SlideShare uma empresa Scribd logo
1 de 13
Baixar para ler offline
APURE Leading Pure
Shanghai GL Environmental Technology Co., Ltd
Model:A10PR
Measuring range
PH: 0.01PH PH: 0.01PH+1Digit 96mm*96mm*132mm(H*W*D)
ORP:1mV ORP:0.1%+1Digit
Temp: 0.1 ℃ Temp:0.1℃ +1Digit 92.5mm*92.5mm(H*W)
Measuring range Accuracy
PH: 0.01PH PH: 0.01PH+1Digit 144mm*144mm*120mm(H*W*D)
ORP:1mV ORP:0.1%+1Digit
Temp: 0.1 ℃ Temp:0.1℃ +1Digit 138mm*138mm(H*W)
●Two 4-20mA outputs independently correspond to PH/ORP or
temperature
Measuring range Accuracy
PH: 0.01PH PH: 0.01PH+1Digit
ORP:1mV ORP:0.1%+1Digit
Temp: 0.1 ℃ Temp:0.1℃ +1Digit
Measuring range Accuracy
PH: 0.01PH PH: 0.01PH+1Digit 144mm*144mm*115mm(H*W*D)
ORP:1mV ORP:0.1%+1Digit
Temp: 0.1 ℃ Temp:0.1℃ +1Digit 138mm*138mm(H*W)
APURE/1 APURE/2
Model:A20PR
Model: RP-3000
Apure industrial online PH&ORP Controller
Model: A30PR
PH: 0~14.00pH
ORP: -1999 ~ 1999 mV
PH: 0~14.00pH
ORP: -1999 ~ 1999 mV
PH: 0~14.00pH
ORP: -1999 ~ 1999 mV
ORP: -1900~1900 mV
PH: 0~14.00pH
Temp: 0-135℃
Temp: 0~135℃ Temp: 0~135℃
Temp: 0~135℃
Apure industrial online PH/ORP Controller:
·48×96mm Microprocessor Water Quality Monitor.
·Compatible with both pH and ORP measurement.
·Compatible with 2-Electrode or 4-Electrode sensors.
·Automatic temperature compensation.
·Compatible with PT1000 and NTC10K temperature probe.
·Linear and non-linear temperature coefficient.
·Isolated current output 4-20mA.
·A30PR offers one relay output, one set point with programmable hystetesis function.
Features:
Features: Features:
Features:
●100-240V or 24VDC; IP65 protection.
pH: Maximum measuring range
of 1 %
mV: Maximum measuring range
of 1%
PT1000/NTC10K compensation
Resolution Accuracy Dimensions
Cut-off dimensions
Resolution Dimensions
Cut-off dimensions
Resolution Resolution Dimensions
Cut-off dimensions
Measurement
●IP65 waterproof and dustproof.
●100-240VAC or 24VDC.
●4-20mA outputs correspond to the PH/ORP
●RS-485(Modbus RTU protocol)Functions (optional)
●Installation: panel or wall mounting
●100-240VAC or 24VDC; IP65 protection.
● PT1000, NTC10K or Manual temp compensation.
● Three limited ON/OFF contact ouptut(relay).
●English and Chinese operation interface, simple operation
●Connect two probes at the same time and measure two PH or PH
and ORP
●Two 4-20mA outputs for pH and ORP.
●Two limited ON/OFF contact output(Relay).
●RS-485 Modbus RTU protocol.
●Installation: panel or wall mounting
and temperature.
●Three limited ON/OFF contact output(Relay).
●RS-485 Modbus RTU protocal(optional).
Shanghai GL Environmental Technology Co., Ltd
APURE/3 APURE/4
Apure industrial online PH sensor Apure industrial online PH sensor
Application: Conventional Sewage treatment,
Swimming Pool treatment, Industrial Process.
Specification:
Measuring range: 0-14pH
Temp range: 0-60℃
Pressure: 0-4bar
Installation: 3/4'' NPT thread
Material: ABS
Application: PH measurement of sewage
treatment with hydrofluoric acid
Specification:
Measuring range:0-14pH
Temp range:0-60℃
Pressure: 0-4bar
Installation: 3/4'' NPT thread
Material: ABS
GRT1010T PH-TI Electrode
For hydrofluoric acid liquid(Colloidal
electroolyte)
GRT1010 Plastic shell PH Electrode
(Colloidal electrolyte)
GRT1010J PTFE shell PH Electrode
(For organic solvent)
GRT1030 Special PH Electrode
For lime solution(Colloidal electroolyte)
GRT1130 High temperature PH electrode
(Jelly electrolyte)
GRT1110 Glass PH Electrode
(Colloidal electrolyte)
GRT1320 High temperature PH Electrode
(Solid electrolyte)
GRT1320A/B Strong Acid(ALKALINE)PH
Electrode(Solid electrolyte)
Application: Sewage treatment, Washing,
Scrubber, Detergents, Industrial Process.
Specification:
Measuring range: 0-14pH
Temp range: 0-60℃
Pressure: 0-4bar
Installation: 3/4'' NPT thread
Material: PTFE
Application: PH measurement of sewage
treatment or clean water treatment
Specification:
Measuring range: 0-14pH
Temp range:0-60℃
Pressure: 0-4bar
Installation: PG13.5 screw thread
Material: Glass
Application: PH measurement of high
temperature water treatment
Specification:
Measuring range:0-14pH
Temp range:0-110℃
Pressure: 0-4bar
Installation: PG13.5 screw thread
Material: Glass
Application: PH measurement of high
temperature water treatment
Specification:
Measuring range:0-14pH
Temp range:0-135℃
Pressure: 0-4bar
Installation: PG13.5 screw thread
Material: Glass
Application: PH measurement of high
temperature strong acid/alkaline water treatment
Specification:
Measuring range:0-14pH
Temp range:0-135℃
Pressure: 0-4bar
Installation: PG13.5 screw thread
Material: Glass
Application: PH measurement of sewage
treatment with lime slurry
Specification:
Measuring range:0-14pH
Temp range: 0-60℃
Pressure: 0-4bar
Installation: 3/4'' NPT thread
Material: ABS
Shanghai GL Environmental Technology Co., Ltd
APURE/5 APURE/6
Apure industrial online ORP sensor APURE PH/ORP sensor mounting sheath
PP1020-100 Submersible sheath
Application: Used for tank installation and
measurement of plastic shell PH electrode
Specification:
Temp range:0-70℃
Pressure: normal pressure
Optional size: 1 m, 2 m, 3m
PP1120-100 Submersible sheath
Application: Used for tank installation and
measurement of glass PH electrode
Specification:
Material: PVC
Temp range:0-70℃
Pressure: normal pressure
Circulation tank type installation sheath
PTB1010;PTB1130(Flow cell)
Application:Circulating (bypass) PH electrode
installation
Specification:
Material: acrylic
Temp range:0-60℃
Pressure: normal pressure
Installation dimension: PTB1010 3/4NPT screw
thread PTB1130 PG13.5 screw thread
Signal amplifier and junction box
Signal amplifier
F4-FD
Order no:60901061
Junction box
F4-05
Order no:60904051
Application: ORP measurement of sewage
treatment with organic solvent
Specification:
Measuring range:-1999mV~1999mV
Temp range:0-60℃
Pressure: 0-4bar
Installation: 3/4'' NPT thread
Material: PTFE
GRT1010P Plastic shell ORP
Electrode(Platinum ring+Agcl)
GRT1010JP ORP Electrode
For organic solvent
(Platinum ring+Agcl)
GRT1110P Glass ORP Electrode
(Platinum ring+Agcl)
GRT1130P High temperature glass
ORPElectrode (Platinum ring+Agcl)
Application:ORP measurement of
general sewage treatment
Specification:
Measuring range:-1999mV~1999mV
Temp range:0-60℃
Pressure: 0-4bar
Installation: 3/4'' NPT therad
Material: ABS
Application: ORP measurement of ambient
temperature sewage
Specification:
Measuring range:-1999mV~1999mv
Temp range: 0-60℃
Pressure: 0-4bar
Installation; PG13.5 screw thread
Material: Glass
Application: ORP measurement of high
temperature sewage
Specification:
Measuring range:-1999mV~1999mv
Temp range:0-110℃
Pressure: 0-4bar
Installation dimension: PG13.5 screw thread
Material: Glass
SS-1130G PF-1120G
Installation dimension: 3/4'' NPT thread
Installation dimension: 3/4'' NPT thread
Optional size: 1 m, 2 m, 3m
Material: PVC (optional: PPR)
PH/ORP/Conductivity Buffer Solution
PH=1.68/4.00/4.01/6.86/7.00/9.18/10.00/10.01
ORP=86MV/256MV
Conductivity=1.3uS/cm, 10uS/cm, 84uS/cm,
100uS/cm, 146.6uS/cm, 1000uS/cm, 1408uS/cm,
1413uS/cm, 12.85mS/cm, 12.88mS/cm,100mS/cm
Capacity:50mL/bottle, 250mL/bottle, 500mL/bottle
Application: For glass PH electrode
Specification:
Material: PVDF; Stainless steel(316L)
Temp range:0-135℃
Pressure: normal pressure
Installation dimension: 3/4’’NPT thread
Shanghai GL Environmental Technology Co., Ltd
APURE/7 APURE/8
Temp
compensation
Accuracy
AUTO ±1%(FS) +Digit 96*96*132mm (H*W*D) 92.5*92.5mm (H*W)
Measuring range
KS-0.01 5-50℃ 0 ~ 0.5Mpa
5-50℃ 0 ~ 0.5Mpa
KS-1 5-50℃ 0 ~ 0.5Mpa
5-120℃ 0 ~ 0.5Mpa
Temp
compensation
Accuracy
AUTO ±1%(FS) +Digit 144*144*115mm
(H*W*D)
92.5*92.5mm (H*W)
Measuring range
KS-0.01 5-50℃ 0 ~ 0.5Mpa
KS-0.1 5-50℃ 0 ~ 0.5Mpa
KS-1 5-50℃ 0 ~ 0.5Mpa
5-120℃ 0 ~ 0.5Mpa
Electrode:KS-1
Electrode:KS-1
Model: A20CD
Electrode:KSS series
KSS-10
KSS-10
Electrode: KSS-10
KS-0.1
•Three limited ON/OFF contact output(relay).
0.1uS/cm-200uS/cm
1uS/cm-20mS/cm
0-9000ppm
10uS/cm-200mS/cm
0.054uS/cm-20uS/cm
1/2 " NPT thread
1/2 " NPT thread
1/2 " NPT thread
1/2 " NPT thread
1/2 " NPT thread
1/2 " NPT thread
1/2 " NPT thread
1/2 " NPT thread
Dimensions
Connection
Dimensions
Connection
Temperature Pressure
Cut-off dimensions
Temperature Pressure
Cut-off dimensions
•Three limited ON/OFF contact ouptut(relay).
Apure Industrial On-line Conductivity/ Resistance/ TDS Controller
•Two 4-20mA outputs correspond to Conductivity/Resistance/TDS
setting and temperature(or one set of 4-20mA &one set of RS485)
•Three limited ON/OFF contact ouptut(relay).
•Password protection.
•NTC10K, PT1000 and Manual temperature
compensation.
Model: A10CD
Apure Industrial On-line Conductivity/Resistance/TDS Controller
1uS/cm-20mS/cm
0-9999ppm
A10CD
Model
Electrode
model
Electrode
model
A20CD
Model
•100-240VAC or 24VDC power supply •100-240VAC or 24VDC power supply.
•Two 4-20MA outputs correspond to conductivity/ Resistance/ TDS setting and temperature
(or one set of 4-20mA &one set of RS485).
• Temperature compensation: PT1000, NTC10K or Manual temp compensation.
• Password protection.
0.05uS/cm-20uS/cm
0.00-20.00MΩ·cm
0.1uS/cm-200uS/cm
10uS/cm-200mS/cm
Shanghai GL Environmental Technology Co., Ltd
APURE/9 APURE/10
Feature:
• One set of 4-20 mA
• One set of relays correspond to
residual chlorine values
•Screen has five ways to display
•IP65 protection level
•220VAC or 24VDC power supply
Model: TC-2200
Model: A10CL
Feature:
• 4-20 mA or RS485 output.
Measuring range Accuracy
CL:
0.01mg/L
±1
%
Power
consumption
Installation
≤ 10W 92.5mm*92.5mm(H*W)
Measuring range
Temp: 0.01 ℃
Applications:
·RO system ·Boiler feed water
·Cooling water ·Ultra-pure water
·Industrial process ·Water treatment
Features:
·Compatible with both conductivity and resistivity measurement
·Compatible with 2-Electrode or 4-Electrode sensors
·Automatic temperature compensation
·Compatible with PT1000 and NTC10K temperature probe
·Linear and non-linear temperature coefficient
·Isolated current output 4~20mA
·A30CD offers one relay output, one set point with programmable hysteresis function.
Compact and Pragmatic
The A30CD conductivity meter utilize microprocessor
technology to provide enhanced capabilities with economical
price. The compact desigh fits in small space installation. The
A30CD features for water quality monitoring and the A30CD
offers one relay output to sufficiently activate your process system. pH sensor & Chlorine sensor
with Flow cell
Manual
·48×96mm Microprocessor Water Quality Meter
Model: TS-270 with flow cell(PP-86)
Model:TS-280
Apure A30CD On-line Conductivity Meter Apure Industrial On-line Chlorine Controller
Apure Industrial On-line Chlorine Controller
Dimensions
Cut-off dimensions
96mm*96mm*132mm(H*W*D)
Flow cell
Resolution
Temperature
0-20.00mg/L
0-60℃
•Three limited ON/OFF contact ouptut(relay).
Chlorine
0-20.00mg/L
PH : 0-14PH
Temp: 0-99 ℃
±1% FS
Accuracy
Resolution
Chlorine
0.01mg/L
PH: 0.01 PH ±0.02 PH
±0.5 ℃ 92mm*92mm(H*W)
Cut-off dimensions
96mm*96mm*132mm(H*W*D)
Dimensions
• IP65 protection.
• 100-240VAC, 24VDC power supply.
Shanghai GL Environmental Technology Co., Ltd
APURE/11 APURE/12
Model: S-730 Model: TD-30
Feature:
• The built-in air bubble elimination
system improves the measurement
precision and stability
• Measurement unit: mg/L and NTU
Model: A10DO Electrode:TS-200 TS-620 TD-400 TD-5000
Meter
model
Accuracy
A10DO 96*96*132mm
(H*W*D)
Electrode
model
Temp resistance Response
time
Accuracy
0.01NTU ±1%FS 242mm*184mm*135mm(H*W*
D)
0.1NTU ±1%FS Installation
Model: A20DO Electrode: TS-200
Feature:
•The built-in air bubble elimination system
improves the measurement precision and stability
• Site calibration can be done by contrast method
Meter
model
Measuring
range
Accuracy
A20DO 0-20.00ppm
(mg/L)
±1.5%(FS) 144*144*115mm (H*W*D) 138*138mm(H*W)
Electrode
model
Temp
resistance
Response
time
Temp compensation
TS-200 0-60℃ 3/4 " NPT thread
Measuring range
TD-30: 0-30g/l 242mm*184mm*135mm(H*W*D)
Installation
Feature:
Apure Industrial On-line Dissolved
Oxygen Controller
Apure Industrial On-line Turbidity
Controller
Apure Industrial On-line Dissolved Oxygen Controller
Apure Industrial On-line Suspended Solids Concentration Controller
Dimensions Cut-off dimensions
Measuring range
Temperature compensation
Submerged or Flow cell
Dimensions
Resolution
Measuring range
RH:≤ 85%
Temperature: 0-60℃
10 sec
Connection
Connection
Dimensions Cut-off dimensions
PT1000 or NTC10K
Resolution Accuracy Dimensions
Submerged
0-60℃
Temperature 1mg/L 1%FS
• Site calibration can be done by contrast
method
TD-5000: 0-5000NTU
TD-400: 0-100.00NTU;
0-400.00NTU
• 4-20mA or RS485 Modbus RTU
• 4-20mA or RS485 Modbus RTU
• 220VAC or 24VDC.
• 220VAC or 24VDC.
• IP65 protection.
• IP65 protection.
Feature:
• For sewage, river water, bredding, agriculture,
aquaculture, etc.
• 4-20mA or RS485 Modbus RTU output.
• Three limited ON/OFF contact output(Relay).
• 100-240VAC or 24VDC, 50/60HZ.
• PT1000, NTC10K, Manual temperature compensation.
• For sewage, river water, bredding, agriculture,
aquaculture, etc.
• 4-20mA or RS485 Modbus RTU output.
• Three limited ON/OFF contact output(Relay).
• 100-240VAC or 24VDC, 50/60HZ.
• PT1000, NTC10K, Manual temperature compensation.
10 sec PT1000 or NTC10K
TS-200 0-60℃ 3/4'' NPT thread
92*92mm (H*W)
±1.5%(FS)
0-20mg/L
(ppm)
Shanghai GL Environmental Technology Co., Ltd
Pressure
Signal ouptut
Pressure
Integrated
Resolution Display
Apure Industrial On-line Salinity Sensor Apure Industrial On-line Ultrasonic Liquid Level Meter
• A set of 4-20mA analog output,two sets of relay
• RS-485 communication(Modbus
RTU agreement)(Optional)
• IP65 protection level
• Sensor material: ABS plastic (PE/PTFE anti-corrosion material is
optional)
Measuring range Measuring range Dead
zone
Range accuracy Pressure
3 atmospheres below
Environment
temperature
1mm -20℃ ~+60℃ Built-in LCD display level
or space distance
•
•
•
•
•
•
•
• Sensor comes with 10m lines(which can be extended to
50m)
Measuring Range
0-20m(can be
customized)
Blind zone
0.25-0.5m ±3mm/ 0.3%
(standard
conditions)
Accuracy
3 atmosphere
below
Display
Pressure
Temperature
Split type 1mm -20~60℃ LCD or space
distance
APURE/13 APURE/14
Structure
Structure
0.5m
0-20M
Apure Industrial On-line Corrosion Rate Digital Sensor
Model: COR-A
Measuring
Range
Temperature
0~5mm/a
±3mm(standard
conditions)
Range resolution
0~80℃
Accuracy Humidity
≤85%
±3%FS
Response
time
Resolution
0.0001mm/a
50S
Power
9-30VDC
Output
Model:AK-7000E
A set of 4-20mA analog output
Four sets of relay switch output
RS-485 communication(Modbus RTU, Optional)
IP65 protection level
220V AC or 24V DC power(Optional)
Installation: split type, meter wall-mounted
Sensor material: ABS plastic (PE/PTFE anti- corrosion
material is optional)
• Installation:one-piece on-site installation and display
• 220V AC or 24VDC(Optional)
Model:AK-2000E
4-20mA&RS485
0-3bar
Pressure
Probe material
Stainless steel, carbon steel, copper
0-10ppm, 0-100ppm,
0-2500oppm, 0-100000ppm
K=0.01, K=0.1,
K=1, K=10
Temp: 0-60.0℃
Mea Resolution:1ppm
Resolution ±0.1℃
Mea Accuracy: ±1%
Accuracy
±1%
4-20mA & RS485 Modbus RTU output
Coefficient
Temp compensation
Auto/Manual
Material
PPS;ABS
0-4Bar
Model: SAL-C
Shanghai GL Environmental Technology Co., Ltd
APURE/15 APURE/16
Integrated type(AYT)
Technical parameter
※DN(mm) Pipeline PTFE lining:
25,32,40,50,65,80,100,125,150,200,250,
300,350,400,450,500,600
※Pipeline rubber lining:-
40,50,65,80,100,125,150,200,300,350,4
00,500,600,800,1000,1200
※Flow direction
Positive, negative, net flow range ratio:
150/1
Repeatability error: ±0.1%
※ Accuracy: 0.3, 0.5, 1.0(class)
※ Temperature
Ordinary rubber lining: -20~60℃
High temp rubber lining: -20~90℃
PTFE lining: -30~100℃
High temp PTFE lining: -30~180℃
※Rated working pressure
Pipeline: DN10~DN65(≤2.5Mpa)
DN80~DN150(≤1.6Mpa);
DN200~DN1200(≤1.0Mpa)
※Conductivity range
Conductivity of measured liquid≥5uS/cm
Split type(AFT)
※Output current and load resistance
4-20mA fully isolated load resistance<750 Ohm
pulse When frequency 0-1KHZ optoelectronic
isolation OCT external power≤35V is connecting,
max current of collector is 25mA.
※Sensor material
Molybdenum-containing stainless steel, Ti, Ta, H, Pt
or other special materials.
※ Protection level
Submerged type: IP68, other IP65.
※Power supply: 85~265V, 45~63HZ
※Straight pipe length
Pipeline upstream≥5DN, downstream≥2DN
※ Connection
Flanged joint is used between flow meter and
distribution pipe, the size should be meet the rule of
B9119-88.
※Explosion-proof: md11BT4
※Environmental temperature: -25~60℃
※Total power consume:≤20W
△Note: RS485 communication is optional. LCD PTFE lining(DN6~DN65), rubber
lining(greater than DN65), 316L sensor, integrated type, no grounded ring(pipe is metal),
can be with RS232, flange mounting, normal pressure is 1.6Mpa.
Electromagnetic
Flow meter
Note Qmin(m3/h) Qmax(m3/h) Dn(mm) Order no
AYT-25S Integrated type 0.5 17.7 25 4112501
AFT-25L Split type 4112502
AYT-32S Integrated type 0.9 29 32 4113201
AFT-32L Split type 4113202
AYT-40S Integrated type 1.4 45.2 40 4114001
AFT-40L Split type 4114002
AYT-50S Integrated type 2.1 70.7 50 4115001
AFT-50L Split type 4115002
AYT-65S Integrated type 4 119 65 4116501
AFT-65L Split type 4116502
AYT-80S Integrated type 5 181 80 4118001
AFT-80L Split type 4118002
AYT-100S Integrated type 8 283 100 41110001
AFT-100L Split type 41110002
AYT-125S Integrated type 13 442 125 41112501
AFT-125L Split type 41112502
AYT-150S Integrated type 19 636 150 41115001
AFT-150L Split type 41115002
AYT-200S Integrated type 30 1130 200 41120001
AFT-200L Split type 41120002
AYT-250S Integrated type 50 1767 250 41125001
AFT-250L Split type 41125002
AYT-300S Integrated type 80 2545 300 41130001
AFT-300L Split type 41130002
AYT-350S Integrated type 100 3464 350 41135001
AFT350L Split type 41135002
AYT-400S Integrated type 140 4524 400 41140001
AFT-400L Split type 41140002
APURE Intelligent Electromagnetic Flow Meter ■Electromagnetic Flow Meter
Data sheet
Shanghai GL Environmental Technology Co., Ltd
APURE/17 APURE/18
Main technical parameters
Measurement principle:
ultrasonic time difference
principle
Measurement range :0~±10m/s
Accuracy: flow rate is better than ±1%
Display :2 x 10 Chinese backlit characters or 2 x 20 character LCD display, support
three languages: Chinese, English and Italian
Flow direction: Positive and negative two-way metering, and measure positive, negative,
net cumulative flow / heat
Data interface; RS485 serial turbulence, can be upgraded by PC computer, support
MODBUS and other protocols
Data recording: positive/negative/net accumulations at 512 days, 128 months and 10
days before automatic memory More than 30 times, power off time and flow rate before
automatic memory and can be automatically or manually added
Signal output: 1 channel 4-20mA current output 1 relay output
1 channel OCT pulse output (programmable between 6~100ms pulse width, default 200ms)
Signal input: 3 channels of 4-20mA analog input, accuracy 01%, can collect pressure, liquid level, warm
sleep and other signals
2-way three-wire PT100 platinum resistor for thermal measurement
Power supply :AC85~264V or DC8~36V
Media turbidity: ≤ 2000pm and small bubble content
Applicable pipe: Carbon steel, stainless steel, copper, PVc, aluminum, fiberglass and other uniform and
dense pipes, allowing lining
Straight pipe section: The sensor installation point is best to meet: upstream 10, downstream 5, from
the pump outlet 300 (D is the diameter)
Measuring
range
Blind
zone
Pressure
±3mm/0.3%(standard
conditions)
3 atmospheres below
Environment temperature
1mm -20℃ ~+60℃ Built-in LCD display level
or space distance
Flow Sensor Standard small sensor L1 DN15~DN100mm, 0~90℃
Standard medium sensor L2 DN50~DN700mm, 0~90℃
Standard large sensor L3 DN300~DN60comm, 0~90℃
High temp small sensor GL1 DN15~DN100mm, 0~160℃
High temp medium sensor GL2 DN50~DN700mm, 0~160℃
High temp large sensor GL3 DN300~DN60comm, 0~160℃
Standard insertion sensor C1 ≥DN80mm
Extended insertion sensor C2 ≥DN80mm
Parallel insertion sensor C3 Customized
Pipe section sensor DN15~DN1000mm
APURE Split Type Open Channel Flow Meter External Clamp Ultrasonic Flow Meter
Resolution
Structure Display
Accuracy
Split type
0-10m 0.25-0.5m
※A set of 4-20mA analog output.
※RS485 communication(Modbus RTU)(optional).
※IP65 protection level.
※220VAC or 24VDC can be choose.
※Installation: Split type, meter wall-mounted.
※Sensor material: ABS plastic(PE/PTFE anti-corrosion material can be choose).
※Sensor comes with 10m cable(max: 50m)
Model: MQ6000
Shanghai GL Environmental Technology Co., Ltd
APURE/19 APURE/20
Figure 1:side mounted
Main technical parameters
【Brand】 Apure
【Measuring range】 200-20000mm
【Medium temperature】 ≤ 100℃
(High temperature can be customized )
【Medium pressure】 ≤ 1.6mpa
【Accuracy】 ±5mm
【Medium density】 ≥ 0.42g/cm³
【Standard flange】 DN20/DN25
【Installation】 Side or top mounted
【Protection level】 IP65
【Measuring materials】 304,316 stainless steel,
PP
【 Optional output 】 remote
transmission, liquid level witch, hart
protocol
Product Selection
Step 1: measuring range
Step 2: Material and installation method
Step 3: shape structure and pressure range
Step 4: Remote transmission and level switch
When there is special need, the user is ordering special
Instructions. Such as density, connection specifications
Model Interface
diameter
Measuring
range
D d H
LZS-15 DN15
5-50L/H
45 20 202
10-100I/h
16-160I/h
25-250I/h
40-400I/h
50-500I/h
60-600I/h
100-1000I/h
LZS-25 DN25
100-1000I/h
60 32 226
160-1600I/h
250-250I/h
LZS-32 DN32 0.4-4m3/h 74 40 288
0.6-6m3/h
LZS-50 DN50
0.4-4m3/h
98 63 341
0.6-6m3/h
1-10m3/h
1.6-16m3/h
LZS-65 DN65
2.5-16m3/h
122 75 430
5-25m3/h
8-40m3/h
12-60m3/h
LZS-100 DN100 14-90m3/h 215 110 555
20-120m3/h
LZS-125 DN125
20-120m3/h
250 140 550
18-150m3/h
20-180m3/h
LZS Float Flow Meter
UHZ Magnetic Flap Level Gauge
Shanghai GL Environmental Technology Co., Ltd
APURE/21 APURE/22
Normal type Explosion-proof type Flat membrane type High temp type
With LED display Clamp type Flange type
Clamp high temp type Differential pressure type
LED high temp type
LCD high temp type
Normal high temp
type
With LCD display
Normal type
Submerged Type Liquid Level Transmitter
Submerged type
PTFE material
Anti-corrosion
Waterproof junction box
For High temperature(coll
ection cylinder)
Material: junction box low copper aluminum alloy; liquid level probe all stainless steel; cable
materialpolyethylene wire, stainless steel tube, PTFE anti-corrosion cable.
Media compatible: Various fluid media compatible with 316L stainless steel
Parameters
Measuring range: 0~10m, 0~100m(can be customized)
Power supply: 12~24VDC
Output signal: 4-20mA, 0-5V, 0-10V
Medium temperature: -40~250℃
Protection:IP68
Pressure overload: 200%FS
Mechanical vibration: 20g(20-5000HZ)
Comprehensive accuracy: 0.5%(standard), 0.2
Stability: ±0.2%FS/Year
Industrial Hammer Pressure Transmitter
Compact Pressure Transmitter
Shanghai GL Environmental Technology Co., Ltd
APURE/23 APURE/24
The digital display instrument cooperates with various types of analog output sensors and transmitters
to measure, transform, display and control physical quantities such as temperature, pressure, flow, liquid
level, composition and displacement.
Overview
The integrated temperature transmitter with thermal resistance and thermocouple is a site-mounted
temperature transmitter unit. It consists of a thermal resistance thermocouple and a temperature
transmitter module. It is a two-wire system with a nonlinear correction circuit that directly measures
liquid, gaseous media and various substances in the industrial process from -150 ° C to 400 ° C.
Temperature signal is converted into a 4-20mADC current output signal linear with the temperature
signal. Display, adjusted recording instrument or computer for distributed control.
Widely used in petroleum machinery, chemical machinery, compressors, electric power, boilers,
natural gas and other automated temperature measurement and control systems.
Performance parameter
Range: -150 ° C ~ 400 ° C
Output signal: 4 ~ 20mA
Supply voltage: 24VDC
Long-term stability: ≤ 0.2% FS / year
Response time: ≤ 20ms
Insertion depth: ≥ 10cm (special requirements can be customized)
Insert diameter: Ф7 (special requirements can be customized)
Nominal pressure: static external pressure on the protective tube at operating temperature
Explosion-proof grade: Exia II CT6 (can be customized)
Power consumption: < 0.5W
1: Ambient temperature: -40 ° C ~ 80 ° C
Working environment 2: relative humidity: 0 ~ 95% RH
3: Mechanical vibration: f < 55HZ, amplitude < 0.15mm
Characteristics
◆Compatible input 2 1 kinds of signals: thermal resistance, thermocouple, standard current, standard
voltage, millivolt, remote pressure gauge resistance, etc. Sensor zero display value, sensor full point
display value, decimal point position can be set freely;
◆Multi-level digital filtering selection, effectively filtering out interference, effectively eliminating
non-critical jump words, realizing alarm, controlling output, transmitting output 0 ~ 1 0 mA、4 ~
2 0 mA 、 0 ~ 2 0 mA 、 0 ~ 5 V、1 ~ 5 V
Model
Content Code Description
XM-601 Digital display
Dimensions
A 160*80*115,80*160*115
B 96*96*60
C 96*48*60,45*96*60
D 72*72*60
E 48*48*115
Panel form
H Horizontal
S Vertical
F Square
Alarm T 2-point alarming
Transmitting Output
A0 No transmitting output
A1 Current output (isolated from
external supply)
A2 Voltage output (isolated from
external supply)
External transmitter power supply (external power supply
input is not isolated, there is isolation
requirement, please specify when
ordering
B0 No external power supply
B1 External power supply 24VDC
B2 External power supply 12VDC
Meter communication
S0 No communication
S1 RS232 communication
S2 RS485 communication
Meter power supply VO 85VCA-265VAC
Intelligent control secondary display instrument
Integrated Temperature sensor / transmitter
Shanghai GL Environmental Technology Co., Ltd

Mais conteúdo relacionado

Mais procurados

1 sinamics components
1 sinamics components1 sinamics components
1 sinamics componentsconfidencial
 
Dewalt Onsite. Novedades y Promociones de la mano de GAMESA.
Dewalt Onsite. Novedades y Promociones de la mano de GAMESA.Dewalt Onsite. Novedades y Promociones de la mano de GAMESA.
Dewalt Onsite. Novedades y Promociones de la mano de GAMESA.Suministros GAME SA
 
Bondstrand serie 2000 m-7000m product data
Bondstrand   serie 2000 m-7000m product dataBondstrand   serie 2000 m-7000m product data
Bondstrand serie 2000 m-7000m product dataMoises Casanova
 
Asme sec viii div 1 s
Asme sec viii div 1 sAsme sec viii div 1 s
Asme sec viii div 1 sJithu John
 
Principles of Pre-reforming Technology
Principles of Pre-reforming TechnologyPrinciples of Pre-reforming Technology
Principles of Pre-reforming TechnologyGerard B. Hawkins
 
ABB Price List -2019..pdf
ABB Price List -2019..pdfABB Price List -2019..pdf
ABB Price List -2019..pdfssuser3f62a8
 
Toy Safety – A Guide to the Latest Updates to ASTM F963
Toy Safety – A Guide to the Latest Updates to ASTM F963Toy Safety – A Guide to the Latest Updates to ASTM F963
Toy Safety – A Guide to the Latest Updates to ASTM F963QIMA
 
Catalog busbar ls 06. busduct presentation e series rev 2
Catalog busbar ls 06. busduct presentation e series rev 2Catalog busbar ls 06. busduct presentation e series rev 2
Catalog busbar ls 06. busduct presentation e series rev 2Dien Ha The
 
Module 6 - Subflowsheet
Module 6 - SubflowsheetModule 6 - Subflowsheet
Module 6 - Subflowsheetriezqaandika
 

Mais procurados (20)

Wps pqr-checklist
Wps pqr-checklistWps pqr-checklist
Wps pqr-checklist
 
Profibus
ProfibusProfibus
Profibus
 
Piping & Instrument Diagram
Piping & Instrument DiagramPiping & Instrument Diagram
Piping & Instrument Diagram
 
1 sinamics components
1 sinamics components1 sinamics components
1 sinamics components
 
Dewalt Onsite. Novedades y Promociones de la mano de GAMESA.
Dewalt Onsite. Novedades y Promociones de la mano de GAMESA.Dewalt Onsite. Novedades y Promociones de la mano de GAMESA.
Dewalt Onsite. Novedades y Promociones de la mano de GAMESA.
 
Bondstrand serie 2000 m-7000m product data
Bondstrand   serie 2000 m-7000m product dataBondstrand   serie 2000 m-7000m product data
Bondstrand serie 2000 m-7000m product data
 
Valve sizing
Valve sizingValve sizing
Valve sizing
 
Asme sec viii div 1 s
Asme sec viii div 1 sAsme sec viii div 1 s
Asme sec viii div 1 s
 
Contactores ABB
Contactores ABBContactores ABB
Contactores ABB
 
Asme sec viii div 1 icb
Asme sec viii div 1 icbAsme sec viii div 1 icb
Asme sec viii div 1 icb
 
Principles of Pre-reforming Technology
Principles of Pre-reforming TechnologyPrinciples of Pre-reforming Technology
Principles of Pre-reforming Technology
 
GBH Enterprises Overview
GBH Enterprises OverviewGBH Enterprises Overview
GBH Enterprises Overview
 
PROFINET - applying the Ethernet standard to industral automation - Peter Th...
PROFINET  - applying the Ethernet standard to industral automation - Peter Th...PROFINET  - applying the Ethernet standard to industral automation - Peter Th...
PROFINET - applying the Ethernet standard to industral automation - Peter Th...
 
ABB Price List -2019..pdf
ABB Price List -2019..pdfABB Price List -2019..pdf
ABB Price List -2019..pdf
 
Toy Safety – A Guide to the Latest Updates to ASTM F963
Toy Safety – A Guide to the Latest Updates to ASTM F963Toy Safety – A Guide to the Latest Updates to ASTM F963
Toy Safety – A Guide to the Latest Updates to ASTM F963
 
Catalog busbar ls 06. busduct presentation e series rev 2
Catalog busbar ls 06. busduct presentation e series rev 2Catalog busbar ls 06. busduct presentation e series rev 2
Catalog busbar ls 06. busduct presentation e series rev 2
 
pipe fitting pnid
pipe fitting pnidpipe fitting pnid
pipe fitting pnid
 
Using HTRI technology within Petro-SIM
Using HTRI technology within Petro-SIM Using HTRI technology within Petro-SIM
Using HTRI technology within Petro-SIM
 
Cable Drum Trailers Manufacturer!
Cable Drum Trailers Manufacturer!Cable Drum Trailers Manufacturer!
Cable Drum Trailers Manufacturer!
 
Module 6 - Subflowsheet
Module 6 - SubflowsheetModule 6 - Subflowsheet
Module 6 - Subflowsheet
 

Semelhante a APURE 2.1 ENG.pdf

Producto 0164-84713
Producto 0164-84713Producto 0164-84713
Producto 0164-84713Raul Mista
 
Explosion Proof Continuous Vertical Liquid Level Sensor (LVCS-FP)
Explosion Proof Continuous Vertical Liquid Level Sensor (LVCS-FP)Explosion Proof Continuous Vertical Liquid Level Sensor (LVCS-FP)
Explosion Proof Continuous Vertical Liquid Level Sensor (LVCS-FP)Thorne & Derrick International
 
i196_2.pdf
i196_2.pdfi196_2.pdf
i196_2.pdfDhpCia
 
Ultrasonic flow meter catalog
Ultrasonic flow meter catalogUltrasonic flow meter catalog
Ultrasonic flow meter catalogKevin Yang
 
Drager Polytron 5700 Fixed Gas Detector - Spec Sheet
Drager Polytron 5700 Fixed Gas Detector - Spec SheetDrager Polytron 5700 Fixed Gas Detector - Spec Sheet
Drager Polytron 5700 Fixed Gas Detector - Spec SheetThorne & Derrick UK
 
EasyTREK / EchoTREK Ultrasonic Level Transmitters for Liquids
EasyTREK / EchoTREK Ultrasonic Level Transmitters for LiquidsEasyTREK / EchoTREK Ultrasonic Level Transmitters for Liquids
EasyTREK / EchoTREK Ultrasonic Level Transmitters for LiquidsArjay Automation
 
ATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdf
ATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdfATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdf
ATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdfENVIMART
 
Dräger Polytron 8700 Fixed Gas Detector - Spec Sheet
Dräger Polytron 8700 Fixed Gas Detector - Spec SheetDräger Polytron 8700 Fixed Gas Detector - Spec Sheet
Dräger Polytron 8700 Fixed Gas Detector - Spec SheetThorne & Derrick UK
 
TIP SERIES Flow Rate Meter & Flow Total Meter with RS-485
TIP SERIES Flow Rate Meter & Flow Total Meter with RS-485TIP SERIES Flow Rate Meter & Flow Total Meter with RS-485
TIP SERIES Flow Rate Meter & Flow Total Meter with RS-485mshasan3
 
T switch - att11-endress+hauser datasheet-thermal flow switch
T switch - att11-endress+hauser datasheet-thermal flow switchT switch - att11-endress+hauser datasheet-thermal flow switch
T switch - att11-endress+hauser datasheet-thermal flow switchENVIMART
 
Drager Polytron 5720 Fixed Gas Detector - Spec Sheet
Drager Polytron 5720 Fixed Gas Detector - Spec SheetDrager Polytron 5720 Fixed Gas Detector - Spec Sheet
Drager Polytron 5720 Fixed Gas Detector - Spec SheetThorne & Derrick UK
 
RBFM Clamp-on Ultrsonic Flow Meter/ Heat Meter Catalog
RBFM Clamp-on Ultrsonic Flow Meter/ Heat Meter CatalogRBFM Clamp-on Ultrsonic Flow Meter/ Heat Meter Catalog
RBFM Clamp-on Ultrsonic Flow Meter/ Heat Meter Catalogjessy Lee RB Flowmeter
 
Kimo MP 200 P Thermo Anemometer Datasheet
Kimo MP 200 P Thermo Anemometer DatasheetKimo MP 200 P Thermo Anemometer Datasheet
Kimo MP 200 P Thermo Anemometer Datasheetalessandrascottfeo
 
Flowfit cca250 endress+hauser datasheet-flow assembly for chlorine and p h:or...
Flowfit cca250 endress+hauser datasheet-flow assembly for chlorine and p h:or...Flowfit cca250 endress+hauser datasheet-flow assembly for chlorine and p h:or...
Flowfit cca250 endress+hauser datasheet-flow assembly for chlorine and p h:or...ENVIMART
 
TIM SERIES Paddle Wheel Flow Meter & Totalizer
TIM SERIES Paddle Wheel Flow Meter & TotalizerTIM SERIES Paddle Wheel Flow Meter & Totalizer
TIM SERIES Paddle Wheel Flow Meter & Totalizermshasan3
 
Omnigrad m tr44 45 endress+hauser datasheet-hygienic, modular rtd assembly-te...
Omnigrad m tr44 45 endress+hauser datasheet-hygienic, modular rtd assembly-te...Omnigrad m tr44 45 endress+hauser datasheet-hygienic, modular rtd assembly-te...
Omnigrad m tr44 45 endress+hauser datasheet-hygienic, modular rtd assembly-te...ENVIMART
 

Semelhante a APURE 2.1 ENG.pdf (20)

Producto 0164-84713
Producto 0164-84713Producto 0164-84713
Producto 0164-84713
 
Explosion Proof Continuous Vertical Liquid Level Sensor (LVCS-FP)
Explosion Proof Continuous Vertical Liquid Level Sensor (LVCS-FP)Explosion Proof Continuous Vertical Liquid Level Sensor (LVCS-FP)
Explosion Proof Continuous Vertical Liquid Level Sensor (LVCS-FP)
 
PMP135
PMP135PMP135
PMP135
 
i196_2.pdf
i196_2.pdfi196_2.pdf
i196_2.pdf
 
Ultrasonic flow meter catalog
Ultrasonic flow meter catalogUltrasonic flow meter catalog
Ultrasonic flow meter catalog
 
Drager Polytron 5700 Fixed Gas Detector - Spec Sheet
Drager Polytron 5700 Fixed Gas Detector - Spec SheetDrager Polytron 5700 Fixed Gas Detector - Spec Sheet
Drager Polytron 5700 Fixed Gas Detector - Spec Sheet
 
EasyTREK / EchoTREK Ultrasonic Level Transmitters for Liquids
EasyTREK / EchoTREK Ultrasonic Level Transmitters for LiquidsEasyTREK / EchoTREK Ultrasonic Level Transmitters for Liquids
EasyTREK / EchoTREK Ultrasonic Level Transmitters for Liquids
 
PMP131
PMP131PMP131
PMP131
 
ATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdf
ATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdfATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdf
ATIs-Q45P-R-pH-ORP-Monitor-Product-Literature.pdf
 
Dräger Polytron 8700 Fixed Gas Detector - Spec Sheet
Dräger Polytron 8700 Fixed Gas Detector - Spec SheetDräger Polytron 8700 Fixed Gas Detector - Spec Sheet
Dräger Polytron 8700 Fixed Gas Detector - Spec Sheet
 
catalogo APURE.pdf
catalogo APURE.pdfcatalogo APURE.pdf
catalogo APURE.pdf
 
TIP SERIES Flow Rate Meter & Flow Total Meter with RS-485
TIP SERIES Flow Rate Meter & Flow Total Meter with RS-485TIP SERIES Flow Rate Meter & Flow Total Meter with RS-485
TIP SERIES Flow Rate Meter & Flow Total Meter with RS-485
 
PTC31 - PTP31
PTC31 - PTP31PTC31 - PTP31
PTC31 - PTP31
 
T switch - att11-endress+hauser datasheet-thermal flow switch
T switch - att11-endress+hauser datasheet-thermal flow switchT switch - att11-endress+hauser datasheet-thermal flow switch
T switch - att11-endress+hauser datasheet-thermal flow switch
 
Drager Polytron 5720 Fixed Gas Detector - Spec Sheet
Drager Polytron 5720 Fixed Gas Detector - Spec SheetDrager Polytron 5720 Fixed Gas Detector - Spec Sheet
Drager Polytron 5720 Fixed Gas Detector - Spec Sheet
 
RBFM Clamp-on Ultrsonic Flow Meter/ Heat Meter Catalog
RBFM Clamp-on Ultrsonic Flow Meter/ Heat Meter CatalogRBFM Clamp-on Ultrsonic Flow Meter/ Heat Meter Catalog
RBFM Clamp-on Ultrsonic Flow Meter/ Heat Meter Catalog
 
Kimo MP 200 P Thermo Anemometer Datasheet
Kimo MP 200 P Thermo Anemometer DatasheetKimo MP 200 P Thermo Anemometer Datasheet
Kimo MP 200 P Thermo Anemometer Datasheet
 
Flowfit cca250 endress+hauser datasheet-flow assembly for chlorine and p h:or...
Flowfit cca250 endress+hauser datasheet-flow assembly for chlorine and p h:or...Flowfit cca250 endress+hauser datasheet-flow assembly for chlorine and p h:or...
Flowfit cca250 endress+hauser datasheet-flow assembly for chlorine and p h:or...
 
TIM SERIES Paddle Wheel Flow Meter & Totalizer
TIM SERIES Paddle Wheel Flow Meter & TotalizerTIM SERIES Paddle Wheel Flow Meter & Totalizer
TIM SERIES Paddle Wheel Flow Meter & Totalizer
 
Omnigrad m tr44 45 endress+hauser datasheet-hygienic, modular rtd assembly-te...
Omnigrad m tr44 45 endress+hauser datasheet-hygienic, modular rtd assembly-te...Omnigrad m tr44 45 endress+hauser datasheet-hygienic, modular rtd assembly-te...
Omnigrad m tr44 45 endress+hauser datasheet-hygienic, modular rtd assembly-te...
 

Último

Arduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptArduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptSAURABHKUMAR892774
 
BSNL Internship Training presentation.pptx
BSNL Internship Training presentation.pptxBSNL Internship Training presentation.pptx
BSNL Internship Training presentation.pptxNiranjanYadav41
 
Internet of things -Arshdeep Bahga .pptx
Internet of things -Arshdeep Bahga .pptxInternet of things -Arshdeep Bahga .pptx
Internet of things -Arshdeep Bahga .pptxVelmuruganTECE
 
Engineering Drawing section of solid
Engineering Drawing     section of solidEngineering Drawing     section of solid
Engineering Drawing section of solidnamansinghjarodiya
 
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor CatchersTechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catcherssdickerson1
 
multiple access in wireless communication
multiple access in wireless communicationmultiple access in wireless communication
multiple access in wireless communicationpanditadesh123
 
Earthing details of Electrical Substation
Earthing details of Electrical SubstationEarthing details of Electrical Substation
Earthing details of Electrical Substationstephanwindworld
 
Sachpazis Costas: Geotechnical Engineering: A student's Perspective Introduction
Sachpazis Costas: Geotechnical Engineering: A student's Perspective IntroductionSachpazis Costas: Geotechnical Engineering: A student's Perspective Introduction
Sachpazis Costas: Geotechnical Engineering: A student's Perspective IntroductionDr.Costas Sachpazis
 
National Level Hackathon Participation Certificate.pdf
National Level Hackathon Participation Certificate.pdfNational Level Hackathon Participation Certificate.pdf
National Level Hackathon Participation Certificate.pdfRajuKanojiya4
 
Main Memory Management in Operating System
Main Memory Management in Operating SystemMain Memory Management in Operating System
Main Memory Management in Operating SystemRashmi Bhat
 
Katarzyna Lipka-Sidor - BIM School Course
Katarzyna Lipka-Sidor - BIM School CourseKatarzyna Lipka-Sidor - BIM School Course
Katarzyna Lipka-Sidor - BIM School Coursebim.edu.pl
 
Configuration of IoT devices - Systems managament
Configuration of IoT devices - Systems managamentConfiguration of IoT devices - Systems managament
Configuration of IoT devices - Systems managamentBharaniDharan195623
 
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdfCCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdfAsst.prof M.Gokilavani
 
Indian Dairy Industry Present Status and.ppt
Indian Dairy Industry Present Status and.pptIndian Dairy Industry Present Status and.ppt
Indian Dairy Industry Present Status and.pptMadan Karki
 
Crystal Structure analysis and detailed information pptx
Crystal Structure analysis and detailed information pptxCrystal Structure analysis and detailed information pptx
Crystal Structure analysis and detailed information pptxachiever3003
 
Ch10-Global Supply Chain - Cadena de Suministro.pdf
Ch10-Global Supply Chain - Cadena de Suministro.pdfCh10-Global Supply Chain - Cadena de Suministro.pdf
Ch10-Global Supply Chain - Cadena de Suministro.pdfChristianCDAM
 
Research Methodology for Engineering pdf
Research Methodology for Engineering pdfResearch Methodology for Engineering pdf
Research Methodology for Engineering pdfCaalaaAbdulkerim
 

Último (20)

Designing pile caps according to ACI 318-19.pptx
Designing pile caps according to ACI 318-19.pptxDesigning pile caps according to ACI 318-19.pptx
Designing pile caps according to ACI 318-19.pptx
 
Arduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.pptArduino_CSE ece ppt for working and principal of arduino.ppt
Arduino_CSE ece ppt for working and principal of arduino.ppt
 
BSNL Internship Training presentation.pptx
BSNL Internship Training presentation.pptxBSNL Internship Training presentation.pptx
BSNL Internship Training presentation.pptx
 
Internet of things -Arshdeep Bahga .pptx
Internet of things -Arshdeep Bahga .pptxInternet of things -Arshdeep Bahga .pptx
Internet of things -Arshdeep Bahga .pptx
 
Engineering Drawing section of solid
Engineering Drawing     section of solidEngineering Drawing     section of solid
Engineering Drawing section of solid
 
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor CatchersTechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
TechTAC® CFD Report Summary: A Comparison of Two Types of Tubing Anchor Catchers
 
multiple access in wireless communication
multiple access in wireless communicationmultiple access in wireless communication
multiple access in wireless communication
 
Earthing details of Electrical Substation
Earthing details of Electrical SubstationEarthing details of Electrical Substation
Earthing details of Electrical Substation
 
Sachpazis Costas: Geotechnical Engineering: A student's Perspective Introduction
Sachpazis Costas: Geotechnical Engineering: A student's Perspective IntroductionSachpazis Costas: Geotechnical Engineering: A student's Perspective Introduction
Sachpazis Costas: Geotechnical Engineering: A student's Perspective Introduction
 
National Level Hackathon Participation Certificate.pdf
National Level Hackathon Participation Certificate.pdfNational Level Hackathon Participation Certificate.pdf
National Level Hackathon Participation Certificate.pdf
 
Main Memory Management in Operating System
Main Memory Management in Operating SystemMain Memory Management in Operating System
Main Memory Management in Operating System
 
Katarzyna Lipka-Sidor - BIM School Course
Katarzyna Lipka-Sidor - BIM School CourseKatarzyna Lipka-Sidor - BIM School Course
Katarzyna Lipka-Sidor - BIM School Course
 
Configuration of IoT devices - Systems managament
Configuration of IoT devices - Systems managamentConfiguration of IoT devices - Systems managament
Configuration of IoT devices - Systems managament
 
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdfCCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
CCS355 Neural Networks & Deep Learning Unit 1 PDF notes with Question bank .pdf
 
Indian Dairy Industry Present Status and.ppt
Indian Dairy Industry Present Status and.pptIndian Dairy Industry Present Status and.ppt
Indian Dairy Industry Present Status and.ppt
 
POWER SYSTEMS-1 Complete notes examples
POWER SYSTEMS-1 Complete notes  examplesPOWER SYSTEMS-1 Complete notes  examples
POWER SYSTEMS-1 Complete notes examples
 
Design and analysis of solar grass cutter.pdf
Design and analysis of solar grass cutter.pdfDesign and analysis of solar grass cutter.pdf
Design and analysis of solar grass cutter.pdf
 
Crystal Structure analysis and detailed information pptx
Crystal Structure analysis and detailed information pptxCrystal Structure analysis and detailed information pptx
Crystal Structure analysis and detailed information pptx
 
Ch10-Global Supply Chain - Cadena de Suministro.pdf
Ch10-Global Supply Chain - Cadena de Suministro.pdfCh10-Global Supply Chain - Cadena de Suministro.pdf
Ch10-Global Supply Chain - Cadena de Suministro.pdf
 
Research Methodology for Engineering pdf
Research Methodology for Engineering pdfResearch Methodology for Engineering pdf
Research Methodology for Engineering pdf
 

APURE 2.1 ENG.pdf

  • 1. APURE Leading Pure Shanghai GL Environmental Technology Co., Ltd
  • 2. Model:A10PR Measuring range PH: 0.01PH PH: 0.01PH+1Digit 96mm*96mm*132mm(H*W*D) ORP:1mV ORP:0.1%+1Digit Temp: 0.1 ℃ Temp:0.1℃ +1Digit 92.5mm*92.5mm(H*W) Measuring range Accuracy PH: 0.01PH PH: 0.01PH+1Digit 144mm*144mm*120mm(H*W*D) ORP:1mV ORP:0.1%+1Digit Temp: 0.1 ℃ Temp:0.1℃ +1Digit 138mm*138mm(H*W) ●Two 4-20mA outputs independently correspond to PH/ORP or temperature Measuring range Accuracy PH: 0.01PH PH: 0.01PH+1Digit ORP:1mV ORP:0.1%+1Digit Temp: 0.1 ℃ Temp:0.1℃ +1Digit Measuring range Accuracy PH: 0.01PH PH: 0.01PH+1Digit 144mm*144mm*115mm(H*W*D) ORP:1mV ORP:0.1%+1Digit Temp: 0.1 ℃ Temp:0.1℃ +1Digit 138mm*138mm(H*W) APURE/1 APURE/2 Model:A20PR Model: RP-3000 Apure industrial online PH&ORP Controller Model: A30PR PH: 0~14.00pH ORP: -1999 ~ 1999 mV PH: 0~14.00pH ORP: -1999 ~ 1999 mV PH: 0~14.00pH ORP: -1999 ~ 1999 mV ORP: -1900~1900 mV PH: 0~14.00pH Temp: 0-135℃ Temp: 0~135℃ Temp: 0~135℃ Temp: 0~135℃ Apure industrial online PH/ORP Controller: ·48×96mm Microprocessor Water Quality Monitor. ·Compatible with both pH and ORP measurement. ·Compatible with 2-Electrode or 4-Electrode sensors. ·Automatic temperature compensation. ·Compatible with PT1000 and NTC10K temperature probe. ·Linear and non-linear temperature coefficient. ·Isolated current output 4-20mA. ·A30PR offers one relay output, one set point with programmable hystetesis function. Features: Features: Features: Features: ●100-240V or 24VDC; IP65 protection. pH: Maximum measuring range of 1 % mV: Maximum measuring range of 1% PT1000/NTC10K compensation Resolution Accuracy Dimensions Cut-off dimensions Resolution Dimensions Cut-off dimensions Resolution Resolution Dimensions Cut-off dimensions Measurement ●IP65 waterproof and dustproof. ●100-240VAC or 24VDC. ●4-20mA outputs correspond to the PH/ORP ●RS-485(Modbus RTU protocol)Functions (optional) ●Installation: panel or wall mounting ●100-240VAC or 24VDC; IP65 protection. ● PT1000, NTC10K or Manual temp compensation. ● Three limited ON/OFF contact ouptut(relay). ●English and Chinese operation interface, simple operation ●Connect two probes at the same time and measure two PH or PH and ORP ●Two 4-20mA outputs for pH and ORP. ●Two limited ON/OFF contact output(Relay). ●RS-485 Modbus RTU protocol. ●Installation: panel or wall mounting and temperature. ●Three limited ON/OFF contact output(Relay). ●RS-485 Modbus RTU protocal(optional). Shanghai GL Environmental Technology Co., Ltd
  • 3. APURE/3 APURE/4 Apure industrial online PH sensor Apure industrial online PH sensor Application: Conventional Sewage treatment, Swimming Pool treatment, Industrial Process. Specification: Measuring range: 0-14pH Temp range: 0-60℃ Pressure: 0-4bar Installation: 3/4'' NPT thread Material: ABS Application: PH measurement of sewage treatment with hydrofluoric acid Specification: Measuring range:0-14pH Temp range:0-60℃ Pressure: 0-4bar Installation: 3/4'' NPT thread Material: ABS GRT1010T PH-TI Electrode For hydrofluoric acid liquid(Colloidal electroolyte) GRT1010 Plastic shell PH Electrode (Colloidal electrolyte) GRT1010J PTFE shell PH Electrode (For organic solvent) GRT1030 Special PH Electrode For lime solution(Colloidal electroolyte) GRT1130 High temperature PH electrode (Jelly electrolyte) GRT1110 Glass PH Electrode (Colloidal electrolyte) GRT1320 High temperature PH Electrode (Solid electrolyte) GRT1320A/B Strong Acid(ALKALINE)PH Electrode(Solid electrolyte) Application: Sewage treatment, Washing, Scrubber, Detergents, Industrial Process. Specification: Measuring range: 0-14pH Temp range: 0-60℃ Pressure: 0-4bar Installation: 3/4'' NPT thread Material: PTFE Application: PH measurement of sewage treatment or clean water treatment Specification: Measuring range: 0-14pH Temp range:0-60℃ Pressure: 0-4bar Installation: PG13.5 screw thread Material: Glass Application: PH measurement of high temperature water treatment Specification: Measuring range:0-14pH Temp range:0-110℃ Pressure: 0-4bar Installation: PG13.5 screw thread Material: Glass Application: PH measurement of high temperature water treatment Specification: Measuring range:0-14pH Temp range:0-135℃ Pressure: 0-4bar Installation: PG13.5 screw thread Material: Glass Application: PH measurement of high temperature strong acid/alkaline water treatment Specification: Measuring range:0-14pH Temp range:0-135℃ Pressure: 0-4bar Installation: PG13.5 screw thread Material: Glass Application: PH measurement of sewage treatment with lime slurry Specification: Measuring range:0-14pH Temp range: 0-60℃ Pressure: 0-4bar Installation: 3/4'' NPT thread Material: ABS Shanghai GL Environmental Technology Co., Ltd
  • 4. APURE/5 APURE/6 Apure industrial online ORP sensor APURE PH/ORP sensor mounting sheath PP1020-100 Submersible sheath Application: Used for tank installation and measurement of plastic shell PH electrode Specification: Temp range:0-70℃ Pressure: normal pressure Optional size: 1 m, 2 m, 3m PP1120-100 Submersible sheath Application: Used for tank installation and measurement of glass PH electrode Specification: Material: PVC Temp range:0-70℃ Pressure: normal pressure Circulation tank type installation sheath PTB1010;PTB1130(Flow cell) Application:Circulating (bypass) PH electrode installation Specification: Material: acrylic Temp range:0-60℃ Pressure: normal pressure Installation dimension: PTB1010 3/4NPT screw thread PTB1130 PG13.5 screw thread Signal amplifier and junction box Signal amplifier F4-FD Order no:60901061 Junction box F4-05 Order no:60904051 Application: ORP measurement of sewage treatment with organic solvent Specification: Measuring range:-1999mV~1999mV Temp range:0-60℃ Pressure: 0-4bar Installation: 3/4'' NPT thread Material: PTFE GRT1010P Plastic shell ORP Electrode(Platinum ring+Agcl) GRT1010JP ORP Electrode For organic solvent (Platinum ring+Agcl) GRT1110P Glass ORP Electrode (Platinum ring+Agcl) GRT1130P High temperature glass ORPElectrode (Platinum ring+Agcl) Application:ORP measurement of general sewage treatment Specification: Measuring range:-1999mV~1999mV Temp range:0-60℃ Pressure: 0-4bar Installation: 3/4'' NPT therad Material: ABS Application: ORP measurement of ambient temperature sewage Specification: Measuring range:-1999mV~1999mv Temp range: 0-60℃ Pressure: 0-4bar Installation; PG13.5 screw thread Material: Glass Application: ORP measurement of high temperature sewage Specification: Measuring range:-1999mV~1999mv Temp range:0-110℃ Pressure: 0-4bar Installation dimension: PG13.5 screw thread Material: Glass SS-1130G PF-1120G Installation dimension: 3/4'' NPT thread Installation dimension: 3/4'' NPT thread Optional size: 1 m, 2 m, 3m Material: PVC (optional: PPR) PH/ORP/Conductivity Buffer Solution PH=1.68/4.00/4.01/6.86/7.00/9.18/10.00/10.01 ORP=86MV/256MV Conductivity=1.3uS/cm, 10uS/cm, 84uS/cm, 100uS/cm, 146.6uS/cm, 1000uS/cm, 1408uS/cm, 1413uS/cm, 12.85mS/cm, 12.88mS/cm,100mS/cm Capacity:50mL/bottle, 250mL/bottle, 500mL/bottle Application: For glass PH electrode Specification: Material: PVDF; Stainless steel(316L) Temp range:0-135℃ Pressure: normal pressure Installation dimension: 3/4’’NPT thread Shanghai GL Environmental Technology Co., Ltd
  • 5. APURE/7 APURE/8 Temp compensation Accuracy AUTO ±1%(FS) +Digit 96*96*132mm (H*W*D) 92.5*92.5mm (H*W) Measuring range KS-0.01 5-50℃ 0 ~ 0.5Mpa 5-50℃ 0 ~ 0.5Mpa KS-1 5-50℃ 0 ~ 0.5Mpa 5-120℃ 0 ~ 0.5Mpa Temp compensation Accuracy AUTO ±1%(FS) +Digit 144*144*115mm (H*W*D) 92.5*92.5mm (H*W) Measuring range KS-0.01 5-50℃ 0 ~ 0.5Mpa KS-0.1 5-50℃ 0 ~ 0.5Mpa KS-1 5-50℃ 0 ~ 0.5Mpa 5-120℃ 0 ~ 0.5Mpa Electrode:KS-1 Electrode:KS-1 Model: A20CD Electrode:KSS series KSS-10 KSS-10 Electrode: KSS-10 KS-0.1 •Three limited ON/OFF contact output(relay). 0.1uS/cm-200uS/cm 1uS/cm-20mS/cm 0-9000ppm 10uS/cm-200mS/cm 0.054uS/cm-20uS/cm 1/2 " NPT thread 1/2 " NPT thread 1/2 " NPT thread 1/2 " NPT thread 1/2 " NPT thread 1/2 " NPT thread 1/2 " NPT thread 1/2 " NPT thread Dimensions Connection Dimensions Connection Temperature Pressure Cut-off dimensions Temperature Pressure Cut-off dimensions •Three limited ON/OFF contact ouptut(relay). Apure Industrial On-line Conductivity/ Resistance/ TDS Controller •Two 4-20mA outputs correspond to Conductivity/Resistance/TDS setting and temperature(or one set of 4-20mA &one set of RS485) •Three limited ON/OFF contact ouptut(relay). •Password protection. •NTC10K, PT1000 and Manual temperature compensation. Model: A10CD Apure Industrial On-line Conductivity/Resistance/TDS Controller 1uS/cm-20mS/cm 0-9999ppm A10CD Model Electrode model Electrode model A20CD Model •100-240VAC or 24VDC power supply •100-240VAC or 24VDC power supply. •Two 4-20MA outputs correspond to conductivity/ Resistance/ TDS setting and temperature (or one set of 4-20mA &one set of RS485). • Temperature compensation: PT1000, NTC10K or Manual temp compensation. • Password protection. 0.05uS/cm-20uS/cm 0.00-20.00MΩ·cm 0.1uS/cm-200uS/cm 10uS/cm-200mS/cm Shanghai GL Environmental Technology Co., Ltd
  • 6. APURE/9 APURE/10 Feature: • One set of 4-20 mA • One set of relays correspond to residual chlorine values •Screen has five ways to display •IP65 protection level •220VAC or 24VDC power supply Model: TC-2200 Model: A10CL Feature: • 4-20 mA or RS485 output. Measuring range Accuracy CL: 0.01mg/L ±1 % Power consumption Installation ≤ 10W 92.5mm*92.5mm(H*W) Measuring range Temp: 0.01 ℃ Applications: ·RO system ·Boiler feed water ·Cooling water ·Ultra-pure water ·Industrial process ·Water treatment Features: ·Compatible with both conductivity and resistivity measurement ·Compatible with 2-Electrode or 4-Electrode sensors ·Automatic temperature compensation ·Compatible with PT1000 and NTC10K temperature probe ·Linear and non-linear temperature coefficient ·Isolated current output 4~20mA ·A30CD offers one relay output, one set point with programmable hysteresis function. Compact and Pragmatic The A30CD conductivity meter utilize microprocessor technology to provide enhanced capabilities with economical price. The compact desigh fits in small space installation. The A30CD features for water quality monitoring and the A30CD offers one relay output to sufficiently activate your process system. pH sensor & Chlorine sensor with Flow cell Manual ·48×96mm Microprocessor Water Quality Meter Model: TS-270 with flow cell(PP-86) Model:TS-280 Apure A30CD On-line Conductivity Meter Apure Industrial On-line Chlorine Controller Apure Industrial On-line Chlorine Controller Dimensions Cut-off dimensions 96mm*96mm*132mm(H*W*D) Flow cell Resolution Temperature 0-20.00mg/L 0-60℃ •Three limited ON/OFF contact ouptut(relay). Chlorine 0-20.00mg/L PH : 0-14PH Temp: 0-99 ℃ ±1% FS Accuracy Resolution Chlorine 0.01mg/L PH: 0.01 PH ±0.02 PH ±0.5 ℃ 92mm*92mm(H*W) Cut-off dimensions 96mm*96mm*132mm(H*W*D) Dimensions • IP65 protection. • 100-240VAC, 24VDC power supply. Shanghai GL Environmental Technology Co., Ltd
  • 7. APURE/11 APURE/12 Model: S-730 Model: TD-30 Feature: • The built-in air bubble elimination system improves the measurement precision and stability • Measurement unit: mg/L and NTU Model: A10DO Electrode:TS-200 TS-620 TD-400 TD-5000 Meter model Accuracy A10DO 96*96*132mm (H*W*D) Electrode model Temp resistance Response time Accuracy 0.01NTU ±1%FS 242mm*184mm*135mm(H*W* D) 0.1NTU ±1%FS Installation Model: A20DO Electrode: TS-200 Feature: •The built-in air bubble elimination system improves the measurement precision and stability • Site calibration can be done by contrast method Meter model Measuring range Accuracy A20DO 0-20.00ppm (mg/L) ±1.5%(FS) 144*144*115mm (H*W*D) 138*138mm(H*W) Electrode model Temp resistance Response time Temp compensation TS-200 0-60℃ 3/4 " NPT thread Measuring range TD-30: 0-30g/l 242mm*184mm*135mm(H*W*D) Installation Feature: Apure Industrial On-line Dissolved Oxygen Controller Apure Industrial On-line Turbidity Controller Apure Industrial On-line Dissolved Oxygen Controller Apure Industrial On-line Suspended Solids Concentration Controller Dimensions Cut-off dimensions Measuring range Temperature compensation Submerged or Flow cell Dimensions Resolution Measuring range RH:≤ 85% Temperature: 0-60℃ 10 sec Connection Connection Dimensions Cut-off dimensions PT1000 or NTC10K Resolution Accuracy Dimensions Submerged 0-60℃ Temperature 1mg/L 1%FS • Site calibration can be done by contrast method TD-5000: 0-5000NTU TD-400: 0-100.00NTU; 0-400.00NTU • 4-20mA or RS485 Modbus RTU • 4-20mA or RS485 Modbus RTU • 220VAC or 24VDC. • 220VAC or 24VDC. • IP65 protection. • IP65 protection. Feature: • For sewage, river water, bredding, agriculture, aquaculture, etc. • 4-20mA or RS485 Modbus RTU output. • Three limited ON/OFF contact output(Relay). • 100-240VAC or 24VDC, 50/60HZ. • PT1000, NTC10K, Manual temperature compensation. • For sewage, river water, bredding, agriculture, aquaculture, etc. • 4-20mA or RS485 Modbus RTU output. • Three limited ON/OFF contact output(Relay). • 100-240VAC or 24VDC, 50/60HZ. • PT1000, NTC10K, Manual temperature compensation. 10 sec PT1000 or NTC10K TS-200 0-60℃ 3/4'' NPT thread 92*92mm (H*W) ±1.5%(FS) 0-20mg/L (ppm) Shanghai GL Environmental Technology Co., Ltd
  • 8. Pressure Signal ouptut Pressure Integrated Resolution Display Apure Industrial On-line Salinity Sensor Apure Industrial On-line Ultrasonic Liquid Level Meter • A set of 4-20mA analog output,two sets of relay • RS-485 communication(Modbus RTU agreement)(Optional) • IP65 protection level • Sensor material: ABS plastic (PE/PTFE anti-corrosion material is optional) Measuring range Measuring range Dead zone Range accuracy Pressure 3 atmospheres below Environment temperature 1mm -20℃ ~+60℃ Built-in LCD display level or space distance • • • • • • • • Sensor comes with 10m lines(which can be extended to 50m) Measuring Range 0-20m(can be customized) Blind zone 0.25-0.5m ±3mm/ 0.3% (standard conditions) Accuracy 3 atmosphere below Display Pressure Temperature Split type 1mm -20~60℃ LCD or space distance APURE/13 APURE/14 Structure Structure 0.5m 0-20M Apure Industrial On-line Corrosion Rate Digital Sensor Model: COR-A Measuring Range Temperature 0~5mm/a ±3mm(standard conditions) Range resolution 0~80℃ Accuracy Humidity ≤85% ±3%FS Response time Resolution 0.0001mm/a 50S Power 9-30VDC Output Model:AK-7000E A set of 4-20mA analog output Four sets of relay switch output RS-485 communication(Modbus RTU, Optional) IP65 protection level 220V AC or 24V DC power(Optional) Installation: split type, meter wall-mounted Sensor material: ABS plastic (PE/PTFE anti- corrosion material is optional) • Installation:one-piece on-site installation and display • 220V AC or 24VDC(Optional) Model:AK-2000E 4-20mA&RS485 0-3bar Pressure Probe material Stainless steel, carbon steel, copper 0-10ppm, 0-100ppm, 0-2500oppm, 0-100000ppm K=0.01, K=0.1, K=1, K=10 Temp: 0-60.0℃ Mea Resolution:1ppm Resolution ±0.1℃ Mea Accuracy: ±1% Accuracy ±1% 4-20mA & RS485 Modbus RTU output Coefficient Temp compensation Auto/Manual Material PPS;ABS 0-4Bar Model: SAL-C Shanghai GL Environmental Technology Co., Ltd
  • 9. APURE/15 APURE/16 Integrated type(AYT) Technical parameter ※DN(mm) Pipeline PTFE lining: 25,32,40,50,65,80,100,125,150,200,250, 300,350,400,450,500,600 ※Pipeline rubber lining:- 40,50,65,80,100,125,150,200,300,350,4 00,500,600,800,1000,1200 ※Flow direction Positive, negative, net flow range ratio: 150/1 Repeatability error: ±0.1% ※ Accuracy: 0.3, 0.5, 1.0(class) ※ Temperature Ordinary rubber lining: -20~60℃ High temp rubber lining: -20~90℃ PTFE lining: -30~100℃ High temp PTFE lining: -30~180℃ ※Rated working pressure Pipeline: DN10~DN65(≤2.5Mpa) DN80~DN150(≤1.6Mpa); DN200~DN1200(≤1.0Mpa) ※Conductivity range Conductivity of measured liquid≥5uS/cm Split type(AFT) ※Output current and load resistance 4-20mA fully isolated load resistance<750 Ohm pulse When frequency 0-1KHZ optoelectronic isolation OCT external power≤35V is connecting, max current of collector is 25mA. ※Sensor material Molybdenum-containing stainless steel, Ti, Ta, H, Pt or other special materials. ※ Protection level Submerged type: IP68, other IP65. ※Power supply: 85~265V, 45~63HZ ※Straight pipe length Pipeline upstream≥5DN, downstream≥2DN ※ Connection Flanged joint is used between flow meter and distribution pipe, the size should be meet the rule of B9119-88. ※Explosion-proof: md11BT4 ※Environmental temperature: -25~60℃ ※Total power consume:≤20W △Note: RS485 communication is optional. LCD PTFE lining(DN6~DN65), rubber lining(greater than DN65), 316L sensor, integrated type, no grounded ring(pipe is metal), can be with RS232, flange mounting, normal pressure is 1.6Mpa. Electromagnetic Flow meter Note Qmin(m3/h) Qmax(m3/h) Dn(mm) Order no AYT-25S Integrated type 0.5 17.7 25 4112501 AFT-25L Split type 4112502 AYT-32S Integrated type 0.9 29 32 4113201 AFT-32L Split type 4113202 AYT-40S Integrated type 1.4 45.2 40 4114001 AFT-40L Split type 4114002 AYT-50S Integrated type 2.1 70.7 50 4115001 AFT-50L Split type 4115002 AYT-65S Integrated type 4 119 65 4116501 AFT-65L Split type 4116502 AYT-80S Integrated type 5 181 80 4118001 AFT-80L Split type 4118002 AYT-100S Integrated type 8 283 100 41110001 AFT-100L Split type 41110002 AYT-125S Integrated type 13 442 125 41112501 AFT-125L Split type 41112502 AYT-150S Integrated type 19 636 150 41115001 AFT-150L Split type 41115002 AYT-200S Integrated type 30 1130 200 41120001 AFT-200L Split type 41120002 AYT-250S Integrated type 50 1767 250 41125001 AFT-250L Split type 41125002 AYT-300S Integrated type 80 2545 300 41130001 AFT-300L Split type 41130002 AYT-350S Integrated type 100 3464 350 41135001 AFT350L Split type 41135002 AYT-400S Integrated type 140 4524 400 41140001 AFT-400L Split type 41140002 APURE Intelligent Electromagnetic Flow Meter ■Electromagnetic Flow Meter Data sheet Shanghai GL Environmental Technology Co., Ltd
  • 10. APURE/17 APURE/18 Main technical parameters Measurement principle: ultrasonic time difference principle Measurement range :0~±10m/s Accuracy: flow rate is better than ±1% Display :2 x 10 Chinese backlit characters or 2 x 20 character LCD display, support three languages: Chinese, English and Italian Flow direction: Positive and negative two-way metering, and measure positive, negative, net cumulative flow / heat Data interface; RS485 serial turbulence, can be upgraded by PC computer, support MODBUS and other protocols Data recording: positive/negative/net accumulations at 512 days, 128 months and 10 days before automatic memory More than 30 times, power off time and flow rate before automatic memory and can be automatically or manually added Signal output: 1 channel 4-20mA current output 1 relay output 1 channel OCT pulse output (programmable between 6~100ms pulse width, default 200ms) Signal input: 3 channels of 4-20mA analog input, accuracy 01%, can collect pressure, liquid level, warm sleep and other signals 2-way three-wire PT100 platinum resistor for thermal measurement Power supply :AC85~264V or DC8~36V Media turbidity: ≤ 2000pm and small bubble content Applicable pipe: Carbon steel, stainless steel, copper, PVc, aluminum, fiberglass and other uniform and dense pipes, allowing lining Straight pipe section: The sensor installation point is best to meet: upstream 10, downstream 5, from the pump outlet 300 (D is the diameter) Measuring range Blind zone Pressure ±3mm/0.3%(standard conditions) 3 atmospheres below Environment temperature 1mm -20℃ ~+60℃ Built-in LCD display level or space distance Flow Sensor Standard small sensor L1 DN15~DN100mm, 0~90℃ Standard medium sensor L2 DN50~DN700mm, 0~90℃ Standard large sensor L3 DN300~DN60comm, 0~90℃ High temp small sensor GL1 DN15~DN100mm, 0~160℃ High temp medium sensor GL2 DN50~DN700mm, 0~160℃ High temp large sensor GL3 DN300~DN60comm, 0~160℃ Standard insertion sensor C1 ≥DN80mm Extended insertion sensor C2 ≥DN80mm Parallel insertion sensor C3 Customized Pipe section sensor DN15~DN1000mm APURE Split Type Open Channel Flow Meter External Clamp Ultrasonic Flow Meter Resolution Structure Display Accuracy Split type 0-10m 0.25-0.5m ※A set of 4-20mA analog output. ※RS485 communication(Modbus RTU)(optional). ※IP65 protection level. ※220VAC or 24VDC can be choose. ※Installation: Split type, meter wall-mounted. ※Sensor material: ABS plastic(PE/PTFE anti-corrosion material can be choose). ※Sensor comes with 10m cable(max: 50m) Model: MQ6000 Shanghai GL Environmental Technology Co., Ltd
  • 11. APURE/19 APURE/20 Figure 1:side mounted Main technical parameters 【Brand】 Apure 【Measuring range】 200-20000mm 【Medium temperature】 ≤ 100℃ (High temperature can be customized ) 【Medium pressure】 ≤ 1.6mpa 【Accuracy】 ±5mm 【Medium density】 ≥ 0.42g/cm³ 【Standard flange】 DN20/DN25 【Installation】 Side or top mounted 【Protection level】 IP65 【Measuring materials】 304,316 stainless steel, PP 【 Optional output 】 remote transmission, liquid level witch, hart protocol Product Selection Step 1: measuring range Step 2: Material and installation method Step 3: shape structure and pressure range Step 4: Remote transmission and level switch When there is special need, the user is ordering special Instructions. Such as density, connection specifications Model Interface diameter Measuring range D d H LZS-15 DN15 5-50L/H 45 20 202 10-100I/h 16-160I/h 25-250I/h 40-400I/h 50-500I/h 60-600I/h 100-1000I/h LZS-25 DN25 100-1000I/h 60 32 226 160-1600I/h 250-250I/h LZS-32 DN32 0.4-4m3/h 74 40 288 0.6-6m3/h LZS-50 DN50 0.4-4m3/h 98 63 341 0.6-6m3/h 1-10m3/h 1.6-16m3/h LZS-65 DN65 2.5-16m3/h 122 75 430 5-25m3/h 8-40m3/h 12-60m3/h LZS-100 DN100 14-90m3/h 215 110 555 20-120m3/h LZS-125 DN125 20-120m3/h 250 140 550 18-150m3/h 20-180m3/h LZS Float Flow Meter UHZ Magnetic Flap Level Gauge Shanghai GL Environmental Technology Co., Ltd
  • 12. APURE/21 APURE/22 Normal type Explosion-proof type Flat membrane type High temp type With LED display Clamp type Flange type Clamp high temp type Differential pressure type LED high temp type LCD high temp type Normal high temp type With LCD display Normal type Submerged Type Liquid Level Transmitter Submerged type PTFE material Anti-corrosion Waterproof junction box For High temperature(coll ection cylinder) Material: junction box low copper aluminum alloy; liquid level probe all stainless steel; cable materialpolyethylene wire, stainless steel tube, PTFE anti-corrosion cable. Media compatible: Various fluid media compatible with 316L stainless steel Parameters Measuring range: 0~10m, 0~100m(can be customized) Power supply: 12~24VDC Output signal: 4-20mA, 0-5V, 0-10V Medium temperature: -40~250℃ Protection:IP68 Pressure overload: 200%FS Mechanical vibration: 20g(20-5000HZ) Comprehensive accuracy: 0.5%(standard), 0.2 Stability: ±0.2%FS/Year Industrial Hammer Pressure Transmitter Compact Pressure Transmitter Shanghai GL Environmental Technology Co., Ltd
  • 13. APURE/23 APURE/24 The digital display instrument cooperates with various types of analog output sensors and transmitters to measure, transform, display and control physical quantities such as temperature, pressure, flow, liquid level, composition and displacement. Overview The integrated temperature transmitter with thermal resistance and thermocouple is a site-mounted temperature transmitter unit. It consists of a thermal resistance thermocouple and a temperature transmitter module. It is a two-wire system with a nonlinear correction circuit that directly measures liquid, gaseous media and various substances in the industrial process from -150 ° C to 400 ° C. Temperature signal is converted into a 4-20mADC current output signal linear with the temperature signal. Display, adjusted recording instrument or computer for distributed control. Widely used in petroleum machinery, chemical machinery, compressors, electric power, boilers, natural gas and other automated temperature measurement and control systems. Performance parameter Range: -150 ° C ~ 400 ° C Output signal: 4 ~ 20mA Supply voltage: 24VDC Long-term stability: ≤ 0.2% FS / year Response time: ≤ 20ms Insertion depth: ≥ 10cm (special requirements can be customized) Insert diameter: Ф7 (special requirements can be customized) Nominal pressure: static external pressure on the protective tube at operating temperature Explosion-proof grade: Exia II CT6 (can be customized) Power consumption: < 0.5W 1: Ambient temperature: -40 ° C ~ 80 ° C Working environment 2: relative humidity: 0 ~ 95% RH 3: Mechanical vibration: f < 55HZ, amplitude < 0.15mm Characteristics ◆Compatible input 2 1 kinds of signals: thermal resistance, thermocouple, standard current, standard voltage, millivolt, remote pressure gauge resistance, etc. Sensor zero display value, sensor full point display value, decimal point position can be set freely; ◆Multi-level digital filtering selection, effectively filtering out interference, effectively eliminating non-critical jump words, realizing alarm, controlling output, transmitting output 0 ~ 1 0 mA、4 ~ 2 0 mA 、 0 ~ 2 0 mA 、 0 ~ 5 V、1 ~ 5 V Model Content Code Description XM-601 Digital display Dimensions A 160*80*115,80*160*115 B 96*96*60 C 96*48*60,45*96*60 D 72*72*60 E 48*48*115 Panel form H Horizontal S Vertical F Square Alarm T 2-point alarming Transmitting Output A0 No transmitting output A1 Current output (isolated from external supply) A2 Voltage output (isolated from external supply) External transmitter power supply (external power supply input is not isolated, there is isolation requirement, please specify when ordering B0 No external power supply B1 External power supply 24VDC B2 External power supply 12VDC Meter communication S0 No communication S1 RS232 communication S2 RS485 communication Meter power supply VO 85VCA-265VAC Intelligent control secondary display instrument Integrated Temperature sensor / transmitter Shanghai GL Environmental Technology Co., Ltd