*An ultrasound artifact is a structure in an image which
does not directly similiar with actual tissue being
scanned.
*Artifact assumes different forms including :
• Structures in the image that are not actually present
• Objects that should be represented but are missing from
the image.
• Structures which are misregistered on the image.
1. REVERBERATION
2. ACOUSTIC SHADOWING
3. ACOUSTIC ENHANCEMENT
4. EDGE SHADOWING
5. BEAM WIDTH ARTIFACT
6. SLICE THICKNESS ARTIFACT
7. SIDE LOBE ARTIFACT
8. MIROR IMAGE
9. DOUBLE IMAGE
10.EQUIPMENT-GENERATED ARTIFACT
11.REFRACTION ARTIFACT
• This is the production of false echoes due to repeated
reflections between two interfaces with a high acoustic
impedance mismatch.
• The echo from the interface is received by the transducer
and displayed on the image.
• Some of the energy in the returned echo is reflected at the
transducer face, and return to the reflecting interface as if
it was a weak transmitted pulse, returning as a second
echo.
• As the time taken for the second echo to arrive is twice
that taken by the depth.
*
• This sequence of reflection and transmission can occur
many times, with the third echo taking three times as long
to return to the transducer and being displayed at three
times the depth, and so on.
• The reverberation echoes will be equally spaced because
the time for each additional echo is multiple of the time of
return of the first echo.
• These reverberation echoes will be strong because of the
high acoustic mismatch.
This artifact will be seen at the skin-transducer interface
and behind bowel gas.
Rectification:
• Increase the amount of gel used.
• Use a stand-off pad.
• Reduce the gain.
• Move the position of the transducer.
• This appears as an area of low amplitude echoes behind an
area of strongly attenuating tissue.
• It is caused by severe attenuation of the beam at an
interface, resulting in very little sound being transmitted
beyond.
• The attenuation can be due to either absorption or
reflection of the sound waves, or a combination of the
two.
• Acoustic shadowing will occur at interfaces with large
acoustic mismatch such as:
Soft tissue and gas
Soft tissue and bone or calculus
*
• This artifact appears as a localized area of increased echo
amplitude behind an area of low attenuation.
• On a scan it will appears as an area of increased brightness,
and can commonly be seen distal to fluid-filled structures
such as the urinary bladder, GB or a cyst.
• The artifact arises due to the application of the time-gain
compensation(TGC) to areas of low attenuating structures
such as fluid.
• It is caused by the low level of attenuation of the beam as it
passes through fluid relative to the greater attenuation of
the beam in the adjacent more solid tissue.
*
• This artifact can often be an useful diagnostic aid,
particularly when scanning a soft-tissue mass or cyst
containing low level echoes.
• These echoes may often cause the structure to disappear
in the image as it blend into the surrounding echo pattern.
• A combination of refraction and reflection occurring at the
edges of rounded structures will result in edge shadowing
artifact.
• It arises due to refraction of the beam caused by both the
curvature of the rounded edges and difference in speed of
two materials.
• When the ultrasound beam reaches the rounded edge of a
structure, reflection will occur, with an angle of incidence
equal to the angle of reflection.
• The outer part of the beam will be totally reflected, but
the reminder of the beam passes through the rounded
structure and is refracted.
*
• This combination of reflection and refraction of the
beam at the edges of a rounded structure results in a
thin strip of tissue behind the edge not being insonated
and causes a shadow.
• This artifact can be demonstrated by scanning a point
reflector in a phantom, where the display will clearly
portray this as a line.
• During routine scanning, the artifact can be seen when
spurious echoes are displayed in an echo-free area.
• Correct positioning of the focal zone will help to reduce
this artifact.
• The focal zone is controlled by electronically narrowing
the beam
*
• These occurs due to the thickness of the beam, and are
similar to beam width artifacts.
• These artifacts will typically be seen in transverse views of
the urinary bladder when structures adjacent to the slice
through the bladder being scanned will be incorporated
into the image.
• These echoes are then displayed as if they were arising
from within the bladder.
• Although the appearance of this artifact is similar to the
beam width artifact, the differentiating factors is that the
reflector causing the slice thickness artifact will not be
seen on the display.
*
• This artifact is a result of inherent characteristics of the
transducer, and apart from trying a different transducer,
cannot be eliminated.
• The energy within the ultrasound beam exists as several
side lobes radiating at a number of angles from a central
lobe.
• Echoes are generated by these side lobes in addition to
the main lobe, but all the returning echoes are assumed
by the transducer to have arisen from the central axis of
the main lobe.
• Side lobe echoes will therefore be misregistered in the
display.
• This artifact can often be seen in area such as the urinary
bladder and may also arise within a cyst.
*
• These artifacts results in a mirror image of a structure
occurring in an ultrasound display.
• They arise due to specular reflection of the beam at a
large smooth interface.
• An area close to a specular reflector will be imaged twice,
once by the original ultrasound beam and once by the
beam after it has reflected off the specular reflector.
• Mirror image artifacts are most commonly seen where
there is a large acoustic mismatch, such as a fluid-air
interface.
*
• Typically this artifact can occur during the scanning of a
full bladder, when air in the rectum behind the bladder
act as specular reflector and mirror image of the
bladder is displayed posteriorly.
• It will then have the appearance of a large cyst behind
the bladder.
• It can also be seen when scanning the liver, and the
diaphragm act as a specular reflector.
• This image is caused by refraction of the beam and may
occur in areas such as the rectus abdominis muscle on the
anterior abdominal wall.
• In the transverse plane the edges of the muscle act as a
lens and the ultrasound beam to be refracted and this
causes the single structure to be interrogated by two
separate refracted beams.
• Two sets of echoes will therefore be returned and these
will cause display of two structures in the image.
*
This results in, for example, two images of the transverse
aorta side by side in the abdomen.
• Incorrect use of the equipment controls can lead to artifact
appearing.
• Misuse of controls such as the gain or TGC can result in
echoes being recorded as too bright or too dark.
• Care must be taken when setting these controls, to ensure
an even brightness throughout the image.
• If the dynamic range control is incorrectly set, this can lead
to an image which has too much contrast, and result in the
loss of subtle echo information.
*
• Gain must be in medium level.
• Blurring of a moving image can occur if the frame rate is
too low or if the persistence is too high.
• It is important to ensure that the frame rate is capable of
recording a moving structure at the speed.
• Use of multiple focal zones can give rise to a prominent
banding effect within the image.
*
• The refraction is the change of the sound direction on
passing from one medium to another.
• In ultrasound, refraction is due to sound velocity
mismatches combines with oblique angles of incidence,
most commonly with convex scanheads.
• When the ultrasound wave crosses at an oblique angle the
interface of two materials, through which the waves
propagate at different velocities, refraction occurs, caused
by bending of the wave beam.
• Refraction artifact cause spatial distortion and loss of
resolution in the image.
Ultrasound artifacts

Mais conteúdo relacionado

Mais procurados

ULTRASOUND PHYSICS
ULTRASOUND PHYSICSULTRASOUND PHYSICS
ULTRASOUND PHYSICSNavni Garg
 
Contrast media & reaction
Contrast media & reactionContrast media & reaction
Contrast media & reactionDr. Mohit Goel
 
Types of transducers in ultrasound diagnostic
Types of transducers in ultrasound diagnosticTypes of transducers in ultrasound diagnostic
Types of transducers in ultrasound diagnosticMohamed M. Elsaied
 
CT numbers, window width and window level
CT numbers, window width and window levelCT numbers, window width and window level
CT numbers, window width and window levelGanesan Yogananthem
 
Ultrasound Artifact
Ultrasound ArtifactUltrasound Artifact
Ultrasound ArtifactAmer Sohail
 
Enteroclysis( small bowel enema)
Enteroclysis( small bowel enema)Enteroclysis( small bowel enema)
Enteroclysis( small bowel enema)Ankit Mishra
 
Types of Ultrasound Probes
Types of Ultrasound ProbesTypes of Ultrasound Probes
Types of Ultrasound Probeshafsatalat1
 
Sinogram and fistulogram
Sinogram and fistulogramSinogram and fistulogram
Sinogram and fistulogramInosRagan
 
Interactions of ultrasound with matter
Interactions of ultrasound with matterInteractions of ultrasound with matter
Interactions of ultrasound with matterSwapnil Shetty
 
Fluoroscopy presentation
Fluoroscopy presentationFluoroscopy presentation
Fluoroscopy presentationHuzaifa Oxford
 
ultrasound physics
ultrasound physicsultrasound physics
ultrasound physicsRakesh Ca
 
Color Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physicsColor Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physicsSyed Yousaf Gilani
 

Mais procurados (20)

ULTRASOUND PHYSICS
ULTRASOUND PHYSICSULTRASOUND PHYSICS
ULTRASOUND PHYSICS
 
Contrast media & reaction
Contrast media & reactionContrast media & reaction
Contrast media & reaction
 
Types of transducers in ultrasound diagnostic
Types of transducers in ultrasound diagnosticTypes of transducers in ultrasound diagnostic
Types of transducers in ultrasound diagnostic
 
CT numbers, window width and window level
CT numbers, window width and window levelCT numbers, window width and window level
CT numbers, window width and window level
 
PRINCIPLES OF ULTRASONOGRAPHY
PRINCIPLES OF ULTRASONOGRAPHYPRINCIPLES OF ULTRASONOGRAPHY
PRINCIPLES OF ULTRASONOGRAPHY
 
Ultrasound Artifact
Ultrasound ArtifactUltrasound Artifact
Ultrasound Artifact
 
Enteroclysis( small bowel enema)
Enteroclysis( small bowel enema)Enteroclysis( small bowel enema)
Enteroclysis( small bowel enema)
 
Ct Generations
Ct  GenerationsCt  Generations
Ct Generations
 
CR & DR
CR & DRCR & DR
CR & DR
 
Types of Ultrasound Probes
Types of Ultrasound ProbesTypes of Ultrasound Probes
Types of Ultrasound Probes
 
Sinogram and fistulogram
Sinogram and fistulogramSinogram and fistulogram
Sinogram and fistulogram
 
Loopogram
LoopogramLoopogram
Loopogram
 
Interactions of ultrasound with matter
Interactions of ultrasound with matterInteractions of ultrasound with matter
Interactions of ultrasound with matter
 
Ultrasound artifacts
Ultrasound artifactsUltrasound artifacts
Ultrasound artifacts
 
Fluoroscopy presentation
Fluoroscopy presentationFluoroscopy presentation
Fluoroscopy presentation
 
BASIC MRI SEQUENCES
BASIC MRI SEQUENCESBASIC MRI SEQUENCES
BASIC MRI SEQUENCES
 
Ultrasound contrast agents
Ultrasound contrast agentsUltrasound contrast agents
Ultrasound contrast agents
 
ultrasound physics
ultrasound physicsultrasound physics
ultrasound physics
 
Color Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physicsColor Doppler Ultrasound - Basic physics
Color Doppler Ultrasound - Basic physics
 
Usg Basics
Usg BasicsUsg Basics
Usg Basics
 

Semelhante a Ultrasound artifacts

Artifacts usg, ct, mri radiology ppt,pdf pk
Artifacts usg, ct, mri radiology  ppt,pdf pkArtifacts usg, ct, mri radiology  ppt,pdf pk
Artifacts usg, ct, mri radiology ppt,pdf pkDr pradeep Kumar
 
Ultrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundUltrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundArjun Reddy
 
ARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxAnayaAnaya14
 
Ultrasound of eye - B scan
Ultrasound of eye - B scan Ultrasound of eye - B scan
Ultrasound of eye - B scan Shruti Laddha
 
Artifacts in echocardiography
Artifacts  in echocardiographyArtifacts  in echocardiography
Artifacts in echocardiographyDilzith Katepogu
 
Physics of ultrasound imaging
Physics of ultrasound imagingPhysics of ultrasound imaging
Physics of ultrasound imagingmtuzer
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxShubham661884
 
ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT  ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT PAWAN JARWAL
 
Probe construction, types and usg artefacts
Probe construction, types and usg artefactsProbe construction, types and usg artefacts
Probe construction, types and usg artefactsVishakha Mittal
 
Artifacts and Bioeffects
Artifacts and Bioeffects Artifacts and Bioeffects
Artifacts and Bioeffects Nusirat Adedewe
 
Basics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wallBasics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wallGamal Agmy
 
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...Gamal Agmy
 
Resolution and different transducers
Resolution and different transducersResolution and different transducers
Resolution and different transducersRaghu Kishore Galla
 
ultrasoundartifacts-200113150706.pdf
ultrasoundartifacts-200113150706.pdfultrasoundartifacts-200113150706.pdf
ultrasoundartifacts-200113150706.pdfGenesisGeronaga
 

Semelhante a Ultrasound artifacts (20)

Artifacts usg, ct, mri radiology ppt,pdf pk
Artifacts usg, ct, mri radiology  ppt,pdf pkArtifacts usg, ct, mri radiology  ppt,pdf pk
Artifacts usg, ct, mri radiology ppt,pdf pk
 
Ultrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasoundUltrasound artifacts and contrast enhanced ultrasound
Ultrasound artifacts and contrast enhanced ultrasound
 
ARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptxARTIFACTS IN ECHO-1.pptx
ARTIFACTS IN ECHO-1.pptx
 
Us artefacts
Us artefactsUs artefacts
Us artefacts
 
Ultrasound of eye - B scan
Ultrasound of eye - B scan Ultrasound of eye - B scan
Ultrasound of eye - B scan
 
Us physics 3
Us physics 3Us physics 3
Us physics 3
 
Ct artiact
Ct artiactCt artiact
Ct artiact
 
Usg
UsgUsg
Usg
 
MRIARTIFACTS.pptx
MRIARTIFACTS.pptxMRIARTIFACTS.pptx
MRIARTIFACTS.pptx
 
Artifacts in echocardiography
Artifacts  in echocardiographyArtifacts  in echocardiography
Artifacts in echocardiography
 
Physics of ultrasound imaging
Physics of ultrasound imagingPhysics of ultrasound imaging
Physics of ultrasound imaging
 
artifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptxartifactsusgctmriradiologypk-190830131406.pptx
artifactsusgctmriradiologypk-190830131406.pptx
 
ULTRA.pptx
ULTRA.pptxULTRA.pptx
ULTRA.pptx
 
ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT  ULTRASONOGRAPHY IN EYE AND ORBIT
ULTRASONOGRAPHY IN EYE AND ORBIT
 
Probe construction, types and usg artefacts
Probe construction, types and usg artefactsProbe construction, types and usg artefacts
Probe construction, types and usg artefacts
 
Artifacts and Bioeffects
Artifacts and Bioeffects Artifacts and Bioeffects
Artifacts and Bioeffects
 
Basics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wallBasics of sonography and anatomy of chest wall
Basics of sonography and anatomy of chest wall
 
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...
Sudanese Chest Sonography Workshop (Basics of sonography and anatomy of chest...
 
Resolution and different transducers
Resolution and different transducersResolution and different transducers
Resolution and different transducers
 
ultrasoundartifacts-200113150706.pdf
ultrasoundartifacts-200113150706.pdfultrasoundartifacts-200113150706.pdf
ultrasoundartifacts-200113150706.pdf
 

Último

ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxAreebaZafar22
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxheathfieldcps1
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...Poonam Aher Patil
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104misteraugie
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactPECB
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Celine George
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphThiyagu K
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingTechSoup
 
Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...
Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...
Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...Shubhangi Sonawane
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.christianmathematics
 
Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxVishalSingh1417
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17Celine George
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxVishalSingh1417
 
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17  How to Extend Models Using Mixin ClassesMixin Classes in Odoo 17  How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17 How to Extend Models Using Mixin ClassesCeline George
 
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptxMaritesTamaniVerdade
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdfQucHHunhnh
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introductionMaksud Ahmed
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfPoh-Sun Goh
 
Making and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdfMaking and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdfChris Hunter
 
Role Of Transgenic Animal In Target Validation-1.pptx
Role Of Transgenic Animal In Target Validation-1.pptxRole Of Transgenic Animal In Target Validation-1.pptx
Role Of Transgenic Animal In Target Validation-1.pptxNikitaBankoti2
 

Último (20)

ICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptxICT Role in 21st Century Education & its Challenges.pptx
ICT Role in 21st Century Education & its Challenges.pptx
 
The basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptxThe basics of sentences session 2pptx copy.pptx
The basics of sentences session 2pptx copy.pptx
 
General Principles of Intellectual Property: Concepts of Intellectual Proper...
General Principles of Intellectual Property: Concepts of Intellectual  Proper...General Principles of Intellectual Property: Concepts of Intellectual  Proper...
General Principles of Intellectual Property: Concepts of Intellectual Proper...
 
Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104Nutritional Needs Presentation - HLTH 104
Nutritional Needs Presentation - HLTH 104
 
Beyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global ImpactBeyond the EU: DORA and NIS 2 Directive's Global Impact
Beyond the EU: DORA and NIS 2 Directive's Global Impact
 
Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17Advanced Views - Calendar View in Odoo 17
Advanced Views - Calendar View in Odoo 17
 
Z Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot GraphZ Score,T Score, Percential Rank and Box Plot Graph
Z Score,T Score, Percential Rank and Box Plot Graph
 
Grant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy ConsultingGrant Readiness 101 TechSoup and Remy Consulting
Grant Readiness 101 TechSoup and Remy Consulting
 
Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...
Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...
Ecological Succession. ( ECOSYSTEM, B. Pharmacy, 1st Year, Sem-II, Environmen...
 
This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.This PowerPoint helps students to consider the concept of infinity.
This PowerPoint helps students to consider the concept of infinity.
 
Unit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptxUnit-IV; Professional Sales Representative (PSR).pptx
Unit-IV; Professional Sales Representative (PSR).pptx
 
How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17How to Give a Domain for a Field in Odoo 17
How to Give a Domain for a Field in Odoo 17
 
Unit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptxUnit-V; Pricing (Pharma Marketing Management).pptx
Unit-V; Pricing (Pharma Marketing Management).pptx
 
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17  How to Extend Models Using Mixin ClassesMixin Classes in Odoo 17  How to Extend Models Using Mixin Classes
Mixin Classes in Odoo 17 How to Extend Models Using Mixin Classes
 
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
2024-NATIONAL-LEARNING-CAMP-AND-OTHER.pptx
 
1029 - Danh muc Sach Giao Khoa 10 . pdf
1029 -  Danh muc Sach Giao Khoa 10 . pdf1029 -  Danh muc Sach Giao Khoa 10 . pdf
1029 - Danh muc Sach Giao Khoa 10 . pdf
 
microwave assisted reaction. General introduction
microwave assisted reaction. General introductionmicrowave assisted reaction. General introduction
microwave assisted reaction. General introduction
 
Micro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdfMicro-Scholarship, What it is, How can it help me.pdf
Micro-Scholarship, What it is, How can it help me.pdf
 
Making and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdfMaking and Justifying Mathematical Decisions.pdf
Making and Justifying Mathematical Decisions.pdf
 
Role Of Transgenic Animal In Target Validation-1.pptx
Role Of Transgenic Animal In Target Validation-1.pptxRole Of Transgenic Animal In Target Validation-1.pptx
Role Of Transgenic Animal In Target Validation-1.pptx
 

Ultrasound artifacts

  • 2. * *An ultrasound artifact is a structure in an image which does not directly similiar with actual tissue being scanned. *Artifact assumes different forms including : • Structures in the image that are not actually present • Objects that should be represented but are missing from the image. • Structures which are misregistered on the image.
  • 3. 1. REVERBERATION 2. ACOUSTIC SHADOWING 3. ACOUSTIC ENHANCEMENT 4. EDGE SHADOWING 5. BEAM WIDTH ARTIFACT 6. SLICE THICKNESS ARTIFACT 7. SIDE LOBE ARTIFACT 8. MIROR IMAGE 9. DOUBLE IMAGE 10.EQUIPMENT-GENERATED ARTIFACT 11.REFRACTION ARTIFACT
  • 4. • This is the production of false echoes due to repeated reflections between two interfaces with a high acoustic impedance mismatch. • The echo from the interface is received by the transducer and displayed on the image. • Some of the energy in the returned echo is reflected at the transducer face, and return to the reflecting interface as if it was a weak transmitted pulse, returning as a second echo. • As the time taken for the second echo to arrive is twice that taken by the depth. *
  • 5. • This sequence of reflection and transmission can occur many times, with the third echo taking three times as long to return to the transducer and being displayed at three times the depth, and so on. • The reverberation echoes will be equally spaced because the time for each additional echo is multiple of the time of return of the first echo. • These reverberation echoes will be strong because of the high acoustic mismatch.
  • 6. This artifact will be seen at the skin-transducer interface and behind bowel gas. Rectification: • Increase the amount of gel used. • Use a stand-off pad. • Reduce the gain. • Move the position of the transducer.
  • 7.
  • 8. • This appears as an area of low amplitude echoes behind an area of strongly attenuating tissue. • It is caused by severe attenuation of the beam at an interface, resulting in very little sound being transmitted beyond. • The attenuation can be due to either absorption or reflection of the sound waves, or a combination of the two. • Acoustic shadowing will occur at interfaces with large acoustic mismatch such as: Soft tissue and gas Soft tissue and bone or calculus *
  • 9.
  • 10.
  • 11. • This artifact appears as a localized area of increased echo amplitude behind an area of low attenuation. • On a scan it will appears as an area of increased brightness, and can commonly be seen distal to fluid-filled structures such as the urinary bladder, GB or a cyst. • The artifact arises due to the application of the time-gain compensation(TGC) to areas of low attenuating structures such as fluid. • It is caused by the low level of attenuation of the beam as it passes through fluid relative to the greater attenuation of the beam in the adjacent more solid tissue. *
  • 12. • This artifact can often be an useful diagnostic aid, particularly when scanning a soft-tissue mass or cyst containing low level echoes. • These echoes may often cause the structure to disappear in the image as it blend into the surrounding echo pattern.
  • 13. • A combination of refraction and reflection occurring at the edges of rounded structures will result in edge shadowing artifact. • It arises due to refraction of the beam caused by both the curvature of the rounded edges and difference in speed of two materials. • When the ultrasound beam reaches the rounded edge of a structure, reflection will occur, with an angle of incidence equal to the angle of reflection. • The outer part of the beam will be totally reflected, but the reminder of the beam passes through the rounded structure and is refracted. *
  • 14. • This combination of reflection and refraction of the beam at the edges of a rounded structure results in a thin strip of tissue behind the edge not being insonated and causes a shadow.
  • 15. • This artifact can be demonstrated by scanning a point reflector in a phantom, where the display will clearly portray this as a line. • During routine scanning, the artifact can be seen when spurious echoes are displayed in an echo-free area. • Correct positioning of the focal zone will help to reduce this artifact. • The focal zone is controlled by electronically narrowing the beam *
  • 16.
  • 17. • These occurs due to the thickness of the beam, and are similar to beam width artifacts. • These artifacts will typically be seen in transverse views of the urinary bladder when structures adjacent to the slice through the bladder being scanned will be incorporated into the image. • These echoes are then displayed as if they were arising from within the bladder. • Although the appearance of this artifact is similar to the beam width artifact, the differentiating factors is that the reflector causing the slice thickness artifact will not be seen on the display. *
  • 18. • This artifact is a result of inherent characteristics of the transducer, and apart from trying a different transducer, cannot be eliminated.
  • 19. • The energy within the ultrasound beam exists as several side lobes radiating at a number of angles from a central lobe. • Echoes are generated by these side lobes in addition to the main lobe, but all the returning echoes are assumed by the transducer to have arisen from the central axis of the main lobe. • Side lobe echoes will therefore be misregistered in the display. • This artifact can often be seen in area such as the urinary bladder and may also arise within a cyst. *
  • 20.
  • 21. • These artifacts results in a mirror image of a structure occurring in an ultrasound display. • They arise due to specular reflection of the beam at a large smooth interface. • An area close to a specular reflector will be imaged twice, once by the original ultrasound beam and once by the beam after it has reflected off the specular reflector. • Mirror image artifacts are most commonly seen where there is a large acoustic mismatch, such as a fluid-air interface. *
  • 22. • Typically this artifact can occur during the scanning of a full bladder, when air in the rectum behind the bladder act as specular reflector and mirror image of the bladder is displayed posteriorly. • It will then have the appearance of a large cyst behind the bladder. • It can also be seen when scanning the liver, and the diaphragm act as a specular reflector.
  • 23.
  • 24. • This image is caused by refraction of the beam and may occur in areas such as the rectus abdominis muscle on the anterior abdominal wall. • In the transverse plane the edges of the muscle act as a lens and the ultrasound beam to be refracted and this causes the single structure to be interrogated by two separate refracted beams. • Two sets of echoes will therefore be returned and these will cause display of two structures in the image. *
  • 25. This results in, for example, two images of the transverse aorta side by side in the abdomen.
  • 26. • Incorrect use of the equipment controls can lead to artifact appearing. • Misuse of controls such as the gain or TGC can result in echoes being recorded as too bright or too dark. • Care must be taken when setting these controls, to ensure an even brightness throughout the image. • If the dynamic range control is incorrectly set, this can lead to an image which has too much contrast, and result in the loss of subtle echo information. *
  • 27. • Gain must be in medium level. • Blurring of a moving image can occur if the frame rate is too low or if the persistence is too high. • It is important to ensure that the frame rate is capable of recording a moving structure at the speed. • Use of multiple focal zones can give rise to a prominent banding effect within the image.
  • 28.
  • 29. * • The refraction is the change of the sound direction on passing from one medium to another. • In ultrasound, refraction is due to sound velocity mismatches combines with oblique angles of incidence, most commonly with convex scanheads. • When the ultrasound wave crosses at an oblique angle the interface of two materials, through which the waves propagate at different velocities, refraction occurs, caused by bending of the wave beam. • Refraction artifact cause spatial distortion and loss of resolution in the image.