Differential Scanning Calorimetry (DSC) is one of the important thermal analytical techniques in which specific physical properties of a material are measures as a function of temperature. It is used both in qualitative and quantitative analysis.
DSC is a technique for measuring the energy necessary to establish a nearly zero temperature difference between a substance and an inert reference material as the two specimens are subjected to identical temperature regimens in an environment heated or cooled at a controlled rate.
This technique was developed by E.S.Watson and M.J.O' Neill in 1964.
The device used to measure this is Calorimeter.
There are two types of DSC systems commonly used:
1. Power compensated DSC
2. Heat -flux DSC
A High resolution of PC-DSC is nowadays widely used known as Hyper DSC.
2. INTRODUCTION
Thermal Techniques
Thermal analysis includes a group of techniques in which specific physical properties of a material
are measured as a function of temperature.
Measurements are usually made with increasing temperature, but measurements made with
decreasing temperatures are also possible.
It is useful in both quantitative an qualitative analysis.
Samples may be identified and characterized by qualitative investigations of their thermal behavior.
Quantitative results are obtained from changes in weight and enthalpy as the sample is heated.
The temperatures of phase changes and reactions as well as heats of reaction are used to determine
the purity of materials.
4. DIFFERENTIAL SCANNING CALORIMETRY
DSC is a technique to establish a nearly zero temperature difference between a
sample and reference is measured as a function of temperature when both are
heated or cooled at a controlled rate.
The technique was developed by E.S. Watson and M.J.O’ Neill in 1964.
The DSC, at present, is the most widely used thermal analysis in the
pharmaceutical field.
In DSC, since power is measured directly, the curves are used quantitatively for the
measurement of the heat of reaction and specific heat etc.
Calorimeter- it is a device used for measuring the heat of chemical reactions or
physical changes as well as heat capacities.
5. What does DSC measures?
DSC measures the amount of energy (heat) absorbed or released by a sample as it is heated or cooled or held at
constant temperature. It also performs precise temperature measurements. DSC is used to analyze:
1. Melting point
2. Crystallization
3. Glass transition
6. 1. Glass transition
Transition from disordered solid to liquid.
Observed in glassy solids.
Tg – glass transition temperature.
2. Crystallization
Material can crystallize.
Tc- crystallization temperature.
3. Melting
Ordered to disordered transition.
Tm- melting temperature.
7. Principles of DSC
The difference in heat supplied to the sample, and the reference material per unit time is recorded and
plotted as dH/dt vs the average temperature to which the sample and reference to be raised.
Endothermic heat flow
Heat flows into the sample (energy required) as a result of either heat capacity (heating) or some
endothermic process (glass transition, melting, evaporation, etc.).
Upward peak in thermogram.
Exothermic heat flow
heat flows out of the sample (energy released) as a result of either heat capacity (cooling) or some
exothermic process (crystallization, oxidation, etc.).
Downward peak in thermogram.
8. INSTRUMENTATION
HEAT-FLUX DSC POWER-COMPENSATION DSC
Sample holders- platinum, aluminum and
stainless steel.
Sample holders- platinum, aluminum and
stainless-steel pans.
Sensors- temperature sensors.
Usually thermocouples which are same for both
sample and reference.
Sensors- platinum resistance thermocouple.
Separate sensors and heaters for both sample and
reference.
Furnace- one block for both reference and
sample cell.
Furnace- separate block for both reference and
sample cell.
• Amplifier- to amplify the signals. Amplifier- to amplify the signals.
• Recorder- provides graph. Recorder- provides graph.
9. HEAT-FLUX DSC
In heat-flux DSC, the difference in heat flow into the
sample and reference is measured while the sample
temperature is changed at a constant rate.
In it, the sample and reference in separate containers
are placed on separate platforms which settle on a
heated metal disc usually made up of Cu-Ni alloy.
Thermocouple are used for monitoring the heat
flow from the metal disc to the sample and
reference.
Inert gas is provided in the system.
11. POWER-COMPENSATION DSC
In power-compensated DSC, the temperatures of the sample and reference are kept equal to each other while both
temperatures are increased or decreased linearly.
It has lower sensitivity than heat-flux DSC, but its response time is more rapid.
It has two independent furnaces- one for heating sample and other for heating reference.
The furnaces are inserted into a large temperature control heat sink.
The power supply is varied which is required to ensure that ΔT= 0 at all times. (temperature is constant).
A thermogram is then produced by noting the difference in power supplied to the two heaters (ΔE).
12.
13. HYPER DSC
The high resolution of PC-DSC or new type of power-compensation DSC provides the best results for an
analysis of melting and crystallization of metals or detection of glass transition temperature (Tg) in
medications.
Fast scan DSC has the ability to perform valid heat flow measurements with fast linear controlled rates
especially by cooling, where the rates are higher than classical PC-DSC..
The benefits of such devices are increased sensitivity at higher rates (which enables a better study of kinetics
in the process), suppression of undesired transformation etc.
It empowers the superior investigation of energy.
This method is particularly appropriate for pharmaceutical industry for testing medicaments at different
temperatures where fast heating rates are necessary to avoid other unwanted reactions.
14. APPLICATIONS OF DSC
1. Enhanced analysis of thermal transitions.
2. Detection of low-level amorphous content.
3. Detection of “true” melting points.
4. Detection of low energy transitions.
5. Identification of substance.
15. REFERENCES
1. Willard; Merritt; “Instrumental Methods of Analysis”; 7th Edition; page no.- 761-762.
2. Skoog D.A; Crouch S.R; Holler F.J; “Principles of Instrumental Analysis”; page no.- 982-984.
3. Kamboj P.C.; Pharmaceutical Analysis; Volume-2; page no.- 984-989.
4. Kodre KV, Attarde SR; “Differential Scanning Calorimetry: A Review”; Research and Reviews: Journal
of Pharmaceutical Analysis; 2014.
5. Ayo Amarachi; Differential Scanning Calorimetry: A Review; International Journal of Applied Biology and
Pharmaceutical Technology; Volume-1; 2020.