Servo Ventilator 300/300A
Service Manual




                                                                  21




                          Adu lt
                                   ic
                          Ped iatr
                                    e
                          Neo nat




                                   15
                                                          261
                                        5        19
                                                          5.0
                                        11        19       259
                                            2
                                                1 . 13      266

                                                 0 . 35     5.1




                                                                               EN S
                                                                       SI EM




E382 E380E 061 01 03 01
Important                                                                 Servo Ventilator 300/300A


     Important
    General                                        Calibration and functional check
    • Unless stated otherwise, the information     • After any service intervention in Servo
      in this Service Manual is valid for:           Ventilator 300/300A, perform a
      – Servo Ventilator 300 (SV 300)                Calibration and Functional check
                                                     according to instructions in the
      – Servo Ventilator 300A (SV 300A).             Operating Manual.
    • The documentation for the Servo
      Ventilator 300/300A consists of:
                                                   Hazard notices
      – Operating Manual
                                                   • Before disassembling or assembling the
      – Service Manual
                                                     Servo Ventilator 300/300A, make sure
      – Circuit Diagram                              that the:
      – Spare Parts Catalogue
                                                     – Gas supply is disconnected.
      – Reference Manual, Computer Interface
                                                     – Mains power cable is disconnected.
    • The Operating Manual is an indispens-
      able complement to the Service Manual          – Mode selector is set to Ventilator off.
      for proper servicing.                            If the mode selector is set in any other
                                                       position, the internal battery will
    • In addition to the Important information         supply power to the PC boards.
      given here and in the related documents,
      always pay attention to applicable local       – The back-up batteries are disconnec-
      and national regulations.                        ted when the power section is open.

    • Responsibility for the safe functioning of     – All gas conveying parts are cleaned
      the equipment reverts to the owner or            according to instructions in the
      user in all cases in which service or            Operating Manual.
      repair has been done by a non-
      professional or by persons who are not
                                                   Service
      employed by or authorized by Siemens,
      and when the equipment is used for           • When working with ESD sensitive
      other than its intended purpose.               components, always use a grounded
                                                     wrist band and a grounded work surface.
    • Data on internal pressures in the Servo
                                                     Adequate service tools must always be
      Ventilator 300/300A are given in Pa (bar).
                                                     used.
      Airway pressures are given in cm H2O.
                                                   • Some parts in the Servo Ventilator 300/
      1 hPa = 1 mbar         1 mbar = 1 hPa
      1 kPa = 10 mbar        1 mbar = 0.1 kPa        300A are comprised by a spare parts
      1 kPa = 0.01 bar       1 bar = 100 kPa         exchange system. Your local Siemens
      1 kPa i 10 cm H2O      1 cm H2O i 0.1 kPa      representative supplies factory
      1 kPa i 0.01 at        1 at ³ 100 kPa          calibrated and tested exchange parts as
      1 kPa i 0.01 kgf/cm2   1 kgf/cm2 i 100 kPa     well as other spare parts indicated in the
      1 kPa i 0.01 kp/cm2    1 kp/cm2 i 100 kPa      spare parts catalogue.
      1 kPa i 0.145 psi      1 psi i 6.9 kPa




2                                         Siemens-Elema AB                 E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                              Important


   Important
  • The Servo Ventilator 300/300A must be        To the responsible service personnel
    serviced at regular intervals by specially   • The contents of this document are not
    trained personnel. The service intervals,      binding. If any significant difference is
    1000 and 3000 operating hours, are             found between the product and this
    described in the Operating Manual. Any         document, please contact Siemens for
    maintenance must be noted in a log             further information.
    book provided for that purpose in
    accordance with national regulations.        • We reserve the right to modify products
    We recommend that service is done as           without amending this document or
    a part of a service contract with              advising the user.
    Siemens.                                     • Only Siemens authorized personnel shall
  • A 1000 hour overhaul must be performed         be permitted to service or repair the
    after 1000 hours of operation or, at the       Servo Ventilator 300/300A. Only
    lastest, every six months. In addition,        Siemens-Elema exchange parts or
    the ventilator shall undergo a technical       genuine spare parts must be used. PC
    safety check (Function check) twice a          boards (spare or exchange parts) must
    year, at six months intervals or according     always be kept in a package for sensitive
    to national regulations.                       electronic devices.

  • A 3000 hour overhaul must be performed       • Siemens will not otherwise assume
    after every 3000 hours of operation or,        responsibility for the materials used, the
    at the latest, once every year.                work performed or any possible
                                                   consequences of same.
  • The internal batteries shall be replaced
    every 3 years according to instructions in   •         The device complies with the
    this Service Manual. The stated battery                requirements of the Medical
    back-up time, approx. 30 minutes, can                  Device Directive 93/42/EEC.
    be guaranteed only if they are used as
    power supply back-up at mains failure.
  • The battery on PC 1587 COMPUTER
    INTERFACE shall be replaced every 5 years
    according to instructions in this Service
    Manual.
              • Old non-functioning batteries
                and O2 cells must be returned
                to the place of purchase or to
                a place where they can be
                disposed of properly.
      Pb        Batteries and O2 cells must
                not be disposed of with
                ordinary waste.




E382 E380E 061 01 03 01                 Siemens-Elema AB                                        3
Notes                      Servo Ventilator 300/300A



Notes




4       Siemens-Elema AB    E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                                   Contents



Contents

1. Introduction ............................................................................       7    1

2. Description of functions....................................................... 21                   2

3. Disassembling and assembling ......................................... 81                            3

4. Service procedures............................................................... 103                4

5. Troubleshooting ..................................................................... 109            5

6. Product change history ........................................................ 113                  6

7. Index ........................................................................................ 115   7

8. Diagrams ................................................................................. 121       8
E382 E380E 061 01 03 01                  Siemens-Elema AB                                           5
Notes                      Servo Ventilator 300/300A



Notes




6       Siemens-Elema AB    E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                           Introduction



                                     1. Introduction
                                     Main units ................................................ 8
                                     Basic principles ........................................ 12    1




E382 E380E 061 01 03 01     Siemens-Elema AB                                                     7
Introduction                                                        Servo Ventilator 300/300A



                                                 Main units
                            1                    The Servo Ventilator 300 can be divided into
                                                 two main units:
1                       2       3                • Control unit (1). The control unit consists
                                                   of the panel section (2) and the control
                                                   section (3).
                                                 • Patient unit (4). The patient unit consists
                                                   of the pneumatic section (5) and the
                                                   power section (6).
                                                 The patient unit and the control unit are
                                                 connected to each other with the inter-
                                                 connection cable (7).




               7




                    5               6
    300-A00X




                            4




    8                                   Siemens-Elema AB                  E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                      Introduction



                                             Control unit
                                             The control unit consists of the control
                                             section which is a housing with a number of
                                             PC boards. The panel section, attached to
                                             the control section, is the other part of the
                                                                                                1
                                             control unit.
                                             With the panel section removed from the
                                             control section, the following parts are
                                             accessible:
                                             • Front panel controls
                                             • PC 1614 PANEL INTERFACE (1) with PC 1588
                                               MICROPROCESSOR MODULE and PAN-PROM

                                             • PC 1745 AUTOMODE PANEL INTERFACE (2). Only
                                               for SV 300A.
                                             • PC 1617 CONTROL INTERCONNECTION (3).
             300-A01X




                        2   1
  300-A02X




                                3




E382 E380E 061 01 03 01             Siemens-Elema AB                                        9
Introduction                                                               Servo Ventilator 300/300A



                                                        All PC boards in the control section (except
                                                        PC 1617 mentioned above) are accessible
                                                        when the control section cover is removed.
1                                                       These PC boards are:
                                                        • PC 1605 REFERENCE & TIMING (1) with PC
                                                          1588 MICROPROCESSOR MODULE and REF-PROM
                                                        • PC 1616 INSPIRATORY CONTROL (2) with
                       1           2                      PC 1588 MICROPROCESSOR MODULE and MIX-
                                                          PROM

                                                        • PC board (3);
                                                          – PC 1665 COMPUTER INTERFACE DUMMY or
                                                          – Optional PC 1587 COMPUTER INTERFACE (3)
                                                            with COM-PROM
                                                        • PC 1608 MONITORING (4) with PC 1588
                   3           4                          MICROPROCESSOR MODULE and MON-PROM.
    300-A03X




                                                        Patient unit
                           1           2                The upper part of the patient unit is the
                                                        pneumatic section. Under the lid of the
                                                        pneumatic section you will find the:
                                                        • Inspiratory channel (1) including O2 cell
                                                        • Expiratory channel (2) including flow
                                                          transducer.
                                                        The inspiratory valves (gas modules) are
                                                        located in three different slots in the
                                                        pneumatic section:
                                                        • Inspiratory valve unit – AIR (3)
                                                        • Inspiratory valve unit – O2 (4)
                                                        • Inspiratory valve unit – Optional (5).
                                                          The valve dummy in the third slot can be
                                                          replaced by an inspiratory valve for
                                                          administration of e. g. NO gas or by
    300-A04X




                                                          equipment for nebulizing drugs (Servo
               3                   4       5              Ultra Nebulizer 345). These products are
                                                          described in separate documents.




    10                                         Siemens-Elema AB                  E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                                                Introduction



                           1       2           3           4        When the pneumatic section casing and the
                                                                    cover plate is removed, the following parts
                                                                    are accessible:
                                                                    • PC 1615 EXPIRATORY FLOW LINEARIZATION (1)
                                                                      with PC 1588 MICROPROCESSOR MODULE and
                                                                                                                        1
                                                                      LIN-PROM

                                                                    • Housing (2) containing PC 1585 CURRENT
                                                                      CONTROL and PC 1586 CURRENT POWER

                                                                    • Expiratory valve (3)
                                                                    • PC 1622 VALVE CONTROL (4)
                                                                    • PC 1611 EXPIRATORY PRESSURE AMPLIFIER (5)
                                                                    • Safety valve (6) with PC 1613 SAFETY VALVE
                                                                      DRIVER
300-A05X




           5           6               7               8            • PC 1611 INSPIRATORY PRESSURE AMPLIFIER (7)
                                                                    • PC 1607    PNEUMATIC INTERCONNECTION (8).




               1               2               3           4        The lower part of the patient unit is the
                                                                    power section. With the pneumatic section
                                                                    separated from the power section the
                                                                    following parts are accessible:
                                                                    • Mains power inlet (1)
                                                                    • External battery inlet (2)
                                                                    • Operating time meter (3)
                                                                    • Batteries (4)
                                                                    • Transformer (5)
300-A06X




                                                                    • Capacitor (6)
                   5           6           7       8
                                                                    • Auxiliary equipment outputs (7)
                                                                    • PC 1618 POWER SUPPLY (8).




E382 E380E 061 01 03 01                                    Siemens-Elema AB                                        11
Introduction                                        Servo Ventilator 300/300A



    Basic principles


1


                                    9          11


                   8                                                   12




                   10



                                                                        2

                   7




                                                                        1
                   6
    300-F71X




                        5          4                3




    12                      Siemens-Elema AB             E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                              Introduction



Gas flow through the patient unit
Inspiratory flow                                  Expiratory flow
1. Gas inlet for air. The connected air must      8. The patient system's expiratory gas tube
   have a pressure between 200 and 650               is connected at the expiratory inlet. The        1
   kPa (2 and 6.5 bar).                              expiratory inlet also contains a moisture
                                                     trap.
2. Gas inlet for O2. The connected O2 must
   have a pressure between 200 and 650            9. The gas flow through the expiratory
   kPa (2 and 6.5 bar).                              channel is measured by the expiratory
                                                     flow transducer. Patient trigger efforts,
3. The flow of the gas delivered to the
                                                     indicated by a decreased continuous
   patient system is regulated by the
                                                     flow, are sensed by this expiratory flow
   inspiratory valves. There is one inspira-
                                                     transducer.
   tory valve unit (gas module) for each
   gas. The inspiratory valves are regulated      10. The expiratory pressure is measured by
   by a feedback control system that is               the expiratory pressure transducer. The
   described in this chapter.                         transducer is protected by a bacteria
                                                      filter. Patient trigger efforts, indicated by
4. The gases are mixed in the inspiratory
                                                      a pressure drop, are sensed by this
   mixing part.
                                                      expiratory pressure transducer.
5. The pressure of the mixed gas delivered
                                                  11. The pressure of the gas (PEEP pressure)
   to the patient (inspiratory pressure) is
                                                      in the patient system is regulated by the
   measured by the inspiratory pressure
                                                      expiratory valve. The expiratory valve is
   transducer. The transducer is protected
                                                      regulated by a feedback control system
   by a bacteria filter.
                                                      that is described in this chapter.
6. The inspiratory pipe leads the mixed gas
                                                  12. The gas from the patient system leaves
   from the inspiratory mixing part to the
                                                      the ventilator via this expiratory outlet.
   patient system. The inspiratory pipe also
                                                      The outlet contains a non-return valve
   contains the safety valve, a holder for the
                                                      which is a part of the patient triggering
   O2 cell and the inspiratory outlet. The
                                                      system.
   springloaded safety valve will open in
   case of a power failure and/or if the inspi-
   ratory pressure exceeds 120 cm H2O. It
   will also open if the set upper pressure
   limit is exceeded by 5 cm H2O.
7. The oxygen concentration inside the
   inspiratory pipe is measured by the O2
   cell. The O2 cell is protected by a
   bacteria filter.




E382 E380E 061 01 03 01                Siemens-Elema AB                                          13
Introduction                                                                          Servo Ventilator 300/300A



    Feedback control system – General

          TRANSDUCER                                  CONTROLLED SYSTEM                             CONTROLLING

1        SIGNAL AMPLIFIER
                                                              COMPARING
                                                                                                      ELEMENT


                                                                                                    (WITH SERVO)
                                                               ELEMENT


                                                CONTROLLING SYSTEM
                  ACTUAL
                   VALUE




                                                                 P

                                                I                                      CONVERTER
                                                                              ERROR                  ACTUATING
                                                                              SIGNAL                  SIGNAL
                                         "COMPENSATOR"           D
         REFERENCE      DESIRED                                      "ERROR
          VARIABLE       VALUE                                  CALCULATOR"


    300-A08E
                                                           PID-CONTROLLER



    Any CONTROLLING SYSTEM may comprise one                   P-action gives a continuous basic positioning.
    single CONTROLLER or, for instance, a cascade
                                                              I-action gives a slowly varying positioning
    of more than one CONTROLLER. Compensating
                                                              compensation for small long time deviation.
    elements may also be added.
                                                              D-action gives a speed-up positioning
                            CONTROLLER symbol used in
                                                              reaction at (fast) changes.
                            this chapter.
                                                              The signals from these parts (from the parts
                                                              used in the actual combination) are added.
    In each CONTROLLER, the actual value normally
                                                              The sum is the total ERROR SIGNAL which may
    from a TRANSDUCER in the CONTROLLED SYSTEM is
                                                              be converted before use as an ACTUATING
    compared to the desired value = REFERENCE
                                                              SIGNAL.
    VARIABLE. The difference between these two
    input values results at the output of the                 Each CONTROLLING ELEMENT (in the ventilator:
    controlling system in an ACTUATING SIGNAL.                inspiratory valves and expiratory valve) is
                                                              positioned by the ACTUATING SIGNAL from its
    Each CONTROLLER may comprise some
                                                              controlling system. Feedback control, which
    combination of the following parts:
                                                              is synonymous with regulation, results in the
    • Proportional action, P                                  controlled valve being moved into such a
                                                              position that the ACTUAL VALUE is kept equal to
    • Integral action, I
                                                              the DESIRED VALUE independently of
    • Derivative action, D.                                   disturbance variables like changes in the
                                                              airway resistance or lung compliance.




    14                                              Siemens-Elema AB                        E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                                                         Introduction



Inspiratory feedback control system

             REFERENCE           TRANSDUCERS                        CONTROLLED SYSTEM                          INSPIRATORY


                                                                                                                                  1
              & TIMING                                                                                            VALVE

                                PRESSURE    FLOW                                                             POSITION VALVE
                                                                                                             SENSOR SOLENOID
                                   GAS        GAS
                                   SUPPLY     FLOW
                                   PRESS.

                                   ACTUAL     ACTUAL                      INSP. VALVE POSITION
            FLOW                   SUPPLY     GAS FLOW
            REF.                   PRESS.     VALUE                       ACTUAL POSITION VALUE
                                   VALUE                                                           ACTUAL CURRENT VALUE
                         GAS                                                                                   CURRENT
                         FLOW                          FLOW PI                 POSITION PD                    CONTROL &
               MIXER     REF.                         CONTROL                   CONTROL                        CURRENT
                                                                    DESIRED                       DESIRED       POWER
                                      DESIRED                       POSITION                      CURRENT
                                      GAS FLOW                      VALUE                         VALUE
           INSPIRATORY                VALUE
             CONTROL
                                                   VALVE POSITION
                                                    CALCULATOR
300-A09E




             Flow
                                                                                                            FLOW CONTROL
             regulation


The inspiratory feedback control system                             •   ACTUAL GAS FLOW VALUE is generated by
uses cascade control with a number of                                   signal converters (FLOW) using signals from
controllers and compensating elements.                                  measuring TRANSDUCERS.
                                                                    • The actuating signal from the FLOW PI
                                                                      CONTROL results in the DESIRED POSITION VALUE
Flow regulation
                                                                      (reference variable) for the POSITION PD
Inspiratory flow regulation occurs during                             CONTROL. The POSITION SENSOR at the
inspiration time in the modes Volume                                  INSPIRATORY VALVE generates a valve position
Control and SIMV (Vol. Contr.) and during                             signal which is used as the ACTUAL POSITION
expiration time in all modes.                                         VALUE in the POSITION PD CONTROL.

When flow regulation is used, the                                   • The actuating signal from the POSITION PD
shadowed main forward path comprises the                              CONTROL results in the DESIRED CURRENT VALUE
following:                                                            (reference variable) for the CURRENT CONTROL
• The microprocessor module in the block                              & CURRENT POWER.
  MODE CPU in the main block REFERENCE &                            • The ACTUAL CURRENT VALUE is an internal
  TIMING generates a FLOW REF. signal.                                feedback signal used in the CURRENT
• The block MIXER, in the main block                                  CONTROL block.

  INSPIRATORY CONTROL, splits the FLOW REF.                         • The actuating signal from the output stage
  signal into one GAS FLOW REF. signal for each                       of the CURRENT CONTROL & CURRENT POWER is a
  inspiratory valve (AIR FLOW REF. and O2 FLOW                        voltage pulse train with pulse width
  REF.). Each GAS FLOW REF. signal is used                            modulation used to actuate the INSPIRATORY
  (after a minor correction within ±5 %) as                           VALVE SOLENOID which generates a certain
  DESIRED GAS FLOW VALUE (reference variable).                        opening of the INSPIRATORY VALVE.




E382 E380E 061 01 03 01                                  Siemens-Elema AB                                                    15
Introduction                                               Servo Ventilator 300/300A



    In addition to the main forward path,
    compensating elements are added as
    described in both inspiratory feedback
1   control system pictures. The compensating
    elements in this system works as follows:
    • Using the input information of ACTUAL
      SUPPLY PRESSURE VALUE and DESIRED GAS FLOW
      VALUE the VALVE POSITION CALCULATOR directly
      generates a rough approximate POSITION
      VALUE which helps to obtain the desired
      position within minimum time.
    • The input information of ACTUAL SUPPLY
      PRESSURE VALUE is used (inverted and
      adequatly scaled) to create one small part
      of the DESIRED CURRENT VALUE. This small part
      of the DESIRED CURRENT VALUE will – in
      combination with the opening force of the
      INSPIRATORY VALVE implied by the pressure of
      the supplied gas – help to balance the
      closing force on the INSPIRATORY VALVE
      implied by the mechanical spring inside
      the VALVE SOLENOID even in closed position
      of the INSPIRATORY VALVE.




    16                                      Siemens-Elema AB    E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                                                               Introduction



Inspiratory feedback control system (cont'd)
             REFERENCE
              & TIMING
           PRESSURE REF.
           DESIRED PRESS.
           VALUE
                                    TRANSDUCERS                           CONTROLLED SYSTEM
                                                                                                                   INSPIRATORY
                                                                                                                      VALVE             1
            INSP. PRESSURE                                                                                       POSITION VALVE
                                    PRESSURE FLOW                                                                SENSOR SOLENOID
           ACTUAL PRESS. VALUE

                  PRESS.               GAS        GAS
                   PID                 SUPPLY     FLOW
                  CONTR.               PRESS.

                                       ACTUAL    ACTUAL
                      FLOW             SUPPLY    GAS FLOW                    INSP. VALVE POSITION
                      REF.             PRESS.    VALUE                         ACTUAL POSITION VALUE
                                       VALUE                                                            ACTUAL CURRENT VALUE
                             GAS                                                                                   CURRENT
                             FLOW                            FLOW PI                  POSITION PD                 CONTROL &
                   MIXER     REF.                           CONTROL                    CONTROL                     CURRENT
                                                                          DESIRED                      DESIRED      POWER
                                                                          POSITION                     CURRENT
                                                DESIRED                   VALUE                        VALUE
           INSPIRATORY                          GAS FLOW
             CONTROL                            VALUE

                                                         VALVE POSITION
                                                          CALCULATOR
300-A10E




               Pressure                                                                                          FLOW CONTROL
               regulation


Pressure regulation
Inspiratory pressure regulation is used only                              • ACTUAL VALUE is the INSP. PRESSURE signal
during inspiration time in all modes except                                 from the INSPIRATORY PRESSURE TRANSDUCER
when flow regulation is used as mentioned
                                                                          • The actuating signal from the PRESSURE PID
previously.
                                                                            CONTROL block is used as a FLOW REFERENCE
When PRESSURE regulation is used, the                                       signal going to the MIXER.
shadowed main forward path includes the
                                                                          The rest of the controlling system, including
PRESSURE PID CONTROL block in the REFERENCE &
                                                                          compensating elements, is the same as in
TIMING main block:
                                                                          flow regulation.
• DESIRED VALUE is the PRESSURE REFERENCE
  signal from (the microprocessor module
  MODE CPU within) the main block REFERENCE
  & TIMING




E382 E380E 061 01 03 01                                         Siemens-Elema AB                                                   17
Introduction                      Servo Ventilator 300/300A



    Notes

1




    20             Siemens-Elema AB    E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                                                         Introduction



Expiratory feedback control system

             TRANSDUCER


               PRESSURE
                                                    CONTROLLED SYSTEM                                        EXPIRATORY
                                                                                                               VALVE

                                                                                                         POSITION VALVE
                                                                                                                                 1
                                                                                                         SENSOR SOLENOID


                    EXP.
                    PRESSURE



                                                        EXP. VALVE POSITION
                                                        ACTUAL POSITION VALUE
                    ACTUAL                                                                 ACTUAL CURRENT VALUE
                    EXPIRATORY
                    PRESSURE
                    VALUE

                                                                                                     CURRENT
                                                                                                     CONTROL
                                 PEEP PID   POSITION           POSITION PD      EXP. VALVE              &
            PEEP LEVEL           CONTROL    REF.                CONTROL         CURRENT REF.         CURRENT
            DESIRED EXPIRATORY               DESIRED                             DESIRED              POWER
            PRESSURE VALUE                   POSITION                            CURRENT
 300-A11E




                                             VALUE                               VALUE
                                                                                                 PRESSURE CONTROL




The expiratory feedback control system                           • The actuating signal from the POSITION PD
uses cascade control. It is active only during                     CONTROL block is the EXP. VALVE CURRENT REF.
expiration time in all modes. It applies                           signal used as DESIRED VALUE for the CURRENT
PRESSURE regulation.                                               CONTROL & CURRENT POWER block.

• DESIRED VALUE is the PEEP LEVEL signal from                    • The ACTUAL CURRENT VALUE is an internal
  the PANEL SECTION (according to front panel                      feedback signal used in the CURRENT
  setting)                                                         CONTROL block.

• ACTUAL VALUE is the EXP. PRESSURE signal                       • The actuating signal from the output stage
  from the expiratory PRESSURE TRANSDUCER.                         of the CURRENT CONTROL & CURRENT POWER is a
                                                                   pulse train with pulse width modulation
• The actuating signal from the PEEP PID
                                                                   used to actuate the EXPIRATORY VALVE
  CONTROL block is the POSITION REF. signal
                                                                   SOLENOID which generates a certain
  used as DESIRED VALUE for the POSITION PD
                                                                   opening of the EXPIRATORY VALVE.
  CONTROL.

• ACTUAL VALUE is the EXP. VALVE POSITION signal
  from the POSITION SENSOR in the EXPIRATORY
  VALVE.




E382 E380E 061 01 03 01                          Siemens-Elema AB                                                          17
                                                                                                                           19
Introduction                                                                                                                                                                                                                                             Servo Ventilator 300/300A



Principle diagram
                EXTERNAL                                                              EXTERNAL
                 CONTROL
                  SIGNALS
                                                                                      OUTPUT
                                                                                      SIGNALS
                                                                                                     CONTROL UNIT                                                                    PATIENT UNIT
                                                                                                                                           FROM
                                                                                                                                           PATIENT
       INPUT FROM                                                        C
   SV 300 INTERNAL
MICROPROCESSORS                                                                       CONTROL
                                        COMPUTER INTERFACE                            SIGNALS
     INPUT SIGNALS
                                                  11

                                                                         M            C                                                                M            L                             G
     INPUT SIGNALS                                                                                                                                           EXP.
                                                MONITORING                            P                                                                     FLOW                         PRESSURE
                                                                                                                                                                                         AMPLIFIER
                                                                                      T                                                                       EXP. PRESSURE
                                                  10                                                                                                   M
                                                                                                                                                                                                      PRESSURE
                                                                                                                                                                        PEEP LEVEL                    CONTROL

                                  REFERENCE & TIMING        T                INSPIRATORY CONTROL         I                                                                       O2 CONC.
                                                                                                                                                                                                 O2 CONC.
                                                                                                                                                                                                                                     9
                 INSP. PRESSURE
                                                                                                                                                                                 AMPLIFIER                       M
                                        PRESSURE
                                                                                                                                                                                                                     EXPIRATORY
                                        PID CONTROL          FLOW REF.                                                                                                                       VALVE CONTROL V         VALVE               X
                                                                                                             O2 FLOW REF.
                                        PRESSURE REF.
                                                                                                 MIXER       AIR FLOW REF.
                                                                SELECTION OF
                                     Pressure
                                                                PRESSURE OR
                                                                FLOW REGULATION
                                     MODE CPU

                                     Flow
                                                             FLOW REF.
                                                                                                                                                                                                                                                                   O2


                                                                                                                                                                                                                                TRANS-
                                                                                                                                                                                                                                DUCERS
                                                        1                2        4                  5
                                                                                                                                                      O2                                                                                             M
                                                                                                                                                     CELL                                                                 FLOW
                                                                                                                                                                                                                          CONTROL            O2 FLOW REF.

                                                                                                                   P
                                                                                                                                    TO
                                                                                                                                    PATIENT                                                                                                    7
                                                                                                                                                                                                      INSPIRATORY VALVE UNIT - O2                      O



                                                                                                                                                                                                                                                                   AIR

                                                                                                                                                                                                                                TRANS-
                                                                                                                                                                                                                                DUCERS

                                                                                                                                                                                                                                                     M
                                                                                                                                                                                                                          FLOW
                                                                                                                                                                                                                          CONTROL            AIR FLOW REF.
                      3

                                                                                                                                                                                                                                                6
                                                                                                                                                                                         H            INSPIRATORY VALVE UNIT - AIR                     A
                                                                                                                                                                             PRESSURE
                     8                                                                                                                                                       AMPLIFIER
                                                                                                                                                                    INSP. PRESSURE



                                                                                                                                                                                     M


                                                                                                                                                                              +5V
                                                                                                                                                                                                                 W
                                                                                                                                                                                                                      EXT. BATTERY
                                                                                                                                                                             ±15V             POWER SUPPLY                                      INTERNAL
     300-A12E




                                                                                                                                                                                                                                                BATTERY
                                                                                                                                                                              +24V                          12        MAINS POWER




18                                                                                                                      Siemens-Elema AB                                                                                                                     E382 E380E 061 01 03 02
Servo Ventilator 300/300A                                                         Introduction



Inspiratory regulation                          Monitoring
1. Choice between pressure regulation and       10. The MONITORING board monitors and
   flow regulation during inspiration is            generates certain displayed parameters.
   made by the MODE CPU dependent on the
   Mode selection.
                                                    It also activates alarm functions.
                                                                                                 1
2. Main parameters used for the flow level      Computer interface
   during inspiration when flow regulation
   is used.                                     11. The optional COMPUTER INTERFACE board is
                                                    used as input/output signal interface and
3. Main parameters used for the pressure            includes functions for computer
   level during inspiration when pressure           communication (RS-232). The PC board
   regulation is used.                              is equipped with the following
4. Inspiration rise time setting determines         connectors:
   the initial inspiration pattern. A succes-       – Master/Slave connection (N80)
   sively increased flow/pressure level may
   be more comfortable for the patient.             – Analog I/O terminal (N81)

5. The O2 concentration delivered to the            – Serial communication ports (N82/N83)
   patient depends on the setting of these          – Analog input & Digital code port (N84).
   O2 concentration control knobs.
6. INSPIRATORY VALVE UNIT– AIR.
   The control system is described in           Power supply
   section "Inspiratory feedback control        12. The POWER SUPPLY board controls and
   system" in this chapter.                         distributes power to the different parts
                                                    of the ventilator. The power source can
7. INSPIRATORY VALVE UNIT– O2.
                                                    be:
   The control system is described in
   section "Inspiratory feedback control            – Mains power
   system" in this chapter.
                                                    – External battery
                                                    – Internal battery.
Expiratory regulation
8. PEEP level setting is used as DESIRED
   EXPIRATORY PRESSURE VALUE. See section
   "Expiratory feedback control system" in
   this chapter.
9. EXPIRATORY VALVE interacts with the VALVE
   CONTROL board as described in section
   "Expiratory feedback control system" in
   this chapter.




E382 E380E 061 01 03 01                Siemens-Elema AB                                    19
                                                                                           21
Introduction                      Servo Ventilator 300/300A



    Notes

1




    20             Siemens-Elema AB    E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                              Description of functions



                                     2. Description of functions
                                     General .................................................... 22
                                     Microprocessor module PC 1588 ............ 23
                                     Version labels .......................................... 24
                                     Control unit – Panel section ..................... 25
                                     1     Front panel ....................................... 25
                                     1A Automode front panel ...................... 27
                                     2     Panel interface .................................. 28
                                                                                                       2
                                     2A Automode panel interface ................ 29
                                     Control unit – Control section .................. 30
                                     3     Computer interface .......................... 30
                                     3A Computer interface dummy ............. 33
                                     4     Monitoring ........................................ 34
                                     5     Optional PC board slot ...................... 48
                                     6     Reference & Timing ......................... 49
                                     7     Inspiratory control ............................. 61
                                     Patient unit – Pneumatic section ............. 62
                                     8     Inspiratory valve unit – Air ................ 62
                                     9     Inspiratory valve unit – O2 . ............... 67
                                     10 Inspiratory valve unit – Optional ....... 67
                                     11 Inspiratory channel ........................... 68
                                     12 Inspiratory pressure .......................... 70
                                     13 Expiratory pressure .......................... 71
                                     14 Expiratory channel ............................ 71
                                     15 Valve control ..................................... 73
                                     16 Expiratory valve ................................ 74
                                     17 Optional interface ............................. 74
                                     18 Expiratory flow linearization .............. 75
                                     19 Cooling system ................................ 76
                                     Patient unit – Power section .................... 76
                                     20 Mains power .................................... 76
                                     21 Power supply ................................... 77
                                     22 Operating power .............................. 79
                                     23 Interconnection cable ....................... 80

E382 E380E 061 01 03 01     Siemens-Elema AB                                                     21
Description of functions                                                  Servo Ventilator 300/300A



    General
    As described in chapter Introduction, the        In this Description of functions, the words
    Servo Ventilator 300 can be divided into two     inspiratory and expiratory are used in the
    Main units; Control unit and Patient unit. The   first place to indicate the site of a part (e g
    two main units are connected to each other       INSPIRATORY CHANNEL).
    with the Interconnection cable.
                                                     The words inspiration and expiration are
    For the purpose to describe all different        used in the first place to indicate the
    functions in the Servo Ventilator 300, the       corresponding time interval during
2   two main units are further divided as
    follows:
                                                     ventilation (e g INSPIRATION TIME)

    Main unit
     Section
         Main block
          Functional block
    Example:
    Control unit
     Panel section
         1 Front panel
          1.1 Patient range selection
          1.2 Airway pressure
          etc.
         2 Panel interface
          2.1 Panel CPU including PAN-PROM
              functions
          2.2 Parameter buffers
          etc.
    Functional descriptions of the Main blocks
    and Functional blocks can be found in this
    chapter. This functional description is based
    on and refers to the Functional block diagram
    that can be found as a fold out on the rear
    cover.




    22                                     Siemens-Elema AB                    E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                  Description of functions



Microprocessor module                              PC 1588 includes a green/red WATCHDOG LED
PC 1588                                            D1 (3). The LED indications are as follows:

A MICROPROCESSOR MODULE, PC1588 (1), is            – Green LED: Microprocessor program is
included on each one of the following five           running.
PC boards:                                         – Red LED: Program error indication.
• PC 1605 – REFERENCE & TIMING                     – LED not lit: Microprocessor program not
• PC 1608 – MONITORING                               running.

• PC 1614 – PANEL INTERFACE                        PC 1588 also includes an exchangable
                                                   program memory, PROM (2). The function             2
• PC 1615 – EXP. FLOW LINEARIZATION                of PC 1588 depends on this PROM which is
• PC 1616 – INSPIRATORY CONTROL.                   different on the five different enumerated
                                                   boards. The PROM functions are mentioned
PC 1588 is mounted with its two rows of            in the description of the following blocks:
connectors (P65 and P66) into the
corresponding connectors (HYB1) on the             • 2.1 Panel CPU (PAN-PROM)
above mentioned PC boards.                         • 4.1 Monitor/Alarm CPU (MON-PROM)
PC 1588 is a programmable digital control          • 6.1 Mode CPU (REF-PROM)
block including:
                                                   • 7.1 Mixer CPU (MIX-PROM)
• Microprocessor
                                                   • 18.1 Linearizing CPU (LIN-PROM)
• Analog – digital conversion
                                                   Note – The MICROPROCESSOR -block on
• Digital – analog conversion                      PC 1587 – COMPUTER INTERFACE, includes two
• Digital I/O.                                     PROM modules (COM-PROM).
                                                   PC 1588 is not used on PC 1587.




                                      xxx
                                      xxx




           1
300-A13X




                          3           2


E382 E380E 061 01 03 01                   Siemens-Elema AB                                      23
Description of functions                                                                Servo Ventilator 300/300A



                                                                    Version labels
                                                                    A label (1) showing all PROM-versions
                                                                    included in the ventilator is attached to the
                                                                    control section. When replacing a PROM,
                                                                    attach the new label (2) that is delivered with
                                                                    the new PROM onto the old label on the
                                                                    control section.
                                                                    Changing PROM-versions in the ventilator is
2                                                                   used to add or change functions in the
                                                                    ventilator. For that reason, it may also be
                                                                    necessary to replace the Operating Manual
                        1                                           in connection with PROM replacements.
                                                                    The Operating Manual, valid for the
                                                                    combination of PROMs used in the
                                                       3            ventilator, is identified with a version
                                                                    number stated both on the manual and on
                                                           2        the label (3). When replacing the Operating
                                            0
                                          30
                                              SV




                                                                    Manual with a new version of the manual,
                  8.x

                            E
                            13
                                E3
                                 X
                                    XX
                  SI al



                                     XX
                 R anu



                                         XX
               VE ting M




                                                                    attach the new label (3) that is delivered with
                     N
                    O

                            ra  e
                                Op



                                            16 V6. E
                                          PC OM 380
                                              08 0X

                                              PR X E
                                            N- X
                                          MO XX X




                                                                    the new Operating Manual onto the old label
                                             XX




                                                                    on the control section.
    300-A14X




    24                                                     Siemens-Elema AB                  E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                               Description of functions



Control unit – Panel section                    1.1 Patient range selection
The Panel section consists of the main          For selection of one of three available
blocks 1 – 2 which are described below.         patient ranges:
                                                • 1 three-position selector. This switch
                                                  generates selection information from two
1 Front panel                                     switch decks delivering one two bit code
This main block contains the front panel          and corresponding two bit complementary
film, in different language versions, and the     code.
various displays and controls with
connection cables and their connectors.
                                                • 3 yellow indicator LEDs.                            2
                                                The following table indicates how Patient
The optional front panel function for SV 300A   range selection affects some other
is described in section 1A AUTOMODE that can    parameters:
be found after 1.8 MODE SELECTION.




 Parameter                                      Adult      Pediatric     Neonate          Unit

 Apnea alarm – delay time                         20           15            10            s

 ”Bias flow” = Constant flow during               32           16             8         ml/s
 expiration time                                   2            1            0.5       l/min
                                                 3000         500            200        ml/s
 Max. inspiratory flow
                                                 180           30             12       l/min
 Max. (preset or measured) inspiratory or
                                                 3999         399            39           ml
 expiratory tidal volume

 FLOW REF overrange limit                         3.3         0.55           0.22          l/s

 Expiratory minute volume alarm:
  Setting range for Upper alarm limit           0 - 60       0 - 60        0-6         l/min
  Setting range for Lower alarm limit           * - 40       * - 40       * - 40       l/min
 * Lowest value for normal (digital) alarm       0.3          0.3          0.06        l/min
 * Lowest value for backup alarm                 0.2          0.2          0.05        l/min
 For Automode: Time duration since last
 patient trigger when Control/Support
                                                  12            8             5            s
 mode switches from Support mode to
 Control mode.




E382 E380E 061 01 03 01                Siemens-Elema AB                                          25
Description of functions                                                 Servo Ventilator 300/300A



    1.2 Airway pressure                             1.3 Respiratory pattern
    For control and monitoring of AIRWAY            For control and/or monitoring of respiratory
    PRESSURE parameters:                            pattern parameters:
    • 4 four-character LED displays.                • 4 four-character LED displays
    • 1 double column LED bargraph.                 • 5 potentiometer knobs:
    • 5 potentiometer knobs:                          – CMV freq. b/min. This control is
                                                        equipped with two parallell
     – Upper press. limit. Push button release
2      function for one-way mechanical stops
       at approx. 60, 80 and 100 cm H2O.
                                                        potentiometers – one for the control
                                                        function and the other for the monitor/
                                                        alarm function.
     – Pressure Control Level above PEEP.
                                                      – Insp. time %. Push button release
       Push button release function for one-
                                                        function for two-way mechanical stops
       way mechanical stops at approx. 30, 60
                                                        at approx. 20, 50 and 70%.
       and 80 cm H2O.
                                                      – Pause time %. Push button release
     – Pressure Support Level above PEEP.
                                                        function for one-way mechanical stop at
       Push button release function for one-
                                                        approx. 20%.
       way mechanical stops at approx. 30, 60
       and 80 cm H2O.                                 – Insp. rise time %. Push button release
                                                        function for one-way mechanical stop at
     – PEEP. Push button release function for
                                                        approx. 1%.
       one-way mechanical stop at approx.
       20 cm H2O.                                     – SIMV freq. b/min.
     – Trigg. sensitivity Level below PEEP.         • 5 yellow indicator LEDs.
    • 5 yellow indicator LEDs.
                                                    1.4 Volume
                                                    For control and/or monitoring of minute
                                                    volume and tidal volume parameters:
                                                    • 2 four-character green LED displays.
                                                      Showing information generated via
                                                      2.1 PANEL CPU.
                                                    • 3 four-character red LED displays.
                                                      Showing information generated via 4.1
                                                      MONITOR/ALARM CPU.

                                                    • 2 fields of double column LED bargraph.
                                                    • 3 potentiometer knobs:
                                                      – Volume
                                                      – Upper alarm limit
                                                      – Lower alarm limit.
                                                    • 3 yellow indicator LEDs.


    26                                     Siemens-Elema AB                   E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                   Description of functions



1.5 O2 concentration                                1.8 Mode selection
For control and monitoring of O2                    For the selection of Ventilation Modes and
concentration parameters:                           control of the Set Parameter Guide
                                                    functions:
• 1 four-character LED display. Showing
  information generated via 2.1 PANEL CPU.          • 1 green Mains indicator LED
• 1 potentiometer knob with push button             • 1 ten-position selector always supplied
  release function for two-way mechanical             with high impedance 24 V. This switch
  stop at approx. 60 %. This control is               generates selection information from two
  equipped with two parallell potentiometers
  – one for the regulation function and the
                                                      switch decks delivering complementary
                                                      code via PC board (PC 1632). The mode
                                                                                                       2
  other for the monitor/alarm function.               Optional can not be selected.
• 1 non-locking, two-way selector for               • 10 different fields each comprising:
  Oxygen breaths and for Start breath.
                                                     – 1 yellow indicator LED
• 2 yellow indicator LEDs.
                                                     – 1 touch sensor.

1.6 Alarms and messages
                                                    1A Automode front panel
For the display of alarms and messages:
                                                    For the control and monitoring of the
• 1 eight-character LED display. Showing            Automode functions:
  information generated via 4.1 MONITOR/
  ALARM CPU.
                                                    • 1 two-position selector for turning the
                                                      Automode On and Off
• 8 different alarm fields each comprising:
                                                    • 2 yellow indicator LEDs.
  – 1 red indicator LED
  – 1 yellow indicator LED
  – 1 touch sensor
• 1 non-locking, two-way selector for Reset
  and for 2 min off. This information is
  handled by the main block 4 MONITORING.


1.7 Pause hold
For the control, via main block 6   REFERENCE   &
TIMING, of pause hold functions:

• 1 non-locking two-way selector for Insp.
  pause hold and for Exp. pause hold.




E382 E380E 061 01 03 01                 Siemens-Elema AB                                         27
Description of functions                                                Servo Ventilator 300/300A



    2 Panel interface                                2.3 Inputs
    This main block consists of the board            Analog multiplexor for input signals to the
    PC 1614 PANEL INTERFACE including its            block 2.1 PANEL CPU.
    PC1588 MICROPROCESSOR MODULE . The board
    is connected (via N18, N19 and N40) to the
    mother board (PC 1617) CONTROL INTERCONNEC-      2.4 LED matrix
    TION and it has connectors (P81 – P99) for       Time multiplexed addressing and driving
    the connection of FRONT PANEL controls.          stages for the block 2.5 LED ARRAY. The input

2   The FRONT PANEL touch sensors as well as
    displays and indicator LEDs are integrated
                                                     data is given from the block 2.1 PANEL CPU.

    parts of this board. Refer to 1 FRONT PANEL in
    this chapter.                                    2.5 LED array
    A connector, N20 (on some older PC-boards,       This block contains the light emitting diodes
    the connector is marked N80), is provided        of the front panel bargraphs and the yellow
    on the board for additional function. This       light emitting diodes at the front panel
    connector can be used to connect the             settings.
    optional PC 1745 AUTOMODE PANEL INTERFACE        Some flashing function in front panel
    used in SV 300A. This function is described      bargraphs are mentioned in 4.1 MONITOR/
    in section 2A AUTOMODE PANEL INTERFACE that      ALARM CPU
    can be found after 2.9 MONITOR DISPLAY.
    .
                                                     2.6 Panel display
    2.1 Panel CPU including PAN-
        PROM functions                               This block contains all the green alpha-
                                                     numerical four character light emitting diode
    This block includes the board PC 1588 with       displays on the front panel. The input data is
    a PAN-PROM. Some details of the PC 1588 are      given from the block 2.1 PANEL CPU.
    mentioned in section Microprocessor
    module PC 1588 in this chapter.
    The PAN-PROM function includes the handling      2.7 Mode select
    of input signals from the main block 1 FRONT     Input stage for the signals from the front
    PANEL, calculation of some displayed values      panel Mode selector. The input is available
    and handling of signals to the displays on       as four bit code and corresponding four bit
    the main block 1 FRONT PANEL.                    complementary code generated via the
                                                     small interconnecting circuit board PC 1632
                                                     on the Mode selector. This block gives the
    2.2 Parameter buffers                            data to the block 2.1 PANEL CPU.
    Input stage for potentiometers on the main
    block 1 FRONT PANEL.




    28                                      Siemens-Elema AB                 E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                Description of functions



2.8 Opto switch                                  2A Automode panel
Time multiplexed driving stage and sensor        interface
receiving stage for all the infra-red sensor     The board PC 1745 AUTOMODE PANEL INTERFACE
touch pads on the front panel. The function      contains electronics for handling of the
is controlled by the block 2.1 PANEL CPU         Automode switch.
including an automatic adaption of the touch
pad sensitivity to the ambient light level.      The status of the switch is given to the
This block gives the data to the block 2.1       blocks 2.1 PANEL CPU, 5 MONITORING and
PANEL CPU.                                       6 REFERENCE & TIMING. The LEDs are driven
                                                 from the main block 4 MONITORING.                  2
2.9 Monitor display
This block contains all the red alpha-
numerical four character light emitting diode
displays as well as eight yellow and eight
red alarm indicating light emitting diodes
below the Alarms and messages display on
the front panel. This block is driven directly
from the main block 4 MONITORING.




E382 E380E 061 01 03 01                Siemens-Elema AB                                       29
Description of functions                                              Servo Ventilator 300/300A



    Control unit – Control section                  There are five connectors (N80 – N84) on
                                                    PC 1587.
    The Control section consists of the main
    blocks 3 – 7 which are described below.
    A ”mother” board, PC 1617 CONTROL
    INTERCONNECTION,provides signal connections
    via connectors (but no electronic functions).



2   3 Computer interface
    The ventilator can be equipped with the
    board PC 1587 COMPUTER INTERFACE to provide
    the interface functions described below.                                  N80
    If these interface functions are not needed,
    the board PC 1665 COMPUTER INTERFACE DUMMY                   N82
                                                                              N81
    can be mounted instead of PC 1587.
    The PC 1665-function is described in section                 N83
    3A COMPUTER INTERFACE DUMMY that can be
    found after section 3.9 LED INDICATORS.
                                                                 N84
    The board PC 1587 COMPUTER INTERFACE was
    replaced during the mandatory upgrades:
    – SV 300, Upgrade 1995-09 (US market)
    – SV 300, Upgrade 1995-10 (Rest of World)       300-A15X

    As the board PC 1587D was included in
    these mandatory upgrades, all SV 300/300A
                                                    Only accessories, supplies or auxiliary
    must be equipped with PC 1587D (or higher).
                                                    equipment listed in Siemens-Elema catalogs
    The main block COMPUTER INTERFACE consists      (”Products and accessories” Order No.
    of the board PC 1587 COMPUTER INTERFACE         90 34 562 E323E and ”Spare and exchange
    which includes one microprocessor and a         parts” Order No. 90 34 570 E323E) must be
    COM-PROM. The COM-PROM is divided in two        connected to or used in conjunction with the
    PROM modules                                    ventilator.
                                                    Warning: Use of accessories and auxiliary
                                                    equipment other than those specified in
                                                    these catalogs may degrade safety and
                                                    performance of the ventilator.




    30                                     Siemens-Elema AB                E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                Description of functions



3.1 Microprocessor CPU                          The voltage level in the battery circuit is
including COM-PROM functions                    monitored resulting in an error indication on
                                                LED D3 (see 3.9 LED indicators) if the
This includes functions for computer
                                                voltage level is below 3.6 V or above 4.5 V.
communication (RS-232) via 3.7 SERIAL
                                                An information message is also displayed on
COMMUNICATION PORTS as described in the
                                                Servo Screen 390 if such unit is connected
Reference Manual "Servo Ventilator 300/
                                                to the SV 300.
300A, Computer Interface (Firmware version
                                                Note – The voltage level measured in the
2.X)", Order No. 63 14 061.
                                                battery circuit is higher than the voltage level
This block also contains a real time clock
with battery backup, driven by a quartz
                                                in the battery itself.
                                                                                                    2
crystal of 32768 Hz. The purpose of this
clock is to time stamp the trend values         3.5 Master/Slave connection
stored in the ventilator. There is also a
                                                15 pole D-sub connector (N80). Can be used
possibility to read the actual time through
                                                for the synchronization of two Servo
the RS-232 communication.
                                                Ventilator 300. Only Siemens connection
                                                cable must be used.
3.2 I/O buffers                                 Pin configuration and signal names can be
                                                found in chapter "8. Diagrams".
Analog and digital input and output stages
for the main block 3 COMPUTER INTERFACE.
                                                3.6 Analog I/O terminal
3.3 A/D, D/A                                    62 pole D-sub connector (N81). Can be used
                                                for connection of monitoring/recording
Analog to digital and digital to analog
                                                equipment.
conversion in connection with the block
3.1 MICROPROCESSOR CPU.                         Input control signals can be connected at
                                                N81 to control some ventilator functions
                                                (e.g. when using Bi-Phasic Ventilation).
3.4 Battery                                     These input signals are routed via two
On board mounted Lithium battery (3.5 V).       sockets, J2 and J3, on PC 1587.
The battery backs up the trend data memory      To enable the control functions, jumpers
and the real time clock in the block            must be mounted in the sockets.
3.1 MICROPROCESSOR CPU when the SV 300 is       Pin configuration, signal names and jumpers
switched off.                                   can be found in chapter "8. Diagrams".
Normal service interval for exchange of the
battery is approx. 5 years. Replacement of
the battery is described in chapter
”4. Service procedures”.




E382 E380E 061 01 03 01                Siemens-Elema AB                                       31
Description of functions                                               Servo Ventilator 300/300A



    3.7 Serial communication ports                  3.9 LED indicators
    26 pole D-sub connectors (N82 and N83) for      There are 9 green/red LED INDICATORS
    RS-232-C data communication. For com-           mounted on PC 1587. The LEDs are visible
    munication protocol, see Reference Manual       if the control section cover is removed.
    "Servo Ventilator 300/300A, Computer
    Interface (Firmware version 2.X)", Order No.
    63 14 061 E380E.
                                                                            D9            D1
    The connectors are identical with one

2   exception; the internal clock can only be set
    via N82.
    Serial communication is indicated on the
    LEDs D6 – D9. See 3.9 LED INDICATORS.
    Pin configuration and signal names can be
    found in chapter "8. Diagrams".


    3.8 Analog input & Digital code
        port
    44 pole D-sub connector (N84). Optional
    input interface.
    Pin configuration and signal names can be
    found in chapter "8. Diagrams".
                                                    300-L24X




                                                    LED functions
                                                    D1: WATCHDOG. During startup the LED will
                                                    flash for a short moment and then be lit red
                                                    for a short moment. It will then be lit green
                                                    again.
                                                    When the startup test is completed, the
                                                    LED indications are as follows:
                                                    – Green LED: Microprocessor program is
                                                      running.
                                                    – Red LED: Program error indication.
                                                    – LED not lit: Microprocessor program not
                                                      running.




    32                                    Siemens-Elema AB                  E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                 Description of functions



D2 – D5: TEST STATUS LED. Error indications      3A Computer interface
during System startup test. The startup test        dummy
is divided in several sequences.
                                                 The PC 1665 COMPUTER INTERFACE DUMMY is
The first sequence is a LED test that starts     mounted instead of PC 1587 COMPUTER
with all LEDs red. The LEDs will then            INTERFACE, in the case that there is no need
change to green, one at the time (starting       for serial communication with the ventilator.
with D2), until all four LEDs are green.
                                                 It is needed as a holder for the PC 1608, as
In the hardware test sequence, D3 can be of      well as for disabling the error signal from the
special interest for troubleshooting.
In the first phase of the hardware test, D3 is
                                                 COMPUTER INTERFACE board and the signals for
                                                 Master – Slave.
                                                                                                     2
lit red for a few seconds indicating COM-PROM
error. In the second phase of the hardware
test, D3 is lit red for a few seconds
indicating BATTERY error.
The hardware test sequence ends showing
all LEDs green if no error was detected and
all LEDs red if any error was detected.
Note – As COM-PROM and BATTERY are the only
spare parts for PC 1587, any other LED error
indications than mentioned above results in
a PC 1587 replacement.
D6 – D7: SERIAL INTERFACE 1. Rapidly flashing
green and red indicates communication on
connector N82. D6 indicates input and D7
indicates output.
D8 – D9: SERIAL INTERFACE 1. Rapidly flashing
green and red indicates communication on
connector N83. D8 indicates input and D9
indicates output.




E382 E380E 061 01 03 01                 Siemens-Elema AB                                       33
Description of functions                                               Servo Ventilator 300/300A



    4 Monitoring                                    4.1 Monitor/Alarm CPU including
                                                        MON-PROM functions
    This main block consists of the board
    PC 1608 MONITORING including its PC 1588        This block includes the board PC 1588 with
    MICROPROCESSOR MODULE and the                   a MON-PROM. Some details of the PC 1588
    4.7 LOUDSPEAKER.                                are mentioned in section Microprocessor
                                                    module PC 1588 in this chapter. The MON-
    New versions of PC 1608 MONITORING and          PROM functions include the (program control)
    MON-PROM  was introduced in production          digital monitoring of certain parameters and
    during 1996 and also as an optional upgrade     the activation of alarm functions:
2   on delivered units. The upgrade was called
    ”SV 300, ALARM SYSTEM ENHANCEMENT KIT 96-05”.   • Front panel indications via the block
    The new versions was:                             2.9 MONITOR DISPLAY
    – PC 1608E                                      • Audible indications via the block
                                                      4.3 SOUND & ALARM CONTROL.
    –   MON-PROM   V6.00
                                                    In the following description (4.1.1 – 4.1.8)
    As the upgrade was optional and not             alarms are mentioned in groups, each group
    mandatory, a number of not upgraded             associated with corresponding Alarm field
    SV 300 are equipped with MON-PROM V5.01         within the Front panel field ”Alarms and
    (and PC 1608A, -B, -C or -D). These units       Messages”.
    must be equipped with a FAILURE ALARM BOX.
    Note – PC 1608E is backwards compatible         For each Alarm field this description of the
    and can be equipped with MON-PROM V5.01.        normal (digital control) alarm contains:
    The FAILURE ALARM BOX must be used also on      • Alarms and Messages - Alarm field
    these units.                                      name
    The alarm functions of the FAILURE ALARM BOX    • Parameter monitored for alarm, PMA
    are integrated on MON-PROM V6.00/PC 1608E         and corresponding Reference value
    and the FAILURE ALARM BOX is not used on
                                                    • Alarm table
    SV 300 equipped with MON-PROM V6.00/
                                                      containing conditions for alarm.
    PC 1608E (or higher).
    For more information about PROM, PC board
    and Operating Manual versions, see chapter
    ”6. Product change history”.




    34                                    Siemens-Elema AB                  E382 E380E 061 01 03 01
Servo Ventilator 300/300A                                                          Description of functions



4.1.1 Airway pressure
PMA:      PI =   INSP. PRESSURE;   signal from main block 12   INSPIRATORY PRESSURE.

          PE =   EXP. PRESSURE;    signal from main block 13   EXPIRATORY PRESSURE.

          LIM_PRESS.L = Signal from the block 6.1        MODE CPU   including   REF-PROM   functions.
Reference value Pmax = Preset Upper pressure limit (on Front panel).
Reference value PEEP = Preset PEEP Level (on Front panel).

Alarm table:
                                                                                                              2
    ”Alarms and messages”                 Alarm condition                Remark
    Airway pressure too high       2
                                          PI > Pmax or                   At true condition 1
                                          PE > Pmax                      INSPIRATION is inhibited
                                                                         and EXPIRATION is started.
    High continuous pressure              PI > PEEP + 15 cm H2O or Alarm if any of the
                                          PE > PEEP + 15 cm H2O    conditions remains true
                                                                   during 15 s.
    Limited pressure                      LIM_PRESS.L = active,          LIM_PRESS.L signal 3 only
                                          (corresponds to condition:     used during PRVC or VS
                                          PREF = Pmax – 5 cm H2O         mode. Alarm if the
                                          and resulting Tidal/Minute     mentioned alarm condition
                                          volume ventilation is less     is true during three
                                          than front panel setting)      consecutive breaths.
1
    Refer to 6.1.2 Respiratory timing - condition (3).
2
    Identical condition is also monitored separately within the main block 2. Panel interface
    and that is used for the following function: At true condition the Airway pressure bargraph
    indication for Upper pressure limit is flashing.
3
    Refer to 6.1.1 Reference generation - PRVC and VS mode.

Monitoring also controls the following opening and closing of the Safety valve:

    Opening condition                     Remark
    PI > Pmax + 6 cm H2O or               The Safety valve opens. If the opening was caused by
    PE > Pmax + 6 cm H2O                  any of these conditions the Backup alarm system
                                          generates continuous audible alarm.

When opened the Safety valve will stay open until following conditions are true:
    Precondition                                         Closing condition
    Safety valve stays open at least during              PI < Pmax + 5 cm H2O and
    400 ms.                                              PE < Pmax + 5 cm H2O
For Airway pressure alarm also see 4.5 Backup alarm system.

E382 E380E 061 01 03 01                       Siemens-Elema AB                                          35
Description of functions                                               Servo Ventilator 300/300A



    4.1.2 O2 concentration
    PMA:    O2 concentration = Signal generated within the block 4.1 using O2 CONC. signal from
            the block 11.3 O2 CELL which is digitally compensated for mean airway pressure and
            barometric pressure according to BAROMETER PRESSURE signal from 4.6 BAROMETER
            PRESSURE TRANSDUCER.

    Reference value Set O2 = Preset O2 concentration (on Front panel).

    Alarm table:
2    ”Alarms and messages” text           Precondition                Alarm condition
     O2 conc too high                     21% < Set O2 < 100%         PMA > Set O2 + 6%


     O2 conc too low                      21% < Set O2 < 24%          PMA < 18%
                                          24% < Set O2 < 96%          PMA < Set O2 – 6%
                                          96% < Set O2                PMA < 90%

     ”Alarms and messages” text           Additional function
     O2 SENSOR                            If the O2 cell is disconnected; High priority alarm
                                          that can be changed to a caution signal.

    If above mentioned O2 concentration alarm condition is true during less than 55 s the alarm
    is not registered in the alarm memory.
    If the O2 concentration setting is changed more than 2 percentage units, the alarm is
    automatically muted for maximum 55 s. This also applies at activation of ”Oxygen breath”.
    For O2 concentration alarm also see 4.5 Backup alarm system.




    4.1.3 Optional alarm field
    On earlier versions of the ventilator, the alarm field name was ”CO2 concentration”.
    This optional alarm field is not used.




    36                                    Siemens-Elema AB                  E382 E380E 061 01 03 01
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual
Sv300service manual

Mais conteúdo relacionado

Mais procurados

Care of ventilated patient
Care of ventilated patientCare of ventilated patient
Care of ventilated patientajishkt
 
Settings Use and Maintenance of Mechanical Ventilator
Settings Use and Maintenance of Mechanical VentilatorSettings Use and Maintenance of Mechanical Ventilator
Settings Use and Maintenance of Mechanical VentilatorSurendran Radjou
 
Mechanical ventilation in emergency
Mechanical ventilation in emergencyMechanical ventilation in emergency
Mechanical ventilation in emergencyMagdy Khames Aly
 
Basic life support
Basic life supportBasic life support
Basic life supportViki Patidar
 
4.hazards of working in the operation room
4.hazards of working in the operation room4.hazards of working in the operation room
4.hazards of working in the operation roomHenok Eshetie
 
Pre anesthetic medications by Meghana C
Pre anesthetic medications by Meghana CPre anesthetic medications by Meghana C
Pre anesthetic medications by Meghana CMEGHANA C
 
Anaesthetic Equipments By Dr. Mehedi Hasan
Anaesthetic Equipments By Dr. Mehedi HasanAnaesthetic Equipments By Dr. Mehedi Hasan
Anaesthetic Equipments By Dr. Mehedi HasanMehedi Hasan
 
ECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenationECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenationprapulla chandra
 
Mechanical Ventilator by Kiran Bhardwaj
Mechanical Ventilator by Kiran BhardwajMechanical Ventilator by Kiran Bhardwaj
Mechanical Ventilator by Kiran BhardwajKiran Bhardwaj
 
AUTOMATED EXTERNAL DEFIBRILLATOR
AUTOMATED EXTERNAL DEFIBRILLATORAUTOMATED EXTERNAL DEFIBRILLATOR
AUTOMATED EXTERNAL DEFIBRILLATORsathish sak
 
Anesthesia & its types
Anesthesia & its typesAnesthesia & its types
Anesthesia & its typesTushar singh
 

Mais procurados (20)

Care of ventilated patient
Care of ventilated patientCare of ventilated patient
Care of ventilated patient
 
Settings Use and Maintenance of Mechanical Ventilator
Settings Use and Maintenance of Mechanical VentilatorSettings Use and Maintenance of Mechanical Ventilator
Settings Use and Maintenance of Mechanical Ventilator
 
Mechanical ventilation in emergency
Mechanical ventilation in emergencyMechanical ventilation in emergency
Mechanical ventilation in emergency
 
Basic life support
Basic life supportBasic life support
Basic life support
 
4.hazards of working in the operation room
4.hazards of working in the operation room4.hazards of working in the operation room
4.hazards of working in the operation room
 
Pre anesthetic medications by Meghana C
Pre anesthetic medications by Meghana CPre anesthetic medications by Meghana C
Pre anesthetic medications by Meghana C
 
General anesthesia
General anesthesiaGeneral anesthesia
General anesthesia
 
Nyha
NyhaNyha
Nyha
 
Anaesthetic Equipments By Dr. Mehedi Hasan
Anaesthetic Equipments By Dr. Mehedi HasanAnaesthetic Equipments By Dr. Mehedi Hasan
Anaesthetic Equipments By Dr. Mehedi Hasan
 
ECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenationECMO - extracorporeal membrane oxygenation
ECMO - extracorporeal membrane oxygenation
 
Modes of mechanical ventilation
Modes of mechanical ventilationModes of mechanical ventilation
Modes of mechanical ventilation
 
Pediatric airway management
Pediatric airway managementPediatric airway management
Pediatric airway management
 
Mechanical Ventilator by Kiran Bhardwaj
Mechanical Ventilator by Kiran BhardwajMechanical Ventilator by Kiran Bhardwaj
Mechanical Ventilator by Kiran Bhardwaj
 
Ventilator machine
Ventilator machineVentilator machine
Ventilator machine
 
General Anesthesia
General AnesthesiaGeneral Anesthesia
General Anesthesia
 
AUTOMATED EXTERNAL DEFIBRILLATOR
AUTOMATED EXTERNAL DEFIBRILLATORAUTOMATED EXTERNAL DEFIBRILLATOR
AUTOMATED EXTERNAL DEFIBRILLATOR
 
Central line
Central line Central line
Central line
 
Extubation
Extubation Extubation
Extubation
 
Anesthesia & its types
Anesthesia & its typesAnesthesia & its types
Anesthesia & its types
 
Mcq exam 2011
Mcq exam 2011Mcq exam 2011
Mcq exam 2011
 

Destaque

Servo 300 function check (each use)
Servo 300 function check (each use)Servo 300 function check (each use)
Servo 300 function check (each use)Ruderocker Billy
 
Servo 300 exchange 0 2 cell
Servo 300 exchange 0 2 cellServo 300 exchange 0 2 cell
Servo 300 exchange 0 2 cellRuderocker Billy
 
Siemens 300 a_service_manual
Siemens 300 a_service_manualSiemens 300 a_service_manual
Siemens 300 a_service_manualRuderocker Billy
 
Siemens 300 a_service_manual
Siemens 300 a_service_manualSiemens 300 a_service_manual
Siemens 300 a_service_manualRuderocker Billy
 
TEXTOS EN MATRIZ LED DE 64X16 CON ARDUINO
TEXTOS EN MATRIZ LED DE 64X16 CON ARDUINO TEXTOS EN MATRIZ LED DE 64X16 CON ARDUINO
TEXTOS EN MATRIZ LED DE 64X16 CON ARDUINO manuel_lito
 
Proyecto Microcontroladores
Proyecto MicrocontroladoresProyecto Microcontroladores
Proyecto Microcontroladoresimera
 
Siemens product & accessory price list
Siemens product & accessory price listSiemens product & accessory price list
Siemens product & accessory price listRuderocker Billy
 
Sistema interfaz visual para el control de la respiración en maniobras autonó...
Sistema interfaz visual para el control de la respiración en maniobras autonó...Sistema interfaz visual para el control de la respiración en maniobras autonó...
Sistema interfaz visual para el control de la respiración en maniobras autonó...Ruderocker Billy
 
Microcontroladores tabla de referencia
Microcontroladores tabla de referenciaMicrocontroladores tabla de referencia
Microcontroladores tabla de referenciaRuderocker Billy
 
RESPIRADOR PARA EMERGENCIAS MEDICAS BioRohm PCA1
RESPIRADOR PARA EMERGENCIAS MEDICAS BioRohm PCA1RESPIRADOR PARA EMERGENCIAS MEDICAS BioRohm PCA1
RESPIRADOR PARA EMERGENCIAS MEDICAS BioRohm PCA1Ruderocker Billy
 
Lenguaje de ensamblador
Lenguaje de ensambladorLenguaje de ensamblador
Lenguaje de ensambladorluisbeltran861
 

Destaque (20)

Servo 300 function check (each use)
Servo 300 function check (each use)Servo 300 function check (each use)
Servo 300 function check (each use)
 
Servo 300 exchange 0 2 cell
Servo 300 exchange 0 2 cellServo 300 exchange 0 2 cell
Servo 300 exchange 0 2 cell
 
Siemens 300 a_service_manual
Siemens 300 a_service_manualSiemens 300 a_service_manual
Siemens 300 a_service_manual
 
como programar un pic
como  programar un piccomo  programar un pic
como programar un pic
 
Siemens 300 a_service_manual
Siemens 300 a_service_manualSiemens 300 a_service_manual
Siemens 300 a_service_manual
 
TEXTOS EN MATRIZ LED DE 64X16 CON ARDUINO
TEXTOS EN MATRIZ LED DE 64X16 CON ARDUINO TEXTOS EN MATRIZ LED DE 64X16 CON ARDUINO
TEXTOS EN MATRIZ LED DE 64X16 CON ARDUINO
 
Servo 300 3000 hour pm
Servo 300 3000 hour pmServo 300 3000 hour pm
Servo 300 3000 hour pm
 
Servo 300 1000 hour pm
Servo 300 1000 hour pmServo 300 1000 hour pm
Servo 300 1000 hour pm
 
Proyecto Microcontroladores
Proyecto MicrocontroladoresProyecto Microcontroladores
Proyecto Microcontroladores
 
Siemens product & accessory price list
Siemens product & accessory price listSiemens product & accessory price list
Siemens product & accessory price list
 
Servo 300 calibaration(2)
Servo 300 calibaration(2)Servo 300 calibaration(2)
Servo 300 calibaration(2)
 
Sistema interfaz visual para el control de la respiración en maniobras autonó...
Sistema interfaz visual para el control de la respiración en maniobras autonó...Sistema interfaz visual para el control de la respiración en maniobras autonó...
Sistema interfaz visual para el control de la respiración en maniobras autonó...
 
Siemens300 300 apart2
Siemens300 300 apart2Siemens300 300 apart2
Siemens300 300 apart2
 
Sv 300 spare parts list
Sv 300 spare parts listSv 300 spare parts list
Sv 300 spare parts list
 
Microcontroladores tabla de referencia
Microcontroladores tabla de referenciaMicrocontroladores tabla de referencia
Microcontroladores tabla de referencia
 
RESPIRADOR PARA EMERGENCIAS MEDICAS BioRohm PCA1
RESPIRADOR PARA EMERGENCIAS MEDICAS BioRohm PCA1RESPIRADOR PARA EMERGENCIAS MEDICAS BioRohm PCA1
RESPIRADOR PARA EMERGENCIAS MEDICAS BioRohm PCA1
 
Matriz de led
Matriz de ledMatriz de led
Matriz de led
 
Tasm
TasmTasm
Tasm
 
Ventilacion mecanica
Ventilacion mecanicaVentilacion mecanica
Ventilacion mecanica
 
Lenguaje de ensamblador
Lenguaje de ensambladorLenguaje de ensamblador
Lenguaje de ensamblador
 

Semelhante a Sv300service manual

Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...ufjjjsejfjksekm
 
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...ufjjdjjkskemme
 
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...ufjdjfjkskefmme
 
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...ufjjdfjkskemfmme
 
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...fujsjefjkskeksmem
 
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...fusjfjskefksmem
 
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...fjskekmmem
 
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...kfdjkskdmm
 
Proface GP2400, GP2500, GP2501, GP2600 manual
Proface GP2400, GP2500, GP2501, GP2600 manualProface GP2400, GP2500, GP2501, GP2600 manual
Proface GP2400, GP2500, GP2501, GP2600 manualvandungdhl
 
Terex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manualTerex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manualfuhsejskekmmm
 
Terex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manualTerex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manualfujsjefkskemm
 
Terex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manualTerex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manualfjjsieskekfmmse
 
Delta ia asda-b2_um_en_20141217
Delta ia asda-b2_um_en_20141217Delta ia asda-b2_um_en_20141217
Delta ia asda-b2_um_en_20141217hoa xuan
 
Liebherr a900 c zw edc f.d. litronic hydraulic excavator service repair manual
Liebherr a900 c zw edc f.d. litronic hydraulic excavator service repair manualLiebherr a900 c zw edc f.d. litronic hydraulic excavator service repair manual
Liebherr a900 c zw edc f.d. litronic hydraulic excavator service repair manualfjseazqkkdmmem
 
Vlt fc 300 operating instructions
Vlt fc 300 operating instructionsVlt fc 300 operating instructions
Vlt fc 300 operating instructionsToàn Huỳnh
 

Semelhante a Sv300service manual (20)

SOLAR PUMP INVERTER CIRCUIT DIAGRAM
SOLAR PUMP INVERTER CIRCUIT DIAGRAMSOLAR PUMP INVERTER CIRCUIT DIAGRAM
SOLAR PUMP INVERTER CIRCUIT DIAGRAM
 
Manual 8560
Manual 8560Manual 8560
Manual 8560
 
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
 
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
 
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
 
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
 
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
 
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
 
P7 user manual
P7 user manualP7 user manual
P7 user manual
 
1305071317146.pdf
1305071317146.pdf1305071317146.pdf
1305071317146.pdf
 
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh430 v 4 stroke air cooled v-twin gasoline engine service repair ma...
 
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
Kawasaki fh381 v 4 stroke air cooled v-twin gasoline engine service repair ma...
 
Proface GP2400, GP2500, GP2501, GP2600 manual
Proface GP2400, GP2500, GP2501, GP2600 manualProface GP2400, GP2500, GP2501, GP2600 manual
Proface GP2400, GP2500, GP2501, GP2600 manual
 
Terex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manualTerex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manual
 
Terex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manualTerex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manual
 
Terex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manualTerex telelift 2306 gladiator agrilift telescopic handler service repair manual
Terex telelift 2306 gladiator agrilift telescopic handler service repair manual
 
Delta ia asda-b2_um_en_20141217
Delta ia asda-b2_um_en_20141217Delta ia asda-b2_um_en_20141217
Delta ia asda-b2_um_en_20141217
 
Liebherr a900 c zw edc f.d. litronic hydraulic excavator service repair manual
Liebherr a900 c zw edc f.d. litronic hydraulic excavator service repair manualLiebherr a900 c zw edc f.d. litronic hydraulic excavator service repair manual
Liebherr a900 c zw edc f.d. litronic hydraulic excavator service repair manual
 
Vlt fc 300 operating instructions
Vlt fc 300 operating instructionsVlt fc 300 operating instructions
Vlt fc 300 operating instructions
 
Hướng dẫn vận hành Danfoss VLT FC 300
Hướng dẫn vận hành Danfoss VLT FC 300Hướng dẫn vận hành Danfoss VLT FC 300
Hướng dẫn vận hành Danfoss VLT FC 300
 

Mais de Ruderocker Billy

Introduccion a los microcontroladores pic y programacion de una matriz de led's
Introduccion a los microcontroladores pic y programacion de una matriz de led'sIntroduccion a los microcontroladores pic y programacion de una matriz de led's
Introduccion a los microcontroladores pic y programacion de una matriz de led'sRuderocker Billy
 
Cirugia model instalacion hidraulica y sanitaria
Cirugia model instalacion hidraulica y sanitaria Cirugia model instalacion hidraulica y sanitaria
Cirugia model instalacion hidraulica y sanitaria Ruderocker Billy
 
Siemens servo screen_390_service_manual
Siemens servo screen_390_service_manualSiemens servo screen_390_service_manual
Siemens servo screen_390_service_manualRuderocker Billy
 
TECHNICAL SPECIFICATIONS SERVO-s
TECHNICAL SPECIFICATIONS SERVO-sTECHNICAL SPECIFICATIONS SERVO-s
TECHNICAL SPECIFICATIONS SERVO-sRuderocker Billy
 
Hoja de datos Servo ventiladores!
Hoja de datos Servo ventiladores!Hoja de datos Servo ventiladores!
Hoja de datos Servo ventiladores!Ruderocker Billy
 
hoja de Datos VENTILACIÓN SERVO-s
hoja de Datos VENTILACIÓN SERVO-shoja de Datos VENTILACIÓN SERVO-s
hoja de Datos VENTILACIÓN SERVO-sRuderocker Billy
 

Mais de Ruderocker Billy (15)

Introduccion a los microcontroladores pic y programacion de una matriz de led's
Introduccion a los microcontroladores pic y programacion de una matriz de led'sIntroduccion a los microcontroladores pic y programacion de una matriz de led's
Introduccion a los microcontroladores pic y programacion de una matriz de led's
 
Diseño sala de cirugia
Diseño sala de cirugiaDiseño sala de cirugia
Diseño sala de cirugia
 
Cirugia model instalacion hidraulica y sanitaria
Cirugia model instalacion hidraulica y sanitaria Cirugia model instalacion hidraulica y sanitaria
Cirugia model instalacion hidraulica y sanitaria
 
Plano
PlanoPlano
Plano
 
Sala de cirugia
Sala de cirugiaSala de cirugia
Sala de cirugia
 
Cirugia model
Cirugia modelCirugia model
Cirugia model
 
Litiasis renal1
Litiasis renal1Litiasis renal1
Litiasis renal1
 
Siemens300 300 apart1
Siemens300 300 apart1Siemens300 300 apart1
Siemens300 300 apart1
 
Siemens servo screen_390_service_manual
Siemens servo screen_390_service_manualSiemens servo screen_390_service_manual
Siemens servo screen_390_service_manual
 
Servo-i ensamble
Servo-i ensambleServo-i ensamble
Servo-i ensamble
 
Servo i service manual
Servo i service manual Servo i service manual
Servo i service manual
 
TECHNICAL SPECIFICATIONS SERVO-s
TECHNICAL SPECIFICATIONS SERVO-sTECHNICAL SPECIFICATIONS SERVO-s
TECHNICAL SPECIFICATIONS SERVO-s
 
Servo-s Ensamble
Servo-s EnsambleServo-s Ensamble
Servo-s Ensamble
 
Hoja de datos Servo ventiladores!
Hoja de datos Servo ventiladores!Hoja de datos Servo ventiladores!
Hoja de datos Servo ventiladores!
 
hoja de Datos VENTILACIÓN SERVO-s
hoja de Datos VENTILACIÓN SERVO-shoja de Datos VENTILACIÓN SERVO-s
hoja de Datos VENTILACIÓN SERVO-s
 

Último

Transaction Management in Database Management System
Transaction Management in Database Management SystemTransaction Management in Database Management System
Transaction Management in Database Management SystemChristalin Nelson
 
Food processing presentation for bsc agriculture hons
Food processing presentation for bsc agriculture honsFood processing presentation for bsc agriculture hons
Food processing presentation for bsc agriculture honsManeerUddin
 
ICS2208 Lecture6 Notes for SL spaces.pdf
ICS2208 Lecture6 Notes for SL spaces.pdfICS2208 Lecture6 Notes for SL spaces.pdf
ICS2208 Lecture6 Notes for SL spaces.pdfVanessa Camilleri
 
Inclusivity Essentials_ Creating Accessible Websites for Nonprofits .pdf
Inclusivity Essentials_ Creating Accessible Websites for Nonprofits .pdfInclusivity Essentials_ Creating Accessible Websites for Nonprofits .pdf
Inclusivity Essentials_ Creating Accessible Websites for Nonprofits .pdfTechSoup
 
Keynote by Prof. Wurzer at Nordex about IP-design
Keynote by Prof. Wurzer at Nordex about IP-designKeynote by Prof. Wurzer at Nordex about IP-design
Keynote by Prof. Wurzer at Nordex about IP-designMIPLM
 
Active Learning Strategies (in short ALS).pdf
Active Learning Strategies (in short ALS).pdfActive Learning Strategies (in short ALS).pdf
Active Learning Strategies (in short ALS).pdfPatidar M
 
How to Add Barcode on PDF Report in Odoo 17
How to Add Barcode on PDF Report in Odoo 17How to Add Barcode on PDF Report in Odoo 17
How to Add Barcode on PDF Report in Odoo 17Celine George
 
Karra SKD Conference Presentation Revised.pptx
Karra SKD Conference Presentation Revised.pptxKarra SKD Conference Presentation Revised.pptx
Karra SKD Conference Presentation Revised.pptxAshokKarra1
 
Earth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatEarth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatYousafMalik24
 
4.16.24 21st Century Movements for Black Lives.pptx
4.16.24 21st Century Movements for Black Lives.pptx4.16.24 21st Century Movements for Black Lives.pptx
4.16.24 21st Century Movements for Black Lives.pptxmary850239
 
Concurrency Control in Database Management system
Concurrency Control in Database Management systemConcurrency Control in Database Management system
Concurrency Control in Database Management systemChristalin Nelson
 
Field Attribute Index Feature in Odoo 17
Field Attribute Index Feature in Odoo 17Field Attribute Index Feature in Odoo 17
Field Attribute Index Feature in Odoo 17Celine George
 
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptx
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptxQ4-PPT-Music9_Lesson-1-Romantic-Opera.pptx
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptxlancelewisportillo
 
Activity 2-unit 2-update 2024. English translation
Activity 2-unit 2-update 2024. English translationActivity 2-unit 2-update 2024. English translation
Activity 2-unit 2-update 2024. English translationRosabel UA
 
Full Stack Web Development Course for Beginners
Full Stack Web Development Course  for BeginnersFull Stack Web Development Course  for Beginners
Full Stack Web Development Course for BeginnersSabitha Banu
 
Influencing policy (training slides from Fast Track Impact)
Influencing policy (training slides from Fast Track Impact)Influencing policy (training slides from Fast Track Impact)
Influencing policy (training slides from Fast Track Impact)Mark Reed
 

Último (20)

Transaction Management in Database Management System
Transaction Management in Database Management SystemTransaction Management in Database Management System
Transaction Management in Database Management System
 
Food processing presentation for bsc agriculture hons
Food processing presentation for bsc agriculture honsFood processing presentation for bsc agriculture hons
Food processing presentation for bsc agriculture hons
 
ICS2208 Lecture6 Notes for SL spaces.pdf
ICS2208 Lecture6 Notes for SL spaces.pdfICS2208 Lecture6 Notes for SL spaces.pdf
ICS2208 Lecture6 Notes for SL spaces.pdf
 
LEFT_ON_C'N_ PRELIMS_EL_DORADO_2024.pptx
LEFT_ON_C'N_ PRELIMS_EL_DORADO_2024.pptxLEFT_ON_C'N_ PRELIMS_EL_DORADO_2024.pptx
LEFT_ON_C'N_ PRELIMS_EL_DORADO_2024.pptx
 
Inclusivity Essentials_ Creating Accessible Websites for Nonprofits .pdf
Inclusivity Essentials_ Creating Accessible Websites for Nonprofits .pdfInclusivity Essentials_ Creating Accessible Websites for Nonprofits .pdf
Inclusivity Essentials_ Creating Accessible Websites for Nonprofits .pdf
 
Keynote by Prof. Wurzer at Nordex about IP-design
Keynote by Prof. Wurzer at Nordex about IP-designKeynote by Prof. Wurzer at Nordex about IP-design
Keynote by Prof. Wurzer at Nordex about IP-design
 
Active Learning Strategies (in short ALS).pdf
Active Learning Strategies (in short ALS).pdfActive Learning Strategies (in short ALS).pdf
Active Learning Strategies (in short ALS).pdf
 
How to Add Barcode on PDF Report in Odoo 17
How to Add Barcode on PDF Report in Odoo 17How to Add Barcode on PDF Report in Odoo 17
How to Add Barcode on PDF Report in Odoo 17
 
Karra SKD Conference Presentation Revised.pptx
Karra SKD Conference Presentation Revised.pptxKarra SKD Conference Presentation Revised.pptx
Karra SKD Conference Presentation Revised.pptx
 
Earth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice greatEarth Day Presentation wow hello nice great
Earth Day Presentation wow hello nice great
 
Raw materials used in Herbal Cosmetics.pptx
Raw materials used in Herbal Cosmetics.pptxRaw materials used in Herbal Cosmetics.pptx
Raw materials used in Herbal Cosmetics.pptx
 
FINALS_OF_LEFT_ON_C'N_EL_DORADO_2024.pptx
FINALS_OF_LEFT_ON_C'N_EL_DORADO_2024.pptxFINALS_OF_LEFT_ON_C'N_EL_DORADO_2024.pptx
FINALS_OF_LEFT_ON_C'N_EL_DORADO_2024.pptx
 
YOUVE GOT EMAIL_FINALS_EL_DORADO_2024.pptx
YOUVE GOT EMAIL_FINALS_EL_DORADO_2024.pptxYOUVE GOT EMAIL_FINALS_EL_DORADO_2024.pptx
YOUVE GOT EMAIL_FINALS_EL_DORADO_2024.pptx
 
4.16.24 21st Century Movements for Black Lives.pptx
4.16.24 21st Century Movements for Black Lives.pptx4.16.24 21st Century Movements for Black Lives.pptx
4.16.24 21st Century Movements for Black Lives.pptx
 
Concurrency Control in Database Management system
Concurrency Control in Database Management systemConcurrency Control in Database Management system
Concurrency Control in Database Management system
 
Field Attribute Index Feature in Odoo 17
Field Attribute Index Feature in Odoo 17Field Attribute Index Feature in Odoo 17
Field Attribute Index Feature in Odoo 17
 
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptx
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptxQ4-PPT-Music9_Lesson-1-Romantic-Opera.pptx
Q4-PPT-Music9_Lesson-1-Romantic-Opera.pptx
 
Activity 2-unit 2-update 2024. English translation
Activity 2-unit 2-update 2024. English translationActivity 2-unit 2-update 2024. English translation
Activity 2-unit 2-update 2024. English translation
 
Full Stack Web Development Course for Beginners
Full Stack Web Development Course  for BeginnersFull Stack Web Development Course  for Beginners
Full Stack Web Development Course for Beginners
 
Influencing policy (training slides from Fast Track Impact)
Influencing policy (training slides from Fast Track Impact)Influencing policy (training slides from Fast Track Impact)
Influencing policy (training slides from Fast Track Impact)
 

Sv300service manual

  • 1. Servo Ventilator 300/300A Service Manual 21 Adu lt ic Ped iatr e Neo nat 15 261 5 19 5.0 11 19 259 2 1 . 13 266 0 . 35 5.1 EN S SI EM E382 E380E 061 01 03 01
  • 2. Important Servo Ventilator 300/300A Important General Calibration and functional check • Unless stated otherwise, the information • After any service intervention in Servo in this Service Manual is valid for: Ventilator 300/300A, perform a – Servo Ventilator 300 (SV 300) Calibration and Functional check according to instructions in the – Servo Ventilator 300A (SV 300A). Operating Manual. • The documentation for the Servo Ventilator 300/300A consists of: Hazard notices – Operating Manual • Before disassembling or assembling the – Service Manual Servo Ventilator 300/300A, make sure – Circuit Diagram that the: – Spare Parts Catalogue – Gas supply is disconnected. – Reference Manual, Computer Interface – Mains power cable is disconnected. • The Operating Manual is an indispens- able complement to the Service Manual – Mode selector is set to Ventilator off. for proper servicing. If the mode selector is set in any other position, the internal battery will • In addition to the Important information supply power to the PC boards. given here and in the related documents, always pay attention to applicable local – The back-up batteries are disconnec- and national regulations. ted when the power section is open. • Responsibility for the safe functioning of – All gas conveying parts are cleaned the equipment reverts to the owner or according to instructions in the user in all cases in which service or Operating Manual. repair has been done by a non- professional or by persons who are not Service employed by or authorized by Siemens, and when the equipment is used for • When working with ESD sensitive other than its intended purpose. components, always use a grounded wrist band and a grounded work surface. • Data on internal pressures in the Servo Adequate service tools must always be Ventilator 300/300A are given in Pa (bar). used. Airway pressures are given in cm H2O. • Some parts in the Servo Ventilator 300/ 1 hPa = 1 mbar 1 mbar = 1 hPa 1 kPa = 10 mbar 1 mbar = 0.1 kPa 300A are comprised by a spare parts 1 kPa = 0.01 bar 1 bar = 100 kPa exchange system. Your local Siemens 1 kPa i 10 cm H2O 1 cm H2O i 0.1 kPa representative supplies factory 1 kPa i 0.01 at 1 at ³ 100 kPa calibrated and tested exchange parts as 1 kPa i 0.01 kgf/cm2 1 kgf/cm2 i 100 kPa well as other spare parts indicated in the 1 kPa i 0.01 kp/cm2 1 kp/cm2 i 100 kPa spare parts catalogue. 1 kPa i 0.145 psi 1 psi i 6.9 kPa 2 Siemens-Elema AB E382 E380E 061 01 03 01
  • 3. Servo Ventilator 300/300A Important Important • The Servo Ventilator 300/300A must be To the responsible service personnel serviced at regular intervals by specially • The contents of this document are not trained personnel. The service intervals, binding. If any significant difference is 1000 and 3000 operating hours, are found between the product and this described in the Operating Manual. Any document, please contact Siemens for maintenance must be noted in a log further information. book provided for that purpose in accordance with national regulations. • We reserve the right to modify products We recommend that service is done as without amending this document or a part of a service contract with advising the user. Siemens. • Only Siemens authorized personnel shall • A 1000 hour overhaul must be performed be permitted to service or repair the after 1000 hours of operation or, at the Servo Ventilator 300/300A. Only lastest, every six months. In addition, Siemens-Elema exchange parts or the ventilator shall undergo a technical genuine spare parts must be used. PC safety check (Function check) twice a boards (spare or exchange parts) must year, at six months intervals or according always be kept in a package for sensitive to national regulations. electronic devices. • A 3000 hour overhaul must be performed • Siemens will not otherwise assume after every 3000 hours of operation or, responsibility for the materials used, the at the latest, once every year. work performed or any possible consequences of same. • The internal batteries shall be replaced every 3 years according to instructions in • The device complies with the this Service Manual. The stated battery requirements of the Medical back-up time, approx. 30 minutes, can Device Directive 93/42/EEC. be guaranteed only if they are used as power supply back-up at mains failure. • The battery on PC 1587 COMPUTER INTERFACE shall be replaced every 5 years according to instructions in this Service Manual. • Old non-functioning batteries and O2 cells must be returned to the place of purchase or to a place where they can be disposed of properly. Pb Batteries and O2 cells must not be disposed of with ordinary waste. E382 E380E 061 01 03 01 Siemens-Elema AB 3
  • 4. Notes Servo Ventilator 300/300A Notes 4 Siemens-Elema AB E382 E380E 061 01 03 01
  • 5. Servo Ventilator 300/300A Contents Contents 1. Introduction ............................................................................ 7 1 2. Description of functions....................................................... 21 2 3. Disassembling and assembling ......................................... 81 3 4. Service procedures............................................................... 103 4 5. Troubleshooting ..................................................................... 109 5 6. Product change history ........................................................ 113 6 7. Index ........................................................................................ 115 7 8. Diagrams ................................................................................. 121 8 E382 E380E 061 01 03 01 Siemens-Elema AB 5
  • 6. Notes Servo Ventilator 300/300A Notes 6 Siemens-Elema AB E382 E380E 061 01 03 01
  • 7. Servo Ventilator 300/300A Introduction 1. Introduction Main units ................................................ 8 Basic principles ........................................ 12 1 E382 E380E 061 01 03 01 Siemens-Elema AB 7
  • 8. Introduction Servo Ventilator 300/300A Main units 1 The Servo Ventilator 300 can be divided into two main units: 1 2 3 • Control unit (1). The control unit consists of the panel section (2) and the control section (3). • Patient unit (4). The patient unit consists of the pneumatic section (5) and the power section (6). The patient unit and the control unit are connected to each other with the inter- connection cable (7). 7 5 6 300-A00X 4 8 Siemens-Elema AB E382 E380E 061 01 03 01
  • 9. Servo Ventilator 300/300A Introduction Control unit The control unit consists of the control section which is a housing with a number of PC boards. The panel section, attached to the control section, is the other part of the 1 control unit. With the panel section removed from the control section, the following parts are accessible: • Front panel controls • PC 1614 PANEL INTERFACE (1) with PC 1588 MICROPROCESSOR MODULE and PAN-PROM • PC 1745 AUTOMODE PANEL INTERFACE (2). Only for SV 300A. • PC 1617 CONTROL INTERCONNECTION (3). 300-A01X 2 1 300-A02X 3 E382 E380E 061 01 03 01 Siemens-Elema AB 9
  • 10. Introduction Servo Ventilator 300/300A All PC boards in the control section (except PC 1617 mentioned above) are accessible when the control section cover is removed. 1 These PC boards are: • PC 1605 REFERENCE & TIMING (1) with PC 1588 MICROPROCESSOR MODULE and REF-PROM • PC 1616 INSPIRATORY CONTROL (2) with 1 2 PC 1588 MICROPROCESSOR MODULE and MIX- PROM • PC board (3); – PC 1665 COMPUTER INTERFACE DUMMY or – Optional PC 1587 COMPUTER INTERFACE (3) with COM-PROM • PC 1608 MONITORING (4) with PC 1588 3 4 MICROPROCESSOR MODULE and MON-PROM. 300-A03X Patient unit 1 2 The upper part of the patient unit is the pneumatic section. Under the lid of the pneumatic section you will find the: • Inspiratory channel (1) including O2 cell • Expiratory channel (2) including flow transducer. The inspiratory valves (gas modules) are located in three different slots in the pneumatic section: • Inspiratory valve unit – AIR (3) • Inspiratory valve unit – O2 (4) • Inspiratory valve unit – Optional (5). The valve dummy in the third slot can be replaced by an inspiratory valve for administration of e. g. NO gas or by 300-A04X equipment for nebulizing drugs (Servo 3 4 5 Ultra Nebulizer 345). These products are described in separate documents. 10 Siemens-Elema AB E382 E380E 061 01 03 01
  • 11. Servo Ventilator 300/300A Introduction 1 2 3 4 When the pneumatic section casing and the cover plate is removed, the following parts are accessible: • PC 1615 EXPIRATORY FLOW LINEARIZATION (1) with PC 1588 MICROPROCESSOR MODULE and 1 LIN-PROM • Housing (2) containing PC 1585 CURRENT CONTROL and PC 1586 CURRENT POWER • Expiratory valve (3) • PC 1622 VALVE CONTROL (4) • PC 1611 EXPIRATORY PRESSURE AMPLIFIER (5) • Safety valve (6) with PC 1613 SAFETY VALVE DRIVER 300-A05X 5 6 7 8 • PC 1611 INSPIRATORY PRESSURE AMPLIFIER (7) • PC 1607 PNEUMATIC INTERCONNECTION (8). 1 2 3 4 The lower part of the patient unit is the power section. With the pneumatic section separated from the power section the following parts are accessible: • Mains power inlet (1) • External battery inlet (2) • Operating time meter (3) • Batteries (4) • Transformer (5) 300-A06X • Capacitor (6) 5 6 7 8 • Auxiliary equipment outputs (7) • PC 1618 POWER SUPPLY (8). E382 E380E 061 01 03 01 Siemens-Elema AB 11
  • 12. Introduction Servo Ventilator 300/300A Basic principles 1 9 11 8 12 10 2 7 1 6 300-F71X 5 4 3 12 Siemens-Elema AB E382 E380E 061 01 03 01
  • 13. Servo Ventilator 300/300A Introduction Gas flow through the patient unit Inspiratory flow Expiratory flow 1. Gas inlet for air. The connected air must 8. The patient system's expiratory gas tube have a pressure between 200 and 650 is connected at the expiratory inlet. The 1 kPa (2 and 6.5 bar). expiratory inlet also contains a moisture trap. 2. Gas inlet for O2. The connected O2 must have a pressure between 200 and 650 9. The gas flow through the expiratory kPa (2 and 6.5 bar). channel is measured by the expiratory flow transducer. Patient trigger efforts, 3. The flow of the gas delivered to the indicated by a decreased continuous patient system is regulated by the flow, are sensed by this expiratory flow inspiratory valves. There is one inspira- transducer. tory valve unit (gas module) for each gas. The inspiratory valves are regulated 10. The expiratory pressure is measured by by a feedback control system that is the expiratory pressure transducer. The described in this chapter. transducer is protected by a bacteria filter. Patient trigger efforts, indicated by 4. The gases are mixed in the inspiratory a pressure drop, are sensed by this mixing part. expiratory pressure transducer. 5. The pressure of the mixed gas delivered 11. The pressure of the gas (PEEP pressure) to the patient (inspiratory pressure) is in the patient system is regulated by the measured by the inspiratory pressure expiratory valve. The expiratory valve is transducer. The transducer is protected regulated by a feedback control system by a bacteria filter. that is described in this chapter. 6. The inspiratory pipe leads the mixed gas 12. The gas from the patient system leaves from the inspiratory mixing part to the the ventilator via this expiratory outlet. patient system. The inspiratory pipe also The outlet contains a non-return valve contains the safety valve, a holder for the which is a part of the patient triggering O2 cell and the inspiratory outlet. The system. springloaded safety valve will open in case of a power failure and/or if the inspi- ratory pressure exceeds 120 cm H2O. It will also open if the set upper pressure limit is exceeded by 5 cm H2O. 7. The oxygen concentration inside the inspiratory pipe is measured by the O2 cell. The O2 cell is protected by a bacteria filter. E382 E380E 061 01 03 01 Siemens-Elema AB 13
  • 14. Introduction Servo Ventilator 300/300A Feedback control system – General TRANSDUCER CONTROLLED SYSTEM CONTROLLING 1 SIGNAL AMPLIFIER COMPARING ELEMENT (WITH SERVO) ELEMENT CONTROLLING SYSTEM ACTUAL VALUE P I CONVERTER ERROR ACTUATING SIGNAL SIGNAL "COMPENSATOR" D REFERENCE DESIRED "ERROR VARIABLE VALUE CALCULATOR" 300-A08E PID-CONTROLLER Any CONTROLLING SYSTEM may comprise one P-action gives a continuous basic positioning. single CONTROLLER or, for instance, a cascade I-action gives a slowly varying positioning of more than one CONTROLLER. Compensating compensation for small long time deviation. elements may also be added. D-action gives a speed-up positioning CONTROLLER symbol used in reaction at (fast) changes. this chapter. The signals from these parts (from the parts used in the actual combination) are added. In each CONTROLLER, the actual value normally The sum is the total ERROR SIGNAL which may from a TRANSDUCER in the CONTROLLED SYSTEM is be converted before use as an ACTUATING compared to the desired value = REFERENCE SIGNAL. VARIABLE. The difference between these two input values results at the output of the Each CONTROLLING ELEMENT (in the ventilator: controlling system in an ACTUATING SIGNAL. inspiratory valves and expiratory valve) is positioned by the ACTUATING SIGNAL from its Each CONTROLLER may comprise some controlling system. Feedback control, which combination of the following parts: is synonymous with regulation, results in the • Proportional action, P controlled valve being moved into such a position that the ACTUAL VALUE is kept equal to • Integral action, I the DESIRED VALUE independently of • Derivative action, D. disturbance variables like changes in the airway resistance or lung compliance. 14 Siemens-Elema AB E382 E380E 061 01 03 01
  • 15. Servo Ventilator 300/300A Introduction Inspiratory feedback control system REFERENCE TRANSDUCERS CONTROLLED SYSTEM INSPIRATORY 1 & TIMING VALVE PRESSURE FLOW POSITION VALVE SENSOR SOLENOID GAS GAS SUPPLY FLOW PRESS. ACTUAL ACTUAL INSP. VALVE POSITION FLOW SUPPLY GAS FLOW REF. PRESS. VALUE ACTUAL POSITION VALUE VALUE ACTUAL CURRENT VALUE GAS CURRENT FLOW FLOW PI POSITION PD CONTROL & MIXER REF. CONTROL CONTROL CURRENT DESIRED DESIRED POWER DESIRED POSITION CURRENT GAS FLOW VALUE VALUE INSPIRATORY VALUE CONTROL VALVE POSITION CALCULATOR 300-A09E Flow FLOW CONTROL regulation The inspiratory feedback control system • ACTUAL GAS FLOW VALUE is generated by uses cascade control with a number of signal converters (FLOW) using signals from controllers and compensating elements. measuring TRANSDUCERS. • The actuating signal from the FLOW PI CONTROL results in the DESIRED POSITION VALUE Flow regulation (reference variable) for the POSITION PD Inspiratory flow regulation occurs during CONTROL. The POSITION SENSOR at the inspiration time in the modes Volume INSPIRATORY VALVE generates a valve position Control and SIMV (Vol. Contr.) and during signal which is used as the ACTUAL POSITION expiration time in all modes. VALUE in the POSITION PD CONTROL. When flow regulation is used, the • The actuating signal from the POSITION PD shadowed main forward path comprises the CONTROL results in the DESIRED CURRENT VALUE following: (reference variable) for the CURRENT CONTROL • The microprocessor module in the block & CURRENT POWER. MODE CPU in the main block REFERENCE & • The ACTUAL CURRENT VALUE is an internal TIMING generates a FLOW REF. signal. feedback signal used in the CURRENT • The block MIXER, in the main block CONTROL block. INSPIRATORY CONTROL, splits the FLOW REF. • The actuating signal from the output stage signal into one GAS FLOW REF. signal for each of the CURRENT CONTROL & CURRENT POWER is a inspiratory valve (AIR FLOW REF. and O2 FLOW voltage pulse train with pulse width REF.). Each GAS FLOW REF. signal is used modulation used to actuate the INSPIRATORY (after a minor correction within ±5 %) as VALVE SOLENOID which generates a certain DESIRED GAS FLOW VALUE (reference variable). opening of the INSPIRATORY VALVE. E382 E380E 061 01 03 01 Siemens-Elema AB 15
  • 16. Introduction Servo Ventilator 300/300A In addition to the main forward path, compensating elements are added as described in both inspiratory feedback 1 control system pictures. The compensating elements in this system works as follows: • Using the input information of ACTUAL SUPPLY PRESSURE VALUE and DESIRED GAS FLOW VALUE the VALVE POSITION CALCULATOR directly generates a rough approximate POSITION VALUE which helps to obtain the desired position within minimum time. • The input information of ACTUAL SUPPLY PRESSURE VALUE is used (inverted and adequatly scaled) to create one small part of the DESIRED CURRENT VALUE. This small part of the DESIRED CURRENT VALUE will – in combination with the opening force of the INSPIRATORY VALVE implied by the pressure of the supplied gas – help to balance the closing force on the INSPIRATORY VALVE implied by the mechanical spring inside the VALVE SOLENOID even in closed position of the INSPIRATORY VALVE. 16 Siemens-Elema AB E382 E380E 061 01 03 01
  • 17. Servo Ventilator 300/300A Introduction Inspiratory feedback control system (cont'd) REFERENCE & TIMING PRESSURE REF. DESIRED PRESS. VALUE TRANSDUCERS CONTROLLED SYSTEM INSPIRATORY VALVE 1 INSP. PRESSURE POSITION VALVE PRESSURE FLOW SENSOR SOLENOID ACTUAL PRESS. VALUE PRESS. GAS GAS PID SUPPLY FLOW CONTR. PRESS. ACTUAL ACTUAL FLOW SUPPLY GAS FLOW INSP. VALVE POSITION REF. PRESS. VALUE ACTUAL POSITION VALUE VALUE ACTUAL CURRENT VALUE GAS CURRENT FLOW FLOW PI POSITION PD CONTROL & MIXER REF. CONTROL CONTROL CURRENT DESIRED DESIRED POWER POSITION CURRENT DESIRED VALUE VALUE INSPIRATORY GAS FLOW CONTROL VALUE VALVE POSITION CALCULATOR 300-A10E Pressure FLOW CONTROL regulation Pressure regulation Inspiratory pressure regulation is used only • ACTUAL VALUE is the INSP. PRESSURE signal during inspiration time in all modes except from the INSPIRATORY PRESSURE TRANSDUCER when flow regulation is used as mentioned • The actuating signal from the PRESSURE PID previously. CONTROL block is used as a FLOW REFERENCE When PRESSURE regulation is used, the signal going to the MIXER. shadowed main forward path includes the The rest of the controlling system, including PRESSURE PID CONTROL block in the REFERENCE & compensating elements, is the same as in TIMING main block: flow regulation. • DESIRED VALUE is the PRESSURE REFERENCE signal from (the microprocessor module MODE CPU within) the main block REFERENCE & TIMING E382 E380E 061 01 03 01 Siemens-Elema AB 17
  • 18. Introduction Servo Ventilator 300/300A Notes 1 20 Siemens-Elema AB E382 E380E 061 01 03 01
  • 19. Servo Ventilator 300/300A Introduction Expiratory feedback control system TRANSDUCER PRESSURE CONTROLLED SYSTEM EXPIRATORY VALVE POSITION VALVE 1 SENSOR SOLENOID EXP. PRESSURE EXP. VALVE POSITION ACTUAL POSITION VALUE ACTUAL ACTUAL CURRENT VALUE EXPIRATORY PRESSURE VALUE CURRENT CONTROL PEEP PID POSITION POSITION PD EXP. VALVE & PEEP LEVEL CONTROL REF. CONTROL CURRENT REF. CURRENT DESIRED EXPIRATORY DESIRED DESIRED POWER PRESSURE VALUE POSITION CURRENT 300-A11E VALUE VALUE PRESSURE CONTROL The expiratory feedback control system • The actuating signal from the POSITION PD uses cascade control. It is active only during CONTROL block is the EXP. VALVE CURRENT REF. expiration time in all modes. It applies signal used as DESIRED VALUE for the CURRENT PRESSURE regulation. CONTROL & CURRENT POWER block. • DESIRED VALUE is the PEEP LEVEL signal from • The ACTUAL CURRENT VALUE is an internal the PANEL SECTION (according to front panel feedback signal used in the CURRENT setting) CONTROL block. • ACTUAL VALUE is the EXP. PRESSURE signal • The actuating signal from the output stage from the expiratory PRESSURE TRANSDUCER. of the CURRENT CONTROL & CURRENT POWER is a pulse train with pulse width modulation • The actuating signal from the PEEP PID used to actuate the EXPIRATORY VALVE CONTROL block is the POSITION REF. signal SOLENOID which generates a certain used as DESIRED VALUE for the POSITION PD opening of the EXPIRATORY VALVE. CONTROL. • ACTUAL VALUE is the EXP. VALVE POSITION signal from the POSITION SENSOR in the EXPIRATORY VALVE. E382 E380E 061 01 03 01 Siemens-Elema AB 17 19
  • 20. Introduction Servo Ventilator 300/300A Principle diagram EXTERNAL EXTERNAL CONTROL SIGNALS OUTPUT SIGNALS CONTROL UNIT PATIENT UNIT FROM PATIENT INPUT FROM C SV 300 INTERNAL MICROPROCESSORS CONTROL COMPUTER INTERFACE SIGNALS INPUT SIGNALS 11 M C M L G INPUT SIGNALS EXP. MONITORING P FLOW PRESSURE AMPLIFIER T EXP. PRESSURE 10 M PRESSURE PEEP LEVEL CONTROL REFERENCE & TIMING T INSPIRATORY CONTROL I O2 CONC. O2 CONC. 9 INSP. PRESSURE AMPLIFIER M PRESSURE EXPIRATORY PID CONTROL FLOW REF. VALVE CONTROL V VALVE X O2 FLOW REF. PRESSURE REF. MIXER AIR FLOW REF. SELECTION OF Pressure PRESSURE OR FLOW REGULATION MODE CPU Flow FLOW REF. O2 TRANS- DUCERS 1 2 4 5 O2 M CELL FLOW CONTROL O2 FLOW REF. P TO PATIENT 7 INSPIRATORY VALVE UNIT - O2 O AIR TRANS- DUCERS M FLOW CONTROL AIR FLOW REF. 3 6 H INSPIRATORY VALVE UNIT - AIR A PRESSURE 8 AMPLIFIER INSP. PRESSURE M +5V W EXT. BATTERY ±15V POWER SUPPLY INTERNAL 300-A12E BATTERY +24V 12 MAINS POWER 18 Siemens-Elema AB E382 E380E 061 01 03 02
  • 21. Servo Ventilator 300/300A Introduction Inspiratory regulation Monitoring 1. Choice between pressure regulation and 10. The MONITORING board monitors and flow regulation during inspiration is generates certain displayed parameters. made by the MODE CPU dependent on the Mode selection. It also activates alarm functions. 1 2. Main parameters used for the flow level Computer interface during inspiration when flow regulation is used. 11. The optional COMPUTER INTERFACE board is used as input/output signal interface and 3. Main parameters used for the pressure includes functions for computer level during inspiration when pressure communication (RS-232). The PC board regulation is used. is equipped with the following 4. Inspiration rise time setting determines connectors: the initial inspiration pattern. A succes- – Master/Slave connection (N80) sively increased flow/pressure level may be more comfortable for the patient. – Analog I/O terminal (N81) 5. The O2 concentration delivered to the – Serial communication ports (N82/N83) patient depends on the setting of these – Analog input & Digital code port (N84). O2 concentration control knobs. 6. INSPIRATORY VALVE UNIT– AIR. The control system is described in Power supply section "Inspiratory feedback control 12. The POWER SUPPLY board controls and system" in this chapter. distributes power to the different parts of the ventilator. The power source can 7. INSPIRATORY VALVE UNIT– O2. be: The control system is described in section "Inspiratory feedback control – Mains power system" in this chapter. – External battery – Internal battery. Expiratory regulation 8. PEEP level setting is used as DESIRED EXPIRATORY PRESSURE VALUE. See section "Expiratory feedback control system" in this chapter. 9. EXPIRATORY VALVE interacts with the VALVE CONTROL board as described in section "Expiratory feedback control system" in this chapter. E382 E380E 061 01 03 01 Siemens-Elema AB 19 21
  • 22. Introduction Servo Ventilator 300/300A Notes 1 20 Siemens-Elema AB E382 E380E 061 01 03 01
  • 23. Servo Ventilator 300/300A Description of functions 2. Description of functions General .................................................... 22 Microprocessor module PC 1588 ............ 23 Version labels .......................................... 24 Control unit – Panel section ..................... 25 1 Front panel ....................................... 25 1A Automode front panel ...................... 27 2 Panel interface .................................. 28 2 2A Automode panel interface ................ 29 Control unit – Control section .................. 30 3 Computer interface .......................... 30 3A Computer interface dummy ............. 33 4 Monitoring ........................................ 34 5 Optional PC board slot ...................... 48 6 Reference & Timing ......................... 49 7 Inspiratory control ............................. 61 Patient unit – Pneumatic section ............. 62 8 Inspiratory valve unit – Air ................ 62 9 Inspiratory valve unit – O2 . ............... 67 10 Inspiratory valve unit – Optional ....... 67 11 Inspiratory channel ........................... 68 12 Inspiratory pressure .......................... 70 13 Expiratory pressure .......................... 71 14 Expiratory channel ............................ 71 15 Valve control ..................................... 73 16 Expiratory valve ................................ 74 17 Optional interface ............................. 74 18 Expiratory flow linearization .............. 75 19 Cooling system ................................ 76 Patient unit – Power section .................... 76 20 Mains power .................................... 76 21 Power supply ................................... 77 22 Operating power .............................. 79 23 Interconnection cable ....................... 80 E382 E380E 061 01 03 01 Siemens-Elema AB 21
  • 24. Description of functions Servo Ventilator 300/300A General As described in chapter Introduction, the In this Description of functions, the words Servo Ventilator 300 can be divided into two inspiratory and expiratory are used in the Main units; Control unit and Patient unit. The first place to indicate the site of a part (e g two main units are connected to each other INSPIRATORY CHANNEL). with the Interconnection cable. The words inspiration and expiration are For the purpose to describe all different used in the first place to indicate the functions in the Servo Ventilator 300, the corresponding time interval during 2 two main units are further divided as follows: ventilation (e g INSPIRATION TIME) Main unit Section Main block Functional block Example: Control unit Panel section 1 Front panel 1.1 Patient range selection 1.2 Airway pressure etc. 2 Panel interface 2.1 Panel CPU including PAN-PROM functions 2.2 Parameter buffers etc. Functional descriptions of the Main blocks and Functional blocks can be found in this chapter. This functional description is based on and refers to the Functional block diagram that can be found as a fold out on the rear cover. 22 Siemens-Elema AB E382 E380E 061 01 03 01
  • 25. Servo Ventilator 300/300A Description of functions Microprocessor module PC 1588 includes a green/red WATCHDOG LED PC 1588 D1 (3). The LED indications are as follows: A MICROPROCESSOR MODULE, PC1588 (1), is – Green LED: Microprocessor program is included on each one of the following five running. PC boards: – Red LED: Program error indication. • PC 1605 – REFERENCE & TIMING – LED not lit: Microprocessor program not • PC 1608 – MONITORING running. • PC 1614 – PANEL INTERFACE PC 1588 also includes an exchangable program memory, PROM (2). The function 2 • PC 1615 – EXP. FLOW LINEARIZATION of PC 1588 depends on this PROM which is • PC 1616 – INSPIRATORY CONTROL. different on the five different enumerated boards. The PROM functions are mentioned PC 1588 is mounted with its two rows of in the description of the following blocks: connectors (P65 and P66) into the corresponding connectors (HYB1) on the • 2.1 Panel CPU (PAN-PROM) above mentioned PC boards. • 4.1 Monitor/Alarm CPU (MON-PROM) PC 1588 is a programmable digital control • 6.1 Mode CPU (REF-PROM) block including: • 7.1 Mixer CPU (MIX-PROM) • Microprocessor • 18.1 Linearizing CPU (LIN-PROM) • Analog – digital conversion Note – The MICROPROCESSOR -block on • Digital – analog conversion PC 1587 – COMPUTER INTERFACE, includes two • Digital I/O. PROM modules (COM-PROM). PC 1588 is not used on PC 1587. xxx xxx 1 300-A13X 3 2 E382 E380E 061 01 03 01 Siemens-Elema AB 23
  • 26. Description of functions Servo Ventilator 300/300A Version labels A label (1) showing all PROM-versions included in the ventilator is attached to the control section. When replacing a PROM, attach the new label (2) that is delivered with the new PROM onto the old label on the control section. Changing PROM-versions in the ventilator is 2 used to add or change functions in the ventilator. For that reason, it may also be necessary to replace the Operating Manual 1 in connection with PROM replacements. The Operating Manual, valid for the combination of PROMs used in the 3 ventilator, is identified with a version number stated both on the manual and on 2 the label (3). When replacing the Operating 0 30 SV Manual with a new version of the manual, 8.x E 13 E3 X XX SI al XX R anu XX VE ting M attach the new label (3) that is delivered with N O ra e Op 16 V6. E PC OM 380 08 0X PR X E N- X MO XX X the new Operating Manual onto the old label XX on the control section. 300-A14X 24 Siemens-Elema AB E382 E380E 061 01 03 01
  • 27. Servo Ventilator 300/300A Description of functions Control unit – Panel section 1.1 Patient range selection The Panel section consists of the main For selection of one of three available blocks 1 – 2 which are described below. patient ranges: • 1 three-position selector. This switch generates selection information from two 1 Front panel switch decks delivering one two bit code This main block contains the front panel and corresponding two bit complementary film, in different language versions, and the code. various displays and controls with connection cables and their connectors. • 3 yellow indicator LEDs. 2 The following table indicates how Patient The optional front panel function for SV 300A range selection affects some other is described in section 1A AUTOMODE that can parameters: be found after 1.8 MODE SELECTION. Parameter Adult Pediatric Neonate Unit Apnea alarm – delay time 20 15 10 s ”Bias flow” = Constant flow during 32 16 8 ml/s expiration time 2 1 0.5 l/min 3000 500 200 ml/s Max. inspiratory flow 180 30 12 l/min Max. (preset or measured) inspiratory or 3999 399 39 ml expiratory tidal volume FLOW REF overrange limit 3.3 0.55 0.22 l/s Expiratory minute volume alarm: Setting range for Upper alarm limit 0 - 60 0 - 60 0-6 l/min Setting range for Lower alarm limit * - 40 * - 40 * - 40 l/min * Lowest value for normal (digital) alarm 0.3 0.3 0.06 l/min * Lowest value for backup alarm 0.2 0.2 0.05 l/min For Automode: Time duration since last patient trigger when Control/Support 12 8 5 s mode switches from Support mode to Control mode. E382 E380E 061 01 03 01 Siemens-Elema AB 25
  • 28. Description of functions Servo Ventilator 300/300A 1.2 Airway pressure 1.3 Respiratory pattern For control and monitoring of AIRWAY For control and/or monitoring of respiratory PRESSURE parameters: pattern parameters: • 4 four-character LED displays. • 4 four-character LED displays • 1 double column LED bargraph. • 5 potentiometer knobs: • 5 potentiometer knobs: – CMV freq. b/min. This control is equipped with two parallell – Upper press. limit. Push button release 2 function for one-way mechanical stops at approx. 60, 80 and 100 cm H2O. potentiometers – one for the control function and the other for the monitor/ alarm function. – Pressure Control Level above PEEP. – Insp. time %. Push button release Push button release function for one- function for two-way mechanical stops way mechanical stops at approx. 30, 60 at approx. 20, 50 and 70%. and 80 cm H2O. – Pause time %. Push button release – Pressure Support Level above PEEP. function for one-way mechanical stop at Push button release function for one- approx. 20%. way mechanical stops at approx. 30, 60 and 80 cm H2O. – Insp. rise time %. Push button release function for one-way mechanical stop at – PEEP. Push button release function for approx. 1%. one-way mechanical stop at approx. 20 cm H2O. – SIMV freq. b/min. – Trigg. sensitivity Level below PEEP. • 5 yellow indicator LEDs. • 5 yellow indicator LEDs. 1.4 Volume For control and/or monitoring of minute volume and tidal volume parameters: • 2 four-character green LED displays. Showing information generated via 2.1 PANEL CPU. • 3 four-character red LED displays. Showing information generated via 4.1 MONITOR/ALARM CPU. • 2 fields of double column LED bargraph. • 3 potentiometer knobs: – Volume – Upper alarm limit – Lower alarm limit. • 3 yellow indicator LEDs. 26 Siemens-Elema AB E382 E380E 061 01 03 01
  • 29. Servo Ventilator 300/300A Description of functions 1.5 O2 concentration 1.8 Mode selection For control and monitoring of O2 For the selection of Ventilation Modes and concentration parameters: control of the Set Parameter Guide functions: • 1 four-character LED display. Showing information generated via 2.1 PANEL CPU. • 1 green Mains indicator LED • 1 potentiometer knob with push button • 1 ten-position selector always supplied release function for two-way mechanical with high impedance 24 V. This switch stop at approx. 60 %. This control is generates selection information from two equipped with two parallell potentiometers – one for the regulation function and the switch decks delivering complementary code via PC board (PC 1632). The mode 2 other for the monitor/alarm function. Optional can not be selected. • 1 non-locking, two-way selector for • 10 different fields each comprising: Oxygen breaths and for Start breath. – 1 yellow indicator LED • 2 yellow indicator LEDs. – 1 touch sensor. 1.6 Alarms and messages 1A Automode front panel For the display of alarms and messages: For the control and monitoring of the • 1 eight-character LED display. Showing Automode functions: information generated via 4.1 MONITOR/ ALARM CPU. • 1 two-position selector for turning the Automode On and Off • 8 different alarm fields each comprising: • 2 yellow indicator LEDs. – 1 red indicator LED – 1 yellow indicator LED – 1 touch sensor • 1 non-locking, two-way selector for Reset and for 2 min off. This information is handled by the main block 4 MONITORING. 1.7 Pause hold For the control, via main block 6 REFERENCE & TIMING, of pause hold functions: • 1 non-locking two-way selector for Insp. pause hold and for Exp. pause hold. E382 E380E 061 01 03 01 Siemens-Elema AB 27
  • 30. Description of functions Servo Ventilator 300/300A 2 Panel interface 2.3 Inputs This main block consists of the board Analog multiplexor for input signals to the PC 1614 PANEL INTERFACE including its block 2.1 PANEL CPU. PC1588 MICROPROCESSOR MODULE . The board is connected (via N18, N19 and N40) to the mother board (PC 1617) CONTROL INTERCONNEC- 2.4 LED matrix TION and it has connectors (P81 – P99) for Time multiplexed addressing and driving the connection of FRONT PANEL controls. stages for the block 2.5 LED ARRAY. The input 2 The FRONT PANEL touch sensors as well as displays and indicator LEDs are integrated data is given from the block 2.1 PANEL CPU. parts of this board. Refer to 1 FRONT PANEL in this chapter. 2.5 LED array A connector, N20 (on some older PC-boards, This block contains the light emitting diodes the connector is marked N80), is provided of the front panel bargraphs and the yellow on the board for additional function. This light emitting diodes at the front panel connector can be used to connect the settings. optional PC 1745 AUTOMODE PANEL INTERFACE Some flashing function in front panel used in SV 300A. This function is described bargraphs are mentioned in 4.1 MONITOR/ in section 2A AUTOMODE PANEL INTERFACE that ALARM CPU can be found after 2.9 MONITOR DISPLAY. . 2.6 Panel display 2.1 Panel CPU including PAN- PROM functions This block contains all the green alpha- numerical four character light emitting diode This block includes the board PC 1588 with displays on the front panel. The input data is a PAN-PROM. Some details of the PC 1588 are given from the block 2.1 PANEL CPU. mentioned in section Microprocessor module PC 1588 in this chapter. The PAN-PROM function includes the handling 2.7 Mode select of input signals from the main block 1 FRONT Input stage for the signals from the front PANEL, calculation of some displayed values panel Mode selector. The input is available and handling of signals to the displays on as four bit code and corresponding four bit the main block 1 FRONT PANEL. complementary code generated via the small interconnecting circuit board PC 1632 on the Mode selector. This block gives the 2.2 Parameter buffers data to the block 2.1 PANEL CPU. Input stage for potentiometers on the main block 1 FRONT PANEL. 28 Siemens-Elema AB E382 E380E 061 01 03 01
  • 31. Servo Ventilator 300/300A Description of functions 2.8 Opto switch 2A Automode panel Time multiplexed driving stage and sensor interface receiving stage for all the infra-red sensor The board PC 1745 AUTOMODE PANEL INTERFACE touch pads on the front panel. The function contains electronics for handling of the is controlled by the block 2.1 PANEL CPU Automode switch. including an automatic adaption of the touch pad sensitivity to the ambient light level. The status of the switch is given to the This block gives the data to the block 2.1 blocks 2.1 PANEL CPU, 5 MONITORING and PANEL CPU. 6 REFERENCE & TIMING. The LEDs are driven from the main block 4 MONITORING. 2 2.9 Monitor display This block contains all the red alpha- numerical four character light emitting diode displays as well as eight yellow and eight red alarm indicating light emitting diodes below the Alarms and messages display on the front panel. This block is driven directly from the main block 4 MONITORING. E382 E380E 061 01 03 01 Siemens-Elema AB 29
  • 32. Description of functions Servo Ventilator 300/300A Control unit – Control section There are five connectors (N80 – N84) on PC 1587. The Control section consists of the main blocks 3 – 7 which are described below. A ”mother” board, PC 1617 CONTROL INTERCONNECTION,provides signal connections via connectors (but no electronic functions). 2 3 Computer interface The ventilator can be equipped with the board PC 1587 COMPUTER INTERFACE to provide the interface functions described below. N80 If these interface functions are not needed, the board PC 1665 COMPUTER INTERFACE DUMMY N82 N81 can be mounted instead of PC 1587. The PC 1665-function is described in section N83 3A COMPUTER INTERFACE DUMMY that can be found after section 3.9 LED INDICATORS. N84 The board PC 1587 COMPUTER INTERFACE was replaced during the mandatory upgrades: – SV 300, Upgrade 1995-09 (US market) – SV 300, Upgrade 1995-10 (Rest of World) 300-A15X As the board PC 1587D was included in these mandatory upgrades, all SV 300/300A Only accessories, supplies or auxiliary must be equipped with PC 1587D (or higher). equipment listed in Siemens-Elema catalogs The main block COMPUTER INTERFACE consists (”Products and accessories” Order No. of the board PC 1587 COMPUTER INTERFACE 90 34 562 E323E and ”Spare and exchange which includes one microprocessor and a parts” Order No. 90 34 570 E323E) must be COM-PROM. The COM-PROM is divided in two connected to or used in conjunction with the PROM modules ventilator. Warning: Use of accessories and auxiliary equipment other than those specified in these catalogs may degrade safety and performance of the ventilator. 30 Siemens-Elema AB E382 E380E 061 01 03 01
  • 33. Servo Ventilator 300/300A Description of functions 3.1 Microprocessor CPU The voltage level in the battery circuit is including COM-PROM functions monitored resulting in an error indication on LED D3 (see 3.9 LED indicators) if the This includes functions for computer voltage level is below 3.6 V or above 4.5 V. communication (RS-232) via 3.7 SERIAL An information message is also displayed on COMMUNICATION PORTS as described in the Servo Screen 390 if such unit is connected Reference Manual "Servo Ventilator 300/ to the SV 300. 300A, Computer Interface (Firmware version Note – The voltage level measured in the 2.X)", Order No. 63 14 061. battery circuit is higher than the voltage level This block also contains a real time clock with battery backup, driven by a quartz in the battery itself. 2 crystal of 32768 Hz. The purpose of this clock is to time stamp the trend values 3.5 Master/Slave connection stored in the ventilator. There is also a 15 pole D-sub connector (N80). Can be used possibility to read the actual time through for the synchronization of two Servo the RS-232 communication. Ventilator 300. Only Siemens connection cable must be used. 3.2 I/O buffers Pin configuration and signal names can be found in chapter "8. Diagrams". Analog and digital input and output stages for the main block 3 COMPUTER INTERFACE. 3.6 Analog I/O terminal 3.3 A/D, D/A 62 pole D-sub connector (N81). Can be used for connection of monitoring/recording Analog to digital and digital to analog equipment. conversion in connection with the block 3.1 MICROPROCESSOR CPU. Input control signals can be connected at N81 to control some ventilator functions (e.g. when using Bi-Phasic Ventilation). 3.4 Battery These input signals are routed via two On board mounted Lithium battery (3.5 V). sockets, J2 and J3, on PC 1587. The battery backs up the trend data memory To enable the control functions, jumpers and the real time clock in the block must be mounted in the sockets. 3.1 MICROPROCESSOR CPU when the SV 300 is Pin configuration, signal names and jumpers switched off. can be found in chapter "8. Diagrams". Normal service interval for exchange of the battery is approx. 5 years. Replacement of the battery is described in chapter ”4. Service procedures”. E382 E380E 061 01 03 01 Siemens-Elema AB 31
  • 34. Description of functions Servo Ventilator 300/300A 3.7 Serial communication ports 3.9 LED indicators 26 pole D-sub connectors (N82 and N83) for There are 9 green/red LED INDICATORS RS-232-C data communication. For com- mounted on PC 1587. The LEDs are visible munication protocol, see Reference Manual if the control section cover is removed. "Servo Ventilator 300/300A, Computer Interface (Firmware version 2.X)", Order No. 63 14 061 E380E. D9 D1 The connectors are identical with one 2 exception; the internal clock can only be set via N82. Serial communication is indicated on the LEDs D6 – D9. See 3.9 LED INDICATORS. Pin configuration and signal names can be found in chapter "8. Diagrams". 3.8 Analog input & Digital code port 44 pole D-sub connector (N84). Optional input interface. Pin configuration and signal names can be found in chapter "8. Diagrams". 300-L24X LED functions D1: WATCHDOG. During startup the LED will flash for a short moment and then be lit red for a short moment. It will then be lit green again. When the startup test is completed, the LED indications are as follows: – Green LED: Microprocessor program is running. – Red LED: Program error indication. – LED not lit: Microprocessor program not running. 32 Siemens-Elema AB E382 E380E 061 01 03 01
  • 35. Servo Ventilator 300/300A Description of functions D2 – D5: TEST STATUS LED. Error indications 3A Computer interface during System startup test. The startup test dummy is divided in several sequences. The PC 1665 COMPUTER INTERFACE DUMMY is The first sequence is a LED test that starts mounted instead of PC 1587 COMPUTER with all LEDs red. The LEDs will then INTERFACE, in the case that there is no need change to green, one at the time (starting for serial communication with the ventilator. with D2), until all four LEDs are green. It is needed as a holder for the PC 1608, as In the hardware test sequence, D3 can be of well as for disabling the error signal from the special interest for troubleshooting. In the first phase of the hardware test, D3 is COMPUTER INTERFACE board and the signals for Master – Slave. 2 lit red for a few seconds indicating COM-PROM error. In the second phase of the hardware test, D3 is lit red for a few seconds indicating BATTERY error. The hardware test sequence ends showing all LEDs green if no error was detected and all LEDs red if any error was detected. Note – As COM-PROM and BATTERY are the only spare parts for PC 1587, any other LED error indications than mentioned above results in a PC 1587 replacement. D6 – D7: SERIAL INTERFACE 1. Rapidly flashing green and red indicates communication on connector N82. D6 indicates input and D7 indicates output. D8 – D9: SERIAL INTERFACE 1. Rapidly flashing green and red indicates communication on connector N83. D8 indicates input and D9 indicates output. E382 E380E 061 01 03 01 Siemens-Elema AB 33
  • 36. Description of functions Servo Ventilator 300/300A 4 Monitoring 4.1 Monitor/Alarm CPU including MON-PROM functions This main block consists of the board PC 1608 MONITORING including its PC 1588 This block includes the board PC 1588 with MICROPROCESSOR MODULE and the a MON-PROM. Some details of the PC 1588 4.7 LOUDSPEAKER. are mentioned in section Microprocessor module PC 1588 in this chapter. The MON- New versions of PC 1608 MONITORING and PROM functions include the (program control) MON-PROM was introduced in production digital monitoring of certain parameters and during 1996 and also as an optional upgrade the activation of alarm functions: 2 on delivered units. The upgrade was called ”SV 300, ALARM SYSTEM ENHANCEMENT KIT 96-05”. • Front panel indications via the block The new versions was: 2.9 MONITOR DISPLAY – PC 1608E • Audible indications via the block 4.3 SOUND & ALARM CONTROL. – MON-PROM V6.00 In the following description (4.1.1 – 4.1.8) As the upgrade was optional and not alarms are mentioned in groups, each group mandatory, a number of not upgraded associated with corresponding Alarm field SV 300 are equipped with MON-PROM V5.01 within the Front panel field ”Alarms and (and PC 1608A, -B, -C or -D). These units Messages”. must be equipped with a FAILURE ALARM BOX. Note – PC 1608E is backwards compatible For each Alarm field this description of the and can be equipped with MON-PROM V5.01. normal (digital control) alarm contains: The FAILURE ALARM BOX must be used also on • Alarms and Messages - Alarm field these units. name The alarm functions of the FAILURE ALARM BOX • Parameter monitored for alarm, PMA are integrated on MON-PROM V6.00/PC 1608E and corresponding Reference value and the FAILURE ALARM BOX is not used on • Alarm table SV 300 equipped with MON-PROM V6.00/ containing conditions for alarm. PC 1608E (or higher). For more information about PROM, PC board and Operating Manual versions, see chapter ”6. Product change history”. 34 Siemens-Elema AB E382 E380E 061 01 03 01
  • 37. Servo Ventilator 300/300A Description of functions 4.1.1 Airway pressure PMA: PI = INSP. PRESSURE; signal from main block 12 INSPIRATORY PRESSURE. PE = EXP. PRESSURE; signal from main block 13 EXPIRATORY PRESSURE. LIM_PRESS.L = Signal from the block 6.1 MODE CPU including REF-PROM functions. Reference value Pmax = Preset Upper pressure limit (on Front panel). Reference value PEEP = Preset PEEP Level (on Front panel). Alarm table: 2 ”Alarms and messages” Alarm condition Remark Airway pressure too high 2 PI > Pmax or At true condition 1 PE > Pmax INSPIRATION is inhibited and EXPIRATION is started. High continuous pressure PI > PEEP + 15 cm H2O or Alarm if any of the PE > PEEP + 15 cm H2O conditions remains true during 15 s. Limited pressure LIM_PRESS.L = active, LIM_PRESS.L signal 3 only (corresponds to condition: used during PRVC or VS PREF = Pmax – 5 cm H2O mode. Alarm if the and resulting Tidal/Minute mentioned alarm condition volume ventilation is less is true during three than front panel setting) consecutive breaths. 1 Refer to 6.1.2 Respiratory timing - condition (3). 2 Identical condition is also monitored separately within the main block 2. Panel interface and that is used for the following function: At true condition the Airway pressure bargraph indication for Upper pressure limit is flashing. 3 Refer to 6.1.1 Reference generation - PRVC and VS mode. Monitoring also controls the following opening and closing of the Safety valve: Opening condition Remark PI > Pmax + 6 cm H2O or The Safety valve opens. If the opening was caused by PE > Pmax + 6 cm H2O any of these conditions the Backup alarm system generates continuous audible alarm. When opened the Safety valve will stay open until following conditions are true: Precondition Closing condition Safety valve stays open at least during PI < Pmax + 5 cm H2O and 400 ms. PE < Pmax + 5 cm H2O For Airway pressure alarm also see 4.5 Backup alarm system. E382 E380E 061 01 03 01 Siemens-Elema AB 35
  • 38. Description of functions Servo Ventilator 300/300A 4.1.2 O2 concentration PMA: O2 concentration = Signal generated within the block 4.1 using O2 CONC. signal from the block 11.3 O2 CELL which is digitally compensated for mean airway pressure and barometric pressure according to BAROMETER PRESSURE signal from 4.6 BAROMETER PRESSURE TRANSDUCER. Reference value Set O2 = Preset O2 concentration (on Front panel). Alarm table: 2 ”Alarms and messages” text Precondition Alarm condition O2 conc too high 21% < Set O2 < 100% PMA > Set O2 + 6% O2 conc too low 21% < Set O2 < 24% PMA < 18% 24% < Set O2 < 96% PMA < Set O2 – 6% 96% < Set O2 PMA < 90% ”Alarms and messages” text Additional function O2 SENSOR If the O2 cell is disconnected; High priority alarm that can be changed to a caution signal. If above mentioned O2 concentration alarm condition is true during less than 55 s the alarm is not registered in the alarm memory. If the O2 concentration setting is changed more than 2 percentage units, the alarm is automatically muted for maximum 55 s. This also applies at activation of ”Oxygen breath”. For O2 concentration alarm also see 4.5 Backup alarm system. 4.1.3 Optional alarm field On earlier versions of the ventilator, the alarm field name was ”CO2 concentration”. This optional alarm field is not used. 36 Siemens-Elema AB E382 E380E 061 01 03 01