Method and Validation basics
—HPLC case study
Hua YIN
)Assessor(
The prequalification programme --Assessor's training | 19-20 January 2011
2 |
Outline
HPLC methodology
- Content of HPLC test procedure
- System Suitability Testing (SST)
- Relative Response Factor (RRF)
Validation of HPLC method
case study
The prequalification programme --Assessor's training | 19-20 January 2011
3 |
Information Sources
FDA CDER reviewer guideline for validation of
chromatographic methods (1994)
WHO TRS 937 Appendix 4 “Analytical Method Validation
2006
ICH Q2(R1) 2005
Compendial General Chapters
Methods and Validation presentation–Lynda Paleshnuik
The prequalification programme --Assessor's training | 19-20 January 2011
4 |
High Performance Liquid Chromatography (HPLC)
HPLC is a separation technique based on a solid stationary phase and
a liquid mobile phase. Separations are achieved by partition,
adsorption, or ion-exchange processes, depending upon the type of
stationary phase used.
 Chiral
 Ion--exchange
 Ion--pair/affinity
 Normal phase
 Reversed phase
 Size exclusion
The reversed-phase HPLC with UV detection is most commonly used
form of HPLC, is selected to illustrate the parameters of HPLC
method and validation.
The prequalification programme --Assessor's training | 19-20 January 2011
5 |
A flow scheme for HPLC
The prequalification programme --Assessor's training | 19-20 January 2011
6 |
Content of HPLC test procedure
Any analytical procedure submitted should be described in
sufficient detail, includes:
Preparation of mobile phase
Chromatographic condition:
– Column: type (e.g., C18 or C8), dimension (length, inner
diameter), particle size (10μm, 5 μm)
– Detector: wavelength
– Injection volume
– column T
– flow rate,
The prequalification programme --Assessor's training | 19-20 January 2011
7 |
Content of HPLC test procedure
Elution procedure: isocratic or gradient elution
Preparation of standards and samples
Operation procedure: sequence of injections
System suitability testing (SST) and criteria
Calculations
QOS 2.3.R.2 analytical procedures and validation summaries
The prequalification programme --Assessor's training | 19-20 January 2011
8 |
Compendial methods
When claim a compendial method, there should be no change in:
 The type of column i.e the stationary phases
 Detector wavelength
 Components in Mobile phase
 System suitability testing and criteria
Adjustments to ratio of components in mobile phase, flow rate, column
temp, dimension of column, particle size (reduction only), may be
necessary to achieve the system suitability criteria. The allowable
variations for each parameter, see Int.Ph 1.14.4 or USP general
chapter <621>.
The prequalification programme --Assessor's training | 19-20 January 2011
9 |
System suitability testing (SST)
Precision:
– Assay: RSD ≤1% (API) or ≤ 2% (FPP), n ≥ 5
– Impurities: in general, RSD ≤ 5% at the limit level, up to 10% or
higher at LOQ, n ≥ 6
Resolution (R): >2
The prequalification programme --Assessor's training | 19-20 January 2011
10 |
System suitability testing (SST)
Tailing factor/peak asymmetry: (≤ 2)
The prequalification programme --Assessor's training | 19-20 January 2011
11 |
System suitability testing (SST)
 Number of theoretical plates (N): column efficiency ≥ 2000
 Gradient elution is one way to increase the N
The prequalification programme --Assessor's training | 19-20 January 2011
12 |
System suitability testing (SST)
A SST should contain:
For Assay:
precision + one or more other parameter
For impurity test:
resolution + precision + one or more other parameter
The prequalification programme --Assessor's training | 19-20 January 2011
13 |
Relative Response Factor (RRF)
Quantitation of Impurities
Against impurity RS’s: when reference standard available
Against API itself
Relative response factor should be considered
The prequalification programme --Assessor's training | 19-20 January 2011
14 |
Relative Response Factor (RRF)
Response factor: the response (e.g. peak area) of drug
substance or related substances per unit weight.
RF= peak area / concentration (mg/ml)
Relative response factor (RRF):
RRF=RF impurity / RF API, OR,
RRF=slope impurity / slope API
The prequalification programme --Assessor's training | 19-20 January 2011
15 |
Relative Response Factor (RRF)
Rifampicine:
y =31.312 x + 4.963
Rifampicine Quinone:
y = 26.198 x + 1.154
RRF= 26.198 / 31.312
=0.84
The prequalification programme --Assessor's training | 19-20 January 2011
16 |
Relative Response Factor (RRF)
To review:
a) RRF calculation, and
b) if RRF is properly used in the final calculation for %
impurity
If RRF within 0.8-1.2, correction may not be necessay
Correction factor= 1/RRF, the reciprocal of the RRF
The prequalification programme --Assessor's training | 19-20 January 2011
17 |
Review points for HPLC method
 is the analytical procedure described in detail including all the parameters ?
 is SST well defined to ensure the consistency of system performance?
 The preparation of solutions:
– assay: concentration of reference standard should be close to the
sample solution
– impurities: concentration of the reference standards should be close to
the limit
 The way of quantitation of impurities
In case API is used as the reference, RRF should be used or justification of
exclusion should be provided.
To check the determination of RRF, check the correction of calculation of
impurities
 confirm/complete the QOS 2.3.R.2
The prequalification programme --Assessor's training | 19-20 January 2011
18 |
Validation – compendial methods
Assay – API
No validation generally required. Exception: specificity for major
impurities not in the monograph.
Assay – FPP
Specificity, accuracy and precision (repeatability).
Purity – API and FPP
Full validation for specified impurities that are not included in the
monograph (specificity, linearity, accuracy, repeatability,
intermediate precision, LOD/LOQ)
Validation of the limit for individual unknowns, if tighter than that in
the monograph: LOQ of the API should be below the limit for
individual unknowns
The prequalification programme --Assessor's training | 19-20 January 2011
19 |
Non-compendial methods
Full validation is required for purity, assay and dissolution methods
(HPLC, UV) :
Specificity
Linearity
Accuracy
Repeatability
Intermediate precision
LOD/LOQ (not required for assay, dissolution)
Robustness (recommended)
The prequalification programme --Assessor's training | 19-20 January 2011
20 |
Specificity
 Blank solution to show no interference
 Placebo to demonstrate the lack of interference from excipients
 Spiked samples to show that all known related substances are
resolved from each other
 Stressed sample of about 10 to 20% degradation is used to
demonstrate the resolution among degradation products
– Check peak purity of drug substance by photodiode array detector (PDA):
eg purity angle is lower than the purity threshold.
 Representative chromatograms should be provided with time scale
and attenuation indicated
The prequalification programme --Assessor's training | 19-20 January 2011
21 |
Linearity / Range
The working sample concentration and samples tested for
accuracy should be in the linear range (concentrations
Vs. Peak areas)
Minimum 5 concentrations
Dilute of stock solution or separate weighings
The prequalification programme --Assessor's training | 19-20 January 2011
22 |
Linearity / Range
Assay : 80-120% of the theoretical content of active
Content Uniformity: 70-130%
Dissolution: ±20% of limits; eg if limits cover from
20% to 90% l.c. (controlled release), linearity should
cover 0-110% of l.c.
Impurities: reporting level to 120% of shelf life limit
Assay/Purity by a single method: reporting level of
the impurities to 120% of assay limit
The prequalification programme --Assessor's training | 19-20 January 2011
23 |
Linearity / Range
Correlation coefficient (r)
API: ≥ 0.998
Impurities: ≥ 0.99
y-Intercept and slope should be indicated together with
plot of the data
The prequalification programme --Assessor's training | 19-20 January 2011
24 |
Accuracy
Assay
API: against an RS of known purity, or via an alternate method of
known accuracy; analysis in triplicate.
FPP: samples/placeboes spiked with API, across the range of 80-120%
of the target concentration, 3 concentrations, in triplicate each.
Report per cent recovery (mean result and RSD): 100±2%
ICH Q2 states: accuracy may be inferred once precision, linearity
and specificity have been established. (Demonstration preferred).
The prequalification programme --Assessor's training | 19-20 January 2011
25 |
Accuracy
Impurities: API/FPP spiked with known impurities
Experienced in PQ:
Across the range of LOQ-150% of the target concentration
(shelf life limit), 3-5 concentrations, in triplicate each.
(LOQ, 50%, 100%, 150%)
Per cent recovery: in general, within 80-120%, depends
on the level of limit
The prequalification programme --Assessor's training | 19-20 January 2011
26 |
Precision
System precision:
– by multiple injections (n ≥5) of a homogeneous sample
(standard solution).
– RSD ≤ 1% is recommended for assay;
– RSD ≤ 5% is recommended for related substances (reference
standards at the limit)
– Indicates the performance of the HPLC system
– As a system suitability test
The prequalification programme --Assessor's training | 19-20 January 2011
27 |
Precision
Repeatability (method precision)
– Multiple measurements of a sample by the same analyst
– A minimum of 6 determinations at the test concentration (6
times of a single batch), or
– 3 levels (80%, 100%, 120%) , 3 repetitions each (combined
with accuracy)
– For Assay: RSD ≤ 2.0%
– For individual impurity above 0.05%, in general, RSD ≤ 10%
The prequalification programme --Assessor's training | 19-20 January 2011
28 |
Precision
Intermediate precision (part of ruggedness)
– Test a sample on multiple days, analysts, equipments
– Repeat the method precision by different analyst in
different equipment using different lot of column on
different days
– RSD should be the same requirement as method precision
Reproducibility (inter-laboratory trial)
– Not requested in the submission
The prequalification programme --Assessor's training | 19-20 January 2011
29 |
LOD/LOQ
 signal to noise ratio: LOD 3:1 , LOQ 10:1
– May vary with lamp aging, model/manufacturer of detector, column
 standard deviation of the response and the slope of the calibration
curve at levels approximating the LOD /LOQ
σ = the standard deviation of the response, base on
– the standard deviation of the blank
– The calibration curve
should be validated by analysis of samples at the limits.
The prequalification programme --Assessor's training | 19-20 January 2011
30 |
LOD/LOQ
LOD: below the reporting threshold
LOQ: at or below the specified limit
Not required for assay/dissolution methods.
Applicant should provide
– the method of determination
– the limits,
– chromotograms
The prequalification programme --Assessor's training | 19-20 January 2011
31 |
Robustness
The method's capability to remain unaffected by small but
deliberate variations in method parameters
– Influence of variations of pH in a mobile phase
– Influence of variations in mobile phase composition
– Different columns (different lots and/or suppliers)
– Temperature
– Flow rate
Evaluate the System suitability parameters
The prequalification programme --Assessor's training | 19-20 January 2011
32 |
Robustness
The prequalification programme --Assessor's training | 19-20 January 2011
33 |
Conclusion
HPLC methods play a critical role in analysis of
pharmaceutical product
Validation of HPLC should demonstrate that the method
is suitable for its intended use
Review the information in dossier against QOS 2.3.R.2
Data for acceptance, release, stability will only be
trustworthy if the methods used are reliable
The prequalification programme --Assessor's training | 19-20 January 2011
34 |
Case Study
Case Study 1--HPLC Method.doc
QOS 2.3.R.doc
Case Study 2 -- Validation of HPLC Method.doc

More Related Content

What's hot

HPLC (RP-HPLC) Method Development for simultaneous estimation of EMP & LNG
HPLC (RP-HPLC) Method Development for simultaneous estimation of EMP & LNGHPLC (RP-HPLC) Method Development for simultaneous estimation of EMP & LNG
HPLC (RP-HPLC) Method Development for simultaneous estimation of EMP & LNGLaxmanBulbule1
 
ANALYTICAL METHOD VALIDATION BY P.RAVISANKAR
ANALYTICAL METHOD VALIDATION BY P.RAVISANKAR ANALYTICAL METHOD VALIDATION BY P.RAVISANKAR
ANALYTICAL METHOD VALIDATION BY P.RAVISANKAR Dr. Ravi Sankar
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instrumentsSanathoiba Singha
 
Degradation study and analysis of impurities
Degradation study and analysis of impuritiesDegradation study and analysis of impurities
Degradation study and analysis of impuritiesYachita Rajwadwala
 
HPLC method development
HPLC method developmentHPLC method development
HPLC method developmentAmy Mehaboob
 
analytical method validation
analytical method validationanalytical method validation
analytical method validationDr. Raja Abhilash
 
Analytical method development
Analytical method developmentAnalytical method development
Analytical method developmentSagar Savale
 
ICH Q2 Analytical Method Validation
ICH Q2  Analytical Method ValidationICH Q2  Analytical Method Validation
ICH Q2 Analytical Method ValidationNaila Kanwal
 
Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)Indus Biotech Pvt.Ltd.
 
Analytical procedures life cycle management
Analytical procedures life cycle managementAnalytical procedures life cycle management
Analytical procedures life cycle managementChandra Prakash Singh
 
Bioanalytical Method Validation
Bioanalytical Method ValidationBioanalytical Method Validation
Bioanalytical Method ValidationSanket Shinde
 
Hplc (basic principles, operation and maintenance)
Hplc (basic principles, operation and maintenance)Hplc (basic principles, operation and maintenance)
Hplc (basic principles, operation and maintenance)alim125135
 
Good Chromatographic Practices
Good Chromatographic PracticesGood Chromatographic Practices
Good Chromatographic PracticesDr. Amsavel A
 
Analytical Method Validation basics by Dr. A. Amsavel
Analytical Method Validation  basics by Dr. A. AmsavelAnalytical Method Validation  basics by Dr. A. Amsavel
Analytical Method Validation basics by Dr. A. AmsavelDr. Amsavel A
 
Analytical method developement
Analytical method developementAnalytical method developement
Analytical method developementNarasimha Sharma
 
qualification of Hptlc
qualification of Hptlcqualification of Hptlc
qualification of HptlcAnkush Sule
 
System suitability testing
System suitability testingSystem suitability testing
System suitability testingVrushali Tambe
 

What's hot (20)

Qualification of HPLC
Qualification of HPLCQualification of HPLC
Qualification of HPLC
 
HPLC (RP-HPLC) Method Development for simultaneous estimation of EMP & LNG
HPLC (RP-HPLC) Method Development for simultaneous estimation of EMP & LNGHPLC (RP-HPLC) Method Development for simultaneous estimation of EMP & LNG
HPLC (RP-HPLC) Method Development for simultaneous estimation of EMP & LNG
 
ANALYTICAL METHOD VALIDATION BY P.RAVISANKAR
ANALYTICAL METHOD VALIDATION BY P.RAVISANKAR ANALYTICAL METHOD VALIDATION BY P.RAVISANKAR
ANALYTICAL METHOD VALIDATION BY P.RAVISANKAR
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
 
Degradation study and analysis of impurities
Degradation study and analysis of impuritiesDegradation study and analysis of impurities
Degradation study and analysis of impurities
 
HPLC method development
HPLC method developmentHPLC method development
HPLC method development
 
analytical method validation
analytical method validationanalytical method validation
analytical method validation
 
Analytical method development
Analytical method developmentAnalytical method development
Analytical method development
 
ICH Q2 Analytical Method Validation
ICH Q2  Analytical Method ValidationICH Q2  Analytical Method Validation
ICH Q2 Analytical Method Validation
 
Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)Analytical method validation by manoj ingale(best ppts)
Analytical method validation by manoj ingale(best ppts)
 
Analytical procedures life cycle management
Analytical procedures life cycle managementAnalytical procedures life cycle management
Analytical procedures life cycle management
 
Bioanalytical Method Validation
Bioanalytical Method ValidationBioanalytical Method Validation
Bioanalytical Method Validation
 
Hplc (basic principles, operation and maintenance)
Hplc (basic principles, operation and maintenance)Hplc (basic principles, operation and maintenance)
Hplc (basic principles, operation and maintenance)
 
Good Chromatographic Practices
Good Chromatographic PracticesGood Chromatographic Practices
Good Chromatographic Practices
 
Analytical Method Validation basics by Dr. A. Amsavel
Analytical Method Validation  basics by Dr. A. AmsavelAnalytical Method Validation  basics by Dr. A. Amsavel
Analytical Method Validation basics by Dr. A. Amsavel
 
ICH Q10 Pharmaceutical Quaity System
ICH Q10 Pharmaceutical Quaity SystemICH Q10 Pharmaceutical Quaity System
ICH Q10 Pharmaceutical Quaity System
 
Analytical method developement
Analytical method developementAnalytical method developement
Analytical method developement
 
cGMP AS PER USFDA
cGMP AS PER USFDAcGMP AS PER USFDA
cGMP AS PER USFDA
 
qualification of Hptlc
qualification of Hptlcqualification of Hptlc
qualification of Hptlc
 
System suitability testing
System suitability testingSystem suitability testing
System suitability testing
 

Viewers also liked

Method-Validation-HPLC-case-study
Method-Validation-HPLC-case-studyMethod-Validation-HPLC-case-study
Method-Validation-HPLC-case-studyShreekant Deshpande
 
Method Validation - Repeatability
Method Validation - RepeatabilityMethod Validation - Repeatability
Method Validation - Repeatabilitylabgo
 
Method Validation - Linearity
Method Validation - LinearityMethod Validation - Linearity
Method Validation - Linearitylabgo
 
Method Validation - ICH /USP Validation, Linearity and Repeatability
Method Validation - ICH /USP Validation, Linearity and Repeatability Method Validation - ICH /USP Validation, Linearity and Repeatability
Method Validation - ICH /USP Validation, Linearity and Repeatability labgo
 
analytical method validation and validation of hplc
analytical method validation and validation of hplcanalytical method validation and validation of hplc
analytical method validation and validation of hplcvenkatesh thota
 
zidovudine hplc method
zidovudine hplc method zidovudine hplc method
zidovudine hplc method Aparna Rajeevi
 
Hplc analysis of nicotinamide, pyridoxine, riboflavin
Hplc analysis of nicotinamide, pyridoxine, riboflavinHplc analysis of nicotinamide, pyridoxine, riboflavin
Hplc analysis of nicotinamide, pyridoxine, riboflavinPriyanka Parkar
 
New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentationHARSHITHA REDDY
 
International Analytical Methodology
International Analytical MethodologyInternational Analytical Methodology
International Analytical Methodologysatenvish
 
Validation basics
Validation basicsValidation basics
Validation basicsAnalysys
 
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...Chromatography & Mass Spectrometry Solutions
 
Hplc by niper pharma analysis
Hplc by niper pharma analysisHplc by niper pharma analysis
Hplc by niper pharma analysispatil17
 
HPLC Instrumentation, Pharmaceutical analysis, HPLC detectorsInstmn sem sharu(4)
HPLC Instrumentation, Pharmaceutical analysis, HPLC detectorsInstmn sem sharu(4)HPLC Instrumentation, Pharmaceutical analysis, HPLC detectorsInstmn sem sharu(4)
HPLC Instrumentation, Pharmaceutical analysis, HPLC detectorsInstmn sem sharu(4)Sharath Hns
 
Computer System Validation
Computer System ValidationComputer System Validation
Computer System ValidationEric Silva
 
Stability studies ICH Q1A-Q1E Guidelines ppt
Stability studies ICH Q1A-Q1E Guidelines pptStability studies ICH Q1A-Q1E Guidelines ppt
Stability studies ICH Q1A-Q1E Guidelines pptAman K Dhamrait
 
Analytical Method Validation
Analytical Method ValidationAnalytical Method Validation
Analytical Method ValidationStefan Holt
 

Viewers also liked (20)

Method-Validation-HPLC-case-study
Method-Validation-HPLC-case-studyMethod-Validation-HPLC-case-study
Method-Validation-HPLC-case-study
 
Method Validation - Repeatability
Method Validation - RepeatabilityMethod Validation - Repeatability
Method Validation - Repeatability
 
Method Validation - Linearity
Method Validation - LinearityMethod Validation - Linearity
Method Validation - Linearity
 
Method Validation - ICH /USP Validation, Linearity and Repeatability
Method Validation - ICH /USP Validation, Linearity and Repeatability Method Validation - ICH /USP Validation, Linearity and Repeatability
Method Validation - ICH /USP Validation, Linearity and Repeatability
 
analytical method validation and validation of hplc
analytical method validation and validation of hplcanalytical method validation and validation of hplc
analytical method validation and validation of hplc
 
Validation
Validation Validation
Validation
 
zidovudine hplc method
zidovudine hplc method zidovudine hplc method
zidovudine hplc method
 
Hplc analysis of nicotinamide, pyridoxine, riboflavin
Hplc analysis of nicotinamide, pyridoxine, riboflavinHplc analysis of nicotinamide, pyridoxine, riboflavin
Hplc analysis of nicotinamide, pyridoxine, riboflavin
 
Chromatography Data System: Comply with Regulations
Chromatography Data System: Comply with RegulationsChromatography Data System: Comply with Regulations
Chromatography Data System: Comply with Regulations
 
New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentation
 
International Analytical Methodology
International Analytical MethodologyInternational Analytical Methodology
International Analytical Methodology
 
Validation basics
Validation basicsValidation basics
Validation basics
 
Hplc validation sud mpharm
Hplc validation sud mpharmHplc validation sud mpharm
Hplc validation sud mpharm
 
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
 
Hplc by niper pharma analysis
Hplc by niper pharma analysisHplc by niper pharma analysis
Hplc by niper pharma analysis
 
HPLC Instrumentation, Pharmaceutical analysis, HPLC detectorsInstmn sem sharu(4)
HPLC Instrumentation, Pharmaceutical analysis, HPLC detectorsInstmn sem sharu(4)HPLC Instrumentation, Pharmaceutical analysis, HPLC detectorsInstmn sem sharu(4)
HPLC Instrumentation, Pharmaceutical analysis, HPLC detectorsInstmn sem sharu(4)
 
Computer System Validation
Computer System ValidationComputer System Validation
Computer System Validation
 
Stability studies ICH Q1A-Q1E Guidelines ppt
Stability studies ICH Q1A-Q1E Guidelines pptStability studies ICH Q1A-Q1E Guidelines ppt
Stability studies ICH Q1A-Q1E Guidelines ppt
 
Analytical Method Validation
Analytical Method ValidationAnalytical Method Validation
Analytical Method Validation
 
HPLC
HPLCHPLC
HPLC
 

Similar to 2 4 method-validation-hplc-case-study

Validation of Analytical Methods.pdf
Validation of Analytical Methods.pdfValidation of Analytical Methods.pdf
Validation of Analytical Methods.pdfanjaneyulu49
 
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)Chandra Prakash Singh
 
Analytical quality by design
Analytical quality by designAnalytical quality by design
Analytical quality by designAnnalisa Forlenza
 
Analytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation ParametersAnalytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation ParametersBussaSravanKumar
 
Analytical Method Development
Analytical Method DevelopmentAnalytical Method Development
Analytical Method DevelopmentBijesh Verma
 
Role of Hippophae rhamnoides L. in the Management of Depression by Regulating...
Role of Hippophae rhamnoides L. in the Management of Depression by Regulating...Role of Hippophae rhamnoides L. in the Management of Depression by Regulating...
Role of Hippophae rhamnoides L. in the Management of Depression by Regulating...BRNSS Publication Hub
 
validation of analytical method used in cleaning
validation of analytical method used in cleaningvalidation of analytical method used in cleaning
validation of analytical method used in cleaningKUNDLAJAYALAKSHMI
 
Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014Ramalingam Badmanaban
 
Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014Ramalingam Badmanaban
 
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13Stephan O. Krause, PhD
 
Quality assurance part_2
Quality assurance part_2Quality assurance part_2
Quality assurance part_2ThorikulHuda2
 
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stroundhydrologywebsite1
 
Quality Assurance and Quality Control
Quality Assurance and Quality ControlQuality Assurance and Quality Control
Quality Assurance and Quality ControlECRD IN
 
Quality assurance and quality control
Quality assurance and quality controlQuality assurance and quality control
Quality assurance and quality controlECRD2015
 
Chapter 12
Chapter 12Chapter 12
Chapter 12ECRD IN
 
Estimation of olopatadine hydrochloride by RP–HPLC and U.V spectrophotometry ...
Estimation of olopatadine hydrochloride by RP–HPLC and U.V spectrophotometry ...Estimation of olopatadine hydrochloride by RP–HPLC and U.V spectrophotometry ...
Estimation of olopatadine hydrochloride by RP–HPLC and U.V spectrophotometry ...pharmaindexing
 
Ich guidelines for validation final
Ich guidelines for validation finalIch guidelines for validation final
Ich guidelines for validation finalsumel ashique
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjPratik434909
 

Similar to 2 4 method-validation-hplc-case-study (20)

Validation of Analytical Methods.pdf
Validation of Analytical Methods.pdfValidation of Analytical Methods.pdf
Validation of Analytical Methods.pdf
 
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)
VALIDATION OF ANALYTICAL PROCEDURES Q2(R2)
 
Analytical quality by design
Analytical quality by designAnalytical quality by design
Analytical quality by design
 
Analytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation ParametersAnalytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation Parameters
 
Analytical Method Development
Analytical Method DevelopmentAnalytical Method Development
Analytical Method Development
 
Role of Hippophae rhamnoides L. in the Management of Depression by Regulating...
Role of Hippophae rhamnoides L. in the Management of Depression by Regulating...Role of Hippophae rhamnoides L. in the Management of Depression by Regulating...
Role of Hippophae rhamnoides L. in the Management of Depression by Regulating...
 
06_IJPBA_1736_18[Research]_REV.pdf
06_IJPBA_1736_18[Research]_REV.pdf06_IJPBA_1736_18[Research]_REV.pdf
06_IJPBA_1736_18[Research]_REV.pdf
 
validation of analytical method used in cleaning
validation of analytical method used in cleaningvalidation of analytical method used in cleaning
validation of analytical method used in cleaning
 
Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014
 
Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014Method validation for drug substances and drug product _remodified_2014
Method validation for drug substances and drug product _remodified_2014
 
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13
Risk-based Analytical Method Validation and Maintenance Strategies SK-Sep13
 
Quality assurance part_2
Quality assurance part_2Quality assurance part_2
Quality assurance part_2
 
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
Download-manuals-water quality-wq-manuals-within-labaqcfindings-1stround
 
Quality Assurance and Quality Control
Quality Assurance and Quality ControlQuality Assurance and Quality Control
Quality Assurance and Quality Control
 
Quality assurance and quality control
Quality assurance and quality controlQuality assurance and quality control
Quality assurance and quality control
 
Chapter 12
Chapter 12Chapter 12
Chapter 12
 
Method development and validation
Method development and validationMethod development and validation
Method development and validation
 
Estimation of olopatadine hydrochloride by RP–HPLC and U.V spectrophotometry ...
Estimation of olopatadine hydrochloride by RP–HPLC and U.V spectrophotometry ...Estimation of olopatadine hydrochloride by RP–HPLC and U.V spectrophotometry ...
Estimation of olopatadine hydrochloride by RP–HPLC and U.V spectrophotometry ...
 
Ich guidelines for validation final
Ich guidelines for validation finalIch guidelines for validation final
Ich guidelines for validation final
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
 

Recently uploaded

The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...chandars293
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...vidya singh
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomdiscovermytutordmt
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...Arohi Goyal
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...hotbabesbook
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Dipal Arora
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...parulsinha
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...tanya dube
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Dipal Arora
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Dipal Arora
 
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Genuine Call Girls
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...astropune
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 

Recently uploaded (20)

The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
 
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
Manyata Tech Park ( Call Girls ) Bangalore ✔ 6297143586 ✔ Hot Model With Sexy...
 
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Jabalpur Just Call 9907093804 Top Class Call Girl Service Available
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Bangalore Just Call 9907093804 Top Class Call Girl Service Available
 
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel roomLucknow Call girls - 8800925952 - 24x7 service with hotel room
Lucknow Call girls - 8800925952 - 24x7 service with hotel room
 
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
All Time Service Available Call Girls Marine Drive 📳 9820252231 For 18+ VIP C...
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Kochi Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
Call Girls Visakhapatnam Just Call 9907093804 Top Class Call Girl Service Ava...
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
Premium Bangalore Call Girls Jigani Dail 6378878445 Escort Service For Hot Ma...
 
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
Best Rate (Patna ) Call Girls Patna ⟟ 8617370543 ⟟ High Class Call Girl In 5 ...
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
Best Rate (Guwahati ) Call Girls Guwahati ⟟ 8617370543 ⟟ High Class Call Girl...
 
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
Pondicherry Call Girls Book Now 9630942363 Top Class Pondicherry Escort Servi...
 
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Ooty Just Call 9907093804 Top Class Call Girl Service Available
 
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
♛VVIP Hyderabad Call Girls Chintalkunta🖕7001035870🖕Riya Kappor Top Call Girl ...
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 

2 4 method-validation-hplc-case-study

  • 1. Method and Validation basics —HPLC case study Hua YIN )Assessor(
  • 2. The prequalification programme --Assessor's training | 19-20 January 2011 2 | Outline HPLC methodology - Content of HPLC test procedure - System Suitability Testing (SST) - Relative Response Factor (RRF) Validation of HPLC method case study
  • 3. The prequalification programme --Assessor's training | 19-20 January 2011 3 | Information Sources FDA CDER reviewer guideline for validation of chromatographic methods (1994) WHO TRS 937 Appendix 4 “Analytical Method Validation 2006 ICH Q2(R1) 2005 Compendial General Chapters Methods and Validation presentation–Lynda Paleshnuik
  • 4. The prequalification programme --Assessor's training | 19-20 January 2011 4 | High Performance Liquid Chromatography (HPLC) HPLC is a separation technique based on a solid stationary phase and a liquid mobile phase. Separations are achieved by partition, adsorption, or ion-exchange processes, depending upon the type of stationary phase used.  Chiral  Ion--exchange  Ion--pair/affinity  Normal phase  Reversed phase  Size exclusion The reversed-phase HPLC with UV detection is most commonly used form of HPLC, is selected to illustrate the parameters of HPLC method and validation.
  • 5. The prequalification programme --Assessor's training | 19-20 January 2011 5 | A flow scheme for HPLC
  • 6. The prequalification programme --Assessor's training | 19-20 January 2011 6 | Content of HPLC test procedure Any analytical procedure submitted should be described in sufficient detail, includes: Preparation of mobile phase Chromatographic condition: – Column: type (e.g., C18 or C8), dimension (length, inner diameter), particle size (10μm, 5 μm) – Detector: wavelength – Injection volume – column T – flow rate,
  • 7. The prequalification programme --Assessor's training | 19-20 January 2011 7 | Content of HPLC test procedure Elution procedure: isocratic or gradient elution Preparation of standards and samples Operation procedure: sequence of injections System suitability testing (SST) and criteria Calculations QOS 2.3.R.2 analytical procedures and validation summaries
  • 8. The prequalification programme --Assessor's training | 19-20 January 2011 8 | Compendial methods When claim a compendial method, there should be no change in:  The type of column i.e the stationary phases  Detector wavelength  Components in Mobile phase  System suitability testing and criteria Adjustments to ratio of components in mobile phase, flow rate, column temp, dimension of column, particle size (reduction only), may be necessary to achieve the system suitability criteria. The allowable variations for each parameter, see Int.Ph 1.14.4 or USP general chapter <621>.
  • 9. The prequalification programme --Assessor's training | 19-20 January 2011 9 | System suitability testing (SST) Precision: – Assay: RSD ≤1% (API) or ≤ 2% (FPP), n ≥ 5 – Impurities: in general, RSD ≤ 5% at the limit level, up to 10% or higher at LOQ, n ≥ 6 Resolution (R): >2
  • 10. The prequalification programme --Assessor's training | 19-20 January 2011 10 | System suitability testing (SST) Tailing factor/peak asymmetry: (≤ 2)
  • 11. The prequalification programme --Assessor's training | 19-20 January 2011 11 | System suitability testing (SST)  Number of theoretical plates (N): column efficiency ≥ 2000  Gradient elution is one way to increase the N
  • 12. The prequalification programme --Assessor's training | 19-20 January 2011 12 | System suitability testing (SST) A SST should contain: For Assay: precision + one or more other parameter For impurity test: resolution + precision + one or more other parameter
  • 13. The prequalification programme --Assessor's training | 19-20 January 2011 13 | Relative Response Factor (RRF) Quantitation of Impurities Against impurity RS’s: when reference standard available Against API itself Relative response factor should be considered
  • 14. The prequalification programme --Assessor's training | 19-20 January 2011 14 | Relative Response Factor (RRF) Response factor: the response (e.g. peak area) of drug substance or related substances per unit weight. RF= peak area / concentration (mg/ml) Relative response factor (RRF): RRF=RF impurity / RF API, OR, RRF=slope impurity / slope API
  • 15. The prequalification programme --Assessor's training | 19-20 January 2011 15 | Relative Response Factor (RRF) Rifampicine: y =31.312 x + 4.963 Rifampicine Quinone: y = 26.198 x + 1.154 RRF= 26.198 / 31.312 =0.84
  • 16. The prequalification programme --Assessor's training | 19-20 January 2011 16 | Relative Response Factor (RRF) To review: a) RRF calculation, and b) if RRF is properly used in the final calculation for % impurity If RRF within 0.8-1.2, correction may not be necessay Correction factor= 1/RRF, the reciprocal of the RRF
  • 17. The prequalification programme --Assessor's training | 19-20 January 2011 17 | Review points for HPLC method  is the analytical procedure described in detail including all the parameters ?  is SST well defined to ensure the consistency of system performance?  The preparation of solutions: – assay: concentration of reference standard should be close to the sample solution – impurities: concentration of the reference standards should be close to the limit  The way of quantitation of impurities In case API is used as the reference, RRF should be used or justification of exclusion should be provided. To check the determination of RRF, check the correction of calculation of impurities  confirm/complete the QOS 2.3.R.2
  • 18. The prequalification programme --Assessor's training | 19-20 January 2011 18 | Validation – compendial methods Assay – API No validation generally required. Exception: specificity for major impurities not in the monograph. Assay – FPP Specificity, accuracy and precision (repeatability). Purity – API and FPP Full validation for specified impurities that are not included in the monograph (specificity, linearity, accuracy, repeatability, intermediate precision, LOD/LOQ) Validation of the limit for individual unknowns, if tighter than that in the monograph: LOQ of the API should be below the limit for individual unknowns
  • 19. The prequalification programme --Assessor's training | 19-20 January 2011 19 | Non-compendial methods Full validation is required for purity, assay and dissolution methods (HPLC, UV) : Specificity Linearity Accuracy Repeatability Intermediate precision LOD/LOQ (not required for assay, dissolution) Robustness (recommended)
  • 20. The prequalification programme --Assessor's training | 19-20 January 2011 20 | Specificity  Blank solution to show no interference  Placebo to demonstrate the lack of interference from excipients  Spiked samples to show that all known related substances are resolved from each other  Stressed sample of about 10 to 20% degradation is used to demonstrate the resolution among degradation products – Check peak purity of drug substance by photodiode array detector (PDA): eg purity angle is lower than the purity threshold.  Representative chromatograms should be provided with time scale and attenuation indicated
  • 21. The prequalification programme --Assessor's training | 19-20 January 2011 21 | Linearity / Range The working sample concentration and samples tested for accuracy should be in the linear range (concentrations Vs. Peak areas) Minimum 5 concentrations Dilute of stock solution or separate weighings
  • 22. The prequalification programme --Assessor's training | 19-20 January 2011 22 | Linearity / Range Assay : 80-120% of the theoretical content of active Content Uniformity: 70-130% Dissolution: ±20% of limits; eg if limits cover from 20% to 90% l.c. (controlled release), linearity should cover 0-110% of l.c. Impurities: reporting level to 120% of shelf life limit Assay/Purity by a single method: reporting level of the impurities to 120% of assay limit
  • 23. The prequalification programme --Assessor's training | 19-20 January 2011 23 | Linearity / Range Correlation coefficient (r) API: ≥ 0.998 Impurities: ≥ 0.99 y-Intercept and slope should be indicated together with plot of the data
  • 24. The prequalification programme --Assessor's training | 19-20 January 2011 24 | Accuracy Assay API: against an RS of known purity, or via an alternate method of known accuracy; analysis in triplicate. FPP: samples/placeboes spiked with API, across the range of 80-120% of the target concentration, 3 concentrations, in triplicate each. Report per cent recovery (mean result and RSD): 100±2% ICH Q2 states: accuracy may be inferred once precision, linearity and specificity have been established. (Demonstration preferred).
  • 25. The prequalification programme --Assessor's training | 19-20 January 2011 25 | Accuracy Impurities: API/FPP spiked with known impurities Experienced in PQ: Across the range of LOQ-150% of the target concentration (shelf life limit), 3-5 concentrations, in triplicate each. (LOQ, 50%, 100%, 150%) Per cent recovery: in general, within 80-120%, depends on the level of limit
  • 26. The prequalification programme --Assessor's training | 19-20 January 2011 26 | Precision System precision: – by multiple injections (n ≥5) of a homogeneous sample (standard solution). – RSD ≤ 1% is recommended for assay; – RSD ≤ 5% is recommended for related substances (reference standards at the limit) – Indicates the performance of the HPLC system – As a system suitability test
  • 27. The prequalification programme --Assessor's training | 19-20 January 2011 27 | Precision Repeatability (method precision) – Multiple measurements of a sample by the same analyst – A minimum of 6 determinations at the test concentration (6 times of a single batch), or – 3 levels (80%, 100%, 120%) , 3 repetitions each (combined with accuracy) – For Assay: RSD ≤ 2.0% – For individual impurity above 0.05%, in general, RSD ≤ 10%
  • 28. The prequalification programme --Assessor's training | 19-20 January 2011 28 | Precision Intermediate precision (part of ruggedness) – Test a sample on multiple days, analysts, equipments – Repeat the method precision by different analyst in different equipment using different lot of column on different days – RSD should be the same requirement as method precision Reproducibility (inter-laboratory trial) – Not requested in the submission
  • 29. The prequalification programme --Assessor's training | 19-20 January 2011 29 | LOD/LOQ  signal to noise ratio: LOD 3:1 , LOQ 10:1 – May vary with lamp aging, model/manufacturer of detector, column  standard deviation of the response and the slope of the calibration curve at levels approximating the LOD /LOQ σ = the standard deviation of the response, base on – the standard deviation of the blank – The calibration curve should be validated by analysis of samples at the limits.
  • 30. The prequalification programme --Assessor's training | 19-20 January 2011 30 | LOD/LOQ LOD: below the reporting threshold LOQ: at or below the specified limit Not required for assay/dissolution methods. Applicant should provide – the method of determination – the limits, – chromotograms
  • 31. The prequalification programme --Assessor's training | 19-20 January 2011 31 | Robustness The method's capability to remain unaffected by small but deliberate variations in method parameters – Influence of variations of pH in a mobile phase – Influence of variations in mobile phase composition – Different columns (different lots and/or suppliers) – Temperature – Flow rate Evaluate the System suitability parameters
  • 32. The prequalification programme --Assessor's training | 19-20 January 2011 32 | Robustness
  • 33. The prequalification programme --Assessor's training | 19-20 January 2011 33 | Conclusion HPLC methods play a critical role in analysis of pharmaceutical product Validation of HPLC should demonstrate that the method is suitable for its intended use Review the information in dossier against QOS 2.3.R.2 Data for acceptance, release, stability will only be trustworthy if the methods used are reliable
  • 34. The prequalification programme --Assessor's training | 19-20 January 2011 34 | Case Study Case Study 1--HPLC Method.doc QOS 2.3.R.doc Case Study 2 -- Validation of HPLC Method.doc