1.Continuous, beat-to-beat blood pressure
  measurement.
  - Hemodynamically unstable pts /ICU pts requiring
  inotropic support
  - Patients undergoing major surgery
2.Frequent arterial blood gas analysis
  -pts with respiratory failure on ventilator
  -severe acid/base disturbance.
Advantages of IBP measurement
 Continuous blood pressure recording
 Accurate blood pressure recording even
   when patients are profoundly hypotensive vs
  NIBP which is difficult or inaccurate
 Real time Visual Display
Disadvantages of IBP measurement
 Potential complications
 Skilled technique reqd
 Expensive
 The radial artery has low complication
  rates compared with other sites.
 It is a superficial artery which aids insertion, and
  also makes it compressible for haemostasis
 The ulnar, brachial, axillary, dorsalis
  pedis, posterial tibial, femoral arteries are
  alternatives.
   Allen’s test is recommended before
    the insertion of a radial arterial line.
   This is used to determine collateral
    circulation between the ulnar and
    radial arteries to the hand
   If ulnar perfusion is poor and a
    cannula occludes the radial artery,
    blood flow to the hand may be
    reduced.
   The test is performed by asking the
    patient to clench their hand. The ulnar
    and radial arteries are occluded with
    digital pressure.
   The hand is unclenched and pressure
    over the ulnar artery is released. If
    there is good collateral perfusion, the
    palm should flush in less than 6
    seconds.


                                               The idea here is to figure out if the ulnar
                                               artery will supply the hand with enough
                                               blood, if the radial artery is blocked with an
                                               a-line.
Arterial cannula
   Made from polytetrafluoroethylene („Teflon‟) to minimize
    the risk of clot formation
   20G (pink) cannula - adult patients
    22G (blue)- paediatrics
    24G (yellow) - neonates and small babies
   Larger gauge cannulae increase the risk of thrombosis,
    smaller cannulae cause damping of the signal.
   The cannula is connected to an arterial giving set.
   Arterial set.
-   Specialized plastic tubing, short and stiff to reduce resonance, connected to a 500 ml bag
    of saline.

   Saline bag
    -500 ml 0.9% saline pressurized to 300 mmHg using a pressure bag, i.e. a pressure higher
    than arterial systolic pressure to prevent backflow from the cannula into the giving set.
    -The arterial set and pressurized saline bag with 2500units Heparin incorporate a
    continuous slow flushing system of 3–4 ml per hour to keep the line free from clots.
     -The arterial set and arterial line should be free from air bubbles.
    - The line is attached to a transducer.
     -Do not allow the saline bag to empty
    ◦   –To maintain patency of arterial cannula.
    ◦   –To prevent air embolism
    ◦   –To maintain accuracy of blood pressure reading
    ◦   –To maintain accuracy of fluid balance chart
    ◦   –To prevent backflow of blood

   Transducer, amplifier and electrical recording equipment.
    -The transducer is zeroed and placed level with the heart.
   Tape and/or steri-strips
   An arm board or towel roll
   Opsite or Tegaderm cover dressing
   Local anesthetic (1% or 2% lidocaine ,lidocaine
    cream)
   Suture material for femoral arterial line placement (2.0
    silk)
   Scissors
   Monitor cable for transducing arterial waveform.
   Benzoin solution
 1. Ensure that all preprocedure steps are
  taken
 2. Assure that pressure tubing with transducer
  is connected to bedside monitor.
 3. Perform the Allen‟s test to assure adequate
  collateral blood flow if using the radial artery.
 4. Wash hands and don gloves
 5. For the radial artery, the arm is
  restrained, palm up, with an armboard to
  hold the wrist dorsiflexed
For the radial artery, the most common
insertion, the arm is restrained, palm
up, with an armboard to hold the wrist
dorsiflexed
5. Apply anesthetic agent (local lidocaine 1-2% or lidocaine cream).
 6. Locate pulsating artery via palpation.
 7. Cleanse area selected for arterial line placement.
 8. Prepare patient for puncture.
 9. Stabilize artery by pulling skin taut.
 10. Puncture skin at 45-60 degree angle for radial artery; 90
   degrees for femoral artery.
 11. Advance catheter when flash of blood is observed in catheter.
 12. Connect to pressure I.V. tubing and check for arterial waveform
   on bedside monitor.
 13. Cleanse area of any blood and allow site to dry.
 14. Apply Benzoin to cleansed area and allow to dry and become
   “tacky.”
 15. Secure arterial line with tape and cover with a Tegaderm
   dressing.
 16. Secure I.V. tubing to prevent it from being caught and pulling on
   arterial catheter. If a femoral arterial line is placed, it should be
   secured with a suture.
 17. Properly dispose of the I.V. sharps and other used materials.
   1. Direct cannulation

   2. Transfixation

   3. Guidewire (Seldinger)
    technique
    Haemorrhage may occur if there are leaks in the system. Connections
     must be tightly secured and the giving set and line closely observed..
    Emboli. Air or thrombo emboli may occur.
     Care should be taken to aspirate air bubbles
    Accidental drug injection may cause severe, irreversible damage to
     the hand.
     -No drugs should be injected via an arterial line
    - The line should be labelled (in red) to reduce the likelihood of this
     occurring
    Arterial vasospasm
    Partial occlusion due to large cannula width, multiple attempts at
     insertion and long duration of use
    Permanent total occlusion
    Sepsis or bacteraemia secondary to infected radial arterial lines is very
     rare (0.13%);
     -local infection is more common.
     -if the area looks inflamed the line site should be changed.
   –Concentration of a drug
    into the tissues served by
    the cannulated artery can
    result in cell death
   –Skin necrosis, severe
    gangrene, limb ischemia,
    amputation & permanent
    disabilities
Mechanism of action
   A transducer is a device that
    reads the fluctuations in pressure
    – it doesn’t matter if it’s arterial, or
    central venous, or PA
   The column of saline in the arterial
    set transmits the pressure changes
    to the diaphragm in the transducer
    The transducer reads the changing
    pressure, and changes it into an
    electrical signal that goes up and
    down as the pressure does which is
    displayed as an arterial waveform
   The transducer connects to the
    bedside monitor with a cable, and
    the wave shows up on the screen,
    going from left to right
   The transducer has to sit in a “transducer holder” – this is the
    white plastic plate that screws onto the rolling pole that holds the
    whole setup.
   The transducer has to be levelled correctly-to make sure that it‟s at
    the fourth intercostal space, at the mid-axillary line
    (Phlebostatic axis)
    Make sure there‟s no air in the line before you hook it up to the
    patient – use the flusher to clear bubbles out of the tubing.
    Zero the line to atmospheric pressure properly
   Choose a screen scale that lets you see the waveform clearly.
   To ensure accuracy of readings
   Flush the device & turn it off to patient but open to
    atmosphere
   These exert pressure on transducer
   This pressure is called zero
   Zero once per shift or if values are questionable
   Ensure flush bag is pumped up
   Once inserted, an
    arterial waveform trace
    should be displayed at
    all times
   This confirms that the
    invasive arterial BP
    monitoring is set up
    correctly, and
    minimizes problems.
:
   The highest point - systolic
    pressure,
     -the lowest is the diastolic.
    Everybody see the little notch
    on the diastolic downslope? –
    there‟s one in each beat.
    A little after the beginning of
    diastole – the start of the
    downward wave – the aortic
    valve flips closed, generating a
    little back-pressure bump:
    called the “dicrotic notch”..
   Now we know how the arterial pressure monitoring
    system works, we need to be able to decide
    whether or not the trace (and BP in
    numerical format) is accurate.
    Failure to notice this may lead to unnecessary, or
    missed treatments for our patients.
   There are 2 main abnormal tracing problems that
    can occur once the monitor gain is set correctly.
   Dampened trace
   Dampening occurs due to:
    ◦   air bubbles
    ◦   overly compliant, distensible tubing
    ◦   catheter kinks
    ◦   clots
    ◦   injection ports
    ◦   low flush bag pressure or no fluid in the flush bag
    ◦   Improper scaling
    ◦   Severe hypotension if everything else is ruled out
   This type of trace Under estimate SBP, over estimate
    DBP
   Resonant trace
   Resonance occurs due to:
    ◦ long tubing
    ◦ overly stiff, non-compliant tubing
    ◦ increased vascular resistance
    ◦ reverberations in tubing causing harmonics that distort
      the trace (i.e. high systolic and low diastolic)
    ◦ non-fully opened stopcock valve
   This type of trace
   Over estimate SBP, under estimate DBP
   Arterial lines measure systolic BP approximately 5
    mmHg higher and the diastolic BP
    approximately 8 mmHg lower compared to non-
    invasive BP (NIBP) measurement
Troubleshooting
    “It takes a year just to learn which way to turn the
    stopcocks!”
    This is really true: some stopcocks point to where
    they‟re open, and some point to where they‟re closed
    – it just takes some time to learn which is which.
   The trick is remembering which way to turn the
    stopcock, and avoiding a mess.
    Don‟t forget to clear the stopcock, recap, and then
    flush the line.
   Keep things nice and sterile.
   This probably means that the artery being monitored
    or gone into spasm.
   You need to think about things that might make this
    happen:
    -Is the patient very cold?
    -Are his extremities poorly perfused?
    -Is he on a “shipload” of pressors, making his
    arterial bed tighten up –
    - Is he “dry” as well?
   Sometimes arteries become unhappy with catheters in
    them, and you just have to convince the team that the
    patient needs a new one placed in another site.
The first thing to think about is:
    1.Is the arterial catheter still in place? Yes? Try drawing
    with a 3cc syringe from the stopcock – if it draws normally,
    then you‟ve got a hardware problem
    2.Cables become loose?
    3.Did the screen scale get accidentally set to, say, 40,
    instead of 150 or 200mm of pressure?- you‟ll only see a flat
    line.

    4.Is it a transducer setup Failure – try a new setup.

   If the line doesn‟t draw –
    -Is there a clot in the hub?
    -Try taking the site dressing down – is the catheter kinked
    going into the patient?
    - Sometimes art-lines just fail – the artery spasms and
    won‟t open up – time for a new site.
   The routine now is 96 hours – make sure that you
    label the line setup when you hang it. Obviously,
    change the line setup if it is contaminated in any
    way.
   Usually this will be pretty obvious:
    the pulse will diminish, or go away altogether. The
    hand may look dusky, or be cold, or lose some
    sensation – remember to assess for coloring,
    sensation, motion, and capillary refill.
   If you think that the a-line is threatening the
    patient’s hand, let the team know right away, and be
    ready to set up for another insertion somewhere else if
    the line is still necessary.
   Compress the site with a sterile 4x4 for at least 5
    minutes, or longer if the patient is anticoagulated.
   Assess the perfusion of the hand.
   Make sure you put the patient on the non-
    invasive cuff at meaningful intervals while you
    talk to the team about replacing the line
   This is usually pretty obvious – the patient is
    hemodynamically stable, needs only one or two
    blood draws in a day, no more need for ABGs
    Disconnect the cable from the monitor which will
    automatically turn off the alarms.
   Take out the sutures in the usual way with a fresh sterile kit.
   Have a gauzepiece ready, pull the catheter, and manually
    compress the site for at least 3 to 5 minutes.
   Make sure the patient‟s hand is still perfused.
   Check for hematoma or bleeding, put a compression
    dressing on the site (not too tight!), which you can then take
    off after about an hour.
   Recheck the site hourly for a few hours afterwards – a
    hematoma could still form, and since there isn‟t a whole lot of
    room in a wrist, you‟d definitely want to know!
Any Questions ????
Thanks

More Related Content

What's hot

LAYNGEAL MASK AIRWAY
LAYNGEAL MASK AIRWAYLAYNGEAL MASK AIRWAY
LAYNGEAL MASK AIRWAYNiresh Raja
 
Anaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyAnaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyUnnikrishnan Prathapadas
 
Pulmonary artery catheter
Pulmonary artery catheterPulmonary artery catheter
Pulmonary artery catheterArun Aru
 
Arterial line analysis
Arterial line analysisArterial line analysis
Arterial line analysissamirelansary
 
Central venous catheterization
Central venous catheterizationCentral venous catheterization
Central venous catheterizationMEEQAT HOSPITAL
 
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewBougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewDhritiman Chakrabarti
 
Cardiac output monitoring
Cardiac output monitoringCardiac output monitoring
Cardiac output monitoringmauryaramgopal
 
Safety features in anesthesia machine
Safety features in anesthesia machineSafety features in anesthesia machine
Safety features in anesthesia machineomar143
 
Invasive blood pressure_monitoring
Invasive blood pressure_monitoringInvasive blood pressure_monitoring
Invasive blood pressure_monitoringUbaidur Rahaman
 
laryngoscope class by Dr Sandeep Singh Jadon ppt.pptx
laryngoscope class by Dr Sandeep Singh Jadon ppt.pptxlaryngoscope class by Dr Sandeep Singh Jadon ppt.pptx
laryngoscope class by Dr Sandeep Singh Jadon ppt.pptxSandeep Singh Jadon
 
Role of Anesthesiologist in Cath Lab
Role of Anesthesiologist in Cath LabRole of Anesthesiologist in Cath Lab
Role of Anesthesiologist in Cath LabAbhijit Nair
 

What's hot (20)

LAYNGEAL MASK AIRWAY
LAYNGEAL MASK AIRWAYLAYNGEAL MASK AIRWAY
LAYNGEAL MASK AIRWAY
 
Anaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyAnaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supply
 
Pulmonary artery catheter
Pulmonary artery catheterPulmonary artery catheter
Pulmonary artery catheter
 
Arterial line analysis
Arterial line analysisArterial line analysis
Arterial line analysis
 
Central venous catheterization
Central venous catheterizationCentral venous catheterization
Central venous catheterization
 
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truviewBougie, trachlite , laryngeal tube , combitube , i gel ,truview
Bougie, trachlite , laryngeal tube , combitube , i gel ,truview
 
Cardiac output monitoring
Cardiac output monitoringCardiac output monitoring
Cardiac output monitoring
 
Cvp line
Cvp lineCvp line
Cvp line
 
PiCCO Monitor
PiCCO Monitor PiCCO Monitor
PiCCO Monitor
 
Safety features in anesthesia machine
Safety features in anesthesia machineSafety features in anesthesia machine
Safety features in anesthesia machine
 
Cardioversion
Cardioversion Cardioversion
Cardioversion
 
Invasive blood pressure_monitoring
Invasive blood pressure_monitoringInvasive blood pressure_monitoring
Invasive blood pressure_monitoring
 
laryngoscope class by Dr Sandeep Singh Jadon ppt.pptx
laryngoscope class by Dr Sandeep Singh Jadon ppt.pptxlaryngoscope class by Dr Sandeep Singh Jadon ppt.pptx
laryngoscope class by Dr Sandeep Singh Jadon ppt.pptx
 
Role of Anesthesiologist in Cath Lab
Role of Anesthesiologist in Cath LabRole of Anesthesiologist in Cath Lab
Role of Anesthesiologist in Cath Lab
 
Central line
Central line Central line
Central line
 
Respiratory monitoring
Respiratory monitoringRespiratory monitoring
Respiratory monitoring
 
Nasopharyngeal Airway.pptx
Nasopharyngeal Airway.pptxNasopharyngeal Airway.pptx
Nasopharyngeal Airway.pptx
 
Rapid sequence intubation
Rapid sequence intubationRapid sequence intubation
Rapid sequence intubation
 
Double lumen tubes
Double lumen tubesDouble lumen tubes
Double lumen tubes
 
CVP Monitoring_Dr. Subrata Kumar_BSMMU_2014
CVP Monitoring_Dr. Subrata Kumar_BSMMU_2014CVP Monitoring_Dr. Subrata Kumar_BSMMU_2014
CVP Monitoring_Dr. Subrata Kumar_BSMMU_2014
 

Similar to Arterial line insertion

Cardiovascular monitoring Part I
Cardiovascular monitoring Part ICardiovascular monitoring Part I
Cardiovascular monitoring Part ISiddhanta Choudhury
 
Lec # 6 hemodynamic monitoring
Lec # 6 hemodynamic monitoringLec # 6 hemodynamic monitoring
Lec # 6 hemodynamic monitoringAli Sheikh
 
Iv Therapy
Iv TherapyIv Therapy
Iv Therapywashinca
 
bpmonitoring.pdf
bpmonitoring.pdfbpmonitoring.pdf
bpmonitoring.pdfshafina27
 
CENTRAL VENOUS CATHETER
CENTRAL VENOUS CATHETERCENTRAL VENOUS CATHETER
CENTRAL VENOUS CATHETERAvijit Prusty
 
centralvenouscatheter-1.pdf
centralvenouscatheter-1.pdfcentralvenouscatheter-1.pdf
centralvenouscatheter-1.pdfisha sharma
 
Free flap postoperative evaluation
Free flap postoperative evaluationFree flap postoperative evaluation
Free flap postoperative evaluationAzis Aimaduddin
 
HAEMODYNAMIC MONITORING – CVP, PAC AND IBP
HAEMODYNAMIC MONITORING – CVP, PAC AND IBPHAEMODYNAMIC MONITORING – CVP, PAC AND IBP
HAEMODYNAMIC MONITORING – CVP, PAC AND IBPGowri Shankar
 
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...prateek gupta
 
Arterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.pptArterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.pptssuser35745f
 
Central venous pressure monitoring
Central venous pressure monitoring Central venous pressure monitoring
Central venous pressure monitoring DR .PALLAVI PATHANIA
 
Intraoperative Monitoring. sandeep. first years.ppt
Intraoperative Monitoring. sandeep. first years.pptIntraoperative Monitoring. sandeep. first years.ppt
Intraoperative Monitoring. sandeep. first years.pptsandeepsandeepkundra
 
Dialysis machines.pptx
Dialysis machines.pptxDialysis machines.pptx
Dialysis machines.pptxDarshanS239776
 

Similar to Arterial line insertion (20)

Cardiovascular monitoring Part I
Cardiovascular monitoring Part ICardiovascular monitoring Part I
Cardiovascular monitoring Part I
 
Lec # 6 hemodynamic monitoring
Lec # 6 hemodynamic monitoringLec # 6 hemodynamic monitoring
Lec # 6 hemodynamic monitoring
 
Iv Therapy
Iv TherapyIv Therapy
Iv Therapy
 
Iv Therapy
Iv TherapyIv Therapy
Iv Therapy
 
bpmonitoring.pdf
bpmonitoring.pdfbpmonitoring.pdf
bpmonitoring.pdf
 
Dialysis machines key features
Dialysis machines key featuresDialysis machines key features
Dialysis machines key features
 
CENTRAL VENOUS CATHETER
CENTRAL VENOUS CATHETERCENTRAL VENOUS CATHETER
CENTRAL VENOUS CATHETER
 
centralvenouscatheter-1.pdf
centralvenouscatheter-1.pdfcentralvenouscatheter-1.pdf
centralvenouscatheter-1.pdf
 
Free flap postoperative evaluation
Free flap postoperative evaluationFree flap postoperative evaluation
Free flap postoperative evaluation
 
Eritro intro01
Eritro intro01Eritro intro01
Eritro intro01
 
HAEMODYNAMIC MONITORING – CVP, PAC AND IBP
HAEMODYNAMIC MONITORING – CVP, PAC AND IBPHAEMODYNAMIC MONITORING – CVP, PAC AND IBP
HAEMODYNAMIC MONITORING – CVP, PAC AND IBP
 
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
 
Hemodynamic monitoring
Hemodynamic monitoringHemodynamic monitoring
Hemodynamic monitoring
 
ICU
ICUICU
ICU
 
Perioprative monitoring
Perioprative monitoringPerioprative monitoring
Perioprative monitoring
 
Arterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.pptArterial_and_CVP_monitoring.ppt
Arterial_and_CVP_monitoring.ppt
 
Central venous pressure monitoring
Central venous pressure monitoring Central venous pressure monitoring
Central venous pressure monitoring
 
Intraoperative Monitoring. sandeep. first years.ppt
Intraoperative Monitoring. sandeep. first years.pptIntraoperative Monitoring. sandeep. first years.ppt
Intraoperative Monitoring. sandeep. first years.ppt
 
Dialysis machines.pptx
Dialysis machines.pptxDialysis machines.pptx
Dialysis machines.pptx
 
Hemodynamic monitorig
Hemodynamic monitorigHemodynamic monitorig
Hemodynamic monitorig
 

Recently uploaded

Advanced Computer Architecture – An Introduction
Advanced Computer Architecture – An IntroductionAdvanced Computer Architecture – An Introduction
Advanced Computer Architecture – An IntroductionDilum Bandara
 
Human Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsHuman Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsMark Billinghurst
 
Commit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easyCommit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easyAlfredo García Lavilla
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity PlanDatabarracks
 
Powerpoint exploring the locations used in television show Time Clash
Powerpoint exploring the locations used in television show Time ClashPowerpoint exploring the locations used in television show Time Clash
Powerpoint exploring the locations used in television show Time Clashcharlottematthew16
 
H2O.ai CEO/Founder: Sri Ambati Keynote at Wells Fargo Day
H2O.ai CEO/Founder: Sri Ambati Keynote at Wells Fargo DayH2O.ai CEO/Founder: Sri Ambati Keynote at Wells Fargo Day
H2O.ai CEO/Founder: Sri Ambati Keynote at Wells Fargo DaySri Ambati
 
CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):comworks
 
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks..."LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...Fwdays
 
Vertex AI Gemini Prompt Engineering Tips
Vertex AI Gemini Prompt Engineering TipsVertex AI Gemini Prompt Engineering Tips
Vertex AI Gemini Prompt Engineering TipsMiki Katsuragi
 
What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024Stephanie Beckett
 
"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr Bagan"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr BaganFwdays
 
Search Engine Optimization SEO PDF for 2024.pdf
Search Engine Optimization SEO PDF for 2024.pdfSearch Engine Optimization SEO PDF for 2024.pdf
Search Engine Optimization SEO PDF for 2024.pdfRankYa
 
DevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platformsDevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platformsSergiu Bodiu
 
WordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your BrandWordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your Brandgvaughan
 
Anypoint Exchange: It’s Not Just a Repo!
Anypoint Exchange: It’s Not Just a Repo!Anypoint Exchange: It’s Not Just a Repo!
Anypoint Exchange: It’s Not Just a Repo!Manik S Magar
 
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxMerck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxLoriGlavin3
 
The Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsThe Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsPixlogix Infotech
 
Leverage Zilliz Serverless - Up to 50X Saving for Your Vector Storage Cost
Leverage Zilliz Serverless - Up to 50X Saving for Your Vector Storage CostLeverage Zilliz Serverless - Up to 50X Saving for Your Vector Storage Cost
Leverage Zilliz Serverless - Up to 50X Saving for Your Vector Storage CostZilliz
 

Recently uploaded (20)

Advanced Computer Architecture – An Introduction
Advanced Computer Architecture – An IntroductionAdvanced Computer Architecture – An Introduction
Advanced Computer Architecture – An Introduction
 
Human Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR SystemsHuman Factors of XR: Using Human Factors to Design XR Systems
Human Factors of XR: Using Human Factors to Design XR Systems
 
Commit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easyCommit 2024 - Secret Management made easy
Commit 2024 - Secret Management made easy
 
How to write a Business Continuity Plan
How to write a Business Continuity PlanHow to write a Business Continuity Plan
How to write a Business Continuity Plan
 
Powerpoint exploring the locations used in television show Time Clash
Powerpoint exploring the locations used in television show Time ClashPowerpoint exploring the locations used in television show Time Clash
Powerpoint exploring the locations used in television show Time Clash
 
H2O.ai CEO/Founder: Sri Ambati Keynote at Wells Fargo Day
H2O.ai CEO/Founder: Sri Ambati Keynote at Wells Fargo DayH2O.ai CEO/Founder: Sri Ambati Keynote at Wells Fargo Day
H2O.ai CEO/Founder: Sri Ambati Keynote at Wells Fargo Day
 
CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):CloudStudio User manual (basic edition):
CloudStudio User manual (basic edition):
 
DMCC Future of Trade Web3 - Special Edition
DMCC Future of Trade Web3 - Special EditionDMCC Future of Trade Web3 - Special Edition
DMCC Future of Trade Web3 - Special Edition
 
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks..."LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
"LLMs for Python Engineers: Advanced Data Analysis and Semantic Kernel",Oleks...
 
Vertex AI Gemini Prompt Engineering Tips
Vertex AI Gemini Prompt Engineering TipsVertex AI Gemini Prompt Engineering Tips
Vertex AI Gemini Prompt Engineering Tips
 
What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024What's New in Teams Calling, Meetings and Devices March 2024
What's New in Teams Calling, Meetings and Devices March 2024
 
"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr Bagan"ML in Production",Oleksandr Bagan
"ML in Production",Oleksandr Bagan
 
Search Engine Optimization SEO PDF for 2024.pdf
Search Engine Optimization SEO PDF for 2024.pdfSearch Engine Optimization SEO PDF for 2024.pdf
Search Engine Optimization SEO PDF for 2024.pdf
 
DevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platformsDevEX - reference for building teams, processes, and platforms
DevEX - reference for building teams, processes, and platforms
 
WordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your BrandWordPress Websites for Engineers: Elevate Your Brand
WordPress Websites for Engineers: Elevate Your Brand
 
Anypoint Exchange: It’s Not Just a Repo!
Anypoint Exchange: It’s Not Just a Repo!Anypoint Exchange: It’s Not Just a Repo!
Anypoint Exchange: It’s Not Just a Repo!
 
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptxMerck Moving Beyond Passwords: FIDO Paris Seminar.pptx
Merck Moving Beyond Passwords: FIDO Paris Seminar.pptx
 
The Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and ConsThe Ultimate Guide to Choosing WordPress Pros and Cons
The Ultimate Guide to Choosing WordPress Pros and Cons
 
E-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptx
E-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptxE-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptx
E-Vehicle_Hacking_by_Parul Sharma_null_owasp.pptx
 
Leverage Zilliz Serverless - Up to 50X Saving for Your Vector Storage Cost
Leverage Zilliz Serverless - Up to 50X Saving for Your Vector Storage CostLeverage Zilliz Serverless - Up to 50X Saving for Your Vector Storage Cost
Leverage Zilliz Serverless - Up to 50X Saving for Your Vector Storage Cost
 

Arterial line insertion

  • 2. Indications  Preparation & Equipments  Positioning  Insertion  Complications  Mechanism of Action  Troubleshooting
  • 3. 1.Continuous, beat-to-beat blood pressure measurement. - Hemodynamically unstable pts /ICU pts requiring inotropic support - Patients undergoing major surgery 2.Frequent arterial blood gas analysis -pts with respiratory failure on ventilator -severe acid/base disturbance.
  • 4. Advantages of IBP measurement  Continuous blood pressure recording  Accurate blood pressure recording even when patients are profoundly hypotensive vs NIBP which is difficult or inaccurate  Real time Visual Display Disadvantages of IBP measurement  Potential complications  Skilled technique reqd  Expensive
  • 5.  The radial artery has low complication rates compared with other sites.  It is a superficial artery which aids insertion, and also makes it compressible for haemostasis  The ulnar, brachial, axillary, dorsalis pedis, posterial tibial, femoral arteries are alternatives.
  • 6.
  • 7. Allen’s test is recommended before the insertion of a radial arterial line.  This is used to determine collateral circulation between the ulnar and radial arteries to the hand  If ulnar perfusion is poor and a cannula occludes the radial artery, blood flow to the hand may be reduced.  The test is performed by asking the patient to clench their hand. The ulnar and radial arteries are occluded with digital pressure.  The hand is unclenched and pressure over the ulnar artery is released. If there is good collateral perfusion, the palm should flush in less than 6 seconds. The idea here is to figure out if the ulnar artery will supply the hand with enough blood, if the radial artery is blocked with an a-line.
  • 8. Arterial cannula  Made from polytetrafluoroethylene („Teflon‟) to minimize the risk of clot formation  20G (pink) cannula - adult patients 22G (blue)- paediatrics 24G (yellow) - neonates and small babies  Larger gauge cannulae increase the risk of thrombosis, smaller cannulae cause damping of the signal.  The cannula is connected to an arterial giving set.
  • 9.
  • 10. Arterial set. - Specialized plastic tubing, short and stiff to reduce resonance, connected to a 500 ml bag of saline.  Saline bag -500 ml 0.9% saline pressurized to 300 mmHg using a pressure bag, i.e. a pressure higher than arterial systolic pressure to prevent backflow from the cannula into the giving set. -The arterial set and pressurized saline bag with 2500units Heparin incorporate a continuous slow flushing system of 3–4 ml per hour to keep the line free from clots. -The arterial set and arterial line should be free from air bubbles. - The line is attached to a transducer. -Do not allow the saline bag to empty ◦ –To maintain patency of arterial cannula. ◦ –To prevent air embolism ◦ –To maintain accuracy of blood pressure reading ◦ –To maintain accuracy of fluid balance chart ◦ –To prevent backflow of blood  Transducer, amplifier and electrical recording equipment. -The transducer is zeroed and placed level with the heart.
  • 11. Tape and/or steri-strips  An arm board or towel roll  Opsite or Tegaderm cover dressing  Local anesthetic (1% or 2% lidocaine ,lidocaine cream)  Suture material for femoral arterial line placement (2.0 silk)  Scissors  Monitor cable for transducing arterial waveform.  Benzoin solution
  • 12.  1. Ensure that all preprocedure steps are taken  2. Assure that pressure tubing with transducer is connected to bedside monitor.  3. Perform the Allen‟s test to assure adequate collateral blood flow if using the radial artery.  4. Wash hands and don gloves  5. For the radial artery, the arm is restrained, palm up, with an armboard to hold the wrist dorsiflexed
  • 13. For the radial artery, the most common insertion, the arm is restrained, palm up, with an armboard to hold the wrist dorsiflexed
  • 14. 5. Apply anesthetic agent (local lidocaine 1-2% or lidocaine cream).  6. Locate pulsating artery via palpation.  7. Cleanse area selected for arterial line placement.  8. Prepare patient for puncture.  9. Stabilize artery by pulling skin taut.  10. Puncture skin at 45-60 degree angle for radial artery; 90 degrees for femoral artery.  11. Advance catheter when flash of blood is observed in catheter.  12. Connect to pressure I.V. tubing and check for arterial waveform on bedside monitor.  13. Cleanse area of any blood and allow site to dry.  14. Apply Benzoin to cleansed area and allow to dry and become “tacky.”  15. Secure arterial line with tape and cover with a Tegaderm dressing.  16. Secure I.V. tubing to prevent it from being caught and pulling on arterial catheter. If a femoral arterial line is placed, it should be secured with a suture.  17. Properly dispose of the I.V. sharps and other used materials.
  • 15. 1. Direct cannulation  2. Transfixation  3. Guidewire (Seldinger) technique
  • 16. Haemorrhage may occur if there are leaks in the system. Connections must be tightly secured and the giving set and line closely observed..  Emboli. Air or thrombo emboli may occur. Care should be taken to aspirate air bubbles  Accidental drug injection may cause severe, irreversible damage to the hand. -No drugs should be injected via an arterial line - The line should be labelled (in red) to reduce the likelihood of this occurring  Arterial vasospasm  Partial occlusion due to large cannula width, multiple attempts at insertion and long duration of use  Permanent total occlusion  Sepsis or bacteraemia secondary to infected radial arterial lines is very rare (0.13%); -local infection is more common. -if the area looks inflamed the line site should be changed.
  • 17.
  • 18. –Concentration of a drug into the tissues served by the cannulated artery can result in cell death  –Skin necrosis, severe gangrene, limb ischemia, amputation & permanent disabilities
  • 19. Mechanism of action  A transducer is a device that reads the fluctuations in pressure – it doesn’t matter if it’s arterial, or central venous, or PA  The column of saline in the arterial set transmits the pressure changes to the diaphragm in the transducer  The transducer reads the changing pressure, and changes it into an electrical signal that goes up and down as the pressure does which is displayed as an arterial waveform  The transducer connects to the bedside monitor with a cable, and the wave shows up on the screen, going from left to right
  • 20. The transducer has to sit in a “transducer holder” – this is the white plastic plate that screws onto the rolling pole that holds the whole setup.  The transducer has to be levelled correctly-to make sure that it‟s at the fourth intercostal space, at the mid-axillary line (Phlebostatic axis)  Make sure there‟s no air in the line before you hook it up to the patient – use the flusher to clear bubbles out of the tubing.  Zero the line to atmospheric pressure properly  Choose a screen scale that lets you see the waveform clearly.
  • 21. To ensure accuracy of readings  Flush the device & turn it off to patient but open to atmosphere  These exert pressure on transducer  This pressure is called zero  Zero once per shift or if values are questionable  Ensure flush bag is pumped up
  • 22. Once inserted, an arterial waveform trace should be displayed at all times  This confirms that the invasive arterial BP monitoring is set up correctly, and minimizes problems.
  • 23. :  The highest point - systolic pressure, -the lowest is the diastolic. Everybody see the little notch on the diastolic downslope? – there‟s one in each beat.  A little after the beginning of diastole – the start of the downward wave – the aortic valve flips closed, generating a little back-pressure bump: called the “dicrotic notch”..
  • 24. Now we know how the arterial pressure monitoring system works, we need to be able to decide whether or not the trace (and BP in numerical format) is accurate.  Failure to notice this may lead to unnecessary, or missed treatments for our patients.  There are 2 main abnormal tracing problems that can occur once the monitor gain is set correctly.
  • 25. Dampened trace
  • 26. Dampening occurs due to: ◦ air bubbles ◦ overly compliant, distensible tubing ◦ catheter kinks ◦ clots ◦ injection ports ◦ low flush bag pressure or no fluid in the flush bag ◦ Improper scaling ◦ Severe hypotension if everything else is ruled out  This type of trace Under estimate SBP, over estimate DBP
  • 27. Resonant trace
  • 28. Resonance occurs due to: ◦ long tubing ◦ overly stiff, non-compliant tubing ◦ increased vascular resistance ◦ reverberations in tubing causing harmonics that distort the trace (i.e. high systolic and low diastolic) ◦ non-fully opened stopcock valve  This type of trace  Over estimate SBP, under estimate DBP
  • 29. Arterial lines measure systolic BP approximately 5 mmHg higher and the diastolic BP approximately 8 mmHg lower compared to non- invasive BP (NIBP) measurement
  • 31. “It takes a year just to learn which way to turn the stopcocks!” This is really true: some stopcocks point to where they‟re open, and some point to where they‟re closed – it just takes some time to learn which is which.  The trick is remembering which way to turn the stopcock, and avoiding a mess.  Don‟t forget to clear the stopcock, recap, and then flush the line.  Keep things nice and sterile.
  • 32. This probably means that the artery being monitored or gone into spasm.  You need to think about things that might make this happen: -Is the patient very cold? -Are his extremities poorly perfused? -Is he on a “shipload” of pressors, making his arterial bed tighten up – - Is he “dry” as well?  Sometimes arteries become unhappy with catheters in them, and you just have to convince the team that the patient needs a new one placed in another site.
  • 33. The first thing to think about is: 1.Is the arterial catheter still in place? Yes? Try drawing with a 3cc syringe from the stopcock – if it draws normally, then you‟ve got a hardware problem 2.Cables become loose? 3.Did the screen scale get accidentally set to, say, 40, instead of 150 or 200mm of pressure?- you‟ll only see a flat line. 4.Is it a transducer setup Failure – try a new setup.  If the line doesn‟t draw – -Is there a clot in the hub? -Try taking the site dressing down – is the catheter kinked going into the patient? - Sometimes art-lines just fail – the artery spasms and won‟t open up – time for a new site.
  • 34. The routine now is 96 hours – make sure that you label the line setup when you hang it. Obviously, change the line setup if it is contaminated in any way.
  • 35. Usually this will be pretty obvious: the pulse will diminish, or go away altogether. The hand may look dusky, or be cold, or lose some sensation – remember to assess for coloring, sensation, motion, and capillary refill.  If you think that the a-line is threatening the patient’s hand, let the team know right away, and be ready to set up for another insertion somewhere else if the line is still necessary.
  • 36. Compress the site with a sterile 4x4 for at least 5 minutes, or longer if the patient is anticoagulated.  Assess the perfusion of the hand.  Make sure you put the patient on the non- invasive cuff at meaningful intervals while you talk to the team about replacing the line
  • 37. This is usually pretty obvious – the patient is hemodynamically stable, needs only one or two blood draws in a day, no more need for ABGs
  • 38. Disconnect the cable from the monitor which will automatically turn off the alarms.  Take out the sutures in the usual way with a fresh sterile kit.  Have a gauzepiece ready, pull the catheter, and manually compress the site for at least 3 to 5 minutes.  Make sure the patient‟s hand is still perfused.  Check for hematoma or bleeding, put a compression dressing on the site (not too tight!), which you can then take off after about an hour.  Recheck the site hourly for a few hours afterwards – a hematoma could still form, and since there isn‟t a whole lot of room in a wrist, you‟d definitely want to know!