Qualification of analytical
instrument of FTIR
Presented by : k. jayalakshmi
D/o k. kristaiah
1st yr M.pharm, pharmaceutical analysis
17-06-2020 1
17-06-2020 2
Introduction
 A simple apparatus to complex computerized systems, is used in the
pharmaceutical industry to acquire data that will help ensure that
products meet their specifications.
 Qualification, calibration, validation, and maintenance are the
parameters used to demonstrate that an instrument is qualified and
under control.
 Generally, the more complex the instrument, or the higher the criticality
of the measurement, the greater the amount of work that is required to
ensure that quality data will be generated. In addition, attention must be
paid to ensuring that data integrity and security are maintained.
 Instruments can generally be classified as belonging to Groups A, B, or
C. It should be noted that the same type of instrument can fit into one or
more categories, depending on its intended use.
17-06-2020 3
17-06-2020 4
Components of data quality
17-06-2020 5
Analytical Instrument Qualification:
AIQ is the collection of documented Evidence that an
instrument performs suitably for its intended purpose.
Analytical Method Validation:
Analytical method validation is the collection of documented
evidence that an analytical procedure is suitable for its intended
use.
System Suitability Tests:
System suitability tests verify that the system will perform in
accordance with the criteria set forth in the procedure.
Quality Control Check Samples:
Many analysts carry out their tests on instruments that have
been standardized by using reference materials and/or calibration
standards.
17-06-2020 6
VALIDATION
 It is a act of demonstrating and documenting that the process operates
effectively and reproducibly to produce a product meeting its
predetermined specifications and quality attributes.
 The aim of validation is to show that the critical steps are under
control and lead continuously to the desirable quality.
CALIBRATION
 Calibration is a comparison between measurements – one of known
magnitude or correctness made or set with one device and another
measurement made in as similar a way as possible with a second
device.
17-06-2020 7
QUALIFICATION:
Qualification is an act or process to assure something complies with some
conditions, standard or specific requirements.
Qualification Phases
The qualification phases consist of mainly 5 phases.
Its start with User requirement specification (URS).
URS is a document which states the laboratory requirement and technical
and operational requirements that should be met.
The other phases consist of
 Design Qualification (DQ)
 Installation Qualification (IQ)
 Operational Qualification (OQ)
 Performance Qualification (PQ)
17-06-2020 8
 Apart from URS, DQ, IQ, OQ and PQ; some other parameters are
to be performed such as Factory acceptance test (FAT) and site
acceptance test (SAT) where applicable.
 When the instrument undergoes major repairs or modifications,
this should be evaluated using change control.
 Relevant IQ,OQ, and/or PQ tests should be repeated to verify that
the instrument continues to operate satisfactorily.
 If an instrument is moved to another location, an assessment
should be made of what, if any, qualification stage should be
repeated.
17-06-2020 9
• Design Qualification:
Documented evidence which shows that the plant design agrees
with the design specifications of the customer.
• Installation Qualification:
Written evidence is given that all parts of equipment are installed according to
the equipment supplier’s and purchase specifications.
• Operational qualification:
Documented evidence which shows that all parts of the plant and equipment
work within their specifications and process parameters.
• Performance Qualification:
Provides documented evidence that all parts of a plant and other processes
produce products of specified quality under conditions of normal production
for a longer period of time.
17-06-2020 10
FTIR
 Fourier transform infrared spectroscopy (FTIR) is a technique which is
used to obtain an infrared spectrum of absorption, emission,
photoconductivity or Raman scattering of a solid, liquid or gas.
 An FTIR spectrometer simultaneously collects spectral data in a wide
spectral range. This confers a significant advantage over a dispersive
spectrometer which measures intensity over a narrow range of
wavelengths at a time.
17-06-2020 11
CALIBRATION
WAVE NUMBER PRESCISION:
• This is performed for substances with well known peak wave number(s)
position such as carbon dioxide, water vapour, polystyrene ,ammonia.
• Test is performed to know whether the exact peak wave numbers are
shown at that time of validation.
• Thus the result is obtained from the difference between the peak wave
numbers position for a substance with a well known peak wave numbers
and the values indicated by the system.
17-06-2020 12
0% TRANSMITTANCE:
• A sample which do not allow the transmission of light is measured in
order to investigate the 0% transmittance.
• This test thus can be used to find out error caused by stray light and
secondary emission spectra.
100% TRANSMITTANCE
• This is investigated by performing analysis with out a sample.
• By performing analysis with out the sample 100% transmittance can be
investigated.
17-06-2020 13
LINEARITY OF CURVE:
A calibration curve for the % transmittance and the
concentration is created and the linearity of the inspected.
REPRODUCIBILITY:
A stable sample is measured twice with in a short period and confirmed
whether the variation in the measurement values such as wave numbers
and transmittance are obtained.
VALIDATION OF FTIR:
• To perform FTIR validation and to confirm
that it is operating properly, diverse IR inspection
was performed by measuring the spectra of polystyrene film.
17-06-2020 14
INSTALLATION OF VALIDATION PROGRAMME
• Software validates the Shimadzu Fourier transform
infrared spectrophotometer IRPrestige-21/IRAffinity1/
FTIR-8000 series by the IR solution software to control them and data
processing
• The Validation program is automatically installed when the IR
solution is installed, therefore the Validation program does not need to
be separately installed.
PERFORMANCE QUALIFICATION
• POWER SPECTRUM
• Power spectrum gives the plot of portion of signal's power(energy per
unit time) falling with in the given frequency bins.
• This test estimates the intensity of power spectrum at a specified
wave numbers .
• When the measured intensity is equal to or larger than the criterion
value, the test is passed.
17-06-2020 15
ACCEPTANCE CRITERIA
17-06-2020 16
RESOLUTION
• The resolution is checked by recording the spectrum of polystyrene film of
approximately 35µm in thickness.
• The difference between percentage transmittance at the absorption
maximum A at 2870 cm-1 and absorption minima B at 2849.5 cm-1 must
be greater than 18.
• The difference between percentage transmittance at the absorption
maximum C at 1589 cm-1 and absorption minima D at 1583 cm-1 must be
greater than 12.
17-06-2020 17
WAVE NUMBER ACCURACY
• The wave number scale is usually calibrated by the
use of several characteristic wave numbers of a polystyrene film.
3060.0(+/-1.5) cm-1
2849.5(+/-1.5) cm-1
1942.9(+/-1.5) cm-1
1601.2(+/-1.0) cm-1
1583.0(+/-1.0) cm-1
1154.5(+/-1.0) cm-1
1028.3(+/-1.0) cm-1
• The software then judges whether the values are within in the allowable
range.
• The program labels the results “PASS” if all the peak numbers are within
the range.
• In case of dispersive spectrophotometer the permissible level of frequency
at 1601.2 cm-1 and at 1028.3 cm-1 should be with in +/- 2.0 cm-1.
17-06-2020 18
WAVE NUMBER REPRODUCIBILTY
• This program specifies three points to measure the peak wave numbers.
• Then it obtains the actual peak wave numbers at each point by measuring
the polystyrene film twice.
• It should satisfy 5 cm-1 around 3000 cm-1of polystyrene absorption wave
number,1 cm-1 around 1000 cm-1.
• The software determines whether the differences between each of two
measurements are within the allowable range and it labels the result PASS if
they are with in the range.
• EP 4.0 doesn’t include this inspection
17-06-2020 19
TRANSMITANCE REPRODUCIBILTY
• This program specifies peak wave number at three points and the
transmittance at each point is measured is twice.
• The transmittance reproducibility should satisfy 0.5%T when the
several points of polystyrene absorption from 3000 cm-1to 1000 cm-1are
measured twice.
• Then it is determined whether the differences between the two data
are within the allowable range and it labels the result PASS if they are
with in the range.
• All of the above furnished data should be represented in
17-06-2020 20
AS PER ASTM E1421-94 Level Zero
• This soft ware complies describes with in the description in the
ASTM(American Society for Testing and materials).
• The FTIR abnormalities or large changes over short term and
long term is assessed by these tests.
• The three parameters checked by this program are:
17-06-2020 21
ENERGY SPECTRUM TEST
 Power spectra obtained in the inspection are compared with
reference data and the spectra are checked for changes over long
periods.
ONE HUNDRED PERCENT LINE TEST
 100% T line spectra are calculated for power spectra & are measured
continuously in inspection and the spectra are checked for changes
over short periods.
POLYSTYRENE TEST
 Evaluation is performed using differences between spectra obtained
for polystyrene film in inspection and the stored reference data.
 When the differences are with in the standard, "pass” results.
 All of the above furnished data should be represented in Validation
Report.
17-06-2020 22
17-06-2020 23

Mais conteúdo relacionado

Mais procurados

Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.
begalavenu
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
RavichandraNadagouda
 

Mais procurados (20)

Qualification of glassware
Qualification of glasswareQualification of glassware
Qualification of glassware
 
Qualification Of HPTLC
Qualification Of HPTLCQualification Of HPTLC
Qualification Of HPTLC
 
validation of analytical method used in cleaning
validation of analytical method used in cleaningvalidation of analytical method used in cleaning
validation of analytical method used in cleaning
 
Qualification of HPLC
Qualification of HPLCQualification of HPLC
Qualification of HPLC
 
Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.Pharmaceutical validation, Quantification, streamlining, validation masterplan.
Pharmaceutical validation, Quantification, streamlining, validation masterplan.
 
Qualification of instrumets
Qualification of instrumetsQualification of instrumets
Qualification of instrumets
 
Validation qualification
Validation qualificationValidation qualification
Validation qualification
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
 
Requalification
RequalificationRequalification
Requalification
 
Qualification of Weights and Measures
Qualification of Weights and MeasuresQualification of Weights and Measures
Qualification of Weights and Measures
 
Qualification of Gas Chromatography
Qualification of Gas Chromatography Qualification of Gas Chromatography
Qualification of Gas Chromatography
 
Qualification of GC & FTIR
Qualification of GC & FTIRQualification of GC & FTIR
Qualification of GC & FTIR
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Computer system validations
Computer system validationsComputer system validations
Computer system validations
 
Qualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility SystemQualification of Dissolution Test Apparatus and Validation of Utility System
Qualification of Dissolution Test Apparatus and Validation of Utility System
 
QUALIFICATION OF HPLC , HPLC
QUALIFICATION OF HPLC , HPLCQUALIFICATION OF HPLC , HPLC
QUALIFICATION OF HPLC , HPLC
 
Qualification of laboratory equipments
Qualification of laboratory equipmentsQualification of laboratory equipments
Qualification of laboratory equipments
 
Qualification of manufacturing equipment.
Qualification of manufacturing equipment.Qualification of manufacturing equipment.
Qualification of manufacturing equipment.
 
Qualification Of HPLC
Qualification Of HPLCQualification Of HPLC
Qualification Of HPLC
 
Validation utility system
Validation utility systemValidation utility system
Validation utility system
 

Semelhante a Qualification of analytical instrument of FTIR

In-Spec UV Vis Standards and Good Lab Practice
In-Spec UV Vis Standards and Good Lab PracticeIn-Spec UV Vis Standards and Good Lab Practice
In-Spec UV Vis Standards and Good Lab Practice
Liza Tallon
 
EPM In-Line Verification 24-04-2014
EPM In-Line Verification 24-04-2014EPM In-Line Verification 24-04-2014
EPM In-Line Verification 24-04-2014
Bert Roos
 
Bruker_Atmosfir Dfenceline brochure_EN (3)
Bruker_Atmosfir Dfenceline brochure_EN (3)Bruker_Atmosfir Dfenceline brochure_EN (3)
Bruker_Atmosfir Dfenceline brochure_EN (3)
Ram Hashmonay, ScD
 
8.4 PERFORMANCE QUALIFICATION PROTOCOL FOR DISPENSING BOOTH (1).pdf
8.4 PERFORMANCE QUALIFICATION PROTOCOL FOR DISPENSING BOOTH (1).pdf8.4 PERFORMANCE QUALIFICATION PROTOCOL FOR DISPENSING BOOTH (1).pdf
8.4 PERFORMANCE QUALIFICATION PROTOCOL FOR DISPENSING BOOTH (1).pdf
abdo badr
 

Semelhante a Qualification of analytical instrument of FTIR (20)

Qualification of INFARED SPECTROSCOPY BY HIMAJA.D
Qualification of INFARED SPECTROSCOPY BY HIMAJA.DQualification of INFARED SPECTROSCOPY BY HIMAJA.D
Qualification of INFARED SPECTROSCOPY BY HIMAJA.D
 
Pharmaceutical validation
Pharmaceutical validationPharmaceutical validation
Pharmaceutical validation
 
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptxqualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx
 
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCQUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLC
 
PV S2 GOKULRAJ QUALIFICATION OF FTIR.pptx
PV S2 GOKULRAJ QUALIFICATION OF FTIR.pptxPV S2 GOKULRAJ QUALIFICATION OF FTIR.pptx
PV S2 GOKULRAJ QUALIFICATION OF FTIR.pptx
 
Validation of IR instrument
Validation of IR instrumentValidation of IR instrument
Validation of IR instrument
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
 
In-Spec UV Vis Standards and Good Lab Practice
In-Spec UV Vis Standards and Good Lab PracticeIn-Spec UV Vis Standards and Good Lab Practice
In-Spec UV Vis Standards and Good Lab Practice
 
Calibration
CalibrationCalibration
Calibration
 
Development of PLC based Transdermal Patch Evaluation System
Development of PLC based Transdermal Patch Evaluation SystemDevelopment of PLC based Transdermal Patch Evaluation System
Development of PLC based Transdermal Patch Evaluation System
 
EPM In-Line Verification 24-04-2014
EPM In-Line Verification 24-04-2014EPM In-Line Verification 24-04-2014
EPM In-Line Verification 24-04-2014
 
Bruker_Atmosfir Dfenceline brochure_EN (3)
Bruker_Atmosfir Dfenceline brochure_EN (3)Bruker_Atmosfir Dfenceline brochure_EN (3)
Bruker_Atmosfir Dfenceline brochure_EN (3)
 
DATION OF EQUIPMENT ICH AND WHO GUIDELINES FOR CALIBRATION AND VALIDATION OF ...
DATION OF EQUIPMENT ICH AND WHO GUIDELINES FOR CALIBRATION AND VALIDATION OF ...DATION OF EQUIPMENT ICH AND WHO GUIDELINES FOR CALIBRATION AND VALIDATION OF ...
DATION OF EQUIPMENT ICH AND WHO GUIDELINES FOR CALIBRATION AND VALIDATION OF ...
 
Cleanroom Qualification
Cleanroom QualificationCleanroom Qualification
Cleanroom Qualification
 
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptxICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
ICH AND WHO GUIDELINES FOR VALIDATION OF EQUIPMENTS.pptx
 
Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)
 
Quality assurance of treatment planning system by Rahim Gohar
Quality assurance of treatment planning system by Rahim GoharQuality assurance of treatment planning system by Rahim Gohar
Quality assurance of treatment planning system by Rahim Gohar
 
EQUIPMENT VALIDATION
EQUIPMENT VALIDATIONEQUIPMENT VALIDATION
EQUIPMENT VALIDATION
 
8.4 PERFORMANCE QUALIFICATION PROTOCOL FOR DISPENSING BOOTH (1).pdf
8.4 PERFORMANCE QUALIFICATION PROTOCOL FOR DISPENSING BOOTH (1).pdf8.4 PERFORMANCE QUALIFICATION PROTOCOL FOR DISPENSING BOOTH (1).pdf
8.4 PERFORMANCE QUALIFICATION PROTOCOL FOR DISPENSING BOOTH (1).pdf
 
Testing/Manufacture/Filter Catalogue
Testing/Manufacture/Filter CatalogueTesting/Manufacture/Filter Catalogue
Testing/Manufacture/Filter Catalogue
 

Mais de KUNDLAJAYALAKSHMI (7)

transfer of technology
transfer of technology transfer of technology
transfer of technology
 
Capillary electrophoresis
Capillary electrophoresisCapillary electrophoresis
Capillary electrophoresis
 
Polymerase chain reaction
Polymerase chain reactionPolymerase chain reaction
Polymerase chain reaction
 
Bio assay of heparin
Bio assay of heparinBio assay of heparin
Bio assay of heparin
 
Biological assay of oxytocin
Biological assay of oxytocinBiological assay of oxytocin
Biological assay of oxytocin
 
Potential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyserPotential elemental impurities ,chns analyser
Potential elemental impurities ,chns analyser
 
Chromatography
ChromatographyChromatography
Chromatography
 

Último

The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
heathfieldcps1
 
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
ZurliaSoop
 
Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functions
KarakKing
 

Último (20)

The basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptxThe basics of sentences session 3pptx.pptx
The basics of sentences session 3pptx.pptx
 
Wellbeing inclusion and digital dystopias.pptx
Wellbeing inclusion and digital dystopias.pptxWellbeing inclusion and digital dystopias.pptx
Wellbeing inclusion and digital dystopias.pptx
 
How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17
 
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
Jual Obat Aborsi Hongkong ( Asli No.1 ) 085657271886 Obat Penggugur Kandungan...
 
ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptx
 
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdfUnit 3 Emotional Intelligence and Spiritual Intelligence.pdf
Unit 3 Emotional Intelligence and Spiritual Intelligence.pdf
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
 
REMIFENTANIL: An Ultra short acting opioid.pptx
REMIFENTANIL: An Ultra short acting opioid.pptxREMIFENTANIL: An Ultra short acting opioid.pptx
REMIFENTANIL: An Ultra short acting opioid.pptx
 
Spatium Project Simulation student brief
Spatium Project Simulation student briefSpatium Project Simulation student brief
Spatium Project Simulation student brief
 
Sociology 101 Demonstration of Learning Exhibit
Sociology 101 Demonstration of Learning ExhibitSociology 101 Demonstration of Learning Exhibit
Sociology 101 Demonstration of Learning Exhibit
 
Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...
Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...
Sensory_Experience_and_Emotional_Resonance_in_Gabriel_Okaras_The_Piano_and_Th...
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.
 
Graduate Outcomes Presentation Slides - English
Graduate Outcomes Presentation Slides - EnglishGraduate Outcomes Presentation Slides - English
Graduate Outcomes Presentation Slides - English
 
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...
 
Towards a code of practice for AI in AT.pptx
Towards a code of practice for AI in AT.pptxTowards a code of practice for AI in AT.pptx
Towards a code of practice for AI in AT.pptx
 
Food safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdfFood safety_Challenges food safety laboratories_.pdf
Food safety_Challenges food safety laboratories_.pdf
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
Salient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functionsSalient Features of India constitution especially power and functions
Salient Features of India constitution especially power and functions
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17
 

Qualification of analytical instrument of FTIR

  • 1. Qualification of analytical instrument of FTIR Presented by : k. jayalakshmi D/o k. kristaiah 1st yr M.pharm, pharmaceutical analysis 17-06-2020 1
  • 3. Introduction  A simple apparatus to complex computerized systems, is used in the pharmaceutical industry to acquire data that will help ensure that products meet their specifications.  Qualification, calibration, validation, and maintenance are the parameters used to demonstrate that an instrument is qualified and under control.  Generally, the more complex the instrument, or the higher the criticality of the measurement, the greater the amount of work that is required to ensure that quality data will be generated. In addition, attention must be paid to ensuring that data integrity and security are maintained.  Instruments can generally be classified as belonging to Groups A, B, or C. It should be noted that the same type of instrument can fit into one or more categories, depending on its intended use. 17-06-2020 3
  • 5. Components of data quality 17-06-2020 5
  • 6. Analytical Instrument Qualification: AIQ is the collection of documented Evidence that an instrument performs suitably for its intended purpose. Analytical Method Validation: Analytical method validation is the collection of documented evidence that an analytical procedure is suitable for its intended use. System Suitability Tests: System suitability tests verify that the system will perform in accordance with the criteria set forth in the procedure. Quality Control Check Samples: Many analysts carry out their tests on instruments that have been standardized by using reference materials and/or calibration standards. 17-06-2020 6
  • 7. VALIDATION  It is a act of demonstrating and documenting that the process operates effectively and reproducibly to produce a product meeting its predetermined specifications and quality attributes.  The aim of validation is to show that the critical steps are under control and lead continuously to the desirable quality. CALIBRATION  Calibration is a comparison between measurements – one of known magnitude or correctness made or set with one device and another measurement made in as similar a way as possible with a second device. 17-06-2020 7
  • 8. QUALIFICATION: Qualification is an act or process to assure something complies with some conditions, standard or specific requirements. Qualification Phases The qualification phases consist of mainly 5 phases. Its start with User requirement specification (URS). URS is a document which states the laboratory requirement and technical and operational requirements that should be met. The other phases consist of  Design Qualification (DQ)  Installation Qualification (IQ)  Operational Qualification (OQ)  Performance Qualification (PQ) 17-06-2020 8
  • 9.  Apart from URS, DQ, IQ, OQ and PQ; some other parameters are to be performed such as Factory acceptance test (FAT) and site acceptance test (SAT) where applicable.  When the instrument undergoes major repairs or modifications, this should be evaluated using change control.  Relevant IQ,OQ, and/or PQ tests should be repeated to verify that the instrument continues to operate satisfactorily.  If an instrument is moved to another location, an assessment should be made of what, if any, qualification stage should be repeated. 17-06-2020 9
  • 10. • Design Qualification: Documented evidence which shows that the plant design agrees with the design specifications of the customer. • Installation Qualification: Written evidence is given that all parts of equipment are installed according to the equipment supplier’s and purchase specifications. • Operational qualification: Documented evidence which shows that all parts of the plant and equipment work within their specifications and process parameters. • Performance Qualification: Provides documented evidence that all parts of a plant and other processes produce products of specified quality under conditions of normal production for a longer period of time. 17-06-2020 10
  • 11. FTIR  Fourier transform infrared spectroscopy (FTIR) is a technique which is used to obtain an infrared spectrum of absorption, emission, photoconductivity or Raman scattering of a solid, liquid or gas.  An FTIR spectrometer simultaneously collects spectral data in a wide spectral range. This confers a significant advantage over a dispersive spectrometer which measures intensity over a narrow range of wavelengths at a time. 17-06-2020 11
  • 12. CALIBRATION WAVE NUMBER PRESCISION: • This is performed for substances with well known peak wave number(s) position such as carbon dioxide, water vapour, polystyrene ,ammonia. • Test is performed to know whether the exact peak wave numbers are shown at that time of validation. • Thus the result is obtained from the difference between the peak wave numbers position for a substance with a well known peak wave numbers and the values indicated by the system. 17-06-2020 12
  • 13. 0% TRANSMITTANCE: • A sample which do not allow the transmission of light is measured in order to investigate the 0% transmittance. • This test thus can be used to find out error caused by stray light and secondary emission spectra. 100% TRANSMITTANCE • This is investigated by performing analysis with out a sample. • By performing analysis with out the sample 100% transmittance can be investigated. 17-06-2020 13
  • 14. LINEARITY OF CURVE: A calibration curve for the % transmittance and the concentration is created and the linearity of the inspected. REPRODUCIBILITY: A stable sample is measured twice with in a short period and confirmed whether the variation in the measurement values such as wave numbers and transmittance are obtained. VALIDATION OF FTIR: • To perform FTIR validation and to confirm that it is operating properly, diverse IR inspection was performed by measuring the spectra of polystyrene film. 17-06-2020 14
  • 15. INSTALLATION OF VALIDATION PROGRAMME • Software validates the Shimadzu Fourier transform infrared spectrophotometer IRPrestige-21/IRAffinity1/ FTIR-8000 series by the IR solution software to control them and data processing • The Validation program is automatically installed when the IR solution is installed, therefore the Validation program does not need to be separately installed. PERFORMANCE QUALIFICATION • POWER SPECTRUM • Power spectrum gives the plot of portion of signal's power(energy per unit time) falling with in the given frequency bins. • This test estimates the intensity of power spectrum at a specified wave numbers . • When the measured intensity is equal to or larger than the criterion value, the test is passed. 17-06-2020 15
  • 17. RESOLUTION • The resolution is checked by recording the spectrum of polystyrene film of approximately 35µm in thickness. • The difference between percentage transmittance at the absorption maximum A at 2870 cm-1 and absorption minima B at 2849.5 cm-1 must be greater than 18. • The difference between percentage transmittance at the absorption maximum C at 1589 cm-1 and absorption minima D at 1583 cm-1 must be greater than 12. 17-06-2020 17
  • 18. WAVE NUMBER ACCURACY • The wave number scale is usually calibrated by the use of several characteristic wave numbers of a polystyrene film. 3060.0(+/-1.5) cm-1 2849.5(+/-1.5) cm-1 1942.9(+/-1.5) cm-1 1601.2(+/-1.0) cm-1 1583.0(+/-1.0) cm-1 1154.5(+/-1.0) cm-1 1028.3(+/-1.0) cm-1 • The software then judges whether the values are within in the allowable range. • The program labels the results “PASS” if all the peak numbers are within the range. • In case of dispersive spectrophotometer the permissible level of frequency at 1601.2 cm-1 and at 1028.3 cm-1 should be with in +/- 2.0 cm-1. 17-06-2020 18
  • 19. WAVE NUMBER REPRODUCIBILTY • This program specifies three points to measure the peak wave numbers. • Then it obtains the actual peak wave numbers at each point by measuring the polystyrene film twice. • It should satisfy 5 cm-1 around 3000 cm-1of polystyrene absorption wave number,1 cm-1 around 1000 cm-1. • The software determines whether the differences between each of two measurements are within the allowable range and it labels the result PASS if they are with in the range. • EP 4.0 doesn’t include this inspection 17-06-2020 19
  • 20. TRANSMITANCE REPRODUCIBILTY • This program specifies peak wave number at three points and the transmittance at each point is measured is twice. • The transmittance reproducibility should satisfy 0.5%T when the several points of polystyrene absorption from 3000 cm-1to 1000 cm-1are measured twice. • Then it is determined whether the differences between the two data are within the allowable range and it labels the result PASS if they are with in the range. • All of the above furnished data should be represented in 17-06-2020 20
  • 21. AS PER ASTM E1421-94 Level Zero • This soft ware complies describes with in the description in the ASTM(American Society for Testing and materials). • The FTIR abnormalities or large changes over short term and long term is assessed by these tests. • The three parameters checked by this program are: 17-06-2020 21
  • 22. ENERGY SPECTRUM TEST  Power spectra obtained in the inspection are compared with reference data and the spectra are checked for changes over long periods. ONE HUNDRED PERCENT LINE TEST  100% T line spectra are calculated for power spectra & are measured continuously in inspection and the spectra are checked for changes over short periods. POLYSTYRENE TEST  Evaluation is performed using differences between spectra obtained for polystyrene film in inspection and the stored reference data.  When the differences are with in the standard, "pass” results.  All of the above furnished data should be represented in Validation Report. 17-06-2020 22