SlideShare uma empresa Scribd logo
1 de 77
Baixar para ler offline
植入性心臟電子儀器地基本原
理與設定
林口長庚 巫龍昇醫師
Voltage, Current, and Impedance
Recap
• Voltage: The force moving the current (V)
– In pacemakers it is a function of the battery
chemistry
• Current: The actual continuing volume of flow of
electricity (I)
– This flow of electrons causes the myocardial cells to
depolarize (to “beat”)
• Impedance: The sum of all resistance to current
flow (R or W or sometimes Z)
– Impedance is a function of the characteristics of the
conductor (wire), the electrode (tip), and the
myocardium
2
Voltage and Current Flow
Electrical Analogies
Spigot (voltage) turned up, lots of
water flows (high current drain)
Spigot (voltage) turned low, little flow
(low current drain)
Water pressure in system is
analogous to voltage –
providing the force to move
the current
Resistance and Current Flow
Electrical Analogies
• Normal resistance – friction caused by the hose and nozzle
More water discharges, but is all of it going to the
nozzle?
• High resistance – a knot results in low total current flow
• Low resistance – leaks in the hose reduce the resistance
Ohm’s Law
• Describes the relationship
between voltage, current,
and resistance
• V = I X R
• I = V / R
• R = V / I
V
I R
V
R
I
V
R
I
R
V
I =
=
=
X
Other terms
• Cathode陰極:
– For example, the electrode on the tip of
a pacing lead
• Anode陽極:
– Examples:
• The “ring” electrode on a
bipolar lead
• The IPG case on a unipolar
system
6
Anode
Cathode
PACING MODES
Pacing Modes
I II III IV
Chamber(s)
Paced
Chamber(s)
Sensed
Response
to Sensing
Rate
Modulation
O = None
A = Atrium
V = Ventricle
D = Dual (A + V)
S = Single (A or V)
O = None
A = Atrium
V = Ventricle
D = Dual (A + V)
S = Single (A or V)
O = None
T = Triggered
I = Inhibited
D = Dual (T + I)
O = None
R = Rate
modulation
Pacemaker modes most often seen:
DDDR/VDD
VVIR
AAIR->DDDR (MVP)
What mode would you use for 3rd Degree Block?
What mode would you use for SSS?
What mode would you use for permanent AF?
Knowledge Checkpoint
What is the rhythm? What type of device does this patient
need?
VVI Mode
I II III
Chamber(s)
Paced
Chamber(s)
Sensed
Response to
Sensing
O = None
A = Atrium
V = Ventricle
D = Dual (A + V)
S = Single (A or V)
O = None
A = Atrium
V = Ventricle
D = Dual (A + V)
S = Single (A or V)
O = None
T = Triggered
I = Inhibited
D = Dual (I + T)
VVI Example
• Chamber paced: Ventricle
• Chamber sensed: Ventricle
• Response to sensing: Inhibition
–VVI 60 = Lower Rate timer of 1000 ms
• Pacing every 1 second if not inhibited
V
P
V
P
V
P
Lower Rate Timer 1000 ms Lower Rate Timer 1000 ms Lower Rate Timer ….
VVI Example (60 bpm)
V
P
V
S
V
P
V
P
Lower rate timer 1000 ms
x
Lower rate timer 1000 ms
•Paces and Senses in the ventricle
•Timed from each QRS
•If it sees a sensed event, it will inhibit the next pace
VOO Mode – Asynchronous Pacing
Chamber paced: Ventricle
Chamber sensed: None
Response to sensing: None
The intrinsic ventricular event
cannot be sensed, and thus, does
not interrupt the pacing interval.
1000 ms1000 ms 1000 ms
V
P
V
P
V
P
V
P
VOO results in fixed-rate pacing in
the ventricle.
Knowledge Checkpoint
What is the rhythm? What type of device does this patient
need?
AAI Mode
I II III
Chamber(s)
Paced
Chamber(s)
Sensed
Response to
Sensing
O = None
A = Atrium
V = Ventricle
D = Dual (A + V)
S = Single (A or V)
O = None
A = Atrium
V = Ventricle
D = Dual (A + V)
S = Single (A or V)
O = None
T = Triggered
I = Inhibited
D = Dual (I + T)
AAI
• Paces in the atrium
• Timed from last P wave
Pacing Interval
A
P
A
P
Lower Rate Timer 1000 ms
Knowledge Checkpoint
What is the rhythm? What type of device does this patient
need?
DDD Mode
I II III
Chamber(s)
Paced
Chamber(s)
Sensed
Response to
Sensing
O = None
A = Atrium
V = Ventricle
D = Dual (A + V)
S = Single (A or V)
O = None
A = Atrium
V = Ventricle
D = Dual (A + V)
S = Single (A or V)
O = None
T = Triggered
I = Inhibited
D = Dual (I + T)
DDD
A
S V
P
A
S V
P
A
P V
S
A
P V
S
• Senses and paces in both
chambers when needed
Knowledge Checkpoint
A
B
C
D
Label each EKG with the faces of pacing (AS-VS, AP-VS, AP-VP,
AS-VP).
Magnet Application for devices
• Magnet application temporarily changes pacing mode to DOO/VOO
Magnet Applied
1 2 3
1
0
0
1
0
0
1
0
0
The 4th letter: “R”
• Question: what does your heart rate do when you
exercise?
• “R” means “Rate response”
• Pacemaker will increase pacing rate in response to
exercise – if patient does not increase his own rate
General guidelines for programming
common pacemaker parameter
parameter situation Chronic setting comments
Lower rate limit General :
minimal pacing desired;
50-70bpm
40-60bpm Use rate hysteresis
Upper rate limit General
Child/athletes
CAD/angina
85% maximal
predicted HR
0.85*(220-age)
(220-age) bpm
110-120bpm
Based on average
levels of activity
May require
programing short
refractory periods
Approximates peak
HR on maximal beta
blocker
Rate hysteresis
General guidelines for programming
common pacemaker parameter
parameter situation Chronic setting comments
Pacing output Fixed voltage
Fixed pulse width
3-4 X pulse width
threshold
2-3X voltage
threshold
Minimizing voltage
output more
efficient
Use autothreshold
function
sensitivity Atrium
ventricle
25-50% threshold
25-50 % threshold
Need <1mV setting
for mode switching
Evaluate
oversensing in
unipolar system
General guidelines for programming
common pacemaker parameter
parameter situation Chronic setting comments
AV delay AV block
Intrinsic conduction
(no HF)
Intrinsic conduction
(CHF)
150-180msec paced AV delay,
sensed AV delay 25-50msec
<paced AV delay
Up to 220msec
Often set even longer AV
delay
Turn on rate adaptive
AV delay in active
patients
Longer AV delay may
compromise
hemodynamics, use AV
hysteresis to promote
intrinsic conduction
Paced induced
dyssynchrony of very
long AV delay
Optimize by doppler,
AV hysteresis
Programmer
Programmer - PAV, SAV
Programmer - Mode Switch
ICD Function
What is the function of an ICD?
• Sense cardiac rhythms
• Detect arrhythmias
• Deliver therapy
• Pace when necessary
Typical Transvenous Lead
Transvenous Leads
Placement
Single Coil Dual Coil
ICD Therapies
• Tachyarrhythmia Therapy
–Cardioversion (CV)Sychronized ro R
–Anti-Tachycardia Pacing (ATP)
–Defibrillation Shock
Low Power
oBradyarrhythmia Therapy
–Pacing Modes
Low Power
High Power
High Power
Bradyarrhythmia Therapy
• Most ICDs offer:
–Single Chamber Pacing
• AAI(R), VVI(R) and VOO
–Dual Chamber Pacing
• DDD(R), DDI(R), DOO and ODO
• Mode Switch
–Separate post-shock pacing programming
• Ensures capture
ICD common setting
• VT zone 160-200bpm
• VF zone >200bpm
• The rate usually set 10bpm slower than the
documented ventricular arrhythmia rate to
reliable detection
• Ventricular sensitivity must allow for the
detection of very low amplitude ventricular
fibrillation signals and typically set to 0.3 mV
ICD basic parameters
parameter function programming comment
VF zone Detection rate for
fastest VF zone: rate of
VF zone must be on
VF zone;
Hemodynamically
unstable VT
Must be on at all
times, only shock
Rx available
VT zone Detection rate for
slower VTs in
multizone
programming
On or off;
Usually 10bpm <
spontaneous VT
ATP and
cardioversion Rx
available
Initial detection
No. of intervals
VF zone : 75% intervals >
VF rate
VT zone: number of
consecutive intervals >
rate limit
VF typically 12/16,
or 18/24 beats >
VF rate
VT typically 8-20
beats >VT rate
ICD basic parameters
parameter function programming comment
Ventricular
sensitivity
Usually 0.18-0.3mV High sensitivity
needed to detect
small EGM in VF
ATP Painless pace
termination of VT
On or off Often initial RX for
VTs <200bpm
Type of sequence
(burst or ramp)
Burst: all paced
intervals the same
Ramp: decrements
between paced
intervals in
sequence
Burst or ramp Ramp therapy
considered more
aggressive but
equal efficacy for
spontaneous
ICD basic parameters
parameter function programming comment
Cardioversion/defib
rillation
Shock therapy Always on in all
zones
Only RX in VF, or
single zone; initial
RX or follows ATP in
VT zones
Energy Magnitude of shock
in joules
VF: DFT =5-10 J
VT: at or above
smallest successful
energy
VT often
terminated with 2-
10 J
41
• 13% - 38% of ICD patients experience significant levels of
psychological stress related to the fear of receiving a shock.6
• Patients receiving shocks reported feeling less healthy, had
lower levels of psychological well-being, and reduced
physical and emotional function.7
• Shock therapy is inversely correlated with quality-of-life.8,9
• Patients receiving ATP first showed a significant increase in
QOL over those receiving shock first.10
Shock Therapy is Related to a Decrease in QOL
6. Hammil SC, et al. J Cardiovasc Electrophysiol, 2000.
7. Namerow PB, et al. PACE, 1999.
8. Carroll DL, et al. Heart Lung, 2005.
9. Irvine J, et al. Am Heart J, 2002.
10. Wathen MS, et al. Circulation, 2004.
Shock Reduction
• Anti-tachycardia pacing
• 延長VT/VF偵測
• 藥物
Anti-Tachycardia Pacing
Re-entry initiated ATP delivered at a rate faster than
tachyarrhythmia.
Wavefronts collide.
Subsequent Pulse: Wavefronts
collide closer to re-entry circuit
Subsequent Pulses: Wavefronts
collide even closer to re-entry
circuit
Arrhythmia
terminated
Anti-tachycardia pacing (ATP)
ATP的風險
Anti-tachycardia pacing(ATP)
Burst v.s Ramp
Programmed Values:
Number of S1 Pulses = 4
Number of Sequences = 4
R- S1% = 91% (less agrassive)
Decrement* = 10 ms
* Decrement between sequences
Programmed Values:
Number of S1 Pulses = 4
Number of Sequences = 4
R-S1% = 91% (less aggrassive)
Decrement* = 10 ms
* Decrement between pulses
* Adds a pulse per sequence
PITAGORA ICD study: Bust is significantly more efficacious than ramp in terminating FVT episodes
Sensing
• Sensing is:
–The process of identifying cardiac depolarizations from an
intracardiac electrogram
Measured Peak-to-Peak
>5 mV for optimal sensing
Detection
• Measured in:
–Beat-to-beat intervals (milliseconds), or
–Beats-per-minute (BPM)
Detection Rate
•Classifies rhythm by detection zone:
–VT = Ventricular Tachycardia +/- FVT
–VF = Ventricular Fibrillation
•Programmable in ranges of rates
Example:VT = 162 bpm – 188 bpm
VF = 188 bpm and faster
Detection
Zone Classifications
ICDexample
Detection
• Measured in:
– Number of intervals to detect (NID), or
– Length of time to detect
•Programmable by:
– Beat or interval counters
• Consecutive
ex: 16 beats within the detect zone
• Probabilistic (percentage or fraction)
ex: 12 out of 16 beats within the detect zone
– Time in seconds
Detect Duration
Detection
Used for detection of VT
Consecutive Counter
Detection
NID = 12/16
Probabilistic Counter
• Can you identify the detect zones?
• Name the rate & duration for each
Detection
Detect Zones
Cardioversion
• Delivers shock on an R-wave
• Aborts if synchronization cannot be obtained due to
arrhythmia termination
Cardioversion
Defibrillation
Programming
*Medtronic Programming Screen
Major Challenges Faced
• Inappropriate therapies
• Unnecessary shocks
• Repetitive shocks
58
59
• 13% - 38% of ICD patients experience significant levels of
psychological stress related to the fear of receiving a shock.6
• Patients receiving shocks reported feeling less healthy, had
lower levels of psychological well-being, and reduced
physical and emotional function.7
• Shock therapy is inversely correlated with quality-of-life.8,9
• Patients receiving ATP first showed a significant increase in
QOL over those receiving shock first.10
Shock Therapy is Related to a Decrease in QOL
6. Hammil SC, et al. J Cardiovasc Electrophysiol, 2000.
7. Namerow PB, et al. PACE, 1999.
8. Carroll DL, et al. Heart Lung, 2005.
9. Irvine J, et al. Am Heart J, 2002.
10. Wathen MS, et al. Circulation, 2004.
辛苦了
謝謝聆聽
Differentiation of SVT from VT
parameter What It does Potentially
useful For
Potential problem
stability Suppressed therapy for
tachyarrhythmia with
variable ventricular rate
Atrial
fibrillation
Underdetection of VT
with irregular rate;
failure to suppress
therapy for SVTs with
regular ventricular
response
onset Suppressed therapy for
tachyarrhythmia that
slowly accelerate
Sinus
tachycardia
Underdetection of
gradually accelerating VT
or VT onset during sinus
tachycardia; failure to
suppress therapy for
sudden onset SVT
Ventricular
electrogram
width
Suppress therapy for
tachyarrhythmia with
narrow ventricular EGM
correlated to narrow QRS
complex
Differentiation
of narrow
complex SVT
from VT
Limited specificity with
BBB; may prevent
therapy for narrow
complex VT
Differentiation of SVT from VT
parameter What It does Potentially
useful For
Potential problem
Ventricular
electrogram
morphology
Suppressed therapy for
tachyarrhythmia with
ventricular EGM
morphology similar to
that in sinus rhythm
D/D SVT from
VT
Limited specificity with
BBB
Atrial to
ventricular
ratio
Compare atrial to
ventricular rate
Atrial
fibrillation
Atrial undersensing can
result in false diagnosis
of VT
SVT Discriminators
• Distinguishes SVTs by analyzing P and R-wave:
–Rate
–Regularity
–AV Association
PR Logic™
– Pattern
SVT Discriminators
• Based on the premise that AF conducts
irregularly to the ventricles
(and VT is a stable, regular rhythm)
• Discriminates regular from irregular
intervals within a detect zone
Stability
SVT Discriminators
UnstableVaries >50 ms from previous 3
Stability = 50 ms
Stability
* in Medtronic devices
SVT Discriminators
• Based on the premise that most VTs are
characterized by a sudden onset
•Evaluates the acceleration of the
ventricular rate
•Discriminates between:
–Gradual rate increase
–Abrupt rate increase
•Determines VT present if rate
increase is abrupt
Onset
SVT Discriminators
• Onset Percentage = 81%
530ms X 81% =
430ms
430ms  460ms = Onset Not Met
Onset
* in Medtronic devices
SVT Discriminators
• Measures and stores the QRS characteristics of a
normal sinus beat
• Identifies SVT vs. VT based on the QRS changes
that occur in most VTs
SINUS RHYTHM VT
Waveform Morphology
SVT Discriminators
• Identifies start and end points of a sensed QRS
complex
• Uses 2 parameters to measures QRS:
–Slew Width
–Slew Threshold
EGM Width
SVT Discriminators
EGM Width
SVT Discriminators
• Records and stores a template of a normal QRS
wave
•Compares stored template with a QRS
occurring within the detection zone
•Withholds detection if 3 of last 8 QRS
complexes match the stored template
– Detects VT/FVT/VF if 6 out of 8 do not match
•Applies to initial detection only
Wavelet
SVT Discriminators
• Template
% difference compared against:
Match Threshold Value
Wavelet
Refractory and Blanking Periods
• Pacemaker sensing occurs when a signal
is large enough to cross the sensing
threshold
1.25 mV Sensitivity
Time
5.0 mV
2.5 mV
1.25 mV
Sensing does not tells us
anything about the origin or
morphology of the sensed
event, only its “size.”
In DDD & VDD modes the
pacemaker will “track” the atrium
AS
VP
Tracking = Pacing the ventricle
after an atrial intrinsic event
Maintains AV Synchrony
Want to limit how fast we pace
Upper Tracking Rate
DDDR 60 / 120 A-A = 500 ms
Upper Activity Rate Limit
Lower Rate Limit
Upper Sensor Rate
• Sensor rate drives the atrial rate up
• In rate responsive, dual chamber modes, the Upper Activity (Sensor) Rate
provides the limit for sensor-driven atrial pacing
PAV PAV
1000 ms
500 ms
Post Shock Pacing

Mais conteúdo relacionado

Mais procurados

Emergency Department ICD and pacemaker issues
Emergency Department ICD and pacemaker issuesEmergency Department ICD and pacemaker issues
Emergency Department ICD and pacemaker issuesSCGH ED CME
 
A D V A N C E D P A C I N G
A D V A N C E D  P A C I N GA D V A N C E D  P A C I N G
A D V A N C E D P A C I N GSherry Knowles
 
Cardiac pacemakers part ii
Cardiac pacemakers part iiCardiac pacemakers part ii
Cardiac pacemakers part iisalah_atta
 
Electrophysiologic basis part3
Electrophysiologic basis part3Electrophysiologic basis part3
Electrophysiologic basis part3salah_atta
 
Basic EP study (part1)
Basic  EP  study (part1)Basic  EP  study (part1)
Basic EP study (part1)SolidaSakhan
 
Anethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devicesAnethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devicesAmr Moustafa Kamel
 
Basic technical concepts in cardiac pacing
Basic technical concepts in cardiac pacingBasic technical concepts in cardiac pacing
Basic technical concepts in cardiac pacingSunil Reddy D
 
Cardiac pace makerspart 1
Cardiac pace makerspart 1Cardiac pace makerspart 1
Cardiac pace makerspart 1salah_atta
 
Evaluation and management of Pacemaker malfunction
Evaluation and management of Pacemaker malfunctionEvaluation and management of Pacemaker malfunction
Evaluation and management of Pacemaker malfunctionPRAVEEN GUPTA
 
ELectrophysiology basics part4
ELectrophysiology basics part4ELectrophysiology basics part4
ELectrophysiology basics part4salah_atta
 
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwarIImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwarPrithvi Puwar
 
Eps basics,part2(lecture)
Eps basics,part2(lecture)Eps basics,part2(lecture)
Eps basics,part2(lecture)salah_atta
 
Left ventricular pressure tracings
Left ventricular pressure tracingsLeft ventricular pressure tracings
Left ventricular pressure tracingsGOPAL GHOSH
 
hemodynamic in cath lab: aortic stenosis and hocm
hemodynamic in cath lab: aortic stenosis and hocmhemodynamic in cath lab: aortic stenosis and hocm
hemodynamic in cath lab: aortic stenosis and hocmrahul arora
 

Mais procurados (20)

Emergency Department ICD and pacemaker issues
Emergency Department ICD and pacemaker issuesEmergency Department ICD and pacemaker issues
Emergency Department ICD and pacemaker issues
 
A D V A N C E D P A C I N G
A D V A N C E D  P A C I N GA D V A N C E D  P A C I N G
A D V A N C E D P A C I N G
 
Cardiac pacemakers part ii
Cardiac pacemakers part iiCardiac pacemakers part ii
Cardiac pacemakers part ii
 
CATHETER ABLATION IN VT
CATHETER ABLATION IN VTCATHETER ABLATION IN VT
CATHETER ABLATION IN VT
 
ICD troubleshooting
ICD troubleshootingICD troubleshooting
ICD troubleshooting
 
Electrophysiologic basis part3
Electrophysiologic basis part3Electrophysiologic basis part3
Electrophysiologic basis part3
 
Basic EP study (part1)
Basic  EP  study (part1)Basic  EP  study (part1)
Basic EP study (part1)
 
Anethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devicesAnethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devices
 
Basic technical concepts in cardiac pacing
Basic technical concepts in cardiac pacingBasic technical concepts in cardiac pacing
Basic technical concepts in cardiac pacing
 
Coronary guidewires
Coronary guidewiresCoronary guidewires
Coronary guidewires
 
Cardiac pace makerspart 1
Cardiac pace makerspart 1Cardiac pace makerspart 1
Cardiac pace makerspart 1
 
Evaluation and management of Pacemaker malfunction
Evaluation and management of Pacemaker malfunctionEvaluation and management of Pacemaker malfunction
Evaluation and management of Pacemaker malfunction
 
ELectrophysiology basics part4
ELectrophysiology basics part4ELectrophysiology basics part4
ELectrophysiology basics part4
 
Cath hemodynamics vir
Cath hemodynamics virCath hemodynamics vir
Cath hemodynamics vir
 
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwarIImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
 
Eps basics,part2(lecture)
Eps basics,part2(lecture)Eps basics,part2(lecture)
Eps basics,part2(lecture)
 
Left ventricular pressure tracings
Left ventricular pressure tracingsLeft ventricular pressure tracings
Left ventricular pressure tracings
 
Temporary pacing
Temporary pacingTemporary pacing
Temporary pacing
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
hemodynamic in cath lab: aortic stenosis and hocm
hemodynamic in cath lab: aortic stenosis and hocmhemodynamic in cath lab: aortic stenosis and hocm
hemodynamic in cath lab: aortic stenosis and hocm
 

Destaque

心臟植入性電子儀器(CIED )護理照護指引- 病房及CCU Troubleshooting_20130907北區
心臟植入性電子儀器(CIED )護理照護指引- 病房及CCU Troubleshooting_20130907北區心臟植入性電子儀器(CIED )護理照護指引- 病房及CCU Troubleshooting_20130907北區
心臟植入性電子儀器(CIED )護理照護指引- 病房及CCU Troubleshooting_20130907北區Taiwan Heart Rhythm Society
 
心臟植入性電子儀器CIED)護理照護指引 - 手術前及手術後_20131019南區
心臟植入性電子儀器CIED)護理照護指引 - 手術前及手術後_20131019南區心臟植入性電子儀器CIED)護理照護指引 - 手術前及手術後_20131019南區
心臟植入性電子儀器CIED)護理照護指引 - 手術前及手術後_20131019南區Taiwan Heart Rhythm Society
 
心臟植入性電子儀器(CIED )護理照護指引 –手術前及手術後_20130914中區
心臟植入性電子儀器(CIED )護理照護指引 –手術前及手術後_20130914中區心臟植入性電子儀器(CIED )護理照護指引 –手術前及手術後_20130914中區
心臟植入性電子儀器(CIED )護理照護指引 –手術前及手術後_20130914中區Taiwan Heart Rhythm Society
 
心臟植入性電子儀器(CIED)護理照護指引-手術中護理照護_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-手術中護理照護_20131019南區心臟植入性電子儀器(CIED)護理照護指引-手術中護理照護_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-手術中護理照護_20131019南區Taiwan Heart Rhythm Society
 
心臟植入性電子儀器(CIED )之適應症 “Indication for CIED”_20130907北區
心臟植入性電子儀器(CIED )之適應症 “Indication for CIED”_20130907北區心臟植入性電子儀器(CIED )之適應症 “Indication for CIED”_20130907北區
心臟植入性電子儀器(CIED )之適應症 “Indication for CIED”_20130907北區Taiwan Heart Rhythm Society
 
心臟植入性電子儀器(CIED )護理照護指引 – 手術前及手術後_20130907北區
心臟植入性電子儀器(CIED )護理照護指引 – 手術前及手術後_20130907北區心臟植入性電子儀器(CIED )護理照護指引 – 手術前及手術後_20130907北區
心臟植入性電子儀器(CIED )護理照護指引 – 手術前及手術後_20130907北區Taiwan Heart Rhythm Society
 
心臟植入性電子儀器(CIED )護理照護指引- 手術中護理照護_20130907北區
心臟植入性電子儀器(CIED )護理照護指引- 手術中護理照護_20130907北區心臟植入性電子儀器(CIED )護理照護指引- 手術中護理照護_20130907北區
心臟植入性電子儀器(CIED )護理照護指引- 手術中護理照護_20130907北區Taiwan Heart Rhythm Society
 
Electrical testing of pacemaker
Electrical testing of pacemakerElectrical testing of pacemaker
Electrical testing of pacemakerRamachandra Barik
 

Destaque (11)

5. 心律調節器
5. 心律調節器5. 心律調節器
5. 心律調節器
 
心臟植入性電子儀器(CIED )護理照護指引- 病房及CCU Troubleshooting_20130907北區
心臟植入性電子儀器(CIED )護理照護指引- 病房及CCU Troubleshooting_20130907北區心臟植入性電子儀器(CIED )護理照護指引- 病房及CCU Troubleshooting_20130907北區
心臟植入性電子儀器(CIED )護理照護指引- 病房及CCU Troubleshooting_20130907北區
 
心臟植入性電子儀器CIED)護理照護指引 - 手術前及手術後_20131019南區
心臟植入性電子儀器CIED)護理照護指引 - 手術前及手術後_20131019南區心臟植入性電子儀器CIED)護理照護指引 - 手術前及手術後_20131019南區
心臟植入性電子儀器CIED)護理照護指引 - 手術前及手術後_20131019南區
 
心臟植入性電子儀器(CIED )護理照護指引 –手術前及手術後_20130914中區
心臟植入性電子儀器(CIED )護理照護指引 –手術前及手術後_20130914中區心臟植入性電子儀器(CIED )護理照護指引 –手術前及手術後_20130914中區
心臟植入性電子儀器(CIED )護理照護指引 –手術前及手術後_20130914中區
 
心臟植入性電子儀器(CIED)護理照護指引-手術中護理照護_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-手術中護理照護_20131019南區心臟植入性電子儀器(CIED)護理照護指引-手術中護理照護_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-手術中護理照護_20131019南區
 
心臟植入性電子儀器(CIED )之適應症 “Indication for CIED”_20130907北區
心臟植入性電子儀器(CIED )之適應症 “Indication for CIED”_20130907北區心臟植入性電子儀器(CIED )之適應症 “Indication for CIED”_20130907北區
心臟植入性電子儀器(CIED )之適應症 “Indication for CIED”_20130907北區
 
心臟植入性電子儀器(CIED )護理照護指引 – 手術前及手術後_20130907北區
心臟植入性電子儀器(CIED )護理照護指引 – 手術前及手術後_20130907北區心臟植入性電子儀器(CIED )護理照護指引 – 手術前及手術後_20130907北區
心臟植入性電子儀器(CIED )護理照護指引 – 手術前及手術後_20130907北區
 
3. 心臟整流去顫器
3. 心臟整流去顫器3. 心臟整流去顫器
3. 心臟整流去顫器
 
心臟植入性電子儀器(CIED )護理照護指引- 手術中護理照護_20130907北區
心臟植入性電子儀器(CIED )護理照護指引- 手術中護理照護_20130907北區心臟植入性電子儀器(CIED )護理照護指引- 手術中護理照護_20130907北區
心臟植入性電子儀器(CIED )護理照護指引- 手術中護理照護_20130907北區
 
1. 認識心律不整
1. 認識心律不整1. 認識心律不整
1. 認識心律不整
 
Electrical testing of pacemaker
Electrical testing of pacemakerElectrical testing of pacemaker
Electrical testing of pacemaker
 

Semelhante a 植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區

心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區Taiwan Heart Rhythm Society
 
How to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptxHow to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptxKush Bhagat
 
Patients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsPatients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsGowri Shankar
 
Aortic acceleration as a noninvasive index of left ventricular contractility ...
Aortic acceleration as a noninvasive index of left ventricular contractility ...Aortic acceleration as a noninvasive index of left ventricular contractility ...
Aortic acceleration as a noninvasive index of left ventricular contractility ...Scintica Instrumentation
 
Cardiac pacemakerspart iii
Cardiac pacemakerspart iiiCardiac pacemakerspart iii
Cardiac pacemakerspart iiisalah_atta
 
Basics of EKG Interpretation.ppt
Basics of EKG Interpretation.pptBasics of EKG Interpretation.ppt
Basics of EKG Interpretation.pptssuser20002c
 
Pacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillatorsPacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillatorsAndrewCrofton
 
Introduction To Electrophysiology
Introduction To ElectrophysiologyIntroduction To Electrophysiology
Introduction To Electrophysiologyjmlafroscia
 
simple ecg learningMEM.pptx
simple ecg learningMEM.pptxsimple ecg learningMEM.pptx
simple ecg learningMEM.pptxDr'manas Pandey
 
basicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.pptbasicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.pptSheliDuya2
 
Cardiac Arrest Rhythm.pdf
Cardiac Arrest Rhythm.pdfCardiac Arrest Rhythm.pdf
Cardiac Arrest Rhythm.pdfssuser76f7cb
 
base-110816084037-phpapp02.pdf
base-110816084037-phpapp02.pdfbase-110816084037-phpapp02.pdf
base-110816084037-phpapp02.pdfssuser61d4e0
 
心臟植入性電子儀器(CIED )護理照護指引- 術中護理照護_20130914中區
心臟植入性電子儀器(CIED )護理照護指引- 術中護理照護_20130914中區心臟植入性電子儀器(CIED )護理照護指引- 術中護理照護_20130914中區
心臟植入性電子儀器(CIED )護理照護指引- 術中護理照護_20130914中區Taiwan Heart Rhythm Society
 

Semelhante a 植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區 (20)

心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區
心臟植入性電子儀器(CIED )之歷史”CIED Overview “_20130914中區
 
How to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptxHow to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptx
 
Patients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsPatients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implications
 
Aortic acceleration as a noninvasive index of left ventricular contractility ...
Aortic acceleration as a noninvasive index of left ventricular contractility ...Aortic acceleration as a noninvasive index of left ventricular contractility ...
Aortic acceleration as a noninvasive index of left ventricular contractility ...
 
Cardiac pacemakerspart iii
Cardiac pacemakerspart iiiCardiac pacemakerspart iii
Cardiac pacemakerspart iii
 
Jp's pacemaker
Jp's pacemakerJp's pacemaker
Jp's pacemaker
 
Ecg
EcgEcg
Ecg
 
Basics of EKG Interpretation.ppt
Basics of EKG Interpretation.pptBasics of EKG Interpretation.ppt
Basics of EKG Interpretation.ppt
 
Pacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillatorsPacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillators
 
Basic Pacing Concepts
Basic Pacing ConceptsBasic Pacing Concepts
Basic Pacing Concepts
 
ECG Basics
ECG BasicsECG Basics
ECG Basics
 
Introduction To Electrophysiology
Introduction To ElectrophysiologyIntroduction To Electrophysiology
Introduction To Electrophysiology
 
simple ecg learningMEM.pptx
simple ecg learningMEM.pptxsimple ecg learningMEM.pptx
simple ecg learningMEM.pptx
 
basicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.pptbasicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.ppt
 
Cardiac Arrest Rhythm.pdf
Cardiac Arrest Rhythm.pdfCardiac Arrest Rhythm.pdf
Cardiac Arrest Rhythm.pdf
 
Pacemaker overview
Pacemaker overviewPacemaker overview
Pacemaker overview
 
Pacemakers
PacemakersPacemakers
Pacemakers
 
EKG analysis.ppt
EKG analysis.pptEKG analysis.ppt
EKG analysis.ppt
 
base-110816084037-phpapp02.pdf
base-110816084037-phpapp02.pdfbase-110816084037-phpapp02.pdf
base-110816084037-phpapp02.pdf
 
心臟植入性電子儀器(CIED )護理照護指引- 術中護理照護_20130914中區
心臟植入性電子儀器(CIED )護理照護指引- 術中護理照護_20130914中區心臟植入性電子儀器(CIED )護理照護指引- 術中護理照護_20130914中區
心臟植入性電子儀器(CIED )護理照護指引- 術中護理照護_20130914中區
 

Mais de Taiwan Heart Rhythm Society

The clinical application of intracardiac echocardiography in cardiac
The clinical application of intracardiac echocardiography in cardiacThe clinical application of intracardiac echocardiography in cardiac
The clinical application of intracardiac echocardiography in cardiacTaiwan Heart Rhythm Society
 
Oral anticoagulants in patients with atrial fibrillation
Oral anticoagulants in patients with atrial fibrillationOral anticoagulants in patients with atrial fibrillation
Oral anticoagulants in patients with atrial fibrillationTaiwan Heart Rhythm Society
 

Mais de Taiwan Heart Rhythm Society (20)

Arrhythmia news 045.pdf
Arrhythmia news 045.pdfArrhythmia news 045.pdf
Arrhythmia news 045.pdf
 
photo.pptx
photo.pptxphoto.pptx
photo.pptx
 
Arrhythmia news no.44
Arrhythmia news no.44Arrhythmia news no.44
Arrhythmia news no.44
 
Thrs arrhythmia news
Thrs arrhythmia newsThrs arrhythmia news
Thrs arrhythmia news
 
Arrhythmia news 042
Arrhythmia news 042Arrhythmia news 042
Arrhythmia news 042
 
Picture
PicturePicture
Picture
 
Arrhythmia news no.41
Arrhythmia news no.41Arrhythmia news no.41
Arrhythmia news no.41
 
Arrhythmia news no.40
Arrhythmia news no.40Arrhythmia news no.40
Arrhythmia news no.40
 
Arrhythmia news 039
Arrhythmia news 039Arrhythmia news 039
Arrhythmia news 039
 
Challenging and Unknown ECGs (2)
Challenging and Unknown ECGs (2)Challenging and Unknown ECGs (2)
Challenging and Unknown ECGs (2)
 
Arrhythmia news 038
Arrhythmia news 038Arrhythmia news 038
Arrhythmia news 038
 
Photos
PhotosPhotos
Photos
 
Arrhythmia news 037
Arrhythmia news 037Arrhythmia news 037
Arrhythmia news 037
 
Arrhythmia news no.36
Arrhythmia news no.36Arrhythmia news no.36
Arrhythmia news no.36
 
The clinical application of intracardiac echocardiography in cardiac
The clinical application of intracardiac echocardiography in cardiacThe clinical application of intracardiac echocardiography in cardiac
The clinical application of intracardiac echocardiography in cardiac
 
Comprehensive management
Comprehensive managementComprehensive management
Comprehensive management
 
Arrhythmia news 035
Arrhythmia news 035Arrhythmia news 035
Arrhythmia news 035
 
Oral anticoagulants in patients with atrial fibrillation
Oral anticoagulants in patients with atrial fibrillationOral anticoagulants in patients with atrial fibrillation
Oral anticoagulants in patients with atrial fibrillation
 
THRS allied professional training course
THRS allied professional training courseTHRS allied professional training course
THRS allied professional training course
 
Pictures
PicturesPictures
Pictures
 

植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區

  • 2. Voltage, Current, and Impedance Recap • Voltage: The force moving the current (V) – In pacemakers it is a function of the battery chemistry • Current: The actual continuing volume of flow of electricity (I) – This flow of electrons causes the myocardial cells to depolarize (to “beat”) • Impedance: The sum of all resistance to current flow (R or W or sometimes Z) – Impedance is a function of the characteristics of the conductor (wire), the electrode (tip), and the myocardium 2
  • 3. Voltage and Current Flow Electrical Analogies Spigot (voltage) turned up, lots of water flows (high current drain) Spigot (voltage) turned low, little flow (low current drain) Water pressure in system is analogous to voltage – providing the force to move the current
  • 4. Resistance and Current Flow Electrical Analogies • Normal resistance – friction caused by the hose and nozzle More water discharges, but is all of it going to the nozzle? • High resistance – a knot results in low total current flow • Low resistance – leaks in the hose reduce the resistance
  • 5. Ohm’s Law • Describes the relationship between voltage, current, and resistance • V = I X R • I = V / R • R = V / I V I R V R I V R I R V I = = = X
  • 6. Other terms • Cathode陰極: – For example, the electrode on the tip of a pacing lead • Anode陽極: – Examples: • The “ring” electrode on a bipolar lead • The IPG case on a unipolar system 6 Anode Cathode
  • 8. I II III IV Chamber(s) Paced Chamber(s) Sensed Response to Sensing Rate Modulation O = None A = Atrium V = Ventricle D = Dual (A + V) S = Single (A or V) O = None A = Atrium V = Ventricle D = Dual (A + V) S = Single (A or V) O = None T = Triggered I = Inhibited D = Dual (T + I) O = None R = Rate modulation Pacemaker modes most often seen: DDDR/VDD VVIR AAIR->DDDR (MVP) What mode would you use for 3rd Degree Block? What mode would you use for SSS? What mode would you use for permanent AF?
  • 9. Knowledge Checkpoint What is the rhythm? What type of device does this patient need?
  • 10. VVI Mode I II III Chamber(s) Paced Chamber(s) Sensed Response to Sensing O = None A = Atrium V = Ventricle D = Dual (A + V) S = Single (A or V) O = None A = Atrium V = Ventricle D = Dual (A + V) S = Single (A or V) O = None T = Triggered I = Inhibited D = Dual (I + T)
  • 11. VVI Example • Chamber paced: Ventricle • Chamber sensed: Ventricle • Response to sensing: Inhibition –VVI 60 = Lower Rate timer of 1000 ms • Pacing every 1 second if not inhibited V P V P V P Lower Rate Timer 1000 ms Lower Rate Timer 1000 ms Lower Rate Timer ….
  • 12. VVI Example (60 bpm) V P V S V P V P Lower rate timer 1000 ms x Lower rate timer 1000 ms •Paces and Senses in the ventricle •Timed from each QRS •If it sees a sensed event, it will inhibit the next pace
  • 13. VOO Mode – Asynchronous Pacing Chamber paced: Ventricle Chamber sensed: None Response to sensing: None The intrinsic ventricular event cannot be sensed, and thus, does not interrupt the pacing interval. 1000 ms1000 ms 1000 ms V P V P V P V P VOO results in fixed-rate pacing in the ventricle.
  • 14. Knowledge Checkpoint What is the rhythm? What type of device does this patient need?
  • 15. AAI Mode I II III Chamber(s) Paced Chamber(s) Sensed Response to Sensing O = None A = Atrium V = Ventricle D = Dual (A + V) S = Single (A or V) O = None A = Atrium V = Ventricle D = Dual (A + V) S = Single (A or V) O = None T = Triggered I = Inhibited D = Dual (I + T)
  • 16. AAI • Paces in the atrium • Timed from last P wave Pacing Interval A P A P Lower Rate Timer 1000 ms
  • 17. Knowledge Checkpoint What is the rhythm? What type of device does this patient need?
  • 18. DDD Mode I II III Chamber(s) Paced Chamber(s) Sensed Response to Sensing O = None A = Atrium V = Ventricle D = Dual (A + V) S = Single (A or V) O = None A = Atrium V = Ventricle D = Dual (A + V) S = Single (A or V) O = None T = Triggered I = Inhibited D = Dual (I + T)
  • 19. DDD A S V P A S V P A P V S A P V S • Senses and paces in both chambers when needed
  • 20. Knowledge Checkpoint A B C D Label each EKG with the faces of pacing (AS-VS, AP-VS, AP-VP, AS-VP).
  • 21. Magnet Application for devices • Magnet application temporarily changes pacing mode to DOO/VOO Magnet Applied 1 2 3 1 0 0 1 0 0 1 0 0
  • 22. The 4th letter: “R” • Question: what does your heart rate do when you exercise? • “R” means “Rate response” • Pacemaker will increase pacing rate in response to exercise – if patient does not increase his own rate
  • 23. General guidelines for programming common pacemaker parameter parameter situation Chronic setting comments Lower rate limit General : minimal pacing desired; 50-70bpm 40-60bpm Use rate hysteresis Upper rate limit General Child/athletes CAD/angina 85% maximal predicted HR 0.85*(220-age) (220-age) bpm 110-120bpm Based on average levels of activity May require programing short refractory periods Approximates peak HR on maximal beta blocker
  • 25. General guidelines for programming common pacemaker parameter parameter situation Chronic setting comments Pacing output Fixed voltage Fixed pulse width 3-4 X pulse width threshold 2-3X voltage threshold Minimizing voltage output more efficient Use autothreshold function sensitivity Atrium ventricle 25-50% threshold 25-50 % threshold Need <1mV setting for mode switching Evaluate oversensing in unipolar system
  • 26. General guidelines for programming common pacemaker parameter parameter situation Chronic setting comments AV delay AV block Intrinsic conduction (no HF) Intrinsic conduction (CHF) 150-180msec paced AV delay, sensed AV delay 25-50msec <paced AV delay Up to 220msec Often set even longer AV delay Turn on rate adaptive AV delay in active patients Longer AV delay may compromise hemodynamics, use AV hysteresis to promote intrinsic conduction Paced induced dyssynchrony of very long AV delay Optimize by doppler,
  • 31. ICD Function What is the function of an ICD? • Sense cardiac rhythms • Detect arrhythmias • Deliver therapy • Pace when necessary
  • 34. ICD Therapies • Tachyarrhythmia Therapy –Cardioversion (CV)Sychronized ro R –Anti-Tachycardia Pacing (ATP) –Defibrillation Shock Low Power oBradyarrhythmia Therapy –Pacing Modes Low Power High Power High Power
  • 35. Bradyarrhythmia Therapy • Most ICDs offer: –Single Chamber Pacing • AAI(R), VVI(R) and VOO –Dual Chamber Pacing • DDD(R), DDI(R), DOO and ODO • Mode Switch –Separate post-shock pacing programming • Ensures capture
  • 36. ICD common setting • VT zone 160-200bpm • VF zone >200bpm • The rate usually set 10bpm slower than the documented ventricular arrhythmia rate to reliable detection • Ventricular sensitivity must allow for the detection of very low amplitude ventricular fibrillation signals and typically set to 0.3 mV
  • 37. ICD basic parameters parameter function programming comment VF zone Detection rate for fastest VF zone: rate of VF zone must be on VF zone; Hemodynamically unstable VT Must be on at all times, only shock Rx available VT zone Detection rate for slower VTs in multizone programming On or off; Usually 10bpm < spontaneous VT ATP and cardioversion Rx available Initial detection No. of intervals VF zone : 75% intervals > VF rate VT zone: number of consecutive intervals > rate limit VF typically 12/16, or 18/24 beats > VF rate VT typically 8-20 beats >VT rate
  • 38. ICD basic parameters parameter function programming comment Ventricular sensitivity Usually 0.18-0.3mV High sensitivity needed to detect small EGM in VF ATP Painless pace termination of VT On or off Often initial RX for VTs <200bpm Type of sequence (burst or ramp) Burst: all paced intervals the same Ramp: decrements between paced intervals in sequence Burst or ramp Ramp therapy considered more aggressive but equal efficacy for spontaneous
  • 39. ICD basic parameters parameter function programming comment Cardioversion/defib rillation Shock therapy Always on in all zones Only RX in VF, or single zone; initial RX or follows ATP in VT zones Energy Magnitude of shock in joules VF: DFT =5-10 J VT: at or above smallest successful energy VT often terminated with 2- 10 J
  • 40. 41 • 13% - 38% of ICD patients experience significant levels of psychological stress related to the fear of receiving a shock.6 • Patients receiving shocks reported feeling less healthy, had lower levels of psychological well-being, and reduced physical and emotional function.7 • Shock therapy is inversely correlated with quality-of-life.8,9 • Patients receiving ATP first showed a significant increase in QOL over those receiving shock first.10 Shock Therapy is Related to a Decrease in QOL 6. Hammil SC, et al. J Cardiovasc Electrophysiol, 2000. 7. Namerow PB, et al. PACE, 1999. 8. Carroll DL, et al. Heart Lung, 2005. 9. Irvine J, et al. Am Heart J, 2002. 10. Wathen MS, et al. Circulation, 2004.
  • 41. Shock Reduction • Anti-tachycardia pacing • 延長VT/VF偵測 • 藥物
  • 42. Anti-Tachycardia Pacing Re-entry initiated ATP delivered at a rate faster than tachyarrhythmia. Wavefronts collide. Subsequent Pulse: Wavefronts collide closer to re-entry circuit Subsequent Pulses: Wavefronts collide even closer to re-entry circuit Arrhythmia terminated
  • 45. Anti-tachycardia pacing(ATP) Burst v.s Ramp Programmed Values: Number of S1 Pulses = 4 Number of Sequences = 4 R- S1% = 91% (less agrassive) Decrement* = 10 ms * Decrement between sequences Programmed Values: Number of S1 Pulses = 4 Number of Sequences = 4 R-S1% = 91% (less aggrassive) Decrement* = 10 ms * Decrement between pulses * Adds a pulse per sequence PITAGORA ICD study: Bust is significantly more efficacious than ramp in terminating FVT episodes
  • 46. Sensing • Sensing is: –The process of identifying cardiac depolarizations from an intracardiac electrogram Measured Peak-to-Peak >5 mV for optimal sensing
  • 47. Detection • Measured in: –Beat-to-beat intervals (milliseconds), or –Beats-per-minute (BPM) Detection Rate •Classifies rhythm by detection zone: –VT = Ventricular Tachycardia +/- FVT –VF = Ventricular Fibrillation •Programmable in ranges of rates Example:VT = 162 bpm – 188 bpm VF = 188 bpm and faster
  • 50. Detection • Measured in: – Number of intervals to detect (NID), or – Length of time to detect •Programmable by: – Beat or interval counters • Consecutive ex: 16 beats within the detect zone • Probabilistic (percentage or fraction) ex: 12 out of 16 beats within the detect zone – Time in seconds Detect Duration
  • 51. Detection Used for detection of VT Consecutive Counter
  • 53. • Can you identify the detect zones? • Name the rate & duration for each Detection Detect Zones
  • 54. Cardioversion • Delivers shock on an R-wave • Aborts if synchronization cannot be obtained due to arrhythmia termination
  • 57. Major Challenges Faced • Inappropriate therapies • Unnecessary shocks • Repetitive shocks 58
  • 58. 59 • 13% - 38% of ICD patients experience significant levels of psychological stress related to the fear of receiving a shock.6 • Patients receiving shocks reported feeling less healthy, had lower levels of psychological well-being, and reduced physical and emotional function.7 • Shock therapy is inversely correlated with quality-of-life.8,9 • Patients receiving ATP first showed a significant increase in QOL over those receiving shock first.10 Shock Therapy is Related to a Decrease in QOL 6. Hammil SC, et al. J Cardiovasc Electrophysiol, 2000. 7. Namerow PB, et al. PACE, 1999. 8. Carroll DL, et al. Heart Lung, 2005. 9. Irvine J, et al. Am Heart J, 2002. 10. Wathen MS, et al. Circulation, 2004.
  • 59.
  • 61. Differentiation of SVT from VT parameter What It does Potentially useful For Potential problem stability Suppressed therapy for tachyarrhythmia with variable ventricular rate Atrial fibrillation Underdetection of VT with irregular rate; failure to suppress therapy for SVTs with regular ventricular response onset Suppressed therapy for tachyarrhythmia that slowly accelerate Sinus tachycardia Underdetection of gradually accelerating VT or VT onset during sinus tachycardia; failure to suppress therapy for sudden onset SVT Ventricular electrogram width Suppress therapy for tachyarrhythmia with narrow ventricular EGM correlated to narrow QRS complex Differentiation of narrow complex SVT from VT Limited specificity with BBB; may prevent therapy for narrow complex VT
  • 62. Differentiation of SVT from VT parameter What It does Potentially useful For Potential problem Ventricular electrogram morphology Suppressed therapy for tachyarrhythmia with ventricular EGM morphology similar to that in sinus rhythm D/D SVT from VT Limited specificity with BBB Atrial to ventricular ratio Compare atrial to ventricular rate Atrial fibrillation Atrial undersensing can result in false diagnosis of VT
  • 63.
  • 64. SVT Discriminators • Distinguishes SVTs by analyzing P and R-wave: –Rate –Regularity –AV Association PR Logic™ – Pattern
  • 65. SVT Discriminators • Based on the premise that AF conducts irregularly to the ventricles (and VT is a stable, regular rhythm) • Discriminates regular from irregular intervals within a detect zone Stability
  • 66. SVT Discriminators UnstableVaries >50 ms from previous 3 Stability = 50 ms Stability * in Medtronic devices
  • 67. SVT Discriminators • Based on the premise that most VTs are characterized by a sudden onset •Evaluates the acceleration of the ventricular rate •Discriminates between: –Gradual rate increase –Abrupt rate increase •Determines VT present if rate increase is abrupt Onset
  • 68. SVT Discriminators • Onset Percentage = 81% 530ms X 81% = 430ms 430ms  460ms = Onset Not Met Onset * in Medtronic devices
  • 69. SVT Discriminators • Measures and stores the QRS characteristics of a normal sinus beat • Identifies SVT vs. VT based on the QRS changes that occur in most VTs SINUS RHYTHM VT Waveform Morphology
  • 70. SVT Discriminators • Identifies start and end points of a sensed QRS complex • Uses 2 parameters to measures QRS: –Slew Width –Slew Threshold EGM Width
  • 72. SVT Discriminators • Records and stores a template of a normal QRS wave •Compares stored template with a QRS occurring within the detection zone •Withholds detection if 3 of last 8 QRS complexes match the stored template – Detects VT/FVT/VF if 6 out of 8 do not match •Applies to initial detection only Wavelet
  • 73. SVT Discriminators • Template % difference compared against: Match Threshold Value Wavelet
  • 74. Refractory and Blanking Periods • Pacemaker sensing occurs when a signal is large enough to cross the sensing threshold 1.25 mV Sensitivity Time 5.0 mV 2.5 mV 1.25 mV Sensing does not tells us anything about the origin or morphology of the sensed event, only its “size.”
  • 75. In DDD & VDD modes the pacemaker will “track” the atrium AS VP Tracking = Pacing the ventricle after an atrial intrinsic event Maintains AV Synchrony Want to limit how fast we pace Upper Tracking Rate
  • 76. DDDR 60 / 120 A-A = 500 ms Upper Activity Rate Limit Lower Rate Limit Upper Sensor Rate • Sensor rate drives the atrial rate up • In rate responsive, dual chamber modes, the Upper Activity (Sensor) Rate provides the limit for sensor-driven atrial pacing PAV PAV 1000 ms 500 ms