2. AIMSAIMS
This study is for imaging aortic
atherosclerotic plaques in Apo E deficient
mice
In addition to study of Apo E
atherosclerotic mice this effort is meant to
provide necessary technical experiences for
rabbit study soon to be approved and
transferred to Galveston
3.
4. MRIMRI
withwith
Contrast (Magnevist)Contrast (Magnevist)
In order to be sure that we are looking to
the aorta,we decided to inject contrast
media
We had to inject Magnevist to mouse while
inside the magnet, so we tried to cannulate
femoral vein, jugular vein, intracardiac
approach
5. In order to decrease the motion effect of
respiration,we had to use respiratory gated
MRI
To have respiratory gating,we needed to
Intubate the mouse and use ventilation
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29. The MR Image characteristicsThe MR Image characteristics
Coil:5 and 9cm
Field of view:6 cm
Matrix(nv):256
Spin Echo:Tr=2500,Te=12, Slice=17-24
Movi(while injecting the Gadelenium) :
Tr=200, Te= 5
54. SPIO InjectionSPIO Injection
With injection of Feridex with maximum
dose(3µmol/kg),we had some problems:
because of small blood content of mouse,
dilution wasn’t enough
It caused too much decrease in T2 , so
although we had approximately good
resolution, we think that by decreasing the
dose we can find better images
55. Future PlansFuture Plans
Injection of 40µmol/kg of Feridex
ECG gated MRI (we have the motion effect
of heart in chest region)
Starting rabbit study