So, Here it's presentation on Qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2. Qualification of analytical instruments. it also includes design qualifications (DQ), Installation qualification(IQ)
Operational qualification(OQ)
Performance qualification(PQ).
The document discusses the qualification of an FTIR analytical instrument. It describes various parameters used to ensure instrument quality such as validation, calibration, and maintenance. Qualification involves collecting evidence that the instrument is suitable for its intended purpose and includes design qualification, installation qualification, operational qualification, and performance qualification phases. The document then discusses specific tests and acceptance criteria for qualifying an FTIR, including tests for resolution, wave number accuracy and reproducibility, transmittance reproducibility, and other tests to evaluate the instrument over time according to ASTM standards.
The document discusses the qualification of high performance thin layer chromatography (HPTLC). It describes the four types of qualification: design qualification, installation qualification, operation qualification, and performance qualification. Design qualification verifies specifications and review methods. Installation qualification documents compliance at installation. Operation qualification documents consistent performance within operating ranges. Performance qualification ascertains the instrument is suitable for specific analytical tasks. The document then provides examples of tests to check HPTLC performance, including linearity of spotting, reproducibility of spotting, and detection capacity.
Autoclave
Principle of Autoclave
Construction of Autoclave
Working of Autoclave
Qualification of Autoclave
Installation Qualification
Operational Qualification
Performance Qualification
References
This document provides guidelines for developing specifications for new drug substances and products according to ICH Q6A. It discusses universal and specific tests/criteria that should be included for drug substances and products, such as identification, description, assay, impurities, dissolution, disintegration, content uniformity, and microbial limits. The document gives acceptance criteria and justification for key tests like dissolution, discussing how to set Q values and limits based on biobatch results and BCS classification. It also provides guidance on other tests for oral liquids, parenterals and solid dosage forms.
The document provides details on qualification and validation procedures for gas chromatography (GC) and Fourier transform infrared spectroscopy (FTIR) systems used in pharmaceutical analysis. It describes calibration tests for both instruments, including tests to check wave number precision and reproducibility, linearity, and temperature accuracy and stability. Validation tests are also outlined, such as resolution checks and tests to evaluate performance against acceptance criteria over time as specified in guidance documents. The document provides a thorough overview of qualification processes for critical analytical instruments.
This document discusses various quality control and documentation procedures in the pharmaceutical industry. It includes 3 key points:
1. It discusses the importance of documentation in defining specifications, methods, providing an audit trail and ensuring authorized personnel have necessary information. This includes documents like specifications, batch production records, SOPs etc.
2. It describes procedures for developing key documents like master formulas, batch manufacturing records, audit plans and reports. This ensures uniform processes and allows tracing of batch history.
3. It discusses quality audits which systematically examine if quality activities comply with arrangements and objectives. This includes internal audits as well as those imposed by regulators or customers.
The document discusses the qualification of an FTIR analytical instrument. It describes various parameters used to ensure instrument quality such as validation, calibration, and maintenance. Qualification involves collecting evidence that the instrument is suitable for its intended purpose and includes design qualification, installation qualification, operational qualification, and performance qualification phases. The document then discusses specific tests and acceptance criteria for qualifying an FTIR, including tests for resolution, wave number accuracy and reproducibility, transmittance reproducibility, and other tests to evaluate the instrument over time according to ASTM standards.
The document discusses the qualification of high performance thin layer chromatography (HPTLC). It describes the four types of qualification: design qualification, installation qualification, operation qualification, and performance qualification. Design qualification verifies specifications and review methods. Installation qualification documents compliance at installation. Operation qualification documents consistent performance within operating ranges. Performance qualification ascertains the instrument is suitable for specific analytical tasks. The document then provides examples of tests to check HPTLC performance, including linearity of spotting, reproducibility of spotting, and detection capacity.
Autoclave
Principle of Autoclave
Construction of Autoclave
Working of Autoclave
Qualification of Autoclave
Installation Qualification
Operational Qualification
Performance Qualification
References
This document provides guidelines for developing specifications for new drug substances and products according to ICH Q6A. It discusses universal and specific tests/criteria that should be included for drug substances and products, such as identification, description, assay, impurities, dissolution, disintegration, content uniformity, and microbial limits. The document gives acceptance criteria and justification for key tests like dissolution, discussing how to set Q values and limits based on biobatch results and BCS classification. It also provides guidance on other tests for oral liquids, parenterals and solid dosage forms.
The document provides details on qualification and validation procedures for gas chromatography (GC) and Fourier transform infrared spectroscopy (FTIR) systems used in pharmaceutical analysis. It describes calibration tests for both instruments, including tests to check wave number precision and reproducibility, linearity, and temperature accuracy and stability. Validation tests are also outlined, such as resolution checks and tests to evaluate performance against acceptance criteria over time as specified in guidance documents. The document provides a thorough overview of qualification processes for critical analytical instruments.
This document discusses various quality control and documentation procedures in the pharmaceutical industry. It includes 3 key points:
1. It discusses the importance of documentation in defining specifications, methods, providing an audit trail and ensuring authorized personnel have necessary information. This includes documents like specifications, batch production records, SOPs etc.
2. It describes procedures for developing key documents like master formulas, batch manufacturing records, audit plans and reports. This ensures uniform processes and allows tracing of batch history.
3. It discusses quality audits which systematically examine if quality activities comply with arrangements and objectives. This includes internal audits as well as those imposed by regulators or customers.
This presentation explains about qualifications of HPTLC, types of qualifications, design qualification , installation qualification ,operational qualification, performance qualification ,documentation of qualification .
It is process of “Establishing documentary evidence that provide a high degree of assurance that a specific process will consistently produce a product meeting its predetermined specifications and quality attributes”.
In the pharmaceutical industry, it is very important that in addition to final testing and compliance of products, it is also assured that the process will consistently produce the expected results.
Validation is action of proving in accordance with the principles of good manufacturing practices, that any procedure, process, equipment, material, activity or system actually leads to expected results.
Cleaning validation is documented evidence with a high degree assurance that one can consistently clean a system or a piece of equipment to predetermined and acceptable limits.
The primary regulatory concern driving the need for cleaning validation is cross contamination of the desired drug substance either by other API from previous batch runs or by residues from the cleaning agents used.
The prime purpose of validating a cleaning process is to ensure compliance with federal and other standard regulations
1. Cross contamination with active ingredients
Contamination of one batch of product with significant levels of residual active ingredients from previous batch cannot be tolerated.
In addition to the obvious problems posed by subjecting consumers or patients to unintended contaminants, potential clinically significant synergistic interactions between pharmacologically active chemicals are a real concern.
2. Contamination with unintended materials or compounds
While inert ingredients used in drug products are generally recognized as safe for human consumption, the routine use, maintenance and cleaning of equipment's provide the potential contamination with such items as equipment parts, lubricants and chemical cleaning agents3. Microbiological contamination
Maintenance , cleaning and storage conditions may provide adventitious microorganisms with the opportunity to proliferate within the processing equipment.
In this slide contains details about Pharmaceutical validation of water system
Presented by: K VENKATSAI PRASAD (Department of pharmaceutical analysis and quality assurance).RIPER, anantapur
In this slide contains Introduction, overview and details of FACTORY ACCEPTANCE TEST
Presented by: P.NARESH (Department of pharmaceutical analysis).RIPER, anantapur
LC-MS is a technique that combines liquid chromatography with mass spectrometry. It separates components in a mixture using HPLC and then uses an ion source like ESI or APCI to ionize the molecules and a mass analyzer like a quadrupole or time-of-flight to separate the ions by mass-to-charge ratio. The separated ions are then detected to identify and quantify each component. Validation of LC-MS systems includes tests of vacuum, mass accuracy, linearity, precision, carryover, and signal-to-noise ratio to ensure proper separation, ionization, detection, and quantification capabilities.
NEW ERA OF DRUG PRODUCT: OPPORTUNITIES AND CHALLENGESganpat420
Abstract
Introduction
Global pharmaceutical industry
Indian pharmaceutical industry
Indian Pharmaceutical Market
Opportunities
Challenges
Conclusion
References
Expiry date , calculation of yields, production record review, change controlSantosh kumar
This presentation discusses key pharmaceutical manufacturing concepts including expiration dating, calculation of yields, production record review, and change control. Expiration dating is determined through stability studies to ensure the stated potency until the end of the product's shelf life. Production record review verifies that manufacturing was properly conducted according to established procedures. Change control manages all proposed alterations to validated processes through a formal review system.
Auditing of vendors and production departmentPRANJAY PATIL
The document discusses vendor audits in the pharmaceutical industry. It provides details on the objectives, parameters, and steps of conducting a vendor audit. The key points are:
- Vendor audits assess a vendor's quality management system, practices, documentation, and adherence to standards to ensure their products and services meet requirements.
- Important parameters reviewed include ISO certifications, manufacturing facilities, packaging and labeling standards, and data handling procedures.
- The goals are to evaluate quality control measures and management commitment to quality standards required by regulations.
- Conducting vendor audits helps reduce costs and risks by gaining insight into supplier processes and compliance.
This document discusses the qualification of manufacturing equipment. It explains that equipment qualification is necessary to ensure equipment works correctly and produces reliable results. There are four types of qualification: design, installation, operational, and performance. Design qualification defines equipment specifications. Installation qualification confirms proper installation. Operational qualification verifies equipment functions as specified. Performance qualification demonstrates consistent performance under routine use. The document then provides details on specific qualification procedures for dry powder mixers and fluidized bed dryers.
Aseptic / sterile- “ A state of control attained by using an aseptic work area and performing activities in a manner that precludes microbiological contamination of the exposed sterile product”
This document discusses the validation of critical utility systems used in pharmaceutical manufacturing facilities. It covers the validation of HVAC systems, water systems, steam systems, compressed air systems, and nitrogen gas systems. For each system, it provides an overview and discusses the user requirements, design qualification, installation qualification, operational qualification, and performance qualification protocols. The validation aims to ensure these utility systems meet quality standards and specifications to support the manufacturing of safe and effective pharmaceutical products.
Auditing of capsule, sterile production and packaging MittalRohit2
The document discusses vendor audits, supplier audits, and audits of sterile product manufacturing facilities. It provides information on:
- The purpose of vendor and supplier audits to assess compliance and reduce costs.
- Key areas evaluated in vendor audits like management responsibility and data integrity.
- Benefits of audits like cost savings, process improvements, and risk reduction.
- Elements of a supplier audit checklist like infrastructure, traceability, and regulatory compliance.
- Additional controls needed for sterile product manufacturing like clean rooms, air filtration, and environmental monitoring.
- Areas examined in audits of sterile facilities including equipment validation, personnel training, and media fill programs.
This document discusses in-process quality control (IPQC) tests for ointments. It describes IPQC as the process of controlling quality parameters during manufacturing from raw materials to final packaging. It then lists and describes 11 common IPQC tests conducted on ointments, including tests for appearance, drug content, pH, sensitivity, spreadability, absorption rate, extrudability, sterility, viscosity, medicament release rate, and uniformity of weight. The tests are designed to ensure the quality, safety and efficacy of ointment products during production.
This document provides standard operating procedures for qualifying various laboratory equipment used in pharmaceutical quality control testing. It includes procedures for calibrating hardness testers, friability test apparatus, tap density testers, disintegration testers, and dissolution test apparatus. The qualification process involves design qualification, installation qualification, operational qualification, and performance qualification to ensure equipment is properly installed, works correctly, and provides expected results. Calibration procedures are also described to verify equipment meets specifications.
The document provides an overview of the qualification process for high performance liquid chromatography (HPLC) equipment, including design qualification (DQ), installation qualification (IQ), operational qualification (OQ), and performance qualification (PQ). It describes the objectives and procedures for each qualification step. Key aspects covered include verifying design specifications, proper installation, operational requirements such as precision, accuracy and noise levels, and ongoing performance monitoring. The goal of qualification is to ensure analytical systems are suitable for their intended use and generate reliable results.
This presentation explains about qualifications of HPTLC, types of qualifications, design qualification , installation qualification ,operational qualification, performance qualification ,documentation of qualification .
It is process of “Establishing documentary evidence that provide a high degree of assurance that a specific process will consistently produce a product meeting its predetermined specifications and quality attributes”.
In the pharmaceutical industry, it is very important that in addition to final testing and compliance of products, it is also assured that the process will consistently produce the expected results.
Validation is action of proving in accordance with the principles of good manufacturing practices, that any procedure, process, equipment, material, activity or system actually leads to expected results.
Cleaning validation is documented evidence with a high degree assurance that one can consistently clean a system or a piece of equipment to predetermined and acceptable limits.
The primary regulatory concern driving the need for cleaning validation is cross contamination of the desired drug substance either by other API from previous batch runs or by residues from the cleaning agents used.
The prime purpose of validating a cleaning process is to ensure compliance with federal and other standard regulations
1. Cross contamination with active ingredients
Contamination of one batch of product with significant levels of residual active ingredients from previous batch cannot be tolerated.
In addition to the obvious problems posed by subjecting consumers or patients to unintended contaminants, potential clinically significant synergistic interactions between pharmacologically active chemicals are a real concern.
2. Contamination with unintended materials or compounds
While inert ingredients used in drug products are generally recognized as safe for human consumption, the routine use, maintenance and cleaning of equipment's provide the potential contamination with such items as equipment parts, lubricants and chemical cleaning agents3. Microbiological contamination
Maintenance , cleaning and storage conditions may provide adventitious microorganisms with the opportunity to proliferate within the processing equipment.
In this slide contains details about Pharmaceutical validation of water system
Presented by: K VENKATSAI PRASAD (Department of pharmaceutical analysis and quality assurance).RIPER, anantapur
In this slide contains Introduction, overview and details of FACTORY ACCEPTANCE TEST
Presented by: P.NARESH (Department of pharmaceutical analysis).RIPER, anantapur
LC-MS is a technique that combines liquid chromatography with mass spectrometry. It separates components in a mixture using HPLC and then uses an ion source like ESI or APCI to ionize the molecules and a mass analyzer like a quadrupole or time-of-flight to separate the ions by mass-to-charge ratio. The separated ions are then detected to identify and quantify each component. Validation of LC-MS systems includes tests of vacuum, mass accuracy, linearity, precision, carryover, and signal-to-noise ratio to ensure proper separation, ionization, detection, and quantification capabilities.
NEW ERA OF DRUG PRODUCT: OPPORTUNITIES AND CHALLENGESganpat420
Abstract
Introduction
Global pharmaceutical industry
Indian pharmaceutical industry
Indian Pharmaceutical Market
Opportunities
Challenges
Conclusion
References
Expiry date , calculation of yields, production record review, change controlSantosh kumar
This presentation discusses key pharmaceutical manufacturing concepts including expiration dating, calculation of yields, production record review, and change control. Expiration dating is determined through stability studies to ensure the stated potency until the end of the product's shelf life. Production record review verifies that manufacturing was properly conducted according to established procedures. Change control manages all proposed alterations to validated processes through a formal review system.
Auditing of vendors and production departmentPRANJAY PATIL
The document discusses vendor audits in the pharmaceutical industry. It provides details on the objectives, parameters, and steps of conducting a vendor audit. The key points are:
- Vendor audits assess a vendor's quality management system, practices, documentation, and adherence to standards to ensure their products and services meet requirements.
- Important parameters reviewed include ISO certifications, manufacturing facilities, packaging and labeling standards, and data handling procedures.
- The goals are to evaluate quality control measures and management commitment to quality standards required by regulations.
- Conducting vendor audits helps reduce costs and risks by gaining insight into supplier processes and compliance.
This document discusses the qualification of manufacturing equipment. It explains that equipment qualification is necessary to ensure equipment works correctly and produces reliable results. There are four types of qualification: design, installation, operational, and performance. Design qualification defines equipment specifications. Installation qualification confirms proper installation. Operational qualification verifies equipment functions as specified. Performance qualification demonstrates consistent performance under routine use. The document then provides details on specific qualification procedures for dry powder mixers and fluidized bed dryers.
Aseptic / sterile- “ A state of control attained by using an aseptic work area and performing activities in a manner that precludes microbiological contamination of the exposed sterile product”
This document discusses the validation of critical utility systems used in pharmaceutical manufacturing facilities. It covers the validation of HVAC systems, water systems, steam systems, compressed air systems, and nitrogen gas systems. For each system, it provides an overview and discusses the user requirements, design qualification, installation qualification, operational qualification, and performance qualification protocols. The validation aims to ensure these utility systems meet quality standards and specifications to support the manufacturing of safe and effective pharmaceutical products.
Auditing of capsule, sterile production and packaging MittalRohit2
The document discusses vendor audits, supplier audits, and audits of sterile product manufacturing facilities. It provides information on:
- The purpose of vendor and supplier audits to assess compliance and reduce costs.
- Key areas evaluated in vendor audits like management responsibility and data integrity.
- Benefits of audits like cost savings, process improvements, and risk reduction.
- Elements of a supplier audit checklist like infrastructure, traceability, and regulatory compliance.
- Additional controls needed for sterile product manufacturing like clean rooms, air filtration, and environmental monitoring.
- Areas examined in audits of sterile facilities including equipment validation, personnel training, and media fill programs.
This document discusses in-process quality control (IPQC) tests for ointments. It describes IPQC as the process of controlling quality parameters during manufacturing from raw materials to final packaging. It then lists and describes 11 common IPQC tests conducted on ointments, including tests for appearance, drug content, pH, sensitivity, spreadability, absorption rate, extrudability, sterility, viscosity, medicament release rate, and uniformity of weight. The tests are designed to ensure the quality, safety and efficacy of ointment products during production.
This document provides standard operating procedures for qualifying various laboratory equipment used in pharmaceutical quality control testing. It includes procedures for calibrating hardness testers, friability test apparatus, tap density testers, disintegration testers, and dissolution test apparatus. The qualification process involves design qualification, installation qualification, operational qualification, and performance qualification to ensure equipment is properly installed, works correctly, and provides expected results. Calibration procedures are also described to verify equipment meets specifications.
The document provides an overview of the qualification process for high performance liquid chromatography (HPLC) equipment, including design qualification (DQ), installation qualification (IQ), operational qualification (OQ), and performance qualification (PQ). It describes the objectives and procedures for each qualification step. Key aspects covered include verifying design specifications, proper installation, operational requirements such as precision, accuracy and noise levels, and ongoing performance monitoring. The goal of qualification is to ensure analytical systems are suitable for their intended use and generate reliable results.
The document discusses the qualification process for UV-Visible spectrophotometers. Qualification ensures instruments are suitable for their intended use and will function properly over time. It involves design, installation, operational, and performance qualification. For a UV-Vis spectrophotometer, key parameters that must be qualified include wavelength accuracy and precision, stray light, resolution, noise, baseline flatness, stability, photometric accuracy, and linearity. The document provides testing protocols for checking each parameter.
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER, FTIR, DSC, HPLCAnupriyaNR
The document discusses the qualification and validation of various analytical techniques used in pharmaceutical quality control including UV-Visible spectrophotometry, Fourier-transform infrared spectroscopy (FTIR), differential scanning calorimetry (DSC), and high performance liquid chromatography (HPLC). It provides details on the design qualifications, installation qualifications, operational qualifications, and performance qualifications required for each technique. The key steps include instrument calibration, determination of accuracy and precision, evaluation of limits, and verification that the instruments are operating as intended over time.
Validation of uv visible spectrophotometer.pptxAnupamaCp2
Validation of UV-Visible spectrophotometer.
It includes installation qualification , design qualification, operational qualification, performance qualification.
The document discusses the qualification of analytical equipment. It describes the types of qualification which include design qualification, installation qualification, operational qualification, and performance qualification. It provides details on qualifying various analytical instruments such as FTIR, GC, and HPLC. Specific parameters to be tested during qualification are discussed for components like the inlet, oven, detector, autosampler, and others, along with typical acceptance criteria. The document emphasizes the importance of qualifying analytical equipment to ensure it is properly installed, operates correctly, and provides expected results.
Qualification of INFARED SPECTROSCOPY BY HIMAJA.Dhimaja donthula
The document presents information on qualifying infrared spectroscopy. It discusses the types of qualification which include design, installation, operational, and performance qualification. Installation qualification ensures the equipment is installed properly. Operational qualification verifies the equipment operates as intended. Performance qualification ensures the equipment performs within specified limits. The document then provides examples of qualification steps for infrared spectrophotometers, including checking parameters like wave number scale, resolution, and noise levels during periodic qualification tests.
This document discusses the qualification of UV-visible spectrophotometry. It begins by defining qualification as an act or process to ensure something complies with conditions, standards, or requirements. There are four types of qualification: design, installation, operational, and performance. UV-visible spectroscopy is concerned with the ultraviolet and visible regions ranging from 200-780 nm. The document outlines parameters for acceptance procedures and performance qualification of a UV-visible spectrophotometer, including wavelength accuracy, stray light, resolution power, noise, baseline flatness, stability, photometric accuracy, and linearity.
Wavelength accuracy is defined as the deviation of a measured wavelength from the known wavelength of an absorption or emission band. Wavelength accuracy and repeatability are important for UV/VIS instruments, as small deviations can cause errors in qualitative and quantitative analysis. Array-based instruments generally have better wavelength stability than scanning instruments, which have moving parts prone to misalignment. Various pharmacopeias specify wavelength accuracy requirements, with some applications requiring sub-nanometer accuracy. Regular calibration using reference standards like holmium oxide is needed to ensure wavelength accuracy is maintained over time.
Calibration and validation of analytical instrumentsSolairajan A
This document discusses the calibration and validation of various analytical instruments used in pharmaceutical analysis. It provides details on calibrating UV-Vis spectrophotometers, IR spectrophotometers, spectrofluorimeters, HPLC, and GC. Calibration ensures instrument readings are accurate against standards, while validation confirms the instrument is correctly installed and operating as intended. The document outlines tests and acceptance criteria for evaluating characteristics like wavelength accuracy, resolution, noise, baseline flatness, sensitivity, flow rate, and linearity during calibration and validation of different analytical instruments.
QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT 1.pptxTaushifulHoque
Quality assurance tests are important to ensure the safe and proper functioning of remote afterloading brachytherapy units. The summary describes the key types of quality assurance tests for brachytherapy units:
1. Mechanical tests check the functioning of sensors, couplings between components, and source positioning accuracy and speed. Electrical tests evaluate interlocks, displays, and console functions.
2. Radiation tests include checks for source leakage and contamination, positioning accuracy and uniformity, temporal accuracy, and source strength verification.
3. Treatment planning system tests assess digitization accuracy and calculation algorithms. Radiation safety tests monitor gamma radiation levels and unit leakage to ensure radiation limits are not exceeded.
QUALIFICATION OF UV-VISIBLE SPECTROPHOTOMETER Ankit Kumar
This document presents a school report on qualifying a UV-Visible spectrophotometer. It discusses the four types of equipment qualification: design qualification, installation qualification, operational qualification, and performance qualification. It describes acceptance procedures and criteria for testing the spectrophotometer's appearance, parts, software version, linearity, noise level, wavelength accuracy, repeatability, stray light, resolution, noise, baseline flatness, and linearity. The document provides references on UV-Vis spectroscopy calibration and validation.
qualification of analytical instruments..M pharmacy 1st year.validationSohailPattan
The document discusses the qualification and calibration of analytical instruments like electronic balances and UV-Visible spectrophotometers. It provides details on the various tests and parameters to be checked during qualification of these instruments to ensure they meet performance requirements. These include tests for baseline flatness, wavelength accuracy, transmittance, absorbance, emission lines, stray light, resolution and photometric linearity. The document also outlines the recommended frequency of qualification and calibration.
This document provides qualification procedures for an electronic balance, pH meter, and UV-visible spectrophotometer. It describes design qualification, installation qualification, and operational qualification tests. For the electronic balance, design qualification includes supplier certification. Installation qualification includes installation and operational tests using reference weights. Operational qualification includes daily measurement of reference weights. For the pH meter and spectrophotometer, design qualification includes selection criteria. Installation qualification includes installation and checks. Operational qualification includes calibration and performance verification tests using standards traceable to national references.
Non-Destructive Testing (NDT) - Industrial Radiography Normal Working Proceduresshahar_sayuti
The document discusses safety and security in industrial radiography. It describes the basic principles and applications of radiography techniques using radiation sources like x-rays and gamma rays. It also discusses the equipment used including radiation devices, safety equipment, and procedures to control external radiation exposure through principles of time, distance and shielding.
University Institute of Pharmaceutical Sciences is a flag bearer of excellence in Pharmaceutical education and research in the country. Here is another initiative to make study material available to everyone worldwide. Based on the new PCI guidelines and syllabus here we have a presentation dealing with qualifications of HPLC which is the " High Performance Liquid Chromatography".
Thank you for reading.
Hope it was of help to you.
UIPS,PU team
validation HPLC and HPTLC D-A star.pptxDevaPundkar
The document discusses the qualification of analytical instruments like HPLC and HPTLC. It describes the four stages of qualification as design qualification, installation qualification, operational qualification and performance qualification. It provides details about each stage, parameters to be tested and acceptance criteria for HPLC and HPTLC. The key aspects of HPLC instrumentation and working are also summarized. The qualification process ensures that the equipment is properly installed, works correctly and provides expected results.
Performance qualification of High performance liquid chromatography Atchaya Thalapathy
This document discusses performance qualification (PQ) of high performance liquid chromatography (HPLC). PQ is part of the instrument qualification process and demonstrates that the HPLC system consistently performs according to specifications. The document outlines several PQ tests for the HPLC including injector precision and linearity, resolution between peaks, solvent delivery flow rate accuracy and precision, and detector parameters. It provides details on how to conduct these tests and acceptance criteria to ensure the HPLC is functioning properly.
The document discusses inspection, non-destructive testing, and statutory requirements for equipment. It defines inspection and describes visual inspection and non-destructive testing techniques like dye penetrant, magnetic particle, ultrasonic, radiography, and eddy current testing. The objectives and steps of an inspection program are outlined. Statutory requirements for equipment in India like the Boiler Regulations, Factories Act, and codes for pressure vessels and tanks are also summarized.
Semelhante a qualification of instrument(UV & FTIR) BY Bhumi Suratiya, M.Pharm sem 2.pptx (20)
ViewShift: Hassle-free Dynamic Policy Enforcement for Every Data LakeWalaa Eldin Moustafa
Dynamic policy enforcement is becoming an increasingly important topic in today’s world where data privacy and compliance is a top priority for companies, individuals, and regulators alike. In these slides, we discuss how LinkedIn implements a powerful dynamic policy enforcement engine, called ViewShift, and integrates it within its data lake. We show the query engine architecture and how catalog implementations can automatically route table resolutions to compliance-enforcing SQL views. Such views have a set of very interesting properties: (1) They are auto-generated from declarative data annotations. (2) They respect user-level consent and preferences (3) They are context-aware, encoding a different set of transformations for different use cases (4) They are portable; while the SQL logic is only implemented in one SQL dialect, it is accessible in all engines.
#SQL #Views #Privacy #Compliance #DataLake
Global Situational Awareness of A.I. and where its headedvikram sood
You can see the future first in San Francisco.
Over the past year, the talk of the town has shifted from $10 billion compute clusters to $100 billion clusters to trillion-dollar clusters. Every six months another zero is added to the boardroom plans. Behind the scenes, there’s a fierce scramble to secure every power contract still available for the rest of the decade, every voltage transformer that can possibly be procured. American big business is gearing up to pour trillions of dollars into a long-unseen mobilization of American industrial might. By the end of the decade, American electricity production will have grown tens of percent; from the shale fields of Pennsylvania to the solar farms of Nevada, hundreds of millions of GPUs will hum.
The AGI race has begun. We are building machines that can think and reason. By 2025/26, these machines will outpace college graduates. By the end of the decade, they will be smarter than you or I; we will have superintelligence, in the true sense of the word. Along the way, national security forces not seen in half a century will be un-leashed, and before long, The Project will be on. If we’re lucky, we’ll be in an all-out race with the CCP; if we’re unlucky, an all-out war.
Everyone is now talking about AI, but few have the faintest glimmer of what is about to hit them. Nvidia analysts still think 2024 might be close to the peak. Mainstream pundits are stuck on the wilful blindness of “it’s just predicting the next word”. They see only hype and business-as-usual; at most they entertain another internet-scale technological change.
Before long, the world will wake up. But right now, there are perhaps a few hundred people, most of them in San Francisco and the AI labs, that have situational awareness. Through whatever peculiar forces of fate, I have found myself amongst them. A few years ago, these people were derided as crazy—but they trusted the trendlines, which allowed them to correctly predict the AI advances of the past few years. Whether these people are also right about the next few years remains to be seen. But these are very smart people—the smartest people I have ever met—and they are the ones building this technology. Perhaps they will be an odd footnote in history, or perhaps they will go down in history like Szilard and Oppenheimer and Teller. If they are seeing the future even close to correctly, we are in for a wild ride.
Let me tell you what we see.
The Ipsos - AI - Monitor 2024 Report.pdfSocial Samosa
According to Ipsos AI Monitor's 2024 report, 65% Indians said that products and services using AI have profoundly changed their daily life in the past 3-5 years.
Codeless Generative AI Pipelines
(GenAI with Milvus)
https://ml.dssconf.pl/user.html#!/lecture/DSSML24-041a/rate
Discover the potential of real-time streaming in the context of GenAI as we delve into the intricacies of Apache NiFi and its capabilities. Learn how this tool can significantly simplify the data engineering workflow for GenAI applications, allowing you to focus on the creative aspects rather than the technical complexities. I will guide you through practical examples and use cases, showing the impact of automation on prompt building. From data ingestion to transformation and delivery, witness how Apache NiFi streamlines the entire pipeline, ensuring a smooth and hassle-free experience.
Timothy Spann
https://www.youtube.com/@FLaNK-Stack
https://medium.com/@tspann
https://www.datainmotion.dev/
milvus, unstructured data, vector database, zilliz, cloud, vectors, python, deep learning, generative ai, genai, nifi, kafka, flink, streaming, iot, edge
Analysis insight about a Flyball dog competition team's performanceroli9797
Insight of my analysis about a Flyball dog competition team's last year performance. Find more: https://github.com/rolandnagy-ds/flyball_race_analysis/tree/main
Predictably Improve Your B2B Tech Company's Performance by Leveraging DataKiwi Creative
Harness the power of AI-backed reports, benchmarking and data analysis to predict trends and detect anomalies in your marketing efforts.
Peter Caputa, CEO at Databox, reveals how you can discover the strategies and tools to increase your growth rate (and margins!).
From metrics to track to data habits to pick up, enhance your reporting for powerful insights to improve your B2B tech company's marketing.
- - -
This is the webinar recording from the June 2024 HubSpot User Group (HUG) for B2B Technology USA.
Watch the video recording at https://youtu.be/5vjwGfPN9lw
Sign up for future HUG events at https://events.hubspot.com/b2b-technology-usa/
STATATHON: Unleashing the Power of Statistics in a 48-Hour Knowledge Extravag...sameer shah
"Join us for STATATHON, a dynamic 2-day event dedicated to exploring statistical knowledge and its real-world applications. From theory to practice, participants engage in intensive learning sessions, workshops, and challenges, fostering a deeper understanding of statistical methodologies and their significance in various fields."
Build applications with generative AI on Google CloudMárton Kodok
We will explore Vertex AI - Model Garden powered experiences, we are going to learn more about the integration of these generative AI APIs. We are going to see in action what the Gemini family of generative models are for developers to build and deploy AI-driven applications. Vertex AI includes a suite of foundation models, these are referred to as the PaLM and Gemini family of generative ai models, and they come in different versions. We are going to cover how to use via API to: - execute prompts in text and chat - cover multimodal use cases with image prompts. - finetune and distill to improve knowledge domains - run function calls with foundation models to optimize them for specific tasks. At the end of the session, developers will understand how to innovate with generative AI and develop apps using the generative ai industry trends.
3. Introduction
Validation: Establishing documented evidence that
provides a high degree of assurance that a process, system,
equipment or assay will consistently produce the desired
results according to predetermined specifications and
quality attributes.
It is a requirement for Good Manufacturing Practices and other
regulatory requirements.
3
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
4. Calibration
Calibration is a necessary component to ensure of the
authenticity of Qualification and Validation.
Calibration is a process that demonstrates a particular
instrument or device produces results within specified
limits, as compared to those produced by a definite standard
over an appropriate range of measurements.
4
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
5. Qualification
Action of proving and documenting that any equipment,
utilities, and systems actually and consistently leads to the
expected results.
Qualification should be completed before process
validation is performed.
Qualification Phases:
The Qualification phases consist of mainly 5 phases.
Its start with User Requirement Specification (URS).
URS is a document which states the laboratory
requirement and technical with operational
requirements that should be met. 5
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
6. The other phases consist of
• Design qualification(DQ)
• Installation qualification(IQ)
• Operational qualification(OQ)
• Performance qualification(PQ)
6
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
7. oDesign Qualification:
Documented evidence which
shows that the plant design agrees
with the design specifications of
the customer.
Operational qualification:
Documented evidence which
shows that all parts of the plant
and equipment work within their
specifications and parameters
Process.
Installation Qualification:
Written evidence is given that all
parts of equipment are installed
according to the equipment
supplier's and purchase
specifications.
o Performance
Qualification: Provides
documented evidence that all
parts of a plant and other
processes produce products of
specified quality under conditions
of normal production for a longer
period of time.
Qualification
7
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
8. Design
Qualification
Installation
Qualification
Operational
Qualification
Performance
Qualification
Timing and Applicability
Prior to purchase of a
new model of
instrument
At installation of each
instrument (new, old, or
existing unqualified)
After installation or
major repair of each
instrument
Periodically at specified intervals for
each instrument
Activities
Assurance of
manufacturer's DQ
Description Fixed parameters Preventive maintenance and repairs
Assurance of adequate
support availability
from manufacturer
Instrument delivery Establish practices to address
operation, calibration, maintenance,
and change control
Instrument's fitness for
use in laboratory
Utilities/facility Environment
Assembly and installation
Network and data storage Secure data storage,
backup, and archive
Installation verification Instrument function
tests
Performance checks
Table:- Timing, Applicability, and Activities for Each Phase of Analytical Instrument Qualification: Activities under
each phase are usually performed as indicated in the table. In some cases, it may be more appropriate to combine a given activity or perform
with another phase. Such activities are shown linked (↔). It is more important that the activity is performed, but not so important under which
phase it is performed.
8
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
9. Class Type of class instrument or
equipment
Process of
qualification
A Simple Magnetic stirrer No qualification
B Moderate pH meter/oven
Partial qualification
routine calibration,
performance check
maintenance
C Complex FT-IR/HPLC/GC
Full qualification-all
elements of qualification
including software
validation
Instruments can generally be classified as belonging to groups A, B, or
C. It should be noted that the same type of instrument can fit into one
or more categories, depending on its intended use.
9
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
11. Supplier must provide documented evidence that the product has
been designed, developed and manufactured in a quality
environment e.g. ISO 9001:2000 certification.
Design Qualification Should be performed before the purchase of
new model of UV Spectrophotometer.
The documents of design qualification should explain:
What the user want the instrument to do?
Examples of functional and operational specifications are:
o Optics: Double beam/ Single beam
o Measurable range: e.g 190-1100 nm
o Wavelength Accuracy: e.g. ±0.5 nm or better
Design Qualification(DQ)
11
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
12. o Criteria for selection of the vendor
o Warranty and maintenance support
o Discount
o Cost of Annual Maintenance Contract (AMC) after
expiry of standard warranty.
Design Qualification(DQ)
12
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
13. Installation Qualification(IQ)
IQ is the documented collection of activities necessary to establish that
an instrument is delivered as designed and specified, and is properly
installed in the selected environment, and that this environment is
suitable for the instrument.
According to USP, the IQ requirements provide evidence that the
hardware and software are properly installed in the desired location.
The instrument has been checked and verified as undamaged.
The appropriate documentation has been supplied and it is of correct
issue and uniquely identified by part number, version number and date.
13
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
14. Operational Qualification(OQ)
OQ is the documented collection of activities necessary to demonstrate
that an instrument will function according to its operational
specification in the selected environment.
Following are the test performed during Operational . Qualification for
UV Spectrophotometer
Wavelength Accuracy
Control of Absorbance
Resolution
Noise
Stability
Limit of Stray Light
14
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
15. Parameter Test Acceptance criteria
Wavelength
accuracy
• Perform to understand deviation of the wavelength
reading from the known wavelength of the band
• Usually, Holmium oxide solution (4% holmium
oxide in 10% perchloric acid) is used
• Holmium oxide solution gives characteristic peaks
at 241.1. 249.9, 278.1, 287.2. 333.5, 345.4 etc.
• ±1 nm in UV region (200-400
nm). ± 2 nm (400-700 nm)
• Three repeated scans of same
peak standard deviation of the
mean must not exceed 0.5 nm
Stray light • This may be due to poor design or faulty
monochromator or may be because of operator,
usually not problem in new instrument but increases
with age of optics and its degradation.
• Higher absorbance ranges (above 1) are more
susceptible.
• For the range of 300-350 nm, aqueous sodium
nitrite (50 g/L) is scanned
The transmittance of the solution
in a 1cm cell should be less than
0.01 or the absorbance value
should be greater than 2
15
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
16. Parameter Test Acceptance criteria
Resolution
power
• Insufficient resolution in any spectrophotometer may loose some
features of fine spectrum or unable to show difference of
absorbance values between two close wavelengths.
• Measure the ratio of the absorbance of a 0.020% (v/v) solution of
toluene in hexane (UV grade) at the maximum and minimum at
about 269 and 266 nm. respectively, using hexane as the
reference.
• The ratio of the
absorbance at 269 nm and
absorbance at 266 nm
should be greater than 1.5
Noise • Noise originating from light source and electronic systems in UV
spectrophotometer and effects the measurement of absorbance in
both lower and higher wavelengths.
• To estimate the noise of spectrophotometer air is scanned (no
sample) for 10 min in absorption mode and Root Mean Square
(RMS) is calculated and recorded at 500 nm.
• The RMS noise should be
less than 0.001 AU
16
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
17. Parameter Test
Acceptance
criteria
Baseline
flatness
• Baseline is used to nullify the effect of environment in measurement by
removing background noise to get true absorption profile of analyte.
• In double beam UV spectrophotometer baseline correction is usually performed
by keeping solvent in one cuvette and test solution in other. This will
automatically correct the baseline and subtract the background noise by solvent.
• This test is performed by scanning air in the absorbance mode in UV region and
highest and lowest deflections in the absorbance unit are recorded.
• The
measurement is
typically
less than 0.01 A
U
Stability • The changes in the components and age are major factors making instability of
response by UV Spectrophotometer resulting error in readings
• The error caused by this factor may be positive or negative (Positive: more than
actual observed, Negative: less than actual response observed)
• For checking stability air is for 60 mins in absorbance mode particular
wavelength at (generally 340 nm) and deflections in the absorbance are
recorded.
• The deflection
is less than
0.002 AU/ hr
17
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
18. Parameter Test Acceptance criteria
Photometric
accuracy
• Photometric accuracy is determined by comparing the
difference between the measured absorbance of the
reference material and the established value.
• Six replicate measurements of
the 0.006% w/v of the potassium
dichromate solution at 235 257,
313 and 350 nm should be less
than 0.5% RSD.
Linearity • Ideally, the absorbance should increase with increase
in the concentration of analyte in solution
• For this various dilutions of Potassium Dichromate
are prepared ranging from 20, 40, 60, 80, and 100
mg/L in 0.005 M sulfuric acid and scanned by
keeping 0.005 M sulphuric acid in reference cuvette.
• The calibration curve is prepared by measuring
absorbance of each solution and correlation
coefficient is calculated.
• Correlation coefficient r > 0.999
18
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
19. Performance Qualification(PQ)
Performance of instrument may change gradually over time because of
normal wear of parts, failure or change of its components.
The purpose of the PQ is to provide evidence that instrument is fit for
its routine use.
Generally, performance of spectrophotometer measurements are
performed under identical condition for the test specimen and the
reference substance.
19
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
22. Supplier must provide documented evidence that the product has been
designed, developed and manufactured in a quality environment e.g.
ISO 9001:2000 certification.
Supplier must provide phone and on-site support in case of defects.
Supplier must provide Information through the internet on availability
of new firmware upgrades.
Design Qualification(DQ)
22
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
23. Attribute Specifications
Detector range
The optical bench shall include a
DTGS(Deuterated triglycine sulfate)
detector with a 7400-350cm-1.
Compressed air interferometer -
Air cooled standard infrared source -
Spectral resolution
The instrument shall have a spectral
resolution not exceeding than 1.07400-
350cm-1.
The below is a sample of technical attributes that may be considered in
design qualification of an FT-IR spectrometer and may be adapted if
necessary.
23
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
24. Installation Qualification(IQ)
Equipment is compared as received with purchase order, including
software, accessories, spare parts and consumables.
Documentation checked for completeness of operating,
maintenance instructions, standard operating procedures for
testing and safety, validation certificates and health and safety
instruments.
Equipment is checked for any damage.
The supplier’s instruction for installation is read.
24
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
25. The instruments are switched on and ensured that all modules
power up and an electronic self test is performed. Any deviations
are recorded.
Software is installed on computer by following the manufacturer’s
recommendation.
Correct software installation is verified.
A backup copy of software is made.
Peripherals (e.g. printers and equipment) modules are configured.
A list with a description of all hardware are identified an made
including drawings where appropriate.
25
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
26. A list with a description of all software installed on the
computer is made.
Equipment's manuals and SOPs are listed.
An installation report is prepared.
26
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
27. Operational Qualification
List of
parameters
which
should be
checked:-
Wavelength accuracy
Wavelength resolution
Wave number scale
Detector Energy Ratio
Signal/noise Ratio
Zero Test
Contamination Test
27
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
28. Wavelength accuracy
Test procedure
Acceptance
limits
Test frequency Remarks
Measurement of
polystyrene
spectrum at 1144,
1680, 2167 and
2307 nm. The
results are
compared with
reference values.
±2 nm Daily Standard should
be traceable to
national standard.
28
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
29. Wavelength resolution
Test procedure Acceptance limits Test frequency
Resolution of
polystyrene at 2870/2851
and at 1589/1583 nm.
T(band 2870 cm-1 - band
2851 cm-1 )18
T(band 1589 cm-1 - band
1583 cm-1 )12
Daily
29
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
30. Wave number Scale
Test procedure
Transmission
minima(cm-1 )
Acceptable
tolerance(cm-1 )
Recording the spectrum
of a polystyrene film
which has transmission
minima(absorption
maxima) at the wave
numbers given in the
next column.
3060.0 ±1.0
2849.5 ±1.0
1942.9 ±1.0
1601.2 ±1.0
1583.0 ±1.0
1154.5 ±1.0
1028.3 ±1.0 30
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
31. Detector Energy Ratio
Test procedure Acceptance limits
The minimum energy ratio value for at
least one of the following is measured
and it is compared to the vendor’s
specifications:
o Energy at 3990 cm-1 /energy at
2000 cm-1
o Energy at 3400 cm-1 /energy at
1300 cm-1
Energy ratio test specifications vary
for each spectrometer configuration
and must be referred to the
manufacturer’s specifications.
31
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
32. Signal/noise Ratio
Test
procedure
Peak to peak
noise between
RMS(root mean
square) noise
between
Limits(%T)
The maximum
noise level is
recorded for
each of the
regions given
in the next two
columns.
• 4050 cm-1 and
3950 cm-1
• 2050 cm-1 and
1950 cm-1
• 050 cm-1 and
950 cm-1
• 550 cm-1 and
450 cm-1
• 4050 cm-1 and
3950 cm-1
• 2050 cm-1 and
1950 cm-1
• 1050 cm-1 and
950 cm-1
• 550 cm-1 and
450 cm-1
Noise level test
specifications vary
for each
spectrometer
configuration and
must be referred to
the manufacturer’s
specifications.
32
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
33. Zero Test
Test procedure Limits(%T)
When using a polystyrene film of approximately 35µm in
thickness as standard at the wavelength of 2925 cm-1 and 700
cm-1, almost complete absorption of the irradiated energy can
be observed.
With this test, the remaining transmission is measured.
As the maximum absorption can be observed at 700 cm-1
negative values may be observed.
The objective of the test is to evaluate if despite the fact that
there is almost complete absorption, energy is still detectable.
Non-valid results are an indication of a non-linear behavior of
the detector and the electronic system.
Results vary
for each
spectrometer
configuration
and must be
referred to the
manufacturer’s
specifications.
33
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
34. Contamination Test
Test procedure Wave number(cm-1 ) Upper limit(A)
The automated function of the
instrument is used to perform
this test (If automation is not
available then a background
spectrum is recorded).
This test checks the presence of
peaks that signal a
contamination problem.
3100 -2800 0.1
1800 -1600 0.1
1400 -1100 0.2
34
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
36. Wavenumber accuracy
The wave number scale is usually calibrated by the use of several
characteristics.
Wave numbers of a polystyrene film:
• 3060.0 (±1.5)cm-1
• 2849.5 (±1.5)cm-1
• 1942.9 (±1.5)cm-1
• 1601.2 (±1.0)cm-1
• 1583.0 (±1.0)cm-1
• 1154.5 (±1.0)cm-1
• 1028.3(±1.0)cm-1
The software then
judges whether the
values are within the
allowable range and the
program labels the
results ‘PASS’ if all the
peak numbers are
within the range. 36
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
37. Wavenumber reproducibility
This program specifies three points to measure the peak wave
numbers.
Then it obtains the actual peak wave numbers at each point by
measuring the polystyrene film twice.
It should satisfy 5 cm-1 around 3000 cm-1 of polystyrene
absorption wave number, 1 cm-1 around 1000 cm-1 .
The software determines whether the difference between each of
two measurements are within the allowable range and it labels
the results ‘PASS’ if they are within the range.
37
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
38. Transmittance reproducibility
This program specifies peak wave numbers at three points and
the transmittance at each point is measured twice.
The transmittance reproducibility should satisfy 0.5 %T when
the several points of polystyrene absorption from 3000 cm-1 to
1000 cm-1 are measured twice.
Then it is determined whether the differences between the two
data are within the range.
The FT-IR abnormalities or large changes over short term and
long term are assessed by these tests.
38
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
39. The two parameters checked by this program are:
Energy spectrum test:-power
spectra obtained in the inspection
are compared with reference data
and the spectra are checked for
changes over long periods. 100%T
line spectra are calculated for power
spectra and are measured
continuously in inspection and the
spectra are checked for changes
over short periods.
A.Polystyrene test:- evaluation is
performed using differences
between spectra obtained for
polystyrene film in inspection and
the stored reference data. When the
differences are within the standard,
it passes the first.
39
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
40. References
1. Gouthami B, Venkateshwarlu G, Venkateshwarlu P. Calibration and
Validation of HPLC, GC and UV-VIS Spectroscopy. Int J Mdm. chem
Appl Sci. 2014,1(4):27-34
2. N.R.Anupriya, “Qualification of UV Visible Spectrophotometry, June
2023, https://www.slideshare.net/AnupriyaNR/qualification-of-
uvvisible-spectrophotometer-ftir-dsc-hplc.
3. Indian Pharmacopoeia Commission , Ghaziabad , Indian Pharmacopoeia
2022, (1): 221-223.
4. Sharma P. Validation in pharmaceutical industry. 2nd ed. Delhi:
Vandana publications; 2013.P:167
40
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
41. References
5. David Rudd, qualification of analytical equipment, method validation
in pharmaceutical analysis, part 2, (chapter 4), 2005
6. D.Gowrisankar, K. Abbulu, O.Bala souri, K.Sujana, Validation and
Calibration of analytical instruments, 2010; 2(2).
41
B
h
u
m
i
S
u
r
a
t
i
y
a
M
p
h
a
r
m
S
e
m
2
Notas do Editor
Stray light is defined as the detected light of any wavelength that is outside the band width of the wavelength selected.
Resolution power- The resolution of the UV-VIS spectrometer is related to its spectral band width. The smaller the band width the finer the resolution. The SBW depends on the slit width and the dispersive power of the monochromator.
Noise - Noise is the measurement affects the accuracy at the both end of the absorbance scale. Photon noise from the light source affects the accuracy of the measurement leads to low absorbance.
Base line flatness - he flat baseline test demonstrates that the ability of the instrument to normalize the light intensity measurement and the spectral out put at different wavelength throughout the spectral range.
Stability - The lamp intensity is a function of the lamp age, temperature fluctuation and wavelength of the measurement. These changes can lead to errors in the value of the measurements, over an extended period of time.
Linearity - The linearity dynamic range of measurement is limited by stray light at high absorbance and by noise at low absorbance. The accuracy of the quantification of the sample depends on the precision and linearity of the measurements.
Linearity - The linearity dynamic range of measurement is limited by stray light at high absorbance and by noise at low absorbance. The accuracy of the quantification of the sample depends on the precision and linearity of the measurements.