    3M™ Littmann® Classic II S.E.
    Stethoscope
    Congratulations on the purchase of your new Littmann
    Classic II S.E. Stethoscope.
    This classic stethoscope with a two-sided chestpiece
    offers high acoustic sensitivity for superior
    performance. A traditional bell is combined with a
    patented tunable diaphragm that conveniently alternates
    between low and high frequency sounds without the
    need to turn over the chestpiece.
    Other features include a patient-friendly nonchill rim
    and diaphragm, excellent acoustic seal and comfortable
    fit with patented 3M™ Littmann® Snap Tight soft-
    sealing eartips, and a comfortably angled, anatomically
    correct headset. Additionally, the single-lumen tubing is
    available in a variety of popular colors to meet
    individual preferences.
    Most of all, know that your Littmann Classic II S.E.
    stethoscope carries the Littmann brand name, the name
    known worldwide for unsurpassed quality. As a trusted
    leader in auscultation technology, the Littmann brand of
    stethoscopes is your assurance of acoustical superiority,
    innovative design and exceptional performance.



1
Instructions for Use                                           Headset Adjustment
Chestpiece                                                                             Your Littmann stethoscope is
                                                                                       designed to provide a
On Littmann stethoscopes with traditional combination                                  comfortable, acoustically
chestpieces, it is necessary to open the side of the                                   sealed ear fit. Notice that the
chestpiece being used. This is accomplished by holding                                 eartubes are set to
the chestpiece stem in one hand and rotating the               Correct                 accommodate the typical
chestpiece with the other hand until a click is felt.                                  anatomy of the ear canal. The
                                                                                       eartips should point in a
Changing Frequencies Using the                                                         forward direction as you insert
Tunable Diaphragm                                                                      them into your ear canals.
Your Littmann Classic II S.E. Stethoscope is equipped
                                                               Incorrect
with a patented tunable diaphragm that enables you to
listen to both low and high frequency sounds without                                   To reduce spring tension in the
turning over the chestpiece.                                                           headset, hold each eartube at
                             Low frequencies: To listen                                the bend near the eartip and
                             to low frequency sounds,                                  gradually pull apart until fully
                             (traditional bell mode)                                   extended.
                             you can use either the                                    To increase spring tension,
                             open bell or tunable                                      grasp the headset with one
                             diaphragm using very                                      hand where the eartubes enter
                             light skin contact.                                       the tubing and squeeze until the
                             High frequencies: To                                      tubing on one eartube touches
                             listen to higher frequency                                the other. Repeat as necessary.
                             sounds, using the tunable                                   For maximum acoustic
                             diaphragm, press firmly                                     performance, comfortable
                             on the chestpiece.                                          patented 3M™ Littmann®
To listen to low and high frequency sounds without                                       Soft-sealing Eartips are
removing and turning over the chestpiece, simply                                         provided with your
alternate between light and firm pressure on the tunable                                 stethoscope. This stethoscope
diaphragm.                                                                               utilizes a unique design for
                                                               attaching the eartip to the eartube. The eartips are


                                                           2
pushed on to the end of the eartube and snapped firmly                                                                                 Visually examine the ring to
into place. To remove, pull firmly on the eartip.                                                                                      ensure that the flexible edge is
                                                                                                                                       smoothly engaged inside the
                                                                                                                                        A
                                                                                                                                            N N




                                                                                                                                    M
                                                                                                                                T




                                                                                                                                                                    Y
Removing the Diaphragm and




                                                                                                                            T




                                                                                                                                                                     T
                                                                                                                                       rim. If necessary, flex the




                                                                                                                                I




                                                                                                                                                                    I
                                                                                                                                L




                                                                                                                                                                L
                                                                                                                                                          A
                                                                                                                                                Q       U




Cleaning the Chestpiece                                                                                                                assembly by pinching the outer
                                                                                                                                       rim edge between the thumb
                         With the diaphragm side up,                                                          and fingers. Repeat this flexing procedure after rotating
                         grasp the rim with the thumbs                                                        the assembly one quarter.
                         and index fingers of both
                                       N
                                           N
                                                           Y
                              A




                                                           L I T
                           T T M




                                                                   TM




                                                                                                                                                                                            To attach the rim/diaphragm
                                                           A




                         hands and roll the rim off the
                              I




                                   L
                                                       U




                                                   Q




                         edge of the chestpiece.                                                                                                                                            assembly to the chestpiece,
                         Remove the diaphragm from                                                                                                    M
                                                                                                                                                            AN
                                                                                                                                                               N
                                                                                                                                                                                            engage the groove of the rim




                                                                                                                                                  T
                                                                                                                                                T




                                                                                                                                                                                   Y
                                                                                                                                                                                            (with the diaphragm attached)




                                                                                                                                                I




                                                                                                                                                                               T
                         the rim and clean the parts in




                                                                                                                                                L




                                                                                                                                                                               I
                                                                                                                                                                           L
                                                                                                                                                                          A
                                                                                                                                                            Q U




soapy water or wipe with alcohol. Chestpiece surfaces
                                                                                                                                                                    TM


                                                                                                                                                                                            around the chestpiece at one
can be wiped with alcohol or soapy water. Dry all parts                                                                                                                                     point and hold it in place with
and surfaces thoroughly before reassembly.                                                                                                                                                  your thumbs.

Assembly and Replacement of the                                                                                                                                                             Slowly roll the rim around and
Tunable Diaphragm                                                                                                                                                                           over the chestpiece edge using
                                                                                                                                A
                                                                                                                                        N
                                                                                                                                            N

                                                                                                                                                                                            both thumbs, moving in
                                                                                                                                                                     T Y
                                                                                                                      I T T M




                                                                                                                                                                          I
                      Using your fingertip, apply a                                                                             L
                                                                                                                                                      Q
                                                                                                                                                                A
                                                                                                                                                                U
                                                                                                                                                                     L
                                                                                                                                                                                            opposite directions around the
                      small amount of talcum powder
                                       MT
                                                                                                                                                                                            chestpiece.
                           U Q

                      to the inside surface of the
                                               A
                                                   L
                                                     Y T I
      T I L




                      flexible edge of the diaphragm.
         T




              N N A M


                      This will facilitate assembly                                                                                    Visually inspect the edge where
                      and maintain smooth and quiet                                                                                    the diaphragm engages the rim
                      low/high frequency alternation.                                                                                  for uniform containment.
                                                                                                                                                    Y
                                                                                                                                                            T
                                                                                                                                                                I
                                                                                                                         M A N N




                                                                                                                                                                         Q U A L




Snap the diaphragm with your fingers to remove excess                                                                                  Minor adjustments may be                        TM
                                                                                                                                T




                                                                                                                                        T
                                                                                                                                            L I

talc.                                                                                                                                  made by slightly pulling and
                                                                                                                                       rolling the rim away from the
                                                                        Insert the flexible edge of the                                diaphragm, allowing the
                       I
                               T T M A
                                                                        diaphragm into the groove of          diaphragm to slip into position.
                                                       N




                                                                        the rim. This is best
              L




                                                               N
              Q




                   U

                                                                        accomplished by starting with
                                                       Y
                             A         L       I T
              TM




                                                                        the rim positioned above the
                                                                        legible side of the diaphragm.

                                                                                                          3
General Use and Maintenance                                      3M Health Care Service Center
                                                                 Bldg. 502, Suite 200
• Avoid extreme heat, cold, solvents, and oils.                  3350 Granada Ave. North
                                                                 Oakdale, MN 55128
• The entire stethoscope can be wiped clean with
  alcohol or soapy water.                                        IN CANADA:
• Eartips can be removed from the eartubes for                   3M Health Care Service Centre
  thorough cleaning.                                             3M Canada Inc.
                                                                 80 Enterprise Drive South
NOTE: Do not immerse your stethoscope in any liquid
                                                                 London, Ontario
or subject it to steam sterilization. If disinfection is
                                                                 Canada N6N1C2
required, the stethoscope may be wiped with a 70%
                                                                 1-800-563-2921
isopropyl alcohol solution.
                                                                 Outside of the U.S. and Canada please contact your local
                                                                 3M subsidiary for maintenance and repair information.
Littmann Stethoscope Service and                                 Explanation of Symbols:
Warranty Program
Your Littmann stethoscope comes with the finest service             ! • Attention, see instructions for use.
and warranty policy in the industry. The Littmann Classic
II S.E. Stethoscope is warranted against any defects in
material and workmanship for a period of three (3) years.                • This product and package do not contain
                                                                         natural rubber latex.
If a material or manufacturing defect is discovered during
the warranty period, repairs will be made without charge
upon the return of the instrument to 3M, except in cases
of obvious abuse or accidental damage.

For maintenance or repair services, write a short note
explaining the repair or service you require, include
your name, address and phone number and send your
stethoscope directly to:




                                                             4

Mais conteúdo relacionado

Semelhante a Cuidados del estetoscopio

Ytesonhuong huongdansudung-littmann-master-clasic
Ytesonhuong huongdansudung-littmann-master-clasicYtesonhuong huongdansudung-littmann-master-clasic
Ytesonhuong huongdansudung-littmann-master-clasicSon Huong Medical Equipment
 
Littmann master cardiology - Ống nghe dành cho chuyên khoa tim mạch.
Littmann master cardiology - Ống nghe dành cho chuyên khoa tim mạch.Littmann master cardiology - Ống nghe dành cho chuyên khoa tim mạch.
Littmann master cardiology - Ống nghe dành cho chuyên khoa tim mạch.Ống Nghe Littmann 3M
 
Ytesonhuong huongdansudung-littmann-classic ii-se
Ytesonhuong huongdansudung-littmann-classic ii-seYtesonhuong huongdansudung-littmann-classic ii-se
Ytesonhuong huongdansudung-littmann-classic ii-seSon Huong Medical Equipment
 
Ytesonhuong huongdansudung-littmann-lightweight ii
Ytesonhuong huongdansudung-littmann-lightweight iiYtesonhuong huongdansudung-littmann-lightweight ii
Ytesonhuong huongdansudung-littmann-lightweight iiSon Huong Medical Equipment
 
Littmann pediatric/ infant - Ống nghe littmann nhi/ sơ sinh
Littmann pediatric/ infant - Ống nghe littmann nhi/ sơ sinhLittmann pediatric/ infant - Ống nghe littmann nhi/ sơ sinh
Littmann pediatric/ infant - Ống nghe littmann nhi/ sơ sinhỐng Nghe Littmann 3M
 

Semelhante a Cuidados del estetoscopio (6)

Ytesonhuong huongdansudung-littmann-master-clasic
Ytesonhuong huongdansudung-littmann-master-clasicYtesonhuong huongdansudung-littmann-master-clasic
Ytesonhuong huongdansudung-littmann-master-clasic
 
Ống nghe littmann master classic II
Ống nghe littmann master classic IIỐng nghe littmann master classic II
Ống nghe littmann master classic II
 
Littmann master cardiology - Ống nghe dành cho chuyên khoa tim mạch.
Littmann master cardiology - Ống nghe dành cho chuyên khoa tim mạch.Littmann master cardiology - Ống nghe dành cho chuyên khoa tim mạch.
Littmann master cardiology - Ống nghe dành cho chuyên khoa tim mạch.
 
Ytesonhuong huongdansudung-littmann-classic ii-se
Ytesonhuong huongdansudung-littmann-classic ii-seYtesonhuong huongdansudung-littmann-classic ii-se
Ytesonhuong huongdansudung-littmann-classic ii-se
 
Ytesonhuong huongdansudung-littmann-lightweight ii
Ytesonhuong huongdansudung-littmann-lightweight iiYtesonhuong huongdansudung-littmann-lightweight ii
Ytesonhuong huongdansudung-littmann-lightweight ii
 
Littmann pediatric/ infant - Ống nghe littmann nhi/ sơ sinh
Littmann pediatric/ infant - Ống nghe littmann nhi/ sơ sinhLittmann pediatric/ infant - Ống nghe littmann nhi/ sơ sinh
Littmann pediatric/ infant - Ống nghe littmann nhi/ sơ sinh
 

Último

Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Celine George
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxheathfieldcps1
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxiammrhaywood
 
Class 11th Physics NEET formula sheet pdf
Class 11th Physics NEET formula sheet pdfClass 11th Physics NEET formula sheet pdf
Class 11th Physics NEET formula sheet pdfAyushMahapatra5
 
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...KokoStevan
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfAdmir Softic
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxDenish Jangid
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdfQucHHunhnh
 
PROCESS RECORDING FORMAT.docx
PROCESS      RECORDING        FORMAT.docxPROCESS      RECORDING        FORMAT.docx
PROCESS RECORDING FORMAT.docxPoojaSen20
 
ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxAreebaZafar22
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAssociation for Project Management
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhikauryashika82
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphThiyagu K
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfJayanti Pande
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Disha Kariya
 
Unit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxUnit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxVishalSingh1417
 
Measures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDMeasures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDThiyagu K
 

Último (20)

Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
 
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptxSOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
SOCIAL AND HISTORICAL CONTEXT - LFTVD.pptx
 
Class 11th Physics NEET formula sheet pdf
Class 11th Physics NEET formula sheet pdfClass 11th Physics NEET formula sheet pdf
Class 11th Physics NEET formula sheet pdf
 
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...
SECOND SEMESTER TOPIC COVERAGE SY 2023-2024 Trends, Networks, and Critical Th...
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 
PROCESS RECORDING FORMAT.docx
PROCESS      RECORDING        FORMAT.docxPROCESS      RECORDING        FORMAT.docx
PROCESS RECORDING FORMAT.docx
 
Advance Mobile Application Development class 07
Advance Mobile Application Development class 07Advance Mobile Application Development class 07
Advance Mobile Application Development class 07
 
ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptx
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across Sectors
 
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in DelhiRussian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
Russian Escort Service in Delhi 11k Hotel Foreigner Russian Call Girls in Delhi
 
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot Graph
 
Web & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdfWeb & Social Media Analytics Previous Year Question Paper.pdf
Web & Social Media Analytics Previous Year Question Paper.pdf
 
Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..Sports & Fitness Value Added Course FY..
Sports & Fitness Value Added Course FY..
 
Unit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxUnit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptx
 
Measures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SDMeasures of Dispersion and Variability: Range, QD, AD and SD
Measures of Dispersion and Variability: Range, QD, AD and SD
 

Cuidados del estetoscopio

  • 1.
  • 2. English 3M™ Littmann® Classic II S.E. Stethoscope Congratulations on the purchase of your new Littmann Classic II S.E. Stethoscope. This classic stethoscope with a two-sided chestpiece offers high acoustic sensitivity for superior performance. A traditional bell is combined with a patented tunable diaphragm that conveniently alternates between low and high frequency sounds without the need to turn over the chestpiece. Other features include a patient-friendly nonchill rim and diaphragm, excellent acoustic seal and comfortable fit with patented 3M™ Littmann® Snap Tight soft- sealing eartips, and a comfortably angled, anatomically correct headset. Additionally, the single-lumen tubing is available in a variety of popular colors to meet individual preferences. Most of all, know that your Littmann Classic II S.E. stethoscope carries the Littmann brand name, the name known worldwide for unsurpassed quality. As a trusted leader in auscultation technology, the Littmann brand of stethoscopes is your assurance of acoustical superiority, innovative design and exceptional performance. 1
  • 3. Instructions for Use Headset Adjustment Chestpiece Your Littmann stethoscope is designed to provide a On Littmann stethoscopes with traditional combination comfortable, acoustically chestpieces, it is necessary to open the side of the sealed ear fit. Notice that the chestpiece being used. This is accomplished by holding eartubes are set to the chestpiece stem in one hand and rotating the Correct accommodate the typical chestpiece with the other hand until a click is felt. anatomy of the ear canal. The eartips should point in a Changing Frequencies Using the forward direction as you insert Tunable Diaphragm them into your ear canals. Your Littmann Classic II S.E. Stethoscope is equipped Incorrect with a patented tunable diaphragm that enables you to listen to both low and high frequency sounds without To reduce spring tension in the turning over the chestpiece. headset, hold each eartube at Low frequencies: To listen the bend near the eartip and to low frequency sounds, gradually pull apart until fully (traditional bell mode) extended. you can use either the To increase spring tension, open bell or tunable grasp the headset with one diaphragm using very hand where the eartubes enter light skin contact. the tubing and squeeze until the High frequencies: To tubing on one eartube touches listen to higher frequency the other. Repeat as necessary. sounds, using the tunable For maximum acoustic diaphragm, press firmly performance, comfortable on the chestpiece. patented 3M™ Littmann® To listen to low and high frequency sounds without Soft-sealing Eartips are removing and turning over the chestpiece, simply provided with your alternate between light and firm pressure on the tunable stethoscope. This stethoscope diaphragm. utilizes a unique design for attaching the eartip to the eartube. The eartips are 2
  • 4. pushed on to the end of the eartube and snapped firmly Visually examine the ring to into place. To remove, pull firmly on the eartip. ensure that the flexible edge is smoothly engaged inside the A N N M T Y Removing the Diaphragm and T T rim. If necessary, flex the I I L L A Q U Cleaning the Chestpiece assembly by pinching the outer rim edge between the thumb With the diaphragm side up, and fingers. Repeat this flexing procedure after rotating grasp the rim with the thumbs the assembly one quarter. and index fingers of both N N Y A L I T T T M TM To attach the rim/diaphragm A hands and roll the rim off the I L U Q edge of the chestpiece. assembly to the chestpiece, Remove the diaphragm from M AN N engage the groove of the rim T T Y (with the diaphragm attached) I T the rim and clean the parts in L I L A Q U soapy water or wipe with alcohol. Chestpiece surfaces TM around the chestpiece at one can be wiped with alcohol or soapy water. Dry all parts point and hold it in place with and surfaces thoroughly before reassembly. your thumbs. Assembly and Replacement of the Slowly roll the rim around and Tunable Diaphragm over the chestpiece edge using A N N both thumbs, moving in T Y I T T M I Using your fingertip, apply a L Q A U L opposite directions around the small amount of talcum powder MT chestpiece. U Q to the inside surface of the A L Y T I T I L flexible edge of the diaphragm. T N N A M This will facilitate assembly Visually inspect the edge where and maintain smooth and quiet the diaphragm engages the rim low/high frequency alternation. for uniform containment. Y T I M A N N Q U A L Snap the diaphragm with your fingers to remove excess Minor adjustments may be TM T T L I talc. made by slightly pulling and rolling the rim away from the Insert the flexible edge of the diaphragm, allowing the I T T M A diaphragm into the groove of diaphragm to slip into position. N the rim. This is best L N Q U accomplished by starting with Y A L I T TM the rim positioned above the legible side of the diaphragm. 3
  • 5. General Use and Maintenance 3M Health Care Service Center Bldg. 502, Suite 200 • Avoid extreme heat, cold, solvents, and oils. 3350 Granada Ave. North Oakdale, MN 55128 • The entire stethoscope can be wiped clean with alcohol or soapy water. IN CANADA: • Eartips can be removed from the eartubes for 3M Health Care Service Centre thorough cleaning. 3M Canada Inc. 80 Enterprise Drive South NOTE: Do not immerse your stethoscope in any liquid London, Ontario or subject it to steam sterilization. If disinfection is Canada N6N1C2 required, the stethoscope may be wiped with a 70% 1-800-563-2921 isopropyl alcohol solution. Outside of the U.S. and Canada please contact your local 3M subsidiary for maintenance and repair information. Littmann Stethoscope Service and Explanation of Symbols: Warranty Program Your Littmann stethoscope comes with the finest service ! • Attention, see instructions for use. and warranty policy in the industry. The Littmann Classic II S.E. Stethoscope is warranted against any defects in material and workmanship for a period of three (3) years. • This product and package do not contain natural rubber latex. If a material or manufacturing defect is discovered during the warranty period, repairs will be made without charge upon the return of the instrument to 3M, except in cases of obvious abuse or accidental damage. For maintenance or repair services, write a short note explaining the repair or service you require, include your name, address and phone number and send your stethoscope directly to: 4