DEFINITION
 UV RADIATION
 PRINCIPLE OF UV-VIS SPECTROMETRY
 THE ABSORPTION SPECTRUM
 TYPES OF TRANSITIONS
 ABSORBANCE LAWS
 INSTRUMENTATION
 EFFECT OF CHROMOPHORE
 WOODWARD-FEISER RULE
 SOLVENT EFFECTS
 APPLICATIONS

Ultraviolet and visible (UV-Vis)
absorption
spectroscopy
is
the
measurement of the attenuation of a
beam of light after it passes through a
sample or after reflection from a sample
surface. Absorption measurements can
be at a single wavelength or over an
extended spectral range.
1.
2.
3.
4.
5.
6.
7.

Detection of functional groups.
Detection of impurities
Qualitative analysis
Quantitative analysis
Single compound without chromophore
Drugs with chromophoric reagent
It is helps to show the relationship
between different groups, it is useful to
detect the conjugation of the compounds
The region beyond red is called infra-red while
that beyond violet is called as ultra –violet. The
wavelength range of uv radiation starts at blue
end of visible light(4000A) & ends at 2000A.
̊




Ultraviolet absorption spectra arise from transition of electron with in a molecule from a lower level to a
higher level.
A molecule absorb ultraviolet radiation of frequency (𝜗), the electron in that molecule undergo transition
from lower to higher energy level.
The energy can be calculated by the equation,

E=h𝜗 erg
E₁-Eₒ= h𝜗
Etotal=Eelectronic+Evibrotional+Erotational

The energies decreases in the following order:
Electronic ⪢Vibrational ⪢ Rotational
Thus the energy of the radiation in the visible range is generally: 36 to 72 kcal/mole while that
in the ultraviolet range goes as high as 143 kcal/mole
When a sample is exposed to light energy that matches
the energy difference between a possible electronic
transition within the molecule, a fraction of the light energy
would be absorbed by the molecule and the electrons
would be promoted to the higher energy state orbital. A
spectrometer records the degree of absorption by a
sample at different wavelengths and the resulting plot of
absorbance (A) versus wavelength (λ) is known as a
spectrum.
The significant features:

λmax (wavelength at which there is a maximum
absorption)
 єmax (The intensity of maximum absorption)


UV-visible spectrum of isoprene showing maximum absorption at 222 nm.
Every time a molecule has a bond, the atoms in a
bond have their atomic orbitals merged to form
molecular orbitals which can be occupied by
electrons of different energy levels. Ground state
molecular orbitals can be excited to anti-bonding
molecular orbitals.
These electrons when imparted with energy in the
form of light radiation get excited from the highest
occupied molecular orbital (HOMO) to the lowest
unoccupied molecular orbital (LUMO) and the
resulting species is known as the excited state or
anti-bonding state.
TYPES OF TRANSITIONS:
In U.V spectroscopy molecule undergo
electronic transition involving σ, π and n
electrons.
 Four types of electronic transition are
possible.
i. σ ⇾ σ* transition
ii. n ⇾ σ* transition
iii. n ⇾ π* transition
iv. π ⇾ π* transition
13
An electron in a bonding σ orbital of a molecule
is excited to the corresponding anti-bonding
orbital by the absorption of radiation.
 To induce a σ ⇾ σ* transition it required LARGE
ENERGY.





Ex: Methane

Methane contain only single C-H bonds it undergo
only σ ⇾ σ* transition only, it gives absorption
maximum at 125nm.
In this type saturated compounds containing atoms
with unshared electron pairs are undergo n ⇾ σ*
transition.
It require less energy than the σ ⇾ σ* type.
Most of the absorption peaks appearing below
200nm.
In the presence of polar solvents the absorption
maximum tend to shift shorter wavelength
Ex: Water , ethanol.
 In this the peaks in U.V region relatively small.
Ex: Methlychloried , Oxygen, Nitrogen.
Most organic compounds are undergo transitions
for n ⇾ π* and π ⇾ π* transition.
 Because energies required for processes bring
the absorption peaks into spectral region.
 Both transition require the presence of an
unsaturated functional group to the ´∏´ orbitals.




Ex: For π ⇾ π* ⧐ Alkenes,
compounds, alkynes

carbonyl

For n ⇾ π* ⧐ carbonyl compounds.
σ∗ (anti-bonding)
Four types of transitions
π∗ (anti-bonding)
σ→σ*
n (non-bonding)

π→π*
n→σ*

π (bonding)
σ (bonding)

n→π*
ABSORBANCE LAWS
BEER’S LAW
“ The intensity of a beam of monochromatic light
decrease exponentially with the increase in concentration
of the absorbing substance” .
Arithmetically;
- dI/ dc ᾱ I
I= Io. eˉkc --------------------------eq (1)
LAMBERT’S LAW

“ When a beam of light is allowed to pass through a
transparent medium, the rate of decrease of
intensity with the thickness of medium is directly
proportional to the intensity of the light”
mathematically;
-dI/ dt ᾱ I
-In . I = kt+b ----------------eq(2)
the combination of eq 1 & 2 we will get
A= Kct
A= ℇct

(K=ℇ)
LIMITATION OF LAWS






The real limitation of the beer’s law is successfully
in describing the absorption behavior of dilute
solution only.
In this regarding it may be considered as a
limiting law.
As degree of interaction depends upon the
contraction, the occurrence of this phenomenon
causes deviations from linear relationship
between absorbance and contraction.
Components of spectrophotometer
 Source
 Monochromator
 Sample compartment
 Detector
 Recorder
Fig.-block diagram of instrumentation of UV-spectrophotometer
amplifier

Read out

Fig.- block diagrammatic representation of UV-spectrophotometer
It is important that the power of the radiation source does
not change abruptly over its wavelength range. The
electrical excitation of deuterium or hydrogen at low
pressure produces a continuous UV spectrum.
Both Deuterium and Hydrogen lamps emit radiation in
the range 160 - 375 nm.
Problem

Due to evaporation of tungsten life period decreases.



It is overcome by using tungsten-halogen lamp.



Halogen gas prevents evaporation of tungsten.
For ultra violet regionHydrogen discharge lamp
 consist of two electrode contain in deuterium filled silica
envelop.
UV-Vis spectrophotometer have both deuterium & tungsten
lamps.
 Selection of lamp is made by moving lamp mounting or
mirror to cause the light fall on Monochromator.
Deuterium lamps: Radiation emitted is 3-5 times more than the hydrogen
discharge lamps.
Xenon discharge lamp: Xenon stored under pressure in 10-30 atmosphere.
All Monochromators contain the following component parts;
• An entrance slit
• A collimating lens
• A dispersing device (a prism or a grating)
• A focusing lens
• An exit slit
 Filters –
a)Glass filters- Made from pieces of colored glass which
transmit limited wave length range of spectrum. Wide band
width 150nm.
b)Gelatin filters- Consist of mixture of dyes placed in gelatin
& sandwiched between glass plates. Band width 25nm.
c)Inter ferometric filters- Band width 15nm
Prisms-Prism bends the monochromatic light.
-Amount of deviation depends on wavelength
-They produce non linear dispersion.
Fig.-mechanism of working of prism.
A variety of sample cells available for UV region. The
choice of sample cell is based on
a) the path length, shape, size
b) the transmission characteristics at the desired
wavelength
c) the relative expense


The cell holding the sample should be transparent to the
wavelength region to be recorded. Quartz or fused silica
cuvettes are required for spectroscopy in the UV region.
Silicate glasses can be used for the manufacture of
cuvettes for use between 350 and 2000nm. The
thickness of the cell is generally 1 cm. cells may be
rectangular in shape or cylindrical with flat ends.
Three common types of detectors are used
I. Barrier layer cell
II. Photo cell detector
III. Photomultiplier , Photo voltaic cells
barrier layer cells
It consist of flat Cu or Fe electrode on which semiconductor such
as selenium is deposited. on the selenium a thin layer of silver or
gold is sputtered over the surface.


Photomultiplier tube
It is generally used as detector in UV- spectrophotometer It is the
combination of photodiode & electron multiplier.
It consist of evacuated tube contains photo- cathode. 9-16 electrodes
known as dynodes.
I0

log(I0/I) = A

I

200

I0

detector

monochromator/
beam splitter optics

I0

referenc
e

UV-VIS sources

sample

Advantage of double beam spectrophotometer:- It is not necessary to
continually replace the blank with the sample or to adjust the auto zero.
The ratio of the powers of the sample & reference is constantly obtained.
It has rapid scanning over the wide wavelength region because of the
above two factors.

λ, nm
Single beam spectrophotometer
Double beam colorimeter
Any Functional group which is
responsible for impairing colour to the
compound is called as chromophore.
Ex: NO2
 Covalently unsaturated groups
responsible for the impairing of the
colures.
Ex: C=C, C=O
 Two types of chromophore
a) Independent
chromophore
b) dependent chromophore

Groups
C–C

1350

C=C

1900

C=O

1900
2800

O–H

1850

NO2

2800

C6H5( PHENYL)

1950
2500

λmax

38
It is the group which itself does not act as a
chromophore but when attached to chromophore it shifts
the absorption maximum towards longer wavelength
along with an increase in intensity of adsorption.
Ex: -OH, -NH2, -OR groups
For example when the auxochrome –NH2 is attached to
the benzene ring, it absorption changes from λmax 255
to 280nm.
TYPES
Two types
a. Bathochromic groups
b. Hypsochromic group
BATHOCHROMIC GROUPS
Those groups which deepen the colour of
chromogen are called bathochromic groups.
Deepening of colour means displacement
to longer wavelength.
yellow⇾ orange ⇾ red ⇾ purple ⇾
violet⇾ blue ⇾ green
Those groups which diminish or lighten the colour of the chromogen
are called hypsochromic groups.
They cause displacement to shorter wavelength.
Ex:- acetylation of –OH or –NH2 groups,
-OCOCH3 and –
NHCOCH3

Hyperchromic



ε

Bathochromic

Hypsochromic

Hypochromic

200 nm

700 nm
It is used for calculating the absorption maxima
 Woodward (1941) gives certain rule for correlating λmax with the
molecular structure
This rule for calculating λmax in conjugated dienes, trienes, polyenes.




Homoannular dienes:cyclic dienes having conjugated double bonds in the same ring.
e.g.

CH3

CH3


Hateroannuler dienes

e.g. Heteroannuler dienes


Endocyclic double bonds
it is the double bond present in ring as shown.

Endocyclic double bond
Exocyclic double bonds
double bond in which one of the double bonded
atom is the part of ring system.

CH2

CH2
Exocyclic double bond










a)Parent values1. acyclic & Hateroannuler conjugated dienes
2.Homoannular conjugated dienes
3.Acyclic trienes
b)Increments1.Each alkyl substituent or ring residue
2.Exocyclic double bond
3.Double bond extending conjugation
4.auxochromes-OR
-SR
-Cl , Br
-NR2
-OCOCH3

215 nm
253 nm
245 nm

5 nm
5 nm
30 nm
6 nm
30 nm
5 nm
60 nm
0 nm
1,4- dimethyl cyclohex-1,3,-diene

H3C

CH3

Parent value for Homoannular dienes = 253 nm
Two alkyl substituent's
2 ˣ 5 = 10 nm
Two ring residues
2 ˣ 5 = 10 nm
Calculated value

= 273 nm
SOLVENT EFFECTS - INTENSITY
Solvents can induce significant changes in the intensity of peaks.

Hyperchromic – Increase in absorption intensity.
Hypochromic – Decrease in absorption intensity.

Absorption characteristics of 2-methylpyridine
λmax

εmax

Hexane

260

2000

Chloroform

263

4500

Ethanol

260

4000

Water

260

4000

Solvent

Ethanol - HCl (1:1)

262

5200


π -> π* transitions leads to more polar excited state that is more easily
stabilized by polar solvent associations (H-bonds). The π* state is more polar
and stabilized more in polar solvent relative to nonpolar one, thus in going
from nonpolar to polar solvent there is a red shift or bathochromic shift
(increase in λmax, decrease in ΔE).



For n -> π* transition, the n state is much more easily stabilized by polar
solvent effects (H-bonds and association), so in going from nonpolar to polar
solvent there is a blue shift or hypsochromic shift (decrease in λmax, increase in
ΔE).
APPLICATIONS:
A.

APPLICATIONS IN ORGANIC COMPOUNDS
1.It is helps to show the relationship between different groups, it is useful to
detect the conjugation of the compounds
2.Detection of geometrical isomers, In case of geometrical isomers compounds,
that trans isomers exhibits λmax at slightly longer wavelength and have larger
extinction coefficient then the cis isomers .
3.Detection of functional groups, it is possible to detect the presence of certain
functional groups with the help of UV Spectrum.
GENERAL APPLICATIONS:
1.Qualitative analysis, UV absorption spectroscopy can characterizes those type of
compounds which absorb UV radiation. Identification is done by comparing the
absorption spectrum with the spectra of known compound.
2. It is useful in Quantitative analysis of the compounds.
3. Detection of impurities, UV absorption spectroscopy is the one of the best
method for detecting impurities in organic compounds.
Tautomeric equilibrium, UV spectroscopy can be used to determine the
percentage of various keto and enol forms present in tautomeric equilibrium.
5. Chemical kinetics, UV spectroscopy can be used to study the kinetics of
reactions.
6. Molecular weight determination, molecular weights of compounds can be
measured by spectroscopy.
7. Analysis of inorganic compounds.
8. Measuring concentration of solution, absorption band can also used to
determine the concentration of compounds in a solution.
9. Inorganic chemistry, absorption spectra have been used in connection with
many problems in inorganic chemistry.
10. It is useful to determine the structure of the chloral.
QUALITATIVE ANALYSIS
Pharmacopoeial identification of drug
(1) By using absorbance & wavelength
(2) By taking absorption ratio
(3) Limit test (b)Structural analysis
Quantitative analysis
Quantitative analysis A)By using beer’s law and using absorptivity value By using
reference standard Multiple standard method
B)Single compound analysis direct analysis Using separation method After
extraction after chromatographic separation Using column chromatography Using
HPLC
Indirect analysis
a)Single compound without chromophore
b) Drugs with chromophoric reagent
1.For analyte which absorb weakly in UV region
2.For avoiding interference
3.Improve selectivity of assay 
4. Determination of composition of complex Mole ratio method Continuous
variation method ( job curve method )
5 . Study of kinetics

Disadvantages:
 Samples should be in solution. Mixture of substances poses difficult to
analyse and requires prior separation.
Interference from the sample’s matrix makes the measurement difficult .
Ultra violet (UV) spectroscopy, introduction,principle instrumentation,different types of uv transitionby mariomakhter@yahoo.com,

Mais conteúdo relacionado

Mais procurados

Ion exchange chromatography
Ion exchange chromatography Ion exchange chromatography
Ion exchange chromatography
Vharsha Haran
 
TLC, thin layer chromatography
TLC, thin layer chromatographyTLC, thin layer chromatography
TLC, thin layer chromatography
shaisejacob
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopy
sharmasuriti
 

Mais procurados (20)

Gas Chromatography
Gas ChromatographyGas Chromatography
Gas Chromatography
 
Uv visible spectroscopy ppt
Uv visible spectroscopy pptUv visible spectroscopy ppt
Uv visible spectroscopy ppt
 
Uv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentationUv visible spectroscopy-instrumentation
Uv visible spectroscopy-instrumentation
 
UV-Visible spectroscopy
UV-Visible spectroscopyUV-Visible spectroscopy
UV-Visible spectroscopy
 
Ion exchange chromatography
Ion exchange chromatography Ion exchange chromatography
Ion exchange chromatography
 
Thin layer Chromatography (TLC)
Thin layer Chromatography (TLC)Thin layer Chromatography (TLC)
Thin layer Chromatography (TLC)
 
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...
IR - Spectroscopy, theory, modes of vibration and sample handling. by Dr. Ume...
 
uv -visible spectroscopy
 uv -visible spectroscopy uv -visible spectroscopy
uv -visible spectroscopy
 
NMR spectroscopy
NMR spectroscopyNMR spectroscopy
NMR spectroscopy
 
Capillary Electrophoresis
Capillary ElectrophoresisCapillary Electrophoresis
Capillary Electrophoresis
 
Application of u.v. spectroscopy
Application of u.v. spectroscopyApplication of u.v. spectroscopy
Application of u.v. spectroscopy
 
Fluorimetry
FluorimetryFluorimetry
Fluorimetry
 
Mass spectroscopy
Mass spectroscopyMass spectroscopy
Mass spectroscopy
 
I R spectroscopy & its application
I R spectroscopy & its application I R spectroscopy & its application
I R spectroscopy & its application
 
TLC, thin layer chromatography
TLC, thin layer chromatographyTLC, thin layer chromatography
TLC, thin layer chromatography
 
Interference In Atomic Absorption Spectroscopy.
Interference In Atomic Absorption Spectroscopy.Interference In Atomic Absorption Spectroscopy.
Interference In Atomic Absorption Spectroscopy.
 
Mass spectrometry
Mass spectrometryMass spectrometry
Mass spectrometry
 
Ion exchange chromatography
Ion  exchange chromatographyIon  exchange chromatography
Ion exchange chromatography
 
atomic absorption spectroscopy
atomic absorption spectroscopyatomic absorption spectroscopy
atomic absorption spectroscopy
 
Fluorescence spectrometry
Fluorescence spectrometryFluorescence spectrometry
Fluorescence spectrometry
 

Destaque

Introduction and Principle of IR spectroscopy
Introduction and Principle of IR spectroscopyIntroduction and Principle of IR spectroscopy
Introduction and Principle of IR spectroscopy
Rajaram Kshetri
 
UV-Visible Spectroscopy
UV-Visible SpectroscopyUV-Visible Spectroscopy
UV-Visible Spectroscopy
Sudha Reddy
 
INTRODUCTION TO SPECTROSCOPY
INTRODUCTION TO SPECTROSCOPYINTRODUCTION TO SPECTROSCOPY
INTRODUCTION TO SPECTROSCOPY
SWAPNIL NIGAM
 
Introduction To Spectroscopy
Introduction To SpectroscopyIntroduction To Spectroscopy
Introduction To Spectroscopy
guest824336
 
tablet presentation
tablet presentationtablet presentation
tablet presentation
Anju K John
 
8.2 integration by parts
8.2 integration by parts8.2 integration by parts
8.2 integration by parts
dicosmo178
 
Circular dichroism
Circular dichroismCircular dichroism
Circular dichroism
Shan Too
 

Destaque (20)

Principle and instrumentation of UV-visible spectrophotometer.
Principle and instrumentation of UV-visible spectrophotometer.Principle and instrumentation of UV-visible spectrophotometer.
Principle and instrumentation of UV-visible spectrophotometer.
 
UV Visible Spectroscopy
UV Visible SpectroscopyUV Visible Spectroscopy
UV Visible Spectroscopy
 
Introduction and Principle of IR spectroscopy
Introduction and Principle of IR spectroscopyIntroduction and Principle of IR spectroscopy
Introduction and Principle of IR spectroscopy
 
UV-Visible Spectroscopy
UV-Visible SpectroscopyUV-Visible Spectroscopy
UV-Visible Spectroscopy
 
Uv seminar ppt
Uv seminar pptUv seminar ppt
Uv seminar ppt
 
Uv-visible spectroscopy
Uv-visible spectroscopyUv-visible spectroscopy
Uv-visible spectroscopy
 
INTRODUCTION TO UV-VISIBLE SPECTROSCOPY
INTRODUCTION TO UV-VISIBLE SPECTROSCOPYINTRODUCTION TO UV-VISIBLE SPECTROSCOPY
INTRODUCTION TO UV-VISIBLE SPECTROSCOPY
 
Theory of IR spectroscopy
Theory of IR spectroscopyTheory of IR spectroscopy
Theory of IR spectroscopy
 
Sampling of solids in IR spectroscopy
Sampling of solids in IR spectroscopySampling of solids in IR spectroscopy
Sampling of solids in IR spectroscopy
 
INTRODUCTION TO SPECTROSCOPY
INTRODUCTION TO SPECTROSCOPYINTRODUCTION TO SPECTROSCOPY
INTRODUCTION TO SPECTROSCOPY
 
UV-VIS Spectroscopy
UV-VIS SpectroscopyUV-VIS Spectroscopy
UV-VIS Spectroscopy
 
Introduction To Spectroscopy
Introduction To SpectroscopyIntroduction To Spectroscopy
Introduction To Spectroscopy
 
Introduction to spectroscopic methods
Introduction to spectroscopic methodsIntroduction to spectroscopic methods
Introduction to spectroscopic methods
 
Infrared instrumentation
Infrared instrumentationInfrared instrumentation
Infrared instrumentation
 
tablet presentation
tablet presentationtablet presentation
tablet presentation
 
Integration By Parts Tutorial & Example- Calculus 2
Integration By Parts Tutorial & Example- Calculus 2Integration By Parts Tutorial & Example- Calculus 2
Integration By Parts Tutorial & Example- Calculus 2
 
Sampling techniques for ir
Sampling techniques for irSampling techniques for ir
Sampling techniques for ir
 
8.2 integration by parts
8.2 integration by parts8.2 integration by parts
8.2 integration by parts
 
Confocal microscopy
Confocal microscopyConfocal microscopy
Confocal microscopy
 
Circular dichroism
Circular dichroismCircular dichroism
Circular dichroism
 

Semelhante a Ultra violet (UV) spectroscopy, introduction,principle instrumentation,different types of uv transitionby mariomakhter@yahoo.com,

Semelhante a Ultra violet (UV) spectroscopy, introduction,principle instrumentation,different types of uv transitionby mariomakhter@yahoo.com, (20)

Uvspectroscopyup 140119003106-phpapp01
Uvspectroscopyup 140119003106-phpapp01Uvspectroscopyup 140119003106-phpapp01
Uvspectroscopyup 140119003106-phpapp01
 
Uvspectroscopyup
UvspectroscopyupUvspectroscopyup
Uvspectroscopyup
 
Seminar on Uv Visible spectroscopy by Amogh G V
Seminar on Uv Visible spectroscopy by Amogh G VSeminar on Uv Visible spectroscopy by Amogh G V
Seminar on Uv Visible spectroscopy by Amogh G V
 
Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...
Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...
Ultraviolet and visible spectrophotometer and Its application in pharmaceutic...
 
uv.pptx
uv.pptxuv.pptx
uv.pptx
 
Unit 5 Spectroscopic Techniques-converted (1) (1).pdf
Unit 5 Spectroscopic Techniques-converted (1) (1).pdfUnit 5 Spectroscopic Techniques-converted (1) (1).pdf
Unit 5 Spectroscopic Techniques-converted (1) (1).pdf
 
UV visible spectroscopy
UV visible spectroscopyUV visible spectroscopy
UV visible spectroscopy
 
Spectroscopy of uv visible
Spectroscopy of uv visibleSpectroscopy of uv visible
Spectroscopy of uv visible
 
UV Spectroscopy- Pharmaceutical Analysis
UV Spectroscopy- Pharmaceutical AnalysisUV Spectroscopy- Pharmaceutical Analysis
UV Spectroscopy- Pharmaceutical Analysis
 
Uv visible spectroscopy
Uv visible spectroscopyUv visible spectroscopy
Uv visible spectroscopy
 
New ppt of uv visible
New ppt of uv visibleNew ppt of uv visible
New ppt of uv visible
 
UV Spectroscopy 1 (1).ppsx
UV Spectroscopy 1 (1).ppsxUV Spectroscopy 1 (1).ppsx
UV Spectroscopy 1 (1).ppsx
 
UV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptxUV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptx
 
UV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptxUV SPECTROSCOPY ppt.pptx
UV SPECTROSCOPY ppt.pptx
 
uv spectelectronic transition in the roscopy
uv spectelectronic transition in the roscopyuv spectelectronic transition in the roscopy
uv spectelectronic transition in the roscopy
 
UV-Visible Spectroscopy
UV-Visible SpectroscopyUV-Visible Spectroscopy
UV-Visible Spectroscopy
 
Uv visible spectroscopy
Uv visible spectroscopyUv visible spectroscopy
Uv visible spectroscopy
 
Uv visible spectroscopy
Uv visible spectroscopyUv visible spectroscopy
Uv visible spectroscopy
 
UV Visible Spectroscopy.pptx
UV Visible Spectroscopy.pptxUV Visible Spectroscopy.pptx
UV Visible Spectroscopy.pptx
 
UV visible Specrosopy and applications
UV visible Specrosopy and applicationsUV visible Specrosopy and applications
UV visible Specrosopy and applications
 

Último

Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
ciinovamais
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
QucHHunhnh
 

Último (20)

TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024
 
Measures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDMeasures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SD
 
Asian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptxAsian American Pacific Islander Month DDSD 2024.pptx
Asian American Pacific Islander Month DDSD 2024.pptx
 
Introduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The BasicsIntroduction to Nonprofit Accounting: The Basics
Introduction to Nonprofit Accounting: The Basics
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17  How to Extend Models Using Mixin ClassesMixin Classes in Odoo 17  How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
 
ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701ComPTIA Overview | Comptia Security+ Book SY0-701
ComPTIA Overview | Comptia Security+ Book SY0-701
 
On National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan FellowsOn National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan Fellows
 
ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.ICT role in 21st century education and it's challenges.
ICT role in 21st century education and it's challenges.
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptx
 
PROCESS RECORDING FORMAT.docx
PROCESS      RECORDING        FORMAT.docxPROCESS      RECORDING        FORMAT.docx
PROCESS RECORDING FORMAT.docx
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy Consulting
 

Ultra violet (UV) spectroscopy, introduction,principle instrumentation,different types of uv transitionby mariomakhter@yahoo.com,

  • 1. Presentation topic – Ultra-violate spectroscopy Submitted by – Moriyom Akhter Md Shah Alam Department of pharmacy World university of Bangladesh
  • 2. DEFINITION  UV RADIATION  PRINCIPLE OF UV-VIS SPECTROMETRY  THE ABSORPTION SPECTRUM  TYPES OF TRANSITIONS  ABSORBANCE LAWS  INSTRUMENTATION  EFFECT OF CHROMOPHORE  WOODWARD-FEISER RULE  SOLVENT EFFECTS  APPLICATIONS 
  • 3. Ultraviolet and visible (UV-Vis) absorption spectroscopy is the measurement of the attenuation of a beam of light after it passes through a sample or after reflection from a sample surface. Absorption measurements can be at a single wavelength or over an extended spectral range.
  • 4. 1. 2. 3. 4. 5. 6. 7. Detection of functional groups. Detection of impurities Qualitative analysis Quantitative analysis Single compound without chromophore Drugs with chromophoric reagent It is helps to show the relationship between different groups, it is useful to detect the conjugation of the compounds
  • 5. The region beyond red is called infra-red while that beyond violet is called as ultra –violet. The wavelength range of uv radiation starts at blue end of visible light(4000A) & ends at 2000A. ̊
  • 6.   Ultraviolet absorption spectra arise from transition of electron with in a molecule from a lower level to a higher level. A molecule absorb ultraviolet radiation of frequency (𝜗), the electron in that molecule undergo transition from lower to higher energy level. The energy can be calculated by the equation, E=h𝜗 erg
  • 7. E₁-Eₒ= h𝜗 Etotal=Eelectronic+Evibrotional+Erotational The energies decreases in the following order: Electronic ⪢Vibrational ⪢ Rotational
  • 8. Thus the energy of the radiation in the visible range is generally: 36 to 72 kcal/mole while that in the ultraviolet range goes as high as 143 kcal/mole
  • 9. When a sample is exposed to light energy that matches the energy difference between a possible electronic transition within the molecule, a fraction of the light energy would be absorbed by the molecule and the electrons would be promoted to the higher energy state orbital. A spectrometer records the degree of absorption by a sample at different wavelengths and the resulting plot of absorbance (A) versus wavelength (λ) is known as a spectrum. The significant features:  λmax (wavelength at which there is a maximum absorption)  єmax (The intensity of maximum absorption)
  • 10.
  • 11.  UV-visible spectrum of isoprene showing maximum absorption at 222 nm.
  • 12. Every time a molecule has a bond, the atoms in a bond have their atomic orbitals merged to form molecular orbitals which can be occupied by electrons of different energy levels. Ground state molecular orbitals can be excited to anti-bonding molecular orbitals. These electrons when imparted with energy in the form of light radiation get excited from the highest occupied molecular orbital (HOMO) to the lowest unoccupied molecular orbital (LUMO) and the resulting species is known as the excited state or anti-bonding state.
  • 13. TYPES OF TRANSITIONS: In U.V spectroscopy molecule undergo electronic transition involving σ, π and n electrons.  Four types of electronic transition are possible. i. σ ⇾ σ* transition ii. n ⇾ σ* transition iii. n ⇾ π* transition iv. π ⇾ π* transition 13
  • 14. An electron in a bonding σ orbital of a molecule is excited to the corresponding anti-bonding orbital by the absorption of radiation.  To induce a σ ⇾ σ* transition it required LARGE ENERGY.    Ex: Methane Methane contain only single C-H bonds it undergo only σ ⇾ σ* transition only, it gives absorption maximum at 125nm.
  • 15. In this type saturated compounds containing atoms with unshared electron pairs are undergo n ⇾ σ* transition. It require less energy than the σ ⇾ σ* type. Most of the absorption peaks appearing below 200nm. In the presence of polar solvents the absorption maximum tend to shift shorter wavelength Ex: Water , ethanol.  In this the peaks in U.V region relatively small. Ex: Methlychloried , Oxygen, Nitrogen.
  • 16. Most organic compounds are undergo transitions for n ⇾ π* and π ⇾ π* transition.  Because energies required for processes bring the absorption peaks into spectral region.  Both transition require the presence of an unsaturated functional group to the ´∏´ orbitals.   Ex: For π ⇾ π* ⧐ Alkenes, compounds, alkynes carbonyl For n ⇾ π* ⧐ carbonyl compounds.
  • 17. σ∗ (anti-bonding) Four types of transitions π∗ (anti-bonding) σ→σ* n (non-bonding) π→π* n→σ* π (bonding) σ (bonding) n→π*
  • 18. ABSORBANCE LAWS BEER’S LAW “ The intensity of a beam of monochromatic light decrease exponentially with the increase in concentration of the absorbing substance” . Arithmetically; - dI/ dc ᾱ I I= Io. eˉkc --------------------------eq (1)
  • 19. LAMBERT’S LAW “ When a beam of light is allowed to pass through a transparent medium, the rate of decrease of intensity with the thickness of medium is directly proportional to the intensity of the light” mathematically; -dI/ dt ᾱ I -In . I = kt+b ----------------eq(2) the combination of eq 1 & 2 we will get A= Kct A= ℇct (K=ℇ)
  • 20. LIMITATION OF LAWS    The real limitation of the beer’s law is successfully in describing the absorption behavior of dilute solution only. In this regarding it may be considered as a limiting law. As degree of interaction depends upon the contraction, the occurrence of this phenomenon causes deviations from linear relationship between absorbance and contraction.
  • 21. Components of spectrophotometer  Source  Monochromator  Sample compartment  Detector  Recorder
  • 22. Fig.-block diagram of instrumentation of UV-spectrophotometer
  • 23. amplifier Read out Fig.- block diagrammatic representation of UV-spectrophotometer
  • 24. It is important that the power of the radiation source does not change abruptly over its wavelength range. The electrical excitation of deuterium or hydrogen at low pressure produces a continuous UV spectrum. Both Deuterium and Hydrogen lamps emit radiation in the range 160 - 375 nm. Problem Due to evaporation of tungsten life period decreases.  It is overcome by using tungsten-halogen lamp.  Halogen gas prevents evaporation of tungsten.
  • 25. For ultra violet regionHydrogen discharge lamp  consist of two electrode contain in deuterium filled silica envelop. UV-Vis spectrophotometer have both deuterium & tungsten lamps.  Selection of lamp is made by moving lamp mounting or mirror to cause the light fall on Monochromator. Deuterium lamps: Radiation emitted is 3-5 times more than the hydrogen discharge lamps. Xenon discharge lamp: Xenon stored under pressure in 10-30 atmosphere.
  • 26. All Monochromators contain the following component parts; • An entrance slit • A collimating lens • A dispersing device (a prism or a grating) • A focusing lens • An exit slit
  • 27.  Filters – a)Glass filters- Made from pieces of colored glass which transmit limited wave length range of spectrum. Wide band width 150nm. b)Gelatin filters- Consist of mixture of dyes placed in gelatin & sandwiched between glass plates. Band width 25nm. c)Inter ferometric filters- Band width 15nm Prisms-Prism bends the monochromatic light. -Amount of deviation depends on wavelength -They produce non linear dispersion.
  • 29. A variety of sample cells available for UV region. The choice of sample cell is based on a) the path length, shape, size b) the transmission characteristics at the desired wavelength c) the relative expense  The cell holding the sample should be transparent to the wavelength region to be recorded. Quartz or fused silica cuvettes are required for spectroscopy in the UV region. Silicate glasses can be used for the manufacture of cuvettes for use between 350 and 2000nm. The thickness of the cell is generally 1 cm. cells may be rectangular in shape or cylindrical with flat ends.
  • 30.
  • 31. Three common types of detectors are used I. Barrier layer cell II. Photo cell detector III. Photomultiplier , Photo voltaic cells barrier layer cells It consist of flat Cu or Fe electrode on which semiconductor such as selenium is deposited. on the selenium a thin layer of silver or gold is sputtered over the surface.
  • 32.  Photomultiplier tube It is generally used as detector in UV- spectrophotometer It is the combination of photodiode & electron multiplier. It consist of evacuated tube contains photo- cathode. 9-16 electrodes known as dynodes.
  • 33.
  • 34. I0 log(I0/I) = A I 200 I0 detector monochromator/ beam splitter optics I0 referenc e UV-VIS sources sample Advantage of double beam spectrophotometer:- It is not necessary to continually replace the blank with the sample or to adjust the auto zero. The ratio of the powers of the sample & reference is constantly obtained. It has rapid scanning over the wide wavelength region because of the above two factors. λ, nm
  • 37. Any Functional group which is responsible for impairing colour to the compound is called as chromophore. Ex: NO2  Covalently unsaturated groups responsible for the impairing of the colures. Ex: C=C, C=O  Two types of chromophore a) Independent chromophore b) dependent chromophore 
  • 39. It is the group which itself does not act as a chromophore but when attached to chromophore it shifts the absorption maximum towards longer wavelength along with an increase in intensity of adsorption. Ex: -OH, -NH2, -OR groups For example when the auxochrome –NH2 is attached to the benzene ring, it absorption changes from λmax 255 to 280nm. TYPES Two types a. Bathochromic groups b. Hypsochromic group
  • 40. BATHOCHROMIC GROUPS Those groups which deepen the colour of chromogen are called bathochromic groups. Deepening of colour means displacement to longer wavelength. yellow⇾ orange ⇾ red ⇾ purple ⇾ violet⇾ blue ⇾ green
  • 41. Those groups which diminish or lighten the colour of the chromogen are called hypsochromic groups. They cause displacement to shorter wavelength. Ex:- acetylation of –OH or –NH2 groups, -OCOCH3 and – NHCOCH3 Hyperchromic  ε Bathochromic Hypsochromic Hypochromic 200 nm 700 nm
  • 42. It is used for calculating the absorption maxima  Woodward (1941) gives certain rule for correlating λmax with the molecular structure This rule for calculating λmax in conjugated dienes, trienes, polyenes.   Homoannular dienes:cyclic dienes having conjugated double bonds in the same ring. e.g. CH3 CH3
  • 43.  Hateroannuler dienes e.g. Heteroannuler dienes  Endocyclic double bonds it is the double bond present in ring as shown. Endocyclic double bond
  • 44. Exocyclic double bonds double bond in which one of the double bonded atom is the part of ring system. CH2 CH2 Exocyclic double bond
  • 45.        a)Parent values1. acyclic & Hateroannuler conjugated dienes 2.Homoannular conjugated dienes 3.Acyclic trienes b)Increments1.Each alkyl substituent or ring residue 2.Exocyclic double bond 3.Double bond extending conjugation 4.auxochromes-OR -SR -Cl , Br -NR2 -OCOCH3 215 nm 253 nm 245 nm 5 nm 5 nm 30 nm 6 nm 30 nm 5 nm 60 nm 0 nm
  • 46. 1,4- dimethyl cyclohex-1,3,-diene H3C CH3 Parent value for Homoannular dienes = 253 nm Two alkyl substituent's 2 ˣ 5 = 10 nm Two ring residues 2 ˣ 5 = 10 nm Calculated value = 273 nm
  • 47. SOLVENT EFFECTS - INTENSITY Solvents can induce significant changes in the intensity of peaks. Hyperchromic – Increase in absorption intensity. Hypochromic – Decrease in absorption intensity. Absorption characteristics of 2-methylpyridine λmax εmax Hexane 260 2000 Chloroform 263 4500 Ethanol 260 4000 Water 260 4000 Solvent Ethanol - HCl (1:1) 262 5200
  • 48.  π -> π* transitions leads to more polar excited state that is more easily stabilized by polar solvent associations (H-bonds). The π* state is more polar and stabilized more in polar solvent relative to nonpolar one, thus in going from nonpolar to polar solvent there is a red shift or bathochromic shift (increase in λmax, decrease in ΔE).  For n -> π* transition, the n state is much more easily stabilized by polar solvent effects (H-bonds and association), so in going from nonpolar to polar solvent there is a blue shift or hypsochromic shift (decrease in λmax, increase in ΔE).
  • 49. APPLICATIONS: A. APPLICATIONS IN ORGANIC COMPOUNDS 1.It is helps to show the relationship between different groups, it is useful to detect the conjugation of the compounds 2.Detection of geometrical isomers, In case of geometrical isomers compounds, that trans isomers exhibits λmax at slightly longer wavelength and have larger extinction coefficient then the cis isomers . 3.Detection of functional groups, it is possible to detect the presence of certain functional groups with the help of UV Spectrum. GENERAL APPLICATIONS: 1.Qualitative analysis, UV absorption spectroscopy can characterizes those type of compounds which absorb UV radiation. Identification is done by comparing the absorption spectrum with the spectra of known compound. 2. It is useful in Quantitative analysis of the compounds. 3. Detection of impurities, UV absorption spectroscopy is the one of the best method for detecting impurities in organic compounds.
  • 50. Tautomeric equilibrium, UV spectroscopy can be used to determine the percentage of various keto and enol forms present in tautomeric equilibrium. 5. Chemical kinetics, UV spectroscopy can be used to study the kinetics of reactions. 6. Molecular weight determination, molecular weights of compounds can be measured by spectroscopy. 7. Analysis of inorganic compounds. 8. Measuring concentration of solution, absorption band can also used to determine the concentration of compounds in a solution. 9. Inorganic chemistry, absorption spectra have been used in connection with many problems in inorganic chemistry. 10. It is useful to determine the structure of the chloral.
  • 51. QUALITATIVE ANALYSIS Pharmacopoeial identification of drug (1) By using absorbance & wavelength (2) By taking absorption ratio (3) Limit test (b)Structural analysis Quantitative analysis Quantitative analysis A)By using beer’s law and using absorptivity value By using reference standard Multiple standard method B)Single compound analysis direct analysis Using separation method After extraction after chromatographic separation Using column chromatography Using HPLC Indirect analysis a)Single compound without chromophore b) Drugs with chromophoric reagent 1.For analyte which absorb weakly in UV region 2.For avoiding interference
  • 52. 3.Improve selectivity of assay  4. Determination of composition of complex Mole ratio method Continuous variation method ( job curve method ) 5 . Study of kinetics Disadvantages:  Samples should be in solution. Mixture of substances poses difficult to analyse and requires prior separation. Interference from the sample’s matrix makes the measurement difficult .