NexWave

User’s Manual
1
NexWave

Table of Contents
Contact Information / Customer Service / Supplies / Support		

4

About the NexWave						5
Safety Information						6
Electrodes and Lead Wires Setup					9
Programming Instructions					12
Preprogrammed Modes						13
Controls								15
Operating Instructions						16
Protocols							17
Indications, Contraindications and Warnings			

31

Precautions and Adverse Reactions				

40

Troubleshooting							41
Specifications and Accessories					42
Electrodes, Batteries, and Cleaning					43
Warranty							44

3
Zynex® Medical
Contact Information

CUSTOMER SERVICE					866-940-7030
Supplies:		

To order more electrodes or other 	accessories

Technical Support:	

Questions or problems with using your device

Device Return:		
			

Order a postage paid return envelope to return your
device at no charge

MAIN OFFICE						800-495-6670
303-703-4906
Billing Questions:	
			
			
			

Questions regarding insurance benefits and
covered benefits for Durable Medical Equipment
(DME) or questions about an Explanation of
Benefits form you received in the mail.

FAX NUMBER						800-495-6695	
303-347-9153
MAILING ADDRESS
Zynex Medical, Inc.
9990 Park Meadows Drive
Lone Tree, CO 80124-6739
USA
EMAIL						info@zynex.com
WEBSITE					www.zynex.com
European Representative
Zynex Europe
Fælledvej 1, Box 19 BDO
DK-5100 Odense C
Denmark

4

0470
About the NexWave
The NexWave is a multiple-mode stimulator which allows users a choice of
treatment options. This device incorporates Interferential Current (IFC), Transcutaneous Electrical Nerve Stimulation (TENS), and Neuromuscular Electrical
Stimulation (NMES).
Interferential Current (IFC)
The left channel (channel 1) produces a signal at a frequency of 4000 Hz. A
slightly higher frequency (4001-4128 Hz) is produced by the right channel
(channel 2). When applied to the skin, the signals from the two channels “interfere” with each other, and create a difference frequency which is felt during
treatment. The skin is 100 times less resistant to electrical stimulation using
this method. Consequently more of the stimulation energy reaches the nerve
and muscle fibers and hence becomes more productive. With more stimulation reaching the affected area, IFC is recommended to be applied 3-4 times
a day for 30-40 minutes per treatment. IFC therapy requires the use of both
channels (four electrodes).
Transcutaneous Electrical Nerve Stimulation (TENS)
TENS delivers electrical impulses through the skin in the range of 1-125 Hz.
TENS treatments can range from 10 minutes to many hours and can be applied as needed for pain relief. Patients can use one or two channels (two or
four electrodes) for TENS therapy.
Neuromuscular Electrical Stimulation (NMES)
NMES delivers electrical impulses to muscular motor points at a frequency of
35 Hz. The output is set to a level which causes muscular contraction. These
impulses are turned on for a period of 10 seconds, and turned off for a period
of 10 seconds, 20 seconds, or 30 seconds. The outputs ramp up for 3 seconds,
remain at the desired level for 10 seconds, ramp down for 1 second, and
remain off for 10, 20, or 30 seconds. Recommended use is 3-4 times per day
for 20-30 minutes per treatment. Patients can use one or two channels (two or
four electrodes) for NMES therapy.
Important: This device must be ordered or prescribed by a licensed physician.

5
Safety Information
Equipment Classification
As per the International Electrotechnical Commission standard IEC
60601-1, and the European standard EN 60601-1, Medical Electrical
Equipment, General Requirements for Basic Safety and Essential Performance, the NexWave is classified as follows:

· Type BF Equipment

A Type BF piece of equipment is one that provides a particular
degree of protection against electric shock, particularly regarding allowable leakage current. Type BF applied part is one that is
floating above ground and is isolated from all other parts of the
equipment.

· Medical Device Directive (MDD) Classification

The NexWave a Class IIa medical device per the European Medical
Device Directive.

· Internally Powered Equipment

When the NexWave is powered by the internal, 9 VDC, MN1604,
battery it is classified as Internally Powered Equipment.

· Class II Medical Equipment

When powered by the external, 12 VDC, medical grade, power
supply, the NexWave is classified as Class II Medical Electrical (ME)
Equipment. Note: When the external power supply is connected
to the NexWave device, the power supply is considered part of
the ME equipment.

· Electromagnetic Compatibility
Conforms to IEC 60601-1-2.

· Temperature

Operating temperature range: 0° - 40° C (32° - 104° F)
Storage temperature range: -20° - 70° C (-4° - 158° F)

· Water Ingress

Ordinary equipment. This device does not have protection
against ingress of water.

· Flammable Anesthetics

This device is not suitable for use in the presence of a flammable
anesthetic mixture with air, or in the presence of a flammable
anesthetic mixture with oxygen or nitrous oxide.

6
Safety Information
Mode of Operation
This device is suitable for continuous operation.

Symbols

Safety symbols shown on the device above are defined below.
On/Off. This symbol indicates that the labeled switch electronically cycles the DC power on and off for part of the equipment.
Note: To disconnect the external power supply, unplug the
power cord of the supplied AC adapter from the AC mains outlet
General Warning Sign. Follow warnings stated in the instruction manual to prevent potential hazards.
Refer to Instruction Manual/Booklet. The operator must read,
understand, and follow all instructions in the accompanying
document including all warnings, cautions, and precautions
before using this medical device.
Type BF Equipment. This symbol indicates that the patient
applied parts (electrodes) are Type BF (floating from ground) offering the user a specific level of safety.
7
Waste Electrical and Electronic Equipment (WEEE). This
product may contain substances known to be hazardous to the
environment or to human health. It should be disposed of properly (for example, at your local waste collection administration or
recycling plant) and in accordance with local ordinances.

Service and Calibration
· Do not remove the cover. There are no user serviceable parts.

Refer all service to authorized personnel. No modification of the
equipment is allowed.

· No preventative inspections are required. Factory testing and

calibration ensure equipment accuracy and response. Contact
Zynex Medical for factory re-calibration if necessary.

8
NexWave

Electrode and Lead Wire Setup

Step 1

Open electrode package and remove electrodes from package.
Keep electrodes on plastic backing.
Note: Only Zynex Medical electrodes are approved for use with
the NexWave. P/N 300027 are provided standard. Sterile electrodes, P/N 300100, are optionally available (for U.S. only). See
Accessories on page 42.
Electrode Connector

Plastic Backing

Electrode

Step 2

Insert lead wire pin connectors into electrode connectors as
shown below. RED lead wire connectors on one side and BLACK
lead-wire connectors on the other side.

IFC modality requires all 4 electrodes (2 channels)
TENS and NMES modalities can
use 2 electrodes (1 channel) or 4
electrodes (2 channels)

9
NexWave

Electrode and Lead Wire Setup (continued)
Step 3

Remove each electrode from the plastic backing and place on
the treatment site according to the type of modality selected.
Electrode Arrangement
When using the IFC modality, RED and BLACK lead wires must
be placed in a crisscrossed pattern as shown in the diagram
below.
TENS or NMES shall not be crisscrossed and can be placed in a
side-by-side pattern.

IFC
Crisscrossed Pattern

10

TENS or NMES
Up and Down Pattern
NexWave

Electrode and Lead Wire Setup (continued)
Step 4

Plug lead wires into the top of the device. The IFC modality
requires both sets of lead wires to be connected (channel 1 and 2).
TENS and NMES can use 1 set (Channel 1) or 2 sets
(Channel 1 and 2) of lead wires.

Step 5

Proceed to page 16 to start treatment or page 12 to program
device.

Channel 1

Channel 2

11
NexWave

Device Programming Instructions
Turn device on by pressing the black On/Standby button
1.	
Select desired Modality by pressing the “IFC”, “TENS”, or “NMES”
	
button once.
2.	
Continue to press the selected modality button until arrow is next to
	
desired program.
3.	
Press Treatment Timer button until desired treatment time is set on
	screen.
Device is now ready to use.

On / Off
Button

Modality
Buttons

Treatment
Timer Button

12
NexWave

Preprogrammed Modes
Interferential Current (IFC)
Mode

Description

Low-High

The frequency of channel 2 sweeps between 4001 Hz
and 4128 Hz every 15 seconds, while the channel 1
frequency remains fixed at 4000 Hz.

Low

The frequency of channel 2 sweeps between 4001 Hz
and 4010 Hz every 15 seconds, while the channel 1
frequency remains fixed at 4000 Hz.

Combo

Combo consists of three 2 minute cycles which repeat
over the duration of the treatment.
1st cycle - Low Mode: as described above.
2nd cycle - High Mode: like Low Mode, but channel 2
sweeps between 4064 Hz and 4128 Hz.
3rd cycle - Muscle Mode: The frequency of channel 2 is
fixed at 4064 Hz, and the frequency of channel 1 is fixed
at 4000 Hz. Both channels are cycled on and off simultaneously at 6 second intervals.

Transcutaneous Electrical Nerve Stimulation (TENS)
Mode

Description

Sweep

The frequency is decreased linearly from 125 Hz to 11
Hz over a 4 second span. During this time, the pulse
width is increased linearly from 120 µs to 300 µs. Then
the frequency is decreased from 10 Hz to 1 Hz over a
two second span. The pulse width is constant at 300 µs.
Then the frequency is increased linearly from 11 Hz to
125 Hz over a four second span. During this time, the
pulse width is decreased linearly from 300 µs to 120 µs.

Lmd

Frequency sweeps between 66.7-100 Hz and back again
over 12 seconds. Pulse width linearly decreases to 150
µs when frequency is sweeping up and increases back
up to the nominal pulse width of 300 µs when frequency is sweeping down.

Hmd

Same as Lmd, but with a frequency shift interval of 2
seconds.

13
Neuromuscular Electrical Stimulation (NMES)
Mode
30:10

Preset 10 seconds on and 30 seconds off time, 35 Hz
frequency, 480 µs pulse width, 3 second ramp-up,
1 second ramp-down, both channels cycle simultaneously.

20:10

Same as above, but with 20 second off time.

10:10

14

Description

Same as above, but with and 10 second off time.
NexWave

Device Controls

Display
Lead Wire Connection

Channel 1

Lead Wire Connection

Channel 2

Uses electricity from
wall socket instead of
9 V battery

A/C Adapter
Input

Turns Unit On/Off

On/Off Button

Channel 1 Intensity Control

Channel 2 Intensity Control

Up and Down Button

Up and Down Button

Unlock Button Controls

IFC, TENS, NMES

Unlock Button

Modality Buttons
Treatment Timer

Treatment Timer Button

15
NexWave

Device Operating Instructions
Start Treatment:
Before starting treatment electrodes must be placed on the treatment site and
lead wires and electrodes connected to the device. See pages 9-11 and 17-30 for
examples.
1.

Turn NexWave on by pressing On/Standby button once

2.

Increase intensity by pressing Channel 1 Up button until a strong
but comfortable stimulation level is felt. Repeat for Channel 2 if both
lead wires are attached and device is in TENS or NMES mode.

3.

Once desired level of stimulation is set, the unit will automatically
shut off at the preset treatment time shown on the display. If Treatment Timer has been set to No Timer then the device will need to
be shut off manually. Refer to Programming Instructions on page 12 to
adjust Treatment Timer.

4.

To turn off device manually, press the On/Standby button.
Note: Intensity level can be increased or decreased with either Channel 1 or Channel 2
Up and Down buttons when device is set to IFC Mode. If “Check Connections!” message is
displayed on the screen, see “Display Alerts” below.

During Treatment:
IMPORTANT: Button controls lock after 20 seconds of inactivity. To unlock
button controls, press Unlock button.
Increase Intensity:
To increase intensity press the Up Intensity button for either channel until the desired
level of stimulation is felt. While in IFC mode, pressing either channel Up Intensity button will increase the stimulation level.
Decrease Intensity:
To decrease intensity press the Down Intensity button for each channel until desired
level of stimulation is felt. While in IFC mode, pressing either channel Down Intensity
button will decrease the stimulation level.
Display Alerts:
Check Connections: Lead wire(s) and/or electrode(s) may not be attached properly.
Check all connections and try again (see page 41). If problem persists, call Technical
Support (see page 4).
Low Battery: Replace battery immediately or connect the A/C adapter.
16

Locked: Stimulation level and mode cannot be changed until Unlock button is pressed.
Radicular Back Pain-Sciatica
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

Treatment Time

TENS		

Swp, Lmd, Hmd		

As needed

ELECTRODE PLACEMENT Option 1
Electrodes should follow the dermatomal pattern related to the affected nerve
root within the spinal segments involved.

TENS Modality
Place the electrodes with the RED and
BLACK ends of the leadwires according to
pattern depicted on figure to the left.

ELECTRODE PLACEMENT Option 2
Electrodes should follow the dermatomal pattern related to the affected nerve
root within the spinal segments involved. OVERLAPPING Techniques may vary.
17
Lumbar Back
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

IFC		
LoHi, Low, Cmb		
TENS		
Swp, Lmd, Hmd		
NMES		
30:10, 20:10, 10:10	
		off:on time

Treatment Time
40 minutes 4 times/day
As needed
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.

IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

TENS / NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.
Using both channels and crisscrossing
the electrodes is optional when using
the TENS or NMES modality.
18
Thoracic Back
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

IFC		
LoHi, Low, Cmb		
TENS		
Swp, Lmd, Hmd		
NMES		
30:10, 20:10, 10:10	
		off:on time

Treatment Time
40 minutes 4 times/day
As needed
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.

IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

TENS / NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.
Using both channels and crisscrossing
the electrodes is optional when using
the TENS or NMES modality.
19
Cervical / Upper Back
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

IFC		
LoHi, Low, Cmb		
TENS		
Swp, Lmd, Hmd		
NMES		
30:10, 20:10, 10:10	
		off:on time

Treatment Time
40 minutes 4 times/day
As needed
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.

IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

TENS / NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.
Using both channels and crisscrossing
the electrodes is optional when using
the TENS or NMES modality.
20
Shoulder
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

IFC		
LoHi, Low, Cmb		
TENS		
Swp, Lmd, Hmd		
NMES		
30:10, 20:10, 10:10	
		off:on time

Treatment Time
40 minutes 4 times/day
As needed
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.

IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

TENS / NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.
Using both channels and crisscrossing
the electrodes is optional when using
the TENS or NMES modality.
21
Elbow
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

Treatment Time

IFC		
TENS		

LoHi, Low, Cmb		
Swp, Lmd, Hmd		

40 minutes 4 times/day
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.
IFC Modality
Place the electrodes with the RED and BLACK ends of the leadwires
according to pattern depicted on figure below.
Dorsal View
same arm
Palmar View
same arm
TENS Modality
Place the electrodes with the RED and BLACK ends of the leadwires
according to pattern depicted on figure below.
Using both channels and crisscrossing the electrodes is optional when using
the TENS or NMES modality.
Dorsal View
same arm

Palmar View
same arm
22
Wrist
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

Treatment Time

IFC		
TENS		

LoHi, Low, Cmb		
Swp, Lmd, Hmd		

40 minutes 4 times/day
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.
IFC Modality
Place the electrodes with the RED and BLACK ends of the leadwires
according to pattern depicted on figure below.

Palmar View
Dorsal View
same arm
same arm
TENS Modality
Place the electrodes with the RED and BLACK ends of the leadwires
according to pattern depicted on figure below.
Using both channels and crisscrossing the electrodes is optional when using
the TENS or NMES modality.

Palmar View
same arm

Dorsal View
same arm
23
Hip
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

IFC		
LoHi, Low, Cmb		
TENS		
Swp, Lmd, Hmd		
NMES		
30:10, 20:10, 10:10	
		off:on time

Treatment Time
40 minutes 4 times/day
As needed
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.

IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

TENS / NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.
Using both channels and crisscrossing
the electrodes is optional when using
the TENS or NMES modality.
24
Back of Leg
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

IFC		
LoHi, Low, Cmb		
TENS		
Swp, Lmd, Hmd		
NMES		
30:10, 20:10, 10:10	
		off:on time

Treatment Time
40 minutes 4 times/day
As needed
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.

IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

TENS / NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.
Using both channels and crisscrossing
the electrodes is optional when using
the TENS or NMES modality.
25
Anterior Knee
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

IFC		
LoHi, Low, Cmb		
TENS		
Swp, Lmd, Hmd		
NMES		
30:10, 20:10, 10:10	
		off:on time

Treatment Time
40 minutes 4 times/day
As needed
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.

IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

TENS / NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.
Using both channels and crisscrossing
the electrodes is optional when using
the TENS or NMES modality.
26
Lateral / Medial Knee
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

IFC		
LoHi, Low, Cmb		
TENS		
Swp, Lmd, Hmd		
NMES		
30:10, 20:10, 10:10	
		off:on time

Treatment Time
40 minutes 4 times/day
As needed
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.

IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

Inside View
Same Leg

Outside

TENS / NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

Inside View
Same Leg

Outside

Using both channels and crisscrossing
the electrodes is optional when using
the TENS or NMES modality.
27
Ankle
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

Treatment Time

IFC		
TENS		

LoHi, Low, Cmb		
Swp, Lmd, Hmd		

40 minutes 4 times/day
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.
IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

ACHILLES TENDON

TENS / NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.
Using both channels and crisscrossing
the electrodes is optional when using
the TENS or NMES modality.
ACHILLES TENDON

Inside View
Same Foot

Inside View
Same Foot

Outside View
Same Foot

Inside View
Same Foot

28

Outside View
Same Foot

Outside View
Same Foot

Inside View
Same Foot

Outside View
Same Foot
Foot
PAIN CONTROL
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

Treatment Time

IFC		
TENS		

LoHi, Low, Cmb		
Swp, Lmd, Hmd		

40 minutes 4 times/day
As needed

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.

IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

TENS / NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.
Using both channels and crisscrossing
the electrodes is optional when using
the TENS or NMES modality.
29
Post-Operative Knee
PAIN RELIEF & EDEMA REDUCTION
Post-operative pain and swelling may be reduced by the application of
electrical stimulation immediately after surgery and continuing as needed.
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

Treatment Time

IFC		
TENS		
TENS		

Cmb			
Swp			
Lmd or Hmd		

40 minutes 4 times/day
As needed for pain & swelling
As needed for pain & swelling

ELECTRODE PLACEMENT
When applying electrodes in the operating room, sterile electrodes must
be used and placed away from incisions as shown below. Electrodes applied outside the operating room do not need to be sterile and should be
placed around the bandaged area in the pattern shown below.
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.
IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

TENS / NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.
Crisscrossing the electrodes is optional
when using the TENS Waveform.

30

Protocol based on the following study: The Effects of Home Interferential Therapy on Post-Operative Pain, Edema, and Range of
Motion of the Knee, by Gregg J. Jarit, MD, Karen J. Mohr, PT, SCS, Robert Waller, BS and Ronald E. Glousman, MD Published in Clinical Journal of Sports Medicine, 13:16-20 © 2003 Lippincott Williams & Wilkins, Inc, Philadelphia
Muscle Spasm
SPASM REDUCTION
Muscle spasm may be reduced by intentionally fatiguing the associated
muscle or muscle group.
INTENSITY LEVEL:
The stimulation level should be set to a strong, but comfortable strength.
DEVICE SET-UP:
Modality	

Mode			

Treatment Time

IFC		
NMES		

Cmb			
10:10 off:on time		

40 minutes 3 to 4 times/day
40 minutes 3 to 4 times/day

ELECTRODE PLACEMENT
Electrodes should surround the targeted treatment site with area of pain
located in the center of the electrode area.

IFC Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

NMES Modality
Place the electrodes with the RED
and BLACK ends of the leadwires
according to pattern depicted on
figure to the left.

31
NexWave

Indications for Use

Indications
·
·

Interferential Current (IFC)

Management and symptomatic relief of chronic intractable, posttraumatic and post-surgical pain.
This stimulator should only be used under supervision for adjunctive
therapy for the treatment of medical diseases and conditions.

Contraindications
·
·
·

This stimulator should not be used on patients with a cardiac
demand pacemaker.
Electrodes should not be placed so that current will be applied to the
carotid sinus (neck) region or transcerebrally (through the head).
This stimulator should not be used whenever pain syndromes are
undiagnosed until etiology is established.

Warnings
·

·
·

When using the separate external power supply, this medical electrical device does not incorporate a power switch to isolate the system
from the AC mains. Unplug the power cord of the AC adapter from
the AC mains outlet in order to positively disconnect from the AC
mains. Ensure that the AC outlet is easily accessible.
The device is capable of generating current densities for electrodes
exceeding 2 mA RMS/cm2 which may require special attention of the
operator.
Do not exceed 0.5 W/cm2 with recommended electrodes, assuming a
load of 1000 Ω.
· 1” round: Maximum stimulation setting = 50 mA
· 2” round: Maximum stimulation setting = 50 mA
· 2” square: Maximum stimulation setting = 50 mA

·

32

Output waveform/power information:
When delivering stimulation energy, the NexWave in IFC mode is a constant current output device. Therefore its output voltage, current, and
power are dependent upon the load as well as the pulse width and frequency of the output waveform. Using the supplied AC adapter to power
the unit, the maximum peak voltage that the NexWave can produce is
50 V (1000 Ω load). The maximum peak current that the NexWave can
produce is 50 mA (1000 Ω load). The maximum output power that the
NexWave can produce is 2.5 W per channel into a 1000 Ω load, 5 W total.
NexWave
Warnings

Interferential Current (IFC)

Warnings (Continued)
· Maximum peak output voltage (500 Ω load) = 25 V
· Maximum peak output current (500 Ω load) = 50 mA
· Maximum peak output voltage (1 KΩ load) = 50 V
· Maximum peak output current (1 KΩ load) = 50 mA
· Maximum peak output voltage (2 KΩ load) = 50 V
· Maximum peak output current (2 KΩ load) = 25 mA
· Frequency: 4000 Hz nominal
· Duty Cycle: 100%
· Maximum RMS voltage (500 Ω load) = 25 V
· Maximum RMS current (500 Ω load) = 50 mA
· Maximum output power (500 Ω load) = 1.25 W per channel
· Maximum RMS voltage (1 KΩ load) = 50 V
· Maximum RMS current (1 KΩ load) = 50 mA
· Maximum output power (1 KΩ load) = 2.5 W per channel
· Maximum RMS voltage (2 KΩ load) = 50 V
· Maximum RMS current (2 KΩ load) = 25 mA
· Maximum output power (2 KΩ load) = 1.25 W per channel
· The safety of Interferential devices for use during pregnancy or birth has
not been established.
· This device is not effective for pain of central origin.
(This includes headache.)
· This device should only be used under the continued supervision of a
trained physician.
· This device does not have curative value.
· This device offers symptomatic treatment such as suppressing the sensation of pain which would otherwise serve as a protective mechanism.
· The user must keep the device out of the reach of children.
· Electronic monitoring equipment (such as ECG monitors and ECG alarms)
may not operate properly when this device is in use.
· The long-term effects of chronic electrical stimulation are unknown.
· Stimulation should not be applied over the carotid sinus nerves,
particularly in patients with known sensitivity to the carotid sinus reflex.
· Stimulation should not be applied over the neck or mouth. Severe spasm
of the laryngeal and pharyngeal muscles may occur and the contractions
may be strong enough to close the airway or cause difficulty in breathing.
· Stimulation should not be applied transthoracically in that the introduction of electrical current into the heart may cause cardiac arrhythmias.
33
NexWave
Warnings

Interferential Current (IFC)

Warnings (Continued)
·
·
·
·

·
·

Stimulation should not be applied transcerebrally.
Stimulation should not be applied over swollen, infected, or inflamed
areas of skin eruptions, e.g. phlebitis, thrombophlebitis, varicose
veins, etc.
Stimulation should not be applied over, or in proximity to,
cancerous lesions.
The device is capable of generating currents and voltages in excess
of 10 mA RMS and 10 V RMS, respectively. A yellow LED indicator is
provided next to each lead wire connector to show that the unit is
delivering energy for any non-zero value of stimulation output
(1-50 mA).
Simultaneous connection of a patient to high frequency surgical
equipment may result in burns at the site of the stimulator electrodes
and possible damage to the stimulator.
Operations within close proximity (within one meter) of shortwave
or microwave therapy equipment may produce instability in the
stimulation output.

Safety References
Zynex® Medical, Inc. is only responsible for the safety, reliability and function
of the device when repairs and adjustments have been made by persons
authorized by Zynex® Medical, Inc., and the device is used in accordance with
the user’s manual. Repairs and technical safety tests shall only be performed
by authorized personnel.

34
NexWave

Indications for Use

Transcutaneous Electrical Nerve Stimulation (TENS)

Indications
·
·

Management and symptomatic relief of chronic intractable, posttraumatic and post-surgical pain.
This stimulator should only be used under supervision for adjunctive
therapy for the treatment of medical diseases and conditions.

Contraindications
·
·
·

This stimulator should not be used on patients with a cardiac demand
pacemaker.
Electrodes should not be placed so that current will be applied to the
carotid sinus (neck) region or transcerebrally (through the head).
This stimulator should not be used whenever pain syndromes are
undiagnosed until etiology is established.

Warnings
·

·
·

·

When using the separate external power supply, this medical electrical device does not incorporate a power switch to isolate the system
from the AC mains. Unplug the power cord of the AC adapter from
the AC mains outlet in order to positively disconnect from the AC
mains. Ensure that the AC outlet is easily is accessible.
The device is capable of generating current densities for electrodes
exceeding 2 mA RMS/cm2 which may require special attention of the
operator.
Do not exceed 0.5 W/cm2 with recommended electrodes, assuming a
load of 1000 Ω.
· 1” round: Maximum stimulation setting = 100 mA
· 2” round: Maximum stimulation setting = 100 mA
· 2” square: Maximum stimulation setting = 100 mA
Output waveform/power information:
When delivering stimulation energy, the NexWave in TENS mode is a
constant current output device. Therefore its output voltage, current,
and power are dependent upon the load as well as the pulse width and
frequency of the output waveform. The maximum peak voltage that
the NexWave can produce is 100 V (1000 Ω load). The maximum peak
current that the NexWave can produce is 100 mA (1000 Ω load). The
maximum output power that the NexWave can produce is 0.15 W per
channel into 1000 Ω load, 0.30 W total.

35
NexWave
Warnings

Transcutaneous Electrical Nerve Stimulation (TENS)

Warnings (Continued)
· Maximum peak output voltage (500 Ω load) = 50 V
· Maximum peak output current (500 Ω load) = 100 mA
· Maximum peak output voltage (1 KΩ load) = 100 V
· Maximum peak output current (1 KΩ load) = 100 mA
· Maximum peak output voltage (2 KΩ load) = 100 V
· Maximum peak output current (2 KΩ load) = 50 mA
· Pulse width range: 120 - 300 µs
· Pulse frequency range: 1 - 125 Hz
· Maximum RMS voltage (500 Ω load) = 7.07 V
· Maximum RMS current (500 Ω load) = 14.14 mA
· Maximum output power (500 Ω load) = 0.10 W per channel
· Maximum RMS voltage (1 KΩ load) = 14.14 V
· Maximum RMS current (1 KΩ load) = 14.14 mA
· Maximum output power (1 KΩ load) = 0.20 W per channel
· Maximum RMS voltage (2 KΩ load) = 14.14 V
· Maximum RMS current (2 KΩ load) = 7.07 mA
· Maximum output power (2 KΩ load) = 0.10 W per channel
· The safety of TENS devices for use during pregnancy or birth has not been
established.
· This device is not effective for pain of central origin. (This includes
headache.)
· This device should only be used under the continued supervision of a
physician.
· This device does not have curative value.
· This device offers symptomatic treatment such as suppressing the sensation of pain which would otherwise serve as a protective mechanism.
· The user must keep the device out of the reach of children.
· Electronic monitoring equipment (such as ECG monitors and ECG alarms)
may not operate properly when this device is in use.
· The long-term effects of chronic electrical stimulation are unknown.
· Stimulation should not be applied over the carotid sinus nerves,
particularly in patients with known sensitivity to the carotid sinus reflex.
· Stimulation should not be applied over the neck or mouth. Severe spasm
of the laryngeal and pharyngeal muscles may occur and the contractions
may be strong enough to close the airway or cause difficulty in breathing.
36
NexWave
Warnings

Transcutaneous Electrical Nerve Stimulation (TENS)

Warnings (Continued)
· Stimulation should not be applied transthoracically in that the introduction
of electrical current into the heart may cause cardiac arrhythmias.
· Stimulation should not be applied transcerebrally.
· Stimulation should not be applied over swollen, infected, or inflamed
areas of skin eruptions, e.g. phlebitis, thrombophlebitis, varicose veins, etc.
Stimulation should not be applied over, or in proximity to, cancerous
· lesions.
The device is capable of generating currents and voltages in excess of 10
mA RMS and 10 V RMS, respectively. A yellow LED indicator is provided
next to each lead wire connector to show that the unit is delivering
energy for any non-zero value of stimulation output (1-100 mA).
Simultaneous connection of a patient to high frequency surgical equip· ment may result in burns at the site of the stimulator electrodes and
possible damage to the stimulator.
Operations within close proximity (within one meter) of shortwave or mi· crowave therapy equipment may produce instability in the
stimulation output.

Safety References
Zynex® Medical, Inc. is only responsible for the safety, reliability and function
of the device when repairs and adjustments have been made by persons
authorized by Zynex® Medical, Inc., and the device is used in accordance with
the user’s manual. Repairs and technical safety tests shall only be performed
by authorized personnel.

37
NexWave

Indications for Use

Transcutaneous Electrical Nerve Stimulation (TENS)

Indications
·
·
·
·
·
·

Muscle re-education
Prevention or retardation of disuse atrophy
Increasing local blood circulation
Maintaining or increasing range of motion
Relaxation of muscle spasms
This stimulator should only be used under supervision for adjunctive
therapy for the treatment of medical diseases and conditions.

Contraindications
· This stimulator should not be used on patients with a cardiac demand
pacemaker.
· Electrodes should not be placed so that current will be applied to the
carotid sinus (neck) region or transcerebrally (through the head).
· This stimulator should not be used whenever pain syndromes are undiagnosed until etiology is established.

Warnings

38

· When using the separate external power supply, this medical electrical
device does not incorporate a power switch to isolate the system from
the AC mains. Unplug the power cord of the AC adapter from the AC
mains outlet in order to positively disconnect from the AC mains. Ensure
that the AC outlet is easily accessible.
· The device is capable of generating current densities for electrodes
exceeding 2 mA rms/cm2 which may require special attention of the
operator.
· Do not exceed 0.5 W/cm2 with recommended electrodes, assuming a
load of 1000 Ω.
· 1” round: Maximum stimulation setting = 100 mA
· 2” round: Maximum stimulation setting = 100 mA
· 2” square: Maximum stimulation setting = 100 mA
· Output waveform/power information:
When delivering stimulation energy, the NexWave in NMES mode is a
constant current output device. Therefore its output voltage, current, and
power are dependent upon the load as well as the pulse width and frequency
of the output waveform. The maximum peak voltage that the NexWave can
produce is 100 V (1000 Ω load).
NexWave
Warnings

Neuromuscular Electrical Stimulation (NMES)

Warnings (Continued)
The maximum output power that the NexWave can produce is 0.168 W per
channel into a 1000 Ω load, 0.336 W total.
· Maximum peak output voltage (500 Ω load) = 50 V
· Maximum peak output current (500 Ω load) = 100 mA
· Maximum peak output voltage (1 KΩ load) = 100 V
· Maximum peak output current (1 KΩ load) = 100 mA
· Maximum peak output voltage (2 KΩ load) = 100 V
· Maximum peak output current (2 KΩ load) = 50 mA
· Pulse width range: 480 µs
· Pulse frequency range: 35 Hz
· Maximum RMS voltage (500 Ω load) = 6.48 V
· Maximum RMS current (500 Ω load) = 12.96 mA
· Maximum output power (500 Ω load) = 0.084 W per channel
· Maximum RMS voltage (1 KΩ load) = 12.96 V
· Maximum RMS current (1 KΩ load) = 12.96 mA
· Maximum output power (1 KΩ load) = 0.168 W per channel
· Maximum RMS voltage (2 KΩ load) = 12.96 V
· Maximum RMS current (2 KΩ load) = 6.48 mA
· Maximum output power (2 KΩ load) = 0.084 W per channel
· The safety of NMES devices for use during pregnancy or birth has not
been established.
· This device is not effective for pain of central origin. (This includes
headache.)
· This device should only be used under the continued supervision of a
physician.
· This device does not have curative value.
· This device offers symptomatic treatment such as suppressing the sensation of pain which would otherwise serve as a protective mechanism.
· The user must keep the device out of the reach of children.
· Electronic monitoring equipment (such as ECG monitors and ECG alarms)
may not operate properly when this device is in use.
· The long-term effects of chronic electrical stimulation are unknown.
· Stimulation should not be applied over the carotid sinus nerves,
particularly in patients with known sensitivity to the carotid sinus reflex.

39
NexWave
Warnings

Neuromuscular Electrical Stimulation (NMES)

Warnings (Continued)
· Stimulation should not be applied over the neck or mouth. Severe spasm
of the laryngeal and pharyngeal muscles may occur and the contractions
may be strong enough to close the airway or cause difficulty in breathing.
· Stimulation should not be applied transthoracically in that the introduction of electrical current into the heart may cause cardiac arrhythmias.
· Stimulation should not be applied transcerebrally.
· Stimulation should not be applied over swollen, infected, or inflamed
areas of skin eruptions, e.g. phlebitis, thrombophlebitis, varicose veins,
etc.
· Stimulation should not be applied over, or in proximity to, cancerous
lesions.
· The device is capable of generating currents and voltages in excess of 10
mA RMS and 10 V RMS, respectively. A yellow LED indicator is provided
next to each lead wire connector to show that the unit is delivering
energy for any non-zero value of stimulation output (1-100 mA).
· Simultaneous connection of a patient to high frequency surgical equipment may result in burns at the site of the stimulator electrodes and
possible damage to the stimulator.
· Operations within close proximity (within one meter) of shortwave or
microwave therapy equipment may produce instability in the stimulation
output.

Safety References
Zynex® Medical, Inc. is only responsible for the safety, reliability and function
of the device when repairs and adjustments have been made by persons
authorized by Zynex® Medical, Inc., and the device is used in accordance with
the user’s manual. Repairs and technical safety tests shall only be performed
by authorized personnel.

40
NexWave

Precautions and Adverse Reactions
All Modes (IFC, TENS, NMES)

Precautions
· Isolated cases of skin irritation may occur at the site of the electrode
placement following long-term application.
· Effectiveness is highly dependent upon patient selection by a person
qualified in management of pain patients.
· Safety of powered muscle stimulators for use during pregnancy has not
been established.
· Caution should be used for patients with suspected or diagnosed heart
problems.
· Caution should be used for patients with suspected or diagnosed
epilepsy.
· Caution should be used in the presence of the following:
1. When there is a tendency to hemorrhage following acute trauma or
fracture.
2. Following recent surgical procedures when muscle contraction may
disrupt the healing process.
3. Over the menstruating or pregnant uterus
4. Over the areas of the skin which lack normal sensation.
· Some patients may experience skin irritation or hypersensitivity due to
the electrical stimulation or electrical conductive medium. The irritation
can usually be reduced by using an alternate conductive medium, or
alternate electrode placement.
· Electrode placement and stimulation settings should be based on the
guidance of the prescribing physician.
· This device should be kept out of reach of children.
· This device should be used only with the leads and electrodes
recommended for use by the manufacturer.
· This device should not be used while driving, operating machinery, or
during any activity in which voluntary muscle contractions may put the
user at undue risk of injury.

Adverse Reactions
· Skin irritation and burns beneath the electrodes are potential adverse
reactions.

41
Troubleshooting
Problem

Solution

Unit stays on even after
treatment ends.

Cannot increase level
from its current set-ting.

Press and release Unlock button or turn
level down 1 mA to unlock this safety
feature. Then increase stimulation to the
desired level/intensity. Intensity level is
always locked after 20 seconds of key press
inactivity.

Do not feel the IF
beating frequency in
the center of the four
electrodes.

Check that the lead wires are connected
correctly to the electrodes (red opposite to
each other, black opposite to each other)

Display shows “Check
Connections”
(elec-trode alarm).

42

Hold Off button down for 2 seconds to shut
unit off. Otherwise unit will shut off automatically after 5 minutes of no stimulation.
Alternatively you can start a new treatment
session after the last one ends.

Check electrode skin contact. Electrodes
should be fresh and stick well. Use water
if necessary to soften the gel for better
contact. Verify that all four electrodes are
connected to lead wires and that both lead
wires are connected to the unit. Connect
the bare metal pins together to shortcircuit the outputs. If the electrode alarm
ceases the problem is with the electrode
connection integrity.
NexWave

Specifications and Accessories
Interferential Current (IFC)
Amplitude:				0-50 mA
Channel 1 frequency:				
4000 Hz
Channel 2 frequency:				
4001- 4128 Hz
I/F Modes:					Low-High, Low, Combo
Muscle Mode:				
64 Hz, 6 sec. on, 6 sec. off, ramp up 		
					
1 sec., and ramp down 0.5 sec.
Waveforms:				Symmetrical biphasic

Transcutaneous Electrical Nerve Stimulation (TENS)
Amplitude:				0-100 mA
Frequency:				1-125 Hz
Pulse Width:				120-300 µs
TENS Modes:				
Sweep, Low Modulated, High Modulated
Waveform:					Symmetrical biphasic

Neuromuscular Electrical Stimulation (NMES)
Amplitude:				0-100 mA
Frequency:				35 Hz
Pulse Width:				480 µs
NMES Modes:				
30:10, 20:10, 10:10
On-Time:					10 sec.
Off-Time:					
30, 20, or 10 sec.
Ramp Up:					3 sec.
Ramp Down:				1 sec.
Waveforms:				Symmetrical biphasic

Other Specifications
Treatment timer:				
10 to 90 minutes, in 10 minute increments, 		
					with no timer setting.
Compliance meter:				
Records total usage time in minutes and 		
					number of times used.
Dimensions:				
2.9 x 4.6 x 1.0 in.
Weight:					5.8 oz. including battery.
Warranty:					
5 year manufacturer’s warranty on
					materials and workmanship.
					Accessories excluded.

Accessories

Lead wires:				
104203, lead wire, black, with black pins.
					
104204, lead wire, gray, with red pins.
Electrodes:				
300027, electrodes, 2” diameter, round, 	
					pkg. of four.
					
00, electrodes, sterile, 2” x 2”, square, pkg. 	
					of four.
Power supply:				
200112, power supply, medical grade, 	
					input 100-240 VAC,
					
50-60 Hz, 0.5 A max., output 12 VDC, 1.25 	
					A max.
Battery:					
130010, battery, 9 V, alkaline, MN1604.

43
NexWave

Electrodes, Batteries, and Cleaning
Electrodes and Skin Care
Proper skin care will help make the use of this device more comfortable and
trouble-free. Prior to treatment, wash the areas where the electrodes will be
placed with mild soap and water and rinse and dry the skin thoroughly. If
necessary, remove excess body hair.
Proper skin care will help make the use of this device more comfortable and
trouble-free. Prior to treatment, wash the areas where the electrodes will be
placed with mild soap and water and rinse and dry the skin thoroughly. If
necessary, remove excess body hair.
Sterile electrodes may be required for some post-operative applications.
Follow directions of the prescribing physician.

Battery Power
One 9 V alkaline battery is used. The battery compartment on the back of the
device opens by sliding the cover downwards. Rechargeable batteries are not
recommended since they have only a short usage time and are not charged
while in the device.
Battery replacement is indicated by the battery icon in the display. Replace
with a new 9 V alkaline battery.
The NexWave operates normally even if the 9 V battery is inserted with
reverse polarity.
Please dispose of used batteries properly.

AC Adapter
The NexWave is supplied with an AC adapter that is plugged into the left side
of the device and then into a 120 or 230 VAC electrical outlet. Note: The supplied AC adapter, Zynex® P/N 200112, mains power requirement is 100-240
VAC, 50-60 Hz, 0.5 A maximum. While plugged in, the device is powered from
the electrical outlet and not the 9 V battery. The battery does not need to be
removed while utilizing the AC adapter.

Device Cleaning
The NexWave case and display window can be cleaned by lightly applying
mild soap to a damp soft cloth or paper towel and using it to wipe the
surfaces clean. Allow the unit to dry thoroughly before using. Do not spray
cleaning solutions directly onto the unit, or immerse it in water or other
liquids.
44
NexWave

Warranty Information
The NexWave device is warranted to be free from defects in material,
workmanship, and structural integrity when subjected to normal domestic
use and service for five years after the original purchase.
During that time, Zynex® Medical, Inc will repair or replace, at its sole
discretion, the NexWave device that has been used in a standard manner. This
warranty does not cover misuse or use contrary to the operating instructions
supplied.
Warranty obligations are limited to replacement or repair of defective parts
and components, at the option of Zynex® Medical, Inc. This Warranty is only
valid for the original purchaser of this product and will not be honored if the
product ownership is transferred or is resold.
To obtain warranty service, please contact Technical Support at the number
listed on page 2 of this manual.

45
Zynex Medical, Inc.
9990 Park Meadows Dr.
Lone Tree, CO 80124
USA
Phone:	800-495-6670
	303-703-4906
Fax:	800-495-6695
	303-347-9153
Website: www.zynex.com
Email: info@zynex.com
European Representative:
Zynex Europe
Fælledvej 1, Box 19 BDO
DK-5100 Odense C
Denmark

300400, Rev.07, 10/15/2012
48

Mais conteúdo relacionado

Mais procurados

Transcutaneous electrical nerve stimulation (TENS)
Transcutaneous electrical nerve stimulation (TENS)Transcutaneous electrical nerve stimulation (TENS)
Transcutaneous electrical nerve stimulation (TENS)Saurab Sharma
 
INTERFERENTIAL THERAPY
INTERFERENTIAL THERAPYINTERFERENTIAL THERAPY
INTERFERENTIAL THERAPYRAJESH MANI
 
Therapeutic Ultrasound SRS
Therapeutic Ultrasound SRSTherapeutic Ultrasound SRS
Therapeutic Ultrasound SRSSreeraj S R
 
high voltage pulsating current
high voltage pulsating currenthigh voltage pulsating current
high voltage pulsating currentAqsa Mushtaq
 
sinusoidal current.ppt
sinusoidal current.pptsinusoidal current.ppt
sinusoidal current.pptBakhtawrAzhr
 
Microwave diathermy for physiotherapists
Microwave diathermy for physiotherapistsMicrowave diathermy for physiotherapists
Microwave diathermy for physiotherapistsJebaraj Fletcher
 
Electric shock and earthing
Electric shock and earthingElectric shock and earthing
Electric shock and earthingRamesh Adhikari
 
Interferential therapy
Interferential therapyInterferential therapy
Interferential therapyShubham Singh
 
Types of currents in Electrotherapy.pptx
Types of currents in Electrotherapy.pptxTypes of currents in Electrotherapy.pptx
Types of currents in Electrotherapy.pptxJamia Millia Islamia
 
Infra red radiations
Infra red radiationsInfra red radiations
Infra red radiationsSidra Aslam
 
Gait biomechanics
Gait biomechanicsGait biomechanics
Gait biomechanicsYumna Ali
 

Mais procurados (20)

TENS
TENSTENS
TENS
 
Interferential therapy
Interferential therapyInterferential therapy
Interferential therapy
 
M. IFT.pptx
M. IFT.pptxM. IFT.pptx
M. IFT.pptx
 
microwave diathermy
microwave diathermy microwave diathermy
microwave diathermy
 
Electrical reaction
Electrical reactionElectrical reaction
Electrical reaction
 
Transcutaneous electrical nerve stimulation (TENS)
Transcutaneous electrical nerve stimulation (TENS)Transcutaneous electrical nerve stimulation (TENS)
Transcutaneous electrical nerve stimulation (TENS)
 
INTERFERENTIAL THERAPY
INTERFERENTIAL THERAPYINTERFERENTIAL THERAPY
INTERFERENTIAL THERAPY
 
MWD
MWDMWD
MWD
 
Therapeutic Ultrasound SRS
Therapeutic Ultrasound SRSTherapeutic Ultrasound SRS
Therapeutic Ultrasound SRS
 
high voltage pulsating current
high voltage pulsating currenthigh voltage pulsating current
high voltage pulsating current
 
Biophysics Basics in Electrotherapy
Biophysics Basics in ElectrotherapyBiophysics Basics in Electrotherapy
Biophysics Basics in Electrotherapy
 
sinusoidal current.ppt
sinusoidal current.pptsinusoidal current.ppt
sinusoidal current.ppt
 
Microwave diathermy for physiotherapists
Microwave diathermy for physiotherapistsMicrowave diathermy for physiotherapists
Microwave diathermy for physiotherapists
 
Russian current
Russian currentRussian current
Russian current
 
Electric shock and earthing
Electric shock and earthingElectric shock and earthing
Electric shock and earthing
 
Traction
TractionTraction
Traction
 
Interferential therapy
Interferential therapyInterferential therapy
Interferential therapy
 
Types of currents in Electrotherapy.pptx
Types of currents in Electrotherapy.pptxTypes of currents in Electrotherapy.pptx
Types of currents in Electrotherapy.pptx
 
Infra red radiations
Infra red radiationsInfra red radiations
Infra red radiations
 
Gait biomechanics
Gait biomechanicsGait biomechanics
Gait biomechanics
 

Destaque

Characteristics of narration
Characteristics of  narrationCharacteristics of  narration
Characteristics of narrationphoebinku
 
44 tushaal
44 tushaal44 tushaal
44 tushaalrtumur
 
Stanford/Arrillaga El Camino Menlo Park Plan
Stanford/Arrillaga El Camino Menlo Park PlanStanford/Arrillaga El Camino Menlo Park Plan
Stanford/Arrillaga El Camino Menlo Park PlanPerla Ni
 
NLCMG - Performance is good, Understanding performance is better
NLCMG - Performance is good, Understanding performance is better NLCMG - Performance is good, Understanding performance is better
NLCMG - Performance is good, Understanding performance is better nlwebperf
 
Yesu Nallavar -Testimony of Sister Nirmala
Yesu Nallavar -Testimony of Sister NirmalaYesu Nallavar -Testimony of Sister Nirmala
Yesu Nallavar -Testimony of Sister NirmalaRaja Venkatesan
 
Python chapter 2
Python chapter 2Python chapter 2
Python chapter 2Raghu nath
 
Normicka's business cards
Normicka's business cardsNormicka's business cards
Normicka's business cardsnormicka
 
Micro Coat Sd0802 C Die Stacking
Micro Coat Sd0802 C Die StackingMicro Coat Sd0802 C Die Stacking
Micro Coat Sd0802 C Die Stackingsforman1
 
Plus Eight Star Web20 Orange Labs
Plus Eight Star Web20 Orange LabsPlus Eight Star Web20 Orange Labs
Plus Eight Star Web20 Orange LabsBenjamin Joffe
 
Business is fun: Shopping in madrid
Business is fun: Shopping in madridBusiness is fun: Shopping in madrid
Business is fun: Shopping in madridsarasacristan
 
Never alone ppt slide
Never alone ppt slideNever alone ppt slide
Never alone ppt sliderenzaldin
 
Mi cat 44 ng - Servicio Tecnico Fagor
Mi cat 44 ng - Servicio Tecnico FagorMi cat 44 ng - Servicio Tecnico Fagor
Mi cat 44 ng - Servicio Tecnico Fagorserviciotecnicofagor
 
SIGEVOlution Spring 2007
SIGEVOlution Spring 2007SIGEVOlution Spring 2007
SIGEVOlution Spring 2007Pier Luca Lanzi
 

Destaque (20)

Break Through
Break ThroughBreak Through
Break Through
 
Characteristics of narration
Characteristics of  narrationCharacteristics of  narration
Characteristics of narration
 
44 tushaal
44 tushaal44 tushaal
44 tushaal
 
Excellence land rover
Excellence land roverExcellence land rover
Excellence land rover
 
Stanford/Arrillaga El Camino Menlo Park Plan
Stanford/Arrillaga El Camino Menlo Park PlanStanford/Arrillaga El Camino Menlo Park Plan
Stanford/Arrillaga El Camino Menlo Park Plan
 
NLCMG - Performance is good, Understanding performance is better
NLCMG - Performance is good, Understanding performance is better NLCMG - Performance is good, Understanding performance is better
NLCMG - Performance is good, Understanding performance is better
 
Teens24
Teens24Teens24
Teens24
 
Yesu Nallavar -Testimony of Sister Nirmala
Yesu Nallavar -Testimony of Sister NirmalaYesu Nallavar -Testimony of Sister Nirmala
Yesu Nallavar -Testimony of Sister Nirmala
 
Python chapter 2
Python chapter 2Python chapter 2
Python chapter 2
 
Normicka's business cards
Normicka's business cardsNormicka's business cards
Normicka's business cards
 
Micro Coat Sd0802 C Die Stacking
Micro Coat Sd0802 C Die StackingMicro Coat Sd0802 C Die Stacking
Micro Coat Sd0802 C Die Stacking
 
Plus Eight Star Web20 Orange Labs
Plus Eight Star Web20 Orange LabsPlus Eight Star Web20 Orange Labs
Plus Eight Star Web20 Orange Labs
 
Rxpay-sepa-germany-01
Rxpay-sepa-germany-01Rxpay-sepa-germany-01
Rxpay-sepa-germany-01
 
Business is fun: Shopping in madrid
Business is fun: Shopping in madridBusiness is fun: Shopping in madrid
Business is fun: Shopping in madrid
 
Never alone ppt slide
Never alone ppt slideNever alone ppt slide
Never alone ppt slide
 
99 app
99 app99 app
99 app
 
Mi cat 44 ng - Servicio Tecnico Fagor
Mi cat 44 ng - Servicio Tecnico FagorMi cat 44 ng - Servicio Tecnico Fagor
Mi cat 44 ng - Servicio Tecnico Fagor
 
Rescue.asd
Rescue.asdRescue.asd
Rescue.asd
 
MBA study material- Ethics
MBA study material- EthicsMBA study material- Ethics
MBA study material- Ethics
 
SIGEVOlution Spring 2007
SIGEVOlution Spring 2007SIGEVOlution Spring 2007
SIGEVOlution Spring 2007
 

Semelhante a ZMPCZM016000.12.01 User's manual

ZMPCRM073000.12.01 User's Manual
ZMPCRM073000.12.01 User's ManualZMPCRM073000.12.01 User's Manual
ZMPCRM073000.12.01 User's ManualPainezee Specialist
 
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.Painezee Specialist
 
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.Painezee Specialist
 
Development of a t dcs device based on polarity interchangeable electrodes
Development of a t dcs device based on polarity interchangeable electrodesDevelopment of a t dcs device based on polarity interchangeable electrodes
Development of a t dcs device based on polarity interchangeable electrodesIkramInf
 
Boston Scientific Maestro 3000
Boston Scientific Maestro 3000Boston Scientific Maestro 3000
Boston Scientific Maestro 3000Christopher Tocci
 
New Ibramed Neurodyn Aussie Sport
New Ibramed Neurodyn Aussie Sport New Ibramed Neurodyn Aussie Sport
New Ibramed Neurodyn Aussie Sport ACN
 
Zmpcrm076000.12.01 Instruction manual
Zmpcrm076000.12.01 Instruction manualZmpcrm076000.12.01 Instruction manual
Zmpcrm076000.12.01 Instruction manualPainezee Specialist
 
Design of Waveform Generator for wearable Intermediate Frequency Therapy Appa...
Design of Waveform Generator for wearable Intermediate Frequency Therapy Appa...Design of Waveform Generator for wearable Intermediate Frequency Therapy Appa...
Design of Waveform Generator for wearable Intermediate Frequency Therapy Appa...IJRESJOURNAL
 
Evespring ST812 Z-Wave flood sendor manual in English
Evespring ST812 Z-Wave flood sendor manual in EnglishEvespring ST812 Z-Wave flood sendor manual in English
Evespring ST812 Z-Wave flood sendor manual in EnglishDomotica daVinci
 
Zmpcrm073000.12.01 User's Manual
Zmpcrm073000.12.01 User's  ManualZmpcrm073000.12.01 User's  Manual
Zmpcrm073000.12.01 User's ManualPainezee Specialist
 

Semelhante a ZMPCZM016000.12.01 User's manual (20)

ZMPCRM073000.12.01 users Manual
ZMPCRM073000.12.01 users ManualZMPCRM073000.12.01 users Manual
ZMPCRM073000.12.01 users Manual
 
ZMPCRM073000.12.01 User's Manual
ZMPCRM073000.12.01 User's ManualZMPCRM073000.12.01 User's Manual
ZMPCRM073000.12.01 User's Manual
 
ZMPCRM073000.12.01 Users Manual
ZMPCRM073000.12.01 Users ManualZMPCRM073000.12.01 Users Manual
ZMPCRM073000.12.01 Users Manual
 
ZMPCRM073000.12.01 users manual
ZMPCRM073000.12.01 users manualZMPCRM073000.12.01 users manual
ZMPCRM073000.12.01 users manual
 
ZMPCRM073000.12.01 Users manual
ZMPCRM073000.12.01 Users manualZMPCRM073000.12.01 Users manual
ZMPCRM073000.12.01 Users manual
 
ZMPCRM073000.12.01 Users Manual
ZMPCRM073000.12.01 Users  ManualZMPCRM073000.12.01 Users  Manual
ZMPCRM073000.12.01 Users Manual
 
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
 
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
ZMPCZM016000.10.02 Nexwave Training for Hyderabad, India.
 
Development of a t dcs device based on polarity interchangeable electrodes
Development of a t dcs device based on polarity interchangeable electrodesDevelopment of a t dcs device based on polarity interchangeable electrodes
Development of a t dcs device based on polarity interchangeable electrodes
 
MAN0824-01
MAN0824-01MAN0824-01
MAN0824-01
 
Boston Scientific Maestro 3000
Boston Scientific Maestro 3000Boston Scientific Maestro 3000
Boston Scientific Maestro 3000
 
New Ibramed Neurodyn Aussie Sport
New Ibramed Neurodyn Aussie Sport New Ibramed Neurodyn Aussie Sport
New Ibramed Neurodyn Aussie Sport
 
Zmpcrm076000.12.01 Instruction manual
Zmpcrm076000.12.01 Instruction manualZmpcrm076000.12.01 Instruction manual
Zmpcrm076000.12.01 Instruction manual
 
Design of Waveform Generator for wearable Intermediate Frequency Therapy Appa...
Design of Waveform Generator for wearable Intermediate Frequency Therapy Appa...Design of Waveform Generator for wearable Intermediate Frequency Therapy Appa...
Design of Waveform Generator for wearable Intermediate Frequency Therapy Appa...
 
27933 d en
27933 d en27933 d en
27933 d en
 
Evespring ST812 Z-Wave flood sendor manual in English
Evespring ST812 Z-Wave flood sendor manual in EnglishEvespring ST812 Z-Wave flood sendor manual in English
Evespring ST812 Z-Wave flood sendor manual in English
 
ZMPCZM018000.12 User's manual
ZMPCZM018000.12 User's  manualZMPCZM018000.12 User's  manual
ZMPCZM018000.12 User's manual
 
ZMPCZM018000.12 User's manual
ZMPCZM018000.12 User's  manualZMPCZM018000.12 User's  manual
ZMPCZM018000.12 User's manual
 
ZMPCZM018000.12 User's manual
ZMPCZM018000.12 User's  manualZMPCZM018000.12 User's  manual
ZMPCZM018000.12 User's manual
 
Zmpcrm073000.12.01 User's Manual
Zmpcrm073000.12.01 User's  ManualZmpcrm073000.12.01 User's  Manual
Zmpcrm073000.12.01 User's Manual
 

Mais de Painezee Specialist

ZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specificationsZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specificationsPainezee Specialist
 
ZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructionsZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructionsPainezee Specialist
 
Neuromove – pathway to recovery
Neuromove – pathway to recoveryNeuromove – pathway to recovery
Neuromove – pathway to recoveryPainezee Specialist
 
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...Painezee Specialist
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeePainezee Specialist
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeePainezee Specialist
 
ZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWaveZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWavePainezee Specialist
 
ZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chartZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chartPainezee Specialist
 
ZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manualZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manualPainezee Specialist
 
ZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manualZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manualPainezee Specialist
 
ZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manualZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manualPainezee Specialist
 
ZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's ManualZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's ManualPainezee Specialist
 
ZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guideZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guidePainezee Specialist
 
ZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol GuidelinesZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol GuidelinesPainezee Specialist
 
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementPainezee Specialist
 

Mais de Painezee Specialist (20)

ZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specificationsZMGYPP100000 Pressure right design_specifications
ZMGYPP100000 Pressure right design_specifications
 
ZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructionsZMGYPP100000 Pressure right product_instructions
ZMGYPP100000 Pressure right product_instructions
 
Neuromove – pathway to recovery
Neuromove – pathway to recoveryNeuromove – pathway to recovery
Neuromove – pathway to recovery
 
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
Zmpczm0170001003 ZMPCZM017000.10.03 Neuromove clinical presentation from Pain...
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
 
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzeeZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
ZMPCZM017000.10.03 Neuromove clinical presentation from PainEzee
 
ZMPCCM111000.03
ZMPCCM111000.03 ZMPCCM111000.03
ZMPCCM111000.03
 
ZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWaveZMPCZM016000.13.04 EC Declaration of Conformity NexWave
ZMPCZM016000.13.04 EC Declaration of Conformity NexWave
 
ZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chartZMPCZM017000.12.03 Neuromove Electrode placement chart
ZMPCZM017000.12.03 Neuromove Electrode placement chart
 
Zywie
ZywieZywie
Zywie
 
ZMPCZM019000.10.01
ZMPCZM019000.10.01ZMPCZM019000.10.01
ZMPCZM019000.10.01
 
ZMPCZM016000.12.07
ZMPCZM016000.12.07ZMPCZM016000.12.07
ZMPCZM016000.12.07
 
ZMPCZM016000.12.06
ZMPCZM016000.12.06ZMPCZM016000.12.06
ZMPCZM016000.12.06
 
ZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manualZMPCZM019000.12.07 NeuroGraph User's manual
ZMPCZM019000.12.07 NeuroGraph User's manual
 
ZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manualZMPCZM019000.12.06 NeuroGames User's manual
ZMPCZM019000.12.06 NeuroGames User's manual
 
ZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manualZMPCZM019000.12.05 Incontinence user's manual
ZMPCZM019000.12.05 Incontinence user's manual
 
ZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's ManualZMPCZM019000.12.04 Head Ache User's Manual
ZMPCZM019000.12.04 Head Ache User's Manual
 
ZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guideZMPCZM019000.12.03 Chiro Scan User's guide
ZMPCZM019000.12.03 Chiro Scan User's guide
 
ZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol GuidelinesZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
ZMPCZM016000.12.16 Electrode placement & Protocol Guidelines
 
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placementZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
ZMPCZM016000.12.14 Hand,Wrist,Elbow & upperarm electrode placement
 

Último

Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Dipal Arora
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Dipal Arora
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋TANUJA PANDEY
 
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeCall Girls Delhi
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...indiancallgirl4rent
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...chandars293
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Servicevidya singh
 
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 

Último (20)

Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
 
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
Call Girls Bhubaneswar Just Call 9907093804 Top Class Call Girl Service Avail...
 
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
VIP Hyderabad Call Girls Bahadurpally 7877925207 ₹5000 To 25K With AC Room 💚😋
 
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 8250077686 Top Class Call Girl Service Available
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...Top Rated  Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
Top Rated Hyderabad Call Girls Erragadda ⟟ 6297143586 ⟟ Call Me For Genuine ...
 
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort ServicePremium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
Premium Call Girls Cottonpet Whatsapp 7001035870 Independent Escort Service
 
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Kochi Just Call 8250077686 Top Class Call Girl Service Available
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Bareilly Just Call 8250077686 Top Class Call Girl Service Available
 

ZMPCZM016000.12.01 User's manual

  • 2. NexWave Table of Contents Contact Information / Customer Service / Supplies / Support 4 About the NexWave 5 Safety Information 6 Electrodes and Lead Wires Setup 9 Programming Instructions 12 Preprogrammed Modes 13 Controls 15 Operating Instructions 16 Protocols 17 Indications, Contraindications and Warnings 31 Precautions and Adverse Reactions 40 Troubleshooting 41 Specifications and Accessories 42 Electrodes, Batteries, and Cleaning 43 Warranty 44 3
  • 3. Zynex® Medical Contact Information CUSTOMER SERVICE 866-940-7030 Supplies: To order more electrodes or other accessories Technical Support: Questions or problems with using your device Device Return: Order a postage paid return envelope to return your device at no charge MAIN OFFICE 800-495-6670 303-703-4906 Billing Questions: Questions regarding insurance benefits and covered benefits for Durable Medical Equipment (DME) or questions about an Explanation of Benefits form you received in the mail. FAX NUMBER 800-495-6695 303-347-9153 MAILING ADDRESS Zynex Medical, Inc. 9990 Park Meadows Drive Lone Tree, CO 80124-6739 USA EMAIL info@zynex.com WEBSITE www.zynex.com European Representative Zynex Europe Fælledvej 1, Box 19 BDO DK-5100 Odense C Denmark 4 0470
  • 4. About the NexWave The NexWave is a multiple-mode stimulator which allows users a choice of treatment options. This device incorporates Interferential Current (IFC), Transcutaneous Electrical Nerve Stimulation (TENS), and Neuromuscular Electrical Stimulation (NMES). Interferential Current (IFC) The left channel (channel 1) produces a signal at a frequency of 4000 Hz. A slightly higher frequency (4001-4128 Hz) is produced by the right channel (channel 2). When applied to the skin, the signals from the two channels “interfere” with each other, and create a difference frequency which is felt during treatment. The skin is 100 times less resistant to electrical stimulation using this method. Consequently more of the stimulation energy reaches the nerve and muscle fibers and hence becomes more productive. With more stimulation reaching the affected area, IFC is recommended to be applied 3-4 times a day for 30-40 minutes per treatment. IFC therapy requires the use of both channels (four electrodes). Transcutaneous Electrical Nerve Stimulation (TENS) TENS delivers electrical impulses through the skin in the range of 1-125 Hz. TENS treatments can range from 10 minutes to many hours and can be applied as needed for pain relief. Patients can use one or two channels (two or four electrodes) for TENS therapy. Neuromuscular Electrical Stimulation (NMES) NMES delivers electrical impulses to muscular motor points at a frequency of 35 Hz. The output is set to a level which causes muscular contraction. These impulses are turned on for a period of 10 seconds, and turned off for a period of 10 seconds, 20 seconds, or 30 seconds. The outputs ramp up for 3 seconds, remain at the desired level for 10 seconds, ramp down for 1 second, and remain off for 10, 20, or 30 seconds. Recommended use is 3-4 times per day for 20-30 minutes per treatment. Patients can use one or two channels (two or four electrodes) for NMES therapy. Important: This device must be ordered or prescribed by a licensed physician. 5
  • 5. Safety Information Equipment Classification As per the International Electrotechnical Commission standard IEC 60601-1, and the European standard EN 60601-1, Medical Electrical Equipment, General Requirements for Basic Safety and Essential Performance, the NexWave is classified as follows: · Type BF Equipment A Type BF piece of equipment is one that provides a particular degree of protection against electric shock, particularly regarding allowable leakage current. Type BF applied part is one that is floating above ground and is isolated from all other parts of the equipment. · Medical Device Directive (MDD) Classification The NexWave a Class IIa medical device per the European Medical Device Directive. · Internally Powered Equipment When the NexWave is powered by the internal, 9 VDC, MN1604, battery it is classified as Internally Powered Equipment. · Class II Medical Equipment When powered by the external, 12 VDC, medical grade, power supply, the NexWave is classified as Class II Medical Electrical (ME) Equipment. Note: When the external power supply is connected to the NexWave device, the power supply is considered part of the ME equipment. · Electromagnetic Compatibility Conforms to IEC 60601-1-2. · Temperature Operating temperature range: 0° - 40° C (32° - 104° F) Storage temperature range: -20° - 70° C (-4° - 158° F) · Water Ingress Ordinary equipment. This device does not have protection against ingress of water. · Flammable Anesthetics This device is not suitable for use in the presence of a flammable anesthetic mixture with air, or in the presence of a flammable anesthetic mixture with oxygen or nitrous oxide. 6
  • 6. Safety Information Mode of Operation This device is suitable for continuous operation. Symbols Safety symbols shown on the device above are defined below. On/Off. This symbol indicates that the labeled switch electronically cycles the DC power on and off for part of the equipment. Note: To disconnect the external power supply, unplug the power cord of the supplied AC adapter from the AC mains outlet General Warning Sign. Follow warnings stated in the instruction manual to prevent potential hazards. Refer to Instruction Manual/Booklet. The operator must read, understand, and follow all instructions in the accompanying document including all warnings, cautions, and precautions before using this medical device. Type BF Equipment. This symbol indicates that the patient applied parts (electrodes) are Type BF (floating from ground) offering the user a specific level of safety. 7
  • 7. Waste Electrical and Electronic Equipment (WEEE). This product may contain substances known to be hazardous to the environment or to human health. It should be disposed of properly (for example, at your local waste collection administration or recycling plant) and in accordance with local ordinances. Service and Calibration · Do not remove the cover. There are no user serviceable parts. Refer all service to authorized personnel. No modification of the equipment is allowed. · No preventative inspections are required. Factory testing and calibration ensure equipment accuracy and response. Contact Zynex Medical for factory re-calibration if necessary. 8
  • 8. NexWave Electrode and Lead Wire Setup Step 1 Open electrode package and remove electrodes from package. Keep electrodes on plastic backing. Note: Only Zynex Medical electrodes are approved for use with the NexWave. P/N 300027 are provided standard. Sterile electrodes, P/N 300100, are optionally available (for U.S. only). See Accessories on page 42. Electrode Connector Plastic Backing Electrode Step 2 Insert lead wire pin connectors into electrode connectors as shown below. RED lead wire connectors on one side and BLACK lead-wire connectors on the other side. IFC modality requires all 4 electrodes (2 channels) TENS and NMES modalities can use 2 electrodes (1 channel) or 4 electrodes (2 channels) 9
  • 9. NexWave Electrode and Lead Wire Setup (continued) Step 3 Remove each electrode from the plastic backing and place on the treatment site according to the type of modality selected. Electrode Arrangement When using the IFC modality, RED and BLACK lead wires must be placed in a crisscrossed pattern as shown in the diagram below. TENS or NMES shall not be crisscrossed and can be placed in a side-by-side pattern. IFC Crisscrossed Pattern 10 TENS or NMES Up and Down Pattern
  • 10. NexWave Electrode and Lead Wire Setup (continued) Step 4 Plug lead wires into the top of the device. The IFC modality requires both sets of lead wires to be connected (channel 1 and 2). TENS and NMES can use 1 set (Channel 1) or 2 sets (Channel 1 and 2) of lead wires. Step 5 Proceed to page 16 to start treatment or page 12 to program device. Channel 1 Channel 2 11
  • 11. NexWave Device Programming Instructions Turn device on by pressing the black On/Standby button 1. Select desired Modality by pressing the “IFC”, “TENS”, or “NMES” button once. 2. Continue to press the selected modality button until arrow is next to desired program. 3. Press Treatment Timer button until desired treatment time is set on screen. Device is now ready to use. On / Off Button Modality Buttons Treatment Timer Button 12
  • 12. NexWave Preprogrammed Modes Interferential Current (IFC) Mode Description Low-High The frequency of channel 2 sweeps between 4001 Hz and 4128 Hz every 15 seconds, while the channel 1 frequency remains fixed at 4000 Hz. Low The frequency of channel 2 sweeps between 4001 Hz and 4010 Hz every 15 seconds, while the channel 1 frequency remains fixed at 4000 Hz. Combo Combo consists of three 2 minute cycles which repeat over the duration of the treatment. 1st cycle - Low Mode: as described above. 2nd cycle - High Mode: like Low Mode, but channel 2 sweeps between 4064 Hz and 4128 Hz. 3rd cycle - Muscle Mode: The frequency of channel 2 is fixed at 4064 Hz, and the frequency of channel 1 is fixed at 4000 Hz. Both channels are cycled on and off simultaneously at 6 second intervals. Transcutaneous Electrical Nerve Stimulation (TENS) Mode Description Sweep The frequency is decreased linearly from 125 Hz to 11 Hz over a 4 second span. During this time, the pulse width is increased linearly from 120 µs to 300 µs. Then the frequency is decreased from 10 Hz to 1 Hz over a two second span. The pulse width is constant at 300 µs. Then the frequency is increased linearly from 11 Hz to 125 Hz over a four second span. During this time, the pulse width is decreased linearly from 300 µs to 120 µs. Lmd Frequency sweeps between 66.7-100 Hz and back again over 12 seconds. Pulse width linearly decreases to 150 µs when frequency is sweeping up and increases back up to the nominal pulse width of 300 µs when frequency is sweeping down. Hmd Same as Lmd, but with a frequency shift interval of 2 seconds. 13
  • 13. Neuromuscular Electrical Stimulation (NMES) Mode 30:10 Preset 10 seconds on and 30 seconds off time, 35 Hz frequency, 480 µs pulse width, 3 second ramp-up, 1 second ramp-down, both channels cycle simultaneously. 20:10 Same as above, but with 20 second off time. 10:10 14 Description Same as above, but with and 10 second off time.
  • 14. NexWave Device Controls Display Lead Wire Connection Channel 1 Lead Wire Connection Channel 2 Uses electricity from wall socket instead of 9 V battery A/C Adapter Input Turns Unit On/Off On/Off Button Channel 1 Intensity Control Channel 2 Intensity Control Up and Down Button Up and Down Button Unlock Button Controls IFC, TENS, NMES Unlock Button Modality Buttons Treatment Timer Treatment Timer Button 15
  • 15. NexWave Device Operating Instructions Start Treatment: Before starting treatment electrodes must be placed on the treatment site and lead wires and electrodes connected to the device. See pages 9-11 and 17-30 for examples. 1. Turn NexWave on by pressing On/Standby button once 2. Increase intensity by pressing Channel 1 Up button until a strong but comfortable stimulation level is felt. Repeat for Channel 2 if both lead wires are attached and device is in TENS or NMES mode. 3. Once desired level of stimulation is set, the unit will automatically shut off at the preset treatment time shown on the display. If Treatment Timer has been set to No Timer then the device will need to be shut off manually. Refer to Programming Instructions on page 12 to adjust Treatment Timer. 4. To turn off device manually, press the On/Standby button. Note: Intensity level can be increased or decreased with either Channel 1 or Channel 2 Up and Down buttons when device is set to IFC Mode. If “Check Connections!” message is displayed on the screen, see “Display Alerts” below. During Treatment: IMPORTANT: Button controls lock after 20 seconds of inactivity. To unlock button controls, press Unlock button. Increase Intensity: To increase intensity press the Up Intensity button for either channel until the desired level of stimulation is felt. While in IFC mode, pressing either channel Up Intensity button will increase the stimulation level. Decrease Intensity: To decrease intensity press the Down Intensity button for each channel until desired level of stimulation is felt. While in IFC mode, pressing either channel Down Intensity button will decrease the stimulation level. Display Alerts: Check Connections: Lead wire(s) and/or electrode(s) may not be attached properly. Check all connections and try again (see page 41). If problem persists, call Technical Support (see page 4). Low Battery: Replace battery immediately or connect the A/C adapter. 16 Locked: Stimulation level and mode cannot be changed until Unlock button is pressed.
  • 16. Radicular Back Pain-Sciatica PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode Treatment Time TENS Swp, Lmd, Hmd As needed ELECTRODE PLACEMENT Option 1 Electrodes should follow the dermatomal pattern related to the affected nerve root within the spinal segments involved. TENS Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. ELECTRODE PLACEMENT Option 2 Electrodes should follow the dermatomal pattern related to the affected nerve root within the spinal segments involved. OVERLAPPING Techniques may vary. 17
  • 17. Lumbar Back PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode IFC LoHi, Low, Cmb TENS Swp, Lmd, Hmd NMES 30:10, 20:10, 10:10 off:on time Treatment Time 40 minutes 4 times/day As needed As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. TENS / NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. 18
  • 18. Thoracic Back PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode IFC LoHi, Low, Cmb TENS Swp, Lmd, Hmd NMES 30:10, 20:10, 10:10 off:on time Treatment Time 40 minutes 4 times/day As needed As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. TENS / NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. 19
  • 19. Cervical / Upper Back PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode IFC LoHi, Low, Cmb TENS Swp, Lmd, Hmd NMES 30:10, 20:10, 10:10 off:on time Treatment Time 40 minutes 4 times/day As needed As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. TENS / NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. 20
  • 20. Shoulder PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode IFC LoHi, Low, Cmb TENS Swp, Lmd, Hmd NMES 30:10, 20:10, 10:10 off:on time Treatment Time 40 minutes 4 times/day As needed As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. TENS / NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. 21
  • 21. Elbow PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode Treatment Time IFC TENS LoHi, Low, Cmb Swp, Lmd, Hmd 40 minutes 4 times/day As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure below. Dorsal View same arm Palmar View same arm TENS Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure below. Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. Dorsal View same arm Palmar View same arm 22
  • 22. Wrist PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode Treatment Time IFC TENS LoHi, Low, Cmb Swp, Lmd, Hmd 40 minutes 4 times/day As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure below. Palmar View Dorsal View same arm same arm TENS Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure below. Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. Palmar View same arm Dorsal View same arm 23
  • 23. Hip PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode IFC LoHi, Low, Cmb TENS Swp, Lmd, Hmd NMES 30:10, 20:10, 10:10 off:on time Treatment Time 40 minutes 4 times/day As needed As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. TENS / NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. 24
  • 24. Back of Leg PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode IFC LoHi, Low, Cmb TENS Swp, Lmd, Hmd NMES 30:10, 20:10, 10:10 off:on time Treatment Time 40 minutes 4 times/day As needed As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. TENS / NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. 25
  • 25. Anterior Knee PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode IFC LoHi, Low, Cmb TENS Swp, Lmd, Hmd NMES 30:10, 20:10, 10:10 off:on time Treatment Time 40 minutes 4 times/day As needed As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. TENS / NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. 26
  • 26. Lateral / Medial Knee PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode IFC LoHi, Low, Cmb TENS Swp, Lmd, Hmd NMES 30:10, 20:10, 10:10 off:on time Treatment Time 40 minutes 4 times/day As needed As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Inside View Same Leg Outside TENS / NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Inside View Same Leg Outside Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. 27
  • 27. Ankle PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode Treatment Time IFC TENS LoHi, Low, Cmb Swp, Lmd, Hmd 40 minutes 4 times/day As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. ACHILLES TENDON TENS / NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. ACHILLES TENDON Inside View Same Foot Inside View Same Foot Outside View Same Foot Inside View Same Foot 28 Outside View Same Foot Outside View Same Foot Inside View Same Foot Outside View Same Foot
  • 28. Foot PAIN CONTROL INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode Treatment Time IFC TENS LoHi, Low, Cmb Swp, Lmd, Hmd 40 minutes 4 times/day As needed ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. TENS / NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Using both channels and crisscrossing the electrodes is optional when using the TENS or NMES modality. 29
  • 29. Post-Operative Knee PAIN RELIEF & EDEMA REDUCTION Post-operative pain and swelling may be reduced by the application of electrical stimulation immediately after surgery and continuing as needed. INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode Treatment Time IFC TENS TENS Cmb Swp Lmd or Hmd 40 minutes 4 times/day As needed for pain & swelling As needed for pain & swelling ELECTRODE PLACEMENT When applying electrodes in the operating room, sterile electrodes must be used and placed away from incisions as shown below. Electrodes applied outside the operating room do not need to be sterile and should be placed around the bandaged area in the pattern shown below. Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. TENS / NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. Crisscrossing the electrodes is optional when using the TENS Waveform. 30 Protocol based on the following study: The Effects of Home Interferential Therapy on Post-Operative Pain, Edema, and Range of Motion of the Knee, by Gregg J. Jarit, MD, Karen J. Mohr, PT, SCS, Robert Waller, BS and Ronald E. Glousman, MD Published in Clinical Journal of Sports Medicine, 13:16-20 © 2003 Lippincott Williams & Wilkins, Inc, Philadelphia
  • 30. Muscle Spasm SPASM REDUCTION Muscle spasm may be reduced by intentionally fatiguing the associated muscle or muscle group. INTENSITY LEVEL: The stimulation level should be set to a strong, but comfortable strength. DEVICE SET-UP: Modality Mode Treatment Time IFC NMES Cmb 10:10 off:on time 40 minutes 3 to 4 times/day 40 minutes 3 to 4 times/day ELECTRODE PLACEMENT Electrodes should surround the targeted treatment site with area of pain located in the center of the electrode area. IFC Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. NMES Modality Place the electrodes with the RED and BLACK ends of the leadwires according to pattern depicted on figure to the left. 31
  • 31. NexWave Indications for Use Indications · · Interferential Current (IFC) Management and symptomatic relief of chronic intractable, posttraumatic and post-surgical pain. This stimulator should only be used under supervision for adjunctive therapy for the treatment of medical diseases and conditions. Contraindications · · · This stimulator should not be used on patients with a cardiac demand pacemaker. Electrodes should not be placed so that current will be applied to the carotid sinus (neck) region or transcerebrally (through the head). This stimulator should not be used whenever pain syndromes are undiagnosed until etiology is established. Warnings · · · When using the separate external power supply, this medical electrical device does not incorporate a power switch to isolate the system from the AC mains. Unplug the power cord of the AC adapter from the AC mains outlet in order to positively disconnect from the AC mains. Ensure that the AC outlet is easily accessible. The device is capable of generating current densities for electrodes exceeding 2 mA RMS/cm2 which may require special attention of the operator. Do not exceed 0.5 W/cm2 with recommended electrodes, assuming a load of 1000 Ω. · 1” round: Maximum stimulation setting = 50 mA · 2” round: Maximum stimulation setting = 50 mA · 2” square: Maximum stimulation setting = 50 mA · 32 Output waveform/power information: When delivering stimulation energy, the NexWave in IFC mode is a constant current output device. Therefore its output voltage, current, and power are dependent upon the load as well as the pulse width and frequency of the output waveform. Using the supplied AC adapter to power the unit, the maximum peak voltage that the NexWave can produce is 50 V (1000 Ω load). The maximum peak current that the NexWave can produce is 50 mA (1000 Ω load). The maximum output power that the NexWave can produce is 2.5 W per channel into a 1000 Ω load, 5 W total.
  • 32. NexWave Warnings Interferential Current (IFC) Warnings (Continued) · Maximum peak output voltage (500 Ω load) = 25 V · Maximum peak output current (500 Ω load) = 50 mA · Maximum peak output voltage (1 KΩ load) = 50 V · Maximum peak output current (1 KΩ load) = 50 mA · Maximum peak output voltage (2 KΩ load) = 50 V · Maximum peak output current (2 KΩ load) = 25 mA · Frequency: 4000 Hz nominal · Duty Cycle: 100% · Maximum RMS voltage (500 Ω load) = 25 V · Maximum RMS current (500 Ω load) = 50 mA · Maximum output power (500 Ω load) = 1.25 W per channel · Maximum RMS voltage (1 KΩ load) = 50 V · Maximum RMS current (1 KΩ load) = 50 mA · Maximum output power (1 KΩ load) = 2.5 W per channel · Maximum RMS voltage (2 KΩ load) = 50 V · Maximum RMS current (2 KΩ load) = 25 mA · Maximum output power (2 KΩ load) = 1.25 W per channel · The safety of Interferential devices for use during pregnancy or birth has not been established. · This device is not effective for pain of central origin. (This includes headache.) · This device should only be used under the continued supervision of a trained physician. · This device does not have curative value. · This device offers symptomatic treatment such as suppressing the sensation of pain which would otherwise serve as a protective mechanism. · The user must keep the device out of the reach of children. · Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate properly when this device is in use. · The long-term effects of chronic electrical stimulation are unknown. · Stimulation should not be applied over the carotid sinus nerves, particularly in patients with known sensitivity to the carotid sinus reflex. · Stimulation should not be applied over the neck or mouth. Severe spasm of the laryngeal and pharyngeal muscles may occur and the contractions may be strong enough to close the airway or cause difficulty in breathing. · Stimulation should not be applied transthoracically in that the introduction of electrical current into the heart may cause cardiac arrhythmias. 33
  • 33. NexWave Warnings Interferential Current (IFC) Warnings (Continued) · · · · · · Stimulation should not be applied transcerebrally. Stimulation should not be applied over swollen, infected, or inflamed areas of skin eruptions, e.g. phlebitis, thrombophlebitis, varicose veins, etc. Stimulation should not be applied over, or in proximity to, cancerous lesions. The device is capable of generating currents and voltages in excess of 10 mA RMS and 10 V RMS, respectively. A yellow LED indicator is provided next to each lead wire connector to show that the unit is delivering energy for any non-zero value of stimulation output (1-50 mA). Simultaneous connection of a patient to high frequency surgical equipment may result in burns at the site of the stimulator electrodes and possible damage to the stimulator. Operations within close proximity (within one meter) of shortwave or microwave therapy equipment may produce instability in the stimulation output. Safety References Zynex® Medical, Inc. is only responsible for the safety, reliability and function of the device when repairs and adjustments have been made by persons authorized by Zynex® Medical, Inc., and the device is used in accordance with the user’s manual. Repairs and technical safety tests shall only be performed by authorized personnel. 34
  • 34. NexWave Indications for Use Transcutaneous Electrical Nerve Stimulation (TENS) Indications · · Management and symptomatic relief of chronic intractable, posttraumatic and post-surgical pain. This stimulator should only be used under supervision for adjunctive therapy for the treatment of medical diseases and conditions. Contraindications · · · This stimulator should not be used on patients with a cardiac demand pacemaker. Electrodes should not be placed so that current will be applied to the carotid sinus (neck) region or transcerebrally (through the head). This stimulator should not be used whenever pain syndromes are undiagnosed until etiology is established. Warnings · · · · When using the separate external power supply, this medical electrical device does not incorporate a power switch to isolate the system from the AC mains. Unplug the power cord of the AC adapter from the AC mains outlet in order to positively disconnect from the AC mains. Ensure that the AC outlet is easily is accessible. The device is capable of generating current densities for electrodes exceeding 2 mA RMS/cm2 which may require special attention of the operator. Do not exceed 0.5 W/cm2 with recommended electrodes, assuming a load of 1000 Ω. · 1” round: Maximum stimulation setting = 100 mA · 2” round: Maximum stimulation setting = 100 mA · 2” square: Maximum stimulation setting = 100 mA Output waveform/power information: When delivering stimulation energy, the NexWave in TENS mode is a constant current output device. Therefore its output voltage, current, and power are dependent upon the load as well as the pulse width and frequency of the output waveform. The maximum peak voltage that the NexWave can produce is 100 V (1000 Ω load). The maximum peak current that the NexWave can produce is 100 mA (1000 Ω load). The maximum output power that the NexWave can produce is 0.15 W per channel into 1000 Ω load, 0.30 W total. 35
  • 35. NexWave Warnings Transcutaneous Electrical Nerve Stimulation (TENS) Warnings (Continued) · Maximum peak output voltage (500 Ω load) = 50 V · Maximum peak output current (500 Ω load) = 100 mA · Maximum peak output voltage (1 KΩ load) = 100 V · Maximum peak output current (1 KΩ load) = 100 mA · Maximum peak output voltage (2 KΩ load) = 100 V · Maximum peak output current (2 KΩ load) = 50 mA · Pulse width range: 120 - 300 µs · Pulse frequency range: 1 - 125 Hz · Maximum RMS voltage (500 Ω load) = 7.07 V · Maximum RMS current (500 Ω load) = 14.14 mA · Maximum output power (500 Ω load) = 0.10 W per channel · Maximum RMS voltage (1 KΩ load) = 14.14 V · Maximum RMS current (1 KΩ load) = 14.14 mA · Maximum output power (1 KΩ load) = 0.20 W per channel · Maximum RMS voltage (2 KΩ load) = 14.14 V · Maximum RMS current (2 KΩ load) = 7.07 mA · Maximum output power (2 KΩ load) = 0.10 W per channel · The safety of TENS devices for use during pregnancy or birth has not been established. · This device is not effective for pain of central origin. (This includes headache.) · This device should only be used under the continued supervision of a physician. · This device does not have curative value. · This device offers symptomatic treatment such as suppressing the sensation of pain which would otherwise serve as a protective mechanism. · The user must keep the device out of the reach of children. · Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate properly when this device is in use. · The long-term effects of chronic electrical stimulation are unknown. · Stimulation should not be applied over the carotid sinus nerves, particularly in patients with known sensitivity to the carotid sinus reflex. · Stimulation should not be applied over the neck or mouth. Severe spasm of the laryngeal and pharyngeal muscles may occur and the contractions may be strong enough to close the airway or cause difficulty in breathing. 36
  • 36. NexWave Warnings Transcutaneous Electrical Nerve Stimulation (TENS) Warnings (Continued) · Stimulation should not be applied transthoracically in that the introduction of electrical current into the heart may cause cardiac arrhythmias. · Stimulation should not be applied transcerebrally. · Stimulation should not be applied over swollen, infected, or inflamed areas of skin eruptions, e.g. phlebitis, thrombophlebitis, varicose veins, etc. Stimulation should not be applied over, or in proximity to, cancerous · lesions. The device is capable of generating currents and voltages in excess of 10 mA RMS and 10 V RMS, respectively. A yellow LED indicator is provided next to each lead wire connector to show that the unit is delivering energy for any non-zero value of stimulation output (1-100 mA). Simultaneous connection of a patient to high frequency surgical equip· ment may result in burns at the site of the stimulator electrodes and possible damage to the stimulator. Operations within close proximity (within one meter) of shortwave or mi· crowave therapy equipment may produce instability in the stimulation output. Safety References Zynex® Medical, Inc. is only responsible for the safety, reliability and function of the device when repairs and adjustments have been made by persons authorized by Zynex® Medical, Inc., and the device is used in accordance with the user’s manual. Repairs and technical safety tests shall only be performed by authorized personnel. 37
  • 37. NexWave Indications for Use Transcutaneous Electrical Nerve Stimulation (TENS) Indications · · · · · · Muscle re-education Prevention or retardation of disuse atrophy Increasing local blood circulation Maintaining or increasing range of motion Relaxation of muscle spasms This stimulator should only be used under supervision for adjunctive therapy for the treatment of medical diseases and conditions. Contraindications · This stimulator should not be used on patients with a cardiac demand pacemaker. · Electrodes should not be placed so that current will be applied to the carotid sinus (neck) region or transcerebrally (through the head). · This stimulator should not be used whenever pain syndromes are undiagnosed until etiology is established. Warnings 38 · When using the separate external power supply, this medical electrical device does not incorporate a power switch to isolate the system from the AC mains. Unplug the power cord of the AC adapter from the AC mains outlet in order to positively disconnect from the AC mains. Ensure that the AC outlet is easily accessible. · The device is capable of generating current densities for electrodes exceeding 2 mA rms/cm2 which may require special attention of the operator. · Do not exceed 0.5 W/cm2 with recommended electrodes, assuming a load of 1000 Ω. · 1” round: Maximum stimulation setting = 100 mA · 2” round: Maximum stimulation setting = 100 mA · 2” square: Maximum stimulation setting = 100 mA · Output waveform/power information: When delivering stimulation energy, the NexWave in NMES mode is a constant current output device. Therefore its output voltage, current, and power are dependent upon the load as well as the pulse width and frequency of the output waveform. The maximum peak voltage that the NexWave can produce is 100 V (1000 Ω load).
  • 38. NexWave Warnings Neuromuscular Electrical Stimulation (NMES) Warnings (Continued) The maximum output power that the NexWave can produce is 0.168 W per channel into a 1000 Ω load, 0.336 W total. · Maximum peak output voltage (500 Ω load) = 50 V · Maximum peak output current (500 Ω load) = 100 mA · Maximum peak output voltage (1 KΩ load) = 100 V · Maximum peak output current (1 KΩ load) = 100 mA · Maximum peak output voltage (2 KΩ load) = 100 V · Maximum peak output current (2 KΩ load) = 50 mA · Pulse width range: 480 µs · Pulse frequency range: 35 Hz · Maximum RMS voltage (500 Ω load) = 6.48 V · Maximum RMS current (500 Ω load) = 12.96 mA · Maximum output power (500 Ω load) = 0.084 W per channel · Maximum RMS voltage (1 KΩ load) = 12.96 V · Maximum RMS current (1 KΩ load) = 12.96 mA · Maximum output power (1 KΩ load) = 0.168 W per channel · Maximum RMS voltage (2 KΩ load) = 12.96 V · Maximum RMS current (2 KΩ load) = 6.48 mA · Maximum output power (2 KΩ load) = 0.084 W per channel · The safety of NMES devices for use during pregnancy or birth has not been established. · This device is not effective for pain of central origin. (This includes headache.) · This device should only be used under the continued supervision of a physician. · This device does not have curative value. · This device offers symptomatic treatment such as suppressing the sensation of pain which would otherwise serve as a protective mechanism. · The user must keep the device out of the reach of children. · Electronic monitoring equipment (such as ECG monitors and ECG alarms) may not operate properly when this device is in use. · The long-term effects of chronic electrical stimulation are unknown. · Stimulation should not be applied over the carotid sinus nerves, particularly in patients with known sensitivity to the carotid sinus reflex. 39
  • 39. NexWave Warnings Neuromuscular Electrical Stimulation (NMES) Warnings (Continued) · Stimulation should not be applied over the neck or mouth. Severe spasm of the laryngeal and pharyngeal muscles may occur and the contractions may be strong enough to close the airway or cause difficulty in breathing. · Stimulation should not be applied transthoracically in that the introduction of electrical current into the heart may cause cardiac arrhythmias. · Stimulation should not be applied transcerebrally. · Stimulation should not be applied over swollen, infected, or inflamed areas of skin eruptions, e.g. phlebitis, thrombophlebitis, varicose veins, etc. · Stimulation should not be applied over, or in proximity to, cancerous lesions. · The device is capable of generating currents and voltages in excess of 10 mA RMS and 10 V RMS, respectively. A yellow LED indicator is provided next to each lead wire connector to show that the unit is delivering energy for any non-zero value of stimulation output (1-100 mA). · Simultaneous connection of a patient to high frequency surgical equipment may result in burns at the site of the stimulator electrodes and possible damage to the stimulator. · Operations within close proximity (within one meter) of shortwave or microwave therapy equipment may produce instability in the stimulation output. Safety References Zynex® Medical, Inc. is only responsible for the safety, reliability and function of the device when repairs and adjustments have been made by persons authorized by Zynex® Medical, Inc., and the device is used in accordance with the user’s manual. Repairs and technical safety tests shall only be performed by authorized personnel. 40
  • 40. NexWave Precautions and Adverse Reactions All Modes (IFC, TENS, NMES) Precautions · Isolated cases of skin irritation may occur at the site of the electrode placement following long-term application. · Effectiveness is highly dependent upon patient selection by a person qualified in management of pain patients. · Safety of powered muscle stimulators for use during pregnancy has not been established. · Caution should be used for patients with suspected or diagnosed heart problems. · Caution should be used for patients with suspected or diagnosed epilepsy. · Caution should be used in the presence of the following: 1. When there is a tendency to hemorrhage following acute trauma or fracture. 2. Following recent surgical procedures when muscle contraction may disrupt the healing process. 3. Over the menstruating or pregnant uterus 4. Over the areas of the skin which lack normal sensation. · Some patients may experience skin irritation or hypersensitivity due to the electrical stimulation or electrical conductive medium. The irritation can usually be reduced by using an alternate conductive medium, or alternate electrode placement. · Electrode placement and stimulation settings should be based on the guidance of the prescribing physician. · This device should be kept out of reach of children. · This device should be used only with the leads and electrodes recommended for use by the manufacturer. · This device should not be used while driving, operating machinery, or during any activity in which voluntary muscle contractions may put the user at undue risk of injury. Adverse Reactions · Skin irritation and burns beneath the electrodes are potential adverse reactions. 41
  • 41. Troubleshooting Problem Solution Unit stays on even after treatment ends. Cannot increase level from its current set-ting. Press and release Unlock button or turn level down 1 mA to unlock this safety feature. Then increase stimulation to the desired level/intensity. Intensity level is always locked after 20 seconds of key press inactivity. Do not feel the IF beating frequency in the center of the four electrodes. Check that the lead wires are connected correctly to the electrodes (red opposite to each other, black opposite to each other) Display shows “Check Connections” (elec-trode alarm). 42 Hold Off button down for 2 seconds to shut unit off. Otherwise unit will shut off automatically after 5 minutes of no stimulation. Alternatively you can start a new treatment session after the last one ends. Check electrode skin contact. Electrodes should be fresh and stick well. Use water if necessary to soften the gel for better contact. Verify that all four electrodes are connected to lead wires and that both lead wires are connected to the unit. Connect the bare metal pins together to shortcircuit the outputs. If the electrode alarm ceases the problem is with the electrode connection integrity.
  • 42. NexWave Specifications and Accessories Interferential Current (IFC) Amplitude: 0-50 mA Channel 1 frequency: 4000 Hz Channel 2 frequency: 4001- 4128 Hz I/F Modes: Low-High, Low, Combo Muscle Mode: 64 Hz, 6 sec. on, 6 sec. off, ramp up 1 sec., and ramp down 0.5 sec. Waveforms: Symmetrical biphasic Transcutaneous Electrical Nerve Stimulation (TENS) Amplitude: 0-100 mA Frequency: 1-125 Hz Pulse Width: 120-300 µs TENS Modes: Sweep, Low Modulated, High Modulated Waveform: Symmetrical biphasic Neuromuscular Electrical Stimulation (NMES) Amplitude: 0-100 mA Frequency: 35 Hz Pulse Width: 480 µs NMES Modes: 30:10, 20:10, 10:10 On-Time: 10 sec. Off-Time: 30, 20, or 10 sec. Ramp Up: 3 sec. Ramp Down: 1 sec. Waveforms: Symmetrical biphasic Other Specifications Treatment timer: 10 to 90 minutes, in 10 minute increments, with no timer setting. Compliance meter: Records total usage time in minutes and number of times used. Dimensions: 2.9 x 4.6 x 1.0 in. Weight: 5.8 oz. including battery. Warranty: 5 year manufacturer’s warranty on materials and workmanship. Accessories excluded. Accessories Lead wires: 104203, lead wire, black, with black pins. 104204, lead wire, gray, with red pins. Electrodes: 300027, electrodes, 2” diameter, round, pkg. of four. 00, electrodes, sterile, 2” x 2”, square, pkg. of four. Power supply: 200112, power supply, medical grade, input 100-240 VAC, 50-60 Hz, 0.5 A max., output 12 VDC, 1.25 A max. Battery: 130010, battery, 9 V, alkaline, MN1604. 43
  • 43. NexWave Electrodes, Batteries, and Cleaning Electrodes and Skin Care Proper skin care will help make the use of this device more comfortable and trouble-free. Prior to treatment, wash the areas where the electrodes will be placed with mild soap and water and rinse and dry the skin thoroughly. If necessary, remove excess body hair. Proper skin care will help make the use of this device more comfortable and trouble-free. Prior to treatment, wash the areas where the electrodes will be placed with mild soap and water and rinse and dry the skin thoroughly. If necessary, remove excess body hair. Sterile electrodes may be required for some post-operative applications. Follow directions of the prescribing physician. Battery Power One 9 V alkaline battery is used. The battery compartment on the back of the device opens by sliding the cover downwards. Rechargeable batteries are not recommended since they have only a short usage time and are not charged while in the device. Battery replacement is indicated by the battery icon in the display. Replace with a new 9 V alkaline battery. The NexWave operates normally even if the 9 V battery is inserted with reverse polarity. Please dispose of used batteries properly. AC Adapter The NexWave is supplied with an AC adapter that is plugged into the left side of the device and then into a 120 or 230 VAC electrical outlet. Note: The supplied AC adapter, Zynex® P/N 200112, mains power requirement is 100-240 VAC, 50-60 Hz, 0.5 A maximum. While plugged in, the device is powered from the electrical outlet and not the 9 V battery. The battery does not need to be removed while utilizing the AC adapter. Device Cleaning The NexWave case and display window can be cleaned by lightly applying mild soap to a damp soft cloth or paper towel and using it to wipe the surfaces clean. Allow the unit to dry thoroughly before using. Do not spray cleaning solutions directly onto the unit, or immerse it in water or other liquids. 44
  • 44. NexWave Warranty Information The NexWave device is warranted to be free from defects in material, workmanship, and structural integrity when subjected to normal domestic use and service for five years after the original purchase. During that time, Zynex® Medical, Inc will repair or replace, at its sole discretion, the NexWave device that has been used in a standard manner. This warranty does not cover misuse or use contrary to the operating instructions supplied. Warranty obligations are limited to replacement or repair of defective parts and components, at the option of Zynex® Medical, Inc. This Warranty is only valid for the original purchaser of this product and will not be honored if the product ownership is transferred or is resold. To obtain warranty service, please contact Technical Support at the number listed on page 2 of this manual. 45
  • 45. Zynex Medical, Inc. 9990 Park Meadows Dr. Lone Tree, CO 80124 USA Phone: 800-495-6670 303-703-4906 Fax: 800-495-6695 303-347-9153 Website: www.zynex.com Email: info@zynex.com European Representative: Zynex Europe Fælledvej 1, Box 19 BDO DK-5100 Odense C Denmark 300400, Rev.07, 10/15/2012 48