BROOKFIELD DIGITAL
VISCOMETER
• PHOTOGALLERY
0
• SOP
1
• VALIDATION
2
APC, S.P.Patel 2
APC, S.P.Patel 3
APC, S.P.Patel 4
APC, S.P.Patel 5
APC, S.P.Patel 6
SOP
1) Turn on the viscometer and allow standing,
must be auto zero. After the few seconds the
screen appears which indicates 2 digits.
2) Now press the key. The screen displays the
instruction to remove spindle. After removing
the spindle and pressing the key the
instrument begins. It is auto zeroed.
3) After approximately 15 second the screen
displays the instruction to replace spindle
4) Attach the spindle to viscometer by screwing
them on the lower shaft using left hand
thread.
APC, S.P.Patel 7
5) Press select spindle key and up and down arrow
keys. When the desired code is displayed
release the arrow key.
6) To select spindle, first press either up and down
key which cause the area to show current speed,
press the set speed key for adjusting the speed.
7) Insert center of this spindle in the test material until
the fluid level is at the immersion groove on the
spindle shaft. Tilt the spindle slightly while
immersing to avoid air entrapment.
8) To measure high viscosity choose a small spindle
and to slow speed if the chosen spindle/speed
result in torque above 100% and then reduce the
value.
APC, S.P.Patel 8
9) Allow time for the individual reading to stabilize,
record the value.
10) Press the motor on/off/escape key to turn off motor
11) The time mode allows the user to record the
reading for fixed period of time or until a set torque
value is attained.
12) Then enter/auto turnkey allows determining
maximum calculated viscosity possible with current
spindle/speed.
13) Pressing the up and down arrow keys will allow the
viscometer data to be examine, pressing and other
key ( except enter/ print key) will bring back
normal display.
14) Turn off the mains after use.
APC, S.P.Patel 9
VALIDATION
IQ
• INSTALLATION
QUALIFICATION
OQ
• OPERATION
QUALIFICATION
PQ
• PERFORMANCE
QUALIFICATION
APC, S.P.Patel 10
IQ: INSTALLATION
QUALIFICATION
 Inspect package for shipping damage
 Confirm that all components have been
supplied as detailed in the Operating
Manual and Packing List
 Assemble lab stand as shown in manual.
 Mount viscometer head to assembled lab
stand as shown in manual.
 Verify appropriate spindle set (LV #1-
#4)(RV/HA/HB #2-#7)
APC, S.P.Patel 11
OQ: OPERATION
QUALIFICATION
 Level viscometer using adjustable feet
 Remove shipping cap on spindle
coupling nut
 Power on Viscometer and follow
instructions displayed to complete auto
zero
 Perform oscillation check.
APC, S.P.Patel 12
PQ: PERFORMANCE
QUALIFICATION
 Verify calibration using a minimum of
one calibrated viscosity standard fluid
according to the procedure described
below
 Centipoise reading should be within the
value stated on the fluid ± the instrument
accuracy and the fluid accuracy.
APC, S.P.Patel 13
Oscillation check
 This check verifies the mechanical
condition of the sensing mechanism in the
viscometer.
 Auto-zero the instrument according to
directions in the manual.
 Do not attach the spindle but do press any
key so the main screen will be displayed.
 The display should show 0.0% torque. The
tolerance for this and all tests is ±0.1%
APC, S.P.Patel 14
1) Start the motor rotating at 10 or 20
RPM.
2) The % reading should initially bounce,
then is should stabilize at 0.0% ±0.1%
3) Stop the motor. When the % display
again stabilizes it should return to 0.0%
±0.1%
APC, S.P.Patel 15
Operational and Calibration
Check
 Determining Measurement Accuracy
 All Brookfield instruments are guaranteed to be
accurate to within 1% of the full scale
range(FSR) in use.
 The full scale range (FSR) is simply the
maximum viscosity that can be measured with
any spindle/speed combination.
 Select a specific spindle and rotational speed.
 To determine the resulting FSR (full scale
range), depress and hold the AUTORANGE key.
APC, S.P.Patel 16
 The % display will show 100% and the
cP display will show the FSR (full scale
range) that can be measured under
current choice of spindle and speed.
 1% of this value is the measurement
accuracy under these conditions.
APC, S.P.Patel 17
 Calibration Check
 In principle, all Brookfield rotational
viscometers are calibrated torque meters.
 Brookfield offers a variety of viscosity
calibration fluids that enable one to check
the calibration and linearity of the instrument
over its operating torque range of 10%-
100% torque.
APC, S.P.Patel 18
 Any spindle and any one of several calibration
fluids may be used to perform a calibration
check.
 Experts recommend using one spindle and one
fluid while taking measurements at multiple
speeds.
 This enables multiple calculations of viscosity
throughout the torque range of the instrument;
 one at low torque, one at medium torque and one at
high torque.
 The FSR, and consequently the instrument
accuracy in terms of centipoise, will be different
at each speed.
APC, S.P.Patel 19
 The calibration fluid must be transferred into
a 600ml low form, Griffin beaker (without
creating any bubbles) and,
 on LV and RV models, the instrument Guard
Leg must be used.
 The beaker, fluid, spindle and Guard Leg
must stabilized at 25.0°C (±0.1°C) before the
calibration check can be performed.
APC, S.P.Patel 20
EX:
Actual viscosity is 5,120cP,
 Allowable error = 1% of the FSR + 1% of the
actual value of the calibration fluid
 The lower and upper viscosity limits,
indicating acceptable calibration, equal the
actual value of the fluid,
 in this case 5,120cP, ± the allowable error.
APC, S.P.Patel 21
RPM MIN
cP
MAX cP
(FSR)
1% of
FSR
%
torque
cP
READIN
G
ALLOWAB
LE
ERROR
12 1000 10000 100 51.2 5120 ± 151
6 2000 20000 200 25.5 5100 ± 251
3 4000 40000 400 12.7 5080 ± 451
Allowable error = 1% of the FSR + 1% of the actual
value of the calibration fluid
APC, S.P.Patel 22
Calibration Worksheet
A B C D E F G H I
Fluid
Value
from
Certificate
cP
1% of
Fluid
Value
cP
Instrument
Model
Spindle RPM FSR
cP
(depre
ss
and
hold
AUTORA
NGE
key)
1% of
FSR
cP
1% of
FSR
cP
Visco
sity
cP
APC, S.P.Patel 23
 The viscosity reading should equal the cP value on the
fluid standard to within the combined accuracies of
the Viscometer and the viscosity standard as calculated
below
Low Limit
= A – (B + G)
Reading
I
High Limit
= A + (B + G)
Pass or Fail
APC, S.P.Patel 24
 Any reading outside these limits may
indicate a Viscometer problem.
 Contact the Brookfield technical sales
department or your local Brookfield
dealer/distributor with test results to
determine the nature of the problem.
APC, S.P.Patel 25
Reference:
 Brookfield DV-ii+ pro, programmable
viscometer, operating manual no. M/03-
165-A0404
 “Solution for a viscous world” brookfield
2007 catalog.
 www.brookfieldengineering.com
APC, S.P.Patel 26
APC, S.P.Patel 27

More Related Content

What's hot

Dissolution testing conventional and controlled release products
Dissolution testing  conventional and controlled release productsDissolution testing  conventional and controlled release products
Dissolution testing conventional and controlled release products
Md Fiaz
 

What's hot (20)

Quality Control Of Packaging Material
Quality Control Of Packaging MaterialQuality Control Of Packaging Material
Quality Control Of Packaging Material
 
Mixing in Pharmaceutical Industry
Mixing in Pharmaceutical Industry Mixing in Pharmaceutical Industry
Mixing in Pharmaceutical Industry
 
particle size & distribution
particle size & distributionparticle size & distribution
particle size & distribution
 
Dissolution testing conventional and controlled release products
Dissolution testing  conventional and controlled release productsDissolution testing  conventional and controlled release products
Dissolution testing conventional and controlled release products
 
Manufacturing of water for injection
Manufacturing of water for injectionManufacturing of water for injection
Manufacturing of water for injection
 
Formulation and development of parenterals
Formulation and development of parenteralsFormulation and development of parenterals
Formulation and development of parenterals
 
Pharmaceutical validation & it's types
 Pharmaceutical validation & it's types Pharmaceutical validation & it's types
Pharmaceutical validation & it's types
 
identification and quantitative determination of additives in pharmaceutical ...
identification and quantitative determination of additives in pharmaceutical ...identification and quantitative determination of additives in pharmaceutical ...
identification and quantitative determination of additives in pharmaceutical ...
 
Pilot plan scale up for semisolid and parenteral by Khushboo kunkulol
Pilot plan scale up for semisolid and parenteral by Khushboo kunkulolPilot plan scale up for semisolid and parenteral by Khushboo kunkulol
Pilot plan scale up for semisolid and parenteral by Khushboo kunkulol
 
Quality Control of Aerosols
Quality Control of AerosolsQuality Control of Aerosols
Quality Control of Aerosols
 
Ampule filling and_sealing_machine.ppt1
Ampule filling and_sealing_machine.ppt1Ampule filling and_sealing_machine.ppt1
Ampule filling and_sealing_machine.ppt1
 
Particle size distribution
Particle size distributionParticle size distribution
Particle size distribution
 
Inprocess as per usp ip bp tablets
Inprocess as per usp ip bp tabletsInprocess as per usp ip bp tablets
Inprocess as per usp ip bp tablets
 
FORMULATION TECHNOLOGY PRACTICAL MANUAL
FORMULATION TECHNOLOGY PRACTICAL MANUALFORMULATION TECHNOLOGY PRACTICAL MANUAL
FORMULATION TECHNOLOGY PRACTICAL MANUAL
 
Dissolution
DissolutionDissolution
Dissolution
 
EVULATION OF PHARMACUTICAL PACAKGING MATERIALS/EVULATION PARAMETERS OF PAKAGI...
EVULATION OF PHARMACUTICAL PACAKGING MATERIALS/EVULATION PARAMETERS OF PAKAGI...EVULATION OF PHARMACUTICAL PACAKGING MATERIALS/EVULATION PARAMETERS OF PAKAGI...
EVULATION OF PHARMACUTICAL PACAKGING MATERIALS/EVULATION PARAMETERS OF PAKAGI...
 
Disintegration
DisintegrationDisintegration
Disintegration
 
Quality control test for packaging material ,qc test for glass,metal,rubber
Quality control test for packaging material ,qc test for glass,metal,rubberQuality control test for packaging material ,qc test for glass,metal,rubber
Quality control test for packaging material ,qc test for glass,metal,rubber
 
PHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTION
PHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTIONPHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTION
PHARMACEUTICAL CALIBRATION, QUALIFICATION AND VALIDATION: AN INTRODUCTION
 
Rapid mixer graqnulator
Rapid mixer graqnulatorRapid mixer graqnulator
Rapid mixer graqnulator
 

Viewers also liked

Viscosity and its determination
Viscosity and its determinationViscosity and its determination
Viscosity and its determination
Umair hanif
 
Viskositas sebagai fungsi temperatur
Viskositas sebagai fungsi temperaturViskositas sebagai fungsi temperatur
Viskositas sebagai fungsi temperatur
Naufa Nur
 
Preparing, Diluting of solutions of different Strengths & safety measures whi...
Preparing, Diluting of solutions of different Strengths & safety measures whi...Preparing, Diluting of solutions of different Strengths & safety measures whi...
Preparing, Diluting of solutions of different Strengths & safety measures whi...
Ekta Belwal
 
Reading ppt
Reading pptReading ppt
Reading ppt
ryfshs49
 

Viewers also liked (20)

Viscometers
ViscometersViscometers
Viscometers
 
Brookfield Dv E Viscometer
Brookfield Dv   E ViscometerBrookfield Dv   E Viscometer
Brookfield Dv E Viscometer
 
Viscosity and its determination
Viscosity and its determinationViscosity and its determination
Viscosity and its determination
 
Viscosity measurement using ostwald viscometer
Viscosity measurement using ostwald viscometerViscosity measurement using ostwald viscometer
Viscosity measurement using ostwald viscometer
 
sop of testing
sop of testingsop of testing
sop of testing
 
harrison
harrisonharrison
harrison
 
Viskositas sebagai fungsi temperatur
Viskositas sebagai fungsi temperaturViskositas sebagai fungsi temperatur
Viskositas sebagai fungsi temperatur
 
Laporan mingguan Sifat Fisik Zat devi
Laporan mingguan Sifat Fisik Zat deviLaporan mingguan Sifat Fisik Zat devi
Laporan mingguan Sifat Fisik Zat devi
 
Sop for ph meter
Sop for ph meterSop for ph meter
Sop for ph meter
 
Preparing, Diluting of solutions of different Strengths & safety measures whi...
Preparing, Diluting of solutions of different Strengths & safety measures whi...Preparing, Diluting of solutions of different Strengths & safety measures whi...
Preparing, Diluting of solutions of different Strengths & safety measures whi...
 
Viscosity Measurement
Viscosity MeasurementViscosity Measurement
Viscosity Measurement
 
viskositas
viskositasviskositas
viskositas
 
Technology Transfer in Pharma Industry
Technology Transfer in Pharma IndustryTechnology Transfer in Pharma Industry
Technology Transfer in Pharma Industry
 
Reading ppt
Reading pptReading ppt
Reading ppt
 
Stanford/Arrillaga El Camino Menlo Park Plan
Stanford/Arrillaga El Camino Menlo Park PlanStanford/Arrillaga El Camino Menlo Park Plan
Stanford/Arrillaga El Camino Menlo Park Plan
 
E Mail Management In Statoil Norsk Arkivrad October 2010
E Mail Management In Statoil   Norsk Arkivrad October 2010E Mail Management In Statoil   Norsk Arkivrad October 2010
E Mail Management In Statoil Norsk Arkivrad October 2010
 
From SOA to SCA and FraSCAti
From SOA to SCA and FraSCAtiFrom SOA to SCA and FraSCAti
From SOA to SCA and FraSCAti
 
Power grid
Power gridPower grid
Power grid
 
Ringers cut 5 knit for pinch point and knuckle impact protection
Ringers cut 5 knit for pinch point and knuckle impact protectionRingers cut 5 knit for pinch point and knuckle impact protection
Ringers cut 5 knit for pinch point and knuckle impact protection
 
Abb v2
Abb v2Abb v2
Abb v2
 

Similar to Brookfield SOP and Validation

Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
fjjsekmsmmes
 
2001 Nissan Pathfinder Service Repair Manual
2001 Nissan Pathfinder Service Repair Manual2001 Nissan Pathfinder Service Repair Manual
2001 Nissan Pathfinder Service Repair Manual
fujjfjksekmdm
 
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
ufjjdjkkskemmd
 

Similar to Brookfield SOP and Validation (20)

Rheological properties using model 900 viscometer
Rheological properties using model 900 viscometerRheological properties using model 900 viscometer
Rheological properties using model 900 viscometer
 
Digital viscometer
Digital viscometer Digital viscometer
Digital viscometer
 
18 OBDG04 ECM Summary Tables
18 OBDG04 ECM Summary Tables18 OBDG04 ECM Summary Tables
18 OBDG04 ECM Summary Tables
 
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
 
2001 Nissan Pathfinder Service Repair Manual
2001 Nissan Pathfinder Service Repair Manual2001 Nissan Pathfinder Service Repair Manual
2001 Nissan Pathfinder Service Repair Manual
 
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
 
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
 
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
 
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
Daewoo doosan dx300 lc excavator service repair manual (serial number 5001 an...
 
Medidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdfMedidor digital de presion 680 UserManual.pdf
Medidor digital de presion 680 UserManual.pdf
 
Pcu230 pi fan_controlwithpf400drives_rev_a
Pcu230 pi fan_controlwithpf400drives_rev_aPcu230 pi fan_controlwithpf400drives_rev_a
Pcu230 pi fan_controlwithpf400drives_rev_a
 
IRJET- Design of Diagnosis Device for Electronic Throttle Body
IRJET-  	  Design of Diagnosis Device for Electronic Throttle BodyIRJET-  	  Design of Diagnosis Device for Electronic Throttle Body
IRJET- Design of Diagnosis Device for Electronic Throttle Body
 
1 a1
1 a11 a1
1 a1
 
lm555.pdf
lm555.pdflm555.pdf
lm555.pdf
 
Pf700 energy savings_lab
Pf700 energy savings_labPf700 energy savings_lab
Pf700 energy savings_lab
 
Hoja de datos lm 555 y configuraciones
Hoja de datos lm 555 y configuraciones Hoja de datos lm 555 y configuraciones
Hoja de datos lm 555 y configuraciones
 
2000 nissan pathfinder service repair manual
2000 nissan pathfinder service repair manual2000 nissan pathfinder service repair manual
2000 nissan pathfinder service repair manual
 
Mitsubishi ac servos melservo j4 solutions vertical form, fill & seal
Mitsubishi ac servos melservo j4 solutions vertical form, fill & sealMitsubishi ac servos melservo j4 solutions vertical form, fill & seal
Mitsubishi ac servos melservo j4 solutions vertical form, fill & seal
 
Mitsubishi ac servos melservo j4 solutions vertical form, fill & seal-die...
Mitsubishi ac servos melservo j4 solutions vertical form, fill & seal-die...Mitsubishi ac servos melservo j4 solutions vertical form, fill & seal-die...
Mitsubishi ac servos melservo j4 solutions vertical form, fill & seal-die...
 
320 c info
320 c info320 c info
320 c info
 

Recently uploaded

Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Victor Rentea
 
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Victor Rentea
 

Recently uploaded (20)

Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...
Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...
Biography Of Angeliki Cooney | Senior Vice President Life Sciences | Albany, ...
 
ICT role in 21st century education and its challenges
ICT role in 21st century education and its challengesICT role in 21st century education and its challenges
ICT role in 21st century education and its challenges
 
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobe
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, AdobeApidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobe
Apidays New York 2024 - Scaling API-first by Ian Reasor and Radu Cotescu, Adobe
 
Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024Finding Java's Hidden Performance Traps @ DevoxxUK 2024
Finding Java's Hidden Performance Traps @ DevoxxUK 2024
 
presentation ICT roal in 21st century education
presentation ICT roal in 21st century educationpresentation ICT roal in 21st century education
presentation ICT roal in 21st century education
 
Exploring Multimodal Embeddings with Milvus
Exploring Multimodal Embeddings with MilvusExploring Multimodal Embeddings with Milvus
Exploring Multimodal Embeddings with Milvus
 
Six Myths about Ontologies: The Basics of Formal Ontology
Six Myths about Ontologies: The Basics of Formal OntologySix Myths about Ontologies: The Basics of Formal Ontology
Six Myths about Ontologies: The Basics of Formal Ontology
 
MS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectorsMS Copilot expands with MS Graph connectors
MS Copilot expands with MS Graph connectors
 
AWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of TerraformAWS Community Day CPH - Three problems of Terraform
AWS Community Day CPH - Three problems of Terraform
 
MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024MINDCTI Revenue Release Quarter One 2024
MINDCTI Revenue Release Quarter One 2024
 
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
Navigating the Deluge_ Dubai Floods and the Resilience of Dubai International...
 
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
Modular Monolith - a Practical Alternative to Microservices @ Devoxx UK 2024
 
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ..."I see eyes in my soup": How Delivery Hero implemented the safety system for ...
"I see eyes in my soup": How Delivery Hero implemented the safety system for ...
 
DBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor PresentationDBX First Quarter 2024 Investor Presentation
DBX First Quarter 2024 Investor Presentation
 
Rising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdf
Rising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdfRising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdf
Rising Above_ Dubai Floods and the Fortitude of Dubai International Airport.pdf
 
Elevate Developer Efficiency & build GenAI Application with Amazon Q​
Elevate Developer Efficiency & build GenAI Application with Amazon Q​Elevate Developer Efficiency & build GenAI Application with Amazon Q​
Elevate Developer Efficiency & build GenAI Application with Amazon Q​
 
Strategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a FresherStrategies for Landing an Oracle DBA Job as a Fresher
Strategies for Landing an Oracle DBA Job as a Fresher
 
Vector Search -An Introduction in Oracle Database 23ai.pptx
Vector Search -An Introduction in Oracle Database 23ai.pptxVector Search -An Introduction in Oracle Database 23ai.pptx
Vector Search -An Introduction in Oracle Database 23ai.pptx
 
Introduction to Multilingual Retrieval Augmented Generation (RAG)
Introduction to Multilingual Retrieval Augmented Generation (RAG)Introduction to Multilingual Retrieval Augmented Generation (RAG)
Introduction to Multilingual Retrieval Augmented Generation (RAG)
 
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot TakeoffStrategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
Strategize a Smooth Tenant-to-tenant Migration and Copilot Takeoff
 

Brookfield SOP and Validation

  • 1. Presentation on APC, S. Patel 1 Guided by: Dr. V.T.Thakkar
  • 2. BROOKFIELD DIGITAL VISCOMETER • PHOTOGALLERY 0 • SOP 1 • VALIDATION 2 APC, S.P.Patel 2
  • 7. SOP 1) Turn on the viscometer and allow standing, must be auto zero. After the few seconds the screen appears which indicates 2 digits. 2) Now press the key. The screen displays the instruction to remove spindle. After removing the spindle and pressing the key the instrument begins. It is auto zeroed. 3) After approximately 15 second the screen displays the instruction to replace spindle 4) Attach the spindle to viscometer by screwing them on the lower shaft using left hand thread. APC, S.P.Patel 7
  • 8. 5) Press select spindle key and up and down arrow keys. When the desired code is displayed release the arrow key. 6) To select spindle, first press either up and down key which cause the area to show current speed, press the set speed key for adjusting the speed. 7) Insert center of this spindle in the test material until the fluid level is at the immersion groove on the spindle shaft. Tilt the spindle slightly while immersing to avoid air entrapment. 8) To measure high viscosity choose a small spindle and to slow speed if the chosen spindle/speed result in torque above 100% and then reduce the value. APC, S.P.Patel 8
  • 9. 9) Allow time for the individual reading to stabilize, record the value. 10) Press the motor on/off/escape key to turn off motor 11) The time mode allows the user to record the reading for fixed period of time or until a set torque value is attained. 12) Then enter/auto turnkey allows determining maximum calculated viscosity possible with current spindle/speed. 13) Pressing the up and down arrow keys will allow the viscometer data to be examine, pressing and other key ( except enter/ print key) will bring back normal display. 14) Turn off the mains after use. APC, S.P.Patel 9
  • 11. IQ: INSTALLATION QUALIFICATION  Inspect package for shipping damage  Confirm that all components have been supplied as detailed in the Operating Manual and Packing List  Assemble lab stand as shown in manual.  Mount viscometer head to assembled lab stand as shown in manual.  Verify appropriate spindle set (LV #1- #4)(RV/HA/HB #2-#7) APC, S.P.Patel 11
  • 12. OQ: OPERATION QUALIFICATION  Level viscometer using adjustable feet  Remove shipping cap on spindle coupling nut  Power on Viscometer and follow instructions displayed to complete auto zero  Perform oscillation check. APC, S.P.Patel 12
  • 13. PQ: PERFORMANCE QUALIFICATION  Verify calibration using a minimum of one calibrated viscosity standard fluid according to the procedure described below  Centipoise reading should be within the value stated on the fluid ± the instrument accuracy and the fluid accuracy. APC, S.P.Patel 13
  • 14. Oscillation check  This check verifies the mechanical condition of the sensing mechanism in the viscometer.  Auto-zero the instrument according to directions in the manual.  Do not attach the spindle but do press any key so the main screen will be displayed.  The display should show 0.0% torque. The tolerance for this and all tests is ±0.1% APC, S.P.Patel 14
  • 15. 1) Start the motor rotating at 10 or 20 RPM. 2) The % reading should initially bounce, then is should stabilize at 0.0% ±0.1% 3) Stop the motor. When the % display again stabilizes it should return to 0.0% ±0.1% APC, S.P.Patel 15
  • 16. Operational and Calibration Check  Determining Measurement Accuracy  All Brookfield instruments are guaranteed to be accurate to within 1% of the full scale range(FSR) in use.  The full scale range (FSR) is simply the maximum viscosity that can be measured with any spindle/speed combination.  Select a specific spindle and rotational speed.  To determine the resulting FSR (full scale range), depress and hold the AUTORANGE key. APC, S.P.Patel 16
  • 17.  The % display will show 100% and the cP display will show the FSR (full scale range) that can be measured under current choice of spindle and speed.  1% of this value is the measurement accuracy under these conditions. APC, S.P.Patel 17
  • 18.  Calibration Check  In principle, all Brookfield rotational viscometers are calibrated torque meters.  Brookfield offers a variety of viscosity calibration fluids that enable one to check the calibration and linearity of the instrument over its operating torque range of 10%- 100% torque. APC, S.P.Patel 18
  • 19.  Any spindle and any one of several calibration fluids may be used to perform a calibration check.  Experts recommend using one spindle and one fluid while taking measurements at multiple speeds.  This enables multiple calculations of viscosity throughout the torque range of the instrument;  one at low torque, one at medium torque and one at high torque.  The FSR, and consequently the instrument accuracy in terms of centipoise, will be different at each speed. APC, S.P.Patel 19
  • 20.  The calibration fluid must be transferred into a 600ml low form, Griffin beaker (without creating any bubbles) and,  on LV and RV models, the instrument Guard Leg must be used.  The beaker, fluid, spindle and Guard Leg must stabilized at 25.0°C (±0.1°C) before the calibration check can be performed. APC, S.P.Patel 20
  • 21. EX: Actual viscosity is 5,120cP,  Allowable error = 1% of the FSR + 1% of the actual value of the calibration fluid  The lower and upper viscosity limits, indicating acceptable calibration, equal the actual value of the fluid,  in this case 5,120cP, ± the allowable error. APC, S.P.Patel 21
  • 22. RPM MIN cP MAX cP (FSR) 1% of FSR % torque cP READIN G ALLOWAB LE ERROR 12 1000 10000 100 51.2 5120 ± 151 6 2000 20000 200 25.5 5100 ± 251 3 4000 40000 400 12.7 5080 ± 451 Allowable error = 1% of the FSR + 1% of the actual value of the calibration fluid APC, S.P.Patel 22
  • 23. Calibration Worksheet A B C D E F G H I Fluid Value from Certificate cP 1% of Fluid Value cP Instrument Model Spindle RPM FSR cP (depre ss and hold AUTORA NGE key) 1% of FSR cP 1% of FSR cP Visco sity cP APC, S.P.Patel 23
  • 24.  The viscosity reading should equal the cP value on the fluid standard to within the combined accuracies of the Viscometer and the viscosity standard as calculated below Low Limit = A – (B + G) Reading I High Limit = A + (B + G) Pass or Fail APC, S.P.Patel 24
  • 25.  Any reading outside these limits may indicate a Viscometer problem.  Contact the Brookfield technical sales department or your local Brookfield dealer/distributor with test results to determine the nature of the problem. APC, S.P.Patel 25
  • 26. Reference:  Brookfield DV-ii+ pro, programmable viscometer, operating manual no. M/03- 165-A0404  “Solution for a viscous world” brookfield 2007 catalog.  www.brookfieldengineering.com APC, S.P.Patel 26