11. CYBERKNIFE EXPERIENCE AT SRC
Installed May 2008
Re-commissioned 2010 with IRIS upgrade
Total Intracranial Patients Treated 466
Total Extracranial Patients Treated 272
TOTAL PATIENTS TREATED 738
12. CYBERKNIFE EXPERIENCE AT SRC
Installed May 2008
Re-commissioned 2010 with IRIS upgrade
AVM/AVOM 1 Breast Met to Brain 41
Trigeminal Neuralgia 62 Renal Met to Brain 12
Vestibular Schwannoma 39 Colon Met to Brain 8
Meningioma 85 Melanoma Met to Brain 6
Pituitary Adenoma 18 Ovarian Met to Brain 1
Glioblastoma 15 Other Metastatic Tumor to Brain 14
Craniopharyngioma 1 Glomus Tumor 3
Hemangioblastoma 2 Astrocytoma/Glioma/GBM 4
Schwannoma 8 Oligodendroglioma/Medulloblastoma 3
Other/Vas/Func Benign Tumors 1 Other Glial Tumors/Other/Unknown 1
Lung Met to Brain 141 Total Intracranial Patients Treated 466
Data Intracranial (5/2008 – 2/2013)
13. CYBERKNIFE EXPERIENCE AT SRC
Data Extracranial
C-spine 8
T-spine 36
L/S-spine 15
Lung 87
Liver 15
Pancreas 5
Head/Neck/ENT 12
Prostate 67
Nasopharynx 1
Other 26
Total Extracranial Patients Treated 272
14. TRACKING MODALITIES
6D Skull
Cranial Lesions
Brain Mets
Trigeminal Neuralgia
Benign Meningiomas
Fiducial Tracking
Body
Prostate
Synchrony
Fiducial Tracking with
respiratory motion
correction
Lung, Liver
X-sight Spine
S,L,T,C Spine
Anything < 5 cm
from spine
X-sight Lung
Lesions >1.5 cm in
periphery of lung
15. 6D SKULL
Alignment center is always set to the center of the skull
Library of 33 pairs of DRRs generated about alignment center
6D correction determined from comparison between live X-rays and DRRS
Similarity measure and rigid transformation based on bony anatomy
Fu et al.: A fast, accurate, and automatic 2D-3D image registration for
image-guided cranial radiosurgery, Med. Phys. 35 (5), May 2008
16. 6D SKULL
6D couch correction is
calculated based on kV X-
rays
Robocouch couch
automatically moves to the
correct position
The Cyberknife robot
adjusts beam targeting
during treatment based
intra-fraction images
(limits: 10 mm, 1.5 degrees)
17. FIDUCIAL TRACKING
Several fiducials surgically implanted in or nearby the tumor
6D tracking requires at least 3 fiducials
20 mm separation, 15°, non-co-linear,
< 5 cm from target
We typically use 0.8 x 3 mm coupled gold markers
18 gauge needle
Fiducials are identified on the CT in MultiPlan and used for
alignment
18. FIDUCIAL TRACKING
“Blobs” are Identified in live X-ray images and compared to a
library of DRRs from the reference CT using a fiducial based
image registration methodology
Intensity thresholds set to live images to bring out blobs
Set of blobs is refined based on expected shape, size, etc.
Ranked by likelyhood
Refine by Inferior Superior location
Blobs with the same I-S position = Same source
Backward project from 2D to 3D space
All potential fiducial configuration candidates compared
to the reference fiducial configuration from the CT
Configurations are ranked and the best fit is used for
alignment
Saw et al.: Implementation of fiducial-based image
registration in the Cyberknife robotic system, Med Dos.
33 (2), 2008
19. FIDUCIAL TRACKING
3 or more fiducial markers are placed inside the tumor with adequate separation
Fiducial pattern is recognized by the Cyberknife imaging system. Marker
locations in the Live X-ray images are compared to expected locations. The
robotic couch automatically repositions the patient.
The Cyberknife makes 6D (X,Y,Z; α,θ,φ) corrections to beam targeting using a
rigid transformation algorithm
Live X-ray images taken during the treatment allows for semi-continuous
monitoring of intra-fraction motion (when not using Synchrony)
20. FIDUCIAL TRACKING
Fiducial Tracking Parameters
Rigid-body distance threshold 1.5 mm
Fiducial spacing threshold 20.0 mm
Colinearity Threshold 15.0°
X-Axis Pairing Tolerance 2.5 mm
Confidence Threshold 60 %
Tracking Range 40 mm
21. SYNCHRONY
Fiducial positions tracked at
discrete points in time
LED Markers monitored in
real time by a camera system
Synchrony establishes a
correlation between external
and internal moments
Robot adjusts beam based on
Synchrony model (translations
only)
22. SYNCHRONY
Breathing Trace
Correlation
Graphs
Coverage of
Breathing
Cycle
Correlation Error
Graph
Nioutsikou et al.: Dosimetric investigation of lung tumor motion compensation with a
robotic respiratory tracking system: An experimental study, Med. Phys. 35 (4), April 2008
Pepin et al.: Correlation and prediction uncertainties in the Cyberknife Synchrony
respiratory tracking system, Med. Phys. 38 (7), July 2011
24. XSIGHT SPINE
Inherent problems aligning spinal anatomy:
Vertebrae can move independent of one another
Rigid transformation may be invalid
Risks associated with surgical fiducial placement
Xsight spine solution: Deformable registration
technique for spine alignment
25. XSIGHT SPINE
Image enhancement
Enhance skeletal structures, suppress soft tissue
DRR generation (17 pairs of DRRs)
ROI placement
Maximum bone information
Skeletal mesh overlayed on spine
2D-3D registration
Spatial transformation base on similarity measure
Local displacement field calculated at each node (81)
3D target location calculated
Maucevic et al.: Technical description, phantom accuracy, and clinical feasibility for
fiducial –free frameless real-time image guided spinal radiosurgery, J. Neurosurg
Spine, 5 October 2006
Furweger et al.: Advances in fiducial-free image-guidance for spinal radiosurgery with
Cyberknife – a phantom study, J. Applied Clinical Med. Phys. 12, (2), Spring 2011
26. XSIGHT SPINE
Difference in spinal anatomy detected between acquired Live X-
ray images and planned DRR images
6D Treatment Couch corrections (X,Y,Z; α,θ,φ) are applied for
initial setup.
The Cyberknife robot adjusts beam targeting during treatment
based intra-fraction images
27. XSIGHT LUNG
Fiducial-less lung tumor tracking
Tracking based on imaging of
the lesion directly
Patient Selection
Target > 15 mm in each axis
Peripherally located
Not obstructed by skeletal
structures
Tracking volume is contoured for
a visual reference
Synchrony used for respiratory
tracking
28. XSIGHT LUNG
Fiducial-less lung tumor tracking
Tracking based on imaging of
the lesion directly
Patient Selection
Target > 15 mm in each axis
Peripherally located
Not obstructed by skeletal
structures
Tracking volume is contoured for
a visual reference
Synchrony used for respiratory
tracking
29. XSIGHT LUNG
Fiducial-less lung tumor tracking
Tracking based on imaging of
the lesion directly
Patient Selection
Target > 15 mm in each axis
Peripherally located
Not obstructed by skeletal
structures
Tracking volume is contoured for
a visual reference
Synchrony used for respiratory
tracking
30. XSIGHT LUNG
Initial patient alignment with Xsight spine
“go to Xsight Lung” Robocouch moves to align to target
Visually confirm that the system truly detects lesion
Build a Synchrony respiratory correlation model
Begin Treatment
Cyberknife adjusts beam targeting during treatment based on
Synchrony and intra-fraction images
38. TREATMENT PLANNING
MultiPlan Version 3.5
Import Image Sets for Planning
CT, MR, PET
Fuse and create contours
Define parameters
Tracking method
Collimation
Conformal vs. Isocentric
Pathset
Optimization and beam reduction
Dose Calculation
Ray trace/Monte Carlo
39. OPTIMIZATION
MU/beam, MU/node
Create shells
VOI Limits
Set global max doses for structures
Objective Steps
Target
Optimize Minimum Dose
Optimize Coverage
Optimize Homogeneity
Critical Structures
Optimize Max Dose
Optimize Mean Dose
40. DOSE CALCULATIONS – RAY TRACING
Calibration Conditions:
dmax = 15 mm
800 SAD
60 mm Fixed Collimator
),(),(
800
),,()/(
2
800 SADcollDMDFSTPR
SAD
DRcollOCRMUD effeff
SAD
RR SAD
800
800
800
SAD
CollFS
)800,60()15,60(
800
800
)15,0,60()/(
2
DMTPROCRMUD
11111)/( 2
MUD
800 mm SAD
cGy/MU
41. DOSE CALCULATIONS – MONTE CARLO
Ray Trace overpredicts dose to PTV
Monte Carlo simulates particle transport and energy
deposition in the patient
Much more accurate dose in presence of heterogeneities
Deng et al.: Commissioning 6 MV photon beams of a stereotactic radiosurgery system for Monte Carlo
treatment planning, Med Phys. 30 (12), December 2003
Wilcox et al.: Comparison of planned dose distributions calculated by monte carlo and ray-trace algorithms
for the treatment of lung tumors with Cyberknife: A preliminary study in 33 patients, Int. J. Radiation
Oncology Biol. Phys. 2009
Wilcox et al.: Stereotactic radiosurgery-radiotherapy: Should Monte Carlo treatment planning be used for all
sites? Practical Radiation Oncology (2011)1, 25
42. EXAMPLE PRESCRIPTION DOSES
Grimm et al.: Dose tolerance limits and dose volume
histogram evaluation for stereotactic body radiotherapy, J.
Applied Clinical Med. Phys. 12, (2), Spring 2011
CRITICAL STRUCTURE TOLERANCE DOSES
Fractions Gy/Fraction Total Dose (Gy)
Prostate 5 7.25 36.25
Trigeminal Neuralgia 1 60 60
Vestibular Schwannoma 3 7 21
Brain mets 3 8 24
1 18 18
Meningioma 5 5 25
Liver 5 7 35
Spine 5 6 30
Lung Ray-Trace 3 20 60
Lung Monte Carlo 3 18 54
46. PHYSICS QA
AQA
End to End Tests (E2E)
Head phantom, Ballcube II
Spine, mini-ballcube
Synchrony motion phantom
Xsight Lung motion phantom
Dose Output, TG-51
Daily
Monthly
Patient Specific QA (PSQA)
IRIS aperture size check
47. AQA
Daily QA check of robot
mastering
Winston Lutz-ish
Concentric circles
AP and Lateral beams
targeted to metal sphere
Aligned with fiducial
tracking
Example EBT3 film Example thresholded images
AQA film
phantom
0°
90°
48. E2E
End to End test
Center ball contoured on CT
70% isodose is centered on the
ball in Multiplan (< 0.1 mm)
Analysis software provided by
Accuray
Delta-Man Adjustments
6D Skull
Fiducial
Xsight spine
49. DOSE CALIBRATION AND QA
TG-51 in water
Atypical TG-51 conditions
Determination of kq
Ref: Toru et al.: Reference
dosimetry condition and
beam quality correction
factor for Cyberknife beam,
Med Phys. 35 (10), October
2008
Monthly calibration check
A14 in solid water phantom
Daily
Birdcage output check
50. PATIENT SPECIFIC QA
Patient-Specific QA performed for
nearly every patient
Patient’s treatment plan (dose rescaled)
delivered to a film-measurement
phantom
Delivered dose is analyzed in RIT and
compared to the prescribed plan by
physics staff and approved prior to
treatment
Verification of Cyberknife targeting and
dose delivery accuracy
Gamma criteria:
5% 1 mm agreement
3% 1 mm in the future with
improved film techniques
Laser-Cut EBT3 gafchromic film
CIRS Anthropomorphic head
phantom with Ballcube II film
insert
54. REFERENCES
Toru et al.: Reference dosimetry condition and beam quality correction factor for
Cyberknife beam, Med Phys. 35 (10), October 2008
Deng et al.: Commissioning 6 MV photon beams of a stereotactic radiosurgery
system for Monte Carlo treatment planning, Med Phys. 30 (12), December 2003
Nioutsikou et al.: Dosimetric investigation of lung tumor motion compensation
with a robotic respiratory tracking system: An experimental study, Med. Phys. 35
(4), April 2008
Sharma et al.: Commissioning and acceptance testing of a Cyberknife linear
accelerator, J. Applied Clinical Med. Phys. 8, (3), Summer 2007
Saw et al.: Implementation of fiducial-based image registration in the
Cyberknife robotic system, Med Dos. 33 (2), 2008
Maucevic et al.: Technical description, phantom accuracy, and clinical feasibility
for fiducial –free frameless real-time image guided spinal radiosurgery, J.
Neurosurg Spine, 5 October 2006
55. REFERENCES
Furweger et al.: Advances in fiducial-free image-guidance for spinal radiosurgery
with Cyberknife – a phantom study, J. Applied Clinical Med. Phys. 12, (2), Spring
2011
Grimm et al.: Dose tolerance limits and dose volume histogram evaluation for
stereotactic body radiotherapy, J. Applied Clinical Med. Phys. 12, (2), Spring 2011
Pepin et al.: Correlation and prediction uncertainties in the Cyberknife
Synchrony respiratory tracking system, Med. Phys. 38 (7), July 2011
Wilcox et al.: Comparison of planned dose distributions calculated by monte
carlo and ray-trace algorithms for the treatment of lung tumors with Cyberknife:
A preliminary study in 33 patients, Int. J. Radiation Oncology Biol. Phys. 2009
Wilcox et al.: Stereotactic radiosurgery-radiotherapy: Should Monte Carlo
treatment planning be used for all sites? Practical Radiation Oncology (2011)1, 25
Chang et al.: An analysis of the accuracy of the CyberKnife: A robotic frameless
stereotactic radiosurgical system, Neurosurgery 52(1)2003
56. REFERENCES
Fu et al.: A fast, accurate, and automatic 2D-3D image registration for image-
guided cranial radiosurgery, Med. Phys. 35 (5), May 2008
Fu et al.: Fiducial-free Lung Tumor Tracking for Cyberknife Radiosurgery, I.J.
Radiation Oncology Biol. Phys. 72 (1), 2979, 2008
Adler, J.R., Chang, S.D., Murphy, M.J., Doty, J., Geis, P., & Hancock,
S.L. (1997). The CyberKnife: A frameless robotic system for radiosurgery.
Stereotactic and Functional Neurosurgery, 69(1–4 Pt. 2),
124–128.
Additional Information and Images taken from:
Accuray Physics Essentials Guide
Accuray Treatment Planning Manual
Accuray Treatment Planning Manual
Technical Training for Radiation Therapists
Accuray Technical Training for Physicians: Full Body Course
57. ACKNOWLEDGEMENTS
Thanks to Saint Raphael’s Cyberknife Team – Physics,
Dosimetry, and Radiation Therapy staff, and the
Smilow Physics group for having me.
Editor's Notes
Increase available beam paths relative to Linac by treating with non-coplanar beam arrangement
Manipulator refers to robot
Programmable positions useful for TG51
Point out on diagram
Heres a visualization of the robotic workspace – or the robots area of travel
SOME of the available paths
Synchrony is an add on to Fiduical tracking
16 degrees off isocenter.
SO heres the treatment flow….
Iterative –
Use Verify by checking points
Discuss intra fraction motion – and how in-between you don’t know! Therapists control image frequency
Moving on to Fiducial tracking
3974 HU.
Mention Pneumothorax as an issue
Mention Breath-hold
Some details on how this works….
Intensity thresholds – so you can imaging thresholding bringing out high density areas – in particular the fiducials
Again – so the typical treatment flow looks something like this
What if can’t detect
What if only 1
Therapists Duties
When working with CK there are several parameters that can be adjusted if needed.
Decisions you make when the patient is on the table
Fid placement doesn’t always meet
X-ray pairing
Rigid Body Maximum deviation of the fiducial from the reference configuration
Confidence – Minimum confidence value in percent
Robotic beam targeting corrections are correlated real-time with patient’s breathing. This allows for continuous tracking and correcting for intra-fraction motion
Linear, 2 Curvilinear, Dual Constrained 4th order polynomial – model depends on detected motion
Correlation error < 2 – our standard
Some details on how this works is that one of the tricks is filtration and enhancement
Image enhancement and filtration
2D Displacements field (constrained by smoothness)
Flow
Interesting way to get around pneumo.
Tracking volume gives therapists a visual reference and also.
Synchrony used here in the same way as with fiducials. – Correlation model made etc.
Upper and Lower bank of tungsten blocks that move to create the various beam sizes.
If you look at the film you can detect this periodicity
When gathering commissioning data its important to completely characterize the beam – only get to put in 1 OCR per collimator szie so you need adequate measurement sampling to get an average profile.
Whats nice is that the planning system gives indication of typical data.
Measured in water with PTW stereotactic diodes 6008 6012
A brief overview on the flow of Planning – Much like all planning systems
Recommended CT set is < 1.5 mm slice thickness
What's different here is setting up the tracking which will be unique to the type of case. Choosing IRIS or fixed,
Overview on treatment planning.,
MU/beam and node – useful for controlling the number of beams used for a treatment
Steps are done as a hierarchy, done in order.
OMD try to get the min dose as close as possible to the goal
OC Maximize the volume getting the goal dose
OH Maximize the volume getting the max dose
Omax dose – minimize dose to be as close to goal as possible
Omean dose – minimize volume getting more than the goal dose.
No volumetric based constraints.
Example Calculation at the calibration position which we expect to have 1cgy/mu
R800 is radial distance at 800 (from CAX )
FS is projected from 800 to SAD to treatment distance
Deff – equivalent path length
People in room much more qualified to give details on this – Holly/ Ravi –
Notice the uncertainly map – lowest at the target where you have the most dose.
Much improved for heterogeneities, especially when there loss of CPE
Worse for small col sized
Depend on anatomical features
We use it for primarily for most Lung treatments., some Liver.
Prescription Doses
Evolving based on Doctors opinions
Maybe bring down lung dose because of chest wall
Typical
CI/nCI – (1.1-1.8)
Hi - (1.1-1.3)
Mention Therapists Duties
AQA which I’ll get into in a second
MORE QA guidelines in TG142 and TG 135 – robotic Radiosurgery
Delta man corrects for any residual error in radiation delivery after mastering. It’s a small adjustment that is the average of 3 main alignment techniques.
Mention Chamber size as an issue.
Nearly every
Targeting- True registration in space, in other words if robot is off it will show up.
Usually > 97% agreement