White Paper | Extremity Cone Beam (CBCT) Imaging
The Value of Weight-Bearing “Functional” CT Scans
Author: John Marzo, M.D.
In clinical orthopedics, advanced imaging like computed axial
tomography (CT) scanning, has become invaluable to the
evaluation and management of patients with musculoskeletal
disease. Bone detail is much better visualized with 2D and 3D
CT renderings of patients with problems like glenoid fracture,
failed shoulder instability surgery, and meniscal root avulsions
like shown in Figures 1, 2, and 3 respectively.
Figure 1
Figure 2
Figure 3
As illustrated, these high-quality images provide multiplanar
two- and three-dimensional visualization for practitioners who
think and work in three dimensions. These images make it
easier for orthopedic physicians to educate their patients,
teach and train medical students, residents, and post-graduate
fellows. A significant limitation of current CT technology is that
current technique forces image acquisition with the subject in
a supine, relaxed position. Some have tried to simulate weight
bearing in a CT scanner by custom designing a rig to apply
longitudinal load through the patient for imaging of the spine
or lower extremity. These methods are at worst, a poor
depiction of functional anatomy, and at best cumbersome and
a less-than-accurate simulation of function. A new cone beam
CT scanner, the OnSight 3D Extremity System, is in
development by Carestream Health (Rochester NY) to acquire
weight-bearing images of the extremities. The prototype is
designed to offer high-quality, portable, low-dose 3D point-of-
care imaging by orthopaedic and sports medicine practices,
hospitals, imaging centers, urgent care facilities and other
healthcare providers.
Figure 4: CARESTREAM OnSight 3D Extremity System
White Paper | Extremity Cone Beam (CBCT) Imaging
2
John Marzo, M.D. and other researchers from the Jacobs
School of Medicine and Biomedical Sciences at the University
at Buffalo are currently performing institutional IRB-approved
clinical trials and basic sciences studies with the prototype
model. These studies are being carried out at the Erie County
Medical Center, Buffalo’s regional orthopedic tertiary care
facility. Based on early data, they are convinced that many
imaging studies should be acquired with subjects in positions
that represent true human function, such as weight bearing on
the lower extremities.
In order to investigate the imaging performance of the Cone
Beam Computed Tomography (CBCT) scanner, the 2D imaging
performance of the CARESTREAM OnSight 3D Extremity
System was compared to the predicate CARESTREAM DRX-1
Detector used with the CARESTREAM DRX-Evolution System.
The 3D volumetric imaging performance of the OnSight system
was compared to a multiple detector computed tomography
(MDCT) scanner ("predicate device"), the Philips Brilliance 64
CT Scanner. The purpose of the study was to demonstrate
equivalent diagnostic image quality between the
investigational and predicate devices, using a RadLex subjective
quality rating scale. The evaluation was performed on equal
numbers of knees, ankles, feet, elbows, and hands from
thirty-three cadaveric human specimens, and from thirteen
living human subjects. Four independent, board-certified
radiologists of varying general reading experience performed
an evaluation of the images/exams captured using both the
investigational and predicate devices. In summary, the
investigational CBCT system produced 2D images with
equivalent diagnostic image quality to the predicate system for
a range of exams, and 3D images were rated equal or better
when compared to the predicate device for a range of exams
on cadaveric specimens and human subjects. Over 80% of all
the 2D images were rated diagnostic or exemplary, whereas
approximately 98% of all 3D images were rated diagnostic or
exemplary. More than 75% of all RadLex rating responses
counted for all 2D images were rated equivalent or favored the
investigational device. Approximately 85% of the RadLex
rating responses counted for the 3D images were rated
equivalent or favored the investigational device. Examples of
representative scans are seen below.
Figure 5: 2D and 3D renderings generated by the CBCT scanner
White Paper | Extremity Cone Beam (CBCT) Imaging
3
Conventional CT measures of patellofemoral alignment include
the congruence angle, patellar tilt angle, and tibial tubercle-
trochlear groove offset distance (TT-TG offset). There are
clearly defined limits of normal for each of these measures,
and they are used by surgeons planning corrective operations
on the patellofemoral joint. When obtained with conventional
CT scan technique, the patient is lying supine on the
examination table with the knee fully extended. The degree of
knee flexion and activity of the quadriceps are known to
influence patellar tracking on the trochlea, but these factors
are removed when images are taken with the patient supine.
The new investigational cone beam CT scanner has been
designed to allow for acquisition of images while the patient is
standing, weight bearing, and in various positions of knee joint
flexion. A clinical study was designed to compare several
commonly used measures of patellofemoral alignment (tilt
angle, congruence angle, and TT-TG offset) on a conventional
CT scanner to the same measures on the prototype extremity
cone beam CT scanner. Based on previous data, the primary
hypothesis was that measures of patellofemoral alignment
obtained from the prototype cone beam CT scanner will be
less than those obtained by the conventional CT scanner. In a
prospective case series, twenty patients with known patella
instability were imaged on both a conventional CT scanner and
on the prototype cone beam CT scanner. Objective
assessments of patellofemoral alignment (tilt angle,
congruence angle, TT-TG offset) were measured on images
obtained from the conventional CT scans and from the
prototype device by two independent reviewers. For both
raters, tilt angle, congruence angle, and TT-TG offset were
statistically significantly decreased on cone beam versus
conventional CT scan.
Table 1: Patella measurements for conventional versus cone beam CT scans by rater
Rater 1 Rater 2 Average for Both Raters
Tilt Angle 28.0 ± 7.3 18.1 ±
12.3
<0.0001 28.2 ± 7.3 18.2 ±
11.3
<0.0001 28.1 ± 7.1 18.2 ±
11.6
<0.0001
Congru-
ence
Angle
22.8 ±
17.3
0.15 ±
31.1
0.001 30.7 ±
20.1
5.8 ± 30.8 <0.0001 26.7 ±
18.1
3.0 ± 30.1 0.0002
TT-TG
Offset
21.4 ± 4.2 12.8 ± 6.3 <0.0001 18.9 ± 4.3 11.8 ± 7.6 0.001 20.1 ± 4.2 12.3 ± 6.3 <0.0001
NOTES:
• All values are mean ± SD.
• CT: computed tomography
• TT-TG: tibial tuberosity trochlear groove
White Paper | Extremity Cone Beam (CBCT) Imaging
carestream.com
©Carestream Health, Inc., 2016. CARESTREAM is a trademark of
Carestream Health. CAT 2000 165 08/16
Our conclusion from this study is that for cases of patella
instability, it may be desirable to obtain images while the
patient is weight bearing on a flexed knee with their
quadriceps muscles active. Improvement in objective measures
of patella alignment should lead to improved clinical and
surgical care of patients with this condition.
A second clinical study is currently underway to take
advantage of the unit’s ability to obtain images in weight
bearing, and will compare measures of ankle stability from the
investigational weight-bearing cone beam computed
tomography scanner to the same measures on gravity stress
X-ray in patients who have supination-external rotation ankle
fractures. The primary hypothesis is that the mean medial clear
space will be different for the investigational weight-bearing
cone beam CT scan versus the gravity stress X-ray. Supination-
external rotation (SER) or Weber B type ankle fractures are
considered the most common type of ankle fracture. An SER
fracture is considered unstable when it is associated with
deltoid ligament rupture. Clinical findings of medial
tenderness, swelling, and ecchymosis were once thought to be
reliable for correctly determining the stability of SER ankle
fractures, but have since proven to be poor predictors of
deltoid ligament disruption. Rupture of the deltoid ligament
allows lateral talar shift, and can be identified by widening of
the medial clear space on standard radiographs as well as
stress radiographs. Many studies have shown that gravity stress
radiographs are necessary and are currently considered to be
the gold standard for diagnosis of SER fractures. Of all the
methods that have been used to diagnose instability and
investigate deltoid ligament integrity---clinical examination,
weight-bearing radiographs, stress radiography, magnetic
resonance imaging (MRI), arthroscopy, ultrasound---none have
been shown to be cost-effective, rapid, reliable, and easy to
use. Gravity stress radiographs are performed without a
uniform method, are known to lack sensitivity, and are very
user dependent. Therefore, more research is needed to
determine which method is most accurate and efficient for SER
fracture diagnosis. Weight bearing on the extremity is thought
to be more representative of the normal use of the ankle joint,
and CT scan is known to deliver excellent bone detail for
interpretation of images. The ability of weight-bearing cone
beam CT scanning to measure medial clear space may present
a new opportunity to evaluate instability of SER ankle
fractures.
These and future studies may validate the value of the OnSight
3D Extremity System. Potential benefits include better quality
images with a lower radiation dose than conventional
computed tomography. The unit is proposed for use in an
orthopedic office, but may have application to the operating
room or at athletic competition sites. The unit is less expensive
than a traditional in-hospital or radiology center CT scanner,
and can be used with existing electrical systems (220V). Most
important however, is the potential to acquire images while
bearing weight and in more functionally relevant positions.
Dr. John Marzo is a physician with UBMD Orthopaedics & Sports Medicine, Associate Professor of Clinical Orthopaedics, Jacobs
School of Medicine and Biomedical Sciences, University at Buffalo and former Medical Director, Buffalo Bills.

Mais conteúdo relacionado

Destaque

Bite-wing and technique errors lecture1
Bite-wing and   technique errors lecture1Bite-wing and   technique errors lecture1
Bite-wing and technique errors lecture1Dr. Tshewang Gyeltshen
 
Teoria e pratica degli appoggi Baumann Alberto Forti
Teoria e pratica degli appoggi Baumann     Alberto FortiTeoria e pratica degli appoggi Baumann     Alberto Forti
Teoria e pratica degli appoggi Baumann Alberto FortiAlberto Forti
 
Digital impressions
Digital impressions Digital impressions
Digital impressions Navreet Bajwa
 
Teoria e pratica della bacchetta e del bastone jaeger Prof.Alberto Forti
Teoria e pratica della bacchetta e del bastone jaeger Prof.Alberto FortiTeoria e pratica della bacchetta e del bastone jaeger Prof.Alberto Forti
Teoria e pratica della bacchetta e del bastone jaeger Prof.Alberto FortiAlberto Forti
 
Cad Cam dentistry and digital impressions
Cad Cam dentistry and digital impressionsCad Cam dentistry and digital impressions
Cad Cam dentistry and digital impressionsVinay Kadavakolanu
 
radiology-bitewing-technique
 radiology-bitewing-technique radiology-bitewing-technique
radiology-bitewing-techniqueParth Thakkar
 
Basics of orthopedic radiology
Basics of orthopedic radiologyBasics of orthopedic radiology
Basics of orthopedic radiologyDrijaz Wazir
 

Destaque (10)

Bite-wing and technique errors lecture1
Bite-wing and   technique errors lecture1Bite-wing and   technique errors lecture1
Bite-wing and technique errors lecture1
 
Teoria e pratica degli appoggi Baumann Alberto Forti
Teoria e pratica degli appoggi Baumann     Alberto FortiTeoria e pratica degli appoggi Baumann     Alberto Forti
Teoria e pratica degli appoggi Baumann Alberto Forti
 
Optical impressions/ dentistry jobs
Optical impressions/ dentistry jobsOptical impressions/ dentistry jobs
Optical impressions/ dentistry jobs
 
Digital impressions
Digital impressions Digital impressions
Digital impressions
 
Cad cam cbct
Cad cam cbctCad cam cbct
Cad cam cbct
 
Teoria e pratica della bacchetta e del bastone jaeger Prof.Alberto Forti
Teoria e pratica della bacchetta e del bastone jaeger Prof.Alberto FortiTeoria e pratica della bacchetta e del bastone jaeger Prof.Alberto Forti
Teoria e pratica della bacchetta e del bastone jaeger Prof.Alberto Forti
 
Cad Cam dentistry and digital impressions
Cad Cam dentistry and digital impressionsCad Cam dentistry and digital impressions
Cad Cam dentistry and digital impressions
 
radiology-bitewing-technique
 radiology-bitewing-technique radiology-bitewing-technique
radiology-bitewing-technique
 
Basics of orthopedic radiology
Basics of orthopedic radiologyBasics of orthopedic radiology
Basics of orthopedic radiology
 
Trauma Radiography
Trauma RadiographyTrauma Radiography
Trauma Radiography
 

Mais de Carestream

Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Carestream
 
Digital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationDigital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationCarestream
 
Tube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareTube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareCarestream
 
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...Carestream
 
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsDose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsCarestream
 
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographySmart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographyCarestream
 
Special Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-raySpecial Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-rayCarestream
 
Special Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentSpecial Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentCarestream
 
The Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityThe Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityCarestream
 
Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Carestream
 
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Carestream
 
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesUsing Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesCarestream
 
Guia obsolescencia seram
Guia obsolescencia seramGuia obsolescencia seram
Guia obsolescencia seramCarestream
 
Cloud Security
Cloud Security Cloud Security
Cloud Security Carestream
 
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Carestream
 
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareWhite Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareCarestream
 
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundStudy: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundCarestream
 
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Carestream
 
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Carestream
 
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Carestream
 

Mais de Carestream (20)

Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
Affordable Digital Upgrade for Medical Imaging - the Benefits and the Return ...
 
Digital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of OperationDigital Tomosynthesis: Theory of Operation
Digital Tomosynthesis: Theory of Operation
 
Tube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization SoftwareTube and Line and Pneumothorax Visualization Software
Tube and Line and Pneumothorax Visualization Software
 
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
EVP Plus Software delivers state-of-the-art image processing for CR and DR sy...
 
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of OperationsDose Efficient Dual Energy Subtraction Radiography - Theory of Operations
Dose Efficient Dual Energy Subtraction Radiography - Theory of Operations
 
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital RadiographySmart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
Smart Noise Cancellation Processing: New Level of Clarity in Digital Radiography
 
Special Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-raySpecial Report: Challenges and Solutions in Pediatric X-ray
Special Report: Challenges and Solutions in Pediatric X-ray
 
Special Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray EquipmentSpecial Report: Getting the Optimal Return on X-ray Equipment
Special Report: Getting the Optimal Return on X-ray Equipment
 
The Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image QualityThe Pursuit of Excellence in Image Quality
The Pursuit of Excellence in Image Quality
 
Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions Whitepaper: Healthcare Data Migration - Top 10 Questions
Whitepaper: Healthcare Data Migration - Top 10 Questions
 
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
Whitepaper: Image Quality Impact of SmartGrid Processing in Bedside Chest Ima...
 
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray TubesUsing Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
Using Carbon Nano-Tube Field Emitters to Miniaturize X-Ray Tubes
 
Guia obsolescencia seram
Guia obsolescencia seramGuia obsolescencia seram
Guia obsolescencia seram
 
Cloud Security
Cloud Security Cloud Security
Cloud Security
 
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
Sunway Medical Centre Installs CARESTREAM Vue PACS to Streamline Imaging Proc...
 
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac CareWhite Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
White Paper: The Benefits of Mobile X-rays in Thoracic and Cardiac Care
 
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in UltrasoundStudy: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
Study: University Clinic, Regensburg, Evaluates Smart Flow in Ultrasound
 
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...Financial Implications for Integrating Carestream OnSight 3D Extremity System...
Financial Implications for Integrating Carestream OnSight 3D Extremity System...
 
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
Whitepaper: Enterprise Access Viewer Connecting technologies, physicians, and...
 
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
Evaluating Enterprise Clinical Data-Management Systems at RSNA 2016
 

Último

ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetneemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Me
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near MeRussian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Me
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Memriyagarg453
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetSambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Memriyagarg453
 
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅gragmanisha42
 
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErnakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh
 
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetJalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.ktanvi103
 
Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510Vipesco
 
Call Girls in Udaipur Girija Udaipur Call Girl ✔ VQRWTO ❤️ 100% offer with...
Call Girls in Udaipur  Girija  Udaipur Call Girl  ✔ VQRWTO ❤️ 100% offer with...Call Girls in Udaipur  Girija  Udaipur Call Girl  ✔ VQRWTO ❤️ 100% offer with...
Call Girls in Udaipur Girija Udaipur Call Girl ✔ VQRWTO ❤️ 100% offer with...mahaiklolahd
 
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetOzhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetbhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetCall Girls Service
 
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girlKolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girlonly4webmaster01
 
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅gragmanisha42
 
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...mahaiklolahd
 
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...chandigarhentertainm
 
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur RajasthanJaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthanindiancallgirl4rent
 

Último (20)

ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
ooty Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetneemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
neemuch Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetraisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
raisen Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Me
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near MeRussian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Me
Russian Call Girls in Noida Pallavi 9711199171 High Class Call Girl Near Me
 
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetBareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Bareilly Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetSambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Sambalpur Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near MeVIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
VIP Call Girls Noida Sia 9711199171 High Class Call Girl Near Me
 
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅
Russian Call Girls Kota * 8250192130 Service starts from just ₹9999 ✅
 
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetErnakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ernakulam Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetJalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Jalna Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
 
Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510Krishnagiri call girls Tamil aunty 7877702510
Krishnagiri call girls Tamil aunty 7877702510
 
Call Girls in Udaipur Girija Udaipur Call Girl ✔ VQRWTO ❤️ 100% offer with...
Call Girls in Udaipur  Girija  Udaipur Call Girl  ✔ VQRWTO ❤️ 100% offer with...Call Girls in Udaipur  Girija  Udaipur Call Girl  ✔ VQRWTO ❤️ 100% offer with...
Call Girls in Udaipur Girija Udaipur Call Girl ✔ VQRWTO ❤️ 100% offer with...
 
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real MeetOzhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
Ozhukarai Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meetbhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
bhubaneswar Call Girls 👙 6297143586 👙 Genuine WhatsApp Number for Real Meet
 
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girlKolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
Kolkata Call Girls Miss Inaaya ❤️ at @30% discount Everyday Call girl
 
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
Call Girl Gorakhpur * 8250192130 Service starts from just ₹9999 ✅
 
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
Call Girl in Bangalore 9632137771 {LowPrice} ❤️ (Navya) Bangalore Call Girls ...
 
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...
❤️Call girls in Jalandhar ☎️9876848877☎️ Call Girl service in Jalandhar☎️ Jal...
 
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur RajasthanJaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
Jaipur Call Girls 9257276172 Call Girl in Jaipur Rajasthan
 

The Value of Weight-Bearing "Functional" CT Scans in medical imaging

  • 1. White Paper | Extremity Cone Beam (CBCT) Imaging The Value of Weight-Bearing “Functional” CT Scans Author: John Marzo, M.D. In clinical orthopedics, advanced imaging like computed axial tomography (CT) scanning, has become invaluable to the evaluation and management of patients with musculoskeletal disease. Bone detail is much better visualized with 2D and 3D CT renderings of patients with problems like glenoid fracture, failed shoulder instability surgery, and meniscal root avulsions like shown in Figures 1, 2, and 3 respectively. Figure 1 Figure 2 Figure 3 As illustrated, these high-quality images provide multiplanar two- and three-dimensional visualization for practitioners who think and work in three dimensions. These images make it easier for orthopedic physicians to educate their patients, teach and train medical students, residents, and post-graduate fellows. A significant limitation of current CT technology is that current technique forces image acquisition with the subject in a supine, relaxed position. Some have tried to simulate weight bearing in a CT scanner by custom designing a rig to apply longitudinal load through the patient for imaging of the spine or lower extremity. These methods are at worst, a poor depiction of functional anatomy, and at best cumbersome and a less-than-accurate simulation of function. A new cone beam CT scanner, the OnSight 3D Extremity System, is in development by Carestream Health (Rochester NY) to acquire weight-bearing images of the extremities. The prototype is designed to offer high-quality, portable, low-dose 3D point-of- care imaging by orthopaedic and sports medicine practices, hospitals, imaging centers, urgent care facilities and other healthcare providers. Figure 4: CARESTREAM OnSight 3D Extremity System
  • 2. White Paper | Extremity Cone Beam (CBCT) Imaging 2 John Marzo, M.D. and other researchers from the Jacobs School of Medicine and Biomedical Sciences at the University at Buffalo are currently performing institutional IRB-approved clinical trials and basic sciences studies with the prototype model. These studies are being carried out at the Erie County Medical Center, Buffalo’s regional orthopedic tertiary care facility. Based on early data, they are convinced that many imaging studies should be acquired with subjects in positions that represent true human function, such as weight bearing on the lower extremities. In order to investigate the imaging performance of the Cone Beam Computed Tomography (CBCT) scanner, the 2D imaging performance of the CARESTREAM OnSight 3D Extremity System was compared to the predicate CARESTREAM DRX-1 Detector used with the CARESTREAM DRX-Evolution System. The 3D volumetric imaging performance of the OnSight system was compared to a multiple detector computed tomography (MDCT) scanner ("predicate device"), the Philips Brilliance 64 CT Scanner. The purpose of the study was to demonstrate equivalent diagnostic image quality between the investigational and predicate devices, using a RadLex subjective quality rating scale. The evaluation was performed on equal numbers of knees, ankles, feet, elbows, and hands from thirty-three cadaveric human specimens, and from thirteen living human subjects. Four independent, board-certified radiologists of varying general reading experience performed an evaluation of the images/exams captured using both the investigational and predicate devices. In summary, the investigational CBCT system produced 2D images with equivalent diagnostic image quality to the predicate system for a range of exams, and 3D images were rated equal or better when compared to the predicate device for a range of exams on cadaveric specimens and human subjects. Over 80% of all the 2D images were rated diagnostic or exemplary, whereas approximately 98% of all 3D images were rated diagnostic or exemplary. More than 75% of all RadLex rating responses counted for all 2D images were rated equivalent or favored the investigational device. Approximately 85% of the RadLex rating responses counted for the 3D images were rated equivalent or favored the investigational device. Examples of representative scans are seen below. Figure 5: 2D and 3D renderings generated by the CBCT scanner
  • 3. White Paper | Extremity Cone Beam (CBCT) Imaging 3 Conventional CT measures of patellofemoral alignment include the congruence angle, patellar tilt angle, and tibial tubercle- trochlear groove offset distance (TT-TG offset). There are clearly defined limits of normal for each of these measures, and they are used by surgeons planning corrective operations on the patellofemoral joint. When obtained with conventional CT scan technique, the patient is lying supine on the examination table with the knee fully extended. The degree of knee flexion and activity of the quadriceps are known to influence patellar tracking on the trochlea, but these factors are removed when images are taken with the patient supine. The new investigational cone beam CT scanner has been designed to allow for acquisition of images while the patient is standing, weight bearing, and in various positions of knee joint flexion. A clinical study was designed to compare several commonly used measures of patellofemoral alignment (tilt angle, congruence angle, and TT-TG offset) on a conventional CT scanner to the same measures on the prototype extremity cone beam CT scanner. Based on previous data, the primary hypothesis was that measures of patellofemoral alignment obtained from the prototype cone beam CT scanner will be less than those obtained by the conventional CT scanner. In a prospective case series, twenty patients with known patella instability were imaged on both a conventional CT scanner and on the prototype cone beam CT scanner. Objective assessments of patellofemoral alignment (tilt angle, congruence angle, TT-TG offset) were measured on images obtained from the conventional CT scans and from the prototype device by two independent reviewers. For both raters, tilt angle, congruence angle, and TT-TG offset were statistically significantly decreased on cone beam versus conventional CT scan. Table 1: Patella measurements for conventional versus cone beam CT scans by rater Rater 1 Rater 2 Average for Both Raters Tilt Angle 28.0 ± 7.3 18.1 ± 12.3 <0.0001 28.2 ± 7.3 18.2 ± 11.3 <0.0001 28.1 ± 7.1 18.2 ± 11.6 <0.0001 Congru- ence Angle 22.8 ± 17.3 0.15 ± 31.1 0.001 30.7 ± 20.1 5.8 ± 30.8 <0.0001 26.7 ± 18.1 3.0 ± 30.1 0.0002 TT-TG Offset 21.4 ± 4.2 12.8 ± 6.3 <0.0001 18.9 ± 4.3 11.8 ± 7.6 0.001 20.1 ± 4.2 12.3 ± 6.3 <0.0001 NOTES: • All values are mean ± SD. • CT: computed tomography • TT-TG: tibial tuberosity trochlear groove
  • 4. White Paper | Extremity Cone Beam (CBCT) Imaging carestream.com ©Carestream Health, Inc., 2016. CARESTREAM is a trademark of Carestream Health. CAT 2000 165 08/16 Our conclusion from this study is that for cases of patella instability, it may be desirable to obtain images while the patient is weight bearing on a flexed knee with their quadriceps muscles active. Improvement in objective measures of patella alignment should lead to improved clinical and surgical care of patients with this condition. A second clinical study is currently underway to take advantage of the unit’s ability to obtain images in weight bearing, and will compare measures of ankle stability from the investigational weight-bearing cone beam computed tomography scanner to the same measures on gravity stress X-ray in patients who have supination-external rotation ankle fractures. The primary hypothesis is that the mean medial clear space will be different for the investigational weight-bearing cone beam CT scan versus the gravity stress X-ray. Supination- external rotation (SER) or Weber B type ankle fractures are considered the most common type of ankle fracture. An SER fracture is considered unstable when it is associated with deltoid ligament rupture. Clinical findings of medial tenderness, swelling, and ecchymosis were once thought to be reliable for correctly determining the stability of SER ankle fractures, but have since proven to be poor predictors of deltoid ligament disruption. Rupture of the deltoid ligament allows lateral talar shift, and can be identified by widening of the medial clear space on standard radiographs as well as stress radiographs. Many studies have shown that gravity stress radiographs are necessary and are currently considered to be the gold standard for diagnosis of SER fractures. Of all the methods that have been used to diagnose instability and investigate deltoid ligament integrity---clinical examination, weight-bearing radiographs, stress radiography, magnetic resonance imaging (MRI), arthroscopy, ultrasound---none have been shown to be cost-effective, rapid, reliable, and easy to use. Gravity stress radiographs are performed without a uniform method, are known to lack sensitivity, and are very user dependent. Therefore, more research is needed to determine which method is most accurate and efficient for SER fracture diagnosis. Weight bearing on the extremity is thought to be more representative of the normal use of the ankle joint, and CT scan is known to deliver excellent bone detail for interpretation of images. The ability of weight-bearing cone beam CT scanning to measure medial clear space may present a new opportunity to evaluate instability of SER ankle fractures. These and future studies may validate the value of the OnSight 3D Extremity System. Potential benefits include better quality images with a lower radiation dose than conventional computed tomography. The unit is proposed for use in an orthopedic office, but may have application to the operating room or at athletic competition sites. The unit is less expensive than a traditional in-hospital or radiology center CT scanner, and can be used with existing electrical systems (220V). Most important however, is the potential to acquire images while bearing weight and in more functionally relevant positions. Dr. John Marzo is a physician with UBMD Orthopaedics & Sports Medicine, Associate Professor of Clinical Orthopaedics, Jacobs School of Medicine and Biomedical Sciences, University at Buffalo and former Medical Director, Buffalo Bills.