Journal of Liquid Chromatography & Related Technologies*, 28: 1253-1263, 2005
Copyright © Taylor & Francis, Inc.
ISSN 1082-6076 print/1520-572X online
DOI: 10.1081/JLC-200053046




                     TM
          UPLC : An Introduction and Review

                              Michael E. Swartz
                 Waters Corporation, Milford, Massachusetts, USA




Abstract: Ultra performance liquid chromatography™ (UPLC) takes advantage of
technological strides made in particle chemistry performance, system optimization,
detector design, and data processing and control. Using sub-2 |j,m particles and
mobile phases at high linear velocities, and instrumentation that operates at higher
pressures than those used in HPLC, dramatic increases in resolution, sensitivity,
and speed of analysis can be obtained. This new category of analytical separation
science retains the practicality and principles of HPLC while creating a step-
function improvement in chromatographic performance.
   This review introduces the theory of UPLC, and summarizes some of the most recent
work in the field.

Keywords: UPLC, Ultra performance liquid chromatography


INTRODUCTION

High performance liquid chromatography (HPLC) has proven to be the
predominant technology used in laboratories worldwide during the past
30-plus years. One of the primary drivers for the growth of this technique
has been the evolution of packing materials used to effect separations.
The underlying principles of this evolution are governed by the van
Deemter equation, with which any student of chromatography is intimately
familiar. flj The van Deemter equation is an empirical formula that describes
the relationship between linear velocity (flow rate) and plate height (HETP,
or column efficiency). And, since particle size is one of the variables, a van
Deemter curve can be used to investigate chromatographic performance.


   Address correspondence to Michael E. Swartz, Waters Corporation, 34 Maple
Street, Milford, MA 01757, USA; E-mail: michael_swartz@ waters.com
1254                                                                                   M. E. Swartz

      For many years, researchers have looked at "fast LC" as a way to speed up
analyses.[2'31 The "need for speed" has been driven by the sheer numbers of
samples in some laboratories (particularly in drug discovery) and the
availability of affordable, easy to use mass spectrometers. Smaller columns
and faster flow rates (amongst other parameters) have been used. Elevated
temperature, having the dual advantages of lowering viscosity, and increasing
mass transfer by increasing the diffusivity of the analytes, has also been
investigated.M However, using conventional particle sizes and pressures, limit-
ations are soon reached and compromises must be made, sacrificing resolution
for time.
      However, as illustrated in Figure 1, as the particle size decreases to less
than 2.5 urn, not only is there a significant gain in efficiency; but the efficiency
doesn't diminish at increased flow rates or linear velocities. By using
smaller particles, speed and peak capacity (number of peaks resolved per
unit time) can be extended to new limits, termed Ultra Performance Liquid
Chromatography, or UPLC . This introduction and review traces some of
the developments and technological advancements made in producing the
first commercially available UPLC instrument.


SMALL PARTICLE CHEMISTRY

The promises of the van Deemter equation cannot be fulfilled without smaller
particles than those traditionally used in HPLC. The design and development


                  10 Mm Porticle ,
                                               HPLC


       H
       E'5
       T
       P .0




                                     2            3         4            5

   f law Kale [ml/mm]:
                                           Linear Velocity [u, mm/sec]
        10 - 1.0 mm      0.04        0.07         0.10      0.13         0.17   0.20       0.24

        10 = 2.1 mm      0.15        0.3          0.45      0.6          0.75   0.9         1.05

        !D » 4.6 mm      0.7         1.4          1.1       2.8          3.5                49


Figure 1. van Deemter plot, illustrating the evolution of particle sizes over the last
three decades.
UPLC : An Introduction and Review                                           1255

of sub-2 |jim particles is a significant challenge, and researchers have been
active in this area for some time to capitalize on their advantages.t5"71
Although high efficiency, non-porous 1.5|j,m particles are commercially
available, they suffer from poor loading capacity and retention due to low
surface area. To maintain retention and capacity similar to HPLC, UPLC
must use novel porous particles that can withstand high pressures. Silica
based particles have good mechanical strength, but can suffer from a
number of disadvantages, which include a limited pH range and tailing of
basic analytes. Polymeric columns can overcome pH limitations, but they
have their own issues, including low efficiencies and limited capacities.
     In 2000, a first generation hybrid chemistry that took advantage of the best
of both the silica and polymeric column worlds was introduced.[8'9-1 Produced
using a classical sol-gel synthesis that incorporates carbon in the form of
methyl groups, these columns are mechanically strong, with high efficiency,
and operate over an extended pH range. But, in order to provide the kind of
enhanced mechanical stability required for UPLC, a second generation
bridged ethane hybrid (BEH) technology was developed.[10] These 1.7|JLm
particles derive their enhanced mechanical stability by bridging the methyl
groups in the silica matrix.
     Packing 1.7 (Jim particles into reproducible and rugged columns was also
a challenge that needed to be overcome. Requirements include a smoother
interior surface of the column hardware, and re-designing the end frits
to retain the small particles and resist clogging. Packed bed uniformity is
also critical, especially if shorter columns are to maintain resolution while
accomplishing the goal of faster separations.
     In addition, at high pressures, frictional heating of the mobile phase can
be quite significant and must be considered.1111 With column diameters
typically used in HPLC (3.0 to 4.6mm), a consequence of frictional heating
is the loss of performance due to temperature induced non uniform flow.
To minimize the effects of frictional heating, smaller diameter columns
(1-2.1 mm) are typically used for UPLC.[12'13]


CAPITALIZING ON SMALLER PARTICLES

Small particles alone do not make it possible to fulfill the promises of the van
Deemter equation (Figure 1). Instrument technology also had to keep pace
to truly take advantage of the increased speed, superior resolution, and
sensitivity afforded by smaller particles. Standard HPLC technology
(pumps, injectors, and detectors) simply doesn't have the horsepower to
take full advantage of sub-2 (Jim particles.
     One-of-a-kind systems, capable of delivering the pressures required to
realize the potential of UPLC have been reported in the literature and
elsewhere.114"16]
1256                                                               M. E. Swartz

     Lee et al. described the design of injection valves and separation reprodu-
cibility,[I4] and the use of a carbon dioxide enhanced slurry packing method on
the capillary scale for the separation of some benzodiazepines, herbicides, and
various pharmaceutical compounds.1171 Jorgenson et al. modified a commer-
cially available HPLC system to operate at 17,500 psi and used 22 cm long
capillaries packed with 1.5|jim Gig-modified particles for the analysis of
proteins.1-15-1
     These reports illustrated that, to take full advantage of low dispersion and
small particle technology to achieve high peak capacity UPLC separations, a
greater pressure range than that achievable by today's HPLC instrumentation
was required. The calculated pressure drop at the optimum flow rate for
maximum efficiency across a 15cm long column packed with 1.7 [Am
particles is about 15,000 psi. Therefore, a pump capable of delivering
solvent smoothly and reproducibly at these pressures, that can compensate
for solvent compressibility, and can operate in both the gradient and
isocratic separation modes, was required.
     Sample introduction is also critical. Conventional injection valves, either
automated or manual, are not designed and hardened to work at extreme
pressure. To protect the column from experiencing extreme pressure fluctu-
ations, the injection process must be relatively pulse-free. The swept
volume of the device also needs to be minimal to reduce potential band
spreading. A fast injection cycle time is needed to fully capitalize on the
speed afforded by UPLC which, in turn, requires a high sample capacity.
Low volume injections with minimal carryover are also required to realize
the increased sensitivity benefits.
     With 1.7 (jum particles, half-height peak widths of less than one second
are obtained, posing significant challenges for the detector. In order to accu-
rately and reproducibly integrate an analyte peak, the detector sampling rate
must be high enough to capture enough data points across the peak. In
addition, the detector cell must have minimal dispersion (volume) to
preserve separation efficiency. Conceptually, the sensitivity increase for
UPLC detection should be 2-3 times higher than with HPLC separations,
depending on the detection technique that is used. Conventional absorbance-
based optical detectors are concentration sensitive detectors and, for UPLC
use, the flow cell volume would have to be reduced in standard UV/Visible
detectors to maintain concentration and signal, while avoiding Beers' Law
limitations.
     In early 2004, the first commercially available UPLC system that
embodied these requirements was described for the separation of various
pharmaceutical related small organic molecules, proteins, and peptides; it is
called the ACQUITY UPLC™ System.[18-20]
     Using UPLC, it is now possible to take full advantage of chromatographic
principles to run separations using shorter columns, and/or higher flow rates
for increased speed, with superior resolution and sensitivity. Figures 2 and 3
UPLC : An Introduction and Review                             1257



                                          E r- -e
                                          bo r^i c
                                                    Q)
                                          £ © oa
                                          ecu"
                                    CD    — .2 3
                                          Q tj £


                                          §st
                                          S u a


                                           i > £




                                          ®I 1
                                           oo cd    o


                                          a i-
                                          |o^
                                          u 2S
                                          s .a I-

                                               *




                                     CD   —



                                          fj   •—   >~>




                                          "!L
                                          C
                                          g
                                               g u f
                                               3 _-0 5
                                          2 '£ y c
                                           • c o
                                          •111
                                          1 ES
                                          •o vo "o

                                          ." o -o .
                                          o si s
                                          S 2I 2
                                    o
                                    6


            2
                                               00   S
                                                    a
                                                          N
                                               H          S

                                               5 & B
0,08




                         004



                         002




                         0.08



                         0.06




                         002



                         0.00


                                0.00   5.00   10.00   15.00   20.00   25,00    30-00    35,00   40.00   45.00   50.00   55-00   60.00
                                                                              Minutes


Figure 3. HPLC vs. UPLC peak capacity. In this gradient peptide map separation, the HPLC (top) separation (on a 5 ^.m CIS column) yields 70
                                                                                                                                                     V.
peaks, or a peak capacity of 143, while the UPLC separation (bottom) run under identical conditions yields 168 peaks, or a peak capacity of 360, a   ^
2.5X increase.                                                                                                                                       3.
                                                                                                                                                     N
UPLC : An Introduction and Review                                               1259

illustrate UPLC in action. In Figure 2 a separation of eight diuretics is accom-
plished in under 1.6 minutes. The same separation on a 2.1 by 100mm 5 |a,m
Cig HPLC column yields comparable resolution, but takes over ten minutes.
For some analyses, however, speed is of secondary importance; peak
capacity and resolution take center stage. Figure 3 shows a peptide map
where the desired goal is to maximize the number of peaks. In this application,
the increased peak capacity (number of peaks resolved per unit time) of
UPLC dramatically improves the quality of the data resulting in a more
definitive map


APPLICATIONS

Chromatographers are accustomed to making compromises; one of the most
common scenarios involves sacrificing resolution for speed. In addition, for
complex samples like natural product extracts, added resolution can provide
more information in the form of additional peaks. Figure 4 shows an HPLC


 0.8AUFS




                                                                           HPLC
                                                                             10jiL
                                                                           Injection
                                         Time in Minutes

2.1 AUFS




                                                            UPLC™
                                                           5^L Injection
                       Time in Minutes


Figure 4. Comparison HPLC and. UPLC for the separation of a ginger root extract.
HPLC conditions: Column: 2.1 by 100mm 5.0|a.m prototype BEH CIS at 28°C. A
25-96%B linear gradient over 10 minutes, at a flow rate of 1.0 mL/min was used.
Mobile phase A was water, B was acetonitrile. UV detection @ 230 nm, 10|j.L
injection. UPLC conditions: Column: 2.1 by 100mm 1.7|j.m ACQUITY BEH CIS
at 28°C. A 50-100%B linear gradient from 1.4 to 3.7 minutes, followed by a hold
until 6.0 minutes, at a flow rate of 0.3 mL/min was used. Mobile phase A was water,
B was acetonitrile. UV detection @ 230 nm, 5 (xL injection.
1260                                                              M. E. Swartz

versus UPLC separation comparison of a ginger root extract sample where
both speed and resolution are improved, as well as an increase in sensitivity.
DryLab software was used to model and redevelop the separation and
transfer it to the ACQUITY UPLC System and BEH chemistry.
     Faster separations can lead to higher throughput and time savings when
running multiple samples. But, a significant amount of time can also be
consumed in developing the method in the first place. Faster, higher resolution
UPLC separations can cut method development time from days, to hours,
or even minutes. Figure 5 is an example of an UPLC separation of several
closely related coumarins and a metabolite that was developed in under an
hour, including UPLC scouting runs for gradient optimization, and individual
runs for elution order identification. These runs were performed in a fraction
of the time that would be necessary by conventional HPLC, resulting in
significant time savings in the method development laboratory.
     As alluded to previously, mass spectrometry has gained widespread
acceptance as an analytical tool for the qualitative and quantitative analysis



       0.03                      3 4




       AUFS




       0.00                                                         1.30



Figure 5. UPLC Separation of Seven Coumarins illustrating fast method develop-
ment. Column: 2.1 by 30mm 1.7 |xm ACQUITY UPLC BEH Cjg @ 35°C. A
20-40%B linear gradient over 1.0 minute, at a flow rate of 0.86mL/min was used.
Mobile phase A was 0.1 % formic acid, B was acetonitrile. UV detection @ 254 nm
and 40 pts/sec. Peaks are in order: 1: 7-hydroxycoumarin-glucuronide, 7-hydroxy-
coumarin, 4-hydroxycoumarin, coumarin, 7-methoxycoumarin, 7-ethoxycoumarin,
and 4-ethoxycoumarin.
• T0= M5 ES*
                                                                        4 U-J                                                                                  BPI
                                                                          I ,,14 41                                                                         ' 56e-l
                                                                    13 68 PI




                       200   4 DO   B [iD   8 [iD   ID DO   12.DD      14 OD     1600   1600   20.00   22.00   2400   2G OD   2B OD   30 DO   3200    34 DO



Figure 6. Separation of rat bile following the administration of midolazam at 5 mg/kg: A) 30 minute separation on a 2.1 by 100mm 3.5 (xm CIS
HPLC Column and B) 30minute separation on a 2.1 by 100mm 1.7 |j.m CIS UPLC column. Reprinted by permission from reference 21.
i-iu-i                                                            wi. &. swartz

of many types of compounds. MS detection is significantly enhanced
by UPLC; increased peak concentrations with reduced chromatographic
dispersion at lower flow rates (no flow splitting) promotes increased source
ionization efficiencies. Jorgenson et al. have shown that higher chromato-
graphic efficiency, resulting from the use of UPLC, translates into better
resolution and higher peak capacity, which is particularly important for the
analysis of peptides and proteins.[15] The increased resolving power made
the resulting data easier to interpret, since more of the MS peaks consisted
of a single compound, and up to a 20-fold improvement in the quality of
the spectral information (vs. nanoelectrospray) was obtained. Lee et al. also
used MS detection for the analysis of low molecular weight compounds
similar to those that might comprise a combinatorial library.[14'17] It was
demonstrated that, in order to address the very narrow peaks produced by
UPLC, it is necessary to use a very high data capture rate MS such as a
TOP or quadrapole with fast scan rates. Lee et al. also pointed out that,
in some instances, related compounds of the same molecular weight
and similar structures could not be differentiated by MS, necessitating
chromatographic resolution on the UPLC time scaled1'
     Plumb et al. have investigated the use of UPLC/MS for the analysis of
metabolites,121'22] and as a tool for differential metabolic pathway profiling
in functional genomic studies.[23] Their data illustrate the benefit obtained
from the extra resolution of UPLC, both in terms of specificity and spectral
quality, revealing new information and reducing the risk of not detecting
potentially important metabolites. Figure 6 shows the benefits of UPLC
versus HPLC for monitoring the in-vivo metabolism of midazolam (an anti-
convulsant) in rat liver bile. This is a challenging separation due to the high
concentration of bile salts that can interfere, and the presence of bilirubin
that can cause ion pairing. The resolution is dramatically improved, and
the number of discreet peaks has more than doubled.

CONCLUSION

UPLC presents the ability to extend and expand the utility of separation
science at a time when many scientists have reached separation barriers,
pushing the limits of conventional HPLC. New chemistry and instrumentation
technology can provide more information per unit of work as UPLC begins
to fulfill the promise of increased speed, resolution, and sensitivity
predicted for liquid chromatography.


ACKNOWLEDGMENTS

The author would like to acknowledge the contributions of the ACQUITY
program team at Waters, particularly Eric Grumbach, Michael Jones,
UPLC : An Introduction and Review                                                 1263

Marianna Kele, Rob Plumb, and Jeff Mazzeo for their contributions to this
manuscript.
   ACQUITY UPLC, UPLC, and Ultra Performance LC are trademarks of
Waters Corporation.


REFERENCES

 1.   van Deemter, J.J.; Zuiderweg, F.J.; Klinkenberg, A. Chem. Eng. Sci. 1956, 5, 271.
 2.   Cheng, Y.-F.; Lu, Z.; Neue, U. Rapid Commun. Mass Spectrom. 2001, 15, 141.
 3.   Tiller, P.R.; Romanyshyn, L.A.; Neue, U.D. Anal. Bioanal. Chem. 2003, 377, 788.
 4.   Neue, U.D.; Mazzeo, J.R. J. Sep. Sci. 2001, 24, 1.
 5.   Jerkovitch, A.D.; Mellors, J.S.; Jorgenson, J.W. LC-GC North Amer. 2003, 21,1.
 6.   Wu, N.; Lippert, J.A.; Lee, M.L. J. Chromotogr. 2001, 911, 1.
 7.   Unger, K.K.; Kumar, D.; Grun, M.; Buchel, G.; Ludtke, S.; Adam, Th.;
      Scumacher, K.; Renker, S. J. Chromatogr. A 2000, 892, 47.
 8.   Neue, U.D.; Walter, T.H.; Alden, B.A.; Jiang, Z.; Fisk, R.P.; Cook, J.T.;
      Close, K.H.; Carmody, J.L.; Grassi, J.M.; Cheng, Y.-F.; Lu, Z.; Crowley, R.J.
      Amer. Lab. 1999, 31, 36.
 9.   Cheng, Y.-F.; Walter, T.H.; Lu, Z.; Iraneta, P.; Gendreau, C.; Neue, U.D.;
      Grassi, J.M.; Carmody, J.L.; O'Gara, J.E.; Fisk, R.P. LC-GC 2000, 18, 1162.
10.   Mazzeo, J.R. Anal. Chem. November 2004, In press.
11.   Halasz, I.; Endele, R.; Asshauer, J. J. Chromatogr. 1975, 112, 37.
12.   MacNair, J.E.; Lewis, K.C.; Jorgenson, J.W. Anal. Chem. 1997, 69, 983.
13.   Colon, L.A.; Citron, J.M.; Anspach, J.A.; Fermier, A.M.; Swinney, K.A. Analyst
      2004, 729, 503.
14.   Wu, N.; Lippert, J.A.; Lee, M.L. J. Chromatogr. A 2001, 911, 1.
15.   Tolley, L.; Jorgenson, J.W.; Mosely, M.A. Anal. Chem. 2001, 73, 2985.
16.   Swinney, K.A.; Fermier, A.M.; Guo, Y.; Oyler, A.R.; Segmuller, B.E.; Dunphy, R.
      Presentation #L1703 at the 28th Intl. Symp. on High Perform. Liq. Phase Sepns. &
      Rel. Techn, Philadelphia, PA, June 2004.
17.   Lippert, J.A.; Xin, B.; Wu, N.; Lee, M.L. J. Microcol. Sepn. 1997, 11, 631.
18.   Swartz, M.E. Presentation at the Pitts, Conf. on Anal. Chem. Appl. Spectrosc,
      Chicago, IL, 2004.
19.   Swartz, M.E.; Murphy, B. Lab Plus Intl. 2004, 18, 6.
20.   Swartz, M.E.; Murphy, B. Pharm.l Formul. Qual. November 2004, In press.
21.   Plumb, R.S.; Castro-Perez, J.; Granger, J.H.; Beattie, I.; Joncour, K.; Wright, A.
      Rapid Commun. Mass Spectrom. 2004, 18, 2331.
22.   Castro-Perez, J.; Plumb, R.; Granger, J.H. Rapid Commun. Mass Spectrom.
      October 2004, Submitted.
23.   Wilson, I.D.; Nicholson, J.K.; Castro-Perez, J.; Granger, J.H.; Smith, B.W.;
      Mazzeo, J.; Plumb, R.S. J. Proteome Res. October 2004, Submitted.


Received November 22, 2004
Accepted December 19, 2004
Manuscript 6549E

Mais conteúdo relacionado

Mais procurados

UPLC - ULTRA PERFORMANCE LIQUID CHROMATOGRAPHY
UPLC - ULTRA PERFORMANCE LIQUID CHROMATOGRAPHYUPLC - ULTRA PERFORMANCE LIQUID CHROMATOGRAPHY
UPLC - ULTRA PERFORMANCE LIQUID CHROMATOGRAPHYGANESH SAI MYNENI
 
ultra performance liquid chromatography ppt.by Manoj Ingale
ultra performance liquid chromatography ppt.by Manoj Ingaleultra performance liquid chromatography ppt.by Manoj Ingale
ultra performance liquid chromatography ppt.by Manoj IngaleIndus Biotech Pvt.Ltd.
 
Difference between HPLC and UPLC
Difference between HPLC and UPLCDifference between HPLC and UPLC
Difference between HPLC and UPLCVrushali Tambe
 
ultra high performance liquid chromatography
ultra high performance  liquid chromatographyultra high performance  liquid chromatography
ultra high performance liquid chromatographyacademic
 
Hplc instrumentation in detail (Practical) Hplc pump inj_column
Hplc instrumentation in detail (Practical) Hplc pump inj_columnHplc instrumentation in detail (Practical) Hplc pump inj_column
Hplc instrumentation in detail (Practical) Hplc pump inj_columnPratikShinde120
 
Ultra performance liquid chromatography
Ultra performance liquid chromatographyUltra performance liquid chromatography
Ultra performance liquid chromatographybiniyapatel
 
Ultra Performance Liquid Chromatography
Ultra Performance Liquid ChromatographyUltra Performance Liquid Chromatography
Ultra Performance Liquid ChromatographyMahewash Sana Pathan
 
Nano liquid chromatography (ncl)
Nano liquid chromatography (ncl)Nano liquid chromatography (ncl)
Nano liquid chromatography (ncl)Tushar Borate
 
Chiral hplc intro
Chiral hplc introChiral hplc intro
Chiral hplc introsameerr98
 
Application of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of PharmaceuticalsApplication of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of PharmaceuticalsShimadzu Scientific Instruments
 

Mais procurados (20)

Uplc
UplcUplc
Uplc
 
UPLC - ULTRA PERFORMANCE LIQUID CHROMATOGRAPHY
UPLC - ULTRA PERFORMANCE LIQUID CHROMATOGRAPHYUPLC - ULTRA PERFORMANCE LIQUID CHROMATOGRAPHY
UPLC - ULTRA PERFORMANCE LIQUID CHROMATOGRAPHY
 
UPLC
UPLCUPLC
UPLC
 
ultra performance liquid chromatography ppt.by Manoj Ingale
ultra performance liquid chromatography ppt.by Manoj Ingaleultra performance liquid chromatography ppt.by Manoj Ingale
ultra performance liquid chromatography ppt.by Manoj Ingale
 
UPLC
UPLCUPLC
UPLC
 
Uplc
UplcUplc
Uplc
 
Uplc ppt
Uplc ppt Uplc ppt
Uplc ppt
 
Difference between HPLC and UPLC
Difference between HPLC and UPLCDifference between HPLC and UPLC
Difference between HPLC and UPLC
 
ultra high performance liquid chromatography
ultra high performance  liquid chromatographyultra high performance  liquid chromatography
ultra high performance liquid chromatography
 
Hplc instrumentation in detail (Practical) Hplc pump inj_column
Hplc instrumentation in detail (Practical) Hplc pump inj_columnHplc instrumentation in detail (Practical) Hplc pump inj_column
Hplc instrumentation in detail (Practical) Hplc pump inj_column
 
Nano LC and UPLC
Nano LC and UPLCNano LC and UPLC
Nano LC and UPLC
 
Mohini patil
Mohini patilMohini patil
Mohini patil
 
Ultra performance liquid chromatography
Ultra performance liquid chromatographyUltra performance liquid chromatography
Ultra performance liquid chromatography
 
Ultra Performance Liquid Chromatography
Ultra Performance Liquid ChromatographyUltra Performance Liquid Chromatography
Ultra Performance Liquid Chromatography
 
Nano liquid chromatography (ncl)
Nano liquid chromatography (ncl)Nano liquid chromatography (ncl)
Nano liquid chromatography (ncl)
 
Chiral hplc intro
Chiral hplc introChiral hplc intro
Chiral hplc intro
 
Hplc cont
Hplc contHplc cont
Hplc cont
 
Bhosale ppt
Bhosale pptBhosale ppt
Bhosale ppt
 
Uplc&nano
Uplc&nanoUplc&nano
Uplc&nano
 
Application of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of PharmaceuticalsApplication of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
Application of a Dual Injection (U)HPLC for the Analysis of Pharmaceuticals
 

Semelhante a UPLC INTRODUCTION AND REVIEW

Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptx
Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptxUltra perfoemence liquid chromatography bhagyashri ppt 23.pptx
Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptxBhagyashriPatil73
 
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...Sandy Simmons
 
Bock Acs Paper Id1119039 08192007 Poster
Bock Acs Paper Id1119039 08192007 PosterBock Acs Paper Id1119039 08192007 Poster
Bock Acs Paper Id1119039 08192007 PosterMarybock
 
Life Cycle Management of Chromatographic Methods for Biopharmaceuticals
Life Cycle Management of Chromatographic Methods for BiopharmaceuticalsLife Cycle Management of Chromatographic Methods for Biopharmaceuticals
Life Cycle Management of Chromatographic Methods for BiopharmaceuticalsWaters Corporation
 
Webinar - Pharmacopeial Modernization: How Will Your Chromatography Workflow ...
Webinar - Pharmacopeial Modernization: How Will Your Chromatography Workflow ...Webinar - Pharmacopeial Modernization: How Will Your Chromatography Workflow ...
Webinar - Pharmacopeial Modernization: How Will Your Chromatography Workflow ...Waters Corporation
 
Analysis of liq liq flow in millichannels
Analysis of liq liq flow in millichannelsAnalysis of liq liq flow in millichannels
Analysis of liq liq flow in millichannelsakash kumar
 
Herbal drugs and fingerprints
Herbal drugs and fingerprintsHerbal drugs and fingerprints
Herbal drugs and fingerprintsSpringer
 
Liquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiencyLiquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiencyguest63ff7d
 
Crimson Publishers-Rapid Microvalve Actuated Electroosmotic Reagent Delivery ...
Crimson Publishers-Rapid Microvalve Actuated Electroosmotic Reagent Delivery ...Crimson Publishers-Rapid Microvalve Actuated Electroosmotic Reagent Delivery ...
Crimson Publishers-Rapid Microvalve Actuated Electroosmotic Reagent Delivery ...CrimsonPublishers-SBB
 
36433 Topic HA W9 R1Number of Pages 1 (Double Spaced)N.docx
36433 Topic HA W9 R1Number of Pages 1 (Double Spaced)N.docx36433 Topic HA W9 R1Number of Pages 1 (Double Spaced)N.docx
36433 Topic HA W9 R1Number of Pages 1 (Double Spaced)N.docxrhetttrevannion
 
Paper presenting practical explanation of kinetic energy separation technolog...
Paper presenting practical explanation of kinetic energy separation technolog...Paper presenting practical explanation of kinetic energy separation technolog...
Paper presenting practical explanation of kinetic energy separation technolog...Carroll Cobb
 
Survey on Declining Curves of Unconventional Wells and Correlation with Key ...
Survey on Declining Curves of Unconventional Wells  and Correlation with Key ...Survey on Declining Curves of Unconventional Wells  and Correlation with Key ...
Survey on Declining Curves of Unconventional Wells and Correlation with Key ...Salman Sadeg Deumah
 
Why polymerics as hplc media and why simulated
Why polymerics as hplc media and why simulatedWhy polymerics as hplc media and why simulated
Why polymerics as hplc media and why simulatedHossein Hodjat
 
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...New Technique for Measuring and Controlling the Permeability of Polymeric Mem...
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...Editor IJCATR
 
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...New Technique for Measuring and Controlling the Permeability of Polymeric Mem...
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...Editor IJCATR
 

Semelhante a UPLC INTRODUCTION AND REVIEW (20)

Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptx
Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptxUltra perfoemence liquid chromatography bhagyashri ppt 23.pptx
Ultra perfoemence liquid chromatography bhagyashri ppt 23.pptx
 
8
88
8
 
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...
Enhancing Sensitivities and Peak Capacities for UHPLC-MS Fast Gradient Analys...
 
Bock Acs Paper Id1119039 08192007 Poster
Bock Acs Paper Id1119039 08192007 PosterBock Acs Paper Id1119039 08192007 Poster
Bock Acs Paper Id1119039 08192007 Poster
 
HPTLC.pptx
HPTLC.pptxHPTLC.pptx
HPTLC.pptx
 
Life Cycle Management of Chromatographic Methods for Biopharmaceuticals
Life Cycle Management of Chromatographic Methods for BiopharmaceuticalsLife Cycle Management of Chromatographic Methods for Biopharmaceuticals
Life Cycle Management of Chromatographic Methods for Biopharmaceuticals
 
Webinar - Pharmacopeial Modernization: How Will Your Chromatography Workflow ...
Webinar - Pharmacopeial Modernization: How Will Your Chromatography Workflow ...Webinar - Pharmacopeial Modernization: How Will Your Chromatography Workflow ...
Webinar - Pharmacopeial Modernization: How Will Your Chromatography Workflow ...
 
UPLC Poster
UPLC Poster UPLC Poster
UPLC Poster
 
Analysis of liq liq flow in millichannels
Analysis of liq liq flow in millichannelsAnalysis of liq liq flow in millichannels
Analysis of liq liq flow in millichannels
 
Hplc and gc analysis
Hplc and gc analysisHplc and gc analysis
Hplc and gc analysis
 
Herbal drugs and fingerprints
Herbal drugs and fingerprintsHerbal drugs and fingerprints
Herbal drugs and fingerprints
 
Liquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiencyLiquid chromatography still striving for high efficiency
Liquid chromatography still striving for high efficiency
 
Crimson Publishers-Rapid Microvalve Actuated Electroosmotic Reagent Delivery ...
Crimson Publishers-Rapid Microvalve Actuated Electroosmotic Reagent Delivery ...Crimson Publishers-Rapid Microvalve Actuated Electroosmotic Reagent Delivery ...
Crimson Publishers-Rapid Microvalve Actuated Electroosmotic Reagent Delivery ...
 
36433 Topic HA W9 R1Number of Pages 1 (Double Spaced)N.docx
36433 Topic HA W9 R1Number of Pages 1 (Double Spaced)N.docx36433 Topic HA W9 R1Number of Pages 1 (Double Spaced)N.docx
36433 Topic HA W9 R1Number of Pages 1 (Double Spaced)N.docx
 
Paper presenting practical explanation of kinetic energy separation technolog...
Paper presenting practical explanation of kinetic energy separation technolog...Paper presenting practical explanation of kinetic energy separation technolog...
Paper presenting practical explanation of kinetic energy separation technolog...
 
Hplc doc
Hplc docHplc doc
Hplc doc
 
Survey on Declining Curves of Unconventional Wells and Correlation with Key ...
Survey on Declining Curves of Unconventional Wells  and Correlation with Key ...Survey on Declining Curves of Unconventional Wells  and Correlation with Key ...
Survey on Declining Curves of Unconventional Wells and Correlation with Key ...
 
Why polymerics as hplc media and why simulated
Why polymerics as hplc media and why simulatedWhy polymerics as hplc media and why simulated
Why polymerics as hplc media and why simulated
 
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...New Technique for Measuring and Controlling the Permeability of Polymeric Mem...
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...
 
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...New Technique for Measuring and Controlling the Permeability of Polymeric Mem...
New Technique for Measuring and Controlling the Permeability of Polymeric Mem...
 

UPLC INTRODUCTION AND REVIEW

  • 1. Journal of Liquid Chromatography & Related Technologies*, 28: 1253-1263, 2005 Copyright © Taylor & Francis, Inc. ISSN 1082-6076 print/1520-572X online DOI: 10.1081/JLC-200053046 TM UPLC : An Introduction and Review Michael E. Swartz Waters Corporation, Milford, Massachusetts, USA Abstract: Ultra performance liquid chromatography™ (UPLC) takes advantage of technological strides made in particle chemistry performance, system optimization, detector design, and data processing and control. Using sub-2 |j,m particles and mobile phases at high linear velocities, and instrumentation that operates at higher pressures than those used in HPLC, dramatic increases in resolution, sensitivity, and speed of analysis can be obtained. This new category of analytical separation science retains the practicality and principles of HPLC while creating a step- function improvement in chromatographic performance. This review introduces the theory of UPLC, and summarizes some of the most recent work in the field. Keywords: UPLC, Ultra performance liquid chromatography INTRODUCTION High performance liquid chromatography (HPLC) has proven to be the predominant technology used in laboratories worldwide during the past 30-plus years. One of the primary drivers for the growth of this technique has been the evolution of packing materials used to effect separations. The underlying principles of this evolution are governed by the van Deemter equation, with which any student of chromatography is intimately familiar. flj The van Deemter equation is an empirical formula that describes the relationship between linear velocity (flow rate) and plate height (HETP, or column efficiency). And, since particle size is one of the variables, a van Deemter curve can be used to investigate chromatographic performance. Address correspondence to Michael E. Swartz, Waters Corporation, 34 Maple Street, Milford, MA 01757, USA; E-mail: michael_swartz@ waters.com
  • 2. 1254 M. E. Swartz For many years, researchers have looked at "fast LC" as a way to speed up analyses.[2'31 The "need for speed" has been driven by the sheer numbers of samples in some laboratories (particularly in drug discovery) and the availability of affordable, easy to use mass spectrometers. Smaller columns and faster flow rates (amongst other parameters) have been used. Elevated temperature, having the dual advantages of lowering viscosity, and increasing mass transfer by increasing the diffusivity of the analytes, has also been investigated.M However, using conventional particle sizes and pressures, limit- ations are soon reached and compromises must be made, sacrificing resolution for time. However, as illustrated in Figure 1, as the particle size decreases to less than 2.5 urn, not only is there a significant gain in efficiency; but the efficiency doesn't diminish at increased flow rates or linear velocities. By using smaller particles, speed and peak capacity (number of peaks resolved per unit time) can be extended to new limits, termed Ultra Performance Liquid Chromatography, or UPLC . This introduction and review traces some of the developments and technological advancements made in producing the first commercially available UPLC instrument. SMALL PARTICLE CHEMISTRY The promises of the van Deemter equation cannot be fulfilled without smaller particles than those traditionally used in HPLC. The design and development 10 Mm Porticle , HPLC H E'5 T P .0 2 3 4 5 f law Kale [ml/mm]: Linear Velocity [u, mm/sec] 10 - 1.0 mm 0.04 0.07 0.10 0.13 0.17 0.20 0.24 10 = 2.1 mm 0.15 0.3 0.45 0.6 0.75 0.9 1.05 !D » 4.6 mm 0.7 1.4 1.1 2.8 3.5 49 Figure 1. van Deemter plot, illustrating the evolution of particle sizes over the last three decades.
  • 3. UPLC : An Introduction and Review 1255 of sub-2 |jim particles is a significant challenge, and researchers have been active in this area for some time to capitalize on their advantages.t5"71 Although high efficiency, non-porous 1.5|j,m particles are commercially available, they suffer from poor loading capacity and retention due to low surface area. To maintain retention and capacity similar to HPLC, UPLC must use novel porous particles that can withstand high pressures. Silica based particles have good mechanical strength, but can suffer from a number of disadvantages, which include a limited pH range and tailing of basic analytes. Polymeric columns can overcome pH limitations, but they have their own issues, including low efficiencies and limited capacities. In 2000, a first generation hybrid chemistry that took advantage of the best of both the silica and polymeric column worlds was introduced.[8'9-1 Produced using a classical sol-gel synthesis that incorporates carbon in the form of methyl groups, these columns are mechanically strong, with high efficiency, and operate over an extended pH range. But, in order to provide the kind of enhanced mechanical stability required for UPLC, a second generation bridged ethane hybrid (BEH) technology was developed.[10] These 1.7|JLm particles derive their enhanced mechanical stability by bridging the methyl groups in the silica matrix. Packing 1.7 (Jim particles into reproducible and rugged columns was also a challenge that needed to be overcome. Requirements include a smoother interior surface of the column hardware, and re-designing the end frits to retain the small particles and resist clogging. Packed bed uniformity is also critical, especially if shorter columns are to maintain resolution while accomplishing the goal of faster separations. In addition, at high pressures, frictional heating of the mobile phase can be quite significant and must be considered.1111 With column diameters typically used in HPLC (3.0 to 4.6mm), a consequence of frictional heating is the loss of performance due to temperature induced non uniform flow. To minimize the effects of frictional heating, smaller diameter columns (1-2.1 mm) are typically used for UPLC.[12'13] CAPITALIZING ON SMALLER PARTICLES Small particles alone do not make it possible to fulfill the promises of the van Deemter equation (Figure 1). Instrument technology also had to keep pace to truly take advantage of the increased speed, superior resolution, and sensitivity afforded by smaller particles. Standard HPLC technology (pumps, injectors, and detectors) simply doesn't have the horsepower to take full advantage of sub-2 (Jim particles. One-of-a-kind systems, capable of delivering the pressures required to realize the potential of UPLC have been reported in the literature and elsewhere.114"16]
  • 4. 1256 M. E. Swartz Lee et al. described the design of injection valves and separation reprodu- cibility,[I4] and the use of a carbon dioxide enhanced slurry packing method on the capillary scale for the separation of some benzodiazepines, herbicides, and various pharmaceutical compounds.1171 Jorgenson et al. modified a commer- cially available HPLC system to operate at 17,500 psi and used 22 cm long capillaries packed with 1.5|jim Gig-modified particles for the analysis of proteins.1-15-1 These reports illustrated that, to take full advantage of low dispersion and small particle technology to achieve high peak capacity UPLC separations, a greater pressure range than that achievable by today's HPLC instrumentation was required. The calculated pressure drop at the optimum flow rate for maximum efficiency across a 15cm long column packed with 1.7 [Am particles is about 15,000 psi. Therefore, a pump capable of delivering solvent smoothly and reproducibly at these pressures, that can compensate for solvent compressibility, and can operate in both the gradient and isocratic separation modes, was required. Sample introduction is also critical. Conventional injection valves, either automated or manual, are not designed and hardened to work at extreme pressure. To protect the column from experiencing extreme pressure fluctu- ations, the injection process must be relatively pulse-free. The swept volume of the device also needs to be minimal to reduce potential band spreading. A fast injection cycle time is needed to fully capitalize on the speed afforded by UPLC which, in turn, requires a high sample capacity. Low volume injections with minimal carryover are also required to realize the increased sensitivity benefits. With 1.7 (jum particles, half-height peak widths of less than one second are obtained, posing significant challenges for the detector. In order to accu- rately and reproducibly integrate an analyte peak, the detector sampling rate must be high enough to capture enough data points across the peak. In addition, the detector cell must have minimal dispersion (volume) to preserve separation efficiency. Conceptually, the sensitivity increase for UPLC detection should be 2-3 times higher than with HPLC separations, depending on the detection technique that is used. Conventional absorbance- based optical detectors are concentration sensitive detectors and, for UPLC use, the flow cell volume would have to be reduced in standard UV/Visible detectors to maintain concentration and signal, while avoiding Beers' Law limitations. In early 2004, the first commercially available UPLC system that embodied these requirements was described for the separation of various pharmaceutical related small organic molecules, proteins, and peptides; it is called the ACQUITY UPLC™ System.[18-20] Using UPLC, it is now possible to take full advantage of chromatographic principles to run separations using shorter columns, and/or higher flow rates for increased speed, with superior resolution and sensitivity. Figures 2 and 3
  • 5. UPLC : An Introduction and Review 1257 E r- -e bo r^i c Q) £ © oa ecu" CD — .2 3 Q tj £ §st S u a i > £ ®I 1 oo cd o a i- |o^ u 2S s .a I- * CD — fj •— >~> "!L C g g u f 3 _-0 5 2 '£ y c • c o •111 1 ES •o vo "o ." o -o . o si s S 2I 2 o 6 2 00 S a N H S 5 & B
  • 6. 0,08 004 002 0.08 0.06 002 0.00 0.00 5.00 10.00 15.00 20.00 25,00 30-00 35,00 40.00 45.00 50.00 55-00 60.00 Minutes Figure 3. HPLC vs. UPLC peak capacity. In this gradient peptide map separation, the HPLC (top) separation (on a 5 ^.m CIS column) yields 70 V. peaks, or a peak capacity of 143, while the UPLC separation (bottom) run under identical conditions yields 168 peaks, or a peak capacity of 360, a ^ 2.5X increase. 3. N
  • 7. UPLC : An Introduction and Review 1259 illustrate UPLC in action. In Figure 2 a separation of eight diuretics is accom- plished in under 1.6 minutes. The same separation on a 2.1 by 100mm 5 |a,m Cig HPLC column yields comparable resolution, but takes over ten minutes. For some analyses, however, speed is of secondary importance; peak capacity and resolution take center stage. Figure 3 shows a peptide map where the desired goal is to maximize the number of peaks. In this application, the increased peak capacity (number of peaks resolved per unit time) of UPLC dramatically improves the quality of the data resulting in a more definitive map APPLICATIONS Chromatographers are accustomed to making compromises; one of the most common scenarios involves sacrificing resolution for speed. In addition, for complex samples like natural product extracts, added resolution can provide more information in the form of additional peaks. Figure 4 shows an HPLC 0.8AUFS HPLC 10jiL Injection Time in Minutes 2.1 AUFS UPLC™ 5^L Injection Time in Minutes Figure 4. Comparison HPLC and. UPLC for the separation of a ginger root extract. HPLC conditions: Column: 2.1 by 100mm 5.0|a.m prototype BEH CIS at 28°C. A 25-96%B linear gradient over 10 minutes, at a flow rate of 1.0 mL/min was used. Mobile phase A was water, B was acetonitrile. UV detection @ 230 nm, 10|j.L injection. UPLC conditions: Column: 2.1 by 100mm 1.7|j.m ACQUITY BEH CIS at 28°C. A 50-100%B linear gradient from 1.4 to 3.7 minutes, followed by a hold until 6.0 minutes, at a flow rate of 0.3 mL/min was used. Mobile phase A was water, B was acetonitrile. UV detection @ 230 nm, 5 (xL injection.
  • 8. 1260 M. E. Swartz versus UPLC separation comparison of a ginger root extract sample where both speed and resolution are improved, as well as an increase in sensitivity. DryLab software was used to model and redevelop the separation and transfer it to the ACQUITY UPLC System and BEH chemistry. Faster separations can lead to higher throughput and time savings when running multiple samples. But, a significant amount of time can also be consumed in developing the method in the first place. Faster, higher resolution UPLC separations can cut method development time from days, to hours, or even minutes. Figure 5 is an example of an UPLC separation of several closely related coumarins and a metabolite that was developed in under an hour, including UPLC scouting runs for gradient optimization, and individual runs for elution order identification. These runs were performed in a fraction of the time that would be necessary by conventional HPLC, resulting in significant time savings in the method development laboratory. As alluded to previously, mass spectrometry has gained widespread acceptance as an analytical tool for the qualitative and quantitative analysis 0.03 3 4 AUFS 0.00 1.30 Figure 5. UPLC Separation of Seven Coumarins illustrating fast method develop- ment. Column: 2.1 by 30mm 1.7 |xm ACQUITY UPLC BEH Cjg @ 35°C. A 20-40%B linear gradient over 1.0 minute, at a flow rate of 0.86mL/min was used. Mobile phase A was 0.1 % formic acid, B was acetonitrile. UV detection @ 254 nm and 40 pts/sec. Peaks are in order: 1: 7-hydroxycoumarin-glucuronide, 7-hydroxy- coumarin, 4-hydroxycoumarin, coumarin, 7-methoxycoumarin, 7-ethoxycoumarin, and 4-ethoxycoumarin.
  • 9. • T0= M5 ES* 4 U-J BPI I ,,14 41 ' 56e-l 13 68 PI 200 4 DO B [iD 8 [iD ID DO 12.DD 14 OD 1600 1600 20.00 22.00 2400 2G OD 2B OD 30 DO 3200 34 DO Figure 6. Separation of rat bile following the administration of midolazam at 5 mg/kg: A) 30 minute separation on a 2.1 by 100mm 3.5 (xm CIS HPLC Column and B) 30minute separation on a 2.1 by 100mm 1.7 |j.m CIS UPLC column. Reprinted by permission from reference 21.
  • 10. i-iu-i wi. &. swartz of many types of compounds. MS detection is significantly enhanced by UPLC; increased peak concentrations with reduced chromatographic dispersion at lower flow rates (no flow splitting) promotes increased source ionization efficiencies. Jorgenson et al. have shown that higher chromato- graphic efficiency, resulting from the use of UPLC, translates into better resolution and higher peak capacity, which is particularly important for the analysis of peptides and proteins.[15] The increased resolving power made the resulting data easier to interpret, since more of the MS peaks consisted of a single compound, and up to a 20-fold improvement in the quality of the spectral information (vs. nanoelectrospray) was obtained. Lee et al. also used MS detection for the analysis of low molecular weight compounds similar to those that might comprise a combinatorial library.[14'17] It was demonstrated that, in order to address the very narrow peaks produced by UPLC, it is necessary to use a very high data capture rate MS such as a TOP or quadrapole with fast scan rates. Lee et al. also pointed out that, in some instances, related compounds of the same molecular weight and similar structures could not be differentiated by MS, necessitating chromatographic resolution on the UPLC time scaled1' Plumb et al. have investigated the use of UPLC/MS for the analysis of metabolites,121'22] and as a tool for differential metabolic pathway profiling in functional genomic studies.[23] Their data illustrate the benefit obtained from the extra resolution of UPLC, both in terms of specificity and spectral quality, revealing new information and reducing the risk of not detecting potentially important metabolites. Figure 6 shows the benefits of UPLC versus HPLC for monitoring the in-vivo metabolism of midazolam (an anti- convulsant) in rat liver bile. This is a challenging separation due to the high concentration of bile salts that can interfere, and the presence of bilirubin that can cause ion pairing. The resolution is dramatically improved, and the number of discreet peaks has more than doubled. CONCLUSION UPLC presents the ability to extend and expand the utility of separation science at a time when many scientists have reached separation barriers, pushing the limits of conventional HPLC. New chemistry and instrumentation technology can provide more information per unit of work as UPLC begins to fulfill the promise of increased speed, resolution, and sensitivity predicted for liquid chromatography. ACKNOWLEDGMENTS The author would like to acknowledge the contributions of the ACQUITY program team at Waters, particularly Eric Grumbach, Michael Jones,
  • 11. UPLC : An Introduction and Review 1263 Marianna Kele, Rob Plumb, and Jeff Mazzeo for their contributions to this manuscript. ACQUITY UPLC, UPLC, and Ultra Performance LC are trademarks of Waters Corporation. REFERENCES 1. van Deemter, J.J.; Zuiderweg, F.J.; Klinkenberg, A. Chem. Eng. Sci. 1956, 5, 271. 2. Cheng, Y.-F.; Lu, Z.; Neue, U. Rapid Commun. Mass Spectrom. 2001, 15, 141. 3. Tiller, P.R.; Romanyshyn, L.A.; Neue, U.D. Anal. Bioanal. Chem. 2003, 377, 788. 4. Neue, U.D.; Mazzeo, J.R. J. Sep. Sci. 2001, 24, 1. 5. Jerkovitch, A.D.; Mellors, J.S.; Jorgenson, J.W. LC-GC North Amer. 2003, 21,1. 6. Wu, N.; Lippert, J.A.; Lee, M.L. J. Chromotogr. 2001, 911, 1. 7. Unger, K.K.; Kumar, D.; Grun, M.; Buchel, G.; Ludtke, S.; Adam, Th.; Scumacher, K.; Renker, S. J. Chromatogr. A 2000, 892, 47. 8. Neue, U.D.; Walter, T.H.; Alden, B.A.; Jiang, Z.; Fisk, R.P.; Cook, J.T.; Close, K.H.; Carmody, J.L.; Grassi, J.M.; Cheng, Y.-F.; Lu, Z.; Crowley, R.J. Amer. Lab. 1999, 31, 36. 9. Cheng, Y.-F.; Walter, T.H.; Lu, Z.; Iraneta, P.; Gendreau, C.; Neue, U.D.; Grassi, J.M.; Carmody, J.L.; O'Gara, J.E.; Fisk, R.P. LC-GC 2000, 18, 1162. 10. Mazzeo, J.R. Anal. Chem. November 2004, In press. 11. Halasz, I.; Endele, R.; Asshauer, J. J. Chromatogr. 1975, 112, 37. 12. MacNair, J.E.; Lewis, K.C.; Jorgenson, J.W. Anal. Chem. 1997, 69, 983. 13. Colon, L.A.; Citron, J.M.; Anspach, J.A.; Fermier, A.M.; Swinney, K.A. Analyst 2004, 729, 503. 14. Wu, N.; Lippert, J.A.; Lee, M.L. J. Chromatogr. A 2001, 911, 1. 15. Tolley, L.; Jorgenson, J.W.; Mosely, M.A. Anal. Chem. 2001, 73, 2985. 16. Swinney, K.A.; Fermier, A.M.; Guo, Y.; Oyler, A.R.; Segmuller, B.E.; Dunphy, R. Presentation #L1703 at the 28th Intl. Symp. on High Perform. Liq. Phase Sepns. & Rel. Techn, Philadelphia, PA, June 2004. 17. Lippert, J.A.; Xin, B.; Wu, N.; Lee, M.L. J. Microcol. Sepn. 1997, 11, 631. 18. Swartz, M.E. Presentation at the Pitts, Conf. on Anal. Chem. Appl. Spectrosc, Chicago, IL, 2004. 19. Swartz, M.E.; Murphy, B. Lab Plus Intl. 2004, 18, 6. 20. Swartz, M.E.; Murphy, B. Pharm.l Formul. Qual. November 2004, In press. 21. Plumb, R.S.; Castro-Perez, J.; Granger, J.H.; Beattie, I.; Joncour, K.; Wright, A. Rapid Commun. Mass Spectrom. 2004, 18, 2331. 22. Castro-Perez, J.; Plumb, R.; Granger, J.H. Rapid Commun. Mass Spectrom. October 2004, Submitted. 23. Wilson, I.D.; Nicholson, J.K.; Castro-Perez, J.; Granger, J.H.; Smith, B.W.; Mazzeo, J.; Plumb, R.S. J. Proteome Res. October 2004, Submitted. Received November 22, 2004 Accepted December 19, 2004 Manuscript 6549E