BASICPRINCIPLESOFMRI
VIJAY NAND CHAUDHARY
Magnetic Resonance
Imaging
MRI MACHINE
Ge Discovery 3T
750 MRI scaner
Seimens
INTRODUCTION
MRI stands for magnetic resonance imaging
MRI technique used in radiology to form picture of anatomy and physiological process of the
body.
MRI scanners use strong magnetic fields, magnetic field gradient and radiowaves to generate
images of the organs.
 Do not involve X-ray or use of ionizing radiation --- like CT Scan and PET scans.
Uses the application NMR – (investigation of chemical and physical properties at molecular
level)
Based on the principle of NMR
HISTORY
In 1997, the first MRI exam was performed on
human being.
FATHER OF MRI
Magnetic resonance imaging inventor ----- RAYMOND
VAHAN DAMADIAN
{ Before talking about MRI principle first look types of
magnets }
Types of Magnet
[1] Permanent magnets
A permanent magnet consists of a material, which
has been magnetized such that it won‟t loose its
magnetic field The field strength is usually very low
and ranges between 0.064T ~ 0.3T, Permanent
magnets have usually an open design, which is
more comfortable for the patient.
Toshiba‟s Access 0.064Tesla system. The Access
was the worlds first open MRI scanner.
Resistive magnets are very large electro magnets ,The magnetic
field is generated by a current, which runs through loops of
wire. The field strength can be up to 0.3 Tesla. They produce a
lot of heat, which requires watercooling. They need a lot of
power to run, and are usually switched off when not in use to
conserve power. They usually have an open design, which
reduces claustrophobia.
[2] Resistive Magnets
Hitachi‟s Airis 0.3 Tesla system.
[3] Superconducting magnets
Today‟s most commonly used magnets are
superconducting magnets. The magnetic field is
generated by a current, which runs through a loop of
wire. The wire is surrounded with a coolant, such as
liquid helium, to reduce the electric resistance of the
wire. At 4 Kelvin (-269º C) electric wire looses its
resistance. Once a system is energized, it won‟t loose
its magnetic field. Superconductivity allows for
systems with very high field strengths up to 12 Tesla.
The ones that are most used in clinical environments
run at 1.5 Tesla. Most superconducting magnets are
bore type magnets
examples of bore type magnets
• RF Coils
RF coils are needed to transmit and receive radio-frequency waves used in MRI scanners. RF coils are one of the
most important components that affect image quality. Current MRI scanners have a range of RF coils suitable to
acquire images of all body parts. There are two types of RF coils: volume coils and surface coils
[1] Volume RF Coils OR Quadrature Coils
The design of a volume coil is usually a saddle
shape, which guarantees a uniform RF field
inside the coil. Volume coils need to have the
area of examination inside the coil. They can
be used for transmit and receive, although
sometimes they are used for receive only the
knee coil is receive only
[2] Surface coils
As the name already implies, surface coils are placed close to the
area under examination such as the temporo - mandibular joint, the
orbits or the shoulder. The coil consists of a single or double loop of
copper wire. They have a high Signal to Noise Ratio (SNR) and allow
for very high resolution imaging. The disadvantage is that they loose
signal uniformity very quickly when you move away from the coil.
[3] Gradient Coils
Gradient coils are a set of wires in the magnet, which enable us
to create additional magnetic fields, which are, in a way,
superimposed on the main magnetic field B0.
Sounds complicated, but it's not really.
There are 3 sets of wires. Each set can create a magnetic field
in a specific direction: Z, X or Y. When a current is fed into the Z
gradient, then a magnetic field is generated in the Z direction .
The same goes for the other gradients
Interesting detail: Everyone knows that MRI can make a lot of
noise during acquisition. The magnetic field, which is
generated, is very strong. Although the gradient coils are very
tightly fixed in a kind of resin, the forces, exhibited by the
gradient coil, are enough to make them vibrate, hence the
noise.
1. The Gz gradient selected an axial slice.
2. The Gy gradient created rows with different phases.
3. The Gx gradient created columns with different frequencies.
[4] Phased Array Coils
Phased array coils consist of multiple surface coils.
Surface coils have the highest SNR but have a limited
sensitive area. By combining 4 or 6 surface coils it is
possible to create a coil with a large sensitive area.
The QD Body Array coil is a volume coil, while the Spine
Array coil is a surface coil. Today most MRI systems come
with Quadrature and phased array coils.
• Other Hardware
A very important part is the Radio Frequency (RF)
chain, which produces the RF signal transmitted into
the patient, and receives the RF signal from the patient.
Actually, the receive coil is a part of the RF chain. The
frequency range used in MRI is the same as used for
radio transmissions. That‟s why MRI scanners are
placed in a Faraday cage to prevent radio waves to
enter the scanner room .
BASIC PRINCIPLE OF MRI
Four basic principle/steps of MRI through which we get MR image---
1. Placing the patient in the magnet
2. Apply/Sending radiofrequency(RF) pulse by coil to the body
3. And then radio frequency pulse stop
4. Receiving signals from the patient as FID from receive coil
5. Transformation of signals into image by complex processing in the computers.
At molecular level about Protons -
Today’s MR imaging is based on PROTON IMAGING.
PROTON -
• Positively charged particle
• Present in nucleus of atom
• Example hydrogen having one proton (H¹⁺)
Proton help in MR imaging
o Positively charged
o Having rotatory movement called spin.
o Randomly moving in human body
o As moving charge generate current and induce small M.F. around It, called Magnetic dipole moment.
 In Human body, proton(H⁺) moves randomly ---
When patient is place in magnet and external magnetic field is applied, protons align themselves
and spin in the direction of Ext. M.F. 1st step
 Some proton align parallel and some anti-parallel to Ext. M.F.
 Proton align parallel to M.F. , not only rotate around itself ( called spin),
but also its axis of rotation form cone called precession (as shown in diagram )
 No. of precession of proton per second called precession frequency(P.F.) ,(hertz)
 Therefore, precession frequency is directly proportion to Ext. M.F.
Hence larmour equation is given as, W₀ α B₀
W₀ = γ B₀
where, W₀ - P.F.
B₀ - M.F.
γ - Gyroscopic ratio, specific to particular nuclei
P.F. of hydrogen at ---
LONGITUDINAL MAGNETIZATION ( L.M.)
 What happen when proton align under the influence of Ext. M.F. ?
• For orientation in space consider X,Y,Z axis.
• Ext. M.F. along Z-axis and Z-axes Is toward patient as well as bore of the magnet.
• As we discuss earlier proton is parallel ( positive) and anti-parallel (negative) to Ext. M.F.
• Force of parallel (positive) and anti-parallel (negative) cancel each other,
but proton spinning to positive side are always more than negative side.
• Force at positive side add up together to form magnetic vector to Z-axis called Longitudinal
magnetization
ds
1 T
1.5 T
3 T
42 MHz
64 MHz
128 MHz
LONGITUDINAL MAGNETISATION
TRASNVERSE MAGNETIZATION ( T.M.)
• As L.M. formed along Z-axes
• Next step send RF pulse 2nd step
• Precessing proton pick up and go to higher energy level and start precessing antiparallel.
• Imbalance results in tilting of magnetization into transverse plane. Called Transverse magnetization.
 In short, RF pulse cause tilting of magnetization into transverse plane.
• Therefore precessing frequency is same as that of RF pulse
• As RF pulse and proton have higher similar frequency protons can pick some energy from RF pulse . This is called
resonance R of MRI.
MR Signal
• Transverse Magnetization has precession frequency.
• It constantly rotates at larmour frequency in transverse plane and induce electric current.
• Receiver RF pulse receive current as MR signal. 3rd step
Strength of signal α Magnitude of T.M.
• MR signals are transformed into MRI image by computers using mathematical methods such as
Fourier Transfromation. 4th step
REVISION
Basic four steps of MR imaging includes—
1. Patient is placed in the magnet-
• All randomly moving protons in patients body align and precess along the external M.F.
• L.M. is formed long the Z-axes.
2. RF pulse is sent---
• Precessing proton pick up energy from RF pulse to go to higher energy level
• Results in reduction in L.M. and formation of T.M. in X-Y plane.
3. MR signal is received---
• T.M. vector precess in transverse plane and generates current.
• Current is received as signal by the RF coil.
4. Image formation ----
• MR signal received by the coil transformed into image by complex mathematical process
such as Fourier Transformation by computers.
Why proton only?
• Other substances can also utilize in MR imaging.
• Requirement are --- nuclei have spin and have odd number of proton.
• Example 13C , 19F, 23Na, 31P can be used for MR imaging.
• HYDROGEN atom has only on proton
• Hydrogen is present in abundance in body water and fat
• H⁺ gives best and most intense signal among all nuclei.
WHAT CAN BE DIAGNOSED BY AN MRI SCAN ?
 Most ailments of brain, including tumor
 Sport injuries
 Most spinal injuries
 Pelvic problems
 Prostrate problems
 Soft tissue and bone pathology
WHO CAN’T HAVE AN MRI SCAN
 A cardiac pacemaker
 Brain aneurysm clips
 A cochlear implant
 Eye implants
 IUD
 Dental fillings and bridge
 Tattoos
In medical imaging, artifacts are misrepresentations of tissue structures produced by
imaging techniques such as ultrasound, X-ray, CT scan, and magnetic resonance imaging
(MRI).
In this section I‟ll discuss the following artifacts:
• Motion related artifacts
• Para-magnetic artifacts
• Phase Wrap artifacts
• Frequency artifacts
• Susceptibility artifacts
• Clipping artifact
• Chemical Shift Artifact
• Spike artifact
• “Zebra” artifac
Image Artifacts
Motion artifacts are caused by phase mis-
mapping of the protons. protons may have
moved, due to respiration, pulsation or
motion, through the gradient magnetic field,
thus acquiring an additional phase
• Motion artifacts • Para-Magnetic Artifacts
Para-magnetic artifacts are caused by
metal (~ iron). Metal deflects the magnetic
field, thus changing the resonance frequency
beyond the range, which is used in MRI.
Image Artifacts
•Phase Wrap Artifacts
Phase Wrap Artifacts Phase wrap artifacts
are caused by mis-mapping of phase.
Phase wrap occurs when the Field Of View
is smaller than the object. The part of the
body outside the FOV will be „wrapped
around‟ into the image
•Frequency Artifacts
Frequency artifacts are caused by „dirty‟ frequencies. Faulty
electronics, external transmitters, RFcage leak, non-shielded
equipment in the scanner room, metal in the patient, or when
the door to the scanner room is left open can generate „dirty‟
frequencies. It usually requires an engineer to solve this kind of
artifact, although, the door to the scanner room can be closed
by non-qualified people as well
Image Artifacts
•Susceptibility Artifacts
Susceptibility is the ability of substances to be
magnetized, for example iron in blood.
Susceptibility artifacts are caused by local
magnetic field inhomogeneity. The different
bonding properties of hydrogen protons in fat and
muscle will cause local magnetic field
inhomogeneity at the tissue boundaries
•Clipping Artifact
Signal clipping or „over flow‟ occurs when
the receiver gain is set to high during the
pre-scan. The maximum received signal is
higher than the value set in the receiver gain
Image Artifacts
•Chemical Shift Artifact
Chemical shift artifacts are caused by different
resonance frequencies of hydrogen in lipids
and hydrogen in water. Chemical shift happens
in the frequency direction.
•Spike Artifact
A spike artifact is caused by one „bad‟ data point in
k-space. one data point in k-space, which is out of
the ordinary. The resulting image show diagonal
lines throughout the image.
Image Artifacts
•“Zebra” Artifact
The “Zebra” artifact
may occur when the
patient touches the coil,
or as a result of phase
wrap. To avoid this
problem you have to
make sure that the
patient is not touching
the receive coil, or use
No-Wrap option.
Image Artifacts
BASICPRINCIPLESOFMRI

Mais conteúdo relacionado

Mais procurados (20)

Mri basics
Mri basicsMri basics
Mri basics
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose Index
 
Icrp60
Icrp60Icrp60
Icrp60
 
Mri basics
Mri basicsMri basics
Mri basics
 
MRI SAFETY.pptx
MRI SAFETY.pptxMRI SAFETY.pptx
MRI SAFETY.pptx
 
Mri coils
Mri coilsMri coils
Mri coils
 
Mri safety – the dos and donts
Mri safety – the dos and dontsMri safety – the dos and donts
Mri safety – the dos and donts
 
Mri physics ii
Mri physics iiMri physics ii
Mri physics ii
 
Basics of mri physics Dr. Muhammad Bin Zulfiqar
Basics of mri physics Dr. Muhammad Bin ZulfiqarBasics of mri physics Dr. Muhammad Bin Zulfiqar
Basics of mri physics Dr. Muhammad Bin Zulfiqar
 
Encoding and image formation
Encoding and image formationEncoding and image formation
Encoding and image formation
 
Physics of 3 Tesla MRI & Silent MRI
Physics of 3 Tesla MRI & Silent MRIPhysics of 3 Tesla MRI & Silent MRI
Physics of 3 Tesla MRI & Silent MRI
 
MRI INSTRUMENTATION/ HARDWARE
MRI INSTRUMENTATION/ HARDWAREMRI INSTRUMENTATION/ HARDWARE
MRI INSTRUMENTATION/ HARDWARE
 
Mri safety sujan karki
Mri safety sujan karkiMri safety sujan karki
Mri safety sujan karki
 
Magnetic resonance imaging
Magnetic resonance imagingMagnetic resonance imaging
Magnetic resonance imaging
 
CT Generations and Artefacts
CT Generations and ArtefactsCT Generations and Artefacts
CT Generations and Artefacts
 
MRI - magnetic Imaging resonance
MRI - magnetic Imaging resonanceMRI - magnetic Imaging resonance
MRI - magnetic Imaging resonance
 
Mri basic principle and sequences
Mri basic principle and sequencesMri basic principle and sequences
Mri basic principle and sequences
 
CT Simulation Procedure
CT Simulation ProcedureCT Simulation Procedure
CT Simulation Procedure
 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pk
 
240 spatial encoding
240 spatial encoding240 spatial encoding
240 spatial encoding
 

Semelhante a BASICPRINCIPLESOFMRI (20)

Magnetic resonance imaging (mri) asit meher ppt
Magnetic resonance imaging (mri) asit meher pptMagnetic resonance imaging (mri) asit meher ppt
Magnetic resonance imaging (mri) asit meher ppt
 
Principle of MRI.pptx
Principle of MRI.pptxPrinciple of MRI.pptx
Principle of MRI.pptx
 
MRI
MRIMRI
MRI
 
Aftab ahmad
Aftab ahmadAftab ahmad
Aftab ahmad
 
MAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdf
MAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdfMAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdf
MAGNETIC RESONANCE IMAGING QUESTION AND ANSWER 3 MARKS.pdf
 
JC on MRI -BASICS.pptx
JC on MRI -BASICS.pptxJC on MRI -BASICS.pptx
JC on MRI -BASICS.pptx
 
MRI-2.pptx
MRI-2.pptxMRI-2.pptx
MRI-2.pptx
 
MRI physics part 1: Basic principle by GKM
MRI physics part 1: Basic principle by GKMMRI physics part 1: Basic principle by GKM
MRI physics part 1: Basic principle by GKM
 
MRI
MRIMRI
MRI
 
MRI Cyrogenics
MRI CyrogenicsMRI Cyrogenics
MRI Cyrogenics
 
lec 2a- MRI.ppt
lec 2a- MRI.pptlec 2a- MRI.ppt
lec 2a- MRI.ppt
 
MRI Scanner, Instrumentation. MDIRT ST. Louis Bamenda, Nchanji Nkeh Keneth
MRI Scanner, Instrumentation. MDIRT ST. Louis Bamenda, Nchanji Nkeh KenethMRI Scanner, Instrumentation. MDIRT ST. Louis Bamenda, Nchanji Nkeh Keneth
MRI Scanner, Instrumentation. MDIRT ST. Louis Bamenda, Nchanji Nkeh Keneth
 
Mripresenation
MripresenationMripresenation
Mripresenation
 
MRI of Brain: Basics
MRI of Brain: BasicsMRI of Brain: Basics
MRI of Brain: Basics
 
Principles of MRI
Principles of MRIPrinciples of MRI
Principles of MRI
 
Magnetic resonance imaging
Magnetic resonance imaging Magnetic resonance imaging
Magnetic resonance imaging
 
Mri hardware
Mri hardwareMri hardware
Mri hardware
 
MRI
MRIMRI
MRI
 
BASIC PRINCIPLES OF MRI Galala.pptx
BASIC PRINCIPLES OF MRI Galala.pptxBASIC PRINCIPLES OF MRI Galala.pptx
BASIC PRINCIPLES OF MRI Galala.pptx
 
MRI
MRIMRI
MRI
 

Último

Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Glomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxGlomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxDr.Nusrat Tariq
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Modelssonalikaur4
 
Call Girl Nagpur Sia 7001305949 Independent Escort Service Nagpur
Call Girl Nagpur Sia 7001305949 Independent Escort Service NagpurCall Girl Nagpur Sia 7001305949 Independent Escort Service Nagpur
Call Girl Nagpur Sia 7001305949 Independent Escort Service NagpurRiya Pathan
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalorenarwatsonia7
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...rajnisinghkjn
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowNehru place Escorts
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfMedicoseAcademics
 
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service SuratCall Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service Suratnarwatsonia7
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...narwatsonia7
 
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
Pharmaceutical Marketting: Unit-5, Pricing
Pharmaceutical Marketting: Unit-5, PricingPharmaceutical Marketting: Unit-5, Pricing
Pharmaceutical Marketting: Unit-5, PricingArunagarwal328757
 
See the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformSee the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformKweku Zurek
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 

Último (20)

Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
 
Glomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptxGlomerular Filtration rate and its determinants.pptx
Glomerular Filtration rate and its determinants.pptx
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
 
Call Girl Nagpur Sia 7001305949 Independent Escort Service Nagpur
Call Girl Nagpur Sia 7001305949 Independent Escort Service NagpurCall Girl Nagpur Sia 7001305949 Independent Escort Service Nagpur
Call Girl Nagpur Sia 7001305949 Independent Escort Service Nagpur
 
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service BangaloreCall Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
Call Girl Bangalore Nandini 7001305949 Independent Escort Service Bangalore
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
Call Girls ITPL Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Electronic City Just Call 7001305949 Top Class Call Girl Service A...
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
 
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowKolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Kolkata Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
 
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service SuratCall Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
 
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
Pharmaceutical Marketting: Unit-5, Pricing
Pharmaceutical Marketting: Unit-5, PricingPharmaceutical Marketting: Unit-5, Pricing
Pharmaceutical Marketting: Unit-5, Pricing
 
See the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformSee the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy Platform
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
 

BASICPRINCIPLESOFMRI

  • 2. MRI MACHINE Ge Discovery 3T 750 MRI scaner Seimens
  • 3. INTRODUCTION MRI stands for magnetic resonance imaging MRI technique used in radiology to form picture of anatomy and physiological process of the body. MRI scanners use strong magnetic fields, magnetic field gradient and radiowaves to generate images of the organs.  Do not involve X-ray or use of ionizing radiation --- like CT Scan and PET scans. Uses the application NMR – (investigation of chemical and physical properties at molecular level) Based on the principle of NMR
  • 4. HISTORY In 1997, the first MRI exam was performed on human being. FATHER OF MRI Magnetic resonance imaging inventor ----- RAYMOND VAHAN DAMADIAN { Before talking about MRI principle first look types of magnets }
  • 5. Types of Magnet [1] Permanent magnets A permanent magnet consists of a material, which has been magnetized such that it won‟t loose its magnetic field The field strength is usually very low and ranges between 0.064T ~ 0.3T, Permanent magnets have usually an open design, which is more comfortable for the patient. Toshiba‟s Access 0.064Tesla system. The Access was the worlds first open MRI scanner.
  • 6. Resistive magnets are very large electro magnets ,The magnetic field is generated by a current, which runs through loops of wire. The field strength can be up to 0.3 Tesla. They produce a lot of heat, which requires watercooling. They need a lot of power to run, and are usually switched off when not in use to conserve power. They usually have an open design, which reduces claustrophobia. [2] Resistive Magnets Hitachi‟s Airis 0.3 Tesla system.
  • 7. [3] Superconducting magnets Today‟s most commonly used magnets are superconducting magnets. The magnetic field is generated by a current, which runs through a loop of wire. The wire is surrounded with a coolant, such as liquid helium, to reduce the electric resistance of the wire. At 4 Kelvin (-269º C) electric wire looses its resistance. Once a system is energized, it won‟t loose its magnetic field. Superconductivity allows for systems with very high field strengths up to 12 Tesla. The ones that are most used in clinical environments run at 1.5 Tesla. Most superconducting magnets are bore type magnets examples of bore type magnets
  • 8. • RF Coils RF coils are needed to transmit and receive radio-frequency waves used in MRI scanners. RF coils are one of the most important components that affect image quality. Current MRI scanners have a range of RF coils suitable to acquire images of all body parts. There are two types of RF coils: volume coils and surface coils [1] Volume RF Coils OR Quadrature Coils The design of a volume coil is usually a saddle shape, which guarantees a uniform RF field inside the coil. Volume coils need to have the area of examination inside the coil. They can be used for transmit and receive, although sometimes they are used for receive only the knee coil is receive only [2] Surface coils As the name already implies, surface coils are placed close to the area under examination such as the temporo - mandibular joint, the orbits or the shoulder. The coil consists of a single or double loop of copper wire. They have a high Signal to Noise Ratio (SNR) and allow for very high resolution imaging. The disadvantage is that they loose signal uniformity very quickly when you move away from the coil.
  • 9. [3] Gradient Coils Gradient coils are a set of wires in the magnet, which enable us to create additional magnetic fields, which are, in a way, superimposed on the main magnetic field B0. Sounds complicated, but it's not really. There are 3 sets of wires. Each set can create a magnetic field in a specific direction: Z, X or Y. When a current is fed into the Z gradient, then a magnetic field is generated in the Z direction . The same goes for the other gradients Interesting detail: Everyone knows that MRI can make a lot of noise during acquisition. The magnetic field, which is generated, is very strong. Although the gradient coils are very tightly fixed in a kind of resin, the forces, exhibited by the gradient coil, are enough to make them vibrate, hence the noise. 1. The Gz gradient selected an axial slice. 2. The Gy gradient created rows with different phases. 3. The Gx gradient created columns with different frequencies.
  • 10. [4] Phased Array Coils Phased array coils consist of multiple surface coils. Surface coils have the highest SNR but have a limited sensitive area. By combining 4 or 6 surface coils it is possible to create a coil with a large sensitive area. The QD Body Array coil is a volume coil, while the Spine Array coil is a surface coil. Today most MRI systems come with Quadrature and phased array coils. • Other Hardware A very important part is the Radio Frequency (RF) chain, which produces the RF signal transmitted into the patient, and receives the RF signal from the patient. Actually, the receive coil is a part of the RF chain. The frequency range used in MRI is the same as used for radio transmissions. That‟s why MRI scanners are placed in a Faraday cage to prevent radio waves to enter the scanner room .
  • 11. BASIC PRINCIPLE OF MRI Four basic principle/steps of MRI through which we get MR image--- 1. Placing the patient in the magnet 2. Apply/Sending radiofrequency(RF) pulse by coil to the body 3. And then radio frequency pulse stop 4. Receiving signals from the patient as FID from receive coil 5. Transformation of signals into image by complex processing in the computers.
  • 12. At molecular level about Protons - Today’s MR imaging is based on PROTON IMAGING. PROTON - • Positively charged particle • Present in nucleus of atom • Example hydrogen having one proton (H¹⁺) Proton help in MR imaging o Positively charged o Having rotatory movement called spin. o Randomly moving in human body o As moving charge generate current and induce small M.F. around It, called Magnetic dipole moment.  In Human body, proton(H⁺) moves randomly --- When patient is place in magnet and external magnetic field is applied, protons align themselves and spin in the direction of Ext. M.F. 1st step
  • 13.  Some proton align parallel and some anti-parallel to Ext. M.F.  Proton align parallel to M.F. , not only rotate around itself ( called spin), but also its axis of rotation form cone called precession (as shown in diagram )  No. of precession of proton per second called precession frequency(P.F.) ,(hertz)  Therefore, precession frequency is directly proportion to Ext. M.F. Hence larmour equation is given as, W₀ α B₀ W₀ = γ B₀ where, W₀ - P.F. B₀ - M.F. γ - Gyroscopic ratio, specific to particular nuclei
  • 14. P.F. of hydrogen at --- LONGITUDINAL MAGNETIZATION ( L.M.)  What happen when proton align under the influence of Ext. M.F. ? • For orientation in space consider X,Y,Z axis. • Ext. M.F. along Z-axis and Z-axes Is toward patient as well as bore of the magnet. • As we discuss earlier proton is parallel ( positive) and anti-parallel (negative) to Ext. M.F. • Force of parallel (positive) and anti-parallel (negative) cancel each other, but proton spinning to positive side are always more than negative side. • Force at positive side add up together to form magnetic vector to Z-axis called Longitudinal magnetization ds 1 T 1.5 T 3 T 42 MHz 64 MHz 128 MHz
  • 16. TRASNVERSE MAGNETIZATION ( T.M.) • As L.M. formed along Z-axes • Next step send RF pulse 2nd step • Precessing proton pick up and go to higher energy level and start precessing antiparallel. • Imbalance results in tilting of magnetization into transverse plane. Called Transverse magnetization.  In short, RF pulse cause tilting of magnetization into transverse plane. • Therefore precessing frequency is same as that of RF pulse • As RF pulse and proton have higher similar frequency protons can pick some energy from RF pulse . This is called resonance R of MRI.
  • 17. MR Signal • Transverse Magnetization has precession frequency. • It constantly rotates at larmour frequency in transverse plane and induce electric current. • Receiver RF pulse receive current as MR signal. 3rd step Strength of signal α Magnitude of T.M. • MR signals are transformed into MRI image by computers using mathematical methods such as Fourier Transfromation. 4th step
  • 18. REVISION Basic four steps of MR imaging includes— 1. Patient is placed in the magnet- • All randomly moving protons in patients body align and precess along the external M.F. • L.M. is formed long the Z-axes. 2. RF pulse is sent--- • Precessing proton pick up energy from RF pulse to go to higher energy level • Results in reduction in L.M. and formation of T.M. in X-Y plane. 3. MR signal is received--- • T.M. vector precess in transverse plane and generates current. • Current is received as signal by the RF coil. 4. Image formation ---- • MR signal received by the coil transformed into image by complex mathematical process such as Fourier Transformation by computers.
  • 19. Why proton only? • Other substances can also utilize in MR imaging. • Requirement are --- nuclei have spin and have odd number of proton. • Example 13C , 19F, 23Na, 31P can be used for MR imaging. • HYDROGEN atom has only on proton • Hydrogen is present in abundance in body water and fat • H⁺ gives best and most intense signal among all nuclei.
  • 20. WHAT CAN BE DIAGNOSED BY AN MRI SCAN ?  Most ailments of brain, including tumor  Sport injuries  Most spinal injuries  Pelvic problems  Prostrate problems  Soft tissue and bone pathology WHO CAN’T HAVE AN MRI SCAN  A cardiac pacemaker  Brain aneurysm clips  A cochlear implant  Eye implants  IUD  Dental fillings and bridge  Tattoos
  • 21.
  • 22. In medical imaging, artifacts are misrepresentations of tissue structures produced by imaging techniques such as ultrasound, X-ray, CT scan, and magnetic resonance imaging (MRI). In this section I‟ll discuss the following artifacts: • Motion related artifacts • Para-magnetic artifacts • Phase Wrap artifacts • Frequency artifacts • Susceptibility artifacts • Clipping artifact • Chemical Shift Artifact • Spike artifact • “Zebra” artifac Image Artifacts
  • 23. Motion artifacts are caused by phase mis- mapping of the protons. protons may have moved, due to respiration, pulsation or motion, through the gradient magnetic field, thus acquiring an additional phase • Motion artifacts • Para-Magnetic Artifacts Para-magnetic artifacts are caused by metal (~ iron). Metal deflects the magnetic field, thus changing the resonance frequency beyond the range, which is used in MRI. Image Artifacts
  • 24. •Phase Wrap Artifacts Phase Wrap Artifacts Phase wrap artifacts are caused by mis-mapping of phase. Phase wrap occurs when the Field Of View is smaller than the object. The part of the body outside the FOV will be „wrapped around‟ into the image •Frequency Artifacts Frequency artifacts are caused by „dirty‟ frequencies. Faulty electronics, external transmitters, RFcage leak, non-shielded equipment in the scanner room, metal in the patient, or when the door to the scanner room is left open can generate „dirty‟ frequencies. It usually requires an engineer to solve this kind of artifact, although, the door to the scanner room can be closed by non-qualified people as well Image Artifacts
  • 25. •Susceptibility Artifacts Susceptibility is the ability of substances to be magnetized, for example iron in blood. Susceptibility artifacts are caused by local magnetic field inhomogeneity. The different bonding properties of hydrogen protons in fat and muscle will cause local magnetic field inhomogeneity at the tissue boundaries •Clipping Artifact Signal clipping or „over flow‟ occurs when the receiver gain is set to high during the pre-scan. The maximum received signal is higher than the value set in the receiver gain Image Artifacts
  • 26. •Chemical Shift Artifact Chemical shift artifacts are caused by different resonance frequencies of hydrogen in lipids and hydrogen in water. Chemical shift happens in the frequency direction. •Spike Artifact A spike artifact is caused by one „bad‟ data point in k-space. one data point in k-space, which is out of the ordinary. The resulting image show diagonal lines throughout the image. Image Artifacts
  • 27. •“Zebra” Artifact The “Zebra” artifact may occur when the patient touches the coil, or as a result of phase wrap. To avoid this problem you have to make sure that the patient is not touching the receive coil, or use No-Wrap option. Image Artifacts