SlideShare a Scribd company logo
1 of 42
Concentration cell system
A solid electrolyte like zirconia exhibits conductivity of oxygen ions at
high temperature. As shown in the figure below, when porous platinum
electrodes are attached to both sides of the zirconia element to be heated
up and gases of different partial oxygen concentrations are brought into
contact with the respective surfaces of the zirconia, the device acts as an
oxygen concentration cell. This phenomenon causes an electromotive
force to be generated between both electrodes according to Nernst’s
equation
The principle of the zirconia oxygen analyzer is as follows:
Zirconia is one of the few compounds that actually become conductive at high
temperatures & more conductive as its temperature increases
Zirconia starts out with a very high resistance at room temperature greater then one
trillion ohms / cm , Having less then 1000 ohms / cm resistance at 1000 degrees
Celsius & become a very good conductor
Heating the element allows different partial oxygen concentrations of the gasses to
come into contact with the opposite side of the zirconia’s creating an oxygen
concentration cell.
These ions travel through the zirconia’s element to the other electrode. At that
point, the Nernst expression applied to calculate the force by measuring the
electromotive force generated between the two electrodes.
 Reference gas on one side and sample
gas on the other side
 Oxygen ions move from the side with
the highest concentration of oxygen to
that with the lowest concentration.
 The movement of ions generates an
EMF (Electro Motive Force) which can
be measured to determine the oxygen
content.
the EMF varies depending
on
 the temperature of the
zirconia sensor and
 the oxygen concentration
of the reference gas (PR),
in the actual device.
• the zirconia sensor is
placed in a constant
temperature oven
• air is generally used as
the reference gas
Limitations
 Flammable gases cannot be used
 Sensor degradation occurs if corrosive gas (fluorine-based gases, chlorine-
based gases, sulfate-based gases)is measured
System Configuration
This system is for monitoring and controlling oxygen concentration in the
combustion gases of a large-size boiler or heating furnace. Clean (dry) air (21%
O2) is used as the reference gas and the span gas for calibration. Zero gas is fed
in from a cylinder during calibration.
System Configuration
This example, represents typical applications in large boilers and heating
furnaces, where there is a need to monitor and control oxygen concentration. The
reference gas and calibration-time span gas are (clean, dry) instrument air. Zero gas
is supplied from a gas cylinder
System uses auto calibration unit, with auto-switching of the calibration gas.
Calibration
The converter is calibrated in such a way that the actual zero and span gases are
measured and those measured values are used to agree with the oxygen
concentrations in the respective gases.
There are three types of calibration procedures available:
(1) Manual calibration conducting zero and span calibrations, or either of these
calibrations in turn.
(2) Semi-automatic calibration which uses the touchpanel or a contact input
signal and conducts calibration operations based on a preset calibration time
and stable time.
(3) Automatic calibration conducted at preset intervals.
Calibration Setting Procedures
Select “Points” (calibration procedure) in the Calibration setup display to display
the “Span – Zero,” “Span, Zero” selection display. In this display, select “Span –
Zero.”
Calibration Gas Concentration Setting
(1) Zero-gas concentration
If zero-gas concentration is selected, the Numeric-data Entry display then
appears.
Use this display to enter an oxygen concentration value for the zero-gas
calibration; if the oxygen concentration is 0.98 vol%O2, enter 00098.
(2) Span-gas concentration
With “Span gas conc” selected in the Calibration setup display, display the
Numeric-data Entry display and enter an oxygen concentration value for the
span-gas calibration; If instrument air is used, enter 02100 for a 21 vol% O2 value.
Manual Calibration
Preliminary
Before performing manual calibration, Setting Unit zero-gas flow valve is fully closed.
Open the zero-gas cylinder pressure regulator so that the secondary pressure equals
measured gas plus approx. 50 kPa (or measured gas pressure plus approx. 150 kPa when a
chack valve is used, maximum pressure rating is 300 kPa)..
Thermox Ametek WDG-IV
The same main principle for YokoJawa oxygen Analyzer using Zirconium
Oxide material
Thermox Ametek WDG-IV
Sensor operation:
A sample is drawn from the process stream by means of an air-operated
aspirator and is immediately returned to the process. A portion of this gas
rises past the oxygen measuring cell.
Assemble sensor’s parts
400-AT-001
Thermox Series 2000 WDG-IV
Manual Calibrate/Manual Verify
To perform a manual calibration or verification, follow these steps:
1. Enter the calibration gas values to match the span and zero calibration gases you
intend to use to calibrate or verify your system
2. Select Initiate Cal or Initiate Verify from the Calibrate key menu, depending on
whether you want to perform a calibration or verification.
3. Select Manual Calibrate or Manual Verify. You are then prompted to apply the 02
span gas to the calibration gas inlet port on the sensor (tubing must be free of oil and
dirt):
4. Press the Enter key once you have applied this calibration gas. The span display
then appears (display line 2 will only appear if you have the combustibles option;
display line 3 will only appear if you have the combined combustibles and
methane option):
5. Press the Enter key once the 02 reading on the far left of the first display line
stabilizes. The reading on the far right of the first display line shows the
setpoint value you entered using the Cal Gas Values menu option. This
setpoint value should match the cal gas cylinder for the 02 span gas.
If your 0, reading on the far left of the display has stabilized, yet is not near the
setpoint value on the right of the display, you should check the cal gas value
you entered and ensure that this value matches the cal gas cylinder for the gas
you applied to the cal gas inlet port of the sensor.
You are then prompted to apply the 02 zero gas to the calibration inlet port of
the sensor:
6. Press the Enter key once you have applied this cal gas. The zero gas display
then appears (display line 2 will only appear if you have the combustibles
option; display line 3 will only appear if you have the combined combustibles
and methane option):
7. Press the Enter key once the zero gas reading(s) on the left of the display have
stabilized. If you don't have the combustibles option, you only need to wait until
the 02 zero gas value stabilizes. The reading on the far right of the first display line
shows the setpoint value you entered for the 02 zero gas using the Cal Gas Values
menu option. This 02 zero gas setpoint value should match the cal gas cylinder for
the 02 zero gas.
Error Messages
Cell is Over Temp
This error message indicates that the senSor temperature is 30 'C or more
over its correct operating temperature. The software shuts off the furnace
until the temperature returns to a normal operating range.
If this message soon goes away, it indicates that the temperature has
returned
to an acceptable range. If there is a problem with the sensor temperature
control system, you will see another error message after this message
appears.
Excessive Cal Error
If a primary calibration cannot correct the sensor temperature within the
software's allowable limits, this message is displayed to indicate the primary
calibration has failed. This message could indicate one of the following
problems:
• Calibration gas setup problem
• Process pressure incorrectly entered
• Failed cell
• Leak or plug in plumbing
Span Gas Range Error
This error message occurs if the calibration span gas does not read within the
software's allowable limits. Once this error occurs, calibration of the system is
aborted.
This message could indicate one of the following problems:
• Calibration gas setup problem
• Process pressure incorrectly entered
•Aspirator air not turned off during calibration (WDG-IV only)
•Failed cell
•Cell wiring reversed (Cell + and Cell - wires)
•Leak in plumbing
•Plugged plumbing
Zero Gas Range Error
This error message occurs if the calibration zero gas does not read within
the software's allowable limits. Once this error occurs, calibration of the
system is aborted. This error indicates one of the following problems:
•Calibration gas setup problem
•Process pressure incorrectly entered.
•Cell wiring reversed (Cell + and Cell- wires)
•Aspirator air not turned off during calibration (WDG-IV only)
•Failed cell (primary calibration should be run at this time)
•Leak or plug in plumbing
Temp Rise Failure
This error message indicates that the sensor has failed to increase in temperature
a minimum of 10 °C in a 60-second time frame during start-up. Once the system
reaches the operating set point, this message will only occur if the sensor falls
below the set point by 15°C and fails to recover within 60 seconds. The problems
that can cause this error are as follows:
•AC line power problem at the sensor
•Open furnace
•Over-temperature protection circuit trip on the sensor board
•Sensor board failure
•Loss of furnace drive signal
•Interconnecting wiring problem
Thermocouple Failure
This error message occurs when the indicated sensor temperature falls below -
70°C. This error message can be the result of one of the following:
• Open thermocouple
• Faulty interconnecting wiring
•Control unit display module failure
T/C (Thermocouple) Circuit Failure
This message appears when the control unit recognizes a sudden drop in
temperature (greater than 100°C). The thermocouple circuit failure message
can be the result of one of the following problems:
•Shorted thermocouple
•Faulty interconnecting wiring
•Control unit display module failure
•Sensor board failure
Paramagnetic Analyzer
High magnetic susceptibility of oxygen as compared to other gases allows it to be
attracted to a magnetic field
Magnetic susceptibility is a measure of the intensity of the magnetization of a
substance when it is placed in a magnetic field
In Simple word Oxygen is attracted to a magnetic field several hundred times
stronger than any other gas
Let’s see the
video
Paramagnetic Analyzer
In the cell, two glass spheres filled with nitrogen gas are suspended with strong
metal.
At first, the spheres are kept in balance in an magnetic field. When oxygen
molecules having a large magnetic susceptibility flow there, the molecules are
pulled toward the stronger magnetic field zone and the spheres are moved away
from the zone.
The resulting deviation of the spheres is detected with the light source, reflecting
mirror and light receiving element, and a current is flowed through the feedback
loop to control so that the spheres can return to the initial balanced state.
The current flowing through the feedback loop is proportional to oxygen
concentration.
Thus, oxygen concentration is converted into an electric signal.
Tunable Diodel Laser measurement system
Tunable Diode Laser (or TDL) measurements are based on absorption spectroscopy.
Analyzer is a TDL system and operates by measuring the amount of laser light that is
absorbed (lost) as it travels through the gas being measured.
In the simplest form a TDL analyzer consists of a laser that produces infrared light, optical
lenses to focus the laser light through the gas to be measured and then on to a detector,
the detector, and electronics that control the laser and translate the detector signal into a
signal representing the gas concentration.
Gas molecules absorb light at specific colors, called absorption lines. This
absorption follows Beers law.
TDL Analyzers are effectively infrared analyzers which obey the Beer-Lambert
Law.
I = I0•e-E.G.L
where I is the radiation intensity after absorption
I0 is the initial radiation intensity
E is the extinction coefficient
G is the gas concentration and L is the path length of the measurement area
Advantages:
• Capability of measuring at very high temperature, high pressures and under
difficult conditions (corrosive, aggressive, high particulate service).
• Most measurements are rapid (5 seconds) and interference free .
Disadvantages:
• Initial installation for the mounting flange is required.
Oxygen gas analyzer

More Related Content

What's hot (20)

Temperature Measurement
Temperature MeasurementTemperature Measurement
Temperature Measurement
 
Module (1.2) psychometrics-air parameters-hvac_by ss-eng. juma
Module (1.2) psychometrics-air parameters-hvac_by ss-eng. jumaModule (1.2) psychometrics-air parameters-hvac_by ss-eng. juma
Module (1.2) psychometrics-air parameters-hvac_by ss-eng. juma
 
Temperature measurement part II
Temperature measurement   part IITemperature measurement   part II
Temperature measurement part II
 
Process Gas Chromatograph - Yokogawa GC8000
Process Gas Chromatograph - Yokogawa GC8000Process Gas Chromatograph - Yokogawa GC8000
Process Gas Chromatograph - Yokogawa GC8000
 
Temperature Measurements
Temperature MeasurementsTemperature Measurements
Temperature Measurements
 
saybolt viscometer
saybolt viscometersaybolt viscometer
saybolt viscometer
 
Temperature measurment devices
Temperature measurment devicesTemperature measurment devices
Temperature measurment devices
 
Oxygen concentrator-Applications and Maintenance
Oxygen concentrator-Applications and MaintenanceOxygen concentrator-Applications and Maintenance
Oxygen concentrator-Applications and Maintenance
 
Pulse oximetry capnography
Pulse oximetry capnography Pulse oximetry capnography
Pulse oximetry capnography
 
Gas analyzers
Gas analyzersGas analyzers
Gas analyzers
 
Flow measurement
Flow measurementFlow measurement
Flow measurement
 
Basic instrumentation
Basic instrumentationBasic instrumentation
Basic instrumentation
 
Spo2 ppt
Spo2 pptSpo2 ppt
Spo2 ppt
 
Temperature measurement
Temperature measurementTemperature measurement
Temperature measurement
 
Level measurement
Level measurementLevel measurement
Level measurement
 
INFRARED TEMPERATURE DETECTOR
INFRARED TEMPERATURE DETECTORINFRARED TEMPERATURE DETECTOR
INFRARED TEMPERATURE DETECTOR
 
Basic controls for boilers
Basic controls for boilersBasic controls for boilers
Basic controls for boilers
 
Temperature measurement
Temperature measurementTemperature measurement
Temperature measurement
 
Level Measurement
Level MeasurementLevel Measurement
Level Measurement
 
Refrigeration
RefrigerationRefrigeration
Refrigeration
 

Similar to Oxygen gas analyzer

VVC AUTO 85.5 - Smog Technician - Level 1 - Baseline testing
VVC AUTO 85.5 - Smog Technician - Level 1 - Baseline testingVVC AUTO 85.5 - Smog Technician - Level 1 - Baseline testing
VVC AUTO 85.5 - Smog Technician - Level 1 - Baseline testingJustin Gatewood
 
Automotive sensory system
Automotive sensory systemAutomotive sensory system
Automotive sensory systemAkshadaBhosale1
 
Chapter 55 Computer Sensors
Chapter 55 Computer SensorsChapter 55 Computer Sensors
Chapter 55 Computer Sensorsmcfalltj
 
E+H-Transmitter sensor matching improves RTD accuracy
E+H-Transmitter sensor matching improves RTD accuracyE+H-Transmitter sensor matching improves RTD accuracy
E+H-Transmitter sensor matching improves RTD accuracyCarotek
 
Automatic gas detection system
Automatic  gas  detection  systemAutomatic  gas  detection  system
Automatic gas detection systemSayanSarkar55
 
Understanding AMS2750E (intermediate) v2
Understanding AMS2750E (intermediate) v2Understanding AMS2750E (intermediate) v2
Understanding AMS2750E (intermediate) v2Peter Sherwin
 
Water Temperature
Water TemperatureWater Temperature
Water TemperatureTheju Paul
 
Fluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer CalibrationFluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer CalibrationTranscat
 
Eurosens fuel flow meters installation
Eurosens fuel flow meters installationEurosens fuel flow meters installation
Eurosens fuel flow meters installationPawel Elenski
 
Pressure Thermometer
Pressure ThermometerPressure Thermometer
Pressure ThermometerDeepak Kumar
 
LPG – Liquefied Petroleum Gas & Gas Detection System
LPG – Liquefied Petroleum Gas & Gas Detection SystemLPG – Liquefied Petroleum Gas & Gas Detection System
LPG – Liquefied Petroleum Gas & Gas Detection SystemMonzer Salahdine
 
Halderman ch076 lecture
Halderman ch076 lectureHalderman ch076 lecture
Halderman ch076 lecturemcfalltj
 
how to find Flash point and fire point
how to find Flash point and fire pointhow to find Flash point and fire point
how to find Flash point and fire pointAbdul Rahman
 

Similar to Oxygen gas analyzer (20)

VVC AUTO 85.5 - Smog Technician - Level 1 - Baseline testing
VVC AUTO 85.5 - Smog Technician - Level 1 - Baseline testingVVC AUTO 85.5 - Smog Technician - Level 1 - Baseline testing
VVC AUTO 85.5 - Smog Technician - Level 1 - Baseline testing
 
Automotive sensory system
Automotive sensory systemAutomotive sensory system
Automotive sensory system
 
Chapter 55 Computer Sensors
Chapter 55 Computer SensorsChapter 55 Computer Sensors
Chapter 55 Computer Sensors
 
Flashar 2014 by ATAC
Flashar 2014 by ATACFlashar 2014 by ATAC
Flashar 2014 by ATAC
 
E+H-Transmitter sensor matching improves RTD accuracy
E+H-Transmitter sensor matching improves RTD accuracyE+H-Transmitter sensor matching improves RTD accuracy
E+H-Transmitter sensor matching improves RTD accuracy
 
themodyamic paper
themodyamic paperthemodyamic paper
themodyamic paper
 
Automatic gas detection system
Automatic  gas  detection  systemAutomatic  gas  detection  system
Automatic gas detection system
 
Ch14
Ch14Ch14
Ch14
 
Presentation1
Presentation1Presentation1
Presentation1
 
Oep
OepOep
Oep
 
Understanding AMS2750E (intermediate) v2
Understanding AMS2750E (intermediate) v2Understanding AMS2750E (intermediate) v2
Understanding AMS2750E (intermediate) v2
 
Caldeira Roca R-20/20
Caldeira Roca R-20/20Caldeira Roca R-20/20
Caldeira Roca R-20/20
 
Water Temperature
Water TemperatureWater Temperature
Water Temperature
 
Fluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer CalibrationFluke Corporation: Gas Custody Transfer Calibration
Fluke Corporation: Gas Custody Transfer Calibration
 
Eurosens fuel flow meters installation
Eurosens fuel flow meters installationEurosens fuel flow meters installation
Eurosens fuel flow meters installation
 
Pressure Thermometer
Pressure ThermometerPressure Thermometer
Pressure Thermometer
 
DSC
DSCDSC
DSC
 
LPG – Liquefied Petroleum Gas & Gas Detection System
LPG – Liquefied Petroleum Gas & Gas Detection SystemLPG – Liquefied Petroleum Gas & Gas Detection System
LPG – Liquefied Petroleum Gas & Gas Detection System
 
Halderman ch076 lecture
Halderman ch076 lectureHalderman ch076 lecture
Halderman ch076 lecture
 
how to find Flash point and fire point
how to find Flash point and fire pointhow to find Flash point and fire point
how to find Flash point and fire point
 

More from Mohamed Sarhan

More from Mohamed Sarhan (7)

Orp analyzer
Orp analyzerOrp analyzer
Orp analyzer
 
Viscosity measurements
Viscosity measurementsViscosity measurements
Viscosity measurements
 
Gas chromatograph analyzer
Gas chromatograph analyzerGas chromatograph analyzer
Gas chromatograph analyzer
 
Moisture Analyzer
Moisture AnalyzerMoisture Analyzer
Moisture Analyzer
 
Oil in water analyzer
Oil in water analyzerOil in water analyzer
Oil in water analyzer
 
Conductivity Analyzer
Conductivity AnalyzerConductivity Analyzer
Conductivity Analyzer
 
pH Analzyer
pH AnalzyerpH Analzyer
pH Analzyer
 

Recently uploaded

VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Bookingdharasingh5698
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlysanyuktamishra911
 
DC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equationDC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equationBhangaleSonal
 
A Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna MunicipalityA Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna MunicipalityMorshed Ahmed Rahath
 
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...roncy bisnoi
 
Employee leave management system project.
Employee leave management system project.Employee leave management system project.
Employee leave management system project.Kamal Acharya
 
Generative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPTGenerative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPTbhaskargani46
 
University management System project report..pdf
University management System project report..pdfUniversity management System project report..pdf
University management System project report..pdfKamal Acharya
 
Minimum and Maximum Modes of microprocessor 8086
Minimum and Maximum Modes of microprocessor 8086Minimum and Maximum Modes of microprocessor 8086
Minimum and Maximum Modes of microprocessor 8086anil_gaur
 
Design For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startDesign For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startQuintin Balsdon
 
Bhosari ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready For ...
Bhosari ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready For ...Bhosari ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready For ...
Bhosari ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready For ...tanu pandey
 
Double Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torqueDouble Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torqueBhangaleSonal
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . pptDineshKumar4165
 
data_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdfdata_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdfJiananWang21
 
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
COST-EFFETIVE  and Energy Efficient BUILDINGS ptxCOST-EFFETIVE  and Energy Efficient BUILDINGS ptx
COST-EFFETIVE and Energy Efficient BUILDINGS ptxJIT KUMAR GUPTA
 
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...soginsider
 
Block diagram reduction techniques in control systems.ppt
Block diagram reduction techniques in control systems.pptBlock diagram reduction techniques in control systems.ppt
Block diagram reduction techniques in control systems.pptNANDHAKUMARA10
 

Recently uploaded (20)

Integrated Test Rig For HTFE-25 - Neometrix
Integrated Test Rig For HTFE-25 - NeometrixIntegrated Test Rig For HTFE-25 - Neometrix
Integrated Test Rig For HTFE-25 - Neometrix
 
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 BookingVIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
VIP Call Girls Ankleshwar 7001035870 Whatsapp Number, 24/07 Booking
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghly
 
DC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equationDC MACHINE-Motoring and generation, Armature circuit equation
DC MACHINE-Motoring and generation, Armature circuit equation
 
A Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna MunicipalityA Study of Urban Area Plan for Pabna Municipality
A Study of Urban Area Plan for Pabna Municipality
 
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
Call Girls Pimpri Chinchwad Call Me 7737669865 Budget Friendly No Advance Boo...
 
Employee leave management system project.
Employee leave management system project.Employee leave management system project.
Employee leave management system project.
 
Generative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPTGenerative AI or GenAI technology based PPT
Generative AI or GenAI technology based PPT
 
University management System project report..pdf
University management System project report..pdfUniversity management System project report..pdf
University management System project report..pdf
 
Call Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort ServiceCall Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
Call Girls in Netaji Nagar, Delhi 💯 Call Us 🔝9953056974 🔝 Escort Service
 
Minimum and Maximum Modes of microprocessor 8086
Minimum and Maximum Modes of microprocessor 8086Minimum and Maximum Modes of microprocessor 8086
Minimum and Maximum Modes of microprocessor 8086
 
Design For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the startDesign For Accessibility: Getting it right from the start
Design For Accessibility: Getting it right from the start
 
Bhosari ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready For ...
Bhosari ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready For ...Bhosari ( Call Girls ) Pune  6297143586  Hot Model With Sexy Bhabi Ready For ...
Bhosari ( Call Girls ) Pune 6297143586 Hot Model With Sexy Bhabi Ready For ...
 
Double Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torqueDouble Revolving field theory-how the rotor develops torque
Double Revolving field theory-how the rotor develops torque
 
Thermal Engineering Unit - I & II . ppt
Thermal Engineering  Unit - I & II . pptThermal Engineering  Unit - I & II . ppt
Thermal Engineering Unit - I & II . ppt
 
data_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdfdata_management_and _data_science_cheat_sheet.pdf
data_management_and _data_science_cheat_sheet.pdf
 
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
COST-EFFETIVE  and Energy Efficient BUILDINGS ptxCOST-EFFETIVE  and Energy Efficient BUILDINGS ptx
COST-EFFETIVE and Energy Efficient BUILDINGS ptx
 
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...
Hazard Identification (HAZID) vs. Hazard and Operability (HAZOP): A Comparati...
 
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar ≼🔝 Delhi door step de...
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar  ≼🔝 Delhi door step de...Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar  ≼🔝 Delhi door step de...
Call Now ≽ 9953056974 ≼🔝 Call Girls In New Ashok Nagar ≼🔝 Delhi door step de...
 
Block diagram reduction techniques in control systems.ppt
Block diagram reduction techniques in control systems.pptBlock diagram reduction techniques in control systems.ppt
Block diagram reduction techniques in control systems.ppt
 

Oxygen gas analyzer

  • 1.
  • 2.
  • 3. Concentration cell system A solid electrolyte like zirconia exhibits conductivity of oxygen ions at high temperature. As shown in the figure below, when porous platinum electrodes are attached to both sides of the zirconia element to be heated up and gases of different partial oxygen concentrations are brought into contact with the respective surfaces of the zirconia, the device acts as an oxygen concentration cell. This phenomenon causes an electromotive force to be generated between both electrodes according to Nernst’s equation
  • 4. The principle of the zirconia oxygen analyzer is as follows: Zirconia is one of the few compounds that actually become conductive at high temperatures & more conductive as its temperature increases Zirconia starts out with a very high resistance at room temperature greater then one trillion ohms / cm , Having less then 1000 ohms / cm resistance at 1000 degrees Celsius & become a very good conductor Heating the element allows different partial oxygen concentrations of the gasses to come into contact with the opposite side of the zirconia’s creating an oxygen concentration cell. These ions travel through the zirconia’s element to the other electrode. At that point, the Nernst expression applied to calculate the force by measuring the electromotive force generated between the two electrodes.
  • 5.  Reference gas on one side and sample gas on the other side  Oxygen ions move from the side with the highest concentration of oxygen to that with the lowest concentration.  The movement of ions generates an EMF (Electro Motive Force) which can be measured to determine the oxygen content.
  • 6. the EMF varies depending on  the temperature of the zirconia sensor and  the oxygen concentration of the reference gas (PR), in the actual device. • the zirconia sensor is placed in a constant temperature oven • air is generally used as the reference gas Limitations  Flammable gases cannot be used  Sensor degradation occurs if corrosive gas (fluorine-based gases, chlorine- based gases, sulfate-based gases)is measured
  • 7.
  • 8.
  • 9.
  • 10.
  • 11. System Configuration This system is for monitoring and controlling oxygen concentration in the combustion gases of a large-size boiler or heating furnace. Clean (dry) air (21% O2) is used as the reference gas and the span gas for calibration. Zero gas is fed in from a cylinder during calibration.
  • 12. System Configuration This example, represents typical applications in large boilers and heating furnaces, where there is a need to monitor and control oxygen concentration. The reference gas and calibration-time span gas are (clean, dry) instrument air. Zero gas is supplied from a gas cylinder System uses auto calibration unit, with auto-switching of the calibration gas.
  • 13.
  • 14.
  • 15.
  • 16. Calibration The converter is calibrated in such a way that the actual zero and span gases are measured and those measured values are used to agree with the oxygen concentrations in the respective gases. There are three types of calibration procedures available: (1) Manual calibration conducting zero and span calibrations, or either of these calibrations in turn. (2) Semi-automatic calibration which uses the touchpanel or a contact input signal and conducts calibration operations based on a preset calibration time and stable time. (3) Automatic calibration conducted at preset intervals.
  • 17. Calibration Setting Procedures Select “Points” (calibration procedure) in the Calibration setup display to display the “Span – Zero,” “Span, Zero” selection display. In this display, select “Span – Zero.”
  • 18. Calibration Gas Concentration Setting (1) Zero-gas concentration If zero-gas concentration is selected, the Numeric-data Entry display then appears. Use this display to enter an oxygen concentration value for the zero-gas calibration; if the oxygen concentration is 0.98 vol%O2, enter 00098. (2) Span-gas concentration With “Span gas conc” selected in the Calibration setup display, display the Numeric-data Entry display and enter an oxygen concentration value for the span-gas calibration; If instrument air is used, enter 02100 for a 21 vol% O2 value.
  • 19. Manual Calibration Preliminary Before performing manual calibration, Setting Unit zero-gas flow valve is fully closed. Open the zero-gas cylinder pressure regulator so that the secondary pressure equals measured gas plus approx. 50 kPa (or measured gas pressure plus approx. 150 kPa when a chack valve is used, maximum pressure rating is 300 kPa)..
  • 20.
  • 21.
  • 22. Thermox Ametek WDG-IV The same main principle for YokoJawa oxygen Analyzer using Zirconium Oxide material
  • 24. Sensor operation: A sample is drawn from the process stream by means of an air-operated aspirator and is immediately returned to the process. A portion of this gas rises past the oxygen measuring cell.
  • 26.
  • 29. Manual Calibrate/Manual Verify To perform a manual calibration or verification, follow these steps: 1. Enter the calibration gas values to match the span and zero calibration gases you intend to use to calibrate or verify your system 2. Select Initiate Cal or Initiate Verify from the Calibrate key menu, depending on whether you want to perform a calibration or verification. 3. Select Manual Calibrate or Manual Verify. You are then prompted to apply the 02 span gas to the calibration gas inlet port on the sensor (tubing must be free of oil and dirt): 4. Press the Enter key once you have applied this calibration gas. The span display then appears (display line 2 will only appear if you have the combustibles option; display line 3 will only appear if you have the combined combustibles and methane option):
  • 30. 5. Press the Enter key once the 02 reading on the far left of the first display line stabilizes. The reading on the far right of the first display line shows the setpoint value you entered using the Cal Gas Values menu option. This setpoint value should match the cal gas cylinder for the 02 span gas. If your 0, reading on the far left of the display has stabilized, yet is not near the setpoint value on the right of the display, you should check the cal gas value you entered and ensure that this value matches the cal gas cylinder for the gas you applied to the cal gas inlet port of the sensor. You are then prompted to apply the 02 zero gas to the calibration inlet port of the sensor:
  • 31. 6. Press the Enter key once you have applied this cal gas. The zero gas display then appears (display line 2 will only appear if you have the combustibles option; display line 3 will only appear if you have the combined combustibles and methane option): 7. Press the Enter key once the zero gas reading(s) on the left of the display have stabilized. If you don't have the combustibles option, you only need to wait until the 02 zero gas value stabilizes. The reading on the far right of the first display line shows the setpoint value you entered for the 02 zero gas using the Cal Gas Values menu option. This 02 zero gas setpoint value should match the cal gas cylinder for the 02 zero gas.
  • 32. Error Messages Cell is Over Temp This error message indicates that the senSor temperature is 30 'C or more over its correct operating temperature. The software shuts off the furnace until the temperature returns to a normal operating range. If this message soon goes away, it indicates that the temperature has returned to an acceptable range. If there is a problem with the sensor temperature control system, you will see another error message after this message appears. Excessive Cal Error If a primary calibration cannot correct the sensor temperature within the software's allowable limits, this message is displayed to indicate the primary calibration has failed. This message could indicate one of the following problems: • Calibration gas setup problem • Process pressure incorrectly entered • Failed cell • Leak or plug in plumbing
  • 33. Span Gas Range Error This error message occurs if the calibration span gas does not read within the software's allowable limits. Once this error occurs, calibration of the system is aborted. This message could indicate one of the following problems: • Calibration gas setup problem • Process pressure incorrectly entered •Aspirator air not turned off during calibration (WDG-IV only) •Failed cell •Cell wiring reversed (Cell + and Cell - wires) •Leak in plumbing •Plugged plumbing Zero Gas Range Error This error message occurs if the calibration zero gas does not read within the software's allowable limits. Once this error occurs, calibration of the system is aborted. This error indicates one of the following problems: •Calibration gas setup problem •Process pressure incorrectly entered. •Cell wiring reversed (Cell + and Cell- wires) •Aspirator air not turned off during calibration (WDG-IV only) •Failed cell (primary calibration should be run at this time) •Leak or plug in plumbing
  • 34. Temp Rise Failure This error message indicates that the sensor has failed to increase in temperature a minimum of 10 °C in a 60-second time frame during start-up. Once the system reaches the operating set point, this message will only occur if the sensor falls below the set point by 15°C and fails to recover within 60 seconds. The problems that can cause this error are as follows: •AC line power problem at the sensor •Open furnace •Over-temperature protection circuit trip on the sensor board •Sensor board failure •Loss of furnace drive signal •Interconnecting wiring problem Thermocouple Failure This error message occurs when the indicated sensor temperature falls below - 70°C. This error message can be the result of one of the following: • Open thermocouple • Faulty interconnecting wiring •Control unit display module failure
  • 35. T/C (Thermocouple) Circuit Failure This message appears when the control unit recognizes a sudden drop in temperature (greater than 100°C). The thermocouple circuit failure message can be the result of one of the following problems: •Shorted thermocouple •Faulty interconnecting wiring •Control unit display module failure •Sensor board failure
  • 36.
  • 37. Paramagnetic Analyzer High magnetic susceptibility of oxygen as compared to other gases allows it to be attracted to a magnetic field Magnetic susceptibility is a measure of the intensity of the magnetization of a substance when it is placed in a magnetic field In Simple word Oxygen is attracted to a magnetic field several hundred times stronger than any other gas Let’s see the video
  • 38. Paramagnetic Analyzer In the cell, two glass spheres filled with nitrogen gas are suspended with strong metal. At first, the spheres are kept in balance in an magnetic field. When oxygen molecules having a large magnetic susceptibility flow there, the molecules are pulled toward the stronger magnetic field zone and the spheres are moved away from the zone. The resulting deviation of the spheres is detected with the light source, reflecting mirror and light receiving element, and a current is flowed through the feedback loop to control so that the spheres can return to the initial balanced state. The current flowing through the feedback loop is proportional to oxygen concentration. Thus, oxygen concentration is converted into an electric signal.
  • 39.
  • 40. Tunable Diodel Laser measurement system Tunable Diode Laser (or TDL) measurements are based on absorption spectroscopy. Analyzer is a TDL system and operates by measuring the amount of laser light that is absorbed (lost) as it travels through the gas being measured. In the simplest form a TDL analyzer consists of a laser that produces infrared light, optical lenses to focus the laser light through the gas to be measured and then on to a detector, the detector, and electronics that control the laser and translate the detector signal into a signal representing the gas concentration.
  • 41. Gas molecules absorb light at specific colors, called absorption lines. This absorption follows Beers law. TDL Analyzers are effectively infrared analyzers which obey the Beer-Lambert Law. I = I0•e-E.G.L where I is the radiation intensity after absorption I0 is the initial radiation intensity E is the extinction coefficient G is the gas concentration and L is the path length of the measurement area Advantages: • Capability of measuring at very high temperature, high pressures and under difficult conditions (corrosive, aggressive, high particulate service). • Most measurements are rapid (5 seconds) and interference free . Disadvantages: • Initial installation for the mounting flange is required.