RADIOACTIVE
COUNTERS
Presented By: Medha Thakur
MSc (Chemistry)
S.N.D.T. University
Contents  Introduction
 Classification Of Detectors
 Gas Ionisation System (GI)
 Types of (GI) counters
 (a) Gas Flow Proportional Counters
 b) Geiger – Muller Counters
 Types of Geiger Counters
 Solid Scintillators (SS)
 Liquid Scintillators (LS)
 Solid State (Semi-Conductor) Counters
Introduction
• various radiation have different
penetrating power & particles are
least penetrative.
• β particles are 10 times as penetrative
as α and γ rays are very penetrative.
• Beacuse of the wide variation in
penetrating power of the various
radiation, detectors employed for
counting the radiation will depend on
the type and energy of the radiation
emitted.
Radiation
Low
Pose a major
Problem for
detection
Medium
Detected by using
low density
materials (Gases)
Very High
Detected by using
denser material
(Solids)
With Penetrating
Power
Classification Of Detectors
Detectors
Gas ionisation
systems
Solid state
scintillation
counters
Liquid
Scintillators
Solid state (semi
– conductors)
counters
Gas Ionisation System (GI)
• Principle:
• When radiation passes through a gas, energy is
passed on to some of the gas molecules. These gas
molecules may ionise to a positive ion.
• e.g. If a suitable electric field is applied, the ions
can be collected and the magnitude of the resulting
currents can be realated to the amount of the
radiation initially responsible for the ionsation.
• The applied strength not only enables collection of
electrones or ions but also allows multiplication.
Types of (GI) counters
GI COUNTERS
(a) GAS FLOW
PROPORTIONAL COUNTERS
(b) GEIGER –
MULLER COUNTERS
(a) GAS FLOW PROPORTIONAL
COUNTERS
• In this type of detectors the applied field is such that the number of
ion pairs collected is proportional to the energy of the incident
radiation.
• This is achieved by placing the source, inside the counter and by
carefully controlling such experimental parameters as the geometry
of the counter, the flow rate of the gas, and the applied voltage.
• The output pulse is proportional to the energy of incident radiation.
Therefore by using a pulse- height analyser, it is possible to
differentiate between different isotopes.
GAS FLOW PROPORTIONAL COUNTER
GAS FLOW PROPORTIONAL COUNTERS
Disadvantages:-
• 1. A very stable high
voltage supply is
needed.
• 2. A very stable, high
gain amplifier is
needed.
• 3. The gas is used for
the counter is very
expensive.
Advantages:-
•1. gives a high and
reproducible detection
efficiency.
•2. used to investigate low
levels of radio-isotopes in
environmental sample.
•3. The sample to be counted
is normally placed inside
the counter, so that the
absorption problems of α
emitters are minimised.
(b) GEIGER – MULLER COUNTERS
Principle:
• In this type of detector, the applied field is increased to such
an extent that the ionisation induced by the acceleration of
initially produced ions and electrons results in virtually all of
the gas in the viscinity of the anode being ionised This is
known as ‘ avalanche effect’.
GEIGER – MULLER COUNTERS
Advantage:
They are simple and
effective.
Disadvantage:
To be compared with a standard.
They work only at high voltages (1000-
1500 V).
Types of Geiger
Counters
GEIGER
COUNTERS
1) END-WINDOW
used for counting solid samples.
It is cylindrical with a thin mica land
window. cylindrical case cathode and
central metal wire anode.
sample is placed directly under the End-
window. absorption of the radiation by the
window, and scattering of the radiation.
2) GEIGER COUNTER for LIQUID
SAMPLES
used for counting liquid samples
made of glass and consist of a central
gas filled tube containing the electrodes
surrounded by cylindrical ‘sheath’ into
which the liquid to be counted is placed.
1) END-WINDOW GEIGER COUNTER
2) GEIGER COUNTER for LIQUID
SAMPLES
Solid Scintillators (SS)
Principle:
• The most significant penetration into solid materials
is achieved by γ rays.
• They cause very little ionisation in gases and hence
are poorly teasured by gas ionsiation methods.
• rays are most efficiently counted using a solid
absorber. which acts as a scintillator.
• This is usually called a ‘Crystal or phosphor’. The
most popular crystal is N2- iodide containing traces of
thallium (t) iodide. The symbol used is Na2|Th|.
CONSTRUCTION
• crystals are cylindrical and a light-tight
aluminium can.
• One end is kept in contact with the PM
tube at which the photocathode is located.
• reflection losses minimised by a films of
transparent silicon oil.
• Counters Used:
• Plastic scintillators are used for β
counting.
• Some specialised counters use NaI(TL) for
Low energy γ and X-ray counting.
The main Criteria for a scintillators:
• 1. Good absorption of incident radiation (i.e.
high density).
• 2. High efficiency of photon production.
• 3. Little/ no reabsorption of the photons by
crystal.
• 4. Emitted photons just have a wavelength
compatible with the PM tube (320-720 nm for
glass window).
• anthracene and stilbene meet these criteria.
Solid Scintillator (SS)
Liquid Scintillators (LS)
Principle:
• GI counters and SS counters enable us to count isotopes which
emmit gamma rays and or medium to high energy β particles.
• But, there are several low energy β emitters for which these
technologies are inappropriate because of low penetration of the
β- particles,
• it is essential that the, isotope to be counted should be internally
mixed with the scintillator and not external to it as for solid
scintillators.
• This is done by dissolving sample wholly or partly in a solvent
containing the scintillator.
Construction and Working
1) Solvents :- Incident radiation should readily for electronically excited
solvent molecules. These readily transfer their energy to the scintillator.
Solvents are usually aromatic hydrocarbons such as xylene, toluene,
dioxane, etc.
2) Scintillator (primary soulte) :- accepts energy from solvent and is
raised to excited state and then de-excited by emitting light photon ( λ
range 300-400 nm) which is counted by PM tube. e.g. Large organic
molecules.
3) Secondary Solute :- Whenever maximum efficiency is required a
secondary solute is used.
They emit light photons of λ range 400-500 nm, where the PM tube is
more sensitive. The excitation energy of the 1° solute molecule is
transferred to the larger 2° solute molecules.
Scintillation
Counters
Advantages
simple to operate.
sensitivity to all types of
ionising radiations.
Distinguishes between photons of
different energies.
Short resolving time & more
efficient, fast and accurate
counting (especially γ rays).
Dis
advantages
A massive PM tube is necessary.
At low energies, the counter
response is not 100 % .
Solid State (Semi-Conductor) Counters
Principle:
• They are exemplified by the Ge (Li) system.
• Detectors of this type consist of an n-type
semiconductor and p-type semiconductor; separated by
an ultra-pure radiation sensitive region.
• The n-type semiconductor contains negative sites (e-s)
for conduction. The P-type S.C. has a deficiency of e-s or
‘Positive hole’.
• The SS ionisation detector requires an element whose
atoms are not only small enough to fit into the
interstitial sites of Ge; but also readily ionise e.g. Li.
Construction and Working
• In Ge (Li) detector, the Li atoms are allowed to migrate or ‘drift’ to
some of the p-sites in a cylinder of p-n Ge.
• Each p-site is deficient in 1e- and each Li is readily ionised and gives
up an e-.
• Hence a p-type S.C. and an n-type S.C. are separated by an ultra-pure
and electrically neutral region (intrinsic region).
• If radiation enters the neutral region (intrinsic), it will cause
ionisation.
• By applying a suitable potential, the e-s can be collected and amplified
to a measurable pulse.
Advantages 1. used for γ ray counting.
2. excellent resolution – 100 times better than NaI (TL)
complexes.
3. Pulse height is directly proportional to the energy of the
incident (gamma) radiation.
Dis-
advantages
1. Li atoms are very mobile therefore the Ge(Li) detector has to
be maintained at low temp
2. poorer detection efficiency because of smaller size [as
compared to NaI (TL)]
3. not as effective as NaI (TL) crystals for the samples with
high countrate.
4. must be kept in vacuum.
Radioactive counters or Detectors

Mais conteúdo relacionado

Mais procurados

Ionization chamber
Ionization chamberIonization chamber
Ionization chamber
Anas Yess
 

Mais procurados (20)

Radioisotopes
RadioisotopesRadioisotopes
Radioisotopes
 
Nuclear detectors
Nuclear detectorsNuclear detectors
Nuclear detectors
 
Fluorescence microscopy
Fluorescence microscopyFluorescence microscopy
Fluorescence microscopy
 
Gm counter
Gm counterGm counter
Gm counter
 
SOILD SCINTILLATION COUNTER PPT.pptx
SOILD SCINTILLATION COUNTER PPT.pptxSOILD SCINTILLATION COUNTER PPT.pptx
SOILD SCINTILLATION COUNTER PPT.pptx
 
Radio isotopes
Radio isotopesRadio isotopes
Radio isotopes
 
Spectrophotometry
SpectrophotometrySpectrophotometry
Spectrophotometry
 
Geiger muller counter
Geiger muller counterGeiger muller counter
Geiger muller counter
 
Ionization chamber
Ionization chamberIonization chamber
Ionization chamber
 
Scintillations principle, working, merits & demerits & applications
Scintillations  principle, working, merits & demerits & applicationsScintillations  principle, working, merits & demerits & applications
Scintillations principle, working, merits & demerits & applications
 
Spectrophotometer
SpectrophotometerSpectrophotometer
Spectrophotometer
 
Flow Cytometry (FCM)
Flow Cytometry (FCM) Flow Cytometry (FCM)
Flow Cytometry (FCM)
 
Gm counter
Gm counterGm counter
Gm counter
 
Chemiluminescence Immunoassay (CLIA) Technique
Chemiluminescence Immunoassay (CLIA) TechniqueChemiluminescence Immunoassay (CLIA) Technique
Chemiluminescence Immunoassay (CLIA) Technique
 
Gm counter& scintillation counter
Gm counter& scintillation counterGm counter& scintillation counter
Gm counter& scintillation counter
 
Flow Cytometry - basics, principles and applications
Flow Cytometry - basics, principles and applicationsFlow Cytometry - basics, principles and applications
Flow Cytometry - basics, principles and applications
 
Absorbance
AbsorbanceAbsorbance
Absorbance
 
Radioimmunoassay (RIA)
Radioimmunoassay (RIA)Radioimmunoassay (RIA)
Radioimmunoassay (RIA)
 
Colorimeter and Spectrophotometer.pptx
Colorimeter and Spectrophotometer.pptxColorimeter and Spectrophotometer.pptx
Colorimeter and Spectrophotometer.pptx
 
Flow Cytometry
Flow CytometryFlow Cytometry
Flow Cytometry
 

Semelhante a Radioactive counters or Detectors

Atomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptxAtomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptx
Ahnaf maznun
 

Semelhante a Radioactive counters or Detectors (20)

Atomic absorption spectrometer 13.4.17
Atomic absorption spectrometer  13.4.17Atomic absorption spectrometer  13.4.17
Atomic absorption spectrometer 13.4.17
 
IR Spectroscopy
IR SpectroscopyIR Spectroscopy
IR Spectroscopy
 
Uv vis-instrument-1
Uv vis-instrument-1Uv vis-instrument-1
Uv vis-instrument-1
 
Radiation detectors
Radiation detectors Radiation detectors
Radiation detectors
 
Ir spectroscopy
Ir spectroscopyIr spectroscopy
Ir spectroscopy
 
Radiopharmaceutical
Radiopharmaceutical Radiopharmaceutical
Radiopharmaceutical
 
Infrared spectroscopy
Infrared spectroscopyInfrared spectroscopy
Infrared spectroscopy
 
Spectral instruments
Spectral instrumentsSpectral instruments
Spectral instruments
 
Pharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptx
Pharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptxPharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptx
Pharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptx
 
Atomic absorption spectrophotometry
Atomic absorption spectrophotometryAtomic absorption spectrophotometry
Atomic absorption spectrophotometry
 
Aas
AasAas
Aas
 
AAS.pptx
AAS.pptxAAS.pptx
AAS.pptx
 
Infrared Spectroscopy
Infrared SpectroscopyInfrared Spectroscopy
Infrared Spectroscopy
 
Radio-active Counters
Radio-active CountersRadio-active Counters
Radio-active Counters
 
Atomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.pptAtomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.ppt
 
Presantation of aas
Presantation of aasPresantation of aas
Presantation of aas
 
Infra red spectroscopy
Infra red spectroscopyInfra red spectroscopy
Infra red spectroscopy
 
Atomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptxAtomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptx
 
Atomic absorption spectrometry (aas)
Atomic absorption spectrometry (aas)Atomic absorption spectrometry (aas)
Atomic absorption spectrometry (aas)
 
M.Sc. Part I 1.3 INFRARED SPECTROPHOTOMETRY PPT.ppsx
M.Sc. Part I  1.3 INFRARED SPECTROPHOTOMETRY PPT.ppsxM.Sc. Part I  1.3 INFRARED SPECTROPHOTOMETRY PPT.ppsx
M.Sc. Part I 1.3 INFRARED SPECTROPHOTOMETRY PPT.ppsx
 

Último

Porella : features, morphology, anatomy, reproduction etc.
Porella : features, morphology, anatomy, reproduction etc.Porella : features, morphology, anatomy, reproduction etc.
Porella : features, morphology, anatomy, reproduction etc.
Silpa
 
development of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virusdevelopment of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virus
NazaninKarimi6
 
Conjugation, transduction and transformation
Conjugation, transduction and transformationConjugation, transduction and transformation
Conjugation, transduction and transformation
Areesha Ahmad
 
Bacterial Identification and Classifications
Bacterial Identification and ClassificationsBacterial Identification and Classifications
Bacterial Identification and Classifications
Areesha Ahmad
 
Module for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learningModule for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learning
levieagacer
 

Último (20)

An introduction on sequence tagged site mapping
An introduction on sequence tagged site mappingAn introduction on sequence tagged site mapping
An introduction on sequence tagged site mapping
 
Thyroid Physiology_Dr.E. Muralinath_ Associate Professor
Thyroid Physiology_Dr.E. Muralinath_ Associate ProfessorThyroid Physiology_Dr.E. Muralinath_ Associate Professor
Thyroid Physiology_Dr.E. Muralinath_ Associate Professor
 
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICEPATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
PATNA CALL GIRLS 8617370543 LOW PRICE ESCORT SERVICE
 
GBSN - Microbiology (Unit 3)
GBSN - Microbiology (Unit 3)GBSN - Microbiology (Unit 3)
GBSN - Microbiology (Unit 3)
 
Velocity and Acceleration PowerPoint.ppt
Velocity and Acceleration PowerPoint.pptVelocity and Acceleration PowerPoint.ppt
Velocity and Acceleration PowerPoint.ppt
 
GBSN - Biochemistry (Unit 1)
GBSN - Biochemistry (Unit 1)GBSN - Biochemistry (Unit 1)
GBSN - Biochemistry (Unit 1)
 
Porella : features, morphology, anatomy, reproduction etc.
Porella : features, morphology, anatomy, reproduction etc.Porella : features, morphology, anatomy, reproduction etc.
Porella : features, morphology, anatomy, reproduction etc.
 
Grade 7 - Lesson 1 - Microscope and Its Functions
Grade 7 - Lesson 1 - Microscope and Its FunctionsGrade 7 - Lesson 1 - Microscope and Its Functions
Grade 7 - Lesson 1 - Microscope and Its Functions
 
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
Locating and isolating a gene, FISH, GISH, Chromosome walking and jumping, te...
 
Call Girls Ahmedabad +917728919243 call me Independent Escort Service
Call Girls Ahmedabad +917728919243 call me Independent Escort ServiceCall Girls Ahmedabad +917728919243 call me Independent Escort Service
Call Girls Ahmedabad +917728919243 call me Independent Escort Service
 
Zoology 5th semester notes( Sumit_yadav).pdf
Zoology 5th semester notes( Sumit_yadav).pdfZoology 5th semester notes( Sumit_yadav).pdf
Zoology 5th semester notes( Sumit_yadav).pdf
 
Dr. E. Muralinath_ Blood indices_clinical aspects
Dr. E. Muralinath_ Blood indices_clinical  aspectsDr. E. Muralinath_ Blood indices_clinical  aspects
Dr. E. Muralinath_ Blood indices_clinical aspects
 
development of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virusdevelopment of diagnostic enzyme assay to detect leuser virus
development of diagnostic enzyme assay to detect leuser virus
 
Stages in the normal growth curve
Stages in the normal growth curveStages in the normal growth curve
Stages in the normal growth curve
 
Factory Acceptance Test( FAT).pptx .
Factory Acceptance Test( FAT).pptx       .Factory Acceptance Test( FAT).pptx       .
Factory Acceptance Test( FAT).pptx .
 
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceuticsPulmonary drug delivery system M.pharm -2nd sem P'ceutics
Pulmonary drug delivery system M.pharm -2nd sem P'ceutics
 
Conjugation, transduction and transformation
Conjugation, transduction and transformationConjugation, transduction and transformation
Conjugation, transduction and transformation
 
Bacterial Identification and Classifications
Bacterial Identification and ClassificationsBacterial Identification and Classifications
Bacterial Identification and Classifications
 
Module for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learningModule for Grade 9 for Asynchronous/Distance learning
Module for Grade 9 for Asynchronous/Distance learning
 
module for grade 9 for distance learning
module for grade 9 for distance learningmodule for grade 9 for distance learning
module for grade 9 for distance learning
 

Radioactive counters or Detectors

  • 1. RADIOACTIVE COUNTERS Presented By: Medha Thakur MSc (Chemistry) S.N.D.T. University
  • 2. Contents  Introduction  Classification Of Detectors  Gas Ionisation System (GI)  Types of (GI) counters  (a) Gas Flow Proportional Counters  b) Geiger – Muller Counters  Types of Geiger Counters  Solid Scintillators (SS)  Liquid Scintillators (LS)  Solid State (Semi-Conductor) Counters
  • 3. Introduction • various radiation have different penetrating power & particles are least penetrative. • β particles are 10 times as penetrative as α and γ rays are very penetrative. • Beacuse of the wide variation in penetrating power of the various radiation, detectors employed for counting the radiation will depend on the type and energy of the radiation emitted. Radiation Low Pose a major Problem for detection Medium Detected by using low density materials (Gases) Very High Detected by using denser material (Solids) With Penetrating Power
  • 4. Classification Of Detectors Detectors Gas ionisation systems Solid state scintillation counters Liquid Scintillators Solid state (semi – conductors) counters
  • 5. Gas Ionisation System (GI) • Principle: • When radiation passes through a gas, energy is passed on to some of the gas molecules. These gas molecules may ionise to a positive ion. • e.g. If a suitable electric field is applied, the ions can be collected and the magnitude of the resulting currents can be realated to the amount of the radiation initially responsible for the ionsation. • The applied strength not only enables collection of electrones or ions but also allows multiplication.
  • 6. Types of (GI) counters GI COUNTERS (a) GAS FLOW PROPORTIONAL COUNTERS (b) GEIGER – MULLER COUNTERS
  • 7. (a) GAS FLOW PROPORTIONAL COUNTERS • In this type of detectors the applied field is such that the number of ion pairs collected is proportional to the energy of the incident radiation. • This is achieved by placing the source, inside the counter and by carefully controlling such experimental parameters as the geometry of the counter, the flow rate of the gas, and the applied voltage. • The output pulse is proportional to the energy of incident radiation. Therefore by using a pulse- height analyser, it is possible to differentiate between different isotopes.
  • 9. GAS FLOW PROPORTIONAL COUNTERS Disadvantages:- • 1. A very stable high voltage supply is needed. • 2. A very stable, high gain amplifier is needed. • 3. The gas is used for the counter is very expensive. Advantages:- •1. gives a high and reproducible detection efficiency. •2. used to investigate low levels of radio-isotopes in environmental sample. •3. The sample to be counted is normally placed inside the counter, so that the absorption problems of α emitters are minimised.
  • 10. (b) GEIGER – MULLER COUNTERS Principle: • In this type of detector, the applied field is increased to such an extent that the ionisation induced by the acceleration of initially produced ions and electrons results in virtually all of the gas in the viscinity of the anode being ionised This is known as ‘ avalanche effect’. GEIGER – MULLER COUNTERS Advantage: They are simple and effective. Disadvantage: To be compared with a standard. They work only at high voltages (1000- 1500 V).
  • 11. Types of Geiger Counters GEIGER COUNTERS 1) END-WINDOW used for counting solid samples. It is cylindrical with a thin mica land window. cylindrical case cathode and central metal wire anode. sample is placed directly under the End- window. absorption of the radiation by the window, and scattering of the radiation. 2) GEIGER COUNTER for LIQUID SAMPLES used for counting liquid samples made of glass and consist of a central gas filled tube containing the electrodes surrounded by cylindrical ‘sheath’ into which the liquid to be counted is placed.
  • 12. 1) END-WINDOW GEIGER COUNTER 2) GEIGER COUNTER for LIQUID SAMPLES
  • 13. Solid Scintillators (SS) Principle: • The most significant penetration into solid materials is achieved by γ rays. • They cause very little ionisation in gases and hence are poorly teasured by gas ionsiation methods. • rays are most efficiently counted using a solid absorber. which acts as a scintillator. • This is usually called a ‘Crystal or phosphor’. The most popular crystal is N2- iodide containing traces of thallium (t) iodide. The symbol used is Na2|Th|.
  • 14. CONSTRUCTION • crystals are cylindrical and a light-tight aluminium can. • One end is kept in contact with the PM tube at which the photocathode is located. • reflection losses minimised by a films of transparent silicon oil. • Counters Used: • Plastic scintillators are used for β counting. • Some specialised counters use NaI(TL) for Low energy γ and X-ray counting. The main Criteria for a scintillators: • 1. Good absorption of incident radiation (i.e. high density). • 2. High efficiency of photon production. • 3. Little/ no reabsorption of the photons by crystal. • 4. Emitted photons just have a wavelength compatible with the PM tube (320-720 nm for glass window). • anthracene and stilbene meet these criteria.
  • 16. Liquid Scintillators (LS) Principle: • GI counters and SS counters enable us to count isotopes which emmit gamma rays and or medium to high energy β particles. • But, there are several low energy β emitters for which these technologies are inappropriate because of low penetration of the β- particles, • it is essential that the, isotope to be counted should be internally mixed with the scintillator and not external to it as for solid scintillators. • This is done by dissolving sample wholly or partly in a solvent containing the scintillator.
  • 17. Construction and Working 1) Solvents :- Incident radiation should readily for electronically excited solvent molecules. These readily transfer their energy to the scintillator. Solvents are usually aromatic hydrocarbons such as xylene, toluene, dioxane, etc. 2) Scintillator (primary soulte) :- accepts energy from solvent and is raised to excited state and then de-excited by emitting light photon ( λ range 300-400 nm) which is counted by PM tube. e.g. Large organic molecules. 3) Secondary Solute :- Whenever maximum efficiency is required a secondary solute is used. They emit light photons of λ range 400-500 nm, where the PM tube is more sensitive. The excitation energy of the 1° solute molecule is transferred to the larger 2° solute molecules.
  • 18. Scintillation Counters Advantages simple to operate. sensitivity to all types of ionising radiations. Distinguishes between photons of different energies. Short resolving time & more efficient, fast and accurate counting (especially γ rays). Dis advantages A massive PM tube is necessary. At low energies, the counter response is not 100 % .
  • 19. Solid State (Semi-Conductor) Counters Principle: • They are exemplified by the Ge (Li) system. • Detectors of this type consist of an n-type semiconductor and p-type semiconductor; separated by an ultra-pure radiation sensitive region. • The n-type semiconductor contains negative sites (e-s) for conduction. The P-type S.C. has a deficiency of e-s or ‘Positive hole’. • The SS ionisation detector requires an element whose atoms are not only small enough to fit into the interstitial sites of Ge; but also readily ionise e.g. Li.
  • 20. Construction and Working • In Ge (Li) detector, the Li atoms are allowed to migrate or ‘drift’ to some of the p-sites in a cylinder of p-n Ge. • Each p-site is deficient in 1e- and each Li is readily ionised and gives up an e-. • Hence a p-type S.C. and an n-type S.C. are separated by an ultra-pure and electrically neutral region (intrinsic region). • If radiation enters the neutral region (intrinsic), it will cause ionisation. • By applying a suitable potential, the e-s can be collected and amplified to a measurable pulse.
  • 21. Advantages 1. used for γ ray counting. 2. excellent resolution – 100 times better than NaI (TL) complexes. 3. Pulse height is directly proportional to the energy of the incident (gamma) radiation. Dis- advantages 1. Li atoms are very mobile therefore the Ge(Li) detector has to be maintained at low temp 2. poorer detection efficiency because of smaller size [as compared to NaI (TL)] 3. not as effective as NaI (TL) crystals for the samples with high countrate. 4. must be kept in vacuum.