Radiation source is mounted on a precisely controlled industrial robot.
- Image guidance system(continuous tracking system)
- Eliminates the need of gating techniques and restrictive head frames
4. Robotic Radiosurgery
Highly precise RT delivery system
- Respiratory tracking
- Fiducial based tracking system
- Intra-fraction motion correction
- Uncomparable dose distribution
- X-ray based image verification
Hypofractionated RT
- High dose short course RT
- Higher BED delivered to target
Ideal for moving targets
- Radiation source is mounted on a precisely controlled industrial robot.
- Image guidance system(continuous tracking system)
- Eliminates the need of gating techniques and restrictive head frames
Copyright @ www.radiosurgery-india.com
5. GTV= Gross tumour volume
CTV= Clinical target volume
IM= Internal margin
SM= Setup margin
PTV= Planning target volume
PTV= IM+SM
ICRU 62
CyberKnife in liver tumour: NO IM & SM<1 mm
In Liver, PTV = GTV/ CTV+ 0-1 mm marginIn Liver, PTV = GTV/ CTV+ 0-1 mm margin
Copyright @ www.radiosurgery-india.com
9. Robot is capable of delivering
radiation from different
100 nodes, with each node is
capable of giving a max 12
different beams.
Usage of these nodes depends
on the treatment room constraints
Copyright @ www.radiosurgery-india.com
10. Table consists of 12 fixed cones and housings of
‘Fixed’ and ‘Iris’ Collimator
Collimator sizes(mm): 5, 7.5, 10, 12.5, 15, 20, 25, 30, 35, 40, 50, 60
Laser Sensor
Copyright @ www.radiosurgery-india.com
13. Cyberknife treatment delivery is based on the following
tracking systems
6D_ Skull tracking system
Fiducials tracking system
Synchrony tracking system
X_sight Spine tracking system
X_sight Lung tracking system
Copyright @ www.radiosurgery-india.com
14. 6D_ Skull tracking system:
used for intra-cranial lesions up to C2
Bony anatomy of the skull is used as reference
for tracking
Copyright @
www.radiosurgery-india.com
15. • Fiducial tracking system: used for soft tissues, where
gold fiducials can be implanted.
Minimum of 3 nos. to be implanted
Copyright @ www.radiosurgery-
20. Respiratory-induced motion
of tumors causes significant
targeting uncertainty
Lung, liver, pancreas,
Prostate, kidney
Traditional radiation
therapy margins are not
optimized for high-dose
radiosurgery
21. Two features to form the basis for accuracy
Fiducials, implanted prior to
treatment
Optical markers on a special
patient vest
Copyright @ www.radiosurgery-india.com
22. Prior to treatment start: creation of dynamic correlation model
Markers are monitored in
real time by a camera system
Imaging system takes positions of fiducials at
discrete points of time
23. Prior to treatment start: creation of dynamic correlation model
Markers are monitored in real time by a
camera system
Imaging system takes positions of fiducials at
discrete points of time
time
displacement
displacement
time
This process repeats throughout the treatment, updating and correcting beam
delivery based upon the patient’s current breathing pattern
Copyright @ www.radiosurgery-india.com
30. The mechanical accuracy of the system is 0.12
mm , according to Accuray
The system maintains sub-millimeter tracking
accuracy, if the patient positions are within the
following limits
Left / Right (Lat) 10 mm
Ant/ Post (Ver) 10 mm
Sup/ Inf ( Long) 10 mm
Roll (Left / Right) 10
Pitch (Head Up / Down) 10
Yaw ( C.W / C.C.W) 30
Copyright @ www.radiosurgery-india.com
32. ‘‘CyberKnife is an extension of Gamma-Knife’CyberKnife is an extension of Gamma-Knife’
- Principles of ‘field arrangement’
- Dose distribution pattern
- Multiple isocentre
-Treatment principles
- Treatment delivery accuracy similar
- Delivered dose in single fractions
- Intra-cranial indications
CK & GK: SimilarityCK & GK: Similarity
‘CyberKnife treatment is as good as Gamma-Knife’ Dr Bernadt Wowra
Copyright @ www.radiosurgery-india.com
33. Cyberknife Vs Gamma-Knife:Cyberknife Vs Gamma-Knife: DissimilarityDissimilarity
GK CK Comments
Immobilization
device
Rigid frame Orfit CK has favorable orfit
RT source Co60 6MV LA GK need to replace sources
every 5/6 yrs
Planning No complex planning Inverse planning Favorable dosimetry in CK
Planning method Simple Complex Even neurosurgeons can plan
in GK
Isodose pres Usually 50% Usually 80-95% GK: more dose heterogeniety
Fractions Single May treat multiple
fraction
Radiobiology favorable in CK
Tumour size Only smaller lesions
can be treated
Larger lesions also
can be treated in
fractionated schedule
Increased indications with CK
Energy source Radiation Electricity GK can work with less
electricity
Verification Not possible Possible Even Intra-fraction movement
can be corrected
Indications Only brain lesions Extra & intra cranial CK more economical
34. SummarySummary
CyberKnife Radiosurgery is an ideal machine for both cranial & extracranial
radiosurgery
Appropriate patient selection is the most important factor
CyberKnife is safe and patient friendly radiotherapy delivery system
Short course, precise, high dose RT may be beneficial in few indications
However, the indications are limited and long-term data of most of the extra-
cranial indication is still to be available
With more clinical data, indications are expanding