SlideShare uma empresa Scribd logo
1 de 16
UNIVERZA V LJUBLJANI
                     ZDRAVSTVENA FAKULTETA
   ODDELEK ZA RADIOLOŠKO TEHNOLOGIJO II. STOPNJE




 OBSEVALNE TEHNIKE NA TELERADIOTERAPEVTSKEM
             ODDELKU ONKOLOŠKEGA INŠTITUTA




Avtor: Andrej Breznik, dipl.inţ.rad.
Mentor: doc. dr. Dejan Ţontar, univ. dipl. fiz.




                             LJUBLJANA, 2011
KAZALO VSEBINE



KAZALO SLIK

Slika 1: Sestava linearnega pospeševalnika ............................................................................... 2
Slika 2: Nameščen tubus in individualna zaščita na obsevalnem aparatu .................................. 4
Slika 3: Porazdelitev globinske doze pri elektronih ................................................................... 4
Slika 4: Večlistni kolimator – MLC ........................................................................................... 5
Slika 5: Porazdelitev globinske doze pri fotonih ....................................................................... 6
Slika 6: LINAC NovalisTX – obsevalni aparat 8 ...................................................................... 8
Slika 7: Primerjava dozne porazdelitve pri različnih obsevalnih tehnikah ................................ 9
Slika 8: Exac Trac 6D robotic IGRT couch sistem .................................................................. 10
Slika 9: Prikaz poravnave z tehniko prekrivanja in deljenega polja pri slikanju z sistemom
Exac Trac.................................................................................................................................. 11
Slika 10: Primerjava obsevalnih planov IMRT in RAPID ARC tehnike obsevanja ................ 13
1 KAJ JE RADIOTERAPIJA?


Radioterapija je medicinska stroka, ki se ukvarja z zdravljenjem malignih in nemalignih
bolezni s pomočjo ionizirajočega sevanja, lahko samostojno ali v kombinaciji z sistemskim in
kirurškim zdravljenjem. Uvrščamo jo med lokalne načine zdravljenja, saj je njen učinek
omejen le na obsevalno območje, kjer se absorbirajo ionizirajoči ţarki. Skupaj s kirurgijo in
sistemskim zdravljenjem je eden izmed treh temeljnih načinov zdravljenja raka.


Glede na poloţaj vira ionizirajočega sevanja v odnosu na obsevanega bolnika delimo
radioterapijo v vsebinskem in organizacijskem delu na teleterapijo (vir je zunaj bolnikovega
telesa) in brahiterapijo (vir je v bolnikovem telesu).


Bolniki so lahko obsevani z fotonskimi ţarkovnimi snopi nizkih (kV) ali visokimi (MV)
energijami in z elektronskimi ţarkovnimi snopi visokih (MeV) energij. Brahiterapevtska
dejavnost se izvaja z napravami za naknadno polnjenje vodil (HDR in LDR), ki jih zdravnik
povprej pod kontrolo različnih slikovnih metod uvede v telo bolnika, z viri ionizirajočega
sevanja.


Kot stroka radioterapija opredeljuje preplet kliničnih znanj in veščin poznavanja značilnosti
malignih in neonkoloških bolezni, ter moţnosti njihovega zdravljenja. Prav tako zavzema
področje radiofizike in radiobiologije, ki poznavata učinke ionizirajočega sevanja v bioloških
sistemih in moţnostih njegove uporabe. Radioterapija vključuje tudi tesno sodelovanje
strokovnjakov, kot so zdravnik specialist radioterapevt (nosilec dejavnosti), medicinski fizik –
dozimetrist (odgovoren za pripravo obsevalnega načrta in umerjanje radioterapevtskih naprav)
in inţenir radiološke tehnologije (upravljalec radioterapevtskih naprav in izvajalec obsevanja)
(Novaković in sod., 2009).


Med osnovne teleterapevtske naprave spadajo radioterapevtski simulator, terapevtska
rentgenska naprava, linearni pospeševalnik in računalniški sistemi za načrtovanje obsevanja in
naprave za preverjanje kakovosti obsevanja (Novaković in sod., 2009).




                                                                                              1
2 LINEARNI POSPEŠEVALNIK - LINAC


V teleterapiji uporabljamo linearne pospeševalnike, ki lahko proizvajajo visokoenergijske
fotone ali elektrone. Sodobni medicinski linearni pospeševalniki pospešujejo elektrone do
kinetičnih energij od 4 MeV do 25 MeV (Novaković in sod., 2009). Linearni pospeševaliki
uporabljajo RF mikrovalovna polja frekvenc od 103 MHz do 104 MHz. Velika večina jih
deluje pri frekvenci mikrovalov 2856 MHz (S-band), kjer so tipične dolţine mikrovalovnih
votlin okrog 10 cm v premeru in 3 cm v dolţini (Majerle, 2003).



                    Izravnalna                                  Bending magnet
                      filtra in
                      sipalne
                        folije                                               Pospeševalna cev

                                                                              Izvor elektornovov - GUN

                      Tarča


                     Primarni in
                     sekundrani
                      kolimator

                           MLC


                         Rtg cev -
                           OBI
                                                                                                   Slikovni
                                                                                                   detektor



                                             EPID
                                            slikovni
                                           detekotor



                                    Slika 1: Sestava linearnega pospeševalnika

VIR:http://www.google.si/imgres?imgurl=http://www.tradevision.com.bd/images/product/27.jpg&imgrefurl=http://www.tradevision.com.bd
/productlist.php%3Fsubdivid%3D7%26subdivname%3DRadiotherapy&h=600&w=600&sz=71&tbnid=pTS2cx5lOJc0tM:&tbnh=135&tbnw
=135&prev=/images%3Fq%3Dvarian%2Blinac&zoom=1&q=varian+linac&hl=sl&usg=__G5f4fyw7_ohDq65TSNfAwj8o5Ks=&sa=X&ei=
                                     jzV_TY7LGsTYsgaam8jsBg&sqi=2&ved=0CD4Q9QEwBA




                                                                                                                                2
Princip delovanja linearnega pospeševalnika je takšen, da v pospeševalno cev injicirane
elektrone z začetno energijo okoli 50 keV vzdolţ le-te pospešimo s pomočjo
visokofrekvenčnega elektromagnetnega valovanja s frekvenco okoli 3000 MHz. V
pospeševalni cevi je vakuum, ki zagotavlja čim manjše število trkov elektronov z molekulami
plina in s tem čim manjše oviranje elektronov. Ob izhodu iz pospeševalne cevi dobimo ozek
snop elektronov s premerom cca 3 mm, ki ga v nadaljevanju s pomočjo magnetnih polj
usmerimo (odklonimo) v navpično smer. Elektronski curek lahko uporabljamo neposredno
(obsevanje z elektroni) ali pa ga izkoristimo za nastanek rentgentskih fotonov z usmeritvijo
elektronov v tarčo (Novaković in sod., 2009).


V principu ločimo dve vrsti obsevanja, glede na vrsto ionizirajočega sevanja:
   -   obsevanje z elektroni,
   -   obsevanje s fotoni.




2.1. Obsevanje z elektronskim snopom

V primeru obsevanja z elektronskim snopom, pospešeni elektroni trčijo ob vstavljeno tanko
kovinsko sipalno folijo, katera ţarek razprši, hkrati pa poskrbi za homogenost obsevalnega
polja, kar pomeni, da je količina elektronov po celotnem preseku obsevalnega polja enaka
»dose flattnes«. Debeline sipalnih folij so takšne, da se elektroni v preteţni meri sipljejo in da
ne prihaja do dodatnega zavornega sevanja. Izhodni snop elektronov omejujemo z t.i.
aplikatorji oz. tubusi in individualnimi zaščitami iz Woodove zlitine, ki posnemajo obliko
tarče. Tipične energije za obsevanje z elektorni so: 6, 9, 12, 15 in 18 MeV. Globinska
porazdelitev doze pa je odvisna od jakosti energije. Pomembno je vedeti, da ima obsevanje z
elektroni hiter padec izodozne krivulje, ko ta doseţe 80%. Na takšen način se lahko izognemo
nepotrebnemu obsevanju t.i. kritičnih oz. zdravih organov. (Novaković in sod., 2009).




                                                                                                3
Slika 2: Nameščen tubus in individualna zaščita na obsevalnem aparatu

                    Vir: http://www.flickr.com/photos/vm1757/562061042/sizes/m/in/photostream/




                        100,0
                                                                                    6 MeV
                                                                                    9 MeV
                         80,0
                                                                                    12 MeV
              PDD [%]




                                                                                    15 MeV
                         60,0
                                                                                    18 MeV

                         40,0

                         20,0

                          0,0
                                0      2          4          6          8          10            12

                                                      globina [cm]

                            Slika 3: Porazdelitev globinske doze pri elektronih


Slika 3 prikazuje razmerje med globino in odstotkom globinske doze (angl. Percent depth
dose – PDD) pri elektronskem snopu.




                                                                                                      4
2.2. Obsevanje s fotonskim snopom

Fotonsko obsevanje poteka po postopku, da elektroni na poti iz pospeševalne cevi trčijo na
vstavljeno tarčo, ki je snov z visokim vrstnim številom Z. V tarči prihaja do zavornega
sevanja in rezultat je nastanek visoko energijskih fotonov primernih za fotonsko obsevanje. V
nadaljevanju fotoni zadanejo izravnalni filter, ki poskrbi, da je ţarek po prehodu skozenj po
svojem preseku čim bolj homogen oz. izravnan. Za dodatno omejitev fotonskih ţarkov skrbi
serija zaslonk, ki ţarek še dodatno oblikujejo. Zadnji element, ki omogoča skoraj poljubno
oblikovanje obsevalnih poljm je t.i. večlistni komatorski sistem MLC (angl. MultiLeaf
Collimator), s katerim oblikujemo fotonska obsevalna polja tako, da ta povzamejo obliko
obolenja, kot ga vidimo iz smeri ţarkov. Sodobni pospeševalniki imajo vgrajen tudi EPID –
elektronski portalni sistem za verifikacijo obsevalnih polj. Pri fotonskem načinu obsevanja se
posluţujemo energijama 6 MV in 15 MV (Novaković in sod., 2009).




                              Slika 4: Večlistni kolimator – MLC
                      Vir: http://varian.mediaroom.com/file.php/301/MLC+-+gold.jpg




                                                                                            5
100
                                                                                 6 MV
                                                                                 15 MV
                    80
                                                                                 Kobalt
          PDD [%]


                    60

                    40

                    20

                     0
                          0              10             20              30                40

                                                 globina [cm]

                              Slika 5: Porazdelitev globinske doze pri fotonih


Slika 5 prikazuje razmerje med globino in odstotkom globinske doze (angl. Percent depth
dose – PDD) pri fotonskem snopu.




                                                                                               6
3 NOVE OBSEVALNE TEHNIKE NA OI


Vključitev novih obsevalnih tehnik v delovno rutino je pogojena z uporabo CT simulatorja in
sodobne podprte računalniške programe, ki s pomočjo algoritmov omogoča izdelavo
natančnega obsevalnega načrta. Z razvojem računalniške podprte tehnologije je sledil tudi
razvoj popolnega računalniškega nadzora oz. krmiljenje linearnih pospeševalnikov, večlistnih
kolimatorjev za oblikovanje obsevalnega polja in računalniško podprtega sistema za
načrtovanje obsevanja (Novaković in sod., 2009).


Prednosti novih obsevalnih tehnik omogoča:
   -   povečanje gradienta med visokodoznim in nizkodoznim območjem,
   -   povečanje konformnosti visokodoznega območja (skladnost s 3D obliko tarče),
   -   natančnejše obsevanje (zmanjšanje odmika lege posameznega polja od referenčne).


Na Onkološkem inštitutu v Ljubljani smo z nakupom novejših obsevalnih aparatov pričeli
obsevati naslednje različne obsevalne tehnike:
   -   IMRT – intizitetno modulirana radioterapija (angl. Intensity-modulated radiation
       therapy)
   -   IGRT – slikovno vodena radioterapija (angl. image-guided radiotherapy)
   -   SRT – stereotaktična radioterapija (angl. stereotactic radiotherapy)
   -   SRS – stereotaktična kirurgija (angl. stereotactic radiosurgery)
   -   RAPIR ARC – volumetrično ločno obsevanje (angl. VMAT - volumetric arc therapy)




                                                                                          7
Infrared sistem                         Silicijeva slikovna detektorja




            Robotic 6D
                                 OBI – On
           couch – Exac
                                   board
               Trac
                                  Imager




                                                                 Room based x
                                                                 ray


                             Slika 6: LINAC NovalisTX – obsevalni aparat 8

                               VIR: http://www.jeffradonc.com/page.php?pID=21




3.1. IMRT – Intenziteno modulirana radioterapija


Intizitetno modulirano obsevanje – IMRT je sodobna obsevalna tehnika, ki zdruţuje vse
doseţke moderne računalnike tehnologije. V primerjavi z navadnimi konvencionalnimi
obsevalnimi tehnikami, lahko z IMRT doseţemo pomembno izboljšanje razporeditve doze v
obsevalnem tkivu. To se kaţe v večji konformnosti, ki jo dopolnjuje strm gradient doze na
robu tarčnega volumna, kar nam omogoča bolj natančno in usmerjeno obsevanje samega
tumorja in manjšo dozno obremenitev zdravih tkiv v okolici tumorja. S tem pripomoremo k
zmanjšanju verjetnosti pojavljanja in restnosti nezaţelenih stranskih učinkov (Strojan in sod.,
2010).


Poznamo več načinov IMRT tehnike, med najpogostejše sodita:
   -     dinamična tehnika (angl. sliding windows) deluje tako, da obsevanje traja ves čas,
         torej kontinuirano s pomočjo premikanja MLC lističev in s tem tudi pogojeno
         spreminjanje obsevalnega polja;


                                                                                             8
-   stopenjska tehnika (angl. step and shoot) deluje na principu »prestavi in sproţi«, kar
       pomeni, da se ţarek sproţi le, kadar obsevalno polje zavzame določeno obliko oz. se
       lističi kolimatorskega sistema zapeljejo v ţe naprej določeno leg.


Iz kliničnega vidika se IMRT od konvencionalnega 3D obsevanja razlikuje v tem, da
omogoča večjo konformnost, kar pomeni, da izodoze natančneje posnemajo 3D obliko
tarčnega volumna in konkavno oblikovanje obsevalnega polja. IMRT tehnika uporablja
računalniško vodeno modulacijo intenzitete s krmiljenjem večlistnih kolimatorjev (Strojan in
sod., 2010).


Prednosti IMRT tehnike pred ostalimi tehnikami:
   -   bolj homogena porazdelitev doze v tarči kot pri 3D konformalnem obsevanju,
   -   niţja dozna obremenitev izven PTV (planirnega terčnega volumna),
   -   omogoča oblikovanje pol z konkavnimi oblikami,
   -   omogoča kompenzacijo nehomogenosti polj.

Prikaz tehnike »sliding windows« IMRT: http://www.youtube.com/watch?v=Llxe9t8wwB0.




                 Slika 7: Primerjava dozne porazdelitve pri različnih obsevalnih tehnikah

                         VIR: IMRT obsevanje ORL bolnika (Kuduzović in Sekereš, 2010)



                                                                                            9
3.2. IGRT – Slikovno vodena radioterapija


Slikovno vodena radioterapija – IGRT nam omogoča izvajanje slikovno vodene metode
kvalitete obsevanja pred vsakim ali celo med samim obsevanjem. Namen IGRT tehnike je
zagotoviti, da je planirni tarčni volumen (PTV) vedno v enakem poloţaju med samim
obsevanjem (Varian Inc., 2011).


Z tehniko IGRT lahko predvsem zmanjšamo napake obsevanja zaradi:
   -   premikanja tarče med obsevanji (intra-fractional),
   -   premikanja tarče med obsevanjem (inter-fractional),
   -   napak pri pozicioniranju pacienta.


Radiološki inţenirji se pri delu posluţujemo različnih komponent obsevalnega aparata in
nekatere pomembne komponenete za izvajanje IGRT tehnike so:
   -   Exac Trac X-ray 6D robotic system,
   -   IR sistem,
   -   On-board Imager (OBI) kV sistem za cone-beam CT (CBCT),
   -   Portal Vision slikovni sistem.




                                Slika 8: Exac Trac 6D robotic IGRT couch sistem

          VIR: http://www.newswise.com/articles/new-image-guided-radiotherapy-system-benefits-high-risk-patients

                                                                                                                   10
Exac Trac 6D robotic IGRT couch se uporablja za uravnavanje nepravilnosti, ki nastanejo
pri vsakodnevnnem pozicioniranju pacienta v poloţaj izocentra. Sestavni deli Exac Tracta so:
dve vgrajeni rentgenski cevi v prostoru linearnega pospeševalnika, dva silicijeva detektorja in
IR sistem za detekcijo rentgenske slike in hkrati poloţaja pacienta.




       Slika 9: Prikaz poravnave z tehniko prekrivanja in deljenega polja pri slikanju z sistemom Exac Trac

               VIR: http://www.varian.com/media/oncology/resources/flash/obi_image_gallery/portal_vision.html




3.2.1 On-board Imager (OBI)


Uporaba On-board Imager (OBI) sistema omogoča bolj učinkovito in primerno dinamično
spremljanje obsevalne tarče – obsevalnega polja ter naslenje podrobnosti:
   -    pravilno pozicioniranje bolnika v primerjavi z izocentrom,
   -    verifikacijo obsevalnih polj v primerjavi z določenimi markerji in terapijo z
        obsevanjem,
   -    pozicioniranje obsevalnega poljna glede na implantirane zlate markerje ali kostne
        strukture,
   -    izboljšan nadzor nad tumorjem pred in med obsevanjem,
   -    zaznavanje sprememb pri premikanju notranjih organov in spremembe pri dihanju,




                                                                                                                11
-   zajete visoko kvalitetne slike omogočajo hitrejšo obravnavo in pregled anatomskih
       razlik,
   -   sprejemanje odločitev o morebitnih premikih izocentra pred obsevanjem,
   -   nizke kV dozne obremenitve omogočajo dnevno preverjanje obsevalnih polj.


S pomočjo OBI sistema in računalniških programov izvedemo korekcijo obsevalnih polj, ti.
»repozicioniranje bolnika«. Repozicioniranje lahko opravimo na dva načina in sicer na:
   a) kostne strukture
           -     omogočanje direktne določitve obsevalne tarče
           -     kostne strukure so zanesljiva orientacija za določitev obsevalne tarče
   b) implantirana zlata zrnca – markerje
           -     prilagoditev izocentra obsevalnega polja na določen implantirani zlati marker
                 (Varian Inc., 2011).



Prikaz IGRT, link: http://www.youtube.com/watch?v=8ySdWkx6m_c&feature=related.




3.3. RapidArc – volumetrično ločno obsevanje


Obsevalna tehnika RapidArc je omogočila velik napredek pri obsevanju bolnikov, saj
izboljšuje skladnost med odmerkom doze in skrajšanim časom obsevanja. RapidArc tehnika
deluje na princupu računalniških algoritmov, ki samodejno premikajo gantrij, kateri obkroţi
bolnika na obsevalni mizi za 360° in hkrati uravnavajo sekvenčno premikanje »sliding
window« MLC lističev in spreminjajočo se intenziteto sevanja (dose rate). Vse to omogoča
kontinuirano in učinkovitejše, ter predvsem hitrejše obsevanje za razliko od ostalih tehnik
(Varian Inc., 2011). RapidArc nam omogoča prenos celotne frakcionirane doze s tehniko
IMRT v eni sami rotaciji gantrija za 360° (Zankowski, 2007).




Glavne prednosti obsevalne tehnike RapidArc so:
   -   simultano gibanje gantrija okoli pacienta za 360°, kar omogoča optimalno
       porazdelitev doze v bolniku,

                                                                                           12
-   simulatano premikanje MLC-jev z načinom t.i. »sliding window«,
   -   nadzor nad hitrostjo premikanja gantrija,
   -   Varianov patentiran »gridded gun«, ki omogoča sprotno spreminjanje intenzitete
       izhodnega snopa iz obsevalnega aparata,
   -   minimalen čas obsevanja (navadno okoli 1-2 min) (Zankowski, 2007).


Prikaz Rapid Arc, link: http://www.youtube.com/watch?v=fMIW2SRTE1I&feature=related.




       Slika 10: Primerjava obsevalnih planov IMRT in RAPID ARC tehnike obsevanja
            VIR: http://www.varian.com/us/oncology/treatments/treatment_techniques/rapidarc/comparison.html




                                                                                                              13
4 LITARATURA

 -   Novaković in sod. (2009). Onkologija. Raziskovanje, diagnostika in zdravljenje raka.
     1. izd. Ljubljana: Mladinska knjiga; 120-154.


 -   Majerle M. (2003). Umerjanje linearnega pospeševalnika Philips SL-75/5 za potrebe
     3D načrtovanja v radioterapiji. Diplomsko delo. Univerza v Ljubljani. Fakuleteta za
     matamatiko in fiziko; 30-58


 -   Strojan in sod. (2010). Onkologija. Intenzitetno modulirano obsevanje (IMRT) zdaj
     tudi na Onkološkem inštututu v Ljubljani. Leto XIV. Onkološki Inštitut; 91-96


 -   Evans et al (2001). Performance assessment of the Gulmay D3300 kilovoltage X-ray
     therapy unit. The British Journal of Radiology. 74 (2001). The British Institute of
     Radiology; 537–547


 -   Zankowski C (2007). Varian’s NewRapidArc™ Delivery. The Next Dimension in
     Speed and Precision. Centerline: 1-5


 -   Advance radiation center of New York (2011). Image radiatoin radiotherapy and
     Rapid Arc technology.
     http://advancedradiationcenters.com/?p=whatisigrt <10.5.2011>.


 -   Varian Inc. (2011). Varian radiation oncology program solution.
     http://www.varian.com/us/oncology/radiation_oncology/ <14.5.2011>.




                                                                                      14

Mais conteúdo relacionado

Mais procurados

SRS & SBRT - Unflattened Beam
SRS & SBRT - Unflattened BeamSRS & SBRT - Unflattened Beam
SRS & SBRT - Unflattened BeamKothanda Raman
 
Gradient Recalled Echo(GRE)
Gradient Recalled Echo(GRE)Gradient Recalled Echo(GRE)
Gradient Recalled Echo(GRE)SUJAN KARKI
 
Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]radiation oncology
 
2D PLANNING IN BRAIN TUMOR
2D PLANNING IN BRAIN TUMOR2D PLANNING IN BRAIN TUMOR
2D PLANNING IN BRAIN TUMORKanhu Charan
 
Rrecent advances in linear accelerators [MR linac]
Rrecent advances in linear accelerators [MR linac]Rrecent advances in linear accelerators [MR linac]
Rrecent advances in linear accelerators [MR linac]Upasna Saxena
 
DEBATE ON HIPPOCAMPAL SPARING IN WHOLE BRAIN RADIATION IN BRAIN METS
DEBATE ON HIPPOCAMPAL SPARING IN WHOLE BRAIN RADIATION IN BRAIN METSDEBATE ON HIPPOCAMPAL SPARING IN WHOLE BRAIN RADIATION IN BRAIN METS
DEBATE ON HIPPOCAMPAL SPARING IN WHOLE BRAIN RADIATION IN BRAIN METSKanhu Charan
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapyjyotimannath
 
Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy Ravindra Shende
 
Icru 29,50 &62
Icru 29,50 &62Icru 29,50 &62
Icru 29,50 &62Dhiman Das
 
Advances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryAdvances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryMiami Cancer Institute
 
multiple filed arrangement in Radiotherapy, Medical College Kolkata
multiple filed arrangement in Radiotherapy, Medical College Kolkatamultiple filed arrangement in Radiotherapy, Medical College Kolkata
multiple filed arrangement in Radiotherapy, Medical College KolkataKazi Manir
 
Seminar 2 .beam modification devices
Seminar 2 .beam modification devicesSeminar 2 .beam modification devices
Seminar 2 .beam modification devicesDr ABDUL WAHEED DAR
 

Mais procurados (20)

SRS & SBRT - Unflattened Beam
SRS & SBRT - Unflattened BeamSRS & SBRT - Unflattened Beam
SRS & SBRT - Unflattened Beam
 
Gradient Recalled Echo(GRE)
Gradient Recalled Echo(GRE)Gradient Recalled Echo(GRE)
Gradient Recalled Echo(GRE)
 
Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]Arc therapy [autosaved] [autosaved]
Arc therapy [autosaved] [autosaved]
 
2D PLANNING IN BRAIN TUMOR
2D PLANNING IN BRAIN TUMOR2D PLANNING IN BRAIN TUMOR
2D PLANNING IN BRAIN TUMOR
 
Immobilizations devices
Immobilizations devicesImmobilizations devices
Immobilizations devices
 
Rrecent advances in linear accelerators [MR linac]
Rrecent advances in linear accelerators [MR linac]Rrecent advances in linear accelerators [MR linac]
Rrecent advances in linear accelerators [MR linac]
 
DEBATE ON HIPPOCAMPAL SPARING IN WHOLE BRAIN RADIATION IN BRAIN METS
DEBATE ON HIPPOCAMPAL SPARING IN WHOLE BRAIN RADIATION IN BRAIN METSDEBATE ON HIPPOCAMPAL SPARING IN WHOLE BRAIN RADIATION IN BRAIN METS
DEBATE ON HIPPOCAMPAL SPARING IN WHOLE BRAIN RADIATION IN BRAIN METS
 
Pet scanning.pptx
Pet scanning.pptxPet scanning.pptx
Pet scanning.pptx
 
Beam Direction
Beam DirectionBeam Direction
Beam Direction
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapy
 
Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy Intensity modulated radiation therapy and Image guided radiation therapy
Intensity modulated radiation therapy and Image guided radiation therapy
 
ICRU 71
ICRU 71ICRU 71
ICRU 71
 
Icru 29,50 &62
Icru 29,50 &62Icru 29,50 &62
Icru 29,50 &62
 
Dose volume histogram
Dose volume histogramDose volume histogram
Dose volume histogram
 
Advances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryAdvances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and Delivery
 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
 
multiple filed arrangement in Radiotherapy, Medical College Kolkata
multiple filed arrangement in Radiotherapy, Medical College Kolkatamultiple filed arrangement in Radiotherapy, Medical College Kolkata
multiple filed arrangement in Radiotherapy, Medical College Kolkata
 
Liver sbrt
Liver sbrtLiver sbrt
Liver sbrt
 
Image Guided Radiation Therapy (IGRT)
Image Guided Radiation Therapy (IGRT)Image Guided Radiation Therapy (IGRT)
Image Guided Radiation Therapy (IGRT)
 
Seminar 2 .beam modification devices
Seminar 2 .beam modification devicesSeminar 2 .beam modification devices
Seminar 2 .beam modification devices
 

Obsevalne tehnike na OI Ljubljana

  • 1. UNIVERZA V LJUBLJANI ZDRAVSTVENA FAKULTETA ODDELEK ZA RADIOLOŠKO TEHNOLOGIJO II. STOPNJE OBSEVALNE TEHNIKE NA TELERADIOTERAPEVTSKEM ODDELKU ONKOLOŠKEGA INŠTITUTA Avtor: Andrej Breznik, dipl.inţ.rad. Mentor: doc. dr. Dejan Ţontar, univ. dipl. fiz. LJUBLJANA, 2011
  • 2. KAZALO VSEBINE KAZALO SLIK Slika 1: Sestava linearnega pospeševalnika ............................................................................... 2 Slika 2: Nameščen tubus in individualna zaščita na obsevalnem aparatu .................................. 4 Slika 3: Porazdelitev globinske doze pri elektronih ................................................................... 4 Slika 4: Večlistni kolimator – MLC ........................................................................................... 5 Slika 5: Porazdelitev globinske doze pri fotonih ....................................................................... 6 Slika 6: LINAC NovalisTX – obsevalni aparat 8 ...................................................................... 8 Slika 7: Primerjava dozne porazdelitve pri različnih obsevalnih tehnikah ................................ 9 Slika 8: Exac Trac 6D robotic IGRT couch sistem .................................................................. 10 Slika 9: Prikaz poravnave z tehniko prekrivanja in deljenega polja pri slikanju z sistemom Exac Trac.................................................................................................................................. 11 Slika 10: Primerjava obsevalnih planov IMRT in RAPID ARC tehnike obsevanja ................ 13
  • 3. 1 KAJ JE RADIOTERAPIJA? Radioterapija je medicinska stroka, ki se ukvarja z zdravljenjem malignih in nemalignih bolezni s pomočjo ionizirajočega sevanja, lahko samostojno ali v kombinaciji z sistemskim in kirurškim zdravljenjem. Uvrščamo jo med lokalne načine zdravljenja, saj je njen učinek omejen le na obsevalno območje, kjer se absorbirajo ionizirajoči ţarki. Skupaj s kirurgijo in sistemskim zdravljenjem je eden izmed treh temeljnih načinov zdravljenja raka. Glede na poloţaj vira ionizirajočega sevanja v odnosu na obsevanega bolnika delimo radioterapijo v vsebinskem in organizacijskem delu na teleterapijo (vir je zunaj bolnikovega telesa) in brahiterapijo (vir je v bolnikovem telesu). Bolniki so lahko obsevani z fotonskimi ţarkovnimi snopi nizkih (kV) ali visokimi (MV) energijami in z elektronskimi ţarkovnimi snopi visokih (MeV) energij. Brahiterapevtska dejavnost se izvaja z napravami za naknadno polnjenje vodil (HDR in LDR), ki jih zdravnik povprej pod kontrolo različnih slikovnih metod uvede v telo bolnika, z viri ionizirajočega sevanja. Kot stroka radioterapija opredeljuje preplet kliničnih znanj in veščin poznavanja značilnosti malignih in neonkoloških bolezni, ter moţnosti njihovega zdravljenja. Prav tako zavzema področje radiofizike in radiobiologije, ki poznavata učinke ionizirajočega sevanja v bioloških sistemih in moţnostih njegove uporabe. Radioterapija vključuje tudi tesno sodelovanje strokovnjakov, kot so zdravnik specialist radioterapevt (nosilec dejavnosti), medicinski fizik – dozimetrist (odgovoren za pripravo obsevalnega načrta in umerjanje radioterapevtskih naprav) in inţenir radiološke tehnologije (upravljalec radioterapevtskih naprav in izvajalec obsevanja) (Novaković in sod., 2009). Med osnovne teleterapevtske naprave spadajo radioterapevtski simulator, terapevtska rentgenska naprava, linearni pospeševalnik in računalniški sistemi za načrtovanje obsevanja in naprave za preverjanje kakovosti obsevanja (Novaković in sod., 2009). 1
  • 4. 2 LINEARNI POSPEŠEVALNIK - LINAC V teleterapiji uporabljamo linearne pospeševalnike, ki lahko proizvajajo visokoenergijske fotone ali elektrone. Sodobni medicinski linearni pospeševalniki pospešujejo elektrone do kinetičnih energij od 4 MeV do 25 MeV (Novaković in sod., 2009). Linearni pospeševaliki uporabljajo RF mikrovalovna polja frekvenc od 103 MHz do 104 MHz. Velika večina jih deluje pri frekvenci mikrovalov 2856 MHz (S-band), kjer so tipične dolţine mikrovalovnih votlin okrog 10 cm v premeru in 3 cm v dolţini (Majerle, 2003). Izravnalna Bending magnet filtra in sipalne folije Pospeševalna cev Izvor elektornovov - GUN Tarča Primarni in sekundrani kolimator MLC Rtg cev - OBI Slikovni detektor EPID slikovni detekotor Slika 1: Sestava linearnega pospeševalnika VIR:http://www.google.si/imgres?imgurl=http://www.tradevision.com.bd/images/product/27.jpg&imgrefurl=http://www.tradevision.com.bd /productlist.php%3Fsubdivid%3D7%26subdivname%3DRadiotherapy&h=600&w=600&sz=71&tbnid=pTS2cx5lOJc0tM:&tbnh=135&tbnw =135&prev=/images%3Fq%3Dvarian%2Blinac&zoom=1&q=varian+linac&hl=sl&usg=__G5f4fyw7_ohDq65TSNfAwj8o5Ks=&sa=X&ei= jzV_TY7LGsTYsgaam8jsBg&sqi=2&ved=0CD4Q9QEwBA 2
  • 5. Princip delovanja linearnega pospeševalnika je takšen, da v pospeševalno cev injicirane elektrone z začetno energijo okoli 50 keV vzdolţ le-te pospešimo s pomočjo visokofrekvenčnega elektromagnetnega valovanja s frekvenco okoli 3000 MHz. V pospeševalni cevi je vakuum, ki zagotavlja čim manjše število trkov elektronov z molekulami plina in s tem čim manjše oviranje elektronov. Ob izhodu iz pospeševalne cevi dobimo ozek snop elektronov s premerom cca 3 mm, ki ga v nadaljevanju s pomočjo magnetnih polj usmerimo (odklonimo) v navpično smer. Elektronski curek lahko uporabljamo neposredno (obsevanje z elektroni) ali pa ga izkoristimo za nastanek rentgentskih fotonov z usmeritvijo elektronov v tarčo (Novaković in sod., 2009). V principu ločimo dve vrsti obsevanja, glede na vrsto ionizirajočega sevanja: - obsevanje z elektroni, - obsevanje s fotoni. 2.1. Obsevanje z elektronskim snopom V primeru obsevanja z elektronskim snopom, pospešeni elektroni trčijo ob vstavljeno tanko kovinsko sipalno folijo, katera ţarek razprši, hkrati pa poskrbi za homogenost obsevalnega polja, kar pomeni, da je količina elektronov po celotnem preseku obsevalnega polja enaka »dose flattnes«. Debeline sipalnih folij so takšne, da se elektroni v preteţni meri sipljejo in da ne prihaja do dodatnega zavornega sevanja. Izhodni snop elektronov omejujemo z t.i. aplikatorji oz. tubusi in individualnimi zaščitami iz Woodove zlitine, ki posnemajo obliko tarče. Tipične energije za obsevanje z elektorni so: 6, 9, 12, 15 in 18 MeV. Globinska porazdelitev doze pa je odvisna od jakosti energije. Pomembno je vedeti, da ima obsevanje z elektroni hiter padec izodozne krivulje, ko ta doseţe 80%. Na takšen način se lahko izognemo nepotrebnemu obsevanju t.i. kritičnih oz. zdravih organov. (Novaković in sod., 2009). 3
  • 6. Slika 2: Nameščen tubus in individualna zaščita na obsevalnem aparatu Vir: http://www.flickr.com/photos/vm1757/562061042/sizes/m/in/photostream/ 100,0 6 MeV 9 MeV 80,0 12 MeV PDD [%] 15 MeV 60,0 18 MeV 40,0 20,0 0,0 0 2 4 6 8 10 12 globina [cm] Slika 3: Porazdelitev globinske doze pri elektronih Slika 3 prikazuje razmerje med globino in odstotkom globinske doze (angl. Percent depth dose – PDD) pri elektronskem snopu. 4
  • 7. 2.2. Obsevanje s fotonskim snopom Fotonsko obsevanje poteka po postopku, da elektroni na poti iz pospeševalne cevi trčijo na vstavljeno tarčo, ki je snov z visokim vrstnim številom Z. V tarči prihaja do zavornega sevanja in rezultat je nastanek visoko energijskih fotonov primernih za fotonsko obsevanje. V nadaljevanju fotoni zadanejo izravnalni filter, ki poskrbi, da je ţarek po prehodu skozenj po svojem preseku čim bolj homogen oz. izravnan. Za dodatno omejitev fotonskih ţarkov skrbi serija zaslonk, ki ţarek še dodatno oblikujejo. Zadnji element, ki omogoča skoraj poljubno oblikovanje obsevalnih poljm je t.i. večlistni komatorski sistem MLC (angl. MultiLeaf Collimator), s katerim oblikujemo fotonska obsevalna polja tako, da ta povzamejo obliko obolenja, kot ga vidimo iz smeri ţarkov. Sodobni pospeševalniki imajo vgrajen tudi EPID – elektronski portalni sistem za verifikacijo obsevalnih polj. Pri fotonskem načinu obsevanja se posluţujemo energijama 6 MV in 15 MV (Novaković in sod., 2009). Slika 4: Večlistni kolimator – MLC Vir: http://varian.mediaroom.com/file.php/301/MLC+-+gold.jpg 5
  • 8. 100 6 MV 15 MV 80 Kobalt PDD [%] 60 40 20 0 0 10 20 30 40 globina [cm] Slika 5: Porazdelitev globinske doze pri fotonih Slika 5 prikazuje razmerje med globino in odstotkom globinske doze (angl. Percent depth dose – PDD) pri fotonskem snopu. 6
  • 9. 3 NOVE OBSEVALNE TEHNIKE NA OI Vključitev novih obsevalnih tehnik v delovno rutino je pogojena z uporabo CT simulatorja in sodobne podprte računalniške programe, ki s pomočjo algoritmov omogoča izdelavo natančnega obsevalnega načrta. Z razvojem računalniške podprte tehnologije je sledil tudi razvoj popolnega računalniškega nadzora oz. krmiljenje linearnih pospeševalnikov, večlistnih kolimatorjev za oblikovanje obsevalnega polja in računalniško podprtega sistema za načrtovanje obsevanja (Novaković in sod., 2009). Prednosti novih obsevalnih tehnik omogoča: - povečanje gradienta med visokodoznim in nizkodoznim območjem, - povečanje konformnosti visokodoznega območja (skladnost s 3D obliko tarče), - natančnejše obsevanje (zmanjšanje odmika lege posameznega polja od referenčne). Na Onkološkem inštitutu v Ljubljani smo z nakupom novejših obsevalnih aparatov pričeli obsevati naslednje različne obsevalne tehnike: - IMRT – intizitetno modulirana radioterapija (angl. Intensity-modulated radiation therapy) - IGRT – slikovno vodena radioterapija (angl. image-guided radiotherapy) - SRT – stereotaktična radioterapija (angl. stereotactic radiotherapy) - SRS – stereotaktična kirurgija (angl. stereotactic radiosurgery) - RAPIR ARC – volumetrično ločno obsevanje (angl. VMAT - volumetric arc therapy) 7
  • 10. Infrared sistem Silicijeva slikovna detektorja Robotic 6D OBI – On couch – Exac board Trac Imager Room based x ray Slika 6: LINAC NovalisTX – obsevalni aparat 8 VIR: http://www.jeffradonc.com/page.php?pID=21 3.1. IMRT – Intenziteno modulirana radioterapija Intizitetno modulirano obsevanje – IMRT je sodobna obsevalna tehnika, ki zdruţuje vse doseţke moderne računalnike tehnologije. V primerjavi z navadnimi konvencionalnimi obsevalnimi tehnikami, lahko z IMRT doseţemo pomembno izboljšanje razporeditve doze v obsevalnem tkivu. To se kaţe v večji konformnosti, ki jo dopolnjuje strm gradient doze na robu tarčnega volumna, kar nam omogoča bolj natančno in usmerjeno obsevanje samega tumorja in manjšo dozno obremenitev zdravih tkiv v okolici tumorja. S tem pripomoremo k zmanjšanju verjetnosti pojavljanja in restnosti nezaţelenih stranskih učinkov (Strojan in sod., 2010). Poznamo več načinov IMRT tehnike, med najpogostejše sodita: - dinamična tehnika (angl. sliding windows) deluje tako, da obsevanje traja ves čas, torej kontinuirano s pomočjo premikanja MLC lističev in s tem tudi pogojeno spreminjanje obsevalnega polja; 8
  • 11. - stopenjska tehnika (angl. step and shoot) deluje na principu »prestavi in sproţi«, kar pomeni, da se ţarek sproţi le, kadar obsevalno polje zavzame določeno obliko oz. se lističi kolimatorskega sistema zapeljejo v ţe naprej določeno leg. Iz kliničnega vidika se IMRT od konvencionalnega 3D obsevanja razlikuje v tem, da omogoča večjo konformnost, kar pomeni, da izodoze natančneje posnemajo 3D obliko tarčnega volumna in konkavno oblikovanje obsevalnega polja. IMRT tehnika uporablja računalniško vodeno modulacijo intenzitete s krmiljenjem večlistnih kolimatorjev (Strojan in sod., 2010). Prednosti IMRT tehnike pred ostalimi tehnikami: - bolj homogena porazdelitev doze v tarči kot pri 3D konformalnem obsevanju, - niţja dozna obremenitev izven PTV (planirnega terčnega volumna), - omogoča oblikovanje pol z konkavnimi oblikami, - omogoča kompenzacijo nehomogenosti polj. Prikaz tehnike »sliding windows« IMRT: http://www.youtube.com/watch?v=Llxe9t8wwB0. Slika 7: Primerjava dozne porazdelitve pri različnih obsevalnih tehnikah VIR: IMRT obsevanje ORL bolnika (Kuduzović in Sekereš, 2010) 9
  • 12. 3.2. IGRT – Slikovno vodena radioterapija Slikovno vodena radioterapija – IGRT nam omogoča izvajanje slikovno vodene metode kvalitete obsevanja pred vsakim ali celo med samim obsevanjem. Namen IGRT tehnike je zagotoviti, da je planirni tarčni volumen (PTV) vedno v enakem poloţaju med samim obsevanjem (Varian Inc., 2011). Z tehniko IGRT lahko predvsem zmanjšamo napake obsevanja zaradi: - premikanja tarče med obsevanji (intra-fractional), - premikanja tarče med obsevanjem (inter-fractional), - napak pri pozicioniranju pacienta. Radiološki inţenirji se pri delu posluţujemo različnih komponent obsevalnega aparata in nekatere pomembne komponenete za izvajanje IGRT tehnike so: - Exac Trac X-ray 6D robotic system, - IR sistem, - On-board Imager (OBI) kV sistem za cone-beam CT (CBCT), - Portal Vision slikovni sistem. Slika 8: Exac Trac 6D robotic IGRT couch sistem VIR: http://www.newswise.com/articles/new-image-guided-radiotherapy-system-benefits-high-risk-patients 10
  • 13. Exac Trac 6D robotic IGRT couch se uporablja za uravnavanje nepravilnosti, ki nastanejo pri vsakodnevnnem pozicioniranju pacienta v poloţaj izocentra. Sestavni deli Exac Tracta so: dve vgrajeni rentgenski cevi v prostoru linearnega pospeševalnika, dva silicijeva detektorja in IR sistem za detekcijo rentgenske slike in hkrati poloţaja pacienta. Slika 9: Prikaz poravnave z tehniko prekrivanja in deljenega polja pri slikanju z sistemom Exac Trac VIR: http://www.varian.com/media/oncology/resources/flash/obi_image_gallery/portal_vision.html 3.2.1 On-board Imager (OBI) Uporaba On-board Imager (OBI) sistema omogoča bolj učinkovito in primerno dinamično spremljanje obsevalne tarče – obsevalnega polja ter naslenje podrobnosti: - pravilno pozicioniranje bolnika v primerjavi z izocentrom, - verifikacijo obsevalnih polj v primerjavi z določenimi markerji in terapijo z obsevanjem, - pozicioniranje obsevalnega poljna glede na implantirane zlate markerje ali kostne strukture, - izboljšan nadzor nad tumorjem pred in med obsevanjem, - zaznavanje sprememb pri premikanju notranjih organov in spremembe pri dihanju, 11
  • 14. - zajete visoko kvalitetne slike omogočajo hitrejšo obravnavo in pregled anatomskih razlik, - sprejemanje odločitev o morebitnih premikih izocentra pred obsevanjem, - nizke kV dozne obremenitve omogočajo dnevno preverjanje obsevalnih polj. S pomočjo OBI sistema in računalniških programov izvedemo korekcijo obsevalnih polj, ti. »repozicioniranje bolnika«. Repozicioniranje lahko opravimo na dva načina in sicer na: a) kostne strukture - omogočanje direktne določitve obsevalne tarče - kostne strukure so zanesljiva orientacija za določitev obsevalne tarče b) implantirana zlata zrnca – markerje - prilagoditev izocentra obsevalnega polja na določen implantirani zlati marker (Varian Inc., 2011). Prikaz IGRT, link: http://www.youtube.com/watch?v=8ySdWkx6m_c&feature=related. 3.3. RapidArc – volumetrično ločno obsevanje Obsevalna tehnika RapidArc je omogočila velik napredek pri obsevanju bolnikov, saj izboljšuje skladnost med odmerkom doze in skrajšanim časom obsevanja. RapidArc tehnika deluje na princupu računalniških algoritmov, ki samodejno premikajo gantrij, kateri obkroţi bolnika na obsevalni mizi za 360° in hkrati uravnavajo sekvenčno premikanje »sliding window« MLC lističev in spreminjajočo se intenziteto sevanja (dose rate). Vse to omogoča kontinuirano in učinkovitejše, ter predvsem hitrejše obsevanje za razliko od ostalih tehnik (Varian Inc., 2011). RapidArc nam omogoča prenos celotne frakcionirane doze s tehniko IMRT v eni sami rotaciji gantrija za 360° (Zankowski, 2007). Glavne prednosti obsevalne tehnike RapidArc so: - simultano gibanje gantrija okoli pacienta za 360°, kar omogoča optimalno porazdelitev doze v bolniku, 12
  • 15. - simulatano premikanje MLC-jev z načinom t.i. »sliding window«, - nadzor nad hitrostjo premikanja gantrija, - Varianov patentiran »gridded gun«, ki omogoča sprotno spreminjanje intenzitete izhodnega snopa iz obsevalnega aparata, - minimalen čas obsevanja (navadno okoli 1-2 min) (Zankowski, 2007). Prikaz Rapid Arc, link: http://www.youtube.com/watch?v=fMIW2SRTE1I&feature=related. Slika 10: Primerjava obsevalnih planov IMRT in RAPID ARC tehnike obsevanja VIR: http://www.varian.com/us/oncology/treatments/treatment_techniques/rapidarc/comparison.html 13
  • 16. 4 LITARATURA - Novaković in sod. (2009). Onkologija. Raziskovanje, diagnostika in zdravljenje raka. 1. izd. Ljubljana: Mladinska knjiga; 120-154. - Majerle M. (2003). Umerjanje linearnega pospeševalnika Philips SL-75/5 za potrebe 3D načrtovanja v radioterapiji. Diplomsko delo. Univerza v Ljubljani. Fakuleteta za matamatiko in fiziko; 30-58 - Strojan in sod. (2010). Onkologija. Intenzitetno modulirano obsevanje (IMRT) zdaj tudi na Onkološkem inštututu v Ljubljani. Leto XIV. Onkološki Inštitut; 91-96 - Evans et al (2001). Performance assessment of the Gulmay D3300 kilovoltage X-ray therapy unit. The British Journal of Radiology. 74 (2001). The British Institute of Radiology; 537–547 - Zankowski C (2007). Varian’s NewRapidArc™ Delivery. The Next Dimension in Speed and Precision. Centerline: 1-5 - Advance radiation center of New York (2011). Image radiatoin radiotherapy and Rapid Arc technology. http://advancedradiationcenters.com/?p=whatisigrt <10.5.2011>. - Varian Inc. (2011). Varian radiation oncology program solution. http://www.varian.com/us/oncology/radiation_oncology/ <14.5.2011>. 14