SlideShare uma empresa Scribd logo
1 de 257
Fundamentals of Instrumentation
& Process Control

Interactive Training Workshop
Introduction to Process Control

       A common misconception in process control is that it is all
       about the controller – that you can force a particular
       process response just by getting the right tuning
       parameters.
       In reality, the controller is just a partner. A process will
       respond to a controller’s commands only in the manner
       which it can. To understand process control you must
       understand the other partners as well: sensors, final
       control elements and the process itself.
       All of these determine what type of response the controller
       is capable of extracting out of the process. It is not the
       other way around.




Copyright © 2007 Control Station, Inc. All Rights Reserved   2
Outline of the Course




Copyright © 2007 Control Station, Inc. All Rights Reserved   3
Introduction to Process Control

       Objectives:
            Why do we need process control?
            What is a process?
            What is process control?
            What is open loop control?
            What is closed loop control?
            What are the modes of closed loop control?
            What are the basic elements of process control?




Copyright © 2007 Control Station, Inc. All Rights Reserved   4
Motivation for Automatic Process Control

                        Safety First:
                             people, environment, equipment


                        The Profit Motive:
                             meeting final product specs
                             minimizing waste production
                             minimizing environmental impact
                             minimizing energy use
                             maximizing overall production rate




Copyright © 2007 Control Station, Inc. All Rights Reserved   5
“Loose” Control Costs Money

                                       65                          operating constraint
                         PV & SP (%)   60                                                                         set point far from
                                                                                                                  profit constraint
                                       55

                                       SP
           more profit




                                       45

                                                                                            Copyright © 2007
                                       40   poor control = large variation in PV       by Control Station, Inc.
                                                                                          All Rights Reserved



                                                                   Time

           It takes more processing to remove impurities, so greatest
           profit is to operate as close to the maximum impurities
           constraint as possible without going over

           It takes more material to make a product thicker, so greatest
           profit is to operate as close to the minimum thickness
           constraint as possible without going under
Copyright © 2007 Control Station, Inc. All Rights Reserved                         6
“Tight” Control is More Profitable

                                       65                              operating constraint                              set point near
                                       60                                                                                profit constraint
                         PV & SP (%)
                                       SP

                                       50
           more profit




                                            tight control = small variation in PV
                                       45

                                       40
                                                         Copyright © 2007 by Control Station, Inc. All Rights Reserved



                                                                      Time


           A well controlled process has less variability in the measured
           process variable (PV), so the process can be operated close
           to the maximum profit constraint.




Copyright © 2007 Control Station, Inc. All Rights Reserved                              7
Terminology for Home Heating Control

         Control Objective
         Measured Process Variable (PV)
         Set Point (SP)
         Controller Output (CO)
         Manipulated Variable
         Disturbances (D)       thermostat
                                                  controller
                                     set point       TC                          TT
                                                                      temperature                                                heat loss
                                                                    sensor/transmitter                                         (disturbance)

                                             control
                                             signal



                                      fuel flow                               furnace
                                                    valve

                                                               Copyright © 2007 by Control Station, Inc. All Rights Reserved




Copyright © 2007 Control Station, Inc. All Rights Reserved              8
What is Process Control

       Terminology:
            The manipulated variable (MV) is a measure of resource
            being fed into the process, for instance how much thermal
            energy.
            A final control element (FCE) is the device that changes
            the value of the manipulated variable.
            The controller output (CO) is the signal from the
            controller to the final control element.
            The process variable (PV) is a measure of the process
            output that changes in response to changes in the
            manipulated variable.
            The set point (SP) is the value at which we wish to
            maintain the process variable at.



Copyright © 2007 Control Station, Inc. All Rights Reserved   9
What is a Process

       A process is broadly defined as an operation
       that uses resources to transform inputs into
       outputs.
       It is the resource that provides the energy into
       the process for the transformation to occur.




Copyright © 2007 Control Station, Inc. All Rights Reserved   10
What is a Process




Copyright © 2007 Control Station, Inc. All Rights Reserved   11
What is a Process

       Each process exhibits a particular dynamic (time
       varying) behavior that governs the
       transformation.
            That is, how do changes in the resource or inputs
            over time affect the transformation.
       This dynamic behavior is determined by the
       physical properties of the inputs, the resource,
       and the process itself.




Copyright © 2007 Control Station, Inc. All Rights Reserved   12
What is a Process

       Can you identify some of the elements that will
       determine the dynamic properties of this
       process?




Copyright © 2007 Control Station, Inc. All Rights Reserved   13
What is Process Control

       Process control is the act of controlling a final
       control element to change the manipulated
       variable to maintain the process variable at a
       desired set point.
            A corollary to our definition of process control is a
            controllable process must behave in a predictable
            manner.
            For a given change in the manipulated variable, the
            process variable must respond in a predictable and
            consistent manner.




Copyright © 2007 Control Station, Inc. All Rights Reserved   14
What is Process Control




Copyright © 2007 Control Station, Inc. All Rights Reserved   15
What is Process Control




Copyright © 2007 Control Station, Inc. All Rights Reserved   16
Section Assessment:
                   Basic Terminology Assessment




Copyright © 2007 Control Station, Inc. All Rights Reserved   17
What is Open Loop Control

       In open loop control the controller output is not
       a function of the process variable.
            In open loop control we are not concerned that a
            particular set point be maintained, the controller
            output is fixed at a value until it is changed by an
            operator.
            Many processes are stable in an open loop control
            mode and will maintain the process variable at a
            value in the absence of a disturbances.
            Disturbances are uncontrolled changes in the
            process inputs or resources.




Copyright © 2007 Control Station, Inc. All Rights Reserved   18
What is Open Loop Control




Copyright © 2007 Control Station, Inc. All Rights Reserved   19
What is Open Loop Control

       Can you think of processes in which open loop
       control is sufficient?




Copyright © 2007 Control Station, Inc. All Rights Reserved   20
What is Closed Loop Control

       In closed loop control the controller output is
       determined by difference between the process
       variable and the set point. Closed loop control is
       also called feedback or regulatory control.
            The output of a closed loop controller is a function of
            the error.
            Error is the deviation of the process variable from
            the set point and is defined as
            E = SP - PV.




Copyright © 2007 Control Station, Inc. All Rights Reserved   21
What is Closed Loop Control




Copyright © 2007 Control Station, Inc. All Rights Reserved   22
What is Closed Loop Control

       From the controllers
       perspective the
       process encompasses
       the RTD, the steam
       control valve, and
       signal processing
       of the PV and
       CO values.




Copyright © 2007 Control Station, Inc. All Rights Reserved   23
What are the Modes of Closed Loop Control

       Closed loop control can be, depending on the
       algorithm that determines the controller output:
            Manual
            On-Off
            PID
            Advanced PID (ratio, cascade, feedforward)
            or Model Based




Copyright © 2007 Control Station, Inc. All Rights Reserved   24
What are the Modes of Closed Loop Control
   Manual Control

       In manual control an operator directly
       manipulates the controller output to the final
       control element to maintain a set point.




Copyright © 2007 Control Station, Inc. All Rights Reserved   25
What are the Modes of Closed Loop Control
   On-Off Control

       On-Off control provides a controller output of
       either on or off in response to error.




Copyright © 2007 Control Station, Inc. All Rights Reserved   26
What are the Modes of Closed Loop Control
   On-Off Control Deadband

       Upon changing the direction of the controller
       output, deadband is the value that must be
       traversed before the controller output will change
       its direction again.




Copyright © 2007 Control Station, Inc. All Rights Reserved   27
What are the Modes of Closed Loop Control
   PID Control

       PID control provides a controller output that
       modulates from 0 to 100% in response to error.




Copyright © 2007 Control Station, Inc. All Rights Reserved   28
What are the Modes of Closed Loop Control
   Time Proportion Control
       Time proportion control is a variant of PID control that
       modulates the on-off time of a final control element that only
       has two command positions.
            To achieve the effect of PID
            control the switching frequency
            of the device is modulated in
            response to error. This is
            achieved by introducing the
            concept of cycle time.


            Cycle Time is the time base of the
           signal the final control element will
          receive from the controller. The PID
          controller determines the final signal
           to the controller by multiplying the
           cycle time by the output of the PID
                         algorithm.




Copyright © 2007 Control Station, Inc. All Rights Reserved   29
What are the Modes of Closed Loop Control
   Cascade Control

       Cascade control uses the output of a primary (master or
       outer) controller to manipulate the set point of a secondary
       (slave or inner) controller as if the slave controller were the
       final control element.




Copyright © 2007 Control Station, Inc. All Rights Reserved   30
What are the Modes of Closed Loop Control
   Cascade Control




Copyright © 2007 Control Station, Inc. All Rights Reserved   31
Basic Elements of Process Control

       Controlling a process requires knowledge of four
       basic elements, the process itself, the sensor
       that measures the process value, the final
       control element that changes the manipulated
       variable, and the controller.




Copyright © 2007 Control Station, Inc. All Rights Reserved   32
Section Assessment:
                   Basic Process Control Assessment




Copyright © 2007 Control Station, Inc. All Rights Reserved   33
Understanding Dynamic Process Behavior


            What We Will Learn in This Section
                Dynamic Process Behavior – What It Is & Why We Care
                What a FOPDT Dynamic Model Represents
                Analyzing Step Test Plot Data to Determine FOPDT
                Dynamic Model Parameters
                Process Gain, Time Constant & Dead Time
                   How To Compute Them From Plot Data
                   How to Use Them For Controller Design and Tuning
                How to Recognize Nonlinear Processes




Copyright © 2007 Control Station, Inc. All Rights Reserved   34
Dynamic Process Behavior and Controller Tuning

              Consider cruise control for a car vs a truck
                 how quickly can each accelerate or decelerate
                 what is the effect of disturbances (wind, hills, etc.)

              Controller (gas flow) manipulations required to maintain
              set point velocity in spite of disturbances (wind, hills)
              are different for a car and truck because the dynamic
              behavior of each "process" is different

              Dynamic behavior → how the measured process variable
              (PV) responds over time to changes in the controller
              output (CO) and disturbances (D)




Copyright © 2007 Control Station, Inc. All Rights Reserved   35
Graphical Modeling of Dynamic Process Data

            To learn about the dynamic behavior of a process, we
            analyze measured process variable (PV) test data

            PV test data can be generated by suddenly changing the
            controller output (CO) signal

            The CO should be moved far and fast enough so that the
            dynamic behavior is clearly revealed as the PV responds

            The dynamic behavior of a process is different as operating
            level changes (nonlinear behavior), so collect data at
            normal operating conditions (design level of operation)




Copyright © 2007 Control Station, Inc. All Rights Reserved   36
Modeling Dynamic Process Behavior

          The best way to understand process data is through modeling

          Modeling means fitting a first order plus dead time (FOPDT)
          dynamic model to the process data:
                          dPV
                    τp             + PV = Kp ×CO(tθp)
                                                   −
                           dt


                where:
                  PV is the measured process variable
                  CO is the controller output signal

          The FOPDT model is simple (low order and linear) so it only
          approximates the behavior of real processes



Copyright © 2007 Control Station, Inc. All Rights Reserved   37
Modeling Dynamic Process Behavior


                   When a first order plus dead time (FOPDT) model
                   is fit to dynamic process data

                              dPV
                        τp             + PV = Kp ×CO(tθp)
                                                       −
                               dt

                   The important parameters that result are:
                      Steady State Process Gain, Kp
                      Overall Process Time Constant, τ p
                      Apparent Dead Time, Өp




Copyright © 2007 Control Station, Inc. All Rights Reserved   38
The FOPDT Model is All Important

          FOPDT model parameters (Kp, τ p and Өp) are used in
          correlations to compute controller tuning values

          Sign of Kp indicates the action of the controller
                (+ Kp ⇒ reverse acting; — Kp ⇒ direct acting)

          Size of τ p indicates the maximum desirable loop sample time
                (be sure sample time T ≤ 0.1τ p )

          Ratio of Өp/τ p indicates whether model predictive control
          such as a Smith predictor would show benefit
                (useful if Өp >τ p )

          Model becomes part of the feed forward, Smith Predictor,
          decoupling and other model-based controllers




Copyright © 2007 Control Station, Inc. All Rights Reserved   39
Step Test Data and Dynamic Process Modeling
                              Manual Mode Step Test on Gravity Drained Tanks




                                                       Copyright © 2007 by Control Station, Inc. All Rights Reserved




                Process starts at steady state in manual mode
                Controller output (CO) signal is stepped to new value
                Process variable (PV) signal must complete the response

Copyright © 2007 Control Station, Inc. All Rights Reserved                40
Process Gain (Kp) from Step Test Data

             Kp describes how far the measured PV travels in
             response to a change in the CO
             A step test starts and ends at steady state, so Kp can
             be computed directly from the plot

                      Steady State Change in the Process Variable,ΔPV
            Kp =
                     Steady State Change in the Controller Output, ΔCO

             where ∆PV and ∆CO are the total change from initial to
             final steady state

             A large process gain, Kp, means that each CO action
             will produce a large PV response




Copyright © 2007 Control Station, Inc. All Rights Reserved   41
Process Gain (Kp) for Gravity-Drained Tanks



                                                                        ∆PV = (2.9 – 1.9) = 1.0 m




                                                                           ∆CO = (60 – 50) = 10 %


                                                      Copyright © 2007 by Control Station, Inc. All Rights Reserved




                    Compute ∆PV and ∆CO as “final minus initial”
                    steady state values

Copyright © 2007 Control Station, Inc. All Rights Reserved              42
Process Gain (Kp) for Gravity-Drained Tanks



                                                                                            ∆PV = 1.0 m



                                                 ∆PV         1.0 m               m
                                          Kp = –––– = ––––– = 0.1 –––
                                                 ∆CO          10%                %



                                                                                           ∆CO = 10 %


                                                       Copyright © 2007 by Control Station, Inc. All Rights Reserved




            Kp has a size (0.1); a sign (+0.1), and units (m/%)


Copyright © 2007 Control Station, Inc. All Rights Reserved               43
Additional Notes on Process Gain
   Measuring the Process Gain

       Process gain as seen by a controller is the
       product of the gains of the sensor, the final
       control element and the process itself.
        Process Gain = Gain Process x Gain Sensor x Gain Final Control Element


       The gain of a controller will be inversely
       proportional to the process gain that it sees
                                      1
               Controller Gain ∝
                                 Process Gain




Copyright © 2007 Control Station, Inc. All Rights Reserved   44
Additional Notes on Process Gain
   Converting Units of Process Gain

       There is one important caveat in this process; the
       gain we have calculated has units of m/%.
       Real world controllers, unlike most software
       simulations, have gain units specified as %/%.
       When calculating the gain for a real controller the
       change in PV needs to be expressed in percent of
       span of the PV as this is how the controller
       calculates error.

                                                 ∆PV             COSpan
                                    Gain =                   x
                                                 ∆CO             PVSpan


Copyright © 2007 Control Station, Inc. All Rights Reserved         45
Additional Notes on Process Gain
   Converting Units of Process Gain

      Assuming the process gain of the gravity
      drained tanks is 0.1 m/% then, to convert
      to %/%

            m span CO       m 0 − 100%        %
        0 .1 ⋅        = 0 .1 ⋅         = 1 .0
            % span PV       % 0 − 10%         %




Copyright © 2007 Control Station, Inc. All Rights Reserved   46
Additional Notes on Process Gain
   Values for Process Gain

       Process gain that the controller sees is influenced
       by two factors other than the process itself, the
       size of the final control element and the span of
       the sensor.
       In the ideal world you would use the full span of
       both final control element and the sensor which
       would give a process gain of 1.0.




Copyright © 2007 Control Station, Inc. All Rights Reserved   47
Additional Notes on Process Gain
   Values for Process Gain

       As a rule of thumb:
            Process gains that are greater than 1 are a result of
            oversized final control elements.
            Process gains less than 1 are a result of sensor spans
            that are too wide.




Copyright © 2007 Control Station, Inc. All Rights Reserved   48
Additional Notes on Process Gain
   Values for Process Gain

       The result of a final control element being too large (high
       gain) is:
            The controller gain will have to be made correspondingly smaller,
            smaller than the controller may accept.
            High gains in the final control element amplify imperfections
            (deadband, stiction), control errors become proportionately
            larger.
       If a sensor has too wide of a span:
            You may experience problems with the quality of the
            measurement.
            The controller gain will have to be made correspondingly larger
            making the controller more jumpy and amplifying signal noise.
            An over spanned sensor can hide an oversized final element.




Copyright © 2007 Control Station, Inc. All Rights Reserved   49
Additional Notes on Process Gain
   Values for Process Gain




             The general rule of thumb is the
             process gain for a self regulating
               process should be between
                      0.5 and 2.0.




Copyright © 2007 Control Station, Inc. All Rights Reserved   50
Process Time Constant τ p ) from Step Test Data
                         (




                                                      Copyright © 2007 by Control Station, Inc. All Rights Reserved




           Time Constant, τ p , describes how fast the measured PV
           responds to changes in the CO
           More specifically, how long it takes the PV to reach 63.2%
           of its total final change (∆PV), starting from when it first
           begins to respond


Copyright © 2007 Control Station, Inc. All Rights Reserved                  51
Process Time Constant τ p ) from Step Test Data
                         (




                                                        Copyright © 2007 by Control Station, Inc. All Rights Reserved




                                  t PVstart

                   1) Locate t PVstart , the time where the PV starts a
                        first clear response to the step change in CO

Copyright © 2007 Control Station, Inc. All Rights Reserved             52
Process Time Constant τ p ) from Step Test Data
                         (


                                  PV63.2
                                                                                                ∆PV total
                                                 63.2% of ∆PV total




                                                        Copyright © 2007 by Control Station, Inc. All Rights Reserved




                                  t PVstart

                   2) Compute 63.2% of the total change in PV as:
                                    PV63.2 = PVinital + 0.632 ∆PV

Copyright © 2007 Control Station, Inc. All Rights Reserved             53
Process Time Constant τ p ) from Step Test Data
                         (


                                  PV63.2




                                             τP
                                                           Copyright © 2007 by Control Station, Inc. All Rights Reserved




                                 t PVstart        t 63.2

                      3) t 63.2 is the time when the PV reaches PV63.2
                      4) Time Constant, τ p , is then: t 63.2 — tPVstart

Copyright © 2007 Control Station, Inc. All Rights Reserved                54
Time Constant (τ p) for Gravity-Drained Tanks




                                                             Copyright © 2007 by Control Station, Inc. All Rights Reserved




                                tPVstart = 4.1

                                      Here, tPVstart = 4.1 min


Copyright © 2007 Control Station, Inc. All Rights Reserved                55
Time Constant (τ p ) for Gravity-Drained Tanks


                              PV63.2 = 2.5
                                                                                                  ∆PV = 1.0 m
                                                   63.2% of ∆PV total




                                                             Copyright © 2007 by Control Station, Inc. All Rights Reserved




                                      4.1

                                 PV63.2 = PVinital + 0.632 ∆PV

                                            = 1.9 + 0.632(1.0) = 2.5 m
Copyright © 2007 Control Station, Inc. All Rights Reserved                56
Time Constant (τ p ) for Gravity-Drained Tanks


                              PV63.2 = 2.5




                                             τP                        τP = 5.7 – 4.1 = 1.6 min
                                                             Copyright © 2007 by Control Station, Inc. All Rights Reserved




                                      4.1            5.7


                      Time Constant, τ p = t 63.2 — tPVstart = 1.6 min
                      τ p must be positive and have units of time
Copyright © 2007 Control Station, Inc. All Rights Reserved                57
Process Dead-Time (Өp) from Step Test Data
           Dead time, Өp, is how much delay occurs from the time when
           the CO step is made until when the measured PV shows a
           first clear response.

           Өp is the sum of these effects:
                 transportation lag, or the time it takes for material to
                 travel from one point to another
                 sample or instrument lag, or the time it takes to collect,
                 analyze or process a measured PV sample
                 higher order processes naturally appear slow to respond
                 and this is treated as dead time


           Dead time, Өp, must be positive and have units of time




Copyright © 2007 Control Station, Inc. All Rights Reserved   58
Dead-Time (Өp) is the Killer of Control

          Tight control grows increasingly difficult as Өp becomes large
          The process time constant is the clock of the process. Dead
          time is large or small relative only to τ p
          When dead time grows such that Өp > τ p , model predictive
          control strategies such as a Smith predictor may show benefit
          For important PVs, work to select, locate and maintain
          instrumentation so as to avoid unnecessary dead time in a loop




Copyright © 2007 Control Station, Inc. All Rights Reserved   59
Process Dead-Time (Өp) from Step Test Data




                                                       Copyright © 2007 by Control Station, Inc. All Rights Reserved




                                            Өp
                           tCOstep
                                     t PVstart

                           Өp = t PVstart        — tCOstep




Copyright © 2007 Control Station, Inc. All Rights Reserved               60
Dead-Time (Өp) for Gravity Drained Tanks




                                                Өp = 4.1 – 3.8 = 0.3 min

                                                       Copyright © 2007 by Control Station, Inc. All Rights Reserved




                    tCOstep = 3.8
                               t PVstart= 4.1

                        Өp = t PVstart      — tCOstep    = 0.3 min


Copyright © 2007 Control Station, Inc. All Rights Reserved               61
The FOPDT Model Parameters

                                               In Summary

        For a change in CO:

              Process Gain, Kp → How Far PV travels

              Time Constant, τ p → How Fast PV responds

              Dead Time, Өp                   → How Much Delay Before PV Responds




Copyright © 2007 Control Station, Inc. All Rights Reserved   62
Hands-On Workshop




           Workshop #1
           Exploring Dynamics of Gravity-Drained Tanks




Copyright © 2007 Control Station, Inc. All Rights Reserved   63
Processes have Time Varying Behavior

         The CO to PV behavior described by an ideal FOPDT model is
         constant, but real processes change every day because:
            surfaces foul or corrode
            mechanical elements like seals or bearings wear
            feedstock quality varies and catalyst activity decays
            environmental conditions like heat and humidity change

         So the values of Kp, τ p and Өp that best describe the dynamic
         behavior of a process today may not be best tomorrow

         As a result, controller performance can degrade with time and
         periodic retuning may be required




Copyright © 2007 Control Station, Inc. All Rights Reserved   64
Processes have Nonlinear Behavior




                                                                  …cause changing (nonlinear)
                                                                   response in real processes
                           equal CO steps…




                                                       Copyright © 2007 by Control Station, Inc. All Rights Reserved




                      The dynamic behavior of most real processes
                      changes as operating level changes



Copyright © 2007 Control Station, Inc. All Rights Reserved               65
Processes have Nonlinear Behavior

                              3) but behavior of real processes
                                 changes with operating level




                                                       2) FOPDT model response is
                                                          constant as operating level changes
                       1) with equal CO steps




                                                         Copyright © 2007 by Control Station, Inc. All Rights Reserved




       A FOPDT model response is constant as operating level changes

       Since the FOPDT model is used for controller design and tuning, a
       process should be modeled at a specific design level of operation!

Copyright © 2007 Control Station, Inc. All Rights Reserved             66
What is a Nonlinear Process

       While linear processes may be the design goal of process
       engineers, the reality is that most processes are nonlinear
       in nature due to nonlinearity in the final control element or
       the process itself.
           Example: A heating process is nonlinear because the rate
           at which heat is transferred between two objects depends
           on the difference in temperature between the objects.
           Example: A valve that is linear in the middle of its
           operating range may become very nonlinear towards its
           limits.




Copyright © 2007 Control Station, Inc. All Rights Reserved   67
What is a Nonlinear Process
   Dealing with Nonlinearity – Set Point Response




Copyright © 2007 Control Station, Inc. All Rights Reserved   68
What is a Nonlinear Process
   Dealing with Nonlinearity

       The robustness of a controller is a measure the
       range of process values over which the controller
       provides stable operation.
       The more nonlinear a process is, the less
       aggressive you must be in your tuning approach
       to maintain robustness.




Copyright © 2007 Control Station, Inc. All Rights Reserved   69
What is Process Action

       Process action is how the process variable changes with
       respect to a change in the controller output. Process
       action is either direct acting or reverse acting.
       The action of a process is defined by the sign of the
       process gain. A process with a positive gain is said to be
       direct acting. A process with a negative gain is said to be
       reverse acting.




Copyright © 2007 Control Station, Inc. All Rights Reserved   70
What is Process Action
   Process and Controller Action

       The action of a process is important because it will
       determine the action of the controller.
            A direct acting process requires a reverse acting
            controller.
            Conversely, a reverse acting process requires a direct
            acting controller.




Copyright © 2007 Control Station, Inc. All Rights Reserved   71
Fundamentals of Instrumentation
& Process Control

Interactive Training Workshop
Introduction to Instrumentation

       The intent of this chapter is neither to teach you how to
       select a particular instrument nor to familiarize you with all
       of the available types of instruments.

       The intent of this material is to provide an introduction to
       commonly measured process variables, including the basic
       terminology and characteristics relevant to each variable’s
       role in a control loop.

       Detailed information and assistance on device selection is
       typically available directly from the instrumentation
       supplier.




Copyright © 2007 Control Station, Inc. All Rights Reserved   73
Introduction to Instrumentation

       Objectives:
         What are sensors and transducers?
         What are the standard instrument signals?
         What are smart transmitters?
         What is a low pass filter?
         What instrument properties affect a process?
         What is input aliasing?
         What is instrument noise?
         How do we measure temperature?
         How do we measure level?
         How do we measure level?
         How do we measure flow?



Copyright © 2007 Control Station, Inc. All Rights Reserved   74
Introduction to Instrumentation


       You cannot control what you
             cannot measure


                            “When you can measure what you are speaking about,
                             and express it in numbers, you know something
                             about it. When you cannot measure it, when you
                             cannot express it in numbers, your knowledge is of a
                             meagre and unsatisfactory kind.”

                                                                  - Lord Kelvin



Copyright © 2007 Control Station, Inc. All Rights Reserved   75
What are Sensors and Transducers

       A sensor is a device that has a characteristic that changes
       in a predictable way when exposed to the stimulus it was
       designed to detect.
       A transducer is a device that converts one form of energy
       into another.




Copyright © 2007 Control Station, Inc. All Rights Reserved   76
What are Standard Instrumentation Signals


       Standard instrument signals for controllers to
       accept as inputs from instrumentation and
       outputs to final control elements are:

            pneumatic
            current loop
            0 to 10 volt




Copyright © 2007 Control Station, Inc. All Rights Reserved   77
What are Standard Instrumentation Signals
   Pneumatic

       3 to 15 psig
            Before 1960, pneumatic signals were used almost
            exclusively to transmit measurement and control
            information.
            Today, it is still common to find 3 to 15 psig used as
            the final signal to a modulating valve.
            Most often an I/P (I to P) transducer is used.
               This converts a 4-20 mA signal (I) into a pressure
               signal (P).




Copyright © 2007 Control Station, Inc. All Rights Reserved   78
What are Standard Instrumentation Signals
   Pneumatic Scaling

       What would our pneumatic signal be if our
       controller output is 40%?
                                  (
          Signal psig = % Controller Output x 12 psig + 3psig     )




Copyright © 2007 Control Station, Inc. All Rights Reserved   79
What are Standard Instrumentation Signals
   Current Loop

       4-20 milliamp
            Current loops are the signal workhorses in our
            processes.
            A DC milliamp current is transmitted through a pair
            of wires from a sensor to a controller or from a
            controller to its final control element.
            Current loops are used because of their immunity to
            noise and the distances that the signal can be
            transmitted.




Copyright © 2007 Control Station, Inc. All Rights Reserved   80
What are Standard Instrumentation Signals
   Current Loop Scaling

       Output Scaling
            Scale outputs for a one-to-one correspondence.
            Controller output is configured for 0% to correspond
            to a 4mA signal and 100% to correspond to a 20mA
            signal.
            The final control element is calibrated so that 4mA
            corresponds to its 0% position or speed and 20mA
            corresponds to its 100% position or speed.




Copyright © 2007 Control Station, Inc. All Rights Reserved   81
What are Standard Instrumentation Signals
   Current Loop Scaling

       Input Scaling
            Scale inputs for a one-to-one correspondence as well.

            Example:
                  If we were using a pressure transducer with a required
                  operating range of 0 psig to 100 psig we would calibrate
                  the instrument such that 0 psig would correspond to 4mA
                  output and 100 psig would correspond to a 20mA output.
                  At the controller we would configure the input such that
                  4mA would correspond to an internal value of 0 psig and
                  10mA would correspond to an internal value of 100 psig.




Copyright © 2007 Control Station, Inc. All Rights Reserved   82
What are Standard Instrumentation Signals
   0 to 10 Volt

       0 to 10 volt is not commonly used in control
       systems because this signal is susceptible to
       induced noise and the distance of the instrument
       or final control element is limited due to voltage
       drop.
       You may find 0-10 volt signals used in control
       systems providing the speed reference to
       variable speed drives.




Copyright © 2007 Control Station, Inc. All Rights Reserved   83
What are Smart Transmitters

       A smart transmitter is a digital device that
       converts the analog information from a sensor
       into digital information, allowing the device to
       simultaneously send and receive information and
       transmit more than a single value.
       Smart transmitters, in general, have the
       following common features:
            Digital Communications
            Configuration
            Re-Ranging
            Signal Conditioning
            Self-Diagnosis




Copyright © 2007 Control Station, Inc. All Rights Reserved   84
What are Smart Transmitters
   Digital Communications

       Smart transmitters are capable of digital communications
       with both a configuration device and a process controller.
       Digital communications have the advantage of being free of
       bit errors, the ability to multiple process values and
       diagnostic information, and the ability to receive
       commands.
       Some smart transmitters use a shared channel for analog
       and digital data (Hart, Honeywell or Modbus over 4-20mA).
        Others use a dedicated communication bus (Profibus,
       Foundation Fieldbus, DeviceNet, Ethernet).




Copyright © 2007 Control Station, Inc. All Rights Reserved   85
What are Smart Transmitters
   Digital Communications



       Most smart instruments wired to
       multi-channel input cards require
       isolated inputs for the digital
       communications to work.




Copyright © 2007 Control Station, Inc. All Rights Reserved   86
What are Smart Transmitters
   Configuration, Signal Conditioning, Self-Diagnosis

       Configuration
            Smart transmitters can be configured with a
            handheld terminal and store the configuration
            settings in non-volatile memory.
       Signal Conditioning
            Smart transmitters can perform noise filtering and
            can provide different signal characterizations.
       Self-Diagnosis
            Smart transmitters also have self-diagnostic
            capability and can report malfunctions that may
            indicate erroneous process values.



Copyright © 2007 Control Station, Inc. All Rights Reserved   87
What Instrument Properties Affect a Process

       The     instrument’s               range and span.
       The     resolution of              the measurement.
       The     instrument’s               accuracy and precision.
       The     instrument’s               dynamics




Copyright © 2007 Control Station, Inc. All Rights Reserved   88
What Instrument Properties Affect a Process
   Range and Span

       The range of a sensor is the lowest and highest
       values it can measure within its specification.
       The span of a sensor is the high end of the
       Range minus the low end of the Range.
       Match Range to Expected Conditions
            Instruments should be selected with a range that
            includes all values a process will normally encounter,
            including expected disturbances and possible failures.




Copyright © 2007 Control Station, Inc. All Rights Reserved   89
What Instrument Properties Affect a Process
   Measurement Resolution

       Resolution is the smallest amount of input
       signal change that the instrument can detect
       reliably.
            Resolution is really a function of the instrument span
            and the controller’s input capability.
            The resolution of a 16 bit conversion is:

                                                                  Input Span
            The resolution of a 12 bit conversion is:
                                                                   65,535
            The bit error.
                                                                  Input Span
                                                                    4,095




Copyright © 2007 Control Station, Inc. All Rights Reserved   90
What Instrument Properties Affect a Process
   Accuracy

       Accuracy of a measurement describes how close
       the measurement approaches the true value of
       the process variable.
            % error over a range
              ± x% over
                          % of full scale
                          % of span
            Absolute over a range
              ± x units over
                          full scale
                          span




Copyright © 2007 Control Station, Inc. All Rights Reserved   91
What Instrument Properties Affect a Process
   Precision

       Precision is the reproducibility with which
       repeated measurements can be made under
       identical conditions.
            This may be referred to as drift.




Copyright © 2007 Control Station, Inc. All Rights Reserved   92
What Instrument Properties Affect a Process
   Accuracy vs. Precision



       Why is precision preferred over accuracy?




Copyright © 2007 Control Station, Inc. All Rights Reserved   93
What Instrument Properties Affect a Process
   Instrumentation Dynamics: Gain and Dead Time

       The gain of an instrument is often called
       sensitivity.
            The sensitivity of a sensor is the ratio of the output
            signal to the change in process variable.
       The dead time of an instrument is the time it
       takes for an instrument to start reacting to
       process change.




Copyright © 2007 Control Station, Inc. All Rights Reserved   94
What Instrument Properties Affect a Process
   Instrumentation Dynamics: Time Constant

       As for processes, one time constant for an instrument is
       the time it takes to provide a signal that represents 63.2%
       of the value of variable it is measuring after a step change
       in the variable.
       Instrument manufacturers may sometimes specify the rise
       time instead of the time constant.
            Rise time is the time it takes for an instrument to
            provide a signal that represents 100% of the value of
            the variable it is measuring after a step change in the
            variable.
            The rise time of an instrument is equal to 5 time
            constants.




Copyright © 2007 Control Station, Inc. All Rights Reserved   95
What is Input Aliasing

       Input aliasing is a phenomenon that occurs from
       digital processing of a signal.
       When a signal is processed digitally it is sampled
       at discrete intervals of time. If the frequency at
       which a signal is sampled is not fast enough the
       digital representation of that signal will not be
       correct.
       Input aliasing is an important consideration in
       digital process control. Processor inputs that
       have configurable sample rates and PID loop
       update times must be set correctly.


Copyright © 2007 Control Station, Inc. All Rights Reserved   96
What is Input Aliasing

                 2 Hz waveform sampled every 0.4 seconds




Copyright © 2007 Control Station, Inc. All Rights Reserved   97
What is Input Aliasing
   Correct Sampling Frequency

       Nyquist Frequency Theorem:
            To correctly sample a waveform it is necessary to
            sample at least twice as fast has the highest
            frequency in the waveform.
            In the digital world this means sampling at 1/20th of
            the period of the waveform.
               Period = 1/frequency




Copyright © 2007 Control Station, Inc. All Rights Reserved   98
What is Input Aliasing
   Correct Sampling Frequency

              2 Hz waveform sampled every 0.025 seconds




Copyright © 2007 Control Station, Inc. All Rights Reserved   99
Workshop Lab:
                   Input Aliasing 1




Copyright © 2007 Control Station, Inc. All Rights Reserved   100
What is Input Aliasing
   Determining the Correct Sampling Interval

       Rules of Thumb
            Set the sample interval for an instrument at 1/10th to
            1/20th of the rise time (1/2 to 1/4th of the time constant).
            Set the sample interval to 1/10th to 1/20th of the process
            time constant.
            Temperature instrumentation (RTDs and thermocouples in
            thermowells) typically have time constants of several
            seconds or more. For these processes sampling intervals
            of 1 second are usually sufficient.
            Pressure and flow instrumentation typically have time
            constants of ½ to 1 second. For these processes sampling
            intervals of 0.1 second are usually sufficient.




Copyright © 2007 Control Station, Inc. All Rights Reserved   101
What is Input Aliasing
   Determining the Correct Sampling Interval




Copyright © 2007 Control Station, Inc. All Rights Reserved   102
Workshop Lab:
                   Input Aliasing 2




Copyright © 2007 Control Station, Inc. All Rights Reserved   103
What is Instrument Noise

       Noise is a variation in a measurement of a
       process variable that does not reflect real
       changes in the process variable.




Copyright © 2007 Control Station, Inc. All Rights Reserved   104
What is Instrument Noise
   Sources

       Noise is generally a result of the technology used
       to sense the process variable.
            Electrical signals used to transmit instrument
            measurements are susceptible to having noise
            induced from other electrical devices.
            Noise can also be caused by wear and tear on the
            mechanical elements of a sensor.
       Noise may be uncontrolled, random variations in
       the process itself.
       Whatever the source, noise distorts the
       measurement signal.



Copyright © 2007 Control Station, Inc. All Rights Reserved   105
What is Instrument Noise
   Effects of Noise

       Noise reduces the accuracy and precision of
       process measurements. Somewhere in the noise
       is the true measurement, but where?
            Noise introduces more uncertainty into the
            measurement.
       Noise also introduces errors in control systems.
       To a controller, fluctuations in the process
       variable due to noise are indistinguishable from
       fluctuations caused by real disturbances.
            Noise in a process variable will be reflected in the
            output of the controller



Copyright © 2007 Control Station, Inc. All Rights Reserved   106
What is Instrument Noise
   Eliminating Noise

       The most effective means of eliminating noise is
       to remove the source.
            Reduce electrically induced noise by following proper
            grounding techniques, using shielded cabling, and
            physically separating the signal cabling form other
            electrical wiring.
            If worn mechanical elements in the sensor are
            causing noise, then repair or replace the sensor.
       When these steps have been taken and excessive
       noise is still a problem in the process variable a
       low pass filter may be appropriate.



Copyright © 2007 Control Station, Inc. All Rights Reserved   107
What is Instrument Noise
   Low Pass Filter

       A low-pass filter allows the low frequency
       components of a signal to pass while attenuating
       the higher frequency components.
                                             Raw (unfiltered) Signal




                                                                   Filtered Signal


Copyright © 2007 Control Station, Inc. All Rights Reserved   108
What is Instrument Noise
   Low Pass Filter

       A low pass filter introduces a first order lag into
       the measurement.
                                             Raw (unfiltered) Signal




                                          Filtered Signal




Copyright © 2007 Control Station, Inc. All Rights Reserved   109
What is Instrument Noise
   Selecting a Low Pass Filter

       By Cut-Off Frequency
            Cut-off frequency is defined as frequency above
            which the filter provides -3dB of signal attenuation.
                                                        Amplitude Out 
                          dB of attenuation = 20 log10                
                                                        Amplitude In 
            An attenuation of 0 dB would mean the signal will
            pass with no reduction in amplitude while a large
            negative dB would indicate a very small amplitude
            ratio.
       Select a cut-off frequency that is above the
       frequency of your process.



Copyright © 2007 Control Station, Inc. All Rights Reserved   110
What is Instrument Noise
   Selecting a Low Pass Filter

       By Time Constant
            Some filters are configured by selecting a time
            constant for the lag response of the filter.
            The relationship between the cut-off frequency and
            the time constant of a low pass is approximately
            given by:



                                                                    1
                              Cut - Off Frequency ≈
                                                             5 Time Constants




Copyright © 2007 Control Station, Inc. All Rights Reserved       111
What is Instrument Noise
   Selecting a Low Pass Filter

       By α Value
            Some filters are selected by a value called alpha (α),
            notably the derivative filter in a PID controller.
               α is an averaging weighting term used in controller
               calculations to impart a first order lag on the
               measured variable.
               α generally has values between 0 and 1. A filter with
               a α = 0 would pass the signal through unfiltered. A
               filter with a α = 1 would filter everything allowing
               nothing to pass through the filter.




Copyright © 2007 Control Station, Inc. All Rights Reserved   112
What is Instrument Noise
   Selecting a Low Pass Filter

       When a filter is specified by cut-off frequency,
       the lower the frequency the greater the filtering
       effect.
       When a filter is specified by time constant, the
       greater the time constant the greater the filtering
       effect.
       When a filter is specified by α, when α=0 no
       filtering is done, when α = 1 no signal passes
       through the fitter.




Copyright © 2007 Control Station, Inc. All Rights Reserved   113
Workshop Lab:
                   Noise Filtering




Copyright © 2007 Control Station, Inc. All Rights Reserved   114
What is Temperature

       Temperature is a measure of degree of the
       hotness or coldness of an object.
       Units of Temperature
            Two most common temperature scales are
            Fahrenheit (°F) and Celsius (°C).
            The reference points are the freezing point and the
            boiling point of water.
               Water freezes at 32°F and boils at 212°F.
               The Celsius scale uses the same reference points
               only it defines the freezing point of water as 0°C and
               the boiling point as 100°C.




Copyright © 2007 Control Station, Inc. All Rights Reserved   115
What is Temperature
   Unit Conversion




               9     
          °F =  x °C  + 32
               5     

                             5
           °C =                          (
                                    x °F - 32                )
                             9
Copyright © 2007 Control Station, Inc. All Rights Reserved   116
What Temperature Instruments Do We Use

       Temperature is one of the most common process
       variables.
       Temperature is most commonly measured by
            Resistance Temperature Devices (RTD)
            Thermocouples
            Infrared (IR) (to a lesser degree)
            Thermistors (may also be found embedded in some
            control equipment)




Copyright © 2007 Control Station, Inc. All Rights Reserved   117
What Temperature Instruments Do We Use
   Thermocouples

       Thermocouples are fabricated from two electrical
       conductors made of two different metal alloys.
            Hot or sensing junction
            Cold or reference junction
            Generate an open-circuit voltage, called the Seebeck
            voltage that is proportional to the temperature
            difference between the sensing (hot) and reference
            (cold) junctions.




Copyright © 2007 Control Station, Inc. All Rights Reserved   118
What Temperature Instruments Do We Use
   Thermocouple Junctions

       There is a misconception of how thermocouples operate.
       The misconception is that the hot junction is the source of
       the output voltage. This is wrong. The voltage is
       generated across the length of the wire.
       Another misconception is that junction voltages are
       generated at the cold end between the special
       thermocouple wire and the copper circuit. Hence, a cold
       junction temperature measurement is required. This
       concept is wrong. The cold end temperature is the
       reference point for measuring the temperature difference
       across the length of the thermocouple circuit.




Copyright © 2007 Control Station, Inc. All Rights Reserved   119
What Temperature Instruments Do We Use
   Thermocouple Junctions

       Hot junction voltage proportional to 210°F.
       Extension wire voltage proportional to 18°F.




Copyright © 2007 Control Station, Inc. All Rights Reserved   120
What Temperature Instruments Do We Use
   Thermocouple Junctions

       Use of the correct extension wire is critical in
       thermocouple applications.
            An incorrect extension will cause the temperature
            differential across the extension leads to be
            introduced as measurement error.
       Does this mean you cannot use copper terminal
       blocks for thermocouples?
            No, it is highly unlikely there will be a temperature
            differential across the terminal block; no error will
            be introduced.




Copyright © 2007 Control Station, Inc. All Rights Reserved   121
What Temperature Instruments Do We Use
   Thermocouple Linearization




Copyright © 2007 Control Station, Inc. All Rights Reserved   122
What Temperature Instruments Do We Use
   Thermocouple Gain




Copyright © 2007 Control Station, Inc. All Rights Reserved   123
What Temperature Instruments Do We Use
   Thermocouple Types




Copyright © 2007 Control Station, Inc. All Rights Reserved   124
What Temperature Instruments Do We Use
   RTDs

       A resistance-temperature detector (RTD) is a temperature
       sensing device whose resistance increases with
       temperature.
       An RTD consists of a wire coil or deposited film of pure
       metal whose resistance at various temperatures has been
       documented.
       RTDs are used when applications require accuracy, long-
       term stability, linearity and repeatability.
       100 Ω Platinum is most common.




Copyright © 2007 Control Station, Inc. All Rights Reserved   125
What Temperature Instruments Do We Use
   RTDs: Importance of α

       α is, roughly, the slope of the RTDs
       resistance curve with respect to
       temperature.
                    R100 − R0
            α =
                      100 x R0
            A 100Ω platinum RTD with a =0.003911 will have a
            resistance of 139.11 ohms at 100°C. A 100 Ω platinum
            RTD with a =0.003850 will have a resistance of 138.50
            ohms at 100°C.
       Common values for the temperature coefficient
       are:
            0.003850, DIN 43760 Standard (also referred to as
            European curve)
            0.003911, American Standard
            0.003926, ITS-90 Standard
Copyright © 2007 Control Station, Inc. All Rights Reserved   126
What Temperature Instruments Do We Use
   RTD Lead Wire Resistance

       The RTD is a resistive device, you must drive a
       current through the device and monitor the
       resulting voltage. Any resistance in the lead
       wires that connect your measurement system to
       the RTD will add error to your readings.
            40 feet of 18 gauge 2
             conductor cable has
            6Ω resistance.
            For a platinum RTD
            with α = 0.00385,
            the resistance equals
            0.6 Ω /(0.385 Ω /°C) = 1.6°C of error.




Copyright © 2007 Control Station, Inc. All Rights Reserved   127
What Temperature Instruments Do We Use
   RTD Lead Wire Resistance

       Error is eliminated through the use of three wire
       RTDs.
            With three leads, two voltage measurements can be
            made.
            When the voltages are subtracted, the result is
            the voltage drop that would have occurred through
            RT alone.




Copyright © 2007 Control Station, Inc. All Rights Reserved   128
What Temperature Instruments Do We Use
   RTD Self Heating Effect

       The current used to excite an RTD causes the RTD to
       internally heat, which appears as an error. Self-heating is
       typically specified as the amount of power that will raise
       the RTD temperature by 1° C, or 1 mW/ °C.
       Self-heating can be minimized by using the smallest
       possible excitation current, but this occurs at the expense
       of lowering the measurable voltages and making the signal
       more susceptible to noise from induced voltages.
       The amount of self-heating also depends heavily on the
       medium in which the RTD is immersed. An RTD can self-
       heat up to 100 times higher in still air than in moving
       water




Copyright © 2007 Control Station, Inc. All Rights Reserved   129
What Temperature Instruments Do We Use
   Thermistors

       A thermistor is similar to an RTD in that it is a
       passive resistance device.
            Thermistors are generally made of semiconductor
            materials giving them much different characteristics.
            Thermistors do not have standardized electrical
            properties like thermocouples or RTDs.




Copyright © 2007 Control Station, Inc. All Rights Reserved   130
What Temperature Instruments Do We Use
   Infrared

       Objects radiate electromagnetic energy.
            The higher the temperature of an object the more
            electromagnetic radiation it emits.
            This radiation occurs within the infrared portion of
            the electromagnetic spectrum.




Copyright © 2007 Control Station, Inc. All Rights Reserved   131
What Temperature Instruments Do We Use
   Infrared Emittance

       The emittance of a real surface is the ratio of the
       thermal radiation emitted by the surface at a
       given temperature to that of the ideal black body
       at the same temperature.
            By definition, a black body has an emittance of 1.
       Another way to think of emissivity is the
       efficiency at which an object radiates thermal
       energy.
       Emittance is a decimal number that ranges
       between 0 and 1 or a percentage between 0%
       and 100%.


Copyright © 2007 Control Station, Inc. All Rights Reserved   132
What Temperature Instruments Do We Use
   Infrared Emittance

       To function properly an infrared temperature
       instrument must take into account the emittance
       of the surface being measured.
       Emittance values can often be found in reference
       tables, but such tables will not take into account
       local conditions of the surface.
       A more practical way to set the emittance is to
       measure the temperature with an RTD or
       thermocouple and set the instrument emissivity
       control so that both readings are the same.




Copyright © 2007 Control Station, Inc. All Rights Reserved   133
What Temperature Instruments Do We Use
   Infrared Field of View

       Infrared temperature instruments are like an optical
       system in that they have a field of view. The field of view
       basically defines the target size at a given distance.
       Field of view may be specified as:
            Angle and focal range (2.3° from 8" to 14')
            Distance to spot size ratio and focal range (25:1 from 8" to 14')
            Spot size at a distance (0.32" diameter spot at 8")
            All of these specifications are equivalent, at 8 inches our field of
            view will be a 0.32" diameter spot (8"/25), at 14 feet our field of
            view will be 6.72" (14' / 25).
       An infrared temperature sensor measures the average
       temperature of everything in its field of view. If the
       surface whose temperature we are measuring does not
       completely fill the field of view we will get inaccurate
       results.



Copyright © 2007 Control Station, Inc. All Rights Reserved   134
What Temperature Instruments Do We Use
   Infrared Spectral Response

       The range of frequencies that an infrared sensor
       can measure is called its spectral response.
       Infrared temperature sensors do not measure the
       entire infrared region.
            Different frequencies (wavelengths) are selected for
            detection to provide certain advantages.
            If you wanted to use an infrared temperature sensor to
            measure the temperature of an object behind glass you
            would need to select an instrument with a spectral
            response in the infrared region in which glass is
            "transparent”.




Copyright © 2007 Control Station, Inc. All Rights Reserved   135
Section Assessment:
                   Temperature




Copyright © 2007 Control Station, Inc. All Rights Reserved   136
What is Pressure

       Pressure is the ratio between a force acting on a
       surface and the area of that surface.
       Pressure is measured in units of force divided by
       area.
            psi (pounds per square inch)
            bar, 1 bar = 14.5 PSI, common for pump ratings
            inH20 (inches of water), 27.680 inH20 = 1 psi,
            common for vacuum systems and tank levels
            mmHg (millimeters of mercury), 760 mmHg = 14.7
            psi, common for vacuum systems




Copyright © 2007 Control Station, Inc. All Rights Reserved   137
What is Pressure
   Absolute, Gauge and Differential

       Gauge pressure is defined relative to atmospheric
       conditions.
            The units of gauge pressure are psig, however gauge
            pressure is often denoted by psi as well.
       Absolute pressure is defined as the pressure
       relative to an absolute vacuum.
            The units of absolute pressure are psia.
       Differential pressure uses a reference point other
       than full vacuum or atmospheric pressure.




Copyright © 2007 Control Station, Inc. All Rights Reserved   138
What is Pressure
   Absolute, Gauge and Differential




Copyright © 2007 Control Station, Inc. All Rights Reserved   139
Section Assessment:
                   Pressure




Copyright © 2007 Control Station, Inc. All Rights Reserved   140
Common Level Sensing Technologies

       Point sensing level probes only sense tank level at a
       discrete level.
            Typically used for high-high or low-low level sensing to prevent plant
            personnel and/or process equipment from being exposed to harmful
            conditions.
            Also used in pairs in processes in which we do not particularly care
            what the exact level in a tank is, only that it is between two points.
       Continuous level probes sense the tank level as a percent
       of span of the probes capabilities.
            Continuous level probes are typically used where we need some type
            of inventory control, where we need to know with some degree of
            confidence what the particular level in a tank is.
            Contact and Non-Contact Types




Copyright © 2007 Control Station, Inc. All Rights Reserved   141
Common Level Sensing Technologies
   Non-Contact: Ultrasonic
       Ultrasonic makes use of sound waves.
       A transducer mounted in the top of a tank transmits sound
       waves in bursts onto the surface of the material to be
       measured. Echoes are reflected back from the surface of
       the material to the transducer and the distance to the
       surface is calculated from the burst-echo timing.
       The key points in applying an ultrasonic transducer are:
            The speed of sound varies with temperature.
            Heavy foam on the surface of the material interferes with the echo.
            An irregular material surface can cause false echoes resulting in
            irregular readings.
            Heavy vapor in the air space can distort the sound waves resulting in
            false reading.




Copyright © 2007 Control Station, Inc. All Rights Reserved   142
Common Level Sensing Technologies
   Non-Contact: Ultrasonic




Copyright © 2007 Control Station, Inc. All Rights Reserved   143
Common Level Sensing Technologies
   Non-Contact: Radar/Microwave

       Radar, or microwave level measurement,
       operates on similar principles to ultrasonic level
       probes but, instead of sound waves,
       electromagnetic waves in the 10GHz range are
       used.
       When properly selected, radar can overcome
       many of the limitations of ultrasonic level probes.
            Be unaffected by temperature changes in the tank air space.
            See through heavy foam to detect the true material level.
            See through heavy vapor in the tank air space to detect true
            material level.




Copyright © 2007 Control Station, Inc. All Rights Reserved   144
Common Level Sensing Technologies
   Non-Contact: Radar/Microwave




Copyright © 2007 Control Station, Inc. All Rights Reserved   145
Common Level Sensing Technologies
   Non-Contact: Nuclear Level Measurement

       Radiation from the source is detected on the
       other side of the tank. Its strength indicates the
       level of the fluid. Point, continuous, and interface
       measurements can be made.
       As no penetration of the vessel is needed there are a
       number of situations that cause nucleonic transmitters to
       be considered over other technologies.
          Nuclear level detection has some drawbacks. One is high
          cost, up to four times that of other technologies. Others
          are the probable requirement for licenses, approvals, and
          periodic inspections; and the difficulty and expense of
          disposing of spent radiation materials.




Copyright © 2007 Control Station, Inc. All Rights Reserved   146
Common Level Sensing Technologies
   Non-Contact: Nuclear Level Measurement




Copyright © 2007 Control Station, Inc. All Rights Reserved   147
Common Level Sensing Technologies
   Contact: Hydrostatic Pressure
       Measurement of level by pressure relies on hydrostatic
       principles.
       Pressure is a unit force over a unit area.
            A cubic foot (12"L x 12"W x 12"H) of water weighs 62.4796 pounds.
            The area that our cubic foot of water occupies is 144 square inches
            (12"L x 12"H), therefore our cubic foot of water exerts a force of
            62.4796 pounds over 144 square inches, or 0.4339 psig for a 12"
            water column.
            1 inch H20 = 0.0316 psig
       It would not matter how many cubic feet of water were
       placed side by side, our pressure would still be 0.4399
       psig. Hydrostatic pressure is only dependent on the height
       of the fluid, not the area that it covers.




Copyright © 2007 Control Station, Inc. All Rights Reserved   148
Common Level Sensing Technologies
   Contact: Hydrostatic Pressure
       How would you handle measuring other fluids using
       pressure?
           Compare the densities.
       For instance, chocolate weighs approximately 80 pounds
       per cubic foot.
           80 divided 62.5 times 0.0316 = 0.0404 psig.
           For a change in level of one inch in a liquid chocolate tank
           the pressure measurement will change by 0.0404 psig.
       Level measurement by pressure requires a constant
       density for accurate measurements.
       If the head pressure in the tank can be other than
       atmospheric, we must use a differential pressure sensor.




Copyright © 2007 Control Station, Inc. All Rights Reserved   149
Common Level Sensing Technologies
   Contact: Hydrostatic Pressure




Copyright © 2007 Control Station, Inc. All Rights Reserved   150
Common Level Sensing Technologies
   Contact: RF/Capacitance
       RF (radio frequency) Capacitance
       level sensors make use of
       electrical characteristics of a
       capacitor to infer the level in a vessel.
          As the material rises in the vessel,
          the capacitance changes.
          The level transducer measures
          this change, linearizes it and transmits the signal to the
          process control system.
          A point level probe will look for a specific change in
          capacitance to determine whether it is on or off.




Copyright © 2007 Control Station, Inc. All Rights Reserved   151
Common Level Sensing Technologies
   Contact: RF/Capacitance
       Requires the material being measured to have a high
       dielectric constant.
          Level measurements are affected by changes in the
          dielectric of the material (moisture content).
       Proper selection requires informing the probe vendor of the
       material to be measure, especially in applications where
       you are measuring conductive materials or have a
       nonmetallic tank.
       Point probe sensitivity can be increased by welding a plate
       on the sensor tip to increase the capacitance, and therefore
       the sensitivity (gain).




Copyright © 2007 Control Station, Inc. All Rights Reserved   152
Common Level Sensing Technologies
   Contact: RF/Capacitance




Copyright © 2007 Control Station, Inc. All Rights Reserved   153
Common Level Sensing Technologies
   Contact: Guided Wave Radar
       Guided wave radar is similar to the non-contact radar
       probes, only a rod or cable is immersed into the material
       like an RF capacitance probe.
       The rod or cable is used to guide the microwave along its
       length, where the rod or cable meets the material to be
       measured a wave reflection is generated. The transit time
       of the wave is used to calculate level very precisely.
       Unlike an RF capacitance probe, a guided wave radar probe
       can measure extremely low dielectric material.




Copyright © 2007 Control Station, Inc. All Rights Reserved   154
Common Level Sensing Technologies
   Contact: Guided Wave Radar




Copyright © 2007 Control Station, Inc. All Rights Reserved   155
Section Assessment:
                   Level




Copyright © 2007 Control Station, Inc. All Rights Reserved   156
What is Flow

       Flow is the motion characteristics of constrained
       fluids (liquids or gases).
       Fluid velocity or mass are typically not measured
       directly.
       Measurements are affected by the properties of
       the fluid, the flow stream, and the physical
       installation.




Copyright © 2007 Control Station, Inc. All Rights Reserved   157
Factors Affecting Flow Measurement

       The critical factors affecting flow measurement
       are:
            Viscosity
            Fluid Type
            Reynolds Number
            Flow Irregularities




Copyright © 2007 Control Station, Inc. All Rights Reserved   158
Factors Affecting Flow Measurement
   Viscosity

       Dynamic or absolute viscosity (η) is a measure of
       the resistance to a fluid to deformation under
       shear stress, or an internal property of a fluid
       that offers resistance to flow.
       Commonly perceived as “thickness” or resistance
       to pouring.
            Water is very “thin” having a relatively low viscosity,
            while molasses is very thick having a relatively high
            viscosity.




Copyright © 2007 Control Station, Inc. All Rights Reserved   159
Factors Affecting Flow Measurement
   Viscosity
       Visualize a flat sheet of glass on a film of oil on top of a flat
       surface.
            A parallel force applied
            to the sheet of glass will
            accelerate it to a final velocity
            dependent only on the
            amount of force applied.
            The oil that is next to the sheet
            of glass will have a velocity
            close to that of the glass; while
            the oil that is next to the stationary
            surface will have a velocity near zero.
            This internal distribution of velocities is due to the internal
            resistance of the fluid to shear stress forces, its viscosity.
       The viscosity of a fluid is the ratio between the per unit
       force to accelerate the plate and the distribution of the
       velocities within the fluid film



Copyright © 2007 Control Station, Inc. All Rights Reserved   160
Factors Affecting Flow Measurement
   Viscosity

       Effect of Temperature
            The dynamic viscosity of a fluid varies with its
            temperature.
            In general, the viscosity of a liquid will decrease with
            increasing temperature while the viscosity of a gas
            will increase with increasing temperature.
            Viscosity measurements are therefore associated
            with a particular temperature




Copyright © 2007 Control Station, Inc. All Rights Reserved   161
Factors Affecting Flow Measurement
   Viscosity

       Viscosity is measured in units of poise or
       Pascal·seconds or stokes.
            Poise or Pascal·seconds are the units of dynamic
            viscosity.
            Stokes are the units of kinematic viscosity.
               Kinematic viscosity is the dynamic viscosity of a fluid
               divided by the density of the fluid.




Copyright © 2007 Control Station, Inc. All Rights Reserved   162
Factors Affecting Flow Measurement
   Viscosity Conversion Chart




Copyright © 2007 Control Station, Inc. All Rights Reserved   163
Factors Affecting Flow Measurement
   Viscosity of Common Fluids




Copyright © 2007 Control Station, Inc. All Rights Reserved   164
Factors Affecting Flow Measurement
   Viscosity

       The fluid streams in most processes can include
       both high and low viscous fluids.
       The viscosity of the fluid under process conditions
       must be taken into account when selecting a flow
       instrument for optimum performance.




Copyright © 2007 Control Station, Inc. All Rights Reserved   165
Factors Affecting Flow Measurement
   Fluid Type

       Newtonian Fluids
            When held at a constant temperature, the viscosity of
            a Non-Newtonian fluid will change with relation to the
            size of the shear force, or it will change over time
            under a constant shear force.
            Water, glycerin, liquid sugar, and corn syrup are
            examples.




Copyright © 2007 Control Station, Inc. All Rights Reserved   166
Factors Affecting Flow Measurement
   Fluid Type

       Non-Newtonian Fluids
            When held at a constant temperature, the viscosity of a
            Non-Newtonian fluid will change with relation to the size
            of the shear force, or it will change over time under a
            constant shear force.
                  Shear Thickening: Peanut Butter, Starch &
                  Water
                  Time Thickening: Milk, Molasses




Copyright © 2007 Control Station, Inc. All Rights Reserved   167
Factors Affecting Flow Measurement
   Reynolds Number

       The Reynolds number is the ratio of inertial
       forces to viscous forces of fluid flow within a pipe
       and is used to determine whether a flow will be
       laminar or turbulent.


                                    Dvρ                      Where : Re = Reynolds number
                        Re=124
                                     η                               D = Pipe diameter in inches
                                                                     v = Fluid velocity in ft/sec
                                                                     ρ = Fluid density in lb/ft3
                                                                     η = Fluid viscosity in cP



Copyright © 2007 Control Station, Inc. All Rights Reserved     168
Factors Affecting Flow Measurement
   Laminar Flow

       Laminar flow occurs at low Reynolds numbers,
       typically Re < 2000, where viscous forces are
       dominant.
       Laminar flow is characterized
       by layers of flow traveling at
       different speeds with virtually
       no mixing between layers.
       The velocity of the flow is highest in the center of
       the pipe and lowest at the walls of the pipe.




Copyright © 2007 Control Station, Inc. All Rights Reserved   169
Factors Affecting Flow Measurement
   Turbulent Flow

       Turbulent flow occurs at high Reynolds numbers,
       typically Re > 4000, where inertial forces are
       dominant.
       Turbulent flow is
       characterized
       by irregular movement
       of the fluid in the pipe.
       There are no definite layers.
       The velocity of the flow is nearly uniform through
       the cross-section of the pipe.



Copyright © 2007 Control Station, Inc. All Rights Reserved   170
Factors Affecting Flow Measurement
   Transitional Flow

       Transitional flow typically occurs at Reynolds
       numbers between 2000 and 4000.
       Flow in this region may be laminar, it may be
       turbulent, or it may exhibit characteristics of
       both.




Copyright © 2007 Control Station, Inc. All Rights Reserved   171
Factors Affecting Flow Measurement
   Reynolds Number




                      Flow instruments must be
                        selected to accurately
                       measure laminar flow or
                            turbulent flow



Copyright © 2007 Control Station, Inc. All Rights Reserved   172
Factors Affecting Flow Measurement
   Flow Irregularities

               Flow irregularities are
               changes in the
               velocity profile of the
               fluid flow caused by
               installation of the flow
               instrument.

                                                             Normal velocity
                                                             profiles are established
                                                             by installing a
                                                             sufficient run of
                                                             straight piping.

Copyright © 2007 Control Station, Inc. All Rights Reserved      173
Common Flowmeter Technologies




Copyright © 2007 Control Station, Inc. All Rights Reserved   174
Common Flowmeter Technologies

       Volumetric Flow
            Magnetic flowmeters
            Positive displacement for viscous liquids
            Orifice Plates
       Mass Flow
            Coriolis flowmeters




Copyright © 2007 Control Station, Inc. All Rights Reserved   175
Common Flowmeter Technologies
   Volumetric Flow

       Volumetric flow is usually inferred.
            Volumetric Flow = Fluid Velocity x Area
               Measured directly with positive displacement flow
               meters.
               Measured indirectly by differential pressure
       Volumetric flow meters provide a signal that is in
       units of volume per unit time.
            gpm (gallons per minute)
            cfm (cubic feet per minute)
            l/hr (liters per hour)




Copyright © 2007 Control Station, Inc. All Rights Reserved   176
Common Flowmeter Technologies
   Volumetric Flow: Positive Displacement

       Operate by capturing the fluid to be measured in
       rotating cavities of a known volume. The volume
       of the cavity and the rate of rotation will give the
       volumetric flow value.




       Unaffected by Reynolds number and work with
       laminar, turbulent, and transitional flow.
       Low viscosity fluids will slip past the gears
       decreasing the accuracy of the flowmeter.


Copyright © 2007 Control Station, Inc. All Rights Reserved   177
Common Flowmeter Technologies
   Volumetric Flow: Magnetic

      Magnetic flowmeters infer the
      velocity of the moving fluid by
      measuring a generated
      voltage.
      Magnetic flowmeters are
      based on Faraday’s law of
      electromagnetic induction - a
      wire moving though a
      magnetic field will generate a
      voltage.
      Requires a conductive fluid.
                      Flow = Pipe Area ⋅ velocity
                                    πD 2
                                                             E = kBDv
                                 2
                           = πr v =      v                      k = Constant
                                     4                         B = Magnetic Flux Density
                             πD 2 E      πDE                   D = Pipe Diameter
                           =           =
                               4 kBD 4kB                       v = Fluid Velocity



Copyright © 2007 Control Station, Inc. All Rights Reserved          178
Common Flowmeter Technologies
   Volumetric Flow: Orifice Plates

                                                                   An orifice plate is basically
                                                                   a thin plate with a hole in
                                                                   the middle. It is usually
                                                                   placed in a pipe in which
                                                                   fluid flows.
                                                                   In practice, the orifice
                                                                   plate is installed in the
                                                                   pipe between two flanges.
                                                                   The orifice constricts the
                                                                   flow of liquid to produce a
                                                                   differential pressure across
                                                                   the plate. Pressure taps on
                                                                   either side of the plate are
                                                                   used to detect the
                                                                   difference.
Copyright © 2007 Control Station, Inc. All Rights Reserved   179
Common Flowmeter Technologies
   Volumetric Flow: Orifice Plates

       Orifice Plate type meters only work well when
       supplied with a fully developed flow profile.
            This is achieved by a long upstream length (20 to 40
            diameters, depending on Reynolds number)
       Orifice plates are small and cheap to install, but
       impose a significant energy loss on the fluid due
       to friction.




Copyright © 2007 Control Station, Inc. All Rights Reserved   180
Common Flowmeter Technologies
   Mass Flow

       Measured directly by measuring the inertial
       effects of the fluids moving mass.
       Mass flow meters provide a signal that is in units
       of mass per unit time.
       Typical units of mass flow are:
            lb/hr (pounds per hour)
            kg/s (kilograms per second)




Copyright © 2007 Control Station, Inc. All Rights Reserved   181
Common Flowmeter Technologies
   Mass Flow: Coriolis

       Coriolis flowmeters are based on the affects of the
       Coriolis force.
       When observing motion from a rotating frame the
       trajectory of motion will appear to be altered by a
       force arising from the rotation.




Copyright © 2007 Control Station, Inc. All Rights Reserved   182
Common Flowmeter Technologies
   Mass Flow: Coriolis




Copyright © 2007 Control Station, Inc. All Rights Reserved   183
Common Flowmeter Technologies
   Mass Flow: Coriolis




Copyright © 2007 Control Station, Inc. All Rights Reserved   184
Common Flowmeter Technologies
   Mass Flow: Coriolis

    Coriolis mass flowmeters
    work by applying a
    vibrating force to a pair
    of curved tubes through
    which fluid passes, in
    effect creating a rotating
    frame of reference.
    The Coriolis Effect of the
    mass passing through
    the tubes creates a force
    on the tubes
    perpendicular to both
    the direction of vibration
    and the direction of flow,
    causing a twist in the
    tubes.


Copyright © 2007 Control Station, Inc. All Rights Reserved   185
Common Flowmeter Technologies
   Flowmeter Turndown

       A specification commonly found with flow sensors
       is turndown or rangeability.
            Turndown is the ratio of maximum to minimum flow
            that can be measured with the specified accuracy.
       For example, a flow meter with a full span of 0 to
       100 gallons per minute with a 10:1 turndown
       could measure a flow as small as 10 gallons per
       minute.
       Trying to measure a smaller flow would result in
       values outside of the manufacturer’s stated
       accuracy claims.


Copyright © 2007 Control Station, Inc. All Rights Reserved   186
Common Flowmeter Technologies
   Installation and Calibration

       As with any instrument installation, optimal flow
       meter performance and accuracy can only be
       achieved through proper installation and
       calibration. Requirements will vary between type
       and manufacturer by some general
       considerations are:
            The length of straight piping required upstream and
            downstream of the instrument. This is usually specified in
            pipe diameters.
            The properties of the fluid/gas to be measured.
            Mounting position of the flowmeter.
                  Flowmeters must be kept full during operation.




Copyright © 2007 Control Station, Inc. All Rights Reserved   187
Common Flowmeter Technologies
   Installation and Calibration

       In general, calibration of a flowmeter requires:
            Entering the sensor characterization constants into
            the flow transmitter.
            Recording the zero flow conditions.
               Many flow meters must be “zeroed” only when full.
            Comparing the meter flow value to an independently
            measured value for the flow stream.
               Measuring the volume or mass of a catch sample and
               comparing the values.
               Use the internal flowmeter totalizer value for
               comparison.




Copyright © 2007 Control Station, Inc. All Rights Reserved   188
Introduction to Final Control Elements

       Objectives:
            What       is a control valve?                         What is a centrifugal
            What       is an actuator?                             pump?
            What       is a positioner?                            What is pump head?
            What       is Cv?                                      Why do we use pump head
                                                                   and not psi?
            What       are valve characteristics?                  What is a pump curve?
            What       is valve deadband?                          What is a system curve?
            What       is stiction?                                What is the system
            What       are the types of valves                     operating point?
                                                                   What is a positive
                                                                   displacement pump?
                                                                   How does a PD pump differ
                                                                   from a centrifugal pump?



Copyright © 2007 Control Station, Inc. All Rights Reserved   189
Introduction to Final Control Elements

       The intent of this chapter is neither to teach you how to
       size a valve or select a pump nor to familiarize you with all
       of the available types of valves and pumps.
       The intent of this chapter is to provide an introduction to
       some of the commonly used pumps and valves, including
       basic terminology and characteristics relevant to their role
       in a control loop.
       Detailed information and assistance on device selection is
       typically available from your instrumentation suppliers.




Copyright © 2007 Control Station, Inc. All Rights Reserved   190
What is a Control Valve

       A control valve is an inline device in a flow
       stream that receives commands from a controller
       and manipulates the flow of a gas or fluid in one
       of three ways:
            Interrupt flow (Shut-Off Service)
            Divert flow to another path in the system (Divert
            Service)
            Regulate the rate of flow (Throttling Service)




Copyright © 2007 Control Station, Inc. All Rights Reserved   191
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)
instrumentation & contoby prashant sakharkar)

Mais conteúdo relacionado

Semelhante a instrumentation & contoby prashant sakharkar)

Semelhante a instrumentation & contoby prashant sakharkar) (20)

Maximizing the return on your control investment meet the experts sessions part2
Maximizing the return on your control investment meet the experts sessions part2Maximizing the return on your control investment meet the experts sessions part2
Maximizing the return on your control investment meet the experts sessions part2
 
Pid controllers
Pid controllersPid controllers
Pid controllers
 
PID controller
PID controllerPID controller
PID controller
 
Control actions
Control actionsControl actions
Control actions
 
Control systems
Control systemsControl systems
Control systems
 
HP0902WhyLoops
HP0902WhyLoopsHP0902WhyLoops
HP0902WhyLoops
 
Toc Sample
Toc SampleToc Sample
Toc Sample
 
PID Controller and its design
PID Controller and its designPID Controller and its design
PID Controller and its design
 
UNIT-V.ppt
UNIT-V.pptUNIT-V.ppt
UNIT-V.ppt
 
ON-OFF CONTROL (2).ppt
ON-OFF CONTROL (2).pptON-OFF CONTROL (2).ppt
ON-OFF CONTROL (2).ppt
 
Set up reduction
Set up reductionSet up reduction
Set up reduction
 
PID Control Basics
PID Control BasicsPID Control Basics
PID Control Basics
 
Pid controllers
Pid controllersPid controllers
Pid controllers
 
Chapter1
Chapter1Chapter1
Chapter1
 
Advanced control foundation tools and techniques
Advanced control foundation   tools and techniquesAdvanced control foundation   tools and techniques
Advanced control foundation tools and techniques
 
Paper id 21201482
Paper id 21201482Paper id 21201482
Paper id 21201482
 
APC Best Practice
APC Best PracticeAPC Best Practice
APC Best Practice
 
Process Control in Chemical Engineering by MATLAB
Process Control in Chemical Engineering by MATLABProcess Control in Chemical Engineering by MATLAB
Process Control in Chemical Engineering by MATLAB
 
Basicontrol
BasicontrolBasicontrol
Basicontrol
 
Vivek mishra197
Vivek mishra197Vivek mishra197
Vivek mishra197
 

Último

Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationnomboosow
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxSayali Powar
 
URLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website AppURLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website AppCeline George
 
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...Marc Dusseiller Dusjagr
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introductionMaksud Ahmed
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...EduSkills OECD
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdfQucHHunhnh
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfciinovamais
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)eniolaolutunde
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformChameera Dedduwage
 
Mastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory InspectionMastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory InspectionSafetyChain Software
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Krashi Coaching
 
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991RKavithamani
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfsanyamsingh5019
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityGeoBlogs
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfchloefrazer622
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingTechSoup
 

Último (20)

Interactive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communicationInteractive Powerpoint_How to Master effective communication
Interactive Powerpoint_How to Master effective communication
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
 
URLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website AppURLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website App
 
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
“Oh GOSH! Reflecting on Hackteria's Collaborative Practices in a Global Do-It...
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introduction
 
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
Presentation by Andreas Schleicher Tackling the School Absenteeism Crisis 30 ...
 
1029-Danh muc Sach Giao Khoa khoi 6.pdf
1029-Danh muc Sach Giao Khoa khoi  6.pdf1029-Danh muc Sach Giao Khoa khoi  6.pdf
1029-Danh muc Sach Giao Khoa khoi 6.pdf
 
Activity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdfActivity 01 - Artificial Culture (1).pdf
Activity 01 - Artificial Culture (1).pdf
 
Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)Software Engineering Methodologies (overview)
Software Engineering Methodologies (overview)
 
A Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy ReformA Critique of the Proposed National Education Policy Reform
A Critique of the Proposed National Education Policy Reform
 
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"Mattingly "AI & Prompt Design: The Basics of Prompt Design"
Mattingly "AI & Prompt Design: The Basics of Prompt Design"
 
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdfTataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
 
Mastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory InspectionMastering the Unannounced Regulatory Inspection
Mastering the Unannounced Regulatory Inspection
 
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
Mattingly "AI & Prompt Design: Structured Data, Assistants, & RAG"
 
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
Kisan Call Centre - To harness potential of ICT in Agriculture by answer farm...
 
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
 
Sanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdfSanyam Choudhary Chemistry practical.pdf
Sanyam Choudhary Chemistry practical.pdf
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activity
 
Arihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdfArihant handbook biology for class 11 .pdf
Arihant handbook biology for class 11 .pdf
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy Consulting
 

instrumentation & contoby prashant sakharkar)

  • 1. Fundamentals of Instrumentation & Process Control Interactive Training Workshop
  • 2. Introduction to Process Control A common misconception in process control is that it is all about the controller – that you can force a particular process response just by getting the right tuning parameters. In reality, the controller is just a partner. A process will respond to a controller’s commands only in the manner which it can. To understand process control you must understand the other partners as well: sensors, final control elements and the process itself. All of these determine what type of response the controller is capable of extracting out of the process. It is not the other way around. Copyright © 2007 Control Station, Inc. All Rights Reserved 2
  • 3. Outline of the Course Copyright © 2007 Control Station, Inc. All Rights Reserved 3
  • 4. Introduction to Process Control Objectives: Why do we need process control? What is a process? What is process control? What is open loop control? What is closed loop control? What are the modes of closed loop control? What are the basic elements of process control? Copyright © 2007 Control Station, Inc. All Rights Reserved 4
  • 5. Motivation for Automatic Process Control Safety First: people, environment, equipment The Profit Motive: meeting final product specs minimizing waste production minimizing environmental impact minimizing energy use maximizing overall production rate Copyright © 2007 Control Station, Inc. All Rights Reserved 5
  • 6. “Loose” Control Costs Money 65 operating constraint PV & SP (%) 60 set point far from profit constraint 55 SP more profit 45 Copyright © 2007 40 poor control = large variation in PV by Control Station, Inc. All Rights Reserved Time It takes more processing to remove impurities, so greatest profit is to operate as close to the maximum impurities constraint as possible without going over It takes more material to make a product thicker, so greatest profit is to operate as close to the minimum thickness constraint as possible without going under Copyright © 2007 Control Station, Inc. All Rights Reserved 6
  • 7. “Tight” Control is More Profitable 65 operating constraint set point near 60 profit constraint PV & SP (%) SP 50 more profit tight control = small variation in PV 45 40 Copyright © 2007 by Control Station, Inc. All Rights Reserved Time A well controlled process has less variability in the measured process variable (PV), so the process can be operated close to the maximum profit constraint. Copyright © 2007 Control Station, Inc. All Rights Reserved 7
  • 8. Terminology for Home Heating Control Control Objective Measured Process Variable (PV) Set Point (SP) Controller Output (CO) Manipulated Variable Disturbances (D) thermostat controller set point TC TT temperature heat loss sensor/transmitter (disturbance) control signal fuel flow furnace valve Copyright © 2007 by Control Station, Inc. All Rights Reserved Copyright © 2007 Control Station, Inc. All Rights Reserved 8
  • 9. What is Process Control Terminology: The manipulated variable (MV) is a measure of resource being fed into the process, for instance how much thermal energy. A final control element (FCE) is the device that changes the value of the manipulated variable. The controller output (CO) is the signal from the controller to the final control element. The process variable (PV) is a measure of the process output that changes in response to changes in the manipulated variable. The set point (SP) is the value at which we wish to maintain the process variable at. Copyright © 2007 Control Station, Inc. All Rights Reserved 9
  • 10. What is a Process A process is broadly defined as an operation that uses resources to transform inputs into outputs. It is the resource that provides the energy into the process for the transformation to occur. Copyright © 2007 Control Station, Inc. All Rights Reserved 10
  • 11. What is a Process Copyright © 2007 Control Station, Inc. All Rights Reserved 11
  • 12. What is a Process Each process exhibits a particular dynamic (time varying) behavior that governs the transformation. That is, how do changes in the resource or inputs over time affect the transformation. This dynamic behavior is determined by the physical properties of the inputs, the resource, and the process itself. Copyright © 2007 Control Station, Inc. All Rights Reserved 12
  • 13. What is a Process Can you identify some of the elements that will determine the dynamic properties of this process? Copyright © 2007 Control Station, Inc. All Rights Reserved 13
  • 14. What is Process Control Process control is the act of controlling a final control element to change the manipulated variable to maintain the process variable at a desired set point. A corollary to our definition of process control is a controllable process must behave in a predictable manner. For a given change in the manipulated variable, the process variable must respond in a predictable and consistent manner. Copyright © 2007 Control Station, Inc. All Rights Reserved 14
  • 15. What is Process Control Copyright © 2007 Control Station, Inc. All Rights Reserved 15
  • 16. What is Process Control Copyright © 2007 Control Station, Inc. All Rights Reserved 16
  • 17. Section Assessment: Basic Terminology Assessment Copyright © 2007 Control Station, Inc. All Rights Reserved 17
  • 18. What is Open Loop Control In open loop control the controller output is not a function of the process variable. In open loop control we are not concerned that a particular set point be maintained, the controller output is fixed at a value until it is changed by an operator. Many processes are stable in an open loop control mode and will maintain the process variable at a value in the absence of a disturbances. Disturbances are uncontrolled changes in the process inputs or resources. Copyright © 2007 Control Station, Inc. All Rights Reserved 18
  • 19. What is Open Loop Control Copyright © 2007 Control Station, Inc. All Rights Reserved 19
  • 20. What is Open Loop Control Can you think of processes in which open loop control is sufficient? Copyright © 2007 Control Station, Inc. All Rights Reserved 20
  • 21. What is Closed Loop Control In closed loop control the controller output is determined by difference between the process variable and the set point. Closed loop control is also called feedback or regulatory control. The output of a closed loop controller is a function of the error. Error is the deviation of the process variable from the set point and is defined as E = SP - PV. Copyright © 2007 Control Station, Inc. All Rights Reserved 21
  • 22. What is Closed Loop Control Copyright © 2007 Control Station, Inc. All Rights Reserved 22
  • 23. What is Closed Loop Control From the controllers perspective the process encompasses the RTD, the steam control valve, and signal processing of the PV and CO values. Copyright © 2007 Control Station, Inc. All Rights Reserved 23
  • 24. What are the Modes of Closed Loop Control Closed loop control can be, depending on the algorithm that determines the controller output: Manual On-Off PID Advanced PID (ratio, cascade, feedforward) or Model Based Copyright © 2007 Control Station, Inc. All Rights Reserved 24
  • 25. What are the Modes of Closed Loop Control Manual Control In manual control an operator directly manipulates the controller output to the final control element to maintain a set point. Copyright © 2007 Control Station, Inc. All Rights Reserved 25
  • 26. What are the Modes of Closed Loop Control On-Off Control On-Off control provides a controller output of either on or off in response to error. Copyright © 2007 Control Station, Inc. All Rights Reserved 26
  • 27. What are the Modes of Closed Loop Control On-Off Control Deadband Upon changing the direction of the controller output, deadband is the value that must be traversed before the controller output will change its direction again. Copyright © 2007 Control Station, Inc. All Rights Reserved 27
  • 28. What are the Modes of Closed Loop Control PID Control PID control provides a controller output that modulates from 0 to 100% in response to error. Copyright © 2007 Control Station, Inc. All Rights Reserved 28
  • 29. What are the Modes of Closed Loop Control Time Proportion Control Time proportion control is a variant of PID control that modulates the on-off time of a final control element that only has two command positions. To achieve the effect of PID control the switching frequency of the device is modulated in response to error. This is achieved by introducing the concept of cycle time. Cycle Time is the time base of the signal the final control element will receive from the controller. The PID controller determines the final signal to the controller by multiplying the cycle time by the output of the PID algorithm. Copyright © 2007 Control Station, Inc. All Rights Reserved 29
  • 30. What are the Modes of Closed Loop Control Cascade Control Cascade control uses the output of a primary (master or outer) controller to manipulate the set point of a secondary (slave or inner) controller as if the slave controller were the final control element. Copyright © 2007 Control Station, Inc. All Rights Reserved 30
  • 31. What are the Modes of Closed Loop Control Cascade Control Copyright © 2007 Control Station, Inc. All Rights Reserved 31
  • 32. Basic Elements of Process Control Controlling a process requires knowledge of four basic elements, the process itself, the sensor that measures the process value, the final control element that changes the manipulated variable, and the controller. Copyright © 2007 Control Station, Inc. All Rights Reserved 32
  • 33. Section Assessment: Basic Process Control Assessment Copyright © 2007 Control Station, Inc. All Rights Reserved 33
  • 34. Understanding Dynamic Process Behavior What We Will Learn in This Section Dynamic Process Behavior – What It Is & Why We Care What a FOPDT Dynamic Model Represents Analyzing Step Test Plot Data to Determine FOPDT Dynamic Model Parameters Process Gain, Time Constant & Dead Time How To Compute Them From Plot Data How to Use Them For Controller Design and Tuning How to Recognize Nonlinear Processes Copyright © 2007 Control Station, Inc. All Rights Reserved 34
  • 35. Dynamic Process Behavior and Controller Tuning Consider cruise control for a car vs a truck how quickly can each accelerate or decelerate what is the effect of disturbances (wind, hills, etc.) Controller (gas flow) manipulations required to maintain set point velocity in spite of disturbances (wind, hills) are different for a car and truck because the dynamic behavior of each "process" is different Dynamic behavior → how the measured process variable (PV) responds over time to changes in the controller output (CO) and disturbances (D) Copyright © 2007 Control Station, Inc. All Rights Reserved 35
  • 36. Graphical Modeling of Dynamic Process Data To learn about the dynamic behavior of a process, we analyze measured process variable (PV) test data PV test data can be generated by suddenly changing the controller output (CO) signal The CO should be moved far and fast enough so that the dynamic behavior is clearly revealed as the PV responds The dynamic behavior of a process is different as operating level changes (nonlinear behavior), so collect data at normal operating conditions (design level of operation) Copyright © 2007 Control Station, Inc. All Rights Reserved 36
  • 37. Modeling Dynamic Process Behavior The best way to understand process data is through modeling Modeling means fitting a first order plus dead time (FOPDT) dynamic model to the process data: dPV τp + PV = Kp ×CO(tθp) − dt where: PV is the measured process variable CO is the controller output signal The FOPDT model is simple (low order and linear) so it only approximates the behavior of real processes Copyright © 2007 Control Station, Inc. All Rights Reserved 37
  • 38. Modeling Dynamic Process Behavior When a first order plus dead time (FOPDT) model is fit to dynamic process data dPV τp + PV = Kp ×CO(tθp) − dt The important parameters that result are: Steady State Process Gain, Kp Overall Process Time Constant, τ p Apparent Dead Time, Өp Copyright © 2007 Control Station, Inc. All Rights Reserved 38
  • 39. The FOPDT Model is All Important FOPDT model parameters (Kp, τ p and Өp) are used in correlations to compute controller tuning values Sign of Kp indicates the action of the controller (+ Kp ⇒ reverse acting; — Kp ⇒ direct acting) Size of τ p indicates the maximum desirable loop sample time (be sure sample time T ≤ 0.1τ p ) Ratio of Өp/τ p indicates whether model predictive control such as a Smith predictor would show benefit (useful if Өp >τ p ) Model becomes part of the feed forward, Smith Predictor, decoupling and other model-based controllers Copyright © 2007 Control Station, Inc. All Rights Reserved 39
  • 40. Step Test Data and Dynamic Process Modeling Manual Mode Step Test on Gravity Drained Tanks Copyright © 2007 by Control Station, Inc. All Rights Reserved Process starts at steady state in manual mode Controller output (CO) signal is stepped to new value Process variable (PV) signal must complete the response Copyright © 2007 Control Station, Inc. All Rights Reserved 40
  • 41. Process Gain (Kp) from Step Test Data Kp describes how far the measured PV travels in response to a change in the CO A step test starts and ends at steady state, so Kp can be computed directly from the plot Steady State Change in the Process Variable,ΔPV Kp = Steady State Change in the Controller Output, ΔCO where ∆PV and ∆CO are the total change from initial to final steady state A large process gain, Kp, means that each CO action will produce a large PV response Copyright © 2007 Control Station, Inc. All Rights Reserved 41
  • 42. Process Gain (Kp) for Gravity-Drained Tanks ∆PV = (2.9 – 1.9) = 1.0 m ∆CO = (60 – 50) = 10 % Copyright © 2007 by Control Station, Inc. All Rights Reserved Compute ∆PV and ∆CO as “final minus initial” steady state values Copyright © 2007 Control Station, Inc. All Rights Reserved 42
  • 43. Process Gain (Kp) for Gravity-Drained Tanks ∆PV = 1.0 m ∆PV 1.0 m m Kp = –––– = ––––– = 0.1 ––– ∆CO 10% % ∆CO = 10 % Copyright © 2007 by Control Station, Inc. All Rights Reserved Kp has a size (0.1); a sign (+0.1), and units (m/%) Copyright © 2007 Control Station, Inc. All Rights Reserved 43
  • 44. Additional Notes on Process Gain Measuring the Process Gain Process gain as seen by a controller is the product of the gains of the sensor, the final control element and the process itself. Process Gain = Gain Process x Gain Sensor x Gain Final Control Element The gain of a controller will be inversely proportional to the process gain that it sees 1 Controller Gain ∝ Process Gain Copyright © 2007 Control Station, Inc. All Rights Reserved 44
  • 45. Additional Notes on Process Gain Converting Units of Process Gain There is one important caveat in this process; the gain we have calculated has units of m/%. Real world controllers, unlike most software simulations, have gain units specified as %/%. When calculating the gain for a real controller the change in PV needs to be expressed in percent of span of the PV as this is how the controller calculates error. ∆PV COSpan Gain = x ∆CO PVSpan Copyright © 2007 Control Station, Inc. All Rights Reserved 45
  • 46. Additional Notes on Process Gain Converting Units of Process Gain Assuming the process gain of the gravity drained tanks is 0.1 m/% then, to convert to %/% m span CO m 0 − 100% % 0 .1 ⋅ = 0 .1 ⋅ = 1 .0 % span PV % 0 − 10% % Copyright © 2007 Control Station, Inc. All Rights Reserved 46
  • 47. Additional Notes on Process Gain Values for Process Gain Process gain that the controller sees is influenced by two factors other than the process itself, the size of the final control element and the span of the sensor. In the ideal world you would use the full span of both final control element and the sensor which would give a process gain of 1.0. Copyright © 2007 Control Station, Inc. All Rights Reserved 47
  • 48. Additional Notes on Process Gain Values for Process Gain As a rule of thumb: Process gains that are greater than 1 are a result of oversized final control elements. Process gains less than 1 are a result of sensor spans that are too wide. Copyright © 2007 Control Station, Inc. All Rights Reserved 48
  • 49. Additional Notes on Process Gain Values for Process Gain The result of a final control element being too large (high gain) is: The controller gain will have to be made correspondingly smaller, smaller than the controller may accept. High gains in the final control element amplify imperfections (deadband, stiction), control errors become proportionately larger. If a sensor has too wide of a span: You may experience problems with the quality of the measurement. The controller gain will have to be made correspondingly larger making the controller more jumpy and amplifying signal noise. An over spanned sensor can hide an oversized final element. Copyright © 2007 Control Station, Inc. All Rights Reserved 49
  • 50. Additional Notes on Process Gain Values for Process Gain The general rule of thumb is the process gain for a self regulating process should be between 0.5 and 2.0. Copyright © 2007 Control Station, Inc. All Rights Reserved 50
  • 51. Process Time Constant τ p ) from Step Test Data ( Copyright © 2007 by Control Station, Inc. All Rights Reserved Time Constant, τ p , describes how fast the measured PV responds to changes in the CO More specifically, how long it takes the PV to reach 63.2% of its total final change (∆PV), starting from when it first begins to respond Copyright © 2007 Control Station, Inc. All Rights Reserved 51
  • 52. Process Time Constant τ p ) from Step Test Data ( Copyright © 2007 by Control Station, Inc. All Rights Reserved t PVstart 1) Locate t PVstart , the time where the PV starts a first clear response to the step change in CO Copyright © 2007 Control Station, Inc. All Rights Reserved 52
  • 53. Process Time Constant τ p ) from Step Test Data ( PV63.2 ∆PV total 63.2% of ∆PV total Copyright © 2007 by Control Station, Inc. All Rights Reserved t PVstart 2) Compute 63.2% of the total change in PV as: PV63.2 = PVinital + 0.632 ∆PV Copyright © 2007 Control Station, Inc. All Rights Reserved 53
  • 54. Process Time Constant τ p ) from Step Test Data ( PV63.2 τP Copyright © 2007 by Control Station, Inc. All Rights Reserved t PVstart t 63.2 3) t 63.2 is the time when the PV reaches PV63.2 4) Time Constant, τ p , is then: t 63.2 — tPVstart Copyright © 2007 Control Station, Inc. All Rights Reserved 54
  • 55. Time Constant (τ p) for Gravity-Drained Tanks Copyright © 2007 by Control Station, Inc. All Rights Reserved tPVstart = 4.1 Here, tPVstart = 4.1 min Copyright © 2007 Control Station, Inc. All Rights Reserved 55
  • 56. Time Constant (τ p ) for Gravity-Drained Tanks PV63.2 = 2.5 ∆PV = 1.0 m 63.2% of ∆PV total Copyright © 2007 by Control Station, Inc. All Rights Reserved 4.1 PV63.2 = PVinital + 0.632 ∆PV = 1.9 + 0.632(1.0) = 2.5 m Copyright © 2007 Control Station, Inc. All Rights Reserved 56
  • 57. Time Constant (τ p ) for Gravity-Drained Tanks PV63.2 = 2.5 τP τP = 5.7 – 4.1 = 1.6 min Copyright © 2007 by Control Station, Inc. All Rights Reserved 4.1 5.7 Time Constant, τ p = t 63.2 — tPVstart = 1.6 min τ p must be positive and have units of time Copyright © 2007 Control Station, Inc. All Rights Reserved 57
  • 58. Process Dead-Time (Өp) from Step Test Data Dead time, Өp, is how much delay occurs from the time when the CO step is made until when the measured PV shows a first clear response. Өp is the sum of these effects: transportation lag, or the time it takes for material to travel from one point to another sample or instrument lag, or the time it takes to collect, analyze or process a measured PV sample higher order processes naturally appear slow to respond and this is treated as dead time Dead time, Өp, must be positive and have units of time Copyright © 2007 Control Station, Inc. All Rights Reserved 58
  • 59. Dead-Time (Өp) is the Killer of Control Tight control grows increasingly difficult as Өp becomes large The process time constant is the clock of the process. Dead time is large or small relative only to τ p When dead time grows such that Өp > τ p , model predictive control strategies such as a Smith predictor may show benefit For important PVs, work to select, locate and maintain instrumentation so as to avoid unnecessary dead time in a loop Copyright © 2007 Control Station, Inc. All Rights Reserved 59
  • 60. Process Dead-Time (Өp) from Step Test Data Copyright © 2007 by Control Station, Inc. All Rights Reserved Өp tCOstep t PVstart Өp = t PVstart — tCOstep Copyright © 2007 Control Station, Inc. All Rights Reserved 60
  • 61. Dead-Time (Өp) for Gravity Drained Tanks Өp = 4.1 – 3.8 = 0.3 min Copyright © 2007 by Control Station, Inc. All Rights Reserved tCOstep = 3.8 t PVstart= 4.1 Өp = t PVstart — tCOstep = 0.3 min Copyright © 2007 Control Station, Inc. All Rights Reserved 61
  • 62. The FOPDT Model Parameters In Summary For a change in CO: Process Gain, Kp → How Far PV travels Time Constant, τ p → How Fast PV responds Dead Time, Өp → How Much Delay Before PV Responds Copyright © 2007 Control Station, Inc. All Rights Reserved 62
  • 63. Hands-On Workshop Workshop #1 Exploring Dynamics of Gravity-Drained Tanks Copyright © 2007 Control Station, Inc. All Rights Reserved 63
  • 64. Processes have Time Varying Behavior The CO to PV behavior described by an ideal FOPDT model is constant, but real processes change every day because: surfaces foul or corrode mechanical elements like seals or bearings wear feedstock quality varies and catalyst activity decays environmental conditions like heat and humidity change So the values of Kp, τ p and Өp that best describe the dynamic behavior of a process today may not be best tomorrow As a result, controller performance can degrade with time and periodic retuning may be required Copyright © 2007 Control Station, Inc. All Rights Reserved 64
  • 65. Processes have Nonlinear Behavior …cause changing (nonlinear) response in real processes equal CO steps… Copyright © 2007 by Control Station, Inc. All Rights Reserved The dynamic behavior of most real processes changes as operating level changes Copyright © 2007 Control Station, Inc. All Rights Reserved 65
  • 66. Processes have Nonlinear Behavior 3) but behavior of real processes changes with operating level 2) FOPDT model response is constant as operating level changes 1) with equal CO steps Copyright © 2007 by Control Station, Inc. All Rights Reserved A FOPDT model response is constant as operating level changes Since the FOPDT model is used for controller design and tuning, a process should be modeled at a specific design level of operation! Copyright © 2007 Control Station, Inc. All Rights Reserved 66
  • 67. What is a Nonlinear Process While linear processes may be the design goal of process engineers, the reality is that most processes are nonlinear in nature due to nonlinearity in the final control element or the process itself. Example: A heating process is nonlinear because the rate at which heat is transferred between two objects depends on the difference in temperature between the objects. Example: A valve that is linear in the middle of its operating range may become very nonlinear towards its limits. Copyright © 2007 Control Station, Inc. All Rights Reserved 67
  • 68. What is a Nonlinear Process Dealing with Nonlinearity – Set Point Response Copyright © 2007 Control Station, Inc. All Rights Reserved 68
  • 69. What is a Nonlinear Process Dealing with Nonlinearity The robustness of a controller is a measure the range of process values over which the controller provides stable operation. The more nonlinear a process is, the less aggressive you must be in your tuning approach to maintain robustness. Copyright © 2007 Control Station, Inc. All Rights Reserved 69
  • 70. What is Process Action Process action is how the process variable changes with respect to a change in the controller output. Process action is either direct acting or reverse acting. The action of a process is defined by the sign of the process gain. A process with a positive gain is said to be direct acting. A process with a negative gain is said to be reverse acting. Copyright © 2007 Control Station, Inc. All Rights Reserved 70
  • 71. What is Process Action Process and Controller Action The action of a process is important because it will determine the action of the controller. A direct acting process requires a reverse acting controller. Conversely, a reverse acting process requires a direct acting controller. Copyright © 2007 Control Station, Inc. All Rights Reserved 71
  • 72. Fundamentals of Instrumentation & Process Control Interactive Training Workshop
  • 73. Introduction to Instrumentation The intent of this chapter is neither to teach you how to select a particular instrument nor to familiarize you with all of the available types of instruments. The intent of this material is to provide an introduction to commonly measured process variables, including the basic terminology and characteristics relevant to each variable’s role in a control loop. Detailed information and assistance on device selection is typically available directly from the instrumentation supplier. Copyright © 2007 Control Station, Inc. All Rights Reserved 73
  • 74. Introduction to Instrumentation Objectives: What are sensors and transducers? What are the standard instrument signals? What are smart transmitters? What is a low pass filter? What instrument properties affect a process? What is input aliasing? What is instrument noise? How do we measure temperature? How do we measure level? How do we measure level? How do we measure flow? Copyright © 2007 Control Station, Inc. All Rights Reserved 74
  • 75. Introduction to Instrumentation You cannot control what you cannot measure “When you can measure what you are speaking about, and express it in numbers, you know something about it. When you cannot measure it, when you cannot express it in numbers, your knowledge is of a meagre and unsatisfactory kind.” - Lord Kelvin Copyright © 2007 Control Station, Inc. All Rights Reserved 75
  • 76. What are Sensors and Transducers A sensor is a device that has a characteristic that changes in a predictable way when exposed to the stimulus it was designed to detect. A transducer is a device that converts one form of energy into another. Copyright © 2007 Control Station, Inc. All Rights Reserved 76
  • 77. What are Standard Instrumentation Signals Standard instrument signals for controllers to accept as inputs from instrumentation and outputs to final control elements are: pneumatic current loop 0 to 10 volt Copyright © 2007 Control Station, Inc. All Rights Reserved 77
  • 78. What are Standard Instrumentation Signals Pneumatic 3 to 15 psig Before 1960, pneumatic signals were used almost exclusively to transmit measurement and control information. Today, it is still common to find 3 to 15 psig used as the final signal to a modulating valve. Most often an I/P (I to P) transducer is used. This converts a 4-20 mA signal (I) into a pressure signal (P). Copyright © 2007 Control Station, Inc. All Rights Reserved 78
  • 79. What are Standard Instrumentation Signals Pneumatic Scaling What would our pneumatic signal be if our controller output is 40%? ( Signal psig = % Controller Output x 12 psig + 3psig ) Copyright © 2007 Control Station, Inc. All Rights Reserved 79
  • 80. What are Standard Instrumentation Signals Current Loop 4-20 milliamp Current loops are the signal workhorses in our processes. A DC milliamp current is transmitted through a pair of wires from a sensor to a controller or from a controller to its final control element. Current loops are used because of their immunity to noise and the distances that the signal can be transmitted. Copyright © 2007 Control Station, Inc. All Rights Reserved 80
  • 81. What are Standard Instrumentation Signals Current Loop Scaling Output Scaling Scale outputs for a one-to-one correspondence. Controller output is configured for 0% to correspond to a 4mA signal and 100% to correspond to a 20mA signal. The final control element is calibrated so that 4mA corresponds to its 0% position or speed and 20mA corresponds to its 100% position or speed. Copyright © 2007 Control Station, Inc. All Rights Reserved 81
  • 82. What are Standard Instrumentation Signals Current Loop Scaling Input Scaling Scale inputs for a one-to-one correspondence as well. Example: If we were using a pressure transducer with a required operating range of 0 psig to 100 psig we would calibrate the instrument such that 0 psig would correspond to 4mA output and 100 psig would correspond to a 20mA output. At the controller we would configure the input such that 4mA would correspond to an internal value of 0 psig and 10mA would correspond to an internal value of 100 psig. Copyright © 2007 Control Station, Inc. All Rights Reserved 82
  • 83. What are Standard Instrumentation Signals 0 to 10 Volt 0 to 10 volt is not commonly used in control systems because this signal is susceptible to induced noise and the distance of the instrument or final control element is limited due to voltage drop. You may find 0-10 volt signals used in control systems providing the speed reference to variable speed drives. Copyright © 2007 Control Station, Inc. All Rights Reserved 83
  • 84. What are Smart Transmitters A smart transmitter is a digital device that converts the analog information from a sensor into digital information, allowing the device to simultaneously send and receive information and transmit more than a single value. Smart transmitters, in general, have the following common features: Digital Communications Configuration Re-Ranging Signal Conditioning Self-Diagnosis Copyright © 2007 Control Station, Inc. All Rights Reserved 84
  • 85. What are Smart Transmitters Digital Communications Smart transmitters are capable of digital communications with both a configuration device and a process controller. Digital communications have the advantage of being free of bit errors, the ability to multiple process values and diagnostic information, and the ability to receive commands. Some smart transmitters use a shared channel for analog and digital data (Hart, Honeywell or Modbus over 4-20mA). Others use a dedicated communication bus (Profibus, Foundation Fieldbus, DeviceNet, Ethernet). Copyright © 2007 Control Station, Inc. All Rights Reserved 85
  • 86. What are Smart Transmitters Digital Communications Most smart instruments wired to multi-channel input cards require isolated inputs for the digital communications to work. Copyright © 2007 Control Station, Inc. All Rights Reserved 86
  • 87. What are Smart Transmitters Configuration, Signal Conditioning, Self-Diagnosis Configuration Smart transmitters can be configured with a handheld terminal and store the configuration settings in non-volatile memory. Signal Conditioning Smart transmitters can perform noise filtering and can provide different signal characterizations. Self-Diagnosis Smart transmitters also have self-diagnostic capability and can report malfunctions that may indicate erroneous process values. Copyright © 2007 Control Station, Inc. All Rights Reserved 87
  • 88. What Instrument Properties Affect a Process The instrument’s range and span. The resolution of the measurement. The instrument’s accuracy and precision. The instrument’s dynamics Copyright © 2007 Control Station, Inc. All Rights Reserved 88
  • 89. What Instrument Properties Affect a Process Range and Span The range of a sensor is the lowest and highest values it can measure within its specification. The span of a sensor is the high end of the Range minus the low end of the Range. Match Range to Expected Conditions Instruments should be selected with a range that includes all values a process will normally encounter, including expected disturbances and possible failures. Copyright © 2007 Control Station, Inc. All Rights Reserved 89
  • 90. What Instrument Properties Affect a Process Measurement Resolution Resolution is the smallest amount of input signal change that the instrument can detect reliably. Resolution is really a function of the instrument span and the controller’s input capability. The resolution of a 16 bit conversion is: Input Span The resolution of a 12 bit conversion is: 65,535 The bit error. Input Span 4,095 Copyright © 2007 Control Station, Inc. All Rights Reserved 90
  • 91. What Instrument Properties Affect a Process Accuracy Accuracy of a measurement describes how close the measurement approaches the true value of the process variable. % error over a range ± x% over  % of full scale  % of span Absolute over a range ± x units over  full scale  span Copyright © 2007 Control Station, Inc. All Rights Reserved 91
  • 92. What Instrument Properties Affect a Process Precision Precision is the reproducibility with which repeated measurements can be made under identical conditions. This may be referred to as drift. Copyright © 2007 Control Station, Inc. All Rights Reserved 92
  • 93. What Instrument Properties Affect a Process Accuracy vs. Precision Why is precision preferred over accuracy? Copyright © 2007 Control Station, Inc. All Rights Reserved 93
  • 94. What Instrument Properties Affect a Process Instrumentation Dynamics: Gain and Dead Time The gain of an instrument is often called sensitivity. The sensitivity of a sensor is the ratio of the output signal to the change in process variable. The dead time of an instrument is the time it takes for an instrument to start reacting to process change. Copyright © 2007 Control Station, Inc. All Rights Reserved 94
  • 95. What Instrument Properties Affect a Process Instrumentation Dynamics: Time Constant As for processes, one time constant for an instrument is the time it takes to provide a signal that represents 63.2% of the value of variable it is measuring after a step change in the variable. Instrument manufacturers may sometimes specify the rise time instead of the time constant. Rise time is the time it takes for an instrument to provide a signal that represents 100% of the value of the variable it is measuring after a step change in the variable. The rise time of an instrument is equal to 5 time constants. Copyright © 2007 Control Station, Inc. All Rights Reserved 95
  • 96. What is Input Aliasing Input aliasing is a phenomenon that occurs from digital processing of a signal. When a signal is processed digitally it is sampled at discrete intervals of time. If the frequency at which a signal is sampled is not fast enough the digital representation of that signal will not be correct. Input aliasing is an important consideration in digital process control. Processor inputs that have configurable sample rates and PID loop update times must be set correctly. Copyright © 2007 Control Station, Inc. All Rights Reserved 96
  • 97. What is Input Aliasing 2 Hz waveform sampled every 0.4 seconds Copyright © 2007 Control Station, Inc. All Rights Reserved 97
  • 98. What is Input Aliasing Correct Sampling Frequency Nyquist Frequency Theorem: To correctly sample a waveform it is necessary to sample at least twice as fast has the highest frequency in the waveform. In the digital world this means sampling at 1/20th of the period of the waveform. Period = 1/frequency Copyright © 2007 Control Station, Inc. All Rights Reserved 98
  • 99. What is Input Aliasing Correct Sampling Frequency 2 Hz waveform sampled every 0.025 seconds Copyright © 2007 Control Station, Inc. All Rights Reserved 99
  • 100. Workshop Lab: Input Aliasing 1 Copyright © 2007 Control Station, Inc. All Rights Reserved 100
  • 101. What is Input Aliasing Determining the Correct Sampling Interval Rules of Thumb Set the sample interval for an instrument at 1/10th to 1/20th of the rise time (1/2 to 1/4th of the time constant). Set the sample interval to 1/10th to 1/20th of the process time constant. Temperature instrumentation (RTDs and thermocouples in thermowells) typically have time constants of several seconds or more. For these processes sampling intervals of 1 second are usually sufficient. Pressure and flow instrumentation typically have time constants of ½ to 1 second. For these processes sampling intervals of 0.1 second are usually sufficient. Copyright © 2007 Control Station, Inc. All Rights Reserved 101
  • 102. What is Input Aliasing Determining the Correct Sampling Interval Copyright © 2007 Control Station, Inc. All Rights Reserved 102
  • 103. Workshop Lab: Input Aliasing 2 Copyright © 2007 Control Station, Inc. All Rights Reserved 103
  • 104. What is Instrument Noise Noise is a variation in a measurement of a process variable that does not reflect real changes in the process variable. Copyright © 2007 Control Station, Inc. All Rights Reserved 104
  • 105. What is Instrument Noise Sources Noise is generally a result of the technology used to sense the process variable. Electrical signals used to transmit instrument measurements are susceptible to having noise induced from other electrical devices. Noise can also be caused by wear and tear on the mechanical elements of a sensor. Noise may be uncontrolled, random variations in the process itself. Whatever the source, noise distorts the measurement signal. Copyright © 2007 Control Station, Inc. All Rights Reserved 105
  • 106. What is Instrument Noise Effects of Noise Noise reduces the accuracy and precision of process measurements. Somewhere in the noise is the true measurement, but where? Noise introduces more uncertainty into the measurement. Noise also introduces errors in control systems. To a controller, fluctuations in the process variable due to noise are indistinguishable from fluctuations caused by real disturbances. Noise in a process variable will be reflected in the output of the controller Copyright © 2007 Control Station, Inc. All Rights Reserved 106
  • 107. What is Instrument Noise Eliminating Noise The most effective means of eliminating noise is to remove the source. Reduce electrically induced noise by following proper grounding techniques, using shielded cabling, and physically separating the signal cabling form other electrical wiring. If worn mechanical elements in the sensor are causing noise, then repair or replace the sensor. When these steps have been taken and excessive noise is still a problem in the process variable a low pass filter may be appropriate. Copyright © 2007 Control Station, Inc. All Rights Reserved 107
  • 108. What is Instrument Noise Low Pass Filter A low-pass filter allows the low frequency components of a signal to pass while attenuating the higher frequency components. Raw (unfiltered) Signal Filtered Signal Copyright © 2007 Control Station, Inc. All Rights Reserved 108
  • 109. What is Instrument Noise Low Pass Filter A low pass filter introduces a first order lag into the measurement. Raw (unfiltered) Signal Filtered Signal Copyright © 2007 Control Station, Inc. All Rights Reserved 109
  • 110. What is Instrument Noise Selecting a Low Pass Filter By Cut-Off Frequency Cut-off frequency is defined as frequency above which the filter provides -3dB of signal attenuation.  Amplitude Out  dB of attenuation = 20 log10    Amplitude In  An attenuation of 0 dB would mean the signal will pass with no reduction in amplitude while a large negative dB would indicate a very small amplitude ratio. Select a cut-off frequency that is above the frequency of your process. Copyright © 2007 Control Station, Inc. All Rights Reserved 110
  • 111. What is Instrument Noise Selecting a Low Pass Filter By Time Constant Some filters are configured by selecting a time constant for the lag response of the filter. The relationship between the cut-off frequency and the time constant of a low pass is approximately given by: 1 Cut - Off Frequency ≈ 5 Time Constants Copyright © 2007 Control Station, Inc. All Rights Reserved 111
  • 112. What is Instrument Noise Selecting a Low Pass Filter By α Value Some filters are selected by a value called alpha (α), notably the derivative filter in a PID controller. α is an averaging weighting term used in controller calculations to impart a first order lag on the measured variable. α generally has values between 0 and 1. A filter with a α = 0 would pass the signal through unfiltered. A filter with a α = 1 would filter everything allowing nothing to pass through the filter. Copyright © 2007 Control Station, Inc. All Rights Reserved 112
  • 113. What is Instrument Noise Selecting a Low Pass Filter When a filter is specified by cut-off frequency, the lower the frequency the greater the filtering effect. When a filter is specified by time constant, the greater the time constant the greater the filtering effect. When a filter is specified by α, when α=0 no filtering is done, when α = 1 no signal passes through the fitter. Copyright © 2007 Control Station, Inc. All Rights Reserved 113
  • 114. Workshop Lab: Noise Filtering Copyright © 2007 Control Station, Inc. All Rights Reserved 114
  • 115. What is Temperature Temperature is a measure of degree of the hotness or coldness of an object. Units of Temperature Two most common temperature scales are Fahrenheit (°F) and Celsius (°C). The reference points are the freezing point and the boiling point of water. Water freezes at 32°F and boils at 212°F. The Celsius scale uses the same reference points only it defines the freezing point of water as 0°C and the boiling point as 100°C. Copyright © 2007 Control Station, Inc. All Rights Reserved 115
  • 116. What is Temperature Unit Conversion 9  °F =  x °C  + 32 5  5 °C = ( x °F - 32 ) 9 Copyright © 2007 Control Station, Inc. All Rights Reserved 116
  • 117. What Temperature Instruments Do We Use Temperature is one of the most common process variables. Temperature is most commonly measured by Resistance Temperature Devices (RTD) Thermocouples Infrared (IR) (to a lesser degree) Thermistors (may also be found embedded in some control equipment) Copyright © 2007 Control Station, Inc. All Rights Reserved 117
  • 118. What Temperature Instruments Do We Use Thermocouples Thermocouples are fabricated from two electrical conductors made of two different metal alloys. Hot or sensing junction Cold or reference junction Generate an open-circuit voltage, called the Seebeck voltage that is proportional to the temperature difference between the sensing (hot) and reference (cold) junctions. Copyright © 2007 Control Station, Inc. All Rights Reserved 118
  • 119. What Temperature Instruments Do We Use Thermocouple Junctions There is a misconception of how thermocouples operate. The misconception is that the hot junction is the source of the output voltage. This is wrong. The voltage is generated across the length of the wire. Another misconception is that junction voltages are generated at the cold end between the special thermocouple wire and the copper circuit. Hence, a cold junction temperature measurement is required. This concept is wrong. The cold end temperature is the reference point for measuring the temperature difference across the length of the thermocouple circuit. Copyright © 2007 Control Station, Inc. All Rights Reserved 119
  • 120. What Temperature Instruments Do We Use Thermocouple Junctions Hot junction voltage proportional to 210°F. Extension wire voltage proportional to 18°F. Copyright © 2007 Control Station, Inc. All Rights Reserved 120
  • 121. What Temperature Instruments Do We Use Thermocouple Junctions Use of the correct extension wire is critical in thermocouple applications. An incorrect extension will cause the temperature differential across the extension leads to be introduced as measurement error. Does this mean you cannot use copper terminal blocks for thermocouples? No, it is highly unlikely there will be a temperature differential across the terminal block; no error will be introduced. Copyright © 2007 Control Station, Inc. All Rights Reserved 121
  • 122. What Temperature Instruments Do We Use Thermocouple Linearization Copyright © 2007 Control Station, Inc. All Rights Reserved 122
  • 123. What Temperature Instruments Do We Use Thermocouple Gain Copyright © 2007 Control Station, Inc. All Rights Reserved 123
  • 124. What Temperature Instruments Do We Use Thermocouple Types Copyright © 2007 Control Station, Inc. All Rights Reserved 124
  • 125. What Temperature Instruments Do We Use RTDs A resistance-temperature detector (RTD) is a temperature sensing device whose resistance increases with temperature. An RTD consists of a wire coil or deposited film of pure metal whose resistance at various temperatures has been documented. RTDs are used when applications require accuracy, long- term stability, linearity and repeatability. 100 Ω Platinum is most common. Copyright © 2007 Control Station, Inc. All Rights Reserved 125
  • 126. What Temperature Instruments Do We Use RTDs: Importance of α α is, roughly, the slope of the RTDs resistance curve with respect to temperature. R100 − R0 α = 100 x R0 A 100Ω platinum RTD with a =0.003911 will have a resistance of 139.11 ohms at 100°C. A 100 Ω platinum RTD with a =0.003850 will have a resistance of 138.50 ohms at 100°C. Common values for the temperature coefficient are: 0.003850, DIN 43760 Standard (also referred to as European curve) 0.003911, American Standard 0.003926, ITS-90 Standard Copyright © 2007 Control Station, Inc. All Rights Reserved 126
  • 127. What Temperature Instruments Do We Use RTD Lead Wire Resistance The RTD is a resistive device, you must drive a current through the device and monitor the resulting voltage. Any resistance in the lead wires that connect your measurement system to the RTD will add error to your readings. 40 feet of 18 gauge 2 conductor cable has 6Ω resistance. For a platinum RTD with α = 0.00385, the resistance equals 0.6 Ω /(0.385 Ω /°C) = 1.6°C of error. Copyright © 2007 Control Station, Inc. All Rights Reserved 127
  • 128. What Temperature Instruments Do We Use RTD Lead Wire Resistance Error is eliminated through the use of three wire RTDs. With three leads, two voltage measurements can be made. When the voltages are subtracted, the result is the voltage drop that would have occurred through RT alone. Copyright © 2007 Control Station, Inc. All Rights Reserved 128
  • 129. What Temperature Instruments Do We Use RTD Self Heating Effect The current used to excite an RTD causes the RTD to internally heat, which appears as an error. Self-heating is typically specified as the amount of power that will raise the RTD temperature by 1° C, or 1 mW/ °C. Self-heating can be minimized by using the smallest possible excitation current, but this occurs at the expense of lowering the measurable voltages and making the signal more susceptible to noise from induced voltages. The amount of self-heating also depends heavily on the medium in which the RTD is immersed. An RTD can self- heat up to 100 times higher in still air than in moving water Copyright © 2007 Control Station, Inc. All Rights Reserved 129
  • 130. What Temperature Instruments Do We Use Thermistors A thermistor is similar to an RTD in that it is a passive resistance device. Thermistors are generally made of semiconductor materials giving them much different characteristics. Thermistors do not have standardized electrical properties like thermocouples or RTDs. Copyright © 2007 Control Station, Inc. All Rights Reserved 130
  • 131. What Temperature Instruments Do We Use Infrared Objects radiate electromagnetic energy. The higher the temperature of an object the more electromagnetic radiation it emits. This radiation occurs within the infrared portion of the electromagnetic spectrum. Copyright © 2007 Control Station, Inc. All Rights Reserved 131
  • 132. What Temperature Instruments Do We Use Infrared Emittance The emittance of a real surface is the ratio of the thermal radiation emitted by the surface at a given temperature to that of the ideal black body at the same temperature. By definition, a black body has an emittance of 1. Another way to think of emissivity is the efficiency at which an object radiates thermal energy. Emittance is a decimal number that ranges between 0 and 1 or a percentage between 0% and 100%. Copyright © 2007 Control Station, Inc. All Rights Reserved 132
  • 133. What Temperature Instruments Do We Use Infrared Emittance To function properly an infrared temperature instrument must take into account the emittance of the surface being measured. Emittance values can often be found in reference tables, but such tables will not take into account local conditions of the surface. A more practical way to set the emittance is to measure the temperature with an RTD or thermocouple and set the instrument emissivity control so that both readings are the same. Copyright © 2007 Control Station, Inc. All Rights Reserved 133
  • 134. What Temperature Instruments Do We Use Infrared Field of View Infrared temperature instruments are like an optical system in that they have a field of view. The field of view basically defines the target size at a given distance. Field of view may be specified as: Angle and focal range (2.3° from 8" to 14') Distance to spot size ratio and focal range (25:1 from 8" to 14') Spot size at a distance (0.32" diameter spot at 8") All of these specifications are equivalent, at 8 inches our field of view will be a 0.32" diameter spot (8"/25), at 14 feet our field of view will be 6.72" (14' / 25). An infrared temperature sensor measures the average temperature of everything in its field of view. If the surface whose temperature we are measuring does not completely fill the field of view we will get inaccurate results. Copyright © 2007 Control Station, Inc. All Rights Reserved 134
  • 135. What Temperature Instruments Do We Use Infrared Spectral Response The range of frequencies that an infrared sensor can measure is called its spectral response. Infrared temperature sensors do not measure the entire infrared region. Different frequencies (wavelengths) are selected for detection to provide certain advantages. If you wanted to use an infrared temperature sensor to measure the temperature of an object behind glass you would need to select an instrument with a spectral response in the infrared region in which glass is "transparent”. Copyright © 2007 Control Station, Inc. All Rights Reserved 135
  • 136. Section Assessment: Temperature Copyright © 2007 Control Station, Inc. All Rights Reserved 136
  • 137. What is Pressure Pressure is the ratio between a force acting on a surface and the area of that surface. Pressure is measured in units of force divided by area. psi (pounds per square inch) bar, 1 bar = 14.5 PSI, common for pump ratings inH20 (inches of water), 27.680 inH20 = 1 psi, common for vacuum systems and tank levels mmHg (millimeters of mercury), 760 mmHg = 14.7 psi, common for vacuum systems Copyright © 2007 Control Station, Inc. All Rights Reserved 137
  • 138. What is Pressure Absolute, Gauge and Differential Gauge pressure is defined relative to atmospheric conditions. The units of gauge pressure are psig, however gauge pressure is often denoted by psi as well. Absolute pressure is defined as the pressure relative to an absolute vacuum. The units of absolute pressure are psia. Differential pressure uses a reference point other than full vacuum or atmospheric pressure. Copyright © 2007 Control Station, Inc. All Rights Reserved 138
  • 139. What is Pressure Absolute, Gauge and Differential Copyright © 2007 Control Station, Inc. All Rights Reserved 139
  • 140. Section Assessment: Pressure Copyright © 2007 Control Station, Inc. All Rights Reserved 140
  • 141. Common Level Sensing Technologies Point sensing level probes only sense tank level at a discrete level. Typically used for high-high or low-low level sensing to prevent plant personnel and/or process equipment from being exposed to harmful conditions. Also used in pairs in processes in which we do not particularly care what the exact level in a tank is, only that it is between two points. Continuous level probes sense the tank level as a percent of span of the probes capabilities. Continuous level probes are typically used where we need some type of inventory control, where we need to know with some degree of confidence what the particular level in a tank is. Contact and Non-Contact Types Copyright © 2007 Control Station, Inc. All Rights Reserved 141
  • 142. Common Level Sensing Technologies Non-Contact: Ultrasonic Ultrasonic makes use of sound waves. A transducer mounted in the top of a tank transmits sound waves in bursts onto the surface of the material to be measured. Echoes are reflected back from the surface of the material to the transducer and the distance to the surface is calculated from the burst-echo timing. The key points in applying an ultrasonic transducer are: The speed of sound varies with temperature. Heavy foam on the surface of the material interferes with the echo. An irregular material surface can cause false echoes resulting in irregular readings. Heavy vapor in the air space can distort the sound waves resulting in false reading. Copyright © 2007 Control Station, Inc. All Rights Reserved 142
  • 143. Common Level Sensing Technologies Non-Contact: Ultrasonic Copyright © 2007 Control Station, Inc. All Rights Reserved 143
  • 144. Common Level Sensing Technologies Non-Contact: Radar/Microwave Radar, or microwave level measurement, operates on similar principles to ultrasonic level probes but, instead of sound waves, electromagnetic waves in the 10GHz range are used. When properly selected, radar can overcome many of the limitations of ultrasonic level probes. Be unaffected by temperature changes in the tank air space. See through heavy foam to detect the true material level. See through heavy vapor in the tank air space to detect true material level. Copyright © 2007 Control Station, Inc. All Rights Reserved 144
  • 145. Common Level Sensing Technologies Non-Contact: Radar/Microwave Copyright © 2007 Control Station, Inc. All Rights Reserved 145
  • 146. Common Level Sensing Technologies Non-Contact: Nuclear Level Measurement Radiation from the source is detected on the other side of the tank. Its strength indicates the level of the fluid. Point, continuous, and interface measurements can be made. As no penetration of the vessel is needed there are a number of situations that cause nucleonic transmitters to be considered over other technologies. Nuclear level detection has some drawbacks. One is high cost, up to four times that of other technologies. Others are the probable requirement for licenses, approvals, and periodic inspections; and the difficulty and expense of disposing of spent radiation materials. Copyright © 2007 Control Station, Inc. All Rights Reserved 146
  • 147. Common Level Sensing Technologies Non-Contact: Nuclear Level Measurement Copyright © 2007 Control Station, Inc. All Rights Reserved 147
  • 148. Common Level Sensing Technologies Contact: Hydrostatic Pressure Measurement of level by pressure relies on hydrostatic principles. Pressure is a unit force over a unit area. A cubic foot (12"L x 12"W x 12"H) of water weighs 62.4796 pounds. The area that our cubic foot of water occupies is 144 square inches (12"L x 12"H), therefore our cubic foot of water exerts a force of 62.4796 pounds over 144 square inches, or 0.4339 psig for a 12" water column. 1 inch H20 = 0.0316 psig It would not matter how many cubic feet of water were placed side by side, our pressure would still be 0.4399 psig. Hydrostatic pressure is only dependent on the height of the fluid, not the area that it covers. Copyright © 2007 Control Station, Inc. All Rights Reserved 148
  • 149. Common Level Sensing Technologies Contact: Hydrostatic Pressure How would you handle measuring other fluids using pressure? Compare the densities. For instance, chocolate weighs approximately 80 pounds per cubic foot. 80 divided 62.5 times 0.0316 = 0.0404 psig. For a change in level of one inch in a liquid chocolate tank the pressure measurement will change by 0.0404 psig. Level measurement by pressure requires a constant density for accurate measurements. If the head pressure in the tank can be other than atmospheric, we must use a differential pressure sensor. Copyright © 2007 Control Station, Inc. All Rights Reserved 149
  • 150. Common Level Sensing Technologies Contact: Hydrostatic Pressure Copyright © 2007 Control Station, Inc. All Rights Reserved 150
  • 151. Common Level Sensing Technologies Contact: RF/Capacitance RF (radio frequency) Capacitance level sensors make use of electrical characteristics of a capacitor to infer the level in a vessel. As the material rises in the vessel, the capacitance changes. The level transducer measures this change, linearizes it and transmits the signal to the process control system. A point level probe will look for a specific change in capacitance to determine whether it is on or off. Copyright © 2007 Control Station, Inc. All Rights Reserved 151
  • 152. Common Level Sensing Technologies Contact: RF/Capacitance Requires the material being measured to have a high dielectric constant. Level measurements are affected by changes in the dielectric of the material (moisture content). Proper selection requires informing the probe vendor of the material to be measure, especially in applications where you are measuring conductive materials or have a nonmetallic tank. Point probe sensitivity can be increased by welding a plate on the sensor tip to increase the capacitance, and therefore the sensitivity (gain). Copyright © 2007 Control Station, Inc. All Rights Reserved 152
  • 153. Common Level Sensing Technologies Contact: RF/Capacitance Copyright © 2007 Control Station, Inc. All Rights Reserved 153
  • 154. Common Level Sensing Technologies Contact: Guided Wave Radar Guided wave radar is similar to the non-contact radar probes, only a rod or cable is immersed into the material like an RF capacitance probe. The rod or cable is used to guide the microwave along its length, where the rod or cable meets the material to be measured a wave reflection is generated. The transit time of the wave is used to calculate level very precisely. Unlike an RF capacitance probe, a guided wave radar probe can measure extremely low dielectric material. Copyright © 2007 Control Station, Inc. All Rights Reserved 154
  • 155. Common Level Sensing Technologies Contact: Guided Wave Radar Copyright © 2007 Control Station, Inc. All Rights Reserved 155
  • 156. Section Assessment: Level Copyright © 2007 Control Station, Inc. All Rights Reserved 156
  • 157. What is Flow Flow is the motion characteristics of constrained fluids (liquids or gases). Fluid velocity or mass are typically not measured directly. Measurements are affected by the properties of the fluid, the flow stream, and the physical installation. Copyright © 2007 Control Station, Inc. All Rights Reserved 157
  • 158. Factors Affecting Flow Measurement The critical factors affecting flow measurement are: Viscosity Fluid Type Reynolds Number Flow Irregularities Copyright © 2007 Control Station, Inc. All Rights Reserved 158
  • 159. Factors Affecting Flow Measurement Viscosity Dynamic or absolute viscosity (η) is a measure of the resistance to a fluid to deformation under shear stress, or an internal property of a fluid that offers resistance to flow. Commonly perceived as “thickness” or resistance to pouring. Water is very “thin” having a relatively low viscosity, while molasses is very thick having a relatively high viscosity. Copyright © 2007 Control Station, Inc. All Rights Reserved 159
  • 160. Factors Affecting Flow Measurement Viscosity Visualize a flat sheet of glass on a film of oil on top of a flat surface. A parallel force applied to the sheet of glass will accelerate it to a final velocity dependent only on the amount of force applied. The oil that is next to the sheet of glass will have a velocity close to that of the glass; while the oil that is next to the stationary surface will have a velocity near zero. This internal distribution of velocities is due to the internal resistance of the fluid to shear stress forces, its viscosity. The viscosity of a fluid is the ratio between the per unit force to accelerate the plate and the distribution of the velocities within the fluid film Copyright © 2007 Control Station, Inc. All Rights Reserved 160
  • 161. Factors Affecting Flow Measurement Viscosity Effect of Temperature The dynamic viscosity of a fluid varies with its temperature. In general, the viscosity of a liquid will decrease with increasing temperature while the viscosity of a gas will increase with increasing temperature. Viscosity measurements are therefore associated with a particular temperature Copyright © 2007 Control Station, Inc. All Rights Reserved 161
  • 162. Factors Affecting Flow Measurement Viscosity Viscosity is measured in units of poise or Pascal·seconds or stokes. Poise or Pascal·seconds are the units of dynamic viscosity. Stokes are the units of kinematic viscosity. Kinematic viscosity is the dynamic viscosity of a fluid divided by the density of the fluid. Copyright © 2007 Control Station, Inc. All Rights Reserved 162
  • 163. Factors Affecting Flow Measurement Viscosity Conversion Chart Copyright © 2007 Control Station, Inc. All Rights Reserved 163
  • 164. Factors Affecting Flow Measurement Viscosity of Common Fluids Copyright © 2007 Control Station, Inc. All Rights Reserved 164
  • 165. Factors Affecting Flow Measurement Viscosity The fluid streams in most processes can include both high and low viscous fluids. The viscosity of the fluid under process conditions must be taken into account when selecting a flow instrument for optimum performance. Copyright © 2007 Control Station, Inc. All Rights Reserved 165
  • 166. Factors Affecting Flow Measurement Fluid Type Newtonian Fluids When held at a constant temperature, the viscosity of a Non-Newtonian fluid will change with relation to the size of the shear force, or it will change over time under a constant shear force. Water, glycerin, liquid sugar, and corn syrup are examples. Copyright © 2007 Control Station, Inc. All Rights Reserved 166
  • 167. Factors Affecting Flow Measurement Fluid Type Non-Newtonian Fluids When held at a constant temperature, the viscosity of a Non-Newtonian fluid will change with relation to the size of the shear force, or it will change over time under a constant shear force. Shear Thickening: Peanut Butter, Starch & Water Time Thickening: Milk, Molasses Copyright © 2007 Control Station, Inc. All Rights Reserved 167
  • 168. Factors Affecting Flow Measurement Reynolds Number The Reynolds number is the ratio of inertial forces to viscous forces of fluid flow within a pipe and is used to determine whether a flow will be laminar or turbulent. Dvρ Where : Re = Reynolds number Re=124 η D = Pipe diameter in inches v = Fluid velocity in ft/sec ρ = Fluid density in lb/ft3 η = Fluid viscosity in cP Copyright © 2007 Control Station, Inc. All Rights Reserved 168
  • 169. Factors Affecting Flow Measurement Laminar Flow Laminar flow occurs at low Reynolds numbers, typically Re < 2000, where viscous forces are dominant. Laminar flow is characterized by layers of flow traveling at different speeds with virtually no mixing between layers. The velocity of the flow is highest in the center of the pipe and lowest at the walls of the pipe. Copyright © 2007 Control Station, Inc. All Rights Reserved 169
  • 170. Factors Affecting Flow Measurement Turbulent Flow Turbulent flow occurs at high Reynolds numbers, typically Re > 4000, where inertial forces are dominant. Turbulent flow is characterized by irregular movement of the fluid in the pipe. There are no definite layers. The velocity of the flow is nearly uniform through the cross-section of the pipe. Copyright © 2007 Control Station, Inc. All Rights Reserved 170
  • 171. Factors Affecting Flow Measurement Transitional Flow Transitional flow typically occurs at Reynolds numbers between 2000 and 4000. Flow in this region may be laminar, it may be turbulent, or it may exhibit characteristics of both. Copyright © 2007 Control Station, Inc. All Rights Reserved 171
  • 172. Factors Affecting Flow Measurement Reynolds Number Flow instruments must be selected to accurately measure laminar flow or turbulent flow Copyright © 2007 Control Station, Inc. All Rights Reserved 172
  • 173. Factors Affecting Flow Measurement Flow Irregularities Flow irregularities are changes in the velocity profile of the fluid flow caused by installation of the flow instrument. Normal velocity profiles are established by installing a sufficient run of straight piping. Copyright © 2007 Control Station, Inc. All Rights Reserved 173
  • 174. Common Flowmeter Technologies Copyright © 2007 Control Station, Inc. All Rights Reserved 174
  • 175. Common Flowmeter Technologies Volumetric Flow Magnetic flowmeters Positive displacement for viscous liquids Orifice Plates Mass Flow Coriolis flowmeters Copyright © 2007 Control Station, Inc. All Rights Reserved 175
  • 176. Common Flowmeter Technologies Volumetric Flow Volumetric flow is usually inferred. Volumetric Flow = Fluid Velocity x Area Measured directly with positive displacement flow meters. Measured indirectly by differential pressure Volumetric flow meters provide a signal that is in units of volume per unit time. gpm (gallons per minute) cfm (cubic feet per minute) l/hr (liters per hour) Copyright © 2007 Control Station, Inc. All Rights Reserved 176
  • 177. Common Flowmeter Technologies Volumetric Flow: Positive Displacement Operate by capturing the fluid to be measured in rotating cavities of a known volume. The volume of the cavity and the rate of rotation will give the volumetric flow value. Unaffected by Reynolds number and work with laminar, turbulent, and transitional flow. Low viscosity fluids will slip past the gears decreasing the accuracy of the flowmeter. Copyright © 2007 Control Station, Inc. All Rights Reserved 177
  • 178. Common Flowmeter Technologies Volumetric Flow: Magnetic Magnetic flowmeters infer the velocity of the moving fluid by measuring a generated voltage. Magnetic flowmeters are based on Faraday’s law of electromagnetic induction - a wire moving though a magnetic field will generate a voltage. Requires a conductive fluid. Flow = Pipe Area ⋅ velocity πD 2 E = kBDv 2 = πr v = v k = Constant 4 B = Magnetic Flux Density πD 2 E πDE D = Pipe Diameter = = 4 kBD 4kB v = Fluid Velocity Copyright © 2007 Control Station, Inc. All Rights Reserved 178
  • 179. Common Flowmeter Technologies Volumetric Flow: Orifice Plates An orifice plate is basically a thin plate with a hole in the middle. It is usually placed in a pipe in which fluid flows. In practice, the orifice plate is installed in the pipe between two flanges. The orifice constricts the flow of liquid to produce a differential pressure across the plate. Pressure taps on either side of the plate are used to detect the difference. Copyright © 2007 Control Station, Inc. All Rights Reserved 179
  • 180. Common Flowmeter Technologies Volumetric Flow: Orifice Plates Orifice Plate type meters only work well when supplied with a fully developed flow profile. This is achieved by a long upstream length (20 to 40 diameters, depending on Reynolds number) Orifice plates are small and cheap to install, but impose a significant energy loss on the fluid due to friction. Copyright © 2007 Control Station, Inc. All Rights Reserved 180
  • 181. Common Flowmeter Technologies Mass Flow Measured directly by measuring the inertial effects of the fluids moving mass. Mass flow meters provide a signal that is in units of mass per unit time. Typical units of mass flow are: lb/hr (pounds per hour) kg/s (kilograms per second) Copyright © 2007 Control Station, Inc. All Rights Reserved 181
  • 182. Common Flowmeter Technologies Mass Flow: Coriolis Coriolis flowmeters are based on the affects of the Coriolis force. When observing motion from a rotating frame the trajectory of motion will appear to be altered by a force arising from the rotation. Copyright © 2007 Control Station, Inc. All Rights Reserved 182
  • 183. Common Flowmeter Technologies Mass Flow: Coriolis Copyright © 2007 Control Station, Inc. All Rights Reserved 183
  • 184. Common Flowmeter Technologies Mass Flow: Coriolis Copyright © 2007 Control Station, Inc. All Rights Reserved 184
  • 185. Common Flowmeter Technologies Mass Flow: Coriolis Coriolis mass flowmeters work by applying a vibrating force to a pair of curved tubes through which fluid passes, in effect creating a rotating frame of reference. The Coriolis Effect of the mass passing through the tubes creates a force on the tubes perpendicular to both the direction of vibration and the direction of flow, causing a twist in the tubes. Copyright © 2007 Control Station, Inc. All Rights Reserved 185
  • 186. Common Flowmeter Technologies Flowmeter Turndown A specification commonly found with flow sensors is turndown or rangeability. Turndown is the ratio of maximum to minimum flow that can be measured with the specified accuracy. For example, a flow meter with a full span of 0 to 100 gallons per minute with a 10:1 turndown could measure a flow as small as 10 gallons per minute. Trying to measure a smaller flow would result in values outside of the manufacturer’s stated accuracy claims. Copyright © 2007 Control Station, Inc. All Rights Reserved 186
  • 187. Common Flowmeter Technologies Installation and Calibration As with any instrument installation, optimal flow meter performance and accuracy can only be achieved through proper installation and calibration. Requirements will vary between type and manufacturer by some general considerations are: The length of straight piping required upstream and downstream of the instrument. This is usually specified in pipe diameters. The properties of the fluid/gas to be measured. Mounting position of the flowmeter. Flowmeters must be kept full during operation. Copyright © 2007 Control Station, Inc. All Rights Reserved 187
  • 188. Common Flowmeter Technologies Installation and Calibration In general, calibration of a flowmeter requires: Entering the sensor characterization constants into the flow transmitter. Recording the zero flow conditions. Many flow meters must be “zeroed” only when full. Comparing the meter flow value to an independently measured value for the flow stream. Measuring the volume or mass of a catch sample and comparing the values. Use the internal flowmeter totalizer value for comparison. Copyright © 2007 Control Station, Inc. All Rights Reserved 188
  • 189. Introduction to Final Control Elements Objectives: What is a control valve? What is a centrifugal What is an actuator? pump? What is a positioner? What is pump head? What is Cv? Why do we use pump head and not psi? What are valve characteristics? What is a pump curve? What is valve deadband? What is a system curve? What is stiction? What is the system What are the types of valves operating point? What is a positive displacement pump? How does a PD pump differ from a centrifugal pump? Copyright © 2007 Control Station, Inc. All Rights Reserved 189
  • 190. Introduction to Final Control Elements The intent of this chapter is neither to teach you how to size a valve or select a pump nor to familiarize you with all of the available types of valves and pumps. The intent of this chapter is to provide an introduction to some of the commonly used pumps and valves, including basic terminology and characteristics relevant to their role in a control loop. Detailed information and assistance on device selection is typically available from your instrumentation suppliers. Copyright © 2007 Control Station, Inc. All Rights Reserved 190
  • 191. What is a Control Valve A control valve is an inline device in a flow stream that receives commands from a controller and manipulates the flow of a gas or fluid in one of three ways: Interrupt flow (Shut-Off Service) Divert flow to another path in the system (Divert Service) Regulate the rate of flow (Throttling Service) Copyright © 2007 Control Station, Inc. All Rights Reserved 191