The inner-shell electron is tightly bound to the nucleus. Thus, K-shell
electrons or inner-shell electrons have a higher binding energy than
outer-shell electrons. Tungsten is a large atom with 74 orbiting
electrons. As atomic complexity increases, the electrons in any given
shell are more tightly bound than in a less complex atom.
When the projectile electron ionizes a target atom by removing a K-
shell electron, a temporary electron void is produced in the K-shell. This
is a highly unnatural state for the target atom and is corrected by an
outer-shell electron falling into the void in the K shell. Tungsten, for
example, has electrons in shells out to the P shell, and when a K-shell
electron is ionized, its position can be filled with electrons from any of
the outer shells. The transition of an orbital electron from an outer shell
to an inner shell is accompanied by the emission of an X-ray.
X-ray Basics
The production of heat and characteristic X-rays involves interactions between the
projectile electrons and the electrons of target atoms. Another type of interaction in
which the projectile electron can lose its kinetic energy is an interaction with the nucleus
of a target atom. In this type of interaction the kinetic energy of the projectile electron is
converted into electromagnetic energy.
A projectile electron that completely avoids the orbital electrons as it passes through a
target atom may come sufficiently close to the nucleus of the atom to come under its
influence. Because the electron is negatively charged and the nucleus is positively
charged, there is an electrostatic force of attraction between them. The closer the
projectile electron gets to the nucleus, the more it is influenced by the electrostatic field
of the nucleus. This field is very strong because the nucleus contains many protons and
the distance between the nucleus and the projectile electron is very small. As the
projectile electron passes by the nucleus, it slows down, changes its course, and leaves
with reduced kinetic energy in a different direction. These types of X-rays are called
Bremsstrahlung X-rays. Bremsstrahlung is a German word for slowing down or braking.
No feedback of generated KV, mA/
open loop
High Skin Dose
Inferior image quality
Close loop regulations
No skin dose
Superior image quality
Micro-controller
Block Diagram Multi-pulse
DC
AC
DC
AC
DC
AC
Line Input
EMI Filter
Inverter High Voltage
Transformer
400 V » UDC = 566 V
IGBT: fres
= 30 - 100 kHz
U T
= 40 - 150 kV
I T
= 0,1 - 1000 mA
P Puls = 4 W - 100 kW
P 60s £ 40 kW
P 10min £ 10 kW
U Tube
ITube
IF
Tube Rotating Anode
Control
fan = 4 - 300 Hz IGBT
S = 10 - 20 kVA
PWM
fres
= 20 - 100 kHz
PH = 100 - 150 W
IGBT
3 »
UDC
mA Regulation/ Tube current
KV Regulation
Power Storage Generators
A grid
Diode
In brief
Hot Tube
HT of Line F Gen
HT of HF Gen
Light weight
Low Cost
KV Generation
Focusing cup
Voltage compensation at filament transformer
 A Crossed grid has two grids sandwiched together
with the result that the strips are usually both
parallel to and at right angles to the long
dimension of the grid.
 A focused grid is one in which the lead strips
are tilted progressively as they move away
from center. Lines through each of these strips
would converge at a point known as the grid
focus.
Grid Primary Transmission
The useful X-Rays which pass straight from
the X-Ray tube to the film are known as
Primary transmission is basically the
percentage of primary X-Rays which pass
through the grid and hit the film, and is thus
measured by using the same apparatus with
and without the grid in place.
Grid cutoff is an undesirable absorption of
primary x-ray beams by grid strips, which
prevents the useful x-rays from reaching the
image receptor. It is caused by improper
grid positioning and most often occurs with
parallel grids.
 The quantity of scattered radiation in an x-ray beam reaching a
receptor can be reduced by separating the patient's body and
receptor surface, as shown below. This separation is known as
an air gap. Scattered radiation leaving a patient's body is more
divergent than the primary x-ray beam. Therefore, scattered
radiation spreads out of the primary beam area. The reduction of
scattered radiation in proportion to primary radiation increases
with air-gap distance. Several factors must be considered when
using this method of scatter reduction. Patient exposure is increased
because of the inverse-square effect. The use of an air gap
introduces magnification. Therefore, a larger receptor size is
required to obtain the same patient area coverage. If the air gap is
obtained by increasing the tube-to-receptor distance, the x-ray
equipment must be operated at a higher output to obtain adequate
receptor exposure.
 A bucky is typically used for table or wall
mounted x-raysystems and holds the x-
ray cassette and grid. A bucky, is a device
found underneath the exam table, a drawer like
device that the cassette and grid is slid into
before shooting x-ray.
Two Pulse generators
Control Console
Control Panel
Image intensifier
Oil in x-ray tube and HV Transformer
for cooling and Insulation
That is stable at high temperatures and has
excellent electrical insulating properties Its
functions are
To insulate,
suppress corona ,
Supress arcing, and to serve as a coolant.
What is Arcing
An AEC system uses a physically thin
radiation
An AEC system uses a physically thin radiation ionization
detector "AEC detector" which is positioned between the
patient being x-rayed and the x-ray film cassette. X-rays
passing through the patient also pass through this "AEC
detector" before they strike the x-ray film.
A weak ionization signal from the AEC detector is
integrated as a ramp shaped voltage waveform. This ramp
signal rises until it matches a pre-set threshold. At this
point the x-ray exposure is terminated. AEC devices are
calibrated to ensure that similar exams have linearity in
exam densities. This is due to the fact that a milliamperage
station is no longer selected and instead relies upon the
ionization within the selected chambers

Mais conteúdo relacionado

Semelhante a BMET X-ray.pdf

UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdfUNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
devicaf983
 

Semelhante a BMET X-ray.pdf (20)

Linear Accelerators.pptx
Linear Accelerators.pptxLinear Accelerators.pptx
Linear Accelerators.pptx
 
X ray production (Emission and Filtration)
X ray production (Emission and Filtration)X ray production (Emission and Filtration)
X ray production (Emission and Filtration)
 
Radiation physics in Dental Radiology...
Radiation physics in Dental Radiology...Radiation physics in Dental Radiology...
Radiation physics in Dental Radiology...
 
UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdfUNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
 
X Ray Production
X Ray ProductionX Ray Production
X Ray Production
 
Production of xrays
Production of xraysProduction of xrays
Production of xrays
 
x-ray-lecture-1-1.pptx
x-ray-lecture-1-1.pptxx-ray-lecture-1-1.pptx
x-ray-lecture-1-1.pptx
 
Production of xray
Production  of  xrayProduction  of  xray
Production of xray
 
X ray physics
X ray physicsX ray physics
X ray physics
 
Medical-Imaging.pdf
Medical-Imaging.pdfMedical-Imaging.pdf
Medical-Imaging.pdf
 
Radiation physics
Radiation physicsRadiation physics
Radiation physics
 
Production and Emission of X-Rays - Sultan LeMarc
Production and Emission of X-Rays - Sultan LeMarcProduction and Emission of X-Rays - Sultan LeMarc
Production and Emission of X-Rays - Sultan LeMarc
 
1.pptx
1.pptx1.pptx
1.pptx
 
Xray production 2
Xray production 2Xray production 2
Xray production 2
 
X-ray production & emission
X-ray production & emissionX-ray production & emission
X-ray production & emission
 
Linac presentation
Linac presentationLinac presentation
Linac presentation
 
U.V Spectroscopy.
U.V Spectroscopy.U.V Spectroscopy.
U.V Spectroscopy.
 
Linear accelerator.pptx
Linear accelerator.pptxLinear accelerator.pptx
Linear accelerator.pptx
 
Basics of radiation and production of x rays
Basics of radiation and production of x raysBasics of radiation and production of x rays
Basics of radiation and production of x rays
 
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
X- Ray physics- X-Ray Tube, Transformer, Generator and Rectifiers by kajalsra...
 

Mais de Yadav Raj

Medical Laboratory Technologist janamaitri
Medical Laboratory Technologist janamaitriMedical Laboratory Technologist janamaitri
Medical Laboratory Technologist janamaitri
Yadav Raj
 
Cyclosporaparasitology medical laboratory technologist .pptx
Cyclosporaparasitology medical laboratory technologist  .pptxCyclosporaparasitology medical laboratory technologist  .pptx
Cyclosporaparasitology medical laboratory technologist .pptx
Yadav Raj
 
Antigen 28th June 2013.ppt
Antigen 28th June 2013.pptAntigen 28th June 2013.ppt
Antigen 28th June 2013.ppt
Yadav Raj
 
Biological Safety Cabinets-Safety training2.ppt
Biological Safety Cabinets-Safety training2.pptBiological Safety Cabinets-Safety training2.ppt
Biological Safety Cabinets-Safety training2.ppt
Yadav Raj
 
Temperature_Controlled_DC_Fan_Using_Ther.docx
Temperature_Controlled_DC_Fan_Using_Ther.docxTemperature_Controlled_DC_Fan_Using_Ther.docx
Temperature_Controlled_DC_Fan_Using_Ther.docx
Yadav Raj
 
CLS Blastomyces dermatitidis.pptx
CLS Blastomyces dermatitidis.pptxCLS Blastomyces dermatitidis.pptx
CLS Blastomyces dermatitidis.pptx
Yadav Raj
 
Antigens And Antibody To.pptx
Antigens And Antibody To.pptxAntigens And Antibody To.pptx
Antigens And Antibody To.pptx
Yadav Raj
 
Antigens And Antibody To.pptx
Antigens And Antibody To.pptxAntigens And Antibody To.pptx
Antigens And Antibody To.pptx
Yadav Raj
 

Mais de Yadav Raj (20)

Medical Laboratory Technologist janamaitri
Medical Laboratory Technologist janamaitriMedical Laboratory Technologist janamaitri
Medical Laboratory Technologist janamaitri
 
Cyclosporaparasitology medical laboratory technologist .pptx
Cyclosporaparasitology medical laboratory technologist  .pptxCyclosporaparasitology medical laboratory technologist  .pptx
Cyclosporaparasitology medical laboratory technologist .pptx
 
Antigen 28th June 2013.ppt
Antigen 28th June 2013.pptAntigen 28th June 2013.ppt
Antigen 28th June 2013.ppt
 
5. Paracoccidioidomyces.pptx
5. Paracoccidioidomyces.pptx5. Paracoccidioidomyces.pptx
5. Paracoccidioidomyces.pptx
 
blood components.pptx
blood components.pptxblood components.pptx
blood components.pptx
 
blood group systems.pptx
blood group systems.pptxblood group systems.pptx
blood group systems.pptx
 
Biological Safety Cabinets-Safety training2.ppt
Biological Safety Cabinets-Safety training2.pptBiological Safety Cabinets-Safety training2.ppt
Biological Safety Cabinets-Safety training2.ppt
 
Temperature_Controlled_DC_Fan_Using_Ther.docx
Temperature_Controlled_DC_Fan_Using_Ther.docxTemperature_Controlled_DC_Fan_Using_Ther.docx
Temperature_Controlled_DC_Fan_Using_Ther.docx
 
Mycobacteria.pptx
Mycobacteria.pptxMycobacteria.pptx
Mycobacteria.pptx
 
Mycobacteria.pptx
Mycobacteria.pptxMycobacteria.pptx
Mycobacteria.pptx
 
hormone final.pptx
hormone final.pptxhormone final.pptx
hormone final.pptx
 
BMI-1.3.ppt
BMI-1.3.pptBMI-1.3.ppt
BMI-1.3.ppt
 
CLS Mycetoma.pptx
CLS Mycetoma.pptxCLS Mycetoma.pptx
CLS Mycetoma.pptx
 
CLS Histoplasmosis.pptx
CLS Histoplasmosis.pptxCLS Histoplasmosis.pptx
CLS Histoplasmosis.pptx
 
CLS Coccidioidomyces.pptx
CLS Coccidioidomyces.pptxCLS Coccidioidomyces.pptx
CLS Coccidioidomyces.pptx
 
CLS Blastomyces dermatitidis.pptx
CLS Blastomyces dermatitidis.pptxCLS Blastomyces dermatitidis.pptx
CLS Blastomyces dermatitidis.pptx
 
Antigens And Antibody To.pptx
Antigens And Antibody To.pptxAntigens And Antibody To.pptx
Antigens And Antibody To.pptx
 
Antigens And Antibody To.pptx
Antigens And Antibody To.pptxAntigens And Antibody To.pptx
Antigens And Antibody To.pptx
 
Widal test 3rd yr.pptx
Widal test 3rd yr.pptxWidal test 3rd yr.pptx
Widal test 3rd yr.pptx
 
Widal test 3rd yr.pptx
Widal test 3rd yr.pptxWidal test 3rd yr.pptx
Widal test 3rd yr.pptx
 

Último

Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
adilkhan87451
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
9953056974 Low Rate Call Girls In Saket, Delhi NCR
 

Último (20)

Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service AvailableCall Girls Vadodara Just Call 8617370543 Top Class Call Girl Service Available
Call Girls Vadodara Just Call 8617370543 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Majestic ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Majestic ⟟  9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Majestic ⟟  9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Majestic ⟟ 9332606886 ⟟ Call Me For Genuine S...
 
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
Call Girls Hosur Just Call 9630942363 Top Class Call Girl Service Available
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
 
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Tirupati Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
Call Girls Vasai Virar Just Call 9630942363 Top Class Call Girl Service Avail...
 
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Rishikesh Just Call 8250077686 Top Class Call Girl Service Available
 
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
💕SONAM KUMAR💕Premium Call Girls Jaipur ↘️9257276172 ↙️One Night Stand With Lo...
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
 
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 8250077686 Top Class Call Girl Service Ava...
 
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
 
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7Call Girls in Gagan Vihar (delhi) call me [🔝  9953056974 🔝] escort service 24X7
Call Girls in Gagan Vihar (delhi) call me [🔝 9953056974 🔝] escort service 24X7
 
Call Girls Guntur Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Guntur  Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Guntur  Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Guntur Just Call 8250077686 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
Call Girls in Delhi Triveni Complex Escort Service(🔝))/WhatsApp 97111⇛47426
 
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
Premium Call Girls In Jaipur {8445551418} ❤️VVIP SEEMA Call Girl in Jaipur Ra...
 
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 

BMET X-ray.pdf

  • 1.
  • 2.
  • 3.
  • 4. The inner-shell electron is tightly bound to the nucleus. Thus, K-shell electrons or inner-shell electrons have a higher binding energy than outer-shell electrons. Tungsten is a large atom with 74 orbiting electrons. As atomic complexity increases, the electrons in any given shell are more tightly bound than in a less complex atom. When the projectile electron ionizes a target atom by removing a K- shell electron, a temporary electron void is produced in the K-shell. This is a highly unnatural state for the target atom and is corrected by an outer-shell electron falling into the void in the K shell. Tungsten, for example, has electrons in shells out to the P shell, and when a K-shell electron is ionized, its position can be filled with electrons from any of the outer shells. The transition of an orbital electron from an outer shell to an inner shell is accompanied by the emission of an X-ray. X-ray Basics
  • 5.
  • 6.
  • 7. The production of heat and characteristic X-rays involves interactions between the projectile electrons and the electrons of target atoms. Another type of interaction in which the projectile electron can lose its kinetic energy is an interaction with the nucleus of a target atom. In this type of interaction the kinetic energy of the projectile electron is converted into electromagnetic energy. A projectile electron that completely avoids the orbital electrons as it passes through a target atom may come sufficiently close to the nucleus of the atom to come under its influence. Because the electron is negatively charged and the nucleus is positively charged, there is an electrostatic force of attraction between them. The closer the projectile electron gets to the nucleus, the more it is influenced by the electrostatic field of the nucleus. This field is very strong because the nucleus contains many protons and the distance between the nucleus and the projectile electron is very small. As the projectile electron passes by the nucleus, it slows down, changes its course, and leaves with reduced kinetic energy in a different direction. These types of X-rays are called Bremsstrahlung X-rays. Bremsstrahlung is a German word for slowing down or braking.
  • 8.
  • 9.
  • 10.
  • 11.
  • 12.
  • 13.
  • 14.
  • 15.
  • 16.
  • 17. No feedback of generated KV, mA/ open loop High Skin Dose Inferior image quality
  • 18.
  • 19. Close loop regulations No skin dose Superior image quality Micro-controller
  • 21. DC AC DC AC DC AC Line Input EMI Filter Inverter High Voltage Transformer 400 V » UDC = 566 V IGBT: fres = 30 - 100 kHz U T = 40 - 150 kV I T = 0,1 - 1000 mA P Puls = 4 W - 100 kW P 60s £ 40 kW P 10min £ 10 kW U Tube ITube IF Tube Rotating Anode Control fan = 4 - 300 Hz IGBT S = 10 - 20 kVA PWM fres = 20 - 100 kHz PH = 100 - 150 W IGBT 3 » UDC
  • 25.
  • 27.
  • 28. Diode
  • 29.
  • 30.
  • 31.
  • 32.
  • 33.
  • 35.
  • 36.
  • 37.
  • 38.
  • 40.
  • 41.
  • 42. HT of Line F Gen HT of HF Gen Light weight Low Cost
  • 44.
  • 45.
  • 46.
  • 47.
  • 48.
  • 49.
  • 50.
  • 51.
  • 52.
  • 53.
  • 54. Focusing cup Voltage compensation at filament transformer
  • 55.
  • 56.
  • 57.
  • 58.
  • 59.
  • 60.
  • 61.
  • 62.
  • 63.
  • 64.
  • 65.
  • 66.
  • 67.
  • 68.
  • 69.
  • 70.
  • 71.
  • 72.
  • 73.
  • 74.
  • 75.
  • 76.
  • 77.  A Crossed grid has two grids sandwiched together with the result that the strips are usually both parallel to and at right angles to the long dimension of the grid.
  • 78.  A focused grid is one in which the lead strips are tilted progressively as they move away from center. Lines through each of these strips would converge at a point known as the grid focus.
  • 79.
  • 80.
  • 81.
  • 82. Grid Primary Transmission The useful X-Rays which pass straight from the X-Ray tube to the film are known as Primary transmission is basically the percentage of primary X-Rays which pass through the grid and hit the film, and is thus measured by using the same apparatus with and without the grid in place.
  • 83.
  • 84.
  • 85.
  • 86.
  • 87.
  • 88.
  • 89. Grid cutoff is an undesirable absorption of primary x-ray beams by grid strips, which prevents the useful x-rays from reaching the image receptor. It is caused by improper grid positioning and most often occurs with parallel grids.
  • 90.
  • 91.
  • 92.
  • 93.  The quantity of scattered radiation in an x-ray beam reaching a receptor can be reduced by separating the patient's body and receptor surface, as shown below. This separation is known as an air gap. Scattered radiation leaving a patient's body is more divergent than the primary x-ray beam. Therefore, scattered radiation spreads out of the primary beam area. The reduction of scattered radiation in proportion to primary radiation increases with air-gap distance. Several factors must be considered when using this method of scatter reduction. Patient exposure is increased because of the inverse-square effect. The use of an air gap introduces magnification. Therefore, a larger receptor size is required to obtain the same patient area coverage. If the air gap is obtained by increasing the tube-to-receptor distance, the x-ray equipment must be operated at a higher output to obtain adequate receptor exposure.
  • 94.
  • 95.  A bucky is typically used for table or wall mounted x-raysystems and holds the x- ray cassette and grid. A bucky, is a device found underneath the exam table, a drawer like device that the cassette and grid is slid into before shooting x-ray.
  • 96.
  • 97.
  • 98.
  • 99.
  • 100.
  • 103.
  • 104.
  • 106.
  • 107.
  • 109.
  • 110.
  • 111.
  • 112.
  • 113. Oil in x-ray tube and HV Transformer for cooling and Insulation That is stable at high temperatures and has excellent electrical insulating properties Its functions are To insulate, suppress corona , Supress arcing, and to serve as a coolant.
  • 114.
  • 116. An AEC system uses a physically thin radiation An AEC system uses a physically thin radiation ionization detector "AEC detector" which is positioned between the patient being x-rayed and the x-ray film cassette. X-rays passing through the patient also pass through this "AEC detector" before they strike the x-ray film. A weak ionization signal from the AEC detector is integrated as a ramp shaped voltage waveform. This ramp signal rises until it matches a pre-set threshold. At this point the x-ray exposure is terminated. AEC devices are calibrated to ensure that similar exams have linearity in exam densities. This is due to the fact that a milliamperage station is no longer selected and instead relies upon the ionization within the selected chambers