pH  & Conductivity Parameter Training Measurement, Maintenance & the Future ISA Boston Section March 15, 2011 U E 2 E 3 E 1 Reference- electrolyte E 6 E 5 E 4 Inner Buffer
What is pH ? ,[object Object],[object Object]
Definition: pH =  - log   [a H+   ]     H 2 0  H +  + OH -  [ a H+   ] * [ a  OH-   ] / [a H2O ] = 10  -14 0  1  2  3  4  5  6  7   8  9  10  11  12  13  14  pH acidic: pH  0-6.9  [H + ] > [OH - ] [H + ] > 10 -7  M neutral:  pH  7.0 [H + ] = [OH - ] [H + ]= 10 -7  M  alkaline: pH  7.1 - 14 [H + ] < [OH - ] [H + ] < 10 -7  M pH basics - pH scale
Some examples of pH values 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 lemon juice orange juice beer cheese milk pure water egg white borax Milk of Magnesia H 2 SO 4  (1N) 4.9% HCl (0.1N) 0.37% acetic acid  (0.1N) 0.6% HCN (0.1N) 0.27% sodium bicarbonate  0.84% (0.1N) potassium ac.  0.98% (0.1N) NH 4 OH   0.017% (0.01N) NH 4 OH   1.7% 1.0N NaOH 4%
How does pH measurement work? pH is a potentiometric measurement via an electrochemical sensor/electrode/probe U= E pH  -E ref  (mV) This potential difference is a function of the solution being measured E pH glass  electrode E ref reference  electrode high impedance pH Meter
Combination pH Sensors
What is special about pH glass? H + H + + + + + + + - - - - - - Acidic Alkaline Glass membrane Glass membrane (0.2 - 0.5mm) Gel  layer ca. 1 µm (outer and inner) positive charge negative  charge internal buffer The surface layer of the glass membrane is the “key performer” in each pH measurement! pH is a measurement of the potential difference between inner and outer layer of glass membrane! This is one reason why pH sensors need to be stored in salt solution when not in use!
[object Object],[object Object],[object Object],[object Object],[object Object],What is a combination electrode? pH meter Reference electrolyte Inner buffer Potential of the glass electrode (E 5 ) can’t be  measured individually. A second (reference) electrode is necessary. The potential of this  electrode must be independent of the  sample solution (buffer) Combined pH electrode- glass and reference electrode integrated E 1 E 2 E 3 E 4 E 5
Reference Junction Biotech pH sensors have a reference junction typically composed of ceramic ,[object Object],[object Object]
The Nernst Equation E =  E 0   +  2.303 R T  log  n F  1. Internal Reference Potential 2. Inner Glass/Solution Potential  3. External Reference Potential 4. LIQUID JUNCTION POTENTIAL
The Nernst Equation E =   E 0   +  2.303 R T   log    n F   2.303 R T  is known as “The Electrode Slope” n F   IDEAL SLOPE = 59.16 mV/pH  unit at 25 °C Slope is temperature dependent
The Nernst Response Curve  mV = 59.16   pH mV E = E o  + 2.3 RT/F log a H + where 2.3 RT/F = 0.198T K  = 59.16 mV @ 25 o C
pH Electrode Calibration Curve pH 0  4  7  10  14 + 500 mv 0 mv - 500 mv millivolts pH 10.00 buffer 0.0 millivolts pH 7.00 buffer -177.5 millivolts IDEAL VALUES!
Understanding Temperature Compensation ,[object Object],[object Object],[object Object],[object Object],[object Object],0  o C 25  o C 30  o C 20  o C 40  o C 10  o C    +/- 0.6 pH Isopotential point
Temperature Error (No Temp. Compensation) Temperature error table for pH signal pH °C 2 3 4 5 6 7 8 9 10 11 12 5 0.30 0.24 0.18 0.12 0.06 0 0.06 0.12 0.18 0.24 0.30 15 0.15 0.12 0.09 0.06 0.03 0 0.03 0.06 0.09 0.12 0.15 25 0 0 0 0 0 0 0 0 0 0 0 35 0.15 0.12 0.09 0.06 0.03 0 0.03 0.06 0.09 0.12 0.15 45 0.30 0.24 0.18 0.12 0.06 0 0.06 0.12 0.18 0.24 0.30 55 0.45 0.36 0.27 0.18 0.09 0 0.09 0.18 0.27 0.36 0.45 65 0.60 0.48 0.36 0.24 0.12 0 0.12 0.24 0.35 0.48 0.60 75 0.75 0.60 0.45 0.30 0.15 0 0.15 0.30 0.45 0.60 0.75 85 0.90 0.72 0.54 0.36 0.18 0 0.18 0.36 0.54 0.72 0.90 No temperature error Temperature error < 0.1 pH units Temperature error > 0.1 but < 0.3 pH units Temperature error = or > 0.3 pH units   ( simplified model with Error = /pH (pr.) - 7/ x /(T (pr.) - 25):10/ x 0.03  )
Causes of Failure or Measurement issues ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object]
[object Object],[object Object],[object Object],[object Object],Causes of Failure or Measurement issues
pH Calibration 101- Bracket your process pH pH 0  4  7  10  14 + 500 mv 0 mv - 500 mv millivolts pH 10.00 buffer 0.0 millivolts pH 7.00 buffer -177.5 millivolts Process pH value ,[object Object],[object Object],[object Object],[object Object]
pH Calibration-Impact of buffers ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object]
pH Calibration /Justification 1. Step Determine Asymmetry Potential / Zero Point 2. Step Determine the Slope pH Buffer 7.00  pH Buffer 4.00   [mV] 200 -200 pH 7 14 [mV] 200 -200 pH 7 14 pH as 4 [%] Slope
pH Calibration ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object]
pH Simulator (mV input) ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object]
How frequently should the sensor be calibrated? ,[object Object],[object Object],[object Object],[object Object]
Grab Samples- Why don’t they agree? ,[object Object],[object Object],[object Object],[object Object],[object Object],37 o C 20 o C Same sample
pH Sensor Installation (How to Install) ,[object Object],15 °
pH Sensor Installation (How  NOT  to Install) ,[object Object],Air Bubble NO! HECK NO!
pH Recap ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object]
What is ORP ? ,[object Object],[object Object],[object Object],[object Object]
What is REDOX measurement tell you? ,[object Object],[object Object],[object Object]
ORP (Oxidation Reduction Potential) ,[object Object],[object Object],[object Object],[object Object],[object Object]
ORP (Oxidation Reduction Potential) ,[object Object],[object Object],[object Object],[object Object],[object Object],E = E 0  + kT log [Oxidants][H + ] [Reductants]
pH effect on ORP ,[object Object],[object Object],E = E 0  + kT log [Oxidants] [H + ] [Reductants]
Temperature effect on ORP ,[object Object],[object Object],E = E 0  + k T  log [Oxidants][H + ] [Reductants]
Questions?
What is Conductivity? ,[object Object],[object Object],[object Object],[object Object],Conductivity is the ability of a solution to carry an electric current.
Conductivity Measurement A conductivity “cell”, sensor, or electrode is an “electro-mechanical” measurement 1 cm 1 cm d = 1 cm Cond Solution Electrode plate VAC
Conductivity Advantages and Limitations ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object]
Electrolytes ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object]
Conductivity- An Electro-mechanical Measurement ,[object Object],[object Object],[object Object],Known Known Back Calculate
Conductance and Resistance ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object]
Conductivity Units of Measure ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object]
Conductivity vs. Concentration ,[object Object],[object Object],[object Object],[object Object],[object Object]
Chemical Concentration Control
Chemical Concentration Control
Cell Geometry The measured resistance will be dependent on the spacing of the electrode – cell geometry     “cell constant” Therefore, units measurement has dimension component, ex. mS/cm 1 cm 1 cm d = 1 cm Cond Solution Electrode plate VAC
Measuring Conductivity ~ H H O Na + Cl - H + O H -
Typical Conductivity Values ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object]
Common Measurement Challenges ,[object Object],[object Object],[object Object]
Concentration ,[object Object]
Conductivity Temperature Effects ~ H + H H O O H - Na + Cl - H + O H - Na + Cl -
Temperature Coefficients ,[object Object],Sample %/  o C (at 25  o C) Salt solution (NaCl) 2.12 5% NaOH 1.72 Dilute Ammonia Solution 1.88 10% HCl 1.32 5% Sulfuric Acid 0.96 98% Sulfuric Acid 2.84 Sugar Syrup 5.64
Sensor Cell Constant 0.1  cm 0.1 cm -1 0.1 cm 1 cm Conductivity Cell Constant  =  Length Area 1 cm 1 cm 2 = =  1 cm -1 INSULATOR ELECTRODE ELECTRODE INNER OUTER 1 cm 1 cm
Contacting Conductivity ,[object Object],[object Object],[object Object],[object Object]
Conductivity Traceable Calibration ,[object Object],[object Object]
Conductivity Instrument Calibration ,[object Object],[object Object],[object Object]
Contacting Conductivity ,[object Object],[object Object],[object Object],[object Object],[object Object],V
Four-Electrode Conductivity Measurement AC Current Source AC Voltage Measurement Drive Electrodes Measuring Electrodes Four Electrode Sensor Four-Electrode Measuring Instrument
Four-Electrode Conductivity Measurement ,[object Object],[object Object],[object Object],[object Object]
Inductive Conductivity  ,[object Object],[object Object],[object Object],[object Object],[object Object],Energized  Measured G D induced current
Inductive Conductivity Measurement ,[object Object],[object Object],[object Object],[object Object],[object Object]
Inductive Sensor Installation ,[object Object],metal synthetic Measuring  value  too low Measuring  value  too high Measuring value  correct 30mm / 1.18” 30mm / 1.18”
Cell Installation -2 electrode ,[object Object],[object Object],[object Object],Recommended Cell Installation... Flow should be directed at the end of the sensor
Cell Installation -2 electrode NOT   Recommended Cell Installation...  INLET OUTLET OUTLET AIR INLET Avoid dead legs and air traps
Cell Installation -4 electrode Recommended  OUTLET INLET Maintain a minimum clearance between sensor and pipe
Cell Installation -4 electrode NOT   Recommended Maintain a minimum clearance between sensor and pipe
Conductivity, Resistivity, TDS Ranges Conductivity 100M  10M  1M  100K  10K  1K  100  10  1 Ultrapure water Deionized water Distilled water Condensate Drinking water Cooling tower water Percentage of acids, bases and salt Waste water Brackish water, Sea water Water for Industrial Process 5% Salinity 2% NaOH 20% HCl 0.01  .1  1  10  100  1000  10k  100k  1000k 0.021  0.4  4.6  46  460  4.6k  46k TDS ppm Conductivity and resistivity are measured at 25  C; TDS is expressed as Sodium Chloride (NaCl) Resistivity   ohm-cm µS-cm
Main Applications and Measuring Range Inductive 4 Elec 2 Elec 0.01  0.1  1.0  10  100  1000  10k  100k  1000k  Conductivity    ( µ S/cm) 100 M  10M  1M  100k  10k  1000  100  10  1  Resistivity   (Ohm-cm) Ultra pure water Pure water Make up water Drinking water Diluted acids, bases, salts Waste water Brackish water Industrial process water Acids, bases Water Processes Biotech/Food and Beverage Chemical   Processes
So….where is analytical measurement today?
Previous Analog Technology Digital technology provides better sound quality
Welcome to the Digital World! Same electrochemical end of the sensor converted to digital signal which is more robust and gives more information
Analog Sensor Technology Most analog sensors provide the user with one piece of information to determine the health of the sensor:  Slope  Limited information for troubleshooting
ISM: The Evolution of the Sensor The processing of sensor diagnostics is fully integrated in the sensor electronics ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],All information is processed in the sensor for connectivity flexibility
How Does Digital Sensor Technology Work? ,[object Object],[object Object],[object Object],[object Object]
iSense™ ISM Asset Suite ,[object Object],[object Object],[object Object],[object Object],iSense allows verification and calibration of pH and DO sensors in lab conditions iSense is key to maximize the benefits of the ISM technology
UniCond ®  Conductivity Sensors with ISM ® ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],UniCond ®  sensors provide breakthrough performance! 50 mm 50 meters
UniCond ®  Conductivity Sensors with ISM ® ,[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],[object Object],UniCond ®  provides accuracy at least 33% better than analog conductivity sensors!
Applications and Measurement Range Expanded range with enhanced accuracy! 0.01  0.1  1.0  10  100  1000  10k  100k  1000k  Conductivity    ( µ S/cm) 100 M  10M  1M  100k  10k  1000  100  10  1  Resistivity   (Ohm-cm) Ultra pure water Pure water Make up water Drinking water Diluted acids, bases, salts Waste water Brackish water Industrial process water Acids, bases CURRENT 4E RANGE CURRENT 2E RANGE  UniCond ®  extends the range of  measurement to cover UPW  to seawater with a single sensor! UPW to seawater with a single UniCond ®  sensor !
[object Object]
Questions?

Mais conteúdo relacionado

Mais procurados

Mais procurados (20)

Conductivity Meter
Conductivity MeterConductivity Meter
Conductivity Meter
 
Basics impurity profiling and degradent characterization[134]
Basics impurity profiling and degradent characterization[134]Basics impurity profiling and degradent characterization[134]
Basics impurity profiling and degradent characterization[134]
 
Elemental Impurities by Shiv Kalia.pptx
Elemental Impurities by Shiv Kalia.pptxElemental Impurities by Shiv Kalia.pptx
Elemental Impurities by Shiv Kalia.pptx
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
pH measurement
pH measurementpH measurement
pH measurement
 
Basics of pH
Basics of pHBasics of pH
Basics of pH
 
Qualification of UV spectrophotometer
Qualification of UV spectrophotometer Qualification of UV spectrophotometer
Qualification of UV spectrophotometer
 
Impurities in drug substance (ich q3 a)
Impurities in drug substance (ich q3 a)Impurities in drug substance (ich q3 a)
Impurities in drug substance (ich q3 a)
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Analytical methods validation as per ich & usp
Analytical methods validation as per ich & uspAnalytical methods validation as per ich & usp
Analytical methods validation as per ich & usp
 
Qualification of glassware
Qualification of glasswareQualification of glassware
Qualification of glassware
 
IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS IMPURITIES OF NEW DRUG PRODUCTS
IMPURITIES OF NEW DRUG PRODUCTS
 
Good Laboratory Practice
Good Laboratory PracticeGood Laboratory Practice
Good Laboratory Practice
 
Rationale for the reporting control of degradation products
Rationale for the reporting control of degradation products Rationale for the reporting control of degradation products
Rationale for the reporting control of degradation products
 
Potentiometry
PotentiometryPotentiometry
Potentiometry
 
Ph, buffer solution ph indicator and ph
Ph, buffer solution ph indicator and phPh, buffer solution ph indicator and ph
Ph, buffer solution ph indicator and ph
 
P h metry
P h metryP h metry
P h metry
 
Analytical method validation as per ich and usp
Analytical method validation as per ich and usp Analytical method validation as per ich and usp
Analytical method validation as per ich and usp
 
Pesticide analysis
Pesticide analysisPesticide analysis
Pesticide analysis
 
Karl Fischer titration,principle,apparatus, titration types,Endpoint detection
Karl Fischer titration,principle,apparatus, titration types,Endpoint detectionKarl Fischer titration,principle,apparatus, titration types,Endpoint detection
Karl Fischer titration,principle,apparatus, titration types,Endpoint detection
 

Destaque

ISA AD2015 ADAMI METTLER TOLEDO V01.0
ISA AD2015 ADAMI METTLER TOLEDO V01.0ISA AD2015 ADAMI METTLER TOLEDO V01.0
ISA AD2015 ADAMI METTLER TOLEDO V01.0
jnic
 
(Efc 4) european federation of corrosion guidelines on electrochemical corros...
(Efc 4) european federation of corrosion guidelines on electrochemical corros...(Efc 4) european federation of corrosion guidelines on electrochemical corros...
(Efc 4) european federation of corrosion guidelines on electrochemical corros...
Muhammad Awais
 
Chapter 4: Corrosion testing (shared using VisualBee)
Chapter 4: Corrosion testing  
(shared using VisualBee)Chapter 4: Corrosion testing  
(shared using VisualBee)
Chapter 4: Corrosion testing (shared using VisualBee)
Nima_s70
 
The Utility of Zeta Potential Measurements in the Characterization of CMP Slu...
The Utility of Zeta Potential Measurements in the Characterization of CMP Slu...The Utility of Zeta Potential Measurements in the Characterization of CMP Slu...
The Utility of Zeta Potential Measurements in the Characterization of CMP Slu...
HORIBA Particle
 
Diploma_I_Applied science(chemistry)U-III Acid & bases
Diploma_I_Applied science(chemistry)U-III Acid & bases Diploma_I_Applied science(chemistry)U-III Acid & bases
Diploma_I_Applied science(chemistry)U-III Acid & bases
Rai University
 
Fluids & electrolytes
Fluids & electrolytesFluids & electrolytes
Fluids & electrolytes
learning
 
Inherently Safer Design
Inherently Safer DesignInherently Safer Design
Inherently Safer Design
Hema Madaka
 
Safety Lifecycle Management - Emerson Exchange 2010 - Meet the Experts
Safety Lifecycle Management - Emerson Exchange 2010 - Meet the Experts Safety Lifecycle Management - Emerson Exchange 2010 - Meet the Experts
Safety Lifecycle Management - Emerson Exchange 2010 - Meet the Experts
Mike Boudreaux
 

Destaque (20)

Conductivity measurement
Conductivity measurementConductivity measurement
Conductivity measurement
 
electrical conductivity meter
electrical conductivity meterelectrical conductivity meter
electrical conductivity meter
 
pH meter design and working principle
pH meter design and working principlepH meter design and working principle
pH meter design and working principle
 
Electrical conductivity
Electrical conductivity Electrical conductivity
Electrical conductivity
 
Introduction to Functional Safety and SIL Certification
Introduction to Functional Safety and SIL CertificationIntroduction to Functional Safety and SIL Certification
Introduction to Functional Safety and SIL Certification
 
Basic Control Valve Sizing and Selection
Basic Control Valve Sizing and SelectionBasic Control Valve Sizing and Selection
Basic Control Valve Sizing and Selection
 
ISA AD2015 ADAMI METTLER TOLEDO V01.0
ISA AD2015 ADAMI METTLER TOLEDO V01.0ISA AD2015 ADAMI METTLER TOLEDO V01.0
ISA AD2015 ADAMI METTLER TOLEDO V01.0
 
GPO Application Dataset
GPO Application DatasetGPO Application Dataset
GPO Application Dataset
 
Electrical measurements - Sultan LeMarc
Electrical measurements - Sultan LeMarcElectrical measurements - Sultan LeMarc
Electrical measurements - Sultan LeMarc
 
(Efc 4) european federation of corrosion guidelines on electrochemical corros...
(Efc 4) european federation of corrosion guidelines on electrochemical corros...(Efc 4) european federation of corrosion guidelines on electrochemical corros...
(Efc 4) european federation of corrosion guidelines on electrochemical corros...
 
Chapter 4: Corrosion testing (shared using VisualBee)
Chapter 4: Corrosion testing  
(shared using VisualBee)Chapter 4: Corrosion testing  
(shared using VisualBee)
Chapter 4: Corrosion testing (shared using VisualBee)
 
GPO Customer Presentation
GPO Customer PresentationGPO Customer Presentation
GPO Customer Presentation
 
external & internal corrosion monitoring
external & internal corrosion monitoringexternal & internal corrosion monitoring
external & internal corrosion monitoring
 
The Utility of Zeta Potential Measurements in the Characterization of CMP Slu...
The Utility of Zeta Potential Measurements in the Characterization of CMP Slu...The Utility of Zeta Potential Measurements in the Characterization of CMP Slu...
The Utility of Zeta Potential Measurements in the Characterization of CMP Slu...
 
Diploma_I_Applied science(chemistry)U-III Acid & bases
Diploma_I_Applied science(chemistry)U-III Acid & bases Diploma_I_Applied science(chemistry)U-III Acid & bases
Diploma_I_Applied science(chemistry)U-III Acid & bases
 
Fluids & electrolytes
Fluids & electrolytesFluids & electrolytes
Fluids & electrolytes
 
Inherently Safer Design
Inherently Safer DesignInherently Safer Design
Inherently Safer Design
 
Usability
UsabilityUsability
Usability
 
Functional integrity certification exida
Functional integrity certification   exidaFunctional integrity certification   exida
Functional integrity certification exida
 
Safety Lifecycle Management - Emerson Exchange 2010 - Meet the Experts
Safety Lifecycle Management - Emerson Exchange 2010 - Meet the Experts Safety Lifecycle Management - Emerson Exchange 2010 - Meet the Experts
Safety Lifecycle Management - Emerson Exchange 2010 - Meet the Experts
 

Semelhante a Analytical Measurements: Troubleshooting, Maintenance and the Future

Steam water analysis system
Steam water analysis systemSteam water analysis system
Steam water analysis system
priyank.modi
 
Calibration of pH meter
Calibration of pH meterCalibration of pH meter
Calibration of pH meter
Janine Samelo
 

Semelhante a Analytical Measurements: Troubleshooting, Maintenance and the Future (20)

Isa saint-louis-advanced-p h-short-course-day-1
Isa saint-louis-advanced-p h-short-course-day-1Isa saint-louis-advanced-p h-short-course-day-1
Isa saint-louis-advanced-p h-short-course-day-1
 
pH Control Solutions - Greg McMillan
pH Control Solutions - Greg McMillanpH Control Solutions - Greg McMillan
pH Control Solutions - Greg McMillan
 
PH meter
PH meterPH meter
PH meter
 
pH Analzyer
pH AnalzyerpH Analzyer
pH Analzyer
 
Introduction to-ph
Introduction to-phIntroduction to-ph
Introduction to-ph
 
Experiment5 chem210
Experiment5 chem210Experiment5 chem210
Experiment5 chem210
 
pH - MEASUREMENT .ppt
pH - MEASUREMENT .pptpH - MEASUREMENT .ppt
pH - MEASUREMENT .ppt
 
Potentio lab report
Potentio lab reportPotentio lab report
Potentio lab report
 
Report on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhiReport on pH and conductivity of different samples from delhi
Report on pH and conductivity of different samples from delhi
 
Industrial Analytical Instruments
Industrial Analytical Instruments Industrial Analytical Instruments
Industrial Analytical Instruments
 
Steam water analysis system
Steam water analysis systemSteam water analysis system
Steam water analysis system
 
Calibration of pH meter
Calibration of pH meterCalibration of pH meter
Calibration of pH meter
 
Analytical-Instrumentation.pptx
Analytical-Instrumentation.pptxAnalytical-Instrumentation.pptx
Analytical-Instrumentation.pptx
 
Measurement of Hydrogen Ion Concentration (pH)
Measurement of Hydrogen Ion Concentration (pH)Measurement of Hydrogen Ion Concentration (pH)
Measurement of Hydrogen Ion Concentration (pH)
 
Chapter 10
Chapter 10Chapter 10
Chapter 10
 
Measurement of hydrogen ion concentration (p h)
Measurement of hydrogen ion concentration (p h)Measurement of hydrogen ion concentration (p h)
Measurement of hydrogen ion concentration (p h)
 
Ph electrodes
Ph electrodesPh electrodes
Ph electrodes
 
pH meter.pdf
pH meter.pdfpH meter.pdf
pH meter.pdf
 
Magical Mystery Tour of High Purity pH Measurements
Magical Mystery Tour of High Purity pH Measurements Magical Mystery Tour of High Purity pH Measurements
Magical Mystery Tour of High Purity pH Measurements
 
Ph Measurement Guide
Ph Measurement GuidePh Measurement Guide
Ph Measurement Guide
 

Último

+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
?#DUbAI#??##{{(☎️+971_581248768%)**%*]'#abortion pills for sale in dubai@
 

Último (20)

Strategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherStrategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a Fresher
 
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ..."I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
 
Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...
Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...
Apidays Singapore 2024 - Building Digital Trust in a Digital Economy by Veron...
 
TrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
TrustArc Webinar - Stay Ahead of US State Data Privacy Law DevelopmentsTrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
TrustArc Webinar - Stay Ahead of US State Data Privacy Law Developments
 
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
+971581248768>> SAFE AND ORIGINAL ABORTION PILLS FOR SALE IN DUBAI AND ABUDHA...
 
Ransomware_Q4_2023. The report. [EN].pdf
Ransomware_Q4_2023. The report. [EN].pdfRansomware_Q4_2023. The report. [EN].pdf
Ransomware_Q4_2023. The report. [EN].pdf
 
ICT role in 21st century education and its challenges
ICT role in 21st century education and its challengesICT role in 21st century education and its challenges
ICT role in 21st century education and its challenges
 
AXA XL - Insurer Innovation Award Americas 2024
AXA XL - Insurer Innovation Award Americas 2024AXA XL - Insurer Innovation Award Americas 2024
AXA XL - Insurer Innovation Award Americas 2024
 
A Beginners Guide to Building a RAG App Using Open Source Milvus
A Beginners Guide to Building a RAG App Using Open Source MilvusA Beginners Guide to Building a RAG App Using Open Source Milvus
A Beginners Guide to Building a RAG App Using Open Source Milvus
 
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
Strategies for Unlocking Knowledge Management in Microsoft 365 in the Copilot...
 
MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024
 
Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...
Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...
Emergent Methods: Multi-lingual narrative tracking in the news - real-time ex...
 
Manulife - Insurer Transformation Award 2024
Manulife - Insurer Transformation Award 2024Manulife - Insurer Transformation Award 2024
Manulife - Insurer Transformation Award 2024
 
2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...2024: Domino Containers - The Next Step. News from the Domino Container commu...
2024: Domino Containers - The Next Step. News from the Domino Container commu...
 
Artificial Intelligence Chap.5 : Uncertainty
Artificial Intelligence Chap.5 : UncertaintyArtificial Intelligence Chap.5 : Uncertainty
Artificial Intelligence Chap.5 : Uncertainty
 
Polkadot JAM Slides - Token2049 - By Dr. Gavin Wood
Polkadot JAM Slides - Token2049 - By Dr. Gavin WoodPolkadot JAM Slides - Token2049 - By Dr. Gavin Wood
Polkadot JAM Slides - Token2049 - By Dr. Gavin Wood
 
DBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor PresentationDBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor Presentation
 
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
Connector Corner: Accelerate revenue generation using UiPath API-centric busi...
 
Data Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt RobisonData Cloud, More than a CDP by Matt Robison
Data Cloud, More than a CDP by Matt Robison
 
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemkeProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
ProductAnonymous-April2024-WinProductDiscovery-MelissaKlemke
 

Analytical Measurements: Troubleshooting, Maintenance and the Future

  • 1. pH & Conductivity Parameter Training Measurement, Maintenance & the Future ISA Boston Section March 15, 2011 U E 2 E 3 E 1 Reference- electrolyte E 6 E 5 E 4 Inner Buffer
  • 2.
  • 3. Definition: pH = - log [a H+ ] H 2 0 H + + OH - [ a H+ ] * [ a OH- ] / [a H2O ] = 10 -14 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 pH acidic: pH 0-6.9 [H + ] > [OH - ] [H + ] > 10 -7 M neutral: pH 7.0 [H + ] = [OH - ] [H + ]= 10 -7 M alkaline: pH 7.1 - 14 [H + ] < [OH - ] [H + ] < 10 -7 M pH basics - pH scale
  • 4. Some examples of pH values 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 lemon juice orange juice beer cheese milk pure water egg white borax Milk of Magnesia H 2 SO 4 (1N) 4.9% HCl (0.1N) 0.37% acetic acid (0.1N) 0.6% HCN (0.1N) 0.27% sodium bicarbonate 0.84% (0.1N) potassium ac. 0.98% (0.1N) NH 4 OH 0.017% (0.01N) NH 4 OH 1.7% 1.0N NaOH 4%
  • 5. How does pH measurement work? pH is a potentiometric measurement via an electrochemical sensor/electrode/probe U= E pH -E ref (mV) This potential difference is a function of the solution being measured E pH glass electrode E ref reference electrode high impedance pH Meter
  • 7. What is special about pH glass? H + H + + + + + + + - - - - - - Acidic Alkaline Glass membrane Glass membrane (0.2 - 0.5mm) Gel layer ca. 1 µm (outer and inner) positive charge negative charge internal buffer The surface layer of the glass membrane is the “key performer” in each pH measurement! pH is a measurement of the potential difference between inner and outer layer of glass membrane! This is one reason why pH sensors need to be stored in salt solution when not in use!
  • 8.
  • 9.
  • 10. The Nernst Equation E = E 0 + 2.303 R T log n F 1. Internal Reference Potential 2. Inner Glass/Solution Potential 3. External Reference Potential 4. LIQUID JUNCTION POTENTIAL
  • 11. The Nernst Equation E = E 0 + 2.303 R T log n F 2.303 R T is known as “The Electrode Slope” n F IDEAL SLOPE = 59.16 mV/pH unit at 25 °C Slope is temperature dependent
  • 12. The Nernst Response Curve  mV = 59.16   pH mV E = E o + 2.3 RT/F log a H + where 2.3 RT/F = 0.198T K = 59.16 mV @ 25 o C
  • 13. pH Electrode Calibration Curve pH 0 4 7 10 14 + 500 mv 0 mv - 500 mv millivolts pH 10.00 buffer 0.0 millivolts pH 7.00 buffer -177.5 millivolts IDEAL VALUES!
  • 14.
  • 15. Temperature Error (No Temp. Compensation) Temperature error table for pH signal pH °C 2 3 4 5 6 7 8 9 10 11 12 5 0.30 0.24 0.18 0.12 0.06 0 0.06 0.12 0.18 0.24 0.30 15 0.15 0.12 0.09 0.06 0.03 0 0.03 0.06 0.09 0.12 0.15 25 0 0 0 0 0 0 0 0 0 0 0 35 0.15 0.12 0.09 0.06 0.03 0 0.03 0.06 0.09 0.12 0.15 45 0.30 0.24 0.18 0.12 0.06 0 0.06 0.12 0.18 0.24 0.30 55 0.45 0.36 0.27 0.18 0.09 0 0.09 0.18 0.27 0.36 0.45 65 0.60 0.48 0.36 0.24 0.12 0 0.12 0.24 0.35 0.48 0.60 75 0.75 0.60 0.45 0.30 0.15 0 0.15 0.30 0.45 0.60 0.75 85 0.90 0.72 0.54 0.36 0.18 0 0.18 0.36 0.54 0.72 0.90 No temperature error Temperature error < 0.1 pH units Temperature error > 0.1 but < 0.3 pH units Temperature error = or > 0.3 pH units ( simplified model with Error = /pH (pr.) - 7/ x /(T (pr.) - 25):10/ x 0.03 )
  • 16.
  • 17.
  • 18.
  • 19.
  • 20. pH Calibration /Justification 1. Step Determine Asymmetry Potential / Zero Point 2. Step Determine the Slope pH Buffer 7.00 pH Buffer 4.00 [mV] 200 -200 pH 7 14 [mV] 200 -200 pH 7 14 pH as 4 [%] Slope
  • 21.
  • 22.
  • 23.
  • 24.
  • 25.
  • 26.
  • 27.
  • 28.
  • 29.
  • 30.
  • 31.
  • 32.
  • 33.
  • 35.
  • 36. Conductivity Measurement A conductivity “cell”, sensor, or electrode is an “electro-mechanical” measurement 1 cm 1 cm d = 1 cm Cond Solution Electrode plate VAC
  • 37.
  • 38.
  • 39.
  • 40.
  • 41.
  • 42.
  • 45. Cell Geometry The measured resistance will be dependent on the spacing of the electrode – cell geometry  “cell constant” Therefore, units measurement has dimension component, ex. mS/cm 1 cm 1 cm d = 1 cm Cond Solution Electrode plate VAC
  • 46. Measuring Conductivity ~ H H O Na + Cl - H + O H -
  • 47.
  • 48.
  • 49.
  • 50. Conductivity Temperature Effects ~ H + H H O O H - Na + Cl - H + O H - Na + Cl -
  • 51.
  • 52. Sensor Cell Constant 0.1 cm 0.1 cm -1 0.1 cm 1 cm Conductivity Cell Constant = Length Area 1 cm 1 cm 2 = = 1 cm -1 INSULATOR ELECTRODE ELECTRODE INNER OUTER 1 cm 1 cm
  • 53.
  • 54.
  • 55.
  • 56.
  • 57. Four-Electrode Conductivity Measurement AC Current Source AC Voltage Measurement Drive Electrodes Measuring Electrodes Four Electrode Sensor Four-Electrode Measuring Instrument
  • 58.
  • 59.
  • 60.
  • 61.
  • 62.
  • 63. Cell Installation -2 electrode NOT Recommended Cell Installation... INLET OUTLET OUTLET AIR INLET Avoid dead legs and air traps
  • 64. Cell Installation -4 electrode Recommended OUTLET INLET Maintain a minimum clearance between sensor and pipe
  • 65. Cell Installation -4 electrode NOT Recommended Maintain a minimum clearance between sensor and pipe
  • 66. Conductivity, Resistivity, TDS Ranges Conductivity 100M 10M 1M 100K 10K 1K 100 10 1 Ultrapure water Deionized water Distilled water Condensate Drinking water Cooling tower water Percentage of acids, bases and salt Waste water Brackish water, Sea water Water for Industrial Process 5% Salinity 2% NaOH 20% HCl 0.01 .1 1 10 100 1000 10k 100k 1000k 0.021 0.4 4.6 46 460 4.6k 46k TDS ppm Conductivity and resistivity are measured at 25  C; TDS is expressed as Sodium Chloride (NaCl) Resistivity ohm-cm µS-cm
  • 67. Main Applications and Measuring Range Inductive 4 Elec 2 Elec 0.01 0.1 1.0 10 100 1000 10k 100k 1000k Conductivity ( µ S/cm) 100 M 10M 1M 100k 10k 1000 100 10 1 Resistivity (Ohm-cm) Ultra pure water Pure water Make up water Drinking water Diluted acids, bases, salts Waste water Brackish water Industrial process water Acids, bases Water Processes Biotech/Food and Beverage Chemical Processes
  • 68. So….where is analytical measurement today?
  • 69. Previous Analog Technology Digital technology provides better sound quality
  • 70. Welcome to the Digital World! Same electrochemical end of the sensor converted to digital signal which is more robust and gives more information
  • 71. Analog Sensor Technology Most analog sensors provide the user with one piece of information to determine the health of the sensor: Slope Limited information for troubleshooting
  • 72.
  • 73.
  • 74.
  • 75.
  • 76.
  • 77. Applications and Measurement Range Expanded range with enhanced accuracy! 0.01 0.1 1.0 10 100 1000 10k 100k 1000k Conductivity ( µ S/cm) 100 M 10M 1M 100k 10k 1000 100 10 1 Resistivity (Ohm-cm) Ultra pure water Pure water Make up water Drinking water Diluted acids, bases, salts Waste water Brackish water Industrial process water Acids, bases CURRENT 4E RANGE CURRENT 2E RANGE UniCond ® extends the range of measurement to cover UPW to seawater with a single sensor! UPW to seawater with a single UniCond ® sensor !
  • 78.

Notas do Editor

  1. Some examples are: Beer 4,5 pH Milk 6,2 pH lemon juice 2,2 pH HCl 0,1 M 1 pH Activity a= concentration* activity factor. This activity factor is often estimated to be close to 1, and means activity~concentration
  2. Describe reference system. Calomel reference (Hg/Hg 2 Cl 2 ) is the general electrochemical standard in place of the standard hydrogen electrode (describe!). Not suitable for industrial use: unstable above 60 o C toxic Preferred system is Ag/AgCl, essentially as good as the Calomel system but high temperature stable.
  3. Walther Nernst developed the equation that bears his name in 1880’s (published 1889) Eo ‘standard’ potential for the electrode R universal gas constant T temperature (absolute, degrees Kelvin) F Faraday constant Valence number of ion not shown since it is 1
  4. Walther Nernst developed the equation that bears his name in 1880’s (published 1889) Eo ‘standard’ potential for the electrode R universal gas constant T temperature (absolute, degrees Kelvin) F Faraday constant Valence number of ion not shown since it is 1
  5. Walther Nernst developed the equation that bears his name in 1880’s (published 1889) Eo ‘standard’ potential for the electrode R universal gas constant T temperature (absolute, degrees Kelvin) F Faraday constant Valence number of ion not shown since it is 1
  6. Some solutions aggressive towards the sensor and cause loss of calibration quicker than others pH 7 + 2.5 needs less calibration than pH 5.2 + 0.2 Quality - who makes it!
  7. Walther Nernst developed the equation that bears his name in 1880’s (published 1889) Eo ‘standard’ potential for the electrode R universal gas constant T temperature (absolute, degrees Kelvin) F Faraday constant Valence number of ion not shown since it is 1
  8. Walther Nernst developed the equation that bears his name in 1880’s (published 1889) Eo ‘standard’ potential for the electrode R universal gas constant T temperature (absolute, degrees Kelvin) F Faraday constant Valence number of ion not shown since it is 1
  9. Walther Nernst developed the equation that bears his name in 1880’s (published 1889) Eo ‘standard’ potential for the electrode R universal gas constant T temperature (absolute, degrees Kelvin) F Faraday constant Valence number of ion not shown since it is 1
  10. Conductivity is affected by temperature since water becomes less viscous at higher temperatures and ions can move more easily—they have greater mobility. Typical mineral ions increase in conductivity by about 2% of value per °C. Temperature as well as water purity can change the conductivity. For this reason, it has become an industry standard to compensate measurements to 25°C. That is, the conductivity value is reported as if the sample were at 25°C. General purpose temperature compensation provides the typical 2% per °C correction.
  11. Conductivity is affected by temperature since water becomes less viscous at higher temperatures and ions can move more easily—they have greater mobility. Typical mineral ions increase in conductivity by about 2% of value per °C. Temperature as well as water purity can change the conductivity. For this reason, it has become an industry standard to compensate measurements to 25°C. That is, the conductivity value is reported as if the sample were at 25°C. General purpose temperature compensation provides the typical 2% per °C correction.
  12. To minimize the problems of the two electrode design, a different method has been developed.
  13. Four-electrode measurement refers to a conductivity sensor incorporating four electrodes into its probe body instead of the usual two. The four-electrode measuring technique is used for highly conductive and/or dirty water samples which would foul the surfaces and/or plug the narrow passages of conventional two-electrode conductivity sensors. Suspended solids, turbidity, silt and oils tend to coat electrode surfaces and accumulate in passages and produce negative conductivity errors with two-electrode conductivity systems. Four-electrode measurement applies an AC current through the sample via two outer drive electrodes. These electrodes may become fouled and the circuit will compensate to maintain the AC current level. Two inner measuring electrodes are used to sense the voltage drop through the portion of solution between them. The circuit makes a high impedance AC voltage measurement, drawing negligible current and making it much less affected by additional resistance due to fouling of the measuring electrode surfaces.
  14. 1. Electrode surface condition for high conductivity measurements with two-electrode systems is critical. The surface must be rough on a microscopic scale in order to provide very intimate contact with the sample. Otherwise a high resistance at the interface would cause low conductivity readings. 2. Coatings which insulate the electrodes of a two-electrode sensor have a direct impact on the accuracy, especially when the conductivity of the sample is high. Moderate insulation of the electrodes of four-electrode sensors has no effect since the voltage measurement is made with virtually no current flow. 3. The flat surface of four-electrode sensors is largely self-cleaning in high flow applications.
  15. Inductive conductivity provides the best tolerance for fouling conditions, with no electrodes in contact with the sample. It can also be used for high purity chemical concentration measurements with its excellent chemical resistance and no wetted metal parts. Insertion and submersion mounting are available.
  16. Conductivity cell installation must assure that the cell is completely immersed in water. No bubbles can be within the annular space between electrodes or erroneously low conductivity (high resistivity) readings will result. Upward flow is desirable so air can easily escape.
  17. When installing in large pipe, do not use a series of reducing bushings that would create a “dead leg” of stagnant water around the cell. Tap directly into the pipe or into a pipe plug in a Tee. Deionization resin beads, if they escape from a column, can become lodged between the electrodes of a conductivity cell and short them, causing erroneously high conductivity (low resistivity) readings.
  18. When installing in large pipe, do not use a series of reducing bushings that would create a “dead leg” of stagnant water around the cell. Tap directly into the pipe or into a pipe plug in a Tee. Deionization resin beads, if they escape from a column, can become lodged between the electrodes of a conductivity cell and short them, causing erroneously high conductivity (low resistivity) readings.
  19. When installing in large pipe, do not use a series of reducing bushings that would create a “dead leg” of stagnant water around the cell. Tap directly into the pipe or into a pipe plug in a Tee. Deionization resin beads, if they escape from a column, can become lodged between the electrodes of a conductivity cell and short them, causing erroneously high conductivity (low resistivity) readings.
  20. Typical ranges of applications are shown in resistivity, conductivity and Total dissolved solids (ppm TDS) units. The TDS scale here is based on the concentration of sodium chloride that would have the identified conductivity and resistivity. Other substances such as calcium carbonate may be use as the basis.
  21. As we all know, In the sensor a small microprocessor has been integrated, that stores and process all relevant data. All this information is then digitally transferred to the instrument without any risk of interferences. The microprocessor is also able to store information like calibration values, identity and timestamp the maximum process temperature. What is even more interesting with ISM is the connection possibilities and how the sensors can perform there own diagnostics in real-time.
  22. With the ISM we have transferred the technology experience and know how that we have collected over the years into the sensor and the sensor head. In the sensor a small microprocessor has been integrated, that stores and process all relevant data. All this information is then digitally transferred to the instrument. With this new technology, the ISM sensors are able to perform its own diagnostics in real-time. All our new transmitters are ready to handle the ISM technology and together with the ISM sensors they can perform true predictive maintenance.
  23. SLIDE CAN BE SKIPPED IN CASE OF TIME PRESSURE There is an overlap in measuring range for inductive and 4E cond. Sensors We position the 4E sensors mainly for chemical, pharma-, and F&amp;B application where the conditions are not to harsh. The inductive sensors are designed for chemical and P&amp;P industrie where you have harsh conditions. Nevertheless, if the customer insist on either a 4E or inductive solution we now can offer both sensor types with best performance. SLIDE CAN BE SKIPPED IN CASE OF TIME PRESSURE