TOTAL QUALITY
MANAGEMENT
Dr Anurag Yadav
Assistant Professor Department of
Biochemistry
Lean Six Sigma Green Belt
CONTENT
 Quality Management
 Errors in Lab
 Element of Quality assurance program
 Control of errors in Variables
 Preanalytical
 Analytical
 Using control material
 Using Patient data
 External quality assessment & Proficiency testing
FUNDAMENTALS
 Total Quality Management has been introduced in the Industrial
level under the pressure of the Production and satisfying
customer with the Quality.
 The principle of Quality Management, assurance & control have
become foundation for Clinical Laboratories.
 Quality is defined as Conformance with the requirement of
users or customers.
 Directly, Quality refers to satisfaction of the needs and
expectations of users or customers.  1) Acceptable
 2) Accessible
 3) Affordable
 4) Appropriate
 Who are our customers?
are often nurses, physicians, patients, patient parties or who
pay the bills.
• Quality also includes:
a) Total Quality Management (TQM)→ an activity to improve pt.
care by having the lab monitor, its work to detect deficiencies
& subsequently correct them
b) Continuous Quality Improvement (CQI) or Performance
Improvement (PI)→ to improve the pt. care by placing the
emphasis on not to make mistakes in the first place
c) Quality Assurance (QA)→ external activities that ensure positive
pt. outcomes. It measures what a lab can do to improve
reliability
 Now, Cost has to be even understood in the context of Quality.
WHY
Quality means conformance with requirements, then Quality Cost
be understood as „cost of Conformance‟ & „Cost Of Non-
conformance‟
 This understanding of quality and cost leads to a new
perspective on the relationship between them.
 Improvement in Quality can lead to reduction in cost
eg: with better analytical quality, lab would be able to reduce
waste.
 Quality improvement occur when problems are eliminated
permanently.
 This can only be done at the management level (85%).
 This emphasis on processes leads to a new view of the
organization as a system of processes.
 Primary importance of these processes for accomplishing the
work of the organization, TQM views the organization as a
support structure rather than as a command structure.
PROBLEM-SOLVING METHODS
 These methods outline distinct steps for
(1) Carefully defining the problem,
(2) Establishing baseline measures of process performance,
(3) Identifying root causes of the problem,
(4) Identifying a remedy for the problem,
(5) Verifying that the remedy actually works,
(6) “Standardizing” or generalizing the solution for routine implementation of
an improved process, and
(7) Establishing on going measures for monitoring and controlling the process.
 Hence making the TQM organization process a flexible rather
than rigid.
TOTAL QUALITY MANAGEMENT OF THE CLINICAL
LABORATORY
 The principle and concept of
TQM has been formulated into
a quality management
procedure.
Traditional
framework
QLP
QC
QA
 QLP includes analytical processes and the general policies,
practices, and procedures that define how work is done.
 QC emphasizes statistical control procedures but also includes
nonstatistical check procedures, such as linearity checks,
reagent and standard checks, and temperature monitors.
 QA, as currently applied, is primarily concerned with broader
measures and monitors of laboratory performance, such as
(1) turnaround time,
(2) specimen identification,
(3) patient identification, and
(4) test utility
 QI- a structured problem solving process to help identify the
root cause of a problem & a remedy for that problem
 QP- provides the planning steps
 The “five-Q” framework also
defines how quality is able to
be managed objectively using
the “scientific method”
 PDCA Cycle ( Plan, Do, Check,
Act)
PDCA
Cycle
Plan
Do
Check
Act
Plan (QP)
DO (QLP)
Check
(QC & QA)
Act (QI)
 The main objective of a laboratory is to provide
reliable, timely and accurate test results. This is only
possible through
 consistent monitoring and evaluation of the laboratory‟s
performance
 the implementation and follow-up of corrective actions for
non-conformance to procedure
• A reputation based on 10,000 good quality result is damaged
by 1 poor quality result
• The establishment of robust testing systems is essential for the
success of laboratory services and the treatment program
• The establishment and continuous monitoring of quality
systems in each testing laboratory is required for reliable and
robust testing
• Provide the assurance of the quality of the test results
• Ensure that the lab operations are coordinated, organised and
standardised
• Are tools to monitor performance
ELEMENTS OF QUALITY ASSURANCE PROGRAM
 Quality assurance is used to represent practices that are
generally recommended for ensuring that desired quality goals.
 Broad spectrum of plans, policies, procedure.
 These techniques monitor particular source of errors, estimate
magnitude of errors, alert lab personnel when indicator
deteriorated.
Commitment
Facilities and Resources
Personnel Competency and Training
Technical Procedures
Problem-Solving Mechanism
The Quality Assurance Cycle
•Data and Lab
Management
•Safety
•Customer
Service
Patient/Client Prep
Sample Collection
Sample Receipt
and Accessioning
Sample Transport
Quality Control
Record Keeping
Reporting
Personnel Competency
Test Evaluations
Testing
26
LABORATORY ERRORS
QUALITY CONTROL IN LABORATORY
• Quality control is a vital part of quality assurance
– All labs benefit from quality control in terms of confidence in
and reproducibility of test results
Quality in labs is mutual responsibility of…..
• Laboratory specialists
• Clinicians
• Public health physicians
29
Objectives of quality in lab
• Support provision of high quality health-care
– Reduce morbidity
– Reduce mortality
– Reduce economic loss
• Ensure credibility of lab
• Generate confidence in lab results
30
Consequences of poor quality
• Inappropriate action
– Over-investigation
– Over-treatment
– Mistreatment
• Inappropriate inaction
– Lack of investigation
– No treatment
• Delayed action
• Loss of credibility of laboratory
• Legal action
31
CONTROLS OF PREANALYTICAL VARIABLES
Responsibility for accurate and timely reporting lies with
lab, but many problems arise before and after
analysis.
To monitor such issues, it is necessary to perform
system analysis and identify types of preanalytical
variables
TYPES OF PREANALYTICAL VARIABLES
1. Test usage & practice Guidelines: appropriate use of testing
procedure to cut the cost bearing. Test which require in
emergency care.
2. Patient Identification: correct identification of patient and
specimen is a major concern. Highest error occur with the
hand written labels. The single id number should be allotted
for patient. Barcode preferred (reduce significant problems).
3. Turnaround Time: delayed & lost test requisition, specimen &
report have been major problem. Automatic computer
generated document & tracking system flow.
4. Laboratory Log: tubes once arrive, various logging and
monitoring system is required. Computerized specimen
receiving and feed back to data-base.
Specimen should be inspected to confirm adequacy of volume,
problems which would interfere assay.
5. Transcription error : if electronic identification and tracking not
been implemented.
6. Patient Preparation: such as recent intake of food, alcohol,
drugs, smoking, exercise, stress, sleep, posture during
specimen collection. Hence proper patient preparation is
essential
7. Specimen collection: technique used will affect the test; eg:
prolong tourniquet application, IV line, Hemolysis, improper
container/preservatives, identification of person itself is
important.
8. Specimen Transport: stability of specimen during transport,(
ABG, Amm, Lact, HCO3, PTH). Controlling the specimen
transport is essential, authority to reject specimen.
9. Centrifuge performance: the calibration, speed, type of tube,
timer & temperature.
10. Specimen separation and aliquoting:
11. Container monitoring: collection tube, pipettes, stoppers,
aliquot tubes. Interfere like cork stopper for calcium (false
elevation 10-50%).
12. Personnel monitoring: trained and supervised staff. Written
procedure manual available at area of work. Fast-track
mechanism for urgent samples. Throughout time monitoring,
weekly/monthly statistics,
However records are necessary for documenting the daily
operation.
QUALITY MANAGEMENT-II
Presenter: Dr Anurag Yadav
Moderator: Dr Malathi M
CONTROL OF ANALYTICAL VARIABLES
Factors influencing analytical variables
EQUIPMENT RELIABILITY:
Meet technical needs, Compatible, User
& maintenance friendly, Cost effective,
Validated
Procedural
reliability using
Standard Operating
Procedures
REAGENTS STABILITY, INTEGRITY AND
EFFICIENCY:
Stable, Efficient, Desired quality,
Continuously available, Validated
SPECIFICITY & SENSITIVITY
OF SELECTED TEST:
Adequate ST, Sufficient SP,
cost effective, compatible
with, available infrastructure
and expertise, interpretable,
meets the needs/ objectives,
validated
PROFICIENCY OF PERSONNEL:
Education, Training, Aptitude,
Competence, Commitment,
Adequate number, CME,
Supervision, Motivation
USE OF APPROPRIATE
CONTROLS:
• Internal: Labs, Calibrated
• External: Supplied by
manufacturer, National,
International
DOCUMENTATION:
All the written policies, plans, procedures,
instructions and records, quality control
procedures and recorded test results involved
in providing a service or the manufacture of a
product
Assessment
ANALYTICAL FACTORS
41
 Analytical variables must be controlled carefully to ensure accurate
measurements by analytical methods
 Reliable analytical methods are obtained through careful process of :
(1) Selection
(2) Evaluation
(3) Implementation
(4) Maintenance
(5) Control
 Clear concepts in relation to Analytical methods- Calibration, trueness,
accuracy, precision, linearity, & limit of detection
CONTROL OF ANALYTICAL VARIABLES
42
CALIBRATION
 The calibration function is the relation between instrument signal(y) and conc. of
analyte (x)
y=f(x)
or, x=f1(y)
 It is set of operations that establish the relationship between values of
quantities indicated by the instrument and the corresponding values realized by
”measurement standards”
 Calibration function may be linear or curved graphically
 In automated clinical chemistry instruments , the relation b/n analyte conc. and
signal is often very stable
43
Calibrators
• Has a known concentration of the substance
(analyte) being measured
• Used to adjust instrument, kit, test system in order
to standardize the assay
• Sometimes called a standard, although usually not
a true standard
• This is not a control
44
TRACEABILITY OF CALIBRATOR
TRUENESS
 Trueness of measurements- closeness of agreement b/n the
average value obtained from a large series of results of
measurements and a true value
 True value- The known, accepted value of a quantifiable property
 Bias- The difference b/n average value & the true value
 Measured value- Result of an individual‟s measurement of a
quantifiable property
47
Accuracy
• Accuracy- the closeness of the agreement
b/n the result of a measurement & a true
conc. of the analyte
48
Precision
• Precision- closeness of agreement b/n
independent results of measurements obtained
under stipulated conditions
• It is the dispersion of replicate measurements
• Inter-assay with multiple runs
• Intra-assay: within a run
49
Precision
• Repeatability (within same run)
• Reproducibility or closeness of results to each other performed under
changed conditions of measurement (time,operator,calibrators, reagent lots)
– Measured by the standard deviation or coefficient of variation
• Std. dev (σ)/mean(X) x 100 = % C.V.
50
Reliability
• The ability to maintain both precision and accuracy
51
LINEARITY
 Linearity refers to the relationship between measured &
expected values over the analytical measurement range
 May be considered in relation to actual or relative analyte
concentration
 Evaluated by plotting measured& expected conc.,
 Testing of linearity is related to assessment of trueness over
the analytical measurement range
 The presence of linearity is a prerequisite for a high degree of
trueness
52
LIMIT OF DETECTION
 Limit of Detection (LoD): lowest concentration of measured that
can be consistently detected in ≥ 95% of samples tested under
routine laboratory conditions and in a given sample matrix
53
Variables that may cause
imprecision
• Equipment
– Multiple instruments
– Pipettes – check precision
– Sporadic maintenance
• Reagents
– Different Lots of Reagents lot-to-lot evaluation
• Staff
– Difference in training, competencies
54
Laboratory Equipment
• All equipment in the laboratory
– Should have instruction manuals regarding proper
use and maintenance requirements
– Should be monitored and recorded for quality
control procedures, function checks, preventative
maintenance and repairs
• These should be documented and filed in separate log
books →55
Laboratory Equipment
• Before putting new equipment or a new method into service – it must be
validated
- This is accomplishment by correlation and/or agreement studies
- The new method or equipment is validated against old method and/or
equipment
• Refrigerators and freezers
- Record serial numbers
- Record temperatures daily
Maintaining correct temperatures is vital to maintaining the integrity of
reagents and should be maintained as per manufacturers‟ instructions.
56
Instrument Maintenance
• CONTROLS and CALIBRATORS
Manufacturers can provide calibrators and internal controls
• REGULAR MAINTENANCE
– Daily and weekly instrument maintenance
– Monthly, six-monthly and annual maintenance as recommended by
supplier
• ROUTINE MAINTENANCE
Ensure that instruments are serviced regularly by a
specialised service engineer and that this maintenance is
documented
57
Instrument Maintenance
Required for:
•Producing reliable test results
•Minimizing instrument breakdown
•Lowering repair costs
•Preventing delays in reporting test results
•Maintaining productivity
•Lengthening instrument life
58
PIPETTE’S IMPACT ON QC
•Pipette accuracy and precision must checked regularly -
the first time of use and periodically thereafter
•If either fails, it is important to follow the manufacturer‟s
instructions for repair and calibration
• Improperly calibrated pipettes will affect our assay and
should be checked for precision and accuracy bi-annually
59
Reagents in Laboratory
• How to store reagents
– Always store according to the manufacturer‟s
recommendations
– Reagents must be dated and initialed upon receipt
– Lot numbers must be recorded in a reagent quality control
record book
– After preparation and/or when placed in service, reagents
must be labeled when put “in service” according to the
manufacturer‟s suggested recommendations
→60
• Reagent parallel testing
– New reagent lots must be checked with old lots using a normal control
before use
– The variability for new lots of reagents compared to the current lot
should not be greater than the variability found for triplicate samples of
the current lot
– Variability should be within 5%
– Results of reagent checks must be recorded, dated and initialed
– Document all lot to lot procedures with date and variability results
Reagents in Laboratory
61
• Restrict all testing procedures to staff with
appropriate technical training
– Testing theory
– Instruments
– Testing procedures
• Perform and document periodic performance
assessments on all testing staff
PERSONNEL (Staff)
62
PERSONNEL (Staff)
 Active participation by everyone working in the system is required
to meet quality standards & continuously improve performance
 It is laboratory director's responsibility to employ sufficient
qualified personnel for the volume and complexity of the work
performed
 Continuing education program should be provided
 All documentation should maintained in personnel file
 Regular meeting to keep staff informed of changes & to solicit
their suggestion for improving the lab. service
63
Documentation
• If you have not documented it,
you have NOT done it …
• If you have documented,
it is a RUMOUR !!!
64
Value of Documentation
• Ensures processes and outcomes are traceable
• Processes can be audited, thus external
assessments can take place
• Tool for training
• Reminds what to do next
65
Key Documents
• Results archive
– File results of the Quality control, calibration sheets, maintenance log report,
backup of patient sample results., etc, in an organised and easily accessible
manner
• Laboratory monitoring
– Documentation of temperature monitoring, reagent control,
accuracy/precision assays, corrective actions, audit reports
66
• Instrumentation
– SOPs
– Equipment files + Manuals
– Service history records of the instrument
– Records of daily, weekly and monthly calibrations and maintenance
Standard Operating Procedures (SOP)
• comprehensively written
document that describes the
laboratory procedure and all
other related issues
• Essential for ensuring
uniformity in laboratory
procedures
SOP for TFT
•The SOP should define test performance, tolerance
limits, reagent preparation, required quality control,
result reporting and references
68
• The SOPM should be written in CLSI format and must be
reviewed and signed annually by the Laboratory director
• Must contain all test methods performed by the laboratory
• The SOPM should be available in the work area. It is the
definitive laboratory reference and is used often for
questions relating to individual test
• Any obsolete procedures should be dated when removed
from SOPM and retained for at least 2 years
69
OUTLINE FOR A CLSI PROCEDURE DOCUMENT
A. Title: intent of the document; concise
B. Purpose or Principle: “the process describes..” info.
regarding theory, clinical implications of examination
C. Procedure instruction: “how to do” a particular steps,
steps involved
D. Related Documents: listing of other procedures used
E. References: reference the source of information
F. Appendixes or attachments: additional information
G. Author(s): author of the document should be
documented
H. Approved Signature: Evidence that the document has
been approved
70
Validation
 is about determining whether
something does what it is supposed
to do
71
Importance of validation
• Validation - before we introduce something
• Re-validation
– after we have changed/modified
– periodic
• Validation is applied to:
– SOP
– reagents
– equipment
– software
72
Factors influencing quality:
Post-analytical
• Right recording and reporting
• Right interpretation
– Range of normal values
• Right turnaround time
• Report to right user
73
Reporting results
• Proper procedure includes:
– All data entry results should be verified by a section head
or supervisor (when available) and reviewed by
management for final interpretation and release of results
– In the event that a report has already been sent out and
needs correction, a new report is issued with updated
report written on it
• The old report remains in the patient file
– Verbal result reports should be documented, listing the
time of the receipt of the report 74
Housekeeping
• Surface decontamination of instruments,
benchtops, biosafety hoods and general tidiness
• Disposal of biohazardous waste
• Monitoring of fridge, freezer and incubator
temperatures
75
STATISTICAL CONTROL OF ANALYTICAL
METHODS
 Statistics definitions :
- Mean, Mode, Median, Standard Deviation
- Coefficient of Variation (CV) is the ratio of the SD to the average. It
is a measure of relative precision expressed as percentage
 Reference ranges :
As per IFCC recommendations,
76
Reference ranges
• What do I do when controls are out of established
reference ranges?
– Results of the normal donor control are expected to be
within the established reference range. If results
exceed reference range limits, follow corrective action:
• Repeat test using same aliquot
• If the results still exceed the limits, do not automatically
invalidate patient results
– Due to the 95% confidence interval, 1 in 20 specimens from healthy
individuals drawn at random can be outside reference range limits
due to biological factors
77
CONTROL MATERIALS
 Specimens that are analyzed for QC purpose are known
as control materials
 Control materials should be available :
-in a stable form
-in aliquots or vials
-for analysis over an extended period of time, at least 1 yr
 Minimal vial-to-vial variation should exist
 Should have preferably the same matrix as the specimen
78
Types of Control Materials
• Assayed
– mean calculated by the manufacturer
– must verify in the laboratory
• Unassayed
– less expensive
– must perform data analysis
• “Homemade” or “In-house”
– pooled sera collected in the laboratory
– characterized
– preserved in small quantities for daily use
79
Managing Control Materials
• Sufficient material from same lot number or serum
pool for one year‟s testing
• May be frozen, freeze-dried, or chemically preserved
according to instructions by the manufacturer
• Requires very accurate reconstitution if this step is
necessary
• Always store as recommended by manufacturer
80
CONTROL CHARTS
 A common method to compare the values observed for control
materials with their known values is the use of control charts
 Simple graphical displays in which the observed values are plotted
versus the time when the observations are made
 Known values are represented by an acceptable range of values
 When plotted points falls within the control limit- method is
performing properly
 When points falls outside control limit – problem may be
developing
81
LEVY- JENNINGS GRAPH
L-Js are the process control graphs wherein the daily
Q.C. values for all levels of controls are plotted
(minimum 20 values) and an inference about the run
is drawn , to decide “in control” or “out of control
run.”
Advantage:
•Simple data analysis and display
•Easy adaptation and integration into existing control
practices
•A low level of false rejection or false alarms
•An improved capability for detecting systematic and
random errors 82
LEVEY - JENNINGS GRAPH
• It should be automatically generated for the
parameters chosen in a QC template
• Should be available for viewing by day,
month, and other fractions of the year
• Use limit criteria either calculated from the data,
or from user defined means and SDs
• Use red for data points that are outliers
83
LJ CHART PLOTTED : AN EXAMPLE
84
Westgard Rules
(Generally used where 2 levels of control
material are analyzed per run)
• 12S rule
• 13S rule
• 22S rule
• R4S rule
• 41S rule
• 10X rule
85
Westgard – 12S Rule
• Warning rule
• One of two control results falls outside ±2SD
• Alerts tech to possible problems
• Not cause for rejecting a run
• Must then evaluate the 13S rule
86
12S Rule = A warning to trigger careful
inspection of the control data
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
12S rule
violation
87
Westgard – 13S Rule
• If either of the two control results falls outside of
±3SD, rule is violated
• Primarily sensitive to random error
• Run must be rejected
• If 13S not violated, check 22S
88
13S Rule = Reject the run when a single control
measurement exceeds the +3SD or -3SD control limit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
13S rule
violation
89
Westgard – 22S Rule
• 2 consecutive control values for the same level fall
outside of ±2SD in the same direction
• Sensitive to systematic error
• Patient results cannot be reported
• Requires corrective action
90
22S Rule = Reject the run when 2 consecutive
control measurements exceed the same
+2SD or -2SD control limit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
22S rule
violation
91
Westgard – R4S Rule
• One control exceeds the mean by –2SD, and the
other control exceeds the mean by +2SD
• The range between the two results will therefore
exceed 4 SD
• Random error has occurred, test run must be
rejected
92
R4S Rule = Reject the run when 1 control
measurement exceed the +2SD and the other
exceeds the -2SD control limit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
R4S rule
violation
93
Westgard – 41S Rule
• Requires control data from previous runs
• Four consecutive QC results for one level of control
are outside ±1SD
• Sensitive to systematic error
94
Westgard – 10X Rule
• Requires control data from previous runs
• Ten consecutive QC results for one level of control
are on one side of the mean (above or below, with
no other requirement on the size of deviation)
• Sensitive to systematic error
95
10x Rule = Reject the run when 10 consecutive
control measurements fall on one side of the mean
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
10x rule
violation
96
CONTROL OF ANALYTICAL QUALITY USING
PATIENT DATA
 Patient result is the final product of lab
procedures, hence the lab shall have the
monitoring results.
 Clinical correlation
 Correlation with other Lab tests
 Intralaboratory duplicates
 Delta checks with previous test results
 Limit checks
EXTERNAL QUALITY ASSESSMENT
 All the control procedures described previously have
focused on monitoring of a single lab
 These procedures constitute internal QC, to
distinguish them from procedures used to compare
the performance of different laboratories are known
as external Quality assessment
 These two procedures are complimentary
 Internal QC- daily monitoring of precision & accuracy
 External QA- long term accuracy of analytical methods
98
• EQA results evaluate performance of the laboratory against other
laboratories participating in the same program
• Different programs do this in different ways.
• When the diff. is significant lab. is alerted
• Results are instrument and protocol specific
• EQA results should be formally documented within the lab and
should be available on request
External Quality Assessment
99
 In India EQA is done by Govt. / private organizations
- National Accreditation Board for Testing & Calibration
Laboratories (NABL) , Dept. of Science &
Technology,GOI.(http://www.nabl-india.org)
- ACBI-CMC External Quality Assurance Scheme,
Christian Medical College, Dept. of Clinical
Biochemistry, Chennai(http://www.cmcvellore.ac.in)
External Quality Assessment in India
100
Proficiency testing
• Laboratories should all enroll and satisfactorily participate in a
performance evaluation/assessment program
– If conventional proficiency testing is not available, the
laboratory must exercise an alternative performance
assessment system for determining the reliability of analytic
testing (sample splitting for inter-laboratory testing)
– If the lab has more than one method-system for performing
tests for an analyte, it must be checked against each other at
least twice a year for correlation of patient results
101
NEW QUALITY INITIATIVES
 Several new quality initiatives have been developed &
implemented to ensure that labs incorporate the principles of
Quality Management & QA in daily operation
- Six Sigma process
- Lean Production
- ISO 9000
- Joint Committee for Traceability in Laboratory Medicine (JCTLM)
Guidelines.
102
103
To ensure that a laboratory practices and provides quality assurance for all
phases of testing, the laboratory should:
 Enroll and satisfactorily participate in a performance
evaluation/assessment program
 Establish reference ranges for analytes being tested
 Document training and competency assessment for their technicians
 Provide review and verification of all results released, including verbal result
reports
→
SUMMARY
SUMMARY
 Quality control is a part of a total laboratory control program under
Total Quality Management
 TQM is responsible for organizational development and management
for improved quality in all aspects
 Five Qs- Qlp,QC,QA,QI,QP constitute the TQM framework
 Quality Control is achieved through proper documented and validated
interventions at Preanalytical, analytical and post analytical stages
 Quality Assurance is internal Quality assessment plus external Quality
assessment
X
104
 No matter how good the quality system is on
paper, quality cannot be achieved if the theory
cannot be translated into practice
Quality costs,
but poor quality costs more …
105
REFERENCE
 Tietz- textbook of clinical chemistry and Molecular
Diagnostics.

More Related Content

What's hot

Pre analytic and postanalytic test management
Pre analytic and postanalytic test managementPre analytic and postanalytic test management
Pre analytic and postanalytic test management
Varsha Shahane
 
Quality control in the medical laboratory
Quality control in the medical laboratoryQuality control in the medical laboratory
Quality control in the medical laboratory
Adnan Jaran
 

What's hot (20)

Quality control
Quality controlQuality control
Quality control
 
INTERFERENCES IN CLINICAL CHEMISTRY ANALYSIS
INTERFERENCES IN CLINICAL CHEMISTRY  ANALYSISINTERFERENCES IN CLINICAL CHEMISTRY  ANALYSIS
INTERFERENCES IN CLINICAL CHEMISTRY ANALYSIS
 
WESTGARD RULES
WESTGARD RULESWESTGARD RULES
WESTGARD RULES
 
Quality control
Quality controlQuality control
Quality control
 
Laboratory Method Verification, March 2017
Laboratory Method Verification, March 2017Laboratory Method Verification, March 2017
Laboratory Method Verification, March 2017
 
Troubleshooting IQC / EQAS
Troubleshooting IQC / EQASTroubleshooting IQC / EQAS
Troubleshooting IQC / EQAS
 
PRE AND POST ANALYTICAL ERRORS
PRE AND POST ANALYTICAL ERRORSPRE AND POST ANALYTICAL ERRORS
PRE AND POST ANALYTICAL ERRORS
 
Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014
 
Pre analytic and postanalytic test management
Pre analytic and postanalytic test managementPre analytic and postanalytic test management
Pre analytic and postanalytic test management
 
Quality Control in Laboratory
Quality Control in LaboratoryQuality Control in Laboratory
Quality Control in Laboratory
 
Quality Control in Clinical Chemistry
Quality Control in Clinical ChemistryQuality Control in Clinical Chemistry
Quality Control in Clinical Chemistry
 
Quality control in the medical laboratory
Quality control in the medical laboratoryQuality control in the medical laboratory
Quality control in the medical laboratory
 
Laboratory accreditation
Laboratory accreditationLaboratory accreditation
Laboratory accreditation
 
Quality control
Quality controlQuality control
Quality control
 
Medical Laboratory Quality Assurance
Medical Laboratory Quality AssuranceMedical Laboratory Quality Assurance
Medical Laboratory Quality Assurance
 
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptxANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
ANALYTICAL VARIABLES IN QUALITY CONTROL.pptx
 
Automation in the Clinical Lab
Automation in the Clinical LabAutomation in the Clinical Lab
Automation in the Clinical Lab
 
quality control in clinical pathology
quality control in clinical pathologyquality control in clinical pathology
quality control in clinical pathology
 
Quality assurance in medical laboratory
Quality assurance in medical laboratoryQuality assurance in medical laboratory
Quality assurance in medical laboratory
 
Laboratory accreditation by iso 15189
Laboratory accreditation by iso 15189Laboratory accreditation by iso 15189
Laboratory accreditation by iso 15189
 

Similar to Total Quality Management (TQM) by Dr Anurag Yadav

Quality control in a virology laboratory.pdf
Quality control in a virology laboratory.pdfQuality control in a virology laboratory.pdf
Quality control in a virology laboratory.pdf
samwel18
 

Similar to Total Quality Management (TQM) by Dr Anurag Yadav (20)

TOTAL QUALITY MANAGEMENT.pptx
TOTAL QUALITY MANAGEMENT.pptxTOTAL QUALITY MANAGEMENT.pptx
TOTAL QUALITY MANAGEMENT.pptx
 
Quality.pptx
Quality.pptxQuality.pptx
Quality.pptx
 
Introduction qa, qc and tqm
Introduction qa, qc and tqmIntroduction qa, qc and tqm
Introduction qa, qc and tqm
 
Laboratory quality management notes
Laboratory quality management notesLaboratory quality management notes
Laboratory quality management notes
 
Clia final regulations
Clia final regulationsClia final regulations
Clia final regulations
 
Lect.2-what is QC -QA_R.pdf
Lect.2-what is QC -QA_R.pdfLect.2-what is QC -QA_R.pdf
Lect.2-what is QC -QA_R.pdf
 
Significance of CAP accreditation
Significance of CAP accreditationSignificance of CAP accreditation
Significance of CAP accreditation
 
Quality control in a virology laboratory.pdf
Quality control in a virology laboratory.pdfQuality control in a virology laboratory.pdf
Quality control in a virology laboratory.pdf
 
Perez.pptx
Perez.pptxPerez.pptx
Perez.pptx
 
Overview of Quality Control and its implementation in the laboratory.pptx
Overview of Quality Control and its implementation in the laboratory.pptxOverview of Quality Control and its implementation in the laboratory.pptx
Overview of Quality Control and its implementation in the laboratory.pptx
 
Quality Control.pptx leadership and management
Quality Control.pptx leadership and managementQuality Control.pptx leadership and management
Quality Control.pptx leadership and management
 
Quality Control, Quality Assurance and Total Quality Management in Dairy Indu...
Quality Control, Quality Assurance and Total Quality Management in Dairy Indu...Quality Control, Quality Assurance and Total Quality Management in Dairy Indu...
Quality Control, Quality Assurance and Total Quality Management in Dairy Indu...
 
Clinical laboratory total quality management (TQM) system
Clinical laboratory total quality management (TQM) systemClinical laboratory total quality management (TQM) system
Clinical laboratory total quality management (TQM) system
 
Lab management
Lab managementLab management
Lab management
 
surgical audit
surgical auditsurgical audit
surgical audit
 
CLO4 LEAN.pptx
CLO4 LEAN.pptxCLO4 LEAN.pptx
CLO4 LEAN.pptx
 
Quality
QualityQuality
Quality
 
Chapter 2 powerpoint
Chapter 2 powerpointChapter 2 powerpoint
Chapter 2 powerpoint
 
Quality Assurance Qual Indicators.ppt
Quality Assurance Qual Indicators.pptQuality Assurance Qual Indicators.ppt
Quality Assurance Qual Indicators.ppt
 
Laboratory quality management, hospital admin hht
Laboratory quality management, hospital admin hhtLaboratory quality management, hospital admin hht
Laboratory quality management, hospital admin hht
 

More from Dr Anurag Yadav

More from Dr Anurag Yadav (20)

Biological oxidation and Electron transport chain (ETC).pptx
Biological oxidation and Electron transport chain (ETC).pptxBiological oxidation and Electron transport chain (ETC).pptx
Biological oxidation and Electron transport chain (ETC).pptx
 
Benefits & Opportunities of research for medical undergraduate students by Dr...
Benefits & Opportunities of research for medical undergraduate students by Dr...Benefits & Opportunities of research for medical undergraduate students by Dr...
Benefits & Opportunities of research for medical undergraduate students by Dr...
 
Lipid chemistry V2 By Dr Anurag Yadav
Lipid chemistry V2 By Dr Anurag YadavLipid chemistry V2 By Dr Anurag Yadav
Lipid chemistry V2 By Dr Anurag Yadav
 
Lipid Chemistry by Dr Anurag Yadav
Lipid Chemistry by Dr Anurag YadavLipid Chemistry by Dr Anurag Yadav
Lipid Chemistry by Dr Anurag Yadav
 
Vitamin C and Vit B1 to B6 by Dr Anurag Yadav
Vitamin C and Vit B1 to B6 by Dr Anurag YadavVitamin C and Vit B1 to B6 by Dr Anurag Yadav
Vitamin C and Vit B1 to B6 by Dr Anurag Yadav
 
Calcium metabolism by Dr Anurag Yadav
Calcium metabolism by Dr Anurag YadavCalcium metabolism by Dr Anurag Yadav
Calcium metabolism by Dr Anurag Yadav
 
Quality control and automation by Dr Anurag Yadav
Quality control and automation by Dr Anurag YadavQuality control and automation by Dr Anurag Yadav
Quality control and automation by Dr Anurag Yadav
 
Carbohydrate chemistry by Dr Anurag Yadav
Carbohydrate chemistry by Dr Anurag YadavCarbohydrate chemistry by Dr Anurag Yadav
Carbohydrate chemistry by Dr Anurag Yadav
 
Iron metabolism by Dr Anurag Yadav
Iron metabolism by Dr Anurag YadavIron metabolism by Dr Anurag Yadav
Iron metabolism by Dr Anurag Yadav
 
Plasma proteins by Dr Anurag Yadav
Plasma proteins by Dr Anurag YadavPlasma proteins by Dr Anurag Yadav
Plasma proteins by Dr Anurag Yadav
 
Glassware and laboratory equipment, biochemistry by Dr. Anurag Yadav
Glassware and laboratory equipment, biochemistry by Dr. Anurag YadavGlassware and laboratory equipment, biochemistry by Dr. Anurag Yadav
Glassware and laboratory equipment, biochemistry by Dr. Anurag Yadav
 
Citric acid cycle (TCA cycle) by Dr. Anurag Yadav
Citric acid cycle (TCA cycle) by Dr. Anurag YadavCitric acid cycle (TCA cycle) by Dr. Anurag Yadav
Citric acid cycle (TCA cycle) by Dr. Anurag Yadav
 
Vitamins A, D, E, K by Dr. Anurag Yadav
Vitamins A, D, E, K by Dr. Anurag YadavVitamins A, D, E, K by Dr. Anurag Yadav
Vitamins A, D, E, K by Dr. Anurag Yadav
 
Amino acids and Protein chemistry by Dr. Anurag Yadav
Amino acids and Protein chemistry by Dr. Anurag YadavAmino acids and Protein chemistry by Dr. Anurag Yadav
Amino acids and Protein chemistry by Dr. Anurag Yadav
 
Cell and cellular structure - Medical Biochemistry by Dr. Anurag Yadav
Cell and cellular structure - Medical Biochemistry by Dr. Anurag YadavCell and cellular structure - Medical Biochemistry by Dr. Anurag Yadav
Cell and cellular structure - Medical Biochemistry by Dr. Anurag Yadav
 
Hemoglobin structure and metabolism by Dr. Anurag Yadav
Hemoglobin structure and metabolism by Dr. Anurag YadavHemoglobin structure and metabolism by Dr. Anurag Yadav
Hemoglobin structure and metabolism by Dr. Anurag Yadav
 
Corona Virus by Dr Anurag Yadav
Corona Virus by Dr Anurag YadavCorona Virus by Dr Anurag Yadav
Corona Virus by Dr Anurag Yadav
 
Enzymes & isoenzymes by Dr. Anurag Yadav
Enzymes & isoenzymes by Dr. Anurag YadavEnzymes & isoenzymes by Dr. Anurag Yadav
Enzymes & isoenzymes by Dr. Anurag Yadav
 
Transcription of DNA to RNA by Dr. Anurag Yadav
Transcription of DNA to RNA by Dr. Anurag YadavTranscription of DNA to RNA by Dr. Anurag Yadav
Transcription of DNA to RNA by Dr. Anurag Yadav
 
DNA repair, DNA Mutation, Gene Expression by Dr. Anurag Yadav
DNA repair, DNA Mutation, Gene Expression by Dr. Anurag YadavDNA repair, DNA Mutation, Gene Expression by Dr. Anurag Yadav
DNA repair, DNA Mutation, Gene Expression by Dr. Anurag Yadav
 

Recently uploaded

Seal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptxSeal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptx
negromaestrong
 
Spellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please PractiseSpellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please Practise
AnaAcapella
 

Recently uploaded (20)

Holdier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdfHoldier Curriculum Vitae (April 2024).pdf
Holdier Curriculum Vitae (April 2024).pdf
 
Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024Mehran University Newsletter Vol-X, Issue-I, 2024
Mehran University Newsletter Vol-X, Issue-I, 2024
 
Key note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdfKey note speaker Neum_Admir Softic_ENG.pdf
Key note speaker Neum_Admir Softic_ENG.pdf
 
On National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan FellowsOn National Teacher Day, meet the 2024-25 Kenan Fellows
On National Teacher Day, meet the 2024-25 Kenan Fellows
 
Unit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptxUnit-IV- Pharma. Marketing Channels.pptx
Unit-IV- Pharma. Marketing Channels.pptx
 
Third Battle of Panipat detailed notes.pptx
Third Battle of Panipat detailed notes.pptxThird Battle of Panipat detailed notes.pptx
Third Battle of Panipat detailed notes.pptx
 
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptxBasic Civil Engineering first year Notes- Chapter 4 Building.pptx
Basic Civil Engineering first year Notes- Chapter 4 Building.pptx
 
Kodo Millet PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
Kodo Millet  PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...Kodo Millet  PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
Kodo Millet PPT made by Ghanshyam bairwa college of Agriculture kumher bhara...
 
Seal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptxSeal of Good Local Governance (SGLG) 2024Final.pptx
Seal of Good Local Governance (SGLG) 2024Final.pptx
 
How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17How to Create and Manage Wizard in Odoo 17
How to Create and Manage Wizard in Odoo 17
 
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdfUGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
UGC NET Paper 1 Mathematical Reasoning & Aptitude.pdf
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptx
 
Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptx
 
Spellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please PractiseSpellings Wk 3 English CAPS CARES Please Practise
Spellings Wk 3 English CAPS CARES Please Practise
 
psychiatric nursing HISTORY COLLECTION .docx
psychiatric  nursing HISTORY  COLLECTION  .docxpsychiatric  nursing HISTORY  COLLECTION  .docx
psychiatric nursing HISTORY COLLECTION .docx
 
Sociology 101 Demonstration of Learning Exhibit
Sociology 101 Demonstration of Learning ExhibitSociology 101 Demonstration of Learning Exhibit
Sociology 101 Demonstration of Learning Exhibit
 
Magic bus Group work1and 2 (Team 3).pptx
Magic bus Group work1and 2 (Team 3).pptxMagic bus Group work1and 2 (Team 3).pptx
Magic bus Group work1and 2 (Team 3).pptx
 
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
TỔNG ÔN TẬP THI VÀO LỚP 10 MÔN TIẾNG ANH NĂM HỌC 2023 - 2024 CÓ ĐÁP ÁN (NGỮ Â...
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...
 
Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...Making communications land - Are they received and understood as intended? we...
Making communications land - Are they received and understood as intended? we...
 

Total Quality Management (TQM) by Dr Anurag Yadav

  • 1. TOTAL QUALITY MANAGEMENT Dr Anurag Yadav Assistant Professor Department of Biochemistry Lean Six Sigma Green Belt
  • 2. CONTENT  Quality Management  Errors in Lab  Element of Quality assurance program  Control of errors in Variables  Preanalytical  Analytical  Using control material  Using Patient data  External quality assessment & Proficiency testing
  • 3. FUNDAMENTALS  Total Quality Management has been introduced in the Industrial level under the pressure of the Production and satisfying customer with the Quality.  The principle of Quality Management, assurance & control have become foundation for Clinical Laboratories.
  • 4.  Quality is defined as Conformance with the requirement of users or customers.  Directly, Quality refers to satisfaction of the needs and expectations of users or customers.  1) Acceptable  2) Accessible  3) Affordable  4) Appropriate
  • 5.  Who are our customers? are often nurses, physicians, patients, patient parties or who pay the bills.
  • 6. • Quality also includes: a) Total Quality Management (TQM)→ an activity to improve pt. care by having the lab monitor, its work to detect deficiencies & subsequently correct them b) Continuous Quality Improvement (CQI) or Performance Improvement (PI)→ to improve the pt. care by placing the emphasis on not to make mistakes in the first place c) Quality Assurance (QA)→ external activities that ensure positive pt. outcomes. It measures what a lab can do to improve reliability
  • 7.  Now, Cost has to be even understood in the context of Quality. WHY Quality means conformance with requirements, then Quality Cost be understood as „cost of Conformance‟ & „Cost Of Non- conformance‟
  • 8.
  • 9.  This understanding of quality and cost leads to a new perspective on the relationship between them.  Improvement in Quality can lead to reduction in cost eg: with better analytical quality, lab would be able to reduce waste.
  • 10.  Quality improvement occur when problems are eliminated permanently.  This can only be done at the management level (85%).  This emphasis on processes leads to a new view of the organization as a system of processes.
  • 11.
  • 12.  Primary importance of these processes for accomplishing the work of the organization, TQM views the organization as a support structure rather than as a command structure.
  • 13. PROBLEM-SOLVING METHODS  These methods outline distinct steps for (1) Carefully defining the problem, (2) Establishing baseline measures of process performance, (3) Identifying root causes of the problem, (4) Identifying a remedy for the problem, (5) Verifying that the remedy actually works, (6) “Standardizing” or generalizing the solution for routine implementation of an improved process, and (7) Establishing on going measures for monitoring and controlling the process.
  • 14.  Hence making the TQM organization process a flexible rather than rigid.
  • 15. TOTAL QUALITY MANAGEMENT OF THE CLINICAL LABORATORY  The principle and concept of TQM has been formulated into a quality management procedure. Traditional framework QLP QC QA
  • 16.
  • 17.  QLP includes analytical processes and the general policies, practices, and procedures that define how work is done.  QC emphasizes statistical control procedures but also includes nonstatistical check procedures, such as linearity checks, reagent and standard checks, and temperature monitors.
  • 18.  QA, as currently applied, is primarily concerned with broader measures and monitors of laboratory performance, such as (1) turnaround time, (2) specimen identification, (3) patient identification, and (4) test utility  QI- a structured problem solving process to help identify the root cause of a problem & a remedy for that problem  QP- provides the planning steps
  • 19.  The “five-Q” framework also defines how quality is able to be managed objectively using the “scientific method”  PDCA Cycle ( Plan, Do, Check, Act) PDCA Cycle Plan Do Check Act
  • 21.  The main objective of a laboratory is to provide reliable, timely and accurate test results. This is only possible through  consistent monitoring and evaluation of the laboratory‟s performance  the implementation and follow-up of corrective actions for non-conformance to procedure
  • 22. • A reputation based on 10,000 good quality result is damaged by 1 poor quality result • The establishment of robust testing systems is essential for the success of laboratory services and the treatment program • The establishment and continuous monitoring of quality systems in each testing laboratory is required for reliable and robust testing
  • 23. • Provide the assurance of the quality of the test results • Ensure that the lab operations are coordinated, organised and standardised • Are tools to monitor performance
  • 24. ELEMENTS OF QUALITY ASSURANCE PROGRAM  Quality assurance is used to represent practices that are generally recommended for ensuring that desired quality goals.  Broad spectrum of plans, policies, procedure.  These techniques monitor particular source of errors, estimate magnitude of errors, alert lab personnel when indicator deteriorated.
  • 25. Commitment Facilities and Resources Personnel Competency and Training Technical Procedures Problem-Solving Mechanism
  • 26. The Quality Assurance Cycle •Data and Lab Management •Safety •Customer Service Patient/Client Prep Sample Collection Sample Receipt and Accessioning Sample Transport Quality Control Record Keeping Reporting Personnel Competency Test Evaluations Testing 26
  • 28. QUALITY CONTROL IN LABORATORY • Quality control is a vital part of quality assurance – All labs benefit from quality control in terms of confidence in and reproducibility of test results
  • 29. Quality in labs is mutual responsibility of….. • Laboratory specialists • Clinicians • Public health physicians 29
  • 30. Objectives of quality in lab • Support provision of high quality health-care – Reduce morbidity – Reduce mortality – Reduce economic loss • Ensure credibility of lab • Generate confidence in lab results 30
  • 31. Consequences of poor quality • Inappropriate action – Over-investigation – Over-treatment – Mistreatment • Inappropriate inaction – Lack of investigation – No treatment • Delayed action • Loss of credibility of laboratory • Legal action 31
  • 32. CONTROLS OF PREANALYTICAL VARIABLES Responsibility for accurate and timely reporting lies with lab, but many problems arise before and after analysis. To monitor such issues, it is necessary to perform system analysis and identify types of preanalytical variables
  • 33. TYPES OF PREANALYTICAL VARIABLES 1. Test usage & practice Guidelines: appropriate use of testing procedure to cut the cost bearing. Test which require in emergency care. 2. Patient Identification: correct identification of patient and specimen is a major concern. Highest error occur with the hand written labels. The single id number should be allotted for patient. Barcode preferred (reduce significant problems).
  • 34. 3. Turnaround Time: delayed & lost test requisition, specimen & report have been major problem. Automatic computer generated document & tracking system flow. 4. Laboratory Log: tubes once arrive, various logging and monitoring system is required. Computerized specimen receiving and feed back to data-base. Specimen should be inspected to confirm adequacy of volume, problems which would interfere assay.
  • 35. 5. Transcription error : if electronic identification and tracking not been implemented. 6. Patient Preparation: such as recent intake of food, alcohol, drugs, smoking, exercise, stress, sleep, posture during specimen collection. Hence proper patient preparation is essential
  • 36. 7. Specimen collection: technique used will affect the test; eg: prolong tourniquet application, IV line, Hemolysis, improper container/preservatives, identification of person itself is important. 8. Specimen Transport: stability of specimen during transport,( ABG, Amm, Lact, HCO3, PTH). Controlling the specimen transport is essential, authority to reject specimen.
  • 37. 9. Centrifuge performance: the calibration, speed, type of tube, timer & temperature. 10. Specimen separation and aliquoting: 11. Container monitoring: collection tube, pipettes, stoppers, aliquot tubes. Interfere like cork stopper for calcium (false elevation 10-50%).
  • 38. 12. Personnel monitoring: trained and supervised staff. Written procedure manual available at area of work. Fast-track mechanism for urgent samples. Throughout time monitoring, weekly/monthly statistics, However records are necessary for documenting the daily operation.
  • 39. QUALITY MANAGEMENT-II Presenter: Dr Anurag Yadav Moderator: Dr Malathi M
  • 41. Factors influencing analytical variables EQUIPMENT RELIABILITY: Meet technical needs, Compatible, User & maintenance friendly, Cost effective, Validated Procedural reliability using Standard Operating Procedures REAGENTS STABILITY, INTEGRITY AND EFFICIENCY: Stable, Efficient, Desired quality, Continuously available, Validated SPECIFICITY & SENSITIVITY OF SELECTED TEST: Adequate ST, Sufficient SP, cost effective, compatible with, available infrastructure and expertise, interpretable, meets the needs/ objectives, validated PROFICIENCY OF PERSONNEL: Education, Training, Aptitude, Competence, Commitment, Adequate number, CME, Supervision, Motivation USE OF APPROPRIATE CONTROLS: • Internal: Labs, Calibrated • External: Supplied by manufacturer, National, International DOCUMENTATION: All the written policies, plans, procedures, instructions and records, quality control procedures and recorded test results involved in providing a service or the manufacture of a product Assessment ANALYTICAL FACTORS 41
  • 42.  Analytical variables must be controlled carefully to ensure accurate measurements by analytical methods  Reliable analytical methods are obtained through careful process of : (1) Selection (2) Evaluation (3) Implementation (4) Maintenance (5) Control  Clear concepts in relation to Analytical methods- Calibration, trueness, accuracy, precision, linearity, & limit of detection CONTROL OF ANALYTICAL VARIABLES 42
  • 43. CALIBRATION  The calibration function is the relation between instrument signal(y) and conc. of analyte (x) y=f(x) or, x=f1(y)  It is set of operations that establish the relationship between values of quantities indicated by the instrument and the corresponding values realized by ”measurement standards”  Calibration function may be linear or curved graphically  In automated clinical chemistry instruments , the relation b/n analyte conc. and signal is often very stable 43
  • 44. Calibrators • Has a known concentration of the substance (analyte) being measured • Used to adjust instrument, kit, test system in order to standardize the assay • Sometimes called a standard, although usually not a true standard • This is not a control 44
  • 46.
  • 47. TRUENESS  Trueness of measurements- closeness of agreement b/n the average value obtained from a large series of results of measurements and a true value  True value- The known, accepted value of a quantifiable property  Bias- The difference b/n average value & the true value  Measured value- Result of an individual‟s measurement of a quantifiable property 47
  • 48. Accuracy • Accuracy- the closeness of the agreement b/n the result of a measurement & a true conc. of the analyte 48
  • 49. Precision • Precision- closeness of agreement b/n independent results of measurements obtained under stipulated conditions • It is the dispersion of replicate measurements • Inter-assay with multiple runs • Intra-assay: within a run 49
  • 50. Precision • Repeatability (within same run) • Reproducibility or closeness of results to each other performed under changed conditions of measurement (time,operator,calibrators, reagent lots) – Measured by the standard deviation or coefficient of variation • Std. dev (σ)/mean(X) x 100 = % C.V. 50
  • 51. Reliability • The ability to maintain both precision and accuracy 51
  • 52. LINEARITY  Linearity refers to the relationship between measured & expected values over the analytical measurement range  May be considered in relation to actual or relative analyte concentration  Evaluated by plotting measured& expected conc.,  Testing of linearity is related to assessment of trueness over the analytical measurement range  The presence of linearity is a prerequisite for a high degree of trueness 52
  • 53. LIMIT OF DETECTION  Limit of Detection (LoD): lowest concentration of measured that can be consistently detected in ≥ 95% of samples tested under routine laboratory conditions and in a given sample matrix 53
  • 54. Variables that may cause imprecision • Equipment – Multiple instruments – Pipettes – check precision – Sporadic maintenance • Reagents – Different Lots of Reagents lot-to-lot evaluation • Staff – Difference in training, competencies 54
  • 55. Laboratory Equipment • All equipment in the laboratory – Should have instruction manuals regarding proper use and maintenance requirements – Should be monitored and recorded for quality control procedures, function checks, preventative maintenance and repairs • These should be documented and filed in separate log books →55
  • 56. Laboratory Equipment • Before putting new equipment or a new method into service – it must be validated - This is accomplishment by correlation and/or agreement studies - The new method or equipment is validated against old method and/or equipment • Refrigerators and freezers - Record serial numbers - Record temperatures daily Maintaining correct temperatures is vital to maintaining the integrity of reagents and should be maintained as per manufacturers‟ instructions. 56
  • 57. Instrument Maintenance • CONTROLS and CALIBRATORS Manufacturers can provide calibrators and internal controls • REGULAR MAINTENANCE – Daily and weekly instrument maintenance – Monthly, six-monthly and annual maintenance as recommended by supplier • ROUTINE MAINTENANCE Ensure that instruments are serviced regularly by a specialised service engineer and that this maintenance is documented 57
  • 58. Instrument Maintenance Required for: •Producing reliable test results •Minimizing instrument breakdown •Lowering repair costs •Preventing delays in reporting test results •Maintaining productivity •Lengthening instrument life 58
  • 59. PIPETTE’S IMPACT ON QC •Pipette accuracy and precision must checked regularly - the first time of use and periodically thereafter •If either fails, it is important to follow the manufacturer‟s instructions for repair and calibration • Improperly calibrated pipettes will affect our assay and should be checked for precision and accuracy bi-annually 59
  • 60. Reagents in Laboratory • How to store reagents – Always store according to the manufacturer‟s recommendations – Reagents must be dated and initialed upon receipt – Lot numbers must be recorded in a reagent quality control record book – After preparation and/or when placed in service, reagents must be labeled when put “in service” according to the manufacturer‟s suggested recommendations →60
  • 61. • Reagent parallel testing – New reagent lots must be checked with old lots using a normal control before use – The variability for new lots of reagents compared to the current lot should not be greater than the variability found for triplicate samples of the current lot – Variability should be within 5% – Results of reagent checks must be recorded, dated and initialed – Document all lot to lot procedures with date and variability results Reagents in Laboratory 61
  • 62. • Restrict all testing procedures to staff with appropriate technical training – Testing theory – Instruments – Testing procedures • Perform and document periodic performance assessments on all testing staff PERSONNEL (Staff) 62
  • 63. PERSONNEL (Staff)  Active participation by everyone working in the system is required to meet quality standards & continuously improve performance  It is laboratory director's responsibility to employ sufficient qualified personnel for the volume and complexity of the work performed  Continuing education program should be provided  All documentation should maintained in personnel file  Regular meeting to keep staff informed of changes & to solicit their suggestion for improving the lab. service 63
  • 64. Documentation • If you have not documented it, you have NOT done it … • If you have documented, it is a RUMOUR !!! 64
  • 65. Value of Documentation • Ensures processes and outcomes are traceable • Processes can be audited, thus external assessments can take place • Tool for training • Reminds what to do next 65
  • 66. Key Documents • Results archive – File results of the Quality control, calibration sheets, maintenance log report, backup of patient sample results., etc, in an organised and easily accessible manner • Laboratory monitoring – Documentation of temperature monitoring, reagent control, accuracy/precision assays, corrective actions, audit reports 66
  • 67. • Instrumentation – SOPs – Equipment files + Manuals – Service history records of the instrument – Records of daily, weekly and monthly calibrations and maintenance
  • 68. Standard Operating Procedures (SOP) • comprehensively written document that describes the laboratory procedure and all other related issues • Essential for ensuring uniformity in laboratory procedures SOP for TFT •The SOP should define test performance, tolerance limits, reagent preparation, required quality control, result reporting and references 68
  • 69. • The SOPM should be written in CLSI format and must be reviewed and signed annually by the Laboratory director • Must contain all test methods performed by the laboratory • The SOPM should be available in the work area. It is the definitive laboratory reference and is used often for questions relating to individual test • Any obsolete procedures should be dated when removed from SOPM and retained for at least 2 years 69
  • 70. OUTLINE FOR A CLSI PROCEDURE DOCUMENT A. Title: intent of the document; concise B. Purpose or Principle: “the process describes..” info. regarding theory, clinical implications of examination C. Procedure instruction: “how to do” a particular steps, steps involved D. Related Documents: listing of other procedures used E. References: reference the source of information F. Appendixes or attachments: additional information G. Author(s): author of the document should be documented H. Approved Signature: Evidence that the document has been approved 70
  • 71. Validation  is about determining whether something does what it is supposed to do 71
  • 72. Importance of validation • Validation - before we introduce something • Re-validation – after we have changed/modified – periodic • Validation is applied to: – SOP – reagents – equipment – software 72
  • 73. Factors influencing quality: Post-analytical • Right recording and reporting • Right interpretation – Range of normal values • Right turnaround time • Report to right user 73
  • 74. Reporting results • Proper procedure includes: – All data entry results should be verified by a section head or supervisor (when available) and reviewed by management for final interpretation and release of results – In the event that a report has already been sent out and needs correction, a new report is issued with updated report written on it • The old report remains in the patient file – Verbal result reports should be documented, listing the time of the receipt of the report 74
  • 75. Housekeeping • Surface decontamination of instruments, benchtops, biosafety hoods and general tidiness • Disposal of biohazardous waste • Monitoring of fridge, freezer and incubator temperatures 75
  • 76. STATISTICAL CONTROL OF ANALYTICAL METHODS  Statistics definitions : - Mean, Mode, Median, Standard Deviation - Coefficient of Variation (CV) is the ratio of the SD to the average. It is a measure of relative precision expressed as percentage  Reference ranges : As per IFCC recommendations, 76
  • 77. Reference ranges • What do I do when controls are out of established reference ranges? – Results of the normal donor control are expected to be within the established reference range. If results exceed reference range limits, follow corrective action: • Repeat test using same aliquot • If the results still exceed the limits, do not automatically invalidate patient results – Due to the 95% confidence interval, 1 in 20 specimens from healthy individuals drawn at random can be outside reference range limits due to biological factors 77
  • 78. CONTROL MATERIALS  Specimens that are analyzed for QC purpose are known as control materials  Control materials should be available : -in a stable form -in aliquots or vials -for analysis over an extended period of time, at least 1 yr  Minimal vial-to-vial variation should exist  Should have preferably the same matrix as the specimen 78
  • 79. Types of Control Materials • Assayed – mean calculated by the manufacturer – must verify in the laboratory • Unassayed – less expensive – must perform data analysis • “Homemade” or “In-house” – pooled sera collected in the laboratory – characterized – preserved in small quantities for daily use 79
  • 80. Managing Control Materials • Sufficient material from same lot number or serum pool for one year‟s testing • May be frozen, freeze-dried, or chemically preserved according to instructions by the manufacturer • Requires very accurate reconstitution if this step is necessary • Always store as recommended by manufacturer 80
  • 81. CONTROL CHARTS  A common method to compare the values observed for control materials with their known values is the use of control charts  Simple graphical displays in which the observed values are plotted versus the time when the observations are made  Known values are represented by an acceptable range of values  When plotted points falls within the control limit- method is performing properly  When points falls outside control limit – problem may be developing 81
  • 82. LEVY- JENNINGS GRAPH L-Js are the process control graphs wherein the daily Q.C. values for all levels of controls are plotted (minimum 20 values) and an inference about the run is drawn , to decide “in control” or “out of control run.” Advantage: •Simple data analysis and display •Easy adaptation and integration into existing control practices •A low level of false rejection or false alarms •An improved capability for detecting systematic and random errors 82
  • 83. LEVEY - JENNINGS GRAPH • It should be automatically generated for the parameters chosen in a QC template • Should be available for viewing by day, month, and other fractions of the year • Use limit criteria either calculated from the data, or from user defined means and SDs • Use red for data points that are outliers 83
  • 84. LJ CHART PLOTTED : AN EXAMPLE 84
  • 85. Westgard Rules (Generally used where 2 levels of control material are analyzed per run) • 12S rule • 13S rule • 22S rule • R4S rule • 41S rule • 10X rule 85
  • 86. Westgard – 12S Rule • Warning rule • One of two control results falls outside ±2SD • Alerts tech to possible problems • Not cause for rejecting a run • Must then evaluate the 13S rule 86
  • 87. 12S Rule = A warning to trigger careful inspection of the control data 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Mean Day +1SD +2SD +3SD -1SD -2SD -3SD 12S rule violation 87
  • 88. Westgard – 13S Rule • If either of the two control results falls outside of ±3SD, rule is violated • Primarily sensitive to random error • Run must be rejected • If 13S not violated, check 22S 88
  • 89. 13S Rule = Reject the run when a single control measurement exceeds the +3SD or -3SD control limit 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Mean Day +1SD +2SD +3SD -1SD -2SD -3SD 13S rule violation 89
  • 90. Westgard – 22S Rule • 2 consecutive control values for the same level fall outside of ±2SD in the same direction • Sensitive to systematic error • Patient results cannot be reported • Requires corrective action 90
  • 91. 22S Rule = Reject the run when 2 consecutive control measurements exceed the same +2SD or -2SD control limit 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Mean Day +1SD +2SD +3SD -1SD -2SD -3SD 22S rule violation 91
  • 92. Westgard – R4S Rule • One control exceeds the mean by –2SD, and the other control exceeds the mean by +2SD • The range between the two results will therefore exceed 4 SD • Random error has occurred, test run must be rejected 92
  • 93. R4S Rule = Reject the run when 1 control measurement exceed the +2SD and the other exceeds the -2SD control limit 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Mean Day +1SD +2SD +3SD -1SD -2SD -3SD R4S rule violation 93
  • 94. Westgard – 41S Rule • Requires control data from previous runs • Four consecutive QC results for one level of control are outside ±1SD • Sensitive to systematic error 94
  • 95. Westgard – 10X Rule • Requires control data from previous runs • Ten consecutive QC results for one level of control are on one side of the mean (above or below, with no other requirement on the size of deviation) • Sensitive to systematic error 95
  • 96. 10x Rule = Reject the run when 10 consecutive control measurements fall on one side of the mean 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 Mean Day +1SD +2SD +3SD -1SD -2SD -3SD 10x rule violation 96
  • 97. CONTROL OF ANALYTICAL QUALITY USING PATIENT DATA  Patient result is the final product of lab procedures, hence the lab shall have the monitoring results.  Clinical correlation  Correlation with other Lab tests  Intralaboratory duplicates  Delta checks with previous test results  Limit checks
  • 98. EXTERNAL QUALITY ASSESSMENT  All the control procedures described previously have focused on monitoring of a single lab  These procedures constitute internal QC, to distinguish them from procedures used to compare the performance of different laboratories are known as external Quality assessment  These two procedures are complimentary  Internal QC- daily monitoring of precision & accuracy  External QA- long term accuracy of analytical methods 98
  • 99. • EQA results evaluate performance of the laboratory against other laboratories participating in the same program • Different programs do this in different ways. • When the diff. is significant lab. is alerted • Results are instrument and protocol specific • EQA results should be formally documented within the lab and should be available on request External Quality Assessment 99
  • 100.  In India EQA is done by Govt. / private organizations - National Accreditation Board for Testing & Calibration Laboratories (NABL) , Dept. of Science & Technology,GOI.(http://www.nabl-india.org) - ACBI-CMC External Quality Assurance Scheme, Christian Medical College, Dept. of Clinical Biochemistry, Chennai(http://www.cmcvellore.ac.in) External Quality Assessment in India 100
  • 101. Proficiency testing • Laboratories should all enroll and satisfactorily participate in a performance evaluation/assessment program – If conventional proficiency testing is not available, the laboratory must exercise an alternative performance assessment system for determining the reliability of analytic testing (sample splitting for inter-laboratory testing) – If the lab has more than one method-system for performing tests for an analyte, it must be checked against each other at least twice a year for correlation of patient results 101
  • 102. NEW QUALITY INITIATIVES  Several new quality initiatives have been developed & implemented to ensure that labs incorporate the principles of Quality Management & QA in daily operation - Six Sigma process - Lean Production - ISO 9000 - Joint Committee for Traceability in Laboratory Medicine (JCTLM) Guidelines. 102
  • 103. 103 To ensure that a laboratory practices and provides quality assurance for all phases of testing, the laboratory should:  Enroll and satisfactorily participate in a performance evaluation/assessment program  Establish reference ranges for analytes being tested  Document training and competency assessment for their technicians  Provide review and verification of all results released, including verbal result reports → SUMMARY
  • 104. SUMMARY  Quality control is a part of a total laboratory control program under Total Quality Management  TQM is responsible for organizational development and management for improved quality in all aspects  Five Qs- Qlp,QC,QA,QI,QP constitute the TQM framework  Quality Control is achieved through proper documented and validated interventions at Preanalytical, analytical and post analytical stages  Quality Assurance is internal Quality assessment plus external Quality assessment X 104
  • 105.  No matter how good the quality system is on paper, quality cannot be achieved if the theory cannot be translated into practice Quality costs, but poor quality costs more … 105
  • 106. REFERENCE  Tietz- textbook of clinical chemistry and Molecular Diagnostics.