Control of Analytical Quality Using
Stable Control Materials of Known
Concentration and Acceptable
Range
By
Dr. Medhat Eldeeb , MD
Lecturer of clinical pathology
What Do You Want to Learn?
Quality Control Basics
Calculations & Statistics
Westgard Rules
Levey-Jennings Charts
QC Product Awareness
Aim of Quality control
-To monitor the performance of
analytical methods.
-To detect analytical errors.
Internal and External Quality Control
• Internal Quality Control Internal quality control is set up within a laboratory to
monitor and ensure the reliability of test results from that laboratory.
• The primary tool for internal quality control is called a control. A control is a
specimen with a predetermined range of result values, called control
values, that is processed in the same manner as a patient sample.
• Control samples are processed with each series or run of patient samples.
If the result of a test on a control sample is different from its known value, this
indicates a problem in the equipment or the methods being used.
Types of analytical errors
Random error due to imprecision
Systematic error due to inaccuracy
Precision:” the repeatability, or reproducibility
of the measurement”
Accuracy: “Closeness of the agreement
between the result of a measurement and a
true value of the measurand”.
What is Random Error?
Random error (also called unsystematic error,
system noise or random variation) has no
pattern.
One minute your readings might be too small.
The next they might be too large.
You can’t predict random error and these errors
are usually unavoidable.
They are unpredictable and can’t be
replicated by repeating the experiment again.
CAUSES OF RANDOM ERRORS
• Bubbles in reagents;
• Instrument instability;
• Temperature variations; and
• Operator variability, such as variation in
pipetting.
• Uncalibrated pipets
• Mixing defect
• Timing defect
What is Systematic Error?
Systematic error (also called systematic bias)
is consistent, repeatable error.
Systematic Errors produce consistent errors,
either a fixed amount (like 1 mg/dL) or a
proportion (like 105% of the true value). If you
repeat the experiment, you’ll get the same
error.
Causes of systematic errors
• change in reagent lot;
• change in calibration;
• assigning the wrong calibrator values;
• reagents that were improperly prepared or are
deteriorating;
• pipettor maintenance error (not adjusted
correctly or misaligned); and/or
• a deteriorating photometric light source in the
instrument.
Systematic vs. Random Errors
Systematic Error
Avoidable error due
to controllable
variables in a
measurement.
Random Errors
Unavoidable errors
that are always
present in any
measurement.
Impossible to
eliminate
Control solutions, control materials
The International Federation of Clinical
Chemistry defines a control solution or control
material as :
a "specimen or solution which is analyzed solely
for quality control purposes, not for
calibration“
Calibration.“The process of testing and adjustment of an instrument,
kit, or test system, to provide a known relationship between the
measurement response and the value of the substance being
measured by the test procedure.” [CLSI]
Quantitative QC - Module 7 13
Calibrators
A substance with a specific
concentration.
Calibrators are used to set
(calibrate) the measuring
points on a scale.
Controls
A substance similar to
patients’ samples that
has an established
concentration.
Controls are used to ensure
the procedure is working
properly.
1 2 3 4 5
1 2 3 4 5
Criteria of Control material
1. Stable, Known Concentration and Acceptable
Range.
2. Box Pricing (The product cost per mL).
3. Available in aliquotes or vials of suitable size,
it is necessary to know the approximate
volume of control to be used each day.
4. Shelf life, . Your quality control shelf life
should match or exceed the laboratory’s
normal usage rate or money will be wasted.
5. Available at least for one year
6. Little vial to vial variability
7. Safe
8. Have a matrix similar to that of the test
specimens
9. Analyte conc. Should be in the normal and
abnormal ranges (similar to medical
situations)
Quantitative QC - Module 7 16
Choosing Control Materials
• values cover medical decision points
• similar to the test sample
• controls are usually available in high, normal, and low ranges
Types of Control Materials
Lyophilized or liquid
Human based or animal based
commercial products or Home-made
Assayed or unassayed
Assayed and Unassayed Controls
• Commercially prepared controls come in either assayed or
unassayed forms. Assayed controls are tested by multiple
methods before sale, and are sold with the results of the
tests.
Assayed controls:
• are more expensive than unassayed controls
• are used to evaluate accuracy and precision
• avoid laboratory errors in determining control values
• may only be suitable for specific methods or conditions
• While the manufacturer's control values can be used to
some extent to measure accuracy, the best measure of
accuracy is certified reference material.
Unassayed controls
• are not tested by the manufacturer before they are sold. The control
values for these materials must be determined by the individual
laboratory. Unassayed controls:
• are less expensive than assayed controls
• are used to evaluate precision only
• avoid manufacturer error in determining control values
• control values are customized to the laboratory's own methods and
conditions
• A final note: although commercially available control materials are
screened for hepatitis antigens and HIV antibodies, control materials
should still be handled with precautions, since they contain biological
materials and could contain infectious agents.
Quantitative QC - Module 7 20
Steps in Implementing Quantitative QC
• Obtain control material
• Run each control 20
times over 30 days
• Calculate mean and +/-1,2,3
Standard Deviations
• Draw L-J chat
• Apply WG rules Mean
1SD
1SD
2SD
3SD
2SD
3SD
The mean (or average) [x] :
Is the laboratory’s best estimate of
the analyte’s true value for a specific
level of control.
Formula 1: Calculating the Mean
X = Σ xn/n
Where:
Σ = sum
xn = each value in the data set
n = the number of values in the data set
Example
• Level I (Normal Control)
• Unassayed Chemistry Control, Lot No. 12345
• Test: K
• Instrument: ABC
• Units: mmol/L
• Control Values are:
• {4.0, 4.1, 4.0, 4.2,4.1, 4.1, 4.2}.
Answer
- The sum [Σ] is 28.7 mmol/L.
- The number of values is 7 (n = 7).
X = 28.7 / 7 = 4.1 mmol/L
X = 28.7 / 7 = 4.1
mmol/L
Standard Deviation [s]
Is a statistic that quantifies how close
numerical values (i.e., QC values) are
in relation to each other.
Formula 2: Calculating a
Standard Deviation [s]
• Level I (Normal Control)
• Unassayed Chemistry Control, Lot No. 12345
• Test: K
• Instrument: ABC
• Units: mmol/L
• Control Values are:
• {4.0, 4.1, 4.0, 4.2,4.1, 4.1, 4.2}.
- The sum [Σ] is 28.7 mmol/L.
- The number of values is 7 (n = 7).
X = 28.7 / 7 = 4.1 mmol/L
• Level I (Normal Control)
• assayed Chemistry Control, Lot No. 12345
• Test: K
• Instrument: ABC
• Units: mmol/L
• Range of control :
• {3.9 - 4.3}.
- The sum [Σ] is 8.2mmol/L.
- The number of values is 2(n = 2).
X = 8.2/ 2= 4.1 mmol/L
Calculating a Standard Deviation
Upper limit = X + 2 SD
4.3 = 4.1 + 2 SD
4.3 - 4.1 = 2 SD
0.2 = 2 SD
SD = 0.1
Calculating Quality Control Limits
• These ranges are used with the mean to construct the
Levey-Jennings chart
• ±1s range is 4.0 to 4.2 mmol/L
• 4.1 – (0.1)(1) = 4.0
• 4.1 + (0.1)(1) = 4.2
• ±2s range is 3.9 to 4.3 mmol/L
• 4.1 – (0.1)(2) = 3.9
• 4.1 + (0.1)(2) = 4.3
• ±3s range is 3.8 to 4.4 mmol/L
• 4.1 – (0.1)(3) = 3.8
• 4.1 + (0.1)(3) = 4.4
Creating a Levey-Jennings Chart
• Standard deviation is commonly used for preparing
Levey-Jennings (L-J or LJ) charts.
• The Levey-Jennings chart is used to graph successive
(run-to-run or day-to-day) quality control values.
• A chart is created for each test and level of control.
• The first step is to calculate decision limits.
• These limits are ±1s, ±2s and ±3s from the mean.
• The mean for the Level I potassium control is 4.1
mmol/L and the standard deviation is 0.1 mmol/L. Next
slide provides examples on how ±1s, ±2s and ±3s
quality control limits are calculated.
Levey-Jennings Chart
Quantitative QC - Module 7 36
Number of Controls
Interpretation depends on number of controls
run with patients’ samples.
• Good: If one control:
– accept results if control is within ± 2SD
unless shift or trend
• Better: If 2 levels of controls
– apply Westgard multirule system
Westgard Rules
12s
• This is a warning rule that is violated when a
single control observation is outside the ±2s
limits.
12s Rule
13s
• This rule identifies unacceptable random error
or possibly the beginning of a large systematic
error. Any QC result outside ±3s violates this
rule.
13s
22s
• This rule identifies systematic error only.
• The criteria for violation of this rule are:
•• Two consecutive QC results
•• Greater than 2s
•• On the same side of the mean
• There are two applications to this rule:
within-run and across runs.
22s within-run
22s across runs
R4s
• This rule identifies random error only, and is
applied only within the current run.
• If there is at least a 4s difference between
control values within a single run, the rule is
violated for random error.
R4s
41s
• The criteria which must be met to violate this
rule are:
• •• Four consecutive results
• •• Greater than 1s
• •• On the same side of the mean
41s
• There are two applications to the 41s rule:
These are within control material (e.g. all Level I
control results)
or
Across control materials (e.g., Level I, II, and III
control results in combination).
41s within control material
41s Across control materials
These rules are violated when there are:
• 7 or 8, or 9, or 10, or 12 control results
• On the same side of the mean
regardless of the specific standard deviation in
which they are located.
• Each of these rules also has two applications:
Within control material (e.g., all Level I control
• results)
or
Across control materials (e.g. Level I, II,and III
control results in combination).
10X Within control material
10X Across control materials
Systematic Error
• Systematic error is evidenced by a change in the mean
of the control values. The change in the mean may be :
• gradual and demonstrated as a trend in control values
or it may be
• abrupt and demonstrated as a shift in control values.
– shift–control on one side of the mean 6 consecutive days
– trend–control moving in one direction– heading toward an
“out of control” value
Quantitative QC - Module 7 56
Levey-Jennings Chart
Shift
MEAN
+1SD
+2SD
-1SD
-2SD
-3SD
+3SD
Days
190.5
192.5
194.5
196.5
188.5
186.5
184.6
Quantitative QC - Module 7 57
Levey-Jennings Chart
Trend
MEAN
+1SD
+2SD
-1SD
-2SD
-3SD
+3SD
Days
190.5
192.5
194.5
196.5
188.5
186.5
184.6
A trend indicates a gradual loss of reliability in the
test system.
Causes of trend:
• Deterioration of the instrument light source
• Gradual accumulation of debris in sample/reagent
tubing
• Gradual accumulation of debris on electrode surfaces
• Aging of reagents
• Gradual deterioration of control materials
• Gradual deterioration of incubation chamber
temperature (enzymes only)
• Gradual deterioration of light filter integrity
• Gradual deterioration of calibration
Shift: Abrupt changes in the control mean are
defined as shifts.
Shifts in QC data represent a sudden and
dramatic positive or negative change in test
system performance.
Shifts may be caused by:
• Sudden failure or change in the light source
• Change in reagent formulation
• Change of reagent lot
• Major instrument maintenance
• Sudden change in incubation temperature
(enzymes only)
• Change in room temperature or humidity
• Failure in the sampling system
• Failure in reagent dispense system
• Inaccurate calibration/recalibration
Quantitative QC - Module 7 61
If QC is out of control
• STOP testing
• identify and correct problem
• repeat testing on patient
samples and controls after correction
• Do not report patient results until
problem is solved and controls
indicate proper performance
Coefficient of Variation [CV]
Is the ratio of the standard deviation
to the mean and is expressed as a
percentage.
Formula 4: Calculating the
Coefficientof Variation [CV]
Applications of CV
The CV allows easier comparisons of the
overall precision.
The Coefficient of Variation can also be used
when comparing instrument performance.
The Coefficient of Variation can also be used
when comparing instrument performance.
Imprecision Differences Due to Instrument or Reagent
Level I (Normal Control)
Chemistry Control
Lot No. 12345
Level I (Normal Control)
Chemistry Control
Lot No. 12345
Instrument #1 / Reagent #1 Instrument #2 / Reagent #2
CV CV
Calcium 6.1% 5.9%
Phosphorus 5.2% 9.9%
Glucose 4.4% 4.2%

Mais conteúdo relacionado

Mais procurados

Quality control in the medical laboratory
Quality control in the medical laboratoryQuality control in the medical laboratory
Quality control in the medical laboratory
Adnan Jaran
 

Mais procurados (20)

Troubleshooting IQC / EQAS
Troubleshooting IQC / EQASTroubleshooting IQC / EQAS
Troubleshooting IQC / EQAS
 
Quality control
Quality controlQuality control
Quality control
 
Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .ppt
 
Troubleshooting iqc eqas 19.07.2018
Troubleshooting iqc eqas 19.07.2018Troubleshooting iqc eqas 19.07.2018
Troubleshooting iqc eqas 19.07.2018
 
Method validation
Method validationMethod validation
Method validation
 
Internal quality control
Internal quality controlInternal quality control
Internal quality control
 
Quality control
Quality controlQuality control
Quality control
 
Lecture , quality_control_in_clinical_biochemistry_laboratory
Lecture , quality_control_in_clinical_biochemistry_laboratoryLecture , quality_control_in_clinical_biochemistry_laboratory
Lecture , quality_control_in_clinical_biochemistry_laboratory
 
Quality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistryQuality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistry
 
Quality Control in Clinical Chemistry
Quality Control in Clinical ChemistryQuality Control in Clinical Chemistry
Quality Control in Clinical Chemistry
 
Quality control in the medical laboratory
Quality control in the medical laboratoryQuality control in the medical laboratory
Quality control in the medical laboratory
 
Method verification
Method verificationMethod verification
Method verification
 
Measurement Uncertainty
Measurement UncertaintyMeasurement Uncertainty
Measurement Uncertainty
 
How to Measure Uncertainty
How to Measure UncertaintyHow to Measure Uncertainty
How to Measure Uncertainty
 
Six Sigma in clinical laboratory
Six Sigma in clinical laboratorySix Sigma in clinical laboratory
Six Sigma in clinical laboratory
 
Quality Control
Quality ControlQuality Control
Quality Control
 
Total Quality Management (TQM) by Dr Anurag Yadav
Total Quality Management (TQM) by Dr Anurag YadavTotal Quality Management (TQM) by Dr Anurag Yadav
Total Quality Management (TQM) by Dr Anurag Yadav
 
Westgard rules
Westgard rulesWestgard rules
Westgard rules
 
Laboratory accreditation by iso 15189
Laboratory accreditation by iso 15189Laboratory accreditation by iso 15189
Laboratory accreditation by iso 15189
 
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
 

Semelhante a Control of analytical quality using stable control materials postgrad

Semelhante a Control of analytical quality using stable control materials postgrad (20)

(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptx(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptx
 
Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012
 
quality control in clinical laboratory
quality control in clinical laboratory quality control in clinical laboratory
quality control in clinical laboratory
 
Share_Total_Quality_Management.pptx
Share_Total_Quality_Management.pptxShare_Total_Quality_Management.pptx
Share_Total_Quality_Management.pptx
 
QUALITY control in hematology by Dr yogeeta.pptx
QUALITY  control in hematology by Dr yogeeta.pptxQUALITY  control in hematology by Dr yogeeta.pptx
QUALITY control in hematology by Dr yogeeta.pptx
 
Quality control
Quality controlQuality control
Quality control
 
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
 
Quality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing LaboratoryQuality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing Laboratory
 
Quality control.pptx
Quality control.pptxQuality control.pptx
Quality control.pptx
 
2. Quality Control Notes
2. Quality Control Notes2. Quality Control Notes
2. Quality Control Notes
 
Quality management system BY M .REJA
Quality management system BY M .REJAQuality management system BY M .REJA
Quality management system BY M .REJA
 
safety_in_laboratory.pdf
safety_in_laboratory.pdfsafety_in_laboratory.pdf
safety_in_laboratory.pdf
 
safety_in_laboratory.pdf
safety_in_laboratory.pdfsafety_in_laboratory.pdf
safety_in_laboratory.pdf
 
Clinical lab qc sethu
Clinical lab qc sethuClinical lab qc sethu
Clinical lab qc sethu
 
Quality Control in Pathological Laboratory
Quality Control in Pathological LaboratoryQuality Control in Pathological Laboratory
Quality Control in Pathological Laboratory
 
Method validation terms quality control and assurance
Method validation terms quality control and assurance Method validation terms quality control and assurance
Method validation terms quality control and assurance
 
qc final final.pptx
qc final final.pptxqc final final.pptx
qc final final.pptx
 
Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014
 
Quality control 12-12-2019
Quality control 12-12-2019Quality control 12-12-2019
Quality control 12-12-2019
 
Quality Control for Medical Laboratories
Quality Control for Medical LaboratoriesQuality Control for Medical Laboratories
Quality Control for Medical Laboratories
 

Último

💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
Sheetaleventcompany
 
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
adilkhan87451
 
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
adilkhan87451
 
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Sheetaleventcompany
 

Último (20)

Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service AvailableCall Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
Call Girls Jaipur Just Call 9521753030 Top Class Call Girl Service Available
 
Call Girls Coimbatore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 8250077686 Top Class Call Girl Service Available
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
Independent Call Girls In Jaipur { 8445551418 } ✔ ANIKA MEHTA ✔ Get High Prof...
 
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near MeTop Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
Top Rated Call Girls Kerala ☎ 8250092165👄 Delivery in 20 Mins Near Me
 
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Varanasi Just Call 8250077686 Top Class Call Girl Service Available
 
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
Jogeshwari ! Call Girls Service Mumbai - 450+ Call Girl Cash Payment 90042684...
 
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
💚Call Girls In Amritsar 💯Anvi 📲🔝8725944379🔝Amritsar Call Girl No💰Advance Cash...
 
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
Call Girls Service Jaipur {9521753030 } ❤️VVIP BHAWNA Call Girl in Jaipur Raj...
 
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
Russian Call Girls Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service...
 
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
Call Girls Service Jaipur {8445551418} ❤️VVIP BHAWNA Call Girl in Jaipur Raja...
 
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
Andheri East ) Call Girls in Mumbai Phone No 9004268417 Elite Escort Service ...
 
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
Call Girls in Lucknow Just Call 👉👉7877925207 Top Class Call Girl Service Avai...
 
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
8980367676 Call Girls In Ahmedabad Escort Service Available 24×7 In Ahmedabad
 
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
Models Call Girls In Hyderabad 9630942363 Hyderabad Call Girl & Hyderabad Esc...
 
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...Top Rated  Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
Top Rated Hyderabad Call Girls Chintal ⟟ 9332606886 ⟟ Call Me For Genuine Se...
 
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
Coimbatore Call Girls in Thudiyalur : 7427069034 High Profile Model Escorts |...
 
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
Call Girls Service Jaipur {9521753030} ❤️VVIP RIDDHI Call Girl in Jaipur Raja...
 
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
9630942363 Genuine Call Girls In Ahmedabad Gujarat Call Girls Service
 
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
Dehradun Call Girls Service {8854095900} ❤️VVIP ROCKY Call Girl in Dehradun U...
 

Control of analytical quality using stable control materials postgrad

  • 1. Control of Analytical Quality Using Stable Control Materials of Known Concentration and Acceptable Range By Dr. Medhat Eldeeb , MD Lecturer of clinical pathology
  • 2. What Do You Want to Learn? Quality Control Basics Calculations & Statistics Westgard Rules Levey-Jennings Charts QC Product Awareness
  • 3. Aim of Quality control -To monitor the performance of analytical methods. -To detect analytical errors.
  • 4. Internal and External Quality Control • Internal Quality Control Internal quality control is set up within a laboratory to monitor and ensure the reliability of test results from that laboratory. • The primary tool for internal quality control is called a control. A control is a specimen with a predetermined range of result values, called control values, that is processed in the same manner as a patient sample. • Control samples are processed with each series or run of patient samples. If the result of a test on a control sample is different from its known value, this indicates a problem in the equipment or the methods being used.
  • 5. Types of analytical errors Random error due to imprecision Systematic error due to inaccuracy Precision:” the repeatability, or reproducibility of the measurement” Accuracy: “Closeness of the agreement between the result of a measurement and a true value of the measurand”.
  • 6.
  • 7. What is Random Error? Random error (also called unsystematic error, system noise or random variation) has no pattern. One minute your readings might be too small. The next they might be too large. You can’t predict random error and these errors are usually unavoidable. They are unpredictable and can’t be replicated by repeating the experiment again.
  • 8. CAUSES OF RANDOM ERRORS • Bubbles in reagents; • Instrument instability; • Temperature variations; and • Operator variability, such as variation in pipetting. • Uncalibrated pipets • Mixing defect • Timing defect
  • 9. What is Systematic Error? Systematic error (also called systematic bias) is consistent, repeatable error. Systematic Errors produce consistent errors, either a fixed amount (like 1 mg/dL) or a proportion (like 105% of the true value). If you repeat the experiment, you’ll get the same error.
  • 10. Causes of systematic errors • change in reagent lot; • change in calibration; • assigning the wrong calibrator values; • reagents that were improperly prepared or are deteriorating; • pipettor maintenance error (not adjusted correctly or misaligned); and/or • a deteriorating photometric light source in the instrument.
  • 11. Systematic vs. Random Errors Systematic Error Avoidable error due to controllable variables in a measurement. Random Errors Unavoidable errors that are always present in any measurement. Impossible to eliminate
  • 12. Control solutions, control materials The International Federation of Clinical Chemistry defines a control solution or control material as : a "specimen or solution which is analyzed solely for quality control purposes, not for calibration“ Calibration.“The process of testing and adjustment of an instrument, kit, or test system, to provide a known relationship between the measurement response and the value of the substance being measured by the test procedure.” [CLSI]
  • 13. Quantitative QC - Module 7 13 Calibrators A substance with a specific concentration. Calibrators are used to set (calibrate) the measuring points on a scale. Controls A substance similar to patients’ samples that has an established concentration. Controls are used to ensure the procedure is working properly. 1 2 3 4 5 1 2 3 4 5
  • 14. Criteria of Control material 1. Stable, Known Concentration and Acceptable Range. 2. Box Pricing (The product cost per mL). 3. Available in aliquotes or vials of suitable size, it is necessary to know the approximate volume of control to be used each day. 4. Shelf life, . Your quality control shelf life should match or exceed the laboratory’s normal usage rate or money will be wasted.
  • 15. 5. Available at least for one year 6. Little vial to vial variability 7. Safe 8. Have a matrix similar to that of the test specimens 9. Analyte conc. Should be in the normal and abnormal ranges (similar to medical situations)
  • 16. Quantitative QC - Module 7 16 Choosing Control Materials • values cover medical decision points • similar to the test sample • controls are usually available in high, normal, and low ranges
  • 17. Types of Control Materials Lyophilized or liquid Human based or animal based commercial products or Home-made Assayed or unassayed
  • 18. Assayed and Unassayed Controls • Commercially prepared controls come in either assayed or unassayed forms. Assayed controls are tested by multiple methods before sale, and are sold with the results of the tests. Assayed controls: • are more expensive than unassayed controls • are used to evaluate accuracy and precision • avoid laboratory errors in determining control values • may only be suitable for specific methods or conditions • While the manufacturer's control values can be used to some extent to measure accuracy, the best measure of accuracy is certified reference material.
  • 19. Unassayed controls • are not tested by the manufacturer before they are sold. The control values for these materials must be determined by the individual laboratory. Unassayed controls: • are less expensive than assayed controls • are used to evaluate precision only • avoid manufacturer error in determining control values • control values are customized to the laboratory's own methods and conditions • A final note: although commercially available control materials are screened for hepatitis antigens and HIV antibodies, control materials should still be handled with precautions, since they contain biological materials and could contain infectious agents.
  • 20. Quantitative QC - Module 7 20 Steps in Implementing Quantitative QC • Obtain control material • Run each control 20 times over 30 days • Calculate mean and +/-1,2,3 Standard Deviations • Draw L-J chat • Apply WG rules Mean 1SD 1SD 2SD 3SD 2SD 3SD
  • 21. The mean (or average) [x] : Is the laboratory’s best estimate of the analyte’s true value for a specific level of control.
  • 22. Formula 1: Calculating the Mean X = Σ xn/n Where: Σ = sum xn = each value in the data set n = the number of values in the data set
  • 23. Example • Level I (Normal Control) • Unassayed Chemistry Control, Lot No. 12345 • Test: K • Instrument: ABC • Units: mmol/L • Control Values are: • {4.0, 4.1, 4.0, 4.2,4.1, 4.1, 4.2}.
  • 24. Answer - The sum [Σ] is 28.7 mmol/L. - The number of values is 7 (n = 7). X = 28.7 / 7 = 4.1 mmol/L X = 28.7 / 7 = 4.1 mmol/L
  • 25. Standard Deviation [s] Is a statistic that quantifies how close numerical values (i.e., QC values) are in relation to each other.
  • 26. Formula 2: Calculating a Standard Deviation [s]
  • 27. • Level I (Normal Control) • Unassayed Chemistry Control, Lot No. 12345 • Test: K • Instrument: ABC • Units: mmol/L • Control Values are: • {4.0, 4.1, 4.0, 4.2,4.1, 4.1, 4.2}.
  • 28. - The sum [Σ] is 28.7 mmol/L. - The number of values is 7 (n = 7). X = 28.7 / 7 = 4.1 mmol/L
  • 29.
  • 30. • Level I (Normal Control) • assayed Chemistry Control, Lot No. 12345 • Test: K • Instrument: ABC • Units: mmol/L • Range of control : • {3.9 - 4.3}.
  • 31. - The sum [Σ] is 8.2mmol/L. - The number of values is 2(n = 2). X = 8.2/ 2= 4.1 mmol/L
  • 32. Calculating a Standard Deviation Upper limit = X + 2 SD 4.3 = 4.1 + 2 SD 4.3 - 4.1 = 2 SD 0.2 = 2 SD SD = 0.1
  • 33. Calculating Quality Control Limits • These ranges are used with the mean to construct the Levey-Jennings chart • ±1s range is 4.0 to 4.2 mmol/L • 4.1 – (0.1)(1) = 4.0 • 4.1 + (0.1)(1) = 4.2 • ±2s range is 3.9 to 4.3 mmol/L • 4.1 – (0.1)(2) = 3.9 • 4.1 + (0.1)(2) = 4.3 • ±3s range is 3.8 to 4.4 mmol/L • 4.1 – (0.1)(3) = 3.8 • 4.1 + (0.1)(3) = 4.4
  • 34. Creating a Levey-Jennings Chart • Standard deviation is commonly used for preparing Levey-Jennings (L-J or LJ) charts. • The Levey-Jennings chart is used to graph successive (run-to-run or day-to-day) quality control values. • A chart is created for each test and level of control. • The first step is to calculate decision limits. • These limits are ±1s, ±2s and ±3s from the mean. • The mean for the Level I potassium control is 4.1 mmol/L and the standard deviation is 0.1 mmol/L. Next slide provides examples on how ±1s, ±2s and ±3s quality control limits are calculated.
  • 36. Quantitative QC - Module 7 36 Number of Controls Interpretation depends on number of controls run with patients’ samples. • Good: If one control: – accept results if control is within ± 2SD unless shift or trend • Better: If 2 levels of controls – apply Westgard multirule system
  • 37. Westgard Rules 12s • This is a warning rule that is violated when a single control observation is outside the ±2s limits.
  • 39. 13s • This rule identifies unacceptable random error or possibly the beginning of a large systematic error. Any QC result outside ±3s violates this rule.
  • 40. 13s
  • 41. 22s • This rule identifies systematic error only. • The criteria for violation of this rule are: •• Two consecutive QC results •• Greater than 2s •• On the same side of the mean • There are two applications to this rule: within-run and across runs.
  • 44. R4s • This rule identifies random error only, and is applied only within the current run. • If there is at least a 4s difference between control values within a single run, the rule is violated for random error.
  • 45. R4s
  • 46. 41s • The criteria which must be met to violate this rule are: • •• Four consecutive results • •• Greater than 1s • •• On the same side of the mean
  • 47. 41s • There are two applications to the 41s rule: These are within control material (e.g. all Level I control results) or Across control materials (e.g., Level I, II, and III control results in combination).
  • 48. 41s within control material
  • 49. 41s Across control materials
  • 50. These rules are violated when there are: • 7 or 8, or 9, or 10, or 12 control results • On the same side of the mean regardless of the specific standard deviation in which they are located.
  • 51. • Each of these rules also has two applications: Within control material (e.g., all Level I control • results) or Across control materials (e.g. Level I, II,and III control results in combination).
  • 52. 10X Within control material
  • 53. 10X Across control materials
  • 54. Systematic Error • Systematic error is evidenced by a change in the mean of the control values. The change in the mean may be : • gradual and demonstrated as a trend in control values or it may be • abrupt and demonstrated as a shift in control values. – shift–control on one side of the mean 6 consecutive days – trend–control moving in one direction– heading toward an “out of control” value
  • 55.
  • 56. Quantitative QC - Module 7 56 Levey-Jennings Chart Shift MEAN +1SD +2SD -1SD -2SD -3SD +3SD Days 190.5 192.5 194.5 196.5 188.5 186.5 184.6
  • 57. Quantitative QC - Module 7 57 Levey-Jennings Chart Trend MEAN +1SD +2SD -1SD -2SD -3SD +3SD Days 190.5 192.5 194.5 196.5 188.5 186.5 184.6
  • 58. A trend indicates a gradual loss of reliability in the test system. Causes of trend: • Deterioration of the instrument light source • Gradual accumulation of debris in sample/reagent tubing • Gradual accumulation of debris on electrode surfaces • Aging of reagents • Gradual deterioration of control materials • Gradual deterioration of incubation chamber temperature (enzymes only) • Gradual deterioration of light filter integrity • Gradual deterioration of calibration
  • 59. Shift: Abrupt changes in the control mean are defined as shifts. Shifts in QC data represent a sudden and dramatic positive or negative change in test system performance.
  • 60. Shifts may be caused by: • Sudden failure or change in the light source • Change in reagent formulation • Change of reagent lot • Major instrument maintenance • Sudden change in incubation temperature (enzymes only) • Change in room temperature or humidity • Failure in the sampling system • Failure in reagent dispense system • Inaccurate calibration/recalibration
  • 61. Quantitative QC - Module 7 61 If QC is out of control • STOP testing • identify and correct problem • repeat testing on patient samples and controls after correction • Do not report patient results until problem is solved and controls indicate proper performance
  • 62. Coefficient of Variation [CV] Is the ratio of the standard deviation to the mean and is expressed as a percentage.
  • 63. Formula 4: Calculating the Coefficientof Variation [CV]
  • 64. Applications of CV The CV allows easier comparisons of the overall precision. The Coefficient of Variation can also be used when comparing instrument performance.
  • 65. The Coefficient of Variation can also be used when comparing instrument performance. Imprecision Differences Due to Instrument or Reagent Level I (Normal Control) Chemistry Control Lot No. 12345 Level I (Normal Control) Chemistry Control Lot No. 12345 Instrument #1 / Reagent #1 Instrument #2 / Reagent #2 CV CV Calcium 6.1% 5.9% Phosphorus 5.2% 9.9% Glucose 4.4% 4.2%