Electrical Testing Of Pacemaker
  & Pacemaker Complications
            Msn Pavan Kumar,DM,NIMS.
Electrical Testing Of Pacemaker
Pacemaker complications
Electrical Testing Of Pacemaker
 Pacemaker components
   Battery
   Pacing impedance
   Pulse generator
    1.   Output circuit
    2.   Sensing circuit
    3.   Timing circuit
    4.   Rate adaptive sensor
    5.   Modes and mode switching
Electrical Testing Of Pacemaker
    Battery :
     Lithium iodine battery
       High energy density ,
         Long shelf life ,
         Predictable loss of battery
     BOL (vol) – 2.8v
      BOL (res) - <1komhs
Electrical Testing Of Pacemaker
 Pacing impedance :
 Pacing impedance refers to the opposition to current flow. Three
  sources contribute to pacing impedance:
     1. Pacing lead conductor coil
     2. Electrode-tissue interface
         Electrode resistance
         Polarization
 Normal lead impedance vary from 250-1200ohms.
 Single impedance value may be of little use with out previous values
  for comparison.
Electrical Testing Of Pacemaker
1.Pulse generator output circuit
 Capture threshold , Pacing threshold , stimulation threshold
 Minimum amount of energy required to constantly cause
  depolarization
 Volts and pulse duration
Electrical Testing Of Pacemaker
1. Pulse generator output circuit


     2.0 v                1.5 v     1v
Electrical Testing Of Pacemaker
 1. Pulse generator output circuit




Site      At implantation Acute                    Chronic
Atrium    <1.5mv                3-5 times          Twice the
                                threshold          Threshold voltage
                          voltage
Ventricle <1mv                       With PW 0.5ms With PW of 0.5ms
Electrical Testing Of Pacemaker
1.Pulse generator output circuit
 High Pacemaker Output can cause
 Reduce longevity
 Diaphragmatic stimulation
 Muscle Sti. in Unipolar pacemakers
 Patient may “feel” heart beat
 Algorithm for checking pacemaker
  output threshold every beat and
  maintaining threshold just above it -
  Auto capture.
Electrical Testing Of Pacemaker
2.Pulse generator sensing circuit :
 Ability of the device to detect intrinsic beat of the heart
 Measured - peak to peak magnitude (mv) & slew rate(mv/ms)
Electrical Testing Of Pacemaker
2.     Pulse generator sensing circuit :
 Reduce Lower Rate below intrinsic rate to inhibit pacing and ensure
  intrinsic activity
 Increase sensitivity setting while observing EGM. The sensitivity
  value at which sensing is lost on the EGM is the sensing threshold.
 Sensitivity threshold safety is twice the attained valve.

                           Sensitivity         Slew rate

     Atrium                1-2mv(0.5mv)        > 0.5 v/s

     ventricle             2-3mv               > 0.75 v/s
Electrical Testing Of Pacemaker
3.     Pulse generator timing circuit :
     a. Lower rate limit (LRL)
     b. Hysteresis rate
     c. Refractory and blanking periods
     d. Ventricular safety pacing interval .
     e. Upper rate response .
Electrical Testing Of Pacemaker
3.Pulse generator timing circuit :
 Lower rate interval - lowest rate that the pacemaker will pace .
 A paced or non-refractory sensed event restarts the rate timer at
  the programmed rate.
Electrical Testing Of Pacemaker
3. Pulse generator timing circuit :

Condition                                 LRL (beats/mt)
Infrequent pauses                         40-50
Chronic persistent bradycardia            60-70
Relative bradycardia detrimental (long QT) 70-80
Detrimental fast heart rates (angina)     50-60
VVI                                       60-70
Electrical Testing Of Pacemaker
3. Pulse generator timing circuit :
 Hysteresis :
 Hysteresis allows the rate to drop below the programmed pacing LRL.
 Advantages of hysteresis :
 1. Encourages native rhythm – maintain AV sync in VVI , prolong
     battery life
 2. Prevent retrograde conduction – avoids pacemaker syndrome
Electrical Testing Of Pacemaker
3. Pulse generator timing circuit :
  AV delay (AVI) – pacemaker equivalent of PR interval.
  Sensed vs paced AVI – paced AVI is programmed at 125-200ms ,
   sensed AV interval is programmed at 20-50ms shorter than paced.
  Dynamic AV delay allow pacemaker to respond to exercise




                          sAVI – 150ms
                          pAVI – 200ms
Electrical Testing Of Pacemaker
3. Pulse generator timing circuit :

 AV delay (AVI)
 Longer AVI :
    Good AV conduction – maintains AV synchrony , long battery life
    Achieved by following methods :Programming longer AVI ,
    managed ventricular pacing , AV delay hysteresis .
 Shorter AVI:
    HOCM – RV apical pacing decreases HOCM gradient
    CRT – usually 80-120ms , for 100%ventricular pacing and optimize CO
    Physiological response to faster heart rates can be answered
Electrical Testing Of Pacemaker
3. Pulse generator timing circuit :

  Refractory and
   blanking periods :
  Refractory period –
   sensing present but no
   action
  Blanking period -
   sensing absent and
   hence no action
Electrical Testing Of Pacemaker
3. Pulse generator timing circuit :  Blanking periods :

Blanking period        Time         Importance
Atrial blanking period 50-100ms     Non programmable ,
                                    Avoid atrial sensing of its own paced
                                    beat
Post ventricular atrial 220ms       Avoid sensing of ventricular beat
blanking period                     Long PVAB decreases detection of
                                    AF,AFL
Ventricular blanking   50-100ms     Non programmable,
period                              Avoid ventricular sensing of its paced
                                    beat
Post atrial ventricular 28ms        if the PAvB period is too long, R on T -
blanking period                     ventricular tachyarrhythmia.
Electrical Testing Of Pacemaker
3. Pulse generator timing circuit :  Refractory period:

 Refractory period                       Importance
 Ventricular refractory period (VRP)     Prevent sensing of T wave .
 Atrial refractory period (ARP)          AVI (120-200ms) .
 Post ventricular atrial refractory period Avoid sensing retrograde P waves
                                           (PMT) , far field R waves .
Electrical Testing Of Pacemaker
 3. Pulse generator timing circuit :
Ventricular safety pacing/ventricular
triggered period/cross talk sensing window :

            Atrial pacing in DDD


         Trigger ventricular sensing
               PAVB - pAVI


            False inhibition of
          ventricular pacing circuit


                Asystole
Electrical Testing Of Pacemaker
4. Pulse generator rate responsive pacing:

 Rate responsive pacing refer to ability of pacemaker to increase its
  lower rate in response to physiological stimulus
 Sinus node dysfunction , AF patients – fail to increase heart rates
 HRR should start with in 10s of exercise , peak at 90 – 120s and
  should return to baseline with in 60 – 120s after exercise.
 Fastest rate at which pacemaker will pace upper rate response.
 If intrinsic atrial rate exceeds URR then wenckebach or 2:1 AVB
 Choosing URR : young patients (150b/mt) , old angina (<110b/mt).
 Various sensors (activity , minute ventilation , QT)
Electrical Testing Of Pacemaker
4. Pulse generator rate responsive pacing:




                                             Wenckebach




                                             2:1 AVB
Electrical Testing Of Pacemaker
4. Pulse generator modes:
Electrical Testing Of Pacemaker
5. Pulse generator modes switching:

                     DDD / VDD
                 Atrial tachyarrythmias


                  Sensed atrial events

                                                DDIR /
               Trigger fast ventricular rates   VVIR

         Palpitations. Dyspnoea. And Fatigue.
Electrical Testing Of Pacemaker
5. Pulse generator modes switching:




 Programming mode switching Mode switching occurs when
 the sensed atrial rate exceeds a programmed atrial tachycardia
 detection rate. By definition, this value must be faster than the URL
 (maximum tracking rate). Atrial tachycardia detection is typically
 programmed to 175-l88bpm or thereabouts.AMS base rate is higher
 than LRI.
Electrical Testing Of Pacemaker
Pacemaker follow up guidelines:        Transtelephonic monitoring guidelines
    Medicare guidelines
         Single chambered pacing              Dual chambered pacing
         1st month q 2 week                   1st month q 2 weeks
         2nd -48th month q 12 week            2nd – 30th month q 12weeks
         49th – 72nd month q 8 week           31st – 48th month q 8 weeks
         73rd month and later q 4 weeks       49th month and later q 4week
  NASPE guidelines
            Single or dual pacing
            1st visit 6 – 8 week post implant , if symptomatic prior to this
            5th month
            From 6th month q 3month
            Battery wear present q 1month
Electrical Testing Of Pacemaker
Pacemaker complications
Pacemaker complications

                       Pacemaker complications
Pocket complications   Lead dislodgement
Pocket hematoma        Pneumothorax /air embolism
Infection              Cardiac perforation
Erosion                Extracardiac stimulation
Wound pain             Venous thrombosis
Allergic reactions     Coronary sinus dissection
                       Twidller syndrome
                       Pacemaker malfunction
Pacemaker complications

Pocket hematoma :
 The risk of haematoma is increased in patients taking antithrombotic
  or anticoagulant drugs (Goldstein et al., 1998).
 Most small hematomas can be managed conservatively with cold
  compress and withdrawal of antiplatelet or antithrombotic agents.
 Occasionally, large hematomas that compromise the suture line or
  skin integrity may have to be surgically evacuated.
 Needle aspiration increases risk of infection and should not be done.
Pacemaker complications

Pocket hematoma :
 In patients requiring oral anticoagulants (warfarin), to take INR of
  about 2.0 at the time of implantation is safe (Belott & Reynolds, 2000).
 Unfractionated heparin or low-molecular-weight heparin are always
  discontinued prior to device implant and ideally avoided for a
  minimum of 24 hours post implantation.
 Administration of anticoagulants can be resumed within 48-72 h after
  implantation if there is no evidence of substantial hematoma formation.
Pacemaker complications
Device-related infections :
 The reported incidence of pacemaker-related infection ranges from
  0.5% to 6% in early series
 The use of prophylactic antibiotics and pocket irrigation with antibiotic
  solutions has decreased the rate of acute infections following pacemaker
  implantations to <1 to 2 percent in most series
 The mortality of persistent infection when infected leads are not
  removed can be as high as 66%.
 DM, malignancy, operator inexperience, advanced age, corticosteroid
  use, anticoagulation, recent device manipulation, CRF, and bacteremia
  from a distant focus of infection.
Pacemaker complications
Device related infection :
 Device infection is defined as either:
   (a) deep infection - infection involving the generator pocket
    and/or the intravenous portion of the leads, with bacteremia,
    requiring device extraction or
   (b) superficial infection - characterized by local inflammation,
    involving the skin but not the generator pocket, and treated with
    oral antibiotics.
Pacemaker complications
Device related infection :




                             2007;49;1851-1859 J. Am. Coll. Cardiol.
Pacemaker complications
Device related infection :




                             2007;49;1851-1859 J. Am. Coll. Cardiol.
Pacemaker complications
Device related infection :




                             2007;49;1851-1859 J. Am. Coll. Cardiol.
Pacemaker complications
Wound pain :
 Infection , Pacemaker implanted too superficially , Pacemaker
  implanted too laterally , Pacemaker allergy .
Skin erosion :
 Incidence has been estimated around 0.8% .Old age , infection.
 Surgical revision of pocket and reimplantation .
Allergic reactions :
 Always rule out infection before coming to diagnosis of allergy
Pacemaker complications
Lead dislodgement:
 Relatively common – 5-10% of
    patients(ICD database 2001)
 Atrial more common than
    ventricular(2-3% vs. 1%)
 Micro dislodgement , macro
    dislodgement
 Increased pacing threshold , failure
    to pace and sense
 Active fixation (decreases risk)
Pacemaker complications
Pneumothorax , :
 Uncommon complication – 1.6-2.6%
 During or 48 hrs after procedure
 Inadvent puncture and laceration of
  subclavian vein , artery or lung
 Related to operator experience and
  underlying anatomy
 Avoided by
   1. Venogram – flouroscpic puncture
   2. Axillary venous access (Martin etal’96)
   3. One way mechanism sheath
Pacemaker complications
Cardiac Perforation :
 Uncommon but potentially serious complication - lower than 1%.
 Acute (<5 days) , subacute(5d-1month) , chronic (>1month)
 Increasing stimulation threshold , RBBB pattern for RV pacing,
  intercostal muscle or diaphragmatic contraction, friction rub, and
  pericarditis, pericardial effusion, or cardiac tamponade.
 CXR , ECHO , CT
 Lead withdrawal and repositioning ; surgical back up
Pacemaker complications
Cardiac Perforation :
Pacemaker complications
Extracardiac stimulation
 The diaphragm or pectoral or intercostal muscles
 Diaphragmatic stimulation - direct stimulation of the diaphragm (left)
  or stimulation of the phrenic nerve (right).
 Early postimplantation period , dislodgment of the pacing lead.
 MC in patients with LV coronary vein branch lead placement for CRT
 Output pacing importance (testing and treatment)
 Pectoral stimulation - incorrect orientation of the pacemaker or a
  current leak from a lead insulation failure or exposed connector.
Pacemaker complications
Venous thrombosis :
 Venous thrombosis occurs in 30% to 50% of patients and only 1-3%
  of patients become symptomatic.
 Manifestations vary from usually asymptomatic, acute symptomatic
  thrombosis, and even SVCS .
 Early or late after pacemaker implantation.
 Predictors of severe stenosis are multiple pacemaker leads , previous
  pacing , double coils , hormone therapy .
 Asymptomatic (no treatment) , symptomatic (anticoagulants –
  endovascular stents – surgical correction ).
Pacemaker complications
Twiddler syndrome:

        Obese women with loose, fatty subcutaneous tissue
      Small size of the implanted generator with a large pocket

              Twisting of pulse generator in long axis


               Lead dislodgement and lead fracture

                        Failure to capture
Pacemaker complications
    Twiddler syndrome:
 The prevelance of this syndrome is
  0.07% (Gungor et al., 2009)
 Rotated along the transverse axis it is
  referred by us as the reel syndrome.
 Pocket should be revised.
 Avoid by
  Limit the pocket size,
   Suture the device to the fascia
   The patients not to manipulate
     their device pocket
Pacemaker malfunction

 Failure to capture
 Failure to output
 Sensing abnormalities(under and over sensing)
 Specific mode complications
  1. Pacemaker related tachycardia
  2. Pacemaker syndrome
Pacemaker malfunction
Failure to capture:
 Pacing artifact present but no evoked potential .
 Causes
   1. Lead dislodgement or perforation
   2. Lead maturation(inflammation/fibrosis)(exit block)
   3. Battery depletion
   4. Circuit failure(coil fracture , insulation defect)
   5. Capture management algorithm failure
   6. Inappropriate programming
   7. Pseudo malfunction
   8. Functional non capture
   9. Metabolic , drugs , cardiomyopathies
Pacemaker malfunction
 Failure to capture:




Electrocardiographic tracing from a patient with a DDDR pacemaker. All ventricular
       pacing artifacts but one failed to result in ventricular depolarization—
                             that is, failure to capture
Pacemaker malfunction
                    Failure to capture:

                     Pacing threshold

     Normal                          Increased
                                                 Dislodgement
                                          Normal Exit block
Battery depletion
Functional non capture
                         Impedance                  Insulation
                                          Decreased failure/break


                                                      Lead fracture
                                          Increased   Loose screw
Pacemaker malfunction
Failure to output:
 Absence of pacing stimuli and hence no capture .
 Causes
   1. Pseudo malfunction - hysteresis , PMT termination , sleep rate
   2. Over sensing - EMI ; T P R over sensing ;
       Myopotential/diaphragmatic ; Cross talk ; Make break signals
   3. Open circuit - lead fracture , loose screw , air in the pocket ,
       incompatible lead .
   4. Battery depletion
   5. Recording artifact.
Pacemaker malfunction
     Failure to output:




 VVIR pacemaker This patient had a pacemaker programmed to a unipolar
  sensing configuration. The sensing of myopotentials led to symptomatic
pauses, and reprogramming the pacemaker to a bipolar sensing configuration
             prevented subsequent myopotential over sensing.
Pacemaker malfunction
                        Failure to
                           output:
                      Application of magnet


  Eliminates pauses                     Pauses persistent


Over sensing                                  Normal Battery
                                                     depletion
Pseudo malfunction
                            Impedance                   Insulation
                                              Decreased failure/break


                                                            Lead fracture
                                              Increased     Loose screw
Pacemaker malfunction
Battery depletion :
 Elective replacement indicators (ERI)
   1. Low voltage(2.1-2.4)
   2. Low pacing rate on magnet application
   3. Elevated battery impedance
   4. Increased pulse width duration
   5. Restricted programmability
   6. Change to simpler pacing mode
 End of life (EOL)
   1. Low voltage(≤2.1vol)
Pacemaker malfunction
Pacemaker undersensing :
 Pacing artifact present but no sensing(sensed beat doesn’t reset cycle)
 Causes are
  1. Defect in signal production – scar /fibrosis , BBB , ectopic ,
      cardioversion , defibrillation , metabolic.
  2. Defect in signal transmission – lead fracture /dislodgement ,
      insulation failure , partial open circuit.
  3. Defect in pacemaker – battery depletion , sensing circuit
      abnormalities , committed DVI.
Pacemaker malfunction
Pacemaker undersensing :




                           VVI pacemaker
Pacemaker malfunction
Pacemaker over sensing :
                                Cross talk :
 Present as failure to pace
                                       High atrial output
 Causes
                                   High ventricular sensitivity
  1. EMI                                   Low VBP
  2. T , P , R over sensing .
  3. Cross talk                   Ventricular sensing of
  4. Myopotential (unipolar)      paced atrial impulse
  5. Make break signals
                                Pts with Poor AV conduction
                                – Ventricular Asystole
Pacemaker malfunction
    Electromagnetic interference :

Source                  Pacer    Inhibition Rate       Asynchronous   Uni/
                        damage              increase   noise          bipolar
Cardioversion/          Y        N          N          N              U/B
Defibrillation
Anit theft devices /    N        Y          N          N              U
Weapon detector
Phone (cell/cordless)   N        Y          Y          Y              U/B
Ablation                Y        Y          Y          N              U/B
Diathermy/              Y        Y          Y          Y              U/B
lithotripsy
FM radio                N        Y          N          Y              U
TV transmitter
MRI/PET                 Y        Y(N)       Y(N)       Y(N)           U/B
Pacemaker malfunction
Pacemaker syndrome :
 Seen in 20% of PPI (5% severe
  symptomatic)
 VVI/DDD/AAI
 Pulsations in neck , fatigue ,
  cough ,chest fullness , headache ,
  chocking sensation , PND ,
  confusion , syncope , pulmonary
  edema.
 Rx : VVI – program hysteresis ,
  or change to DDD ; DDD –atrial
  lead reprogrammed or changed
Pacemaker malfunction
Pacemaker mediated tachycardia :


                                 Dual chamber
                                 VPC , intact retrograde
                                  conduction , PVARP<VA .
                                 Px , Rx :
                                  1. PVARP > VA
                                  2. Long PVARP after VPC
                                  3. Absent atrial sensing after VPC
Pacemaker Testing
Pacemaker Testing
Pacemaker Testing
Thank you
Hey:Share with me please.

Mais conteúdo relacionado

Mais procurados

Echocardiography of Mitral regurgitation
Echocardiography of Mitral regurgitationEchocardiography of Mitral regurgitation
Echocardiography of Mitral regurgitationDr. Muhammad AzAm Shah
 
Assessment of shunt by cardiac catheterization
Assessment of shunt by cardiac catheterizationAssessment of shunt by cardiac catheterization
Assessment of shunt by cardiac catheterizationRamachandra Barik
 
Follow up and management of pacemaker programming and
Follow up and management of pacemaker programming andFollow up and management of pacemaker programming and
Follow up and management of pacemaker programming anddrskd6
 
ventricular tachycardia (VT) Localisation
ventricular tachycardia (VT) Localisationventricular tachycardia (VT) Localisation
ventricular tachycardia (VT) LocalisationMalleswara rao Dangeti
 
ELectrophysiology basics part4
ELectrophysiology basics part4ELectrophysiology basics part4
ELectrophysiology basics part4salah_atta
 
Diastolic Dysfunction 2016
Diastolic Dysfunction 2016Diastolic Dysfunction 2016
Diastolic Dysfunction 2016drabhishekbabbu
 
Cardiac pacemakers part ii
Cardiac pacemakers part iiCardiac pacemakers part ii
Cardiac pacemakers part iisalah_atta
 
Basics of Pacemaker Dr Hafeesh Fazulu Pushpagiri
Basics of Pacemaker Dr Hafeesh Fazulu PushpagiriBasics of Pacemaker Dr Hafeesh Fazulu Pushpagiri
Basics of Pacemaker Dr Hafeesh Fazulu PushpagiriHafeesh Fazulu
 
Basics of electrophysiology
Basics of electrophysiologyBasics of electrophysiology
Basics of electrophysiologyPinkesh Parmar
 
Echo assessment of Aortic Regurgitation
Echo assessment of Aortic RegurgitationEcho assessment of Aortic Regurgitation
Echo assessment of Aortic RegurgitationMashiul Alam
 
Cardiac resynchronization therapy
Cardiac resynchronization therapyCardiac resynchronization therapy
Cardiac resynchronization therapyRamachandra Barik
 
Eps basics,part2(lecture)
Eps basics,part2(lecture)Eps basics,part2(lecture)
Eps basics,part2(lecture)salah_atta
 
Cardiac Resynchronization therapy.pptx
Cardiac Resynchronization therapy.pptxCardiac Resynchronization therapy.pptx
Cardiac Resynchronization therapy.pptxdesktoppc
 
Ecg criteria of chamber enlargement
Ecg criteria of chamber enlargementEcg criteria of chamber enlargement
Ecg criteria of chamber enlargementAdarsh
 

Mais procurados (20)

Electrophysiology AVNRT
Electrophysiology AVNRTElectrophysiology AVNRT
Electrophysiology AVNRT
 
Echocardiography of Mitral regurgitation
Echocardiography of Mitral regurgitationEchocardiography of Mitral regurgitation
Echocardiography of Mitral regurgitation
 
Pacemaker ECGs. Yasmeen Kamal
Pacemaker ECGs. Yasmeen KamalPacemaker ECGs. Yasmeen Kamal
Pacemaker ECGs. Yasmeen Kamal
 
Assessment of shunt by cardiac catheterization
Assessment of shunt by cardiac catheterizationAssessment of shunt by cardiac catheterization
Assessment of shunt by cardiac catheterization
 
ICD troubleshooting
ICD troubleshootingICD troubleshooting
ICD troubleshooting
 
Follow up and management of pacemaker programming and
Follow up and management of pacemaker programming andFollow up and management of pacemaker programming and
Follow up and management of pacemaker programming and
 
ventricular tachycardia (VT) Localisation
ventricular tachycardia (VT) Localisationventricular tachycardia (VT) Localisation
ventricular tachycardia (VT) Localisation
 
ELectrophysiology basics part4
ELectrophysiology basics part4ELectrophysiology basics part4
ELectrophysiology basics part4
 
Diastolic Dysfunction 2016
Diastolic Dysfunction 2016Diastolic Dysfunction 2016
Diastolic Dysfunction 2016
 
Cardiac pacemakers part ii
Cardiac pacemakers part iiCardiac pacemakers part ii
Cardiac pacemakers part ii
 
Basics of Pacemaker Dr Hafeesh Fazulu Pushpagiri
Basics of Pacemaker Dr Hafeesh Fazulu PushpagiriBasics of Pacemaker Dr Hafeesh Fazulu Pushpagiri
Basics of Pacemaker Dr Hafeesh Fazulu Pushpagiri
 
Basics of electrophysiology
Basics of electrophysiologyBasics of electrophysiology
Basics of electrophysiology
 
Echo assessment of Aortic Regurgitation
Echo assessment of Aortic RegurgitationEcho assessment of Aortic Regurgitation
Echo assessment of Aortic Regurgitation
 
Cardiac resynchronization therapy
Cardiac resynchronization therapyCardiac resynchronization therapy
Cardiac resynchronization therapy
 
Eps basics,part2(lecture)
Eps basics,part2(lecture)Eps basics,part2(lecture)
Eps basics,part2(lecture)
 
Electrophysiology AVRT
Electrophysiology AVRTElectrophysiology AVRT
Electrophysiology AVRT
 
Implication of 3D Mapping in EP
Implication of 3D Mapping in EP Implication of 3D Mapping in EP
Implication of 3D Mapping in EP
 
Cardiac Resynchronization therapy.pptx
Cardiac Resynchronization therapy.pptxCardiac Resynchronization therapy.pptx
Cardiac Resynchronization therapy.pptx
 
Fraction flow reserve
Fraction flow reserveFraction flow reserve
Fraction flow reserve
 
Ecg criteria of chamber enlargement
Ecg criteria of chamber enlargementEcg criteria of chamber enlargement
Ecg criteria of chamber enlargement
 

Destaque

Management of a patient with pacemaker
Management of a patient with pacemakerManagement of a patient with pacemaker
Management of a patient with pacemakersalman habeeb
 
A D V A N C E D P A C I N G
A D V A N C E D  P A C I N GA D V A N C E D  P A C I N G
A D V A N C E D P A C I N GSherry Knowles
 
Cardiac pacemakerspart iii
Cardiac pacemakerspart iiiCardiac pacemakerspart iii
Cardiac pacemakerspart iiisalah_atta
 
Dr hardik temporary pacemaker preview (1)
Dr hardik temporary pacemaker  preview (1)Dr hardik temporary pacemaker  preview (1)
Dr hardik temporary pacemaker preview (1)Hardik Patel
 
Cardiac pace makerspart 1
Cardiac pace makerspart 1Cardiac pace makerspart 1
Cardiac pace makerspart 1salah_atta
 
Pacemakers ppt1
Pacemakers  ppt1Pacemakers  ppt1
Pacemakers ppt1tulasi27
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slidesmaltiaziz24
 

Destaque (14)

Management of a patient with pacemaker
Management of a patient with pacemakerManagement of a patient with pacemaker
Management of a patient with pacemaker
 
A D V A N C E D P A C I N G
A D V A N C E D  P A C I N GA D V A N C E D  P A C I N G
A D V A N C E D P A C I N G
 
Cardiac pacemakerspart iii
Cardiac pacemakerspart iiiCardiac pacemakerspart iii
Cardiac pacemakerspart iii
 
Dr hardik temporary pacemaker preview (1)
Dr hardik temporary pacemaker  preview (1)Dr hardik temporary pacemaker  preview (1)
Dr hardik temporary pacemaker preview (1)
 
Cardiac pace makerspart 1
Cardiac pace makerspart 1Cardiac pace makerspart 1
Cardiac pace makerspart 1
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
Pacemaker Rhythms
Pacemaker RhythmsPacemaker Rhythms
Pacemaker Rhythms
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
Pacemaker basics
Pacemaker basicsPacemaker basics
Pacemaker basics
 
Pacemakers
PacemakersPacemakers
Pacemakers
 
Pacemakers ppt1
Pacemakers  ppt1Pacemakers  ppt1
Pacemakers ppt1
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slides
 
Cardiac Pacemaker
Cardiac PacemakerCardiac Pacemaker
Cardiac Pacemaker
 
pacemaker
pacemakerpacemaker
pacemaker
 

Semelhante a Electrical testing of pacemaker

心臟植入性電子儀器(CIED)的基本原理及設定
心臟植入性電子儀器(CIED)的基本原理及設定心臟植入性電子儀器(CIED)的基本原理及設定
心臟植入性電子儀器(CIED)的基本原理及設定Taiwan Heart Rhythm Society
 
Patients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsPatients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsGowri Shankar
 
Pacemaker & cabg
Pacemaker & cabgPacemaker & cabg
Pacemaker & cabgEhab Khairy
 
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區Taiwan Heart Rhythm Society
 
Introduction To Electrophysiology
Introduction To ElectrophysiologyIntroduction To Electrophysiology
Introduction To Electrophysiologyjmlafroscia
 
Electro Cardio Gram - Basics of ECG
Electro Cardio Gram - Basics of ECG Electro Cardio Gram - Basics of ECG
Electro Cardio Gram - Basics of ECG Dr. Dinesh C. Sharma
 
Cardiac rhythm management
Cardiac rhythm managementCardiac rhythm management
Cardiac rhythm managementshivam keshari
 
How to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptxHow to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptxKush Bhagat
 
INTRACARDIAC DEVICES AND ITS COMPLICATION seminar..pptx
INTRACARDIAC DEVICES AND ITS COMPLICATION seminar..pptxINTRACARDIAC DEVICES AND ITS COMPLICATION seminar..pptx
INTRACARDIAC DEVICES AND ITS COMPLICATION seminar..pptxddocofdera
 
Pacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillatorsPacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillatorsAndrewCrofton
 
Basics of EKG Interpretation.ppt
Basics of EKG Interpretation.pptBasics of EKG Interpretation.ppt
Basics of EKG Interpretation.pptssuser20002c
 
Basics of Electrocardiography, Arrhythmia & Pacemaker
Basics of Electrocardiography, Arrhythmia & PacemakerBasics of Electrocardiography, Arrhythmia & Pacemaker
Basics of Electrocardiography, Arrhythmia & PacemakerPallab Nath
 
Pacemaker - EC8073 Medical Electronics - Hints for Slow Learner
Pacemaker - EC8073 Medical Electronics - Hints for Slow LearnerPacemaker - EC8073 Medical Electronics - Hints for Slow Learner
Pacemaker - EC8073 Medical Electronics - Hints for Slow LearnerMathavan N
 
basicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.pptbasicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.pptSheliDuya2
 
Anaesthesia for patient with pacemaker
Anaesthesia for patient with pacemakerAnaesthesia for patient with pacemaker
Anaesthesia for patient with pacemakerHASSAN RASHID
 
Cardiac Arrest Rhythm.pdf
Cardiac Arrest Rhythm.pdfCardiac Arrest Rhythm.pdf
Cardiac Arrest Rhythm.pdfssuser76f7cb
 
Ep diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrtEp diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrtRahul Chalwade
 
Anesthesia and cardiac pacemaker
Anesthesia  and   cardiac pacemakerAnesthesia  and   cardiac pacemaker
Anesthesia and cardiac pacemakerVishal Cb
 
temporary pacemaker easy to learn for PICU
temporary pacemaker easy to learn for PICUtemporary pacemaker easy to learn for PICU
temporary pacemaker easy to learn for PICUAbdul Ahad Korar
 

Semelhante a Electrical testing of pacemaker (20)

心臟植入性電子儀器(CIED)的基本原理及設定
心臟植入性電子儀器(CIED)的基本原理及設定心臟植入性電子儀器(CIED)的基本原理及設定
心臟植入性電子儀器(CIED)的基本原理及設定
 
Patients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsPatients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implications
 
Pacemaker & cabg
Pacemaker & cabgPacemaker & cabg
Pacemaker & cabg
 
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
 
Introduction To Electrophysiology
Introduction To ElectrophysiologyIntroduction To Electrophysiology
Introduction To Electrophysiology
 
Electro Cardio Gram - Basics of ECG
Electro Cardio Gram - Basics of ECG Electro Cardio Gram - Basics of ECG
Electro Cardio Gram - Basics of ECG
 
Cardiac rhythm management
Cardiac rhythm managementCardiac rhythm management
Cardiac rhythm management
 
How to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptxHow to read ICD episode_Dr Kush Bhagat.pptx
How to read ICD episode_Dr Kush Bhagat.pptx
 
INTRACARDIAC DEVICES AND ITS COMPLICATION seminar..pptx
INTRACARDIAC DEVICES AND ITS COMPLICATION seminar..pptxINTRACARDIAC DEVICES AND ITS COMPLICATION seminar..pptx
INTRACARDIAC DEVICES AND ITS COMPLICATION seminar..pptx
 
Pacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillatorsPacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillators
 
Basics of EKG Interpretation.ppt
Basics of EKG Interpretation.pptBasics of EKG Interpretation.ppt
Basics of EKG Interpretation.ppt
 
Basics of Electrocardiography, Arrhythmia & Pacemaker
Basics of Electrocardiography, Arrhythmia & PacemakerBasics of Electrocardiography, Arrhythmia & Pacemaker
Basics of Electrocardiography, Arrhythmia & Pacemaker
 
Pacemaker - EC8073 Medical Electronics - Hints for Slow Learner
Pacemaker - EC8073 Medical Electronics - Hints for Slow LearnerPacemaker - EC8073 Medical Electronics - Hints for Slow Learner
Pacemaker - EC8073 Medical Electronics - Hints for Slow Learner
 
basicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.pptbasicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.ppt
 
Anaesthesia for patient with pacemaker
Anaesthesia for patient with pacemakerAnaesthesia for patient with pacemaker
Anaesthesia for patient with pacemaker
 
Cardiac Arrest Rhythm.pdf
Cardiac Arrest Rhythm.pdfCardiac Arrest Rhythm.pdf
Cardiac Arrest Rhythm.pdf
 
Ep diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrtEp diagnosis and ablation of avnrt
Ep diagnosis and ablation of avnrt
 
Anesthesia and cardiac pacemaker
Anesthesia  and   cardiac pacemakerAnesthesia  and   cardiac pacemaker
Anesthesia and cardiac pacemaker
 
ECG
ECGECG
ECG
 
temporary pacemaker easy to learn for PICU
temporary pacemaker easy to learn for PICUtemporary pacemaker easy to learn for PICU
temporary pacemaker easy to learn for PICU
 

Mais de Ramachandra Barik

Intensive care of congenital heart disease.pptx
Intensive care of congenital heart disease.pptxIntensive care of congenital heart disease.pptx
Intensive care of congenital heart disease.pptxRamachandra Barik
 
Management of Hypetension.pptx
Management of Hypetension.pptxManagement of Hypetension.pptx
Management of Hypetension.pptxRamachandra Barik
 
CRISPR and cardiovascular diseases.pdf
CRISPR and cardiovascular diseases.pdfCRISPR and cardiovascular diseases.pdf
CRISPR and cardiovascular diseases.pdfRamachandra Barik
 
Pacemaker Pocket Infection After Splenectomy
Pacemaker Pocket Infection After SplenectomyPacemaker Pocket Infection After Splenectomy
Pacemaker Pocket Infection After SplenectomyRamachandra Barik
 
A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large D...
A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large D...A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large D...
A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large D...Ramachandra Barik
 
Trio of Rheumatic Mitral Stenosis, Right Posterior Septal Accessory Pathway a...
Trio of Rheumatic Mitral Stenosis, Right Posterior Septal Accessory Pathway a...Trio of Rheumatic Mitral Stenosis, Right Posterior Septal Accessory Pathway a...
Trio of Rheumatic Mitral Stenosis, Right Posterior Septal Accessory Pathway a...Ramachandra Barik
 
Anticoagulation therapy during pregnancy
Anticoagulation therapy during pregnancyAnticoagulation therapy during pregnancy
Anticoagulation therapy during pregnancyRamachandra Barik
 
Intracoronary optical coherence tomography
Intracoronary optical coherence tomographyIntracoronary optical coherence tomography
Intracoronary optical coherence tomographyRamachandra Barik
 
A roadmap for the human development
A roadmap for the human developmentA roadmap for the human development
A roadmap for the human developmentRamachandra Barik
 
Left ventricular false tendons
Left ventricular false tendonsLeft ventricular false tendons
Left ventricular false tendonsRamachandra Barik
 

Mais de Ramachandra Barik (20)

Willens's syndrome.pptx
Willens's syndrome.pptxWillens's syndrome.pptx
Willens's syndrome.pptx
 
Intensive care of congenital heart disease.pptx
Intensive care of congenital heart disease.pptxIntensive care of congenital heart disease.pptx
Intensive care of congenital heart disease.pptx
 
Management of Hypetension.pptx
Management of Hypetension.pptxManagement of Hypetension.pptx
Management of Hypetension.pptx
 
CRISPR and cardiovascular diseases.pdf
CRISPR and cardiovascular diseases.pdfCRISPR and cardiovascular diseases.pdf
CRISPR and cardiovascular diseases.pdf
 
Pacemaker Pocket Infection After Splenectomy
Pacemaker Pocket Infection After SplenectomyPacemaker Pocket Infection After Splenectomy
Pacemaker Pocket Infection After Splenectomy
 
Piccolo Duct Occluder.pdf
Piccolo Duct Occluder.pdfPiccolo Duct Occluder.pdf
Piccolo Duct Occluder.pdf
 
MISPLACED ECG LEADS.pptx
MISPLACED ECG LEADS.pptxMISPLACED ECG LEADS.pptx
MISPLACED ECG LEADS.pptx
 
A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large D...
A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large D...A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large D...
A Case of Device Closure of an Eccentric Atrial Septal Defect Using a Large D...
 
Arrythmia-IV.pptx
Arrythmia-IV.pptxArrythmia-IV.pptx
Arrythmia-IV.pptx
 
Arrythmia-III.pptx
Arrythmia-III.pptxArrythmia-III.pptx
Arrythmia-III.pptx
 
Arrythmia-II.pptx
Arrythmia-II.pptxArrythmia-II.pptx
Arrythmia-II.pptx
 
Arrythmia-I.pptx
Arrythmia-I.pptxArrythmia-I.pptx
Arrythmia-I.pptx
 
Trio of Rheumatic Mitral Stenosis, Right Posterior Septal Accessory Pathway a...
Trio of Rheumatic Mitral Stenosis, Right Posterior Septal Accessory Pathway a...Trio of Rheumatic Mitral Stenosis, Right Posterior Septal Accessory Pathway a...
Trio of Rheumatic Mitral Stenosis, Right Posterior Septal Accessory Pathway a...
 
Anticoagulation therapy during pregnancy
Anticoagulation therapy during pregnancyAnticoagulation therapy during pregnancy
Anticoagulation therapy during pregnancy
 
Coronary guidewire
Coronary guidewireCoronary guidewire
Coronary guidewire
 
Intracoronary optical coherence tomography
Intracoronary optical coherence tomographyIntracoronary optical coherence tomography
Intracoronary optical coherence tomography
 
Brugada syndrome
Brugada syndromeBrugada syndrome
Brugada syndrome
 
A roadmap for the human development
A roadmap for the human developmentA roadmap for the human development
A roadmap for the human development
 
Intra aortic balloon pump
Intra aortic balloon pumpIntra aortic balloon pump
Intra aortic balloon pump
 
Left ventricular false tendons
Left ventricular false tendonsLeft ventricular false tendons
Left ventricular false tendons
 

Electrical testing of pacemaker

  • 1. Electrical Testing Of Pacemaker & Pacemaker Complications Msn Pavan Kumar,DM,NIMS.
  • 2. Electrical Testing Of Pacemaker Pacemaker complications
  • 3. Electrical Testing Of Pacemaker  Pacemaker components  Battery  Pacing impedance  Pulse generator 1. Output circuit 2. Sensing circuit 3. Timing circuit 4. Rate adaptive sensor 5. Modes and mode switching
  • 4. Electrical Testing Of Pacemaker  Battery : Lithium iodine battery  High energy density ,  Long shelf life ,  Predictable loss of battery BOL (vol) – 2.8v  BOL (res) - <1komhs
  • 5. Electrical Testing Of Pacemaker  Pacing impedance :  Pacing impedance refers to the opposition to current flow. Three sources contribute to pacing impedance: 1. Pacing lead conductor coil 2. Electrode-tissue interface  Electrode resistance  Polarization  Normal lead impedance vary from 250-1200ohms.  Single impedance value may be of little use with out previous values for comparison.
  • 6. Electrical Testing Of Pacemaker 1.Pulse generator output circuit  Capture threshold , Pacing threshold , stimulation threshold  Minimum amount of energy required to constantly cause depolarization  Volts and pulse duration
  • 7. Electrical Testing Of Pacemaker 1. Pulse generator output circuit 2.0 v 1.5 v 1v
  • 8. Electrical Testing Of Pacemaker 1. Pulse generator output circuit Site At implantation Acute Chronic Atrium <1.5mv 3-5 times Twice the threshold Threshold voltage voltage Ventricle <1mv With PW 0.5ms With PW of 0.5ms
  • 9. Electrical Testing Of Pacemaker 1.Pulse generator output circuit  High Pacemaker Output can cause Reduce longevity Diaphragmatic stimulation Muscle Sti. in Unipolar pacemakers Patient may “feel” heart beat  Algorithm for checking pacemaker output threshold every beat and maintaining threshold just above it - Auto capture.
  • 10. Electrical Testing Of Pacemaker 2.Pulse generator sensing circuit :  Ability of the device to detect intrinsic beat of the heart  Measured - peak to peak magnitude (mv) & slew rate(mv/ms)
  • 11. Electrical Testing Of Pacemaker 2. Pulse generator sensing circuit :  Reduce Lower Rate below intrinsic rate to inhibit pacing and ensure intrinsic activity  Increase sensitivity setting while observing EGM. The sensitivity value at which sensing is lost on the EGM is the sensing threshold.  Sensitivity threshold safety is twice the attained valve. Sensitivity Slew rate Atrium 1-2mv(0.5mv) > 0.5 v/s ventricle 2-3mv > 0.75 v/s
  • 12. Electrical Testing Of Pacemaker 3. Pulse generator timing circuit : a. Lower rate limit (LRL) b. Hysteresis rate c. Refractory and blanking periods d. Ventricular safety pacing interval . e. Upper rate response .
  • 13. Electrical Testing Of Pacemaker 3.Pulse generator timing circuit :  Lower rate interval - lowest rate that the pacemaker will pace .  A paced or non-refractory sensed event restarts the rate timer at the programmed rate.
  • 14. Electrical Testing Of Pacemaker 3. Pulse generator timing circuit : Condition LRL (beats/mt) Infrequent pauses 40-50 Chronic persistent bradycardia 60-70 Relative bradycardia detrimental (long QT) 70-80 Detrimental fast heart rates (angina) 50-60 VVI 60-70
  • 15. Electrical Testing Of Pacemaker 3. Pulse generator timing circuit :  Hysteresis :  Hysteresis allows the rate to drop below the programmed pacing LRL.  Advantages of hysteresis : 1. Encourages native rhythm – maintain AV sync in VVI , prolong battery life 2. Prevent retrograde conduction – avoids pacemaker syndrome
  • 16. Electrical Testing Of Pacemaker 3. Pulse generator timing circuit :  AV delay (AVI) – pacemaker equivalent of PR interval.  Sensed vs paced AVI – paced AVI is programmed at 125-200ms , sensed AV interval is programmed at 20-50ms shorter than paced.  Dynamic AV delay allow pacemaker to respond to exercise sAVI – 150ms pAVI – 200ms
  • 17. Electrical Testing Of Pacemaker 3. Pulse generator timing circuit :  AV delay (AVI)  Longer AVI :  Good AV conduction – maintains AV synchrony , long battery life  Achieved by following methods :Programming longer AVI , managed ventricular pacing , AV delay hysteresis .  Shorter AVI:  HOCM – RV apical pacing decreases HOCM gradient  CRT – usually 80-120ms , for 100%ventricular pacing and optimize CO  Physiological response to faster heart rates can be answered
  • 18. Electrical Testing Of Pacemaker 3. Pulse generator timing circuit :  Refractory and blanking periods :  Refractory period – sensing present but no action  Blanking period - sensing absent and hence no action
  • 19. Electrical Testing Of Pacemaker 3. Pulse generator timing circuit :  Blanking periods : Blanking period Time Importance Atrial blanking period 50-100ms Non programmable , Avoid atrial sensing of its own paced beat Post ventricular atrial 220ms Avoid sensing of ventricular beat blanking period Long PVAB decreases detection of AF,AFL Ventricular blanking 50-100ms Non programmable, period Avoid ventricular sensing of its paced beat Post atrial ventricular 28ms if the PAvB period is too long, R on T - blanking period ventricular tachyarrhythmia.
  • 20. Electrical Testing Of Pacemaker 3. Pulse generator timing circuit :  Refractory period: Refractory period Importance Ventricular refractory period (VRP) Prevent sensing of T wave . Atrial refractory period (ARP) AVI (120-200ms) . Post ventricular atrial refractory period Avoid sensing retrograde P waves (PMT) , far field R waves .
  • 21. Electrical Testing Of Pacemaker 3. Pulse generator timing circuit : Ventricular safety pacing/ventricular triggered period/cross talk sensing window : Atrial pacing in DDD Trigger ventricular sensing PAVB - pAVI False inhibition of ventricular pacing circuit Asystole
  • 22. Electrical Testing Of Pacemaker 4. Pulse generator rate responsive pacing:  Rate responsive pacing refer to ability of pacemaker to increase its lower rate in response to physiological stimulus  Sinus node dysfunction , AF patients – fail to increase heart rates  HRR should start with in 10s of exercise , peak at 90 – 120s and should return to baseline with in 60 – 120s after exercise.  Fastest rate at which pacemaker will pace upper rate response.  If intrinsic atrial rate exceeds URR then wenckebach or 2:1 AVB  Choosing URR : young patients (150b/mt) , old angina (<110b/mt).  Various sensors (activity , minute ventilation , QT)
  • 23. Electrical Testing Of Pacemaker 4. Pulse generator rate responsive pacing: Wenckebach 2:1 AVB
  • 24. Electrical Testing Of Pacemaker 4. Pulse generator modes:
  • 25. Electrical Testing Of Pacemaker 5. Pulse generator modes switching: DDD / VDD Atrial tachyarrythmias Sensed atrial events DDIR / Trigger fast ventricular rates VVIR Palpitations. Dyspnoea. And Fatigue.
  • 26. Electrical Testing Of Pacemaker 5. Pulse generator modes switching: Programming mode switching Mode switching occurs when the sensed atrial rate exceeds a programmed atrial tachycardia detection rate. By definition, this value must be faster than the URL (maximum tracking rate). Atrial tachycardia detection is typically programmed to 175-l88bpm or thereabouts.AMS base rate is higher than LRI.
  • 27. Electrical Testing Of Pacemaker Pacemaker follow up guidelines: Transtelephonic monitoring guidelines Medicare guidelines Single chambered pacing Dual chambered pacing 1st month q 2 week 1st month q 2 weeks 2nd -48th month q 12 week 2nd – 30th month q 12weeks 49th – 72nd month q 8 week 31st – 48th month q 8 weeks 73rd month and later q 4 weeks 49th month and later q 4week NASPE guidelines Single or dual pacing 1st visit 6 – 8 week post implant , if symptomatic prior to this 5th month From 6th month q 3month Battery wear present q 1month
  • 28. Electrical Testing Of Pacemaker Pacemaker complications
  • 29. Pacemaker complications Pacemaker complications Pocket complications Lead dislodgement Pocket hematoma Pneumothorax /air embolism Infection Cardiac perforation Erosion Extracardiac stimulation Wound pain Venous thrombosis Allergic reactions Coronary sinus dissection Twidller syndrome Pacemaker malfunction
  • 30. Pacemaker complications Pocket hematoma :  The risk of haematoma is increased in patients taking antithrombotic or anticoagulant drugs (Goldstein et al., 1998).  Most small hematomas can be managed conservatively with cold compress and withdrawal of antiplatelet or antithrombotic agents.  Occasionally, large hematomas that compromise the suture line or skin integrity may have to be surgically evacuated.  Needle aspiration increases risk of infection and should not be done.
  • 31. Pacemaker complications Pocket hematoma :  In patients requiring oral anticoagulants (warfarin), to take INR of about 2.0 at the time of implantation is safe (Belott & Reynolds, 2000).  Unfractionated heparin or low-molecular-weight heparin are always discontinued prior to device implant and ideally avoided for a minimum of 24 hours post implantation.  Administration of anticoagulants can be resumed within 48-72 h after implantation if there is no evidence of substantial hematoma formation.
  • 32. Pacemaker complications Device-related infections :  The reported incidence of pacemaker-related infection ranges from 0.5% to 6% in early series  The use of prophylactic antibiotics and pocket irrigation with antibiotic solutions has decreased the rate of acute infections following pacemaker implantations to <1 to 2 percent in most series  The mortality of persistent infection when infected leads are not removed can be as high as 66%.  DM, malignancy, operator inexperience, advanced age, corticosteroid use, anticoagulation, recent device manipulation, CRF, and bacteremia from a distant focus of infection.
  • 33. Pacemaker complications Device related infection :  Device infection is defined as either:  (a) deep infection - infection involving the generator pocket and/or the intravenous portion of the leads, with bacteremia, requiring device extraction or  (b) superficial infection - characterized by local inflammation, involving the skin but not the generator pocket, and treated with oral antibiotics.
  • 34. Pacemaker complications Device related infection : 2007;49;1851-1859 J. Am. Coll. Cardiol.
  • 35. Pacemaker complications Device related infection : 2007;49;1851-1859 J. Am. Coll. Cardiol.
  • 36. Pacemaker complications Device related infection : 2007;49;1851-1859 J. Am. Coll. Cardiol.
  • 37. Pacemaker complications Wound pain :  Infection , Pacemaker implanted too superficially , Pacemaker implanted too laterally , Pacemaker allergy . Skin erosion :  Incidence has been estimated around 0.8% .Old age , infection.  Surgical revision of pocket and reimplantation . Allergic reactions :  Always rule out infection before coming to diagnosis of allergy
  • 38. Pacemaker complications Lead dislodgement:  Relatively common – 5-10% of patients(ICD database 2001)  Atrial more common than ventricular(2-3% vs. 1%)  Micro dislodgement , macro dislodgement  Increased pacing threshold , failure to pace and sense  Active fixation (decreases risk)
  • 39. Pacemaker complications Pneumothorax , :  Uncommon complication – 1.6-2.6%  During or 48 hrs after procedure  Inadvent puncture and laceration of subclavian vein , artery or lung  Related to operator experience and underlying anatomy  Avoided by 1. Venogram – flouroscpic puncture 2. Axillary venous access (Martin etal’96) 3. One way mechanism sheath
  • 40. Pacemaker complications Cardiac Perforation :  Uncommon but potentially serious complication - lower than 1%.  Acute (<5 days) , subacute(5d-1month) , chronic (>1month)  Increasing stimulation threshold , RBBB pattern for RV pacing, intercostal muscle or diaphragmatic contraction, friction rub, and pericarditis, pericardial effusion, or cardiac tamponade.  CXR , ECHO , CT  Lead withdrawal and repositioning ; surgical back up
  • 42. Pacemaker complications Extracardiac stimulation  The diaphragm or pectoral or intercostal muscles  Diaphragmatic stimulation - direct stimulation of the diaphragm (left) or stimulation of the phrenic nerve (right).  Early postimplantation period , dislodgment of the pacing lead.  MC in patients with LV coronary vein branch lead placement for CRT  Output pacing importance (testing and treatment)  Pectoral stimulation - incorrect orientation of the pacemaker or a current leak from a lead insulation failure or exposed connector.
  • 43. Pacemaker complications Venous thrombosis :  Venous thrombosis occurs in 30% to 50% of patients and only 1-3% of patients become symptomatic.  Manifestations vary from usually asymptomatic, acute symptomatic thrombosis, and even SVCS .  Early or late after pacemaker implantation.  Predictors of severe stenosis are multiple pacemaker leads , previous pacing , double coils , hormone therapy .  Asymptomatic (no treatment) , symptomatic (anticoagulants – endovascular stents – surgical correction ).
  • 44. Pacemaker complications Twiddler syndrome: Obese women with loose, fatty subcutaneous tissue Small size of the implanted generator with a large pocket Twisting of pulse generator in long axis Lead dislodgement and lead fracture Failure to capture
  • 45. Pacemaker complications Twiddler syndrome:  The prevelance of this syndrome is 0.07% (Gungor et al., 2009)  Rotated along the transverse axis it is referred by us as the reel syndrome.  Pocket should be revised.  Avoid by Limit the pocket size,  Suture the device to the fascia  The patients not to manipulate their device pocket
  • 46. Pacemaker malfunction  Failure to capture  Failure to output  Sensing abnormalities(under and over sensing)  Specific mode complications 1. Pacemaker related tachycardia 2. Pacemaker syndrome
  • 47. Pacemaker malfunction Failure to capture:  Pacing artifact present but no evoked potential .  Causes 1. Lead dislodgement or perforation 2. Lead maturation(inflammation/fibrosis)(exit block) 3. Battery depletion 4. Circuit failure(coil fracture , insulation defect) 5. Capture management algorithm failure 6. Inappropriate programming 7. Pseudo malfunction 8. Functional non capture 9. Metabolic , drugs , cardiomyopathies
  • 48. Pacemaker malfunction Failure to capture: Electrocardiographic tracing from a patient with a DDDR pacemaker. All ventricular pacing artifacts but one failed to result in ventricular depolarization— that is, failure to capture
  • 49. Pacemaker malfunction Failure to capture: Pacing threshold Normal Increased Dislodgement Normal Exit block Battery depletion Functional non capture Impedance Insulation Decreased failure/break Lead fracture Increased Loose screw
  • 50. Pacemaker malfunction Failure to output:  Absence of pacing stimuli and hence no capture .  Causes 1. Pseudo malfunction - hysteresis , PMT termination , sleep rate 2. Over sensing - EMI ; T P R over sensing ; Myopotential/diaphragmatic ; Cross talk ; Make break signals 3. Open circuit - lead fracture , loose screw , air in the pocket , incompatible lead . 4. Battery depletion 5. Recording artifact.
  • 51. Pacemaker malfunction  Failure to output: VVIR pacemaker This patient had a pacemaker programmed to a unipolar sensing configuration. The sensing of myopotentials led to symptomatic pauses, and reprogramming the pacemaker to a bipolar sensing configuration prevented subsequent myopotential over sensing.
  • 52. Pacemaker malfunction Failure to output: Application of magnet Eliminates pauses Pauses persistent Over sensing Normal Battery depletion Pseudo malfunction Impedance Insulation Decreased failure/break Lead fracture Increased Loose screw
  • 53. Pacemaker malfunction Battery depletion :  Elective replacement indicators (ERI) 1. Low voltage(2.1-2.4) 2. Low pacing rate on magnet application 3. Elevated battery impedance 4. Increased pulse width duration 5. Restricted programmability 6. Change to simpler pacing mode  End of life (EOL) 1. Low voltage(≤2.1vol)
  • 54. Pacemaker malfunction Pacemaker undersensing :  Pacing artifact present but no sensing(sensed beat doesn’t reset cycle)  Causes are 1. Defect in signal production – scar /fibrosis , BBB , ectopic , cardioversion , defibrillation , metabolic. 2. Defect in signal transmission – lead fracture /dislodgement , insulation failure , partial open circuit. 3. Defect in pacemaker – battery depletion , sensing circuit abnormalities , committed DVI.
  • 56. Pacemaker malfunction Pacemaker over sensing : Cross talk :  Present as failure to pace High atrial output  Causes High ventricular sensitivity 1. EMI Low VBP 2. T , P , R over sensing . 3. Cross talk Ventricular sensing of 4. Myopotential (unipolar) paced atrial impulse 5. Make break signals Pts with Poor AV conduction – Ventricular Asystole
  • 57. Pacemaker malfunction  Electromagnetic interference : Source Pacer Inhibition Rate Asynchronous Uni/ damage increase noise bipolar Cardioversion/ Y N N N U/B Defibrillation Anit theft devices / N Y N N U Weapon detector Phone (cell/cordless) N Y Y Y U/B Ablation Y Y Y N U/B Diathermy/ Y Y Y Y U/B lithotripsy FM radio N Y N Y U TV transmitter MRI/PET Y Y(N) Y(N) Y(N) U/B
  • 58. Pacemaker malfunction Pacemaker syndrome :  Seen in 20% of PPI (5% severe symptomatic)  VVI/DDD/AAI  Pulsations in neck , fatigue , cough ,chest fullness , headache , chocking sensation , PND , confusion , syncope , pulmonary edema.  Rx : VVI – program hysteresis , or change to DDD ; DDD –atrial lead reprogrammed or changed
  • 59. Pacemaker malfunction Pacemaker mediated tachycardia :  Dual chamber  VPC , intact retrograde conduction , PVARP<VA .  Px , Rx : 1. PVARP > VA 2. Long PVARP after VPC 3. Absent atrial sensing after VPC
  • 64. Hey:Share with me please.