Protocol




    Imaging the Deep Brain of
  Freely Moving Transgenic Mice




                                  Ver 1.0
Imaging the Deep Brain of Freely Moving Transgenic Mice


          Materials and Methods

          Mouse Model                    Adult fluorescent mice (B6 background)
          Fluorophore/Marker             YFP
          Promoter                       Thy-1

          Note: There are a number of commercially available transgenic animals expressing
          fluorescent proteins under the direction of neural- or glial-specific promoters. For further
          information, please contact applications support at VisualSonics (apps@visualsonics.com),
          or visit The Jackson Laboratory website:
          http://jaxmice.jax.org/research/fluorescent_proteins_or_lacZ.html


          Equipment

               Cellvizio® LAB 488
               ImageCellTM software
               QuantiKitTM 488 calibration kit
               NeuroPak


          Laboratory material

          Anesthesia and appropriate delivery            Betadine
          method (i.e. vaporizer or syringe)             Topical analgesic
                                                         Topical antibiotic
          Hair clippers                                  Scalpel
          Gauze                                          Surgical scissors
          PBS or 0.9% NaCl                               Forceps
          Anesthetic (gas or solution)                   Stereotaxic brain atlas
          Dissecting microscope                          Stereotactic frame (for mouse) with electrode and
                                                         implant holders
          Ice                                            Ear bars
          70% ethanol solution                           Needle
                                                         Drilling tool (with small drill bits: 0.25mm – 0.5mm)
                                                         Dental cement/acrylic
                                                         Lead pencil
                                                         Bone wax




Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice                                                Page 1
Protocol

          Step 1: Implantation of guide-cannula (Day 1)

          NOTE: See also Appendix I for visual instructions

              1. Anesthetize the mouse according to the method of the lab. Pinch the hind foot - the
                 mouse is ready when it is not responsive.

              2. Clean all surgical instruments with 70% ethanol solution to disinfect.

              3. Shave the head to expose skin.

              4. Fix the mouse head into the stereotaxic frame with ear bars and bite plate. Ensure
                 that the animal is breathing normally.

              5. Make a longitudinal incision in the middle of the head with scalpel or scissors to
                 expose Lambda, Bregma, and the approximate X, Y position of the target structure.
                 Clean the skull with PBS or 0.9% NaCl. Wipe with betadine and 70% alcohol.
                 Depending on your facility, add topical analgesic and/or topical antibiotic.

              6. Under a dissecting microscope, use forceps to apply pressure and ensure that the
                 skull remains immobile. Identify anatomical landmarks Lambda (caudal aspect) and
                 Bregma (frontal aspect). Mark them with the lead pencil. Refer to Appendix I for
                 visual instructions.

              7. Use the drilling tool to make 3 holes for the anchorage microscrews contralateral to
                 the image target. The diameter of the holes should be adapted to the size of the
                 screws. To ensure stability of the guide-cannula, one microscrew should be fixed in
                 the frontal bone as far away as possible from Bregma, one in the parietal bone, and
                 one on the occipital bone. Allow the top of the microscrews to remain above the
                 skull: they will better serve as anchorage points if dental cement can be put all
                 around them. Refer to Appendix I for visual instructions.

          Tip:    Gently blow bone dust away rather than wiping it away with a damp piece of gauze.

          Positioning of the skull

              1. Fix a needle onto the electrode holder of the stereotactic frame. Move the tip of the
                 needle using the micromanipulators of the stereotaxic frame to ensure that Bregma
                 and Lambda are aligned along the X, Y axis. If they are not aligned, reposition the
                 skull using the ear bar micromanipulators until these points are perfectly aligned.
                 Similarly, note the Z positions of Lambda and Bregma and reposition the skull using
                 the ear bar and bite plate micromanipulators until the Z coordinates are the same.

              2. Move 1mm away from Lambda on each side. Note the Z for these two points. Adjust
                 skull position to get these points at the same level. This procedure aims to ensure
                 that the skull is perfectly placed, a need to perform a good probe placement.




Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice                                        Page 2
3. Reposition the needle tip at Bregma and note its position by placing the tip of the
                 needle exactly on this point. Note all coordinates for this point (X, Y, Z). This is the
                 reference point (0, 0, 0).

              4. Use the brain atlas to determine the appropriate coordinates of the structure that
                 you want to reach, according to Bregma. Use the coordinates to calculate the target
                 position (X, Y). Using the micromanipulators of the stereotaxic frame, move the tip
                 of the needle to reach this point. Mark the skull with the pencil on this precise
                 location.

              5. Use the drilling tool to make a hole in the marked location. A hole at least 1,250µm
                 in diameter is required to allow insertion of the guide-cannula.

          Tip:    If a blood vessel is damaged and blood begins to leak, use dry gauze to absorb the
                  blood and rinse with ice-cold PBS or 0.9% NaCl.

              6. Use a standard 25 G needle or pointed forceps to punch through and remove the
                 meninges. This will facilitate insertion of the dedicated implant.

              7. Fix a guide-cannula implant onto the dedicated implant holder provided in the
                 NeuroPak kit. Exchange the stereotaxic electrode holder with the dedicated implant
                 holder and rotate such that the implant is 45 degrees from the midline as described
                 on the following diagrams.




                                                               Schematic representation of skull and
                                                               implant for insertion of guide cannula within
                                                               the left hemisphere




                                                               Schematic representation of skull and
                                                               implant for insertion of guide cannula within
                                                               the right hemisphere




Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice                                              Page 3
Lateral view of the implant turned
                                                                      45 degrees away from the midline
                                                                      (implantation on the left side).
                                                                      Note    that   the   bevel   allows
                                                                      visualization of the guide-cannula
                                                                      as it enters the predrilled hole
                                                                      during implantation.

                                                                      The guide-cannula of the implant is
                                                                      represented in blue, while the red
                                                                      dot represents the anchor column
                                                                      to which the CerboFlex™ will be
                                                                      secured.




              8. Using the micromanipulators on the stereotaxic frame, position the center of the
                 guide-cannula above Bregma. Note the X, Y and Z0 positions. This is the new
                 reference coordinate (0,0,0). Using the micromanipulators on the stereotaxic frame,
                 place the guide-cannula in the X and Y positions according to the stereotaxic atlas,
                 above the image target (should be centered in the hole previously done).

              9. Place the guide-cannula directly onto the brain. Note the Z1 position.

              10. Insert the guide-cannula into the brain. For subcortical areas, insert the tube 300
                  micrometers below the brain surface. For deeper brain areas, insert the tube deeper
                  below the brain surface (600 micrometers) Note the final position Z2. Knowing Z0
                  for Bregma and final Z2 position allows precise calculation of the depth of
                  implantation within the brain from Bregma (= distance of the exit part of the cannula
                  from Bregma).

              11. This point will be used later to reach the target with precision.

              12. Apply a small amount of bone wax on the underside of the implant around the base
                  of the guide-cannula to avoid bleeding and prevent dental cement from entering the
                  guide-cannula.

          Note: This step is not required but it should also help to prevent brain infection by closing
                the hole around the cannula.

          Prepare the dental cement
          (as described in the preparation guide, follow the safety instructions).

              1. Begin by adding small amounts of dental cement. Begin by the screws then, as the
                 cement begins to polymerize, finish by the immediate periphery of the cannula and
                 include the implant. Be very careful to avoid blockage of the guide-cannula with
                 cement (It should not be of concern if the end of the guide-cannula is inserted
                 correctly into the brain). Allow the cement dry for 10 minutes.

          NOTE: Avoid contact between the skin/fur and the dental cement.



Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice                                           Page 4
2. Once the cement has hardened, disconnect the guide-cannula implant from the
                 implant holder.

              3. Insert the anti-contamination plug into the guide-cannula to prevent infection.
                 Suture the skin around the base of the implanted guide-cannula.

              4. Remove the animal from the stereotaxic frame and allow the animal to recover from
                 the surgery for at least one week prior to imaging. Follow guidelines from your
                 institution for post operative animal care (use of analgesics, antibiotics, etc.).


          Step 2: In vivo imaging
          (Day 7)

              1. Turn on the Cellvizio LAB LSU488 imaging system and the connected computer.
                 Insert and lock the proximal end of the CerboFlex™ into the connector on the LSU.
                 For detailed instructions regarding the Cellvizio LAB hardware, consult the hardware
                 user manual.

              2. Launch the ImageCell program. For detailed instructions regarding the ImageCell
                 software, consult the software user manual. Ensure that the CerboFlex microprobe is
                 detected by turning on the laser. If the probe is not detected, remove it from the
                 LSU, insert the CerboFlex installation CD (included in the NeuroPak™ kit) into the
                 computer and follow the onscreen instructions. Launch ImageCell anew and start the
                 laser to ensure that the microprobe is detected. Turn on the laser and leave it on for
                 at least 20 minutes immediately prior to the imaging session.

              3. Prior to mounting the CerboFlex into the electrode holder of the stereotaxic frame,
                 calibrate the system using the QuantiKit 488 calibration kit. Set up the desired
                 storage folder and prefix for the images to be acquired. Rinse the tip in 70% EtOH
                 and then rinse with distilled water.

              4. Fix the CerboFlex into the electrode holder of the stereotaxic frame. Remove the
                 anti-contamination plug. Carefully align the tip of the CerboFlex with the center of
                 the guide-cannula implant. Ensure that the orientation of both the fiberoptic
                 microprobe and the guide-cannula are such that both parts will fit together perfectly.
                 Be very careful to avoid any contact between the tip and the walls of the cannula
                 (risk of damage). If the parts are not aligned, refer to the ‘positioning of the skull’
                 section above. Ensure that the Z-lock screw on the CerboFlex is sufficiently loose to
                 allow the CerboFlex to fit around the stabilization column of the guide-cannula
                 implant.

              5. Using dissecting microscope and the micromanipulators on the stereotaxic frame,
                 center the tip of the CerboFlex above the entrance to the guide-cannula. Be very
                 careful to avoid any contact between the tip and the walls of the cannula (risk of
                 damage). Lower the CerboFlex until the tip of the microprobe reaches the entrance
                 of the tube. Note the Z3 position.




Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice                                          Page 5
6. Calculate the target position distance from this point (Z3) according to cannula
                 length and the depth of the cannula base in the brain. Knowing that the guide-
                 cannula is 7.5 mm long and that the end of the tube is inserted 300 or 600
                 micrometers into the brain, calculate how far away from the entrance of the cannula
                 the CerboFlex™ should be lowered to reach the targeted cells (Calculation: 7.5 mm
                 + distance of target from base of the cannula).

              7. Slowly and incrementally insert the CerboFlex through the cannula and into the brain
                 tissue. If there is any resistance while inserting the CerboFlex into the cannula, Z-
                 lock screw on the CerboFlex is sufficiently loose to allow the CerboFlex to fit around
                 the stabilization column of the guide-cannula implant and ensure that there is
                 nothing blocking the cannula using a needle under stereotaxic guidance to clean the
                 port. Turn on the laser to guide the CerboFlex to the image target under image
                 guidance.

          Tip:    The image target may move slightly once reached. Once cells are identified, turn the
                  laser off and wait 5 minutes to let the tissue to adjust to the probe. If movement is
                  detected in the field of view, very fine adjustment of the CerboFlex in the Z direction
                  toward brain surface may be helpful.

              8. Carefully secure the CerboFlex onto the stabilization column by tightening the Z-lock
                 screw.

          Tip:    The field of view may have moved slightly when tightening the Z-lock screw. Turn
                  the laser off and wait 5 minutes. If the field of view changed, adjust the position of
                  the CerboFlex. Repeat this process until the field of view remains stable.

              9. Remove the animal from the stereotactic frame and allow it to recover from the
                 anesthetic. Put it in a resting place to recover from anaesthesia or directly within
                 your experimetal set up. Ensure that the CerboFlex does not become twisted by
                 monitoring the animal and manually manipulating the animal if necessary.

              10. From this point, images can be captured continuously or intermittently using time
                  lapse acquisition according to your needs.


          Step 4: Removal of the CerboFlex

              1. At the end of the imaging session, turn off the laser beam.

              2. Anaesthetize and position the animal into the stereotactic frame to remove the
                 CerboFlex. Use caution and slowly remove the CerboFlex using the
                 micromanipulators of the stereotaxic frame to prevent damage to the probe tip. It is
                 equally important to use caution when removing the probe as when inserting the
                 probe.

              3. Clean the CerboFlex immediately thoroughly using the QuantiKit according to
                 instructions outlined in the user guide.




Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice                                           Page 6

Mais conteúdo relacionado

Mais procurados (20)

Binocular vision
Binocular visionBinocular vision
Binocular vision
 
Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.Autorefractometry: principle and procedure.
Autorefractometry: principle and procedure.
 
Synoptophore 2
Synoptophore 2Synoptophore 2
Synoptophore 2
 
Retinoscopy
RetinoscopyRetinoscopy
Retinoscopy
 
Suppression third yr arya
Suppression third yr aryaSuppression third yr arya
Suppression third yr arya
 
Binocular vision
Binocular visionBinocular vision
Binocular vision
 
Binocular vision and vision perception
Binocular vision and vision perceptionBinocular vision and vision perception
Binocular vision and vision perception
 
Fusion
FusionFusion
Fusion
 
Binocular vision
Binocular visionBinocular vision
Binocular vision
 
Introduction to accommodative and binocular anomalies
Introduction to accommodative and binocular anomaliesIntroduction to accommodative and binocular anomalies
Introduction to accommodative and binocular anomalies
 
Binocular vision
Binocular visionBinocular vision
Binocular vision
 
Slit lamp ..
Slit lamp ..Slit lamp ..
Slit lamp ..
 
Sensory evaluation of squint
Sensory  evaluation of squintSensory  evaluation of squint
Sensory evaluation of squint
 
Slit lamp in Ophthalmology
Slit lamp in OphthalmologySlit lamp in Ophthalmology
Slit lamp in Ophthalmology
 
Electrophysiology of the Eye
Electrophysiology of the EyeElectrophysiology of the Eye
Electrophysiology of the Eye
 
Slit lamp
Slit lamp Slit lamp
Slit lamp
 
Stereoscopic Vision
Stereoscopic VisionStereoscopic Vision
Stereoscopic Vision
 
Streak retinoscope
Streak retinoscopeStreak retinoscope
Streak retinoscope
 
Slit lamp biomicroscopy
Slit lamp biomicroscopySlit lamp biomicroscopy
Slit lamp biomicroscopy
 
Autorefractometer
AutorefractometerAutorefractometer
Autorefractometer
 

Semelhante a Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice

8. Cephalometric Radiography.pptx
8. Cephalometric Radiography.pptx8. Cephalometric Radiography.pptx
8. Cephalometric Radiography.pptxManuelKituzi
 
Lt0520 d peek bio-pushlock knotless anchor for bankart repair
Lt0520 d   peek bio-pushlock knotless anchor for bankart repairLt0520 d   peek bio-pushlock knotless anchor for bankart repair
Lt0520 d peek bio-pushlock knotless anchor for bankart repairdrnaula
 
Darine mottin 2007_6442-57_spie bios
Darine mottin  2007_6442-57_spie biosDarine mottin  2007_6442-57_spie bios
Darine mottin 2007_6442-57_spie biosStéphane MOTTIN
 
Seminar extra oral radiograph
Seminar extra oral radiographSeminar extra oral radiograph
Seminar extra oral radiographNegi Singh
 
Cephalometrics in orthodontics /certified fixed orthodontic courses by India...
Cephalometrics in orthodontics  /certified fixed orthodontic courses by India...Cephalometrics in orthodontics  /certified fixed orthodontic courses by India...
Cephalometrics in orthodontics /certified fixed orthodontic courses by India...Indian dental academy
 
Keratometry and Dynamic Retinoscopy
Keratometry and Dynamic RetinoscopyKeratometry and Dynamic Retinoscopy
Keratometry and Dynamic RetinoscopyKamal Luitel
 
OPG and Extraoral radiography
OPG and Extraoral radiographyOPG and Extraoral radiography
OPG and Extraoral radiographyNeha Sharma
 
biometry-160212135415 (1).pdf
biometry-160212135415 (1).pdfbiometry-160212135415 (1).pdf
biometry-160212135415 (1).pdfssuser0f453c
 
Biometry & Iol calculations
Biometry & Iol calculationsBiometry & Iol calculations
Biometry & Iol calculationsrakesh jaiswal
 
Employing Electrophysiology and Optogenetics to Measure and Manipulate Neuron...
Employing Electrophysiology and Optogenetics to Measure and Manipulate Neuron...Employing Electrophysiology and Optogenetics to Measure and Manipulate Neuron...
Employing Electrophysiology and Optogenetics to Measure and Manipulate Neuron...InsideScientific
 
Acs0206 Parotidectomy
Acs0206 ParotidectomyAcs0206 Parotidectomy
Acs0206 Parotidectomymedbookonline
 
Extraoral radiograph lecture
Extraoral radiograph lectureExtraoral radiograph lecture
Extraoral radiograph lectureLama K Banna
 
The microscope and its historical development
The microscope and its historical developmentThe microscope and its historical development
The microscope and its historical developmentArmand Anthony
 

Semelhante a Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice (20)

8. Cephalometric Radiography.pptx
8. Cephalometric Radiography.pptx8. Cephalometric Radiography.pptx
8. Cephalometric Radiography.pptx
 
Basic of cephalometric rupesh
Basic of cephalometric rupeshBasic of cephalometric rupesh
Basic of cephalometric rupesh
 
Biometry.pptx
Biometry.pptxBiometry.pptx
Biometry.pptx
 
Cephalometrics
CephalometricsCephalometrics
Cephalometrics
 
Lt0520 d peek bio-pushlock knotless anchor for bankart repair
Lt0520 d   peek bio-pushlock knotless anchor for bankart repairLt0520 d   peek bio-pushlock knotless anchor for bankart repair
Lt0520 d peek bio-pushlock knotless anchor for bankart repair
 
Darine mottin 2007_6442-57_spie bios
Darine mottin  2007_6442-57_spie biosDarine mottin  2007_6442-57_spie bios
Darine mottin 2007_6442-57_spie bios
 
Myelogram pdf
Myelogram pdfMyelogram pdf
Myelogram pdf
 
Seminar extra oral radiograph
Seminar extra oral radiographSeminar extra oral radiograph
Seminar extra oral radiograph
 
Cephalometrics in orthodontics /certified fixed orthodontic courses by India...
Cephalometrics in orthodontics  /certified fixed orthodontic courses by India...Cephalometrics in orthodontics  /certified fixed orthodontic courses by India...
Cephalometrics in orthodontics /certified fixed orthodontic courses by India...
 
Anaesthetic techniques
Anaesthetic techniquesAnaesthetic techniques
Anaesthetic techniques
 
EXTRA ORAL RADIOGRAPHY
EXTRA ORAL RADIOGRAPHYEXTRA ORAL RADIOGRAPHY
EXTRA ORAL RADIOGRAPHY
 
Keratometry and Dynamic Retinoscopy
Keratometry and Dynamic RetinoscopyKeratometry and Dynamic Retinoscopy
Keratometry and Dynamic Retinoscopy
 
OPG and Extraoral radiography
OPG and Extraoral radiographyOPG and Extraoral radiography
OPG and Extraoral radiography
 
biometry-160212135415 (1).pdf
biometry-160212135415 (1).pdfbiometry-160212135415 (1).pdf
biometry-160212135415 (1).pdf
 
Biometry & Iol calculations
Biometry & Iol calculationsBiometry & Iol calculations
Biometry & Iol calculations
 
Employing Electrophysiology and Optogenetics to Measure and Manipulate Neuron...
Employing Electrophysiology and Optogenetics to Measure and Manipulate Neuron...Employing Electrophysiology and Optogenetics to Measure and Manipulate Neuron...
Employing Electrophysiology and Optogenetics to Measure and Manipulate Neuron...
 
Acs0206 Parotidectomy
Acs0206 ParotidectomyAcs0206 Parotidectomy
Acs0206 Parotidectomy
 
Extraoral radiograph lecture
Extraoral radiograph lectureExtraoral radiograph lecture
Extraoral radiograph lecture
 
The microscope and its historical development
The microscope and its historical developmentThe microscope and its historical development
The microscope and its historical development
 
Basic cephalometrics
Basic cephalometrics Basic cephalometrics
Basic cephalometrics
 

Mais de FUJIFILM VisualSonics Inc.

White Paper: In vivo Fiberoptic Fluorescence Microscopy in freely behaving mice
White Paper: In vivo Fiberoptic Fluorescence Microscopy in freely behaving miceWhite Paper: In vivo Fiberoptic Fluorescence Microscopy in freely behaving mice
White Paper: In vivo Fiberoptic Fluorescence Microscopy in freely behaving miceFUJIFILM VisualSonics Inc.
 
In Vivo Detection of Nanoparticles in Mouse Cancer Using an Integrated Photoa...
In Vivo Detection of Nanoparticles in Mouse Cancer Using an Integrated Photoa...In Vivo Detection of Nanoparticles in Mouse Cancer Using an Integrated Photoa...
In Vivo Detection of Nanoparticles in Mouse Cancer Using an Integrated Photoa...FUJIFILM VisualSonics Inc.
 
Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...
Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...
Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...FUJIFILM VisualSonics Inc.
 
Cellvizio® LAB Fibered Fluorescence Microscope and Microdialysis Research
Cellvizio®  LAB Fibered Fluorescence Microscope and  Microdialysis Research Cellvizio®  LAB Fibered Fluorescence Microscope and  Microdialysis Research
Cellvizio® LAB Fibered Fluorescence Microscope and Microdialysis Research FUJIFILM VisualSonics Inc.
 
Application Note - Neuroscience: Peripheral Nerve Imaging
Application Note - Neuroscience: Peripheral Nerve ImagingApplication Note - Neuroscience: Peripheral Nerve Imaging
Application Note - Neuroscience: Peripheral Nerve ImagingFUJIFILM VisualSonics Inc.
 
Imaging Vascular Oxygen Saturation in an Acute Hindlimb Ischemia
Imaging Vascular Oxygen Saturation in an Acute Hindlimb IschemiaImaging Vascular Oxygen Saturation in an Acute Hindlimb Ischemia
Imaging Vascular Oxygen Saturation in an Acute Hindlimb IschemiaFUJIFILM VisualSonics Inc.
 
Application Brief: Tumor Microenvironment Imaging with Photoacoustic Technology
Application Brief: Tumor Microenvironment Imaging with Photoacoustic TechnologyApplication Brief: Tumor Microenvironment Imaging with Photoacoustic Technology
Application Brief: Tumor Microenvironment Imaging with Photoacoustic TechnologyFUJIFILM VisualSonics Inc.
 

Mais de FUJIFILM VisualSonics Inc. (20)

Vevo LAZR-X Brochure mkt02906 v1.0
Vevo LAZR-X Brochure mkt02906 v1.0Vevo LAZR-X Brochure mkt02906 v1.0
Vevo LAZR-X Brochure mkt02906 v1.0
 
Mkt51280 vevo3100 brochure 2.0
Mkt51280 vevo3100 brochure 2.0Mkt51280 vevo3100 brochure 2.0
Mkt51280 vevo3100 brochure 2.0
 
Bibliography: Cardiovascular Research
Bibliography: Cardiovascular ResearchBibliography: Cardiovascular Research
Bibliography: Cardiovascular Research
 
Bibliography: Cancer Ver 1.2
Bibliography: Cancer Ver 1.2Bibliography: Cancer Ver 1.2
Bibliography: Cancer Ver 1.2
 
Bibliography - Nephrology
Bibliography - NephrologyBibliography - Nephrology
Bibliography - Nephrology
 
Application Brief - Cancer Angiogenisis
Application Brief - Cancer AngiogenisisApplication Brief - Cancer Angiogenisis
Application Brief - Cancer Angiogenisis
 
Application Brief - Breast Cancer Research
Application Brief - Breast Cancer ResearchApplication Brief - Breast Cancer Research
Application Brief - Breast Cancer Research
 
White Paper: In vivo Fiberoptic Fluorescence Microscopy in freely behaving mice
White Paper: In vivo Fiberoptic Fluorescence Microscopy in freely behaving miceWhite Paper: In vivo Fiberoptic Fluorescence Microscopy in freely behaving mice
White Paper: In vivo Fiberoptic Fluorescence Microscopy in freely behaving mice
 
Vevo 2100 Brochure
Vevo 2100 BrochureVevo 2100 Brochure
Vevo 2100 Brochure
 
Application Note: Angiogenesis
Application Note: AngiogenesisApplication Note: Angiogenesis
Application Note: Angiogenesis
 
Vevo2100vs770
Vevo2100vs770Vevo2100vs770
Vevo2100vs770
 
Comparison: Vevo 2100 vs Vevo 770
Comparison: Vevo 2100 vs Vevo 770Comparison: Vevo 2100 vs Vevo 770
Comparison: Vevo 2100 vs Vevo 770
 
In Vivo Detection of Nanoparticles in Mouse Cancer Using an Integrated Photoa...
In Vivo Detection of Nanoparticles in Mouse Cancer Using an Integrated Photoa...In Vivo Detection of Nanoparticles in Mouse Cancer Using an Integrated Photoa...
In Vivo Detection of Nanoparticles in Mouse Cancer Using an Integrated Photoa...
 
Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...
Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...
Development and Validation of a Combined Photoacoustic Micro-Ultrasound Syste...
 
Cellvizio® LAB Fibered Fluorescence Microscope and Microdialysis Research
Cellvizio®  LAB Fibered Fluorescence Microscope and  Microdialysis Research Cellvizio®  LAB Fibered Fluorescence Microscope and  Microdialysis Research
Cellvizio® LAB Fibered Fluorescence Microscope and Microdialysis Research
 
Application Note - Neuroscience: Peripheral Nerve Imaging
Application Note - Neuroscience: Peripheral Nerve ImagingApplication Note - Neuroscience: Peripheral Nerve Imaging
Application Note - Neuroscience: Peripheral Nerve Imaging
 
Application Brief: Abdominal Aortic Aneurysm
Application Brief: Abdominal Aortic AneurysmApplication Brief: Abdominal Aortic Aneurysm
Application Brief: Abdominal Aortic Aneurysm
 
Application Brief: Nephrology
Application Brief: NephrologyApplication Brief: Nephrology
Application Brief: Nephrology
 
Imaging Vascular Oxygen Saturation in an Acute Hindlimb Ischemia
Imaging Vascular Oxygen Saturation in an Acute Hindlimb IschemiaImaging Vascular Oxygen Saturation in an Acute Hindlimb Ischemia
Imaging Vascular Oxygen Saturation in an Acute Hindlimb Ischemia
 
Application Brief: Tumor Microenvironment Imaging with Photoacoustic Technology
Application Brief: Tumor Microenvironment Imaging with Photoacoustic TechnologyApplication Brief: Tumor Microenvironment Imaging with Photoacoustic Technology
Application Brief: Tumor Microenvironment Imaging with Photoacoustic Technology
 

Último

VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escortsaditipandeya
 
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Haridwar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...narwatsonia7
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...narwatsonia7
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...chandars293
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...perfect solution
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...hotbabesbook
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeCall Girls Delhi
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiAlinaDevecerski
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...indiancallgirl4rent
 
Call Girls Siliguri Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 8250077686 Top Class Call Girl Service AvailableDipal Arora
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escortsvidya singh
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...astropune
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...narwatsonia7
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Call Girls in Nagpur High Profile
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableDipal Arora
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...parulsinha
 

Último (20)

VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore EscortsVIP Call Girls Indore Kirti 💚😋  9256729539 🚀 Indore Escorts
VIP Call Girls Indore Kirti 💚😋 9256729539 🚀 Indore Escorts
 
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Haridwar Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Haridwar Just Call 8250077686 Top Class Call Girl Service Available
 
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...Bangalore Call Girls Nelamangala Number 9332606886  Meetin With Bangalore Esc...
Bangalore Call Girls Nelamangala Number 9332606886 Meetin With Bangalore Esc...
 
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Cuttack Just Call 9907093804 Top Class Call Girl Service Available
 
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Ooty Just Call 8250077686 Top Class Call Girl Service Available
 
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟  9332606886 ⟟ Call Me For G...
Top Rated Bangalore Call Girls Ramamurthy Nagar ⟟ 9332606886 ⟟ Call Me For G...
 
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
The Most Attractive Hyderabad Call Girls Kothapet 𖠋 6297143586 𖠋 Will You Mis...
 
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
College Call Girls in Haridwar 9667172968 Short 4000 Night 10000 Best call gi...
 
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
Night 7k to 12k Chennai City Center Call Girls 👉👉 7427069034⭐⭐ 100% Genuine E...
 
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any TimeTop Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
Top Quality Call Girl Service Kalyanpur 6378878445 Available Call Girls Any Time
 
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls DelhiRussian Escorts Girls  Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
Russian Escorts Girls Nehru Place ZINATHI 🔝9711199012 ☪ 24/7 Call Girls Delhi
 
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
(Rocky) Jaipur Call Girl - 09521753030 Escorts Service 50% Off with Cash ON D...
 
Call Girls Siliguri Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 8250077686 Top Class Call Girl Service AvailableCall Girls Siliguri Just Call 8250077686 Top Class Call Girl Service Available
Call Girls Siliguri Just Call 8250077686 Top Class Call Girl Service Available
 
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore EscortsCall Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
Call Girls Horamavu WhatsApp Number 7001035870 Meeting With Bangalore Escorts
 
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Coimbatore Just Call 9907093804 Top Class Call Girl Service Available
 
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
Best Rate (Hyderabad) Call Girls Jahanuma ⟟ 8250192130 ⟟ High Class Call Girl...
 
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...Top Rated Bangalore Call Girls Mg Road ⟟   9332606886 ⟟ Call Me For Genuine S...
Top Rated Bangalore Call Girls Mg Road ⟟ 9332606886 ⟟ Call Me For Genuine S...
 
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
Book Paid Powai Call Girls Mumbai 𖠋 9930245274 𖠋Low Budget Full Independent H...
 
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service AvailableCall Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
Call Girls Dehradun Just Call 9907093804 Top Class Call Girl Service Available
 
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
(Low Rate RASHMI ) Rate Of Call Girls Jaipur ❣ 8445551418 ❣ Elite Models & Ce...
 

Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice

  • 1. Protocol Imaging the Deep Brain of Freely Moving Transgenic Mice Ver 1.0
  • 2. Imaging the Deep Brain of Freely Moving Transgenic Mice Materials and Methods Mouse Model Adult fluorescent mice (B6 background) Fluorophore/Marker YFP Promoter Thy-1 Note: There are a number of commercially available transgenic animals expressing fluorescent proteins under the direction of neural- or glial-specific promoters. For further information, please contact applications support at VisualSonics (apps@visualsonics.com), or visit The Jackson Laboratory website: http://jaxmice.jax.org/research/fluorescent_proteins_or_lacZ.html Equipment  Cellvizio® LAB 488  ImageCellTM software  QuantiKitTM 488 calibration kit  NeuroPak Laboratory material Anesthesia and appropriate delivery Betadine method (i.e. vaporizer or syringe) Topical analgesic Topical antibiotic Hair clippers Scalpel Gauze Surgical scissors PBS or 0.9% NaCl Forceps Anesthetic (gas or solution) Stereotaxic brain atlas Dissecting microscope Stereotactic frame (for mouse) with electrode and implant holders Ice Ear bars 70% ethanol solution Needle Drilling tool (with small drill bits: 0.25mm – 0.5mm) Dental cement/acrylic Lead pencil Bone wax Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice Page 1
  • 3. Protocol Step 1: Implantation of guide-cannula (Day 1) NOTE: See also Appendix I for visual instructions 1. Anesthetize the mouse according to the method of the lab. Pinch the hind foot - the mouse is ready when it is not responsive. 2. Clean all surgical instruments with 70% ethanol solution to disinfect. 3. Shave the head to expose skin. 4. Fix the mouse head into the stereotaxic frame with ear bars and bite plate. Ensure that the animal is breathing normally. 5. Make a longitudinal incision in the middle of the head with scalpel or scissors to expose Lambda, Bregma, and the approximate X, Y position of the target structure. Clean the skull with PBS or 0.9% NaCl. Wipe with betadine and 70% alcohol. Depending on your facility, add topical analgesic and/or topical antibiotic. 6. Under a dissecting microscope, use forceps to apply pressure and ensure that the skull remains immobile. Identify anatomical landmarks Lambda (caudal aspect) and Bregma (frontal aspect). Mark them with the lead pencil. Refer to Appendix I for visual instructions. 7. Use the drilling tool to make 3 holes for the anchorage microscrews contralateral to the image target. The diameter of the holes should be adapted to the size of the screws. To ensure stability of the guide-cannula, one microscrew should be fixed in the frontal bone as far away as possible from Bregma, one in the parietal bone, and one on the occipital bone. Allow the top of the microscrews to remain above the skull: they will better serve as anchorage points if dental cement can be put all around them. Refer to Appendix I for visual instructions. Tip: Gently blow bone dust away rather than wiping it away with a damp piece of gauze. Positioning of the skull 1. Fix a needle onto the electrode holder of the stereotactic frame. Move the tip of the needle using the micromanipulators of the stereotaxic frame to ensure that Bregma and Lambda are aligned along the X, Y axis. If they are not aligned, reposition the skull using the ear bar micromanipulators until these points are perfectly aligned. Similarly, note the Z positions of Lambda and Bregma and reposition the skull using the ear bar and bite plate micromanipulators until the Z coordinates are the same. 2. Move 1mm away from Lambda on each side. Note the Z for these two points. Adjust skull position to get these points at the same level. This procedure aims to ensure that the skull is perfectly placed, a need to perform a good probe placement. Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice Page 2
  • 4. 3. Reposition the needle tip at Bregma and note its position by placing the tip of the needle exactly on this point. Note all coordinates for this point (X, Y, Z). This is the reference point (0, 0, 0). 4. Use the brain atlas to determine the appropriate coordinates of the structure that you want to reach, according to Bregma. Use the coordinates to calculate the target position (X, Y). Using the micromanipulators of the stereotaxic frame, move the tip of the needle to reach this point. Mark the skull with the pencil on this precise location. 5. Use the drilling tool to make a hole in the marked location. A hole at least 1,250µm in diameter is required to allow insertion of the guide-cannula. Tip: If a blood vessel is damaged and blood begins to leak, use dry gauze to absorb the blood and rinse with ice-cold PBS or 0.9% NaCl. 6. Use a standard 25 G needle or pointed forceps to punch through and remove the meninges. This will facilitate insertion of the dedicated implant. 7. Fix a guide-cannula implant onto the dedicated implant holder provided in the NeuroPak kit. Exchange the stereotaxic electrode holder with the dedicated implant holder and rotate such that the implant is 45 degrees from the midline as described on the following diagrams. Schematic representation of skull and implant for insertion of guide cannula within the left hemisphere Schematic representation of skull and implant for insertion of guide cannula within the right hemisphere Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice Page 3
  • 5. Lateral view of the implant turned 45 degrees away from the midline (implantation on the left side). Note that the bevel allows visualization of the guide-cannula as it enters the predrilled hole during implantation. The guide-cannula of the implant is represented in blue, while the red dot represents the anchor column to which the CerboFlex™ will be secured. 8. Using the micromanipulators on the stereotaxic frame, position the center of the guide-cannula above Bregma. Note the X, Y and Z0 positions. This is the new reference coordinate (0,0,0). Using the micromanipulators on the stereotaxic frame, place the guide-cannula in the X and Y positions according to the stereotaxic atlas, above the image target (should be centered in the hole previously done). 9. Place the guide-cannula directly onto the brain. Note the Z1 position. 10. Insert the guide-cannula into the brain. For subcortical areas, insert the tube 300 micrometers below the brain surface. For deeper brain areas, insert the tube deeper below the brain surface (600 micrometers) Note the final position Z2. Knowing Z0 for Bregma and final Z2 position allows precise calculation of the depth of implantation within the brain from Bregma (= distance of the exit part of the cannula from Bregma). 11. This point will be used later to reach the target with precision. 12. Apply a small amount of bone wax on the underside of the implant around the base of the guide-cannula to avoid bleeding and prevent dental cement from entering the guide-cannula. Note: This step is not required but it should also help to prevent brain infection by closing the hole around the cannula. Prepare the dental cement (as described in the preparation guide, follow the safety instructions). 1. Begin by adding small amounts of dental cement. Begin by the screws then, as the cement begins to polymerize, finish by the immediate periphery of the cannula and include the implant. Be very careful to avoid blockage of the guide-cannula with cement (It should not be of concern if the end of the guide-cannula is inserted correctly into the brain). Allow the cement dry for 10 minutes. NOTE: Avoid contact between the skin/fur and the dental cement. Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice Page 4
  • 6. 2. Once the cement has hardened, disconnect the guide-cannula implant from the implant holder. 3. Insert the anti-contamination plug into the guide-cannula to prevent infection. Suture the skin around the base of the implanted guide-cannula. 4. Remove the animal from the stereotaxic frame and allow the animal to recover from the surgery for at least one week prior to imaging. Follow guidelines from your institution for post operative animal care (use of analgesics, antibiotics, etc.). Step 2: In vivo imaging (Day 7) 1. Turn on the Cellvizio LAB LSU488 imaging system and the connected computer. Insert and lock the proximal end of the CerboFlex™ into the connector on the LSU. For detailed instructions regarding the Cellvizio LAB hardware, consult the hardware user manual. 2. Launch the ImageCell program. For detailed instructions regarding the ImageCell software, consult the software user manual. Ensure that the CerboFlex microprobe is detected by turning on the laser. If the probe is not detected, remove it from the LSU, insert the CerboFlex installation CD (included in the NeuroPak™ kit) into the computer and follow the onscreen instructions. Launch ImageCell anew and start the laser to ensure that the microprobe is detected. Turn on the laser and leave it on for at least 20 minutes immediately prior to the imaging session. 3. Prior to mounting the CerboFlex into the electrode holder of the stereotaxic frame, calibrate the system using the QuantiKit 488 calibration kit. Set up the desired storage folder and prefix for the images to be acquired. Rinse the tip in 70% EtOH and then rinse with distilled water. 4. Fix the CerboFlex into the electrode holder of the stereotaxic frame. Remove the anti-contamination plug. Carefully align the tip of the CerboFlex with the center of the guide-cannula implant. Ensure that the orientation of both the fiberoptic microprobe and the guide-cannula are such that both parts will fit together perfectly. Be very careful to avoid any contact between the tip and the walls of the cannula (risk of damage). If the parts are not aligned, refer to the ‘positioning of the skull’ section above. Ensure that the Z-lock screw on the CerboFlex is sufficiently loose to allow the CerboFlex to fit around the stabilization column of the guide-cannula implant. 5. Using dissecting microscope and the micromanipulators on the stereotaxic frame, center the tip of the CerboFlex above the entrance to the guide-cannula. Be very careful to avoid any contact between the tip and the walls of the cannula (risk of damage). Lower the CerboFlex until the tip of the microprobe reaches the entrance of the tube. Note the Z3 position. Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice Page 5
  • 7. 6. Calculate the target position distance from this point (Z3) according to cannula length and the depth of the cannula base in the brain. Knowing that the guide- cannula is 7.5 mm long and that the end of the tube is inserted 300 or 600 micrometers into the brain, calculate how far away from the entrance of the cannula the CerboFlex™ should be lowered to reach the targeted cells (Calculation: 7.5 mm + distance of target from base of the cannula). 7. Slowly and incrementally insert the CerboFlex through the cannula and into the brain tissue. If there is any resistance while inserting the CerboFlex into the cannula, Z- lock screw on the CerboFlex is sufficiently loose to allow the CerboFlex to fit around the stabilization column of the guide-cannula implant and ensure that there is nothing blocking the cannula using a needle under stereotaxic guidance to clean the port. Turn on the laser to guide the CerboFlex to the image target under image guidance. Tip: The image target may move slightly once reached. Once cells are identified, turn the laser off and wait 5 minutes to let the tissue to adjust to the probe. If movement is detected in the field of view, very fine adjustment of the CerboFlex in the Z direction toward brain surface may be helpful. 8. Carefully secure the CerboFlex onto the stabilization column by tightening the Z-lock screw. Tip: The field of view may have moved slightly when tightening the Z-lock screw. Turn the laser off and wait 5 minutes. If the field of view changed, adjust the position of the CerboFlex. Repeat this process until the field of view remains stable. 9. Remove the animal from the stereotactic frame and allow it to recover from the anesthetic. Put it in a resting place to recover from anaesthesia or directly within your experimetal set up. Ensure that the CerboFlex does not become twisted by monitoring the animal and manually manipulating the animal if necessary. 10. From this point, images can be captured continuously or intermittently using time lapse acquisition according to your needs. Step 4: Removal of the CerboFlex 1. At the end of the imaging session, turn off the laser beam. 2. Anaesthetize and position the animal into the stereotactic frame to remove the CerboFlex. Use caution and slowly remove the CerboFlex using the micromanipulators of the stereotaxic frame to prevent damage to the probe tip. It is equally important to use caution when removing the probe as when inserting the probe. 3. Clean the CerboFlex immediately thoroughly using the QuantiKit according to instructions outlined in the user guide. Protocol: Imaging the Deep Brain of Freely Moving Transgenic Mice Page 6