Distortion artifacts in MRI and
their correction
Vibha Chaswal, Ph.D.
MRI Gradient fields
• The gradient fields are superimposed over the static
magnetic field. For transverse images:
• Z gradient–selects the slice, and is applied whilst the 90° RF
pulse is on
• X gradient–creates a frequency change in emitted signal
producing lines (or columns) of data, and is applied when
the echo RF is being measured
• Y gradient–produces a change in the phase of rotation in
different voxels, and is applied between the 90° and 180°
RF pulses at a different strength during each cycle. The
number of phase steps (e.g. 256) affects scan time.
• “Frequency encoding” and “Phase encoding” directions (X
and Y) are interchangeable.
Gradient spatial encoding
Pulse sequencing diagram
MRI distortion Artifacts
A whole range……..
Equipment-related
• Non-uniformity of signal
• Wraparound (phase wrap,
aliasing)
• Peripheral signal artifacts
• Slice overlap (cross-excitation)
• Noise and RF interference
• Truncation (ringing, Gibbs
artifact)
• Phase smearing

Patient-related
• Signal void due to freezing
• Partial volume averaging
• Chemical shift
• Motion artifacts (blurring,
ghosting and pulsatility)
• Flow artifacts
• Susceptibility and image
distortion
• Magic angle
B0 inhomogeneity and susceptibility
• Extremely homogeneous static magnetic field is required
around isocenter of the magnet for MRI.
• B0 homogeneity of the empty magnet:
typical value of 1ppm (0.001%) in a 50 cm diameter
spherical volume.
• susceptibility = local changes in Bo due to a substance’s
own magnetic properties
• Artifacts occur at interfaces between substances of different
susceptibilities, e.g. air/bone/metal/ calcification/hemorrhage;
implants and previous surgery, tooth fillings, mascara, ear-rings,
hair-slides
– worse at higher field strengths
B0 inhomogeneity
• Decreased frequency-selective fat-saturation efficacy of pulses
• Effect more pronounced the further the image slice is positioned from
the isocenter of the magnet

(a) off-center acquisition with insufficient fat
suppression (arrows)

(b) Identical acquisition at isocenter

Reduction:
improve field shimming
Optimal positioning of the patient in z-direction
Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65, 2008
B0 inhomogeneity
•

Relatively small frequency shifts lead to substantial reduction of the
transversal steady-state magnetization and, thus, to band-shaped signal losses
in the image

Reduction:
Move the artifact out of the region of interest (apply frequency offset to nominal
resonant frequency)
Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65, 2008
Susceptibility: Metal artifacts
•

An important manifestation of a susceptibility-related artifact is the signal
loss in gradient-echo images around metallic implants or surgical clips

Right hip prosthesis is seen as region of signal drop
out (closed arrows)
Figure from: Mechlin et al, AJR 143, December 1984

STIR axial MRI image of a patient, arrows B point to
artifacts due to right hip prosthesis.
Figure from: Dr. Geller et al, Malignant Peripheral Nerve
Sheath Tumors (MPNST), ESUN, june 2006, 3(3)
http://sarcomahelp.org/learning_center/mpnst.html

Reduction:
Decrease Echo time, TE, of pulse sequence
Increase receiver band-width
Susceptibility: Geometric distortion
artifact
•
•

Usually in echo-planar imaging (EPI)
At interfaces between soft-tissue and bone/air e.g., in the
base of the skull or in the head-and-neck region

Reduction:
a) decrease the echo-spacing of the
read out train e.g., by increasing
the receiver bandwidth or by
parallel imaging techniques
b) Use fast-spin echo technique

(b) EPI acquisition without parallel imaging, severe distortion
artifacts are visible (arrows) and (c) EPI acquisition with
acceleration factor 2, distortion artifacts are still present but
considerably reduced (arrows).

Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65,
2008
B1 inhomogeneity
• Spatial inhomogeneity of the B1 field results in
flip-angle deviations depending on the spatial
position
• Reduced signal intensity in these areas or to
altered contrast particularly in FLASH sequences
whose T1-weighting depends on the flip angle
Dielectric resonance effects/ RF
interference/ standing wave effect
•

Constructive or destructive interferences of the transmitted RF field may be observed
resulting in either regional (e.g., central) brightening or regional signal loss,
respectively.

Signal loss due to wavelength effect
Reduction:
a) Effect can be mitigated by positioning a dielectric cushion close to the ROI, or
b) Manually modify the RF-transmitter amplitude in order to reduce B1-inhomogeneityinduced signal loss
Chemical Shift Artifact
• For MRI in vivo, two important groups of
molecules : protons in water and protons in
fat tissue.
• Larmor frequencies difference of these
protons ~ 3.5 ppm.
• chemical shift leads to a slight geometric shift
of the relative position of water and fat
protons in readout direction.
Chemical Shift Artifact Reduction

image-compromising superposition of
fat tissue onto other tissues
Reduction:
a)manually increase the RF-transmitter amplitude and apply image post-processing
filters to obtain more uniform image intensities.
b)Increase (~double) receiver bandwidth
c)Based on anatomy switch readout and phase encoding directions
Blood Flow related artifact
• Magnetohydrodynamic effect: The flow of blood ions
perpendicular to the strong static magnetic field, B0, gives rise to
induced voltages and currents;

“(a) Signal loss in the pulmonary vessels in non-gated 3-T single-shot fast-spin-echo
(FSE) acquisition possibly related to the magnetohydrodynamic effect. (b) Non-gated
1.5-T single-shot FSE acquisition demonstrating visual signal in pulmonary vessels. (c)
ECG-gated 3-T single-shot FSE acquisition of the same volunteer as in (a) with data
acquisition during diastole; the vessel signal is restored.” …….Deitrich et al, EJR
65(2008)
SNR related artifacts
• Artifacts to Noise ratio (ANR)
• Artifacts are masked by increased statistical
noise.
• Gibbs Ringing: caused by data clipping at the
edges of K-space (raw data readings are still over
noise level at the borders of acquired K-space)
Gibbs artifact

Reduction:
Increase spatial resolution or,
Apply reconstruction filters (Hanning filter)
Resources
Many more Artifacts but time restriction
⇒Resources for MR Artifacts ad their reduction
⇒MR-TIP: online database (open forum) from University
of Illinois, Excellent resource
⇒ Review article by Dietrich et.al,”Artifacts in 3-T MRI:
Physical Background and reduction strategies” EJR, Vol
65, 2008
Thank You

Mais conteúdo relacionado

Mais procurados

Artefacts mr & ct
Artefacts   mr & ctArtefacts   mr & ct
Artefacts mr & ctRakesh Ca
 
Gradient echo pulse sequence swastik
Gradient echo pulse sequence  swastikGradient echo pulse sequence  swastik
Gradient echo pulse sequence swastikBipul Poudel
 
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha Avinesh Shrestha
 
Perfusion MRI (DSC and DCE perfusion techniques) for radiology residents
Perfusion MRI (DSC and DCE perfusion techniques) for radiology residentsPerfusion MRI (DSC and DCE perfusion techniques) for radiology residents
Perfusion MRI (DSC and DCE perfusion techniques) for radiology residentsRiham Dessouky
 
Magnetic Resonance Diffusion
Magnetic Resonance DiffusionMagnetic Resonance Diffusion
Magnetic Resonance DiffusionRahman Ud Din
 
MRI Abdomen ( Dynamic Study - Tri-phasic of Liver )
MRI Abdomen ( Dynamic Study - Tri-phasic of Liver )MRI Abdomen ( Dynamic Study - Tri-phasic of Liver )
MRI Abdomen ( Dynamic Study - Tri-phasic of Liver )Abd Elrhman ElTurkish
 
Gradient echo pulse sequence and its application
Gradient echo pulse sequence and its applicationGradient echo pulse sequence and its application
Gradient echo pulse sequence and its applicationJayanti Gyawali
 
Presentation1, mr pulse sequences.
Presentation1, mr pulse sequences.Presentation1, mr pulse sequences.
Presentation1, mr pulse sequences.Abdellah Nazeer
 
FACTORS AFFECTING THE SIGNAL-TO-NOISE RATIO.pptx
FACTORS AFFECTING THE SIGNAL-TO-NOISE RATIO.pptxFACTORS AFFECTING THE SIGNAL-TO-NOISE RATIO.pptx
FACTORS AFFECTING THE SIGNAL-TO-NOISE RATIO.pptxRohit Bansal
 
CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis Upakar Paudel
 
Basics of mri physics Dr. Muhammad Bin Zulfiqar
Basics of mri physics Dr. Muhammad Bin ZulfiqarBasics of mri physics Dr. Muhammad Bin Zulfiqar
Basics of mri physics Dr. Muhammad Bin ZulfiqarDr. Muhammad Bin Zulfiqar
 

Mais procurados (20)

Artefacts mr & ct
Artefacts   mr & ctArtefacts   mr & ct
Artefacts mr & ct
 
Gradient echo pulse sequence swastik
Gradient echo pulse sequence  swastikGradient echo pulse sequence  swastik
Gradient echo pulse sequence swastik
 
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha
 
Perfusion MRI (DSC and DCE perfusion techniques) for radiology residents
Perfusion MRI (DSC and DCE perfusion techniques) for radiology residentsPerfusion MRI (DSC and DCE perfusion techniques) for radiology residents
Perfusion MRI (DSC and DCE perfusion techniques) for radiology residents
 
Magnetic Resonance Diffusion
Magnetic Resonance DiffusionMagnetic Resonance Diffusion
Magnetic Resonance Diffusion
 
CT artifact
CT artifact CT artifact
CT artifact
 
MRI physics
MRI physicsMRI physics
MRI physics
 
Basic of mri
Basic of mriBasic of mri
Basic of mri
 
MRI Abdomen ( Dynamic Study - Tri-phasic of Liver )
MRI Abdomen ( Dynamic Study - Tri-phasic of Liver )MRI Abdomen ( Dynamic Study - Tri-phasic of Liver )
MRI Abdomen ( Dynamic Study - Tri-phasic of Liver )
 
Gradient echo pulse sequence and its application
Gradient echo pulse sequence and its applicationGradient echo pulse sequence and its application
Gradient echo pulse sequence and its application
 
Presentation1, mr pulse sequences.
Presentation1, mr pulse sequences.Presentation1, mr pulse sequences.
Presentation1, mr pulse sequences.
 
FACTORS AFFECTING THE SIGNAL-TO-NOISE RATIO.pptx
FACTORS AFFECTING THE SIGNAL-TO-NOISE RATIO.pptxFACTORS AFFECTING THE SIGNAL-TO-NOISE RATIO.pptx
FACTORS AFFECTING THE SIGNAL-TO-NOISE RATIO.pptx
 
CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION
 
CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis CT Procedure OF Abdomen & Pelvis
CT Procedure OF Abdomen & Pelvis
 
Doppler physics by Dr Nani
Doppler physics by Dr NaniDoppler physics by Dr Nani
Doppler physics by Dr Nani
 
Artifacts
ArtifactsArtifacts
Artifacts
 
MRI Artefacts
MRI ArtefactsMRI Artefacts
MRI Artefacts
 
Basics of mri physics Dr. Muhammad Bin Zulfiqar
Basics of mri physics Dr. Muhammad Bin ZulfiqarBasics of mri physics Dr. Muhammad Bin Zulfiqar
Basics of mri physics Dr. Muhammad Bin Zulfiqar
 
BASIC MRI SEQUENCES
BASIC MRI SEQUENCESBASIC MRI SEQUENCES
BASIC MRI SEQUENCES
 
CT Artifacts
CT ArtifactsCT Artifacts
CT Artifacts
 

Destaque (20)

Cardiac MRI
Cardiac MRICardiac MRI
Cardiac MRI
 
Gre&SE
Gre&SEGre&SE
Gre&SE
 
Cardiac mri
Cardiac mriCardiac mri
Cardiac mri
 
Mr spectroscopy
Mr spectroscopyMr spectroscopy
Mr spectroscopy
 
Factors controlling the beam of x-ray
Factors controlling the beam of x-rayFactors controlling the beam of x-ray
Factors controlling the beam of x-ray
 
Special Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORTSpecial Procedures: TBI, TSET and IORT
Special Procedures: TBI, TSET and IORT
 
Cardiac MRI principle
Cardiac MRI principleCardiac MRI principle
Cardiac MRI principle
 
What is spatial Resolution
What is spatial ResolutionWhat is spatial Resolution
What is spatial Resolution
 
Radiographic factor
Radiographic factorRadiographic factor
Radiographic factor
 
Advances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and DeliveryAdvances in Brachytherapy Treatment Planning and Delivery
Advances in Brachytherapy Treatment Planning and Delivery
 
Mri physics-uk
Mri physics-ukMri physics-uk
Mri physics-uk
 
Mri artifacts gamal mahdaly
Mri artifacts gamal mahdalyMri artifacts gamal mahdaly
Mri artifacts gamal mahdaly
 
Contrast based artifacts
Contrast based artifactsContrast based artifacts
Contrast based artifacts
 
Fluroscopy
Fluroscopy Fluroscopy
Fluroscopy
 
Dark room equipments and entrance
Dark room equipments and entranceDark room equipments and entrance
Dark room equipments and entrance
 
Cardiac MRI
Cardiac MRICardiac MRI
Cardiac MRI
 
Mr fluoroscopy
Mr fluoroscopyMr fluoroscopy
Mr fluoroscopy
 
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
Nguyen khoi viet cardiac mri for the evaluation of ischemic heart disease jfi...
 
Pns New
Pns NewPns New
Pns New
 
076 cardiac magnetic resonance imaging
076 cardiac magnetic resonance imaging076 cardiac magnetic resonance imaging
076 cardiac magnetic resonance imaging
 

Semelhante a Distortion Artifacts in MRI and their correction

ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxKeerthan39
 
Recent Advances In MRI
Recent Advances In MRIRecent Advances In MRI
Recent Advances In MRIUpakar Paudel
 
Technical aspects of dental CBCT - Dr. Kavan Gandhi
Technical aspects of dental CBCT - Dr. Kavan GandhiTechnical aspects of dental CBCT - Dr. Kavan Gandhi
Technical aspects of dental CBCT - Dr. Kavan GandhiKAVAN GANDHI
 
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTSDIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTSMichaelAkortsu1
 
Nuclear Magnification Resonance Imaging
Nuclear Magnification Resonance ImagingNuclear Magnification Resonance Imaging
Nuclear Magnification Resonance ImagingDr ATHUL CHANDRA.M
 
Tissue harmonic imaging
Tissue harmonic imaging Tissue harmonic imaging
Tissue harmonic imaging MAMTA PANDA
 
Echo physics and instrumentation
Echo physics and instrumentationEcho physics and instrumentation
Echo physics and instrumentationguru here
 
Navneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to itNavneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to itNavneet Ranjan
 
CT Physics
CT PhysicsCT Physics
CT PhysicsRMLIMS
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxShubham661884
 
TGS Arcis- Canada Curvature Attributes and Applications Poster
TGS Arcis- Canada Curvature Attributes and Applications PosterTGS Arcis- Canada Curvature Attributes and Applications Poster
TGS Arcis- Canada Curvature Attributes and Applications PosterTGS
 

Semelhante a Distortion Artifacts in MRI and their correction (20)

mri artefacts.pptx
mri artefacts.pptxmri artefacts.pptx
mri artefacts.pptx
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
Mri artifacts
Mri artifactsMri artifacts
Mri artifacts
 
X-ray.ppt
X-ray.pptX-ray.ppt
X-ray.ppt
 
ppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptxppMRIARTIFACTS.pptx
ppMRIARTIFACTS.pptx
 
Swi seminar final
Swi seminar finalSwi seminar final
Swi seminar final
 
Recent Advances In MRI
Recent Advances In MRIRecent Advances In MRI
Recent Advances In MRI
 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
 
Technical aspects of dental CBCT - Dr. Kavan Gandhi
Technical aspects of dental CBCT - Dr. Kavan GandhiTechnical aspects of dental CBCT - Dr. Kavan Gandhi
Technical aspects of dental CBCT - Dr. Kavan Gandhi
 
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTSDIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
DIFFERENT TYPES OF RADIOGRAPHY CT_ARTIFACTS
 
Nuclear Magnification Resonance Imaging
Nuclear Magnification Resonance ImagingNuclear Magnification Resonance Imaging
Nuclear Magnification Resonance Imaging
 
Tissue harmonic imaging
Tissue harmonic imaging Tissue harmonic imaging
Tissue harmonic imaging
 
Echo physics and instrumentation
Echo physics and instrumentationEcho physics and instrumentation
Echo physics and instrumentation
 
Ultasonic testing
Ultasonic testingUltasonic testing
Ultasonic testing
 
Navneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to itNavneet.mr hardware & artifacts related to it
Navneet.mr hardware & artifacts related to it
 
CT Physics
CT PhysicsCT Physics
CT Physics
 
Ct image artifacts
Ct image artifactsCt image artifacts
Ct image artifacts
 
Physics basic
Physics basicPhysics basic
Physics basic
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
 
TGS Arcis- Canada Curvature Attributes and Applications Poster
TGS Arcis- Canada Curvature Attributes and Applications PosterTGS Arcis- Canada Curvature Attributes and Applications Poster
TGS Arcis- Canada Curvature Attributes and Applications Poster
 

Mais de Miami Cancer Institute

IGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing ProcedureIGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing ProcedureMiami Cancer Institute
 
Ion chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and EvaluationIon chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and EvaluationMiami Cancer Institute
 
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QAArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QAMiami Cancer Institute
 
Linear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QALinear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QAMiami Cancer Institute
 
TPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation OncologyTPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation OncologyMiami Cancer Institute
 
Mutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registrationMutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registrationMiami Cancer Institute
 
Imaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationImaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationMiami Cancer Institute
 
TG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit CalculationsTG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit CalculationsMiami Cancer Institute
 

Mais de Miami Cancer Institute (14)

IGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing ProcedureIGRT: MVCBCT Calibration and Acceptance testing Procedure
IGRT: MVCBCT Calibration and Acceptance testing Procedure
 
Ion chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and EvaluationIon chambers Acceptance Testing and Evaluation
Ion chambers Acceptance Testing and Evaluation
 
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QAArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
ArcCheck 3D-Diode array evaluation, commissioning, testing for VMAT QA
 
Linear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QALinear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QA
 
Shielding Design
Shielding DesignShielding Design
Shielding Design
 
Stereotactic Radiosurgery at UIHC
Stereotactic Radiosurgery at UIHCStereotactic Radiosurgery at UIHC
Stereotactic Radiosurgery at UIHC
 
TPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation OncologyTPS modeling and Beam commissioning at UIHC Radiation Oncology
TPS modeling and Beam commissioning at UIHC Radiation Oncology
 
Radiotherapy Treatment Simulation
Radiotherapy Treatment SimulationRadiotherapy Treatment Simulation
Radiotherapy Treatment Simulation
 
Image Contrast, Noise, Resolution
Image Contrast, Noise, ResolutionImage Contrast, Noise, Resolution
Image Contrast, Noise, Resolution
 
Mutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registrationMutual Information Algorithm applied to rigid registration
Mutual Information Algorithm applied to rigid registration
 
Causes of Noise in PET imaging
Causes of Noise in PET imagingCauses of Noise in PET imaging
Causes of Noise in PET imaging
 
Imaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verificationImaging for Radiotherapy delivery and verification
Imaging for Radiotherapy delivery and verification
 
Clinical Reference Dosimetry
Clinical Reference DosimetryClinical Reference Dosimetry
Clinical Reference Dosimetry
 
TG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit CalculationsTG-51 and Clinical Monitor Unit Calculations
TG-51 and Clinical Monitor Unit Calculations
 

Último

Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaPooja Gupta
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...narwatsonia7
 
Call Girl Nagpur Sia 7001305949 Independent Escort Service Nagpur
Call Girl Nagpur Sia 7001305949 Independent Escort Service NagpurCall Girl Nagpur Sia 7001305949 Independent Escort Service Nagpur
Call Girl Nagpur Sia 7001305949 Independent Escort Service NagpurRiya Pathan
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...saminamagar
 
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiNehru place Escorts
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...narwatsonia7
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Modelssonalikaur4
 
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️saminamagar
 
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...narwatsonia7
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptxDr.Nusrat Tariq
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfMedicoseAcademics
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbaisonalikaur4
 

Último (20)

Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
 
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
Call Girls Frazer Town Just Call 7001305949 Top Class Call Girl Service Avail...
 
Call Girl Nagpur Sia 7001305949 Independent Escort Service Nagpur
Call Girl Nagpur Sia 7001305949 Independent Escort Service NagpurCall Girl Nagpur Sia 7001305949 Independent Escort Service Nagpur
Call Girl Nagpur Sia 7001305949 Independent Escort Service Nagpur
 
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...call girls in Connaught Place  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
call girls in Connaught Place DELHI 🔝 >༒9540349809 🔝 genuine Escort Service ...
 
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in green park  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in green park DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
 
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hebbal Just Call 7001305949 Top Class Call Girl Service Available
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service ChennaiCall Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
Call Girls Service Chennai Jiya 7001305949 Independent Escort Service Chennai
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
Housewife Call Girls Bangalore - Call 7001305949 Rs-3500 with A/C Room Cash o...
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
 
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️call girls in munirka  DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
call girls in munirka DELHI 🔝 >༒9540349809 🔝 genuine Escort Service 🔝✔️✔️
 
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
Housewife Call Girls Hsr Layout - Call 7001305949 Rs-3500 with A/C Room Cash ...
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptx
 
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original PhotosBook Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
Book Call Girls in Yelahanka - For 7001305949 Cheap & Best with original Photos
 
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service MumbaiVIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
VIP Call Girls Mumbai Arpita 9910780858 Independent Escort Service Mumbai
 

Distortion Artifacts in MRI and their correction

  • 1. Distortion artifacts in MRI and their correction Vibha Chaswal, Ph.D.
  • 2. MRI Gradient fields • The gradient fields are superimposed over the static magnetic field. For transverse images: • Z gradient–selects the slice, and is applied whilst the 90° RF pulse is on • X gradient–creates a frequency change in emitted signal producing lines (or columns) of data, and is applied when the echo RF is being measured • Y gradient–produces a change in the phase of rotation in different voxels, and is applied between the 90° and 180° RF pulses at a different strength during each cycle. The number of phase steps (e.g. 256) affects scan time. • “Frequency encoding” and “Phase encoding” directions (X and Y) are interchangeable.
  • 3. Gradient spatial encoding Pulse sequencing diagram
  • 4. MRI distortion Artifacts A whole range…….. Equipment-related • Non-uniformity of signal • Wraparound (phase wrap, aliasing) • Peripheral signal artifacts • Slice overlap (cross-excitation) • Noise and RF interference • Truncation (ringing, Gibbs artifact) • Phase smearing Patient-related • Signal void due to freezing • Partial volume averaging • Chemical shift • Motion artifacts (blurring, ghosting and pulsatility) • Flow artifacts • Susceptibility and image distortion • Magic angle
  • 5. B0 inhomogeneity and susceptibility • Extremely homogeneous static magnetic field is required around isocenter of the magnet for MRI. • B0 homogeneity of the empty magnet: typical value of 1ppm (0.001%) in a 50 cm diameter spherical volume. • susceptibility = local changes in Bo due to a substance’s own magnetic properties • Artifacts occur at interfaces between substances of different susceptibilities, e.g. air/bone/metal/ calcification/hemorrhage; implants and previous surgery, tooth fillings, mascara, ear-rings, hair-slides – worse at higher field strengths
  • 6. B0 inhomogeneity • Decreased frequency-selective fat-saturation efficacy of pulses • Effect more pronounced the further the image slice is positioned from the isocenter of the magnet (a) off-center acquisition with insufficient fat suppression (arrows) (b) Identical acquisition at isocenter Reduction: improve field shimming Optimal positioning of the patient in z-direction Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65, 2008
  • 7. B0 inhomogeneity • Relatively small frequency shifts lead to substantial reduction of the transversal steady-state magnetization and, thus, to band-shaped signal losses in the image Reduction: Move the artifact out of the region of interest (apply frequency offset to nominal resonant frequency) Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65, 2008
  • 8. Susceptibility: Metal artifacts • An important manifestation of a susceptibility-related artifact is the signal loss in gradient-echo images around metallic implants or surgical clips Right hip prosthesis is seen as region of signal drop out (closed arrows) Figure from: Mechlin et al, AJR 143, December 1984 STIR axial MRI image of a patient, arrows B point to artifacts due to right hip prosthesis. Figure from: Dr. Geller et al, Malignant Peripheral Nerve Sheath Tumors (MPNST), ESUN, june 2006, 3(3) http://sarcomahelp.org/learning_center/mpnst.html Reduction: Decrease Echo time, TE, of pulse sequence Increase receiver band-width
  • 9. Susceptibility: Geometric distortion artifact • • Usually in echo-planar imaging (EPI) At interfaces between soft-tissue and bone/air e.g., in the base of the skull or in the head-and-neck region Reduction: a) decrease the echo-spacing of the read out train e.g., by increasing the receiver bandwidth or by parallel imaging techniques b) Use fast-spin echo technique (b) EPI acquisition without parallel imaging, severe distortion artifacts are visible (arrows) and (c) EPI acquisition with acceleration factor 2, distortion artifacts are still present but considerably reduced (arrows). Dietrich et. al., Artifacts in 3-T MRI: Physical background and reduction strategies, EJR, 65, 2008
  • 10. B1 inhomogeneity • Spatial inhomogeneity of the B1 field results in flip-angle deviations depending on the spatial position • Reduced signal intensity in these areas or to altered contrast particularly in FLASH sequences whose T1-weighting depends on the flip angle
  • 11. Dielectric resonance effects/ RF interference/ standing wave effect • Constructive or destructive interferences of the transmitted RF field may be observed resulting in either regional (e.g., central) brightening or regional signal loss, respectively. Signal loss due to wavelength effect Reduction: a) Effect can be mitigated by positioning a dielectric cushion close to the ROI, or b) Manually modify the RF-transmitter amplitude in order to reduce B1-inhomogeneityinduced signal loss
  • 12. Chemical Shift Artifact • For MRI in vivo, two important groups of molecules : protons in water and protons in fat tissue. • Larmor frequencies difference of these protons ~ 3.5 ppm. • chemical shift leads to a slight geometric shift of the relative position of water and fat protons in readout direction.
  • 13. Chemical Shift Artifact Reduction image-compromising superposition of fat tissue onto other tissues Reduction: a)manually increase the RF-transmitter amplitude and apply image post-processing filters to obtain more uniform image intensities. b)Increase (~double) receiver bandwidth c)Based on anatomy switch readout and phase encoding directions
  • 14. Blood Flow related artifact • Magnetohydrodynamic effect: The flow of blood ions perpendicular to the strong static magnetic field, B0, gives rise to induced voltages and currents; “(a) Signal loss in the pulmonary vessels in non-gated 3-T single-shot fast-spin-echo (FSE) acquisition possibly related to the magnetohydrodynamic effect. (b) Non-gated 1.5-T single-shot FSE acquisition demonstrating visual signal in pulmonary vessels. (c) ECG-gated 3-T single-shot FSE acquisition of the same volunteer as in (a) with data acquisition during diastole; the vessel signal is restored.” …….Deitrich et al, EJR 65(2008)
  • 15. SNR related artifacts • Artifacts to Noise ratio (ANR) • Artifacts are masked by increased statistical noise. • Gibbs Ringing: caused by data clipping at the edges of K-space (raw data readings are still over noise level at the borders of acquired K-space)
  • 16. Gibbs artifact Reduction: Increase spatial resolution or, Apply reconstruction filters (Hanning filter)
  • 17. Resources Many more Artifacts but time restriction ⇒Resources for MR Artifacts ad their reduction ⇒MR-TIP: online database (open forum) from University of Illinois, Excellent resource ⇒ Review article by Dietrich et.al,”Artifacts in 3-T MRI: Physical Background and reduction strategies” EJR, Vol 65, 2008

Notas do Editor

  1. Spin echo
  2. Happens If the wavelength of the RF field is in the same order of magnitude as the geometric dimensions of the imaged object
  3. This is because the spatial position in MRI is encoded using varying Larmor frequencies induced by the applied magnetic gradient pulses.
  4. 1.5 T machines had high ANR and Gibbs was suppressed because of high noise. More in 3T
  5. Filter smooths the image and reduces the signal at the edges of K-space