2. 2 Fluke Calibration Fluke 5790B Extended Specifications
General Specifications
Warm-up Time ............................................ 30 minutes
Relative Humidity
Operating ................................................ 45 % to 50 °C 75 % to 45 °C 95 % to 30 °C
Storage.................................................... <95 % non-condensing
Altitude
Operating ................................................ 3,050 meters (10,000 feet)
Non-Operating ........................................ 12,200 meters (40,000 feet)
Temperature
Operating ................................................ 0 °C to 50 °C
Calibration .............................................. 15 °C to 35 °C
Storage.................................................... -40 °C to 70 °C
Electromagnetic Compatibility (EMC)
International ........................................... IEC 61326-1: Controlled Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is
necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments
other than domestic and those directly connected to a low-
voltage power supply network that supplies buildings used
for domestic purposes. There may be potential difficulties in
ensuring electromagnetic compatibility in other environments
due to conducted and radiated disturbances.
Emissions that exceed the levels required by CISPR 11 can
occur when the equipment is connected to a test object.
Korea (KCC) ............................................. Class A Equipment (Industrial Broadcasting & Communication
Equipment)
Class A: Equipment meets requirements for industrial
electromagnetic wave equipment and the seller or user
should take notice of it. This equipment is intended for use in
business environments and not to be used in homes.
USA (FCC) ................................................ 47 CFR 15 subpart B. This product is considered an exempt device
per clause 15.103.
Surge........................................................... ANSI C62.41-1980, Category A
Reliability.................................................... MIL-T-2880D, paragraph 3.13.3
Size
Height ..................................................... 17.8 cm (7 in) standard rackmount + 1.5 cm (0.6 in)
Width ...................................................... 43.2 cm (17 in)
Depth....................................................... 63 cm (24.8 in)
Maximum Power Requirements
5790B...................................................... 95 VA
With Wideband Option............................ 120 VA
Weight
5790B...................................................... 24 kg (53 lb)
With Wideband ....................................... 24.5 kg (54 lb)
Line Power .................................................. 50 Hz/60 Hz; 100 V - 120 V, 220 V - 240 V
Safety .......................................................... IEC 61010-1: Overvoltage Category II, Pollution Degree 2
IEC 61010-2-030: Measurement 1000 V CAT O
Remote Interfaces ....................................... RS-232, IEEE-488
Confidence Level......................................... 99 % unless otherwise specified.
DC Zero Cal.................................................. Perform the dc zero calibration every 30 days. In addition, perform
the dc zero calibration after powering up the unit the first time
after unpacking following a shipment or if exposed to an
environmental change of greater than 5 °C.
3. 3 Fluke Calibration Fluke 5790B Extended Specifications
Resolution and Range Limits
Voltage Range
Autorange Limits
[1]
Resolution
Upper Lower Filter Fast Filter Med/Slow
2.2 mV 2.2 mV 600 μV 0.1 μV 0.1 μV
7 mV 7 mV 1.9 mV 0.1 μV 0.1 μV
22 mV 22 mV 6 mV 0.1 μV 0.1 μV
70 mV 70 mV 19 mV 0.1 μV 0.1 μV
220 mV 220 mV 60 mV 0.1 μV 0.1 μV
700 mV 700 mV 190 mV 1.0 μV 0.1 μV
2.2 V 2.2 V 600 mV 1.0 μV 0.1 μV
7 V 7 V 1.9 V 10 μV 1.0 μV
22 V 22 V 6 V 10 μV 1.0 μV
70 V 70 V 19 V 100 μV 10 μV
220 V 220 V 60 V 100 μV 10 μV
700 V 700 V 190 V 1.0 mV 100 μV
1000 V 1050 V 600 V 1.0 mV 100 μV
[1] In locked ranges, readings may be made approximately 1 % beyond the autorange limits.
Electrical Specifications
The Product specifications describe the Absolute Instrumental Uncertainty of the Product. The Product
specifications include stability, temperature, and humidity; within specified limits, linearity, line and load
regulation, and the reference standard measurement uncertainty. The Product specifications are provided at a 99
% confidence level, k=2.58, normally distributed, unless otherwise stated.
The relative specifications are provided for enhanced accuracy applications. To calculate an enhanced absolute
specification from the relative specification, it is necessary to combine the uncertainty of your external standards
with the pertinent relative specifications. Specifications are valid after allowing a warm-up period of 30 minutes,
or twice the time the Product has been turned off.
AC Absolute Voltage
Voltage
Range
Frequency Range
Absolute ±5 °C of Calibration Temperature
AC/DC Transfer
Mode ± ppm
2 Years
Measurement Mode ± (ppm of Reading + μV)
90 Days 1 Year 2 Years
2.2 mV
10 Hz - 20 Hz
20 Hz- 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
1700 + 1.3
740 + 1.3
420 + 1.3
810 + 2.0
1200 + 2.5
2300 + 4.0
2400 + 6.0
3200 + 6.0
1700 + 1.3
740 + 1.3
420 + 1.3
810 + 2.0
1200 + 2.5
2300 + 4.0
2400 + 8.0
3500 + 8.0
1700 + 1.3
740 + 1.3
420 + 1.3
820 + 2.0
1200 + 2.5
2300 + 4.0
2600 + 8.0
5000 + 8.0
7 mV
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
850 + 1.3
370 + 1.3
210 + 1.3
400 + 2.0
600 + 2.5
1200 + 4.0
1300 + 6.0
2000 + 6.0
850 + 1.3
370 + 1.3
210 + 1.3
400 + 2.0
600 + 2.5
1200 + 4.0
1300 + 8.0
2300 + 8.0
850 + 1.3
370 + 1.3
210 + 1.3
410 + 2.0
610 + 2.5
1200 + 4.0
1400 + 8.0
3600 + 8.0
22 mV
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
290 + 1.3
180 + 1.3
110 + 1.3
210 + 2.0
310 + 2.5
810 + 4.0
860 + 6.0
1400 + 6.0
290 + 1.3
190 + 1.3
110 + 1.3
210 + 2.0
310 + 2.5
810 + 4.0
890 + 8.0
1700 + 8.0
290 + 1.3
190 + 1.3
110 + 1.3
210 + 2.0
310 + 2.5
820 + 4.0
1000 + 8.0
2600 + 8.0
70 mV
10 Hz - 20 Hz
[1]
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
240 + 1.5
120 + 1.5
64 + 1.5
120 + 2.0
260 + 2.5
510 + 4.0
660 + 6.0
1100 + 6.0
240 + 1.5
120 + 1.5
65 + 1.5
130 + 2.0
260 + 2.5
510 + 4.0
670 + 8.0
1100 + 8.0
240 + 1.5
130 + 1.5
69 + 1.5
130 + 2.0
260 + 2.5
530 + 4.0
680 + 8.0
1300 + 8.0
8. 8 Fluke Calibration Fluke 5790B Extended Specifications
DC Absolute Voltage
Voltage Range
Absolute ± 5 °C of Calibration Temperature
Measurement Mode ± (ppm of Reading + μV)
90 Days 1 Year 2 Years
220 mV 37 + 1.5 38 + 1.5 43 + 1.5
700 mV 31 + 1.5 33 + 1.5 38 + 1.5
2.2 V 22 24 29
7 V 22 24 29
22 V 25 27 31
70 V 30 32 39
220 V 29 31 38
700 V 39 41 47
1000 V 37 38 44
Note: DC specification valid only when dc input signal is averaged with an equal and opposite dc input signal to eliminate dc
offset errors. The use of Input 1 for dc inputs is not recommended due to the inherent thermal EMFs in a “N”
connector. See Operators Manual for details.
DC Relative Voltage
Voltage Range
Relative ± 5 °C of Calibration Temperature
Measurement Mode ± (ppm of Reading + μV)
90 Days 1 Year 2 Years
220 mV 22 + 1.5 24 + 1.5 31 + 1.5
700 mV 22 + 1.5 24 + 1.5 31 + 1.5
2.2 V 20 22 27
7 V 20 22 27
22 V 20 22 27
70 V 23 25 34
220 V 23 25 34
700 V 24 25 31
1000 V 24 25 31
Note: DC specification valid only when dc input signal is averaged with an equal and opposite dc input signal to eliminate dc
offset errors. The use of Input 1 for dc inputs is not recommended due to the inherent thermal EMFs in a “N”
connector. See Operators Manual for details.
10. 10 Fluke Calibration Fluke 5790B Extended Specifications
Voltage
Range Frequency Range
24 Hour AC
Stability ± 1 °C
Slow Filter Peak-
Peak ± μV
Temperature Coefficient
[1]
Input
Resistance
[2]10 °C to 40 °C
0 °C to 10 °C
40 °C to 50 °C
ppm / °C
7 V
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 5 1.5 3.0
40 Hz - 20 kHz 5 1.5 3.0
20 kHz - 50 kHz 5 2.0 3.0 50 kΩ
50 kHz - 100 kHz 10 5.0 8.0
100 kHz - 300 kHz 18 15.0 15.0
300 kHz - 500 kHz 30 30.0 30.0
500 kHz - 1 MHz 90 65.0 65.0
22 V
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 5 1.5 3.0
40 Hz - 20 kHz 5 1.5 3.0
20 kHz - 50 kHz 5 2.0 3.0 50 kΩ
50 kHz - 100 kHz 10 5.0 8.0
100 kHz - 300 kHz 18 15.0 15.0
300 kHz - 500 kHz 30 30.0 30.0
500 kHz - 1 MHz 90 65.0 65.0
70 V
[3]
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 5 1.5 3.0
40 Hz - 20 kHz 5 1.5 3.0
20 kHz - 50 kHz 5 2.0 3.0 50 kΩ
50 kHz - 100 kHz 18 5.0 8.0
100 kHz - 300 kHz 36 15.0 15.0
300 kHz - 500 kHz 48 40.0 40.0
500 kHz - 1 MHz 120 75.0 75.0
220 V
[3]
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 5 1.5 3.0
40 Hz - 20 kHz 5 1.5 3.0
20 kHz - 50 kHz 5 2.0 3.0 50 kΩ
50 kHz - 100 kHz 18 5.0 8.0
100 kHz - 300 kHz 36 15.0 15.0
300 kHz - 500 kHz 48 40.0 40.0
700 V
10 Hz - 20 Hz
[4]
8 1.5 4.0
20 Hz - 40 Hz 5 1.5 4.0
40 Hz – 20 kHz 5 1.5 4.0 500 kΩ
20 kHz – 50 kHz 18 5.0 7.0
50 kHz - 100 kHz 36 15.0 15.0
1000 V
10 Hz - 20 Hz
[4]
8 1.5 4.0
20 Hz - 40 Hz 5 1.5 4.0
40 Hz - 20 kHz 5 1.5 4.0 500 kΩ
20 kHz - 50 kHz
[5]
18 5.0 7.0
50 kHz - 100 kHz
[5]
36 15.0 15.0
[1] Add to uncertainty when more than 5 °C from calibration temperature.
[2] Input capacitance approximately 100 pF.
[3] Inputs with a V*Hz product >2.2 E7 are unspecified.
[4] Typical specification, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.
[5] Inputs that are >30 kHz and >750 V are typical, as determined by sourcing with the Fluke 5205A Precision Power
Amplifier.
11. 11 Fluke Calibration Fluke 5790B Extended Specifications
DC Secondary Performance
Voltage Range
Temperature Coefficient
[1]
Input
Resistance
[2]10 °C to 40 °C
0 °C to 10 °C
40 °C to 50 °C
ppm / °C
220 mV 1.5 3.0 10 MΩ
700 mV 1.5 3.0 10 MΩ
2.2 V 1.5 3.0 10 MΩ
7 V 1.5 3.0 50 kΩ
22 V 1.5 3.0 50 kΩ
70 V 1.5 3.0 50 kΩ
220 V 1.5 3.0 50 kΩ
700 V 1.5 4.0 500 kΩ
1000 V 1.5 4.0 500 kΩ
[1] Add to uncertainty when more than 5 °C from calibration temperature.
[2] Input capacitance approximately 100 pF.
Note: DC specification valid only when dc input signal is averaged with an equal and opposite dc input signal to eliminate dc
offset errors. The use of Input 1 for dc inputs is not recommended due to the inherent thermal EMFs in a “N” connector.
See Operators Manual for details.
12. 12 Fluke Calibration Fluke 5790B Extended Specifications
Operating Characteristics
Maximum Non-destructive Input..................... 1200 V rms
Guard Isolation................................................ 10 V peak
Volt-Hertz Product .......................................... 1 x 10
8
Frequency Accuracy (from 0 °C to 50 °C)
10 Hz - 120 Hz ............................................ 100 ppm + 10 digits
Above 120 Hz.............................................. 100 ppm + 2 digits
Frequency Resolution ..................................... 1.00 Hz to 119.99 Hz
0.1200 kHz to 1.1999 kHz
1.200 kHz to 11.999 kHz
12.00 kHz to 119.99 kHz
0.1200 MHz to 1.0000 MHz
1.0000 MHz to 1.1999 MHz (Wideband only)
1.200 MHz to 11.999 MHz (Wideband only)
12.00 MHz to 30.00 MHz (Wideband only)
30.00 MHz to 50.00 MHz (Wideband 5790B/5 & 5790B/AF only)
Reading Rate
<40 Hz ........................................................ 2 seconds per reading
40 Hz........................................................... 2 seconds decreasing linearly to 1 second at 200 Hz
>200 Hz ...................................................... 1 second per reading
Maximum Settling Time to Full Specifications (in range lock)
Filter Off ...................................................... 1 sample
dc ............................................................ 6 seconds
<200 Hz .................................................. 8 seconds
>200 Hz .................................................. 4 seconds
Filter Fast .................................................... 4 averaged samples
dc ............................................................ 10 seconds
<200 Hz .................................................. 16 seconds
>200 Hz .................................................. 8 seconds
Filter Medium.............................................. 16 averaged samples
dc ............................................................ 22 seconds
<200 Hz .................................................. 32 seconds
>200 Hz .................................................. 16 seconds
Filter Slow................................................... 32 averaged samples
dc ............................................................ 40 seconds
<200 Hz .................................................. 64 seconds
>200 Hz .................................................. 32 seconds
Filter Buffer Restart Limits:
Fine: Fast: 10 counts
Medium/Slow
<220 mV ............................................. 10 counts
>220 mV ............................................. 100 counts
Medium: Fast: 100 counts
Medium/Slow
<220 mV ............................................. 100 counts
>220 mV ............................................. 1000 counts
Course: Fast: 1000 counts
Medium/Slow
<220 mV ............................................. 1000 counts
>220 mV ............................................. 10000 counts
Specified for sinewave with THD less than 1 %
AUX Input Characteristics
The AUX input can be used with the Fluke A40/A40A Series Current Shunts to make relative ac current
measurements. The 5790A-7001 A40/A40A Current Shunt Adapter and Cable are required. For optimal current
measurements using shunts, see the Operators Manual.
Input Resistance.......................................... 91 Ω ± 1 %
Operating Input Voltage .............................. 250 mV to 500 mV
Maximum Non-Destructive Input................. 50 V rms