SlideShare uma empresa Scribd logo
1 de 31
HUMIDIFICATION
MANISH MASIH
Critical Care Therapist
Intensive Care Unit
Duncan Hopsital
Raxaul
OBJECTIVE
•Describe how airway heat and moisture exchange
normally occurs.
•State the effect dry gases have on the respiratory
tract.
•State when to humidify and warm inspired gas.
•Describe how various types of humidifiers work.
•Identify the indications, contraindications, and
hazards that pertain to humidification during
mechanical ventilation.
•Describe how to monitor patients receiving
humidity therapy.
•Describe how to identify and resolve common
problems with humidification systems.
HUMIDITY THERAPY
•Humidity is the quantity of moisture in
air or gas that is caused by the addition
of water in a gaseous state, or vapor.
Also called molecular water or invisible
moisture.
•Clinical Uses of Humidity
• To provide 100% body humidity of the
inspired gas for patients with ET tubes or
tracheostomy tubes
• To humidify dry therapeutic gases
NORMAL AIRWAY HUMIDIFICATION
• The nose warms, humidifies, and filters inspired
air.
• The pharynx, trachea, and bronchial tree also
warm, humidify, and filter inspired air.
• By the time the inspired air reaches the
oropharynx, it has been warmed to approximately
34°C and is 80% to 90% saturated with H2O.
• By the time the inspired air reaches the carina, it
has been warmed to body temperature (37°C) and
is 100% saturated.
• when the inspired air is fully saturated (100%) at
37°C, it holds 44 mg H2O per liter of gas and
exerts a water vapor pressure of 47 mmHg.
• As inspired gas moves into the lungs, it achieves
BTPS conditions (body temperature, 37°C;
barometric pressure; saturated with water
ABSOLUTE AND RELATIVE HUMIDITY
1. Absolute humidity is the amount of water in
a given volume of gas; its measurement is
expressed in milligrams per liter.
2. Relative humidity is a ratio between the
amount of water in a given volume of gas
and the maximum amount it is capable of
holding at that temperature (capacity). Its
measurement is expressed as a percentage
and is obtained with a hygrometer.
3. Relative humidity= absolute humidity
capacityx100
ISOTHERMIC SATURATION
BOUNDARY(ISB)
 As inspired gas moves into the lungs, it achieves
BTPS conditions (body temperature, 37°C;
barometric pressure; saturated with water
vapor[100% relative humidity at 37°C]). This point,
normally approximately 5cm below the carina, is
called the isothermic saturation boundary(ISB).
 Above the ISB, temperature and humidity decrease
during inspiration and increase during exhalation.
 Below the ISB, temperature and relative humidity
remain constant (BTPS).
 Numerous factors can shift the ISB deeper into the
lungs.
 The ISB shifts distally when a person breathes through the
mouth rather than the nose; when the person breaths cold,
dry air; when the upper airway is bypassed (breathing
through an artificial tracheal airway); or when the minute
ventilation is higher than normal.
 When this shift of ISB occurs, additional surfaces of the
airway are recruited to meet the heat and humidity
requirements of the lung.
 This recruitment of airways that do not typically provide
this level of heat and humidity can have a negative impact
on epithelial integrity.
 These shifts of the ISB can compromise the body’s normal
heat and moisture exchange mechanisms, and humidity
therapy is indicated.
TYPES OF HUMIDIFIERS
Humidifiers are either active (actively adding heat or
water or both to the device-patient interface) or passive
(recycling exhaled heat and humidity from the patient).
Active humidifiers typically include
1. Bubble humidifiers
2. Passover humidifers
3. Nebulizers of bland aerosols, and
4. Vapourizers
Passive humidifiers refer to typical heat and moisture
exchangers (HMEs).
INDICATION
Primary
Humidifying dry medical gases
Overcoming humidity deficit created when upper
airway is bypassed
Secondary
Managing hypothermia
Treating bronchospasm caused by cold air
CLINICAL SIGNS AND
SYMPTOMS OF INADEQUATE
AIRWAY HUMIDIFICATION
Atelectasis
Dry, nonproductive cough
Increased airway resistance
Increased incidence of infection
Increased work of breathing
Patient complaint of substernal pain and airway dryness
Thick, dehydrated secretions
HUMIDIFIERS
Humidifiers add molecular water to gas
Factors that affect performance:
Temperature
Surface area
Time of contact
Thermal mass
HUMIDIFIER PERFORMANCE
Temperature: as temperature increases,
capacity (potential humidity) will increase
Surface area: the greater the surface area, the
more potential for evaporation
Contact time: the longer the time of contact, the
more time for evaporation to occur
Thermal mass: the more water in the humidifier,
the more potential for transfer of heat
TYPES OF HUMIDIFIERS
BUBBLE HUMIDIFIER
Gas passes through tube to bottom of water
reservoir
Gas bubbles through reservoir
Unheated bubbles through humidifier
Provides humidity for oxygen therapy
BUBBLE HUMIDIFIER (CONT.)
PASSOVER HUMIDIFIERS
Direct gas over liquid or over surface saturated
by liquid
Types:
Simple reservoir model
Wick units
Membrane devices
PASSOVER HUMIDIFIERS
(CONT.)
Simple reservoir
Gas flows over surface
of volume of water
Usually used as
heated system to
provide humidity to
mechanically
ventilated patients
HEATED HUMIDIFIER
Adding heat increases absolute humidity of gas
Heating of gas increases risk of condensation in
tubing, causing obstruction
Risk is decreased by using large-bore tubing
Heated wire circuits are used to decrease
amount of condensation in circuit
HEAT MOISTURE EXCHANGER
Considered to be passive humidifier
Traps heat and humidity in expired gas
Has been used to provide humidity for
spontaneously and mechanically ventilated
patients
Types of heat moisture exchangers (HMEs):
Simple condenser
Hygroscopic condenser
Hydrophobic condenser
Simple condenser
humidifier
Contains condenser
element to trap heat and
humidity of expired gas
Retains about 50% of
expired heat and humidity
Maximum absolute
humidity is 18 to 28 mg/L
Hygroscopic heat
exchanger
Uses condenser element
made of paper, wool,
or foam
Material includes a salt
Maximum absolute
humidity is 22 to 34 mg/L
Active Heat Moisture
Exchangers
Add heat or humidity (or
both) to inspired gas
External heat and
moisture is introduced into
inspired gas
Capable of providing
100% relative humidity at
BTPS
CONTRAINDICATIONS
There are no contraindication to providing physiologic
conditioning of inspired gas during mechanical ventilation.
However, an HME is contraindicated in the following
circumstances
For patients with thick, copious, or bloody secretions
For patients with an expired tidal volume less than 70% of
the delivered tidal volume (eg., patients with large
bronchopleural fistulas or incompetent or absent
endotracheal tube cuffs)
For patients whose body temperature is less than 32°C
For patients with high spontaneous minute volumes (>10
L/min)
For patients receiving in-line aerosol drug treatments (an
HME must be removed from the patient circuit during
treatments)
HAZARDS AND
COMPLICATIONS
Hazards and complication associated with the use of
heated humidifier (HH) and HME devices during
mechanical ventilation include the following:
1. High flow rates during disconnect may aerosolize
contaminated condensate in heated humidifer circuit
2. Increased work of breathing (HME or HH)
3. Elevated airway pressures caused by condensation
(HH)
4. Inadvertent overfilling resulting in unintended tracheal
lavage (HH)
5. Inadvertent tracheal lavage from pooled condensate in
circuit (HH)
MONITORING
The humidifier should be inspected during the patient-ventilator
system check, and condensate should be inspected and replaced if
secretions have contaminated the insert or filter.
The following should be recorded during equipment inspection:
1. During routine use on an intubated patient, an HH should be set to
deliver inspired gas at 33°C ± 2°C and should provide a minimum of
30 mg/L of water vapor.
2. Inspired gas temperature should be monitored at or near the patient’s
airway opening (HH)
3. Specific temperature may vary with the patient’s condition; airway
temperature should never exceed 37°C.
4. For heated wire circuits used with infants, the probe must be placed
outside the incubator or away from the radiant warmer.
5. The high temperature alarm should be set no higher than 37°C, and
the low setting should not be less than 30°C.
6. Water level and function of automatic feed system (if applicable)
should be monitored.
PROBLEM SOLVING AND
TROUBLESHOOTING
Common problems with humidification systems
include dealing with condensation, avoiding cross
contamination, and ensuring proper conditioning of
the inspired gas.
Condensation
As the gas cools, its water vapor capacity
decreases, resulting in condensation. Factors
influencing the amount of condensation include
1. The temperature difference across the system
(humidifier to airway)
2. The ambient temperature
3. The gas flow
4. The set airway temperature; and
Condensation can disrupt or occlude gas flow through the circuit,
potentially altering fractional inspired oxygen or ventilator function or
both.
Condensate can work its way toward the patient and be aspirated. For
these reasons, circuits must be positioned to drain condensate away
from the patient and must be checked often, and excess condensate
must be drained from heated humidifier breathing circuits on a regular
basis.
Several techniques are used to minimize problems with breathing circuit
condensate. One common method is to place water traps at low points
in the circuit(both the inspiratory and the expiratory limbs of ventilator
circuits).
CROSS CONTAMINATION
Aerosols and condensate from ventilator circuits
are known sources of bacterial colonization.
However, advances in both circuit and humidifier
technology have reduced the risk of nosocomial
infection when these systems are used.
Heated wire circuits reduce production and
pooling of condensate within the circuit.
In addition, the high reservoir temperature in
humidifiers are bactericidal.
IMPORTANT POINTS CONCERNING
HUMIDIFIERS
1. Most nonheated humidifiers have a pressure pop-off valve set at 2
psi, after the device is set up, the tubing of the oxygen delivery
device (e.g., cannula, mask ) should be kinked to obstruct flow. If the
pop-off sounds, there are no leaks. If no sound is heard, all
connections, as well as the humidifier top, should be tightened.
2. Water levels of all humidifiers should be maintained at the levels
marked on the humidifier jar to ensure maximum humidity output.
3. Condensation occurs in the tubing of heated humidifiers. This water
should be discarded in a trash container or basin and should never
be put back into the humidifier.
4. Warm moist areas, such as heated humidifiers, are a breeding
ground for microorganisms (especially pseudomonas species). The
humidifier should be replaced every 24 hours(2).
5. Without a heated wire circuit, the humidifier may need to be heated
to as much as 50°C for the gas temperature to approximate body
temperature (37°C) by the time it reaches the patient’s upper airway.
As the highly saturated and warm gas passes through the ventilator
circuit, ambient air surrounding the circuit tubing cools(3).
REFERENCES
EGANS, Fundamentals of Respiratory Care.
Respiratory care exam review, Gary Persing
Clinical Application of Mechanical ventilation Susan
Pilbeam’s.

Mais conteúdo relacionado

Mais procurados

Autogenic Drainage
Autogenic DrainageAutogenic Drainage
Autogenic Drainage
vinuravaliya
 

Mais procurados (20)

Mechanical ventlation
Mechanical ventlationMechanical ventlation
Mechanical ventlation
 
AIRWAY CLEARANCE TECHNIQUES
AIRWAY CLEARANCE TECHNIQUESAIRWAY CLEARANCE TECHNIQUES
AIRWAY CLEARANCE TECHNIQUES
 
INCENTIVE SPIROMETER (1).pptx
INCENTIVE SPIROMETER (1).pptxINCENTIVE SPIROMETER (1).pptx
INCENTIVE SPIROMETER (1).pptx
 
Incentive spirometry ppt
Incentive spirometry pptIncentive spirometry ppt
Incentive spirometry ppt
 
Pep devices
Pep devicesPep devices
Pep devices
 
Pneumonectomy
PneumonectomyPneumonectomy
Pneumonectomy
 
Chest physiotherapy in ICU
Chest physiotherapy in ICUChest physiotherapy in ICU
Chest physiotherapy in ICU
 
Peep & cpap
Peep & cpapPeep & cpap
Peep & cpap
 
Broncho hygienic techniques.
Broncho   hygienic techniques. Broncho   hygienic techniques.
Broncho hygienic techniques.
 
Exercise tolerance testing
Exercise tolerance testingExercise tolerance testing
Exercise tolerance testing
 
IPPB
IPPBIPPB
IPPB
 
Airway clearance techniques
Airway clearance techniquesAirway clearance techniques
Airway clearance techniques
 
ICU management
ICU managementICU management
ICU management
 
Weaning from mechanical ventilation
Weaning from mechanical ventilationWeaning from mechanical ventilation
Weaning from mechanical ventilation
 
Aerosol therapy
Aerosol therapyAerosol therapy
Aerosol therapy
 
physiotherapy in icu patients
physiotherapy in icu patientsphysiotherapy in icu patients
physiotherapy in icu patients
 
Inspiratory muscle training
Inspiratory muscle trainingInspiratory muscle training
Inspiratory muscle training
 
Coughing and huffing
Coughing and huffingCoughing and huffing
Coughing and huffing
 
Pt management in icu
Pt management in icuPt management in icu
Pt management in icu
 
Autogenic Drainage
Autogenic DrainageAutogenic Drainage
Autogenic Drainage
 

Destaque

15 humidification fundamentals
15   humidification fundamentals15   humidification fundamentals
15 humidification fundamentals
nogat
 
Humidifiers, nebulizers (atomizers) and mucolytics
Humidifiers, nebulizers (atomizers)  and mucolyticsHumidifiers, nebulizers (atomizers)  and mucolytics
Humidifiers, nebulizers (atomizers) and mucolytics
Ritoban C
 
Refrigeration and Air conditioning
Refrigeration and Air conditioningRefrigeration and Air conditioning
Refrigeration and Air conditioning
SLA1987
 
Cu06997 lecture 5_reynolds_and_r
Cu06997 lecture 5_reynolds_and_rCu06997 lecture 5_reynolds_and_r
Cu06997 lecture 5_reynolds_and_r
Henk Massink
 
10.2 - First law of Thermodynamics and PV graphs
10.2 - First law of Thermodynamics and PV graphs10.2 - First law of Thermodynamics and PV graphs
10.2 - First law of Thermodynamics and PV graphs
simonandisa
 

Destaque (20)

15 humidification fundamentals
15   humidification fundamentals15   humidification fundamentals
15 humidification fundamentals
 
Humidifiers, nebulizers (atomizers) and mucolytics
Humidifiers, nebulizers (atomizers)  and mucolyticsHumidifiers, nebulizers (atomizers)  and mucolytics
Humidifiers, nebulizers (atomizers) and mucolytics
 
Humidifiers in anaesthesia and critical care
Humidifiers in anaesthesia and critical careHumidifiers in anaesthesia and critical care
Humidifiers in anaesthesia and critical care
 
Air Conditioning & Hvac Systems
Air Conditioning & Hvac SystemsAir Conditioning & Hvac Systems
Air Conditioning & Hvac Systems
 
HME
HME HME
HME
 
Humidity
HumidityHumidity
Humidity
 
Humidity
HumidityHumidity
Humidity
 
Refrigeration and Air conditioning
Refrigeration and Air conditioningRefrigeration and Air conditioning
Refrigeration and Air conditioning
 
PSYCHROMETRY
PSYCHROMETRYPSYCHROMETRY
PSYCHROMETRY
 
Air conditioning system
Air conditioning systemAir conditioning system
Air conditioning system
 
ECG Exercise Tolerance Test
ECG  Exercise Tolerance  TestECG  Exercise Tolerance  Test
ECG Exercise Tolerance Test
 
EET (Exercise Tolerance Test )
EET (Exercise Tolerance Test )EET (Exercise Tolerance Test )
EET (Exercise Tolerance Test )
 
The Science of HMEs
The Science of HMEsThe Science of HMEs
The Science of HMEs
 
Identification of textile fibers
Identification of textile fibersIdentification of textile fibers
Identification of textile fibers
 
Diploma Fire Safety Engineering
Diploma Fire Safety EngineeringDiploma Fire Safety Engineering
Diploma Fire Safety Engineering
 
14 chapter 5
14 chapter 514 chapter 5
14 chapter 5
 
Employees health safety institute hydrostatics
Employees health safety institute hydrostaticsEmployees health safety institute hydrostatics
Employees health safety institute hydrostatics
 
Cu06997 lecture 5_reynolds_and_r
Cu06997 lecture 5_reynolds_and_rCu06997 lecture 5_reynolds_and_r
Cu06997 lecture 5_reynolds_and_r
 
HUMIDIFICATION BY: MELODY H. MANALO BSRT :)
HUMIDIFICATION BY: MELODY H. MANALO BSRT :)HUMIDIFICATION BY: MELODY H. MANALO BSRT :)
HUMIDIFICATION BY: MELODY H. MANALO BSRT :)
 
10.2 - First law of Thermodynamics and PV graphs
10.2 - First law of Thermodynamics and PV graphs10.2 - First law of Thermodynamics and PV graphs
10.2 - First law of Thermodynamics and PV graphs
 

Semelhante a Humidification

Education - The Science of HMEs
Education - The Science of HMEsEducation - The Science of HMEs
Education - The Science of HMEs
Steve Koontz
 
Sem 2 bs1 ventilation
Sem 2 bs1 ventilationSem 2 bs1 ventilation
Sem 2 bs1 ventilation
Est
 

Semelhante a Humidification (20)

Humidifier & scavenging system
Humidifier & scavenging systemHumidifier & scavenging system
Humidifier & scavenging system
 
Humidification in Anaesthesia
Humidification in AnaesthesiaHumidification in Anaesthesia
Humidification in Anaesthesia
 
Humidification & inhalation therapy
Humidification & inhalation therapyHumidification & inhalation therapy
Humidification & inhalation therapy
 
HUMIDIFICATION AND NEBULIZATION.pptx
HUMIDIFICATION AND NEBULIZATION.pptxHUMIDIFICATION AND NEBULIZATION.pptx
HUMIDIFICATION AND NEBULIZATION.pptx
 
Humidifiers for Ventilators- Uses and Maintenance
Humidifiers for Ventilators- Uses and MaintenanceHumidifiers for Ventilators- Uses and Maintenance
Humidifiers for Ventilators- Uses and Maintenance
 
HUMIDIFICATION AND NEBULISATION PPT.pptx
HUMIDIFICATION AND NEBULISATION PPT.pptxHUMIDIFICATION AND NEBULISATION PPT.pptx
HUMIDIFICATION AND NEBULISATION PPT.pptx
 
Humidification & Nebulization.pptx
Humidification & Nebulization.pptxHumidification & Nebulization.pptx
Humidification & Nebulization.pptx
 
Humidifiers.pptx
Humidifiers.pptxHumidifiers.pptx
Humidifiers.pptx
 
Humidification
HumidificationHumidification
Humidification
 
Humidication in NICU - Dr Padmesh - Neonatology
Humidication in NICU - Dr Padmesh - NeonatologyHumidication in NICU - Dr Padmesh - Neonatology
Humidication in NICU - Dr Padmesh - Neonatology
 
oxygen_therapy_.ppt
oxygen_therapy_.pptoxygen_therapy_.ppt
oxygen_therapy_.ppt
 
humidifier .pptx
humidifier .pptxhumidifier .pptx
humidifier .pptx
 
Education - The Science of HMEs
Education - The Science of HMEsEducation - The Science of HMEs
Education - The Science of HMEs
 
Clean room
Clean roomClean room
Clean room
 
Physiology of nose
Physiology of nose Physiology of nose
Physiology of nose
 
PRELIMINARY DESIGN AND DEVELOPMENT OF A MEDICAL HUMIDIFIER
PRELIMINARY DESIGN AND DEVELOPMENT OF A MEDICAL HUMIDIFIERPRELIMINARY DESIGN AND DEVELOPMENT OF A MEDICAL HUMIDIFIER
PRELIMINARY DESIGN AND DEVELOPMENT OF A MEDICAL HUMIDIFIER
 
Sem 2 bs1 ventilation
Sem 2 bs1 ventilationSem 2 bs1 ventilation
Sem 2 bs1 ventilation
 
Hvac
HvacHvac
Hvac
 
Psychrometric processes
Psychrometric processesPsychrometric processes
Psychrometric processes
 
Humidification.pptx
Humidification.pptxHumidification.pptx
Humidification.pptx
 

Humidification

  • 1. HUMIDIFICATION MANISH MASIH Critical Care Therapist Intensive Care Unit Duncan Hopsital Raxaul
  • 2. OBJECTIVE •Describe how airway heat and moisture exchange normally occurs. •State the effect dry gases have on the respiratory tract. •State when to humidify and warm inspired gas. •Describe how various types of humidifiers work. •Identify the indications, contraindications, and hazards that pertain to humidification during mechanical ventilation. •Describe how to monitor patients receiving humidity therapy. •Describe how to identify and resolve common problems with humidification systems.
  • 3. HUMIDITY THERAPY •Humidity is the quantity of moisture in air or gas that is caused by the addition of water in a gaseous state, or vapor. Also called molecular water or invisible moisture. •Clinical Uses of Humidity • To provide 100% body humidity of the inspired gas for patients with ET tubes or tracheostomy tubes • To humidify dry therapeutic gases
  • 4. NORMAL AIRWAY HUMIDIFICATION • The nose warms, humidifies, and filters inspired air. • The pharynx, trachea, and bronchial tree also warm, humidify, and filter inspired air. • By the time the inspired air reaches the oropharynx, it has been warmed to approximately 34°C and is 80% to 90% saturated with H2O. • By the time the inspired air reaches the carina, it has been warmed to body temperature (37°C) and is 100% saturated. • when the inspired air is fully saturated (100%) at 37°C, it holds 44 mg H2O per liter of gas and exerts a water vapor pressure of 47 mmHg. • As inspired gas moves into the lungs, it achieves BTPS conditions (body temperature, 37°C; barometric pressure; saturated with water
  • 5.
  • 6. ABSOLUTE AND RELATIVE HUMIDITY 1. Absolute humidity is the amount of water in a given volume of gas; its measurement is expressed in milligrams per liter. 2. Relative humidity is a ratio between the amount of water in a given volume of gas and the maximum amount it is capable of holding at that temperature (capacity). Its measurement is expressed as a percentage and is obtained with a hygrometer. 3. Relative humidity= absolute humidity capacityx100
  • 7. ISOTHERMIC SATURATION BOUNDARY(ISB)  As inspired gas moves into the lungs, it achieves BTPS conditions (body temperature, 37°C; barometric pressure; saturated with water vapor[100% relative humidity at 37°C]). This point, normally approximately 5cm below the carina, is called the isothermic saturation boundary(ISB).  Above the ISB, temperature and humidity decrease during inspiration and increase during exhalation.  Below the ISB, temperature and relative humidity remain constant (BTPS).  Numerous factors can shift the ISB deeper into the lungs.
  • 8.  The ISB shifts distally when a person breathes through the mouth rather than the nose; when the person breaths cold, dry air; when the upper airway is bypassed (breathing through an artificial tracheal airway); or when the minute ventilation is higher than normal.  When this shift of ISB occurs, additional surfaces of the airway are recruited to meet the heat and humidity requirements of the lung.  This recruitment of airways that do not typically provide this level of heat and humidity can have a negative impact on epithelial integrity.  These shifts of the ISB can compromise the body’s normal heat and moisture exchange mechanisms, and humidity therapy is indicated.
  • 9. TYPES OF HUMIDIFIERS Humidifiers are either active (actively adding heat or water or both to the device-patient interface) or passive (recycling exhaled heat and humidity from the patient). Active humidifiers typically include 1. Bubble humidifiers 2. Passover humidifers 3. Nebulizers of bland aerosols, and 4. Vapourizers Passive humidifiers refer to typical heat and moisture exchangers (HMEs).
  • 10. INDICATION Primary Humidifying dry medical gases Overcoming humidity deficit created when upper airway is bypassed Secondary Managing hypothermia Treating bronchospasm caused by cold air
  • 11. CLINICAL SIGNS AND SYMPTOMS OF INADEQUATE AIRWAY HUMIDIFICATION Atelectasis Dry, nonproductive cough Increased airway resistance Increased incidence of infection Increased work of breathing Patient complaint of substernal pain and airway dryness Thick, dehydrated secretions
  • 12.
  • 13. HUMIDIFIERS Humidifiers add molecular water to gas Factors that affect performance: Temperature Surface area Time of contact Thermal mass
  • 14. HUMIDIFIER PERFORMANCE Temperature: as temperature increases, capacity (potential humidity) will increase Surface area: the greater the surface area, the more potential for evaporation Contact time: the longer the time of contact, the more time for evaporation to occur Thermal mass: the more water in the humidifier, the more potential for transfer of heat
  • 16. BUBBLE HUMIDIFIER Gas passes through tube to bottom of water reservoir Gas bubbles through reservoir Unheated bubbles through humidifier Provides humidity for oxygen therapy
  • 18. PASSOVER HUMIDIFIERS Direct gas over liquid or over surface saturated by liquid Types: Simple reservoir model Wick units Membrane devices
  • 19. PASSOVER HUMIDIFIERS (CONT.) Simple reservoir Gas flows over surface of volume of water Usually used as heated system to provide humidity to mechanically ventilated patients
  • 20. HEATED HUMIDIFIER Adding heat increases absolute humidity of gas Heating of gas increases risk of condensation in tubing, causing obstruction Risk is decreased by using large-bore tubing Heated wire circuits are used to decrease amount of condensation in circuit
  • 21. HEAT MOISTURE EXCHANGER Considered to be passive humidifier Traps heat and humidity in expired gas Has been used to provide humidity for spontaneously and mechanically ventilated patients Types of heat moisture exchangers (HMEs): Simple condenser Hygroscopic condenser Hydrophobic condenser
  • 22. Simple condenser humidifier Contains condenser element to trap heat and humidity of expired gas Retains about 50% of expired heat and humidity Maximum absolute humidity is 18 to 28 mg/L Hygroscopic heat exchanger Uses condenser element made of paper, wool, or foam Material includes a salt Maximum absolute humidity is 22 to 34 mg/L
  • 23. Active Heat Moisture Exchangers Add heat or humidity (or both) to inspired gas External heat and moisture is introduced into inspired gas Capable of providing 100% relative humidity at BTPS
  • 24. CONTRAINDICATIONS There are no contraindication to providing physiologic conditioning of inspired gas during mechanical ventilation. However, an HME is contraindicated in the following circumstances For patients with thick, copious, or bloody secretions For patients with an expired tidal volume less than 70% of the delivered tidal volume (eg., patients with large bronchopleural fistulas or incompetent or absent endotracheal tube cuffs) For patients whose body temperature is less than 32°C For patients with high spontaneous minute volumes (>10 L/min) For patients receiving in-line aerosol drug treatments (an HME must be removed from the patient circuit during treatments)
  • 25. HAZARDS AND COMPLICATIONS Hazards and complication associated with the use of heated humidifier (HH) and HME devices during mechanical ventilation include the following: 1. High flow rates during disconnect may aerosolize contaminated condensate in heated humidifer circuit 2. Increased work of breathing (HME or HH) 3. Elevated airway pressures caused by condensation (HH) 4. Inadvertent overfilling resulting in unintended tracheal lavage (HH) 5. Inadvertent tracheal lavage from pooled condensate in circuit (HH)
  • 26. MONITORING The humidifier should be inspected during the patient-ventilator system check, and condensate should be inspected and replaced if secretions have contaminated the insert or filter. The following should be recorded during equipment inspection: 1. During routine use on an intubated patient, an HH should be set to deliver inspired gas at 33°C ± 2°C and should provide a minimum of 30 mg/L of water vapor. 2. Inspired gas temperature should be monitored at or near the patient’s airway opening (HH) 3. Specific temperature may vary with the patient’s condition; airway temperature should never exceed 37°C. 4. For heated wire circuits used with infants, the probe must be placed outside the incubator or away from the radiant warmer. 5. The high temperature alarm should be set no higher than 37°C, and the low setting should not be less than 30°C. 6. Water level and function of automatic feed system (if applicable) should be monitored.
  • 27. PROBLEM SOLVING AND TROUBLESHOOTING Common problems with humidification systems include dealing with condensation, avoiding cross contamination, and ensuring proper conditioning of the inspired gas. Condensation As the gas cools, its water vapor capacity decreases, resulting in condensation. Factors influencing the amount of condensation include 1. The temperature difference across the system (humidifier to airway) 2. The ambient temperature 3. The gas flow 4. The set airway temperature; and
  • 28. Condensation can disrupt or occlude gas flow through the circuit, potentially altering fractional inspired oxygen or ventilator function or both. Condensate can work its way toward the patient and be aspirated. For these reasons, circuits must be positioned to drain condensate away from the patient and must be checked often, and excess condensate must be drained from heated humidifier breathing circuits on a regular basis. Several techniques are used to minimize problems with breathing circuit condensate. One common method is to place water traps at low points in the circuit(both the inspiratory and the expiratory limbs of ventilator circuits).
  • 29. CROSS CONTAMINATION Aerosols and condensate from ventilator circuits are known sources of bacterial colonization. However, advances in both circuit and humidifier technology have reduced the risk of nosocomial infection when these systems are used. Heated wire circuits reduce production and pooling of condensate within the circuit. In addition, the high reservoir temperature in humidifiers are bactericidal.
  • 30. IMPORTANT POINTS CONCERNING HUMIDIFIERS 1. Most nonheated humidifiers have a pressure pop-off valve set at 2 psi, after the device is set up, the tubing of the oxygen delivery device (e.g., cannula, mask ) should be kinked to obstruct flow. If the pop-off sounds, there are no leaks. If no sound is heard, all connections, as well as the humidifier top, should be tightened. 2. Water levels of all humidifiers should be maintained at the levels marked on the humidifier jar to ensure maximum humidity output. 3. Condensation occurs in the tubing of heated humidifiers. This water should be discarded in a trash container or basin and should never be put back into the humidifier. 4. Warm moist areas, such as heated humidifiers, are a breeding ground for microorganisms (especially pseudomonas species). The humidifier should be replaced every 24 hours(2). 5. Without a heated wire circuit, the humidifier may need to be heated to as much as 50°C for the gas temperature to approximate body temperature (37°C) by the time it reaches the patient’s upper airway. As the highly saturated and warm gas passes through the ventilator circuit, ambient air surrounding the circuit tubing cools(3).
  • 31. REFERENCES EGANS, Fundamentals of Respiratory Care. Respiratory care exam review, Gary Persing Clinical Application of Mechanical ventilation Susan Pilbeam’s.