BASIC PRINCIPLE OF
GAMMA CAMERA
Upakar Paudel (B.Sc
MIT) UCMSTH
INTRODUCTION
• Nuclear medicine is a branch of medicine and medical imaging
that uses small amount of radiotracer/radiopharmaceuticals
to diagnose disease and treat disease.
• Nuclear medicine differs from other medical imaging
modalities in the sense that CT and MR scans are anatomically
based but nuclear medicine looks at physiology of all the
organ system. So, we can follow the physiological processes as
they occur in the living humans using these
radiopharmaceuticals through use of appropriate imaging
system.
HISTORY
 1895 : Discovery of x-ray by Roentgen
 1896 : Discovery of radioactivity by Bequerel
 1898 : Production of radium by Curie
 1927 :Use of radon to measure the blood transit
 1945 : Invention of nuclear reactor
 1951 : Rectilinear scanner to acquire images
 1958 : Invention of Anger camera
 1964 : Use of Tc-99m (I-131 only prior to 1964)
Tc-99m: metastable (T1/2 = 6.01 hr) pure γ decay (E = 140
keV), flexible for labeling.
I-131: electrons and 364 keV photons, thyroid disorders
only
 1970 : Derivation of image reconstruction algorithm for
tomography (CT, SPECT, PET)
The First Anger camera (1958)
Activity
• The quantity of radioactive material, expressed as the number
of radioactive atoms undergoing nuclear transformation per
unit time, is called activity (A).
Decay constant
• Number of atoms decaying per unit time is proportional to the
number of unstable atoms
• Constant of proportionality is the decay constant ()
– -dN/dt =  N
– A =  N
Half life
• Useful parameter related to the decay constant; defined as
the time required for the number of radioactive atoms in a
sample to decrease by one half.
• Physical half-life and decay constant are inversely related and
unique for each radionuclide
Nuclear transformation
• When the atomic nucleus undergoes spontaneous
transformation, called radioactive decay, radiation is emitted
If the daughter nucleus is stable, this spontaneous
transformation ends
If the daughter is unstable, the process continues until a
stable nuclide is reached
• Most radionuclides decay in one or more of the following
ways:
(a) alpha decay
(b) Beta decay (+or -)
(c) Gamma emission
energyntransitioHeYX 24
2
4A
2Z
A
Z  

energyβYX -A
1Z
A
Z   
energyβYX A
1-Z
A
Z  

energyYeX A
1-Z
-A
Z  
Alpha Decay
Beta minus decay
Beta plus decay
Electron capture decay
Gamma Emission
 Radionuclides emitting gamma radiation during their decay
are potential imaging agents for nuclear medicine providing
their gamma energies are between 80 and 200 KeV.
 This is the ideal energy range for gamma cameras since lower
energies undergo tissue absorption and higher energies are
not absorbed by thin NaI(Tl) detector used in gamma camera
construction.
 Gamma emission results from any radioactive decay and
radioactive decay followed by isomeric transition of
metastable radionuclide to daughter product.
 Clinical Value :
Gamma radiation within the energies 90 to 200 KeV is
optimal for gamma camera imaging.
Isomeric Transition
energyXX A
Z
Am
Z 
•During radioactive decay, a daughter may be formed in an
excited state.
•Gamma rays are emitted as the daughter nucleus
transitions from the excited state to a lower-energy state.
•Some excited states may have a half-lives ranging up to
more than 600 years.
•Clinical Value :
Those isotopes that decay by isomeric transition
provide a source of activity that is short lived yielding a
pure gamma emitter with a minimal patient radiation
dose.
Gamma Camera
• Developed by Hal Anger at Berkeley in 1957 therefore also
called Anger camera
• An electronic device that detects gamma rays emitted by
radio pharmaceutical (e.g technetium 99m (Tc-99m) that have
been introduced into the body as tracers.
• Once a radiopharmaceutical has been administered, it is
necessary to detect the gamma ray emissions in order to
attain the functional information.
• The position of the source of the radioactivity can be plotted
and displayed on a TV monitor or photographic film.
• Either digital or analog.
• Internal radiation is administered by means of a
pharmaceutical which is labeled with a radioactive isotope /
tracer / radiopharmaceutical, is either injected, ingested, or
inhaled.
• The radioactive isotope decays, resulting in the emission of
gamma rays.
• The Gamma camera collects gamma rays that are emitted
from within the patient, enabling us to reconstruct a picture
of where the gamma rays originated. From this, we can
determine how a particular organ or system is functioning.
• The gamma camera can be used in planar imaging to acquire
2-dimensional images
Gamma Camera
Components of Gamma Camera
1) The collimator
2) Detector crystal
3) Photomultiplier tube array
4) Position logic circuits
5) Gantry
Collimator
 Made of perforated or folded usually lead or tungsten and is
interposed between the patient and the scintillation crystal.
 Is about ½ - 2 inches thick slab.
 Allows the gamma camera to localize accurately the
radionuclide in the patient’s body.
 Performs this function by absorbing and stopping most
radiation except that arriving almost perpendicular to the
detector face.
 Of all the photons emitted by an administered
radiopharmaceutical, more than 99% are wasted and not
recorded by the gamma camera, less than 1% are used to
generate the desired image.
 Thus it is rate limiting step in the imaging chain of gamma
camera technology.
 Consist of single or many holes
 The lead walls between the holes are referred as septa.
Collimator
Types of Collimator
 Pin hole collimator
 Multi hole collimator
-parallel hole collimator
-converging collimator
-diverging collimator
Pin Hole Collimator
 Collimator consists of small pinhole aperture in a piece of
Lead, (most common),Tungsten, platinum aperture is located
at the end of a lead cone , typically 20-25cm from detector.
 Size of pinhole can be varied by using removable inserts.
 Radiation must pass through the pinhole aperture to be
imaged and the image is always inverted on the scintillation
crystal.
 Used primarily for magnification imaging of small objects
 Poor sensitivity.
 Are routinely used for very high resolution images of small
organs , such as thyroid and for certain skeletal regions such
as hips or wrists.
Increasing source-to-detector distance leads to :
1) increasing FOV
2) decreasing image size
3) lower R
4) lower sensitivity
Pin Hole Collimator
Multi Hole Collimator
 Most widely used multihole collimator in nuclear medicine
laboratories.
 Consists of thousands of parallel holes with la long axis
perpendicular to the plane of the scintillation crystal.
 The lead walls between the holes – septa.
 Holes may be round, square, triangular or hexagonal.
 Septa absorbs most gamma rays that do not emanate from
the direction of interest.
 For high energy gamma rays- thicker septa are used than for a
low energy rays.
 Septa are designed in such a way that septal penetration by
unwanted gamma rays does not exceed 10% to 25%.
 Collimators are available with different lengths and different
widths of septa.
 Longer the septa, the better is the resolution but lower is the
count rate for ( Sensitivity) for a given amount of radionuclide.
 The count rate is inversely proportional to the square of the
collimator hole length.
 There is an inherent compromise between the spatial
resolution and the efficiency ( sensitivity) of the collimator.
 If length of the septa decreased, the detected count rate
increases and resolution decreases.
 Size of object = size of image i.e. neither magnification nor
minification gain.
Multi Hole Collimator
 Most common designs are Low Energy All-Purpose (LEAP),
Low Energy High-Resolution (LEHR) of less than 140 KeV and
Medium of less than 260 KeV and High Energy collimators of
less than 400 KeV.
 LEAP collimators have holes with a large diameter. The
sensitivity is relatively high & resolution is moderate (average
sensitivity500,000 cpm for a 1-uCi source & resolution is
1.0cm at 10cm from the patent side of the collimator)
 LEHR collimators have higher resolution images than the LEAP.
Holes are smaller & deeper. The sensitivity is approx. 185,000
cpm for 1-uCi source, and the resolution is higher 0.65cm at
10cm from the patient side of the collimator.
Multi Hole Collimator
 Medium Energy Collimators are used for medium energy
photons of nuclides such as Krypton81, Gallium67,
Indium111.
 High Energy Collimators are used for Iodine131 and F-18FDG.
 These collimators have thicker septa than LEAP and LEHR
collimators (mainly used with Technetium 99m) in order to
reduce septal penetration by the higher energy photons.
Multi Hole Collimator
Converging Collimator
 Converging collimator have an array of tapered holes that
converge at a point (usually 50 cm) in front of collimator
(Focal point)
 This convergence forms a magnified image.
 Resolution(high at surface) and decreases with distance
 Sensitivity- slowly increases as source is moved from
collimator face to focal plane and then decreases.
 Good for imaging smaller objects.
Diverging Collimator
 Has an array of tapered holes that diverge from hypothetical
point behind crystal( 40-50 cm).
 Generally, the use of a diverging collimator increases the
imaged area by about 30% over that obtained with a parallel
hole collimator.
 The image so obtained is minified.
 Both the sensitivity and resolution worsens as the object of
interest moves away from the collimator.
 Used particularly on cameras with small crystal faces to image
large organs such as lungs.
Diverging Collimator
Specialized Collimator
 Fan beam collimator- hybrid of parallel and converging
collimator ,used in SPECT
 Multiple pinhole collimator -50% more sensitivity than
parallel collimator at same spatial resolution.
 Rotating slant hole collimator- variation of the Parallel hole
,has all tunnels slanted at a specific angle , generates an
oblique view for better visualization of an organ, which is
blocked by other parts of body & can be positioned close to
the body for the maximum gain in resolution.
Scintillation Detector
 Uses Sodium iodide crystals activated with thallium (0.1-0.4
mole %) coupled to PMT as detector .
 The crystal surface may be circular and up to about 22 inches(
10- 21.5 inches) in diameter or it may be square or
rectangular.
 Crystals are usually ¼ - 5/8 thick( usually 3/8 inch)
 The crystal has an aluminum housing that protects it from
moisture, extraneous light, and minor physical damage.
 Larger the crystal surface diameter, the large is the field of
view.
 Thicker the crystal, worse is the spatial resolution.
 With thinner crystal , the overall sensitivity (count) decreases
by 10 % but approx. 30% increase in spatial resolution.
Properties of scintillation crystal
 High efficiency for stopping gamma rays.
 Stopping should be without scatter.
 High conversion of gamma rays into light.
 Wave length of light should match response of PMT’s.
 Crystal should be transparent to the emitted light.
 Crystals should be mechanically robust.
 Length of scintillation should be short.
 High iodine content (Z=53) & high density
(3.67g/cm3)provides high QDE
 Is sensitive to photons of energy 50 to 300 keV
 High conversion efficiency , 13 % of deposited energy is
converted to light – best energy resolution.
 Sensitivity>85% at140 keV
 Transparent to its own scintillation emission
 Emits light very promptly, decay constant – 230 nsec
 Can be grown relatively inexpensively in large plates
 Spectral matching between the emitted light and the
sensitivity of the PMT
 However they are fragile and crack easily if subjected to rapid
temperature change
Properties of scintillation crystal
What happens within crystals..
 Unfortunately, only a small fraction of the energy lost by a
gamma ray is converted into light, typically 10%.
 Interaction of the gamma ray with the crystal may result in
ejection of an orbital electron (photoelectric absorption),
producing a pulse of fluorescent light (scintillation event)
proportional in intensity to the energy of the gamma ray.
 Some photons may pass through the crystal without causing
light event.
 Some scatter out and deposit some energy ( Compton
Scattering)
 In most thallium activated sodium iodide crystals, about 20-30
light photons are released for each KeV of energy absobed.
 About 30% of the light from each event reaches the PMTs so
amplification is necessary.
 Low-energy radionuclides do not show much difference in
sensitivity between the two. As the energy of the radioisotope
increases, the difference in sensitivity increases
 Modern camera have rectangular crystal (60x40 cm) which
provides increased FOV
 Is surrounded by highly reflective material TiO2 or Mag.Oxide
to maximize light output
 At higher photon energies (>=300 KeV) , needs large volume
of NaI(Tl) for adequate detection efficiency
What happens within crystals..
Light Guides
 While the front face and sides of the crystal are canned,
usually with aluminum sufficiently thin so as not to attenuate
the incoming gamma rays unduly, the rear crystal surface
needs a transparent interface between the crystal and PMTs.
 This is usually provided by a Pyrex optical plate or light guide
a few centimeters in thickness.
 Array of PMT is coupled optically to the back face of the
crystal with a silicon based adhesive or grease.
 Light Guides(Lucite light pipe)- employed between the
detector crystal and PM tubes.
 It increase light collection efficiency , improve the uniformity
of light collection as a function of position.
Photo Multiplier Tube
 Is an electronic vacuum tube containing a light sensitivity
photocathode, 10 to 12 electrodes called dynodes and an
anode
 While PMTs with a photocathode diameter of 3 inches are
used mainly, it is also necessary to use some 2 inch diameter
tubes.
 It performs two functions-conversion of light photon to
electrical signal & Signal amplification
 PMT is attached to the back of the crystal directly on the
crystal, connected to the crystal by light pipes or optically
coupled to the crystal with silicon like material.
 PMT detects and amplifies the signal.(106 -108)
 Gamma camera consists of array of PMT(1st gamma camera –
7 PMT)
 Cameras with 37,55,61,75 or 91 tube are common ( 40-100 ).
 The greater the no. of PMTs, the greater is the resolution.
 Number is determined by the size & shape of both crystal &
each individual PMT.
 More the number better is the spatial resolution & linearity
 With thicker crystal, PMTs are farther away from the
scintillation point and are unable to determine the
coordinates as accurately, thus reduced spatial resolution.
 Current tubes have hexagonal cross section to cover more
area for efficient detection of scintillation photons
Photo Multiplier Tube
 When a photon is absorbed by the crystal, a no of PMT
around the specific point will see the light & produce
electrical pulse .
 Amplitude of pulse in a given PMT is directly proportional to
amount of light received by photocathode
 PMT closest to scintillation event will give largest output pulse
 Localization of the event in the final image depends on the
amount of light sensed by each PMT and thus on the pattern
of PMT voltage output.
Position and Summing Circuit
 The summation signal for each scintillation event is then
formed by weighing the output of each tube.
 This signal has three components: spatial coordinates on x-
and y- axes as well as a signal (z) related to the intensity
(energy).
 The x- and y- coordinates may go directly to instrumentation
for display on the CRT or may be recorded in the computer.
 Z (energy) pulse is obtained by adding the signal from all PMTs
& is proportional to total energy deposited in crystal.
 The signal intensity is processed by the PHA.
Position and Summing Circuit
 The basic principle of the PHA is to discard signals from
background and scattered radiation or radiation from
interfering isotopes so that only photons known to come from
the photopeak of the isotope being imaged are recorded.
 The PHA discriminates between events occuring in the crystal
that will be displayed or stored in the computer and events
that will be rejected.
 The PHA can make this discrimination because the energy
deposited by a scintillation event in the crystal bears a linear
relation to the voltage signal emerging from the PMTs.
 The pulse height analyzer allows the operator to select only
the signals from those gammas in which the height of the Z
signal, that is, gamma ray energy, has a certain value or range
of values.
Pulse Height Analyzer (PHA)
 If many useful gammas are not to be excluded from the
image, a range of energies must be allowed through the PHA,
and typically a window equal to 20% of the peak energy value
is used
 For 99mTc with a gamma ray of 140 KeV those signals with
energies between 126 and 154 keV are judged to be
acceptable.
 When using radionuclides ( gallium 67) that emit gamma rays
at different energies, multiple window analyzers need to be
employed.
 Typically a maximum of three sets of windows is available.
 On newer cameras, the signal processing circuitry such as
preamplifiers and PHAs is located on the base of each PMY so
that there is little signal distortion between the camera head
and the console.
Pulse Height Analyzer (PHA)
 Most gamma camera allow for a fine adjustment known as
automatic peaking of the isotope.
 Occasionally, an asymmetric window is used to improve
resolution by eliminating some of the Compton scatter.
 Image exposure time is selected by console control and
usually compromises :
• a preset count, a preset time and a preset information density
for the image a accumulation.
Console Controls
Display
• The final x and y signals generated by the positional circuitry
are accepted by either an analog film formatter or digitized to
take part in computer display system.
• The analog film formatter uses a cathode ray tube that has an
extreamly fine dot dimension.
• The light from the dot is recorded on the flm directly on the
film to produce an image.
• Single dot represents a single gamma photon within the
chosen energy window.
• A good quality image requires at least 1 million of these dots
equivalent to 1 million of the accepted gamma events.
Summary
• A collimator accepts orthogonal gamma events from the
patient.
• A large NaI crystals is the scintillation detector.
• PM tubes forms a hexagonal or rectangular array on its
surface.
• The signals from those goes to charge amplifiers.
• The signals are logarithmically amplified and accepted by the
positional circuitary which computes x and y axes of gamma
events.
• PHA enables this signal if x and y signal is a photo peak event.
• The x and y signals forms the display.
References…
• Essentials of Nuclear Medicine Imaging, 5th
Edition, Fred A. Mettler & Milton J. Guiberteau.
• The Essential Physics of Medical Imaging, 2nd
Edition, Bushberg.
• The Physics of Diagnostic Imaging, 2nd Edition,
Dowsett, Kenny & Johnston
• Various websites.
Basic principle of gamma camera

Mais conteúdo relacionado

Mais procurados

Basic Pulse Sequences In MRI
Basic Pulse Sequences In MRIBasic Pulse Sequences In MRI
Basic Pulse Sequences In MRIUpakar Paudel
 
CT Physics
CT PhysicsCT Physics
CT PhysicsRMLIMS
 
Everything Radionuclides - Nuclear Medicine
Everything Radionuclides - Nuclear MedicineEverything Radionuclides - Nuclear Medicine
Everything Radionuclides - Nuclear MedicineVictor Ekpo
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectorssandip suman
 
CT Generation (Generation of CT)
CT Generation (Generation of CT)CT Generation (Generation of CT)
CT Generation (Generation of CT)Upakar Paudel
 
SPECT: Single Photon Emission Computed Tomography
SPECT: Single Photon Emission Computed TomographySPECT: Single Photon Emission Computed Tomography
SPECT: Single Photon Emission Computed TomographyVictor Ekpo
 
Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imgingSahith Reddy
 
Multi slice ct ppt
Multi slice ct pptMulti slice ct ppt
Multi slice ct pptAnand Rk
 
Digital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging SystemDigital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging SystemMuhammad Arif Afridi
 
Nuclear medicine 2
Nuclear medicine 2Nuclear medicine 2
Nuclear medicine 2Rad Tech
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionparthajyotidas11
 

Mais procurados (20)

CT History
CT HistoryCT History
CT History
 
Basic Pulse Sequences In MRI
Basic Pulse Sequences In MRIBasic Pulse Sequences In MRI
Basic Pulse Sequences In MRI
 
soft tissue radiography
soft tissue radiographysoft tissue radiography
soft tissue radiography
 
CT Physics
CT PhysicsCT Physics
CT Physics
 
Everything Radionuclides - Nuclear Medicine
Everything Radionuclides - Nuclear MedicineEverything Radionuclides - Nuclear Medicine
Everything Radionuclides - Nuclear Medicine
 
Nuclear imaging
Nuclear imagingNuclear imaging
Nuclear imaging
 
Ct tube and detectors
Ct tube and detectorsCt tube and detectors
Ct tube and detectors
 
Ct instrument
Ct instrumentCt instrument
Ct instrument
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
CT Generation (Generation of CT)
CT Generation (Generation of CT)CT Generation (Generation of CT)
CT Generation (Generation of CT)
 
SPECT: Single Photon Emission Computed Tomography
SPECT: Single Photon Emission Computed TomographySPECT: Single Photon Emission Computed Tomography
SPECT: Single Photon Emission Computed Tomography
 
Nuclear medicine
Nuclear medicineNuclear medicine
Nuclear medicine
 
Flouroscopic imging
Flouroscopic imgingFlouroscopic imging
Flouroscopic imging
 
Multi slice ct ppt
Multi slice ct pptMulti slice ct ppt
Multi slice ct ppt
 
Digital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging SystemDigital Fluoroscopy Imaging System
Digital Fluoroscopy Imaging System
 
Nuclear medicine 2
Nuclear medicine 2Nuclear medicine 2
Nuclear medicine 2
 
Automatic exposure control
Automatic exposure controlAutomatic exposure control
Automatic exposure control
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
 
CT detectors
CT detectorsCT detectors
CT detectors
 
Cobalt & linac
Cobalt & linacCobalt & linac
Cobalt & linac
 

Semelhante a Basic principle of gamma camera

GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptxGAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptxTaushifulHoque
 
Gamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfGamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfNehaMishra787239
 
Gamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfGamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfNehaMishra787239
 
Gamma Camera.pptx
Gamma Camera.pptxGamma Camera.pptx
Gamma Camera.pptxMISSCOM1
 
Intro to radiography 1_2(NDT)
Intro to radiography 1_2(NDT)Intro to radiography 1_2(NDT)
Intro to radiography 1_2(NDT)Ravi Shekhar
 
Nuclear imaging in dentistry
Nuclear imaging in dentistryNuclear imaging in dentistry
Nuclear imaging in dentistryDr Jinki Singha
 
Radiography Testing for Btech metallurgical and materirials science engineering
Radiography Testing for Btech metallurgical and materirials science engineeringRadiography Testing for Btech metallurgical and materirials science engineering
Radiography Testing for Btech metallurgical and materirials science engineeringshyamkumarrakoti1
 
GAMMA SPECT PET.pptx
GAMMA SPECT PET.pptxGAMMA SPECT PET.pptx
GAMMA SPECT PET.pptxVishwasLS
 
Nucleaer imaging
Nucleaer imagingNucleaer imaging
Nucleaer imagingRupal Patle
 
CT SCAN equipment.pptx
CT SCAN equipment.pptxCT SCAN equipment.pptx
CT SCAN equipment.pptxRohit Bansal
 
Physics of Nuclear Medicine, SPECT and PET.ppt
Physics of Nuclear Medicine, SPECT and PET.pptPhysics of Nuclear Medicine, SPECT and PET.ppt
Physics of Nuclear Medicine, SPECT and PET.pptHassan Chattha
 
Rad 206 p04
Rad 206 p04Rad 206 p04
Rad 206 p04sehlawi
 
Intro of co 60 unit
Intro of co 60 unitIntro of co 60 unit
Intro of co 60 unitNileshK8
 
Single Positron Emission Computed Tomography (SPECT)
Single Positron Emission Computed Tomography (SPECT)Single Positron Emission Computed Tomography (SPECT)
Single Positron Emission Computed Tomography (SPECT)shohel rana
 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentationMumba Chilimboyi
 

Semelhante a Basic principle of gamma camera (20)

GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptxGAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
GAMmA cajjjjjjjjjjjujjjjjjjjjjjjjjjjjjjjjjjjmera old.pptx
 
Gamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfGamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdf
 
Gamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdfGamma Camera by sameer ahmad.pdf
Gamma Camera by sameer ahmad.pdf
 
Muhammad yaseen
Muhammad yaseenMuhammad yaseen
Muhammad yaseen
 
Gamma Camera.pptx
Gamma Camera.pptxGamma Camera.pptx
Gamma Camera.pptx
 
Intro to radiography 1_2(NDT)
Intro to radiography 1_2(NDT)Intro to radiography 1_2(NDT)
Intro to radiography 1_2(NDT)
 
Nuclear imaging in dentistry
Nuclear imaging in dentistryNuclear imaging in dentistry
Nuclear imaging in dentistry
 
Radiography Testing.ppt
Radiography Testing.pptRadiography Testing.ppt
Radiography Testing.ppt
 
Radiography Testing for Btech metallurgical and materirials science engineering
Radiography Testing for Btech metallurgical and materirials science engineeringRadiography Testing for Btech metallurgical and materirials science engineering
Radiography Testing for Btech metallurgical and materirials science engineering
 
GAMMA SPECT PET.pptx
GAMMA SPECT PET.pptxGAMMA SPECT PET.pptx
GAMMA SPECT PET.pptx
 
Nucleaer imaging
Nucleaer imagingNucleaer imaging
Nucleaer imaging
 
CT SCAN equipment.pptx
CT SCAN equipment.pptxCT SCAN equipment.pptx
CT SCAN equipment.pptx
 
Physics of Nuclear Medicine, SPECT and PET.ppt
Physics of Nuclear Medicine, SPECT and PET.pptPhysics of Nuclear Medicine, SPECT and PET.ppt
Physics of Nuclear Medicine, SPECT and PET.ppt
 
Radio pharmaceuticals
Radio pharmaceuticalsRadio pharmaceuticals
Radio pharmaceuticals
 
Rad 206 p04
Rad 206 p04Rad 206 p04
Rad 206 p04
 
U.V Spectroscopy.
U.V Spectroscopy.U.V Spectroscopy.
U.V Spectroscopy.
 
Intro of co 60 unit
Intro of co 60 unitIntro of co 60 unit
Intro of co 60 unit
 
Single Positron Emission Computed Tomography (SPECT)
Single Positron Emission Computed Tomography (SPECT)Single Positron Emission Computed Tomography (SPECT)
Single Positron Emission Computed Tomography (SPECT)
 
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTSUnit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
Unit 1 ppt notes /BM8702 /RADIOLOGICAL EQUIPMENTS
 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentation
 

Mais de Upakar Paudel

Radiology Contrast media and its Types
Radiology Contrast media and its TypesRadiology Contrast media and its Types
Radiology Contrast media and its TypesUpakar Paudel
 
Measurement of Radiation (Thimble Ionization Chamber, Free air Ionization Cha...
Measurement of Radiation (Thimble Ionization Chamber, Free air Ionization Cha...Measurement of Radiation (Thimble Ionization Chamber, Free air Ionization Cha...
Measurement of Radiation (Thimble Ionization Chamber, Free air Ionization Cha...Upakar Paudel
 
CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis Upakar Paudel
 
HSG DCS and Sailography
HSG DCS  and SailographyHSG DCS  and Sailography
HSG DCS and SailographyUpakar Paudel
 
Sterilization & Disinfection
Sterilization &  DisinfectionSterilization &  Disinfection
Sterilization & DisinfectionUpakar Paudel
 
Abnormal sign in chest X- Ray
Abnormal sign in chest X- RayAbnormal sign in chest X- Ray
Abnormal sign in chest X- RayUpakar Paudel
 
Sensitometry or characteristic curve
Sensitometry or characteristic curve   Sensitometry or characteristic curve
Sensitometry or characteristic curve Upakar Paudel
 
CT Procedure of Thorax (CT Chest)
CT Procedure of Thorax (CT Chest)CT Procedure of Thorax (CT Chest)
CT Procedure of Thorax (CT Chest)Upakar Paudel
 
Recent Advances In MRI
Recent Advances In MRIRecent Advances In MRI
Recent Advances In MRIUpakar Paudel
 
How to write a thesis
How to write a thesisHow to write a thesis
How to write a thesisUpakar Paudel
 

Mais de Upakar Paudel (12)

Radiology Contrast media and its Types
Radiology Contrast media and its TypesRadiology Contrast media and its Types
Radiology Contrast media and its Types
 
Measurement of Radiation (Thimble Ionization Chamber, Free air Ionization Cha...
Measurement of Radiation (Thimble Ionization Chamber, Free air Ionization Cha...Measurement of Radiation (Thimble Ionization Chamber, Free air Ionization Cha...
Measurement of Radiation (Thimble Ionization Chamber, Free air Ionization Cha...
 
CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis
 
Digital Radiography
Digital RadiographyDigital Radiography
Digital Radiography
 
HSG DCS and Sailography
HSG DCS  and SailographyHSG DCS  and Sailography
HSG DCS and Sailography
 
Sterilization & Disinfection
Sterilization &  DisinfectionSterilization &  Disinfection
Sterilization & Disinfection
 
Abnormal sign in chest X- Ray
Abnormal sign in chest X- RayAbnormal sign in chest X- Ray
Abnormal sign in chest X- Ray
 
Sensitometry or characteristic curve
Sensitometry or characteristic curve   Sensitometry or characteristic curve
Sensitometry or characteristic curve
 
CT Procedure of Thorax (CT Chest)
CT Procedure of Thorax (CT Chest)CT Procedure of Thorax (CT Chest)
CT Procedure of Thorax (CT Chest)
 
Recent Advances In MRI
Recent Advances In MRIRecent Advances In MRI
Recent Advances In MRI
 
How to write a thesis
How to write a thesisHow to write a thesis
How to write a thesis
 
ERCP PROCEDURE
ERCP PROCEDURE ERCP PROCEDURE
ERCP PROCEDURE
 

Último

Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Deliverynehamumbai
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...narwatsonia7
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...Neha Kaur
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...aartirawatdelhi
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...jageshsingh5554
 
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service KochiLow Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service KochiSuhani Kapoor
 
Chandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableChandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableDipal Arora
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...CALL GIRLS
 

Último (20)

Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Aurangabad Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on DeliveryCall Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
Call Girls Colaba Mumbai ❤️ 9920874524 👈 Cash on Delivery
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
Top Rated Bangalore Call Girls Mg Road ⟟ 8250192130 ⟟ Call Me For Genuine Sex...
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
VIP Russian Call Girls in Varanasi Samaira 8250192130 Independent Escort Serv...
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
Night 7k to 12k Navi Mumbai Call Girl Photo 👉 BOOK NOW 9833363713 👈 ♀️ night ...
 
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
 
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 9907093804 Top Class Call Girl Service Available
 
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
VIP Service Call Girls Sindhi Colony 📳 7877925207 For 18+ VIP Call Girl At Th...
 
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service KochiLow Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
Low Rate Call Girls Kochi Anika 8250192130 Independent Escort Service Kochi
 
Chandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD availableChandrapur Call girls 8617370543 Provides all area service COD available
Chandrapur Call girls 8617370543 Provides all area service COD available
 
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
Call Girls Service Surat Samaira ❤️🍑 8250192130 👄 Independent Escort Service ...
 

Basic principle of gamma camera

  • 1. BASIC PRINCIPLE OF GAMMA CAMERA Upakar Paudel (B.Sc MIT) UCMSTH
  • 2. INTRODUCTION • Nuclear medicine is a branch of medicine and medical imaging that uses small amount of radiotracer/radiopharmaceuticals to diagnose disease and treat disease. • Nuclear medicine differs from other medical imaging modalities in the sense that CT and MR scans are anatomically based but nuclear medicine looks at physiology of all the organ system. So, we can follow the physiological processes as they occur in the living humans using these radiopharmaceuticals through use of appropriate imaging system.
  • 3. HISTORY  1895 : Discovery of x-ray by Roentgen  1896 : Discovery of radioactivity by Bequerel  1898 : Production of radium by Curie  1927 :Use of radon to measure the blood transit  1945 : Invention of nuclear reactor  1951 : Rectilinear scanner to acquire images  1958 : Invention of Anger camera  1964 : Use of Tc-99m (I-131 only prior to 1964) Tc-99m: metastable (T1/2 = 6.01 hr) pure γ decay (E = 140 keV), flexible for labeling. I-131: electrons and 364 keV photons, thyroid disorders only  1970 : Derivation of image reconstruction algorithm for tomography (CT, SPECT, PET)
  • 4. The First Anger camera (1958)
  • 5. Activity • The quantity of radioactive material, expressed as the number of radioactive atoms undergoing nuclear transformation per unit time, is called activity (A). Decay constant • Number of atoms decaying per unit time is proportional to the number of unstable atoms • Constant of proportionality is the decay constant () – -dN/dt =  N – A =  N
  • 6. Half life • Useful parameter related to the decay constant; defined as the time required for the number of radioactive atoms in a sample to decrease by one half. • Physical half-life and decay constant are inversely related and unique for each radionuclide
  • 7. Nuclear transformation • When the atomic nucleus undergoes spontaneous transformation, called radioactive decay, radiation is emitted If the daughter nucleus is stable, this spontaneous transformation ends If the daughter is unstable, the process continues until a stable nuclide is reached • Most radionuclides decay in one or more of the following ways: (a) alpha decay (b) Beta decay (+or -) (c) Gamma emission
  • 8. energyntransitioHeYX 24 2 4A 2Z A Z    energyβYX -A 1Z A Z    energyβYX A 1-Z A Z    energyYeX A 1-Z -A Z   Alpha Decay Beta minus decay Beta plus decay Electron capture decay
  • 9. Gamma Emission  Radionuclides emitting gamma radiation during their decay are potential imaging agents for nuclear medicine providing their gamma energies are between 80 and 200 KeV.  This is the ideal energy range for gamma cameras since lower energies undergo tissue absorption and higher energies are not absorbed by thin NaI(Tl) detector used in gamma camera construction.  Gamma emission results from any radioactive decay and radioactive decay followed by isomeric transition of metastable radionuclide to daughter product.  Clinical Value : Gamma radiation within the energies 90 to 200 KeV is optimal for gamma camera imaging.
  • 10. Isomeric Transition energyXX A Z Am Z  •During radioactive decay, a daughter may be formed in an excited state. •Gamma rays are emitted as the daughter nucleus transitions from the excited state to a lower-energy state. •Some excited states may have a half-lives ranging up to more than 600 years. •Clinical Value : Those isotopes that decay by isomeric transition provide a source of activity that is short lived yielding a pure gamma emitter with a minimal patient radiation dose.
  • 11. Gamma Camera • Developed by Hal Anger at Berkeley in 1957 therefore also called Anger camera • An electronic device that detects gamma rays emitted by radio pharmaceutical (e.g technetium 99m (Tc-99m) that have been introduced into the body as tracers. • Once a radiopharmaceutical has been administered, it is necessary to detect the gamma ray emissions in order to attain the functional information. • The position of the source of the radioactivity can be plotted and displayed on a TV monitor or photographic film. • Either digital or analog.
  • 12. • Internal radiation is administered by means of a pharmaceutical which is labeled with a radioactive isotope / tracer / radiopharmaceutical, is either injected, ingested, or inhaled. • The radioactive isotope decays, resulting in the emission of gamma rays. • The Gamma camera collects gamma rays that are emitted from within the patient, enabling us to reconstruct a picture of where the gamma rays originated. From this, we can determine how a particular organ or system is functioning. • The gamma camera can be used in planar imaging to acquire 2-dimensional images Gamma Camera
  • 13. Components of Gamma Camera 1) The collimator 2) Detector crystal 3) Photomultiplier tube array 4) Position logic circuits 5) Gantry
  • 14. Collimator  Made of perforated or folded usually lead or tungsten and is interposed between the patient and the scintillation crystal.  Is about ½ - 2 inches thick slab.  Allows the gamma camera to localize accurately the radionuclide in the patient’s body.  Performs this function by absorbing and stopping most radiation except that arriving almost perpendicular to the detector face.
  • 15.  Of all the photons emitted by an administered radiopharmaceutical, more than 99% are wasted and not recorded by the gamma camera, less than 1% are used to generate the desired image.  Thus it is rate limiting step in the imaging chain of gamma camera technology.  Consist of single or many holes  The lead walls between the holes are referred as septa. Collimator
  • 16. Types of Collimator  Pin hole collimator  Multi hole collimator -parallel hole collimator -converging collimator -diverging collimator
  • 17. Pin Hole Collimator  Collimator consists of small pinhole aperture in a piece of Lead, (most common),Tungsten, platinum aperture is located at the end of a lead cone , typically 20-25cm from detector.  Size of pinhole can be varied by using removable inserts.  Radiation must pass through the pinhole aperture to be imaged and the image is always inverted on the scintillation crystal.  Used primarily for magnification imaging of small objects  Poor sensitivity.  Are routinely used for very high resolution images of small organs , such as thyroid and for certain skeletal regions such as hips or wrists.
  • 18. Increasing source-to-detector distance leads to : 1) increasing FOV 2) decreasing image size 3) lower R 4) lower sensitivity Pin Hole Collimator
  • 19. Multi Hole Collimator  Most widely used multihole collimator in nuclear medicine laboratories.  Consists of thousands of parallel holes with la long axis perpendicular to the plane of the scintillation crystal.  The lead walls between the holes – septa.  Holes may be round, square, triangular or hexagonal.  Septa absorbs most gamma rays that do not emanate from the direction of interest.  For high energy gamma rays- thicker septa are used than for a low energy rays.  Septa are designed in such a way that septal penetration by unwanted gamma rays does not exceed 10% to 25%.
  • 20.  Collimators are available with different lengths and different widths of septa.  Longer the septa, the better is the resolution but lower is the count rate for ( Sensitivity) for a given amount of radionuclide.  The count rate is inversely proportional to the square of the collimator hole length.  There is an inherent compromise between the spatial resolution and the efficiency ( sensitivity) of the collimator.  If length of the septa decreased, the detected count rate increases and resolution decreases.  Size of object = size of image i.e. neither magnification nor minification gain. Multi Hole Collimator
  • 21.
  • 22.
  • 23.
  • 24.  Most common designs are Low Energy All-Purpose (LEAP), Low Energy High-Resolution (LEHR) of less than 140 KeV and Medium of less than 260 KeV and High Energy collimators of less than 400 KeV.  LEAP collimators have holes with a large diameter. The sensitivity is relatively high & resolution is moderate (average sensitivity500,000 cpm for a 1-uCi source & resolution is 1.0cm at 10cm from the patent side of the collimator)  LEHR collimators have higher resolution images than the LEAP. Holes are smaller & deeper. The sensitivity is approx. 185,000 cpm for 1-uCi source, and the resolution is higher 0.65cm at 10cm from the patient side of the collimator. Multi Hole Collimator
  • 25.  Medium Energy Collimators are used for medium energy photons of nuclides such as Krypton81, Gallium67, Indium111.  High Energy Collimators are used for Iodine131 and F-18FDG.  These collimators have thicker septa than LEAP and LEHR collimators (mainly used with Technetium 99m) in order to reduce septal penetration by the higher energy photons. Multi Hole Collimator
  • 26. Converging Collimator  Converging collimator have an array of tapered holes that converge at a point (usually 50 cm) in front of collimator (Focal point)  This convergence forms a magnified image.  Resolution(high at surface) and decreases with distance  Sensitivity- slowly increases as source is moved from collimator face to focal plane and then decreases.  Good for imaging smaller objects.
  • 27.
  • 28. Diverging Collimator  Has an array of tapered holes that diverge from hypothetical point behind crystal( 40-50 cm).  Generally, the use of a diverging collimator increases the imaged area by about 30% over that obtained with a parallel hole collimator.  The image so obtained is minified.  Both the sensitivity and resolution worsens as the object of interest moves away from the collimator.  Used particularly on cameras with small crystal faces to image large organs such as lungs.
  • 30. Specialized Collimator  Fan beam collimator- hybrid of parallel and converging collimator ,used in SPECT  Multiple pinhole collimator -50% more sensitivity than parallel collimator at same spatial resolution.  Rotating slant hole collimator- variation of the Parallel hole ,has all tunnels slanted at a specific angle , generates an oblique view for better visualization of an organ, which is blocked by other parts of body & can be positioned close to the body for the maximum gain in resolution.
  • 31. Scintillation Detector  Uses Sodium iodide crystals activated with thallium (0.1-0.4 mole %) coupled to PMT as detector .  The crystal surface may be circular and up to about 22 inches( 10- 21.5 inches) in diameter or it may be square or rectangular.  Crystals are usually ¼ - 5/8 thick( usually 3/8 inch)  The crystal has an aluminum housing that protects it from moisture, extraneous light, and minor physical damage.  Larger the crystal surface diameter, the large is the field of view.  Thicker the crystal, worse is the spatial resolution.  With thinner crystal , the overall sensitivity (count) decreases by 10 % but approx. 30% increase in spatial resolution.
  • 32. Properties of scintillation crystal  High efficiency for stopping gamma rays.  Stopping should be without scatter.  High conversion of gamma rays into light.  Wave length of light should match response of PMT’s.  Crystal should be transparent to the emitted light.  Crystals should be mechanically robust.  Length of scintillation should be short.  High iodine content (Z=53) & high density (3.67g/cm3)provides high QDE  Is sensitive to photons of energy 50 to 300 keV
  • 33.  High conversion efficiency , 13 % of deposited energy is converted to light – best energy resolution.  Sensitivity>85% at140 keV  Transparent to its own scintillation emission  Emits light very promptly, decay constant – 230 nsec  Can be grown relatively inexpensively in large plates  Spectral matching between the emitted light and the sensitivity of the PMT  However they are fragile and crack easily if subjected to rapid temperature change Properties of scintillation crystal
  • 34. What happens within crystals..  Unfortunately, only a small fraction of the energy lost by a gamma ray is converted into light, typically 10%.  Interaction of the gamma ray with the crystal may result in ejection of an orbital electron (photoelectric absorption), producing a pulse of fluorescent light (scintillation event) proportional in intensity to the energy of the gamma ray.  Some photons may pass through the crystal without causing light event.  Some scatter out and deposit some energy ( Compton Scattering)  In most thallium activated sodium iodide crystals, about 20-30 light photons are released for each KeV of energy absobed.
  • 35.  About 30% of the light from each event reaches the PMTs so amplification is necessary.  Low-energy radionuclides do not show much difference in sensitivity between the two. As the energy of the radioisotope increases, the difference in sensitivity increases  Modern camera have rectangular crystal (60x40 cm) which provides increased FOV  Is surrounded by highly reflective material TiO2 or Mag.Oxide to maximize light output  At higher photon energies (>=300 KeV) , needs large volume of NaI(Tl) for adequate detection efficiency What happens within crystals..
  • 36. Light Guides  While the front face and sides of the crystal are canned, usually with aluminum sufficiently thin so as not to attenuate the incoming gamma rays unduly, the rear crystal surface needs a transparent interface between the crystal and PMTs.  This is usually provided by a Pyrex optical plate or light guide a few centimeters in thickness.  Array of PMT is coupled optically to the back face of the crystal with a silicon based adhesive or grease.  Light Guides(Lucite light pipe)- employed between the detector crystal and PM tubes.  It increase light collection efficiency , improve the uniformity of light collection as a function of position.
  • 37. Photo Multiplier Tube  Is an electronic vacuum tube containing a light sensitivity photocathode, 10 to 12 electrodes called dynodes and an anode  While PMTs with a photocathode diameter of 3 inches are used mainly, it is also necessary to use some 2 inch diameter tubes.  It performs two functions-conversion of light photon to electrical signal & Signal amplification  PMT is attached to the back of the crystal directly on the crystal, connected to the crystal by light pipes or optically coupled to the crystal with silicon like material.  PMT detects and amplifies the signal.(106 -108)
  • 38.  Gamma camera consists of array of PMT(1st gamma camera – 7 PMT)  Cameras with 37,55,61,75 or 91 tube are common ( 40-100 ).  The greater the no. of PMTs, the greater is the resolution.  Number is determined by the size & shape of both crystal & each individual PMT.  More the number better is the spatial resolution & linearity  With thicker crystal, PMTs are farther away from the scintillation point and are unable to determine the coordinates as accurately, thus reduced spatial resolution.  Current tubes have hexagonal cross section to cover more area for efficient detection of scintillation photons Photo Multiplier Tube
  • 39.
  • 40.  When a photon is absorbed by the crystal, a no of PMT around the specific point will see the light & produce electrical pulse .  Amplitude of pulse in a given PMT is directly proportional to amount of light received by photocathode  PMT closest to scintillation event will give largest output pulse  Localization of the event in the final image depends on the amount of light sensed by each PMT and thus on the pattern of PMT voltage output. Position and Summing Circuit
  • 41.  The summation signal for each scintillation event is then formed by weighing the output of each tube.  This signal has three components: spatial coordinates on x- and y- axes as well as a signal (z) related to the intensity (energy).  The x- and y- coordinates may go directly to instrumentation for display on the CRT or may be recorded in the computer.  Z (energy) pulse is obtained by adding the signal from all PMTs & is proportional to total energy deposited in crystal.  The signal intensity is processed by the PHA. Position and Summing Circuit
  • 42.  The basic principle of the PHA is to discard signals from background and scattered radiation or radiation from interfering isotopes so that only photons known to come from the photopeak of the isotope being imaged are recorded.  The PHA discriminates between events occuring in the crystal that will be displayed or stored in the computer and events that will be rejected.  The PHA can make this discrimination because the energy deposited by a scintillation event in the crystal bears a linear relation to the voltage signal emerging from the PMTs.  The pulse height analyzer allows the operator to select only the signals from those gammas in which the height of the Z signal, that is, gamma ray energy, has a certain value or range of values. Pulse Height Analyzer (PHA)
  • 43.  If many useful gammas are not to be excluded from the image, a range of energies must be allowed through the PHA, and typically a window equal to 20% of the peak energy value is used  For 99mTc with a gamma ray of 140 KeV those signals with energies between 126 and 154 keV are judged to be acceptable.  When using radionuclides ( gallium 67) that emit gamma rays at different energies, multiple window analyzers need to be employed.  Typically a maximum of three sets of windows is available.  On newer cameras, the signal processing circuitry such as preamplifiers and PHAs is located on the base of each PMY so that there is little signal distortion between the camera head and the console. Pulse Height Analyzer (PHA)
  • 44.  Most gamma camera allow for a fine adjustment known as automatic peaking of the isotope.  Occasionally, an asymmetric window is used to improve resolution by eliminating some of the Compton scatter.  Image exposure time is selected by console control and usually compromises : • a preset count, a preset time and a preset information density for the image a accumulation. Console Controls
  • 45. Display • The final x and y signals generated by the positional circuitry are accepted by either an analog film formatter or digitized to take part in computer display system. • The analog film formatter uses a cathode ray tube that has an extreamly fine dot dimension. • The light from the dot is recorded on the flm directly on the film to produce an image. • Single dot represents a single gamma photon within the chosen energy window. • A good quality image requires at least 1 million of these dots equivalent to 1 million of the accepted gamma events.
  • 46. Summary • A collimator accepts orthogonal gamma events from the patient. • A large NaI crystals is the scintillation detector. • PM tubes forms a hexagonal or rectangular array on its surface. • The signals from those goes to charge amplifiers. • The signals are logarithmically amplified and accepted by the positional circuitary which computes x and y axes of gamma events. • PHA enables this signal if x and y signal is a photo peak event. • The x and y signals forms the display.
  • 47. References… • Essentials of Nuclear Medicine Imaging, 5th Edition, Fred A. Mettler & Milton J. Guiberteau. • The Essential Physics of Medical Imaging, 2nd Edition, Bushberg. • The Physics of Diagnostic Imaging, 2nd Edition, Dowsett, Kenny & Johnston • Various websites.