Introduction
to
Electron Microscopy
1
R. Basori
How to Study Nanomaterial ?
Processing
Synthesis
Properties
Applications
Micro/Nano
Structures
Characteriza-
tion
Materials
(Device
Fabrication)
2
Sample Characterization Techniques
Microscopic Techniques Spectroscopic Techniques
(Interaction of matter with
Electron)
(Interaction of matter with
Photon/light)
Scanning Electron Microscope (SEM)
Transmission Electron Microscope (TEM)
X-ray photoelectron spectroscopy (XPS)
 UV-Visible
FTIR, Raman etc….
3
Why electron microscopy?
4
The overall design of an electron
microscope is similar to that of a light
microscope
5
6
7
DIFFERENCES BETWEEN OM AND EM OPTICAL MICROSCOPE ELECTRON
8
Fundamentals of Electron Optics
Refraction and reflection of Electron beam when encountering the
region of potential difference.
( ) 1
( ) 2
Sin r V
Sin i V

An electron passing from a lower potential V1 to higher potential
V2 undergoes refraction as defined by Snell’s law.
i
r
V1 V1
V2 V2
i
V1 V1
V2 V2
V1<V2
V1>V2
r’
r
9
( ) 1
( ) 2
Sin r V
Sin i V

1
2
V
i Sin
V


1
2
V
i Sin
V


Refraction,
Reflection, r’ = i
Fundamentals of Electron Optics
Electron beam on passing through a region of potential difference
with V1<V2, experience a retardation making an angle of refraction
greater than incident angle.
10
Why Electron Microscope ?
Characteristic of a Microscope
 Magnification
 Resolution
 Depth of Focus
• Higher magnification
11
Magnification (M):
Size of image
Actual size of the object
M=25/f
12
Resolution: Power of resolving or identifying two closely spaced
particles separately
Limit of Resolution Where, NA numerical aperture of the
microscope
sin
NA n 

nrefractive index of the media
Resolving Power of a Microscope
R∞
1
D R high , D low  good microscope
D

0.61
D
NA


Optical microscope of λ=400 -700 nm, D>200
13
6.63 x10-34
Limit of resolution,
( ) D
D
D
Resolving Power
1
D

14
15
Depth of Field/Focus:
Another important aspect to resolution is the axial (or longitudinal) resolving
power of an objective, which is measured parallel to the optical axis and is most
often referred to as depth of field/focus.
When considering resolution in optical microscopy, a majority of the emphasis is placed
on point-to-point lateral resolution in the plane perpendicular to the optical axis.
It decreases with increased magnification. Sharpness of an
image depends upon the numerical aperture ‘NA’ and the
refractive index n of the lens material.
16
Optical Microscope Electron Microscope
Electron Microscope
Electron Microscope 17
K
L
M
Elastically
scattered electron
Inelastically
scattered electron
Unscattered electron
Nucleus
Electron shells
Ejected electrons
Electron beam
ΔE
E-ΔE
E
E
E
High-energy electron going through the sample
Interaction of unscattered
and elastically scattered
electrons will give image.
18
(Topography)
19
Scanning Electron
Microscope
20
A Scanning Electron Microscope (SEM) is an instrument for observing
and analyzing the surface microstructure of a bulk sample using a finely
focused beam of electrons.
SEM can achieve 1 – 4 nm imaging resolution depending on the
instruments.
Primary Applications:
•Surface topography / morphology
•Composition analysis
•Contrast
•And more ……
What is Scanning Electron Microscope ?
21
Basic components of electron microscopes
22
Rotor
Rotary Pump
Air is drawn into the pump through an inlet
and it is compressed.
Consists of a stationary part, Stator and a
moving part, Rotor, assembled inside a
casing.
This rotor is driven by an electric motor at
a constant speed.
Vacuum System :
23
Diffusion Pump
Vacuum System :
It consists of a chamber housing a oil vessel with a
heater, a chimney and a nozzle. On chamber's outer
surface, cooling coils carrying water are wound.
The heater vaporizes the oil and these hot vapors
rise into the vapor chimney.
The hot vapors are deflected downwards by an
annular nozzle or a jet assembly mounted at the top
of the chimney.
This jet, moving downwards at supersonic speeds, imparts momentum to
randomly moving gas molecules in the chamber.
This momentum deflects the molecules towards the pump outlet.
A backing pump is constantly used to remove the gas molecules.
The hot oil condenses on cold walls and returns to the vessel at bottom. 24
Turbo Molecular Pump
Vacuum System :
It consists of alternate layers of stator and rotor discs.
The rotor rotates at a very high RPM.
The blades are mounted at an optimum angle, on
both stator and rotor.
This high speed rotation imparts momentum to the
gas particles upon collision with the rotor discs.
The high speed molecules are directed towards
the exit using the stator discs.
25
26
Mean Free Path and Pressure (λ):
1 – 2 m
27
•“Gun Valve” separating the upper column from the rest of
the micro sample exchange; appropriate vacuum
•Use gloves when mounting samples and transferring
them to the column
•Samples are dry and free of excessive outgassing.
Precaution:
28
• There are three main types of gun –
Thermionic Gun - electrons are emitted from a filament when heated
sufficiently by passing current through it.
Field Emission Gun (FEG) - a very strong electric field is used to
extract electron from a material filament.
Thermal Field (Schottky) Emission Gun – electrons are ejected for
combined action of both heating and electric field.
Electron Source/Gun
29
Purpose:
Produce electrons
Roughly shape the beam
Set the electron energy (accelerating voltag
Thermionic Gun
•Filament is heated and begins to produce electrons.
•Electrons leave the filament tip with a negative potential so accelerate
towards the earthed anode and into the microscope column.
•A small negative bias on the Wehnelt then focuses the beam to a
crossover which acts as the electron source.
Example: Tungsten (W), Lanthanum hexaboride (LaB6)
30
Tungsten filament
assembly
Tungsten filament
LaB6 filament tip
• Can also use LaB6 crystal grown to a tip – gives a brighter
beam than W for same kV.
Thermionic Sources
Tungsten: Low work function
High Melting temperature ~ 3700 K
Emit electron at 2700K
31
•An extraction voltage of around 2kV is applied
to the first anode to create an intense electric
field to allow electrons to escape from the tip.
•The second anode is then used to accelerate
the electrons into the microscope at the
required energy.
•Combination of the two anodes focuses the beam into
a crossover creating a fine beam source.
• A very strong electric field (~109Vm-1) draws
electrons from a very fine metallic tip (usually
W).
FEG source – ZrO/W tip
Field Emission Gun (FEG)
32
Field Emission Gun (FEG)
Major Advantages:
•High resolution
•Very long potential source lifetime
• Low operating temperature (~300K)
• High brightness
• High current density (~1010 A/m2)
• Small source size < 0.01 um
• Highly spatially coherent, small energy
spread
Major Disadvantages:
•Small source size
•Not good for large area specimen, easy lose
current density
•Lowest maximum probe current.
•Poor current stability.
•Require ultra high vacuum in gun area.
In field emission, electrons tunnel through a
potential barrier, rather than escaping over it
as in thermionic emission.
33
Thermal Field (Schottky) Emission Gun
Combined heating and electric field (hybrid
source)
34
Comparison of Sources
35
36
Thermionic Sources
Major Advantages:
•High resolution
•Very long potential source lifetime
• Low operating temperature (~300K)
• High brightness
• High current density (~1010 A/m2)
• Small source size < 0.01 um
• Highly spatially coherent, small energy
spread
Major Disadvantages:
•Small source size
•Not good for large area specimen, easy lose
current density
•Lowest maximum probe current.
•Poor current stability.
•Require ultra high vacuum in gun area.
Field Emission Gun (FEG)
Thermal Field (Schottky)
Emission Gun
37
EM field exerts a force on a moving electron in a direction normal to both the field
and the propagation direction of the electrons.
Condenser Lens
Objective Lens
Two types of electromagnetic
lenses lenses in EM
V
V
38
Function of Condenser Lenses: Spot size
39
Function of Objective Lenses: Focus
40
Ideal and Actual lenses: Rays ideally focused to same plane. But, in reality
single focal plane does not occur.
41
Since the focal length ‘f’ of a lens is dependent on the strength of the
lens, it follows that different wavelengths will be focused to different
positions. Chromatic aberration of a lens is seen as fringes around the
image due to a “zone” of focus.
Lens Defects (Chromatic Aberration)
E +ΔE
42
The simplest way to correct for chromatic aberration is to use
illumination of a single wavelength!
Correction of Chromatic Aberration
In light optics chromatic aberration can be corrected by combining a
converging lens with a diverging lens. This is known as a “doublet” lens
Increase accelerating voltage which will
decrease spreading of electron energy
and make almost monochromatic
High accelerating voltage
43
Lens Defects (Spherical Aberration)
The fact that wavelengths enter and
leave the lens field at different angles
results in a defect known as spherical
aberration.
Peripheral Beam
Paraxial Beam
Aperture
Solution: Add a limiting aperture
44
Lens Defects (Aperture Diffraction)
45
Lens Defects (Astigmatism)
Reasons:
•Slight imperfection in manufacturing
pole pieces and copper winding
46
Astigmatism Correction:
Objective lend
adjustment
47
Apertures
Advantages
Increase contrast by blocking
scattered Electrons
Decrease effects of chromatic
and spherical aberration by
cutting off edges of a lens
Disadvantages
Decrease resolution due to
effects of diffraction
Decrease resolution by
reducing half angle of
illumination
Decrease illumination by
blocking scattered electrons
48
Scanning coils
49
Back scattered
electron detector
Secondary electron
detector
Source
Electromagnetic
lenses,
Apertures
etc.
Specimen
Detectors
50
51
(Topography)
52
53
to maintain the
conservation of energy and momentum
54
Incident beam
Auger Electron:
55
56
Interaction Volume
The ‘interaction volume’ is the area of the sample excited by the electron
beam to produce a signal.
The penetration of the electron beam into
the sample is affected by the accelerating
voltage used, the higher the kV the greater
the penetration.
The effective interaction volume can be
calculated using the electron range, R:-
1.67
0
0.89
0.0276
( )
AE
R m
Z



Where
A atomic weight (g/mole),
Z atomic number,
ρ density (in g/cm3)
Eo energy of the primary electron beam (in kV).
Accelerating voltage
(kV)
Primary Electron
Range (µm)
50 3.1
30 0.99
5 0.16
1 0.01 (10 nm)
Fe  A=55.85, Z=26, ρ=7.78 g/cm3
57
Interaction Volume (R)
58
1 kV 5 kV
15 kV
Accelerating Voltage
The dimensions of
the interaction
volume increases
with accelerating
voltage
The probability of
elastic scattering is
inversely proportional
to beam energy
59
Depth based information
Surface based information
Accelerating voltage reduction  reduced interaction volume  surface information 60
61
C, 6 Fe, 26
Au, 79
Atomic Number
The dimensions of the
interaction volume will
decrease with higher
atomic number
elements.
The rate of energy
loss of the electron
beam increases with
atomic number and
thus electrons do not
penetrate as deeply
into the sample
The probability of elastic
scattering increases with
A – causing the
interaction volume to
widen.
Accelerating voltage, 5kV
62
The angle at which the beam
strikes the specimen and the
distance from the surface are
important factors in how much of
signal escapes from the specimen.
63
Elements of different atomic number
SE or BSE emitted from different element are different , create
different contrast in image 64
65
Primary Beam
66
+
+
+
+
+
Faraday Cage
+300 V
67
Image Formation in SEM
68
69
‘Light’ region is made up predominantly of Fe.
(i.e. the heaviest element)
‘Grey’ region is made up predominantly of Ca.
‘Dark’ region is made up predominantly of Si
and Al. (i.e. the lightest elements)
Energy Dispersive Spectroscopy on the SEM
• There were 3 distinct regions in the Backscattered Image
Back scattered electrons are deflected more by heavier atoms leading to a brighter
contrast in BEI images – the lighter the region the heavier the element present.
The Y-axis shows the counts (number of X-rays
received and processed by the detector) and the
X-axis shows the energy level of those counts.
70
An electron microscope is operating with an accelerating voltage
~100KV under vacuum of ~10-8 mbar. If half angle of incident beam
cone is 0.01 radian, what will be the resolution of the electron
microscope? [sin(0.01rad)=0.00999, h=6.62x10-34m2 kg / s, e=1.6x10-19C
, m=9.1x10-31kg]
For electron energy of 10kV, calculate the interaction volume of Fe
having A=55.85, Z=26, ρ=7.78 g/cm3
Where
A atomic weight (g/mole),
Z atomic number,
ρ density (in g/cm3)
Eo energy of the primary electron beam (in kV).
71
Thank You !!
72

Mais conteúdo relacionado

Mais procurados

Scanning electron microscopy mubbu
Scanning electron microscopy mubbuScanning electron microscopy mubbu
Scanning electron microscopy mubbu
Mubashshir Arif
 
( EDAX)energy Dispersive xray spectroscopy
( EDAX)energy Dispersive xray spectroscopy ( EDAX)energy Dispersive xray spectroscopy
( EDAX)energy Dispersive xray spectroscopy
RushikeshGParit
 
Scanning tunneling microscope (STM)
 Scanning tunneling microscope (STM) Scanning tunneling microscope (STM)
Scanning tunneling microscope (STM)
Balsam Ata
 
Raman spectroscopy of carbon-nanotubes
Raman spectroscopy of carbon-nanotubesRaman spectroscopy of carbon-nanotubes
Raman spectroscopy of carbon-nanotubes
ajay singh
 

Mais procurados (20)

X ray photoelectron spectroscopy (xps)
X ray photoelectron spectroscopy (xps)X ray photoelectron spectroscopy (xps)
X ray photoelectron spectroscopy (xps)
 
Sputtering process
Sputtering processSputtering process
Sputtering process
 
Xps - X-Ray Photoelectron Spectroscopy
Xps - X-Ray Photoelectron SpectroscopyXps - X-Ray Photoelectron Spectroscopy
Xps - X-Ray Photoelectron Spectroscopy
 
X ray photoelectron spectroscopy
X ray photoelectron spectroscopyX ray photoelectron spectroscopy
X ray photoelectron spectroscopy
 
Scanning electron microscopy mubbu
Scanning electron microscopy mubbuScanning electron microscopy mubbu
Scanning electron microscopy mubbu
 
Transmission electron microscope, high resolution tem and selected area elect...
Transmission electron microscope, high resolution tem and selected area elect...Transmission electron microscope, high resolution tem and selected area elect...
Transmission electron microscope, high resolution tem and selected area elect...
 
Auger Electron Spectroscopy
Auger Electron SpectroscopyAuger Electron Spectroscopy
Auger Electron Spectroscopy
 
Scanning Electron Microscopy (SEM) lecture
Scanning Electron Microscopy (SEM) lectureScanning Electron Microscopy (SEM) lecture
Scanning Electron Microscopy (SEM) lecture
 
Optical properties of 2D materials
Optical properties of 2D materialsOptical properties of 2D materials
Optical properties of 2D materials
 
( EDAX)energy Dispersive xray spectroscopy
( EDAX)energy Dispersive xray spectroscopy ( EDAX)energy Dispersive xray spectroscopy
( EDAX)energy Dispersive xray spectroscopy
 
Materials Characterization Technique Lecture Notes
Materials Characterization Technique Lecture NotesMaterials Characterization Technique Lecture Notes
Materials Characterization Technique Lecture Notes
 
Scanning electron microscopy
Scanning electron microscopyScanning electron microscopy
Scanning electron microscopy
 
X ray photoelecton spectroscopy
X ray photoelecton spectroscopy X ray photoelecton spectroscopy
X ray photoelecton spectroscopy
 
Scanning tunneling microscope (STM)
 Scanning tunneling microscope (STM) Scanning tunneling microscope (STM)
Scanning tunneling microscope (STM)
 
TRANSMISSION ELECTRON MICROSCOPY
TRANSMISSION ELECTRON MICROSCOPYTRANSMISSION ELECTRON MICROSCOPY
TRANSMISSION ELECTRON MICROSCOPY
 
Raman spectroscopy of carbon-nanotubes
Raman spectroscopy of carbon-nanotubesRaman spectroscopy of carbon-nanotubes
Raman spectroscopy of carbon-nanotubes
 
Scanning Electron Microscope- Energy - Dispersive X -Ray Microanalysis (Sem E...
Scanning Electron Microscope- Energy - Dispersive X -Ray Microanalysis (Sem E...Scanning Electron Microscope- Energy - Dispersive X -Ray Microanalysis (Sem E...
Scanning Electron Microscope- Energy - Dispersive X -Ray Microanalysis (Sem E...
 
X-Ray Photo-electron Spectroscopy
X-Ray Photo-electron SpectroscopyX-Ray Photo-electron Spectroscopy
X-Ray Photo-electron Spectroscopy
 
Instrumentation presentation - Auger Electron Spectroscopy (AES)
Instrumentation presentation - Auger Electron Spectroscopy (AES)Instrumentation presentation - Auger Electron Spectroscopy (AES)
Instrumentation presentation - Auger Electron Spectroscopy (AES)
 
SEM,TEM & AFM
SEM,TEM & AFMSEM,TEM & AFM
SEM,TEM & AFM
 

Semelhante a Scanning Electron Microscope.pptx

Semelhante a Scanning Electron Microscope.pptx (20)

Transmission Electron Microscope
Transmission Electron MicroscopeTransmission Electron Microscope
Transmission Electron Microscope
 
Sem mahfooz
Sem mahfoozSem mahfooz
Sem mahfooz
 
Chapter-3. Electron Microscopy.pptx
Chapter-3. Electron Microscopy.pptxChapter-3. Electron Microscopy.pptx
Chapter-3. Electron Microscopy.pptx
 
Electron microscopy
Electron microscopyElectron microscopy
Electron microscopy
 
E mintro
E mintroE mintro
E mintro
 
2018 HM-scanning electron microscope
2018 HM-scanning electron microscope2018 HM-scanning electron microscope
2018 HM-scanning electron microscope
 
Electron microscopy
Electron microscopyElectron microscopy
Electron microscopy
 
WORKING PRINCIPLE OF SEM.pptx
WORKING PRINCIPLE OF SEM.pptxWORKING PRINCIPLE OF SEM.pptx
WORKING PRINCIPLE OF SEM.pptx
 
Scanning Electron Microscope
Scanning Electron Microscope Scanning Electron Microscope
Scanning Electron Microscope
 
ELECTRON MICROSCOPY (TEM & SEM)
ELECTRON MICROSCOPY (TEM & SEM)ELECTRON MICROSCOPY (TEM & SEM)
ELECTRON MICROSCOPY (TEM & SEM)
 
Dr nh mat.char notes
Dr nh   mat.char  notesDr nh   mat.char  notes
Dr nh mat.char notes
 
Sem Poster
Sem PosterSem Poster
Sem Poster
 
Sem md ppt
Sem md pptSem md ppt
Sem md ppt
 
Electron Microscopy
Electron MicroscopyElectron Microscopy
Electron Microscopy
 
Introduction to Electron Microscopyy.pdf
Introduction to Electron Microscopyy.pdfIntroduction to Electron Microscopyy.pdf
Introduction to Electron Microscopyy.pdf
 
Sem
SemSem
Sem
 
Sem
SemSem
Sem
 
Scanning Electron Microscopy - SEM
Scanning Electron Microscopy - SEM Scanning Electron Microscopy - SEM
Scanning Electron Microscopy - SEM
 
Electron Microscopy - Scanning electron microscope, Transmission Electron Mic...
Electron Microscopy - Scanning electron microscope, Transmission Electron Mic...Electron Microscopy - Scanning electron microscope, Transmission Electron Mic...
Electron Microscopy - Scanning electron microscope, Transmission Electron Mic...
 
scanning electron microscope (SEM)
scanning electron microscope (SEM)scanning electron microscope (SEM)
scanning electron microscope (SEM)
 

Último

Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global Impact
PECB
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
heathfieldcps1
 
An Overview of Mutual Funds Bcom Project.pdf
An Overview of Mutual Funds Bcom Project.pdfAn Overview of Mutual Funds Bcom Project.pdf
An Overview of Mutual Funds Bcom Project.pdf
SanaAli374401
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
QucHHunhnh
 
Gardella_Mateo_IntellectualProperty.pdf.
Gardella_Mateo_IntellectualProperty.pdf.Gardella_Mateo_IntellectualProperty.pdf.
Gardella_Mateo_IntellectualProperty.pdf.
MateoGardella
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
kauryashika82
 

Último (20)

Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global Impact
 
ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptx
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activity
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptx
 
Unit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxUnit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptx
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptxINDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
INDIA QUIZ 2024 RLAC DELHI UNIVERSITY.pptx
 
Advance Mobile Application Development class 07
Advance Mobile Application Development class 07Advance Mobile Application Development class 07
Advance Mobile Application Development class 07
 
An Overview of Mutual Funds Bcom Project.pdf
An Overview of Mutual Funds Bcom Project.pdfAn Overview of Mutual Funds Bcom Project.pdf
An Overview of Mutual Funds Bcom Project.pdf
 
Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptx
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across Sectors
 
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
Explore beautiful and ugly buildings. Mathematics helps us create beautiful d...
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot Graph
 
Gardella_Mateo_IntellectualProperty.pdf.
Gardella_Mateo_IntellectualProperty.pdf.Gardella_Mateo_IntellectualProperty.pdf.
Gardella_Mateo_IntellectualProperty.pdf.
 
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdf
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
 

Scanning Electron Microscope.pptx

  • 2. How to Study Nanomaterial ? Processing Synthesis Properties Applications Micro/Nano Structures Characteriza- tion Materials (Device Fabrication) 2
  • 3. Sample Characterization Techniques Microscopic Techniques Spectroscopic Techniques (Interaction of matter with Electron) (Interaction of matter with Photon/light) Scanning Electron Microscope (SEM) Transmission Electron Microscope (TEM) X-ray photoelectron spectroscopy (XPS)  UV-Visible FTIR, Raman etc…. 3
  • 5. The overall design of an electron microscope is similar to that of a light microscope 5
  • 6. 6
  • 7. 7 DIFFERENCES BETWEEN OM AND EM OPTICAL MICROSCOPE ELECTRON
  • 8. 8 Fundamentals of Electron Optics Refraction and reflection of Electron beam when encountering the region of potential difference. ( ) 1 ( ) 2 Sin r V Sin i V  An electron passing from a lower potential V1 to higher potential V2 undergoes refraction as defined by Snell’s law. i r V1 V1 V2 V2 i V1 V1 V2 V2 V1<V2 V1>V2 r’ r
  • 9. 9 ( ) 1 ( ) 2 Sin r V Sin i V  1 2 V i Sin V   1 2 V i Sin V   Refraction, Reflection, r’ = i Fundamentals of Electron Optics Electron beam on passing through a region of potential difference with V1<V2, experience a retardation making an angle of refraction greater than incident angle.
  • 10. 10
  • 11. Why Electron Microscope ? Characteristic of a Microscope  Magnification  Resolution  Depth of Focus • Higher magnification 11
  • 12. Magnification (M): Size of image Actual size of the object M=25/f 12
  • 13. Resolution: Power of resolving or identifying two closely spaced particles separately Limit of Resolution Where, NA numerical aperture of the microscope sin NA n   nrefractive index of the media Resolving Power of a Microscope R∞ 1 D R high , D low  good microscope D  0.61 D NA   Optical microscope of λ=400 -700 nm, D>200 13
  • 14. 6.63 x10-34 Limit of resolution, ( ) D D D Resolving Power 1 D  14
  • 15. 15
  • 16. Depth of Field/Focus: Another important aspect to resolution is the axial (or longitudinal) resolving power of an objective, which is measured parallel to the optical axis and is most often referred to as depth of field/focus. When considering resolution in optical microscopy, a majority of the emphasis is placed on point-to-point lateral resolution in the plane perpendicular to the optical axis. It decreases with increased magnification. Sharpness of an image depends upon the numerical aperture ‘NA’ and the refractive index n of the lens material. 16
  • 17. Optical Microscope Electron Microscope Electron Microscope Electron Microscope 17
  • 18. K L M Elastically scattered electron Inelastically scattered electron Unscattered electron Nucleus Electron shells Ejected electrons Electron beam ΔE E-ΔE E E E High-energy electron going through the sample Interaction of unscattered and elastically scattered electrons will give image. 18
  • 21. A Scanning Electron Microscope (SEM) is an instrument for observing and analyzing the surface microstructure of a bulk sample using a finely focused beam of electrons. SEM can achieve 1 – 4 nm imaging resolution depending on the instruments. Primary Applications: •Surface topography / morphology •Composition analysis •Contrast •And more …… What is Scanning Electron Microscope ? 21
  • 22. Basic components of electron microscopes 22
  • 23. Rotor Rotary Pump Air is drawn into the pump through an inlet and it is compressed. Consists of a stationary part, Stator and a moving part, Rotor, assembled inside a casing. This rotor is driven by an electric motor at a constant speed. Vacuum System : 23
  • 24. Diffusion Pump Vacuum System : It consists of a chamber housing a oil vessel with a heater, a chimney and a nozzle. On chamber's outer surface, cooling coils carrying water are wound. The heater vaporizes the oil and these hot vapors rise into the vapor chimney. The hot vapors are deflected downwards by an annular nozzle or a jet assembly mounted at the top of the chimney. This jet, moving downwards at supersonic speeds, imparts momentum to randomly moving gas molecules in the chamber. This momentum deflects the molecules towards the pump outlet. A backing pump is constantly used to remove the gas molecules. The hot oil condenses on cold walls and returns to the vessel at bottom. 24
  • 25. Turbo Molecular Pump Vacuum System : It consists of alternate layers of stator and rotor discs. The rotor rotates at a very high RPM. The blades are mounted at an optimum angle, on both stator and rotor. This high speed rotation imparts momentum to the gas particles upon collision with the rotor discs. The high speed molecules are directed towards the exit using the stator discs. 25
  • 26. 26
  • 27. Mean Free Path and Pressure (λ): 1 – 2 m 27
  • 28. •“Gun Valve” separating the upper column from the rest of the micro sample exchange; appropriate vacuum •Use gloves when mounting samples and transferring them to the column •Samples are dry and free of excessive outgassing. Precaution: 28
  • 29. • There are three main types of gun – Thermionic Gun - electrons are emitted from a filament when heated sufficiently by passing current through it. Field Emission Gun (FEG) - a very strong electric field is used to extract electron from a material filament. Thermal Field (Schottky) Emission Gun – electrons are ejected for combined action of both heating and electric field. Electron Source/Gun 29 Purpose: Produce electrons Roughly shape the beam Set the electron energy (accelerating voltag
  • 30. Thermionic Gun •Filament is heated and begins to produce electrons. •Electrons leave the filament tip with a negative potential so accelerate towards the earthed anode and into the microscope column. •A small negative bias on the Wehnelt then focuses the beam to a crossover which acts as the electron source. Example: Tungsten (W), Lanthanum hexaboride (LaB6) 30
  • 31. Tungsten filament assembly Tungsten filament LaB6 filament tip • Can also use LaB6 crystal grown to a tip – gives a brighter beam than W for same kV. Thermionic Sources Tungsten: Low work function High Melting temperature ~ 3700 K Emit electron at 2700K 31
  • 32. •An extraction voltage of around 2kV is applied to the first anode to create an intense electric field to allow electrons to escape from the tip. •The second anode is then used to accelerate the electrons into the microscope at the required energy. •Combination of the two anodes focuses the beam into a crossover creating a fine beam source. • A very strong electric field (~109Vm-1) draws electrons from a very fine metallic tip (usually W). FEG source – ZrO/W tip Field Emission Gun (FEG) 32
  • 33. Field Emission Gun (FEG) Major Advantages: •High resolution •Very long potential source lifetime • Low operating temperature (~300K) • High brightness • High current density (~1010 A/m2) • Small source size < 0.01 um • Highly spatially coherent, small energy spread Major Disadvantages: •Small source size •Not good for large area specimen, easy lose current density •Lowest maximum probe current. •Poor current stability. •Require ultra high vacuum in gun area. In field emission, electrons tunnel through a potential barrier, rather than escaping over it as in thermionic emission. 33
  • 34. Thermal Field (Schottky) Emission Gun Combined heating and electric field (hybrid source) 34
  • 36. 36 Thermionic Sources Major Advantages: •High resolution •Very long potential source lifetime • Low operating temperature (~300K) • High brightness • High current density (~1010 A/m2) • Small source size < 0.01 um • Highly spatially coherent, small energy spread Major Disadvantages: •Small source size •Not good for large area specimen, easy lose current density •Lowest maximum probe current. •Poor current stability. •Require ultra high vacuum in gun area. Field Emission Gun (FEG) Thermal Field (Schottky) Emission Gun
  • 37. 37 EM field exerts a force on a moving electron in a direction normal to both the field and the propagation direction of the electrons.
  • 38. Condenser Lens Objective Lens Two types of electromagnetic lenses lenses in EM V V 38
  • 39. Function of Condenser Lenses: Spot size 39
  • 40. Function of Objective Lenses: Focus 40
  • 41. Ideal and Actual lenses: Rays ideally focused to same plane. But, in reality single focal plane does not occur. 41
  • 42. Since the focal length ‘f’ of a lens is dependent on the strength of the lens, it follows that different wavelengths will be focused to different positions. Chromatic aberration of a lens is seen as fringes around the image due to a “zone” of focus. Lens Defects (Chromatic Aberration) E +ΔE 42
  • 43. The simplest way to correct for chromatic aberration is to use illumination of a single wavelength! Correction of Chromatic Aberration In light optics chromatic aberration can be corrected by combining a converging lens with a diverging lens. This is known as a “doublet” lens Increase accelerating voltage which will decrease spreading of electron energy and make almost monochromatic High accelerating voltage 43
  • 44. Lens Defects (Spherical Aberration) The fact that wavelengths enter and leave the lens field at different angles results in a defect known as spherical aberration. Peripheral Beam Paraxial Beam Aperture Solution: Add a limiting aperture 44
  • 45. Lens Defects (Aperture Diffraction) 45
  • 46. Lens Defects (Astigmatism) Reasons: •Slight imperfection in manufacturing pole pieces and copper winding 46
  • 48. Apertures Advantages Increase contrast by blocking scattered Electrons Decrease effects of chromatic and spherical aberration by cutting off edges of a lens Disadvantages Decrease resolution due to effects of diffraction Decrease resolution by reducing half angle of illumination Decrease illumination by blocking scattered electrons 48
  • 50. Back scattered electron detector Secondary electron detector Source Electromagnetic lenses, Apertures etc. Specimen Detectors 50
  • 51. 51
  • 53. 53
  • 54. to maintain the conservation of energy and momentum 54
  • 56. 56
  • 57. Interaction Volume The ‘interaction volume’ is the area of the sample excited by the electron beam to produce a signal. The penetration of the electron beam into the sample is affected by the accelerating voltage used, the higher the kV the greater the penetration. The effective interaction volume can be calculated using the electron range, R:- 1.67 0 0.89 0.0276 ( ) AE R m Z    Where A atomic weight (g/mole), Z atomic number, ρ density (in g/cm3) Eo energy of the primary electron beam (in kV). Accelerating voltage (kV) Primary Electron Range (µm) 50 3.1 30 0.99 5 0.16 1 0.01 (10 nm) Fe  A=55.85, Z=26, ρ=7.78 g/cm3 57
  • 59. 1 kV 5 kV 15 kV Accelerating Voltage The dimensions of the interaction volume increases with accelerating voltage The probability of elastic scattering is inversely proportional to beam energy 59
  • 60. Depth based information Surface based information Accelerating voltage reduction  reduced interaction volume  surface information 60
  • 61. 61
  • 62. C, 6 Fe, 26 Au, 79 Atomic Number The dimensions of the interaction volume will decrease with higher atomic number elements. The rate of energy loss of the electron beam increases with atomic number and thus electrons do not penetrate as deeply into the sample The probability of elastic scattering increases with A – causing the interaction volume to widen. Accelerating voltage, 5kV 62
  • 63. The angle at which the beam strikes the specimen and the distance from the surface are important factors in how much of signal escapes from the specimen. 63
  • 64. Elements of different atomic number SE or BSE emitted from different element are different , create different contrast in image 64
  • 65. 65
  • 69. 69
  • 70. ‘Light’ region is made up predominantly of Fe. (i.e. the heaviest element) ‘Grey’ region is made up predominantly of Ca. ‘Dark’ region is made up predominantly of Si and Al. (i.e. the lightest elements) Energy Dispersive Spectroscopy on the SEM • There were 3 distinct regions in the Backscattered Image Back scattered electrons are deflected more by heavier atoms leading to a brighter contrast in BEI images – the lighter the region the heavier the element present. The Y-axis shows the counts (number of X-rays received and processed by the detector) and the X-axis shows the energy level of those counts. 70
  • 71. An electron microscope is operating with an accelerating voltage ~100KV under vacuum of ~10-8 mbar. If half angle of incident beam cone is 0.01 radian, what will be the resolution of the electron microscope? [sin(0.01rad)=0.00999, h=6.62x10-34m2 kg / s, e=1.6x10-19C , m=9.1x10-31kg] For electron energy of 10kV, calculate the interaction volume of Fe having A=55.85, Z=26, ρ=7.78 g/cm3 Where A atomic weight (g/mole), Z atomic number, ρ density (in g/cm3) Eo energy of the primary electron beam (in kV). 71

Notas do Editor

  1. As the term says we are utilizing transmission mode of the electrons to find out the features inside the materials. So in this particular case, we wants the material to be transparent to the electron so that electrons get transmitted through the material and provide information about the materials.
  2. Wehnelt electrode - an electrode in the electron gun assembly of some thermionic devices, used for focusing and control of the electron beam. A Wehnelt acts as a control grid and it also serves as a convergent electrostatic lens. An electron emitter is positioned directly above the Wehnelt aperture, and an anode is located below the Wehnelt. The anode is biased to a high positive voltage (typically +1 to +30 kV) relative to the emitter so as to accelerate electrons from the emitter towards the anode, thus creating an electron beam that passes through the Wehnelt aperture.
  3. After coating tungsten tip with zirconium oxide operating temperature of tungsten reduces from 2700 K to 1800 K.
  4. To get high field strength with low applied bias, field emitting tips are made sharp.
  5. Changing converging point of the beam, some part of the beam is going to be cut and we will get small spot size i. e. low current and high resolution.
  6. The second lens that we have is called objective lens. Based on changing the lens strength we can change the poit at which the system focuses.
  7. Energy dispersive x-ray spectroscopy is techniques used for elemental analysis or chemical composition of the elements