IOSR Journal of Electrical and Electronics Engineering (IOSR-JEEE)
e-ISSN: 2278-1676 Volume 5, Issue 2 (Mar. - Apr. 2013), PP 12-20
www.iosrjournals.org
www.iosrjournals.org 12 | Page
Design of a micro device to electrically separate and count wbc
from whole blood
Soumya Gangopadhyay1
,AVimala Juliet2
1,2
(Department of Instrumentation and ControlEngineering,SRM University, India)
Abstract:This work presents a micro device which separates and counts Leukocytes (white blood cell or WBC)
from other two type of blood cells, Erythrocytes (red blood cell or RBC), and Thrombocytes (Platelet). The
operation is performed using four basic parts, a micro mixer which mixes whole blood and diluent to dilute the
blood, a size dependent electrical cell separator which separates WBC from other blood cells , a micro nozzle to
increase the flow velocity of cells, a counter unit to count the number of WBC in sample. The size dependent cell
sorting is done by electrical pillar arrays. The focusing of incoming cells to the counter is done by focusing
electrodes. The main aim of this work is to count the separated white blood cells and it is done by a MEMS
basedCoulter counter. Simple structure and ease of operation is the two main features of this device. The design
of the device has been done by simulating different unit of the device in ComsolMultiphysics. Simulation results
were satisfactory and each unit is able to perform its work as per requirement.
Keywords-Coulter counter, deterministic lateral displacement, Dielectrophoresis (DEP), micro mixer and
micro nozzle.
I. Introduction
There are efforts on cell separation and counting systems [1], [2], [3], [4]. There are different systems
to separate the blood cells e.g. fluorescence spectroscopy, magnetophoretic cell separation, physical pillar based
separation etc. But each of these systems have their own problems like fluorescence spectroscopy can’t count
the cells separately it gives an average count of cells and most importantly it is not possible to fabricate it in a
micro device. The magnetophoretic cell separator uses two permanent magnets which is not possible to place in
close range of micro device. The physical pillar based cell separator can do size based cell separation and can be
fabricated in a micro device, but after some use of the device the cells tend to stick to the channel floor and
create clogging [1]. In this work dielectrophoresis has been used for separating the blood cells which would
allow the cells neither to settle down to the channel floor nor to slow down during the process. In previous micro
Coulter counter devices there was a problem, there was a chance that more than one cells could get lumped
together and the counter would have been counting them as one [4]. In this work focusing electrodes has been
used for focusing the cell flow so that no two cells can get lumped together and the counter would count them
individually. In this work the dilution system has been implemented in the device. Hence no dilution needs to be
done before injecting the blood to the device.This device is a combination of the cell separation and counting
system which increases the efficiency to count the number of the most important cell in human blood which
causes most of the infectious diseases.
The three blood cells in human blood are respectively Erythrocyte (red blood cell or RBC), Leukocyte
(white blood cell or WBC) and Thrombocyte (Platelet). TABLE 1 compares the size of the three main blood
cells
TABLE 1. Comparison of wbc, rbc and platelet.
Blood Cells Cell Diameter (µm)
WBC 7.72 – 15
RBC 6 - 8
Platelet 2.65 – 2.9
TABLE 2. Description of different wbc.
Type Approx. % in
adults
Diameter (µm)
Neutrophil 62 10 - 12
Eosinophil 2.3 10 - 12
Basophil 0.4 12 - 15
Lymphocyte 30 7 - 15
Monocyte 5.3 7.72 – 9.99
Design of a micro device to separate and count wbc from whole blood
www.iosrjournals.org 13 | Page
WBC is of five main types neutrophil, eosinophil, basophil, lymphocyte and monocyte. TABLE 2 describes
their size and amount of presence in blood among the total leukocytes.
As the size of WBC ranges from 7.72 to 15 µm, so the cells having lesser diameter than 7.72µm is to be rejected
and the greater or equal diameter cell is to be selected and forwarded for the counting purpose.
II. Device Principle
2.1. Micro Mixer Unit
Blood needs to get diluted before it can be used for any test because if it is not diluted, the test solution
becomes over populated by cells and the count can’t be precise. Hence the blood is 1st
diluted to a certain
percentage to reduce the number of cells per unit sample. Another reason to dilute the blood is to prevent
clotting of blood in the channel.This unit mixes the blood with PBS (Phospheted Buffer Saline) to dilute the
blood.
This unit has a simple structure as shown in figure.1 in which the blood and PBS is injected into the
device from two different inlets and directed to a junction where the two fluid streams get mixed and then
moved through a zigzag long channel to give both the fluids to get mixed well and then finally the mixture or
the diluted blood is released in a specific point of the cell separator unit.
In case of the fabricated device the dilution factor is to be used is 1:100 i.e. 0.1ml of whole blood is to be mixed
with 9.9ml of PBS to prepare a 10ml of diluted blood. The naturel count of WBC in blood is 4500 to 10800 /µl.
After dilution the count reduces to 45 to 108/µl. It reduces the time to count the cells for the
device. 𝐷𝑖𝑙𝑢𝑡𝑖𝑜𝑛 𝐹𝑎𝑐𝑡𝑜𝑟 =
𝑆𝑎𝑚𝑝𝑙𝑒 𝑉𝑜𝑙𝑢𝑚𝑒
𝑇𝑜𝑡𝑎𝑙 𝑉𝑜𝑙𝑢𝑚𝑒
(1)
Figure1 Micro mixer unit.
2.2. Cell Separation Unit
This unit consists of a two dimensional array of circular spot electrodeswhere each of the row of the
array is shifted by ΔX. Each row consists of alternate pairs of VDEP and GND electrodes. DEP force is generated
by these electrodes at a particular voltage and frequency of a square wave AC signal.
Figure 2 Arrangement of the spot electrodes.
Design of a micro device to separate and count wbc from whole blood
www.iosrjournals.org 14 | Page
Working equation for deterministic lateral displacement is as follows:
𝑌 = 𝜋 − 𝛼 𝑟𝑒 +
𝑋−𝑟
2
sin 𝛼(2)
𝑟𝑐
∗
=
∆𝑋−
𝑋−𝑟
2
𝜋− 𝛼
sin 𝛼(3)
Where Y is the lateral displacement of the cell along the flow path. If Y>ΔX then the cell would follow
thezigzag path between the electrodes and if Y<ΔX then the cell would follow the displacement mode. Now re is
the diameter of the cell which decides the value of Y i.e. cells with smaller diameter will flow in zigzag mode
and cells with larger diameter will flow in deterministic lateral displacement mode.rcdepends on the applied
voltage on the electrodes.rc
*
is the critical diameter of the cell i.e. if the celldiameteris less than rc
*
then the cell
will flow in zigzag mode andifitisgreater than rc
*
then it would flow in deterministic lateral displacement mode.
This kind of displacement is called deterministic displacement because the lateral displacement Y of the cell can
be determined by the formula equation (2)[2].
The working equation for DEP force is:
𝐹𝐷𝐸𝑃 = 2𝜋𝑅3
𝜀 𝑚 𝑅𝐸 𝐹𝐶𝑀 𝛻𝐸 𝑅𝑀𝑆
2
(4)
𝐹𝐶𝑀 =
𝜀 𝑝
∗ − 𝜀 𝑚
∗
𝜀 𝑝
∗ − 2𝜀 𝑚
∗ (5)
𝜀 𝑚
∗
= 𝜀 𝑚 −
j𝜍 𝑚
𝜔
(6)
𝜀 𝑝
∗
= 𝜀 𝑝 −
j𝜍 𝑝
𝜔
(7)
Where, ∇ERMS is the non-uniformity factor of the applied electric field i.e. the rms value of the non-uniform
electric gradient between the electrodes, FCM is the Clausius-Mosotti factor which depends on the complex
permittivity of the medium and the cell which are respectively εm
*
and εp
*
. In this case complex permittivity has
taken in account because the conductivity and the angular frequency of the applied electric field effects the
electric permittivity of the cell and the medium as in the equation 6 and 7. Now the complex permittivity effects
the FCM. In case of this device the particles should get repelled by the electric pillars hence we need to get a
negative DEP force to achieve the required repelling force.To create a negative DEP force the FCM should be
negative because it decides the direction of the force with respect to the electric pillars [6].
There is a phenomenon called β dispersion in which the cell accepts only the high frequency signals in
range of kHz to few MHz to pass through it and disperses the high range of frequencies. So the angular
frequency should be high enough, so that the cell would allow the high frequency signal to pass through it and
the cells would get effected by the electric field.
Figure3Nozzle structure.
2.3. Micro Nozzle Unit
This unit is a micro fabricated nozzle. Which increases the flow rate of the fluid cell mixture. In case of a nozzle
the flow resistance ζn is responsible for the increase in flow rate. ζn can be calculated by the equation 8 (as the
Reynolds’s number is very low ):
𝜁𝑛 =
𝐴 𝑛
𝑅 𝑒
(8)
𝑅𝑒 =
𝑣 𝑜 𝐷 𝑜
ν
(9)
𝐴 𝑛 =
19
𝑛 𝑜
0.5 𝑡𝑎𝑛 𝛼 0.75 (10)
Where Re is the Reynolds’s number of the flow, υ is the kinematic viscosity and no = (D1
2
/ Do
2
).As the fluid
gets resistance in its path of flow the pressure starts to drop and according to the law of conservation of energy
Design of a micro device to separate and count wbc from whole blood
www.iosrjournals.org 15 | Page
the flow rate of the fluid increases to compensate the loss of energy [5]. The increase of flow rate is required to
get faster response from the device.
The nozzle section also contains alternate positive and ground electrodes at its two sides. These electrodes are
used for focusing the flow of the cells.
2.4. Cell Counting Unit
This unit uses aperture impedance method. In this method there is an aperture and there are two
electrodes at both sides of that aperture. The configuration has been shown in figure 4. The two electrodes are
connected through the liquid medium. An AC voltage is applied to the electrodes. Whenever there is a cell in
between them there is a change of resistance in the circuit (ΔR).
Figure 4 Electrode over the channel with a cell in between them in flowing medium.
The change in resistance is represented by:
Δ𝑅 =
𝜌
𝐴2 𝑉𝑝(11)
Where A is the cross sectional area of the channel, ρ is the specific resistance of the medium, Vpis the
volume of the cell. A and ρ remains constant hence the magnitude of ΔR depends on Vp[3].
III. Device Simulation Result
Each unit of the device has been simulated with COMSOL Multiphysics 4.3a.
3.1. Micro Mixer Unit
This unit has been simulated with laminar flow domain. Here velocity for blood and PBS has been used
1.1µm/sec and 2.1µm/sec respectively. As it results in a uniform flow rate and proper mixing of blood and PBS.
The resulting velocity of the diluted blood is 10.61µm/sec.
(a)
Design of a micro device to separate and count wbc from whole blood
www.iosrjournals.org 16 | Page
(b)
Figure 5 (a) Simulation of flow velocity in micro-mixer, (b) Simulation of change in Reynolds’s number along
the channel.
Figure 6 Flow velocity along the channel of micro-mixer.
3.2.Cell Separation Unit
The separation unit has been simulated by using domains like electric current, particle tracing and
laminar flow.
In electric current domain the alternate electrodes in a row has been assigned as terminals with a voltage signal
of 20V and other electrodes as ground. The analysis has been done in frequency analysis study with 20 kHz
frequency square wave signal to get an AC signal at the electrodes.
(a)
Design of a micro device to separate and count wbc from whole blood
www.iosrjournals.org 17 | Page
(b)
Figure 7 (a) Electric field along a row of the electrode arrangement. (b) Electric field along a column of the
electrode arrangement.
Figure 8 Simulated electrode arrangement in the device with arrows indicating the direction of the electric field.
In particle tracing domain the particles has been used to represent blood cells. In laminar flow domain an inlet
flow of 10.61µm/sec has been used for the diluted blood flow as in this range of velocity the cell gets the proper
separation time. In case of higher velocity the cells just gets washed off from the surface.
At 20 V the electric field covers the electrode and increases the diameter of the electrodes. So that is
how we can change the diameter of the electrodes to get the right value of Y which would decide the mode of
flow of the cells along the channel.With no electric field applied, the value of Y is equal to ΔX and rc
*
= 0.2µm.
In simulation the channel dimensions has been taken as 500 X 500 µm (to test the electrodes functionality in
case of the actual device the width would be sufficient but the length of the channel must be increased to get the
proper separation). The required FCM to create the negative DEP force is -0.2353.
The particle tracing domain analysis simulation results can’t be given in this paper because the animation is
impossible to represent in a paper.
3.3. Micro Nozzle Unit
This unit has been simulated using only the laminar flow domain.
The inlet dimension of the nozzle is 60 X 40 µm and the outlet dimension is 30 x 20 µm. The inlet flow velocity
of the nozzle is 0.41134 x 10-6
m/s and outlet flow velocity is 3.45 x 10-6
m/s.This arrangement shows a pressure
drop of 0.0152Pa. The flow resistance co-efficient is 0.226.
Design of a micro device to separate and count wbc from whole blood
www.iosrjournals.org 18 | Page
Figure 9Simulation of the cell focusing electrodes at the micro nozzle and the counter electrodes.
(a) (b)
Figure10 (a) Pressure drop along the channel, (b) Flow velocity magnitude along the micro nozzle.
(a) (b)
Figure 11 (a)Trajectory of the cell along the channel for different electric field along y-axis,(b) Trajectory of cell
along the channel when electric field along y-axis is 4 x 105
V/m and electric field along z-axis has been varied.
From Fig.11 (a) it is evident that when 4 x 105
V/m electric field is applied to the focusing electrodes along y-
axis, cells are following a more linear path than the other applied electric fields. So 7V square wave AC voltage
is applied to the focusing electrodes applying force along y-axis. Form figure 11 (b) it can be seen that when
another set of focusing electrode has been used to apply an electric field along z-axis the particles becomes more
non-linear than without this focusing electrode. But as the test has been done on simulated cells not on the actual
biological cell. Hence it is expected that in case of the actual biological cell the focusing electrodes at z-axis
would be required and an electric field of 3 x 105
V/m should be applied to the electrodes. The two sets of
focusing electrodes would act as a lens to focus the cells.
Hence the cell focusing electrodes are excited by square wave AC signal of 7V and 20 kHz frequency to keep
the cell flow in a linear path and most preferably along the central line of the nozzle.
Design of a micro device to separate and count wbc from whole blood
www.iosrjournals.org 19 | Page
3.3. Cell CountingUnit
This unit has been simulated using laminar flow, electric current and particle tracing physics.
Figure 12 Cell counter circuit.
The circuit used for sensing the change in resistance has been given in the Fig. 12. The cell counter unit is
connected in a voltage divider circuit and the output of that voltage divider is connected with an amplifier, so
that if any resistance change occurs in the counter unit the amplifier would give a pulse equivalent to the
resistance change.
Figure 13 Trajectory of the cell along the channel after the micro-nozzle unit for different electric filed applied
to the y-axis.
From Fig. 13 it can be seen that the cell motion along the channel is linear when the applied electric field is 3 x
105
V/m. But at 4 x 105
V/m as the cell comes nearer to the counter electrodes, it gets more focused at the center
of the channel width as desired. Hence 4 x 105
V/m electric field at y-axis is appropriate for the counter
electrodes.
The inlet flow of this unit is as same as the outlet of the micro nozzle. The electrodes used for sensing the
change in resistance are excited by 20 kHz square wave AC signal as with a voltage of 7V.
From figure 14 it can be seen that when cell is not present in between the electrodes the voltage at the middle of
the channel is much higher than the times when the cell is present. As the cell is an obstraction to the electric
field, the voltage when cell is present ranges from 0 to 0.5 V. This change in voltage can be detected by the
circuit and the out put can be given accordingly. Every time a cell passes through the gapbetween the electrodes
this changeoccures and a pulse is generated. Thus it can detect the cells and the number of pulses give the count
of wbc.
Design of a micro device to separate and count wbc from whole blood
www.iosrjournals.org 20 | Page
Figure 14 Change in voltage in between the electrodes due to prescence of cell of different sizes.
IV. Conclusion
The MEMS based device has been simulated using COMSOL Multiphysics 4.3a. The results were
satisfactory. To get proper results physical properties of human blood and PBS has been used for the simulated
fluids. The cell separation unit has been simulated with the proper voltage excitation. The micro nozzle part is
increasing the flow velocity as desirable and its cell focusing electrodes are allowing the flow of cell to be
linearized. But due to lack of availability of fabrication facility, the device is still in design phase. Hence the
results have been derived from the simulation rather than from the fabricated device. The results presented in the
paper are only simulation results and practical results are unavailable. The circuit to count the Leukocytes has
been designed and would be implemented practically and calibrated only after the device gets fabricated. Also
the length of the separation channel can be decided only after fabrication and testing of the fabricated device. It
is expected that the fabrication facility would be arranged in future and this device would get fabricated as soon
as possible. It is believed that this device would help doctors as well as common people to diagnose their
Leukocyte related diseases quicker and with more ease.
References
[1] SiyangZhmg, Raylene Yung, Yu-Chong Tail, Harvey Kasdan“Determinstic lateral displacement mems device for continuous blood
cellseparation”, Caltech Micromachining Lab, California Institute of Technology, USA,StanfordUniversity, USAInsDiagnostics,
International Remote Imaging Systems, Inc., USA, 8732-5/05 IEEE851-854 2007
[2] Young-Ho Cho, “Bio-Inspired MEMS Devices forElectrical Cell Separation and Mechanical Cell Characterization”, Digital
Nanolocomotion Center, Korea Advanced Institute of Science and Technology (KAIST) 373-1 Guseong-dong, Yuseong-gu,
Daejeon 305-701 Republic ofKorea 978-1-4244-1858-9/07 IEEE 514 - 5182007
[3] Daisuke Satakea,, Hiroyuki Ebia, NarihiroOkub, KoichiroMatsudaa, HidekuniTakaoc, MitsuakiAshikic, Makoto Ishidac, “A sensor
for blood cell counter using MEMS technology”, Fundamental Technology ResearchDepartment, HORIBA Ltd., 2 Miyanohigashi,
Kisshoin,Minami-ku, Kyoto 601-8510, Japan,Medical Electronic Systems, HORIBA Ltd., 2 Miyanohigashi, Kisshoin, Minami-ku,
Kyoto 601-8510, Japan,Department of Electrical and Electronic Engineering, Toyohashi University of Technology, 1-1,
Hibarigaoka, Tempaku-cho,Toyohashi 441-8580, Japan.Sensors and Actuators B 83 77–81 2002
[4] M. Korampally, J. D. Benson, Y. Wu, 1 J. K. Critser, M. Almasri, “MEMS based Coulter counter for cell sizing”,Department of
Electricaland Computer Engineering, and Department of Veterinary,Pathobiology, University of Missouri, Columbia, Missouri,
65201, USA. Proc. of SPIE Vol. 6886, 68860A, 0277-786X/08/$18 · doi: 10.1117/12.763741 2008 1-7
[5] X.N.Jiang ,Y.Zhou, Y.Li, Y.Yang, X.Y.Huang , and C.Y.Liu. “Experiments and analysis for micronozzle/diffuser flow and
microvalveless pumps”, School of MPE, NanyangTechnological University Singapore639798,Dept. of Precision Instruments and
Mechanology, TsinghuaUniversity, Beijing, 100084,China.0-7803-3829-4/97 1997 IEEE 369 - 372
[6] S. Burgarella, M. Bianchessiand M. De Fazio
,“
Numerical modeling of dielectrophoretic forces acting uponbiological cells in
silicon lab-on-chip devices
” ,
STMicroelectronics, Advanced System Technology, R&I e-Health ,Corresponding author:
STMicroelectronics, Centro Dir. Colleoni, Pal. Dialettica, Via Cardano 1, 20041AgrateBrianza (MI), Italy.Excerpt from the
Proceedings of the COMSOL Users Conference Grenoble 2007

Mais conteúdo relacionado

Mais procurados

Automated cell counters: principle and types
Automated cell counters: principle and typesAutomated cell counters: principle and types
Automated cell counters: principle and typesSivaranjini N
 
Automation in hematology
Automation in hematologyAutomation in hematology
Automation in hematologyDr Siddartha
 
Automated blood cell counter by yasin arafat
Automated blood cell counter by yasin arafatAutomated blood cell counter by yasin arafat
Automated blood cell counter by yasin arafatYasin Arafat
 
Automation in haematology
Automation in haematologyAutomation in haematology
Automation in haematologyraj kumar
 
Automated cell counter & its quality control
Automated cell counter & its quality controlAutomated cell counter & its quality control
Automated cell counter & its quality controlSaikat Mandal
 
Automated cell counters
Automated  cell countersAutomated  cell counters
Automated cell countersRicha Sharma
 
Automated blood cell counter by prof. dr. ma mohsin
Automated blood cell counter by prof. dr. ma mohsinAutomated blood cell counter by prof. dr. ma mohsin
Automated blood cell counter by prof. dr. ma mohsinMuhammad Tareque Hasan
 
Automation in clinical hematology
Automation in clinical hematologyAutomation in clinical hematology
Automation in clinical hematologyTitto Rahim
 
Histosch cw ieng
Histosch cw iengHistosch cw ieng
Histosch cw iengbabu3151
 
Automation in haematology bernard
Automation in haematology   bernardAutomation in haematology   bernard
Automation in haematology bernardBosco Mbonimpa
 
DETECTING AND COUNTING THE NO. OF WHITE BLOOD CELLS IN BLOOD SAMPLE IMAGES BY...
DETECTING AND COUNTING THE NO. OF WHITE BLOOD CELLS IN BLOOD SAMPLE IMAGES BY...DETECTING AND COUNTING THE NO. OF WHITE BLOOD CELLS IN BLOOD SAMPLE IMAGES BY...
DETECTING AND COUNTING THE NO. OF WHITE BLOOD CELLS IN BLOOD SAMPLE IMAGES BY...IJEEE
 
Reticulocyte count
Reticulocyte countReticulocyte count
Reticulocyte countShabab Ali
 

Mais procurados (18)

Automated cell counters: principle and types
Automated cell counters: principle and typesAutomated cell counters: principle and types
Automated cell counters: principle and types
 
Automation in hematology
Automation in hematologyAutomation in hematology
Automation in hematology
 
Automated blood cell counter by yasin arafat
Automated blood cell counter by yasin arafatAutomated blood cell counter by yasin arafat
Automated blood cell counter by yasin arafat
 
Coulter Counter
Coulter CounterCoulter Counter
Coulter Counter
 
Automation in haematology
Automation in haematologyAutomation in haematology
Automation in haematology
 
Automated cell counter & its quality control
Automated cell counter & its quality controlAutomated cell counter & its quality control
Automated cell counter & its quality control
 
Histograms
HistogramsHistograms
Histograms
 
Ss4
Ss4Ss4
Ss4
 
pdf
pdfpdf
pdf
 
Hematology auto analyzer
Hematology auto analyzerHematology auto analyzer
Hematology auto analyzer
 
Automated cell counters
Automated  cell countersAutomated  cell counters
Automated cell counters
 
Automated blood cell counter by prof. dr. ma mohsin
Automated blood cell counter by prof. dr. ma mohsinAutomated blood cell counter by prof. dr. ma mohsin
Automated blood cell counter by prof. dr. ma mohsin
 
Automation in clinical hematology
Automation in clinical hematologyAutomation in clinical hematology
Automation in clinical hematology
 
Histosch cw ieng
Histosch cw iengHistosch cw ieng
Histosch cw ieng
 
Automation in haematology bernard
Automation in haematology   bernardAutomation in haematology   bernard
Automation in haematology bernard
 
Blood cell counters
Blood cell countersBlood cell counters
Blood cell counters
 
DETECTING AND COUNTING THE NO. OF WHITE BLOOD CELLS IN BLOOD SAMPLE IMAGES BY...
DETECTING AND COUNTING THE NO. OF WHITE BLOOD CELLS IN BLOOD SAMPLE IMAGES BY...DETECTING AND COUNTING THE NO. OF WHITE BLOOD CELLS IN BLOOD SAMPLE IMAGES BY...
DETECTING AND COUNTING THE NO. OF WHITE BLOOD CELLS IN BLOOD SAMPLE IMAGES BY...
 
Reticulocyte count
Reticulocyte countReticulocyte count
Reticulocyte count
 

Destaque

Performance Evaluation of Routing Protocols for MAC Layer Models
Performance Evaluation of Routing Protocols for MAC Layer ModelsPerformance Evaluation of Routing Protocols for MAC Layer Models
Performance Evaluation of Routing Protocols for MAC Layer ModelsIOSR Journals
 
Correlation Preserving Indexing Based Text Clustering
Correlation Preserving Indexing Based Text ClusteringCorrelation Preserving Indexing Based Text Clustering
Correlation Preserving Indexing Based Text ClusteringIOSR Journals
 
Detecting the High Level Similarities in Software Implementation Process Usin...
Detecting the High Level Similarities in Software Implementation Process Usin...Detecting the High Level Similarities in Software Implementation Process Usin...
Detecting the High Level Similarities in Software Implementation Process Usin...IOSR Journals
 
Temperature and Azimuth angle variation effect on the Building Integrated Pho...
Temperature and Azimuth angle variation effect on the Building Integrated Pho...Temperature and Azimuth angle variation effect on the Building Integrated Pho...
Temperature and Azimuth angle variation effect on the Building Integrated Pho...IOSR Journals
 
Morpho-Palynological Studies On The Angiospermic Fern Cyrtomium Cyrotideum C....
Morpho-Palynological Studies On The Angiospermic Fern Cyrtomium Cyrotideum C....Morpho-Palynological Studies On The Angiospermic Fern Cyrtomium Cyrotideum C....
Morpho-Palynological Studies On The Angiospermic Fern Cyrtomium Cyrotideum C....IOSR Journals
 
Disposition Pattern of Sulphur Functional Groups in High Sulphur Ledo Coals o...
Disposition Pattern of Sulphur Functional Groups in High Sulphur Ledo Coals o...Disposition Pattern of Sulphur Functional Groups in High Sulphur Ledo Coals o...
Disposition Pattern of Sulphur Functional Groups in High Sulphur Ledo Coals o...IOSR Journals
 
Transient Stability Enhancement of Grid by Using Fuzzy Logic Based Statcom
Transient Stability Enhancement of Grid by Using Fuzzy Logic Based StatcomTransient Stability Enhancement of Grid by Using Fuzzy Logic Based Statcom
Transient Stability Enhancement of Grid by Using Fuzzy Logic Based StatcomIOSR Journals
 
In Vitro Gas Production Parameters and Characteristics of Four Types of Sweet...
In Vitro Gas Production Parameters and Characteristics of Four Types of Sweet...In Vitro Gas Production Parameters and Characteristics of Four Types of Sweet...
In Vitro Gas Production Parameters and Characteristics of Four Types of Sweet...IOSR Journals
 
Open Source Software Survivability Analysis Using Communication Pattern Valid...
Open Source Software Survivability Analysis Using Communication Pattern Valid...Open Source Software Survivability Analysis Using Communication Pattern Valid...
Open Source Software Survivability Analysis Using Communication Pattern Valid...IOSR Journals
 
Environmental health Effect and Air Pollution from cigarette smokers in Cross...
Environmental health Effect and Air Pollution from cigarette smokers in Cross...Environmental health Effect and Air Pollution from cigarette smokers in Cross...
Environmental health Effect and Air Pollution from cigarette smokers in Cross...IOSR Journals
 
An Efficient Biological Sequence Compression Technique Using LUT and Repeat ...
An Efficient Biological Sequence Compression Technique Using  LUT and Repeat ...An Efficient Biological Sequence Compression Technique Using  LUT and Repeat ...
An Efficient Biological Sequence Compression Technique Using LUT and Repeat ...IOSR Journals
 
Allelopathic effect of Albizia saman F. Muell on three widely cultivated Indi...
Allelopathic effect of Albizia saman F. Muell on three widely cultivated Indi...Allelopathic effect of Albizia saman F. Muell on three widely cultivated Indi...
Allelopathic effect of Albizia saman F. Muell on three widely cultivated Indi...IOSR Journals
 
Portfolio Cost Management in Offshore Software Development Outsourcing Relat...
Portfolio Cost Management in Offshore Software Development  Outsourcing Relat...Portfolio Cost Management in Offshore Software Development  Outsourcing Relat...
Portfolio Cost Management in Offshore Software Development Outsourcing Relat...IOSR Journals
 
Multi Wavelet for Image Retrival Based On Using Texture and Color Querys
Multi Wavelet for Image Retrival Based On Using Texture and  Color QuerysMulti Wavelet for Image Retrival Based On Using Texture and  Color Querys
Multi Wavelet for Image Retrival Based On Using Texture and Color QuerysIOSR Journals
 
Generation and Implementation of Barker and Nested Binary codes
Generation and Implementation of Barker and Nested Binary codesGeneration and Implementation of Barker and Nested Binary codes
Generation and Implementation of Barker and Nested Binary codesIOSR Journals
 
Influence of Different Soil Management Practices On Soil Properties and Its I...
Influence of Different Soil Management Practices On Soil Properties and Its I...Influence of Different Soil Management Practices On Soil Properties and Its I...
Influence of Different Soil Management Practices On Soil Properties and Its I...IOSR Journals
 
Power Optimization in MIMO-CN with MANET
Power Optimization in MIMO-CN with MANETPower Optimization in MIMO-CN with MANET
Power Optimization in MIMO-CN with MANETIOSR Journals
 

Destaque (20)

Performance Evaluation of Routing Protocols for MAC Layer Models
Performance Evaluation of Routing Protocols for MAC Layer ModelsPerformance Evaluation of Routing Protocols for MAC Layer Models
Performance Evaluation of Routing Protocols for MAC Layer Models
 
Correlation Preserving Indexing Based Text Clustering
Correlation Preserving Indexing Based Text ClusteringCorrelation Preserving Indexing Based Text Clustering
Correlation Preserving Indexing Based Text Clustering
 
Detecting the High Level Similarities in Software Implementation Process Usin...
Detecting the High Level Similarities in Software Implementation Process Usin...Detecting the High Level Similarities in Software Implementation Process Usin...
Detecting the High Level Similarities in Software Implementation Process Usin...
 
Temperature and Azimuth angle variation effect on the Building Integrated Pho...
Temperature and Azimuth angle variation effect on the Building Integrated Pho...Temperature and Azimuth angle variation effect on the Building Integrated Pho...
Temperature and Azimuth angle variation effect on the Building Integrated Pho...
 
A0510107
A0510107A0510107
A0510107
 
Morpho-Palynological Studies On The Angiospermic Fern Cyrtomium Cyrotideum C....
Morpho-Palynological Studies On The Angiospermic Fern Cyrtomium Cyrotideum C....Morpho-Palynological Studies On The Angiospermic Fern Cyrtomium Cyrotideum C....
Morpho-Palynological Studies On The Angiospermic Fern Cyrtomium Cyrotideum C....
 
Disposition Pattern of Sulphur Functional Groups in High Sulphur Ledo Coals o...
Disposition Pattern of Sulphur Functional Groups in High Sulphur Ledo Coals o...Disposition Pattern of Sulphur Functional Groups in High Sulphur Ledo Coals o...
Disposition Pattern of Sulphur Functional Groups in High Sulphur Ledo Coals o...
 
Transient Stability Enhancement of Grid by Using Fuzzy Logic Based Statcom
Transient Stability Enhancement of Grid by Using Fuzzy Logic Based StatcomTransient Stability Enhancement of Grid by Using Fuzzy Logic Based Statcom
Transient Stability Enhancement of Grid by Using Fuzzy Logic Based Statcom
 
H0434555
H0434555H0434555
H0434555
 
In Vitro Gas Production Parameters and Characteristics of Four Types of Sweet...
In Vitro Gas Production Parameters and Characteristics of Four Types of Sweet...In Vitro Gas Production Parameters and Characteristics of Four Types of Sweet...
In Vitro Gas Production Parameters and Characteristics of Four Types of Sweet...
 
Open Source Software Survivability Analysis Using Communication Pattern Valid...
Open Source Software Survivability Analysis Using Communication Pattern Valid...Open Source Software Survivability Analysis Using Communication Pattern Valid...
Open Source Software Survivability Analysis Using Communication Pattern Valid...
 
Environmental health Effect and Air Pollution from cigarette smokers in Cross...
Environmental health Effect and Air Pollution from cigarette smokers in Cross...Environmental health Effect and Air Pollution from cigarette smokers in Cross...
Environmental health Effect and Air Pollution from cigarette smokers in Cross...
 
An Efficient Biological Sequence Compression Technique Using LUT and Repeat ...
An Efficient Biological Sequence Compression Technique Using  LUT and Repeat ...An Efficient Biological Sequence Compression Technique Using  LUT and Repeat ...
An Efficient Biological Sequence Compression Technique Using LUT and Repeat ...
 
Allelopathic effect of Albizia saman F. Muell on three widely cultivated Indi...
Allelopathic effect of Albizia saman F. Muell on three widely cultivated Indi...Allelopathic effect of Albizia saman F. Muell on three widely cultivated Indi...
Allelopathic effect of Albizia saman F. Muell on three widely cultivated Indi...
 
Portfolio Cost Management in Offshore Software Development Outsourcing Relat...
Portfolio Cost Management in Offshore Software Development  Outsourcing Relat...Portfolio Cost Management in Offshore Software Development  Outsourcing Relat...
Portfolio Cost Management in Offshore Software Development Outsourcing Relat...
 
Multi Wavelet for Image Retrival Based On Using Texture and Color Querys
Multi Wavelet for Image Retrival Based On Using Texture and  Color QuerysMulti Wavelet for Image Retrival Based On Using Texture and  Color Querys
Multi Wavelet for Image Retrival Based On Using Texture and Color Querys
 
Generation and Implementation of Barker and Nested Binary codes
Generation and Implementation of Barker and Nested Binary codesGeneration and Implementation of Barker and Nested Binary codes
Generation and Implementation of Barker and Nested Binary codes
 
Influence of Different Soil Management Practices On Soil Properties and Its I...
Influence of Different Soil Management Practices On Soil Properties and Its I...Influence of Different Soil Management Practices On Soil Properties and Its I...
Influence of Different Soil Management Practices On Soil Properties and Its I...
 
G0444246
G0444246G0444246
G0444246
 
Power Optimization in MIMO-CN with MANET
Power Optimization in MIMO-CN with MANETPower Optimization in MIMO-CN with MANET
Power Optimization in MIMO-CN with MANET
 

Semelhante a Design of a micro device to electrically separate and count wbc from whole blood

automationinhematologypart1-160820102053.pptx
automationinhematologypart1-160820102053.pptxautomationinhematologypart1-160820102053.pptx
automationinhematologypart1-160820102053.pptxGPBelwal
 
A novel active charge equalisation method for series connected batteries
A novel active charge equalisation method for series connected batteriesA novel active charge equalisation method for series connected batteries
A novel active charge equalisation method for series connected batterieseSAT Publishing House
 
UROP Proposal Spring 2014 (CTC)
UROP Proposal Spring 2014 (CTC)UROP Proposal Spring 2014 (CTC)
UROP Proposal Spring 2014 (CTC)Anthony Han
 
AUTOMATION IN HAEMATOLOGY.pptx
AUTOMATION IN HAEMATOLOGY.pptxAUTOMATION IN HAEMATOLOGY.pptx
AUTOMATION IN HAEMATOLOGY.pptxFraishu
 
A simple instrumentation system for separation of whole blood components usin...
A simple instrumentation system for separation of whole blood components usin...A simple instrumentation system for separation of whole blood components usin...
A simple instrumentation system for separation of whole blood components usin...eSAT Publishing House
 
Finite Element Simulation of Microfluidic Biochip for High Throughput Hydrody...
Finite Element Simulation of Microfluidic Biochip for High Throughput Hydrody...Finite Element Simulation of Microfluidic Biochip for High Throughput Hydrody...
Finite Element Simulation of Microfluidic Biochip for High Throughput Hydrody...TELKOMNIKA JOURNAL
 
DIELECTROPHORETIC DEFORMATION OF ERYTHROCYTES ON TRANSPARENT INDIUM TIN OXIDE...
DIELECTROPHORETIC DEFORMATION OF ERYTHROCYTES ON TRANSPARENT INDIUM TIN OXIDE...DIELECTROPHORETIC DEFORMATION OF ERYTHROCYTES ON TRANSPARENT INDIUM TIN OXIDE...
DIELECTROPHORETIC DEFORMATION OF ERYTHROCYTES ON TRANSPARENT INDIUM TIN OXIDE...Larry O'Connell
 
BIOE 276 Final Report
BIOE 276 Final ReportBIOE 276 Final Report
BIOE 276 Final ReportZina Kurian
 
Physics of bio electrical impedance analysis
Physics of bio electrical impedance analysisPhysics of bio electrical impedance analysis
Physics of bio electrical impedance analysisAlexander Decker
 
A low power front end analog multiplexing unit for 12 lead ecg signal acquisi...
A low power front end analog multiplexing unit for 12 lead ecg signal acquisi...A low power front end analog multiplexing unit for 12 lead ecg signal acquisi...
A low power front end analog multiplexing unit for 12 lead ecg signal acquisi...VLSICS Design
 
A1 Project Poster 2016-General Template
A1 Project Poster 2016-General TemplateA1 Project Poster 2016-General Template
A1 Project Poster 2016-General TemplateXianqi Yuan
 
MPAT (MPL 101T) - Capillary Electrophoresis.pptx
MPAT (MPL 101T) - Capillary Electrophoresis.pptxMPAT (MPL 101T) - Capillary Electrophoresis.pptx
MPAT (MPL 101T) - Capillary Electrophoresis.pptxShraddhaRaut43
 

Semelhante a Design of a micro device to electrically separate and count wbc from whole blood (20)

AUTOMATION IN HEMATOLOGY
AUTOMATION IN HEMATOLOGYAUTOMATION IN HEMATOLOGY
AUTOMATION IN HEMATOLOGY
 
automationinhematologypart1-160820102053.pptx
automationinhematologypart1-160820102053.pptxautomationinhematologypart1-160820102053.pptx
automationinhematologypart1-160820102053.pptx
 
I04845861
I04845861I04845861
I04845861
 
CellCounter_DAE_V0.0.pptx
CellCounter_DAE_V0.0.pptxCellCounter_DAE_V0.0.pptx
CellCounter_DAE_V0.0.pptx
 
A novel active charge equalisation method for series connected batteries
A novel active charge equalisation method for series connected batteriesA novel active charge equalisation method for series connected batteries
A novel active charge equalisation method for series connected batteries
 
UROP Proposal Spring 2014 (CTC)
UROP Proposal Spring 2014 (CTC)UROP Proposal Spring 2014 (CTC)
UROP Proposal Spring 2014 (CTC)
 
Ec 6001 medical electronics
Ec 6001  medical electronicsEc 6001  medical electronics
Ec 6001 medical electronics
 
Ecis Applied Biophysics
Ecis Applied BiophysicsEcis Applied Biophysics
Ecis Applied Biophysics
 
11
1111
11
 
AUTOMATION IN HAEMATOLOGY.pptx
AUTOMATION IN HAEMATOLOGY.pptxAUTOMATION IN HAEMATOLOGY.pptx
AUTOMATION IN HAEMATOLOGY.pptx
 
A simple instrumentation system for separation of whole blood components usin...
A simple instrumentation system for separation of whole blood components usin...A simple instrumentation system for separation of whole blood components usin...
A simple instrumentation system for separation of whole blood components usin...
 
Finite Element Simulation of Microfluidic Biochip for High Throughput Hydrody...
Finite Element Simulation of Microfluidic Biochip for High Throughput Hydrody...Finite Element Simulation of Microfluidic Biochip for High Throughput Hydrody...
Finite Element Simulation of Microfluidic Biochip for High Throughput Hydrody...
 
ABC of automated CBC
ABC of automated CBCABC of automated CBC
ABC of automated CBC
 
DIELECTROPHORETIC DEFORMATION OF ERYTHROCYTES ON TRANSPARENT INDIUM TIN OXIDE...
DIELECTROPHORETIC DEFORMATION OF ERYTHROCYTES ON TRANSPARENT INDIUM TIN OXIDE...DIELECTROPHORETIC DEFORMATION OF ERYTHROCYTES ON TRANSPARENT INDIUM TIN OXIDE...
DIELECTROPHORETIC DEFORMATION OF ERYTHROCYTES ON TRANSPARENT INDIUM TIN OXIDE...
 
BIOE 276 Final Report
BIOE 276 Final ReportBIOE 276 Final Report
BIOE 276 Final Report
 
Physics of bio electrical impedance analysis
Physics of bio electrical impedance analysisPhysics of bio electrical impedance analysis
Physics of bio electrical impedance analysis
 
A low power front end analog multiplexing unit for 12 lead ecg signal acquisi...
A low power front end analog multiplexing unit for 12 lead ecg signal acquisi...A low power front end analog multiplexing unit for 12 lead ecg signal acquisi...
A low power front end analog multiplexing unit for 12 lead ecg signal acquisi...
 
A1 Project Poster 2016-General Template
A1 Project Poster 2016-General TemplateA1 Project Poster 2016-General Template
A1 Project Poster 2016-General Template
 
MPAT (MPL 101T) - Capillary Electrophoresis.pptx
MPAT (MPL 101T) - Capillary Electrophoresis.pptxMPAT (MPL 101T) - Capillary Electrophoresis.pptx
MPAT (MPL 101T) - Capillary Electrophoresis.pptx
 
wbc_rbc
wbc_rbcwbc_rbc
wbc_rbc
 

Mais de IOSR Journals (20)

A011140104
A011140104A011140104
A011140104
 
M0111397100
M0111397100M0111397100
M0111397100
 
L011138596
L011138596L011138596
L011138596
 
K011138084
K011138084K011138084
K011138084
 
J011137479
J011137479J011137479
J011137479
 
I011136673
I011136673I011136673
I011136673
 
G011134454
G011134454G011134454
G011134454
 
H011135565
H011135565H011135565
H011135565
 
F011134043
F011134043F011134043
F011134043
 
E011133639
E011133639E011133639
E011133639
 
D011132635
D011132635D011132635
D011132635
 
C011131925
C011131925C011131925
C011131925
 
B011130918
B011130918B011130918
B011130918
 
A011130108
A011130108A011130108
A011130108
 
I011125160
I011125160I011125160
I011125160
 
H011124050
H011124050H011124050
H011124050
 
G011123539
G011123539G011123539
G011123539
 
F011123134
F011123134F011123134
F011123134
 
E011122530
E011122530E011122530
E011122530
 
D011121524
D011121524D011121524
D011121524
 

Último

Verification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptxVerification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptxchumtiyababu
 
Double Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torqueDouble Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torqueBhangaleSonal
 
A Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna MunicipalityA Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna MunicipalityMorshed Ahmed Rahath
 
School management system project Report.pdf
School management system project Report.pdfSchool management system project Report.pdf
School management system project Report.pdfKamal Acharya
 
A CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptx
A CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptxA CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptx
A CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptxmaisarahman1
 
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills KuwaitKuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwaitjaanualu31
 
Introduction to Serverless with AWS Lambda
Introduction to Serverless with AWS LambdaIntroduction to Serverless with AWS Lambda
Introduction to Serverless with AWS LambdaOmar Fathy
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . pptDineshKumar4165
 
kiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal loadkiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal loadhamedmustafa094
 
Wadi Rum luxhotel lodge Analysis case study.pptx
Wadi Rum luxhotel lodge Analysis case study.pptxWadi Rum luxhotel lodge Analysis case study.pptx
Wadi Rum luxhotel lodge Analysis case study.pptxNadaHaitham1
 
HOA1&2 - Module 3 - PREHISTORCI ARCHITECTURE OF KERALA.pptx
HOA1&2 - Module 3 - PREHISTORCI ARCHITECTURE OF KERALA.pptxHOA1&2 - Module 3 - PREHISTORCI ARCHITECTURE OF KERALA.pptx
HOA1&2 - Module 3 - PREHISTORCI ARCHITECTURE OF KERALA.pptxSCMS School of Architecture
 
GEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLE
GEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLEGEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLE
GEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLEselvakumar948
 
Hospital management system project report.pdf
Hospital management system project report.pdfHospital management system project report.pdf
Hospital management system project report.pdfKamal Acharya
 
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdfAldoGarca30
 
Standard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power PlayStandard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power PlayEpec Engineered Technologies
 
Thermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.pptThermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.pptDineshKumar4165
 
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...drmkjayanthikannan
 
Unleashing the Power of the SORA AI lastest leap
Unleashing the Power of the SORA AI lastest leapUnleashing the Power of the SORA AI lastest leap
Unleashing the Power of the SORA AI lastest leapRishantSharmaFr
 

Último (20)

FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced LoadsFEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
FEA Based Level 3 Assessment of Deformed Tanks with Fluid Induced Loads
 
Verification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptxVerification of thevenin's theorem for BEEE Lab (1).pptx
Verification of thevenin's theorem for BEEE Lab (1).pptx
 
Double Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torqueDouble Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torque
 
A Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna MunicipalityA Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna Municipality
 
School management system project Report.pdf
School management system project Report.pdfSchool management system project Report.pdf
School management system project Report.pdf
 
A CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptx
A CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptxA CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptx
A CASE STUDY ON CERAMIC INDUSTRY OF BANGLADESH.pptx
 
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills KuwaitKuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
Kuwait City MTP kit ((+919101817206)) Buy Abortion Pills Kuwait
 
Introduction to Serverless with AWS Lambda
Introduction to Serverless with AWS LambdaIntroduction to Serverless with AWS Lambda
Introduction to Serverless with AWS Lambda
 
Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7
Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7
Call Girls in South Ex (delhi) call me [🔝9953056974🔝] escort service 24X7
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . ppt
 
kiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal loadkiln thermal load.pptx kiln tgermal load
kiln thermal load.pptx kiln tgermal load
 
Wadi Rum luxhotel lodge Analysis case study.pptx
Wadi Rum luxhotel lodge Analysis case study.pptxWadi Rum luxhotel lodge Analysis case study.pptx
Wadi Rum luxhotel lodge Analysis case study.pptx
 
HOA1&2 - Module 3 - PREHISTORCI ARCHITECTURE OF KERALA.pptx
HOA1&2 - Module 3 - PREHISTORCI ARCHITECTURE OF KERALA.pptxHOA1&2 - Module 3 - PREHISTORCI ARCHITECTURE OF KERALA.pptx
HOA1&2 - Module 3 - PREHISTORCI ARCHITECTURE OF KERALA.pptx
 
GEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLE
GEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLEGEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLE
GEAR TRAIN- BASIC CONCEPTS AND WORKING PRINCIPLE
 
Hospital management system project report.pdf
Hospital management system project report.pdfHospital management system project report.pdf
Hospital management system project report.pdf
 
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
1_Introduction + EAM Vocabulary + how to navigate in EAM.pdf
 
Standard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power PlayStandard vs Custom Battery Packs - Decoding the Power Play
Standard vs Custom Battery Packs - Decoding the Power Play
 
Thermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.pptThermal Engineering -unit - III & IV.ppt
Thermal Engineering -unit - III & IV.ppt
 
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
Unit 4_Part 1 CSE2001 Exception Handling and Function Template and Class Temp...
 
Unleashing the Power of the SORA AI lastest leap
Unleashing the Power of the SORA AI lastest leapUnleashing the Power of the SORA AI lastest leap
Unleashing the Power of the SORA AI lastest leap
 

Design of a micro device to electrically separate and count wbc from whole blood

  • 1. IOSR Journal of Electrical and Electronics Engineering (IOSR-JEEE) e-ISSN: 2278-1676 Volume 5, Issue 2 (Mar. - Apr. 2013), PP 12-20 www.iosrjournals.org www.iosrjournals.org 12 | Page Design of a micro device to electrically separate and count wbc from whole blood Soumya Gangopadhyay1 ,AVimala Juliet2 1,2 (Department of Instrumentation and ControlEngineering,SRM University, India) Abstract:This work presents a micro device which separates and counts Leukocytes (white blood cell or WBC) from other two type of blood cells, Erythrocytes (red blood cell or RBC), and Thrombocytes (Platelet). The operation is performed using four basic parts, a micro mixer which mixes whole blood and diluent to dilute the blood, a size dependent electrical cell separator which separates WBC from other blood cells , a micro nozzle to increase the flow velocity of cells, a counter unit to count the number of WBC in sample. The size dependent cell sorting is done by electrical pillar arrays. The focusing of incoming cells to the counter is done by focusing electrodes. The main aim of this work is to count the separated white blood cells and it is done by a MEMS basedCoulter counter. Simple structure and ease of operation is the two main features of this device. The design of the device has been done by simulating different unit of the device in ComsolMultiphysics. Simulation results were satisfactory and each unit is able to perform its work as per requirement. Keywords-Coulter counter, deterministic lateral displacement, Dielectrophoresis (DEP), micro mixer and micro nozzle. I. Introduction There are efforts on cell separation and counting systems [1], [2], [3], [4]. There are different systems to separate the blood cells e.g. fluorescence spectroscopy, magnetophoretic cell separation, physical pillar based separation etc. But each of these systems have their own problems like fluorescence spectroscopy can’t count the cells separately it gives an average count of cells and most importantly it is not possible to fabricate it in a micro device. The magnetophoretic cell separator uses two permanent magnets which is not possible to place in close range of micro device. The physical pillar based cell separator can do size based cell separation and can be fabricated in a micro device, but after some use of the device the cells tend to stick to the channel floor and create clogging [1]. In this work dielectrophoresis has been used for separating the blood cells which would allow the cells neither to settle down to the channel floor nor to slow down during the process. In previous micro Coulter counter devices there was a problem, there was a chance that more than one cells could get lumped together and the counter would have been counting them as one [4]. In this work focusing electrodes has been used for focusing the cell flow so that no two cells can get lumped together and the counter would count them individually. In this work the dilution system has been implemented in the device. Hence no dilution needs to be done before injecting the blood to the device.This device is a combination of the cell separation and counting system which increases the efficiency to count the number of the most important cell in human blood which causes most of the infectious diseases. The three blood cells in human blood are respectively Erythrocyte (red blood cell or RBC), Leukocyte (white blood cell or WBC) and Thrombocyte (Platelet). TABLE 1 compares the size of the three main blood cells TABLE 1. Comparison of wbc, rbc and platelet. Blood Cells Cell Diameter (µm) WBC 7.72 – 15 RBC 6 - 8 Platelet 2.65 – 2.9 TABLE 2. Description of different wbc. Type Approx. % in adults Diameter (µm) Neutrophil 62 10 - 12 Eosinophil 2.3 10 - 12 Basophil 0.4 12 - 15 Lymphocyte 30 7 - 15 Monocyte 5.3 7.72 – 9.99
  • 2. Design of a micro device to separate and count wbc from whole blood www.iosrjournals.org 13 | Page WBC is of five main types neutrophil, eosinophil, basophil, lymphocyte and monocyte. TABLE 2 describes their size and amount of presence in blood among the total leukocytes. As the size of WBC ranges from 7.72 to 15 µm, so the cells having lesser diameter than 7.72µm is to be rejected and the greater or equal diameter cell is to be selected and forwarded for the counting purpose. II. Device Principle 2.1. Micro Mixer Unit Blood needs to get diluted before it can be used for any test because if it is not diluted, the test solution becomes over populated by cells and the count can’t be precise. Hence the blood is 1st diluted to a certain percentage to reduce the number of cells per unit sample. Another reason to dilute the blood is to prevent clotting of blood in the channel.This unit mixes the blood with PBS (Phospheted Buffer Saline) to dilute the blood. This unit has a simple structure as shown in figure.1 in which the blood and PBS is injected into the device from two different inlets and directed to a junction where the two fluid streams get mixed and then moved through a zigzag long channel to give both the fluids to get mixed well and then finally the mixture or the diluted blood is released in a specific point of the cell separator unit. In case of the fabricated device the dilution factor is to be used is 1:100 i.e. 0.1ml of whole blood is to be mixed with 9.9ml of PBS to prepare a 10ml of diluted blood. The naturel count of WBC in blood is 4500 to 10800 /µl. After dilution the count reduces to 45 to 108/µl. It reduces the time to count the cells for the device. 𝐷𝑖𝑙𝑢𝑡𝑖𝑜𝑛 𝐹𝑎𝑐𝑡𝑜𝑟 = 𝑆𝑎𝑚𝑝𝑙𝑒 𝑉𝑜𝑙𝑢𝑚𝑒 𝑇𝑜𝑡𝑎𝑙 𝑉𝑜𝑙𝑢𝑚𝑒 (1) Figure1 Micro mixer unit. 2.2. Cell Separation Unit This unit consists of a two dimensional array of circular spot electrodeswhere each of the row of the array is shifted by ΔX. Each row consists of alternate pairs of VDEP and GND electrodes. DEP force is generated by these electrodes at a particular voltage and frequency of a square wave AC signal. Figure 2 Arrangement of the spot electrodes.
  • 3. Design of a micro device to separate and count wbc from whole blood www.iosrjournals.org 14 | Page Working equation for deterministic lateral displacement is as follows: 𝑌 = 𝜋 − 𝛼 𝑟𝑒 + 𝑋−𝑟 2 sin 𝛼(2) 𝑟𝑐 ∗ = ∆𝑋− 𝑋−𝑟 2 𝜋− 𝛼 sin 𝛼(3) Where Y is the lateral displacement of the cell along the flow path. If Y>ΔX then the cell would follow thezigzag path between the electrodes and if Y<ΔX then the cell would follow the displacement mode. Now re is the diameter of the cell which decides the value of Y i.e. cells with smaller diameter will flow in zigzag mode and cells with larger diameter will flow in deterministic lateral displacement mode.rcdepends on the applied voltage on the electrodes.rc * is the critical diameter of the cell i.e. if the celldiameteris less than rc * then the cell will flow in zigzag mode andifitisgreater than rc * then it would flow in deterministic lateral displacement mode. This kind of displacement is called deterministic displacement because the lateral displacement Y of the cell can be determined by the formula equation (2)[2]. The working equation for DEP force is: 𝐹𝐷𝐸𝑃 = 2𝜋𝑅3 𝜀 𝑚 𝑅𝐸 𝐹𝐶𝑀 𝛻𝐸 𝑅𝑀𝑆 2 (4) 𝐹𝐶𝑀 = 𝜀 𝑝 ∗ − 𝜀 𝑚 ∗ 𝜀 𝑝 ∗ − 2𝜀 𝑚 ∗ (5) 𝜀 𝑚 ∗ = 𝜀 𝑚 − j𝜍 𝑚 𝜔 (6) 𝜀 𝑝 ∗ = 𝜀 𝑝 − j𝜍 𝑝 𝜔 (7) Where, ∇ERMS is the non-uniformity factor of the applied electric field i.e. the rms value of the non-uniform electric gradient between the electrodes, FCM is the Clausius-Mosotti factor which depends on the complex permittivity of the medium and the cell which are respectively εm * and εp * . In this case complex permittivity has taken in account because the conductivity and the angular frequency of the applied electric field effects the electric permittivity of the cell and the medium as in the equation 6 and 7. Now the complex permittivity effects the FCM. In case of this device the particles should get repelled by the electric pillars hence we need to get a negative DEP force to achieve the required repelling force.To create a negative DEP force the FCM should be negative because it decides the direction of the force with respect to the electric pillars [6]. There is a phenomenon called β dispersion in which the cell accepts only the high frequency signals in range of kHz to few MHz to pass through it and disperses the high range of frequencies. So the angular frequency should be high enough, so that the cell would allow the high frequency signal to pass through it and the cells would get effected by the electric field. Figure3Nozzle structure. 2.3. Micro Nozzle Unit This unit is a micro fabricated nozzle. Which increases the flow rate of the fluid cell mixture. In case of a nozzle the flow resistance ζn is responsible for the increase in flow rate. ζn can be calculated by the equation 8 (as the Reynolds’s number is very low ): 𝜁𝑛 = 𝐴 𝑛 𝑅 𝑒 (8) 𝑅𝑒 = 𝑣 𝑜 𝐷 𝑜 ν (9) 𝐴 𝑛 = 19 𝑛 𝑜 0.5 𝑡𝑎𝑛 𝛼 0.75 (10) Where Re is the Reynolds’s number of the flow, υ is the kinematic viscosity and no = (D1 2 / Do 2 ).As the fluid gets resistance in its path of flow the pressure starts to drop and according to the law of conservation of energy
  • 4. Design of a micro device to separate and count wbc from whole blood www.iosrjournals.org 15 | Page the flow rate of the fluid increases to compensate the loss of energy [5]. The increase of flow rate is required to get faster response from the device. The nozzle section also contains alternate positive and ground electrodes at its two sides. These electrodes are used for focusing the flow of the cells. 2.4. Cell Counting Unit This unit uses aperture impedance method. In this method there is an aperture and there are two electrodes at both sides of that aperture. The configuration has been shown in figure 4. The two electrodes are connected through the liquid medium. An AC voltage is applied to the electrodes. Whenever there is a cell in between them there is a change of resistance in the circuit (ΔR). Figure 4 Electrode over the channel with a cell in between them in flowing medium. The change in resistance is represented by: Δ𝑅 = 𝜌 𝐴2 𝑉𝑝(11) Where A is the cross sectional area of the channel, ρ is the specific resistance of the medium, Vpis the volume of the cell. A and ρ remains constant hence the magnitude of ΔR depends on Vp[3]. III. Device Simulation Result Each unit of the device has been simulated with COMSOL Multiphysics 4.3a. 3.1. Micro Mixer Unit This unit has been simulated with laminar flow domain. Here velocity for blood and PBS has been used 1.1µm/sec and 2.1µm/sec respectively. As it results in a uniform flow rate and proper mixing of blood and PBS. The resulting velocity of the diluted blood is 10.61µm/sec. (a)
  • 5. Design of a micro device to separate and count wbc from whole blood www.iosrjournals.org 16 | Page (b) Figure 5 (a) Simulation of flow velocity in micro-mixer, (b) Simulation of change in Reynolds’s number along the channel. Figure 6 Flow velocity along the channel of micro-mixer. 3.2.Cell Separation Unit The separation unit has been simulated by using domains like electric current, particle tracing and laminar flow. In electric current domain the alternate electrodes in a row has been assigned as terminals with a voltage signal of 20V and other electrodes as ground. The analysis has been done in frequency analysis study with 20 kHz frequency square wave signal to get an AC signal at the electrodes. (a)
  • 6. Design of a micro device to separate and count wbc from whole blood www.iosrjournals.org 17 | Page (b) Figure 7 (a) Electric field along a row of the electrode arrangement. (b) Electric field along a column of the electrode arrangement. Figure 8 Simulated electrode arrangement in the device with arrows indicating the direction of the electric field. In particle tracing domain the particles has been used to represent blood cells. In laminar flow domain an inlet flow of 10.61µm/sec has been used for the diluted blood flow as in this range of velocity the cell gets the proper separation time. In case of higher velocity the cells just gets washed off from the surface. At 20 V the electric field covers the electrode and increases the diameter of the electrodes. So that is how we can change the diameter of the electrodes to get the right value of Y which would decide the mode of flow of the cells along the channel.With no electric field applied, the value of Y is equal to ΔX and rc * = 0.2µm. In simulation the channel dimensions has been taken as 500 X 500 µm (to test the electrodes functionality in case of the actual device the width would be sufficient but the length of the channel must be increased to get the proper separation). The required FCM to create the negative DEP force is -0.2353. The particle tracing domain analysis simulation results can’t be given in this paper because the animation is impossible to represent in a paper. 3.3. Micro Nozzle Unit This unit has been simulated using only the laminar flow domain. The inlet dimension of the nozzle is 60 X 40 µm and the outlet dimension is 30 x 20 µm. The inlet flow velocity of the nozzle is 0.41134 x 10-6 m/s and outlet flow velocity is 3.45 x 10-6 m/s.This arrangement shows a pressure drop of 0.0152Pa. The flow resistance co-efficient is 0.226.
  • 7. Design of a micro device to separate and count wbc from whole blood www.iosrjournals.org 18 | Page Figure 9Simulation of the cell focusing electrodes at the micro nozzle and the counter electrodes. (a) (b) Figure10 (a) Pressure drop along the channel, (b) Flow velocity magnitude along the micro nozzle. (a) (b) Figure 11 (a)Trajectory of the cell along the channel for different electric field along y-axis,(b) Trajectory of cell along the channel when electric field along y-axis is 4 x 105 V/m and electric field along z-axis has been varied. From Fig.11 (a) it is evident that when 4 x 105 V/m electric field is applied to the focusing electrodes along y- axis, cells are following a more linear path than the other applied electric fields. So 7V square wave AC voltage is applied to the focusing electrodes applying force along y-axis. Form figure 11 (b) it can be seen that when another set of focusing electrode has been used to apply an electric field along z-axis the particles becomes more non-linear than without this focusing electrode. But as the test has been done on simulated cells not on the actual biological cell. Hence it is expected that in case of the actual biological cell the focusing electrodes at z-axis would be required and an electric field of 3 x 105 V/m should be applied to the electrodes. The two sets of focusing electrodes would act as a lens to focus the cells. Hence the cell focusing electrodes are excited by square wave AC signal of 7V and 20 kHz frequency to keep the cell flow in a linear path and most preferably along the central line of the nozzle.
  • 8. Design of a micro device to separate and count wbc from whole blood www.iosrjournals.org 19 | Page 3.3. Cell CountingUnit This unit has been simulated using laminar flow, electric current and particle tracing physics. Figure 12 Cell counter circuit. The circuit used for sensing the change in resistance has been given in the Fig. 12. The cell counter unit is connected in a voltage divider circuit and the output of that voltage divider is connected with an amplifier, so that if any resistance change occurs in the counter unit the amplifier would give a pulse equivalent to the resistance change. Figure 13 Trajectory of the cell along the channel after the micro-nozzle unit for different electric filed applied to the y-axis. From Fig. 13 it can be seen that the cell motion along the channel is linear when the applied electric field is 3 x 105 V/m. But at 4 x 105 V/m as the cell comes nearer to the counter electrodes, it gets more focused at the center of the channel width as desired. Hence 4 x 105 V/m electric field at y-axis is appropriate for the counter electrodes. The inlet flow of this unit is as same as the outlet of the micro nozzle. The electrodes used for sensing the change in resistance are excited by 20 kHz square wave AC signal as with a voltage of 7V. From figure 14 it can be seen that when cell is not present in between the electrodes the voltage at the middle of the channel is much higher than the times when the cell is present. As the cell is an obstraction to the electric field, the voltage when cell is present ranges from 0 to 0.5 V. This change in voltage can be detected by the circuit and the out put can be given accordingly. Every time a cell passes through the gapbetween the electrodes this changeoccures and a pulse is generated. Thus it can detect the cells and the number of pulses give the count of wbc.
  • 9. Design of a micro device to separate and count wbc from whole blood www.iosrjournals.org 20 | Page Figure 14 Change in voltage in between the electrodes due to prescence of cell of different sizes. IV. Conclusion The MEMS based device has been simulated using COMSOL Multiphysics 4.3a. The results were satisfactory. To get proper results physical properties of human blood and PBS has been used for the simulated fluids. The cell separation unit has been simulated with the proper voltage excitation. The micro nozzle part is increasing the flow velocity as desirable and its cell focusing electrodes are allowing the flow of cell to be linearized. But due to lack of availability of fabrication facility, the device is still in design phase. Hence the results have been derived from the simulation rather than from the fabricated device. The results presented in the paper are only simulation results and practical results are unavailable. The circuit to count the Leukocytes has been designed and would be implemented practically and calibrated only after the device gets fabricated. Also the length of the separation channel can be decided only after fabrication and testing of the fabricated device. It is expected that the fabrication facility would be arranged in future and this device would get fabricated as soon as possible. It is believed that this device would help doctors as well as common people to diagnose their Leukocyte related diseases quicker and with more ease. References [1] SiyangZhmg, Raylene Yung, Yu-Chong Tail, Harvey Kasdan“Determinstic lateral displacement mems device for continuous blood cellseparation”, Caltech Micromachining Lab, California Institute of Technology, USA,StanfordUniversity, USAInsDiagnostics, International Remote Imaging Systems, Inc., USA, 8732-5/05 IEEE851-854 2007 [2] Young-Ho Cho, “Bio-Inspired MEMS Devices forElectrical Cell Separation and Mechanical Cell Characterization”, Digital Nanolocomotion Center, Korea Advanced Institute of Science and Technology (KAIST) 373-1 Guseong-dong, Yuseong-gu, Daejeon 305-701 Republic ofKorea 978-1-4244-1858-9/07 IEEE 514 - 5182007 [3] Daisuke Satakea,, Hiroyuki Ebia, NarihiroOkub, KoichiroMatsudaa, HidekuniTakaoc, MitsuakiAshikic, Makoto Ishidac, “A sensor for blood cell counter using MEMS technology”, Fundamental Technology ResearchDepartment, HORIBA Ltd., 2 Miyanohigashi, Kisshoin,Minami-ku, Kyoto 601-8510, Japan,Medical Electronic Systems, HORIBA Ltd., 2 Miyanohigashi, Kisshoin, Minami-ku, Kyoto 601-8510, Japan,Department of Electrical and Electronic Engineering, Toyohashi University of Technology, 1-1, Hibarigaoka, Tempaku-cho,Toyohashi 441-8580, Japan.Sensors and Actuators B 83 77–81 2002 [4] M. Korampally, J. D. Benson, Y. Wu, 1 J. K. Critser, M. Almasri, “MEMS based Coulter counter for cell sizing”,Department of Electricaland Computer Engineering, and Department of Veterinary,Pathobiology, University of Missouri, Columbia, Missouri, 65201, USA. Proc. of SPIE Vol. 6886, 68860A, 0277-786X/08/$18 · doi: 10.1117/12.763741 2008 1-7 [5] X.N.Jiang ,Y.Zhou, Y.Li, Y.Yang, X.Y.Huang , and C.Y.Liu. “Experiments and analysis for micronozzle/diffuser flow and microvalveless pumps”, School of MPE, NanyangTechnological University Singapore639798,Dept. of Precision Instruments and Mechanology, TsinghuaUniversity, Beijing, 100084,China.0-7803-3829-4/97 1997 IEEE 369 - 372 [6] S. Burgarella, M. Bianchessiand M. De Fazio ,“ Numerical modeling of dielectrophoretic forces acting uponbiological cells in silicon lab-on-chip devices ” , STMicroelectronics, Advanced System Technology, R&I e-Health ,Corresponding author: STMicroelectronics, Centro Dir. Colleoni, Pal. Dialettica, Via Cardano 1, 20041AgrateBrianza (MI), Italy.Excerpt from the Proceedings of the COMSOL Users Conference Grenoble 2007