Company Name
User Requirements Specification for pH meter Model BVCFTR
Edition: 1.00
                                           Page 1 of 6




              User Requirement Specification for pH Meter



                                pH Meter Model BVCFTR
                                Equipment Number : 01234


 Applies to                                     Supersedes :
 Matrix :
 Section :                                      Copy no. :




 References :

 User Manual for pH Meter Model BVCFTR.
 Operation Instruction for pH Meter Model BVCFTR Document number: 12345.
 USP References - General Chapters <791> pH.
 European pharmacopeia,2,2.3: Potentiometric determination of pH.


 Prepared by department no.: XXX

 Validation Improvement Specialist
                                       (date)        Fatima Grangeiro Landim (FGL)



 Reviewed and Approved by:

 Quality Control Labs Manager
                                       (date)         QC Manager


 Quality Assurance
                                       (date)        Quality Assurance Officer (Initial)




                                       (date)
Company Name
User Requirements Specification for pH meter Model BVCFTR
Edition: 1.00
                                           Page 2 of 6


                 Table of Contents

                 1    Purpose and Scope ............................................................................. 3
                 2    Responsibilities .................................................................................... 3
                 3    Definitions and Abbreviations ............................................................... 3
                 4    Process and Equipment Description .................................................... 3
                 5    User Requirements .............................................................................. 4
                      5.1 Design Requirements ................................................................... 4
                      5.2 Functional Requirements .............................................................. 5
                 6    Documentation and Regulation ............................................................ 5
                 7    Support ................................................................................................ 5
Company Name
User Requirements Specification for pH meter Model BVCFTR
Edition: 1.00
                                           Page 3 of 6




                 1 Purpose and Scope
                 This URS details the requirements of the pH Meter (Model BVCFTR) which will
                 be used for the determination of the pH of solutions over the approximate range
                 of pH 2 to pH 12.
                 The pH meter will be used routinely within the QC labs at room 14, building 20
                 at the Company LLLL.

                 The further qualification activities for the pH meter will be documented in IQ/OQ
                 and PQ protocols. The qualification of the pH meter will include qualification of
                 the temperature accuracy, voltage accuracy, pH accuracy, sign/noise ration and
                 further testing.
                 This is a controlled document. Additional copies are obtained through the
                 issuing company authority. In the event of copying locally, mark each document
                 with a stamp “Uncontrolled copy”.


                 2 Responsibilities
     Initials        Department                   Responsibility     Title

     FGL             QC Microbiology dept.        Author             Validation
                     Number XXX                                      Specialist
     XXX             QC Microbiology dept.        Approval           QC Manager
                     Number XXX
     XXX             QA dept. Number XXX          QA Approval        QA Officer



                 3 Definitions and Abbreviations
     Abbreviation    Full text/ Explanation

     URS             User Requirement Specification
     VMP             Validation Master Plan

     SOP             Standard Operating Procedure
     USP             U.S. Pharmacopoeia


                 4 Process and Equipment Description
                 The degree of acidity or alkalinity of a substance is determined by pH
                 measurement. The QC labs routinely measure the acidity or alkalinity of
                 samples and substances, and accurate pH measurement are essential in a
                 number of different routine situations within the QC Labs.
Company Name
User Requirements Specification for pH meter Model BVCFTR
Edition: 1.00
                                           Page 4 of 6


                 Some of the equipment features are described below:

                     1. High-contrast display with large characters for easy readability
                     2. RS232 interface as standard;
                     3. Choice of print formats (Short / Standard / GLP);
                     4. Sample ID;
                     5. Sensor ID;
                     6. User ID;
                     7. Data storage (1000 measurements / 400 calibrations / 40 methods);
                     8. Multilingual (English / German / French / Spanish);
                     9. Function that enables the time and date to be included on the printout;
                     10. Electrode holder;


                 5 User Requirements
                 The pH meter must meet the following User requirements.

     References     Requirements
     URS1           Must store at least 4 calibration points.
                    Must calculate and display asymmetry and slope values of the calibration
     URS2
                    curve.
     URS3           Alert Alarm, if occur a power failure and /or calibration data is lost.
     URS4           Large, backlit display showing clearly all relevant information.
     URS5           Date/time and sensor and sample ID fully recorded for each measurement.
                    To be able to document user ID and sensor serial number for each
     URS6
                    measurement.
                    The user interface displays menu and system settings as well as all error
     URS7
                    messages and warnings.
     URS8           The electrode (s) must be suitable for use in a pH range between 0 and 12.


                 5.1 Design Requirements

                 The pH meter must meet the following Design requirements.

     References     Requirements
     URS9           Must display pH to 0.001 pH units.
     URS10          Must have a footprint of no more than 0.6m².
     URS11          Must display a temperature to 0.1°C.
     URS12          The electrode must include a temperature sensor.
     URS13          All electrodes must have serials numbers permanently attached.
     URS14          Must operate using a 220-240V, 50-60 Hz power supply.
     URS15          Must have an interface data output to an external printer or software system.
Company Name
User Requirements Specification for pH meter Model BVCFTR
Edition: 1.00
                                           Page 5 of 6


     References     Requirements
                    All the electrodes used in the PH meter must have serial number and the
     URS16
                    name of the manufacturer on them.
                    Measurement of the zero point of the electrode shall appear on the body of
     URS17
                    the electrode.


                 5.2 Functional Requirements

                 The pH meter must meet the following Functional Requirements:

     References     Requirements
                    Accurate temperature measurement must be within ± 0.02 pH units at 25°C
     URS18
                    for the calibration curve.
     URS19          Reproducibility of pH measurement must be within ± 0.02 pH units at 25°C.
     URS20          Accurate temperature measurement to ± 0.5 °C between 20°C and 25°C.


                 6 Documentation and Regulation
                 The pH meter requires the following documentation.

     References     Requirements
     URS21 *        Manufacturer Manuals (operation & Maintenance).
     URS22          Manufacturer Quality Documentation.
     URS23          Declaration of conformity.
     URS24          All calibration certificates.
     URS25          Manufacturer is an approved QA company supplier.
     URS26          Technical Specifications.
                    All documentation supplied by supplier/manufacturer must also contain
     URS27
                    information about the electrodes and buffer solutions.
     URS28          List of recommended spare parts.

                 * The content of the equipment manuals must meet the acceptance criteria of
                 this User Requirement and further Qualification activities shall be
                 included/executed/ and documented in order to verify whether the equipment
                 manufacturer supply the user with Appropriate information, and if there is
                 consistency between documentation.


                 7 Support

     References     Requirements
     URS29          Minimum 12 months warranty.
Company Name
User Requirements Specification for pH meter Model BVCFTR
Edition: 1.00
                                           Page 6 of 6


     References     Requirements
                    Spares and accessories (pH electrode, buffer sachets, field case, probe,
     URS30
                    etc).
                    Supplier must provide service and repair for the pH meter (training in how to
     URS31          use the pH meter should be supplied by manufacturer/vendor).
Company name
User Requirement Specification for pH meter Model BVCFTR
Edition: 1.00
                                                                          Page 1 of 1




                                     Change History
Edition     Effective date Description of change           Revised without
                                                           changes
                                                           Date / Sign.

1.0         January 2011    Initial version

Mais conteúdo relacionado

Mais procurados

Validation of hvac
Validation of hvacValidation of hvac
Validation of hvac
K Manivannan
 

Mais procurados (20)

Validation master plan
Validation master planValidation master plan
Validation master plan
 
Validation utility system
Validation utility systemValidation utility system
Validation utility system
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
 
Validation of hvac
Validation of hvacValidation of hvac
Validation of hvac
 
Qualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemQualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility System
 
Pharmaceutical validation of water system
Pharmaceutical validation of  water system Pharmaceutical validation of  water system
Pharmaceutical validation of water system
 
Validation and qualification
Validation and qualificationValidation and qualification
Validation and qualification
 
Qualification of tablet compression (machine) & Capsule filling machine
Qualification of tablet compression (machine) & Capsule filling machineQualification of tablet compression (machine) & Capsule filling machine
Qualification of tablet compression (machine) & Capsule filling machine
 
SITE ACCEPTANCE TEST
SITE ACCEPTANCE TESTSITE ACCEPTANCE TEST
SITE ACCEPTANCE TEST
 
Qualification Of HPLC
Qualification Of HPLCQualification Of HPLC
Qualification Of HPLC
 
Qualification of high performance liquid chromatography
Qualification of high performance liquid chromatographyQualification of high performance liquid chromatography
Qualification of high performance liquid chromatography
 
Qualification of analytical instrument of FTIR
Qualification of analytical instrument of FTIR Qualification of analytical instrument of FTIR
Qualification of analytical instrument of FTIR
 
IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS
 
Qualification of UV spectrophotometer
Qualification of UV spectrophotometer Qualification of UV spectrophotometer
Qualification of UV spectrophotometer
 
Qualification of manufacturing equipment.
Qualification of manufacturing equipment.Qualification of manufacturing equipment.
Qualification of manufacturing equipment.
 
Computerized system validation
Computerized system validationComputerized system validation
Computerized system validation
 
Computer System Validation
Computer System ValidationComputer System Validation
Computer System Validation
 
Impurities in drug substance (ich q3 a)
Impurities in drug substance (ich q3 a)Impurities in drug substance (ich q3 a)
Impurities in drug substance (ich q3 a)
 
Vendor qualification
Vendor qualification Vendor qualification
Vendor qualification
 
Qualification of Analytical Equipments
Qualification of Analytical EquipmentsQualification of Analytical Equipments
Qualification of Analytical Equipments
 

Semelhante a User Requirement For P H Meter.Sample Copy Only.

User Requirement For Moisture Analyser.Pdf.
User Requirement For Moisture Analyser.Pdf.User Requirement For Moisture Analyser.Pdf.
User Requirement For Moisture Analyser.Pdf.
nenalandim
 
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Malarvili14
 
b.Types of validation.gayathri - gayathri murali.pptx
b.Types of validation.gayathri - gayathri murali.pptxb.Types of validation.gayathri - gayathri murali.pptx
b.Types of validation.gayathri - gayathri murali.pptx
Aarthi358332
 

Semelhante a User Requirement For P H Meter.Sample Copy Only. (20)

User Requirement For Moisture Analyser.Pdf.
User Requirement For Moisture Analyser.Pdf.User Requirement For Moisture Analyser.Pdf.
User Requirement For Moisture Analyser.Pdf.
 
TD_SMARTPAT-PH-1590_en_160303_4002735602_R04.pdf
TD_SMARTPAT-PH-1590_en_160303_4002735602_R04.pdfTD_SMARTPAT-PH-1590_en_160303_4002735602_R04.pdf
TD_SMARTPAT-PH-1590_en_160303_4002735602_R04.pdf
 
Fluke Process Calibration Tools Catalog
Fluke Process Calibration Tools CatalogFluke Process Calibration Tools Catalog
Fluke Process Calibration Tools Catalog
 
Blood gas analyser
Blood gas analyserBlood gas analyser
Blood gas analyser
 
Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)
 
Handling of Refernce Standards_Dr.A.Amsavel
Handling of Refernce Standards_Dr.A.Amsavel Handling of Refernce Standards_Dr.A.Amsavel
Handling of Refernce Standards_Dr.A.Amsavel
 
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptxICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
 
Ion Meter from Metrohm
Ion Meter from MetrohmIon Meter from Metrohm
Ion Meter from Metrohm
 
pH / Ion meter from Metrohm
pH / Ion meter from MetrohmpH / Ion meter from Metrohm
pH / Ion meter from Metrohm
 
Sop 1 20140911
Sop 1 20140911Sop 1 20140911
Sop 1 20140911
 
Rotronic Catalog 2018
Rotronic Catalog 2018Rotronic Catalog 2018
Rotronic Catalog 2018
 
Dr2700 data sheet
Dr2700 data sheetDr2700 data sheet
Dr2700 data sheet
 
E+H - Global Offering for LIfe Sciences
E+H - Global Offering for LIfe SciencesE+H - Global Offering for LIfe Sciences
E+H - Global Offering for LIfe Sciences
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
 
Qualification of instrumets
Qualification of instrumetsQualification of instrumets
Qualification of instrumets
 
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
Calibration Intermediate Check of Measuring and Voulmetric Equipemnt, Mr. Sub...
 
validation ppt.pptx
 validation ppt.pptx validation ppt.pptx
validation ppt.pptx
 
b.Types of validation.gayathri - gayathri murali.pptx
b.Types of validation.gayathri - gayathri murali.pptxb.Types of validation.gayathri - gayathri murali.pptx
b.Types of validation.gayathri - gayathri murali.pptx
 
Brooks Instrument Flow Control 2014 Product Guide - Northern California
Brooks Instrument Flow Control 2014 Product Guide - Northern CaliforniaBrooks Instrument Flow Control 2014 Product Guide - Northern California
Brooks Instrument Flow Control 2014 Product Guide - Northern California
 
Modern pharmaceutics
Modern pharmaceutics Modern pharmaceutics
Modern pharmaceutics
 

User Requirement For P H Meter.Sample Copy Only.

  • 1. Company Name User Requirements Specification for pH meter Model BVCFTR Edition: 1.00 Page 1 of 6 User Requirement Specification for pH Meter pH Meter Model BVCFTR Equipment Number : 01234 Applies to Supersedes : Matrix : Section : Copy no. : References : User Manual for pH Meter Model BVCFTR. Operation Instruction for pH Meter Model BVCFTR Document number: 12345. USP References - General Chapters <791> pH. European pharmacopeia,2,2.3: Potentiometric determination of pH. Prepared by department no.: XXX Validation Improvement Specialist (date) Fatima Grangeiro Landim (FGL) Reviewed and Approved by: Quality Control Labs Manager (date) QC Manager Quality Assurance (date) Quality Assurance Officer (Initial) (date)
  • 2. Company Name User Requirements Specification for pH meter Model BVCFTR Edition: 1.00 Page 2 of 6 Table of Contents 1 Purpose and Scope ............................................................................. 3 2 Responsibilities .................................................................................... 3 3 Definitions and Abbreviations ............................................................... 3 4 Process and Equipment Description .................................................... 3 5 User Requirements .............................................................................. 4 5.1 Design Requirements ................................................................... 4 5.2 Functional Requirements .............................................................. 5 6 Documentation and Regulation ............................................................ 5 7 Support ................................................................................................ 5
  • 3. Company Name User Requirements Specification for pH meter Model BVCFTR Edition: 1.00 Page 3 of 6 1 Purpose and Scope This URS details the requirements of the pH Meter (Model BVCFTR) which will be used for the determination of the pH of solutions over the approximate range of pH 2 to pH 12. The pH meter will be used routinely within the QC labs at room 14, building 20 at the Company LLLL. The further qualification activities for the pH meter will be documented in IQ/OQ and PQ protocols. The qualification of the pH meter will include qualification of the temperature accuracy, voltage accuracy, pH accuracy, sign/noise ration and further testing. This is a controlled document. Additional copies are obtained through the issuing company authority. In the event of copying locally, mark each document with a stamp “Uncontrolled copy”. 2 Responsibilities Initials Department Responsibility Title FGL QC Microbiology dept. Author Validation Number XXX Specialist XXX QC Microbiology dept. Approval QC Manager Number XXX XXX QA dept. Number XXX QA Approval QA Officer 3 Definitions and Abbreviations Abbreviation Full text/ Explanation URS User Requirement Specification VMP Validation Master Plan SOP Standard Operating Procedure USP U.S. Pharmacopoeia 4 Process and Equipment Description The degree of acidity or alkalinity of a substance is determined by pH measurement. The QC labs routinely measure the acidity or alkalinity of samples and substances, and accurate pH measurement are essential in a number of different routine situations within the QC Labs.
  • 4. Company Name User Requirements Specification for pH meter Model BVCFTR Edition: 1.00 Page 4 of 6 Some of the equipment features are described below: 1. High-contrast display with large characters for easy readability 2. RS232 interface as standard; 3. Choice of print formats (Short / Standard / GLP); 4. Sample ID; 5. Sensor ID; 6. User ID; 7. Data storage (1000 measurements / 400 calibrations / 40 methods); 8. Multilingual (English / German / French / Spanish); 9. Function that enables the time and date to be included on the printout; 10. Electrode holder; 5 User Requirements The pH meter must meet the following User requirements. References Requirements URS1 Must store at least 4 calibration points. Must calculate and display asymmetry and slope values of the calibration URS2 curve. URS3 Alert Alarm, if occur a power failure and /or calibration data is lost. URS4 Large, backlit display showing clearly all relevant information. URS5 Date/time and sensor and sample ID fully recorded for each measurement. To be able to document user ID and sensor serial number for each URS6 measurement. The user interface displays menu and system settings as well as all error URS7 messages and warnings. URS8 The electrode (s) must be suitable for use in a pH range between 0 and 12. 5.1 Design Requirements The pH meter must meet the following Design requirements. References Requirements URS9 Must display pH to 0.001 pH units. URS10 Must have a footprint of no more than 0.6m². URS11 Must display a temperature to 0.1°C. URS12 The electrode must include a temperature sensor. URS13 All electrodes must have serials numbers permanently attached. URS14 Must operate using a 220-240V, 50-60 Hz power supply. URS15 Must have an interface data output to an external printer or software system.
  • 5. Company Name User Requirements Specification for pH meter Model BVCFTR Edition: 1.00 Page 5 of 6 References Requirements All the electrodes used in the PH meter must have serial number and the URS16 name of the manufacturer on them. Measurement of the zero point of the electrode shall appear on the body of URS17 the electrode. 5.2 Functional Requirements The pH meter must meet the following Functional Requirements: References Requirements Accurate temperature measurement must be within ± 0.02 pH units at 25°C URS18 for the calibration curve. URS19 Reproducibility of pH measurement must be within ± 0.02 pH units at 25°C. URS20 Accurate temperature measurement to ± 0.5 °C between 20°C and 25°C. 6 Documentation and Regulation The pH meter requires the following documentation. References Requirements URS21 * Manufacturer Manuals (operation & Maintenance). URS22 Manufacturer Quality Documentation. URS23 Declaration of conformity. URS24 All calibration certificates. URS25 Manufacturer is an approved QA company supplier. URS26 Technical Specifications. All documentation supplied by supplier/manufacturer must also contain URS27 information about the electrodes and buffer solutions. URS28 List of recommended spare parts. * The content of the equipment manuals must meet the acceptance criteria of this User Requirement and further Qualification activities shall be included/executed/ and documented in order to verify whether the equipment manufacturer supply the user with Appropriate information, and if there is consistency between documentation. 7 Support References Requirements URS29 Minimum 12 months warranty.
  • 6. Company Name User Requirements Specification for pH meter Model BVCFTR Edition: 1.00 Page 6 of 6 References Requirements Spares and accessories (pH electrode, buffer sachets, field case, probe, URS30 etc). Supplier must provide service and repair for the pH meter (training in how to URS31 use the pH meter should be supplied by manufacturer/vendor).
  • 7. Company name User Requirement Specification for pH meter Model BVCFTR Edition: 1.00 Page 1 of 1 Change History Edition Effective date Description of change Revised without changes Date / Sign. 1.0 January 2011 Initial version