HPLC Method and Validation 
basics 
Regulatory guidelines— 
Shreekant Deshpande 
Senior Scientist 
Eutech Sci Ser Inc
Outline 
• HPLC methodology 
Content of HPLC test procedure 
System Suitability Testing (SST) 
Relative Response Factor (RRF) 
• Validation of HPLC method 
• Case study
Information Sources 
• FDA CDER reviewer guideline for validation of 
chromatographic methods (1994) 
• WHO TRS 937 Appendix 4 “Analytical Method Validation 
2006 
• ICH Q2(R1) 2005 
• Compendial General Chapters 
• Methods and Validation presentation–Lynda Paleshnuik 
• Methods and Validation basics–Hua Yin
High Performance Liquid Chromatography (HPLC) 
HPLC is a separation technique based on a solid stationary phase 
and a liquid mobile phase. Separations are achieved by partition, 
adsorption, or ion-exchange processes, depending upon the type of 
stationary phase used. 
• Chiral 
• Ion--exchange 
• Ion--pair/affinity 
• Normal phase 
• Reversed phase 
• Size exclusion 
The reversed-phase HPLC with UV detection is most commonly 
used form of HPLC, is selected to illustrate the parameters of 
HPLC method and validation.
A flow scheme for HPLC
Content of HPLC test procedure 
Any analytical procedure submitted should be described in 
sufficient detail, includes: 
• Materials/Chemicals 
• Preparation of mobile phase 
• Chromatographic condition: 
• Column: type (e.g., C18 or C8), dimension (length, inner diameter), 
particle size (10μm, 5 μm) 
• Detector: wavelength 
• Injection volume 
• column Temp 
• flow rate,
Content of HPLC test procedure 
• Elution procedure: isocratic or gradient elution 
• Preparation of standards and samples 
• Operation procedure: sequence of injections 
• System suitability testing (SST) and criteria 
• Calculations 
QOS 2.3.R.2 analytical procedures and validation summaries
Compendial methods 
When claim a compendial method, there should be no change in: 
• The type of column i.e the stationary phases 
• Detector wavelength 
• Components in Mobile phase 
• System suitability testing and criteria 
Adjustments to ratio of components in mobile phase, flow rate, 
column temp, dimension of column, particle size (reduction only), 
may be necessary to achieve the system suitability criteria. 
The allowable variations for each parameter, see Int.Ph 1.14.4 or 
USP general chapter <621>.
System suitability testing (SST) 
• Precision: 
• Assay: RSD ≤1% (API) or ≤ 2% (FPP), n ≥ 5 
• Impurities: in general, RSD ≤ 5% at the limit level, up to 10% or 
higher at LOQ, n ≥ 6 
• Resolution (R): >2
System suitability testing (SST) 
• Tailing factor/peak asymmetry: (≤ 2)
System suitability testing (SST) 
• Number of theoretical plates (N): column efficiency ≥ 2000 
• Gradient elution is one way to increase the N
System suitability testing (SST) 
A SST should contain: 
• For Assay: 
precision + one or more other parameter 
• For impurity test: 
resolution + precision + one or more other parameter
Relative Response Factor (RRF) 
Quantitation of Impurities/ derivatives 
• Where the loss of analyte is inevitable, Use IS and RRF! 
• Against impurity RS’s: when reference standard available 
• Against API itself 
Relative response factor should be considered
Relative Response Factor (RRF) 
• Response factor: the response (e.g. peak area) of drug 
substance or related substances per unit weight. 
RF= peak area / concentration (mg/ml) 
• Relative response factor (RRF): 
RRF=RF impurity / RF API, OR, 
RRF=slope impurity / slope API
Relative Response Factor (RRF) 
Rifampicine: 
y =31.312 x + 4.963 
Rifampicine Quinone: 
y = 26.198 x + 1.154 
RRF= 26.198 / 31.312 
=0.84
Relative Response Factor (RRF) 
• To review: 
a) RRF calculation, and 
b) if RRF is properly used in the final calculation for % 
impurity 
If RRF within 0.8-1.2, correction may not be necessay 
• Correction factor= 1/RRF, the reciprocal of the RRF
Review points for HPLC method 
• is the analytical procedure described in detail including all the parameters ? 
• is SST well defined to ensure the consistency of system performance? 
• The preparation of solutions: 
• assay: concentration of reference standard should be close to the sample 
solution 
• Quantitation: Sample concentration should fall under standard curve 
• impurities: concentration of the reference standards should be close to the limit 
• The way of quantitation of impurities 
In case API is used as the reference, RRF should be used or justification of 
exclusion should be provided. 
To check the determination of RRF, check the correction of calculation of 
impurities 
• confirm/complete the QOS 2.3.R.2
Validation – compendial methods 
Assay – API 
No validation generally required. Exception: specificity for major 
impurities not in the monograph. 
Assay – FPP 
Specificity, accuracy and precision (repeatability). 
Purity – API and FPP 
Full validation for specified impurities that are not included in the 
monograph (specificity, linearity, accuracy, repeatability, 
intermediate precision, LOD/LOQ) 
Validation of the limit for individual unknowns, if tighter than that in 
the monograph: LOQ of the API should be below the limit for 
individual unknowns
Non-compendial methods 
Full validation is required for purity, assay and dissolution methods 
(HPLC, UV) : 
Specificity 
Linearity 
Accuracy 
Repeatability 
Intermediate precision 
LOD/LOQ (not required for assay, dissolution) 
Robustness (recommended)
Specificity 
• Blank solution to show no interference 
• Placebo to demonstrate the lack of interference from excipients 
• Spiked samples to show that all known related substances are 
resolved from each other 
• Stressed sample of about 10 to 20% degradation is used to 
demonstrate the resolution among degradation products 
• Check peak purity of drug substance by photodiode array detector (PDA): eg 
purity angle is lower than the purity threshold. 
• Representative chromatograms should be provided with time scale 
and attenuation indicated
Linearity / Range 
• The working sample concentration and samples tested for 
accuracy should be in the linear range (concentrations Vs. 
Peak areas) 
• Minimum 5 concentrations 
• Dilute of stock solution or separate weighings
Linearity / Range 
• Assay : 80-120% of the theoretical content of active 
• Content Uniformity: 70-130% 
• Dissolution: ±20% of limits; eg if limits cover from 20% to 
90% l.c. (controlled release), linearity should cover 0-110% 
of l.c (Label Claim). 
• Impurities: reporting level to 120% of shelf life limit 
• Assay/Purity by a single method: reporting level of the 
impurities to 120% of assay limit
Linearity / Range 
Correlation coefficient (r) 
API: ≥ 0.998 
Impurities: ≥ 0.99 
y-Intercept and slope should be indicated together with plot of 
the data
Accuracy 
Assay 
API: against an RS of known purity, or via an alternate method of 
known accuracy; analysis in triplicate. 
FPP: samples/placeboes spiked with API, across the range of 80- 
120% of the target concentration, 3 concentrations, in triplicate 
each. 
Report per cent recovery (mean result and RSD): 100±2% 
ICH Q2 states: accuracy may be inferred once precision, linearity 
and specificity have been established.
Accuracy 
Impurities: API/FPP spiked with known impurities 
Experienced in PQ: 
Across the range of LOQ to150% of the target concentration 
(shelf life limit), 3-5 concentrations, in triplicate each. (LOQ, 
50%, 100%, 150%) 
Per cent recovery: in general, within 80-120%, depends on 
the level of limit
Precision 
• System precision: 
• by multiple injections (n ≥5) of a homogeneous sample 
(standard solution). 
• RSD ≤ 1% is recommended for assay; 
• RSD ≤ 5% is recommended for related substances 
(reference standards at the limit) 
• Indicates the performance of the HPLC system 
• As a system suitability test
Precision 
• Repeatability (method precision) 
• Multiple measurements of a sample by the same analyst 
• A minimum of 6 determinations at the test concentration (6 
times of a single batch), or 
• 3 levels (80%, 100%, 120%) , 3 repetitions each (combined 
with accuracy) 
• For Assay: RSD ≤ 2.0% 
• For individual impurity above 0.05%, in general, RSD ≤ 10%
Precision 
• Intermediate precision (part of ruggedness) 
• Test a sample on multiple days, analysts, equipments 
• Repeat the method precision by different analyst in different 
equipment using different lot of column on different days 
• RSD should be the same requirement as method precision 
• Reproducibility (inter-laboratory trial) 
• Not requested in the submission
LOD/LOQ 
• signal to noise ratio: LOD 3:1 , LOQ 10:1 
• May vary with lamp aging, model/manufacturer of detector, column 
• standard deviation of the response and the slope of the calibration 
curve at levels approximating the LOD /LOQ 
σ = the standard deviation of the response, base on 
• the standard deviation of the blank 
• The calibration curve 
should be validated by analysis of samples at the limits.
LOD/LOQ 
• LOD: below the reporting threshold 
• LOQ: at or below the specified limit 
Not required for assay/dissolution methods. 
• Applicant should provide 
• the method of determination 
• the limits, 
• chromotograms
Robustness 
• The method's capability to remain unaffected by small but 
deliberate variations in method parameters 
• Influence of variations of pH in a mobile phase 
• Influence of variations in mobile phase composition 
• Different columns (different lots and/or suppliers) 
• Temperature 
• Flow rate 
• Evaluate the System suitability parameters
Robustness 
:Parameters 
Change in column temperature ± 5°C 
Change in flow rate ± 0. 2ml /min 
Change in mobile phase Buffer pH ± 0. 2units 
%Change in organic composition ± 2.0 
:Acceptance Criteria 
The system suitability parameters should pass for all the 
,conditions 
All known and Unknown impurities shall be separated from 
.each other; in sample spiked with impurities
Conclusion 
• HPLC methods play a critical role in analysis of 
pharmaceutical product 
• Validation of HPLC should demonstrate that the method is 
suitable for its intended use 
• Review the information in dossier against QOS 2.3.R.2 
• Data for acceptance, release, stability will only be 
trustworthy if the methods used are reliable
Thank you

Mais conteúdo relacionado

Mais procurados

Mais procurados (20)

Ion pair chromatography final
Ion pair chromatography finalIon pair chromatography final
Ion pair chromatography final
 
Human liver microsomes & rat liver microsomes
Human liver microsomes & rat liver microsomesHuman liver microsomes & rat liver microsomes
Human liver microsomes & rat liver microsomes
 
Analytical method development
Analytical method developmentAnalytical method development
Analytical method development
 
General considerations and method development in ce,
General considerations and method development in ce,General considerations and method development in ce,
General considerations and method development in ce,
 
Bioanalysis of drugs from biological samples
Bioanalysis of drugs from biological samplesBioanalysis of drugs from biological samples
Bioanalysis of drugs from biological samples
 
System suitability testing
System suitability testingSystem suitability testing
System suitability testing
 
Bioanlytical method development
Bioanlytical method developmentBioanlytical method development
Bioanlytical method development
 
IMPURITIES AND STABILITY STUDIES
IMPURITIES AND STABILITY STUDIESIMPURITIES AND STABILITY STUDIES
IMPURITIES AND STABILITY STUDIES
 
Ion pair chromatography for pharmacy students
Ion pair chromatography for pharmacy studentsIon pair chromatography for pharmacy students
Ion pair chromatography for pharmacy students
 
HPLC Method Development & Method Validation (mr.s)
HPLC Method Development & Method Validation (mr.s)HPLC Method Development & Method Validation (mr.s)
HPLC Method Development & Method Validation (mr.s)
 
Bioanalytical method validation
Bioanalytical method validationBioanalytical method validation
Bioanalytical method validation
 
method validation as per emea.pptx
method validation as per emea.pptxmethod validation as per emea.pptx
method validation as per emea.pptx
 
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
Analytical method- Content, Development, validation, Transfer & Life Cycle Ma...
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
ICH USFDA Validation Differences
ICH USFDA Validation DifferencesICH USFDA Validation Differences
ICH USFDA Validation Differences
 
Calibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical InstrumentsCalibration and Qualification of Analytical Instruments
Calibration and Qualification of Analytical Instruments
 
LC-MS in bioactivity screening and proteomics
LC-MS in bioactivity screening and proteomicsLC-MS in bioactivity screening and proteomics
LC-MS in bioactivity screening and proteomics
 
M10 bioanalytical method validation
M10 bioanalytical method validationM10 bioanalytical method validation
M10 bioanalytical method validation
 
System suitability parameters assessment by HPLC
System suitability parameters assessment by HPLCSystem suitability parameters assessment by HPLC
System suitability parameters assessment by HPLC
 
Potential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyserPotential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyser
 

Semelhante a Method-Validation-HPLC-case-study

Dissolution-Method validation _ppt_slide
Dissolution-Method validation _ppt_slideDissolution-Method validation _ppt_slide
Dissolution-Method validation _ppt_slide
Bhanu Prakash N
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Pratik434909
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Pratik434909
 
Assay-Method validation-PPT _slide
Assay-Method validation-PPT _slideAssay-Method validation-PPT _slide
Assay-Method validation-PPT _slide
Bhanu Prakash N
 
Related Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slideRelated Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slide
Bhanu Prakash N
 

Semelhante a Method-Validation-HPLC-case-study (20)

Analytical Method Validation.pptx
Analytical Method Validation.pptxAnalytical Method Validation.pptx
Analytical Method Validation.pptx
 
validation and calibration of HPLC
validation and calibration of HPLCvalidation and calibration of HPLC
validation and calibration of HPLC
 
Analytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation ParametersAnalytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation Parameters
 
Dissolution-Method validation _ppt_slide
Dissolution-Method validation _ppt_slideDissolution-Method validation _ppt_slide
Dissolution-Method validation _ppt_slide
 
Calibaration and validation of hplc
Calibaration and validation of hplcCalibaration and validation of hplc
Calibaration and validation of hplc
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
 
Development and Validation of a RP-HPLC method
Development and Validation of a RP-HPLC methodDevelopment and Validation of a RP-HPLC method
Development and Validation of a RP-HPLC method
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
2 4 method-validation-hplc-case-study
2 4 method-validation-hplc-case-study2 4 method-validation-hplc-case-study
2 4 method-validation-hplc-case-study
 
Hplc
HplcHplc
Hplc
 
HPLC validation.ppt
HPLC validation.pptHPLC validation.ppt
HPLC validation.ppt
 
hplc
 hplc hplc
hplc
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Assay-Method validation-PPT _slide
Assay-Method validation-PPT _slideAssay-Method validation-PPT _slide
Assay-Method validation-PPT _slide
 
Validation of Analytical method.ppt
Validation of Analytical method.pptValidation of Analytical method.ppt
Validation of Analytical method.ppt
 
Accuracy and precision presentation
Accuracy and precision presentationAccuracy and precision presentation
Accuracy and precision presentation
 
Related Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slideRelated Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slide
 
ICP QC protocol
ICP  QC  protocolICP  QC  protocol
ICP QC protocol
 
Quantitation techniques used in chromatography
Quantitation techniques used in chromatographyQuantitation techniques used in chromatography
Quantitation techniques used in chromatography
 

Último

College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
perfect solution
 

Último (20)

Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...Top Rated Bangalore Call Girls Richmond Circle ⟟  9332606886 ⟟ Call Me For Ge...
Top Rated Bangalore Call Girls Richmond Circle ⟟ 9332606886 ⟟ Call Me For Ge...
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ludhiana Just Call 9907093804 Top Class Call Girl Service Available
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
(👑VVIP ISHAAN ) Russian Call Girls Service Navi Mumbai🖕9920874524🖕Independent...
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 

Method-Validation-HPLC-case-study

  • 1. HPLC Method and Validation basics Regulatory guidelines— Shreekant Deshpande Senior Scientist Eutech Sci Ser Inc
  • 2. Outline • HPLC methodology Content of HPLC test procedure System Suitability Testing (SST) Relative Response Factor (RRF) • Validation of HPLC method • Case study
  • 3. Information Sources • FDA CDER reviewer guideline for validation of chromatographic methods (1994) • WHO TRS 937 Appendix 4 “Analytical Method Validation 2006 • ICH Q2(R1) 2005 • Compendial General Chapters • Methods and Validation presentation–Lynda Paleshnuik • Methods and Validation basics–Hua Yin
  • 4. High Performance Liquid Chromatography (HPLC) HPLC is a separation technique based on a solid stationary phase and a liquid mobile phase. Separations are achieved by partition, adsorption, or ion-exchange processes, depending upon the type of stationary phase used. • Chiral • Ion--exchange • Ion--pair/affinity • Normal phase • Reversed phase • Size exclusion The reversed-phase HPLC with UV detection is most commonly used form of HPLC, is selected to illustrate the parameters of HPLC method and validation.
  • 5. A flow scheme for HPLC
  • 6. Content of HPLC test procedure Any analytical procedure submitted should be described in sufficient detail, includes: • Materials/Chemicals • Preparation of mobile phase • Chromatographic condition: • Column: type (e.g., C18 or C8), dimension (length, inner diameter), particle size (10μm, 5 μm) • Detector: wavelength • Injection volume • column Temp • flow rate,
  • 7. Content of HPLC test procedure • Elution procedure: isocratic or gradient elution • Preparation of standards and samples • Operation procedure: sequence of injections • System suitability testing (SST) and criteria • Calculations QOS 2.3.R.2 analytical procedures and validation summaries
  • 8. Compendial methods When claim a compendial method, there should be no change in: • The type of column i.e the stationary phases • Detector wavelength • Components in Mobile phase • System suitability testing and criteria Adjustments to ratio of components in mobile phase, flow rate, column temp, dimension of column, particle size (reduction only), may be necessary to achieve the system suitability criteria. The allowable variations for each parameter, see Int.Ph 1.14.4 or USP general chapter <621>.
  • 9. System suitability testing (SST) • Precision: • Assay: RSD ≤1% (API) or ≤ 2% (FPP), n ≥ 5 • Impurities: in general, RSD ≤ 5% at the limit level, up to 10% or higher at LOQ, n ≥ 6 • Resolution (R): >2
  • 10. System suitability testing (SST) • Tailing factor/peak asymmetry: (≤ 2)
  • 11. System suitability testing (SST) • Number of theoretical plates (N): column efficiency ≥ 2000 • Gradient elution is one way to increase the N
  • 12. System suitability testing (SST) A SST should contain: • For Assay: precision + one or more other parameter • For impurity test: resolution + precision + one or more other parameter
  • 13. Relative Response Factor (RRF) Quantitation of Impurities/ derivatives • Where the loss of analyte is inevitable, Use IS and RRF! • Against impurity RS’s: when reference standard available • Against API itself Relative response factor should be considered
  • 14. Relative Response Factor (RRF) • Response factor: the response (e.g. peak area) of drug substance or related substances per unit weight. RF= peak area / concentration (mg/ml) • Relative response factor (RRF): RRF=RF impurity / RF API, OR, RRF=slope impurity / slope API
  • 15. Relative Response Factor (RRF) Rifampicine: y =31.312 x + 4.963 Rifampicine Quinone: y = 26.198 x + 1.154 RRF= 26.198 / 31.312 =0.84
  • 16. Relative Response Factor (RRF) • To review: a) RRF calculation, and b) if RRF is properly used in the final calculation for % impurity If RRF within 0.8-1.2, correction may not be necessay • Correction factor= 1/RRF, the reciprocal of the RRF
  • 17. Review points for HPLC method • is the analytical procedure described in detail including all the parameters ? • is SST well defined to ensure the consistency of system performance? • The preparation of solutions: • assay: concentration of reference standard should be close to the sample solution • Quantitation: Sample concentration should fall under standard curve • impurities: concentration of the reference standards should be close to the limit • The way of quantitation of impurities In case API is used as the reference, RRF should be used or justification of exclusion should be provided. To check the determination of RRF, check the correction of calculation of impurities • confirm/complete the QOS 2.3.R.2
  • 18. Validation – compendial methods Assay – API No validation generally required. Exception: specificity for major impurities not in the monograph. Assay – FPP Specificity, accuracy and precision (repeatability). Purity – API and FPP Full validation for specified impurities that are not included in the monograph (specificity, linearity, accuracy, repeatability, intermediate precision, LOD/LOQ) Validation of the limit for individual unknowns, if tighter than that in the monograph: LOQ of the API should be below the limit for individual unknowns
  • 19. Non-compendial methods Full validation is required for purity, assay and dissolution methods (HPLC, UV) : Specificity Linearity Accuracy Repeatability Intermediate precision LOD/LOQ (not required for assay, dissolution) Robustness (recommended)
  • 20. Specificity • Blank solution to show no interference • Placebo to demonstrate the lack of interference from excipients • Spiked samples to show that all known related substances are resolved from each other • Stressed sample of about 10 to 20% degradation is used to demonstrate the resolution among degradation products • Check peak purity of drug substance by photodiode array detector (PDA): eg purity angle is lower than the purity threshold. • Representative chromatograms should be provided with time scale and attenuation indicated
  • 21. Linearity / Range • The working sample concentration and samples tested for accuracy should be in the linear range (concentrations Vs. Peak areas) • Minimum 5 concentrations • Dilute of stock solution or separate weighings
  • 22. Linearity / Range • Assay : 80-120% of the theoretical content of active • Content Uniformity: 70-130% • Dissolution: ±20% of limits; eg if limits cover from 20% to 90% l.c. (controlled release), linearity should cover 0-110% of l.c (Label Claim). • Impurities: reporting level to 120% of shelf life limit • Assay/Purity by a single method: reporting level of the impurities to 120% of assay limit
  • 23. Linearity / Range Correlation coefficient (r) API: ≥ 0.998 Impurities: ≥ 0.99 y-Intercept and slope should be indicated together with plot of the data
  • 24. Accuracy Assay API: against an RS of known purity, or via an alternate method of known accuracy; analysis in triplicate. FPP: samples/placeboes spiked with API, across the range of 80- 120% of the target concentration, 3 concentrations, in triplicate each. Report per cent recovery (mean result and RSD): 100±2% ICH Q2 states: accuracy may be inferred once precision, linearity and specificity have been established.
  • 25. Accuracy Impurities: API/FPP spiked with known impurities Experienced in PQ: Across the range of LOQ to150% of the target concentration (shelf life limit), 3-5 concentrations, in triplicate each. (LOQ, 50%, 100%, 150%) Per cent recovery: in general, within 80-120%, depends on the level of limit
  • 26. Precision • System precision: • by multiple injections (n ≥5) of a homogeneous sample (standard solution). • RSD ≤ 1% is recommended for assay; • RSD ≤ 5% is recommended for related substances (reference standards at the limit) • Indicates the performance of the HPLC system • As a system suitability test
  • 27. Precision • Repeatability (method precision) • Multiple measurements of a sample by the same analyst • A minimum of 6 determinations at the test concentration (6 times of a single batch), or • 3 levels (80%, 100%, 120%) , 3 repetitions each (combined with accuracy) • For Assay: RSD ≤ 2.0% • For individual impurity above 0.05%, in general, RSD ≤ 10%
  • 28. Precision • Intermediate precision (part of ruggedness) • Test a sample on multiple days, analysts, equipments • Repeat the method precision by different analyst in different equipment using different lot of column on different days • RSD should be the same requirement as method precision • Reproducibility (inter-laboratory trial) • Not requested in the submission
  • 29. LOD/LOQ • signal to noise ratio: LOD 3:1 , LOQ 10:1 • May vary with lamp aging, model/manufacturer of detector, column • standard deviation of the response and the slope of the calibration curve at levels approximating the LOD /LOQ σ = the standard deviation of the response, base on • the standard deviation of the blank • The calibration curve should be validated by analysis of samples at the limits.
  • 30. LOD/LOQ • LOD: below the reporting threshold • LOQ: at or below the specified limit Not required for assay/dissolution methods. • Applicant should provide • the method of determination • the limits, • chromotograms
  • 31. Robustness • The method's capability to remain unaffected by small but deliberate variations in method parameters • Influence of variations of pH in a mobile phase • Influence of variations in mobile phase composition • Different columns (different lots and/or suppliers) • Temperature • Flow rate • Evaluate the System suitability parameters
  • 32. Robustness :Parameters Change in column temperature ± 5°C Change in flow rate ± 0. 2ml /min Change in mobile phase Buffer pH ± 0. 2units %Change in organic composition ± 2.0 :Acceptance Criteria The system suitability parameters should pass for all the ,conditions All known and Unknown impurities shall be separated from .each other; in sample spiked with impurities
  • 33. Conclusion • HPLC methods play a critical role in analysis of pharmaceutical product • Validation of HPLC should demonstrate that the method is suitable for its intended use • Review the information in dossier against QOS 2.3.R.2 • Data for acceptance, release, stability will only be trustworthy if the methods used are reliable

Notas do Editor

  1. TRS: Technical Report series CDER: Center for Drug Evaluation and Research
  2. QOS 2.3.R.2 : Quality Overall Summery 2.3