SlideShare uma empresa Scribd logo
1 de 30
HPLC Problems, Probable Causes, and
Remedies
Problem No. 1: No Peaks/Very Small Peaks
• Probable Cause
1) Detector lamp off.
2) Loose/broken wire between detector and integrator or recorder.
3) No mobile phase flow.
4) No sample/deteriorated sample/ wrong sample.
5) Settings too high on detector or recorder.
• Remedy/Comments
1) Turn lamp on.
2) Check electrical connections and cables.
3) Pump off/ valve (knob) open
4) Be sure automatic sampler vials have sufficient liquid and no air bubbles in the
sample. Evaluate system performance with fresh standard to confirm sample as
source of problem.
5) Check attenuation or gain settings. Check lamp status. Auto-zero if necessary.
Problem No. 2: No Flow
• Probable Cause
1) Pump off.
2) Flow interrupted/obstructed.
3) Leak.
4) Air trapped in pump head. (Revealed by pressure fluctuations.)
• Remedy/Comments
1) Start pump.
2) Check mobile phase level in reservoir(s). Check flow throughout system. Examine sample loop for
obstruction or air lock. Make sure mobile phase components are miscible and mobile phase is properly
degassed.
3) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if
necessary.
4) Disconnect tubing at guard column (if present) or analytical column inlet. Check for flow. Purge pump at high
flow rate (e.g., 5-10 mL/min.), prime system if necessary. (Prime each pump head separately.) If system has
check valve, loosen valve to allow air to escape. If problem persists, flush system with 100% methanol or
isopropanol. If problem still persists, contact system manufacturer.
Problem No. 3: No Pressure/Pressure Lower
• Probable Cause
1) Leak.
2) Mobile phase flow interrupted/obstructed.
3) Air trapped in pump head. (Revealed by pressure fluctuations.)
4) Leak at column inlet end fitting.
5) Air trapped elsewhere in system.
6) Worn pump seal causing leaks around pump head.
7) Faulty check valve.
8) Faulty pump seals.
• Remedy/Comments
1) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary.
2) Check mobile phase level in reservoir(s). Check flow throughout system. Examine sample loop for obstruction or air lock.
Make sure mobile phase components are miscible and mobile phase is properly degassed.
3) Disconnect tubing at guard column (if present) or analytical column inlet. Check for flow. Purge pump at high flow rate
(e.g., 10 mL/min.), prime system if necessary. (Prime each pump head separately.) If system has check valve, loosen valve
to allow air to escape.
4) Reconnect column and pump solvent at double the flow rate. If pressure is still low, check for leaks at inlet fitting or
column end fitting.
• Disconnect guard and analytical column and purge system. Reconnect column(s). If problem persists, flush system with
100% methanol or isopropanol.
• Replace seal. If problem persists, replace piston and seal.
• Rebuild or replace valve.
• Replace seals.
Problem No. 4: Pressure Higher
• Probable Cause
• Problem in pump, injector, in-line filter, or tubing.
• Obstructed guard column or analytical column.
• Remedy/Comments
• Remove guard column and analytical column from system. Replace with unions and 0.010'' I.D. or larger tubing to reconnect
injector to detector. Run pump at 2-5 mL/min. If pressure is minimal, see Cause 2. If not, isolate cause by systematically
eliminating system components, starting with detector, then in-line filter, and working back to pump. Replace filter in pump
if present.
• Remove guard column (if present) and check pressure. Replace guard column if necessary. If analytical column is
obstructed, reverse and flush the column, while disconnected from the detector. If problem persists, column may be
clogged with strongly retained contaminants. Use appropriate restoration procedure (Table 1). If problem still
persists, change inlet frit or replace column.
Problem No. 5: Variable Retention Times
• Probable Cause
1) Leak.
2) Change in mobile phase composition. (Small changes can lead to large changes in retention times.)
3) Air trapped in pump. (Retention times increase and decrease at random times.)
4) Column temperature fluctuations (especially evident in ion exchange systems).
5) Column overloading. (Retention times usually decrease as mass of solute injected on column exceeds column capacity.)
6) Sample solvent incompatible with mobile phase.
7) Column problem. (Not a common cause of erratic retention. As a column ages, retention times gradually decrease.)
• Remedy/Comments
1) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary.
2) Check make-up of mobile phase. If mobile phase is machine mixed using proportioning values, hand mix and supply from
one reservoir.
3) Purge air from pump head or check valves. Change pump seals if necessary. Be sure mobile phase is degassed.
4) Use reliable column oven. (Note: higher column temperatures increase column efficiency. For optimum results, heat
eluant before introducing it onto column.)
5) Inject smaller volume (e.g., 10 μL vs. 100 μL) or inject the same volume after 1:10 or 1:100 dilutions of sample.
6) Adjust solvent. Whenever possible, inject samples in mobile phase.
7) Substitute new column of same type to confirm column as cause. Discard old column if restoration procedures fail.
Problem No. 6: Loss of Resolution
• Probable Cause
1) Mobile phase contaminated/deteriorated (causing retention times and/or selectivity to change).
2) Obstructed guard or analytical column.
• Remedy/Comments
1) Prepare fresh mobile phase.
2) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is
obstructed, reverse and flush. If problem persists, column may be clogged with strongly retained contaminants. Use
appropriate restoration procedure (Table 1). If problem still persists, change inlet frit or replace column.
Problem No. 7: Split Peaks
• Probable Cause
1) Contamination on guard or analytical column inlet.
2) Partially blocked frit.
3) Small (uneven) void at column inlet.
4) Sample solvent incompatible with mobile phase.
• Remedy/Comments
1) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is
obstructed, reverse and flush. If problem persists, column may be clogged with strongly retained contaminants. Use
appropriate restoration procedure (Table1 ). If problem still persists, inlet frit is probably (partially) plugged. Change
frit or replace column.
2) Replace frit (see above).
3) Repack top of column with pellicular particles of same bonded phase functionality. Continue using the column in reverse
flow direction.
4) Adjust solvent. Whenever possible, inject samples in mobile phase.
Problem No. 8: Tailing Peaks
• Probable Cause
1) Guard or analytical column contaminated/worn out.
2) Mobile phase contaminated/deteriorated.
3) Interfering components in sample.
• Remedy/Comments
1) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is source
of problem, use appropriate restoration procedure (Table 1). If problem persists, replace column.
2) Check make-up of mobile phase.
3) Check column performance with standards.
Problem No. 9: Fronting Peaks
• Probable Cause
1) Column overloaded.
2) Sample solvent incompatible with mobile phase.
3) Shoulder or gradual baseline rise before a main peak may be another sample component.
• Remedy/Comments
1) Inject smaller volume (e.g., 10 μL vs. 100 μL). Dilute the sample 1:10 or 1:100 fold in case of mass overload.
2) Adjust solvent. Whenever possible, inject samples in mobile phase. Flush polar bonded phase column with 50 column
volumes HPLC grade ethyl acetate at 2-3 times the standard flow rate, then with intermediate polarity solvent prior to
analysis.
3) Increase efficiency or change selectivity of system to improve resolution. Try another column type if necessary (e.g.,
switch from nonpolar C18 to polar cyano phase).
Problem No. 10: Baseline Noise (regular)
• Probable Cause
1) Air in mobile phase, detector cell, or pump.
2) Pump pulsations.
3) Incomplete mobile phase mixing.
4) Temperature effect (column at high temperature, detector unheated).
5) Other electronic equipment on same line.
6) Leak.
• Remedy/Comments
1) Degas mobile phase. Flush system to remove air from detector cell or pump.
2) Incorporate pulse damper into system.
3) Mix mobile phase by hand or use less viscous solvent.
4) Reduce differential or add heat exchanger.
5) Isolate LC, detector, recorder to determine if source of problem is external. Correct as necessary.
6) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary.
Problem No. 11: Negative Peak
• Probable Cause
1) Recorder leads reversed.
2) Refractive index of solute less than that of mobile phase (RI detector).
3) Sample solvent and mobile phase differ greatly in composition (vacancy peaks).
4) Mobile phase more absorptive than sample components to UV wavelength.
• Remedy/Comments
1) Check polarity.
2) Use mobile phase with lower refractive index, or reverse recorder leads.
3) Adjust or change sample solvent. Dilute sample in mobile phase whenever possible.
4) a. Change polarity when using indirect UV detection, or
b. Change UV wavelength or use mobile phase that does not adsorb chosen wavelength.
Problem No. 12: Ghost Peak
• Probable Cause
1) Contamination in injector or column.
2) Late eluting peak (usually broad) present in sample.
• Remedy/Comments
1) Flush injector between analyses (a good routine practice). If necessary, run strong solvent through column
to remove late eluters. Include final wash step in gradient analyses, to remove strongly retained compounds.
2) a. Check sample preparation.
b. Include (step) gradient to quickly elute component.
Table-1
• Table-1. Column Restoration Procedures
• Silica-Based Reversed Phase Column
(alkyl (C8, C18, etc.), phenyl, or diphenyl column, SUPELCOSIL LC-PAH column)
•A. Water Soluble Samples Flush with the following:
• Flush with warm (60 °C) distilled water
• 50 mL methanol
• 50 mL acetonitrile
• 25 mL methanol
• 25 mL mobile phase
• Evaluate column performance.
•B. Samples Not Soluble in Water Flush with the
following:
• 50 mL 2-propanol
• 50 mL methylene chloride
• 50 mL hexane
• 25 mL isopropanol
• 25 mL mobile phase
• Evaluate column performance.
THANK YOU

Mais conteúdo relacionado

Mais procurados

HPLC INTRO AND TROUBLESHOOTING
HPLC INTRO AND TROUBLESHOOTINGHPLC INTRO AND TROUBLESHOOTING
HPLC INTRO AND TROUBLESHOOTINGSirshendu Roy
 
HPLC Sample Preparation and Instrumentatio
HPLC Sample Preparation and InstrumentatioHPLC Sample Preparation and Instrumentatio
HPLC Sample Preparation and InstrumentatioAdnan Sohail Akhtar
 
Development and Validation of a RP-HPLC method
Development and Validation of a RP-HPLC methodDevelopment and Validation of a RP-HPLC method
Development and Validation of a RP-HPLC methodUshaKhanal3
 
Chromatographic Column chemistry
Chromatographic Column chemistryChromatographic Column chemistry
Chromatographic Column chemistrykiranreddy munnangi
 
Chiral chromatography & ion pair chromatography
Chiral chromatography & ion pair chromatographyChiral chromatography & ion pair chromatography
Chiral chromatography & ion pair chromatographyHemantBansode2
 
HPLC Principle,Instrumentation and Application
HPLC Principle,Instrumentation and ApplicationHPLC Principle,Instrumentation and Application
HPLC Principle,Instrumentation and ApplicationSureshKumarKandode
 
Hplc (basic principles, operation and maintenance)
Hplc (basic principles, operation and maintenance)Hplc (basic principles, operation and maintenance)
Hplc (basic principles, operation and maintenance)alim125135
 
Optimum Performance Laminar Chromatography (OPLC)
Optimum Performance Laminar Chromatography (OPLC) Optimum Performance Laminar Chromatography (OPLC)
Optimum Performance Laminar Chromatography (OPLC) D.R. Chandravanshi
 
System suitability parameters assessment by HPLC
System suitability parameters assessment by HPLCSystem suitability parameters assessment by HPLC
System suitability parameters assessment by HPLCAnirban Barik
 
Human liver microsomes & rat liver microsomes
Human liver microsomes & rat liver microsomesHuman liver microsomes & rat liver microsomes
Human liver microsomes & rat liver microsomesgaurav sharma
 
Analytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USPAnalytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USPKushal Shah
 

Mais procurados (20)

HPLC Column
HPLC ColumnHPLC Column
HPLC Column
 
HPLC INTRO AND TROUBLESHOOTING
HPLC INTRO AND TROUBLESHOOTINGHPLC INTRO AND TROUBLESHOOTING
HPLC INTRO AND TROUBLESHOOTING
 
HPLC Sample Preparation and Instrumentatio
HPLC Sample Preparation and InstrumentatioHPLC Sample Preparation and Instrumentatio
HPLC Sample Preparation and Instrumentatio
 
Hplc
Hplc Hplc
Hplc
 
Gc troubleshooting
Gc troubleshootingGc troubleshooting
Gc troubleshooting
 
Development and Validation of a RP-HPLC method
Development and Validation of a RP-HPLC methodDevelopment and Validation of a RP-HPLC method
Development and Validation of a RP-HPLC method
 
Chromatographic Column chemistry
Chromatographic Column chemistryChromatographic Column chemistry
Chromatographic Column chemistry
 
Columns in HPLC
Columns in HPLCColumns in HPLC
Columns in HPLC
 
Chiral chromatography & ion pair chromatography
Chiral chromatography & ion pair chromatographyChiral chromatography & ion pair chromatography
Chiral chromatography & ion pair chromatography
 
Hplc instrumentation
Hplc instrumentationHplc instrumentation
Hplc instrumentation
 
HPLC-COLUMNS
HPLC-COLUMNS HPLC-COLUMNS
HPLC-COLUMNS
 
Hplc method development
Hplc method developmentHplc method development
Hplc method development
 
HPLC Principle,Instrumentation and Application
HPLC Principle,Instrumentation and ApplicationHPLC Principle,Instrumentation and Application
HPLC Principle,Instrumentation and Application
 
Hplc (basic principles, operation and maintenance)
Hplc (basic principles, operation and maintenance)Hplc (basic principles, operation and maintenance)
Hplc (basic principles, operation and maintenance)
 
Optimum Performance Laminar Chromatography (OPLC)
Optimum Performance Laminar Chromatography (OPLC) Optimum Performance Laminar Chromatography (OPLC)
Optimum Performance Laminar Chromatography (OPLC)
 
Uplc&nano
Uplc&nanoUplc&nano
Uplc&nano
 
Scope, objectives & applications of instrumental methods of analysis
Scope, objectives & applications of instrumental methods of analysisScope, objectives & applications of instrumental methods of analysis
Scope, objectives & applications of instrumental methods of analysis
 
System suitability parameters assessment by HPLC
System suitability parameters assessment by HPLCSystem suitability parameters assessment by HPLC
System suitability parameters assessment by HPLC
 
Human liver microsomes & rat liver microsomes
Human liver microsomes & rat liver microsomesHuman liver microsomes & rat liver microsomes
Human liver microsomes & rat liver microsomes
 
Analytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USPAnalytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USP
 

Semelhante a HPLC Problems, Probable Causes, and Remedies2020.pptx

hplc-120629111345-phpapp01.ppt
hplc-120629111345-phpapp01.ppthplc-120629111345-phpapp01.ppt
hplc-120629111345-phpapp01.pptZORAIZ HAIDER
 
Tips and tricks_hplc_troubleshooting
Tips and tricks_hplc_troubleshootingTips and tricks_hplc_troubleshooting
Tips and tricks_hplc_troubleshootingFlavio De Jorge
 
HPLC Instrumentation & Techniques_basics.ppt
HPLC Instrumentation & Techniques_basics.pptHPLC Instrumentation & Techniques_basics.ppt
HPLC Instrumentation & Techniques_basics.pptBussaSravanKumar
 
Informative Lecture about Atomic Absorption Spectrometer.pptx
Informative Lecture about Atomic Absorption Spectrometer.pptxInformative Lecture about Atomic Absorption Spectrometer.pptx
Informative Lecture about Atomic Absorption Spectrometer.pptxHemant Patil
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatographyVaishali Dudhabale
 
New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentationHARSHITHA REDDY
 
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...Vandana Devesh Sharma
 
ICS 5000+ Trouble-shooting guide
ICS 5000+ Trouble-shooting guideICS 5000+ Trouble-shooting guide
ICS 5000+ Trouble-shooting guideElliot Morgan
 
introduccion a los Inyectores para cromatografía de gases
introduccion a los Inyectores para cromatografía de gasesintroduccion a los Inyectores para cromatografía de gases
introduccion a los Inyectores para cromatografía de gasesLgIrvin
 
Care and use manual waters x bridge columns
Care and use manual waters x bridge columnsCare and use manual waters x bridge columns
Care and use manual waters x bridge columnsJohn Omondi
 
Gagandutt sharma(HPLC) [Pharmaceutics].pdf
Gagandutt sharma(HPLC) [Pharmaceutics].pdfGagandutt sharma(HPLC) [Pharmaceutics].pdf
Gagandutt sharma(HPLC) [Pharmaceutics].pdfRitikSharma849463
 
Gagandutt sharma (HPLC) ..pptx
Gagandutt sharma (HPLC) ..pptxGagandutt sharma (HPLC) ..pptx
Gagandutt sharma (HPLC) ..pptxRitikSharma849463
 

Semelhante a HPLC Problems, Probable Causes, and Remedies2020.pptx (20)

HPLC_basics.ppt
HPLC_basics.pptHPLC_basics.ppt
HPLC_basics.ppt
 
HPLC
HPLCHPLC
HPLC
 
Hplc
HplcHplc
Hplc
 
hplc-120629111345-phpapp01.ppt
hplc-120629111345-phpapp01.ppthplc-120629111345-phpapp01.ppt
hplc-120629111345-phpapp01.ppt
 
Tips and tricks_hplc_troubleshooting
Tips and tricks_hplc_troubleshootingTips and tricks_hplc_troubleshooting
Tips and tricks_hplc_troubleshooting
 
HPLC Instrumentation & Techniques_basics.ppt
HPLC Instrumentation & Techniques_basics.pptHPLC Instrumentation & Techniques_basics.ppt
HPLC Instrumentation & Techniques_basics.ppt
 
Informative Lecture about Atomic Absorption Spectrometer.pptx
Informative Lecture about Atomic Absorption Spectrometer.pptxInformative Lecture about Atomic Absorption Spectrometer.pptx
Informative Lecture about Atomic Absorption Spectrometer.pptx
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
 
HPLC.pptx
HPLC.pptxHPLC.pptx
HPLC.pptx
 
New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentation
 
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
 
HPLC
HPLC HPLC
HPLC
 
Hplc.pptx
Hplc.pptxHplc.pptx
Hplc.pptx
 
ICS 5000+ Trouble-shooting guide
ICS 5000+ Trouble-shooting guideICS 5000+ Trouble-shooting guide
ICS 5000+ Trouble-shooting guide
 
Hplc by ikrm khan and groupe
Hplc by ikrm khan and groupeHplc by ikrm khan and groupe
Hplc by ikrm khan and groupe
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
 
introduccion a los Inyectores para cromatografía de gases
introduccion a los Inyectores para cromatografía de gasesintroduccion a los Inyectores para cromatografía de gases
introduccion a los Inyectores para cromatografía de gases
 
Care and use manual waters x bridge columns
Care and use manual waters x bridge columnsCare and use manual waters x bridge columns
Care and use manual waters x bridge columns
 
Gagandutt sharma(HPLC) [Pharmaceutics].pdf
Gagandutt sharma(HPLC) [Pharmaceutics].pdfGagandutt sharma(HPLC) [Pharmaceutics].pdf
Gagandutt sharma(HPLC) [Pharmaceutics].pdf
 
Gagandutt sharma (HPLC) ..pptx
Gagandutt sharma (HPLC) ..pptxGagandutt sharma (HPLC) ..pptx
Gagandutt sharma (HPLC) ..pptx
 

Último

Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service Nashik
Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service NashikLow Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service Nashik
Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
Call Girls in Vashi Escorts Services - 7738631006
Call Girls in Vashi Escorts Services - 7738631006Call Girls in Vashi Escorts Services - 7738631006
Call Girls in Vashi Escorts Services - 7738631006Pooja Nehwal
 
Shikrapur Call Girls Most Awaited Fun 6297143586 High Profiles young Beautie...
Shikrapur Call Girls Most Awaited Fun  6297143586 High Profiles young Beautie...Shikrapur Call Girls Most Awaited Fun  6297143586 High Profiles young Beautie...
Shikrapur Call Girls Most Awaited Fun 6297143586 High Profiles young Beautie...tanu pandey
 
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...amitlee9823
 
Book Sex Workers Available Pune Call Girls Yerwada 6297143586 Call Hot India...
Book Sex Workers Available Pune Call Girls Yerwada  6297143586 Call Hot India...Book Sex Workers Available Pune Call Girls Yerwada  6297143586 Call Hot India...
Book Sex Workers Available Pune Call Girls Yerwada 6297143586 Call Hot India...Call Girls in Nagpur High Profile
 
Top Rated Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
Top Rated  Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...Top Rated  Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
Top Rated Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...Call Girls in Nagpur High Profile
 
(PARI) Alandi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(PARI) Alandi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(PARI) Alandi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(PARI) Alandi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escortsranjana rawat
 
Top Rated Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
Top Rated  Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...Top Rated  Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
Top Rated Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...Call Girls in Nagpur High Profile
 
Lubrication and it's types and properties of the libricabt
Lubrication and it's types and properties of the libricabtLubrication and it's types and properties of the libricabt
Lubrication and it's types and properties of the libricabtdineshkumar430venkat
 
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...MOHANI PANDEY
 
Call Now ≽ 9953056974 ≼🔝 Call Girls In Yusuf Sarai ≼🔝 Delhi door step delevry≼🔝
Call Now ≽ 9953056974 ≼🔝 Call Girls In Yusuf Sarai ≼🔝 Delhi door step delevry≼🔝Call Now ≽ 9953056974 ≼🔝 Call Girls In Yusuf Sarai ≼🔝 Delhi door step delevry≼🔝
Call Now ≽ 9953056974 ≼🔝 Call Girls In Yusuf Sarai ≼🔝 Delhi door step delevry≼🔝9953056974 Low Rate Call Girls In Saket, Delhi NCR
 
Deira Dubai Escorts +0561951007 Escort Service in Dubai by Dubai Escort Girls
Deira Dubai Escorts +0561951007 Escort Service in Dubai by Dubai Escort GirlsDeira Dubai Escorts +0561951007 Escort Service in Dubai by Dubai Escort Girls
Deira Dubai Escorts +0561951007 Escort Service in Dubai by Dubai Escort GirlsEscorts Call Girls
 
Makarba ( Call Girls ) Ahmedabad ✔ 6297143586 ✔ Hot Model With Sexy Bhabi Rea...
Makarba ( Call Girls ) Ahmedabad ✔ 6297143586 ✔ Hot Model With Sexy Bhabi Rea...Makarba ( Call Girls ) Ahmedabad ✔ 6297143586 ✔ Hot Model With Sexy Bhabi Rea...
Makarba ( Call Girls ) Ahmedabad ✔ 6297143586 ✔ Hot Model With Sexy Bhabi Rea...Naicy mandal
 
VVIP Pune Call Girls Karve Nagar (7001035870) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Karve Nagar (7001035870) Pune Escorts Nearby with Comple...VVIP Pune Call Girls Karve Nagar (7001035870) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Karve Nagar (7001035870) Pune Escorts Nearby with Comple...Call Girls in Nagpur High Profile
 
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)kojalkojal131
 
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...motiram463
 
Call Girls Dubai Slut Wife O525547819 Call Girls Dubai Gaped
Call Girls Dubai Slut Wife O525547819 Call Girls Dubai GapedCall Girls Dubai Slut Wife O525547819 Call Girls Dubai Gaped
Call Girls Dubai Slut Wife O525547819 Call Girls Dubai Gapedkojalkojal131
 
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样qaffana
 
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...amitlee9823
 

Último (20)

Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service Nashik
Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service NashikLow Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service Nashik
Low Rate Call Girls Nashik Vedika 7001305949 Independent Escort Service Nashik
 
Call Girls in Vashi Escorts Services - 7738631006
Call Girls in Vashi Escorts Services - 7738631006Call Girls in Vashi Escorts Services - 7738631006
Call Girls in Vashi Escorts Services - 7738631006
 
Shikrapur Call Girls Most Awaited Fun 6297143586 High Profiles young Beautie...
Shikrapur Call Girls Most Awaited Fun  6297143586 High Profiles young Beautie...Shikrapur Call Girls Most Awaited Fun  6297143586 High Profiles young Beautie...
Shikrapur Call Girls Most Awaited Fun 6297143586 High Profiles young Beautie...
 
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
Call Girls Banashankari Just Call 👗 7737669865 👗 Top Class Call Girl Service ...
 
Book Sex Workers Available Pune Call Girls Yerwada 6297143586 Call Hot India...
Book Sex Workers Available Pune Call Girls Yerwada  6297143586 Call Hot India...Book Sex Workers Available Pune Call Girls Yerwada  6297143586 Call Hot India...
Book Sex Workers Available Pune Call Girls Yerwada 6297143586 Call Hot India...
 
Top Rated Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
Top Rated  Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...Top Rated  Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
Top Rated Pune Call Girls Katraj ⟟ 6297143586 ⟟ Call Me For Genuine Sex Serv...
 
(PARI) Alandi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(PARI) Alandi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(PARI) Alandi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(PARI) Alandi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
 
🔝 9953056974🔝 Delhi Call Girls in Ajmeri Gate
🔝 9953056974🔝 Delhi Call Girls in Ajmeri Gate🔝 9953056974🔝 Delhi Call Girls in Ajmeri Gate
🔝 9953056974🔝 Delhi Call Girls in Ajmeri Gate
 
Top Rated Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
Top Rated  Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...Top Rated  Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
Top Rated Pune Call Girls Shirwal ⟟ 6297143586 ⟟ Call Me For Genuine Sex Ser...
 
Lubrication and it's types and properties of the libricabt
Lubrication and it's types and properties of the libricabtLubrication and it's types and properties of the libricabt
Lubrication and it's types and properties of the libricabt
 
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...
Get Premium Pimple Saudagar Call Girls (8005736733) 24x7 Rate 15999 with A/c ...
 
Call Now ≽ 9953056974 ≼🔝 Call Girls In Yusuf Sarai ≼🔝 Delhi door step delevry≼🔝
Call Now ≽ 9953056974 ≼🔝 Call Girls In Yusuf Sarai ≼🔝 Delhi door step delevry≼🔝Call Now ≽ 9953056974 ≼🔝 Call Girls In Yusuf Sarai ≼🔝 Delhi door step delevry≼🔝
Call Now ≽ 9953056974 ≼🔝 Call Girls In Yusuf Sarai ≼🔝 Delhi door step delevry≼🔝
 
Deira Dubai Escorts +0561951007 Escort Service in Dubai by Dubai Escort Girls
Deira Dubai Escorts +0561951007 Escort Service in Dubai by Dubai Escort GirlsDeira Dubai Escorts +0561951007 Escort Service in Dubai by Dubai Escort Girls
Deira Dubai Escorts +0561951007 Escort Service in Dubai by Dubai Escort Girls
 
Makarba ( Call Girls ) Ahmedabad ✔ 6297143586 ✔ Hot Model With Sexy Bhabi Rea...
Makarba ( Call Girls ) Ahmedabad ✔ 6297143586 ✔ Hot Model With Sexy Bhabi Rea...Makarba ( Call Girls ) Ahmedabad ✔ 6297143586 ✔ Hot Model With Sexy Bhabi Rea...
Makarba ( Call Girls ) Ahmedabad ✔ 6297143586 ✔ Hot Model With Sexy Bhabi Rea...
 
VVIP Pune Call Girls Karve Nagar (7001035870) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Karve Nagar (7001035870) Pune Escorts Nearby with Comple...VVIP Pune Call Girls Karve Nagar (7001035870) Pune Escorts Nearby with Comple...
VVIP Pune Call Girls Karve Nagar (7001035870) Pune Escorts Nearby with Comple...
 
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)
(=Towel) Dubai Call Girls O525547819 Call Girls In Dubai (Fav0r)
 
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
(👉Ridhima)👉VIP Model Call Girls Mulund ( Mumbai) Call ON 9967824496 Starting ...
 
Call Girls Dubai Slut Wife O525547819 Call Girls Dubai Gaped
Call Girls Dubai Slut Wife O525547819 Call Girls Dubai GapedCall Girls Dubai Slut Wife O525547819 Call Girls Dubai Gaped
Call Girls Dubai Slut Wife O525547819 Call Girls Dubai Gaped
 
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样
哪里办理美国宾夕法尼亚州立大学毕业证(本硕)psu成绩单原版一模一样
 
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...
Kothanur Call Girls Service: 🍓 7737669865 🍓 High Profile Model Escorts | Bang...
 

HPLC Problems, Probable Causes, and Remedies2020.pptx

  • 1. HPLC Problems, Probable Causes, and Remedies
  • 2. Problem No. 1: No Peaks/Very Small Peaks
  • 3. • Probable Cause 1) Detector lamp off. 2) Loose/broken wire between detector and integrator or recorder. 3) No mobile phase flow. 4) No sample/deteriorated sample/ wrong sample. 5) Settings too high on detector or recorder. • Remedy/Comments 1) Turn lamp on. 2) Check electrical connections and cables. 3) Pump off/ valve (knob) open 4) Be sure automatic sampler vials have sufficient liquid and no air bubbles in the sample. Evaluate system performance with fresh standard to confirm sample as source of problem. 5) Check attenuation or gain settings. Check lamp status. Auto-zero if necessary.
  • 4. Problem No. 2: No Flow
  • 5. • Probable Cause 1) Pump off. 2) Flow interrupted/obstructed. 3) Leak. 4) Air trapped in pump head. (Revealed by pressure fluctuations.) • Remedy/Comments 1) Start pump. 2) Check mobile phase level in reservoir(s). Check flow throughout system. Examine sample loop for obstruction or air lock. Make sure mobile phase components are miscible and mobile phase is properly degassed. 3) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary. 4) Disconnect tubing at guard column (if present) or analytical column inlet. Check for flow. Purge pump at high flow rate (e.g., 5-10 mL/min.), prime system if necessary. (Prime each pump head separately.) If system has check valve, loosen valve to allow air to escape. If problem persists, flush system with 100% methanol or isopropanol. If problem still persists, contact system manufacturer.
  • 6. Problem No. 3: No Pressure/Pressure Lower
  • 7. • Probable Cause 1) Leak. 2) Mobile phase flow interrupted/obstructed. 3) Air trapped in pump head. (Revealed by pressure fluctuations.) 4) Leak at column inlet end fitting. 5) Air trapped elsewhere in system. 6) Worn pump seal causing leaks around pump head. 7) Faulty check valve. 8) Faulty pump seals. • Remedy/Comments 1) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary. 2) Check mobile phase level in reservoir(s). Check flow throughout system. Examine sample loop for obstruction or air lock. Make sure mobile phase components are miscible and mobile phase is properly degassed. 3) Disconnect tubing at guard column (if present) or analytical column inlet. Check for flow. Purge pump at high flow rate (e.g., 10 mL/min.), prime system if necessary. (Prime each pump head separately.) If system has check valve, loosen valve to allow air to escape. 4) Reconnect column and pump solvent at double the flow rate. If pressure is still low, check for leaks at inlet fitting or column end fitting.
  • 8. • Disconnect guard and analytical column and purge system. Reconnect column(s). If problem persists, flush system with 100% methanol or isopropanol. • Replace seal. If problem persists, replace piston and seal. • Rebuild or replace valve. • Replace seals.
  • 9. Problem No. 4: Pressure Higher
  • 10. • Probable Cause • Problem in pump, injector, in-line filter, or tubing. • Obstructed guard column or analytical column. • Remedy/Comments • Remove guard column and analytical column from system. Replace with unions and 0.010'' I.D. or larger tubing to reconnect injector to detector. Run pump at 2-5 mL/min. If pressure is minimal, see Cause 2. If not, isolate cause by systematically eliminating system components, starting with detector, then in-line filter, and working back to pump. Replace filter in pump if present. • Remove guard column (if present) and check pressure. Replace guard column if necessary. If analytical column is obstructed, reverse and flush the column, while disconnected from the detector. If problem persists, column may be clogged with strongly retained contaminants. Use appropriate restoration procedure (Table 1). If problem still persists, change inlet frit or replace column.
  • 11. Problem No. 5: Variable Retention Times
  • 12. • Probable Cause 1) Leak. 2) Change in mobile phase composition. (Small changes can lead to large changes in retention times.) 3) Air trapped in pump. (Retention times increase and decrease at random times.) 4) Column temperature fluctuations (especially evident in ion exchange systems). 5) Column overloading. (Retention times usually decrease as mass of solute injected on column exceeds column capacity.) 6) Sample solvent incompatible with mobile phase. 7) Column problem. (Not a common cause of erratic retention. As a column ages, retention times gradually decrease.) • Remedy/Comments 1) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary. 2) Check make-up of mobile phase. If mobile phase is machine mixed using proportioning values, hand mix and supply from one reservoir. 3) Purge air from pump head or check valves. Change pump seals if necessary. Be sure mobile phase is degassed. 4) Use reliable column oven. (Note: higher column temperatures increase column efficiency. For optimum results, heat eluant before introducing it onto column.) 5) Inject smaller volume (e.g., 10 μL vs. 100 μL) or inject the same volume after 1:10 or 1:100 dilutions of sample. 6) Adjust solvent. Whenever possible, inject samples in mobile phase. 7) Substitute new column of same type to confirm column as cause. Discard old column if restoration procedures fail.
  • 13. Problem No. 6: Loss of Resolution
  • 14. • Probable Cause 1) Mobile phase contaminated/deteriorated (causing retention times and/or selectivity to change). 2) Obstructed guard or analytical column. • Remedy/Comments 1) Prepare fresh mobile phase. 2) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is obstructed, reverse and flush. If problem persists, column may be clogged with strongly retained contaminants. Use appropriate restoration procedure (Table 1). If problem still persists, change inlet frit or replace column.
  • 15. Problem No. 7: Split Peaks
  • 16. • Probable Cause 1) Contamination on guard or analytical column inlet. 2) Partially blocked frit. 3) Small (uneven) void at column inlet. 4) Sample solvent incompatible with mobile phase. • Remedy/Comments 1) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is obstructed, reverse and flush. If problem persists, column may be clogged with strongly retained contaminants. Use appropriate restoration procedure (Table1 ). If problem still persists, inlet frit is probably (partially) plugged. Change frit or replace column. 2) Replace frit (see above). 3) Repack top of column with pellicular particles of same bonded phase functionality. Continue using the column in reverse flow direction. 4) Adjust solvent. Whenever possible, inject samples in mobile phase.
  • 17. Problem No. 8: Tailing Peaks
  • 18. • Probable Cause 1) Guard or analytical column contaminated/worn out. 2) Mobile phase contaminated/deteriorated. 3) Interfering components in sample. • Remedy/Comments 1) Remove guard column (if present) and attempt analysis. Replace guard column if necessary. If analytical column is source of problem, use appropriate restoration procedure (Table 1). If problem persists, replace column. 2) Check make-up of mobile phase. 3) Check column performance with standards.
  • 19. Problem No. 9: Fronting Peaks
  • 20. • Probable Cause 1) Column overloaded. 2) Sample solvent incompatible with mobile phase. 3) Shoulder or gradual baseline rise before a main peak may be another sample component. • Remedy/Comments 1) Inject smaller volume (e.g., 10 μL vs. 100 μL). Dilute the sample 1:10 or 1:100 fold in case of mass overload. 2) Adjust solvent. Whenever possible, inject samples in mobile phase. Flush polar bonded phase column with 50 column volumes HPLC grade ethyl acetate at 2-3 times the standard flow rate, then with intermediate polarity solvent prior to analysis. 3) Increase efficiency or change selectivity of system to improve resolution. Try another column type if necessary (e.g., switch from nonpolar C18 to polar cyano phase).
  • 21. Problem No. 10: Baseline Noise (regular)
  • 22. • Probable Cause 1) Air in mobile phase, detector cell, or pump. 2) Pump pulsations. 3) Incomplete mobile phase mixing. 4) Temperature effect (column at high temperature, detector unheated). 5) Other electronic equipment on same line. 6) Leak. • Remedy/Comments 1) Degas mobile phase. Flush system to remove air from detector cell or pump. 2) Incorporate pulse damper into system. 3) Mix mobile phase by hand or use less viscous solvent. 4) Reduce differential or add heat exchanger. 5) Isolate LC, detector, recorder to determine if source of problem is external. Correct as necessary. 6) Check system for loose fittings. Check pump for leaks, salt buildup, unusual noises. Change pump seals if necessary.
  • 23. Problem No. 11: Negative Peak
  • 24. • Probable Cause 1) Recorder leads reversed. 2) Refractive index of solute less than that of mobile phase (RI detector). 3) Sample solvent and mobile phase differ greatly in composition (vacancy peaks). 4) Mobile phase more absorptive than sample components to UV wavelength. • Remedy/Comments 1) Check polarity. 2) Use mobile phase with lower refractive index, or reverse recorder leads. 3) Adjust or change sample solvent. Dilute sample in mobile phase whenever possible. 4) a. Change polarity when using indirect UV detection, or b. Change UV wavelength or use mobile phase that does not adsorb chosen wavelength.
  • 25. Problem No. 12: Ghost Peak
  • 26. • Probable Cause 1) Contamination in injector or column. 2) Late eluting peak (usually broad) present in sample. • Remedy/Comments 1) Flush injector between analyses (a good routine practice). If necessary, run strong solvent through column to remove late eluters. Include final wash step in gradient analyses, to remove strongly retained compounds. 2) a. Check sample preparation. b. Include (step) gradient to quickly elute component.
  • 27. Table-1 • Table-1. Column Restoration Procedures • Silica-Based Reversed Phase Column (alkyl (C8, C18, etc.), phenyl, or diphenyl column, SUPELCOSIL LC-PAH column) •A. Water Soluble Samples Flush with the following: • Flush with warm (60 °C) distilled water • 50 mL methanol • 50 mL acetonitrile • 25 mL methanol • 25 mL mobile phase • Evaluate column performance.
  • 28. •B. Samples Not Soluble in Water Flush with the following: • 50 mL 2-propanol • 50 mL methylene chloride • 50 mL hexane • 25 mL isopropanol • 25 mL mobile phase • Evaluate column performance.
  • 29.