SlideShare uma empresa Scribd logo
1 de 89
Dharam Prakash Saran
1
2
Intra aortic balloon counter pulsation( IABP):
Most common and widely available methods of mechanical
circulatory support
Temporary support for the left ventricle by mechanically
displacing blood within the aorta
Concepts:
- Systolic unloading
- Diastolic augmentation
Traditionally used in surgical and non surgical patients
with cardiogenic shock
3
Indications for IABP
1. Cardiogenic shock:
2. In association with CABG :
Preoperative insertion
- Patients with severe LV dysfunction
- Patients with intractable ischemic arrhythmias
Postoperative insertion
- Postcardiotomy cardiogenic shock
- Associated with acute MI
- Mechanical complications of MI - MR , VSD
3. In association with nonsurgical revascularization:
-Hemodynamically unstable infarct patients
-High risk coronary interventions
- severe LV dysfunction, LMCA, complex coronary artery disease
4. Stabilization of cardiac transplant recipient before insertion of VAD
Post infarction angina
Ventricular arrhythmias relathed to ischemia
4
CONTRAINDICATIONS
Absolute-
▫ Significant aortic regurgitation
▫ Aortic dissection
▫ Aortic stents
▫ Bilateral femoral popliteal bypass grafts for severe
PVD
Relative -
▫ Abdominal aortic aneurysm
▫ Uncontrolled septicemia
▫ Uncontrolled bleeding diathesis
▫ Severe bilateral peripheral vascular disease
6
7
8
Interpreting IABP waveforms
9
IABP –instrumentation and
techniques
10
The IAB Counter pulsation system
- two principal parts
 A flexible catheter -2 lumen
• first - for distal aspiration/flushing or pressure monitoring
• second - for the periodic delivery and removal of helium gas to a
closed balloon.
 A mobile console
• system for helium transfer
• computer for control of the inflation and deflation cycle
11
12
Expected changes with IABP support in hemodynamic profile in
patients with Cardiogenic shock
- Decrease in SBP by 20 %
- Increase in aortic Diastolic Press. by 30 % ( raise coronary blood flow)
- Increase in MAP
- Reduction of the HR by 20%
-Decrease in the mean PCWP by 20 %
- Elevation in the COP by 20%
13
IABP catheter:
 10-20 cm long polyurethane bladder
 25cc to 50cc capacity
 Optimal 85% of aorta occluded (not 100%)
 The shaft of the balloon catheter contains 2 lumens:
- one allows for gas exchange from console to
balloon
- second lumen
- for catheter delivery over a guide wire
- for monitoring of central aortic pressure
after installation.
14
Intra Aortic Balloon
15
Balloon sizing
 Sizing based on patients
height
 Four common balloon
sizes
 Balloon length and
diameter increases with
each larger size
 40 cm³ balloon is most
commonly used
 Paediatric balloons also
available : sizes 2.5, 5.0,
12.0 and 20 cm³
Balloon size Height
50 cm³ > 6 feet
40 cm³ 5 feet 4 inch
to 6 feet
34 cm³ 5 feet to 5
feet 4 inch
25 cm³ < 5 feet
16
Benefits of larger volume IABs
• More blood volume displacement
• More diastolic augmentation
• More systolic unloading
17
Introducer needle
• Guide wire
• Vessel dilators
• Sheath
• IABP (34 or 40cc)
• Gas tubing
• 60-mL syringe
• Three-way stopcock
IABP Kit Contents
18
Connect ECG
Set up pressure lines
Femoral access – followed by insertion of the supplied
sheath
0.030 inch supplied J-shaped guide wire to the level of the
aortic arch (LAO view)
- IABP insertion
19
Take the entire catheter and T handle as one unit
(DO NOT disconnect one-way
valve when removing the
extracorporeal tubing from the
tray.)
Pull out the T- handle only as shown
20
• Remove stylet/aspirate/Flush
• Insert the balloon only over the guide wire
• Hold the catheter close to skin insertion point
• Advance in small steps of 1 to 2 cm at a time and
stop if any resistance.
• The IABP should advance freely
Inserting the Balloon catheter
- Many vascular complications occur during insertion itself
- Resistance during insertion either indicates PVOD, or dissection
- Kinking of IABP » improper inflation/deflation
21
- The end of the balloon should be just distal (1-2 cm) to the takeoff of the
left subclavian artery
- Position should be confirmed by fluoroscopy or chest x-ray
Positioning
22
Connecting to console:
- Connect helium gas tube to the console via a long extender
- Open helium tank.
- The central lumen of the catheter is flushed and connected to pressure
tubing with 3 way and then to a pressure transducer to allow for
monitoring of central aortic pressure.
- Zero the transducer
Initial set-up:
- Once connected properly the console would show ECG and pressure
waveforms.
- Check Basal mean pressure
- Make sure the setting is at “auto”
- Usually IABP started at 1:1 or 1:2 augmentation
- Usually Augmentation is kept at maxim
23
24
25
Trigger modes
Trigger :
- Event the pump uses to identify the onset of cardiac cycle (systole)
- Pump must have consistent trigger in order to provide patient assist
- If selected trigger not detected, counter pulsation will interrupted
1.ECG
- uses the slope of QR segment to detect triggering point
2. AP(Arterial pressure wave)
- Systolic upstroke of the arterial pressure wave form is the trigger
26
ECG signal – most common
• Inflation
- middle of T wave
• Deflation
– peak of R wave
• Pacer (v/a)
• Arterial waveform
• An intrinsic pump rate
(VF, CPB)
27
28
29
30
Increased coronary
perfusion
The “normal” augmented waveform
31
Not all Sub optimal augmentation is due to Timing
errors/kinks
32
Factors affecting diastolic augmentation
Patient
- Heart rate
- Mean arterial pressure
- Stroke volume
- Systemic vascular resistance
Intra aortic balloon catheter
- IAB in sheath
- IAB not unfolded
- IAB position
- Kink in the IAB catheter
- IAB leak
- Low helium concentration
Intra aortic balloon pump
- Timing
- Position of IAB augmentation control
33
How to check waveform is acceptable ?
 First change from 1:1 to 1:2 augmentation
 Check the dicrotic notch
 See if augmentation starts at that point
This should produce a sharp “V” at inflation.
34
How to check waveform is acceptable ?
 First change from 1:1 to 1:2 augmentation
 Check the dicrotic notch
 See if augmentation starts at that point
This should produce a sharp “V” at inflation.
 Check if diastolic augmented wave is › systolic wave
35
How to check waveform is acceptable ?
 First change from 1:1 to 1:2 augmentation
 Check the dicrotic notch
 See if augmentation starts at that point
This should produce a sharp “V” at inflation.
 Check if diastolic augmented wave is › systolic wave
 Confirm if end diastolic wave
following the augmented wave
is less than an non augmented
wave.
 Is Deflation slope ok
36
37
Late Inflation
 Inflation of the IAB markedly after closure of the aortic valve.
 Waveform Characteristics:
• Inflation of IAB after the dicrotic notch.
• Absence of sharp V.
• Sub optimal diastolic augmentation
38
Early Deflation
 Premature deflation of the IAB during the diastolic phase.
39
Late Deflation
 Late deflation of the IAB during the diastolic phase.
 Waveform Characteristics:
• Assisted aortic end diastolic pressure may be equal to the
unassisted aortic end diastolic pressure.
• Rate of rise of assisted systole is prolonged.
• Diastolic augmentation may appear widened
40
41
Variation in balloon pressure wave forms
Increased duration of
plateau due to longer
diastolic phase
Decreased duration
of plateau due to
shortened diastolic
phase
42
Variation in balloon pressure wave forms
Varying R-R intervals
result in irregular
plateau durations
43
Variation in balloon pressure wave forms
Increased height
or amplitude of
the waveform
Decreased height
or amplitude of the
waveform
44
Variation in balloon pressure wave forms
Gas leak
Leak in the closed system causing the
balloon pressure waveform to fall below
zero baseline..
- due to a loose connection
- a leak in the IAB catheter
- H2O condensation in the external tubing
- a patient who is tachycardiac and febrile which causes increased gas
diffusion through the IAB membrane
45
Catheter Kink
Rounded balloon pressure waveform
- Loss of plateau resulting from a kink or
obstruction of shuttle gas
- Kink in the catheter tubing
- Improper IAB catheter position
- Sheath not being pulled back to allow
inflation of the IAB
- IAB is too large for the aorta
- IAB is not fully unwrapped
- H2O condensation in the external tubing
46
“Balloon too
large”
syndrome
47
Patient Management During IABP support
 Anticoagulation-- maintain apTT at 50 to 70 seconds
 CXR daily – to R/O IAB migration
 Check lower limb pulses - 2 hourly.
- If not palpable » ? - vascular obstruction
- thrombus, embolus, or dissection
(urgent surgical consultation)
 Prophylactic antibiotics NOT INDICATED
 Hip flexion is restricted, and the head of the bed should not be
elevated beyond 30°.
48
 Never leave in standby by mode for more than 20 minutes >
thrombus formation
 Daily
– Haemoglobin (risk of bleeding or haemolysis)
– Platelet count (risk of thrombocytopenia)
– Renal function (risk of acute kidney injury secondary to distal
migration of IABP catheter)
 Wean off the IABP as early as possible as longer duration is associated
with higher incidence of limb complications
Patient Management During IABP support
49
50
Weaning of IABP
Timing of weaning:
- Patient should be stable for 12 – 24 hours
- Decrease inotropic support
- Decrease pump ratio
– From 1:1 to 1:2 or 1:3
- Decrease augmentation
- Monitor patient closely
– If patient becomes unstable, weaning should be
immediately discontinued
51
IABP Removal
-Discontinue heparin 1 hour prior to removal
-Disconnect the IAB catheter from the IAB pump
- Patient blood pressure will collapse the balloon membrane for withdrawal
- Withdraw the IAB catheter through the introducer sheath until resistance
is met.
- NEVER attempt to withdraw the balloon membrane through the
introducer sheath.
-Remove the IAB catheter and the introducer sheath as a unit
- Check for adequacy of limb perfusion after hemostasis is achieved.
ACC/AHA 2013Practice Guidelines (ESC
2012- IIb)
52
ESC 2014 GUIDELINES
53
IABP IN UA/NSTEMI (2007, 2012
update)
(Current Practices)(Class IIa LOE-©)
54
The placement of an IABP could be useful in
patients with recurrent ischemia despite
maximal medical management and in those
with hemodynamic instability until coronary
angiography and revascularization can be
completed.
TRIALS OF IABP
55
Benchmark registry
• n = 17,000( june 1996- aug 2000)
• 203 hospitals- 90% US
• 18.8% of IABP used for cardiogenic shock
• Device related death – 0.5%
• Major complication – 2.6%
• Minor complication – 4.2%
56
Balloon-pump assisted Coronary Intervention
Study (BCIS-1):
• The first randomized controlled trial of elective
Intra-Aortic Balloon Pump (IABP) insertion prior to
high-risk PCI vs. PCI with no planned IABP use
• 17 UK centres
• n=301 (150 in each arm)
57
58IABP in high risk PCI(BCIS-1 TRIAL)
N= 301 Elective IABP(
151)
No elective
IABP(150)
P VALUE
MACE 15.2% 16% 0.85
All cause mortality
at 6 mths
4.6% 7.4% 0.32
Major procedural
complications
1.3% 10.7% <0.001
Major or minor
bleeding
19.2% 11.3% 0.06
Access site
complications
3.3% 0% 0.06
Patients (n = 301) had severe left ventricular dysfunction (ejection
fraction ≤ 30%) and extensive coronary disease (Jeopardy Score ≥
8/12); those with contraindications to or class I indications for IABP
therapy were excluded
JAMA. 2010;304(8):867-874
59
Conclusions of long term results of
BCIS1 trial(2012-2013)
In patients with severe ischemic cardiomyopathy treated
with PCI, all cause-mortality was 33% at 51 months
(median)
Elective IABP use during PCI was associated with an
observed 34% reduction in long-term all-cause mortality
60
Counterpulsation Reduces Infarct Size Acute Myocardial
Infarction (CRISP AMI) trial.
Intra-aortic balloon pump counterpulsation prior to PCI in
patients with ST segment elevation MI without shock does
not reduce infarct size as measured by MRI
Shock trial and registry
61
62
N-302 pts
N- 152 pts N- 150 pts
63
SHOCK Trial
Primary and Secondary Endpoints
0
20
40
60
80
30 Days 6 months
Immediate
Revascularization
Strategy
Medical Stabilization
as an Initial Strategy
Primary
Endpoint
Secondary
Endpoint
Mortality(%)
46.7
%
56.0
% 50.3
%
63.1
%
P=.11
P= .027
Hochman et al, NEJM 1999; 341:625.
64
Impact of thrombolysis, intra-aortic balloon
pump counterpulsation, and their
combination in cardiogenic shock
complicating acute myocardial infarction: a
report from the SHOCK Trial Registry
65
SHOCK Registry: Impact of Thrombolytics
and IABP
0
20
40
60
80
47%
52%
%
P<0.0001
63%
77%
Thrombolytics
+ IABP
No
Thrombolytics
+ IABP
Thrombolytics
+ No IABP
Neither
Hochman et al, NEJM 1999; 341:625
66
Intraaortic Balloon Support for
Myocardial Infarction with
Cardiogenic Shock
IABP Shock II Trial
Conclusion
• The use of intraaortic balloon counterpulsation
did not significantly reduce 30-day mortality in
patients with cardiogenic shock complicating
acute myocardial infarction for whom an early
revascularization strategy was planned.
67
MCQs
68
• 1. Major physiological effects of counter pulsation
include?
▫ A) increased coronary artery perfusion, increased preload,
decreased after load, decreased myocardial oxygen consumption
▫ B) increased coronary artery perfusion, increased preload,
increased after load, decreased myocardial oxygen consumption
▫ C) increased coronary artery perfusion, decreased preload,
decreased after load, increased myocardial oxygen consumption
▫ D) increased coronary artery perfusion, decreased preload,
decreased after load, decreased myocardial oxygen consumption
69
2. The dicrotic notch on the arterial wave form
reflects
A) aortic valve opening
B) aortic valve closure
C) isovolumetric contraction
D)rapid ejection
70
3. Expected changes with IABP support in hemodynamic
profile in patients with Cardiogenic shock include all
except?
A) Decrease in SBP by 20 %
B) Increase in aortic DP by 30 %
C) Decrease in MAP by 10%
D) Reduction of the HR by 20%
E)Decrease in the mean PCWP by 20 %
71
4. late inflation of the balloon can result in?
A) premature augmentation
B) increased augmentation
C) decreased augmentation
D) increased coronary perfusion
72
5. A rounded balloon pressure wave form
indicate?
A) helium leak
B) power failure
C) hypovolemia
D) balloon occluding the aorta
73
6. width of balloon pressure wave form
corresponds to
A) length of systole
B) length of diastole
C) arterial pressure
D) helium level
74
7. true statement
a) Dicrotic notch- land mark used to set deflation
b) Deflation is timed to occur during period of iso
volumetric contraction
c) Most common trigger used is arterial pressure
wave method
d) Internal trigger mode is acceptable to use in a
patient with normal sinus rhythm
75
8. true statement
A) pacing spikes are automatically rejected in ECG
triggered modes
B) pacing trigger modes can be used in a patient
of 50% paced rhythm
C) Varying R-R interval result in regular plateau
durations in Balloon pressureWave form
76
9. Identify the tracing abnormality
77
10. Identify the tracing abnormality
78
79
• 1. Major physiological effects of counter pulsation
include?
▫ A) increased coronary artery perfusion, increased preload,
decreased after load, decreased myocardial oxygen consumption
▫ B) increased coronary artery perfusion, increased preload,
increased after load, decreased myocardial oxygen consumption
▫ C) increased coronary artery perfusion, decreased preload,
decreased after load, increased myocardial oxygen consumption
▫ D) increased coronary artery perfusion, decreased preload,
decreased after load, decreased myocardial oxygen consumption
80
2. the dicrotic notch on the arterial wave form
reflects
A) aortic valve opening
B) aortic valve closure
C) isovolumetric contraction
D)rapid ejection
81
3. Expected changes with IABP support in hemodynamic
profile in patients with Cardiogenic shock include all
except?
•A) Decrease in SBP by 20 %
•B) Increase in aortic DP by 30 %
•C) Decrease in MAP by 10%
•D) Reduction of the HR by 20%
•E)Decrease in the mean PCWP by 20 %
82
4. late inflation of the balloon can result in?
• A) premature augmentation
• B) increased augmentation
• C) decreased augmentation
• D) increased coronary perfusion
83
5. A rounded balloon pressure wave form
indicate?
• A) helium leak
• B) power failure
• C) hypovolemia
• D) balloon occluding the aorta
84
6. width of balloon pressure wave form
corresponds to
• A) length of systole
• B) length of diastole
• C) arterial pressure
• D) helium level
85
7. true statement
a) Dicrotic notch- land mark used to set deflation
b) Deflation is timed to occur during period of iso
volumetric contraction
c) Most common trigger used is arterial pressure
wave method
d) Internal trigger mode is acceptable to use in a
patient with normal sinus rhythm
86
8. true statement
A) pacing spikes are automatically rejected in ECG
triggered modes
B) pacing trigger modes can be used in a patient
of 50% paced rhyth
C) Varying R-R interval result in regular plateau
durations in Balloon press. Wave form
87
9.
88
10.
89

Mais conteúdo relacionado

Mais procurados

Iabp instrumentation, indications and complications
Iabp  instrumentation, indications and complicationsIabp  instrumentation, indications and complications
Iabp instrumentation, indications and complicationsteja bayapalli
 
Intra Aortic Balloon Pump
Intra Aortic Balloon PumpIntra Aortic Balloon Pump
Intra Aortic Balloon Pumpdstroup461
 
Iabp instrumentation, indications and complications
Iabp  instrumentation, indications and complicationsIabp  instrumentation, indications and complications
Iabp instrumentation, indications and complicationsManu Jacob
 
Transesophageal echocardiography
Transesophageal echocardiographyTransesophageal echocardiography
Transesophageal echocardiographyAmit Gulati
 
IABP- intra Aortic Balloon Pump
IABP- intra Aortic Balloon PumpIABP- intra Aortic Balloon Pump
IABP- intra Aortic Balloon PumpSOHAM SHAH
 
cardioplegia delivery management, perfusion safety
cardioplegia delivery management, perfusion safetycardioplegia delivery management, perfusion safety
cardioplegia delivery management, perfusion safetyIda Simanjuntak
 
ECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenationECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenationprapulla chandra
 
Ultrafiltration during cardiopulmonary_bypass
Ultrafiltration during cardiopulmonary_bypassUltrafiltration during cardiopulmonary_bypass
Ultrafiltration during cardiopulmonary_bypassdr amarja nagre
 
Pulmonary artery catheter
Pulmonary artery catheterPulmonary artery catheter
Pulmonary artery catheterrajkumarsrihari
 
Basic principles of myocardial proctection
Basic principles of myocardial proctectionBasic principles of myocardial proctection
Basic principles of myocardial proctectionRaja Lahiri
 
BALLOON AORTIC VALVULOPLASTY
BALLOON AORTIC VALVULOPLASTYBALLOON AORTIC VALVULOPLASTY
BALLOON AORTIC VALVULOPLASTYPraveen Nagula
 
INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING vikramnaidu2311
 

Mais procurados (20)

Iabp instrumentation, indications and complications
Iabp  instrumentation, indications and complicationsIabp  instrumentation, indications and complications
Iabp instrumentation, indications and complications
 
ECMO
ECMOECMO
ECMO
 
Cardioplegia
CardioplegiaCardioplegia
Cardioplegia
 
IABP when , why and how ?
IABP when , why and how ?IABP when , why and how ?
IABP when , why and how ?
 
Intra Aortic Balloon Pump
Intra Aortic Balloon PumpIntra Aortic Balloon Pump
Intra Aortic Balloon Pump
 
Transesophageal echocardiography(TEE)
Transesophageal echocardiography(TEE)Transesophageal echocardiography(TEE)
Transesophageal echocardiography(TEE)
 
Iabp instrumentation, indications and complications
Iabp  instrumentation, indications and complicationsIabp  instrumentation, indications and complications
Iabp instrumentation, indications and complications
 
Transesophageal echocardiography
Transesophageal echocardiographyTransesophageal echocardiography
Transesophageal echocardiography
 
Lvad
LvadLvad
Lvad
 
IABP- intra Aortic Balloon Pump
IABP- intra Aortic Balloon PumpIABP- intra Aortic Balloon Pump
IABP- intra Aortic Balloon Pump
 
OPCAB
OPCABOPCAB
OPCAB
 
Cardioversion
Cardioversion Cardioversion
Cardioversion
 
cardioplegia delivery management, perfusion safety
cardioplegia delivery management, perfusion safetycardioplegia delivery management, perfusion safety
cardioplegia delivery management, perfusion safety
 
ECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenationECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenation
 
Ultrafiltration during cardiopulmonary_bypass
Ultrafiltration during cardiopulmonary_bypassUltrafiltration during cardiopulmonary_bypass
Ultrafiltration during cardiopulmonary_bypass
 
Pulmonary artery catheter
Pulmonary artery catheterPulmonary artery catheter
Pulmonary artery catheter
 
Basic principles of myocardial proctection
Basic principles of myocardial proctectionBasic principles of myocardial proctection
Basic principles of myocardial proctection
 
Iabp
IabpIabp
Iabp
 
BALLOON AORTIC VALVULOPLASTY
BALLOON AORTIC VALVULOPLASTYBALLOON AORTIC VALVULOPLASTY
BALLOON AORTIC VALVULOPLASTY
 
INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING INTRA-ARTERIAL BLOOD PRESSURE MONITORING
INTRA-ARTERIAL BLOOD PRESSURE MONITORING
 

Semelhante a IABP

Intra aortic balloon pump and ECMO
Intra aortic balloon pump and ECMOIntra aortic balloon pump and ECMO
Intra aortic balloon pump and ECMORachel Jeevakirubai
 
Intra Aortic Balloon Pump (IABP) 2009.ppt
Intra Aortic Balloon Pump (IABP) 2009.pptIntra Aortic Balloon Pump (IABP) 2009.ppt
Intra Aortic Balloon Pump (IABP) 2009.ppttaimourali64
 
Ventricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac CathetherizationVentricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac Cathetherizationvasanth7pv
 
Cardiovascular monitoring Part II
Cardiovascular monitoring Part IICardiovascular monitoring Part II
Cardiovascular monitoring Part IISiddhanta Choudhury
 
Fontan circulation
Fontan circulationFontan circulation
Fontan circulationShivani Rao
 
Swan-Ganz-catheterisation_amit-panjwani.pdf
Swan-Ganz-catheterisation_amit-panjwani.pdfSwan-Ganz-catheterisation_amit-panjwani.pdf
Swan-Ganz-catheterisation_amit-panjwani.pdframbhoopal1
 
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...prateek gupta
 
Advanced treatment in hf ppt
Advanced treatment in hf pptAdvanced treatment in hf ppt
Advanced treatment in hf pptShalini Garg
 
iabp troubleshooting and management.pdf
iabp  troubleshooting and management.pdfiabp  troubleshooting and management.pdf
iabp troubleshooting and management.pdfNugrahaPranadipta
 
Overview of Intra-Aortic Balloon Pump (IABP)
Overview of Intra-Aortic Balloon Pump (IABP)Overview of Intra-Aortic Balloon Pump (IABP)
Overview of Intra-Aortic Balloon Pump (IABP)Suheil Dhanse
 
Ccpa catheter basics07medicine
Ccpa catheter basics07medicineCcpa catheter basics07medicine
Ccpa catheter basics07medicineanjika
 
Iabp principle, hemodynamic, timing, weaning 2016 background asmiha,isman edit
Iabp principle, hemodynamic, timing, weaning 2016  background asmiha,isman editIabp principle, hemodynamic, timing, weaning 2016  background asmiha,isman edit
Iabp principle, hemodynamic, timing, weaning 2016 background asmiha,isman editIsman Firdaus
 
Dr jeevraj cabg management
Dr jeevraj cabg managementDr jeevraj cabg management
Dr jeevraj cabg managementjeevraj24
 

Semelhante a IABP (20)

Intra aortic balloon pump and ECMO
Intra aortic balloon pump and ECMOIntra aortic balloon pump and ECMO
Intra aortic balloon pump and ECMO
 
IABP - DEV BUCHE
IABP - DEV BUCHEIABP - DEV BUCHE
IABP - DEV BUCHE
 
Intra Aortic Balloon Pump (IABP) 2009.ppt
Intra Aortic Balloon Pump (IABP) 2009.pptIntra Aortic Balloon Pump (IABP) 2009.ppt
Intra Aortic Balloon Pump (IABP) 2009.ppt
 
Ventricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac CathetherizationVentricular assist device of cardiac Cathetherization
Ventricular assist device of cardiac Cathetherization
 
Cardiovascular monitoring Part II
Cardiovascular monitoring Part IICardiovascular monitoring Part II
Cardiovascular monitoring Part II
 
Fontan circulation
Fontan circulationFontan circulation
Fontan circulation
 
IABP chetan
IABP chetanIABP chetan
IABP chetan
 
Swan-Ganz-catheterisation_amit-panjwani.pdf
Swan-Ganz-catheterisation_amit-panjwani.pdfSwan-Ganz-catheterisation_amit-panjwani.pdf
Swan-Ganz-catheterisation_amit-panjwani.pdf
 
icmo.pptx
icmo.pptxicmo.pptx
icmo.pptx
 
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...central venous pressure and intra-arterial blood pressure monitoring. invasiv...
central venous pressure and intra-arterial blood pressure monitoring. invasiv...
 
Advanced treatment in hf ppt
Advanced treatment in hf pptAdvanced treatment in hf ppt
Advanced treatment in hf ppt
 
iabp troubleshooting and management.pdf
iabp  troubleshooting and management.pdfiabp  troubleshooting and management.pdf
iabp troubleshooting and management.pdf
 
IABP troubleshooting
IABP troubleshootingIABP troubleshooting
IABP troubleshooting
 
Overview of Intra-Aortic Balloon Pump (IABP)
Overview of Intra-Aortic Balloon Pump (IABP)Overview of Intra-Aortic Balloon Pump (IABP)
Overview of Intra-Aortic Balloon Pump (IABP)
 
IABP
IABPIABP
IABP
 
Ccpa catheter basics07medicine
Ccpa catheter basics07medicineCcpa catheter basics07medicine
Ccpa catheter basics07medicine
 
IABP
IABPIABP
IABP
 
Iabp principle, hemodynamic, timing, weaning 2016 background asmiha,isman edit
Iabp principle, hemodynamic, timing, weaning 2016  background asmiha,isman editIabp principle, hemodynamic, timing, weaning 2016  background asmiha,isman edit
Iabp principle, hemodynamic, timing, weaning 2016 background asmiha,isman edit
 
Dr jeevraj cabg management
Dr jeevraj cabg managementDr jeevraj cabg management
Dr jeevraj cabg management
 
ICU management of ECMO pt
ICU management of ECMO ptICU management of ECMO pt
ICU management of ECMO pt
 

Último

Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfMedicoseAcademics
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Servicesonalikaur4
 
Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.ANJALI
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any TimeCall Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Timevijaych2041
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceNehru place Escorts
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbaisonalikaur4
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...narwatsonia7
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Bookingnarwatsonia7
 
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersBook Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersnarwatsonia7
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...narwatsonia7
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptxDr.Nusrat Tariq
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...rajnisinghkjn
 
Noida Sector 135 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few C...
Noida Sector 135 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few C...Noida Sector 135 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few C...
Noida Sector 135 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few C...rajnisinghkjn
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAAjennyeacort
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaPooja Gupta
 
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Me
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near MeHigh Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Me
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Menarwatsonia7
 

Último (20)

Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
 
Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.Statistical modeling in pharmaceutical research and development.
Statistical modeling in pharmaceutical research and development.
 
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hsr Layout Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any TimeCall Girls Viman Nagar 7001305949 All Area Service COD available Any Time
Call Girls Viman Nagar 7001305949 All Area Service COD available Any Time
 
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort ServiceCollege Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
College Call Girls Vyasarpadi Whatsapp 7001305949 Independent Escort Service
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service MumbaiLow Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
Low Rate Call Girls Mumbai Suman 9910780858 Independent Escort Service Mumbai
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
 
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jp Nagar Just Call 7001305949 Top Class Call Girl Service Available
 
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...
Russian Call Girls Gunjur Mugalur Road : 7001305949 High Profile Model Escort...
 
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment BookingCall Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
Call Girl Koramangala | 7001305949 At Low Cost Cash Payment Booking
 
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbersBook Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
Book Call Girls in Kasavanahalli - 7001305949 with real photos and phone numbers
 
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
Call Girls Kanakapura Road Just Call 7001305949 Top Class Call Girl Service A...
 
Glomerular Filtration and determinants of glomerular filtration .pptx
Glomerular Filtration and  determinants of glomerular filtration .pptxGlomerular Filtration and  determinants of glomerular filtration .pptx
Glomerular Filtration and determinants of glomerular filtration .pptx
 
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
Dwarka Sector 6 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few Cl...
 
Noida Sector 135 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few C...
Noida Sector 135 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few C...Noida Sector 135 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few C...
Noida Sector 135 Call Girls ( 9873940964 ) Book Hot And Sexy Girls In A Few C...
 
97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA97111 47426 Call Girls In Delhi MUNIRKAA
97111 47426 Call Girls In Delhi MUNIRKAA
 
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service NoidaCall Girls Service Noida Maya 9711199012 Independent Escort Service Noida
Call Girls Service Noida Maya 9711199012 Independent Escort Service Noida
 
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Me
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near MeHigh Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Me
High Profile Call Girls Mavalli - 7001305949 | 24x7 Service Available Near Me
 

IABP

  • 2. 2 Intra aortic balloon counter pulsation( IABP): Most common and widely available methods of mechanical circulatory support Temporary support for the left ventricle by mechanically displacing blood within the aorta Concepts: - Systolic unloading - Diastolic augmentation Traditionally used in surgical and non surgical patients with cardiogenic shock
  • 3. 3 Indications for IABP 1. Cardiogenic shock: 2. In association with CABG : Preoperative insertion - Patients with severe LV dysfunction - Patients with intractable ischemic arrhythmias Postoperative insertion - Postcardiotomy cardiogenic shock - Associated with acute MI - Mechanical complications of MI - MR , VSD 3. In association with nonsurgical revascularization: -Hemodynamically unstable infarct patients -High risk coronary interventions - severe LV dysfunction, LMCA, complex coronary artery disease 4. Stabilization of cardiac transplant recipient before insertion of VAD Post infarction angina Ventricular arrhythmias relathed to ischemia
  • 4. 4
  • 5. CONTRAINDICATIONS Absolute- ▫ Significant aortic regurgitation ▫ Aortic dissection ▫ Aortic stents ▫ Bilateral femoral popliteal bypass grafts for severe PVD Relative - ▫ Abdominal aortic aneurysm ▫ Uncontrolled septicemia ▫ Uncontrolled bleeding diathesis ▫ Severe bilateral peripheral vascular disease
  • 6. 6
  • 7. 7
  • 10. 10 The IAB Counter pulsation system - two principal parts  A flexible catheter -2 lumen • first - for distal aspiration/flushing or pressure monitoring • second - for the periodic delivery and removal of helium gas to a closed balloon.  A mobile console • system for helium transfer • computer for control of the inflation and deflation cycle
  • 11. 11
  • 12. 12 Expected changes with IABP support in hemodynamic profile in patients with Cardiogenic shock - Decrease in SBP by 20 % - Increase in aortic Diastolic Press. by 30 % ( raise coronary blood flow) - Increase in MAP - Reduction of the HR by 20% -Decrease in the mean PCWP by 20 % - Elevation in the COP by 20%
  • 13. 13 IABP catheter:  10-20 cm long polyurethane bladder  25cc to 50cc capacity  Optimal 85% of aorta occluded (not 100%)  The shaft of the balloon catheter contains 2 lumens: - one allows for gas exchange from console to balloon - second lumen - for catheter delivery over a guide wire - for monitoring of central aortic pressure after installation.
  • 15. 15 Balloon sizing  Sizing based on patients height  Four common balloon sizes  Balloon length and diameter increases with each larger size  40 cm³ balloon is most commonly used  Paediatric balloons also available : sizes 2.5, 5.0, 12.0 and 20 cm³ Balloon size Height 50 cm³ > 6 feet 40 cm³ 5 feet 4 inch to 6 feet 34 cm³ 5 feet to 5 feet 4 inch 25 cm³ < 5 feet
  • 16. 16 Benefits of larger volume IABs • More blood volume displacement • More diastolic augmentation • More systolic unloading
  • 17. 17 Introducer needle • Guide wire • Vessel dilators • Sheath • IABP (34 or 40cc) • Gas tubing • 60-mL syringe • Three-way stopcock IABP Kit Contents
  • 18. 18 Connect ECG Set up pressure lines Femoral access – followed by insertion of the supplied sheath 0.030 inch supplied J-shaped guide wire to the level of the aortic arch (LAO view) - IABP insertion
  • 19. 19 Take the entire catheter and T handle as one unit (DO NOT disconnect one-way valve when removing the extracorporeal tubing from the tray.) Pull out the T- handle only as shown
  • 20. 20 • Remove stylet/aspirate/Flush • Insert the balloon only over the guide wire • Hold the catheter close to skin insertion point • Advance in small steps of 1 to 2 cm at a time and stop if any resistance. • The IABP should advance freely Inserting the Balloon catheter - Many vascular complications occur during insertion itself - Resistance during insertion either indicates PVOD, or dissection - Kinking of IABP » improper inflation/deflation
  • 21. 21 - The end of the balloon should be just distal (1-2 cm) to the takeoff of the left subclavian artery - Position should be confirmed by fluoroscopy or chest x-ray Positioning
  • 22. 22 Connecting to console: - Connect helium gas tube to the console via a long extender - Open helium tank. - The central lumen of the catheter is flushed and connected to pressure tubing with 3 way and then to a pressure transducer to allow for monitoring of central aortic pressure. - Zero the transducer Initial set-up: - Once connected properly the console would show ECG and pressure waveforms. - Check Basal mean pressure - Make sure the setting is at “auto” - Usually IABP started at 1:1 or 1:2 augmentation - Usually Augmentation is kept at maxim
  • 23. 23
  • 24. 24
  • 25. 25 Trigger modes Trigger : - Event the pump uses to identify the onset of cardiac cycle (systole) - Pump must have consistent trigger in order to provide patient assist - If selected trigger not detected, counter pulsation will interrupted 1.ECG - uses the slope of QR segment to detect triggering point 2. AP(Arterial pressure wave) - Systolic upstroke of the arterial pressure wave form is the trigger
  • 26. 26 ECG signal – most common • Inflation - middle of T wave • Deflation – peak of R wave • Pacer (v/a) • Arterial waveform • An intrinsic pump rate (VF, CPB)
  • 27. 27
  • 28. 28
  • 29. 29
  • 31. 31 Not all Sub optimal augmentation is due to Timing errors/kinks
  • 32. 32 Factors affecting diastolic augmentation Patient - Heart rate - Mean arterial pressure - Stroke volume - Systemic vascular resistance Intra aortic balloon catheter - IAB in sheath - IAB not unfolded - IAB position - Kink in the IAB catheter - IAB leak - Low helium concentration Intra aortic balloon pump - Timing - Position of IAB augmentation control
  • 33. 33 How to check waveform is acceptable ?  First change from 1:1 to 1:2 augmentation  Check the dicrotic notch  See if augmentation starts at that point This should produce a sharp “V” at inflation.
  • 34. 34 How to check waveform is acceptable ?  First change from 1:1 to 1:2 augmentation  Check the dicrotic notch  See if augmentation starts at that point This should produce a sharp “V” at inflation.  Check if diastolic augmented wave is › systolic wave
  • 35. 35 How to check waveform is acceptable ?  First change from 1:1 to 1:2 augmentation  Check the dicrotic notch  See if augmentation starts at that point This should produce a sharp “V” at inflation.  Check if diastolic augmented wave is › systolic wave  Confirm if end diastolic wave following the augmented wave is less than an non augmented wave.  Is Deflation slope ok
  • 36. 36
  • 37. 37 Late Inflation  Inflation of the IAB markedly after closure of the aortic valve.  Waveform Characteristics: • Inflation of IAB after the dicrotic notch. • Absence of sharp V. • Sub optimal diastolic augmentation
  • 38. 38 Early Deflation  Premature deflation of the IAB during the diastolic phase.
  • 39. 39 Late Deflation  Late deflation of the IAB during the diastolic phase.  Waveform Characteristics: • Assisted aortic end diastolic pressure may be equal to the unassisted aortic end diastolic pressure. • Rate of rise of assisted systole is prolonged. • Diastolic augmentation may appear widened
  • 40. 40
  • 41. 41 Variation in balloon pressure wave forms Increased duration of plateau due to longer diastolic phase Decreased duration of plateau due to shortened diastolic phase
  • 42. 42 Variation in balloon pressure wave forms Varying R-R intervals result in irregular plateau durations
  • 43. 43 Variation in balloon pressure wave forms Increased height or amplitude of the waveform Decreased height or amplitude of the waveform
  • 44. 44 Variation in balloon pressure wave forms Gas leak Leak in the closed system causing the balloon pressure waveform to fall below zero baseline.. - due to a loose connection - a leak in the IAB catheter - H2O condensation in the external tubing - a patient who is tachycardiac and febrile which causes increased gas diffusion through the IAB membrane
  • 45. 45 Catheter Kink Rounded balloon pressure waveform - Loss of plateau resulting from a kink or obstruction of shuttle gas - Kink in the catheter tubing - Improper IAB catheter position - Sheath not being pulled back to allow inflation of the IAB - IAB is too large for the aorta - IAB is not fully unwrapped - H2O condensation in the external tubing
  • 47. 47 Patient Management During IABP support  Anticoagulation-- maintain apTT at 50 to 70 seconds  CXR daily – to R/O IAB migration  Check lower limb pulses - 2 hourly. - If not palpable » ? - vascular obstruction - thrombus, embolus, or dissection (urgent surgical consultation)  Prophylactic antibiotics NOT INDICATED  Hip flexion is restricted, and the head of the bed should not be elevated beyond 30°.
  • 48. 48  Never leave in standby by mode for more than 20 minutes > thrombus formation  Daily – Haemoglobin (risk of bleeding or haemolysis) – Platelet count (risk of thrombocytopenia) – Renal function (risk of acute kidney injury secondary to distal migration of IABP catheter)  Wean off the IABP as early as possible as longer duration is associated with higher incidence of limb complications Patient Management During IABP support
  • 49. 49
  • 50. 50 Weaning of IABP Timing of weaning: - Patient should be stable for 12 – 24 hours - Decrease inotropic support - Decrease pump ratio – From 1:1 to 1:2 or 1:3 - Decrease augmentation - Monitor patient closely – If patient becomes unstable, weaning should be immediately discontinued
  • 51. 51 IABP Removal -Discontinue heparin 1 hour prior to removal -Disconnect the IAB catheter from the IAB pump - Patient blood pressure will collapse the balloon membrane for withdrawal - Withdraw the IAB catheter through the introducer sheath until resistance is met. - NEVER attempt to withdraw the balloon membrane through the introducer sheath. -Remove the IAB catheter and the introducer sheath as a unit - Check for adequacy of limb perfusion after hemostasis is achieved.
  • 52. ACC/AHA 2013Practice Guidelines (ESC 2012- IIb) 52
  • 54. IABP IN UA/NSTEMI (2007, 2012 update) (Current Practices)(Class IIa LOE-©) 54 The placement of an IABP could be useful in patients with recurrent ischemia despite maximal medical management and in those with hemodynamic instability until coronary angiography and revascularization can be completed.
  • 56. Benchmark registry • n = 17,000( june 1996- aug 2000) • 203 hospitals- 90% US • 18.8% of IABP used for cardiogenic shock • Device related death – 0.5% • Major complication – 2.6% • Minor complication – 4.2% 56
  • 57. Balloon-pump assisted Coronary Intervention Study (BCIS-1): • The first randomized controlled trial of elective Intra-Aortic Balloon Pump (IABP) insertion prior to high-risk PCI vs. PCI with no planned IABP use • 17 UK centres • n=301 (150 in each arm) 57
  • 58. 58IABP in high risk PCI(BCIS-1 TRIAL) N= 301 Elective IABP( 151) No elective IABP(150) P VALUE MACE 15.2% 16% 0.85 All cause mortality at 6 mths 4.6% 7.4% 0.32 Major procedural complications 1.3% 10.7% <0.001 Major or minor bleeding 19.2% 11.3% 0.06 Access site complications 3.3% 0% 0.06 Patients (n = 301) had severe left ventricular dysfunction (ejection fraction ≤ 30%) and extensive coronary disease (Jeopardy Score ≥ 8/12); those with contraindications to or class I indications for IABP therapy were excluded JAMA. 2010;304(8):867-874
  • 59. 59 Conclusions of long term results of BCIS1 trial(2012-2013) In patients with severe ischemic cardiomyopathy treated with PCI, all cause-mortality was 33% at 51 months (median) Elective IABP use during PCI was associated with an observed 34% reduction in long-term all-cause mortality
  • 60. 60 Counterpulsation Reduces Infarct Size Acute Myocardial Infarction (CRISP AMI) trial. Intra-aortic balloon pump counterpulsation prior to PCI in patients with ST segment elevation MI without shock does not reduce infarct size as measured by MRI
  • 61. Shock trial and registry 61
  • 62. 62 N-302 pts N- 152 pts N- 150 pts
  • 63. 63 SHOCK Trial Primary and Secondary Endpoints 0 20 40 60 80 30 Days 6 months Immediate Revascularization Strategy Medical Stabilization as an Initial Strategy Primary Endpoint Secondary Endpoint Mortality(%) 46.7 % 56.0 % 50.3 % 63.1 % P=.11 P= .027 Hochman et al, NEJM 1999; 341:625.
  • 64. 64 Impact of thrombolysis, intra-aortic balloon pump counterpulsation, and their combination in cardiogenic shock complicating acute myocardial infarction: a report from the SHOCK Trial Registry
  • 65. 65 SHOCK Registry: Impact of Thrombolytics and IABP 0 20 40 60 80 47% 52% % P<0.0001 63% 77% Thrombolytics + IABP No Thrombolytics + IABP Thrombolytics + No IABP Neither Hochman et al, NEJM 1999; 341:625
  • 66. 66 Intraaortic Balloon Support for Myocardial Infarction with Cardiogenic Shock IABP Shock II Trial
  • 67. Conclusion • The use of intraaortic balloon counterpulsation did not significantly reduce 30-day mortality in patients with cardiogenic shock complicating acute myocardial infarction for whom an early revascularization strategy was planned. 67
  • 69. • 1. Major physiological effects of counter pulsation include? ▫ A) increased coronary artery perfusion, increased preload, decreased after load, decreased myocardial oxygen consumption ▫ B) increased coronary artery perfusion, increased preload, increased after load, decreased myocardial oxygen consumption ▫ C) increased coronary artery perfusion, decreased preload, decreased after load, increased myocardial oxygen consumption ▫ D) increased coronary artery perfusion, decreased preload, decreased after load, decreased myocardial oxygen consumption 69
  • 70. 2. The dicrotic notch on the arterial wave form reflects A) aortic valve opening B) aortic valve closure C) isovolumetric contraction D)rapid ejection 70
  • 71. 3. Expected changes with IABP support in hemodynamic profile in patients with Cardiogenic shock include all except? A) Decrease in SBP by 20 % B) Increase in aortic DP by 30 % C) Decrease in MAP by 10% D) Reduction of the HR by 20% E)Decrease in the mean PCWP by 20 % 71
  • 72. 4. late inflation of the balloon can result in? A) premature augmentation B) increased augmentation C) decreased augmentation D) increased coronary perfusion 72
  • 73. 5. A rounded balloon pressure wave form indicate? A) helium leak B) power failure C) hypovolemia D) balloon occluding the aorta 73
  • 74. 6. width of balloon pressure wave form corresponds to A) length of systole B) length of diastole C) arterial pressure D) helium level 74
  • 75. 7. true statement a) Dicrotic notch- land mark used to set deflation b) Deflation is timed to occur during period of iso volumetric contraction c) Most common trigger used is arterial pressure wave method d) Internal trigger mode is acceptable to use in a patient with normal sinus rhythm 75
  • 76. 8. true statement A) pacing spikes are automatically rejected in ECG triggered modes B) pacing trigger modes can be used in a patient of 50% paced rhythm C) Varying R-R interval result in regular plateau durations in Balloon pressureWave form 76
  • 77. 9. Identify the tracing abnormality 77
  • 78. 10. Identify the tracing abnormality 78
  • 79. 79
  • 80. • 1. Major physiological effects of counter pulsation include? ▫ A) increased coronary artery perfusion, increased preload, decreased after load, decreased myocardial oxygen consumption ▫ B) increased coronary artery perfusion, increased preload, increased after load, decreased myocardial oxygen consumption ▫ C) increased coronary artery perfusion, decreased preload, decreased after load, increased myocardial oxygen consumption ▫ D) increased coronary artery perfusion, decreased preload, decreased after load, decreased myocardial oxygen consumption 80
  • 81. 2. the dicrotic notch on the arterial wave form reflects A) aortic valve opening B) aortic valve closure C) isovolumetric contraction D)rapid ejection 81
  • 82. 3. Expected changes with IABP support in hemodynamic profile in patients with Cardiogenic shock include all except? •A) Decrease in SBP by 20 % •B) Increase in aortic DP by 30 % •C) Decrease in MAP by 10% •D) Reduction of the HR by 20% •E)Decrease in the mean PCWP by 20 % 82
  • 83. 4. late inflation of the balloon can result in? • A) premature augmentation • B) increased augmentation • C) decreased augmentation • D) increased coronary perfusion 83
  • 84. 5. A rounded balloon pressure wave form indicate? • A) helium leak • B) power failure • C) hypovolemia • D) balloon occluding the aorta 84
  • 85. 6. width of balloon pressure wave form corresponds to • A) length of systole • B) length of diastole • C) arterial pressure • D) helium level 85
  • 86. 7. true statement a) Dicrotic notch- land mark used to set deflation b) Deflation is timed to occur during period of iso volumetric contraction c) Most common trigger used is arterial pressure wave method d) Internal trigger mode is acceptable to use in a patient with normal sinus rhythm 86
  • 87. 8. true statement A) pacing spikes are automatically rejected in ECG triggered modes B) pacing trigger modes can be used in a patient of 50% paced rhyth C) Varying R-R interval result in regular plateau durations in Balloon press. Wave form 87
  • 88. 9. 88