
Alarms Alarms
Airway pressure (upper): Autoset alarm limits Invasive ventilation, controlled
Invasive ventilation: 16 – 120 cmH2O specifications: modes only
Non-invasive ventilation 16 - 40 cmH2O High airway pressure: Mean peak pressure
+10 cmH2O or at least
35 cmH2O.
Expired minute volume 0.5 - 60 l/min
(Upper alarm limit): Upper minute volume: Expiratory minute volume
+ 50%.
Expired Minute Volume 0.5 – 40 l/min
(Lower alarm limit): Lower minute volume: Expiratory minute volume
– 50%.
No patient effort (Apnea) alarm 15 – 45 s
Upper respiratory frequency: Breathing frequency + 40%.
Automatic return to support mode on patient triggering
Lower respiratory frequency: Breathing frequency – 40%.
No consistent patient effort Yes, described in User's manual High end expiratory pressure: Mean end expiratory pressure
Respiratory frequency: 1 – 160 breaths/min + 5 cmH2O.
High end expiratory pressure: 0 – 55 cmH2O Low end expiratory pressure: Mean end expiratory pressure
Low end expiratory pressure: 0 – 47 cmH2O. – 3 cmH2O.
Note. Setting the alarm to 0
(zero) is equal to alarm off.
High continuous pressure: Set PEEP level +15 cmH2O
exceeded for more than 15
seconds.
O2 concentration: Set value ±5 vol% or 18 vol%
Gas supply: Below 200 kPa / 2.0 bar /
29 PSI and
above 650 kPa / 6.5 bar / 94 PSI
Battery: Limited battery capacity:
10 min.
No battery capacity: less than
3 min.
Low battery voltage.
Leakage out of range in NIV: Yes. Described in the User’s
manual.
Technical: Yes. Described in the User’s
manual.
7. | Critical Care | SERVO-s | 7 |
TECHNICAL SPECIFICATIONS
SERVO-s
Displayed Displayed
Trended Trended
Monitoring value value* Monitoring value value*
Breathing frequency: Yes Yes Time Constant**: Yes No
Spontaneous breaths per Yes Yes P0.1 measurement**: Yes Yes
minute (RRsp): Work of Breathing patient**: Yes Yes
Peak Airway Pressure: Yes Yes Work of Breathing ventilator**: Yes Yes
Mean Airway Pressure: Yes Yes Shallow Breathing Index (SBI)**: Yes Yes
Pause Airway Pressure: Yes Yes Supply pressure (O2 and air): Yes No
End Expiratory Pressure: Yes Yes Battery remaining time: Yes No
Inspired Tidal Volume: Yes Yes Barometric pressure: Yes No
Expired Tidal Volume: Yes Yes * Stored trend values for up to 24 hours
Inspired Minute Volume: Yes Yes ** Optional
Expired Minute Volume: Yes Yes
Leakage fraction in NIV (%): Yes Yes
Waveform and loop presentations
Ti/Ttot: Yes No
Real time waveforms - up to 3 waveforms can be displayed
I:E ratio: Yes No simultaneously:
Total PEEP**: Yes No Pressure
O2 Concentration (measured): Yes Yes Flow
MVe sp / MVe: Yes No Volume
Spontaneous Exp. Minute Yes Yes
Volume (MVe sp): Loops:
End Expiratory Flow: Yes Yes Volume / Pressure
Static Compliance**: Yes Yes Flow / Volume
Dynamic Compliance**: Yes Yes
Inspiratory Resistance**: Yes Yes
Expiratory Resistance**: Yes Yes Log function
Elastance**: Yes Yes Event log: Alarms.
Ventilator settings.
Apnea periods.
Immediate functions.
Service log: Technical alarms.
Test results.
Preventive maintenance.
Service history.
Configuration log.
8. | 8 | SERVO-s | Critical Care |
TECHNICAL SPECIFICATIONS
SERVO-s
Parameter: Setting range: Parameter: Setting range:
Inspiratory tidal volume (ml): 100 – 2000 Oxygen breaths: 100% for 1 minute
Inspiratory minute volume 0.5 – 60 Start breath: Initiation of 1 breath (In SIMV
(l/min): mode initiation of 1 mandatory
Apnea, time to alarm (s): 15 – 45 breath)
PC/PS above PEEP (cmH2O) 0 – (120 - PEEP) Pause hold: Insp. or exp (0 – 30 seconds)
PC/PS above PEEP in NIV 0 – (32 - PEEP) Alarm silence/reset: 2 minute silence and reset of
(cmH2O) latched alarms
PEEP (cmH2O): 0 – 50 Compliance compensation: On/Off
PEEP in NIV (cmH2O): 2 – 20 Backup ventilation On/Off
CMV frequency (breaths/min): 4 – 100
SIMV frequency (breaths/min): 1 – 60 Backup settings Setting range:
Breath cycle time, SIMV (s): 1 – 15
Parameter:
PHigh (cmH2O): (PEEP+1) – 50
THigh (s): 0.2 – 10 Inspiratory tidal volume (ml): 100 - 2000
TPEEP (s): 0.2 – 10 PC above PEEP (cmH2O) 5 – (120 - PEEP)
O2 concentration (%): 21 – 100 PC above PEEP in NIV 5 – (32 - PEEP)
I:E ratio: 1:10 – 4:1 CMV frequency (breaths/min): 4 – 100
TInsp (s): 0.1 – 5 I:E ratio: 1:10 – 4:1
TPause (% of breath cycle time) 0 – 30 TInsp (s): 0.1 – 5
Flow trigger sensitivity level 0 – 100%
(fraction of bias flow):
Suction Support
Press. trigg sensitivity (cmH2O): -20 – 0
Pre oxygenation time: Max. 2 min
Insp. rise time (% of breath 0 – 20
cycle time): Post oxygenation time: Max. 1 min
Insp. rise time (s): 0 – 0.4 Suction phase time: No maximum level
Insp. cycle off (% of peak flow): 1 – 70 Adjustable oxygen level: 21 – 100 %
Insp. cycle off in NIV (% of peak 10 – 70
flow):
Communication/Interface
Serial port: RS-232C - isolated. For data
communication via the
Communication Interface
Emulator (CIE).
9. | Critical Care | SERVO-s | 9 |
TECHNICAL SPECIFICATIONS
SERVO-s
Non invasive ventilation (optional) SERVO-s Mobile Cart (optional)
Max. leakage compensation 65 l/min Weight: 25 kg
level:
Dimensions: H 830 mm (with handles
Leakage overrange detection: Automatic 1.040 mm)
Disconnect detection: Automatic x L 650 mm x W 500 mm
Disconnect flow: Configurable (see dimensional drawing)
Low: 7.5 l/min
High: 40 l/min Gas cylinder restrainer (optional)
Disabled Deactivates disconnect Max load: 2 x 5-liter bottles
detection
Connect detection: Manual, or automatic via bias Compressor Mini (optional)
flow See separate data sheet.
Alarm output connector (optional) SERVO-s Shelf Base (optional)
Connector: 4-pole Modular connector Weight: 0.1 kg
Ratings: Max 40 V DC, Max 500 mA, Dimensions: H 8 mm x L 160 mm x W 60 mm
Max 20 W (see dimensional drawing)
Service Aeroneb Nebulizer Systems (optional)
Regular maintenance: Once every 12 months or at See separate data sheets.
least after 5000 operating hours.
ORDERING INFORMATION
Note: For inaccuracies and more detailed technical
specifications please refer to the User’s manual. SERVO-s, Ventilator and accessories: See separate information:
“SERVO-, System Version 5.0 — System Flow Chart” (Order no:
66 70 112).
10. | 10 | SERVO-s | Critical Care |
TECHNICAL SPECIFICATIONS
SERVO-s
Dimensional drawings
SERVO-s on Mobile Cart SERVO-s on Shelf base
1350 mm
295 mm
520 mm
1040 mm
803 mm
210 mm
380 mm 300 mm
500 mm 605 mm
Mobile Cart SERVO-s Shelf Base SERVO-s
8 mm
1040 mm
830 mm
60 mm
500 mm 650 mm
160 mm