Subjective refraction  By                                                praveen
Subjective refraction technique rely on the patient’s response to obtain the refractive correction that gives the best visual acuity . If all the refractive errors were simply spherical ,subjective refraction will be easy. Determining the astigmatic portion of the correction is more complex .
Different subjective refraction techniques are employed .09323330039 paul
Cross-cylinder technique[jackson cross cylinder ]  First step in cross cylinder refraction is adjusting the sphere to yield best visual acuity without accommodation . Place the prescription the patient is wearing ,if any ,into a trial frame or phoropter . Fog the eye to be examined with plus sphere while the patient views a visual acuity chart. Then decrease the fog until best visual acuity is obtained
If no cylindrical correction is present initially ,the cross cylinder may still be used ,placed arbitrarily at 90 0 and 180 0 . To check the presence of astigmatism . If  a preferred flip position is found ,cylinder  is added with axis parallel to the respective plus or minus axis of the cross cylinder until the two flip choices are equal .
If no preference is found with the cross- cylinder axis at 90 0 and 180 0 ,45 0 and 135 0  should always be checked before assuming that no astigmatism is present . Once any cylinder  power is found , axis and power are refined in the usual manner .
Cylinder axis is always refined first . This sequence is necessary  because the correct axis can be found in the presence of an in correct power . But the full cylinder  power will not be found in the presence of an incorrect axis .
To refine the axis place the cross cylinder with its principal meridians 45 0 away from the principal meridians of the correcting cylinder . Present the patient with alternative flip choices and inquiring which is” blackest and sharpest “. Rotate the axis of the correcting cylinder toward the corresponding plus or minus axis of the cross cylinder .
Low-power cylinder are rotated in large steps ,high powered cylinder are rotated in small steps ,repeat the procedure until the flip choices appear equal . To refine cylinder power ,align the cross – cylinder axis with the principal meridians of the correcting lens . As the examiner changes cylinder  power according to the patient’s response . The spherical equivalent of the refractive correction should remain constant .
That is for every 0.50 D of cylinder power change ,the sphere is changed 0.25 D in the opposite direction . If the cylinder  power is changed a large amount ,the sphere power should be readjusted for best visual acuity . Continue to refine cylinder power until choices appear “about the same “to the patient ,that will be the correct endpoint .
Remember always to use the proper cross cylinder for the patient’s visual acuity level . For eg :- +- 0.25 D cross cylinder is commonly used with visual acuity levels of 20/30 and better . A high cross cylinder [+- 0.50 D or +- 1.00 D ] should be used with poorer vision .
Following steps are used in cross –cylinder refraction . Adjust sphere to the most plus or least minus that gives best visual acuity . Use test figure one or two lines larger than the patient’s BCVA , because the introduction of the JCC produce blur . If cylinder correction is not already present ,look for astigmatism .
Refine axis first  Refine cylinder power . Refine sphere ,cylinder axis ,and cylinder power .
Tricking   Avoid confusion with previous choices by giving different number to subsequent choices  Which is better ,one or two ,three or four? If the patient persists in always choosing either the first or the second number ,reverse the order .
Clock Dial  When the clock dial is presented to the patient under sufficient fog . The examiner’s first job is to determine the axis of the correcting cylinder . This is done by first asking the patient if he or she can see three lines in any or all of the spokes . And then ask to report in which of the spokes the three lines are the sharpest or most distinct .
Most clock dial charts equipped with numbers similar to those on the face of a clock . Expected  response of the patient is that the spoke from 12 to 6 ,from 1 to 7 ,from 2 to 8 o’clock is most distinct . To determine the axis of the correcting cylinder ,the smaller of the two number reported by the is multiplied by 30.
For example ,if the patient reports that the 12 t 6 o’clock spoke  is most distinct the examiner would place the axis of the correcting cylinder at 180 degrees. The examiner then begins adding minus cylinder power ,0.25 D at a time . Questioning the patient each time as to the relative sharpness of the lines in the spokes representing the two principal meridians of the eye .
If sufficient minus cylinder power is added ,the horizontal and vertical focal lines will be located in the same plane . Then the patient will report that the 12 to 6 and the 3-to -9 spokes are equally distinct . This procedure is referred to as “collapsing the conoid of sturm “ the horizontal and vertical focal line are replaced by a point image .
Monocular  subjective refraction  Monocular subjective refraction consists of the following procedure  Determining the cylindrical correction under fog.  Refining the cylindrical correction without fog . Determining the spherical end point . Following these procedure ,one or more binocular balancing tests are performed .
Determining the cylinder under fog  Suitable fogging lenses will have to be put into place before beginning the monocular  subjective refraction . The left eye is occluded ,a block of letter is introduced at the 6-m distance . Plus power in front of the right eye is reduced ,0.25 at a time until the patient can read all of the 20 /40 letters . Ensures that the eye is fogged by approximately 1.00 D ,so the entire conoid of sturm is in front of the retina .
The astigmatic chart is presented ,and the axis of the correcting cylinder is determined . If clock dial is used ,the patient is asked first whether three lines can be seen in any or all of the spokes . If he can see then ask the patient to report in which of the spokes the three lines are the most distinct .
Minus cylinder is added to the indicated axis 0.25 D at a time until equality is obtained . Additional minus cylinder power is then added 0.25 D at a time to obtain a reversal . Then the examiner returning to the lowest-power cylinder that brings about equality .
The right is then occluded and the procedure are repeated for the left eye .
Refining the cylinder without fog  On completion of the astigmatic chart test . The right eye is defogged ,reducing plus lens power 0.25 D at a time ,until the best visual acuity line [20/20 or 20/15 ] is reads. When the removal of additional plus power [or addition of minus power ]fails to improve visual acuity ,the spherical lens power is returned to the maximum plus or minimum minus power that resulted in the best visual acuity for that eye .
Beginning with the cylinder power and axis found with the astigmatic chart under fog . The practitioner first refines cylinder axis and then cylinder power .
Crossed –cylinder test for Axis
Crossed –cylinder Test for power
Additional cylinder check tests  Additional check test for cylinder power . Many patient tended to reject cylinder power in the crossed- cylinder test even though it has been evident both in retinoscopy and in the astigmatic chart test . This occurs ,the patient is asked to watch the smallest readable row of letters .
An additional -0.25 D cylinder is placed in the refractor ,and the patient is asked to report which of the two views appears to be more distinct . If the additional -0.25 D cylinder does not improve the clarity of the letters ,the original cylinder power is left in the refractor .
If the additional cylinder power causes the letters to be more distinct ,the examiner can choose to leave the additional cylindrical power in the refractor.
Additional check test of axis  The additional cylinder check test for axis is sometime referred to as bracketing . It is useful mainly when the power of the correcting cylinder is 1.00 D or more . The patient is asked to view a row of 20/20 or 20/15 letter  And ask to report when the cylinder lens in the refractor is slowly rotated.
When the blur is reported ,the change in the axis is noted mentally . And the test is repeated with the same instructions . For example : if the original correcting cylinder axis is located at 180 degrees . The patient reports a blur at 15 degrees and blur at 165 degrees ,then the examiner can assume that the orginal axis was correct .
Determining the spherical end point  Each eye is fogged to 0.75 or 1.00 D and defogged to best acuity. Maximum plus power of best visual acuity . Examiner should present a block of letters extending from 20/40 to 20/15. We should make a mental note of the patient’s acuity for each eye through the +0.75 or +1.00 D fogging lens prior to defogging .
It is important that accommodation be relaxed in arriving at the monocular end point . For example : +1.00 D fog may blur the right eye to 20/30 ,but it may blur the left eye only to 20/20 . This evidence that the left eye is underplussed [or over minused ] ,and the subjective end point for both eye should be determined again .
Patient instruction  To determine the monocular endpoint correct patient instruction are important . The examiner should understand that the patient’s subjective evaluation of the clarity or distinctness of the letters is not the important consideration . The important ,overriding consideration is the ability of the patient to resolve the letters .
As plus power is decreased 0.25 D at a time . When the point has been reached where an additional decrease in plus power of 0.25 D does not make any more letter readable. On the other hand ,if the examiner allows the patient to respond in terms of clarity of the letter or in terms of which lens is preferred .
Many young patient will continue to accommodate with each 0.25 D of reduction in plus ,with the result that the end point will be completely invalid. When minus lens power is added to the point that accommodation is necessary to keep the letters in sharp focus on the retina  Many patient will notice that the letter appear to be smaller .
This is purely optical effect ,known as accommodative micropsia . it is right to ask the patient if the letter looks smaller to avoid overminusing the patient . If patient report that the letter are not smaller should not  be taken as evidence .
Binocular balance  The purpose of binocular balancing tests is not to balance the visual acuity but to balance the state of accommodation of the two eyes . If the corrected visual acuity is same in both eye ,the balancing procedure may consist of comparison of the visual acuity for the two eyes . If the correcting visual acuity is not same in both eyes ,then a method not based on visual acuity must be used .
Procedure used for balancing the state of accommodation for the two eyes are often referred to as equalization tests or as binocular balancing tests .
Balancing with little or no fog  On completion of the monocular subjective refraction ,the lenses are left in the refractor . The occluders has to remove from both eyes . The patient’s attention is called to a block of letters at 6 m. Plus lens power is added in front of both eyes until the 20/20 letter are blurred but the 20/25 letter are easily resolved .
This normally requires an increase in plus or a decrease in minus of 0.25 to 0.50 D. Then ask to compare the clarity of the 20/25 letters for the two eyes . Using either prism dissociation or alternative occlusion .
Prism dissociation  The examiner  place 3      of base –down prism in front of the right eye and 3     of base-up prism in front of the left eye . The patient will see two charts separated –vertically . The upper chart seen by right eye . Patient attention is called to the 20 /25 letters .
Then asked to report whether the letter are more distinct or easier to read in the upper chart or the lower chart . If the equally distinct for the two eyes, the accommodative state of the two eyes is considered to be balanced ,and the test is over .
If the patient reports a difference in clarity of the letters for the two eyes. 0.25 D is added in front of the eye with the better vision and the test is repeated . Often the patient fails to report equal clarity for any lens combination . This problem can be solved by instructing the patient to report “which of the lens combination causes the upper and lower charts to be more nearly equal .
Another way of resolving this difficulty is to give better acuity to the dominant eye . Once the patient’s acuity is balanced at 20/25 the patient is defogged binocularly.
Alternate occlusion  Patient is instructed to compare alternate views of the chart while the each eye is alternately occluded. Occluder in the refractor can be used for occlusion . Fogging to 20/25 . Adding 0.25 D to the better eye ,are done in the same manner of prism dissociation test .
Check Test for the Binocular End point  Once the examiner is satisfied with the binocular balance ,binocular end point can be verified . Patient’s attention is called to the 20 /20 letter . Add +0.25 D sphere to both eye and ask for any difference in the clarity of the letters.
The same question is again  asked after adding 0.25 D sphere . Again third +0.25 D sphere is added ,ask of the difference in the clarity . Expected response are that the 20/20 letters will be “slightly blurred “. With the first 0.25 D of plus . “badly blurred “ with the second 0.25D. “blurred  out “ with the third 0.25 D .
Bichrome test  As a monocular End –point test  The bichrome  test must be done in an almost completely darkened room . Starting with the result of monocular subjective finding . +0.50D or + 0.75 D of spherical power is placed in front of each eye. The red and green filter is placed on the projector.
The patient is asked to report which of the letters ,those on the red background or those in the green background –are “sharper blacker ,or more distinct “. The patient is expected to report that the letters on the red background are more distinct than those on the green background .
As the plus power is reduced 0.25 D at a time . At some point the patient should report that the letter on  red and green background are equally distinct . As plus  power is further reduced ,the patient should say those on the green side are more distinct than the red side .
If the original monocular subjective endpoint was correct ,the patient will typically report that the red letters are more distinct with +0.75,+0.50,+0.25 D fog . When all the fog is removed the letter will appear to be equally distinct on both the red green sides .
Prescribing for children  Prescribing visual correction for children often has two goals : Providing a focused retinal image . Achieving the optimal balance between accommodation and convergence . In some cases subjective refraction is impossible because child’s  inability to cooperate. The optimal refraction in an infant  requires the paralysis of accommodation with complete cycloplegia .
The presence of strabismus may modify normal prescribing guidelines .
Myopia [general guidelines for correction of significant childhood myopia] Cycloplegic refraction are mandatory . Full refractive error ,including cylinder ,should be corrected .young children tolerate cylinder well. Some will under correct myopia ,other may even use bifocals with or without atropine ,on the theory that prolonged accommodation hastens .
Intentional undercorrection of a myopicaesotrope to decrease the angle of deviation is rarely tolerated . Indentional overcorrection of a myopic error can be of some value in controlling an intermittent  exodeviation . Parents should be educated about the natural progression of myopia and the need for frequent refraction and possible prescription changes .
Contact lenses may be desirable in older children to avoid  the problem of image minification found with high- minus lenses.
Hyperopia When hyperopia and esotropia coexist ,initial management includes full correction of the cycloplegic refractive error . In school age child ,the full refractive correction may cause blurring of distance vision because of the inability to relax accommodation fully . The amount of correction may have to be reduced for the child to accept the glasses.

Mais conteúdo relacionado

Mais procurados

Soft toric Contact Lens
Soft toric Contact LensSoft toric Contact Lens
Soft toric Contact LensManish Dahal
 
Ophthalmic dispensing
Ophthalmic dispensingOphthalmic dispensing
Ophthalmic dispensingKewal Hirwani
 
Binocular refraction techniques, binocular balancing & binocular
Binocular refraction techniques, binocular balancing & binocularBinocular refraction techniques, binocular balancing & binocular
Binocular refraction techniques, binocular balancing & binocularsabina paudel
 
Static retinoscopy by Neha D. Jadhav B.Optometry 2nd yr
Static retinoscopy by Neha D. Jadhav B.Optometry 2nd yrStatic retinoscopy by Neha D. Jadhav B.Optometry 2nd yr
Static retinoscopy by Neha D. Jadhav B.Optometry 2nd yrNeha Jadhav
 
Dynamic retinoscopy
Dynamic retinoscopy Dynamic retinoscopy
Dynamic retinoscopy Rajeshwori
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refractionsanju_95
 
Objective Refraction and Subjective Refraction
Objective Refraction and Subjective RefractionObjective Refraction and Subjective Refraction
Objective Refraction and Subjective RefractionAnuMusyakhwo7
 
Measuring interpupillary distance
Measuring interpupillary distanceMeasuring interpupillary distance
Measuring interpupillary distanceIndra Prasad Sharma
 
Soft Toric Contact Lens
Soft Toric Contact LensSoft Toric Contact Lens
Soft Toric Contact LensTahseen Jawaid
 
Corneal Topography
Corneal TopographyCorneal Topography
Corneal TopographyRaman Gupta
 
Jackson cross cylinder
Jackson cross cylinderJackson cross cylinder
Jackson cross cylinderameen Rashid
 
DUOCHROM TEST.pptx
DUOCHROM TEST.pptxDUOCHROM TEST.pptx
DUOCHROM TEST.pptxKhulesh Sahu
 

Mais procurados (20)

Jackson cross cylinder
Jackson cross cylinderJackson cross cylinder
Jackson cross cylinder
 
Retinoscopy
RetinoscopyRetinoscopy
Retinoscopy
 
Soft toric Contact Lens
Soft toric Contact LensSoft toric Contact Lens
Soft toric Contact Lens
 
Ophthalmic dispensing
Ophthalmic dispensingOphthalmic dispensing
Ophthalmic dispensing
 
Binocular refraction techniques, binocular balancing & binocular
Binocular refraction techniques, binocular balancing & binocularBinocular refraction techniques, binocular balancing & binocular
Binocular refraction techniques, binocular balancing & binocular
 
Binocular balancing
Binocular balancing Binocular balancing
Binocular balancing
 
Static retinoscopy by Neha D. Jadhav B.Optometry 2nd yr
Static retinoscopy by Neha D. Jadhav B.Optometry 2nd yrStatic retinoscopy by Neha D. Jadhav B.Optometry 2nd yr
Static retinoscopy by Neha D. Jadhav B.Optometry 2nd yr
 
Dynamic retinoscopy
Dynamic retinoscopy Dynamic retinoscopy
Dynamic retinoscopy
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
 
Objective Refraction and Subjective Refraction
Objective Refraction and Subjective RefractionObjective Refraction and Subjective Refraction
Objective Refraction and Subjective Refraction
 
AC/A
AC/AAC/A
AC/A
 
Aberrometry
AberrometryAberrometry
Aberrometry
 
Measuring interpupillary distance
Measuring interpupillary distanceMeasuring interpupillary distance
Measuring interpupillary distance
 
Soft Toric Contact Lens
Soft Toric Contact LensSoft Toric Contact Lens
Soft Toric Contact Lens
 
Corneal Topography
Corneal TopographyCorneal Topography
Corneal Topography
 
Pediatric contact lens
Pediatric contact lensPediatric contact lens
Pediatric contact lens
 
Dynamic retinoscopy
Dynamic retinoscopyDynamic retinoscopy
Dynamic retinoscopy
 
Jackson cross cylinder
Jackson cross cylinderJackson cross cylinder
Jackson cross cylinder
 
Vergences of the eye
Vergences of the eyeVergences of the eye
Vergences of the eye
 
DUOCHROM TEST.pptx
DUOCHROM TEST.pptxDUOCHROM TEST.pptx
DUOCHROM TEST.pptx
 

Semelhante a Subjective refraction by optom praveen

Semelhante a Subjective refraction by optom praveen (20)

Astigmatic fan
 Astigmatic fan Astigmatic fan
Astigmatic fan
 
14 cross cylinder
14 cross cylinder14 cross cylinder
14 cross cylinder
 
asigmatic clock and fan .pdf
asigmatic clock and fan                .pdfasigmatic clock and fan                .pdf
asigmatic clock and fan .pdf
 
Methods of subjective refraction
Methods of subjective refractionMethods of subjective refraction
Methods of subjective refraction
 
subjective verification of refraction
subjective verification of refractionsubjective verification of refraction
subjective verification of refraction
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
 
Cross Cylinder 4th Yr.pptx
Cross Cylinder 4th Yr.pptxCross Cylinder 4th Yr.pptx
Cross Cylinder 4th Yr.pptx
 
Retinoscopy.pptx
Retinoscopy.pptxRetinoscopy.pptx
Retinoscopy.pptx
 
V551#9 prac#2review2010 [autosaved]
V551#9 prac#2review2010 [autosaved]V551#9 prac#2review2010 [autosaved]
V551#9 prac#2review2010 [autosaved]
 
JCC -Jackson Cross Cylinder
JCC -Jackson Cross CylinderJCC -Jackson Cross Cylinder
JCC -Jackson Cross Cylinder
 
retinoscopy
retinoscopyretinoscopy
retinoscopy
 
JACKSON CROSS CYLINDER
JACKSON CROSS CYLINDERJACKSON CROSS CYLINDER
JACKSON CROSS CYLINDER
 
refractive technique
refractive techniquerefractive technique
refractive technique
 
Refraction and Retinoscopy
Refraction and RetinoscopyRefraction and Retinoscopy
Refraction and Retinoscopy
 
Refraction Method by Siddhartha Khandewal ( Click below for Online Lecture)
Refraction Method by Siddhartha Khandewal ( Click below for Online  Lecture)Refraction Method by Siddhartha Khandewal ( Click below for Online  Lecture)
Refraction Method by Siddhartha Khandewal ( Click below for Online Lecture)
 
lecture(3) phoropter.ppt
lecture(3) phoropter.pptlecture(3) phoropter.ppt
lecture(3) phoropter.ppt
 
Jackson Cross Cylinder
Jackson Cross CylinderJackson Cross Cylinder
Jackson Cross Cylinder
 
Technique of Jackson cross cylinder
Technique of Jackson cross cylinderTechnique of Jackson cross cylinder
Technique of Jackson cross cylinder
 
Iol
IolIol
Iol
 
Keratometer Slides
Keratometer SlidesKeratometer Slides
Keratometer Slides
 

Último

(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...TanyaAhuja34
 
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...Rashmi Entertainment
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsMedicoseAcademics
 
❤️Panchkula Call Girls☎️9809698092☎️ Call Girl service in Panchkula☎️ Panchku...
❤️Panchkula Call Girls☎️9809698092☎️ Call Girl service in Panchkula☎️ Panchku...❤️Panchkula Call Girls☎️9809698092☎️ Call Girl service in Panchkula☎️ Panchku...
❤️Panchkula Call Girls☎️9809698092☎️ Call Girl service in Panchkula☎️ Panchku...Sheetaleventcompany
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana GuptaLifecare Centre
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowtanudubay92
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...gragneelam30
 
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...Sheetaleventcompany
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxSwetaba Besh
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Janvi Singh
 
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...Namrata Singh
 
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...Namrata Singh
 
Hyderabad Call Girls Service ❤️ 7783825323 Independent Escort Service Hyderabad
Hyderabad Call Girls Service ❤️ 7783825323 Independent Escort Service HyderabadHyderabad Call Girls Service ❤️ 7783825323 Independent Escort Service Hyderabad
Hyderabad Call Girls Service ❤️ 7783825323 Independent Escort Service HyderabadSheetaleventcompany
 
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...Sheetaleventcompany
 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...call girls hydrabad
 
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...Sheetaleventcompany
 
🍑👄Hyderabad Escorts Service☎️7783825323🍑👄 Call Girl service in Hyderabad☎️Hyd...
🍑👄Hyderabad Escorts Service☎️7783825323🍑👄 Call Girl service in Hyderabad☎️Hyd...🍑👄Hyderabad Escorts Service☎️7783825323🍑👄 Call Girl service in Hyderabad☎️Hyd...
🍑👄Hyderabad Escorts Service☎️7783825323🍑👄 Call Girl service in Hyderabad☎️Hyd...Sheetaleventcompany
 
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...chanderprakash5506
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxSwetaba Besh
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...rajnisinghkjn
 

Último (20)

(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
(RIYA)🎄Airhostess Call Girl Jaipur Call Now 8445551418 Premium Collection Of ...
 
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
❤️ Chandigarh Call Girls☎️98151-579OO☎️ Call Girl service in Chandigarh ☎️ Ch...
 
Circulatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanismsCirculatory Shock, types and stages, compensatory mechanisms
Circulatory Shock, types and stages, compensatory mechanisms
 
❤️Panchkula Call Girls☎️9809698092☎️ Call Girl service in Panchkula☎️ Panchku...
❤️Panchkula Call Girls☎️9809698092☎️ Call Girl service in Panchkula☎️ Panchku...❤️Panchkula Call Girls☎️9809698092☎️ Call Girl service in Panchkula☎️ Panchku...
❤️Panchkula Call Girls☎️9809698092☎️ Call Girl service in Panchkula☎️ Panchku...
 
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
7 steps How to prevent Thalassemia : Dr Sharda Jain & Vandana Gupta
 
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book nowChennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
Chennai ❣️ Call Girl 6378878445 Call Girls in Chennai Escort service book now
 
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
Call Girls Bangalore - 450+ Call Girl Cash Payment 💯Call Us 🔝 6378878445 🔝 💃 ...
 
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
Low Cost Call Girls Bangalore {9179660964} ❤️VVIP NISHA Call Girls in Bangalo...
 
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptxANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF REPRODUCTIVE SYSTEM.pptx
 
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
Call Girls in Lucknow Just Call 👉👉 8875999948 Top Class Call Girl Service Ava...
 
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...Kolkata Call Girls Shobhabazar  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Gir...
Kolkata Call Girls Shobhabazar 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Gir...
 
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...Kolkata Call Girls Naktala  💯Call Us 🔝 8005736733 🔝 💃  Top Class Call Girl Se...
Kolkata Call Girls Naktala 💯Call Us 🔝 8005736733 🔝 💃 Top Class Call Girl Se...
 
Hyderabad Call Girls Service ❤️ 7783825323 Independent Escort Service Hyderabad
Hyderabad Call Girls Service ❤️ 7783825323 Independent Escort Service HyderabadHyderabad Call Girls Service ❤️ 7783825323 Independent Escort Service Hyderabad
Hyderabad Call Girls Service ❤️ 7783825323 Independent Escort Service Hyderabad
 
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
💚Chandigarh Call Girls Service 💯Piya 📲🔝8868886958🔝Call Girls In Chandigarh No...
 
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
Call girls Service Phullen / 9332606886 Genuine Call girls with real Photos a...
 
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
❤️Amritsar Escorts Service☎️9815674956☎️ Call Girl service in Amritsar☎️ Amri...
 
🍑👄Hyderabad Escorts Service☎️7783825323🍑👄 Call Girl service in Hyderabad☎️Hyd...
🍑👄Hyderabad Escorts Service☎️7783825323🍑👄 Call Girl service in Hyderabad☎️Hyd...🍑👄Hyderabad Escorts Service☎️7783825323🍑👄 Call Girl service in Hyderabad☎️Hyd...
🍑👄Hyderabad Escorts Service☎️7783825323🍑👄 Call Girl service in Hyderabad☎️Hyd...
 
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
Russian Call Girls In Pune 👉 Just CALL ME: 9352988975 ✅❤️💯low cost unlimited ...
 
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF RESPIRATORY SYSTEM.pptx
 
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
👉 Chennai Sexy Aunty’s WhatsApp Number 👉📞 7427069034 👉📞 Just📲 Call Ruhi Colle...
 

Subjective refraction by optom praveen

  • 2. Subjective refraction technique rely on the patient’s response to obtain the refractive correction that gives the best visual acuity . If all the refractive errors were simply spherical ,subjective refraction will be easy. Determining the astigmatic portion of the correction is more complex .
  • 3. Different subjective refraction techniques are employed .09323330039 paul
  • 4. Cross-cylinder technique[jackson cross cylinder ] First step in cross cylinder refraction is adjusting the sphere to yield best visual acuity without accommodation . Place the prescription the patient is wearing ,if any ,into a trial frame or phoropter . Fog the eye to be examined with plus sphere while the patient views a visual acuity chart. Then decrease the fog until best visual acuity is obtained
  • 5. If no cylindrical correction is present initially ,the cross cylinder may still be used ,placed arbitrarily at 90 0 and 180 0 . To check the presence of astigmatism . If a preferred flip position is found ,cylinder is added with axis parallel to the respective plus or minus axis of the cross cylinder until the two flip choices are equal .
  • 6. If no preference is found with the cross- cylinder axis at 90 0 and 180 0 ,45 0 and 135 0 should always be checked before assuming that no astigmatism is present . Once any cylinder power is found , axis and power are refined in the usual manner .
  • 7. Cylinder axis is always refined first . This sequence is necessary because the correct axis can be found in the presence of an in correct power . But the full cylinder power will not be found in the presence of an incorrect axis .
  • 8. To refine the axis place the cross cylinder with its principal meridians 45 0 away from the principal meridians of the correcting cylinder . Present the patient with alternative flip choices and inquiring which is” blackest and sharpest “. Rotate the axis of the correcting cylinder toward the corresponding plus or minus axis of the cross cylinder .
  • 9. Low-power cylinder are rotated in large steps ,high powered cylinder are rotated in small steps ,repeat the procedure until the flip choices appear equal . To refine cylinder power ,align the cross – cylinder axis with the principal meridians of the correcting lens . As the examiner changes cylinder power according to the patient’s response . The spherical equivalent of the refractive correction should remain constant .
  • 10. That is for every 0.50 D of cylinder power change ,the sphere is changed 0.25 D in the opposite direction . If the cylinder power is changed a large amount ,the sphere power should be readjusted for best visual acuity . Continue to refine cylinder power until choices appear “about the same “to the patient ,that will be the correct endpoint .
  • 11. Remember always to use the proper cross cylinder for the patient’s visual acuity level . For eg :- +- 0.25 D cross cylinder is commonly used with visual acuity levels of 20/30 and better . A high cross cylinder [+- 0.50 D or +- 1.00 D ] should be used with poorer vision .
  • 12. Following steps are used in cross –cylinder refraction . Adjust sphere to the most plus or least minus that gives best visual acuity . Use test figure one or two lines larger than the patient’s BCVA , because the introduction of the JCC produce blur . If cylinder correction is not already present ,look for astigmatism .
  • 13. Refine axis first Refine cylinder power . Refine sphere ,cylinder axis ,and cylinder power .
  • 14. Tricking Avoid confusion with previous choices by giving different number to subsequent choices Which is better ,one or two ,three or four? If the patient persists in always choosing either the first or the second number ,reverse the order .
  • 15. Clock Dial When the clock dial is presented to the patient under sufficient fog . The examiner’s first job is to determine the axis of the correcting cylinder . This is done by first asking the patient if he or she can see three lines in any or all of the spokes . And then ask to report in which of the spokes the three lines are the sharpest or most distinct .
  • 16.
  • 17. Most clock dial charts equipped with numbers similar to those on the face of a clock . Expected response of the patient is that the spoke from 12 to 6 ,from 1 to 7 ,from 2 to 8 o’clock is most distinct . To determine the axis of the correcting cylinder ,the smaller of the two number reported by the is multiplied by 30.
  • 18. For example ,if the patient reports that the 12 t 6 o’clock spoke is most distinct the examiner would place the axis of the correcting cylinder at 180 degrees. The examiner then begins adding minus cylinder power ,0.25 D at a time . Questioning the patient each time as to the relative sharpness of the lines in the spokes representing the two principal meridians of the eye .
  • 19. If sufficient minus cylinder power is added ,the horizontal and vertical focal lines will be located in the same plane . Then the patient will report that the 12 to 6 and the 3-to -9 spokes are equally distinct . This procedure is referred to as “collapsing the conoid of sturm “ the horizontal and vertical focal line are replaced by a point image .
  • 20. Monocular subjective refraction Monocular subjective refraction consists of the following procedure Determining the cylindrical correction under fog. Refining the cylindrical correction without fog . Determining the spherical end point . Following these procedure ,one or more binocular balancing tests are performed .
  • 21. Determining the cylinder under fog Suitable fogging lenses will have to be put into place before beginning the monocular subjective refraction . The left eye is occluded ,a block of letter is introduced at the 6-m distance . Plus power in front of the right eye is reduced ,0.25 at a time until the patient can read all of the 20 /40 letters . Ensures that the eye is fogged by approximately 1.00 D ,so the entire conoid of sturm is in front of the retina .
  • 22. The astigmatic chart is presented ,and the axis of the correcting cylinder is determined . If clock dial is used ,the patient is asked first whether three lines can be seen in any or all of the spokes . If he can see then ask the patient to report in which of the spokes the three lines are the most distinct .
  • 23. Minus cylinder is added to the indicated axis 0.25 D at a time until equality is obtained . Additional minus cylinder power is then added 0.25 D at a time to obtain a reversal . Then the examiner returning to the lowest-power cylinder that brings about equality .
  • 24. The right is then occluded and the procedure are repeated for the left eye .
  • 25. Refining the cylinder without fog On completion of the astigmatic chart test . The right eye is defogged ,reducing plus lens power 0.25 D at a time ,until the best visual acuity line [20/20 or 20/15 ] is reads. When the removal of additional plus power [or addition of minus power ]fails to improve visual acuity ,the spherical lens power is returned to the maximum plus or minimum minus power that resulted in the best visual acuity for that eye .
  • 26. Beginning with the cylinder power and axis found with the astigmatic chart under fog . The practitioner first refines cylinder axis and then cylinder power .
  • 29. Additional cylinder check tests Additional check test for cylinder power . Many patient tended to reject cylinder power in the crossed- cylinder test even though it has been evident both in retinoscopy and in the astigmatic chart test . This occurs ,the patient is asked to watch the smallest readable row of letters .
  • 30. An additional -0.25 D cylinder is placed in the refractor ,and the patient is asked to report which of the two views appears to be more distinct . If the additional -0.25 D cylinder does not improve the clarity of the letters ,the original cylinder power is left in the refractor .
  • 31. If the additional cylinder power causes the letters to be more distinct ,the examiner can choose to leave the additional cylindrical power in the refractor.
  • 32. Additional check test of axis The additional cylinder check test for axis is sometime referred to as bracketing . It is useful mainly when the power of the correcting cylinder is 1.00 D or more . The patient is asked to view a row of 20/20 or 20/15 letter And ask to report when the cylinder lens in the refractor is slowly rotated.
  • 33. When the blur is reported ,the change in the axis is noted mentally . And the test is repeated with the same instructions . For example : if the original correcting cylinder axis is located at 180 degrees . The patient reports a blur at 15 degrees and blur at 165 degrees ,then the examiner can assume that the orginal axis was correct .
  • 34. Determining the spherical end point Each eye is fogged to 0.75 or 1.00 D and defogged to best acuity. Maximum plus power of best visual acuity . Examiner should present a block of letters extending from 20/40 to 20/15. We should make a mental note of the patient’s acuity for each eye through the +0.75 or +1.00 D fogging lens prior to defogging .
  • 35. It is important that accommodation be relaxed in arriving at the monocular end point . For example : +1.00 D fog may blur the right eye to 20/30 ,but it may blur the left eye only to 20/20 . This evidence that the left eye is underplussed [or over minused ] ,and the subjective end point for both eye should be determined again .
  • 36. Patient instruction To determine the monocular endpoint correct patient instruction are important . The examiner should understand that the patient’s subjective evaluation of the clarity or distinctness of the letters is not the important consideration . The important ,overriding consideration is the ability of the patient to resolve the letters .
  • 37. As plus power is decreased 0.25 D at a time . When the point has been reached where an additional decrease in plus power of 0.25 D does not make any more letter readable. On the other hand ,if the examiner allows the patient to respond in terms of clarity of the letter or in terms of which lens is preferred .
  • 38. Many young patient will continue to accommodate with each 0.25 D of reduction in plus ,with the result that the end point will be completely invalid. When minus lens power is added to the point that accommodation is necessary to keep the letters in sharp focus on the retina Many patient will notice that the letter appear to be smaller .
  • 39. This is purely optical effect ,known as accommodative micropsia . it is right to ask the patient if the letter looks smaller to avoid overminusing the patient . If patient report that the letter are not smaller should not be taken as evidence .
  • 40. Binocular balance The purpose of binocular balancing tests is not to balance the visual acuity but to balance the state of accommodation of the two eyes . If the corrected visual acuity is same in both eye ,the balancing procedure may consist of comparison of the visual acuity for the two eyes . If the correcting visual acuity is not same in both eyes ,then a method not based on visual acuity must be used .
  • 41. Procedure used for balancing the state of accommodation for the two eyes are often referred to as equalization tests or as binocular balancing tests .
  • 42. Balancing with little or no fog On completion of the monocular subjective refraction ,the lenses are left in the refractor . The occluders has to remove from both eyes . The patient’s attention is called to a block of letters at 6 m. Plus lens power is added in front of both eyes until the 20/20 letter are blurred but the 20/25 letter are easily resolved .
  • 43. This normally requires an increase in plus or a decrease in minus of 0.25 to 0.50 D. Then ask to compare the clarity of the 20/25 letters for the two eyes . Using either prism dissociation or alternative occlusion .
  • 44. Prism dissociation The examiner place 3 of base –down prism in front of the right eye and 3 of base-up prism in front of the left eye . The patient will see two charts separated –vertically . The upper chart seen by right eye . Patient attention is called to the 20 /25 letters .
  • 45. Then asked to report whether the letter are more distinct or easier to read in the upper chart or the lower chart . If the equally distinct for the two eyes, the accommodative state of the two eyes is considered to be balanced ,and the test is over .
  • 46. If the patient reports a difference in clarity of the letters for the two eyes. 0.25 D is added in front of the eye with the better vision and the test is repeated . Often the patient fails to report equal clarity for any lens combination . This problem can be solved by instructing the patient to report “which of the lens combination causes the upper and lower charts to be more nearly equal .
  • 47. Another way of resolving this difficulty is to give better acuity to the dominant eye . Once the patient’s acuity is balanced at 20/25 the patient is defogged binocularly.
  • 48. Alternate occlusion Patient is instructed to compare alternate views of the chart while the each eye is alternately occluded. Occluder in the refractor can be used for occlusion . Fogging to 20/25 . Adding 0.25 D to the better eye ,are done in the same manner of prism dissociation test .
  • 49. Check Test for the Binocular End point Once the examiner is satisfied with the binocular balance ,binocular end point can be verified . Patient’s attention is called to the 20 /20 letter . Add +0.25 D sphere to both eye and ask for any difference in the clarity of the letters.
  • 50. The same question is again asked after adding 0.25 D sphere . Again third +0.25 D sphere is added ,ask of the difference in the clarity . Expected response are that the 20/20 letters will be “slightly blurred “. With the first 0.25 D of plus . “badly blurred “ with the second 0.25D. “blurred out “ with the third 0.25 D .
  • 51. Bichrome test As a monocular End –point test The bichrome test must be done in an almost completely darkened room . Starting with the result of monocular subjective finding . +0.50D or + 0.75 D of spherical power is placed in front of each eye. The red and green filter is placed on the projector.
  • 52.
  • 53. The patient is asked to report which of the letters ,those on the red background or those in the green background –are “sharper blacker ,or more distinct “. The patient is expected to report that the letters on the red background are more distinct than those on the green background .
  • 54. As the plus power is reduced 0.25 D at a time . At some point the patient should report that the letter on red and green background are equally distinct . As plus power is further reduced ,the patient should say those on the green side are more distinct than the red side .
  • 55. If the original monocular subjective endpoint was correct ,the patient will typically report that the red letters are more distinct with +0.75,+0.50,+0.25 D fog . When all the fog is removed the letter will appear to be equally distinct on both the red green sides .
  • 56. Prescribing for children Prescribing visual correction for children often has two goals : Providing a focused retinal image . Achieving the optimal balance between accommodation and convergence . In some cases subjective refraction is impossible because child’s inability to cooperate. The optimal refraction in an infant requires the paralysis of accommodation with complete cycloplegia .
  • 57. The presence of strabismus may modify normal prescribing guidelines .
  • 58. Myopia [general guidelines for correction of significant childhood myopia] Cycloplegic refraction are mandatory . Full refractive error ,including cylinder ,should be corrected .young children tolerate cylinder well. Some will under correct myopia ,other may even use bifocals with or without atropine ,on the theory that prolonged accommodation hastens .
  • 59. Intentional undercorrection of a myopicaesotrope to decrease the angle of deviation is rarely tolerated . Indentional overcorrection of a myopic error can be of some value in controlling an intermittent exodeviation . Parents should be educated about the natural progression of myopia and the need for frequent refraction and possible prescription changes .
  • 60. Contact lenses may be desirable in older children to avoid the problem of image minification found with high- minus lenses.
  • 61. Hyperopia When hyperopia and esotropia coexist ,initial management includes full correction of the cycloplegic refractive error . In school age child ,the full refractive correction may cause blurring of distance vision because of the inability to relax accommodation fully . The amount of correction may have to be reduced for the child to accept the glasses.