O & M Manual
Model Q46CT
Toroidal Conductivity
Monitor
Home Office European Office
Analytical Technology, Inc. ATI (UK) Limited
6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park
Collegeville, PA 19426 Delph New Road, Delph
Phone: 800-959-0299 Saddleworth OL3 5DE
610-917-0991 Phone: +44 (0)1457-873-318
Fax: 610-917-0992 Fax: + 44 (0)1457-874-468
Email: sales@analyticaltechnology.com Email: sales@atiuk.com
2
O&M Manual
Rev-G (8/17)
Table of Contents
PART 1 - INTRODUCTION.....................................4
1.1 General ................................................. 4
1.2 Features................................................ 4
1.3 Q46CT System Specifications.............. 5
1.4 Q46CT Performance Specifications ..... 7
PART 2 – ANALYZER MOUNTING .......................8
2.1 General ................................................. 8
2.2 Wall or Pipe Mount ............................... 9
2.3 Panel Mount Monitor.......................... 11
PART 3 – SENSOR MOUNTING......................... 12
3.1 General ............................................... 12
3.2 Tee Mounting...................................... 13
3.3 Submersion Mounting......................... 14
3.4 In-Line Installation............................... 15
PART 4 – ELECTRICAL INSTALLATION .......... 17
4.1 General ............................................... 17
4.2 Power Connection .............................. 17
4.3 Relay Connection ............................... 19
4.4 Optional Output or Relay Connections20
4.5 Sensor Wiring ..................................... 21
4.6 Direct Sensor Connection................... 23
4.7 Junction Box Connection.................... 24
PART 5 – CONFIGURATION .............................. 25
5.1 User Interface ..................................... 25
5.11 Keys ................................................ 26
5.12 Display ............................................ 26
5.2 Software.............................................. 27
5.22 Measure Menu [MEASURE] ........... 30
5.23 Calibration Menu [CAL] ........................ 31
5.24 Configuration Menu [CONFIG]....... 31
5.25 Control Menu................................... 35
5.26 Diagnostics Menu [DIAG] .................... 38
PART 6 – CALIBRATION.................................... 41
6.1 Overview and Methods....................... 41
6.11 1-Point Cal Explained ..................... 41
6.12 Zero Cal Cal Explained ................... 41
6.2 Performing Sensor Zero Cal............... 41
6.3 Performing 1-Point Cal ....................... 42
6.4 Temperature Calibration..................... 43
PART 7 – PID CONTROLLER DETAILS ............ 45
7.1 PID Description................................... 45
7.2 PID Algorithm...................................... 45
7.3 Classical PID Tuning .......................... 47
7.4 Manual PID Override Control ............. 47
7.5 Common PID Pitfalls........................... 47
PART 8 – SYSTEM MAINTENANCE .................. 49
8.1 System Checks................................... 49
8.2 Instrument Checks.............................. 49
8.3 Sensor Tests....................................... 49
8.4 Display Messages............................... 50
SPARE PARTS.................................................... 53
3
O&M Manual
Rev-G (8/17)
Table of Figures
FIGURE 1 - Q46 ENCLOSURE DIMENSIONS........................................................................................................8
FIGURE 2 - WALL OR PIPE MOUNT BRACKET.....................................................................................................9
FIGURE 3 - WALL MOUNTING DIAGRAM...........................................................................................................10
FIGURE 4 - PIPE MOUNTING DIAGRAM ............................................................................................................10
FIGURE 5 - AC PWR PANEL MOUNT CUTOUT ..................................................................................................11
FIGURE 6 - TOROIDAL SENSOR DIMENSIONS...................................................................................................12
FIGURE 7 – SENSOR AND FLOW TEE ASSEMBLY..............................................................................................13
FIGURE 8 - SENSOR SUBMERSION MOUNT......................................................................................................14
FIGURE 9 - FLOW TEE PROCESS PIPING .........................................................................................................15
FIGURE 10 - IN-LINE PROCESS PIPING............................................................................................................16
FIGURE 11 - Q46 CONNECTION LAYOUT.........................................................................................................18
FIGURE 12 - RELAY CONNECTIONS.................................................................................................................19
FIGURE 13 - OPTIONAL RELAY BOARD WIRING ...............................................................................................20
FIGURE 14 - OPTIONAL ANALOG OUTPUT WIRING ...........................................................................................20
FIGURE 15 - Q46CT SENSOR CONNECTIONS .................................................................................................22
FIGURE 16 - SENSOR CABLE PREPARATION....................................................................................................23
FIGURE 17 - JUNCTION BOX INTERCONNECT WIRING.......................................................................................24
FIGURE 18 - USER INTERFACE .......................................................................................................................25
FIGURE 19 - SOFTWARE MAP.........................................................................................................................29
FIGURE 20 - CONTROL RELAY EXAMPLE.........................................................................................................37
FIGURE 21 - ALARM RELAY EXAMPLE .............................................................................................................37
FIGURE 22 - NACL REFERENCE SOLUTION CALIBRATION CHART .....................................................................42
FIGURE 23 - PID EQUATION...........................................................................................................................45
FIGURE 24 - PT100 RTD RESISTANCE TABLE.................................................................................................50
4
O&M Manual
Rev-G (8/17)
Part 1 - Introduction
1.1 General
The Model Q46CT Conductivity monitor/analyzer provides a versatile system for monitoring and
control of conductivity over a range of 1 µS/cm to 2.000 Siemens/cm. The display may be
programmed for ranges of 0-2000 uS, 0-2.000 mS, 0-20.00 mS, 0-200 mS, 0-2000 mS, or 0-
2.000 S. A lower line display indicates the equivalent TDS value in mg/l, using a user
programmable conversion factor. Q46CT monitors use a toroidal conductivity sensor that
measures conductivity inductively. This type of sensor is particularly well suited to
measurements in solutions that contain materials that would foul a normal conductivity sensor. It
is very good for high conductivity measurements, but should not be used if normal conductivity is
below 100 uS.
For special applications, the Q46CT has a few common pre-programmed tables that allow the
user to monitor “percent concentration”. Selectable tables include NaCl (0-20%), HCl (0-18%),
NaOH (0-15%), and KOH (0-15%). If data is available, you can enter your own table data into
the Q46CT to provide a customized concentration display for other chemicals in solution.
Monitors are available in two electronic versions, a universal AC powered monitor for operation
from 90-260 VAC and a 12-24 VDC unit. Both versions provide two 4-20 mA analog outputs and
3 SPDT relays. One analog output may be configured for PID control and one of the relays may
be configured to provide a remote trouble indication.
The Q46CT is available with a few options to expand the capabilities of the monitor. Users may
select either a third analog 4-20 mA output or an expansion card that provides three additional
low voltage SPST relays. The expansion relays are not isolated from each other.
This monitor also provides digital communication capability. Users may order Q46 systems with
either Profibus DP, Modbus RTU, or Ethernet communications boards factory installed.
WARNING: Not following operating instructions may impair safety.
1.2 Features
 Power options for either AC or low voltage DC operation.
 High accuracy, high sensitivity system, measures from 1 to 2,000,000 uS through 7 internal
automatic ranges. User display ranges include 2000 us, 2.000 mS, 20.00 mS, or 200.0 mS, 2000
mS, or 2.000 S.
 Output Hold, Output Simulate, Output Alarm, and Output Delay. All forced changes in output
condition include bumpless transfer to provide gradual return to on-line signal levels and to avoid
system control shocks on both analog outputs.
 Selectable PID controller on main analog output. PID includes manual operation feature, and
diagnostic “stuck-controller” timer feature for relay notification of control problems.
ATI Q46CT Conductivity System Part 1– Introduction
5
O&M Manual
Rev-G (8/17)
 Digital communication option for Profibus-DP, Modbus-RTU. More versions pending. See Q46
Communications Manual for Specificaitons.
 Configurable for TDS display and signal output on one analog output.
 Two analog outputs track conductivity and temperature, conductivity and conductivity, or
conductivity and TDS. Both analog outputs can be individually programmed to fail to specific
values.
 Large, high contrast, custom LCD display with LED back light provides excellent readability in any
light conditions. The second line of display utilizes 5x7 dot matrix characters for clear message
display.
 Toroidal conductivity sensor. Sensing electrodes are molded in a noryl body and completely
isolated from the measured solution. Conductivity measurements can be made in almost any
aqueous solution without being affected by high solids, oils, and other materials that foul
contacting sensors.
 Quick and easy one-point calibration method and sensor zero-cal. To provide high accuracy, all
calibration methods include stability monitors that check temperature and main parameter stability
before accepting data.
 High accuracy Pt1000 temperature input. Temperature element can be user calibrated.
 Security lock feature to prevent unauthorized tampering with transmitter settings. All settings can
be viewed while locked, but they cannot be changed.
1.3 Q46CT System Specifications
Displayed Parameters Main input, 0 uS to 2S (2,000,000 uS)
Sensor temperature, -10.0 to 210.0 °C (23 to 410 ºF)
Loop current, 4.00 to 20.00 mA
Sensor slope
Model number and software version
PID Controller Status
Timer Status
Main Parameter Ranges Automatic or Manual selection of display range
0 to 2000 uS
0.0 to 2.000 mS
0.00 to 20.00 mS
0.0 to 200.0 mS
0 to 2000 mS
0.000 to 2.000 S
Display 0.75” (19.1 mm) high 4-digit main display with sign
12-digit secondary display, 0.3" (7.6 mm) 5x7 dot matrix.
Integral LED back-light for visibility in the dark.
Power 90 - 260 VAC, 50-60 Hz, 10 VA max
12-24 VDC, 500 mA max.
ATI Q46CT Conductivity System Part 1– Introduction
6
O&M Manual
Rev-G (8/17)
Enclosure NEMA 4X, polycarbonate, stainless steel hardware,
weatherproof and corrosion resistant.
Mounting Options Wall or pipe mount bracket standard. Bracket suitable for either
1.5” or 2” I.D. U-Bolts for pipe mounting.
Panel mount adapter optional.
Conduit Openings Five ½” NPT openings, Adapter can be removed to provide a 1”
NPT opening in the bottom of the enclosure. Gland seals
provided but not installed.
Relays, Electromechanical: Three SPDT, 6 amp @ 250 VAC, 5 amp @ 24 VDC contacts.
Software selection for setpoint, phase, delay, deadband, hi-lo
alarm, and failsafe.A-B indicators on main LCD, and C indicator
on lower display.
Analog Outputs Two 4-20 mA outputs. Output one programmable for pH or PID.
Output 2 programmable for pH or Temperature. Max load 500
Ohms for each output. Outputs ground isolated and isolated
from each other. An additional 3
rd
analog option is available.
Keypad 4-key membrane type with tactile feedback, polycarbonate with
UV coating
Weight 2 lb. (0.9 kg)
Ambient Temperature Analyzer Service, -20 to 60 °C (-4 to 140 ºF)
Storage, -30 to 70 °C (-22 to 158 ºF)
Ambient Humidity 0 to 95%, indoor/outdoor use, non-condensing.
Altitude Up to 2000 m (6562 Ft.)
Electrical Certification Ordinary Location, cCSAus (CSA and UL standards - both
approved by CSA), pollution degree 2, installation category 2
EMI/RFI Influence Designed to EN 61326-1
Output Isolation 600 V galvanic isolation
Q25CT Conductivity sensor
Sensor Toroidal sensor with encapsulated temp. element.
Measuring Range 0.000 to 2.000 S/cm
Temperature Element Pt1000 RTD
Wetted Materials Noryl
Optional Flow Tee Special 2” CPVC tee with alignment slot
Sensor Cable 20 ft. (6 m.) standard.
ATI Q46CT Conductivity System Part 1– Introduction
7
O&M Manual
Rev-G (8/17)
Pressure Range 0-150 PSIG
Temperature Range 0-105° C
Maximum Flow Rate 10 feet (3 meters) per second
Max. Sensor-Analyzer
Distance 200 feet (60 m.)
1.4 Q46CT Performance Specifications
Accuracy 0.5% of user range, or better (± 2 μS)
Repeatability 0.2% of user range, or better (± 2 μS)
Sensitivity 0.05% of user ranges (± 2 μS)
Stability 0.2% of user range per 24 hours, non-cumulative
Warm-up Time 3 seconds to rated performance (electronics only)
Supply Voltage Effects ± 0.05% of user range
Instrument Response Time 6 seconds to 90% of step input at lowest setting
Temperature Drift Span or zero, 0.04% of span/°C
Equipment bearing this marking may not be discarded by traditional methods in the
European community after August 12 2005 per EU Directive 2002/96/EC. End
users must return old equipment to the manufacturer for proper disposal.
8
O&M Manual
Rev-G (8/17)
Part 2 – Analyzer Mounting
2.1 General
All Q46 Series instruments offer maximum mounting flexibility. A bracket is included with each
unit that allows mounting to walls or pipes. In all cases, choose a location that is readily
accessible for calibrations. Also consider that it may be necessary to utilize a location where
solutions can be used during the calibration process. To take full advantage of the high contrast
display, mount the instrument in a location where the display can be viewed from various angles
and long distances.
Locate the instrument in close proximity to the point of sensor installation - this will allow easy
access during calibration. The standard cable length of the conductivity sensor is 20 feet. For
sensor cables longer than 20 feet, use the optional junction box (07-0100) and sensor
interconnect cable (31-0057).
Figure 1 - Q46 Enclosure Dimensions
ATI Q46CT Conductivity System Part 2– Analyzer Mounting
9
O&M Manual
Rev-G (8/17)
2.2 Wall or Pipe Mount
A PVC mounting bracket with attachment screws is supplied with each transmitter. The multi-
purpose bracket is attached to the rear of the enclosure using the four flat head screws. The
instrument is then attached to the wall using the four outer mounting holes in the bracket. These
holes are slotted to accommodate two sizes of u-bolt that may be used to pipe mount the unit.
Slots will accommodate u-bolts designed for 1½“ or 2” pipe. The actual center to center
dimensions for the u-bolts are shown in the drawing. Note that these slots are for u-bolts with ¼-
20 threads. The 1½” pipe u-bolt (2” I.D. clearance) is available from ATI in type 304 stainless
steel under part number (47-0005)
Figure 2 - Wall or Pipe Mount Bracket
ATI Q46CT Conductivity System Part 2– Analyzer Mounting
10
O&M Manual
Rev-G (8/17)
Figure 3 - Wall Mounting Diagram
Figure 4 - Pipe Mounting Diagram
Note: Analyzer shown with optional
Profibus Connector mounted to
side of enclosure.
ATI Q46CT Conductivity System Part 2– Analyzer Mounting
11
O&M Manual
Rev-G (8/17)
2.3 Panel Mount Monitor
Panel mounting uses the panel mounting flange molded into the rear section of the enclosure.
Figure 5 provides dimensions for the panel cutout required for mounting.
The panel mounting bracket kit must be ordered separately (part number 05-0094). This kit
contains a metal retainer bracket that attaches to the rear of the enclosure, 4 screws for
attachment of this bracket, and a sealing gasket to insure that the panel mounted monitor
provides a water tight seal when mounted to a panel.
The sealing gasket must first be attached to the enclosure. The gasket contains an adhesive on
one side so that it remains in place on the enclosure. Remove the protective paper from the
adhesive side of the gasket and slide the gasket over the back of the enclosure so that the
adhesive side lines up with the back of the enclosure flange. Once in place, you can proceed to
mount the monitor in the panel.
Figure 5 - AC Pwr Panel Mount Cutout
12
O&M Manual
Rev-G (8/17)
Part 3 – Sensor Mounting
3.1 General
The Q25CT Conductivity Sensor is designed for industrial and municipal process applications.
Mounting options include flow-through or submersion. The sensor-to-analyzer distance must not
exceed 200 feet (60 m.).
In submersion applications, there must be a minimum of 2” clearance around and below
the toroidal sensor. Placing the sensor close to wall or other objects will result in a loss of
accuracy.
Calibrate the sensor before placing it into the process. See Part 7 of this manual for detailed
calibration instructions.
Figure 6 below shows the dimensions of the toroidal sensor.
Figure 6 - Toroidal Sensor Dimensions
ATI Q46CT Conductivity System Part 3 – Sensor Mounting
13
O&M Manual
Rev-G (8/17)
3.2 Tee Mounting
The toroidal sensors are mounted in a 2" pipe using an optional pipe adapter. The tee fitting is
keyed so the sensor is oriented in the process as shown in Figure 7, below. The sensor bore
opening should be aligned so that flow passes directly through the open sensor bore. A positioning
notch is located on the upper sensor collar to aid in this alignment. This orientation is used to
ensure a representative sample is being measured and to keep the sensor bore clean.
Note that sensor must be zero-calibrated before final mounting in tee.
2-WIRE TOROIDAL
CONDUCTIVITY TRANSMITTER
Figure 7 – Sensor and Flow Tee Assembly
ATI Q46CT Conductivity System Part 3 – Sensor Mounting
14
O&M Manual
Rev-G (8/17)
3.3 Submersion Mounting
When using this conductivity sensor for submersion applications, mount the sensor to the end of a
1” mounting pipe using a 1” coupling. ATI’s (00-0628) mounting assembly shown in 11 is available
for submersible applications. This assembly is designed to mount to standard handrails and
facilitates insertion and removal of the sensor.
Figure 8 - Sensor Submersion Mount
ATI Q46CT Conductivity System Part 3 – Sensor Mounting
15
O&M Manual
Rev-G (8/17)
3.4 In-Line Installation
Toroidal Conductivity sensors may be installed directly into a flowing pipe system provided that
the water does not contain a lot of entrained air. A 2” flow tee assembly is available for this
purpose. It is best to install the sensor in a vertical pipe section with water flowing upward. This
assures that air pockets cannot develop at the sensor. If installed in a horizontal run of pipe,
place the sensor at the 3 or 9 o’clock position. Never mount the sensor on the top or bottom of
the pipe. It is also good practice to install a bypass system around the sensor for maintenance
and calibration purposes.
Figure 9 - Flow Tee Process Piping
ATI Q46CT Conductivity System Part 3 – Sensor Mounting
16
O&M Manual
Rev-G (8/17)
Figure 10 - In-Line Process Piping
17
O&M Manual
Rev-G (8/17)
Part 4 – Electrical Installation
4.1 General
The Q46 is powered in one of two ways, depending on the version purchased. The 12-24 VDC
powered analyzer requires a customer supplied DC power supply. The 90-260 VAC version
requires line power. Please verify the type of unit before connecting any power.
WARNING: Do not connect AC line power to the DC version. Severe damage will result.
Important Notes:
1. Use wiring practices that conform to all national, state and local electrical codes. For
proper safety as well as stable measuring performance, it is important that the earth
ground connection be made to a solid ground point from TB7. The AC power supply
contains a single 630mA slo-blo fuse (Wickmann/Littlefuse #372-0630). The fuse F1
is located adjacent to TB7 and is easily replaceable.
2. Do NOT run sensor cables or instrument 4-20 mA output wiring in the same conduit
that contains AC power wiring. AC power wiring should be run in a dedicated conduit
to prevent electrical noise from coupling with the instrumentation signals.
3. This analyzer must be installed by specifically trained personnel in accordance with
relevant local codes and instructions contained in this operating manual. Observe
the analyzer's technical specifications and input ratings. Proper electrical
disconnection means must be provided prior to the electrical power connected to this
instrument, such as a circuit breaker - rated 250 VAC, 2 A minimum. If one line of
the line power mains is not neutral, use a double-pole mains switch to disconnect the
analyzer.
4. Repeated problems with lightning strikes damaging sensitive instrumentation are
often attributed to poorly bonded earth grounds in the instrument power source. The
protection schemes incorporated into this analyzer cannot operate to maximum
efficiency unless the ground connection is at its’ absolute lowest impedance.
There is no standard ground resistance universally recognized. Many agencies
recommend a ground resistance value of 5 ohms or less. The NEC recommends an
impedance to ground of less than 25 ohms, and less than 5 ohms where sensitive
equipment is installed. Power sources feeding sensitive instruments like the Q46
should have the lowest possible impedance to ground.
4.2 Power Connection
Verify the AC power supply requirement before installing. Also verify that power is fully
disconnected before attempting to wire.
Q46 systems are supplied with 5 cable gland fittings for sealing cable entries.
Connect HOT, NEUTRAL, and GROUND to the matching designations on terminal strip TB7.
ATI Q46CT Conductivity System Part 4 – Electrical Installation
18
O&M Manual
Rev-G (8/17)
The two analog outputs for the standard system are present at terminal TB1. The loop-load
limitation in this configuration is 500 Ohms maximum for each output. Also note that these two
outputs are completely isolated from each other to insure that ground loops do not result from the
connection of both outputs to the same device such as a PLC or DCS.
A ribbon cable connects the power supply assembly with the microprocessor assembly located in
the front section of the enclosure. This cable may be unplugged from the front section of the
monitor if service is needed, but should normally be left in place during installation.
Figure 11 - Q46 Connection Layout
WARNING
Disconnect line power voltage BEFORE connecting line power
wires to Terminal TB7 of the power supply. The power supply
accepts only standard three-wire single phase power. AC power
supplies are configured for 100-240 VAC operation at the factory
at time of order, and the power supply is labeled as such. Do
NOT connect voltages other than the labeled requirement to the
input.
ATI Q46CT Conductivity System Part 4 – Electrical Installation
19
O&M Manual
Rev-G (8/17)
4.3 Relay Connection
Three SPDT relays are provided on the power supply board. None of the relay contacts are
powered. The user must supply the proper power to the contacts. For applications that require
the same switched operating voltage as the Q46 (115 or 230 V), power may be jumpered from
the power input terminals at TB7. Relay wiring is connected at TB4, TB5, and TB6 as shown
below. Note that the relay contact markings are shown in the NORMAL mode. Programming a
relay for “Failsafe” operation reverses the NO and NC positions in this diagram (Figure 12).
Figure 12 - Relay Connections
ATI Q46CT Conductivity System Part 4 – Electrical Installation
20
O&M Manual
Rev-G (8/17)
4.4 Optional Output or Relay Connections
TB2, is used to connect to the optional 3-relay card (Figure 13) OR the optional third analog
output Out#3, (Figure 14). The Q46 can be configured for only one of these optional features,
and the hardware for either option must be factory installed.
Figure 13 - Optional Relay Board Wiring
Figure 14 - Optional Analog Output Wiring
ATI Q46CT Conductivity System Part 4 – Electrical Installation
21
O&M Manual
Rev-G (8/17)
4.5 Sensor Wiring
The sensor cable can be quickly connected to the Q46 terminal strip by matching the wire colors
on the cable to the color designations on the label in the monitor. A junction box is also available
to provide a break point for long sensor cable runs. Route signal cable away from AC power
lines, adjustable frequency drives, motors, or other noisy electrical signal lines. Do not run
sensor or signal cables in conduit that contains AC power lines or motor leads.
When installing conductivity monitors and sensors, we recommend keeping the sensor cable as
short as is practical. This will minimize potential noise problems.
Dad DANGER: DO NOT connect sensor cable to power lines. Serious injury may result.
Take care to route sensor cable away from AC power lines, adjustable frequency drives, motors,
or other noisy electrical signal lines. Do not run signal lines in the same conduit as AC power
lines. Run signal cable in dedicated metal conduit if possible. For optimum electrical noise
protection, run an earth ground wire to the ground terminal in the transmitter.
Refer to Figure 11 & Figure 15 for illustrative details on electrical installation.
ATI Q46CT Conductivity System Part 4 – Electrical Installation
22
O&M Manual
Rev-G (8/17)
Figure 15 - Q46CT Sensor Connections
ATI Q46CT Conductivity System Part 4 – Electrical Installation
23
O&M Manual
Rev-G (8/17)
4.6 Direct Sensor Connection
Sensor connections are made in accordance with Figure 15. The sensor cable can be routed into
the enclosure through one of cord-grips supplied with the unit. Routing sensor wiring through
conduit is only recommended if a junction box is to be used. Some loose cable is needed near
the installation point so that the sensor can be inserted and removed easily depending on the
installation type.
Cord-grips used for sealing the cable should be snugly tightened after electrical connections have
been made to prevent moisture incursion. When stripping cables, leave adequate length for
connections in the transmitter enclosure as shown below. The standard 20 ft. sensor cable
normally supplied with the system is already stripped and ready for wiring. This cable can be cut
to a shorter length if desired to remove extra cable in a given installation. Do not cut the cable so
short as to make installation and removal of the sensor difficult.
Figure 16 - Sensor Cable Preparation
ATI Q46CT Conductivity System Part 4 – Electrical Installation
24
O&M Manual
Rev-G (8/17)
4.7 Junction Box Connection
When sensor separation from the monitor is needed and to be greater than 20 feet, the sensor
junction box is required (07-0100). Wire according to Figure 17 with 3 paired, individually shielded
22 AWG cable (31-0068).
CAUTION: When using a junction box and sensor interconnect cable, the RED SHIELD
must be isolated from the WHITE and GREEN SHIELDS. Failure to maintain
isolation with the RED SHIELD will result in measurement instability.
Figure 17 - Junction Box Interconnect Wiring
25
O&M Manual
Rev-G (8/17)
Part 5 – Configuration
5.1 User Interface
The user interface for the Q46 Series instrument consists of a custom display and a membrane
keypad. All functions are accessed from this user interface (no internal jumpers, pots, etc.).
Figure 18 - User Interface
MENU ICONS
UNITS
12-CHARACTER
SECONDARY
DISPLAY
MEMBRANE
KEYPAD
MENU
ESC
ENTER
A
B
DIAG
FAIL
HOLD
CAL
CONF
MENU ICONS
UNITS
12-CHARACTER
SECONDARY
DISPLAY
MEMBRANE
KEYPAD
ENTER KEY
LEFT ARROW
KEY
4-DIGIT
MAIN DISPLAY
MENU/ESCAPE
KEY
UP ARROW
KEY
SIGN
RELAY/LO-BAT
INDICATOR
4-KEY USER
INTERFACE
RELAY
INDICATOR
ATI Q46CT Conductivity System Part 5 – Configuration
26
O&M Manual
Rev-G (8/17)
5.11 Keys
All user configurations occur through the use of four membrane keys. These keys are used as
follows:
MENU/ESC To scroll through the menu section headers or to escape from anywhere in
software. The escape sequence allows the user to back out of any changes in
a logical manner. Using the escape key aborts all changes to the current
screen and backs the user out one level in the software tree. The manual will
refer to this key as either MENU or ESC, depending upon its particular
function.
UP (arrow) To scroll through individual list or display items and to change number values.
LEFT (arrow) To move the cursor from right to left during changes to a number value.
ENTER To select a menu section or list item for change and to store any change.
5.12 Display
The large custom display provides clear information for general measurement use and user
configuration. There are three main areas of the display: the main parameter display, the
secondary message line, and the icon area.
Main Parameter During normal operation, the main parameter display indicates the present
process input with sign and units. This main display may be configured to
display any of the main measurements that the system provides. During
configuration, this area displays other useful set-up information to the user.
Lower Line During normal operation, the lower line of the display indicates user-selected
secondary measurements that the system is making. This also includes
calibration data from the last calibration sequence and the transmitter model
number and software version. During configuration, the lower line displays
menu items and set-up prompts to the user. Finally, the lower line will display
error messages when necessary. For a description of all display messages,
refer to Section 8.5.
ATI Q46CT Conductivity System Part 5 – Configuration
27
O&M Manual
Rev-G (8/17)
Icon Area The icon area contains display icons that assist the user in set-up and indicate
important states of system functions. The CAL, CONFIG, CNTRL, and DIAG
icons are used to tell the user what branch of the software tree the user is in
while scrolling through the menu items. This improves software map
navigation dramatically. Upon entry into a menu, the title is displayed (such as
CAL), and then the title disappears to make way for the actual menu item.
However, the icon stays on.
HOLD The HOLD icon indicates that the current output of the transmitter has been put
into output hold. In this case, the output is locked to the last input value
measured when the HOLD function was entered. HOLD values are retained
even if the unit power is cycled.
FAIL The FAIL icon indicates that the system diagnostic function has detected a
problem that requires immediate attention. This icon is automatically cleared
once the problem has been resolved.
Relay Area A/B The relay area contains two icons that indicate the state of the system relays.
Relay C is normally configured for FAIL indication, so it is only displayed on the
lower MEASURE display line.
5.2 Software
The software of the Q46CT is organized in an easy to follow menu-based system. All user
settings are organized under five menu sections: Measure, Calibration [CAL], Configuration
[CONFIG], Control [CONTROL] and Diagnostics [DIAG].
Note: The default Measure Menu is display-only and has no menu icon.
A
B
ATI Q46CT Conductivity System Part 5 – Configuration
28
O&M Manual
Rev-G (8/17)
5.21 Software Navigation
Within the CAL, CONFIG, CONTROL, and DIAG menu sections is a list of selectable items.
Once a menu section (such as CONFIG) has been selected with the MENU key, the user can
access the item list in this section by pressing either the ENTER key or the UP arrow key. The
list items can then be scrolled through using the UP arrow key. Once the last item is reached, the
list wraps around and the first list item is shown again. The items in the menu sections are
organized such that more frequently used functions are first, while more permanent function
settings are later in the list. See Figure 19 for a visual description of the software.
Each list item allows a change to a stored system variable. List items are designed in one of two
forms: simple single variable, or multiple variable sequences. In the single variable format, the
user can quickly modify one parameter - for example, changing temperature display units from °F
to °C. In the multiple variable sequence, variables are changed as the result of some process.
For example, the calibration of conductivity generally requires more than one piece of information
to be entered. The majority of the menu items in the software consist of the single variable format
type.
Any data that may be changed will be flashing. This flashing indicates user entry mode and is
initiated by pressing the ENTER key. The UP arrow key will increase a flashing digit from 0 to 9.
The LEFT arrow key moves the flashing digit from right to left. Once the change has been
completed, pressing ENTER again stores the variable and stops the flashing. Pressing ESC
aborts the change and also exits user entry mode.
The starting (default) screen is always the Measure Menu. The UP arrow key is used to select
the desired display. From anywhere in this section the user can press the MENU key to select
one of the four Menu Sections.
The UP arrow icon next to all list items on the display is a reminder to scroll through the list using
the UP arrow key.
To select a list item for modification, first select the proper menu with the MENU key. Scroll to the
list item with the UP arrow key and then press the ENTER key. This tells the system that the user
wishes to perform a change on that item. For single item type screens, once the user presses the
ENTER key, part or all of the variable will begin to flash, indicating that the user may modify that
variable using the arrow keys. However, if the instrument is locked, the transmitter will display
the message Locked! and will not enter user entry mode. The instrument must be unlocked by
entering the proper code value to allow authorized changes to user entered values. Once the
variable has been reset, pressing the ENTER key again causes the change to be stored and the
flashing to stop. The message Accepted! will be displayed if the change is within pre-defined
variable limits. If the user decides not to modify the value after it has already been partially
changed, pressing the ESC key aborts the modification and returns the entry to its original stored
value.
In a menu item which is a multiple variable sequence type, once the ENTER key is pressed there
may be several prompts and sequences that are run to complete the modification. The ESC key
can always be used to abort the sequence without changing any stored variables.
ATI Q46CT Conductivity System Part 5 – Configuration
29
O&M Manual
Rev-G (8/17)
MEASURE CAL CONFIG DIAG
or or or
Cal Cond
1
PID 0% #1 Set Hold
Cal Temp
1
PID 100% #1
1
PID Setpoint #1
1
PID Prop #1
Fault List
Sim Out
Fail Out #1
Fail Val #1
Fail Out #2
Offset
Temperature
1
PID % Output
Loop Current (#1)
LIST
ITEMS
Cal Zeros
Loop Current (#2)
Set Default
4
Loop Current (#3)
CONTROL
Fail Val #2
Failsafe
1
PID Int #1
1
PID Deriv #1
Set 4mA (#1)
Set 20mA (#1)
or
Entry Lock
Set Delay
Contrast
Ref Temp
I out 1 Mode
I out 2 Mode
Relay B Mode
Temp Units
Set 4mA (#2)
Set 20mA (#2)
TDS
5
Aux rly=
1
PID Timer
Cell Factor
MENU
SECTIONS
Software Version
Main Display
Zero Filter
Slope
Relay A Mode
4
I out 3 Mode
Relay C Mode
5
Relay E Mode
5
Relay D Mode
5
Relay F Mode
4
Set 4mA (#3)
4
Set 20mA (#3)
6
Setpnt A (or A-HI, A-LO)
6
Hyst A (or A-HI, A-LO)
6
Delay A (or A-HI, A-LO)
Phase A
Setpnt B
Hyst B
Delay B
Phase B
Setpnt C
Hyst C
Delay C
Phase C
5
Setpnt D
5
Hyst D
5
Delay D
5
Phase D
5
Setpnt E
5
Hyst E
5
Delay E
5
Phase E
5
Setpnt F
5
Hyst F
5
D elay F
5
Phase F
4
Fail Out #3
4
Fail Val #3
Backlight
Notes:
(1) If Relay A,B,C,D,E,F is set to FAIL mode, relay settings are not
displayed in menu.
(2) The annunciator for Relay C is shown in the MEASURE/
temperature display
PID is enabled
Linear Temp Comp Selected
Table Temp Comp Selected
Optional third 4-20 output installed
Optional 3-relay card installed (D,E,F)
If Relay A is set to ALARM mode, the settings are divided into
2 groups of HI and LO points.
1
2
3
4
5
6
7
If Comm Mode is set to a selection other than none,
additional Comm menus will show.
Start Delay
TC Factor
Cond Range
TC Method
2
Lin T-Comp
3
TC Comp Tbl
TDS Factor
Comm Mode
7
Comm Address
Figure 19 - Software Map
ATI Q46CT Conductivity System Part 5 – Configuration
30
O&M Manual
Rev-G (8/17)
5.22 Measure Menu [MEASURE]
The default menu for the system is the display-only menu MEASURE. This menu is a display-only
measurement menu, and has no changeable list items. When left alone, the instrument will
automatically return to this menu after approximately 30 minutes. While in the default menu, the
UP arrow allows the user to scroll through the secondary variables on the lower line of the
display. A brief description of the fields in the basic transmitter version is as follows:
TRANSMITTER MEAS SCREENS:
25.7°C Temperature display. Can be displayed in °C or °F, depending on user
selection. A small “m” to the left side of the temperature indicates the
transmitter has automatically jumped to a manual 25°C setting due to a
failure with the temperature signal input. An L
C
symbol to the left of the
temperature will appear if relay C activates.
TDS = 200 ppm Total Dissolved Solids (TDS). Displays TDS of process.
100% 20.00 mA PID Status screen (if enabled.) Shows the present controller output level
on left, and actual transmitter current on the right. The controller can be
placed in manual while viewing this screen by pressing and holding the
ENTER key for 5 seconds until a small flashing “m” appears on the
screen. At that point the controller output can be adjusted up or down
using the UP and LEFT arrow keys. To return to automatic operation,
press and hold the ENTER key for 5 seconds and the “M” will disappear.
#1 4.00 mA Analyzer output current # 1.
#2 12.00 mA Analyzer output current # 2.
#3 20.00 mA Analyzer output current # 3 (if option included.)
Aux relay=D,E,F Auxilliary relay annunciators (if option included.)
Slope The slope is an indication of the sensor output as compared to the
theoretical output if the sensor were perfect. The slope value is
calculated after completion of a calibration as described in Section 7 of
this manual.
Q46CT v4.00 Transmitter software version number.
Note: A display test (all segments ON) can be actuated by pressing and holding the
ENTER key while viewing the model/version number on the lower line of the
display.
The MEASURE screens are intended to be used as a very quick means of looking up critical
values during operation or troubleshooting.
ATI Q46CT Conductivity System Part 5 – Configuration
31
O&M Manual
Rev-G (8/17)
5.23 Calibration Menu [CAL]
The calibration menu contains items for frequent calibration of user parameters. There are five
items in this list: Cal Cond, Cell Temp, Cal Const, and Cal Zeros.
Cal Cond The conductivity calibration function allows the user to adjust the
ttransmitter offset and span reading to match reference buffers, or to
adjust the sensor offset to match the sample reading. See Part 7 -
Calibration for more details.
Cal Temp The temperature calibration function allows the user to adjust the offset
of the temperature response by a small factor of ± 5°C. The temperature
input is factory calibrated to very high accuracy. However, long cable
lengths and junction boxes may degrade the accuracy of the temperature
measurement in some extreme situations. Therefore, this feature is
provided as an adjustment. See Part 7 - Calibration for more details.
Cal Zeros This function calibrates all range zero-points to the specific sensor being
used. This function is only required to be performed once at initial start-
up or when the sensor has been replaced. See Part 7 - Calibration for
more details.
Cell Factor This function allows the user to directly enter the factory measured cell
constant for the Q25CT sensor. When this feature is used, calibration
with reference solutions is not necessary. The default setting is 1.00.
Note that this value is actually calculated from an actual liquid calibration,
so running a conductivity calibration will change this value.
TC Factor This function allows the user to directly enter the factory measured
temperature offset value marked on the sensor label. The default for this
setting is 7.50 and most sensors are fairly close to that number.
Cond Range This function allows the user to set the display range of the transmitter
for a specific application. Once set, all output functions use this display
range to establish configuration settings. The default range is 0-200.0
mS. Press ENTER to initiate user entry mode, and the value will flash.
Use the arrow key to modify the range for the desired range and then
press ENTER.
5.24 Configuration Menu [CONFIG]
The Configuration Menu contains all of the general user settings:
Entry Lock This function allows the user to lock out unauthorized tampering with
instrument settings. All settings may be viewed while the instrument is
locked, but they cannot be modified. The Entry Lock feature is a toggle-
type setting; that is, entering the correct code will lock the transmitter and
entering the correct code again will unlock it. The code is preset at a
fixed value. Press ENTER to initiate user entry mode and the first digit
will flash. Use arrow keys to modify value. See end of manual for the
Q46CT lock/unlock code. Press ENTER to toggle lock setting once
code is correct. Incorrect codes do not change state of lock condition.
ATI Q46CT Conductivity System Part 5 – Configuration
32
O&M Manual
Rev-G (8/17)
Set Delay The delay function sets the amount of damping on the instrument. This
function allows the user to apply a first order time delay function to the
conductivity measurements being made. Both the display and the output
value are affected by the degree of damping. Functions such as
calibration are not affected by this parameter. The calibration routines
contain their own filtering and stability monitoring functions to minimize
the calibration timing. Press ENTER to initiate user entry mode, and the
value will flash. Use the arrow keys to modify value; range is 0.1 to 9.9
minutes. Press ENTER to store the new value.
Contrast This function sets the contrast level for the display. The custom display
is designed with a wide temperature range, Super-Twist Nematic (STN)
fluid. Press ENTER to initiate user entry mode, and the value will flash.
Use arrow keys to modify the value; range is 0 to 8 (0 being lightest).
Press ENTER to update and store the new value.
Main Display This function allows the user to change the measurement in
the primary display area, selecting between conductivity or Auto-Ranging
conductivity.
Press ENTER to initiate user entry mode, and the entire value will flash.
Use the UP arrow key to modify the desired display value. Press
ENTER to store the new value.
Zero Filter This function forces the reading to zero when reading is below the
entered value. For example, if the entered value were 2.1 mS, the
display would read 0.0 mS when the measured value reached 2.0 mS.
This feature is useful in blanking out very small and meaningless zero
variability.
Ref Temp The reference temperature function sets the basis point for the linear
temperature compensation methods. In most cases this setting should
be left at the default of 25.0 °C.
Press ENTER to initiate user entry mode, and the entire value will flash.
Use the arrow keys to modify the desired value; range is 0.0°C to
50.0°C. Press ENTER to update and store the new value. This setting
appears in the Software Menu only if “Temp Mode” is set to Lin.
TC Method This function allows selection of the method used for the temperature
compensation in the instrument. The choices are 1-Lin or 2-Table. The
Lin (linear) method is by far the most common and should be used
unless special measurements are being made.
Under the 2-Table selection, you have the ability to enter your own
custom temperature compensation table. You can enter up to 6
multipliers from your own temperature compensation data if required for
special applications.
NOTE: Do not set the Temp Mode to a value other than Lin unless you are
confident in the temperature data developed for your special application.
ATI Q46CT Conductivity System Part 5 – Configuration
33
O&M Manual
Rev-G (8/17)
Lin T-Comp This function sets the correction slope value for the linear temperature
compensation method and is used when the “Temp Mode” is set to Lin.
Linear compensation is the method recommended for most aqueous
solutions, and the value is typically 2.00 %/°C (25°C reference
temperature) for neutral water. This is the factory default and it provides
the best compensation for most aqueous solutions. Other typical ranges
include:
Acids: 1.0 to 1.6%/°C
Bases: 1.8 to 2.0%/°C
Salts: 2.2 to 3.0%/°C
NOTE: If the temperature units are changed between °C and °F (see
Temp Units in this Section), the default setting for this output will change
between 2.00 %/°C and 1.11%/°F accordingly.
Other compensation slopes for uncommon solutions may be found in
chemical handbooks (such as the CRC). Press ENTER to initiate user
entry mode, and the entire value will flash. Use the arrow keys to modify
the desired value; entry range is 0.000%/°C (no compensation) to
4.000%/°C. Press ENTER to store the new value.
TDS Factor This function sets the linear relationship of the TDS (total dissolved
solids) reading to the conductivity measurement. The actual units for the
slope are in mg/L/μS. The default value is 00.49 mg/L/μS.
Press ENTER to initiate user entry mode, and the entire value will flash.
Use the arrow keys to modify the desired value; range is 00.00 mg/L/μS
to 99.99 mg/L/μS. Press ENTER to update and store the new value.
Com Mode Sets digital communication mode of analyzer. Optional digital
communication card must be plugged into the power supply slot for this
function to work. Press ENTER to initiate user entry mode, and the
entire value will flash. Use the UP arrow key to modify the desired value;
selections include 1-None, 2- P-DP for Profibus DP, 3 – Modbus, 4 –
Ethernet IP. Press ENTER to store the new value.
Com Address Sets bus address for digital communication mode of analyzer. Optional
digital communication card must be plugged into the power supply slot
for this function to work. Press ENTER to initiate user entry mode, and
the entire value will flash. Use the UP arrow key to modify the desired
value. Range is 1-125. Press ENTER to store the value.
Iout#1 Mode This assigns the 4-20 mA output #1 to either S (by selecting 1) or for
PID output (by selecting 2)
Iout#2 Mode This assigns the 4-20 mA output # 2 to Temperature (by selecting 1), S
(by selecting 2) or TDS mg/L (by selecting 3)
*Iout#3 Mode OPTIONAL. This function sets analog output #3 for temperature,
conductivity, or TDS. Press ENTER to initiate user entry mode, and the
entire value will flash. Use the UP arrow key to modify the desired value;
selections include 1-C/F for temperature, 2-uS, or 3-mg/l. Press ENTER
to store the new value.
ATI Q46CT Conductivity System Part 5 – Configuration
34
O&M Manual
Rev-G (8/17)
Relay A Mode Relay A can be used in three different ways: as a setpoint control, as a
fail alarm, or as a HI-LO alarm band. The three settings for Rly A Mode
are CON, FAIL and AL.
The CON setting enables normal control operation for Relay A, with
settings for setpoint, hysteresis, delay and phasing appearing in the
CONFIG menu automatically. See Fig. 12 for further details.
The FAIL setting enables the fail alarm mode for Relay A. Relay A will
then trip on any condition that causes the FAIL icon to be displayed on
the LCD. Using this mode allows the User to send alarm indications to
other remote devices.
The AL setting allows two setpoints to be selected for the same relay,
producing a HI-LO alarm band. In this mode, Relay A will trip inside or
outside of the band, depending upon the Phase selected. See Figure 19
for further details.
Relay B Mode Relay B and C can be used in two ways: as a setpoint control, or as an
Relay C Mode alarm. The two settings for Relay B Mode are CON and FAIL. Relay B
also is set up for 2 “clean modes”, but these modes are generally not
used for toroidal conductivity. They are used to activate automatic
sensor cleaning systems that are generally not required.
The CON setting enables normal setpoint operation for Relay B/C.
Relay B/C then operates identically to Relay A, with settings for setpoint,
hysteresis, delay and phasing appearing in the CONFIG menu
automatically.
The FAIL setting enables the fail alarm mode for Relay B/C. Relay B/C
will then trip on any condition that causes the FAIL icon to be displayed
on the LCD. Note that the Relay C indicator shows up only on the lower
screen of the display next to the temperature reading. This is because
the default setting for relay C is the FAIL setting. Using this mode allows
the User to send alarm indications to other remote devices.
*Relay D Mode OPTIONAL. Relays D, E, and F can be used in two ways: as a setpoint
*Relay E Mode control, or as an alarm. The two settings for Relay B Mode are CON and
*Relay F Mode FAIL.
The CON setting enables normal setpoint operation for Relay B. Relay B
then operates identically to Relay A, with settings for setpoint, hysteresis,
delay and phasing appearing in the CONFIG menu automatically.
Temp Units This function sets the display units for temperature measurement. Press
ENTER to initiate user entry mode, and the entire value will flash. Use
the UP arrow key to modify the desired display value. The choices are
°F and °C. Press ENTER to store the new value.
ATI Q46CT Conductivity System Part 5 – Configuration
35
O&M Manual
Rev-G (8/17)
5.25 Control Menu
The Control Menu contains all of the output control user settings:
Set PID 0% If the PID is enabled, this function sets the minimum and maximum
Set PID 100% controller end points. Unlike the standard 4-20 mA output, the controller
[Iout1=PID] does not “scale” output values across the endpoints. Rather, the
endpoints determine where the controller would normally force minimum
or maximum output in an attempt to recover the setpoint (even though
the controller can achieve 0% or 100% anywhere within the range.)
If the 0% point is lower than the 100% point, then the controller action
will be “reverse” acting. That is, the output of the controller will increase
if the measured value is less than the setpoint, and the output will
decrease if the measured value is larger than the setpoint. Flipping the
stored values in these points will reverse the action of the controller to
“direct” mode.
The entry value is limited to a value within the range specified in “Set
Range”, and the 0% and the 100% point must be separated by at least
1% of this range Use the LEFT arrow key to select the first digit to be
modified. Then use the UP and LEFT arrow keys to select the desired
numerical value. Press ENTER to store the new value.
PID Setpnt The measured value which the controller is attempting to maintain by
[Iout1=PID] adjusting output value. It is the nature of the PID controller that it never
actually gets to the exact value and stops. The controller is continually
making smaller and smaller adjustments as the measured value gets
near the setpoint.
PID Prop Proportional gain factor. The proportional gain value is a multiplier on
[Iout1=PID] the controller error (difference between measured value and setpoint
value.) Increasing this value will make the controller more responsive.
PID Int Integral is the number of “repeats-per-minute” of the action of the
[Iout1=PID] controller. It is the number of times per minute that the controller acts on
the input error. At a setting of 2.0 rpm, there are two repeats every
minute. If the integral is set to zero, a fixed offset value is added to the
controller (manual reset.) Increasing this value will make the controller
more responsive.
PID Deriv Derivative is a second order implementation of Integral, used to suppress
[Iout1=PID] “second-order” effects from process variables. These variables may
include items like pumps or mixers that may have minor impacts on the
measured value. The derivative factor is rarely used and therefore it is
best in most cases to leave it at the default value. Increasing this value
will make the controller more responsive.
Set 4 mA (#1) These functions set the main 4 and 20 mA current loop #1 output points
Set 20 mA (#1) for the analyzer when output 1 is in normal, non-PID, mode of operation.
[Iout1=conductivity] The value stored for the 4 mA point may be higher or lower than the
value stored for the 20 mA point. The entry values are limited to values
within the range specified in “Set Range”, and the 4 mA and the 20 mA
point must be separated by at least 1% of this range Use the LEFT arrow
ATI Q46CT Conductivity System Part 5 – Configuration
36
O&M Manual
Rev-G (8/17)
key to select the first digit to be modified. Then use the UP and LEFT
arrow keys to select the desired numerical value. Press ENTER to store
the new value..
*Set 4 mA (#2) These functions set the second 4 mA and 20 mA current loop output
*Set 20 mA (#2) points for the transmitter. The output may be set to track temperature
[temp/conductivity] (default), or conductivity. The values stored for the 4 mA point may be
higher or lower than the value stored for the 20 mA point.
The entry value is limited to a value between 0 and 110°C if it is set for
temperature, within the range specified in “Set Range” if the output is set
to track conductivity. The 4 mA and the 20 mA point must be at least 20
units away from each other. Press ENTER to initiate user entry mode,
and the value will flash. Use arrow keys to modify value. Press ENTER
to store the new value.
*Set 4 mA #3 OPTIONAL These functions set the optional third 4 mA and 20 mA
*Set 20 mA #3 current loop output points for the analyzer. The output may be set to
[temp/uS/TDS] track temperature (default), conductivity, or TDS. The values stored for
the 4 mA point may be higher or lower than the value stored for the 20
mA point.
The range is between 0 and 110°C if it is set for temperature, within the
range specified in “Set Range” if the output is set to track conductivity,
and must be within the display range established by the conductivity
range if set to track TDS. The 4 mA and the 20 mA point must be at
least 20 units away from each other. Press ENTER to initiate user entry
mode, and the value will flash. Use arrow keys to modify value. Press
ENTER to store the new value.
*A Setpoint This function establishes the conductivity trip point for relay A. The entry
value is limited to a value within the range specified in “Set Range”. Use
the LEFT arrow key to select the first digit to be modified. Then use the
UP and LEFT arrow keys to select the desired numerical value. Press
ENTER to store the new value.
*A Hysteresis This function establishes the hysteresis, or “deadband”, for Relay A.
Hysteresis is most often used to control relay chattering; however, it may
also be used in control schemes to separate the ON/OFF trip points of
the relay. Press ENTER to initiate user entry mode, and the value will
flash. Use the arrow keys to modify value. Press ENTER to store the
new value.
*A Delay This function places an additional amount of time delay on the trip point
for relay A. This delay is in addition to the main delay setting for the
controller. The entry value is limited to a value between 0 and 999
seconds. Press ENTER to initiate user entry mode, and the value will
flash. Use arrow keys to modify value; range is 0 to 999 seconds. Press
ENTER to store the new value.
*A Phasing This function establishes the direction of the relay trip. When phase is
HI, the relay operates in a direct mode. Therefore, the relay energizes
and the LCD indicator illuminates when the conductivity value exceeds
the setpoint. When the phase is LO, the relay energizes and the LCD
indicator illuminates when the conductivity level drops below the
setpoint. The failsafe setting does have an impact on this logic. The
ATI Q46CT Conductivity System Part 5 – Configuration
37
O&M Manual
Rev-G (8/17)
description here assumes the failsafe setting is OFF. Press ENTER to
initiate user entry mode, and the entire value will flash. Use the UP
arrow key to modify the desired value; selections include HI for direct
operation or LO for reverse operation. Press ENTER to store the new
value. See Figure 20 below for a visual description of a typical control
relay application.
Figure 20 - Control Relay Example
If Relay A Mode is set to Alarm Mode, AL, then the following settings will
appear in the Config Menu list automatically. In this mode, two setpoints
can be selected on the same relay, to create an alarm band. Phase HI
selection causes the relay to energize outside of the band, and Phase
LO causes the relay to energize inside of the band. This feature enables
one relay to be used as a control relay while the other is used as a HI-LO
Alarm relay at the same time. Setpoint A-LO must be set lower than
Setpoint A-HI. When AL mode is first selected, Setpoint A-LO is
defaulted to 0. Figure 21 is a visual description of a typical alarm relay
application.
Figure 21 - Alarm Relay Example
When value rises to ≥ 1.000 mS, relay closes. When value falls to ≤ 1.000 mS, relay opens.
When value rises to ≥ 1.000 mS, relay closes,
until value falls back to ≤ 0.950 mS.
Settings:
When value falls to ≤ 1.000 µS, relay closes,
until value rises back to ≥ 1.050 µS.
When value rises to ≥ 0.500 µS, relay closes,
until value falls back to ≤ 0.450 µS.
When value falls to ≤ 0.500 mS, relay closes, until
value rises back to ≥ 0.550 mS.
*Setpnt A-HI
*Hyst A-HI
*Delay A-HI
*Setpnt A-LO
*Hyst A-LO
*Delay A-LO
When value rises to ≥ 1.000 mS, relay opens. When value falls to ≤ 1.000 mS, relay closes.
}
1.000 mS
0.950 mS
PHASE: HI
ON
HYSTERESIS
OR
“DEAD BAND”
X
OFF
}
1.050 mS
1.000 mS
PHASE: LO
OFF
HYSTERESIS
OR
“DEAD BAND”
X
ON
}
1.000 mS
0.950 mS
PHASE: HI
HYST-HI
X
ON
0.550 mS
0.500 mS
OFF
X
}HYST-LO
ON
}
1.050 mS
1.000 mS
PHASE: LO
HYST-HI
X
OFF
0.500 mS
0.450 mS
ON
X
}HYST-LO
OFF
Settings: Setpoint: 1.000 mS Hyst: 0.050
Delay: 000 Failsafe: OFF
Setpoint A-HI: 1.000 mS Setpoint A-LO: 0.500 mS
Hyst A-HI: 0.050 Hyst A-LO: 0.05
Delay A-HI: 000 Delay A-LO: 000
ATI Q46CT Conductivity System Part 5 – Configuration
38
O&M Manual
Rev-G (8/17)
If Relay B Mode is set to CON (see Relay B Mode), then Relay B will
function identically to Relay A. Relay B settings appear in the CONFIG
menu list automatically.
If Relay C Mode is set to CON (see Relay C Mode), then Relay C will
function identically to Relay A. Relay C settings appear in the CONFIG
menu list automatically.
If optional Relay D, E, or F Mode is set to CON (see Relay D,E,F
Modes), then the Relay will function identically to Relay A. Relay
settings appear in the CONFIG menu list automatically.
5.26 Diagnostics Menu [DIAG]
The diagnostics menu contains all of the user settings that are specific to the system diagnostic
functions, as well as functions that aid in troubleshooting application problems.
Set Hold The Set Hold function locks the current loop output values on the present
process value, and halts operation of the PID controller. This function
can be used prior to calibration, or when removing the sensor from the
process, to hold the output in a known state. Once HOLD is released,
the outputs return to their normal state of following the process input.
The transfer out of HOLD is bumpless on the both analog outputs - that
is, the transfer occurs in a smooth manner rather than as an abrupt
change. An icon on the display indicates the HOLD state, and the HOLD
state is retained even if power is cycled. Press ENTER to initiate user
entry mode, and entire value will flash. Use the UP arrow key to modify
the desired value, selections are ON for engaging the HOLD function,
and OFF to disengage the function. Press ENTER to store the new
value. Note - the Set Hold function holds BOTH current levels, as well
as ALL relay settings.
The Set Hold function can also hold at an output value specified by the
user. To customize the hold value, first turn the HOLD function on.
Press the ESC key to go to the DIAG Menu and scroll to Sim Output
using the UP arrow key. Press ENTER. Follow the instructions under
Sim Output (see following page).
CAUTION: There is no time-out on the hold feature. Once placed
into hold mode, return to normal operation must be done manually.
Fault List The Fault List screen is a read-only screen that allows the user to display
the cause of the highest priority failure. The screen indicates the number
of faults present in the system and a message detailing the highest
priority fault present. Note that some faults can result in multiple
displayed failures due to the high number of internal tests occurring. As
faults are corrected, they are immediately cleared.
*B Setpoint
*B Hysteresis
*B Delay
*B Phasing
C Setpoint
C Hysteresis
C Delay
C Phasing
D,E,F Setpoint
D,E,FHyster
D,E,F Delay
D,E,F Phasing
ATI Q46CT Conductivity System Part 5 – Configuration
39
O&M Manual
Rev-G (8/17)
Faults are not stored; therefore, they are immediately removed if power
is cycled. If the problem causing the faults still exists, however, faults will
be displayed again after power is re-applied and a period of time elapses
during which the diagnostic system re-detects them. The exception
to this rule is the calibration failure. When a calibration fails, no corrupt
data is stored. Therefore, the system continues to function normally on
the data that was present before the calibration was attempted.
After 30 minutes or if power to the analyzer is cycled, the failure for
calibration will be cleared until calibration is attempted again. If the
problem still exists, the calibration failure will re-occur. Press ENTER to
initiate view of the highest priority failure. The display will automatically
return to normal after a few seconds.
PID Timer This function sets a timer to monitor the amount of time the PID
controller remains at 0% or 100%. This function only appears if the PID
controller is enabled. If the timer is set to 0000, the feature is effectively
disabled. If the timer value is set to any number other zero, a FAIL
condition will occur if the PID controller remains at 0% or 100% for the
timer value. If one of the relays is set to FAIL mode, this failure condition
can be signaled by a changing relay contact.
Press ENTER to initiate user entry mode, and the entire value will flash.
Use the UP arrow key to modify desired value; range of value is 0-9999
seconds. Press ENTER to store the new value.
Sim Out The Sim Out function allows the user to simulate the chlorine level of the
instrument in the user selected display range. The user enters a ppm
value directly onto the screen, and the output responds as if it were
actually receiving the signal from the sensor. This allows the user to
check the function of attached monitoring equipment during set-up or
troubleshooting. Escaping this screen returns the unit to normal
operation. Press ENTER to initiate the user entry mode, and the right-
most digit of the value will flash. Use arrow keys to modify desired value.
The starting display value in SIM mode will be the last read value of the
input. The output will be under control of the SIM screen until the ESC
key is pressed. The instrument will automatically terminate the simulated
output after 30 minutes and return to normal operation unless the
“HOLD” function is engaged.
Note: If the HOLD function is engaged before the Sim Output function is
engaged, the simulated output will remain the same even when the ESC
key is pressed. Disengage the HOLD function to return to normal output.
Fail Out #1 This function enables the user to define a specified value that the main
current output will go to under fault conditions. When enabled to ON, the
output may be forced to the current value set in Fail Val (next item.) With
the Fail Out setting of ON, and a Fail Val setting of 6.5 mA, any alarm
condition will cause the current loop output to drop outside the normal
operating range to exactly 6.5 mA, indicating a system failure that
requires attention.
Press ENTER to initiate user entry mode, and the entire value will flash.
Use the UP arrow key to modify desired value; selections are ON, OFF.
Press ENTER to store the new value.
ATI Q46CT Conductivity System Part 5 – Configuration
40
O&M Manual
Rev-G (8/17)
Fail Val #1 Sets the output failure value for Iout#1. When Fail Out above is set to
ON, this function sets value of the current loop under a FAIL condition.
The output may be forced to any current value between 4-20 mA.
Press ENTER to initiate user entry mode, and the entire value will flash.
Use the UP arrow key to modify desired value; selections are between
4mA, and 20mA. Press ENTER to store the new value.
Fail Val #2 This function sets the value of current loop output #2 under a FAIL
condition. The settings and operation are identical to Fail Out for output
#1.
*Fail Out #3 OPTIONAL. This function sets the fail-mode of current loop output #3
under a FAIL condition. The settings and operation are identical to Fail
Out for output #1.
*Fail Val #3 OPTIONAL. This function sets the value of current loop output #3 under
a FAIL condition. The settings and operation are identical to Fail Out for
output #1.
Backlight This function has three options. ON – On all the time, OFF – Off all the
time, AL – Alarm (Default). This function flashes the backlight on and off
whenever the Fail icon is displayed.
Failsafe This function allows the user to set the optional system relays to a
failsafe condition. In a failsafe condition, the relay logic is reversed so
that the relay is electrically energized in a normal operating state. By
doing this, the relay will not only change state when, for example, a
chlorine limit is exceeded, but also when power is lost to the controller.
When failsafe is selected to be ON, the normally-open contacts of the
relay will be closed during normal operation. In an attempt to make this
configuration less confusing, the LCD icon logic is reversed with this
setting, and the icon is OFF under this normal condition. Therefore,
when the trip condition occurs, the closed N.O. contacts will be opened
(relay de-energized), and the LCD icon will illuminate. In addition, a
power fail would also cause the contacts to open.
Start Delay This function is designed to minimize control or alarm issues arising from
temporary power loss. When power goes down, the monitor records the
analog output values and the status of relays and PID functions. When
power is restored, the analog values and relays will be held at the pre-
power loss values for a defined period of time. This “start delay” may be
programmed for periods from 0-9.9 minutes. This function is set to 0.0
minutes by default and must be activated by the user if desired by setting
a positive time value
Set Default The Set Default function allows the user to return the instrument back to
factory default data for all user settings or for just the calibration default.
It is intended to be used as a last resort troubleshooting procedure. All
user settings or the calibration settings are returned to the original factory
values. Hidden factory calibration data remains unchanged. Press
ENTER to initiate user entry mode and select either All or CAL with the
UP arrow key.
41
O&M Manual
Rev-G (8/17)
Part 6 – Calibration
6.1 Overview and Methods
Calibration of the Q46CT is required to accurately match the sensor characteristics to the
monitor/analyzer. Since the output of the conductivity sensor does not degrade over time, it is
typically only required that the sensor be calibrated at initial installation and then cleaned
periodically to maintain proper system accuracy.
It is important for the user to establish a periodic cleaning and calibration-check schedule for
sensor maintenance to maintain high system accuracy. Since the conductivity of a solution is
greatly affected by temperature, proper settings for thermal compensation are critical for accurate
operation. Before calibrating the instrument for the very first time, it is important to select the
proper operating parameters in the configuration menus for temperature compensation methods.
When using conductivity calibration standards for a wet calibration, take care not to inadvertently
contaminate the reference solution; always thoroughly clean the sensor, rinsing off in tap water,
and then finish rinsing in pure or de-ionized water. In addition, note that calibration solutions less
than 200 μS or greater than 100 mS can be very unstable. Moving the sensor back and forth
between different value conductivity reference solutions can quickly contaminate the solutions
and render them inaccurate.
6.11 1-Point Cal Explained
The 1-point calibration method is generally known as the "grab sample" calibration method. In
the 1-point calibration method, the sensor may be removed from the application and placed into a
reference solution. It may also be left in the measurement process and calibrated by reference.
The 1-point calibration adjusts the sensor slope to match the exact calibration point. Readings
beyond that point are then extrapolated from the determined slope of the calibration line. Since
the sensor slope does not degrade over time, frequent re-calibration is unnecessary. Calibration
accuracy can be optimized by calibrating with a reference solution which is close to the values
typically measured.
6.12 Zero Cal Cal Explained
The sensor offset must be set for the system only on initial sensor installation, or when the cable
length has been altered. The Zero Cal method establishes all of the sensor offset points for the
instrument’s 6 ranges of operation.
6.2 Performing Sensor Zero Cal
The sensor offset MUST be set for the system on initial sensor installation, or when the cable
length has been altered. However, it can easily be adjusted at any time by re-calibrating the
sensor in air. The sensor zero-calibration generally has little effect in measurements above about
50 mS, but it can have a significant effect in measurements below about 1 mS. If the sensor zero
cal is to be performed, it must be done BEFORE the 1-point reference calibration.
To begin the sensor zero cal, verify that the sensor is connected and clean and dry. It should be
placed in the air with the electrodes at least 1 foot away from any nearby objects. Holding it is not
recommended – place on table or just hang.
ATI Q46CT Conductivity System Part 6 – Calibration
42
O&M Manual
Rev-G (8/17)
Procedure
1. Remove sensor from process and clean thoroughly. Dry sensor and position on table or
hang in air (in air is best.) If on table, let end of sensor hang over edge of table.
2. Scroll to the CAL menu section using the MENU key and press ENTER or the UP arrow key.
Scroll to the menu Zero Cal.
3. Press the ENTER key. The screen will prompt the user to postion the sensor in air.
4. Press the ENTER key. The screen will automatically scroll through all ranges and establish
and store the proper zero points.
6.3 Performing 1-Point Cal
This calibration method is intended to be used as an on-line calibration method or a wet-cal with
reference solutions. During calibration, the system will display the current conductivity reading,
and the user can manually enter a reference value from a reference solution bottle or a
comparative reference instrument.
For wet calibrations, the user may use pre-made calibration references (also available from ATI)
or a NaCl solution may be made using pure, dried NaCl crystals and one liter of high purity, de-
ionized, CO2-free water as mixed in the table shown in Figure 22 - NaCl Reference Solution
Calibration Chart
All table data is at 25°C - therefore, the sensor must be at this temperature to calibrate properly
using the table data. If another reference calibration solution is being used, be sure to note
temperature of reference solution before calibration. Since the sensor must ideally be at the
specified temperature, wet calibrations can be difficult to perform accurately.
NaCl Reference Solution for
Calibration (25°C)
μS/cm NaCl (gm)
100 0.05
200 0.10
500 0.25
1000 0.50
2000 1.01
3000 1.53
4000 2.06
5000 2.61
8000 4.34
10000 5.56
20000 11.59
Figure 22 - NaCl Reference Solution Calibration Chart
ATI Q46CT Conductivity System Part 6 – Calibration
43
O&M Manual
Rev-G (8/17)
During the 1-point calibration, the system will automatically pick the correct range for the
calibration reference if the Q46CT is in the AUTO range. If the Q46CT is in a normal display
mode, the user must be careful to calibrate with a solution that falls into the manual range
selected. If the calibration solution is outside the manual range, an error will result.
Procedure
1. If a zero calibration on the sensor is also to be performed, that must be done FIRST. The zero
calibration process can have an impact on the result of the 1-point calibration. So if a zero cal is
required, do that procedure and return here.
2. Determine whether the calibration will be done on-line or with the sensor removed and placed into
a reference solution. If the sensor is removed from the application, rinse and clean.
3. If the sensor has been removed and placed into a solution, allow the sensor to temperature
equilibrate with the solution as much as possible. With the sensor coming from an application
that differs greatly in temperature, the user may have to wait as much as 20 minutes. If the
sensor is on-line, the user may want to set the output HOLD feature prior to calibration to lock out
any output fluctuations.
4. Scroll to the CAL menu section using the MENU key and press ENTER or the UP arrow key.
Scroll until Cal Cond is displayed. Press ENTER.
5. The screen will prompt the user to place the sensor into the reference solution (ideally this has
already been done to achieve temperature equilibrium.) Once sensor is ready, press ENTER.
6. The system now begins acquiring data for the calibration value. As data is gathered, the units for
conductivity and temperature may flash. Flashing units indicate that this parameter is unstable.
The calibration data point acquisition will stop only when the data remains stable for a pre-
determined amount of time. This can be overridden by pressing ENTER. If the data remains
unstable for 10 minutes, the calibration will fail and the message Cal Unstable will be displayed.
7. The screen will display the last measured conductivity value and a message will be displayed
prompting the user for the reference value. The user must then modify the screen value with the
arrow keys and press ENTER. The system then performs the proper checks.
8. If accepted, the screen will display the message PASS with the slope value, then it will return to
the main measurement display. If the calibration fails, a message indicating the cause of the
failure will be displayed and the FAIL icon will be turned on.
6.4 Temperature Calibration
The temperature input is factory calibrated for the highest accuracy. Temperature calibration is
not recommended; however, it is provided for applications in which very long cable lengths are
needed. For example, at 50 feet, readings may be off ± 0.2°C. The temperature calibration
sequence is essentially a 1-point offset calibration that allows adjustments of approximately
± 5°C.
The sensor temperature may be calibrated on line, or the sensor can be removed from the
process and placed into a known solution temperature reference. In any case, it is critical that the
sensor be allowed to reach temperature equilibrium with the solution in order to provide the
highest accuracy. When moving the sensor between widely different temperature conditions, it
may be necessary to allow as much as one hour before the calibration sequence is initiated. If
the sensor is on-line, the user may want to set the output HOLD (see Section 5.26) feature prior
to calibration to lock out any output fluctuations.
ATI Q46CT Conductivity System Part 6 – Calibration
44
O&M Manual
Rev-G (8/17)
Procedure
1. Scroll to the CAL menu section using the MENU key and press ENTER or the UP arrow key.
2. Press the UP arrow key until Cal Temp is displayed.
3. Press the ENTER key. The message Place sensor in solution then press ENTER will be
displayed. Move the sensor into the calibration reference (if it hasn’t been moved already)
and wait for temperature equilibrium to be achieved. Press ENTER to begin the calibration
sequence.
4. The calibration data gathering process will begin. The message Wait will flash as data is
accumulated and analyzed. The °C or °F symbol may flash periodically if the reading is too
unstable.
5. The message Adjust temp value then press ENTER will be displayed, and the right-most
digit will begin to flash, indicating that the value can be modified. Using the UP and LEFT
arrow keys, modify the value to the known ref solution temperature. Adjustments up to ± 5 °C
from the factory calibrated temperature are allowed. Press ENTER.
6. Once completed, the display will indicate PASS or FAIL. If the unit fails, the temperature
adjustment may be out of range, the sensor may not have achieved complete temperature
equilibrium, or there may be a problem with the temperature element. In the event of
calibration failure, it is recommended to attempt the calibration again immediately.
45
O&M Manual
Rev-G (8/17)
Part 7 – PID Controller Details
7.1 PID Description
PID control, like many other control schemes, are used in chemical control to improve the
efficiency of chemical addition or control. By properly tuning the control loop that controls
chemical addition, only the amount of chemical that is truly required is added to the system,
saving money. The savings can be substantial when compared to a system which may be simply
adding chemical at a constant rate to maintain some minimal addition under even the worst case
conditions. The PID output controller is highly advantageous over simple control schemes that
just utilize direct (proportional only) 4-20 mA output connections for control, since the PID
controller can automatically adjust the “rate” of recovery based on the error between the setpoint
and the measured value – which can be a substantial efficiency improvement..
The PID controller is basically designed to provide a “servo” action on the 4-20 mA output to
control a process. If the user requires that a measured process stay as close as possible to a
specific setpoint value, the controller output will change from 0% to 100% in an effort to keep the
process at the setpoint. To affect this control, the controller must be used with properly selected
control elements (valves, proper chemicals, etc.) that enable the controller to add or subtract
chemical rapidly enough. This is not only specific to pumps and valves, but also to line sizes,
delays in the system, etc.
This section is included to give a brief description of tuning details for the PID controller, and is
not intended to be an exhaustive analysis of the complexities of PID loop tuning. Numerous
sources are available for specialized methods of tuning that are appropriate for a specific
application.
7.2 PID Algorithm
As most users of PID controllers realize, the terminology for the actual algorithm terms and even
the algorithms themselves can vary between different manufacturers. This is important to
recognize as early as possible, since just plugging in similar values from one controller into
another can result in dramatically different results. There are various basic forms of PID
algorithms that are commonly seen, and the implementation here is the most common version;
The ISA algorithm (commonly referred to as the “ideal” algorithm.)








  dt
t
de
D
t
d
t
e
I
t
e
P
output
)
(
)
(
)
(
1
)
(
Where:
output = controller output
P = proportional gain
I = integral gain
D = derivative gain
t = time
e(t) = controller error (e=measured variable – setpoint)
Figure 23 - PID Equation
ATI Q46CT Conductivity System Part 7 – PID Controller Details
46
O&M Manual
Rev-G (8/17)
The most notable feature of the algorithm is the fact the proportional gain term affects all
components directly (unlike some other algorithms - like the “series” form.) If a pre-existing
controller utilizes the same form of the algorithm shown above, it is likely similar settings can for
made if the units on the settings are exactly the same. Be careful of this, as many times the units
are the reciprocals of each other (i.e. reps-per-min, sec-per-rep.)
PID stands for “proportional, integral, derivative.” These terms describe the three elements of the
complete controller action, and each contributes a specific reaction in the control process. The
PID controller is designed to be primarily used in a “closed-loop” control scheme, where the
output of the controller directly affects the input through some control device, such as a pump,
valve, etc.
Although the three components of the PID are described in the setting area (section 4.25), here
are more general descriptions of what each of the PID elements contribute to the overall action of
the controller.
P Proportional gain. With no “I” or “D” contribution, the controller output is simply a factor of
the proportional gain multiplied by the input error (difference between the measured input
and the controller setpoint.) Because a typical chemical control loop cannot react
instantaneously to a correction signal, proportional gain is typically not efficient by itself –
it must be combined with some integral action to be useful. Set the P term to a number
between 2-4 to start. Higher numbers will cause the controller action to be quicker.
I Integral gain. Integral gain is what allows the controller to eventually drive the input error
to zero – providing accuracy to the control loop. It must be used to affect the accuracy in
the servo action of the controller. Like proportional gain, increasing integral gain results
in the control action happening quicker. Set the I term to a number between 3-5 to start
(1-2 more than P). Like proportional gain, increasing the integral term will cause the
controller action to be quicker.
D Derivative gain. The addition of derivative control can be problematic in many
applications, because it greatly contributes to oscillatory behavior. In inherently slow
chemical control process’, differential control is generally added in very small amounts to
suppress erratic actions in the process that are non-continuous, such as pumps and
valves clicking on and off. However, as a starting point for chemical process control, its
best to leave the “D” term set to 0.
Based on these descriptions, the focus on tuning for chemical applications really only involves
adjustment of “P” and “I” in most cases. However, increasing both increases the response of the
controller. The difference is in the time of recovery. Although combinations of high “P’s” and low
“I” will appear to operate the same as combinations of low “P’s” and high “I’s”, there will be a
difference in rate of recovery and stability. Because of the way the algorithm is structured, large
“P’s” can have a larger impact to instability, because the proportional gain term impacts all the
other terms directly. Therefore, keep proportional gain lower to start and increase integral gain to
achieve the effect required.
Many of the classical tuning techniques have the user start with all values at 0, and then increase
the P term until oscillations occur. The P value is then reduced to ½ of the oscillatory value, and
the I term is increased to give the desired response. This can be done with the Q46CT controller,
with the exception that the I term should start no lower than 1.0.
If it appears that even large amounts of integral gain (>20) don’t appreciably increase the desired
response, drop I back to about 1.0, and increase P by 1.00, and start increasing I again. In most
chemical control schemes, I will be approximately 3 times the value of P.
ATI Q46CT Conductivity System Part 7 – PID Controller Details
47
O&M Manual
Rev-G (8/17)
7.3 Classical PID Tuning
Unlike many high speed position applications where PID loops are commonly used, the chemical
feed application employed by this instrument does not require intense mathematical exercise to
determine tuning parameters for the PID. In fact, the risk of instability is far greater with overly
tuned PID control schemes. In addition, many of the classical mathematical exercises can be
damaging or wasteful in the use of chemicals when the process is bumped with large amounts of
input error to seek a response curve. Because of this, the general adjustment guidelines
described in section 7.2 are sufficient for almost all application tuning for this instrument. Beyond
this, many sources are available for classical tuning methods.
7.4 Manual PID Override Control
The Q46 PID output function allows the user to take manual control of the PID control signal.
This is often useful when starting up a control loop, or in the event that you wish to bump the
system manually to measure system response time.
To access the manual PID control, you must be in the MEASURE mode of operation and you
must have the PID output displayed on the lower line. This line will indicate “XX.X% XX.X mA”
with the X values simply indicating the current values. With this display on the screen, press and
hold the ENTER key for about 5 seconds. You will see a small “m” show up between the % value
and the mA value. This indicates you are now in manual mode.
Once in manual, you may increase the PID output by pressing the UP arrow or you may decrease
the output by pressing the LEFT arrow. This will allow you to drive the PID output to any desired
setting.
To revert to normal PID control, press and hold the ENTER key again until the “m” indicator
disappears.
7.5 Common PID Pitfalls
The most common problem occurring in PID control applications involves the false belief that
proper settings on only the PID controller can balance any process to an efficient level.
Close-loop control can only be effective if all elements in the loop are properly selected for the
application, and the process behavior is properly understood. Luckily, the nature of simple
chemical control process’ are generally slow in nature. Therefore, even a de-tuned controller
(one that responds somewhat slowly) can still provide substantial improvements to setpoint
control. In fact, damaging oscillatory behavior is far more likely in tightly tuned controllers where
the user attempted to increase response too much.
When deciding on a PID control scheme, it is important to initially review all elements of the
process. Sticking valves, undersized pumps, or delays in reaction times associated with chemical
addition can have a dramatic effect on the stability of the control loop. When controlling a
chemical mix or reaction, the sensor should be placed in a location that ensures proper mixing or
reaction time has occurred.
The easiest processes to control with closed-loop schemes are generally linear, and symmetrical,
in nature. For example, controlling level in tank where the opening of valve for a fixed period of
time corresponds linearly to the amount that flows into a tank. Chemical control process’ can be
more problematic when the nature of the setpoint value is non-linear relative to the input of
chemical added. For example, pH control of a process may appear linear only in a certain range
of operation, and become highly exponential at the extreme ranges of the measuring scale. In
addition, if a chemical process is not symmetrical, that means it responds differentially to the
ATI Q46CT Conductivity System Part 7 – PID Controller Details
48
O&M Manual
Rev-G (8/17)
addition and subtraction of chemical. It is important in these applications to study steady-state
impact as well as step-change impact to process changes. In other words, once the process has
apparently been tuned under normal operating conditions, the user should attempt to force a
dramatic change to the input to study how the output reacts. If this is difficult to do with the actual
process input (the recommended method), the user can place the control in manual at an
extreme control point such as 5% or 95%, and release it in manual. The recovery should not be
overly oscillatory. If so, the loop needs to be de-tuned to deal with that condition (reduce P
and/or I.)
49
O&M Manual
Rev-G (8/17)
Part 8 – System Maintenance
8.1 System Checks
1. If the FAIL icon is flashing on the display, check the Fault List to determine the cause of the
failure. To access the Fault List, press the MENU/ESC key until the DIAG menu appears.
Then press the UP arrow key until the Fault List appears. Press the ENTER key to access
the Fault List, and the highest priority fault message will be displayed. For a list of all
messages and possible causes/solutions, refer to Section 9.5.
2. In ALL environments, connect an earth ground jumper to earth terminal connection on
transmitter.
3. Perform a one-point calibration prior to sensor installation.
4. Check sensor cable color to terminal strip markings.
5. For highly unstable behavior, remove sensor from the process and measure the process
solution in a plastic beaker. If the reading now stabilizes, place wire in beaker solution and
actual process solution to determine if a ground loop exists.
6. If the instrument 4-20 mA output is connected into other control systems, disconnect output
loop from system load and run through a handheld DMM to monitor current. Verify that the
system operates correctly in this mode first.
8.2 Instrument Checks
1. Remove sensor completely and connect 1100 Ohms from the GREEN to BLACK (green
bundle) on the analyzer input leads. The temperature reading should display approximately
25 °C and the conductivity reading should display approximately 0.0 uS.
2. Reconnect the sensor and leave dry in air. With a DMM, measure the AC voltage between
RED and BLACK sensor leads. The DMM should read between 100 mVrms and 300 mVrms
@ about 10 kHz. The display should show some value close to 0 uS if the sensor has been
properly zero calibrated.
8.3 Sensor Tests
Toroidal sensors can be tested with a digital voltmeter (DVM) to determine if a major sensor
problem exists. Follow the steps below to verify sensor integrity:
A. Disconnect the nine sensor wires from the terminal strip on the transmitter. Check
sensor plastic covering for any mechanical damage to plastic covering.
B. Connect a DVM between the RED and BLACK wires in the red jacket pair. With the DVM
set to measure resistance, you should measure between 0.4 and 2.0 Ohms.
C. Connect a DVM between the WHITE and BLACK wires in the white jacket pair. With the
DVM set to measure resistance, you should measure between 0.4 and 2.0 Ohms.
ATI Q46CT Conductivity System Part 8 – System Maintenance
50
O&M Manual
Rev-G (8/17)
D. Connect a DVM between the WHITE wire from the white jacket pair, and the RED wire
from the red jacket pair. With the DVM set to measure resistance, you should measure
an open circuit.
E. Connect a DVM between the GREEN and BLACK wires in the green jacket pair. These
are the RTD leads, and you should find a resistance value that depends on the
temperature. The table below lists the resistance values for various temperatures.
Temperature °C Resistance 
0 1000
5 1019
10 1039
15 1058
20 1078
25 1097
30 1117
35 1136
40 1155
45 1175
50 1194
Figure 24 - Pt100 RTD Resistance Table
8.4 Display Messages
The Q46 Series instruments provide a number of diagnostic messages that indicate problems
during normal operation and calibration. These messages appear as prompts on the secondary
line of the display or as items on the Fault List .
The following messages will appear as prompts:
MESSAGE DESCRIPTION POSSIBLE CORRECTION
Cal Unstable Calibration problem, data too
unstable to calibrate.
Clean sensor, get fresh cal solutions, allow
temperature and conductivity readings to fully
stabilize, do not handle sensor or cable during
calibration.
Slope HIGH Sensor slope from calibration is
greater than 400%.
Get fresh cal solutions, allow temperature and
conductivity readings to fully stabilize, check for
correct buffer values
Slope LOW Sensor slope from calibration is
less than 20%.
Clean sensor, get fresh cal solutions, allow
temperature and conductivity readings to fully
stabilize, check for correct buffer values.
Offset HIGH Sensor offset from calibration is too
high
Clean sensor, get fresh cal solutions, allow
temperature and conductivity readings to fully
stabilize, check for correct solution values.
ATI Q46CT Conductivity System Part 8 – System Maintenance
51
O&M Manual
Rev-G (8/17)
Out of Range Input value is outside selected
range of the specific list item being
configured.
Check manual for limits of the function to be
configured.
Locked! Transmitter security setting is
locked.
Enter security code to allow modifications to
settings.
Unlocked! Transmitter security has just been
unlocked.
Displayed just after security code has been
entered.
TC-F25 lock! The TC won’t calibrate because
there is something wrong with the
connection to the temperature
element in the sensor.
Perform Sensor Tests and Instrument Tests to
confirm operation of TC in sensor. Check all
connections between sensor and instrument.
The following messages will appear as items on the Fault List:
MESSAGE DESCRIPTION POSSIBLE CORRECTION
Sensor High The raw signal from the sensor is too
high.
Check wiring connections to sensor.
Sensor Low The raw signal from the sensor is too
low.
Check wiring connections to sensor.
Cond too High The conductivity reading is > 2000 mS. The conductivity reading is over operating
range limits. Move to a higher range or
select “Auto” from Main Display in
CONFIG menu.
Temp High The temperature reading is > 210 ºC. The temperature reading is over operating
limits. Check wiring and expected temp
level. Perform RTD test as described in
sensor manual. Recalibrate sensor
temperature element if necessary.
Temp Low The temperature reading is < -10 ºC The temperature reading is under
operating limits. Check wiring and
expected temp level. Perform RTD test as
described in sensor manual. Recalibrate
sensor temperature element if necessary.
TC Error TC may be open or shorted. Check sensor wiring and perform RTD test
as described in sensor manual.
Cond Cal Fail Failure of conductivity calibration. Clean sensor, get fresh cal solutions and
redo calibration. Verify conductivity value
of application or reference solutions.
Perform sensor tests as described in
sensor manual. Replace sensor if still
failure.
EEprom Fail Internal nonvolatile memory failure System failure, consult factory.
ATI Q46CT Conductivity System Part 8 – System Maintenance
52
O&M Manual
Rev-G (8/17)
Chcksum Fail Internal software storage error. System failure, consult factory.
Display Fail Internal display driver fail. System failure, consult factory.
Range Cal Fail Failure of factory temperature
calibration.
Consult factory.
53
O&M Manual
Rev-G (8/17)
Spare Parts
Part No. Description
03-0396 Q46CT Front Lid Assembly
07-0334 Q46CT Monitor, 100-240 VAC
07-0335 Q46CT Monitor, 12-24 VDC
07-0336 Q46CT Monitor, 100-240 VAC with Profibus
07-0337 Q46CT Monitor, 12-24 VDC with Profibus
03-0407 Q46 P/S Assy, 100-240VAC
03-0408 Q46 P/S Assy, 100-240VAC with 3
rd
4-20mA output
03-0409 Q46 P/S Assy, 100-240VAC with 3 relay exp. board
03-0410 Q46 P/S Assy, 12-24 VDC
03-0411 Q46 P/S Assy, 12-24 VDC with 3
rd
4-20mA output
03-0412 Q46 P/S Assy, 12-24 VDC with 3 relay exp. Board
38-0072 Terminal block plug, 3 position (relays)
38-0073 Terminal block plug, 4 position (outputs)
38-0074 Terminal block plug, 3 position (ground)
38-0081 Terminal block plug, 3 position (power)
38-0084 Terminal block plug, 3 position (power) – VDC version*
*NOTE: prior to Dec 2018, VDC (power) Terminal block used the (38-0081)
23-0029 Fuse, 630mA, 250V, TR-5 (for AC and DC Analyzers)
07-0100 Junction Box
63-0099 Toroidal Sensor, Noryl, 20 ft. cable
31-0068 Sensor Interconnect cable
09-0046 Conductivity Standard – 84 microSiemens, 500 mL
09-0047 Conductivity Standard – 447 microSiemens, 500 mL
09-0048 Conductivity Standard – 1,500 microSiemens, 500 mL
09-0049 Conduvtivity Standard – 8,974 microSiemens, 500 mL
09-0050 Conductivity Standard – 80,000 microSiemens, 500 mL
Lock/Unlock Code: 1467
PRODUCT WARRANTY
Analytical Technology, Inc. (Manufacturer) warrants to the Customer that if any part(s) of the
Manufacturer's equipment proves to be defective in materials or workmanship within the earlier of 18
months of the date of shipment or 12 months of the date of start-up, such defective parts will be repaired
or replaced free of charge. Inspection and repairs to products thought to be defective within the warranty
period will be completed at the Manufacturer's facilities in Collegeville, PA. Products on which warranty
repairs are required shall be shipped freight prepaid to the Manufacturer. The product(s) will be returned
freight prepaid and allowed if it is determined by the manufacturer that the part(s) failed due to defective
materials or workmanship.
This warranty does not cover consumable items, batteries, or wear items subject to periodic
replacement including lamps and fuses.
Gas sensors carry a 12 months from date of shipment warranty and are subject to inspection for
evidence of misuse, abuse, alteration, improper storage, or extended exposure to excessive gas
concentrations. Should inspection indicate that sensors have failed due to any of the above, the warranty
shall not apply.
The Manufacturer assumes no liability for consequential damages of any kind, and the buyer by
acceptance of this equipment will assume all liability for the consequences of its use or misuse by the
Customer, his employees, or others. A defect within the meaning of this warranty is any part of any piece
of a Manufacturer's product which shall, when such part is capable of being renewed, repaired, or
replaced, operate to condemn such piece of equipment.
This warranty is in lieu of all other warranties ( including without limiting the generality of the
foregoing warranties of merchantability and fitness for a particular purpose), guarantees, obligations or
liabilities expressed or implied by the Manufacturer or its representatives and by statute or rule of law.
This warranty is void if the Manufacturer's product(s) has been subject to misuse or abuse, or has
not been operated or stored in accordance with instructions, or if the serial number has been removed.
Analytical Technology, Inc. makes no other warranty expressed or implied except as stated
above.
Q46CT-Toroidal-Conductivity-Monitor-Manual.pdf

Mais conteúdo relacionado

Semelhante a Q46CT-Toroidal-Conductivity-Monitor-Manual.pdf

Research Inventy: International Journal of Engineering and Science
Research Inventy: International Journal of Engineering and ScienceResearch Inventy: International Journal of Engineering and Science
Research Inventy: International Journal of Engineering and Scienceresearchinventy
 
Research Inventy : International Journal of Engineering and Science is publis...
Research Inventy : International Journal of Engineering and Science is publis...Research Inventy : International Journal of Engineering and Science is publis...
Research Inventy : International Journal of Engineering and Science is publis...researchinventy
 
1 mrk511232 ben-e_en_product_guide__rec670_1.2_pre-configured
1 mrk511232 ben-e_en_product_guide__rec670_1.2_pre-configured1 mrk511232 ben-e_en_product_guide__rec670_1.2_pre-configured
1 mrk511232 ben-e_en_product_guide__rec670_1.2_pre-configuredJavier Cuzco
 
Door control embedded system using accelerometer sensor
Door control embedded system using accelerometer sensor Door control embedded system using accelerometer sensor
Door control embedded system using accelerometer sensor Vikrant Gupta
 
IRJET-Smart System for Food Industries and Bakeries
IRJET-Smart System for Food Industries and BakeriesIRJET-Smart System for Food Industries and Bakeries
IRJET-Smart System for Food Industries and BakeriesIRJET Journal
 
IRJET- Items’ Filling System Prototype with Sorting System According to the C...
IRJET- Items’ Filling System Prototype with Sorting System According to the C...IRJET- Items’ Filling System Prototype with Sorting System According to the C...
IRJET- Items’ Filling System Prototype with Sorting System According to the C...IRJET Journal
 
Q46CT-Toroidal-Conductivity-Monitor.pdf
Q46CT-Toroidal-Conductivity-Monitor.pdfQ46CT-Toroidal-Conductivity-Monitor.pdf
Q46CT-Toroidal-Conductivity-Monitor.pdfENVIMART
 
PD-1076_hardware_manual_hw1.10_rev1.07.pdf
PD-1076_hardware_manual_hw1.10_rev1.07.pdfPD-1076_hardware_manual_hw1.10_rev1.07.pdf
PD-1076_hardware_manual_hw1.10_rev1.07.pdfMohdAsriTaip1
 
DETECTING POWER GRID SYNCHRONISATION FAILURE ON SENSING BAD VOLTAGE OR FREQUE...
DETECTING POWER GRID SYNCHRONISATION FAILURE ON SENSING BAD VOLTAGE OR FREQUE...DETECTING POWER GRID SYNCHRONISATION FAILURE ON SENSING BAD VOLTAGE OR FREQUE...
DETECTING POWER GRID SYNCHRONISATION FAILURE ON SENSING BAD VOLTAGE OR FREQUE...Pradeep Avanigadda
 
testing purpose for ic testing and othettastinng
testing purpose for ic testing and othettastinngtesting purpose for ic testing and othettastinng
testing purpose for ic testing and othettastinngPvaveen
 
IRJET- Solar Powered Transmission Line Inspection Robot
IRJET- Solar Powered Transmission Line Inspection RobotIRJET- Solar Powered Transmission Line Inspection Robot
IRJET- Solar Powered Transmission Line Inspection RobotIRJET Journal
 
classic control and PLC training report
classic control and PLC training reportclassic control and PLC training report
classic control and PLC training reportahmed abdelnasser
 
intelligent braking system report
intelligent braking system reportintelligent braking system report
intelligent braking system reportSumit Kumar
 
IRJET- IoT based Industrial Automation
IRJET-  	  IoT based Industrial AutomationIRJET-  	  IoT based Industrial Automation
IRJET- IoT based Industrial AutomationIRJET Journal
 

Semelhante a Q46CT-Toroidal-Conductivity-Monitor-Manual.pdf (20)

Research Inventy: International Journal of Engineering and Science
Research Inventy: International Journal of Engineering and ScienceResearch Inventy: International Journal of Engineering and Science
Research Inventy: International Journal of Engineering and Science
 
Research Inventy : International Journal of Engineering and Science is publis...
Research Inventy : International Journal of Engineering and Science is publis...Research Inventy : International Journal of Engineering and Science is publis...
Research Inventy : International Journal of Engineering and Science is publis...
 
1 mrk511232 ben-e_en_product_guide__rec670_1.2_pre-configured
1 mrk511232 ben-e_en_product_guide__rec670_1.2_pre-configured1 mrk511232 ben-e_en_product_guide__rec670_1.2_pre-configured
1 mrk511232 ben-e_en_product_guide__rec670_1.2_pre-configured
 
Door control embedded system using accelerometer sensor
Door control embedded system using accelerometer sensor Door control embedded system using accelerometer sensor
Door control embedded system using accelerometer sensor
 
IRJET-Smart System for Food Industries and Bakeries
IRJET-Smart System for Food Industries and BakeriesIRJET-Smart System for Food Industries and Bakeries
IRJET-Smart System for Food Industries and Bakeries
 
IRJET- Items’ Filling System Prototype with Sorting System According to the C...
IRJET- Items’ Filling System Prototype with Sorting System According to the C...IRJET- Items’ Filling System Prototype with Sorting System According to the C...
IRJET- Items’ Filling System Prototype with Sorting System According to the C...
 
Q46CT-Toroidal-Conductivity-Monitor.pdf
Q46CT-Toroidal-Conductivity-Monitor.pdfQ46CT-Toroidal-Conductivity-Monitor.pdf
Q46CT-Toroidal-Conductivity-Monitor.pdf
 
PD-1076_hardware_manual_hw1.10_rev1.07.pdf
PD-1076_hardware_manual_hw1.10_rev1.07.pdfPD-1076_hardware_manual_hw1.10_rev1.07.pdf
PD-1076_hardware_manual_hw1.10_rev1.07.pdf
 
Pt650 d英文说明书 (1)
Pt650 d英文说明书 (1)Pt650 d英文说明书 (1)
Pt650 d英文说明书 (1)
 
Presentaton on Plc & Scada
Presentaton on Plc & ScadaPresentaton on Plc & Scada
Presentaton on Plc & Scada
 
embedded system
embedded system embedded system
embedded system
 
DETECTING POWER GRID SYNCHRONISATION FAILURE ON SENSING BAD VOLTAGE OR FREQUE...
DETECTING POWER GRID SYNCHRONISATION FAILURE ON SENSING BAD VOLTAGE OR FREQUE...DETECTING POWER GRID SYNCHRONISATION FAILURE ON SENSING BAD VOLTAGE OR FREQUE...
DETECTING POWER GRID SYNCHRONISATION FAILURE ON SENSING BAD VOLTAGE OR FREQUE...
 
testing purpose for ic testing and othettastinng
testing purpose for ic testing and othettastinngtesting purpose for ic testing and othettastinng
testing purpose for ic testing and othettastinng
 
Q gx developer_e
Q gx developer_eQ gx developer_e
Q gx developer_e
 
IRJET- Solar Powered Transmission Line Inspection Robot
IRJET- Solar Powered Transmission Line Inspection RobotIRJET- Solar Powered Transmission Line Inspection Robot
IRJET- Solar Powered Transmission Line Inspection Robot
 
Eurotherm US Product Selection Guide
Eurotherm US Product Selection GuideEurotherm US Product Selection Guide
Eurotherm US Product Selection Guide
 
classic control and PLC training report
classic control and PLC training reportclassic control and PLC training report
classic control and PLC training report
 
6600ingles
6600ingles6600ingles
6600ingles
 
intelligent braking system report
intelligent braking system reportintelligent braking system report
intelligent braking system report
 
IRJET- IoT based Industrial Automation
IRJET-  	  IoT based Industrial AutomationIRJET-  	  IoT based Industrial Automation
IRJET- IoT based Industrial Automation
 

Mais de ENVIMART

Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdfZinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdfNickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdfManganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdfLead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdfIron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdfIron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdfChromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdfCopper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdfCopper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdfCopper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdfCobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdfCadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdfArsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdfAluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdfENVIMART
 
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdfATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdfENVIMART
 
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdfATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdfENVIMART
 
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdfATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdfENVIMART
 
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdfATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdfENVIMART
 
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdfATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdfENVIMART
 
B12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdfB12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdfENVIMART
 

Mais de ENVIMART (20)

Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdfZinc Continuous Analysis from SEIBOLD-Wasser.pdf
Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
 
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdfNickel Continuous Analysis from SEIBOLD-Wasser.pdf
Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
 
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdfManganese Continuous Analysis from SEIBOLD-Wasser.pdf
Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
 
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdfLead Continuous Analysis from SEIBOLD-Wasser.pdf
Lead Continuous Analysis from SEIBOLD-Wasser.pdf
 
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdfIron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Manganese Continuous Analysis from SEIBOLD-Wasser.pdf
 
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdfIron Continuous Analysis from SEIBOLD-Wasser.pdf
Iron Continuous Analysis from SEIBOLD-Wasser.pdf
 
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdfChromium Continuous Analysis from SEIBOLD-Wasser.pdf
Chromium Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdfCopper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Zinc Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdfCopper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Nickel Continuous Analysis from SEIBOLD-Wasser.pdf
 
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdfCopper Continuous Analysis from SEIBOLD-Wasser.pdf
Copper Continuous Analysis from SEIBOLD-Wasser.pdf
 
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdfCobalt Continuous Analysis from SEIBOLD-Wasser.pdf
Cobalt Continuous Analysis from SEIBOLD-Wasser.pdf
 
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdfCadmium Continuous Analysis from SEIBOLD-Wasser.pdf
Cadmium Continuous Analysis from SEIBOLD-Wasser.pdf
 
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdfArsenic Continuous Analysis from SEIBOLD-Wasser.pdf
Arsenic Continuous Analysis from SEIBOLD-Wasser.pdf
 
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdfAluminium Continuous Analysis from SEIBOLD-Wasser.pdf
Aluminium Continuous Analysis from SEIBOLD-Wasser.pdf
 
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdfATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Product-Literature.pdf
 
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdfATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
ATIs-B12-2-Wire-Gas-Transmitter-Support-Drawings.pdf
 
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdfATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
ATIs-B12-Wet-Gas-Detector-Product-Literature.pdf
 
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdfATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
ATIs-B12-Wet-Gas-Detector-Support-Drawings.pdf
 
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdfATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
ATIs-B14-Gas-Alarm-Module-Product-Literature.pdf
 
B12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdfB12-2-Wire-Gas-Transmitter.pdf
B12-2-Wire-Gas-Transmitter.pdf
 

Último

Discussing how the scientific mechanisms of the CBD work. CBD mechanisms, pro...
Discussing how the scientific mechanisms of the CBD work. CBD mechanisms, pro...Discussing how the scientific mechanisms of the CBD work. CBD mechanisms, pro...
Discussing how the scientific mechanisms of the CBD work. CBD mechanisms, pro...pensoftservices
 
LCCXG global forum, April 2024, Lydie-Line Paroz
LCCXG global forum, April 2024,  Lydie-Line ParozLCCXG global forum, April 2024,  Lydie-Line Paroz
LCCXG global forum, April 2024, Lydie-Line ParozOECD Environment
 
DETAILED LESSON FINAL DEMO TEACHING.docx
DETAILED LESSON FINAL DEMO TEACHING.docxDETAILED LESSON FINAL DEMO TEACHING.docx
DETAILED LESSON FINAL DEMO TEACHING.docxImilyAcma
 
The Road to Cali. Dr. Aleksandar Rankovic
The Road to Cali. Dr. Aleksandar RankovicThe Road to Cali. Dr. Aleksandar Rankovic
The Road to Cali. Dr. Aleksandar Rankovicpensoftservices
 
CCXG global forum, April 2024, Alban Kitous
CCXG global forum, April 2024,  Alban KitousCCXG global forum, April 2024,  Alban Kitous
CCXG global forum, April 2024, Alban KitousOECD Environment
 
https://www.facebook.com/people/Proper-Keto-Capsules-UK/61557989683758/
https://www.facebook.com/people/Proper-Keto-Capsules-UK/61557989683758/https://www.facebook.com/people/Proper-Keto-Capsules-UK/61557989683758/
https://www.facebook.com/people/Proper-Keto-Capsules-UK/61557989683758/dikjog
 
Biodiversity and Health. Prof. Richard Kock.
Biodiversity and Health. Prof. Richard Kock.Biodiversity and Health. Prof. Richard Kock.
Biodiversity and Health. Prof. Richard Kock.pensoftservices
 
CCXG global forum, April 2024, Mia Ryan
CCXG global forum, April 2024,  Mia RyanCCXG global forum, April 2024,  Mia Ryan
CCXG global forum, April 2024, Mia RyanOECD Environment
 
Get inspired by SYMBA Project: promoting Industrial Symbiosis
Get inspired by SYMBA Project: promoting Industrial SymbiosisGet inspired by SYMBA Project: promoting Industrial Symbiosis
Get inspired by SYMBA Project: promoting Industrial Symbiosissymbaprojecteu
 
CCXG global forum, April 2024, Annett Möhner
CCXG global forum, April 2024,  Annett MöhnerCCXG global forum, April 2024,  Annett Möhner
CCXG global forum, April 2024, Annett MöhnerOECD Environment
 
Objectives and main international biodiversity events during the upcoming Hun...
Objectives and main international biodiversity events during the upcoming Hun...Objectives and main international biodiversity events during the upcoming Hun...
Objectives and main international biodiversity events during the upcoming Hun...pensoftservices
 
CCXG global forum, April 2024, Brian Motherway and Paolo Frankl
CCXG global forum, April 2024,  Brian Motherway and Paolo FranklCCXG global forum, April 2024,  Brian Motherway and Paolo Frankl
CCXG global forum, April 2024, Brian Motherway and Paolo FranklOECD Environment
 
Your Compost Is Not Compostable -- Tips on What Best Rots
Your Compost Is Not Compostable -- Tips on What Best RotsYour Compost Is Not Compostable -- Tips on What Best Rots
Your Compost Is Not Compostable -- Tips on What Best RotsDeidre Pike
 
Oro -dental Preparation.pdfsksjssjsjsssss
Oro -dental Preparation.pdfsksjssjsjsssssOro -dental Preparation.pdfsksjssjsjsssss
Oro -dental Preparation.pdfsksjssjsjsssssMustafaHussain83
 
CCXG global forum, April 2024, MJ Mace
CCXG global forum, April 2024,   MJ MaceCCXG global forum, April 2024,   MJ Mace
CCXG global forum, April 2024, MJ MaceOECD Environment
 
CCXG global forum, April 2024, Thomas Spencer
CCXG global forum, April 2024,  Thomas SpencerCCXG global forum, April 2024,  Thomas Spencer
CCXG global forum, April 2024, Thomas SpencerOECD Environment
 
The Entropy Law and the Economic Process
The Entropy Law and the Economic ProcessThe Entropy Law and the Economic Process
The Entropy Law and the Economic ProcessJoão Soares
 
Ecosystem and Environment, GE3451 Environmental Science
Ecosystem and Environment, GE3451 Environmental ScienceEcosystem and Environment, GE3451 Environmental Science
Ecosystem and Environment, GE3451 Environmental Sciencepriyamanohari1
 
CCXG global forum, April 2024, Niklas Höhne
CCXG global forum, April 2024,  Niklas HöhneCCXG global forum, April 2024,  Niklas Höhne
CCXG global forum, April 2024, Niklas HöhneOECD Environment
 

Último (20)

Discussing how the scientific mechanisms of the CBD work. CBD mechanisms, pro...
Discussing how the scientific mechanisms of the CBD work. CBD mechanisms, pro...Discussing how the scientific mechanisms of the CBD work. CBD mechanisms, pro...
Discussing how the scientific mechanisms of the CBD work. CBD mechanisms, pro...
 
LCCXG global forum, April 2024, Lydie-Line Paroz
LCCXG global forum, April 2024,  Lydie-Line ParozLCCXG global forum, April 2024,  Lydie-Line Paroz
LCCXG global forum, April 2024, Lydie-Line Paroz
 
DETAILED LESSON FINAL DEMO TEACHING.docx
DETAILED LESSON FINAL DEMO TEACHING.docxDETAILED LESSON FINAL DEMO TEACHING.docx
DETAILED LESSON FINAL DEMO TEACHING.docx
 
The Road to Cali. Dr. Aleksandar Rankovic
The Road to Cali. Dr. Aleksandar RankovicThe Road to Cali. Dr. Aleksandar Rankovic
The Road to Cali. Dr. Aleksandar Rankovic
 
Biopesticide. pptx.
Biopesticide. pptx.Biopesticide. pptx.
Biopesticide. pptx.
 
CCXG global forum, April 2024, Alban Kitous
CCXG global forum, April 2024,  Alban KitousCCXG global forum, April 2024,  Alban Kitous
CCXG global forum, April 2024, Alban Kitous
 
https://www.facebook.com/people/Proper-Keto-Capsules-UK/61557989683758/
https://www.facebook.com/people/Proper-Keto-Capsules-UK/61557989683758/https://www.facebook.com/people/Proper-Keto-Capsules-UK/61557989683758/
https://www.facebook.com/people/Proper-Keto-Capsules-UK/61557989683758/
 
Biodiversity and Health. Prof. Richard Kock.
Biodiversity and Health. Prof. Richard Kock.Biodiversity and Health. Prof. Richard Kock.
Biodiversity and Health. Prof. Richard Kock.
 
CCXG global forum, April 2024, Mia Ryan
CCXG global forum, April 2024,  Mia RyanCCXG global forum, April 2024,  Mia Ryan
CCXG global forum, April 2024, Mia Ryan
 
Get inspired by SYMBA Project: promoting Industrial Symbiosis
Get inspired by SYMBA Project: promoting Industrial SymbiosisGet inspired by SYMBA Project: promoting Industrial Symbiosis
Get inspired by SYMBA Project: promoting Industrial Symbiosis
 
CCXG global forum, April 2024, Annett Möhner
CCXG global forum, April 2024,  Annett MöhnerCCXG global forum, April 2024,  Annett Möhner
CCXG global forum, April 2024, Annett Möhner
 
Objectives and main international biodiversity events during the upcoming Hun...
Objectives and main international biodiversity events during the upcoming Hun...Objectives and main international biodiversity events during the upcoming Hun...
Objectives and main international biodiversity events during the upcoming Hun...
 
CCXG global forum, April 2024, Brian Motherway and Paolo Frankl
CCXG global forum, April 2024,  Brian Motherway and Paolo FranklCCXG global forum, April 2024,  Brian Motherway and Paolo Frankl
CCXG global forum, April 2024, Brian Motherway and Paolo Frankl
 
Your Compost Is Not Compostable -- Tips on What Best Rots
Your Compost Is Not Compostable -- Tips on What Best RotsYour Compost Is Not Compostable -- Tips on What Best Rots
Your Compost Is Not Compostable -- Tips on What Best Rots
 
Oro -dental Preparation.pdfsksjssjsjsssss
Oro -dental Preparation.pdfsksjssjsjsssssOro -dental Preparation.pdfsksjssjsjsssss
Oro -dental Preparation.pdfsksjssjsjsssss
 
CCXG global forum, April 2024, MJ Mace
CCXG global forum, April 2024,   MJ MaceCCXG global forum, April 2024,   MJ Mace
CCXG global forum, April 2024, MJ Mace
 
CCXG global forum, April 2024, Thomas Spencer
CCXG global forum, April 2024,  Thomas SpencerCCXG global forum, April 2024,  Thomas Spencer
CCXG global forum, April 2024, Thomas Spencer
 
The Entropy Law and the Economic Process
The Entropy Law and the Economic ProcessThe Entropy Law and the Economic Process
The Entropy Law and the Economic Process
 
Ecosystem and Environment, GE3451 Environmental Science
Ecosystem and Environment, GE3451 Environmental ScienceEcosystem and Environment, GE3451 Environmental Science
Ecosystem and Environment, GE3451 Environmental Science
 
CCXG global forum, April 2024, Niklas Höhne
CCXG global forum, April 2024,  Niklas HöhneCCXG global forum, April 2024,  Niklas Höhne
CCXG global forum, April 2024, Niklas Höhne
 

Q46CT-Toroidal-Conductivity-Monitor-Manual.pdf

  • 1. O & M Manual Model Q46CT Toroidal Conductivity Monitor Home Office European Office Analytical Technology, Inc. ATI (UK) Limited 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park Collegeville, PA 19426 Delph New Road, Delph Phone: 800-959-0299 Saddleworth OL3 5DE 610-917-0991 Phone: +44 (0)1457-873-318 Fax: 610-917-0992 Fax: + 44 (0)1457-874-468 Email: sales@analyticaltechnology.com Email: sales@atiuk.com
  • 2. 2 O&M Manual Rev-G (8/17) Table of Contents PART 1 - INTRODUCTION.....................................4 1.1 General ................................................. 4 1.2 Features................................................ 4 1.3 Q46CT System Specifications.............. 5 1.4 Q46CT Performance Specifications ..... 7 PART 2 – ANALYZER MOUNTING .......................8 2.1 General ................................................. 8 2.2 Wall or Pipe Mount ............................... 9 2.3 Panel Mount Monitor.......................... 11 PART 3 – SENSOR MOUNTING......................... 12 3.1 General ............................................... 12 3.2 Tee Mounting...................................... 13 3.3 Submersion Mounting......................... 14 3.4 In-Line Installation............................... 15 PART 4 – ELECTRICAL INSTALLATION .......... 17 4.1 General ............................................... 17 4.2 Power Connection .............................. 17 4.3 Relay Connection ............................... 19 4.4 Optional Output or Relay Connections20 4.5 Sensor Wiring ..................................... 21 4.6 Direct Sensor Connection................... 23 4.7 Junction Box Connection.................... 24 PART 5 – CONFIGURATION .............................. 25 5.1 User Interface ..................................... 25 5.11 Keys ................................................ 26 5.12 Display ............................................ 26 5.2 Software.............................................. 27 5.22 Measure Menu [MEASURE] ........... 30 5.23 Calibration Menu [CAL] ........................ 31 5.24 Configuration Menu [CONFIG]....... 31 5.25 Control Menu................................... 35 5.26 Diagnostics Menu [DIAG] .................... 38 PART 6 – CALIBRATION.................................... 41 6.1 Overview and Methods....................... 41 6.11 1-Point Cal Explained ..................... 41 6.12 Zero Cal Cal Explained ................... 41 6.2 Performing Sensor Zero Cal............... 41 6.3 Performing 1-Point Cal ....................... 42 6.4 Temperature Calibration..................... 43 PART 7 – PID CONTROLLER DETAILS ............ 45 7.1 PID Description................................... 45 7.2 PID Algorithm...................................... 45 7.3 Classical PID Tuning .......................... 47 7.4 Manual PID Override Control ............. 47 7.5 Common PID Pitfalls........................... 47 PART 8 – SYSTEM MAINTENANCE .................. 49 8.1 System Checks................................... 49 8.2 Instrument Checks.............................. 49 8.3 Sensor Tests....................................... 49 8.4 Display Messages............................... 50 SPARE PARTS.................................................... 53
  • 3. 3 O&M Manual Rev-G (8/17) Table of Figures FIGURE 1 - Q46 ENCLOSURE DIMENSIONS........................................................................................................8 FIGURE 2 - WALL OR PIPE MOUNT BRACKET.....................................................................................................9 FIGURE 3 - WALL MOUNTING DIAGRAM...........................................................................................................10 FIGURE 4 - PIPE MOUNTING DIAGRAM ............................................................................................................10 FIGURE 5 - AC PWR PANEL MOUNT CUTOUT ..................................................................................................11 FIGURE 6 - TOROIDAL SENSOR DIMENSIONS...................................................................................................12 FIGURE 7 – SENSOR AND FLOW TEE ASSEMBLY..............................................................................................13 FIGURE 8 - SENSOR SUBMERSION MOUNT......................................................................................................14 FIGURE 9 - FLOW TEE PROCESS PIPING .........................................................................................................15 FIGURE 10 - IN-LINE PROCESS PIPING............................................................................................................16 FIGURE 11 - Q46 CONNECTION LAYOUT.........................................................................................................18 FIGURE 12 - RELAY CONNECTIONS.................................................................................................................19 FIGURE 13 - OPTIONAL RELAY BOARD WIRING ...............................................................................................20 FIGURE 14 - OPTIONAL ANALOG OUTPUT WIRING ...........................................................................................20 FIGURE 15 - Q46CT SENSOR CONNECTIONS .................................................................................................22 FIGURE 16 - SENSOR CABLE PREPARATION....................................................................................................23 FIGURE 17 - JUNCTION BOX INTERCONNECT WIRING.......................................................................................24 FIGURE 18 - USER INTERFACE .......................................................................................................................25 FIGURE 19 - SOFTWARE MAP.........................................................................................................................29 FIGURE 20 - CONTROL RELAY EXAMPLE.........................................................................................................37 FIGURE 21 - ALARM RELAY EXAMPLE .............................................................................................................37 FIGURE 22 - NACL REFERENCE SOLUTION CALIBRATION CHART .....................................................................42 FIGURE 23 - PID EQUATION...........................................................................................................................45 FIGURE 24 - PT100 RTD RESISTANCE TABLE.................................................................................................50
  • 4. 4 O&M Manual Rev-G (8/17) Part 1 - Introduction 1.1 General The Model Q46CT Conductivity monitor/analyzer provides a versatile system for monitoring and control of conductivity over a range of 1 µS/cm to 2.000 Siemens/cm. The display may be programmed for ranges of 0-2000 uS, 0-2.000 mS, 0-20.00 mS, 0-200 mS, 0-2000 mS, or 0- 2.000 S. A lower line display indicates the equivalent TDS value in mg/l, using a user programmable conversion factor. Q46CT monitors use a toroidal conductivity sensor that measures conductivity inductively. This type of sensor is particularly well suited to measurements in solutions that contain materials that would foul a normal conductivity sensor. It is very good for high conductivity measurements, but should not be used if normal conductivity is below 100 uS. For special applications, the Q46CT has a few common pre-programmed tables that allow the user to monitor “percent concentration”. Selectable tables include NaCl (0-20%), HCl (0-18%), NaOH (0-15%), and KOH (0-15%). If data is available, you can enter your own table data into the Q46CT to provide a customized concentration display for other chemicals in solution. Monitors are available in two electronic versions, a universal AC powered monitor for operation from 90-260 VAC and a 12-24 VDC unit. Both versions provide two 4-20 mA analog outputs and 3 SPDT relays. One analog output may be configured for PID control and one of the relays may be configured to provide a remote trouble indication. The Q46CT is available with a few options to expand the capabilities of the monitor. Users may select either a third analog 4-20 mA output or an expansion card that provides three additional low voltage SPST relays. The expansion relays are not isolated from each other. This monitor also provides digital communication capability. Users may order Q46 systems with either Profibus DP, Modbus RTU, or Ethernet communications boards factory installed. WARNING: Not following operating instructions may impair safety. 1.2 Features  Power options for either AC or low voltage DC operation.  High accuracy, high sensitivity system, measures from 1 to 2,000,000 uS through 7 internal automatic ranges. User display ranges include 2000 us, 2.000 mS, 20.00 mS, or 200.0 mS, 2000 mS, or 2.000 S.  Output Hold, Output Simulate, Output Alarm, and Output Delay. All forced changes in output condition include bumpless transfer to provide gradual return to on-line signal levels and to avoid system control shocks on both analog outputs.  Selectable PID controller on main analog output. PID includes manual operation feature, and diagnostic “stuck-controller” timer feature for relay notification of control problems.
  • 5. ATI Q46CT Conductivity System Part 1– Introduction 5 O&M Manual Rev-G (8/17)  Digital communication option for Profibus-DP, Modbus-RTU. More versions pending. See Q46 Communications Manual for Specificaitons.  Configurable for TDS display and signal output on one analog output.  Two analog outputs track conductivity and temperature, conductivity and conductivity, or conductivity and TDS. Both analog outputs can be individually programmed to fail to specific values.  Large, high contrast, custom LCD display with LED back light provides excellent readability in any light conditions. The second line of display utilizes 5x7 dot matrix characters for clear message display.  Toroidal conductivity sensor. Sensing electrodes are molded in a noryl body and completely isolated from the measured solution. Conductivity measurements can be made in almost any aqueous solution without being affected by high solids, oils, and other materials that foul contacting sensors.  Quick and easy one-point calibration method and sensor zero-cal. To provide high accuracy, all calibration methods include stability monitors that check temperature and main parameter stability before accepting data.  High accuracy Pt1000 temperature input. Temperature element can be user calibrated.  Security lock feature to prevent unauthorized tampering with transmitter settings. All settings can be viewed while locked, but they cannot be changed. 1.3 Q46CT System Specifications Displayed Parameters Main input, 0 uS to 2S (2,000,000 uS) Sensor temperature, -10.0 to 210.0 °C (23 to 410 ºF) Loop current, 4.00 to 20.00 mA Sensor slope Model number and software version PID Controller Status Timer Status Main Parameter Ranges Automatic or Manual selection of display range 0 to 2000 uS 0.0 to 2.000 mS 0.00 to 20.00 mS 0.0 to 200.0 mS 0 to 2000 mS 0.000 to 2.000 S Display 0.75” (19.1 mm) high 4-digit main display with sign 12-digit secondary display, 0.3" (7.6 mm) 5x7 dot matrix. Integral LED back-light for visibility in the dark. Power 90 - 260 VAC, 50-60 Hz, 10 VA max 12-24 VDC, 500 mA max.
  • 6. ATI Q46CT Conductivity System Part 1– Introduction 6 O&M Manual Rev-G (8/17) Enclosure NEMA 4X, polycarbonate, stainless steel hardware, weatherproof and corrosion resistant. Mounting Options Wall or pipe mount bracket standard. Bracket suitable for either 1.5” or 2” I.D. U-Bolts for pipe mounting. Panel mount adapter optional. Conduit Openings Five ½” NPT openings, Adapter can be removed to provide a 1” NPT opening in the bottom of the enclosure. Gland seals provided but not installed. Relays, Electromechanical: Three SPDT, 6 amp @ 250 VAC, 5 amp @ 24 VDC contacts. Software selection for setpoint, phase, delay, deadband, hi-lo alarm, and failsafe.A-B indicators on main LCD, and C indicator on lower display. Analog Outputs Two 4-20 mA outputs. Output one programmable for pH or PID. Output 2 programmable for pH or Temperature. Max load 500 Ohms for each output. Outputs ground isolated and isolated from each other. An additional 3 rd analog option is available. Keypad 4-key membrane type with tactile feedback, polycarbonate with UV coating Weight 2 lb. (0.9 kg) Ambient Temperature Analyzer Service, -20 to 60 °C (-4 to 140 ºF) Storage, -30 to 70 °C (-22 to 158 ºF) Ambient Humidity 0 to 95%, indoor/outdoor use, non-condensing. Altitude Up to 2000 m (6562 Ft.) Electrical Certification Ordinary Location, cCSAus (CSA and UL standards - both approved by CSA), pollution degree 2, installation category 2 EMI/RFI Influence Designed to EN 61326-1 Output Isolation 600 V galvanic isolation Q25CT Conductivity sensor Sensor Toroidal sensor with encapsulated temp. element. Measuring Range 0.000 to 2.000 S/cm Temperature Element Pt1000 RTD Wetted Materials Noryl Optional Flow Tee Special 2” CPVC tee with alignment slot Sensor Cable 20 ft. (6 m.) standard.
  • 7. ATI Q46CT Conductivity System Part 1– Introduction 7 O&M Manual Rev-G (8/17) Pressure Range 0-150 PSIG Temperature Range 0-105° C Maximum Flow Rate 10 feet (3 meters) per second Max. Sensor-Analyzer Distance 200 feet (60 m.) 1.4 Q46CT Performance Specifications Accuracy 0.5% of user range, or better (± 2 μS) Repeatability 0.2% of user range, or better (± 2 μS) Sensitivity 0.05% of user ranges (± 2 μS) Stability 0.2% of user range per 24 hours, non-cumulative Warm-up Time 3 seconds to rated performance (electronics only) Supply Voltage Effects ± 0.05% of user range Instrument Response Time 6 seconds to 90% of step input at lowest setting Temperature Drift Span or zero, 0.04% of span/°C Equipment bearing this marking may not be discarded by traditional methods in the European community after August 12 2005 per EU Directive 2002/96/EC. End users must return old equipment to the manufacturer for proper disposal.
  • 8. 8 O&M Manual Rev-G (8/17) Part 2 – Analyzer Mounting 2.1 General All Q46 Series instruments offer maximum mounting flexibility. A bracket is included with each unit that allows mounting to walls or pipes. In all cases, choose a location that is readily accessible for calibrations. Also consider that it may be necessary to utilize a location where solutions can be used during the calibration process. To take full advantage of the high contrast display, mount the instrument in a location where the display can be viewed from various angles and long distances. Locate the instrument in close proximity to the point of sensor installation - this will allow easy access during calibration. The standard cable length of the conductivity sensor is 20 feet. For sensor cables longer than 20 feet, use the optional junction box (07-0100) and sensor interconnect cable (31-0057). Figure 1 - Q46 Enclosure Dimensions
  • 9. ATI Q46CT Conductivity System Part 2– Analyzer Mounting 9 O&M Manual Rev-G (8/17) 2.2 Wall or Pipe Mount A PVC mounting bracket with attachment screws is supplied with each transmitter. The multi- purpose bracket is attached to the rear of the enclosure using the four flat head screws. The instrument is then attached to the wall using the four outer mounting holes in the bracket. These holes are slotted to accommodate two sizes of u-bolt that may be used to pipe mount the unit. Slots will accommodate u-bolts designed for 1½“ or 2” pipe. The actual center to center dimensions for the u-bolts are shown in the drawing. Note that these slots are for u-bolts with ¼- 20 threads. The 1½” pipe u-bolt (2” I.D. clearance) is available from ATI in type 304 stainless steel under part number (47-0005) Figure 2 - Wall or Pipe Mount Bracket
  • 10. ATI Q46CT Conductivity System Part 2– Analyzer Mounting 10 O&M Manual Rev-G (8/17) Figure 3 - Wall Mounting Diagram Figure 4 - Pipe Mounting Diagram Note: Analyzer shown with optional Profibus Connector mounted to side of enclosure.
  • 11. ATI Q46CT Conductivity System Part 2– Analyzer Mounting 11 O&M Manual Rev-G (8/17) 2.3 Panel Mount Monitor Panel mounting uses the panel mounting flange molded into the rear section of the enclosure. Figure 5 provides dimensions for the panel cutout required for mounting. The panel mounting bracket kit must be ordered separately (part number 05-0094). This kit contains a metal retainer bracket that attaches to the rear of the enclosure, 4 screws for attachment of this bracket, and a sealing gasket to insure that the panel mounted monitor provides a water tight seal when mounted to a panel. The sealing gasket must first be attached to the enclosure. The gasket contains an adhesive on one side so that it remains in place on the enclosure. Remove the protective paper from the adhesive side of the gasket and slide the gasket over the back of the enclosure so that the adhesive side lines up with the back of the enclosure flange. Once in place, you can proceed to mount the monitor in the panel. Figure 5 - AC Pwr Panel Mount Cutout
  • 12. 12 O&M Manual Rev-G (8/17) Part 3 – Sensor Mounting 3.1 General The Q25CT Conductivity Sensor is designed for industrial and municipal process applications. Mounting options include flow-through or submersion. The sensor-to-analyzer distance must not exceed 200 feet (60 m.). In submersion applications, there must be a minimum of 2” clearance around and below the toroidal sensor. Placing the sensor close to wall or other objects will result in a loss of accuracy. Calibrate the sensor before placing it into the process. See Part 7 of this manual for detailed calibration instructions. Figure 6 below shows the dimensions of the toroidal sensor. Figure 6 - Toroidal Sensor Dimensions
  • 13. ATI Q46CT Conductivity System Part 3 – Sensor Mounting 13 O&M Manual Rev-G (8/17) 3.2 Tee Mounting The toroidal sensors are mounted in a 2" pipe using an optional pipe adapter. The tee fitting is keyed so the sensor is oriented in the process as shown in Figure 7, below. The sensor bore opening should be aligned so that flow passes directly through the open sensor bore. A positioning notch is located on the upper sensor collar to aid in this alignment. This orientation is used to ensure a representative sample is being measured and to keep the sensor bore clean. Note that sensor must be zero-calibrated before final mounting in tee. 2-WIRE TOROIDAL CONDUCTIVITY TRANSMITTER Figure 7 – Sensor and Flow Tee Assembly
  • 14. ATI Q46CT Conductivity System Part 3 – Sensor Mounting 14 O&M Manual Rev-G (8/17) 3.3 Submersion Mounting When using this conductivity sensor for submersion applications, mount the sensor to the end of a 1” mounting pipe using a 1” coupling. ATI’s (00-0628) mounting assembly shown in 11 is available for submersible applications. This assembly is designed to mount to standard handrails and facilitates insertion and removal of the sensor. Figure 8 - Sensor Submersion Mount
  • 15. ATI Q46CT Conductivity System Part 3 – Sensor Mounting 15 O&M Manual Rev-G (8/17) 3.4 In-Line Installation Toroidal Conductivity sensors may be installed directly into a flowing pipe system provided that the water does not contain a lot of entrained air. A 2” flow tee assembly is available for this purpose. It is best to install the sensor in a vertical pipe section with water flowing upward. This assures that air pockets cannot develop at the sensor. If installed in a horizontal run of pipe, place the sensor at the 3 or 9 o’clock position. Never mount the sensor on the top or bottom of the pipe. It is also good practice to install a bypass system around the sensor for maintenance and calibration purposes. Figure 9 - Flow Tee Process Piping
  • 16. ATI Q46CT Conductivity System Part 3 – Sensor Mounting 16 O&M Manual Rev-G (8/17) Figure 10 - In-Line Process Piping
  • 17. 17 O&M Manual Rev-G (8/17) Part 4 – Electrical Installation 4.1 General The Q46 is powered in one of two ways, depending on the version purchased. The 12-24 VDC powered analyzer requires a customer supplied DC power supply. The 90-260 VAC version requires line power. Please verify the type of unit before connecting any power. WARNING: Do not connect AC line power to the DC version. Severe damage will result. Important Notes: 1. Use wiring practices that conform to all national, state and local electrical codes. For proper safety as well as stable measuring performance, it is important that the earth ground connection be made to a solid ground point from TB7. The AC power supply contains a single 630mA slo-blo fuse (Wickmann/Littlefuse #372-0630). The fuse F1 is located adjacent to TB7 and is easily replaceable. 2. Do NOT run sensor cables or instrument 4-20 mA output wiring in the same conduit that contains AC power wiring. AC power wiring should be run in a dedicated conduit to prevent electrical noise from coupling with the instrumentation signals. 3. This analyzer must be installed by specifically trained personnel in accordance with relevant local codes and instructions contained in this operating manual. Observe the analyzer's technical specifications and input ratings. Proper electrical disconnection means must be provided prior to the electrical power connected to this instrument, such as a circuit breaker - rated 250 VAC, 2 A minimum. If one line of the line power mains is not neutral, use a double-pole mains switch to disconnect the analyzer. 4. Repeated problems with lightning strikes damaging sensitive instrumentation are often attributed to poorly bonded earth grounds in the instrument power source. The protection schemes incorporated into this analyzer cannot operate to maximum efficiency unless the ground connection is at its’ absolute lowest impedance. There is no standard ground resistance universally recognized. Many agencies recommend a ground resistance value of 5 ohms or less. The NEC recommends an impedance to ground of less than 25 ohms, and less than 5 ohms where sensitive equipment is installed. Power sources feeding sensitive instruments like the Q46 should have the lowest possible impedance to ground. 4.2 Power Connection Verify the AC power supply requirement before installing. Also verify that power is fully disconnected before attempting to wire. Q46 systems are supplied with 5 cable gland fittings for sealing cable entries. Connect HOT, NEUTRAL, and GROUND to the matching designations on terminal strip TB7.
  • 18. ATI Q46CT Conductivity System Part 4 – Electrical Installation 18 O&M Manual Rev-G (8/17) The two analog outputs for the standard system are present at terminal TB1. The loop-load limitation in this configuration is 500 Ohms maximum for each output. Also note that these two outputs are completely isolated from each other to insure that ground loops do not result from the connection of both outputs to the same device such as a PLC or DCS. A ribbon cable connects the power supply assembly with the microprocessor assembly located in the front section of the enclosure. This cable may be unplugged from the front section of the monitor if service is needed, but should normally be left in place during installation. Figure 11 - Q46 Connection Layout WARNING Disconnect line power voltage BEFORE connecting line power wires to Terminal TB7 of the power supply. The power supply accepts only standard three-wire single phase power. AC power supplies are configured for 100-240 VAC operation at the factory at time of order, and the power supply is labeled as such. Do NOT connect voltages other than the labeled requirement to the input.
  • 19. ATI Q46CT Conductivity System Part 4 – Electrical Installation 19 O&M Manual Rev-G (8/17) 4.3 Relay Connection Three SPDT relays are provided on the power supply board. None of the relay contacts are powered. The user must supply the proper power to the contacts. For applications that require the same switched operating voltage as the Q46 (115 or 230 V), power may be jumpered from the power input terminals at TB7. Relay wiring is connected at TB4, TB5, and TB6 as shown below. Note that the relay contact markings are shown in the NORMAL mode. Programming a relay for “Failsafe” operation reverses the NO and NC positions in this diagram (Figure 12). Figure 12 - Relay Connections
  • 20. ATI Q46CT Conductivity System Part 4 – Electrical Installation 20 O&M Manual Rev-G (8/17) 4.4 Optional Output or Relay Connections TB2, is used to connect to the optional 3-relay card (Figure 13) OR the optional third analog output Out#3, (Figure 14). The Q46 can be configured for only one of these optional features, and the hardware for either option must be factory installed. Figure 13 - Optional Relay Board Wiring Figure 14 - Optional Analog Output Wiring
  • 21. ATI Q46CT Conductivity System Part 4 – Electrical Installation 21 O&M Manual Rev-G (8/17) 4.5 Sensor Wiring The sensor cable can be quickly connected to the Q46 terminal strip by matching the wire colors on the cable to the color designations on the label in the monitor. A junction box is also available to provide a break point for long sensor cable runs. Route signal cable away from AC power lines, adjustable frequency drives, motors, or other noisy electrical signal lines. Do not run sensor or signal cables in conduit that contains AC power lines or motor leads. When installing conductivity monitors and sensors, we recommend keeping the sensor cable as short as is practical. This will minimize potential noise problems. Dad DANGER: DO NOT connect sensor cable to power lines. Serious injury may result. Take care to route sensor cable away from AC power lines, adjustable frequency drives, motors, or other noisy electrical signal lines. Do not run signal lines in the same conduit as AC power lines. Run signal cable in dedicated metal conduit if possible. For optimum electrical noise protection, run an earth ground wire to the ground terminal in the transmitter. Refer to Figure 11 & Figure 15 for illustrative details on electrical installation.
  • 22. ATI Q46CT Conductivity System Part 4 – Electrical Installation 22 O&M Manual Rev-G (8/17) Figure 15 - Q46CT Sensor Connections
  • 23. ATI Q46CT Conductivity System Part 4 – Electrical Installation 23 O&M Manual Rev-G (8/17) 4.6 Direct Sensor Connection Sensor connections are made in accordance with Figure 15. The sensor cable can be routed into the enclosure through one of cord-grips supplied with the unit. Routing sensor wiring through conduit is only recommended if a junction box is to be used. Some loose cable is needed near the installation point so that the sensor can be inserted and removed easily depending on the installation type. Cord-grips used for sealing the cable should be snugly tightened after electrical connections have been made to prevent moisture incursion. When stripping cables, leave adequate length for connections in the transmitter enclosure as shown below. The standard 20 ft. sensor cable normally supplied with the system is already stripped and ready for wiring. This cable can be cut to a shorter length if desired to remove extra cable in a given installation. Do not cut the cable so short as to make installation and removal of the sensor difficult. Figure 16 - Sensor Cable Preparation
  • 24. ATI Q46CT Conductivity System Part 4 – Electrical Installation 24 O&M Manual Rev-G (8/17) 4.7 Junction Box Connection When sensor separation from the monitor is needed and to be greater than 20 feet, the sensor junction box is required (07-0100). Wire according to Figure 17 with 3 paired, individually shielded 22 AWG cable (31-0068). CAUTION: When using a junction box and sensor interconnect cable, the RED SHIELD must be isolated from the WHITE and GREEN SHIELDS. Failure to maintain isolation with the RED SHIELD will result in measurement instability. Figure 17 - Junction Box Interconnect Wiring
  • 25. 25 O&M Manual Rev-G (8/17) Part 5 – Configuration 5.1 User Interface The user interface for the Q46 Series instrument consists of a custom display and a membrane keypad. All functions are accessed from this user interface (no internal jumpers, pots, etc.). Figure 18 - User Interface MENU ICONS UNITS 12-CHARACTER SECONDARY DISPLAY MEMBRANE KEYPAD MENU ESC ENTER A B DIAG FAIL HOLD CAL CONF MENU ICONS UNITS 12-CHARACTER SECONDARY DISPLAY MEMBRANE KEYPAD ENTER KEY LEFT ARROW KEY 4-DIGIT MAIN DISPLAY MENU/ESCAPE KEY UP ARROW KEY SIGN RELAY/LO-BAT INDICATOR 4-KEY USER INTERFACE RELAY INDICATOR
  • 26. ATI Q46CT Conductivity System Part 5 – Configuration 26 O&M Manual Rev-G (8/17) 5.11 Keys All user configurations occur through the use of four membrane keys. These keys are used as follows: MENU/ESC To scroll through the menu section headers or to escape from anywhere in software. The escape sequence allows the user to back out of any changes in a logical manner. Using the escape key aborts all changes to the current screen and backs the user out one level in the software tree. The manual will refer to this key as either MENU or ESC, depending upon its particular function. UP (arrow) To scroll through individual list or display items and to change number values. LEFT (arrow) To move the cursor from right to left during changes to a number value. ENTER To select a menu section or list item for change and to store any change. 5.12 Display The large custom display provides clear information for general measurement use and user configuration. There are three main areas of the display: the main parameter display, the secondary message line, and the icon area. Main Parameter During normal operation, the main parameter display indicates the present process input with sign and units. This main display may be configured to display any of the main measurements that the system provides. During configuration, this area displays other useful set-up information to the user. Lower Line During normal operation, the lower line of the display indicates user-selected secondary measurements that the system is making. This also includes calibration data from the last calibration sequence and the transmitter model number and software version. During configuration, the lower line displays menu items and set-up prompts to the user. Finally, the lower line will display error messages when necessary. For a description of all display messages, refer to Section 8.5.
  • 27. ATI Q46CT Conductivity System Part 5 – Configuration 27 O&M Manual Rev-G (8/17) Icon Area The icon area contains display icons that assist the user in set-up and indicate important states of system functions. The CAL, CONFIG, CNTRL, and DIAG icons are used to tell the user what branch of the software tree the user is in while scrolling through the menu items. This improves software map navigation dramatically. Upon entry into a menu, the title is displayed (such as CAL), and then the title disappears to make way for the actual menu item. However, the icon stays on. HOLD The HOLD icon indicates that the current output of the transmitter has been put into output hold. In this case, the output is locked to the last input value measured when the HOLD function was entered. HOLD values are retained even if the unit power is cycled. FAIL The FAIL icon indicates that the system diagnostic function has detected a problem that requires immediate attention. This icon is automatically cleared once the problem has been resolved. Relay Area A/B The relay area contains two icons that indicate the state of the system relays. Relay C is normally configured for FAIL indication, so it is only displayed on the lower MEASURE display line. 5.2 Software The software of the Q46CT is organized in an easy to follow menu-based system. All user settings are organized under five menu sections: Measure, Calibration [CAL], Configuration [CONFIG], Control [CONTROL] and Diagnostics [DIAG]. Note: The default Measure Menu is display-only and has no menu icon. A B
  • 28. ATI Q46CT Conductivity System Part 5 – Configuration 28 O&M Manual Rev-G (8/17) 5.21 Software Navigation Within the CAL, CONFIG, CONTROL, and DIAG menu sections is a list of selectable items. Once a menu section (such as CONFIG) has been selected with the MENU key, the user can access the item list in this section by pressing either the ENTER key or the UP arrow key. The list items can then be scrolled through using the UP arrow key. Once the last item is reached, the list wraps around and the first list item is shown again. The items in the menu sections are organized such that more frequently used functions are first, while more permanent function settings are later in the list. See Figure 19 for a visual description of the software. Each list item allows a change to a stored system variable. List items are designed in one of two forms: simple single variable, or multiple variable sequences. In the single variable format, the user can quickly modify one parameter - for example, changing temperature display units from °F to °C. In the multiple variable sequence, variables are changed as the result of some process. For example, the calibration of conductivity generally requires more than one piece of information to be entered. The majority of the menu items in the software consist of the single variable format type. Any data that may be changed will be flashing. This flashing indicates user entry mode and is initiated by pressing the ENTER key. The UP arrow key will increase a flashing digit from 0 to 9. The LEFT arrow key moves the flashing digit from right to left. Once the change has been completed, pressing ENTER again stores the variable and stops the flashing. Pressing ESC aborts the change and also exits user entry mode. The starting (default) screen is always the Measure Menu. The UP arrow key is used to select the desired display. From anywhere in this section the user can press the MENU key to select one of the four Menu Sections. The UP arrow icon next to all list items on the display is a reminder to scroll through the list using the UP arrow key. To select a list item for modification, first select the proper menu with the MENU key. Scroll to the list item with the UP arrow key and then press the ENTER key. This tells the system that the user wishes to perform a change on that item. For single item type screens, once the user presses the ENTER key, part or all of the variable will begin to flash, indicating that the user may modify that variable using the arrow keys. However, if the instrument is locked, the transmitter will display the message Locked! and will not enter user entry mode. The instrument must be unlocked by entering the proper code value to allow authorized changes to user entered values. Once the variable has been reset, pressing the ENTER key again causes the change to be stored and the flashing to stop. The message Accepted! will be displayed if the change is within pre-defined variable limits. If the user decides not to modify the value after it has already been partially changed, pressing the ESC key aborts the modification and returns the entry to its original stored value. In a menu item which is a multiple variable sequence type, once the ENTER key is pressed there may be several prompts and sequences that are run to complete the modification. The ESC key can always be used to abort the sequence without changing any stored variables.
  • 29. ATI Q46CT Conductivity System Part 5 – Configuration 29 O&M Manual Rev-G (8/17) MEASURE CAL CONFIG DIAG or or or Cal Cond 1 PID 0% #1 Set Hold Cal Temp 1 PID 100% #1 1 PID Setpoint #1 1 PID Prop #1 Fault List Sim Out Fail Out #1 Fail Val #1 Fail Out #2 Offset Temperature 1 PID % Output Loop Current (#1) LIST ITEMS Cal Zeros Loop Current (#2) Set Default 4 Loop Current (#3) CONTROL Fail Val #2 Failsafe 1 PID Int #1 1 PID Deriv #1 Set 4mA (#1) Set 20mA (#1) or Entry Lock Set Delay Contrast Ref Temp I out 1 Mode I out 2 Mode Relay B Mode Temp Units Set 4mA (#2) Set 20mA (#2) TDS 5 Aux rly= 1 PID Timer Cell Factor MENU SECTIONS Software Version Main Display Zero Filter Slope Relay A Mode 4 I out 3 Mode Relay C Mode 5 Relay E Mode 5 Relay D Mode 5 Relay F Mode 4 Set 4mA (#3) 4 Set 20mA (#3) 6 Setpnt A (or A-HI, A-LO) 6 Hyst A (or A-HI, A-LO) 6 Delay A (or A-HI, A-LO) Phase A Setpnt B Hyst B Delay B Phase B Setpnt C Hyst C Delay C Phase C 5 Setpnt D 5 Hyst D 5 Delay D 5 Phase D 5 Setpnt E 5 Hyst E 5 Delay E 5 Phase E 5 Setpnt F 5 Hyst F 5 D elay F 5 Phase F 4 Fail Out #3 4 Fail Val #3 Backlight Notes: (1) If Relay A,B,C,D,E,F is set to FAIL mode, relay settings are not displayed in menu. (2) The annunciator for Relay C is shown in the MEASURE/ temperature display PID is enabled Linear Temp Comp Selected Table Temp Comp Selected Optional third 4-20 output installed Optional 3-relay card installed (D,E,F) If Relay A is set to ALARM mode, the settings are divided into 2 groups of HI and LO points. 1 2 3 4 5 6 7 If Comm Mode is set to a selection other than none, additional Comm menus will show. Start Delay TC Factor Cond Range TC Method 2 Lin T-Comp 3 TC Comp Tbl TDS Factor Comm Mode 7 Comm Address Figure 19 - Software Map
  • 30. ATI Q46CT Conductivity System Part 5 – Configuration 30 O&M Manual Rev-G (8/17) 5.22 Measure Menu [MEASURE] The default menu for the system is the display-only menu MEASURE. This menu is a display-only measurement menu, and has no changeable list items. When left alone, the instrument will automatically return to this menu after approximately 30 minutes. While in the default menu, the UP arrow allows the user to scroll through the secondary variables on the lower line of the display. A brief description of the fields in the basic transmitter version is as follows: TRANSMITTER MEAS SCREENS: 25.7°C Temperature display. Can be displayed in °C or °F, depending on user selection. A small “m” to the left side of the temperature indicates the transmitter has automatically jumped to a manual 25°C setting due to a failure with the temperature signal input. An L C symbol to the left of the temperature will appear if relay C activates. TDS = 200 ppm Total Dissolved Solids (TDS). Displays TDS of process. 100% 20.00 mA PID Status screen (if enabled.) Shows the present controller output level on left, and actual transmitter current on the right. The controller can be placed in manual while viewing this screen by pressing and holding the ENTER key for 5 seconds until a small flashing “m” appears on the screen. At that point the controller output can be adjusted up or down using the UP and LEFT arrow keys. To return to automatic operation, press and hold the ENTER key for 5 seconds and the “M” will disappear. #1 4.00 mA Analyzer output current # 1. #2 12.00 mA Analyzer output current # 2. #3 20.00 mA Analyzer output current # 3 (if option included.) Aux relay=D,E,F Auxilliary relay annunciators (if option included.) Slope The slope is an indication of the sensor output as compared to the theoretical output if the sensor were perfect. The slope value is calculated after completion of a calibration as described in Section 7 of this manual. Q46CT v4.00 Transmitter software version number. Note: A display test (all segments ON) can be actuated by pressing and holding the ENTER key while viewing the model/version number on the lower line of the display. The MEASURE screens are intended to be used as a very quick means of looking up critical values during operation or troubleshooting.
  • 31. ATI Q46CT Conductivity System Part 5 – Configuration 31 O&M Manual Rev-G (8/17) 5.23 Calibration Menu [CAL] The calibration menu contains items for frequent calibration of user parameters. There are five items in this list: Cal Cond, Cell Temp, Cal Const, and Cal Zeros. Cal Cond The conductivity calibration function allows the user to adjust the ttransmitter offset and span reading to match reference buffers, or to adjust the sensor offset to match the sample reading. See Part 7 - Calibration for more details. Cal Temp The temperature calibration function allows the user to adjust the offset of the temperature response by a small factor of ± 5°C. The temperature input is factory calibrated to very high accuracy. However, long cable lengths and junction boxes may degrade the accuracy of the temperature measurement in some extreme situations. Therefore, this feature is provided as an adjustment. See Part 7 - Calibration for more details. Cal Zeros This function calibrates all range zero-points to the specific sensor being used. This function is only required to be performed once at initial start- up or when the sensor has been replaced. See Part 7 - Calibration for more details. Cell Factor This function allows the user to directly enter the factory measured cell constant for the Q25CT sensor. When this feature is used, calibration with reference solutions is not necessary. The default setting is 1.00. Note that this value is actually calculated from an actual liquid calibration, so running a conductivity calibration will change this value. TC Factor This function allows the user to directly enter the factory measured temperature offset value marked on the sensor label. The default for this setting is 7.50 and most sensors are fairly close to that number. Cond Range This function allows the user to set the display range of the transmitter for a specific application. Once set, all output functions use this display range to establish configuration settings. The default range is 0-200.0 mS. Press ENTER to initiate user entry mode, and the value will flash. Use the arrow key to modify the range for the desired range and then press ENTER. 5.24 Configuration Menu [CONFIG] The Configuration Menu contains all of the general user settings: Entry Lock This function allows the user to lock out unauthorized tampering with instrument settings. All settings may be viewed while the instrument is locked, but they cannot be modified. The Entry Lock feature is a toggle- type setting; that is, entering the correct code will lock the transmitter and entering the correct code again will unlock it. The code is preset at a fixed value. Press ENTER to initiate user entry mode and the first digit will flash. Use arrow keys to modify value. See end of manual for the Q46CT lock/unlock code. Press ENTER to toggle lock setting once code is correct. Incorrect codes do not change state of lock condition.
  • 32. ATI Q46CT Conductivity System Part 5 – Configuration 32 O&M Manual Rev-G (8/17) Set Delay The delay function sets the amount of damping on the instrument. This function allows the user to apply a first order time delay function to the conductivity measurements being made. Both the display and the output value are affected by the degree of damping. Functions such as calibration are not affected by this parameter. The calibration routines contain their own filtering and stability monitoring functions to minimize the calibration timing. Press ENTER to initiate user entry mode, and the value will flash. Use the arrow keys to modify value; range is 0.1 to 9.9 minutes. Press ENTER to store the new value. Contrast This function sets the contrast level for the display. The custom display is designed with a wide temperature range, Super-Twist Nematic (STN) fluid. Press ENTER to initiate user entry mode, and the value will flash. Use arrow keys to modify the value; range is 0 to 8 (0 being lightest). Press ENTER to update and store the new value. Main Display This function allows the user to change the measurement in the primary display area, selecting between conductivity or Auto-Ranging conductivity. Press ENTER to initiate user entry mode, and the entire value will flash. Use the UP arrow key to modify the desired display value. Press ENTER to store the new value. Zero Filter This function forces the reading to zero when reading is below the entered value. For example, if the entered value were 2.1 mS, the display would read 0.0 mS when the measured value reached 2.0 mS. This feature is useful in blanking out very small and meaningless zero variability. Ref Temp The reference temperature function sets the basis point for the linear temperature compensation methods. In most cases this setting should be left at the default of 25.0 °C. Press ENTER to initiate user entry mode, and the entire value will flash. Use the arrow keys to modify the desired value; range is 0.0°C to 50.0°C. Press ENTER to update and store the new value. This setting appears in the Software Menu only if “Temp Mode” is set to Lin. TC Method This function allows selection of the method used for the temperature compensation in the instrument. The choices are 1-Lin or 2-Table. The Lin (linear) method is by far the most common and should be used unless special measurements are being made. Under the 2-Table selection, you have the ability to enter your own custom temperature compensation table. You can enter up to 6 multipliers from your own temperature compensation data if required for special applications. NOTE: Do not set the Temp Mode to a value other than Lin unless you are confident in the temperature data developed for your special application.
  • 33. ATI Q46CT Conductivity System Part 5 – Configuration 33 O&M Manual Rev-G (8/17) Lin T-Comp This function sets the correction slope value for the linear temperature compensation method and is used when the “Temp Mode” is set to Lin. Linear compensation is the method recommended for most aqueous solutions, and the value is typically 2.00 %/°C (25°C reference temperature) for neutral water. This is the factory default and it provides the best compensation for most aqueous solutions. Other typical ranges include: Acids: 1.0 to 1.6%/°C Bases: 1.8 to 2.0%/°C Salts: 2.2 to 3.0%/°C NOTE: If the temperature units are changed between °C and °F (see Temp Units in this Section), the default setting for this output will change between 2.00 %/°C and 1.11%/°F accordingly. Other compensation slopes for uncommon solutions may be found in chemical handbooks (such as the CRC). Press ENTER to initiate user entry mode, and the entire value will flash. Use the arrow keys to modify the desired value; entry range is 0.000%/°C (no compensation) to 4.000%/°C. Press ENTER to store the new value. TDS Factor This function sets the linear relationship of the TDS (total dissolved solids) reading to the conductivity measurement. The actual units for the slope are in mg/L/μS. The default value is 00.49 mg/L/μS. Press ENTER to initiate user entry mode, and the entire value will flash. Use the arrow keys to modify the desired value; range is 00.00 mg/L/μS to 99.99 mg/L/μS. Press ENTER to update and store the new value. Com Mode Sets digital communication mode of analyzer. Optional digital communication card must be plugged into the power supply slot for this function to work. Press ENTER to initiate user entry mode, and the entire value will flash. Use the UP arrow key to modify the desired value; selections include 1-None, 2- P-DP for Profibus DP, 3 – Modbus, 4 – Ethernet IP. Press ENTER to store the new value. Com Address Sets bus address for digital communication mode of analyzer. Optional digital communication card must be plugged into the power supply slot for this function to work. Press ENTER to initiate user entry mode, and the entire value will flash. Use the UP arrow key to modify the desired value. Range is 1-125. Press ENTER to store the value. Iout#1 Mode This assigns the 4-20 mA output #1 to either S (by selecting 1) or for PID output (by selecting 2) Iout#2 Mode This assigns the 4-20 mA output # 2 to Temperature (by selecting 1), S (by selecting 2) or TDS mg/L (by selecting 3) *Iout#3 Mode OPTIONAL. This function sets analog output #3 for temperature, conductivity, or TDS. Press ENTER to initiate user entry mode, and the entire value will flash. Use the UP arrow key to modify the desired value; selections include 1-C/F for temperature, 2-uS, or 3-mg/l. Press ENTER to store the new value.
  • 34. ATI Q46CT Conductivity System Part 5 – Configuration 34 O&M Manual Rev-G (8/17) Relay A Mode Relay A can be used in three different ways: as a setpoint control, as a fail alarm, or as a HI-LO alarm band. The three settings for Rly A Mode are CON, FAIL and AL. The CON setting enables normal control operation for Relay A, with settings for setpoint, hysteresis, delay and phasing appearing in the CONFIG menu automatically. See Fig. 12 for further details. The FAIL setting enables the fail alarm mode for Relay A. Relay A will then trip on any condition that causes the FAIL icon to be displayed on the LCD. Using this mode allows the User to send alarm indications to other remote devices. The AL setting allows two setpoints to be selected for the same relay, producing a HI-LO alarm band. In this mode, Relay A will trip inside or outside of the band, depending upon the Phase selected. See Figure 19 for further details. Relay B Mode Relay B and C can be used in two ways: as a setpoint control, or as an Relay C Mode alarm. The two settings for Relay B Mode are CON and FAIL. Relay B also is set up for 2 “clean modes”, but these modes are generally not used for toroidal conductivity. They are used to activate automatic sensor cleaning systems that are generally not required. The CON setting enables normal setpoint operation for Relay B/C. Relay B/C then operates identically to Relay A, with settings for setpoint, hysteresis, delay and phasing appearing in the CONFIG menu automatically. The FAIL setting enables the fail alarm mode for Relay B/C. Relay B/C will then trip on any condition that causes the FAIL icon to be displayed on the LCD. Note that the Relay C indicator shows up only on the lower screen of the display next to the temperature reading. This is because the default setting for relay C is the FAIL setting. Using this mode allows the User to send alarm indications to other remote devices. *Relay D Mode OPTIONAL. Relays D, E, and F can be used in two ways: as a setpoint *Relay E Mode control, or as an alarm. The two settings for Relay B Mode are CON and *Relay F Mode FAIL. The CON setting enables normal setpoint operation for Relay B. Relay B then operates identically to Relay A, with settings for setpoint, hysteresis, delay and phasing appearing in the CONFIG menu automatically. Temp Units This function sets the display units for temperature measurement. Press ENTER to initiate user entry mode, and the entire value will flash. Use the UP arrow key to modify the desired display value. The choices are °F and °C. Press ENTER to store the new value.
  • 35. ATI Q46CT Conductivity System Part 5 – Configuration 35 O&M Manual Rev-G (8/17) 5.25 Control Menu The Control Menu contains all of the output control user settings: Set PID 0% If the PID is enabled, this function sets the minimum and maximum Set PID 100% controller end points. Unlike the standard 4-20 mA output, the controller [Iout1=PID] does not “scale” output values across the endpoints. Rather, the endpoints determine where the controller would normally force minimum or maximum output in an attempt to recover the setpoint (even though the controller can achieve 0% or 100% anywhere within the range.) If the 0% point is lower than the 100% point, then the controller action will be “reverse” acting. That is, the output of the controller will increase if the measured value is less than the setpoint, and the output will decrease if the measured value is larger than the setpoint. Flipping the stored values in these points will reverse the action of the controller to “direct” mode. The entry value is limited to a value within the range specified in “Set Range”, and the 0% and the 100% point must be separated by at least 1% of this range Use the LEFT arrow key to select the first digit to be modified. Then use the UP and LEFT arrow keys to select the desired numerical value. Press ENTER to store the new value. PID Setpnt The measured value which the controller is attempting to maintain by [Iout1=PID] adjusting output value. It is the nature of the PID controller that it never actually gets to the exact value and stops. The controller is continually making smaller and smaller adjustments as the measured value gets near the setpoint. PID Prop Proportional gain factor. The proportional gain value is a multiplier on [Iout1=PID] the controller error (difference between measured value and setpoint value.) Increasing this value will make the controller more responsive. PID Int Integral is the number of “repeats-per-minute” of the action of the [Iout1=PID] controller. It is the number of times per minute that the controller acts on the input error. At a setting of 2.0 rpm, there are two repeats every minute. If the integral is set to zero, a fixed offset value is added to the controller (manual reset.) Increasing this value will make the controller more responsive. PID Deriv Derivative is a second order implementation of Integral, used to suppress [Iout1=PID] “second-order” effects from process variables. These variables may include items like pumps or mixers that may have minor impacts on the measured value. The derivative factor is rarely used and therefore it is best in most cases to leave it at the default value. Increasing this value will make the controller more responsive. Set 4 mA (#1) These functions set the main 4 and 20 mA current loop #1 output points Set 20 mA (#1) for the analyzer when output 1 is in normal, non-PID, mode of operation. [Iout1=conductivity] The value stored for the 4 mA point may be higher or lower than the value stored for the 20 mA point. The entry values are limited to values within the range specified in “Set Range”, and the 4 mA and the 20 mA point must be separated by at least 1% of this range Use the LEFT arrow
  • 36. ATI Q46CT Conductivity System Part 5 – Configuration 36 O&M Manual Rev-G (8/17) key to select the first digit to be modified. Then use the UP and LEFT arrow keys to select the desired numerical value. Press ENTER to store the new value.. *Set 4 mA (#2) These functions set the second 4 mA and 20 mA current loop output *Set 20 mA (#2) points for the transmitter. The output may be set to track temperature [temp/conductivity] (default), or conductivity. The values stored for the 4 mA point may be higher or lower than the value stored for the 20 mA point. The entry value is limited to a value between 0 and 110°C if it is set for temperature, within the range specified in “Set Range” if the output is set to track conductivity. The 4 mA and the 20 mA point must be at least 20 units away from each other. Press ENTER to initiate user entry mode, and the value will flash. Use arrow keys to modify value. Press ENTER to store the new value. *Set 4 mA #3 OPTIONAL These functions set the optional third 4 mA and 20 mA *Set 20 mA #3 current loop output points for the analyzer. The output may be set to [temp/uS/TDS] track temperature (default), conductivity, or TDS. The values stored for the 4 mA point may be higher or lower than the value stored for the 20 mA point. The range is between 0 and 110°C if it is set for temperature, within the range specified in “Set Range” if the output is set to track conductivity, and must be within the display range established by the conductivity range if set to track TDS. The 4 mA and the 20 mA point must be at least 20 units away from each other. Press ENTER to initiate user entry mode, and the value will flash. Use arrow keys to modify value. Press ENTER to store the new value. *A Setpoint This function establishes the conductivity trip point for relay A. The entry value is limited to a value within the range specified in “Set Range”. Use the LEFT arrow key to select the first digit to be modified. Then use the UP and LEFT arrow keys to select the desired numerical value. Press ENTER to store the new value. *A Hysteresis This function establishes the hysteresis, or “deadband”, for Relay A. Hysteresis is most often used to control relay chattering; however, it may also be used in control schemes to separate the ON/OFF trip points of the relay. Press ENTER to initiate user entry mode, and the value will flash. Use the arrow keys to modify value. Press ENTER to store the new value. *A Delay This function places an additional amount of time delay on the trip point for relay A. This delay is in addition to the main delay setting for the controller. The entry value is limited to a value between 0 and 999 seconds. Press ENTER to initiate user entry mode, and the value will flash. Use arrow keys to modify value; range is 0 to 999 seconds. Press ENTER to store the new value. *A Phasing This function establishes the direction of the relay trip. When phase is HI, the relay operates in a direct mode. Therefore, the relay energizes and the LCD indicator illuminates when the conductivity value exceeds the setpoint. When the phase is LO, the relay energizes and the LCD indicator illuminates when the conductivity level drops below the setpoint. The failsafe setting does have an impact on this logic. The
  • 37. ATI Q46CT Conductivity System Part 5 – Configuration 37 O&M Manual Rev-G (8/17) description here assumes the failsafe setting is OFF. Press ENTER to initiate user entry mode, and the entire value will flash. Use the UP arrow key to modify the desired value; selections include HI for direct operation or LO for reverse operation. Press ENTER to store the new value. See Figure 20 below for a visual description of a typical control relay application. Figure 20 - Control Relay Example If Relay A Mode is set to Alarm Mode, AL, then the following settings will appear in the Config Menu list automatically. In this mode, two setpoints can be selected on the same relay, to create an alarm band. Phase HI selection causes the relay to energize outside of the band, and Phase LO causes the relay to energize inside of the band. This feature enables one relay to be used as a control relay while the other is used as a HI-LO Alarm relay at the same time. Setpoint A-LO must be set lower than Setpoint A-HI. When AL mode is first selected, Setpoint A-LO is defaulted to 0. Figure 21 is a visual description of a typical alarm relay application. Figure 21 - Alarm Relay Example When value rises to ≥ 1.000 mS, relay closes. When value falls to ≤ 1.000 mS, relay opens. When value rises to ≥ 1.000 mS, relay closes, until value falls back to ≤ 0.950 mS. Settings: When value falls to ≤ 1.000 µS, relay closes, until value rises back to ≥ 1.050 µS. When value rises to ≥ 0.500 µS, relay closes, until value falls back to ≤ 0.450 µS. When value falls to ≤ 0.500 mS, relay closes, until value rises back to ≥ 0.550 mS. *Setpnt A-HI *Hyst A-HI *Delay A-HI *Setpnt A-LO *Hyst A-LO *Delay A-LO When value rises to ≥ 1.000 mS, relay opens. When value falls to ≤ 1.000 mS, relay closes. } 1.000 mS 0.950 mS PHASE: HI ON HYSTERESIS OR “DEAD BAND” X OFF } 1.050 mS 1.000 mS PHASE: LO OFF HYSTERESIS OR “DEAD BAND” X ON } 1.000 mS 0.950 mS PHASE: HI HYST-HI X ON 0.550 mS 0.500 mS OFF X }HYST-LO ON } 1.050 mS 1.000 mS PHASE: LO HYST-HI X OFF 0.500 mS 0.450 mS ON X }HYST-LO OFF Settings: Setpoint: 1.000 mS Hyst: 0.050 Delay: 000 Failsafe: OFF Setpoint A-HI: 1.000 mS Setpoint A-LO: 0.500 mS Hyst A-HI: 0.050 Hyst A-LO: 0.05 Delay A-HI: 000 Delay A-LO: 000
  • 38. ATI Q46CT Conductivity System Part 5 – Configuration 38 O&M Manual Rev-G (8/17) If Relay B Mode is set to CON (see Relay B Mode), then Relay B will function identically to Relay A. Relay B settings appear in the CONFIG menu list automatically. If Relay C Mode is set to CON (see Relay C Mode), then Relay C will function identically to Relay A. Relay C settings appear in the CONFIG menu list automatically. If optional Relay D, E, or F Mode is set to CON (see Relay D,E,F Modes), then the Relay will function identically to Relay A. Relay settings appear in the CONFIG menu list automatically. 5.26 Diagnostics Menu [DIAG] The diagnostics menu contains all of the user settings that are specific to the system diagnostic functions, as well as functions that aid in troubleshooting application problems. Set Hold The Set Hold function locks the current loop output values on the present process value, and halts operation of the PID controller. This function can be used prior to calibration, or when removing the sensor from the process, to hold the output in a known state. Once HOLD is released, the outputs return to their normal state of following the process input. The transfer out of HOLD is bumpless on the both analog outputs - that is, the transfer occurs in a smooth manner rather than as an abrupt change. An icon on the display indicates the HOLD state, and the HOLD state is retained even if power is cycled. Press ENTER to initiate user entry mode, and entire value will flash. Use the UP arrow key to modify the desired value, selections are ON for engaging the HOLD function, and OFF to disengage the function. Press ENTER to store the new value. Note - the Set Hold function holds BOTH current levels, as well as ALL relay settings. The Set Hold function can also hold at an output value specified by the user. To customize the hold value, first turn the HOLD function on. Press the ESC key to go to the DIAG Menu and scroll to Sim Output using the UP arrow key. Press ENTER. Follow the instructions under Sim Output (see following page). CAUTION: There is no time-out on the hold feature. Once placed into hold mode, return to normal operation must be done manually. Fault List The Fault List screen is a read-only screen that allows the user to display the cause of the highest priority failure. The screen indicates the number of faults present in the system and a message detailing the highest priority fault present. Note that some faults can result in multiple displayed failures due to the high number of internal tests occurring. As faults are corrected, they are immediately cleared. *B Setpoint *B Hysteresis *B Delay *B Phasing C Setpoint C Hysteresis C Delay C Phasing D,E,F Setpoint D,E,FHyster D,E,F Delay D,E,F Phasing
  • 39. ATI Q46CT Conductivity System Part 5 – Configuration 39 O&M Manual Rev-G (8/17) Faults are not stored; therefore, they are immediately removed if power is cycled. If the problem causing the faults still exists, however, faults will be displayed again after power is re-applied and a period of time elapses during which the diagnostic system re-detects them. The exception to this rule is the calibration failure. When a calibration fails, no corrupt data is stored. Therefore, the system continues to function normally on the data that was present before the calibration was attempted. After 30 minutes or if power to the analyzer is cycled, the failure for calibration will be cleared until calibration is attempted again. If the problem still exists, the calibration failure will re-occur. Press ENTER to initiate view of the highest priority failure. The display will automatically return to normal after a few seconds. PID Timer This function sets a timer to monitor the amount of time the PID controller remains at 0% or 100%. This function only appears if the PID controller is enabled. If the timer is set to 0000, the feature is effectively disabled. If the timer value is set to any number other zero, a FAIL condition will occur if the PID controller remains at 0% or 100% for the timer value. If one of the relays is set to FAIL mode, this failure condition can be signaled by a changing relay contact. Press ENTER to initiate user entry mode, and the entire value will flash. Use the UP arrow key to modify desired value; range of value is 0-9999 seconds. Press ENTER to store the new value. Sim Out The Sim Out function allows the user to simulate the chlorine level of the instrument in the user selected display range. The user enters a ppm value directly onto the screen, and the output responds as if it were actually receiving the signal from the sensor. This allows the user to check the function of attached monitoring equipment during set-up or troubleshooting. Escaping this screen returns the unit to normal operation. Press ENTER to initiate the user entry mode, and the right- most digit of the value will flash. Use arrow keys to modify desired value. The starting display value in SIM mode will be the last read value of the input. The output will be under control of the SIM screen until the ESC key is pressed. The instrument will automatically terminate the simulated output after 30 minutes and return to normal operation unless the “HOLD” function is engaged. Note: If the HOLD function is engaged before the Sim Output function is engaged, the simulated output will remain the same even when the ESC key is pressed. Disengage the HOLD function to return to normal output. Fail Out #1 This function enables the user to define a specified value that the main current output will go to under fault conditions. When enabled to ON, the output may be forced to the current value set in Fail Val (next item.) With the Fail Out setting of ON, and a Fail Val setting of 6.5 mA, any alarm condition will cause the current loop output to drop outside the normal operating range to exactly 6.5 mA, indicating a system failure that requires attention. Press ENTER to initiate user entry mode, and the entire value will flash. Use the UP arrow key to modify desired value; selections are ON, OFF. Press ENTER to store the new value.
  • 40. ATI Q46CT Conductivity System Part 5 – Configuration 40 O&M Manual Rev-G (8/17) Fail Val #1 Sets the output failure value for Iout#1. When Fail Out above is set to ON, this function sets value of the current loop under a FAIL condition. The output may be forced to any current value between 4-20 mA. Press ENTER to initiate user entry mode, and the entire value will flash. Use the UP arrow key to modify desired value; selections are between 4mA, and 20mA. Press ENTER to store the new value. Fail Val #2 This function sets the value of current loop output #2 under a FAIL condition. The settings and operation are identical to Fail Out for output #1. *Fail Out #3 OPTIONAL. This function sets the fail-mode of current loop output #3 under a FAIL condition. The settings and operation are identical to Fail Out for output #1. *Fail Val #3 OPTIONAL. This function sets the value of current loop output #3 under a FAIL condition. The settings and operation are identical to Fail Out for output #1. Backlight This function has three options. ON – On all the time, OFF – Off all the time, AL – Alarm (Default). This function flashes the backlight on and off whenever the Fail icon is displayed. Failsafe This function allows the user to set the optional system relays to a failsafe condition. In a failsafe condition, the relay logic is reversed so that the relay is electrically energized in a normal operating state. By doing this, the relay will not only change state when, for example, a chlorine limit is exceeded, but also when power is lost to the controller. When failsafe is selected to be ON, the normally-open contacts of the relay will be closed during normal operation. In an attempt to make this configuration less confusing, the LCD icon logic is reversed with this setting, and the icon is OFF under this normal condition. Therefore, when the trip condition occurs, the closed N.O. contacts will be opened (relay de-energized), and the LCD icon will illuminate. In addition, a power fail would also cause the contacts to open. Start Delay This function is designed to minimize control or alarm issues arising from temporary power loss. When power goes down, the monitor records the analog output values and the status of relays and PID functions. When power is restored, the analog values and relays will be held at the pre- power loss values for a defined period of time. This “start delay” may be programmed for periods from 0-9.9 minutes. This function is set to 0.0 minutes by default and must be activated by the user if desired by setting a positive time value Set Default The Set Default function allows the user to return the instrument back to factory default data for all user settings or for just the calibration default. It is intended to be used as a last resort troubleshooting procedure. All user settings or the calibration settings are returned to the original factory values. Hidden factory calibration data remains unchanged. Press ENTER to initiate user entry mode and select either All or CAL with the UP arrow key.
  • 41. 41 O&M Manual Rev-G (8/17) Part 6 – Calibration 6.1 Overview and Methods Calibration of the Q46CT is required to accurately match the sensor characteristics to the monitor/analyzer. Since the output of the conductivity sensor does not degrade over time, it is typically only required that the sensor be calibrated at initial installation and then cleaned periodically to maintain proper system accuracy. It is important for the user to establish a periodic cleaning and calibration-check schedule for sensor maintenance to maintain high system accuracy. Since the conductivity of a solution is greatly affected by temperature, proper settings for thermal compensation are critical for accurate operation. Before calibrating the instrument for the very first time, it is important to select the proper operating parameters in the configuration menus for temperature compensation methods. When using conductivity calibration standards for a wet calibration, take care not to inadvertently contaminate the reference solution; always thoroughly clean the sensor, rinsing off in tap water, and then finish rinsing in pure or de-ionized water. In addition, note that calibration solutions less than 200 μS or greater than 100 mS can be very unstable. Moving the sensor back and forth between different value conductivity reference solutions can quickly contaminate the solutions and render them inaccurate. 6.11 1-Point Cal Explained The 1-point calibration method is generally known as the "grab sample" calibration method. In the 1-point calibration method, the sensor may be removed from the application and placed into a reference solution. It may also be left in the measurement process and calibrated by reference. The 1-point calibration adjusts the sensor slope to match the exact calibration point. Readings beyond that point are then extrapolated from the determined slope of the calibration line. Since the sensor slope does not degrade over time, frequent re-calibration is unnecessary. Calibration accuracy can be optimized by calibrating with a reference solution which is close to the values typically measured. 6.12 Zero Cal Cal Explained The sensor offset must be set for the system only on initial sensor installation, or when the cable length has been altered. The Zero Cal method establishes all of the sensor offset points for the instrument’s 6 ranges of operation. 6.2 Performing Sensor Zero Cal The sensor offset MUST be set for the system on initial sensor installation, or when the cable length has been altered. However, it can easily be adjusted at any time by re-calibrating the sensor in air. The sensor zero-calibration generally has little effect in measurements above about 50 mS, but it can have a significant effect in measurements below about 1 mS. If the sensor zero cal is to be performed, it must be done BEFORE the 1-point reference calibration. To begin the sensor zero cal, verify that the sensor is connected and clean and dry. It should be placed in the air with the electrodes at least 1 foot away from any nearby objects. Holding it is not recommended – place on table or just hang.
  • 42. ATI Q46CT Conductivity System Part 6 – Calibration 42 O&M Manual Rev-G (8/17) Procedure 1. Remove sensor from process and clean thoroughly. Dry sensor and position on table or hang in air (in air is best.) If on table, let end of sensor hang over edge of table. 2. Scroll to the CAL menu section using the MENU key and press ENTER or the UP arrow key. Scroll to the menu Zero Cal. 3. Press the ENTER key. The screen will prompt the user to postion the sensor in air. 4. Press the ENTER key. The screen will automatically scroll through all ranges and establish and store the proper zero points. 6.3 Performing 1-Point Cal This calibration method is intended to be used as an on-line calibration method or a wet-cal with reference solutions. During calibration, the system will display the current conductivity reading, and the user can manually enter a reference value from a reference solution bottle or a comparative reference instrument. For wet calibrations, the user may use pre-made calibration references (also available from ATI) or a NaCl solution may be made using pure, dried NaCl crystals and one liter of high purity, de- ionized, CO2-free water as mixed in the table shown in Figure 22 - NaCl Reference Solution Calibration Chart All table data is at 25°C - therefore, the sensor must be at this temperature to calibrate properly using the table data. If another reference calibration solution is being used, be sure to note temperature of reference solution before calibration. Since the sensor must ideally be at the specified temperature, wet calibrations can be difficult to perform accurately. NaCl Reference Solution for Calibration (25°C) μS/cm NaCl (gm) 100 0.05 200 0.10 500 0.25 1000 0.50 2000 1.01 3000 1.53 4000 2.06 5000 2.61 8000 4.34 10000 5.56 20000 11.59 Figure 22 - NaCl Reference Solution Calibration Chart
  • 43. ATI Q46CT Conductivity System Part 6 – Calibration 43 O&M Manual Rev-G (8/17) During the 1-point calibration, the system will automatically pick the correct range for the calibration reference if the Q46CT is in the AUTO range. If the Q46CT is in a normal display mode, the user must be careful to calibrate with a solution that falls into the manual range selected. If the calibration solution is outside the manual range, an error will result. Procedure 1. If a zero calibration on the sensor is also to be performed, that must be done FIRST. The zero calibration process can have an impact on the result of the 1-point calibration. So if a zero cal is required, do that procedure and return here. 2. Determine whether the calibration will be done on-line or with the sensor removed and placed into a reference solution. If the sensor is removed from the application, rinse and clean. 3. If the sensor has been removed and placed into a solution, allow the sensor to temperature equilibrate with the solution as much as possible. With the sensor coming from an application that differs greatly in temperature, the user may have to wait as much as 20 minutes. If the sensor is on-line, the user may want to set the output HOLD feature prior to calibration to lock out any output fluctuations. 4. Scroll to the CAL menu section using the MENU key and press ENTER or the UP arrow key. Scroll until Cal Cond is displayed. Press ENTER. 5. The screen will prompt the user to place the sensor into the reference solution (ideally this has already been done to achieve temperature equilibrium.) Once sensor is ready, press ENTER. 6. The system now begins acquiring data for the calibration value. As data is gathered, the units for conductivity and temperature may flash. Flashing units indicate that this parameter is unstable. The calibration data point acquisition will stop only when the data remains stable for a pre- determined amount of time. This can be overridden by pressing ENTER. If the data remains unstable for 10 minutes, the calibration will fail and the message Cal Unstable will be displayed. 7. The screen will display the last measured conductivity value and a message will be displayed prompting the user for the reference value. The user must then modify the screen value with the arrow keys and press ENTER. The system then performs the proper checks. 8. If accepted, the screen will display the message PASS with the slope value, then it will return to the main measurement display. If the calibration fails, a message indicating the cause of the failure will be displayed and the FAIL icon will be turned on. 6.4 Temperature Calibration The temperature input is factory calibrated for the highest accuracy. Temperature calibration is not recommended; however, it is provided for applications in which very long cable lengths are needed. For example, at 50 feet, readings may be off ± 0.2°C. The temperature calibration sequence is essentially a 1-point offset calibration that allows adjustments of approximately ± 5°C. The sensor temperature may be calibrated on line, or the sensor can be removed from the process and placed into a known solution temperature reference. In any case, it is critical that the sensor be allowed to reach temperature equilibrium with the solution in order to provide the highest accuracy. When moving the sensor between widely different temperature conditions, it may be necessary to allow as much as one hour before the calibration sequence is initiated. If the sensor is on-line, the user may want to set the output HOLD (see Section 5.26) feature prior to calibration to lock out any output fluctuations.
  • 44. ATI Q46CT Conductivity System Part 6 – Calibration 44 O&M Manual Rev-G (8/17) Procedure 1. Scroll to the CAL menu section using the MENU key and press ENTER or the UP arrow key. 2. Press the UP arrow key until Cal Temp is displayed. 3. Press the ENTER key. The message Place sensor in solution then press ENTER will be displayed. Move the sensor into the calibration reference (if it hasn’t been moved already) and wait for temperature equilibrium to be achieved. Press ENTER to begin the calibration sequence. 4. The calibration data gathering process will begin. The message Wait will flash as data is accumulated and analyzed. The °C or °F symbol may flash periodically if the reading is too unstable. 5. The message Adjust temp value then press ENTER will be displayed, and the right-most digit will begin to flash, indicating that the value can be modified. Using the UP and LEFT arrow keys, modify the value to the known ref solution temperature. Adjustments up to ± 5 °C from the factory calibrated temperature are allowed. Press ENTER. 6. Once completed, the display will indicate PASS or FAIL. If the unit fails, the temperature adjustment may be out of range, the sensor may not have achieved complete temperature equilibrium, or there may be a problem with the temperature element. In the event of calibration failure, it is recommended to attempt the calibration again immediately.
  • 45. 45 O&M Manual Rev-G (8/17) Part 7 – PID Controller Details 7.1 PID Description PID control, like many other control schemes, are used in chemical control to improve the efficiency of chemical addition or control. By properly tuning the control loop that controls chemical addition, only the amount of chemical that is truly required is added to the system, saving money. The savings can be substantial when compared to a system which may be simply adding chemical at a constant rate to maintain some minimal addition under even the worst case conditions. The PID output controller is highly advantageous over simple control schemes that just utilize direct (proportional only) 4-20 mA output connections for control, since the PID controller can automatically adjust the “rate” of recovery based on the error between the setpoint and the measured value – which can be a substantial efficiency improvement.. The PID controller is basically designed to provide a “servo” action on the 4-20 mA output to control a process. If the user requires that a measured process stay as close as possible to a specific setpoint value, the controller output will change from 0% to 100% in an effort to keep the process at the setpoint. To affect this control, the controller must be used with properly selected control elements (valves, proper chemicals, etc.) that enable the controller to add or subtract chemical rapidly enough. This is not only specific to pumps and valves, but also to line sizes, delays in the system, etc. This section is included to give a brief description of tuning details for the PID controller, and is not intended to be an exhaustive analysis of the complexities of PID loop tuning. Numerous sources are available for specialized methods of tuning that are appropriate for a specific application. 7.2 PID Algorithm As most users of PID controllers realize, the terminology for the actual algorithm terms and even the algorithms themselves can vary between different manufacturers. This is important to recognize as early as possible, since just plugging in similar values from one controller into another can result in dramatically different results. There are various basic forms of PID algorithms that are commonly seen, and the implementation here is the most common version; The ISA algorithm (commonly referred to as the “ideal” algorithm.)           dt t de D t d t e I t e P output ) ( ) ( ) ( 1 ) ( Where: output = controller output P = proportional gain I = integral gain D = derivative gain t = time e(t) = controller error (e=measured variable – setpoint) Figure 23 - PID Equation
  • 46. ATI Q46CT Conductivity System Part 7 – PID Controller Details 46 O&M Manual Rev-G (8/17) The most notable feature of the algorithm is the fact the proportional gain term affects all components directly (unlike some other algorithms - like the “series” form.) If a pre-existing controller utilizes the same form of the algorithm shown above, it is likely similar settings can for made if the units on the settings are exactly the same. Be careful of this, as many times the units are the reciprocals of each other (i.e. reps-per-min, sec-per-rep.) PID stands for “proportional, integral, derivative.” These terms describe the three elements of the complete controller action, and each contributes a specific reaction in the control process. The PID controller is designed to be primarily used in a “closed-loop” control scheme, where the output of the controller directly affects the input through some control device, such as a pump, valve, etc. Although the three components of the PID are described in the setting area (section 4.25), here are more general descriptions of what each of the PID elements contribute to the overall action of the controller. P Proportional gain. With no “I” or “D” contribution, the controller output is simply a factor of the proportional gain multiplied by the input error (difference between the measured input and the controller setpoint.) Because a typical chemical control loop cannot react instantaneously to a correction signal, proportional gain is typically not efficient by itself – it must be combined with some integral action to be useful. Set the P term to a number between 2-4 to start. Higher numbers will cause the controller action to be quicker. I Integral gain. Integral gain is what allows the controller to eventually drive the input error to zero – providing accuracy to the control loop. It must be used to affect the accuracy in the servo action of the controller. Like proportional gain, increasing integral gain results in the control action happening quicker. Set the I term to a number between 3-5 to start (1-2 more than P). Like proportional gain, increasing the integral term will cause the controller action to be quicker. D Derivative gain. The addition of derivative control can be problematic in many applications, because it greatly contributes to oscillatory behavior. In inherently slow chemical control process’, differential control is generally added in very small amounts to suppress erratic actions in the process that are non-continuous, such as pumps and valves clicking on and off. However, as a starting point for chemical process control, its best to leave the “D” term set to 0. Based on these descriptions, the focus on tuning for chemical applications really only involves adjustment of “P” and “I” in most cases. However, increasing both increases the response of the controller. The difference is in the time of recovery. Although combinations of high “P’s” and low “I” will appear to operate the same as combinations of low “P’s” and high “I’s”, there will be a difference in rate of recovery and stability. Because of the way the algorithm is structured, large “P’s” can have a larger impact to instability, because the proportional gain term impacts all the other terms directly. Therefore, keep proportional gain lower to start and increase integral gain to achieve the effect required. Many of the classical tuning techniques have the user start with all values at 0, and then increase the P term until oscillations occur. The P value is then reduced to ½ of the oscillatory value, and the I term is increased to give the desired response. This can be done with the Q46CT controller, with the exception that the I term should start no lower than 1.0. If it appears that even large amounts of integral gain (>20) don’t appreciably increase the desired response, drop I back to about 1.0, and increase P by 1.00, and start increasing I again. In most chemical control schemes, I will be approximately 3 times the value of P.
  • 47. ATI Q46CT Conductivity System Part 7 – PID Controller Details 47 O&M Manual Rev-G (8/17) 7.3 Classical PID Tuning Unlike many high speed position applications where PID loops are commonly used, the chemical feed application employed by this instrument does not require intense mathematical exercise to determine tuning parameters for the PID. In fact, the risk of instability is far greater with overly tuned PID control schemes. In addition, many of the classical mathematical exercises can be damaging or wasteful in the use of chemicals when the process is bumped with large amounts of input error to seek a response curve. Because of this, the general adjustment guidelines described in section 7.2 are sufficient for almost all application tuning for this instrument. Beyond this, many sources are available for classical tuning methods. 7.4 Manual PID Override Control The Q46 PID output function allows the user to take manual control of the PID control signal. This is often useful when starting up a control loop, or in the event that you wish to bump the system manually to measure system response time. To access the manual PID control, you must be in the MEASURE mode of operation and you must have the PID output displayed on the lower line. This line will indicate “XX.X% XX.X mA” with the X values simply indicating the current values. With this display on the screen, press and hold the ENTER key for about 5 seconds. You will see a small “m” show up between the % value and the mA value. This indicates you are now in manual mode. Once in manual, you may increase the PID output by pressing the UP arrow or you may decrease the output by pressing the LEFT arrow. This will allow you to drive the PID output to any desired setting. To revert to normal PID control, press and hold the ENTER key again until the “m” indicator disappears. 7.5 Common PID Pitfalls The most common problem occurring in PID control applications involves the false belief that proper settings on only the PID controller can balance any process to an efficient level. Close-loop control can only be effective if all elements in the loop are properly selected for the application, and the process behavior is properly understood. Luckily, the nature of simple chemical control process’ are generally slow in nature. Therefore, even a de-tuned controller (one that responds somewhat slowly) can still provide substantial improvements to setpoint control. In fact, damaging oscillatory behavior is far more likely in tightly tuned controllers where the user attempted to increase response too much. When deciding on a PID control scheme, it is important to initially review all elements of the process. Sticking valves, undersized pumps, or delays in reaction times associated with chemical addition can have a dramatic effect on the stability of the control loop. When controlling a chemical mix or reaction, the sensor should be placed in a location that ensures proper mixing or reaction time has occurred. The easiest processes to control with closed-loop schemes are generally linear, and symmetrical, in nature. For example, controlling level in tank where the opening of valve for a fixed period of time corresponds linearly to the amount that flows into a tank. Chemical control process’ can be more problematic when the nature of the setpoint value is non-linear relative to the input of chemical added. For example, pH control of a process may appear linear only in a certain range of operation, and become highly exponential at the extreme ranges of the measuring scale. In addition, if a chemical process is not symmetrical, that means it responds differentially to the
  • 48. ATI Q46CT Conductivity System Part 7 – PID Controller Details 48 O&M Manual Rev-G (8/17) addition and subtraction of chemical. It is important in these applications to study steady-state impact as well as step-change impact to process changes. In other words, once the process has apparently been tuned under normal operating conditions, the user should attempt to force a dramatic change to the input to study how the output reacts. If this is difficult to do with the actual process input (the recommended method), the user can place the control in manual at an extreme control point such as 5% or 95%, and release it in manual. The recovery should not be overly oscillatory. If so, the loop needs to be de-tuned to deal with that condition (reduce P and/or I.)
  • 49. 49 O&M Manual Rev-G (8/17) Part 8 – System Maintenance 8.1 System Checks 1. If the FAIL icon is flashing on the display, check the Fault List to determine the cause of the failure. To access the Fault List, press the MENU/ESC key until the DIAG menu appears. Then press the UP arrow key until the Fault List appears. Press the ENTER key to access the Fault List, and the highest priority fault message will be displayed. For a list of all messages and possible causes/solutions, refer to Section 9.5. 2. In ALL environments, connect an earth ground jumper to earth terminal connection on transmitter. 3. Perform a one-point calibration prior to sensor installation. 4. Check sensor cable color to terminal strip markings. 5. For highly unstable behavior, remove sensor from the process and measure the process solution in a plastic beaker. If the reading now stabilizes, place wire in beaker solution and actual process solution to determine if a ground loop exists. 6. If the instrument 4-20 mA output is connected into other control systems, disconnect output loop from system load and run through a handheld DMM to monitor current. Verify that the system operates correctly in this mode first. 8.2 Instrument Checks 1. Remove sensor completely and connect 1100 Ohms from the GREEN to BLACK (green bundle) on the analyzer input leads. The temperature reading should display approximately 25 °C and the conductivity reading should display approximately 0.0 uS. 2. Reconnect the sensor and leave dry in air. With a DMM, measure the AC voltage between RED and BLACK sensor leads. The DMM should read between 100 mVrms and 300 mVrms @ about 10 kHz. The display should show some value close to 0 uS if the sensor has been properly zero calibrated. 8.3 Sensor Tests Toroidal sensors can be tested with a digital voltmeter (DVM) to determine if a major sensor problem exists. Follow the steps below to verify sensor integrity: A. Disconnect the nine sensor wires from the terminal strip on the transmitter. Check sensor plastic covering for any mechanical damage to plastic covering. B. Connect a DVM between the RED and BLACK wires in the red jacket pair. With the DVM set to measure resistance, you should measure between 0.4 and 2.0 Ohms. C. Connect a DVM between the WHITE and BLACK wires in the white jacket pair. With the DVM set to measure resistance, you should measure between 0.4 and 2.0 Ohms.
  • 50. ATI Q46CT Conductivity System Part 8 – System Maintenance 50 O&M Manual Rev-G (8/17) D. Connect a DVM between the WHITE wire from the white jacket pair, and the RED wire from the red jacket pair. With the DVM set to measure resistance, you should measure an open circuit. E. Connect a DVM between the GREEN and BLACK wires in the green jacket pair. These are the RTD leads, and you should find a resistance value that depends on the temperature. The table below lists the resistance values for various temperatures. Temperature °C Resistance  0 1000 5 1019 10 1039 15 1058 20 1078 25 1097 30 1117 35 1136 40 1155 45 1175 50 1194 Figure 24 - Pt100 RTD Resistance Table 8.4 Display Messages The Q46 Series instruments provide a number of diagnostic messages that indicate problems during normal operation and calibration. These messages appear as prompts on the secondary line of the display or as items on the Fault List . The following messages will appear as prompts: MESSAGE DESCRIPTION POSSIBLE CORRECTION Cal Unstable Calibration problem, data too unstable to calibrate. Clean sensor, get fresh cal solutions, allow temperature and conductivity readings to fully stabilize, do not handle sensor or cable during calibration. Slope HIGH Sensor slope from calibration is greater than 400%. Get fresh cal solutions, allow temperature and conductivity readings to fully stabilize, check for correct buffer values Slope LOW Sensor slope from calibration is less than 20%. Clean sensor, get fresh cal solutions, allow temperature and conductivity readings to fully stabilize, check for correct buffer values. Offset HIGH Sensor offset from calibration is too high Clean sensor, get fresh cal solutions, allow temperature and conductivity readings to fully stabilize, check for correct solution values.
  • 51. ATI Q46CT Conductivity System Part 8 – System Maintenance 51 O&M Manual Rev-G (8/17) Out of Range Input value is outside selected range of the specific list item being configured. Check manual for limits of the function to be configured. Locked! Transmitter security setting is locked. Enter security code to allow modifications to settings. Unlocked! Transmitter security has just been unlocked. Displayed just after security code has been entered. TC-F25 lock! The TC won’t calibrate because there is something wrong with the connection to the temperature element in the sensor. Perform Sensor Tests and Instrument Tests to confirm operation of TC in sensor. Check all connections between sensor and instrument. The following messages will appear as items on the Fault List: MESSAGE DESCRIPTION POSSIBLE CORRECTION Sensor High The raw signal from the sensor is too high. Check wiring connections to sensor. Sensor Low The raw signal from the sensor is too low. Check wiring connections to sensor. Cond too High The conductivity reading is > 2000 mS. The conductivity reading is over operating range limits. Move to a higher range or select “Auto” from Main Display in CONFIG menu. Temp High The temperature reading is > 210 ºC. The temperature reading is over operating limits. Check wiring and expected temp level. Perform RTD test as described in sensor manual. Recalibrate sensor temperature element if necessary. Temp Low The temperature reading is < -10 ºC The temperature reading is under operating limits. Check wiring and expected temp level. Perform RTD test as described in sensor manual. Recalibrate sensor temperature element if necessary. TC Error TC may be open or shorted. Check sensor wiring and perform RTD test as described in sensor manual. Cond Cal Fail Failure of conductivity calibration. Clean sensor, get fresh cal solutions and redo calibration. Verify conductivity value of application or reference solutions. Perform sensor tests as described in sensor manual. Replace sensor if still failure. EEprom Fail Internal nonvolatile memory failure System failure, consult factory.
  • 52. ATI Q46CT Conductivity System Part 8 – System Maintenance 52 O&M Manual Rev-G (8/17) Chcksum Fail Internal software storage error. System failure, consult factory. Display Fail Internal display driver fail. System failure, consult factory. Range Cal Fail Failure of factory temperature calibration. Consult factory.
  • 53. 53 O&M Manual Rev-G (8/17) Spare Parts Part No. Description 03-0396 Q46CT Front Lid Assembly 07-0334 Q46CT Monitor, 100-240 VAC 07-0335 Q46CT Monitor, 12-24 VDC 07-0336 Q46CT Monitor, 100-240 VAC with Profibus 07-0337 Q46CT Monitor, 12-24 VDC with Profibus 03-0407 Q46 P/S Assy, 100-240VAC 03-0408 Q46 P/S Assy, 100-240VAC with 3 rd 4-20mA output 03-0409 Q46 P/S Assy, 100-240VAC with 3 relay exp. board 03-0410 Q46 P/S Assy, 12-24 VDC 03-0411 Q46 P/S Assy, 12-24 VDC with 3 rd 4-20mA output 03-0412 Q46 P/S Assy, 12-24 VDC with 3 relay exp. Board 38-0072 Terminal block plug, 3 position (relays) 38-0073 Terminal block plug, 4 position (outputs) 38-0074 Terminal block plug, 3 position (ground) 38-0081 Terminal block plug, 3 position (power) 38-0084 Terminal block plug, 3 position (power) – VDC version* *NOTE: prior to Dec 2018, VDC (power) Terminal block used the (38-0081) 23-0029 Fuse, 630mA, 250V, TR-5 (for AC and DC Analyzers) 07-0100 Junction Box 63-0099 Toroidal Sensor, Noryl, 20 ft. cable 31-0068 Sensor Interconnect cable 09-0046 Conductivity Standard – 84 microSiemens, 500 mL 09-0047 Conductivity Standard – 447 microSiemens, 500 mL 09-0048 Conductivity Standard – 1,500 microSiemens, 500 mL 09-0049 Conduvtivity Standard – 8,974 microSiemens, 500 mL 09-0050 Conductivity Standard – 80,000 microSiemens, 500 mL Lock/Unlock Code: 1467
  • 54. PRODUCT WARRANTY Analytical Technology, Inc. (Manufacturer) warrants to the Customer that if any part(s) of the Manufacturer's equipment proves to be defective in materials or workmanship within the earlier of 18 months of the date of shipment or 12 months of the date of start-up, such defective parts will be repaired or replaced free of charge. Inspection and repairs to products thought to be defective within the warranty period will be completed at the Manufacturer's facilities in Collegeville, PA. Products on which warranty repairs are required shall be shipped freight prepaid to the Manufacturer. The product(s) will be returned freight prepaid and allowed if it is determined by the manufacturer that the part(s) failed due to defective materials or workmanship. This warranty does not cover consumable items, batteries, or wear items subject to periodic replacement including lamps and fuses. Gas sensors carry a 12 months from date of shipment warranty and are subject to inspection for evidence of misuse, abuse, alteration, improper storage, or extended exposure to excessive gas concentrations. Should inspection indicate that sensors have failed due to any of the above, the warranty shall not apply. The Manufacturer assumes no liability for consequential damages of any kind, and the buyer by acceptance of this equipment will assume all liability for the consequences of its use or misuse by the Customer, his employees, or others. A defect within the meaning of this warranty is any part of any piece of a Manufacturer's product which shall, when such part is capable of being renewed, repaired, or replaced, operate to condemn such piece of equipment. This warranty is in lieu of all other warranties ( including without limiting the generality of the foregoing warranties of merchantability and fitness for a particular purpose), guarantees, obligations or liabilities expressed or implied by the Manufacturer or its representatives and by statute or rule of law. This warranty is void if the Manufacturer's product(s) has been subject to misuse or abuse, or has not been operated or stored in accordance with instructions, or if the serial number has been removed. Analytical Technology, Inc. makes no other warranty expressed or implied except as stated above.