PCI & AimRadial 2018 | FFR using 4Fr catheters: Can Slender Technique Work Here? - Ian C Gilchrist
1. FFR using 4Fr catheters: Can
Slender Technique Work Here?
Ian C Gilchrist, MD, FACC, FSCAI
Professor of Medicine
Penn State’s Hershey Medical Center
Heart & Vascular Institute
Hershey, PA
USA
2. icg2018
4 French Feasibility
Yadav, Gilchrist. JACC 2014;64(11_S)
Pop Test
Pulse-Duplicator
Area 4-F catheter = 3.14 mm2
Area 0.014” wire = 0.10 mm2
Mean difference = 0.006 [-0.052-0.064]
Legalery, et al. Am J Cardiol 2003;91:1075–1078
4F Diagnostic vs 7F Guide FFR
∆(4FFFR-7FrFFR)
µFFR
3. icg2018
4F vs 6F Diagnostic Catheter FFR
61 patients
4F FFR
FFR 0.70-0.90
Cut point 0.80
6F FFR
Paired Comparison
Moriyama N, et al. COFFEE Trial. JACC I 2018;;11:412-14
4. icg2018
Example: Taking it a Down a Notch
50 year old woman with recurrent atypical pain
RCA stent last year via left radial out of town
Stress test was equivocal
Small, light framed woman
Definitely wants a wrist approach again
Right radial
5F sheath
Option for intervention
icg 2017
5. icg2018
0.035” wire stopped at sheath
Right forearm angiogram via sheath
High origin of radial from brachial artery
High-branching radial
& spasm
13. icg2018
4-F Technique
Less forgiving than larger French
Use fresh catheter & rapidly introduce wire
Meticulous FFR/iFR technique needed
• Confirm lumen is well flushed
• Transducer or catheter is not-damped
• Dicrotic notches on tracing?
Lack of immediate interventional
capacity may temper patient choice
14. icg2018
Improvements Needed
Better understanding of wire/internal catheter
interface to improve coeffiecent of friction
4-F guides or equivalent (FFR catheters?)
Potential for smaller catheters if sensor also
placed on distal catheter
Goal: Use of the smallest diameter access into the radial artery required
for diagnosis will maximize long-term radial artery function