Oxygen
Manufacturing,
Storage,Supply,Therapy and Toxicity
by –pawan kumr rai
JR Anaesthesiology
DR.RMLIMS LUCKNOW
How Oxygen is produced?
• Fractional Distillation/ Cryogenic Separation
• Physical Separation
Pressure swing absorption,
Membrane filter technology and
Electrolysis.
Cryogenic Separation
• Carl Von Linde in 1895
• The idea behind the process is to
lower the temperature of the air such
that nitrogen and oxygen separate
based on their boiling points.
• If higher purity oxygen or liquid
oxygen is required, further
distillation is required.
Pressure swing adsorption(PSA) /
Molecular Sieve
• PSA devices take atmospheric air into a pressurized tank.
• Inside the tank are zeolites.
Storage
• Units of Pressure:
1 atmosphere = 14.7 psi
760 mmHg
1030 cm H20
1000 mbar
100 KPa
RsCr 220
9
Bulk Oxygen Systems
(high pressure system)
• Large capacity
• Liquid oxygen
• Small capacity (or back up)
Cylinders of different sizes
A-J , A is the smallest size and J is the largest size
D,E,H :Used most commonly in hospital settings
Components of cylinder
• Consists of:
– Body
– Valve
– Port
– Stem
– Pressure relief devices
Body
•MRI compatible cylinders are made of aluminium
• Modern cylinders are made of alloy of MOLYBDENUM +
STEEL +/- CHROMIUM
It is used to increase strength and to minimise weight and
wall thickness
Port: It is the point of exit for the gas .
Pressure Relief Devices:
– Every cylinder is fitted with pressure relief devices whose purpose is to vent the
cylinder’s contents to the atmosphere , if the pressure of enclosed gas increases to
dangerous levels
– Types: • Rupture Disc
• Fusible Plug
• Combination of Both
• Pressure Relief valve (spring loaded)
Cylinder valves are of 2 types
–packed type and diaphragm type
• In order to ensure that the correct cylinder is attached to
the appropriate hanger yoke of the anesthesia
machine/workstation.
A series of pins on the hanger yoke is made to fit into the
corresponding indentations(pits/holes) drilled into
cylinder valve
• Holes are positioned in an arc below the outlet port
•The pins are 4.75mm in diameter and 6mm long except
for pin number 7 which is slightly thicker and placed at
the centre (between port 3 and 4)
Pin Index Safety System
RsCr 220
Diameter Index Safety System (DISS)
• Safety Systems
• DISS use noninterchange-able,
threaded fittings to connect gas-
powered devices to station outlets
50 P.S.I (INTERMEDIATE PRESSURE SYSTEM)
RsCr 220
Oxygen cylinder Markings
•Periodic visual inspection for any distortion, cracks, dents, leaks, etc.
•Tensile test on 1 out of every 100 cylinders manufactured. Strips of the
cylinder are cut and stretched.
•Flattening test, Bend test and impact test are carried out on at least one in
every 100 cylinders
•Hydraulic test—the cylinder is tested to 1.66 times it service pressure.
This helps detect any leaks and to determine the retention of structural
strength. A cylinder must be subjected to internal hydrostatic pressure
testing at least every 5 years.
TESTING OF CYLINDER
• Factors for determination of duration
• “E” cylinder = 0.28
• “H” cylinder = 3.14
Duration of flow = Pressure x Cylinder factor
Flow
Liquid system - factor :860
Content=Amount of gas in liquid = Weight of liquid x 860
2.5
Duration of flow = Contents
Flow
What precautions should be taken while
transporting a cylinder?
ƒ
 Upright position ƒ, Wear protective footwear/safety gloves
 ƒDo not lift by protective cap/guard as you may damage valve ƒ
 Ensure all cylinders are properly labeled as to the contents ƒ
Do not subject cylinders to temperature extremes.
ƒ
ƒ
Cylinders should be handled only by trained person
Cylinder valves, regulator,Gauges, and fitting should never come into contact of oils , greases ,
organic lubricants etc .
Cylinders should not be subjected to extreme of temperatures .
Connection to piping , regulator, and other equipments should always be kept tight to prevent
leakage.
No part of the cylinders should be tainted , painted or modified by users .
A cylinder should never come into contact of electricity .
Before use ,cylinder should be identified by the label.
The color of the cylinder should not be relied upon for identification .
RULES FOR SAFE USE OF CYLINDERS
Only DOT or ICC approved cylinder should be used .
A pressure regulator should always be used before any fitting is applied to cylinder valve
, the valve should be opened slowly and briefly to remove dust or any other
particle ( cracking) .
A sealing washer or gasket in good condition should always be used with small cylinder
valve .
 If more than one seal is used the pin on yoke may not extrude far enough to engage
mating whole or safety pin index.
A cylinder valve should be opened slowly ,because if gases passes quickly into the space
between valve and the yoke or the regulator , the rapid recompression in this space will
generate large amount of heat ,because there is no space for dissipation of heat , this
constitutes an adiabatic process. Particle of dust, grease, etc present in that space may be
ignited by the heat causing a flash fire or explosion.
Do not drag slide or roll- use cylinder cart
Cylinder Manifold System
Intermediate
pressure system
High pressure
system
One of the most serious mishaps that occurred was depletion of the oxygen supply.
The result was delivery of 100% anesthetic gas.
Numerous inventions have been devised to prevent this-
• Oxygen failure
safety device.1.
• Oxygen failure
supply alarm.2.
Oxygen pressure failure devices
Operates in a threshold manner: either open or shut.
Pressure Sensor Shut-off Valve:
 Oxygen pressure moves the piston and pin upward and the valve opens for N2O
When pressure of oxygen falls below preset value, force of the valve return spring completely closes the valve
Based on a proportioning principle rather than a threshold principle. Pressure
of N2O falls in Propotion of decrease of Oxygen.
Total cutoff seen at <12 psig.
Seat nozzel assembly connected to a spring loaded conical tapered piston.
Oxygen Failure Protection Device
The oxygen flush receives oxygen from the pipeline inlet or cylinder pressure
regulator and directs a high unmetered flow directly to the common gas outlet.
It is commonly labeled “02+.”
It can be activated regardless of whether the master switch is turned ON or OFF.
A flow between 35 and 75 L/minute must be delivered.
Oxygen flush
“Single purpose, self closing device & recessed”
BALL AND SPRING
VALVE
Mechanical device : Link 25 system
Pneumatic device : Oxygen Ratio Monitor Controller (ORMC)
Electronically controlled : Penlon Ltd
ANTI HYPOXIA DEVICES
Mechanical Linking of O2 & N2O.
O2 – 14 tooth sprocket, & N2O – 29 tooth sprocket,
connected by metal chain.
• Both knobs turn together – to maintain min 25%
O2 conc.
 • Disadvantage – if 3rd gas is administered (eg
He,air) – hypoxic mixture may be delivered.
Mechanical device : Link 25 system
• Mandatory Min O2 Flow (MMOF):- A MMOF – 50-250 ml/min – must as
soon as the master switch is turned on.
•O2 flow – cant be reduced below that.
• Min O2 Ratio :- ASTM Stds require that; A Min O2 conc of 21% in the
FGF at common gas outlet.
• Pressure is reduced to a 50 psi working pressure in the patient’s
room where it can be used to operate respiratory equipment.
• Provide connections for gas-delivery devices such as flowmeters and
mechanical ventilators.
Station outlet
Therapy
•Oxygen is also like a drug.
• It must be used meticulously & deligently.
• If abused it can cause complications.
•
FATHER OF MODERN
O2 THERAPY
J S HALDANE
INDICATIONS FOR O2 THERAPY
1.Hypoxia – when PaO2 comes down to 60mmhg, like low cardiac output
states,M.I,anemia,hemodilution,CO poisoning,acute hypermetabolic states.
2.Trapped gases – like obstruction.
3.Special situations – like anesthesia.
Indications of hyperbaric oxygen therapy
Poisoning: CO
Infections: clostridium myonecrosis
Acute ischaemia:crush injury
Chronic ischaemia: radiation necrosis
Ischaemic ulcers: diabetic ulcers
1.Nasal catheters.
8-14 G ,passed through into oropharynx.
Reservoir of O2 builds up in pharynx.
O2 DELIVERY SYSTEMS:
1.Low flow or variable performance systems.
2.High flow fixed performance systems.
LOW FLOW SYSTEMS:
O2 Flowrate (L/min) FIO2
1 0.24
2 0.28
3 0.32
4 0.36
5 0.40
6 0.44
Demerits • Mucosal irritation (uncomfortable) • Gastric dilatation (especially with high flows)
2.Nasal prongs:
-The prongs protrude 1cm into nares.
-Well tolerated.
-Useless in mouth breathers.
-No reservoir system.
3.Simple O2 masks:
- Covers the nose & mouth.
-4-6 lits/min.
O2 Flowrate (L/min) FIO2
5-6 0.40
6-7 0.50
7-8 0.60
4.Mask with reservoir bags:
-High FiO2 with 5-6 lits of O2.
1.VENTURI MASKS:
- Works on Bernoulli principle.
- Can deliver O2 at accurate concs.
HIGH FLOW SYSTEMS:
- Total flow is high with or without entrainment
3. O2 tent:
-Used in pediatric practise.
-10-12 lits/min.
-Fire accidents.
2. O2 hood:
-Used in bed bound infants not intubated.
-Made of plexi glass.
-Humidified O2, 3 times M.V.
4.JET VENTILATION:
- High frequency oscillation.
Drying of
mucus
membrane
Depression of
ventilation in
COPD
Atelectasis due
to absorption
collapse
O2 toxicity
HAZARDS OF O2 THERAPY
OXYGEN TOXICITY
• Increase in oxygen concentration leads to free radical formation
• Reacts with cellular DNA proteins, lipids and membrane.
1.Pulmonary oxygen toxicity
(LorrainSmith effect)
3.C.N.S O2 toxicity (Paul-Bert effect)
2.Retrolental fibroplasia
1.Pulmonary oxygen toxicity (LorrainSmith effect):
Symptoms: substernal pain,
irresistable cough,
dyspnoea,V.C dec.,
compliance dec.,
pulmonary interstitial edema leading to fibrosis.
Safe dose of oxygen
100%O2 given not more than12 hours
-80% O2 for not more than 24hrs.
- 60%O2 not more than 36hrs.
2.Retrolental fibroplasia –
Occurs when PaO2 is more than 80mmhg for more than 3 hrs in new born.
Very premature babies are more susceptible. - O2 saturation must be around 90-92 %.
3.C.N.S O2 toxicity (Paul-Bert effect): -
Seizures or convulsions
Depressed ventilatory drive
Coma
Brain death and eventually death
Take home message
No drug is more important than oxygen.
Medical oxygen is of high purity >99.5%, manufacturing of oxygen by fractional
distillation and PSA techniques give us O2 of >99.5% purity.
Cylinders has various specifications marked over it, must be checked before.
Medical gas supply system has 3 different pneumatic zones, having various safety
features.
Oxygen is like a drug, life saving when used meticulously otherwise toxic.
SAFE DOSE- 100% not>12hrs, 80%not>24hrs, 60%not>36hrs.
THANKS THANK YOU
Defintions
Psi: Pounds per square inch
Psig: Pounds per square inch gauge
GAUGE PRESSURE
It is the difference between the measured pressure and the atmospheric pressure Most gauges
are measured to read zero at atmospheric pressure
• Psia: Pounds per square inch absolute
ABSOLUTE PRESSURE It is based on reference point of zero in a perfect vacuum Psia = Psig
+ atmospheric pressure Eg: at sea level- atmospheric pressure is 0 but psia is 14.7psia
•Critical temperature: – It is defined as the temperature below which no gas can be liquefied
irrespective of which pressure is applied
• Critical Pressure: – Defined as the minimum pressure required to liquefy a gas at its critical
temperature
Oxygen  MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT

Mais conteúdo relacionado

Mais procurados

Breathing circuit's
Breathing circuit'sBreathing circuit's
Breathing circuit'sImran Sheikh
 
NON OPERATING ROOM ANAESTHESIA
NON OPERATING ROOM ANAESTHESIA NON OPERATING ROOM ANAESTHESIA
NON OPERATING ROOM ANAESTHESIA Kundan Ghimire
 
The basic anaesthesia machine
The basic anaesthesia machineThe basic anaesthesia machine
The basic anaesthesia machinehrishi bharali
 
Fiberoptic intubation
Fiberoptic  intubationFiberoptic  intubation
Fiberoptic intubationWesam Mousa
 
Anaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyAnaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyUnnikrishnan Prathapadas
 
Breathing System
Breathing SystemBreathing System
Breathing SystemKhalid
 
Medical gas supplies (Anesthetic machine)
Medical gas supplies (Anesthetic machine)Medical gas supplies (Anesthetic machine)
Medical gas supplies (Anesthetic machine)Mohammad Abdeljawad
 
anaesthesia Breathing circuits and its classification and functional analysis
anaesthesia Breathing circuits and its classification and functional analysisanaesthesia Breathing circuits and its classification and functional analysis
anaesthesia Breathing circuits and its classification and functional analysisprateek gupta
 
Anaesthesia for cardiac patient undergoing non cardiac surgery
Anaesthesia for cardiac patient undergoing non cardiac surgeryAnaesthesia for cardiac patient undergoing non cardiac surgery
Anaesthesia for cardiac patient undergoing non cardiac surgeryDhritiman Chakrabarti
 
Interscalene & supraclavicular nerve blocks
Interscalene  & supraclavicular nerve blocksInterscalene  & supraclavicular nerve blocks
Interscalene & supraclavicular nerve blocksDavis Kurian
 
Physics In Anaesthesia
Physics In AnaesthesiaPhysics In Anaesthesia
Physics In AnaesthesiaNARENDRA PATIL
 
Delivering only intended gases from the anaesthesia workstation
Delivering only intended gases from the anaesthesia workstationDelivering only intended gases from the anaesthesia workstation
Delivering only intended gases from the anaesthesia workstationDhritiman Chakrabarti
 
Anesthesia machine
Anesthesia machineAnesthesia machine
Anesthesia machineamar pandey
 

Mais procurados (20)

Breathing circuit's
Breathing circuit'sBreathing circuit's
Breathing circuit's
 
Anaesthesia machine 2
Anaesthesia machine 2Anaesthesia machine 2
Anaesthesia machine 2
 
Vaporizers
Vaporizers Vaporizers
Vaporizers
 
NON OPERATING ROOM ANAESTHESIA
NON OPERATING ROOM ANAESTHESIA NON OPERATING ROOM ANAESTHESIA
NON OPERATING ROOM ANAESTHESIA
 
The basic anaesthesia machine
The basic anaesthesia machineThe basic anaesthesia machine
The basic anaesthesia machine
 
Fiberoptic intubation
Fiberoptic  intubationFiberoptic  intubation
Fiberoptic intubation
 
Anaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supplyAnaesthesia gas cylinders & pipeline gas supply
Anaesthesia gas cylinders & pipeline gas supply
 
Breathing systems
Breathing systemsBreathing systems
Breathing systems
 
Double lumen tubes
Double lumen tubesDouble lumen tubes
Double lumen tubes
 
Breathing System
Breathing SystemBreathing System
Breathing System
 
Medical gas supplies (Anesthetic machine)
Medical gas supplies (Anesthetic machine)Medical gas supplies (Anesthetic machine)
Medical gas supplies (Anesthetic machine)
 
The canister (the absorber)
The canister (the absorber)The canister (the absorber)
The canister (the absorber)
 
anaesthesia Breathing circuits and its classification and functional analysis
anaesthesia Breathing circuits and its classification and functional analysisanaesthesia Breathing circuits and its classification and functional analysis
anaesthesia Breathing circuits and its classification and functional analysis
 
Dexmedetomidine
DexmedetomidineDexmedetomidine
Dexmedetomidine
 
Anaesthesia for cardiac patient undergoing non cardiac surgery
Anaesthesia for cardiac patient undergoing non cardiac surgeryAnaesthesia for cardiac patient undergoing non cardiac surgery
Anaesthesia for cardiac patient undergoing non cardiac surgery
 
Interscalene & supraclavicular nerve blocks
Interscalene  & supraclavicular nerve blocksInterscalene  & supraclavicular nerve blocks
Interscalene & supraclavicular nerve blocks
 
Physics In Anaesthesia
Physics In AnaesthesiaPhysics In Anaesthesia
Physics In Anaesthesia
 
Delivering only intended gases from the anaesthesia workstation
Delivering only intended gases from the anaesthesia workstationDelivering only intended gases from the anaesthesia workstation
Delivering only intended gases from the anaesthesia workstation
 
Anesthesia machine
Anesthesia machineAnesthesia machine
Anesthesia machine
 
Inhalational Agents
Inhalational AgentsInhalational Agents
Inhalational Agents
 

Semelhante a Oxygen MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT

Presentation on aneasthetic machine edited 21st july (2)
Presentation on aneasthetic machine edited 21st july (2)Presentation on aneasthetic machine edited 21st july (2)
Presentation on aneasthetic machine edited 21st july (2)MD. HASANUL HAQUE SAGOR
 
Anaesthesia machine by Dr. shailendra
Anaesthesia machine by Dr. shailendra Anaesthesia machine by Dr. shailendra
Anaesthesia machine by Dr. shailendra Shailendra Satpute
 
ANESTHESIA MACHINE PPT .pptx
ANESTHESIA MACHINE PPT .pptxANESTHESIA MACHINE PPT .pptx
ANESTHESIA MACHINE PPT .pptxPrarthana83
 
MEDICAL GAS PIPELINE SYSTEMS AND CYLINDERS.pptx
MEDICAL GAS PIPELINE SYSTEMS AND CYLINDERS.pptxMEDICAL GAS PIPELINE SYSTEMS AND CYLINDERS.pptx
MEDICAL GAS PIPELINE SYSTEMS AND CYLINDERS.pptxMyDelhi1
 
anaesthesia machine ppt 2-1.pptx
anaesthesia machine ppt 2-1.pptxanaesthesia machine ppt 2-1.pptx
anaesthesia machine ppt 2-1.pptxjayantigupta9
 
Safety in welding
Safety in weldingSafety in welding
Safety in weldingIlhamR222
 
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptxANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptxUnnikrishnan Prathapadas
 
Safety Anesthesia Work Station
Safety Anesthesia Work StationSafety Anesthesia Work Station
Safety Anesthesia Work Stationteja bayapalli
 
Safety feature of BOYLE'S anesthesia machine
Safety feature of BOYLE'S anesthesia machineSafety feature of BOYLE'S anesthesia machine
Safety feature of BOYLE'S anesthesia machineZIKRULLAH MALLICK
 
Safe Handling of Compressed Gases by UCSD
Safe Handling of Compressed Gases by UCSDSafe Handling of Compressed Gases by UCSD
Safe Handling of Compressed Gases by UCSDAtlantic Training, LLC.
 
Medical gas supply.pipelines and cylinders
Medical gas supply.pipelines and cylindersMedical gas supply.pipelines and cylinders
Medical gas supply.pipelines and cylindersdrsauravdas1977
 
HIGH PRESSURE SYSTEM.pptx
HIGH PRESSURE SYSTEM.pptxHIGH PRESSURE SYSTEM.pptx
HIGH PRESSURE SYSTEM.pptxanjulkumarsingh
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machineimran80
 
ANESTHESIA_WORKSTATION_(1).pptx
ANESTHESIA_WORKSTATION_(1).pptxANESTHESIA_WORKSTATION_(1).pptx
ANESTHESIA_WORKSTATION_(1).pptxViswanathSelvam1
 
Oxygen concentrator-Applications and Maintenance
Oxygen concentrator-Applications and MaintenanceOxygen concentrator-Applications and Maintenance
Oxygen concentrator-Applications and Maintenanceshashi sinha
 
Medical gas supply. (central supply)
Medical gas supply.   (central supply)Medical gas supply.   (central supply)
Medical gas supply. (central supply)BKMC
 
Anaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmcAnaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmcDeepak Choudhary
 

Semelhante a Oxygen MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT (20)

anaesthesia machine(1).pptx
anaesthesia machine(1).pptxanaesthesia machine(1).pptx
anaesthesia machine(1).pptx
 
Presentation on aneasthetic machine edited 21st july (2)
Presentation on aneasthetic machine edited 21st july (2)Presentation on aneasthetic machine edited 21st july (2)
Presentation on aneasthetic machine edited 21st july (2)
 
Anaesthesia machine by Dr. shailendra
Anaesthesia machine by Dr. shailendra Anaesthesia machine by Dr. shailendra
Anaesthesia machine by Dr. shailendra
 
ANESTHESIA MACHINE PPT .pptx
ANESTHESIA MACHINE PPT .pptxANESTHESIA MACHINE PPT .pptx
ANESTHESIA MACHINE PPT .pptx
 
MEDICAL GAS PIPELINE SYSTEMS AND CYLINDERS.pptx
MEDICAL GAS PIPELINE SYSTEMS AND CYLINDERS.pptxMEDICAL GAS PIPELINE SYSTEMS AND CYLINDERS.pptx
MEDICAL GAS PIPELINE SYSTEMS AND CYLINDERS.pptx
 
anaesthesia machine ppt 2-1.pptx
anaesthesia machine ppt 2-1.pptxanaesthesia machine ppt 2-1.pptx
anaesthesia machine ppt 2-1.pptx
 
Leak Tests.pptx
Leak Tests.pptxLeak Tests.pptx
Leak Tests.pptx
 
Safety in welding
Safety in weldingSafety in welding
Safety in welding
 
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptxANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
ANAESTHESIA GAS CYLINDERS & PIPELINE GAS SUPPLY (2).pptx
 
Safety Anesthesia Work Station
Safety Anesthesia Work StationSafety Anesthesia Work Station
Safety Anesthesia Work Station
 
Safety feature of BOYLE'S anesthesia machine
Safety feature of BOYLE'S anesthesia machineSafety feature of BOYLE'S anesthesia machine
Safety feature of BOYLE'S anesthesia machine
 
Safe Handling of Compressed Gases by UCSD
Safe Handling of Compressed Gases by UCSDSafe Handling of Compressed Gases by UCSD
Safe Handling of Compressed Gases by UCSD
 
Medical gas supply.pipelines and cylinders
Medical gas supply.pipelines and cylindersMedical gas supply.pipelines and cylinders
Medical gas supply.pipelines and cylinders
 
HIGH PRESSURE SYSTEM.pptx
HIGH PRESSURE SYSTEM.pptxHIGH PRESSURE SYSTEM.pptx
HIGH PRESSURE SYSTEM.pptx
 
PIPELINE.pptx
PIPELINE.pptxPIPELINE.pptx
PIPELINE.pptx
 
Anaesthesia machine
Anaesthesia machineAnaesthesia machine
Anaesthesia machine
 
ANESTHESIA_WORKSTATION_(1).pptx
ANESTHESIA_WORKSTATION_(1).pptxANESTHESIA_WORKSTATION_(1).pptx
ANESTHESIA_WORKSTATION_(1).pptx
 
Oxygen concentrator-Applications and Maintenance
Oxygen concentrator-Applications and MaintenanceOxygen concentrator-Applications and Maintenance
Oxygen concentrator-Applications and Maintenance
 
Medical gas supply. (central supply)
Medical gas supply.   (central supply)Medical gas supply.   (central supply)
Medical gas supply. (central supply)
 
Anaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmcAnaesthesia machine 2014 spmc
Anaesthesia machine 2014 spmc
 

Mais de Dr.RMLIMS lucknow

New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentationDr.RMLIMS lucknow
 
Anesthetic management in copd
Anesthetic management in copdAnesthetic management in copd
Anesthetic management in copdDr.RMLIMS lucknow
 
Anatomy of the lower respiratory tract
Anatomy of the lower respiratory tractAnatomy of the lower respiratory tract
Anatomy of the lower respiratory tractDr.RMLIMS lucknow
 
ARTERIAL BLOOD GASES INTERPRETATION
ARTERIAL BLOOD GASES INTERPRETATIONARTERIAL BLOOD GASES INTERPRETATION
ARTERIAL BLOOD GASES INTERPRETATIONDr.RMLIMS lucknow
 
Monitoring Modality in anesthesia
Monitoring Modality  in anesthesiaMonitoring Modality  in anesthesia
Monitoring Modality in anesthesiaDr.RMLIMS lucknow
 
COPD:CLINICAL REVIEW AND ANESTHESIA CONSIDERATION
COPD:CLINICAL REVIEW AND ANESTHESIA CONSIDERATIONCOPD:CLINICAL REVIEW AND ANESTHESIA CONSIDERATION
COPD:CLINICAL REVIEW AND ANESTHESIA CONSIDERATIONDr.RMLIMS lucknow
 
Anticoagulantaandregionalanesthesia review
Anticoagulantaandregionalanesthesia reviewAnticoagulantaandregionalanesthesia review
Anticoagulantaandregionalanesthesia reviewDr.RMLIMS lucknow
 
inhalational agents:brief review
inhalational agents:brief reviewinhalational agents:brief review
inhalational agents:brief reviewDr.RMLIMS lucknow
 

Mais de Dr.RMLIMS lucknow (13)

New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentation
 
Oxygen therapy final
Oxygen therapy  finalOxygen therapy  final
Oxygen therapy final
 
Pneumonectomy
PneumonectomyPneumonectomy
Pneumonectomy
 
Anesthetic management in copd
Anesthetic management in copdAnesthetic management in copd
Anesthetic management in copd
 
Olv nandhu
Olv nandhuOlv nandhu
Olv nandhu
 
Physiology of respiration
Physiology of respirationPhysiology of respiration
Physiology of respiration
 
Anatomy of the lower respiratory tract
Anatomy of the lower respiratory tractAnatomy of the lower respiratory tract
Anatomy of the lower respiratory tract
 
Cns physiology
Cns physiologyCns physiology
Cns physiology
 
ARTERIAL BLOOD GASES INTERPRETATION
ARTERIAL BLOOD GASES INTERPRETATIONARTERIAL BLOOD GASES INTERPRETATION
ARTERIAL BLOOD GASES INTERPRETATION
 
Monitoring Modality in anesthesia
Monitoring Modality  in anesthesiaMonitoring Modality  in anesthesia
Monitoring Modality in anesthesia
 
COPD:CLINICAL REVIEW AND ANESTHESIA CONSIDERATION
COPD:CLINICAL REVIEW AND ANESTHESIA CONSIDERATIONCOPD:CLINICAL REVIEW AND ANESTHESIA CONSIDERATION
COPD:CLINICAL REVIEW AND ANESTHESIA CONSIDERATION
 
Anticoagulantaandregionalanesthesia review
Anticoagulantaandregionalanesthesia reviewAnticoagulantaandregionalanesthesia review
Anticoagulantaandregionalanesthesia review
 
inhalational agents:brief review
inhalational agents:brief reviewinhalational agents:brief review
inhalational agents:brief review
 

Último

Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptxDr.Nusrat Tariq
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...saminamagar
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaPooja Gupta
 
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Me
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near MeHigh Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Me
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Menarwatsonia7
 
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingCall Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingNehru place Escorts
 
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
See the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformSee the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformKweku Zurek
 
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...narwatsonia7
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...narwatsonia7
 
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any TimeCall Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Timevijaych2041
 
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...narwatsonia7
 
Pharmaceutical Marketting: Unit-5, Pricing
Pharmaceutical Marketting: Unit-5, PricingPharmaceutical Marketting: Unit-5, Pricing
Pharmaceutical Marketting: Unit-5, PricingArunagarwal328757
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfMedicoseAcademics
 
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...Nehru place Escorts
 
Hematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes FunctionsHematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes FunctionsMedicoseAcademics
 

Último (20)

Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptx
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
 
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Me
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near MeHigh Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Me
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Me
 
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment BookingCall Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
Call Girls Service Nandiambakkam | 7001305949 At Low Cost Cash Payment Booking
 
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
See the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy PlatformSee the 2,456 pharmacies on the National E-Pharmacy Platform
See the 2,456 pharmacies on the National E-Pharmacy Platform
 
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in paharganj DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
 
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
Russian Call Girl Brookfield - 7001305949 Escorts Service 50% Off with Cash O...
 
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any TimeCall Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Time
 
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...
 
Pharmaceutical Marketting: Unit-5, Pricing
Pharmaceutical Marketting: Unit-5, PricingPharmaceutical Marketting: Unit-5, Pricing
Pharmaceutical Marketting: Unit-5, Pricing
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
 
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...
Call Girls Service in Virugambakkam - 7001305949 | 24x7 Service Available Nea...
 
Hematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes FunctionsHematology and Immunology - Leukocytes Functions
Hematology and Immunology - Leukocytes Functions
 

Oxygen MANUFACTRE STORAGE PREPERATION AND CLINICAL ASPECT

  • 1. Oxygen Manufacturing, Storage,Supply,Therapy and Toxicity by –pawan kumr rai JR Anaesthesiology DR.RMLIMS LUCKNOW
  • 2. How Oxygen is produced? • Fractional Distillation/ Cryogenic Separation • Physical Separation Pressure swing absorption, Membrane filter technology and Electrolysis.
  • 3. Cryogenic Separation • Carl Von Linde in 1895 • The idea behind the process is to lower the temperature of the air such that nitrogen and oxygen separate based on their boiling points. • If higher purity oxygen or liquid oxygen is required, further distillation is required.
  • 4. Pressure swing adsorption(PSA) / Molecular Sieve • PSA devices take atmospheric air into a pressurized tank. • Inside the tank are zeolites.
  • 5. Storage • Units of Pressure: 1 atmosphere = 14.7 psi 760 mmHg 1030 cm H20 1000 mbar 100 KPa
  • 6.
  • 7.
  • 8.
  • 9. RsCr 220 9 Bulk Oxygen Systems (high pressure system) • Large capacity • Liquid oxygen • Small capacity (or back up) Cylinders of different sizes A-J , A is the smallest size and J is the largest size D,E,H :Used most commonly in hospital settings
  • 10.
  • 11. Components of cylinder • Consists of: – Body – Valve – Port – Stem – Pressure relief devices Body •MRI compatible cylinders are made of aluminium • Modern cylinders are made of alloy of MOLYBDENUM + STEEL +/- CHROMIUM It is used to increase strength and to minimise weight and wall thickness
  • 12. Port: It is the point of exit for the gas . Pressure Relief Devices: – Every cylinder is fitted with pressure relief devices whose purpose is to vent the cylinder’s contents to the atmosphere , if the pressure of enclosed gas increases to dangerous levels – Types: • Rupture Disc • Fusible Plug • Combination of Both • Pressure Relief valve (spring loaded) Cylinder valves are of 2 types –packed type and diaphragm type
  • 13.
  • 14. • In order to ensure that the correct cylinder is attached to the appropriate hanger yoke of the anesthesia machine/workstation. A series of pins on the hanger yoke is made to fit into the corresponding indentations(pits/holes) drilled into cylinder valve • Holes are positioned in an arc below the outlet port •The pins are 4.75mm in diameter and 6mm long except for pin number 7 which is slightly thicker and placed at the centre (between port 3 and 4) Pin Index Safety System
  • 15.
  • 16. RsCr 220 Diameter Index Safety System (DISS) • Safety Systems • DISS use noninterchange-able, threaded fittings to connect gas- powered devices to station outlets 50 P.S.I (INTERMEDIATE PRESSURE SYSTEM)
  • 18. •Periodic visual inspection for any distortion, cracks, dents, leaks, etc. •Tensile test on 1 out of every 100 cylinders manufactured. Strips of the cylinder are cut and stretched. •Flattening test, Bend test and impact test are carried out on at least one in every 100 cylinders •Hydraulic test—the cylinder is tested to 1.66 times it service pressure. This helps detect any leaks and to determine the retention of structural strength. A cylinder must be subjected to internal hydrostatic pressure testing at least every 5 years. TESTING OF CYLINDER
  • 19. • Factors for determination of duration • “E” cylinder = 0.28 • “H” cylinder = 3.14 Duration of flow = Pressure x Cylinder factor Flow Liquid system - factor :860 Content=Amount of gas in liquid = Weight of liquid x 860 2.5 Duration of flow = Contents Flow
  • 20. What precautions should be taken while transporting a cylinder? ƒ  Upright position ƒ, Wear protective footwear/safety gloves  ƒDo not lift by protective cap/guard as you may damage valve ƒ  Ensure all cylinders are properly labeled as to the contents ƒ Do not subject cylinders to temperature extremes. ƒ ƒ
  • 21. Cylinders should be handled only by trained person Cylinder valves, regulator,Gauges, and fitting should never come into contact of oils , greases , organic lubricants etc . Cylinders should not be subjected to extreme of temperatures . Connection to piping , regulator, and other equipments should always be kept tight to prevent leakage. No part of the cylinders should be tainted , painted or modified by users . A cylinder should never come into contact of electricity . Before use ,cylinder should be identified by the label. The color of the cylinder should not be relied upon for identification . RULES FOR SAFE USE OF CYLINDERS
  • 22. Only DOT or ICC approved cylinder should be used . A pressure regulator should always be used before any fitting is applied to cylinder valve , the valve should be opened slowly and briefly to remove dust or any other particle ( cracking) . A sealing washer or gasket in good condition should always be used with small cylinder valve .  If more than one seal is used the pin on yoke may not extrude far enough to engage mating whole or safety pin index. A cylinder valve should be opened slowly ,because if gases passes quickly into the space between valve and the yoke or the regulator , the rapid recompression in this space will generate large amount of heat ,because there is no space for dissipation of heat , this constitutes an adiabatic process. Particle of dust, grease, etc present in that space may be ignited by the heat causing a flash fire or explosion. Do not drag slide or roll- use cylinder cart
  • 23. Cylinder Manifold System Intermediate pressure system High pressure system
  • 24.
  • 25. One of the most serious mishaps that occurred was depletion of the oxygen supply. The result was delivery of 100% anesthetic gas. Numerous inventions have been devised to prevent this- • Oxygen failure safety device.1. • Oxygen failure supply alarm.2. Oxygen pressure failure devices
  • 26.
  • 27. Operates in a threshold manner: either open or shut. Pressure Sensor Shut-off Valve:  Oxygen pressure moves the piston and pin upward and the valve opens for N2O When pressure of oxygen falls below preset value, force of the valve return spring completely closes the valve
  • 28. Based on a proportioning principle rather than a threshold principle. Pressure of N2O falls in Propotion of decrease of Oxygen. Total cutoff seen at <12 psig. Seat nozzel assembly connected to a spring loaded conical tapered piston. Oxygen Failure Protection Device
  • 29.
  • 30.
  • 31. The oxygen flush receives oxygen from the pipeline inlet or cylinder pressure regulator and directs a high unmetered flow directly to the common gas outlet. It is commonly labeled “02+.” It can be activated regardless of whether the master switch is turned ON or OFF. A flow between 35 and 75 L/minute must be delivered. Oxygen flush “Single purpose, self closing device & recessed”
  • 33. Mechanical device : Link 25 system Pneumatic device : Oxygen Ratio Monitor Controller (ORMC) Electronically controlled : Penlon Ltd ANTI HYPOXIA DEVICES Mechanical Linking of O2 & N2O. O2 – 14 tooth sprocket, & N2O – 29 tooth sprocket, connected by metal chain. • Both knobs turn together – to maintain min 25% O2 conc.  • Disadvantage – if 3rd gas is administered (eg He,air) – hypoxic mixture may be delivered. Mechanical device : Link 25 system
  • 34. • Mandatory Min O2 Flow (MMOF):- A MMOF – 50-250 ml/min – must as soon as the master switch is turned on. •O2 flow – cant be reduced below that. • Min O2 Ratio :- ASTM Stds require that; A Min O2 conc of 21% in the FGF at common gas outlet.
  • 35. • Pressure is reduced to a 50 psi working pressure in the patient’s room where it can be used to operate respiratory equipment. • Provide connections for gas-delivery devices such as flowmeters and mechanical ventilators. Station outlet
  • 36. Therapy •Oxygen is also like a drug. • It must be used meticulously & deligently. • If abused it can cause complications. • FATHER OF MODERN O2 THERAPY J S HALDANE
  • 37.
  • 38. INDICATIONS FOR O2 THERAPY 1.Hypoxia – when PaO2 comes down to 60mmhg, like low cardiac output states,M.I,anemia,hemodilution,CO poisoning,acute hypermetabolic states. 2.Trapped gases – like obstruction. 3.Special situations – like anesthesia.
  • 39. Indications of hyperbaric oxygen therapy Poisoning: CO Infections: clostridium myonecrosis Acute ischaemia:crush injury Chronic ischaemia: radiation necrosis Ischaemic ulcers: diabetic ulcers
  • 40. 1.Nasal catheters. 8-14 G ,passed through into oropharynx. Reservoir of O2 builds up in pharynx. O2 DELIVERY SYSTEMS: 1.Low flow or variable performance systems. 2.High flow fixed performance systems. LOW FLOW SYSTEMS: O2 Flowrate (L/min) FIO2 1 0.24 2 0.28 3 0.32 4 0.36 5 0.40 6 0.44 Demerits • Mucosal irritation (uncomfortable) • Gastric dilatation (especially with high flows)
  • 41. 2.Nasal prongs: -The prongs protrude 1cm into nares. -Well tolerated. -Useless in mouth breathers. -No reservoir system. 3.Simple O2 masks: - Covers the nose & mouth. -4-6 lits/min. O2 Flowrate (L/min) FIO2 5-6 0.40 6-7 0.50 7-8 0.60
  • 42. 4.Mask with reservoir bags: -High FiO2 with 5-6 lits of O2.
  • 43. 1.VENTURI MASKS: - Works on Bernoulli principle. - Can deliver O2 at accurate concs. HIGH FLOW SYSTEMS: - Total flow is high with or without entrainment
  • 44. 3. O2 tent: -Used in pediatric practise. -10-12 lits/min. -Fire accidents. 2. O2 hood: -Used in bed bound infants not intubated. -Made of plexi glass. -Humidified O2, 3 times M.V. 4.JET VENTILATION: - High frequency oscillation.
  • 45. Drying of mucus membrane Depression of ventilation in COPD Atelectasis due to absorption collapse O2 toxicity HAZARDS OF O2 THERAPY
  • 46. OXYGEN TOXICITY • Increase in oxygen concentration leads to free radical formation • Reacts with cellular DNA proteins, lipids and membrane. 1.Pulmonary oxygen toxicity (LorrainSmith effect) 3.C.N.S O2 toxicity (Paul-Bert effect) 2.Retrolental fibroplasia
  • 47. 1.Pulmonary oxygen toxicity (LorrainSmith effect): Symptoms: substernal pain, irresistable cough, dyspnoea,V.C dec., compliance dec., pulmonary interstitial edema leading to fibrosis. Safe dose of oxygen 100%O2 given not more than12 hours -80% O2 for not more than 24hrs. - 60%O2 not more than 36hrs.
  • 48. 2.Retrolental fibroplasia – Occurs when PaO2 is more than 80mmhg for more than 3 hrs in new born. Very premature babies are more susceptible. - O2 saturation must be around 90-92 %. 3.C.N.S O2 toxicity (Paul-Bert effect): - Seizures or convulsions Depressed ventilatory drive Coma Brain death and eventually death
  • 49.
  • 50. Take home message No drug is more important than oxygen. Medical oxygen is of high purity >99.5%, manufacturing of oxygen by fractional distillation and PSA techniques give us O2 of >99.5% purity. Cylinders has various specifications marked over it, must be checked before. Medical gas supply system has 3 different pneumatic zones, having various safety features. Oxygen is like a drug, life saving when used meticulously otherwise toxic. SAFE DOSE- 100% not>12hrs, 80%not>24hrs, 60%not>36hrs.
  • 51.
  • 53.
  • 54.
  • 55.
  • 56. Defintions Psi: Pounds per square inch Psig: Pounds per square inch gauge GAUGE PRESSURE It is the difference between the measured pressure and the atmospheric pressure Most gauges are measured to read zero at atmospheric pressure • Psia: Pounds per square inch absolute ABSOLUTE PRESSURE It is based on reference point of zero in a perfect vacuum Psia = Psig + atmospheric pressure Eg: at sea level- atmospheric pressure is 0 but psia is 14.7psia •Critical temperature: – It is defined as the temperature below which no gas can be liquefied irrespective of which pressure is applied • Critical Pressure: – Defined as the minimum pressure required to liquefy a gas at its critical temperature